Galilei Corneal Tomography for Screening of Refractive Surgery Candidates: A Review of the Literature, Part II

Majid Moshirfar 1,2, Mahsaw N Motlagh 3, Michael S Murri 1, Hamed Momeni-Moghaddam 4, Yasmyne C Ronquillo 2, Phillip C Hoopes 2
PMCID: PMC6778460  PMID: 31598521

Abstract

Corneal topography is the most widely used technology for examining the anterior corneal surface. Scheimpflug imaging is a newer technique that allows for measurement of both the anterior and posterior corneal surface, which allows for three-dimensional reconstruction of the cornea. This is of particular interest and value in the field of cataract and refractive surgery. The Galilei camera is a commercially available dual Scheimpflug system that combines curvature data from Placido disc-based corneal topography with elevation data from Scheimpflug technology. The addition of Placido disc topography makes the Galilei unique from its more popular counterpart, the Pentacam, which was discussed in Part I. Compared to the Pentacam, and however, the Galilei analyzer is a newer system that has emerged as a valuable screening tool given its dual Scheimpflug capability. In the first article of this series, the authors summarized the refractive indices available on the Pentacam system with the purpose of identifying the best diagnostic parameters for keratoconus. Similarly, the purpose of this article is to summarize corneal surface indices available on the Galilei system and evaluate their use in screening of the refractive surgery candidate. Since post-operative keratectasia is still prevalent, this paper aims to identify the most clinically relevant indices that may be used in pre-operative evaluation.

Key Words: Corneal Biomechanics, Ectasia, Refractive Surgery, Pentacam; Galilei, Corvis, Keratoconus, Scheimpflug Imaging

INTRODUCTION

Function of Dual Scheimpflug Imaging

As discussed in Part I of this article series, anterior segment imaging has drastically changed over the last 25 years. Since its introduction in the 1980s, topography has vastly altered the sphere of pre-operative screening for refractive surgery candidates [1, 2]. Tomographic devices have many functions including evaluation of cataracts, calculations of intraocular lenses (IOLs) power, guiding of surgical plans, and longitudinal follow-up of patients with known disease. In the realm of refractive surgery, the fundamental purpose of any tomographic device is to appropriately screen surgical candidates and facilitate high quality outcomes for patients. In this review, we focus on the Galilei dual Scheimpflug system (Ziemer Ophthalmic Systems AG, Port, Switzerland), which is one of the newer tomography devices available for refractive and cataract surgery. The Galilei analyzer is a uniquely elegant system that combines corneal curvature data from Placido disc topography with elevation data from Scheimpflug technology. The device itself has a rotating camera head that can capture two Scheimpflug slit images and Placido disc simultaneously [3, 4]. By combining height data from the slit images, the camera can track decentration from the inevitable micromovements of the eye. A benefit of the Galilei compared to other Scheimpflug analyzers such as Pentacam (Oculus Optikgeraete GmbH; Wetzlar, Germany) or Orbscan II (Bausch & Lomb, Orbtek Inc., Utah, USA) is its ability to reduce motion error by correcting eye motion and cyclotorsion through a software algorithm. Moreover, height data from both Scheimpflug and Placido are fitted to the anterior corneal surface data [5]. Data for the posterior corneal surface is derived from the images of the dual Scheimpflug analyzer. Another advantage of the Galilei comes in the form of corneal aberration analysis. The system calculates wavefront aberrations for both the anterior and posterior corneal surface independent of aberrations from the lens [1]. This data is then displayed alongside root mean square indices calculated over a 6.0-mm optical zone centered over the pupil [1].

When compared to other tomography devices, dual Scheimpflug imaging improves posterior corneal measurements and provides for excellent accuracy in assessing corneal thickness [6-12]. Moreover, from a practical standpoint, the Galilei can perform all the measurements a refractive surgeon may require prior to surgery. The convenience of having this “one-stop shop” is advantageous to the patient experience and for clinical workflow. Another benefit of the Galilei is calculation of total corneal power using ray tracing, which refracts incoming parallel rays through the anterior and posterior corneal surface. The calculated power and axis of astigmatism from these measurements are immensely helpful in pre-operative planning of IOLs and limbal relaxing incisions. Studies have shown that Galilei has superior IOL power prediction with a lower error when compared to other Scheimpflug devices [13, 14]. Despite demonstrating superior repeatability in thickness and higher order aberrations [14-20], there is less data available for the Galilei because it is a newer system. This is a limitation as longitudinal analysis, and further repeatability studies are required to ascertain the actual utility of the Galilei device in clinical practice.

Application for Screening Refractive Surgery Candidates

The Galilei device offers advantages in both accuracy and reproducibility that are valuable in understanding the corneal surface [7, 15, 21, 22]. Galilei has the highest level of repeatability compared to other Scheimpflug systems [23]. As a dual Scheimpflug analyzer, the Galilei eliminates the parallax of a single image and compensates for off-center measurements. Thus, theoretically, the Galilei camera can provide higher sensitivity and specificity in detecting early ectatic changes on the corneal surface. When considering refractive screening indices, anterior corneal measurements from Placido imaging aid in improving sensitivity, while the posterior measurements derived from Scheimpflug imaging improve specificity. By integrating these data points, the Galilei has the unique potential of becoming an excellent screening tool in clinical practice.

As discussed in Part I of this article series, the age-old question remains: what are we screening for? While many classification systems for keratoconus (KC) exist [24-33], the community of refractive surgeons has failed to agree on a universal definition of early ectasia [4]. This history of this debate is discussed at length in Part I of this article series. As such, we continue to use pre-keratoconus as a unified term that includes form-fruste KC, subclinical KC, borderline KC, suspect KC, and early KC. In this article, we present a review of the literature as it pertains to the Galilei system and highlight the clinical application of screening indices in discerning normal and ectatic corneas.

METHODS

A literature review was performed using various databases, including PubMed, Mendeley, Ovid, Elsevier, and Science Direct. For the database search the primary search term included “Galilei”, which was connected to descriptors such as “LASIK”, “progression parameters”, “screening”, “comparison”, “dual Scheimpflug”, “tomography”, “topography”, “keratoconus”, “subclinical keratoconus”, “evaluation”, “G6”, “G4”, “G2”, “analyzer”, “index”, “indices” and various others. Peer-reviewed and scholarly resources, including original scientific articles as well as review articles were included. Publications between 1900 and 2019 were included in this review. Articles were screened for relevance and significance based on their abstracts. Those that were identified as appropriate for this review were included. Additional searches were made to find relevant literature through Mendeley, Ovid, Elsevier, and ScienceDirect. All articles that were deemed relevant to this topic were included in this review.

As described in Part I, studies that evaluate refractive indices rely on the area under the curve (AUC) to determine diagnostic accuracy. Therefore, Galilei parameters with AUC >0.900 were judged suitable for screening of KC, while parameters with AUC >0.800 were selected for screening of pre-keratoconus. The selected parameters that met these requirements were then incorporated in creating proposed cut-off thresholds. Indices that met these criteria in at least two studies were then averaged based on the cut-off value proposed by the individual study. Clinically accepted normal index ranges for the parameters evaluated in this article are shown in Table 1. In Table 2 the highlighted parameters indicate the selected cut-off values that were averaged for screening of KC. However, given the limited studies evaluating Galilei with cases of pre-keratoconus, we elected to highlight the parameters with an AUC >0.800 in Table 3 as there was insufficient data to develop proposed thresholds.

Table 1.

Accepted Abnormality Thresholds of Keratoconus Indices for the Galilei System Based on Literature Review

Parameter Description Abnormal Threshold
AAI Asymmetry of asphericity over the corneal surface correlates to rate of curvature change > 25
CSI Difference between the area-corrected corneal power between two rings on the corneal surface > 1.00
DSI Greatest difference between any two 45-degree corneal sectors > 3.50
IAI Average of corneal power variations along each meridian of the corneal surface > 0.50
I-S Dioptric asymmetry between the inferior and superior corneal hemispheres > 1.4
KPI Compilation index that describes the percent probability of keratoconus > 30
OSI Greatest difference between any two oppositely positioned 45-degree corneal sectors > 2.10
PPK Probability index that describes the optimal threshold for detecting keratoconus > 45
SDP Standard deviation of total corneal power > 2.00
SAI Average of differences in corneal power between opposite points on the corneal surface > 0.50
SRI Sum of power variations along ten central rings over corneal surface, characterizes local fluctuations > 1.55

Abbreviations: AAI: Asphericity Asymmetry Index; CSI: Center/Surround Index; DSI: Differential Sector Index; I-S: Inferior-Superior Index; IAI: Irregular Astigmatism Index; KPI: Keratoconus Probability Index; SDP: Standard Deviation of Corneal Power; OSI: Opposite Sector Index; PPK: Percentage Probability of Keratoconus; SAI: Surface Asymmetry Index; SRI: Surface Regularity Index.

Table 2.

Summary of Galilei Keratoconus Parameters in Detecting Clinical Keratoconus

Study Cut-off Value Sensitivity Specificity AUC
AAI
Smadja et al [ 5 ] 34.5 1.000 0.9995 -
CLMI
Shetty et al [ 34 ] - - - 0.966
Mahmoud et al [ 48 ] - 0.892 0.988 0.950
Mahmoud et al [ 37 ] - 0.994 0.996 0.995
Kocamiş et al [ 39 ] 1.82 0.890 0.940 0.920
CSI
Demir et al [ 35 ] - - - 0.918
Feizi et al [ 36 ] 0.700* 0.920 0.950 0.951
Shetty et al [ 34 ] 0.700 0.135 0.143 0.906
DSI
Demir et al [ 35 ] - - - 0.989
Feizi et al [ 36 ] 3.26 0.980 0.910 0.977
Shetty et al [ 34 ] 3.26* 0.904 0.976 0.983
IAI
Demir et al [ 35 ] - - - 0.960
Feizi et al [ 4 ] 0.580* 0.920 0.920 0.974
Shetty et al [ 3 ] 0.580 0.673 1.000 0.973
I-S
Demir et al [ 6 ] - - - 0.968
Feizi et al [ 4 ] 2.33 0.900 0.980 0.971
Shetty et al [ 3 ] 2.33* 0.885 0.952 0.980
KPI
Feizi et al [ 4 ] 18.55* 1.000 0.990 0.999
Shetty et al [ 3 ] 18.55 0.962 0.952 0.993
Mahmoud et al [ 37 ] - 0.850 1.000 -
Kprob
Feizi et al [ 4 ] 25.55* 0.980 0.990 0.998
Shetty et al [ 3 ] 25.55 0.962 0.952 0.993
OSI
Demir et al [ 6 ] - - - 0.987
Feizi et al [ 4 ] 2.04* 0.960 0.990 0.988
Shetty et al [ 3 ] 2.04 0.923 0.952 0.983
PPK
Shetty et al [3] 45.0 1.000 0.286 0.968
SAI
Demir et al [ 6 ] - - - 0.998
Feizi et al [ 4 ] 1.25* 1.000 0.990 0.999
Shetty et al [ 3 ] 1.25 0.923 0.952 0.984
SDP
Feizi et al [ 4 ] 1.93* 0.920 1.000 0.993
Shetty et al [ 3 ] 1.93 0.981 0.833 0.986
SRI
Demir et al [ 6 ] - - - 0.942
Feizi et al [ 4 ] 1.52 0.900 1.000 0.982
Shetty et al [ 3 ] 1.52* 0.654 1.000 0.992
TCP-Central
Reddy et al [ 32 ] 45.40 0.870 0.960 0.940
Demir et al [ 6 ] - - - 0.984
Feizi et al [ 38 ] 45.69* 0.894 0.995 0.986
TCP-Flat
Reddy et al [ 32 ] 44.50 0.690 0.990 0.790
Feizi et al [ 38 ] 44.04* 0.732 0.980 0.915
TCP-Steep
Reddy et al [ 32 ] 46.10 0.980 0.970 0.990
Feizi et al [ 38 ] 46.47* 0.944 0.990 0.994
Total HOA
Kocamiş et al [ 39 ] 0.873 0.840 0.810 0.820
Total RMS
Reddy et al [ 32 ] 2.05 0.920 0.910 0.970
Trefoil Z (3, -3)
Reddy et al [ 32 ] 0.31 0.500 0.980 0.700
Vertical Coma
Reddy et al [ 32 ] -1.07 0.630 1.000 0.800
Kocamiş et al [ 39 ] -0.312* 0.850 0.840 0.850
3rd-order RMS
Reddy et al [ 32 ] 1.1 0.810 0.980 0.910
4th-order RMS
Reddy et al [ 32 ] 0.47 0.820 0.950 0.930
5th-order RMS
Reddy et al [ 32 ] 0.15 0.840 0.940 0.900
6th-order RMS
Reddy et al [ 32 ] 0.06 0.820 0.900 0.890

Abbreviations: AAI: Asphericity Asymmetry Index; AUC: Area under the Curve; CLMI: Cone Location and Magnitude Index; CSI: Center/Surround Index; DSI: Differential Sector Index; HOA: Higher Order Aberration; I-S: Inferior-Superior Index; IAI: Irregular Astigmatism Index; KPI: Keratoconus Probability Index; Kprob: Keratoconus Probability; OSI: Opposite Sector Index; PPK: Percentage Probability of Keratoconus; TCP: Total Corneal Power; RMS: Root Mean Square SAI: Surface Asymmetry Index; SDP: Standard Deviation of Corneal Power; SRI: Surface Regularity Index. *denotes cut-off point with best area under the curve if more than one study evaluated index accuracy. Highlighted indices represent the parameters that were included in the final evaluation of proposed thresholds.

Table 3.

Summary of Galilei Keratoconus Parameters in Detecting Pre-Keratoconus

Study Cut-off Value Sensitivity Specificity AUC
AAI
Smadja et al [ 5 ] 21.5 0.936 0.972 -
Golan et al [ 7 ] 13.5 0.774 0.670 0.750
CLMI
Shetty et al [ 3 ] - - - 0.314
CSI
Feizi et al [ 4 ] 0.99 0.217 1.000 0.534
Shetty et al [ 3 ] 0.90* 0.973 0.047 0.556
DSI
Feizi et al [ 4 ] 1.725* 0.739 0.556 0.646
Shetty et al [ 3 ] 1.730 0.541 0.698 0.627
IAI
Golan et al [ 7 ] 0.400 0.516 0.736 0.625
Feizi et al [ 4 ] 0.445 0.739 0.541 0.664
Shetty et al [ 3 ] 0.450* 0.541 0.814 0.858
I-S
Feizi et al [ 4 ] 1.60* 0.348 0.895 0.597
Shetty et al [ 3 ] 1.60 0.108 0.907 0.595
KPI
Feizi et al [ 4 ] 5.00* 0.565 0.835 0.710
Shetty et al [ 3 ] 5.00 0.568 0.581 0.629
Kprob
Feizi et al [ 4 ] 11.60* 0.391 0.955 0.669
Shetty et al [ 3 ] 11.60 0.297 0.860 0.626
OSI
Feizi et al [ 4 ] 1.850 0.304 0.985 0.637
Shetty et al [ 3 ] 1.850 0.216 0.953 0.510
Golan et al [ 7 ] 0.671* 0.839 0.450 0.658
PPK
Shetty et al [ 3 ] 25.0 0.811 0.093 0.318
SAI
Feizi et al [ 4 ] 0.895 0.435 0.917 0.644
Shetty et al [ 3 ] 0.895* 0.432 0.907 0.656
SDP
Feizi et al [ 4 ] 1.065 0.565 0.857 0.692
Shetty et al [ 3 ] 1.065* 0.892 0.814 0.916
SRI
Feizi et al [ 4 ] 0.735 0.826 0.511 0.679
Shetty et al [ 3 ] 0.735* 0.676 0.860 0.875
TCP-Central
Reddy et al [ 32 ] 44.12 0.170 0.670 0.510
TCP-Flat
Reddy et al [ 32 ] 43.3 0.830 0.400 0.580
TCP-Steep
Reddy et al [ 32 ] 42.8 0.870 0.360 0.630
Total RMS
Reddy et al [ 32 ] 1.15 0.960 0.580 0.820
Trefoil Z (3,-3)
Reddy et al [ 32 ] -0.08 0.610 0.580 0.540
Vertical Coma
Reddy et al [ 32 ] -0.54 0.570 0.900 0.730
3rd-order RMS
Reddy et al [ 32 ] 0.78 0.710 0.900 0.830
4th-order RMS
Reddy et al [ 32 ] 0.37 0.710 0.830 0.790
5th-order RMS
Reddy et al [ 32 ] 0.09 0.670 0.820 0.800
6th-order RMS
Reddy et al [ 32 ] 0.05 0.620 0.780 0.740

Abbreviations: AAI: Asphericity Asymmetry Index; AUC: Area under the Curve; CLMI: Cone Location and Magnitude Index; CSI: Center/Surround Index; DSI: Differential Sector Index; I-S: Inferior-Superior Index; IAI: Irregular Astigmatism Index; KPI: Keratoconus Probability Index; Kprob: Keratoconus Probability; OSI: Opposite Sector Index; PPK: Percentage Probability of Keratoconus; RMS: Root Mean Square SAI: Surface Asymmetry Index; TCP: Total Corneal Power; SDP: Standard Deviation of Corneal Power; SRI: Surface Regularity Index. *denotes cut-off point with best area under the curve if more than one study evaluated index accuracy. Highlighted indices represent the parameters that had an AUC >0.800, which is the minimum diagnostic accuracy recommended for screening of pre-keratoconus.

Indices for Refractive Screening

Like other corneal imaging systems, the Galilei device has indices for topographic, topometric, pachymetric, and corneal wavefront data. Our focus is to define and review the specific keratoconus indices included in Galilei software that are most valuable to clinical practice. A unique feature of the Galilei system is that it automatically evaluates certain parameters in the prediction of KC. Galilei’s output reports are extremely user friendly and include the Refractive Report, the Keratoconus Report, the Wavefront Report, and the IOL Power Report [2]. While no formal thresholds have been set to define abnormal, we have provided the clinically accepted normal index ranges in Table 1. Based on the available literature, a side-by-side comparison of screening indices along with their respective sensitivity, specificity, and area under the curve (AUC) for both clinical and subclinical KC can be found in Table 2 and Table 3 respectively. A simplified version of these screening indices along with our recommended cut-off values is found in Table 4.

Table 4.

The Galilei Clinical “Cheat Sheet”: Suggested Cut-off Values for Keratoconus Indices in Screening Clinical Keratoconus and Pre-Keratoconus

Parameter Clinical Keratoconus Pre-Keratoconus
Cut-Off Value Cut-Off Value
CSI 0.70 -
DSI 3.26 -
IAI 0.58 0.450*
I-S 2.33 -
KPI 18.55 -
Kprob 25.55 -
OSI 2.04 -
SAI 1.25 -
SDP 1.93 1.065*
SRI 1.52 0.735*
TCP-Central 45.55 -
TCP-Steep 46.29 -

This table is to serve as a quick screening tool. Formal guidelines are not available at this time and further validation studies are required. Nevertheless, if multiple parameters are calculated to be beyond the recommended threshold values, then it should increase the index of suspicion and trigger further testing for the patient.

Abbreviations: CSI: Center/Surround Index; DSI: Differential Sector Index; I-S: Inferior-Superior Index; IAI: Irregular Astigmatism Index; KPI: Keratoconus Probability Index; Kprob: Keratoconus Probability; OSI: Opposite Sector Index; TCP: Total Corneal Power; SAI: Surface Asymmetry Index; SDP: Standard Deviation of Corneal Power; SRI: Surface Regularity Index. *These are the only indices that demonstrated diagnostic accuracy appropriate for pre-keratoconus (area under the curve >0.800); they do not represent averages as further validation studies are required, however, the authors elected to have them displayed in this table for the convenience of the clinician.

The following subsections detail the several indices available through the Galilei system for screening. As mentioned previously, there are few investigations of Galilei as it is a newer imaging system. Cognizant of this, we encourage the clinician to pay special attention to the following studies as they were best in evaluating the Galilei device based on selection criteria, methods, and study size: Shetty et al (number [n] = 42 normal, 51 KC, 37 pre-keratoconus) [3], Feizi et al (n = 136 normal, 51 KC, 23 pre-keratoconus) [4], Smadja et al (n = 177 normal, 148 KC, 47 pre-keratoconus) [5], Demir et al (n = 151 normal, 67 KC) [6], and the recent study by Golan et al (n = 178 normal, 31 pre-keratoconus) [7].

Throughout each subsection, the clinician is encouraged to keep in mind that although almost all indices are validated for detection of KC, the opposite holds true in the discrimination of pre-keratoconus. Moreover, based on the aforementioned studies, we recommend the following indices as the most effective for detection of pre-keratoconus and recommend the clinician’s special attention to asphericity asymmetry index (AAI) and surface regularity (SRI). While AAI lacks published values demonstrating its diagnostic accuracy as assessed by AUC, it demonstrates superior sensitivity and specificity that warrants its consideration for pre-keratoconus scrrening. Moreover, based on our review of the literature, standard deviation of corneal power (SDP) demonstrated the single highest AUC for detection of pre-keratoconus (Table 3).

Keratoconus Predictive Indices

Asphericity Asymmetry Index

The asphericity asymmetry index (AAI), also known as the Kranemann-Arce index, is a parameter that quantifies the asymmetry of asphericity over the corneal surface [8]. The AAI is calculated as the magnitude of difference between the maximum negative best-fit toric aspheric (BFTA) reference surface value and the maximum positive BFTA elevation value (Fig. 1) [5, 8]. Symmetric aspheric meridians have elevation values closer to zero, while asymmetric aspheric meridians have higher variability in the change of curvature within each hemimeridian.

Figure 1.

Figure 1

The Asphericity Asymmetry Index (AAI) Quantifies the Asymmetry of Asphericity over the Corneal Surface. As Shown by Smadja et al. [5, 11], it is Best Assessed using the Best-fit Toric Aspheric (BFTA) Maps of the Galilei System. AAI is Calculated as the difference between the Maximum Negative and Positive BFTA elevation Values

Therefore, higher values of AAI correlate with increased rates of curvature change and correspond well with the amount of corneal coma [8]. Posterior AAI is a relatively new, manually derived parameter that utilizes BFTA map as well; it is calculated as the absolute value of the highest negative and positive elevations within the posterior corneal zone [7, 9]. While no formal threshold has been set, it is generally accepted that normal posterior AAI values are below 20-25 micrometer (µm) [10]. With the growing consensus that posterior corneal changes may occur first in the development of KC, posterior AAI may be a valuable index to detect pre-keratoconus.

In a recent study, posterior AAI was selected as the most discriminant variable in the detection of suspicious corneas [5]. By using a binary automated decision tree, their study showed posterior AAI to have excellent sensitivity (100%) and specificity (99.5%) in the detection of clinical KC [5]. While diagnostic accuracy decreased slightly for the detection of pre-keratoconus, it was still the most discriminant index among fifty-five Galilei parameters. Interestingly, anterior AAI did not share the same significance in discriminating pre-keratoconus and KC, and in a separate study, anterior AAI did not show significant differences between normal and pre-keratoconic eyes [11]. This finding strengthens the growing consensus that posterior corneal changes occur first in the development of KC [12-17].

More recently, no individual index had satisfactory discriminatory accuracy in detecting pre-keratoconus [7]. However, posterior AAI was included in a combination index that had an excellent AUC of 0.960 (sensitivity 90.3%, specificity 92.6%) [9]. While Smadja and colleagues [5] found an optimal cut-off value of 21.5 µm for posterior AAI, Golan et al. [7] found an optimized cut-off of 13.5 µm. These disparities may in part be attributable to each study’s selection criteria and definition of pre-keratoconus. Despite promising initial results, further validation studies with large population datasets are required to affirm the utility of AAI as a reliable screening index.

Center/Surround Index

The Center/Surround Index (CSI) is a quantitative index that characterizes the difference in corneal power between two areas of the cornea. To fully understand CSI, the corneal surface is first divided into eight arbitrary sectors measuring a 45° angle as shown in Fig. 2. For each sector the mean axial keratometric power is then calculated. This process repeats itself until each possible pattern of sector distribution has been applied to the corneal surface [18-20]. Described succinctly by Koch et al., this means that if keratometry is analyzed for 256 semimeridians, then there are 32 possible patterns (256 semimeridians/8 sectors: 32 patterns) [19]. Based on the 45° sectors, CSI represents the difference in the average area-corrected corneal power between a central area (3.0-mm diameter) and an annulus surrounding the central area (3.0-6.0mm diameter) [6]. In normal eyes or with regular astigmatism CSI values are low while in keratoconus CSI is high, thus making CSI a sensitive index for identification of centrally located steepening (Fig. 3) [21, 40].

Figure 2.

Figure 2

The Center/Surround Index (CSI), Differential Sector Index (DSI), and Opposite Sector Index (OSI) are Calculated based on Eight Arbitrary 45-Degree Sectors of the Corneal Surface. For each Sector Mean Axial Power is Calculated. Sectoral Patterning of the Corneal Surface is Repeated until each Possible Pattern of Sector Distribution is Applied. CSI is the difference of the Area-corrected Corneal Power between a Central 3mm Ring and a 6mm Annulus that Surrounds the Central Ring. DSI is the Greatest difference between any Two Sectors. OSI is the Greatest difference between any Two Opposite Sectors

Figure 3.

Figure 3

Relationship between Normal Corneal Surface, Astigmatism, and Steepening with Keratoconus Predictive Indices

Maeda and colleagues were one of the first to report CSI as a valuable index in differentiating among normal corneas and KC [40]. Moreover, CSI has a significant correlation with visual function, making it a valuable parameter in assessing overall acuity [10]. CSI obtained by Galilei had an excellent diagnostic accuracy in distinguishing keratoconic eyes [6]. Feizi and colleagues reported an AUC of 0.951 for CSI [4]. However, CSI failed as an individual parameter in discriminating pre-keratoconus with a best AUC of 0.534 [4]. These findings were similar to the study by Shetty et al., which reported CSI as a reliable index for identifying KC, but an insensitive diagnostic test for the detection of pre-keratoconus cases [3]. Therefore, based on the available literature we conclude that CSI is a reliable parameter for the diagnosis of KC but should not be used as a standalone index when diagnosing pre-keratoconus.

Differential Sector Index

The differential sector index (DSI) is an index related to the degree of asymmetry present on the corneal surface [19]. As with CSI, DSI is calculated based on the 8-sector pattern of the corneal surface (Fig. 2). With this in mind, DSI characterizes the maximum difference between any two sectors. Thus, with any increase in surface irregularity, the DSI expectedly increases. DSI is a sensitive index for the identification of peripherally located steepening and in corneas with regular astigmatism as shown in Fig. 3 [40].

The studies that evaluate DSI are limited. Similar to other Galilei parameters, DSI is a highly accurate index in the identification of clinical KC [3, 4, 6]. However, in the current literature, DSI is not reported to be a reliable parameter for the identification of pre-keratoconus. Studies have reported AUC <0.650, which is not suitable for clinical standards [3, 4]. Given the available literature, we make the same recommendation for DSI as we did for CSI: an excellent parameter for use in clinical KC, but the index should not be relied on individually for diagnosis of pre-keratoconus.

Higher Order Aberrations and Wavefront Analysis

While higher order aberrations (HOAs) are available in a multitude of topography systems, the Galilei system has demonstrated excellent and perhaps superior repeatability with these measurements [22, 41-43]. Wavefront data from corneal surface analysis is expressed in the form of Zernike polynomials [23-26], and each corneal surface is expressed as the sum of Zernike polynomials. These are generally referred to as corneal aberrations. The Galilei system automatically computes corneal wavefront data and HOAs based on the corneal elevation profile. When compared to single Scheimpflug systems, the Galilei system has less variability in reported Zernike terms [42], likely due to the added benefit of having a dual-channel camera. Another benefit of using the Zernike expansion is that the coefficient of each mode represents the root mean square (RMS) wavefront error that is attributable to that particular mode [26]. Thus, larger coefficient values are associated with modes that contribute a greater amount to the total RMS of the system.

The total HOAs are conveniently displayed in the Wavefront Report, which is automatically calculated by the Galilei system. An example of the Zernike polynomial expansion is shown in Fig. 4. The second-order corneal aberrations are the Zernike coefficients for astigmatism and defocus [2, 26]. Third-order corneal aberrations represent the vertical and horizontal values for trefoil and coma [2, 26]. Coma is a valuable indicator for the progression of KC and has been documented as an index that reflects early changes in the corneal surface [27]. Moreover, higher values for vertical coma and large coma RMS values are associated with KC and possible pre-keratoconus [13, 27-31]. Fourth-order HOAs include the quatrefoil and 4th order astigmatism values, which also tend to increase in magnitude if the cornea is deformed [26, 27, 29]. The most clinically relevant fourth-order aberration is the spherical aberration (SA) as it is documented to have a normative range of +0.15 µm to +0.30 µm [2]. Therefore deviations in this range may indicate variations of disease.

Figure 4.

Figure 4

Wavefront Aberrations as Represented by a Zernike Expansion, which is a Weight Sum of Zernike Polynomials where each Polynomial Corresponds to an Aberration Coefficient. Wavefront Aberrations Quantify the Length between a Plane Wavefront Entering the Pupil and the Corresponding Wavefront of Light that Exits a Point on the Retina

Given the variation of HOAs in corneal disease, several studies have explored its value as a screening parameter for KC. Reddy and associates identified several parameters capable of serving as discriminant indices, which included 3rd order, 4th order, 5th order, 6th order, and total higher-order RMS [32]. Furthermore, the RMS for 3rd order, 4th order, 5th order, 6th order, and total higher-order RMS were found to be significantly different between normal eyes and pre-keratoconus [32]. However, only 3rd order and total RMS had acceptable AUC values of 0.830 and 0.820, respectively, in differentiating these two patient groups. Interestingly, in a recent study by Golan and colleagues, none of the HOAs were significantly different between normal eyes and pre-keratoconus [7]. As noted before, it is possible that these disparate conclusions are attributable to the inclusion criteria for pre-keratoconus defined by each study.

Recently, vertical coma was significantly different between normal and pre-keratoconus corneas [11]. However, they did not analyze AUC or discuss the respective sensitivity and specificity for each parameter. Nevertheless, some reports corroborate the importance of vertical coma in its ability to discriminate pre-keratoconus [16, 28]. Vertical coma was significantly greater in progressive versus non-progressive KC [33]. This may indicate that HOAs play a role in analyzing the evolution of KC over time. Moreover, Tellouck and colleagues found that changes in vertical coma occurred before any changes of the posterior corneal surface [38], which may also point towards the diagnostic utility of HOAs in screening of pre-keratoconus. Subsequent validation studies are required to elucidate the clinical value of vertical coma and other HOAs for screening of corneal refractive surgery candidates.

Irregular Astigmatism Index

The irregular astigmatism index (IAI) describes variation in measured axial power between central rings along any given meridian [19]. IAI is calculated as the average sum of area-corrected keratometric power variations along every meridian for the entire analyzed corneal surface (Fig. 5) [40]. IAI had excellent diagnostic accuracy for KC in multiple studies [3, 4, 6, 9].

Figure 5.

Figure 5

Irregular astigmatism index (IAI) is calculated as the average sum of the inter-ring area-corrected corneal power along every meridian along the corneal surface

For pre-keratoconus, there is no consistent agreement in the literature. Feizi and colleagues reported an AUC of 0.664 [4], which is very similar to the AUC of 0.625 [7]. However, Shetty and associates found a much higher AUC of 0.858 despite using a similar cut-off value as shown in Table 3 [3]. As all three studies were conducted in various regions, it is also possible that inherent population demographics may also play a role in this difference. Thus, our recommendation is to use IAI with caution for pre-keratoconus and to consider a combination of indices when making screening refractive surgery candidates. As with other parameters, IAI as an individual index is capable of distinguishing KC according to the current literature.

Inferior-Superior Index

First described by Rabinowitz, the inferior-superior (I-S) index characterizes the dioptric asymmetry between the inferior and superior hemispheres [44]. Rabinowitz and McDonnell modified the threshold values and established criteria that stated an I-S index ≥ 1.4 diopter (D) is susceptible to KC [34]. The I-S index obtained by Galilei is a valuable diagnostic parameter for KC [3, 4, 6]. For pre-keratoconus, however, the diagnostic accuracy is weak which points towards the limitation of the index in diagnosing these cases. Interestingly, however, for pre-keratoconus Shetty et al. reported a mere 10.8% sensitivity with a 90.7% specificity [3]. Thus, while an ideal screening parameter has a higher sensitivity to rule-out disease, the I-S index may be better equipped to rule-in ectatic disease based on its relatively greater specificity.

Opposite Sector Index

As previously described, the opposite sector index (OSI) is calculated based on the 45° sector patterns applied to the corneal surface. OSI is equal to the greatest difference between any two opposite sectors (Fig. 2) [6, 36]. Maeda and Klyce were the first to demonstrate the value of OSI and described its sensitivity in identifying peripherally located corneal steepening (Fig. 3) [36].

Demir et al. were the first to describe the diagnostic accuracy of OSI for KC [6]. Feizi et al. corroborated these findings with a similar AUC, as noted in Table 2 [4]. More recently, Shetty and associates further validated OSI as an index for frank KC [3]. However, in all three studies, OSI was not a reliable parameter for distinguishing eyes with pre-keratoconus. Therefore OSI, while a valuable screening index for KC, should not be used alone when considering pre-keratoconus.

Surface Asymmetry Index

The surface asymmetry index (SAI) is determined by the differences in keratometric power between opposite points distributed on 128 meridians [19]. In simpler terms, SAI characterizes the average of differences in corneal power. Multiple studies have validated the diagnostic utility of SAI in distinguishing clinical KC, which is shown in Table 2 [3, 6]. In the prospective cohort study by Feizi et al., SAI was identified as the best parameter to discriminate clinical KC among all Galilei indices [4]. While as a stand-alone index, SAI was not able to discriminate pre-keratoconus, it was able to do so alongside posterior best-fit sphere in a two-step tree analysis (100% sensitivity, 91.3% specificity) [4]. Given these findings, SAI may play a future role in the diagnosis of pre-keratoconus through a multivariate index.

Surface Regularity Index

The surface regularity index (SRI) characterizes the local irregularities of the corneal surface. SRI is calculated as the sum of power variation along 256 semimeridians on ten central rings over the corneal surface [19]. Thus, similar to DSI, any increase in corneal surface irregularity will manifest as an increase in SRI. Conversely, if the SRI value is equal to zero, the corneal surface is perfectly smooth [6]. SRI was first described by Wilson and Klyce and represents the localized fluctuations of the corneal surface [35]. While there are no formally described ranges, any value of SRI below 1.55 is generally accepted as normal [13, 45].

Galilei computed SRI has excellent diagnostic sensitivity and accuracy for identification of clinical KC (Table 2) [4, 6]. Shetty and colleagues confirmed the diagnostic utility of SRI in their recent study; interestingly, they reported an AUC of 0.875 for detecting pre-keratoconus, which is the highest reported for SRI in the available literature [3]. This is comparable to the Belin-Ambrósio Enhanced Ectasia Display Total Deviation Value (BAD_D) available on the Pentacam system. While a large amount of published data exists for BAD_D, there is inadequate validation for SRI. Studies are required to extrapolate the use of SRI for pre-keratoconus similar to BAD_D.

Total Corneal Power and Standard Deviation of Corneal Power

The Total Corneal Power (TCP) represents the average of corneal power for every detected point in a selected region of interest. In order to calculate TCP, the Galilei system utilizes ray tracing of the anterior surface, posterior surface, and pachymetry data [2]. Moreover, TCP can be calculated as the average TCP over steep (TCP-steep) and flat (TCP-flat) meridians, or the average over a central 4.0-mm zone (TCP-central). SDP is simply the value of the standard deviation of TCP. TCP has excellent reproducibility and is a reliable parameter when assessing patients for refractive surgery [42]. Even after surgical procedures, it has been documented that Galilei maintains high intra-observer repeatability for corneal power [46].

TCP-steep was also shown to have an excellent AUC of 0.990 in differentiating KC [32]. This was closely followed by TCP-central with an AUC of 0.94 [32]. Interestingly, however, TCP-flat had only a fair AUC of 0.790. For pre-keratoconus, the AUCs decreased markedly and were weak discriminant parameters [32]. These findings are similar to those of Demir et al., who also concluded that TCP has excellent diagnostic accuracy in distinguishing frank KC as demonstrated in Table 2 [6]. Feizi et al. also concluded that all TCP parameters have excellent diagnostic ability, but that TCP-steep had the highest AUC of 0.994 (94.4% sensitivity, 99.0% specificity) [38]. SDP has also been shown to have high sensitivity for identifying KC [3, 4]. In the recent study, SDP had excellent diagnostic accuracy as an individual parameter in identifying pre-keratoconus (Table 3) [3]. In fact, it was superior to the BAD_D of Pentacam, which had an AUC of 0.887. These studies underline the importance of further exploration of Galilei computed TCP and SDP for screening of refractive surgery candidates and detection of KC.

Keratoconus Probability Indices

Cone Location and Magnitude Index

The Cone Location and Magnitude Index (CLMI) was first described by Mahmoud and Roberts [37]. The CLMI relies on an area-corrected average steepest 2 mm-diameter circle within the central 8 mm-diameter anterior curvature map. From this, a curvature difference, M1, is calculated as the difference between all points outside the circle and all points inside the circle. A second circle that is centered 180 degrees away in angular position is analyzed in the same manner, resulting in curvature difference M2. CLMI is then calculated based on M1 and M2. In much simpler terms, the CLMI characterizes the steepest area of curvature, and the magnitude of the index identifies the difference between the steepest area and the rest of the curvature map [37]. This is schematically represented in Fig. 6. CLMI was developed as an index that could be calculated on different topographic systems. The index has shown good repeatability among various devices [41, 47]. Mauger et al. also calls attention to the value of posterior CLMI, which may serve as a valuable measure of asymmetric corneal steepening [47].

Figure 6.

Figure 6

Schematic Representation of the Cone Location and Magnitude Index (CLMI) [46,48] . A first Circle (C1) is 2 mm in Diameter over the Steepest Region of the Corneal Surface. From this a Curvature difference (M1) is Calculated as the Difference between the Points outside and inside C1. A second circle (C2) is Centered 180-Degrees Away from C1. In the same Fashion as M1, A Second Curvature difference, M2, is Calculated. CLMI is then Calculated based on M1 and M2, which Aims to Characterize the Steepest Area of Curvature Relative to the Rest of the Corneal Surface Map

Recently, Mahmoud and colleagues modified the CLMI to include posterior surface and corneal thickness data. This addition improved the sensitivity and specificity of the CLMI index from 89.2% sensitivity and 98.8% specificity to 99.4% and 99.6%, respectively [48]. This improvement emphasizes the underlying theme that a robust, multivariate index is superior to an individual parameter in detecting KC. The validation studies for CLMI on the Galilei system are limited. Based on the available literature CLMI it is an excellent screening index for KC, but inadequate for detection of pre-keratoconus as shown in Tables 2 and 3 [3].

Keratoconus Prediction Index

First described by Maeda et al., the Keratoconus Prediction Index (KPI) is a compilation index that describes the percent probability of KC based on analysis of the anterior corneal surface [40, 49]. KPI utilizes tomographic data to characterize the probability of having KC. Parameters included in this multivariate index are simulated keratometry, DSI, OSI, CSI, SAI, IAI, and percent area-analyzed. The formula for KPI is found in Formula 1. Based on the Galilei system, KPI from 0 to 10% corresponds to normal or suspicious corneas; KPI from 20 to 30% corresponds to keratoconic or suspicious corneas, and KPI >30% is indicative of pellucid marginal degeneration (PMD) [2, 3].

KPI was capable of differentiating KC from warpage due to keratoplasty, photorefractive keratectomy, radial keratotomy, and contact lens wear [40]. The discriminant functions of KPI are in large part because it is a multivariate parameter that includes a variety of tomographic data. For KC, Maeda and colleagues were able to differentiate KC with a perfect sensitivity (100%) and excellent specificity (96.0%) [40]. This finding is further supported by recent studies that also report excellent diagnostic accuracy for KC [3, 4].

However, for pre-keratoconus, KPI is largely reported to have a moderate to fair diagnostic accuracy as seen in Table 3 [3, 4]. In contrast, none of the KC probability indices, which includes KPI, were significantly different between normal and pre-keratoconus eyes [7]. Again, it is apparent that further validation studies are required to determine the appropriate role of KPI in screening refractive surgery candidates. As a multivariate index, we are more inclined to recommend KPI for screening purposes; however, given the limited studies available, it should not be used alone in screening patients.

KPI=0.30+0.01(-41.23-0.15*DSI+1.18*OSI+1.49*CSI+4.13*SAI-0.60*SimK1+1.08*SimK2-3.74*IAI+0.10*AA)

Formula 1: Keratoconus Prediction Index (KPI) Formula as Described by Maeda et al [40]. Abbreviations: DSI: Differential Sector Index; OSI: Opposite Sector Index; CSI: Center/Surround Index; SAI: Surface Asymmetry Index; SimK1: Simulated Keratometry 1; SimK2: Simulated Keratometry 1; IAI: Irregular Astigmatism Index.

Keratoconus Probability

The Keratoconus Probability (Kprob) is an index that characterizes sensitivity and specificity of the reported KPI based on a normative and keratoconic database [37]. Expectedly, Kprob has a inverse relationship with visual function [10]. Feizi et al. were the first to report the utility of Kprob in their prospective study evaluating clinical KC and pre-keratoconus. They found an excellent AUC of 0.998 for KC, but an inferior AUC of 0.669 when distinguishing eyes with pre-keratoconus [4]. This result was similarly obtained by Shetty and colleagues, who reported an AUC of 0.993 for KC and a similar AUC of 0.626 for pre-keratoconus [3]. For this reason, we recommend the use of Kprob only for discriminating clinical KC.

Percentage Probability of Keratoconus

Mahmoud and colleagues were also the first to describe Percent Probability of Keratoconus (PPK), which is defined as the optimal threshold for detecting KC [37]. The PPK is calculated from a validated equation that incorporates CLMI using axial data. The recommended cutoff value for clinical KC is 45.0%, while the cutoff value for pre-keratoconus is 20.0% [37]. Like the other probability indices, PPK can discriminate KC with excellent diagnostic accuracy but fails in distinguishing pre-keratoconus KC from normal eyes [2, 3].

DISCUSSION

Application and Interpretation of Galilei Indices

As with the Pentacam camera, the majority of screening indices on the Galilei system play a valuable role in discerning eyes with KC. When assessing pre-operative risk, it is often the pre-keratoconus cases that are of particular interest. Based on this comprehensive review, physicians should not rely solely on a single parameter to guide the course of treatment or surgical eligibility. Instead, we recommend a step-wise approach that begins with the patient history and incorporates tomographic data. By constructing the bigger clinical picture, clinicians are best able to stratify patients and turn them away from surgery if there is sufficient clinical concern.

Based on this review, nearly all Galilei calculated KC indices are capable of distinguishing clinical KC with an AUC >0.900. For pre-keratoconus, the only standalone parameter that meets an AUC of 0.900 is SDP. However, as this was only validated in one study, close attention to AAI and SRI is recommended as they have fair to good diagnostic accuracy; however, in the studies thus far these indices have not demonstrated strong enough diagnostic accuracy to be accepted as individual screening parameters. Moreover, given the limited amount of studies, further investigations are required to appreciate the diagnostic utility of these indices in full. Our recommended optimized cut-off values for each index is highlighted in Table 4. Unfortunately, the lack of studies interferes with our ability to recommend pre-keratoconus thresholds which is dissimilar from the other articles in this three-part series. Nevertheless, we have identified the parameters that have demonstrated an AUC >0.800 in Table 3 to draw special attention to the indices that are likely to have the biggest role in future refractive screening.

These recommendations can be used as a quick reference tool for daily clinical practice. For pre-keratoconus, our recommendation is to evaluate the parameters in total. For example, if a patient meets criteria for pre-keratoconus across multiple indices, then the index of suspicion that disease is present should be higher. Studies have shown that the best indices for screening are multivariate parameters that combine pachymetric and tomographic data. We anticipate that future studies will ultimately confirm this and that combination indices will prove to have the highest diagnostic accuracy.

Another application of the dual Scheimpflug system is its superior accuracy in predicting IOL power prediction [50]. Beyond its role in planning and screening for refractive surgery, the Galilei system has a clinical role in tracking outcomes and refractive error. In a recent study, it was identified as a superior system in evaluating corneal stabilization following corneal crosslinking [51]. The camera has also demonstrated excellent repeatability even after laser ablation [46], and thus its application includes monitoring post-operative changes that may be indicative of pre-keratoconus.

Limitations

At present, studies assessing the diagnostic accuracy of Galilei indices are limited and there are inconsistencies in the literature. Validation studies will further elucidate the Galilei indices and their role in screening of the refractive surgery candidates. Most importantly, the ability to accurately diagnose pre-keratoconus, during the screening process will lead to safer and better outcomes. As mentioned in the Pentacam review, creating a universal definition for these particular cases will also facilitate better evaluation of diagnostic indices.

The majority of published research compares Galilei to other Scheimpflug devices. However, there are varying levels of agreement amongst different devices. While many studies have found good concordance [52-59], others have recommended not to interchangeably use measurements calculated by various devices [60-63]. Further, some studies describe agreement for specific parameters, while finding significant differences for others [54, 64]. Given the disparate findings in the literature, there is insufficient data to conclude the devices’ measurements are interchangeable at this time. Even though systems like Pentacam and TMS-5 operate on the same Scheimpflug principles, the devices cannot reliably be used interchangeably. Development of indices that can be used on any device such as CLMI will help reduce the impact of this limitation on future studies.

While the current review focuses on the KC predictive indices calculated by the Galilei system, it is important to mention the value of elevation data in screening refractive surgery candidates. Galilei computed elevation data still plays an important role when evaluating for KC. As described by Jafarinasab and colleagues, both anterior and posterior elevation can discern clinical KC and pre-keratoconus [65]. Interestingly, their study found that a 3-mm zone discriminates KC, while the 7-mm zone optimally distinguishes pre-keratoconus. This finding may indicate that the early ectatic changes occur in the periphery, yet as the disease progresses, the central involvement becomes more prominent.

Looking Ahead

Recent studies indicate that combination indices have the highest diagnostic accuracy [9, 66]. It is likely that the future will include these metrics for screening the refractive surgery patient. As with Pentacam, we do not recommend the use of an individual index for discerning KC. Instead, the best approach in the clinical setting is to evaluate multiple indices. While individual indices, such as AAI, have diagnostic utility, the lack of validation studies hinders our ability to recommend a single index for refractive screening.

The additive value of optical coherence tomography and ultrasound assessments can lead to improved diagnostic accuracy of combination parameters [66]. Biomechanical evaluation when evaluating ectasia risk can be incorporated into screening. With the advent of technologies that assess biomechanical properties, the armamentarium of screening methods continues to grow [67]. While we address biomechanical properties independently in the third article of this series, the highest yield screening method is one that combines all available corneal surface data.

There are already many studies that have demonstrated the value of applying neural networks, and this is likely where the future of refractive screening is headed [4, 5, 9, 49]. While it will be decades until both patient and physician are comfortable with automated analysis, the application of artificial intelligence, specifically deep learning algorithms, may prove to be the best approach for diagnosing pre-keratoconus. Nevertheless, the value of clinical picture cannot be overstated. The refractive indices provided by Galilei and other Scheimpflug analyzers are best interpreted in the clinical context.

CONCLUSION

Corneal surface indices provided by the Galilei system are reliable parameters in the identification of keratoconus. However, as evidenced by our review, there is insufficient data to conclude the reliability of these parameters in screening of pre-keratoconus. In the first article of this series we arrived at a similar conclusion for the Pentacam system. When comparing these two camera systems, the literature is far more robust for the Pentacam camera. Thus, future investigations are required to determine the true diagnostic accuracy of Galilei indices in the diagnosis of pre-keratoconus. Based on the current literature, the best approach for screening of the refractive surgery candidate is a combination of refractive indices aimed towards prevention of pre-keratoconus and postoperative iatrogenic ectasia. In the third article of this series, the authors introduce biomechanical parameters that can also be used in supplementing corneal surface data. Ultimately, the greatest diagnostic accuracy is expected from a parameter that combines pachymetric, keratometric, elevation, and biomechanical data from various modalities of corneal imaging.

DISCLOSURE

Ethical issues have been completely observed by the authors. All named authors meet the International Committee of Medical Journal Editors (ICMJE) criteria for authorship of this manuscript, take responsibility for the integrity of the work as a whole, and have given final approval for the version to be published. No conflict of interest has been presented. Funding/Support: None

ACKNOWLEDGEMENTS

The authors would like to thank Dr. Tirth Shah and Dr. David Skanchy for their support and leadership in preparation of this review article.

References

  • 1.Galilei G4 Operator Manual. Port. Switzerland: Zeimer Ophthalmic Systems AG; 2016. [Google Scholar]
  • 2.Arce C. GALILEI: Map Interpretation Guide. Software V. Port. Switzerland: Zeimer Ophthalmic Systems AG; 2011. [Google Scholar]
  • 3.Shetty R, Rao H, Khamar P, Sainani K, Vunnava K, Jayadev C, et al. Keratoconus Screening Indices and Their Diagnostic Ability to Distinguish Normal From Ectatic Corneas. Am J Ophthalmol. 2017;181:140–8. doi: 10.1016/j.ajo.2017.06.031. [DOI] [PubMed] [Google Scholar]
  • 4.Feizi S, Yaseri M, Kheiri B. Predictive Ability of Galilei to Distinguish Subclinical Keratoconus and Keratoconus from Normal Corneas. J Ophthalmic Vis Res. 2016;11(1):8–16. doi: 10.4103/2008-322X.180707. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Smadja D, Touboul D, Cohen A, Doveh E, Santhiago MR, et al. Detection of subclinical keratoconus using an automated decision tree classification. Am J Ophthalmol. 2013;156(2):237–46 e1. doi: 10.1016/j.ajo.2013.03.034. [DOI] [PubMed] [Google Scholar]
  • 6.Demir S, Sonmez B, Yeter V, Ortak H. Comparison of normal and keratoconic corneas by Galilei Dual-Scheimpflug Analyzer. Cont Lens Anterior Eye. 2013;36(5):219–25. doi: 10.1016/j.clae.2013.04.001. [DOI] [PubMed] [Google Scholar]
  • 7.Golan O, Hwang ES, Lang P, et al. Differences in Posterior Corneal Features Between Normal Corneas and Subclinical Keratoconus. J Refract Surg. 2018;34(10):664–70. doi: 10.3928/1081597X-20180823-02. [DOI] [PubMed] [Google Scholar]
  • 8.Arce C. Qualitative and quantitative analysis of aspheric symmetry and asymmetry on corneal surfaces. Boston: MA: ASCRS Symposium and Congress; 2010. [Google Scholar]
  • 9.Golan O, Piccinini AL, Hwang ES, De Oca Gonzalez IM, Krauthammer M, Khandelwal SS, et al. Distinguishing Highly Asymmetric Keratoconus Eyes Using Dual Scheimpflug/Placido Analysis. Am J Ophthalmol. 2019;201:46–53. doi: 10.1016/j.ajo.2019.01.023. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Bayraktar Bilen N, Hepsen IF, Arce CG. Correlation between visual function and refractive, topographic, pachymetric and aberrometric data in eyes with keratoconus. Int J Ophthalmol. 2016;9(8):1127–33. doi: 10.18240/ijo.2016.08.07. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Smadja D, Touboul D, Colin J. Comparative Evaluation of Elevation, Keratometric, Pachymetric and Wavefront Parameters in Normal Eyes, Subclinical Keratoconus and Keratoconus with a Dual Scheimpflug Analyzer. Int J Kerat Ect Cor Dis. 2012;1(3):158–166. [Google Scholar]
  • 12.Schlegel Z, Hoang-Xuan T, Gatinel D. Comparison of and correlation between anterior and posterior corneal elevation maps in normal eyes and keratoconus-suspect eyes. J Cataract Refract Surg. 2008;34(5):789–95. doi: 10.1016/j.jcrs.2007.12.036. [DOI] [PubMed] [Google Scholar]
  • 13.Pinero DP, Nieto JC, Lopez-Miguel A. Characterization of corneal structure in keratoconus. J Cataract Refract Surg. 2012;38(12):2167–83. doi: 10.1016/j.jcrs.2012.10.022. [DOI] [PubMed] [Google Scholar]
  • 14.de Sanctis U, Loiacono C, Richiardi L, Turco D, Mutani B, Grignolo FM. Sensitivity and specificity of posterior corneal elevation measured by Pentacam in discriminating keratoconus/subclinical keratoconus. Ophthalmology. 2008;115(9):1534–9. doi: 10.1016/j.ophtha.2008.02.020. [DOI] [PubMed] [Google Scholar]
  • 15.Kamiya K, Ishii R, Shimizu K, Igarashi A. Evaluation of corneal elevation, pachymetry and keratometry in keratoconic eyes with respect to the stage of Amsler-Krumeich classification. Br J Ophthalmol. 2014;98(4):459–63. doi: 10.1136/bjophthalmol-2013-304132. [DOI] [PubMed] [Google Scholar]
  • 16.Saad A, Gatinel D. Topographic and tomographic properties of forme fruste keratoconus corneas. Invest Ophthalmol Vis Sci. 2010;51(11):5546–55. doi: 10.1167/iovs.10-5369. [DOI] [PubMed] [Google Scholar]
  • 17.Arbelaez MC, Versaci F, Vestri G, Barboni P, Savini G. Use of a support vector machine for keratoconus and subclinical keratoconus detection by topographic and tomographic data. Ophthalmology. 2012;119(11):2231–8. doi: 10.1016/j.ophtha.2012.06.005. [DOI] [PubMed] [Google Scholar]
  • 18.Boyd S, Gutierrez A, McCulley J. Atlas and text of corneal pathology and surgery. Ciudad de Panamá: Highlights Medical Publishers; 2011. [Google Scholar]
  • 19.Febbraro J-L, Khan H, Koch D. Surgical Correction of Astigmatism. Cham: Springer International Publishing; 2018. [Google Scholar]
  • 20.Barbara A e. Controversies in the Management of Keratoconus. Cham: Springer International Publishing; 2019. [Google Scholar]
  • 21.Twa M. Irregular Astigmatism: Diagnosis and Treatment. Optom Vis Sci. 2009;86(10):1209. [Google Scholar]
  • 22.Hernandez-Camarena JC, Chirinos-Saldana P, Navas A, Ramirez-Miranda A, de la Mota A, Jimenez-Corona A, et al. Repeatability, reproducibility, and agreement between three different Scheimpflug systems in measuring corneal and anterior segment biometry. J Refract Surg. 2014;30(9):616–21. doi: 10.3928/1081597X-20140815-02. [DOI] [PubMed] [Google Scholar]
  • 23.McMahon TT, Szczotka-Flynn L, Barr JT, Anderson RJ, Slaughter ME, Lass JH, et al. A new method for grading the severity of keratoconus: the Keratoconus Severity Score (KSS) Cornea. 2006;25(7):794–800. doi: 10.1097/01.ico.0000226359.26678.d1. [DOI] [PubMed] [Google Scholar]
  • 24.Bayhan HA, Aslan Bayhan S, Muhafiz E, Can I. Repeatability of aberrometric measurements in normal and keratoconus eyes using a new Scheimpflug-Placido topographer. J Cataract Refract Surg. 2014;40(2):269–75. doi: 10.1016/j.jcrs.2013.07.046. [DOI] [PubMed] [Google Scholar]
  • 25.Xu Z, Li W, Jiang J, Zhuang X, Chen W, Peng M, et al. Characteristic of entire corneal topography and tomography for the detection of sub-clinical keratoconus with Zernike polynomials using Pentacam. Sci Rep. 2017;7(1):16486. doi: 10.1038/s41598-017-16568-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Lakshminarayanan V, Fleck A. Zernike polynomials: a guide. Journal of Modern Optics. 2011;58(7):545–61. [Google Scholar]
  • 27.Gobbe M, Guillon M. Corneal wavefront aberration measurements to detect keratoconus patients. Cont Lens Anterior Eye. 2005;28(2):57–66. doi: 10.1016/j.clae.2004.12.001. [DOI] [PubMed] [Google Scholar]
  • 28.Buhren J, Kuhne C, Kohnen T. Defining subclinical keratoconus using corneal first-surface higher-order aberrations. Am J Ophthalmol. 2007;143(3):381–9. doi: 10.1016/j.ajo.2006.11.062. [DOI] [PubMed] [Google Scholar]
  • 29.L Alió J, H Shabayek M. Corneal Higher Order Aberrations: A Method to Grade Keratoconus. Journal of Refractive Surgery. 2006;22(6):539–45. doi: 10.3928/1081-597X-20060601-05. [DOI] [PubMed] [Google Scholar]
  • 30.Barbero S, Marcos S, Merayo-Lloves J, Moreno-Barriuso E. Validation of the estimation of corneal aberrations from videokeratography in keratoconus. J Refract Surg. 2002;18(3):263–70. doi: 10.3928/1081-597X-20020501-09. [DOI] [PubMed] [Google Scholar]
  • 31.Lim L, Wei RH, Chan WK, Tan DT. Evaluation of higher order ocular aberrations in patients with keratoconus. J Refract Surg. 2007;23(8):825–8. doi: 10.3928/1081-597X-20071001-13. [DOI] [PubMed] [Google Scholar]
  • 32.Reddy JC, Rapuano CJ, Cater JR, Suri K, Nagra PK, Hammersmith KM. Comparative evaluation of dual Scheimpflug imaging parameters in keratoconus, early keratoconus, and normal eyes. J Cataract Refract Surg. 2014;40(4):582–92. doi: 10.1016/j.jcrs.2013.08.061. [DOI] [PubMed] [Google Scholar]
  • 33.Tellouck J, Touboul D, Santhiago MR, Tellouck L, Paya C, Smadja D. Evolution Profiles of Different Corneal Parameters in Progressive Keratoconus. Cornea. 2016;35(6):807–13. doi: 10.1097/ICO.0000000000000833. [DOI] [PubMed] [Google Scholar]
  • 34.Rabinowitz YS, Rasheed K. KISA% index: a quantitative videokeratography algorithm embodying minimal topographic criteria for diagnosing keratoconus. J Cataract Refract Surg. 1999;25(10):1327–35. doi: 10.1016/s0886-3350(99)00195-9. [DOI] [PubMed] [Google Scholar]
  • 35.Wilson SE, Lin DT, Klyce SD. Corneal topography of keratoconus. Cornea. 1991;10(1):2–8. [PubMed] [Google Scholar]
  • 36.Maeda N, Klyce SD, Smolek MK. Comparison of methods for detecting keratoconus using videokeratography. Arch Ophthalmol. 1995;113(7):870–4. doi: 10.1001/archopht.1995.01100070044023. [DOI] [PubMed] [Google Scholar]
  • 37.Mahmoud AM, Roberts CJ, Lembach RG, Twa MD, Herderick EE, McMahon TT, et al. CLMI: the cone location and magnitude index. Cornea. 2008;27(4):480–7. doi: 10.1097/ICO.0b013e31816485d3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Feizi S, Delfazayebaher S, Javadi MA, Karimian F, Ownagh V, Sadeghpour F. Mean Posterior Corneal Power and Astigmatism in Normal Versus Keratoconic Eyes. J Ophthalmic Vis Res. 2018;13(2):93–100. doi: 10.4103/jovr.jovr_19_17. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Kocamis SI, Cakmak HB, Cagil N, Toklu Y. Investigation of the Efficacy of the Cone Location and Magnitude Index in the Diagnosis of Keratoconus. Semin Ophthalmol. 2016;31(3):203–9. doi: 10.3109/08820538.2014.914234. [DOI] [PubMed] [Google Scholar]
  • 40.Maeda N, Klyce SD, Smolek MK, Thompson HW. Automated keratoconus screening with corneal topography analysis. Invest Ophthalmol Vis Sci. 1994;35(6):2749–57. [PubMed] [Google Scholar]
  • 41.Shetty R, Arora V, Jayadev C, Nuijts RM, Kumar M, Puttaiah NK, et al. Repeatability and agreement of three Scheimpflug-based imaging systems for measuring anterior segment parameters in keratoconus. Invest Ophthalmol Vis Sci. 2014;55(8):5263–8. doi: 10.1167/iovs.14-15055. [DOI] [PubMed] [Google Scholar]
  • 42.Wang L, Shirayama M, Koch DD. Repeatability of corneal power and wavefront aberration measurements with a dual-Scheimpflug Placido corneal topographer. J Cataract Refract Surg. 2010;36(3):425–30. doi: 10.1016/j.jcrs.2009.09.034. [DOI] [PubMed] [Google Scholar]
  • 43.Aramberri J, Araiz L, Garcia A, Illarramendi I, Olmos J, Oyanarte I, et al. Dual versus single Scheimpflug camera for anterior segment analysis: Precision and agreement. J Cataract Refract Surg. 2012;38(11):1934–49. doi: 10.1016/j.jcrs.2012.06.049. [DOI] [PubMed] [Google Scholar]
  • 44.Rabinowitz YS. Videokeratographic indices to aid in screening for keratoconus. J Refract Surg. 1995;11(5):371–9. doi: 10.3928/1081-597X-19950901-14. [DOI] [PubMed] [Google Scholar]
  • 45.Wilson SE, Klyce SD. Advances in the analysis of corneal topography. Surv Ophthalmol. 1991;35(4):269–77. doi: 10.1016/0039-6257(91)90047-j. [DOI] [PubMed] [Google Scholar]
  • 46.Kim EJ, Montes de Oca I, Wang L, Weikert MP, Koch DD, Khandelwal SS. Repeatability of posterior and total corneal curvature measurements with a dual Scheimpflug-Placido tomographer. J Cataract Refract Surg. 2015;41(12):2731–8. doi: 10.1016/j.jcrs.2015.07.035. [DOI] [PubMed] [Google Scholar]
  • 47.Mauger TF, Mahmoud AM, Roberts CJ, Chheda LV, Kuennen RA, Hendershot AJ, et al. Comparison of Placido, Scheimpflug and Combined Dual Scheimpflug-Placido Technologies in Evaluating Anterior and Posterior CLMI, SimK's as well as Kmax, in Keratoconic and Postrefractive Surgery Ectasia. Int J Keratoconus Ectatic Corneal Dis. 2012;1(1):44–52. [Google Scholar]
  • 48.Mahmoud AM, Nunez MX, Blanco C, Koch DD, Wang L, Weikert MP, et al. Expanding the cone location and magnitude index to include corneal thickness and posterior surface information for the detection of keratoconus. Am J Ophthalmol. 2013;156(6):1102–11. doi: 10.1016/j.ajo.2013.07.018. [DOI] [PubMed] [Google Scholar]
  • 49.Maeda N, Klyce SD, Smolek MK. Neural network classification of corneal topography Preliminary demonstration. Invest Ophthalmol Vis Sci. 1995;36(7):1327–35. [PubMed] [Google Scholar]
  • 50.Savini G, Barboni P, Carbonelli M, Hoffer KJ. Accuracy of a dual Scheimpflug analyzer and a corneal topography system for intraocular lens power calculation in unoperated eyes. J Cataract Refract Surg. 2011;37(1):72–6. doi: 10.1016/j.jcrs.2010.08.036. [DOI] [PubMed] [Google Scholar]
  • 51.Sadoughi MM, Feizi S, Delfazayebaher S, Baradaran-Rafii A, Einollahi B, Shahabi C. Corneal Changes After Collagen Crosslinking for Keratoconus Using Dual Scheimpflug Imaging. J Ophthalmic Vis Res. 2015;10(4):358–63. doi: 10.4103/2008-322X.176894. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52.Fahd DC, Cherfan CG, Raad C, Asouad M, Awwad ST. Assessment of anterior and posterior corneal indices using two Scheimpflug analyzers. Arq Bras Oftalmol. 2014;77(1):17–20. doi: 10.5935/0004-2749.20140006. [DOI] [PubMed] [Google Scholar]
  • 53.Menassa N, Kaufmann C, Goggin M, Job OM, Bachmann LM, Thiel MA. Comparison and reproducibility of corneal thickness and curvature readings obtained by the Galilei and the Orbscan II analysis systems. J Cataract Refract Surg. 2008;34(10):1742–7. doi: 10.1016/j.jcrs.2008.06.024. [DOI] [PubMed] [Google Scholar]
  • 54.Baradaran-Rafii A, Motevasseli T, Yazdizadeh F, Karimian F, Fekri S, Baradaran-Rafii A. Comparison between Two Scheimpflug Anterior Segment Analyzers. J Ophthalmic Vis Res. 2017;12(1):23–9. doi: 10.4103/jovr.jovr_104_16. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 55.Salouti R, Nowroozzadeh MH, Zamani M, Ghoreyshi M, Salouti R. Comparison of anterior chamber depth measurements using Galilei, HR Pentacam, and Orbscan II. Optometry. 2010;81(1):35–9. doi: 10.1016/j.optm.2009.04.100. [DOI] [PubMed] [Google Scholar]
  • 56.Salouti R, Nowroozzadeh MH, Zamani M, Fard AH, Niknam S. Comparison of anterior and posterior elevation map measurements between 2 Scheimpflug imaging systems. J Cataract Refract Surg. 2009;35(5):856–62. doi: 10.1016/j.jcrs.2009.01.008. [DOI] [PubMed] [Google Scholar]
  • 57.de Jong T, Sheehan MT, Koopmans SA, Jansonius NM. Posterior corneal shape: Comparison of height data from 3 corneal topographers. J Cataract Refract Surg. 2017;43(4):518–24. doi: 10.1016/j.jcrs.2017.03.021. [DOI] [PubMed] [Google Scholar]
  • 58.Lee YW, Choi CY, Yoon GY. Comparison of dual rotating Scheimpflug-Placido, swept-source optical coherence tomography, and Placido-scanning-slit systems. J Cataract Refract Surg. 2015;41(5):1018–29. doi: 10.1016/j.jcrs.2014.08.040. [DOI] [PubMed] [Google Scholar]
  • 59.Park SH, Choi SK, Lee D, Jun EJ, Kim JH. Corneal thickness measurement using Orbscan, Pentacam, Galilei, and ultrasound in normal and post-femtosecond laser in situ keratomileusis eyes. Cornea. 2012;31(9):978–82. doi: 10.1097/ICO.0b013e31823d03fc. [DOI] [PubMed] [Google Scholar]
  • 60.Huang J, Savini G, Wang C, Lu W, Gao R, Li Y, et al. Precision of corneal thickness measurements obtained using the scheimpflug-placido imaging and agreement with ultrasound pachymetry. J Ophthalmol. 2015;2015:328798. doi: 10.1155/2015/328798. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 61.Crawford AZ, Patel DV, McGhee CN. Comparison and repeatability of keratometric and corneal power measurements obtained by Orbscan II, Pentacam, and Galilei corneal tomography systems. Am J Ophthalmol. 2013;156(1):53–60. doi: 10.1016/j.ajo.2013.01.029. [DOI] [PubMed] [Google Scholar]
  • 62.Hsieh YH, Weng TH, Chou YC, Wu KL, Liang CM, Tai MC. Agreement of post-LASIK corneal power and corneal thickness measurements by pentacam and GALILEI corneal tomography systems. J Chin Med Assoc. 2018 doi: 10.1016/j.jcma.2018.07.005. [DOI] [PubMed] [Google Scholar]
  • 63.Anayol MA, Guler E, Yagci R, Sekeroglu MA, Ylmazoglu M, Trhs H, et al. Comparison of central corneal thickness, thinnest corneal thickness, anterior chamber depth, and simulated keratometry using galilei, Pentacam, and Sirius devices. Cornea. 2014;33(6):582–6. doi: 10.1097/ICO.0000000000000119. [DOI] [PubMed] [Google Scholar]
  • 64.Jahadi Hosseini HR, Katbab A, Khalili MR, Abtahi MB. Comparison of corneal thickness measurements using Galilei, HR Pentacam, and ultrasound. Cornea. 2010;29(10):1091–5. doi: 10.1097/ICO.0b013e3181cf98e5. [DOI] [PubMed] [Google Scholar]
  • 65.Jafarinasab MR, Shirzadeh E, Feizi S, Karimian F, Akaberi A, Hasanpour H. Sensitivity and specificity of posterior and anterior corneal elevation measured by orbscan in diagnosis of clinical and subclinical keratoconus. J Ophthalmic Vis Res. 2015;10(1):10–5. doi: 10.4103/2008-322X.156085. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 66.Hwang ES, Perez-Straziota CE, Kim SW, Santhiago MR, Randleman JB. Distinguishing Highly Asymmetric Keratoconus Eyes Using Combined Scheimpflug and Spectral-Domain OCT Analysis. Ophthalmology. 2018;125(12):1862–71. doi: 10.1016/j.ophtha.2018.06.020. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 67.Belin MW, Khachikian SS. New devices and clinical implications for measuring corneal thickness. Clin Exp Ophthalmol. 2006;34(8):729–31. doi: 10.1111/j.1442-9071.2006.01395.x. [DOI] [PubMed] [Google Scholar]

Articles from Medical Hypothesis, Discovery and Innovation in Ophthalmology are provided here courtesy of International Virtual Ophthalmic Research Center