Abstract
Background
We evaluated the effects of pneumatic tube system (PTS) transport rates and distances on routine hematology and coagulation analysis. PTS effects on centrifuged blood samples were also examined.
Method
The study was completed at Dicle University Hospital, which has the longest pneumatic tube system in Turkey. Blood samples were collected at three different locations within the hospital and an emergency department, and delivered to the central laboratory by the PTS or a human carrier. Samples were transported at different rates and over varying distances. Each specimen's potassium (K) and lactic dehydrogenase (LDH) levels, in both the serum and plasma, were tracked to monitor hemolysis. Measurements of LDH and K were obtained using heparin or citrate.
Result
A positive correlation was observed between distance and hemolysis in serum samples transported at 4.2 m/sec, and at 3.1 m/sec for more than 2200 m (r = 0.774 and r = 0.766, respectively). Distance and hemolysis were also correlated in non‐centrifuged samples (r = 0.871). The alterations in plasma LDH and K levels at different rates and PTS lengths were not statistically significant.
Conclusion
The rate of hemolysis in PTS transported samples, dependent on PTS length and rate, may seriously affect routine tests of non‐centrifuged samples.
Keywords: tube system, rate, length, hemolysis
INTRODUCTION
Pneumatic tube systems (PTSs) are widely used in hospitals for convenience and reduced turnaround times, but they have also been implicated in hemolysis in transported blood specimens 1. During transport in a PTS, blood samples are often subjected to high speeds and rapid acceleration and deceleration, which can lead to hemolysis. Hemolysis, in turn, potentially affects laboratory parameters, such as lactate dehydrogenase (LDH) and potassium (K) levels 2, 3. Using gel in collection test tubes may provide protection against hemolysis 4, and, therefore, centrifuged sera are likely less prone to hemolysis during PTS transport. Several studies have assessed the level of hemolysis observed in blood samples transported through a PTS, but none have evaluated the role of transport distance and rate. In our study, we report the effects of different PTS rates and distances on routine hematology and coagulation test samples obtained from healthy volunteers. We also evaluate PTS effects on centrifuged samples.
MATERIALS AND METHODS
Tube System Design
The PTS installed at Dicle University Hospital is a computer‐controlled, branched system of 11‐cm diameter tubes supplied by Swisslog Healthcare Solutions Division, Germany (trading under the name Simeks in Turkey). Three hospitals—the children's hospital, the cardiology institute, and the oncology hospital—are connected to the central laboratory via the pneumatic tube delivery system, which is the longest system in Turkey. System specifications are detailed in Table 1. The system is capable of running at multiple speeds and requires several transfer stations. The force of the PTS is provided by positive and negative air pressure. Samples are placed in plastic containers, or “tube carriers,” made of high‐impact resistant polycarbonate. The tube carriers contain removable padded liners and the sample carriers (“pods”) have protective circular felt or “O” rings on the outside.
Table 1.
Descriptive Properties of the Pneumatic Tube Delivery System
| Emergency service | Oncology hospital | Cardiology institute | Children hospital | |
|---|---|---|---|---|
| Distance to central laboratory (m) | 500 | 1,380 | 1,450 | 1,100 |
| Monthly volume of visit (n) | 2,170 | 2,080 | 1,100 | 740 |
| Transport duration (min) | 6 | 12 | 10 | 11 |
| Transfer station (n) | 1 | 2 | 1 | 2 |
Specimens and Transport
In this study, we used blood samples from 40 healthy volunteers. Each volunteer gave ten vacutainer tubes of blood obtained from a single venipuncture performed by a phlebotomist. These samples were sent to the central laboratory from four different locations, three hospitals and an emergency department, via the PTS and a porter. Blood tubes contained separating gel without anticoagulant, heparin, K2EDTA, or citrate (BD Medical Systems, Franklin Lakes, NJ). The first group, containing all sample types, was immediately transported by a porter. K and LDH levels were measured and accepted as reference values. The second group of samples, containing separating gel without anticoagulant, was divided into two subgroups—one was sent to the central laboratory via the PTS before centrifugation and one after. The third group of samples, with anticoagulant, was sent to the central laboratory via the PTS. All samples were transported via the PTS at two different rates, 3.1 m/sec and 4.2 m/sec.
Sample Analysis
Assays were performed within 1 hr of collection, after centrifugation of the paired specimens at 1500 g for 10 min. Hemolysis was monitored by tracking K concentration and LDH activity. K and LDH levels were determined in both the serum and the plasma using heparin and citrate and an autoanalyzer, Architect 1600c (Abbott Laboratories, Irving, TX). Only LDH levels were measured in tubes with K2EDTA. Hemogram parameters—including white blood cell counts, hemoglobin, red blood cell counts, and platelets—were measured using Cell‐Dyn 3700 (Abbott Labs, Dallas, TX). Assays for prothrombin time (PT), activated partial thromboplastin time (PTT), and fibrinogen were measured on a SysmexCA‐7000 coagulation analyzer (Sysmex, Kobe, Japan) using Thromborel S, actin, and thrombin reagent (Dade Behring, Marburg, Germany), respectively.
Statistical Analysis
The nonparametric Mann–Whitney U test was used to evaluate the significance of differences between groups. Results were considered statistically significant given a P‐value of < 0.05. Correlations were computed using Spearman's test.
RESULTS
K and LDH levels increased significantly in samples transported at a rate of 4.2 m/sec. At a rate of 3.1 m/sec, the increase was not significantly different for those delivered by the PTS and those delivered by hand. However, when sample transportation was repeated twice at 3.1 m/sec, for a total distance of more than 2200 m, a positive correlation between distance and hemolysis (r = 0.716, r = 0.628, respectively) was observed. K and LDH increases were then also significantly higher for PTS samples than for hand‐delivered samples (Fig. 1). A positive correlation between distance and levels of K and LDH was also found in noncentrifuged serum samples (r = 0.871 and r = 0.774, respectively) transported at a rate of 4.2 m/sec.
Figure 1.

Serum lactic dehydrogenase (LDH) (U/L) and potassium (K) (mEq/L) values in different rates and lengths transports by PTS before centrifugation.
No statistically significant differences between PTS samples and samples delivered by hand were observed in routine complete blood cell counts, white cell differential parameters, or markers of platelet activation. There were also no statistically significant differences in PT, activated PTT, and fibrinogen levels between samples delivered by PTS and by hand. Plasma LDH and K levels did not significantly change, given different PTS rates and lengths. The severity of hemolysis (determined by levels of LDH and K) was not significantly different in centrifuged samples transported by PTS and human carrier, regardless of rate and distance (Tables 2 and 3).
Table 2.
Mean values of Hemogram, Coagulation, and Hemolysis Parameters in Different Rates and Lengths of the Pneumatic Tube Delivery System
| Rate of PTS | 3.1 (m/sec) | 4.2 (m/sec) | |||||||
|---|---|---|---|---|---|---|---|---|---|
| Length of PTS (m) | § | 500 | 1,100 | 1,450 | 2,900 | 500 | 1,100 | 1,450 | 2,900 |
| WBC (K/μl) | 6.31 | 6.45 | 6.62 | 5.92 | 6.33 | 6.25 | 6.20 | 6.04 | 6.14 |
| RBC (M/ μl) | 5.03 | 4.99 | 5.01 | 5.04 | 4.98 | 4.81 | 4.90 | 5.04 | 4.85 |
| HGB (g/dl) | 12.9 | 12.7 | 12.7 | 12.9 | 13.0 | 12.8 | 12.7 | 12.9 | 12.8 |
| PLT (K/μl) | 219 | 212 | 225 | 205 | 207 | 218 | 213 | 222 | 227 |
| PTT (sec) | 11.5 | 11.5 | 11.6 | 11.7 | 11.3 | 11.2 | 11.2 | 11.4 | 11.3 |
| aPTT (sec) | 28.7 | 28.4 | 30.9 | 27.3 | 28.6 | 29.1 | 29.3 | 27.8 | 28.4 |
| K* (mEq/L) | 3.8 | 3.8 | 3.8 | 3.9 | 3.9 | 3.8 | 3.9 | 3.9 | 4.0 |
| K† (mEq/L) | 3.8 | 3.8 | 3.9 | 3.9 | 3.9 | 3.8 | 3.9 | 4.0 | 4.0 |
| K‡ (mEq/L) | 3.8 | 3.8 | 3.8 | 3.8 | 3.9 | 3.8 | 3.8 | 3.9 | 3.9 |
| LDH* (U/L) | 215 | 217 | 222 | 221 | 228 | 218 | 225 | 232 | 245 |
| LDH† (U/L) | 217 | 223 | 229 | 232 | 237 | 219 | 231 | 235 | 248 |
| LDH‡ (U/L) | 213 | 216 | 224 | 221 | 218 | 217 | 226 | 228 | 223 |
*Plasma obtained by heparin.
†Plasma obtained by citrate.
‡Serum obtained by centrifugation before PTS.
§Transported by a carrier.
Table 3.
Mean Values of K and LDH in Different Rates and Lengths of the Pneumatic Tube Delivery System
| Rate of PTS | ‡ | 3.1 (m/sec) | 4.2 (m/sec) | ||||||
|---|---|---|---|---|---|---|---|---|---|
| Length of PTS (m) | 500 | 1,100 | 1,450 | 2,900 | 500 | 1,100 | 1,450 | 2,900 | |
| K* (mEq/L) | 3.8 | 3.8 | 3.8 | 3.8 | 3.9 | 3.8 | 3.8 | 3.9 | 3.9 |
| K† (mEq/L) | 3.8 | 3.9 | 4.0 | 4.0 | 4.3§ | 4.0 | 4.1 | 4.2§ | 4.7§ |
| LDH* (U/L) | 213 | 216 | 224 | 221 | 218 | 217 | 226 | 228 | 223 |
| LDH† (U/L) | 221 | 235 | 241 | 253 | 352§ | 259§ | 289§ | 372§ | 406§ |
*Serum obtained by centrifugation before PTS.
†Serum obtained by centrifugation after PTS.
‡Transported by a carrier.
§ P < 0.05 compared with samples transported by a carrier.
DISCUSSION
Using a PTS system to transport blood samples can significantly reduce turn around times for test results without reducing sample quality 5. Laboratories should be aware that defects may arise in tube systems, which could cause rapid sample deceleration and excessive hemolysis 6. The pressure applied to samples during transport should also be considered, as it plays a pivotal role in hemolysis. The length of the PTS and the rate of transport increase the pressure on the erythrocytes, resulting in hemolysis 7. These considerations must be taken into account when designing new PTSs. Different hospitals will have varying system configurations and use different sample types. Sodi et al. recommended that each hospital investigate their own system to assess whether hemolysis is a recurring problem in any of the transported sample types 8.
In our hospitals, the longest transport distance is 1,450 m to the central laboratory. The manufacturer warned us about the possibility of PTS‐related hemolysis. According to the literature, the mean PTS rate is approximately 7.6 m/sec, but the optimal length is unknown 9. When we increased the rate of tube transport, hemolysis rates also increased. The optimum rate, calculated from the Figure 1, is 3.1 m/sec, assuming a travel distance below 2200 m. We were not satisfied with these results, as samples that are processed incorrectly (for example, without barcodes or with incorrect barcodes), or whose carriers are sent to wrong locations in the PTS, may elongate the transport distance. After centrifugation of gel containing tubes, hemolysis rates were the same as the reference values, independent of the length of the PTS.
Interestingly, the hemolysis rate of plasma samples increased very slowly as the PTS transport rate and distance increased (Table 2). In nonclotting samples, the hydrodynamic structure of blood may compensate for the pressure of environmental factors, but when clotting occurs in blood without gel separation, the hemolysis rate increases as clots hit the tube wall during transport.
Some researchers found that pneumatic tube sample transport impairs platelet aggregation 10, aspirin responsiveness 11, and spectrophotometric analysis of cerebrospinal fluid samples 12. The rate and length of the PTS may be affecting test reliability; more detailed studies are required.
In conclusion, transport through a PTS has no clinically significant effect on hematology and coagulation results when serum is centrifuged prior to transport. Each hospital should investigate their own system to assess whether hemolysis is a problem for any of their sample types under PTS conditions.
REFERENCES
- 1. Stair TO, Howell JM, Fitzgerald DJ, Bailey SC, Bastasch MD. Hemolysis of blood specimens transported from ED to laboratory by pneumatic tube. Am J Emerg Med 1995;13:484. [DOI] [PubMed] [Google Scholar]
- 2. Poznanski W, Smith F, Bodley F. Implementation of a pneumatic‐tube system for transport of blood specimens. Am J Clin Pathol 1978;70:291–295. [DOI] [PubMed] [Google Scholar]
- 3. Zanner R, Moser N, Blobner M, Luppa PB. Transport of blood gas samples: Is the pneumatic tube system safe?. Anaesthesist 2006;55:1099–1104. [DOI] [PubMed] [Google Scholar]
- 4. Sodi R, Darn SM, Stott A. Pneumatic tube system induced haemolysis: Assessing sample type susceptibility to haemolysis. Ann Clin Biochem 2004;41:237–240. [DOI] [PubMed] [Google Scholar]
- 5. Fernandes CM, Worster A, Eva K, Hill S, McCallum C. Pneumatic tube delivery system for blood samples reduces turnaround times without affecting sample quality. J Emerg Nurs 2006;32:139–143. [DOI] [PubMed] [Google Scholar]
- 6. Ellis G. An episode of increased hemolysis due to a defective pneumatic air tube delivery system. Clin Biochem 2009;42:1265–1269. [DOI] [PubMed] [Google Scholar]
- 7. Throckmorton AL, Kapadia JY, Chopski SG, Bhavsar SS, Moskowitz WB, Gullquist SD, Gangemi JJ, Haggerty CM, Yoganathan AP. Numerical, hydraulic, and hemolytic evaluation of an intravascular axial flow blood pump to mechanically support Fontan patients. Ann Biomed Eng 2011;39:324–336. [DOI] [PubMed] [Google Scholar]
- 8. Sodi R, Darn SM, Stott A. Pneumatic tube system induced haemolysis: Assessing sample type susceptibility to haemolysis. Ann Clin Biochem 2004;41:237–240. [DOI] [PubMed] [Google Scholar]
- 9. Kratz A, Salem RO, Van Cott EM. Effects of a pneumatic tube system on routine and novel hematology and coagulation parameters in healthy volunteers. Arch Pathol Lab Med 2007;131:293–296. [DOI] [PubMed] [Google Scholar]
- 10. Hübner U, Böckel‐Frohnhöfer N, Hummel B, Geisel J. The effect of a pneumatic tube transport system on platelet aggregation using optical aggregometry and the PFA‐100. Clin Lab 2010;56:59–64. [PubMed] [Google Scholar]
- 11. Bolliger D, Seeberger MD, Tanaka KA, Dell‐Kuster S, Gregor M, Zenklusen U, Grapow M, Tsakiris DA, Filipovic M. Pre‐analytical effects of pneumatic tube transport on impedance platelet aggregometry. Platelets 2009;20:458–465. [DOI] [PubMed] [Google Scholar]
- 12. Wenham PR, Hanson T, Ashby JP. Interference in spectrophotometric analysis of cerebrospinal fluid by haemolysis induced by transport through a pneumatic tube system. Ann Clin Biochem 2001;38:371–375. [DOI] [PubMed] [Google Scholar]
