Skip to main content
Journal of Clinical Laboratory Analysis logoLink to Journal of Clinical Laboratory Analysis
. 2014 Nov 10;29(5):383–386. doi: 10.1002/jcla.21783

Evaluation of a Rapid Immunochromatographic Treponemal Antibody Test Comparing the Treponema Pallidum Particle Agglutination Assay

Jong‐Han Lee 1, Chae Seung Lim 1, Min‐Geol Lee 2, Hyon‐Suk Kim 3,
PMCID: PMC6807139  PMID: 25385043

Abstract

Background

In addition to conventional tests, several methods for detection of treponema‐specific antibodies in clinical settings have been recently introduced. We aim to comparatively evaluate a rapid immunochromatographic test (ICT) for Treponema pallidum specific antibody (SD Bioline Syphilis 3.0) and the T. pallidum particle agglutination (TPPA) assay.

Methods

In all, 132 serum samples from 78 syphilis patients and 54 syphilis‐negative controls were analyzed. SD Bioline Syphilis 3.0 test (Standard Diagnostic, Inc., Yongin, Korea) was evaluated and compared to Serodia TPPA assay (Fujirebio, Inc., Tokyo, Japan). All discrepant results between the two assays were repeatedly tested and evaluated by the fluorescent treponemal antibody‐absorption (FTA‐ABS) assay. Test reproducibility and 95% limit of detection of SD Bioline Syphilis 3.0 were determined across three different lots for seven consecutive days in triplicate. Interference due to autoantibodies and pregnancy was also tested.

Results

Percent agreement between SD Bioline Syphilis 3.0 and TPPA assays was 99.2%. Sensitivity and specificity were 100%, respectively. In TPPA assay, test‐to‐test, day‐to‐day, and lot‐to‐lot variations were not identified until 1:320 titer (eightfold dilutions). There was no interference due to the presence of antinuclear antibodies or samples or pregnancy.

Conclusions

Percent agreement of SD Syphilis 3.0 and TPPA was very good. Sensitivity and specificity were appropriate for T. pallidum antibody detection. Thus, a rapid ICT could be suitable for syphilis antibody detection.

Keywords: Syphilis, rapid immunochromatographic test, Treponema pallidum particle agglutination (TPPA), agreement


Syphilis is one of the major infectious diseases and was recently found in about 11 million adults among developing countries in 2005 1. In Korea, the prevalence of syphilis has rapidly decreased since the 1970s, and it was approximately 0.2% in 2000. However, recent studies have suggested a slight increase in the prevalence, which later resulted in a plateau 2. Despite this finding, primary and secondary syphilis in immunocompromised patients have increased worldwide in the last decade 3. Therefore, appropriate screening, diagnosis, and treatment protocols are needed and well established 4, 5, 6. A recently highlighted concern is that traditional diagnostic approaches should be replaced by a reverse syphilis testing algorithm. The reverse algorithm is a treponemal‐specific test applied as a screening tool followed by a quantitative nontreponemal assay that diagnoses active disease and monitors patients’ treatment response 7. Of the various treponemal assays, the Treponema pallidum particle agglutination (TPPA) assay is more widely used than the fluorescent treponemal antibody‐absorption (FTA‐ABS) test for confirmatory testing of syphilis 8. The aim of this study was to evaluate a rapid immunochromatographic test (ICT), SD Bioline Syphilis 3.0, for its clinical usage.

A total of 132 serum samples from 78 patients with confirmed syphilis and 54 healthy nonsyphilis control specimens were collected and preserved at −70°C until analyses. Accurate analyses of the patients according to the series of syphilis stage could not be carried out owing to certain characteristics and infrequency of the disease in Korea. A rapid ICT, SD Bioline Syphilis 3.0 (Standard Diagnostic, Inc., Yongin, Korea), was evaluated and compared to the Serodia TPPA assay (Fujirebio, Inc., Tokyo, Japan). SD Bioline Syphilis 3.0 is a solid‐phase ICT used for the detection of antibodies of T. pallidum. The kit contains reagents and membrane strips that are precoated with three kinds of recombinant T. pallidum antigens, including rTpN15, rTpN17, and rTpN47, on the test band region. Binding of these antigens to the antibodies present in the patient's sample and the secondary colloidal gold conjugate leads to the formation of visible bands. Serodia TPPA assay is based on the agglutination of colored gelatin particles that have been sensitized (coated) with T. pallidum (Nichols strain) antigens. For quantitative results, the antibody titer was determined as the final dilution showing a positive result. A final dilution of 1:80 or more was considered as a positive reaction. Discrepancy between SD Bioline Syphilis 3.0 and TPPA assay results were repeatedly tested by different technologists who simultaneously performed the FTA‐ABS assay. The FTA‐ABS assay is performed using a commercially available test kit (Zeus Scientific, Raritan, NJ) in which T. pallidum substrate is fixed on a glass slide according to the manufacturer's protocol. Percent agreement was calculated as kappa values, which were used to categorize the result as very good (0.81–1.0), good (0.61–0.8), moderate (0.41–0.6), fair (0.21–0.4), or poor (0–0.2) 9. The sensitivity and specificity of the assay were analyzed by defining true positive cases that were based on the results of TPPA and FTA‐ABS assays and the patients’ medical chart review. Reproducibility of SD Bioline Syphilis 3.0 test was analyzed according to modified CLSI guidelines EP12‐A2 10. Reproducibility and limit of detection were determined by repeating the tests with two‐fold serial dilutions of three kinds of pooled positive serum samples diluted up to 1:128 (1:1, 1:2, 1:4, 1:8, 1:16, 1:32, 1:64, and 1:128), reflecting the treponemal antibody concentrations. We performed the tests in triplicate over a period of seven consecutive days using three different lots (lot numbers: 025103A, 025118C, and 025120A). In all, a total of 504 tests were carried out for this evaluation. Fifteen antinuclear antibody (ANA) positive samples and ten samples from pregnant women were additionally tested for production of any false reactions by SD Bioline Syphilis 3.0.

The percent agreement between SD Bioline Syphilis 3.0 and Serodia TPPA was 99.2% (kappa: 0.984, 95% CI: 0.954–1.015). Only one discrepant result was observed among 132 subjects (Table 1). This discrepant sample showed a negative SD Bioline Syphilis 3.0 and positive result (1:320) of Serodia TPPA assays, respectively, and the sample showed minimally reactive with the FTA‐ABS test. This sample originated from a 61‐year‐old male patient who was suspected of herpes simplex virus infection or drug effect rather than syphilis infection based on his dermatologist's review. The sensitivity and specificity of SD Bioline Syphilis 3.0 was 100% (95% CI: 95.3–100), 100% (95% CI: 93.3–100) and Serodia TPPA assay was 100% (95% CI: 95.3–100), 98.2% (95% CI: 90.1–99.7), respectively. The reproducibility among replica specimens, days, and lots showed 100% positivity with the pooled TPPA‐positive serum at titer values of 1:2560 (onefold), 1:1280 (twofold), 1:640 (fourfold), and 1:320 (eightfold). However, titer values of 1:160 (16‐fold), 1:80 (32‐fold), 1:40 (64‐fold), 1:20 (128‐fold), 1:10 (256‐fold), and 1:5 (512‐fold) of the pooled serum showed 88.9%, 85.7%, 25.4%, 1.6%, 1.6%, and 0% positivity, respectively. The 95% limit of detection of SD Bioline Syphilis 3.0 was between 1:40 titer (64‐fold) and 1:20 titer (128‐fold), and 50% limit of detection was between 1:80 titers (32‐fold) and 1:40 titers (64‐fold) of TPPA, respectively. No false‐positive results were obtained due to the presence of ANAs in 15 patients and samples from ten pregnant women.

Table 1.

Percent Agreement and Kappa Values of Rapid SD Bioline and TPPA Assays

TPPA True syphilis infection
+ +
SD Bioline‐Syphilis 3.0 + 78 0 78 0
1 53 0 54

Percent agreement: 99.2%.

Kappa value: 0.984.

95% confidence interval: 0.954–1.015.

The strength of agreement is considered to be very good.

TPPA, Treponema pallidum particle agglutination.

In secondary and latent stages of syphilis except primary syphilis, the treponemal test is very sensitive and approximately 84% of cases of syphilis remain treponemal test positive for life 8. In clinical settings, the commonly used methods for detection of T. pallidum antibodies are T. pallidum hemagglutination (TPHA), TPPA, and FTA‐ABS assays. TPPA has been reported nearly similar sensitivity as the FTA‐ABS assay in primary syphilis cases, and it is as useful as the rapid plasma regain (RPR) test in monitoring therapy 11. It is less subjective than FTA‐ABS assay and the results are easier to interpret than the classical microhemagglutination assay for antibodies to T. pallidum (MHA‐TP) 11. TPPA assay was also reported to be useful in the diagnosis of neurosyphilis using cerebrospinal fluid (CSF) samples 12. The rapid ICT for syphilis detection has some advantages, such as completion of the test within 30 min without the use of complex equipments and technical skills. We found that the concordance rate of SD Bioline Syphilis 3.0 and Serodia TPPA assays was very good. The overall diagnostic performances of these two assays were the same or even superior to that of the TPPA assay. A previous multicenter evaluation reported that SD Bioline Syphilis 3.0 showed a diagnostic sensitivity of 95% (95% CI: 92.8–97.1) and a specificity of 94.9% (95% CI: 92.7–97.1) 13. It might depend on the study population or syphilis stages. Our results showed the highest sensitivity and specificity compared to the results of any of the previous multicenter reports 13. One report from the Reproductive Health and Research Department of WHO comparatively evaluated six different syphilis rapid diagnostic tests by using TPHA assay as a reference; SD Bioline Syphilis 3.0 showed 100% sensitivity and specificity, similar to the result of our study 14. In addition, no false‐positive reactions due to the presence of ANAs or samples from pregnant women were observed in our study. There was only a single case of discrepant result between SD Bioline 3.0 and TPPA assays. The discrepant case produced a negative result of the rapid SD Bioline 3.0 assay and a positive result of the TPPA assay, and the FTA‐ABS test showed minimally reactive. These discrepant results were obtained in a sample from a 61‐year‐old male patient who had dermatological problems with no syphilis history confirmed by his dermatologist. Therefore, we considered that this patient was not infected with syphilis and TPPA, and FTA‐ABS results were dual false positive. Serological false‐positive results of syphilis test are common and have been documented in the past 15, 16, 17, 18, 19, 20. The main causes of false‐positive results included systemic lupus erythematosus 15, diabetes mellitus 16, HIV infection 17, 18, leprosy 19, and prozone phenomenon 20. We did not notice any definite false interference of the rapid ICT used in our study. However, false‐positive results of the treponemal tests were reported as often transient, and the underlying causes have not been fully understood 8. In United States and Europe, several recent reports have suggested that treponemal‐specific tests should be used as the first‐line serologic screening test of syphilis for their high sensitivity and specificity 5, 6, 7. However, Centers for Disease Control and Prevention (CDC) still have recommended screening with a nontreponemal test such as RPR test 4. Unfortunately, in this study, we could not categorize the patients according to the syphilis stages. Treponemal tests can vary in their reactivity in early primary syphilis, but in secondary and latent stages of syphilis the treponemal tests are usually 100% reactive 8.

In conclusion, SD Bioline Syphilis 3.0 showed an excellent performance compared to the TPPA assay. It can be suggested that this rapid ICT can be applied in clinical settings considering a simple test procedure, and rapid result with comparable performance to the previous treponemal tests.

ACKNOWLEDGEMENT

We are grateful to Standard Diagnostic, Inc. for kindly providing the rapid test kits used in this study.

REFERENCES

  • 1. World Health Organization . Prevalence and Incidence of Selected Sexually Transmitted Infections, Chlamydia trachomatis, Neisseria gonorrhoeae, Syphilis and Trichomonas vaginalis: Methods and Results Used by WHO to Generate 2005 estimates. Geneva, Switzerland: World Health Organization. [Google Scholar]
  • 2. Cho YH, Kim HO, Lee JB, Lee MG. Syphilis prevalence has rapidly decreased in South Korea. Sex Transm Infect 2003;79:323–324. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Mattei PL, Beachkofsky TM, Gilson RT, Wisco OJ. Syphilis: A reemerging infection. Am Fam Physician 2012;86:433–440. [PubMed] [Google Scholar]
  • 4. Centers for Disease CP . Discordant results from reverse sequence syphilis screening—Five laboratories, United States, 2006–2010. Morb Mortal Wkly Rep 2011;60:133–137. [PubMed] [Google Scholar]
  • 5. Egglestone SI, Turner AJ. Serological diagnosis of syphilis. PHLS Syphilis Serology Working Group. Commun Dis Public Health 2000;3:158–162. [PubMed] [Google Scholar]
  • 6. French P, Gomberg M, Janier M, et al. IUSTI: 2008 European Guidelines on the management of syphilis. Int J STD AIDS 2009;20:300–309. [DOI] [PubMed] [Google Scholar]
  • 7. Loeffelholz MJ, Binnicker MJ. It is time to use treponema‐specific antibody screening tests for diagnosis of syphilis. J Clin Microbiol 2012;50:2–6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Detrick B, Hamilton RG, Folds JD. Manual of Molecular and Clinical Laboratory Immunology, seventh ed., Washington, DC: ASM Press; 2006, p 477–492. [Google Scholar]
  • 9. Landis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics 1977;33:159–174. [PubMed] [Google Scholar]
  • 10. Clinical and Laboratory Standards Institute . User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline, second edition. CLSI Document EP12‐A2. Wayne, PA: Clinical and Laboratory Standards Institute; 2008. [Google Scholar]
  • 11. Castro RR, Prieto ES, Santo I, Azevedo J, Exposto FL. Evaluation of the passive particle agglutination test in the serodiagnosis and follow‐up of syphilis. Am J Clin Pathol 2001;116:581–585. [DOI] [PubMed] [Google Scholar]
  • 12. Castro R, Prieto ES, Aguas MJ, et al. Evaluation of the Treponema pallidum particle agglutination technique (TP.PA) in the diagnosis of neurosyphilis. J Clin Lab Anal 2006;20:233–238. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Herring AJ, Ballard RC, Pope V, et al. A multi‐centre evaluation of nine rapid, point‐of‐care syphilis tests using archived sera. Sex Transm Infect 2006; 82(Suppl 5):v7–v12. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14. Broutet N. Strategy for the elimination of congenital syphilis. Available from: http://www.gfmer.ch/PGC_RH_2005/pdf/Syphilis.pdf. Access date September 22, 2014.
  • 15. Contreras MA, Andreu JL, Isasi C, Barbadillo C. False positive treponemal test result in a patient with active systemic lupus erythematosus. J Rheumatol 2000;27:2059. [PubMed] [Google Scholar]
  • 16. Brauner A, Carlsson B, Sundkvist G, Ostenson CG. False‐positive treponemal serology in patients with diabetes mellitus. J Diabetes Complications 1994;8:57–62. [DOI] [PubMed] [Google Scholar]
  • 17. Augenbraun MH, DeHovitz JA, Feldman J, Clarke L, Landesman S, Minkoff HM. Biological false‐positive syphilis test results for women infected with human immunodeficiency virus. Clin Infect Dis 1994;19:1040–1044. [DOI] [PubMed] [Google Scholar]
  • 18. Rompalo AM, Cannon RO, Quinn TC, Hook EW, 3rd . Association of biologic false‐positive reactions for syphilis with human immunodeficiency virus infection. J Infect Dis 1992;165:1124–1126. [DOI] [PubMed] [Google Scholar]
  • 19. Cabrini JM, Bottasso OA, Margasin S, Schujman L, Mangiaterra L, Morini JC. Biological false positive test for syphilis in lepromatous leprosy patients with concomitant hepatitis B virus infection. J Investig Allergol Clin Immunol 1991;1:45–48. [PubMed] [Google Scholar]
  • 20. Berkowitz K, Baxi L, Fox HE. False‐negative syphilis screening: The prozone phenomenon, nonimmune hydrops, and diagnosis of syphilis during pregnancy. Am J Obstet Gynecol 1990;163:975–977. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Clinical Laboratory Analysis are provided here courtesy of Wiley

RESOURCES