Arterial Stiffness: Different Metrics, Different Meanings

B Spronck 1, J D Humphrey 2,1
PMCID: PMC6808013  PMID: 30985880

Abstract

Findings from basic science and clinical studies agree that arterial stiffness is fundamental to both the mechanobiology and the biomechanics that dictate vascular health and disease. There is, therefore, an appropriately growing literature on arterial stiffness. Perusal of the literature reveals, however, that many different methods and metrics are used to quantify arterial stiffness, and reported values often differ by orders of magnitude and have different meanings. Without clear definitions and an understanding of possible inter-relations therein, it is increasingly difficult to integrate results from the literature to glean true understanding. In this paper, we briefly review methods that are used to infer values of arterial stiffness that span studies on isolated cells, excised intact vessels, and clinical assessments. We highlight similarities and differences and identify a single theoretical approach that can be used across scales and applications and thus could help to unify future results. We conclude by emphasizing the need to move toward a synthesis of many disparate reports, for only in this way will we be able to move from our current fragmented understanding to a true appreciation of how vascular cells maintain, remodel, or repair the arteries that are fundamental to cardiovascular properties and function.

Keywords: aorta, stress, atomic force microscopy, pulse wave velocity, elasticity

Introduction

Mechanical factors have long been known to play vital roles in arterial health, disease, and treatment. Studies dating back to the late 19th century showed that altered blood flow and pressure typically result in changes in luminal radius and wall thickness [1,2], thus foreshadowing the need to quantify the mechanoregulation of arterial structure and function. It was not until the 1970s, however, that it was discovered how such changes occur. Experiments on cultured endothelial cells and vascular smooth muscle cells revealed flow- and stretch-induced changes in secreted proteins that resulted from altered gene expression [3,4]. Hence was born the field of vascular mechanobiology, reviews of the first three decades of which can be found elsewhere [5,6]. Importantly, careful studies of the aorta during this same defining period revealed that circumferential lamellar tension (∼2 N/m), and by inference circumferential lamellar stress (∼150 kPa), is similar within the medial layer across multiple mammalian species [7]. This finding, coupled with subsequent observations that aortic material stiffness at mean arterial pressure is also similar (∼500 kPa) across invertebrates and vertebrates [8], strongly suggests the existence of a “mechanical homeostasis”—that is, vascular cells seek to establish and then maintain particular mechanical quantities near target (homeostatic) values. Indeed, findings at subcellular, cellular, tissue, and organ levels suggest that such homeostatic processes exist across many spatial and temporal scales [9].

Two seminal contributions by Fung and his colleagues advanced our ability to quantify arterial wall stress and the associated stiffness, which are important both for assessing the mechanics of the wall and understanding its mechanobiology. First, Fung observed via uniaxial experiments on soft tissues that material stiffness (in this case, a change in the first Piola-Kirchhoff stress with respect to stretch) relates nearly linearly to the stress itself [10]. Importantly, this finding suggests directly that the first Piola-Kirchhoff stress increases nearly exponentially with stretch. Indeed, a similar observation had previously been reported for the overall structural stiffness of the pressurized eye based on its pressure–volume relation [11]. Motivated by these findings, Fung later hypothesized the existence of an exponential stored energy function W that yields an exponential relationship between the second Piola-Kirchhoff stress S and Green strain E, where S=W/E. Now known as Fung elasticity, this hyperelastic function can be written as W=c(exp(Q)1), where c is a material parameter and the scalar function Q depends quadratically on E [12]. An associated metric of material stiffness, in referential form, is thus C=S/E=2W/EE. Notwithstanding the importance of Fung's constitutive hypothesis in quantifying the mechanical behavior of many soft tissues and solving associated initial-boundary value problems, the observation that material stiffness relates linearly to stress implies that it should be challenging to delineate whether mechanobiological responses correlate better with stress or stiffness, consistent with the aforementioned observations that homeostatic values of aortic stress and material stiffness are similar across species [7,8]. Second, Fung showed that the existence of residual stress in arteries dramatically affects the calculated distribution of Cauchy stress t across the arterial wall [13], which in combination with basal smooth muscle cell tone results in a nearly homogenized transmural distribution [12]. This finding supports the concept that vascular cells seek to establish and then maintain mechanical stress or stiffness near a homeostatic target, regardless of location within the wall. Importantly, it now seems that different homeostatic targets may exist for different cell types that populate different layers of the wall [14], thus reinforcing the concept of a cell-specific mechanical homeostasis that can manifest at the vessel level as well.

The importance of arterial stiffening in human disease was anticipated as early as the turn of the 20th century by Sir Wm Osler, one of four founding fathers of Johns Hopkins medicine, but confirmation had to await a seminal clinical study at Paris, France that revealed that an increased pulse wave velocity (PWV) is an initiator and indicator of diverse cardiovascular, neurovascular, and renovascular disease and thus all-cause mortality [15]. Carotid-to-femoral PWV is now considered to be the gold standard for clinical assessment of arterial stiffness [16], which increases with diabetes, hypertension, natural aging, particular connective tissue disorders, and so forth [1719]. This metric of stiffness is best measured by dividing the vascular centerline distance between two recording sites by the time that it takes the (foot of the) pressure pulse wave to travel between these sites (“foot-to-foot” PWV). Intuitive understanding of the PWV often comes from the Moens–Korteweg equation, derived in the 1870s and commonly written as PWV=Eh/2ρa, where E is an intrinsic (linear) material stiffness, h is the wall thickness, a is the luminal radius, and ρ is the mass density of the fluid. Understood in this way, we see that PWV depends on both the intrinsic material stiffness and the geometry of the wall, hence rendering it an integrated (along the centerline distance) measure of structural, not material, stiffness.

Not surprisingly, there continues to be increasingly greater interest in measuring and understanding arterial stiffness in basic science studies and clinical assessments. Quantifying stiffness helps us to understand critical questions related to, among other issues, fundamental aspects of vascular mechanobiology, effects of a distensible wall on the hemodynamics, disease progression, and even the efficacy of particular clinical treatments. In this paper, we briefly review and contrast different methods for measuring and interpreting arterial stiffness both to emphasize that which has been learned and that which remains unclear. In particular, we emphasize that many different metrics of arterial stiffness reported in the literature have different theoretical underpinnings and thus different meanings. Caution is thus warranted when comparing values of stiffness across studies as well as when interpreting fundamental implications of a particular metric on either the cell biology or the systems physiology.

Approaches

We first review the theoretical basis for some of the experimental methods that are commonly used to infer material properties of the arterial wall and its constituent parts. The presentation is organized by physical scale, not chronological introduction within the field. Regardless of scale, it is critical to delineate concepts of material versus structural stiffness, and similarly, it is essential to note that values of stiffness depend on the type (tensile, compressive, or shear) and magnitude (small or large) of the loading as well as the mechanical state in which the experimental loads are imposed, especially if the sample is otherwise unloaded during testing. Unfortunately, many papers often simply report a value for stiffness without emphasizing the precise definition or experimental conditions, hence obscuring the range of applicability. Here, we attempt to delineate some of these issues.

Atomic Force Microscopy.

Invented in the early 1980s at IBM, atomic force microscopy (AFM) uses a laser to detect small deflections of a cantilever probe as it interacts with the surface of a specimen. The atomic force microscope can be used in various modes to assess the topography of a surface or to probe mechanical properties. With regard to the latter, one can perform precise indentation (compression) tests or functionalize the probe to enable tractions (tension) to be applied at the surface. As implied by its name, the most precise atomic force microscopes measure nanoscale forces and motions, thus it is not surprising that AFM has been used extensively in biology, biophysics, and bioengineering, often to assess aspects of cell stiffness, cell-matrix interactions, or local matrix stiffness. With larger probes, one can also assess bulk tissue stiffness. Associated data consist primarily of the applied force (inferred from knowledge of the cantilever properties and deflections) and the imposed motion of the probe, often the depth of penetration into the surface. Quantities such as stress and stiffness can be determined by solving the associated initial-boundary value problem.

Data on indentation force (f) and depth (δ) are often interpreted using a classical 19th century solution by Hertz for the localized indentation of a semi-infinite half-space that exhibits a linearly elastic and isotropic material behavior under small strains and rotations [20]. In this way, one can infer an associated Young's modulus E, a measure of the intrinsic compressive material stiffness for a specialized class of material behaviors. Albeit often not noted, this classical Hertz solution also assumes that the half-space is unloaded prior to the indentation by the probe. Of course, all arteries and their attendant cells are under significant loads in vivo and often are loaded in vitro in cell or organ culture, hence any Young's modulus inferred from the Hertz equation must be viewed cautiously. In particular, it has been shown both analytically [21] and numerically [22] that the indentation force-depth relationship, and thus inferred stiffness, depends strongly on the pre-existing state of stress/strain in the material. Because analytical findings typically provide more intuitive insight than numerical results, note that an exact solution relevant to AFM can be obtained using a theory of small deformations superimposed on large (Appendix A) for the special case of an initially isotropic planar specimen subjected to finite equibiaxial in-plane stretching prior to a superimposed small indentation in the out-of-plane direction [23]. In this case, the indentation force f and (small) depth of penetration δ are related linearly via f=αδ where the “transverse structural stiffness” α depends nonlinearly on the finite in-plane stretch experienced by the specimen as well as on the intrinsic material stiffness of the material and the geometry of the rigid probe. For example, for a neo-Hookean material behavior defined by a stored energy function W=μ(trC3), where μ is the shear modulus and C=FTF is the right Cauchy–Green tensor, with F being the deformation gradient tensor, it can be shown that the transverse stiffness, having units of N/m, can be written as [21]

α=8roμ(λ9+λ6+3λ31)λ4(λ3+1) (1)

where 2ro is the diameter (do) of a flat-ended cylindrical probe and λ is the equibiaxial in-plane stretch. A Hertz-type solution can be recovered as λ1, for which α16roμ=4doE/3, with Young's modulus E(=6μ) having units of N/m2. Hence, E can be computed easily given the diameter of the probe and the slope of the force-depth data (f,δ) for an otherwise unloaded specimen. Of course, because of inherent uncertainty in all experimental data, the stiffness parameter(s) should be estimated using least-squares regressions of data via an over-determined system of equations. Moreover, in the case of a nonlinear behavior, the intrinsic material stiffness (e.g., shear modulus μ for a neo-Hookean response) should be inferred from force-depth data over a range of in-plane stretches that are relevant to the biological or physiological regime of interest in order to characterize well the overall material behavior. Noting that adherent cells spread out and develop significant cytoskeletal stress when cultured on any substrate, which is to say that they develop a finite nearly in-plane state of stress or stretch, it has been shown that a small-on-large framework can be used to assess cell stiffness from AFM for more general exponential stored energy functions [24].

Indeed, although a neo-Hookean relation admits a simple analytical solution that allows one to intuit effects of finite in-plane stretches on AFM measured stiffness (Eq. (1)), the Fung exponential model is more appropriate for most soft tissues and even cells. Appendix B summarizes a general Fung model and Appendix C provides results for a small-on-large solution for AFM based on an isotropic Fung exponential model. As expected, the numerical implementation is straightforward. Nevertheless, results from AFM for assessing vascular cell stiffness [25,26] and arterial properties [2730] typically have been interpreted in terms of a single Young's modulus inferred from the Hertz solution, which tacitly ignores the important nonlinear dependence on in-plane prestretch and the nonlinear (often exponential) material behavior. Hence, the inferred values of stiffness are typically well below in vivo values and could be misleading regarding in vivo mechanobiology.

Although not discussed in detail here, additional methods are also used to assess cell stiffness, including magnetic twisting cytometry (MTC) and optical tweezers [31]. MTC is similar to AFM except that one either affixes onto or embeds within a living cell a ferromagnetic bead that can be cyclically twisted using a magnetic field. Associated data are often interpreted in terms of the so-called storage (G) and loss (G) shear moduli, which are basic descriptors of a one-dimensional (1D) linearly viscoelastic behavior over small strains and rotations [32]. For example, MTC was used to show that the stiffness of vascular smooth muscle cells (i.e., elastic storage and dissipative loss moduli) increases in aging [33]. Similar to the Hertz solution, however, this inference of material properties does not explicitly account for the underlying nonlinear material behaviors or the finite deformations that adherent cells experience in vitro when spreading on a surface or to which they are likely exposed in vivo. Again, a small-on-large approach for data analysis could be more appropriate.

Biaxial Biomechanical Testing.

Arteries are subjected in vivo to complex multiaxial loads due to blood pressure and flow as well as axial prestresses that develop largely due to somatic growth [12]. Flow-induced shear stresses are important mechanobiologically and can be determined by solving equations that govern the hemodynamics (e.g., Navier–Stokes solutions within the context of fluid–solid interactions), yet they are typically five orders of magnitude smaller (1.5 Pa) than the in-plane intramural stresses (∼150 kPa) and thus are neglected in most analyses of the wall stress field. The in-plane circumferential (tθθ) and axial (tzz) components of Cauchy stress are tensile, whereas the out-of-plane Cauchy radial (trr) stress is compressive, on the order of −15 kPa, and typically dictated largely by the traction boundary conditions. For these reasons, biaxial loading has long been preferred for studying the biomechanical properties and function of blood vessels, particularly arteries and veins [34]. Associated pressure-diameter and axial force-length data provide direct insight into the structural stiffness of these vessels. Material stiffness can be inferred from a global equilibrium solution that relates the loads that are measured during standard biaxial tests on excised cylindrical samples to components of the Cauchy stress as [12]

P=rirotθθtrrrdr,L=πriro(2tzztrrtθθ)rdr (2)

where r[ri,ro], with ri and ro being the inner and the outer radius of the sample, respectively, P is the distending pressure, and L is the reduced axial load. If only the mean (i.e., radially averaged, ) values of stress are of interest, then these two integral relations can be replaced with algebraic ones, tθθσθθ=Pri/h and tzzσzz=(L+πri2P)/πh(2ri+h), where h=rori is the wall thickness. Mean values of stress are surprisingly useful because of the aforementioned effect of residual stress in homogenizing the transmural distribution of stress. Regardless, some investigators infer stiffness from plots of stress versus stretch or strain (yielding so-called tangent moduli), but stiffness depends on the full multiaxial deformation and is best computed from an appropriate constitutive relation as noted above for the referential stiffness (S/E). Given the complex microstructure of the vascular wall, most investigators now prefer the so-called fiber-family constitutive models [3537] to characterize multiaxial data, though the Fung exponential provides good fits to data in many cases (recall that Appendix B reviews a general orthotropic Fung relation). Best-fit values of material parameters for any appropriate constitutive relation can be determined via nonlinear regression of pressure-diameter and axial force-length data, with data from multiple biaxial protocols typically combined to improve the parameter estimation, again via (nonlinear) least squares regression. In this regard, it is important to note that many studies nevertheless report only pressure-diameter data at a single value of axial stretch and often do not measure the associated axial force. Such data are essentially one-dimensional and not useful for calculating in vivo relevant biaxial stress or stiffness [38].

There are, in addition, a few other issues regarding the in vivo applicability of constitutive relations that are inferred in vitro. First, values of arterial stiffness change with pressure over a cardiac cycle, hence the theory of small deformations superimposed on large has been used to compute “single” values of the spatial material stiffness C that are often representative over a cardiac cycle and thereby render computations of fluid-solid-interactions more efficient [35]. In component form, we have (see Appendix A)

Cijkl=2δikFlAoFjBoWCAB|Co+2δjkFiAoFlBoWCAB|Co+4FiAoFjBoFkPoFlQo2WCABCPQ|Co (3)

where δij is the Kronecker delta and a superscript o denotes a deformation quantity that is associated with an original finite deformation about which the small superimposed deformation occurs. For example, the large deformation could be for an artery from a traction-free reference configuration to a finitely distended and extended in vivo configuration near mean arterial pressure about which relatively small motions occur over a cardiac cycle. Second, there is a pressing need to understand better how central arterial stiffness and resistance vessel function interact to affect both local and global hemodynamics and cardiovascular function [19,39]. Third, the effects of perivascular tethering can be as important as the structural stiffness (i.e., material stiffness and geometry) in affecting the hemodynamics though there has not been much quantification of this effect [40,41]. Fourth, constitutive relations that are inferred in vitro are generally established under quasi-static conditions, whereas in vivo loading is pulsatile and hence intrinsically dynamic. Associated differences in strain-rate may influence calculated values of measured stiffness [42,43]. Finally, although well suited for nearly cylindrical samples, standard distension–extension tests are not sufficient for more complex arterial geometries, particularly those manifesting in disease. Fortunately, new methods are emerging that enable local material properties to be inferred using full-field strain measurements and inverse methods for material characterization [44,45].

Although not discussed in detail here, there are also many reports of vascular function and properties based on either uniaxial loading tests (e.g., Ref. [46]) or ring myography (e.g., Ref. [25]). Originally conceived to study isometric contraction, ring tests are performed by placing a short ring-like sample of the vascular wall on two mounting fixtures that are separated by a finite distance. The sample thus stretches primarily in the circumferential direction as the fixtures are separated and the sample deforms from a circular to an oval to a more uniaxial geometry within the central region of measurement. Measurement of the force acting on (passive) or generated by (active) the sample thus allows one to estimate 1D stress-stretch information or to construct dose-response curves while holding a sample at a fixed separation distance (isometric). This approach is particularly valuable for high-throughput comparisons of different drugs or their doses, but the associated contraction of the smooth muscle is under nonphysiological loading [47]. Inference of passive or active stiffness from these ring tests is further compromised by the lack of an exact solution to the full boundary value problem (finite bending and uniaxial extension of an annulus) and the lack of biaxial loading [12,34]. In particular, both the radial and axial stretches reduce below one as circumferential stretch increases with the separation of the mounting fixtures. Axial stretch is typically much greater than unity in vivo and is important to both the mechanics and the mechanobiology [48]. Finally, in-plane biaxial tests can be performed on excised samples (e.g., Ref. [49]) and are generally very informative for they can largely mimic the in vivo state of stress, with the exception of radial compression due to a distending blood pressure. The primary caveat with in-plane biaxial tests is that large specimens can retain some of their natural cylindrical curvature in the unloaded state, thus requiring combined bending and extension to load the sample biaxially. Uniaxial tests performed on different samples with different gross orientations can avoid some issues of residual curvature, but associated protocols are limited and cannot explore directly the inherent coupling that is important in dictating the multiaxial mechanical behavior.

Pulse Wave Velocity and Distensibility.

Whereas AFM, MTC, biaxial testing, ring myography, and allied methods are employed in vitro on excised samples, which generally enables significant experimental control, clinical studies necessarily require less invasive in vivo methods, measurements, and metrics. As noted earlier, the current clinical gold standard metric of arterial stiffness is the carotid-to-femoral pulse wave velocity, which is a measured quantity that represents a structural response that is integrated over a particular vascular path. There are, in addition, more local metrics of structural stiffness that are used clinically. One such metric is the so-called distensibility

D=(dsysnddiasn)/ddiasn(PsysPdias) (4)

where d is the luminal diameter, with sys and dias denoting values at systole and diastole, respectively. The parameter n = 1 or 2, yielding values of D that differ numerically by approximately a factor of 2. Taking n=2 implies using cross-sectional area rather than diameter, which potentially avoids problems with low spatial imaging resolution. Note further that the Bramwell–Hill form for pulse wave velocity can be written as PWV=1/ρD for n=2, hence localizing the value of PWV. Another local measure of structural stiffness (the pressure–strain or Peterson modulus) can be computed as

EP=(PsysPdias)ddias/(dsysddias) (5)

Clinical findings confirm that local measures such as EP can correlate well with the more global PWV [50], hence suggesting that these metrics are complementary. We end this overview by noting that PWV, D, and EP all intrinsically depend on the operating point (i.e., blood pressure) about which they are measured/calculated [51,52]. In a research setting, such dependence can sometimes be controlled post hoc using regression methods, but this is less feasible in clinical practice. Regression may be especially problematic in studies of hypertension where (a change in) blood pressure may have both direct (through nonlinear mechanics) and indirect (causing arterial remodeling) effects on the measured metric of stiffness, a distinction that cannot be made with regression analyses [53]. This issue motivated considerable research for “pressure-corrected” metrics of stiffness. As should be clear from the preceding text, however, arterial mechanics is highly nonlinear and fully capturing it requires complex constitutive relations with many parameters, with value(s) of stiffness always dependent on the level of stress. Nevertheless, simple metrics have been proposed that, within individual limitations, work surprisingly well [54,55]. Examples include the cardio-ankle vascular indices (CAVI and CAVI0), which are metrics of global structural stiffness that essentially correct a measured foot-to-foot PWV for its pressure dependence via another metric of local structural stiffness (β and β0), namely

CAVI=lnPsysPdiasPWV2·2ρPsysPdias,CAVI0=PWV2·2ρPdiaslnPdiasPref (6)
β=lnPsysPdiasddiasdsysddias,andβ0=βlnPdiasPref (7)

with Pref being an arbitrary but constant reference value of pressure [56]. These so-called pressure-corrected metrics are motivated by the observation that blood pressure relates approximately exponentially with diameter [55]. CAVI0 and β0 represent modifications to the original metrics (CAVI and β) to correspond better with an actual exponential relationship; additionally, CAVI0 assumes PWV to scale with diastolic instead of mean blood pressure, which seems more in line with experimental observations on the pressure dependency of PWV [51,56].

Of course, PWV can also be estimated using computational methods given much more information: the spatially distributed geometry, material properties, and boundary conditions. Such calculations avoid implicit assumptions inherent in the Moens–Korteweg and related equations and they enable informative parametric studies, as, for example, how vascular taper or distal resistances affect the pulse wave. Unsteady 1D and three-dimensional models have been used to compute PWV [57,58], but again additional intuitive insight can be gleaned from analytical solutions. Toward this end, the theory of small deformations superimposed on large has also been used to obtain analytical results for a simple cylindrical geometry [59]. This solution shows explicitly that both the diastolic distension and the axial prestretch are important contributors to the computed values of PWV; in the limit as the strains become small, this relation recovers the Moens–Korteweg equation. As in the case of the AFM, however, this analysis shows clearly that complexities due to nonlinear material behaviors and finite deformations play important roles in determining the precise value of stiffness that is computed.

Comparisons

Fung and colleagues used nonlinear regression to determine best-fit values of the seven material parameters in an exponential stored energy function W (Eqs. (A1) and (A2) in Appendix B) from both uniaxial (radial compression [60]) and biaxial (pressure-diameter-axial force-stretch [13]) data for a rabbit thoracic artery. Not surprisingly, the parameter values differed for the predominantly compressive versus tensile behaviors. Figure 1 compares these results together; note the anisotropy and strong nonlinearity in the tensile (circumferential and axial) behaviors and the asymmetry between the tensile and compressive behaviors even at modest strains. Results are not shown for greater compressive strains since the original tests used modest levels of compression and the associated analytical solution reveals potential bifurcations in the equilibrium solutions at higher compressive strains. Importantly, given that the slope of the stress-stretch curves reflects (but does not define) the stretch-dependent spatial material stiffness, note the expected tremendous differences between the low values of material stiffness near the unloaded configuration (low stress) and high values near the in vivo configuration (physiologic stress). Figure 2 shows calculated circumferential behaviors for different degrees of fixed axial stretch from one to the in vivo value (1.691 for this particular rabbit artery). This result reveals the strong biaxial coupling, with circumferential stress and stiffness affected dramatically by the value of axial stretch. The aforementioned ring myography and uniaxial tests disregard such coupling and thus can underestimate the actual stiffness dramatically.

Figure 3 shows calculated results for a simulated AFM indentation test with the material modeled with the same Fung-exponential type of constitutive relation and the indentation performed at different levels of fixed equibiaxial in-plane stretch. Since the analytical small-on-large solution holds for isotropic material behaviors [21], we first estimated new values of the material parameters in the Fung-exponential that yield an isotropic response similar to the mean of the anisotropic response shown in Fig. 1. Again, it can be seen that there is a strongly coupled response between different directions of loading, here out-of-plane versus in-plane. Most published works on arterial wall and vascular cell stiffness disregard this coupling effect and use AFM to test samples, in the absence of a pre-existing in-plane stress, to compute Young's modulus that strictly holds only for small strains. Such a situation is not physiological and again is expected to underestimate the actual in vivo stiffness dramatically.

To emphasize further both the multiaxial nature of stress and stiffness, and couplings therein, we show three-dimensional plots in Fig. 4 for simple cases of combined circumferential and axial extensions of Fung elastic materials. That is, assuming principal homogeneous deformations and incompressibility, the stored energy function and thus associated wall stress and stiffness depend on the in-plane principal stretches alone. The primary in-plane components of the mean Cauchy stress can thus be thought of conceptually in terms of components of the left stretch tensor V=FFT, namely σθθ=σ^θθ(λθ,λz) and σzz=σ^zz(λθ,λz). As it can be seen in the figure, for both isotropic (left columns; cf. Fig. 3) and anisotropic (right columns; cf. Figs. 1 and 2) behaviors, circumferential and axial deformations contribute similarly to the overall elastic energy storage and thus stress and material stiffness, here in a spatial rather than referential description since spatial quantities (defined relative to the current configuration) are most relevant in vivo.

Discussion

It is now widely accepted that both cell- and matrix-level stiffness are fundamental to mechanobiological responses within the vasculature, including modulation of cell phenotype. Similarly, it is widely accepted that tissue-level stiffness is fundamental to the hemodynamics, particularly propagation of the arterial pressure wave that dictates many aspects of end organ health or disease. There is, therefore, an appropriately growing literature on arterial stiffness. Yet, the many different reported metrics have different meanings because of different types of loads that are imposed relative to different biomechanical states. We summarize in Table 1 some of these methods and metrics and, for purposes of illustration, we list in Table 2 some values of stiffness resulting from these different methods. As it can be seen, reported values of stiffness (some compressive, some shear, most tensile) differ by orders of magnitude, as should be expected for highly nonlinear material behaviors when assessed relative to different configurations, ranging from otherwise unloaded to in vivo relevant. Assuming that the associated calculations were performed correctly, each of these different values of stiffness should be viewed as reliable. The key question, which we do not attempt to answer here, is therefore: How can we extract from these disparate metrics, having different meanings, a unified understanding of vascular cell mechanobiology and biomechanics and their roles in dictating vascular health or disease progression? We submit that focused effort should be directed toward answering this question, for without such our ability to use basic science findings to inform clinical decisions will remain largely wanting since understanding will remain largely fragmented and incomplete.

Table 1.

Arterial stiffness—measurement modalities, metrics, and characteristics

Modality Application Configuration for stiffness measurement Vascular axes of stiffness measurement Mode of stiffness measurement Loading rate Typical output metric (s)
Atomic force microscopy In vitro Unloaded1 Radial or axial Compressive Quasi-static E (M)
Wire myography In vitro Uniaxially loaded Circumferential Tensile Quasi-static2a E (M), C (S)
Pressure myography In vitro Uniaxially or biaxially loaded Circumferential and axial Tensile Quasi-static2b E (M), W (M), C (S)
Biomechanical biaxial testing In vitro Biaxially loaded Circumferential and axial Tensile Quasi-static W (M), C (S)
Surface transit time PWV In vivo Loaded Tensile Dynamic PWV (S)
Magnetic Resonance Imaging In vivo Loaded Circumferential Tensile Dynamic C (S), D (S), PWV (S)3,4
Ultrasound echotracking In vivo Loaded Circumferential Tensile Dynamic E (M), C (S),D (S), PWV (S)3

C, compliance coefficient; D, distensibility; E, Young's modulus; W, strain energy; PWV, pulse wave velocity. M, material stiffness metric; S, structural stiffness metric.

1

Typically unloaded, but has been performed on tissue maintained in a biaxially loaded state [61];

2

typically quasi-static, but has been performed under dynamic loading (a[62], b[63]);

3

calculated using the Bramwell–Hill relationship;

4

measured by estimating transit time.

Table 2.

Values of aortic stiffness (mouse) depend strongly on methodology

Vessel/mouse Modality Definition/state Value, units, age Reference
Thoracic aorta, C57BL/6J AFM In Vitro, unloaded, cut open, radially indented from luminal side (endothelium intact)/compressive E = 3.1 kPa (2 months) [29]
E = 3.6 kPa (6 months)
(age 2–18 months)
E = 16.9 kPa (12 months)
E = 21.8 kPa (18 months)
Suprarenal abdominal aorta, C57/Sv129 AFM In Vitro, unloaded or pressurized to 100 mmHg and elongated to In Vivo axial stretch, ring, axially indented/compressive E = 18.7 kPa (unloaded) [61]
(age 10–13 months) E = 12.3–76.4 kPa (loaded; bimodal distribution)
Aorta, C57BL/6J AFM In Vitro, unloaded, cut open, radially indented from luminal side (endothelium removed)/compressive E = 24 kPa [28]
(age 11 months)
Ascending thoracic aorta, C57BL/6J AFM In Vitro, unloaded, cut open, radially indented from luminal side (endothelium intact)/compressive E = 2.8–12.7 kPa (0.5 months)1 [30]
(age 0.5–3.5 months) E = 5.0–38.8 kPa (2 months)1
E = 4.4-36.7 kPa (3.5 months)1
Ascending thoracic aorta, C57BL/6J Biaxial testing In Vitro, loaded, intact, pressurized to 128 mmHg, elongated to In Vivo axial stretch/tensile Cθθθθ = 2.76 MPa [64]
Czzzz = 2.26 MPa
(age 15.2±0.1 weeks)
Suprarenal abdominal aorta, C57BL/6J Biaxial testing In Vitro, loaded, intact, pressurized to 100 mmHg, elongated to In Vivo axial stretch/tensile Sθθ/∂Eθθ = 1.33 MPa [65]
Szz/∂Ezz = 0.082 MPa
(age 5–6 months)
Carotid-to-femoral arterial bed, C57BL/6J Applanation tonometry In Vivo, 4–5% sevoflurane or 75 mg/kg sodium pentobarbital anesthesia, noninvasive/PWV PWV = 3.96 m/s (sevoflurane)2 [66]
(age 5.6±0.2 months) PWV = 2.89 m/s (sodium pentobarbital)2
Abdominal aorta, C57BL/6 Ultrasound echotracking In Vivo, 125 mg/kg tribromoethanol anesthesia, noninvasive/PWV PWV = 2.70 m/s3 [67]
(age 3–4 months)
Aorta (regionally dependent), C57BL/6J Ultrasound echotracking/pressure catheter In Vivo, 1.5% PWV = 5.2 m/s4 [68]
Isoflurane anesthesia, noninvasive (ultrasound), invasive (catheter)/PWV PWV = 3.0 m/s5
(age 3 months)
PWV = 3.5 m/s6

E, Young's modulus.

1

Computationally separated, numbers denote intimal/medial moduli. AFM, atomic force microscopy; PWV, pulse wave velocity; Cθθθθ and Czzzz, linearized circumferential and axial spatial material stiffness obtained using the theory of small-on-large. ∂Sθθ/∂Eθθ and ∂Szz/∂Ezz, referential material stiffness defined as the derivative of second Piola-Kirchhoff stress with respect to Green strain;

2

carotid-to-femoral transit-time PWV;

3

PWV in the window of an ultrasound probe;

4

aortic arch-to-femoral bifurcation transit-time PWV (ultrasound);

5

abdominal transit-time PWV (blood pressure catheter, 2 cm path length);

6

distensibility-based local abdominal PWV obtained from Bramwell–Hill equation. In studies where interventions were performed, control groups are displayed here.

In some ways, we have emphasized the obvious—the value of stiffness depends on its definition (material versus structural), the configuration to which it refers (referential-unloaded, versus spatial-current), and the conditions under which it is evaluated (compression versus tension versus shear in an otherwise unloaded state or not). Yet, we are not aware of a prior consistent discussion of methods used in vascular mechanics across scales from atomic force microscopy to biaxial tests on cylindrical segments to in vivo measurements. As we have noted, a number of metrics reported in the literature are based on solutions from classical elasticity because of the associated simplicity, not the theoretical relevance. Strictly speaking, therefore, most of these results are applicable only for linearly elastic isotropic responses under small strains and rotations when measured about an unloaded state—conditions not applicable to mechanobiologically, physiologically, or clinically relevant situations. Although we noted two analytical examples wherein a material stiffness applicable to a nonlinear behavior can be compared directly in the limit to a small strain Young's modulus [21,59], it is generally problematic and uninformative to compare findings based on Hertz, Moens–Korteweg, or similar equations with those based on nonlinear constitutive relations or appropriate linearizations thereof, including the theory of small deformations superimposed on large. Yet, values of Young's modulus continue to be reported and, in some cases, results are similarly presented based on concepts from linearized viscoelasticity, including storage and loss moduli. Of course, one could argue that such results can be insightful when one consistently compares values across studies using the same methods and metrics whether they strictly hold theoretically or not in configurations relevant to in vivo conditions or not. For example, a compressive Young's modulus inferred from AFM has been reported by multiple groups to increase for isolated smooth muscle cells in aging and hypertension relative to that in normalcy [25,69], hence these data are trying to tell us something—the question is, What? That is, although results may be reproducible and reliable, the key question should focus on their possible relevance in vivo. Similar methods have been used to show that the nuclear protein lamin-A scales with tissue stiffness [70], with tissues ranging from compliant (adipose or liver) to stiff (ligaments and bones). Again, the results are reproducible, reliable, and provocative, though not evaluated at in vivo values of stiffness. How such results should be interpreted or compared to values that are relevant to the in vivo condition remains an open question.

We submit that in vivo relevant conditions and metrics should be used when possible. Toward this end, the theory of small deformations superimposed on large deformations [71] can serve as a theoretically appropriate method to compute linearized values of stiffness while accounting for underlying nonlinear material behaviors and in vivo relevant finite deformations, with applicability including interpretations of AFM data [21], defining biaxial material stiffness in intact excised vessels for simulations of hemodynamics [35], and computing PWV either analytically [59] or numerically [58]. With regard to the need to measure and report metrics that are relevant to the in vivo conditions, it would be prudent to remember the words of Y. C. Fung written ∼50 years ago. First, “The main difficulty [problem] lies in the customary use of infinitesimal theory of elasticity to the media which normally exhibit finite deformations” [10] and “the greatest need lies in the direction of collecting data in multiaxial loading conditions and formulating a [constitutive] theory for the general rheological behavior of living tissues…” [72].

In conclusion, it is critical to quantify arterial stiffness—material and structural—because deviations from normal values associate with both the phenotypic modulation of vascular cells and the clinical severity of disease or disease risk. That said, there is also a fundamental conceptual issue that must be considered carefully as we seek to advance our understanding of the underlying mechanobiology. Although vascular cells clearly attempt to establish and then maintain certain mechanical quantities near homeostatic values [9,73], the continuum quantities of stress and strain, and metrics such as material stiffness that are derived from them, are actually mathematical concepts, not physical realities [74]. Hence, even though it appears that mean wall stress and stiffness are normally regulated near homeostatic targets across mammalian species [7,8,75], we should not expect a cell to necessarily respond to a stress (i.e., a linear transformation, or tensor, that transforms an outward unit vector into a traction vector at a point). Rather, it is more likely that forces acting at the molecular level change the conformation of important biomolecules and thereby stimulate cell signaling and downstream gene products. There is, therefore, a pressing need to understand better the micromechanics of mechanosensing by cells and the associated mechanoregulation of matrix [76] and to associate such phenomena with convenient continuum metrics such as stress and stiffness. Nevertheless, until, and possibly after, such multiscale understanding is achieved, direct correlations of mechanobiological and (patho)physiological responses with wall stress and stiffness should continue to be identified. Toward this end, an increased use of concepts of nondimensionalization and allometric scaling [77] should also become a priority. There is, therefore, a need for continued development of new concepts and techniques in vascular mechanics and mechanobiology and we conclude with words of Fung in his foreword to the inaugural issue of the journal Biomechanics and Modeling in Mechanobiology in 2002 [78]—“let us enjoy the work.”

Acknowledgment

JDH also gratefully acknowledges the Y.C. Fung Young Investigator Award from ASME that he received in 1990.

Appendix A. Small on Large

The polymath A. Cauchy knew equations of nonlinear elasticity in the 1820s, but analytical solutions to this class of problems had to await the semi-inverse approach of R. Rivlin in the late 1940s. Because of the inherent complexities, analytical solutions to problems in nonlinear elasticity were yet possible for only a relatively small class of motions [79] and many employed the advances in finite element methods that soon arrived [80]. It is often difficult to develop intuition from numerical solutions of highly nonlinear problems, however, thus the concept of “small deformations superimposed on large deformations” became useful in extending the range of possible analytical solutions. As noted herein, small-on-large solutions have been found useful in interpreting experimental results associated with atomic force microscopy [21], quantifying mechanical properties from biaxial tests on excised arteries [35], and relating the structural stiffness that can be inferred in vivo from measurements of pulse wave velocity to the material stiffness inferred from in vitro tests [59]. It seems appropriate, therefore, to briefly outline steps of this approach; the interested reader is referred elsewhere for more details [71].

Briefly, let the location of a material particle in an original configuration be denoted by X and in a finitely deformed configuration by x. In addition, let the location of this particle in a configuration that is close to the finitely deformed one be denoted by y. The total deformation gradient is thus F=(y/x)(x/X), which can be written as F=FsFo, where subscript o denotes the original finite deformation and subscript s denotes the superimposed small deformation. It proves convenient to write y=x+u, where u is a small displacement vector. Hence, Fs=x/x+u/x which we write as Fs=I+h, where h is a displacement gradient with h=tr(hhT)1. The right Cauchy–Green tensor can then be written as C=FTF=[I+hFo]TI+hFo=FoTFo+FoThTFo+FoThFo+FoThThFoCo+FoThT+hFo=Co+FoT2εFo if we neglect terms that are higher order in the (small) displacement gradient and we recognize the small strain tensor ε=(h+hT)/2. Hence, the right Cauchy–Green tensor in the final configuration simply equals that in an intermediate (finitely deformed) configuration, denoted by subscript o, plus a part that is linear in the small displacement gradient.

Just as the total deformation gradient F=I+hFo=Fo+hFo consists of a contribution that is due to a finite deformation to an intermediate configuration plus an addition, one can also assume that the total second Piola-Kirchhoff stress can be computed as S=So+S*, where the first contribution is associated with the original finite deformation and the second with a superimposed small deformation, namely S*(S/C)o:C* with S=W/Co and C*=CCo=FoT2εFo from above. Importantly, it can be shown that we can write the total Cauchy stress t, for an incompressible response and to order h, as [35] t=to+ht^o+t^ohTp*I+FoS^*FoT, where t^o is the extra (i.e., deformation-dependent) part of the Cauchy stress in the finitely deformed intermediate configuration and p* is a Lagrange multiplier that enforces isochoric motions during the small superimposed deformation. Finally, using the result above for the second Piola-Kirchhoff stress in terms of the stored energy function W, and recognizing that a small displacement gradient can be written in terms of the infinitesimal strain and rotation (i.e., h=h+hT/2+(hhT)/2, or, h=ε+Ω), we arrive at our final result: t+p*I=to+Cε+DΩ, where the spatial material stiffness (fourth order) tensor C can be written for an artery as given in component form in the third equation in the main text. Noting that the infinitesimal rotation tensor Ω vanishes with assumed principal deformations, this derivation reveals clearly that the spatial material stiffness associated with a superimposed small deformation depends directly and strongly on the prestress (or prestretch) and the nonlinear properties represented by W, which are amplified nonlinearly by the initial finite deformation when evaluated in the current (spatial) configuration [35].

Appendix B. Fung Elasticity

The illustrative solutions in Figs. 14 were obtained using the same Fung exponential strain energy function to facilitate comparisons across methods. This function is defined by [12]

W=12ceQ1 (A1)

where, in terms of principal Green strains,

Q=b1EΘΘ2+b2EZZ2+b3ERR2+2b4EΘΘEZZ+2b5EZZERR+2b6ERREΘΘ (A2)

with c and bj (j=1,2,,6) being material parameters that need to be determined from nonlinear regressions of data. For principal deformations, as of interest herein, the Green strains Eii=(λi21)/2 for (no sum on) i=Θ,Z,R.

Uniaxial Compression.

Uniaxial radial compression testing comparable to that reported by Chuong and Fung [60] was simulated by prescribing radial stretches λr from 1 to 0.85 (Fig. 1, negative abscissa). Incompressibility was assumed, and circumferential and axial directions were assumed to be traction-free consistent with the reported experiment, though it would have been better, albeit difficult, to have induced a finite in-plane deformation prior to the radial compression. Nonetheless, given the assumption of a homogeneous deformation, the in-plane Cauchy stresses in the actual experiment were tθθ=tzz=0. Constitutively, Cauchy stress was calculated from W as

tii=p+λi2WEii,withnosumoni=θ,z,rorΘ,Z,R (A3)

with p being a Lagrange multiplier enforcing incompressibility. From the traction-free conditions and homogeneous state of deformation, the Lagrange multiplier p for compression tests (pcomp) can be determined from Eq. (A3) using either

pcomp=λθ2WEΘΘorpcomp=λz2WEZZ (A4)

after which either λθ was determined as λθ=1/λzλr with λz determined iteratively to satisfy tzz=0 or vice versa for the axial stretch. Numerical simulations revealed equivalent outcomes using these two constraints.

Simulations of the compression experiment were then performed using reported values of the constitutive parameters for a rabbit thoracic artery (sample 11, incompressible case) given in Ref. [60]: c=43.12kPa, b1=0.8230, b2=0.9125, b3=1×107, b4=1.1237, b5=0.4125, and b6=0.3768.

Biaxial Stretching.

Equibiaxial in-plane stretch testing similar to that performed by Vande Geest et al. [49] was simulated by prescribing λθ=λz from 1.0 to 1.8 (Fig. 1, positive abscissa), with λr=1/λθλz. Cauchy stress was then obtained using Eq. (A3), with p (ptension) determined from trr=0, because of the traction-free condition on the top and bottom surfaces and the assumed homogeneous deformation, as

ptension=λr2WERR (A5)

Simulations of tensile loading were performed using values of the constitutive parameters, again for a rabbit thoracic artery [13]: c=22.40kPa,b1=1.0672, b2=0.4775, b3=0.0499, b4=0.0903, b5=0.0585, and b6=0.0042. Note the differences between the values for compression and tension despite the same functional form of W and the same sample tested by the same investigators.

Additional in-plane biaxial tension experiments (tension in θ and z directions), with λθλz but trr=0, were simulated with the same tensile constitutive parameters (c, b1b6) that were used in the prior section. Such loading simulates possible in-plane biaxial testing [49] but also mimics mean values of stress in possible distension–extension tests [38]. Indeed, corresponding in vivo loading states were determined by simulating pressure-diameter testing as follows. Unloaded inner and outer radii were Ri=li/2π and Ro=lo/2π, with li=8.75mm and lo=12.5mm the inner and outer unloaded circumferences given in Ref. [13], with unloaded thickness H=RoRi. Loaded inner radius ri was varied iteratively and loaded outer radius ro was computed as

ro=ri2+Ro2Ri2λz (A6)

with loaded thickness h=rori. The mean circumferential stretch λθ is then

λθ=ri+h2Ri+H2 (A7)

Prescribing axial stretch λz allows the radial stretch λr to be determined from incompressibility (λr=1/λθλz). Components of Cauchy stress again follow from Eq. (A3) and p from Eq. (A5). Luminal distending pressure P is now obtained from the mean value of Cauchy stress and geometry as

P=h·σθθri (A8)

Increasing values of ri thus yield desired increases in pressure (e.g., up to 120 mmHg).

Appendix C. Simulations of Atomic Force Microscopy for Fung Elasticity

The effect of in-plane prestretch on out-of-plane indentation testing, as in AFM, was simulated using the solution in Ref. [21]. For these derivations to hold, the in-plane stretch must be equibiaxial and the material must be isotropic. For consistency with our other simulations (Appendix B), we again used a Fung exponential strain energy function, though one that is parameterized isotropically, namely

b1=b2=b3andb4=b5=b6 (A9)

To obtain parameter values that model tension similar to the mean anisotropic response shown in Fig. 1, we simulated equibiaxial testing of an isotropic Fung elastic material (described by unique parameters c, b1, and b4), and fitted these three parameters to the mean of the anisotropic case. Specifically, we calculated the homogeneous (mean) equibiaxial Cauchy stresses σθθiso=σzziso, and minimized

e(c,b1,b4)=k=1N[σθθanisoσθθiso(c,b1,b4)k2+σzzanisoσzziso(c,b1,b4)k2] (A10)

with σθθanisoσzzaniso being the anisotropic Cauchy stresses shown in Fig. 1 (positive abscissa), with k a data point index and N the total number of data points (i.e., equilibrium configurations assessed during testing). This procedure yielded c=26.97kPa, b1=b2=b3=0.6685, and b4=b5=b6=0.2823, here for a “virtual” rabbit thoracic artery.

To use the formulation in Ref. [21], we reformulated the isotropic form of W in terms of invariants of the Green strain tensor E, namely IE and IIE, and then wroteW in terms of invariants of the right Cauchy–Green tensor C=2E+I. Toward this end, we used the full formulation for Q [12] where

Q=b1EΘΘ2+b2EZZ2+b3ERR2+2b4EΘΘEZZ+2b5EZZERR+2b6ERREΘΘ+b7EΘZ2+EZΘ2+b8EZR2+ERZ2+b9ERΘ2+EΘR2 (A11)

which for isotropy (Eq. (A9) and b7=b8=b9), with b7=b1b4, can be written as

Q=aEΘΘ+EZZ+ERR2+b(EΘΘEZZ+EZZERR+ERREΘΘERΘEΘREΘZEZΘEZRERZ) (A12)

or more compactly

Q=aIE2+bIIE (A13)

with

IE=EΘΘ+EZZ+ERRIIE=EΘΘEZZ+EZZERR+ERREΘΘERΘEΘREΘZEZΘEZRERZ (A14)

because E is a symmetric tensor. Note, too, that the two (exponential) material parameters for isotropy are a=b1 and b=2b4b1=2b7 (with b1>b4). Next, recognize that the right Cauchy–Green tensor C, with invariants IC and IIC

IC=2EΘΘ+EZZ+ERR+3=2IE+3 (A15)

and

IIC=4EΘΘEZZ+EZZERR+ERREΘΘ4EΘZ2+EZR2+EΘR2+4EΘΘ+EZZ+ERR+3=4IE+4IIE+3 (A16)

Re-arranging Eqs. (A15) and (A16) and inserting into Eq. (A13) yields

Q=14aIC232a+12bIC+14bIIC+94a+34b (A17)

Given W(Q) written in this form, the equations in Ref. [21], written as functions of W/IC and W/IIC, can be used directly to simulate AFM indentations for an isotropic Fung elastic material, similar to the calculation discussed in the text for the simpler neo-Hookean material.

Fig. 1.

Fig. 1

Intrinsic differences in tensile and compressive Fung-elastic behaviors. Tensile stress-stretch responses are shown for a simulated equibiaxial stretching experiment (λθ=λz>1) in circumferential (σθθ versus λθ) and axial (σzz versus λz) directions for mean values of stress (i.e., averaged through the thickness of the sample); note the nonlinear anisotropic behavior. A stress-stretch response is also shown for a simulated uniaxial compression experiment in the radial direction (λr<1): mean radial Cauchy stress as a function of radial stretch (σrr versus λr). The assumed constitutive behaviors were reported by Chuong and Fung [13,60].

Fig. 2.

Fig. 2

Circumferential behavior depends strongly on axial stretch. Stored energy W (panel a), mean circumferential Cauchy stress σθθ (panel b), and linearized circumferential material stiffness Cθθθθ (panel c) are shown as a function of circumferential stretch (abscissa) as well as for different values of (constant) axial stretch from 1.0 to 1.7 (8 lines). Black dots correspond to values at a distending pressure of 120 mmHg. Note that mean Cauchy stresses (i.e., integrated through the thickness) are on the order of 150 kPa under physiologic loads, whereas stiffness values are on the order of 1–2 MPa, particularly in a physiologic range of loads.

Fig. 3.

Fig. 3

In-plane prestretch has an important effect on out-of-plane indentation stiffness. The schematic drawing shows a potential arterial wall sample isolated for AFM testing. This planar sample could similarly represent an adherent cell after spreading on a surface. Panel a shows the effect of equibiaxial stretching (λθ=λzλ>1) on mean (integrate through the thickness) in-plane Cauchy stress for an assumed isotropic behavior. Panel b shows the relationship between indentation force (f) and indentation depth (δ) for values of in-plane stretch λ increasing from 1.0 to 1.8 in steps of 0.1. The slope of these lines, effectively the transverse structural stiffness α, is plotted as a function of in-plane equibiaxial stretch λ in panel c. Closed dots (panels a and c) indicate approximate values of the in vivo in-plane stretch (λθ=λz=1.70); open dots indicate the absence of in-plane stretch (λθ=λz=1.00), consistent with many reports in the literature. The thick line in panel b similarly corresponds to λθ=λz=1.70. Note the marked effect of in-plane stretching on the out-of-plane (transverse) stiffness as measured, for example, in atomic force microscopy.

Fig. 4.

Fig. 4

Linearized spatial material stiffness depends strongly and biaxially on operating point. Panels, from top to bottom, show stored strain energy (W), mean circumferential stress (σθθ), mean axial stress (σzz), circumferential material stiffness (Cθθθθ), and axial material stiffness (Czzzz), all as a function of axial stretch (λz) and circumferential stretch (λθ). The left two columns correspond to an isotropic Fung relation (Appendix C); the right two columns to an anisotropic Fung relation (Appendix B). Closed dots indicate a possible in vivo operating point at systole (λz=1.69; λθ=1.69/1.63 (isotropic/anisotropic), corresponding to a blood pressure of ∼120 mmHg). Open dots indicate an unloaded state (λz=λθ=1.00). Note the large, two orders of magnitude, difference in Cθθθθ between unloaded (Cθθθθ=0.018/0.024MPa (isotropic/anisotropic)) and loaded (Cθθθθ=1.9/2.0MPa (isotropic/anisotropic)) states. Black lines on three-dimensional surface plots are projections of the (gray) λθ and λz gridlines.

Contributor Information

B. Spronck, Department of Biomedical Engineering, , Yale University, , New Haven, CT 06520

J. D. Humphrey, Fellow ASME , Department of Biomedical Engineering, , Yale University, , New Haven, CT 06520; , Vascular Biology and Therapeutics Program, , Yale School of Medicine, , New Haven, CT 06520 , e-mail: jay.humphrey@yale.edu.

Funding Data

  • US NIH (Funder ID: 10.13039/100000002; Grant Nos. R01 HL105297, P01 HL134605).

  • Netherlands Organization for Scientific Research (NWO) (Funder ID: 10.13039/501100003246; Rubicon Grant 452172006).

References

  • [1]. Roy, C. S. , 1881, “ The Elastic Properties of the Arterial Wall,” J. Physiol., 3(2), pp. 125–159. 10.1113/jphysiol.1881.sp000088 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [2]. Thoma, R. , 1893, Untersuchungen Über Die Histogenese Und Histomechanik Des Gefäßsystems, Enke, Stuttgart, Germany. [Google Scholar]
  • [3]. Rosen, L. A. , Hollis, T. M. , and Sharma, M. G. , 1974, “ Alterations in Bovine Endothelial Histidine Decarboxylase Activity Following Exposure to Shearing Stresses,” Exp. Mol. Pathol., 20(3), pp. 329–343. 10.1016/0014-4800(74)90064-1 [DOI] [PubMed] [Google Scholar]
  • [4]. Leung, D. Y. , Glagov, S. , and Mathews, M. B. , 2003, “ Cyclic Stretching Stimulates Synthesis of Matrix Components by Arterial Smooth Muscle Cells In Vitro,” Science, 191(4226), pp. 475–477. 10.1126/science.128820 [DOI] [PubMed] [Google Scholar]
  • [5]. Haga, J. H. , Li, Y. S. , and Chien, S. , 2007, “ Molecular Basis of the Effects of Mechanical Stretch on Vascular Smooth Muscle Cells,” J. Biomech., 40(5), pp. 947–960. 10.1016/j.jbiomech.2006.04.011 [DOI] [PubMed] [Google Scholar]
  • [6]. Davies, P. F. , 2009, “ Hemodynamic Shear Stress and the Endothelium in Cardiovascular Pathophysiology,” Nat. Clin. Pract. Cardiovasc. Med., 6(1), pp. 16–26. 10.1038/ncpcardio1397 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [7]. Wolinsky, H. , and Glagov, S. , 1967, “ A Lamellar Unit of Aortic Medial Structure and Function in Mammals,” Circ. Res., 20(1), pp. 99–111. 10.1161/01.RES.20.1.99 [DOI] [PubMed] [Google Scholar]
  • [8]. Shadwick, R. E. , 1999, “ Mechanical Design in Arteries,” J. Exp. Biol., 202(Pt. 23), pp. 3305–3313.http://jeb.biologists.org/content/202/23/3305 [DOI] [PubMed] [Google Scholar]
  • [9]. Humphrey, J. D. , 2008, “ Vascular Adaptation and Mechanical Homeostasis at Tissue, Cellular, and Sub-Cellular Levels,” Cell Biochem. Biophys., 50(2), pp. 53–78. 10.1007/s12013-007-9002-3 [DOI] [PubMed] [Google Scholar]
  • [10]. Fung, Y. C. , 1967, “ Elasticity of Soft Tissues in Simple Elongation,” Am. J. Physiol., 213(6), pp. 1532–1544. 10.1152/ajplegacy.1967.213.6.1532 [DOI] [PubMed] [Google Scholar]
  • [11]. McEwen, W. K. , and St Helen, R. , 1965, “ Rheology of the Human Sclera. Unifying Formulation of Ocular Rigidity,” Ophthalmologica, 150(5), pp. 321–346. 10.1159/000304862 [DOI] [PubMed] [Google Scholar]
  • [12]. Humphrey, J. D. , 2002, Cardiovascular Solid Mechanics: Cells, Tissues, and Organs, Springer, NY. [Google Scholar]
  • [13]. Chuong, C. J. , and Fung, Y. C. , 1986, “ On Residual Stresses in Arteries,” ASME J. Biomech. Eng., 108(2), pp. 189–192. 10.1115/1.3138600 [DOI] [PubMed] [Google Scholar]
  • [14]. Bellini, C. , Ferruzzi, J. , Roccabianca, S. , Di Martino, E. S. , and Humphrey, J. D. , 2014, “ A Microstructurally Motivated Model of Arterial Wall Mechanics With Mechanobiological Implications,” Ann. Biomed. Eng., 42(3), pp. 488–502. 10.1007/s10439-013-0928-x [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [15]. Laurent, S. , Boutouyrie, P. , Asmar, R. , Gautier, I. , Laloux, B. , Guize, L. , Ducimetiere, P. , and Benetos, A. , 2001, “ Aortic Stiffness Is an Independent Predictor of All-Cause and Cardiovascular Mortality in Hypertensive Patients,” Hypertension, 37(5), pp. 1236–1241. 10.1161/01.HYP.37.5.1236 [DOI] [PubMed] [Google Scholar]
  • [16]. Laurent, S. , Cockcroft, J. , Van Bortel, L. , Boutouyrie, P. , Giannattasio, C. , Hayoz, D. , Pannier, B. , Vlachopoulos, C. , Wilkinson, I. , Struijker-Boudier, H. , and European Network for Non-Invasive Investigation of Large, A., 2006, “ Expert Consensus Document on Arterial Stiffness: Methodological Issues and Clinical Applications,” Eur. Heart J., 27(21), pp. 2588–2605. 10.1093/eurheartj/ehl254 [DOI] [PubMed] [Google Scholar]
  • [17]. McEniery, C. M. , Yasmin, Hall, I. R. , Qasem, A. , Wilkinson, I. B. , Cockcroft, J. R. , and Investigators ACCT, 2005, “ Normal Vascular Aging: Differential Effects on Wave Reflection and Aortic Pulse Wave Velocity: The Anglo-Cardiff Collaborative Trial (ACCT),” J. Am. Coll. Cardiol., 46(9), pp. 1753–1760. 10.1016/j.jacc.2005.07.037 [DOI] [PubMed] [Google Scholar]
  • [18]. Van Bortel, L. M. , Laurent, S. , Boutouyrie, P. , Chowienczyk, P. , Cruickshank, J. K. , De Backer, T. , Filipovsky, J. , Huybrechts, S. , Mattace-Raso, F. U. , Protogerou, A. D. , Schillaci, G. , Segers, P. , Vermeersch, S. , Weber, T. , and Artery, S. , European Society of Hypertension Working Group on Vascular Function, and European Network for Noninvasive Investigation of Large Arteries, 2012, “ Expert Consensus Document on the Measurement of Aortic Stiffness in Daily Practice Using Carotid-Femoral Pulse Wave Velocity,” J. Hypertens., 30(3), pp. 445–448. 10.1097/HJH.0b013e32834fa8b0 [DOI] [PubMed] [Google Scholar]
  • [19]. Laurent, S. , and Boutouyrie, P. , 2015, “ The Structural Factor of Hypertension: Large and Small Artery Alterations,” Circ. Res., 116(6), pp. 1007–1021. 10.1161/CIRCRESAHA.116.303596 [DOI] [PubMed] [Google Scholar]
  • [20]. Vinckier, A. , and Semenza, G. , 1998, “ Measuring Elasticity of Biological Materials by Atomic Force Microscopy,” FEBS Lett., 430(1–2), pp. 12–16. 10.1016/S0014-5793(98)00592-4 [DOI] [PubMed] [Google Scholar]
  • [21]. Humphrey, J. , Halperin, H. R. , and Yin, F. C. , 1991, “ Small Indentation Superimposed on a Finite Equibiaxial Stretch. Implications for Cardiac Mechanics,” ASME J. Appl. Mech., 58(4), pp. 1108–1111. 10.1115/1.2897696 [DOI] [Google Scholar]
  • [22]. Costa, K. D. , and Yin, F. C. , 1999, “ Analysis of Indentation: Implications for Measuring Mechanical Properties With Atomic Force Microscopy,” ASME J. Biomech. Eng., 121(5), pp. 462–471. 10.1115/1.2835074 [DOI] [PubMed] [Google Scholar]
  • [23]. Beatty, M. , and Usmani, S. , 1975, “ On the Indentation of a Highly Elastic Half-Space,” Q. J. Mech. Appl. Math., 28(1), pp. 47–62. 10.1093/qjmam/28.1.47 [DOI] [Google Scholar]
  • [24]. Na, S. , Sun, Z. , Meininger, G. A. , and Humphrey, J. D. , 2004, “ On Atomic Force Microscopy and the Constitutive Behavior of Living Cells,” Biomech. Model. Mechanobiol., 3(2), pp. 75–84. 10.1007/s10237-004-0051-x [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [25]. Sehgel, N. L. , Zhu, Y. , Sun, Z. , Trzeciakowski, J. P. , Hong, Z. , Hunter, W. C. , Vatner, D. E. , Meininger, G. A. , and Vatner, S. F. , 2013, “ Increased Vascular Smooth Muscle Cell Stiffness: A Novel Mechanism for Aortic Stiffness in Hypertension,” Am. J. Physiol. Heart Circ. Physiol., 305(9), pp. H1281–H1287. 10.1152/ajpheart.00232.2013 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [26]. Desai, A. , Geraghty, S. , and Dean, D. , 2019, “ Effects of Blocking Integrin Beta1 and N-Cadherin Cellular Interactions on Mechanical Properties of Vascular Smooth Muscle Cells,” J. Biomech., 82, pp. 337–345. 10.1016/j.jbiomech.2018.11.004 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [27]. Lindeman, J. H. , Ashcroft, B. A. , Beenakker, J. W. , van Es, M. , Koekkoek, N. B. , Prins, F. A. , Tielemans, J. F. , Abdul-Hussien, H. , Bank, R. A. , and Oosterkamp, T. H. , 2010, “ Distinct Defects in Collagen Microarchitecture Underlie Vessel-Wall Failure in Advanced Abdominal Aneurysms and Aneurysms in Marfan Syndrome,” Proc. Natl. Acad. Sci. U. S. A., 107(2), pp. 862–865. 10.1073/pnas.0910312107 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [28]. Weisbrod, R. M. , Shiang, T. , Al Sayah, L. , Fry, J. L. , Bajpai, S. , Reinhart-King, C. A. , Lob, H. E. , Santhanam, L. , Mitchell, G. , Cohen, R. A. , and Seta, F. , 2013, “ Arterial Stiffening Precedes Systolic Hypertension in Diet-Induced Obesity,” Hypertension, 62(6), pp. 1105–1110. 10.1161/HYPERTENSIONAHA.113.01744 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [29]. Liu, S. L. , Bae, Y. H. , Yu, C. , Monslow, J. , Hawthorne, E. A. , Castagnino, P. , Branchetti, E. , Ferrari, G. , Damrauer, S. M. , Pure, E. , and Assoian, R. K. , 2015, “ Matrix Metalloproteinase-12 Is an Essential Mediator of Acute and Chronic Arterial Stiffening,” Sci. Rep., 5, p. 17189. 10.1038/srep17189 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [30]. Lee, J. J. , Galatioto, J. , Rao, S. , Ramirez, F. , and Costa, K. D. , 2016, “ Losartan Attenuates Degradation of Aorta and Lung Tissue Micromechanics in a Mouse Model of Severe Marfan Syndrome,” Ann. Biomed. Eng., 44(10), pp. 2994–3006. 10.1007/s10439-016-1616-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [31]. Laurent, V. M. , Hénon, S. , Planus, E. , Fodil, R. , Balland, M. , Isabey, D. , and Gallet, FOIS. , 2002, “ Assessment of Mechanical Properties of Adherent Living Cells by Bead Micromanipulation: Comparison of Magnetic Twisting Cytometry Vs Optical Tweezers,” ASME J. Biomech. Eng., 124(4), pp. 408–421. 10.1115/1.1485285 [DOI] [PubMed] [Google Scholar]
  • [32]. Humphrey, J. D. , and O'Rourke, S. L. , 2015, An Introduction to Biomechanics: Solids and Fluids, Analysis and Design, Springer, New York. [Google Scholar]
  • [33]. Zhu, W. , Kim, B. C. , Wang, M. , Huang, J. , Isak, A. , Bexiga, N. M. , Monticone, R. , Ha, T. , Lakatta, E. G. , and An, S. S. , 2018, “ TGFβ1 Reinforces Arterial Aging in the Vascular Smooth Muscle Cell Through a Long-Range Regulation of the Cytoskeletal Stiffness,” Sci. Rep., 8(1), p. 2668. 10.1038/s41598-018-20763-w [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [34]. Cox, R. H. , 1983, “ Comparison of Arterial-Wall Mechanics Using Ring and Cylindrical Segments,” Am. J. Physiol., 244(2), pp. H298–H303. 10.1152/ajpheart.1983.244.2.H298 [DOI] [PubMed] [Google Scholar]
  • [35]. Baek, S. , Gleason, R. , Rajagopal, K. , and Humphrey, J. , 2007, “ Theory of Small on Large: Potential Utility in Computations of Fluid–Solid Interactions in Arteries,” Comput. Methods Appl. Mech. Eng., 196(31–32), pp. 3070–3078. 10.1016/j.cma.2006.06.018 [DOI] [Google Scholar]
  • [36]. Holzapfel, G. A. , and Ogden, R. W. , 2010, “ Constitutive Modelling of Arteries,” Proc. R. Soc. A, 466(2118), pp. 1551–1597. 10.1098/rspa.2010.0058 [DOI] [Google Scholar]
  • [37]. Schroeder, F. , Polzer, S. , Slazansky, M. , Man, V. , and Skacel, P. , 2018, “ Predictive Capabilities of Various Constitutive Models for Arterial Tissue,” J. Mech. Behav. Biomed. Mater., 78, pp. 369–380. 10.1016/j.jmbbm.2017.11.035 [DOI] [PubMed] [Google Scholar]
  • [38]. Ferruzzi, J. , Bersi, M. R. , and Humphrey, J. D. , 2013, “ Biomechanical Phenotyping of Central Arteries in Health and Disease: Advantages of and Methods for Murine Models,” Ann. Biomed. Eng., 41(7), pp. 1311–1330. 10.1007/s10439-013-0799-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [39]. Humphrey, J. D. , Harrison, D. G. , Figueroa, C. A. , Lacolley, P. , and Laurent, S. , 2016, “ Central Artery Stiffness in Hypertension and Aging: A Problem With Cause and Consequence,” Circ. Res., 118(3), pp. 379–381. 10.1161/CIRCRESAHA.115.307722 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [40]. Humphrey, J. D. , and Na, S. , 2002, “ Elastodynamics and Arterial Wall Stress,” Ann. Biomed. Eng., 30(4), pp. 509–523. 10.1114/1.1467676 [DOI] [PubMed] [Google Scholar]
  • [41]. Hodis, S. , and Zamir, M. , 2011, “ Pulse Wave Velocity as a Diagnostic Index: The Pitfalls of Tethering Versus Stiffening of the Arterial Wall,” J. Biomech., 44(7), pp. 1367–1373. 10.1016/j.jbiomech.2010.12.029 [DOI] [PubMed] [Google Scholar]
  • [42]. Learoyd, B. M. , and Taylor, M. G. , 1966, “ Alterations With Age in the Viscoelastic Properties of Human Arterial Walls,” Circ. Res., 18(3), pp. 278–292. 10.1161/01.RES.18.3.278 [DOI] [PubMed] [Google Scholar]
  • [43]. Lichtenstein, O. , Safar, M. E. , Mathieu, E. , Poitevin, P. , and Levy, B. I. , 1998, “ Static and Dynamic Mechanical Properties of the Carotid Artery From Normotensive and Hypertensive Rats,” Hypertension, 32(2), pp. 346–350. 10.1161/01.HYP.32.2.346 [DOI] [PubMed] [Google Scholar]
  • [44]. Farzaneh, S. , Trabelsi, O. , and Avril, S. , 2019, “ Inverse Identification of Local Stiffness Across Ascending Thoracic Aortic Aneurysms,” Biomech. Model. Mechanobiol., 18(1), pp. 137–153. [DOI] [PubMed] [Google Scholar]
  • [45]. Bersi, M. R. , Bellini, C. , Humphrey, J. D. , and Avril, S. , 2019, “ Local Variations in Material and Structural Properties Characterize Murine Thoracic Aortic Aneurysm Mechanics,” Biomech. Model. Mechanobiol., 18(1), pp. 203–218. 10.1007/s10237-018-1073-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [46]. Holzapfel, G. A. , 2006, “ Determination of Material Models for Arterial Walls From Uniaxial Extension Tests and Histological Structure,” J. Theor. Biol., 238(2), pp. 290–302. 10.1016/j.jtbi.2005.05.006 [DOI] [PubMed] [Google Scholar]
  • [47]. Caulk, A. W. , Humphrey, J. , and Murtada, S. I. , 2019, “ Fundamental Roles of Axial Stretch in Isometric and Isobaric Evaluations of Vascular Contractility,” ASME J. Biomech. Eng., 141(3), pp. 031008. 10.1115/1.4042171 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [48]. Humphrey, J. D. , Eberth, J. F. , Dye, W. W. , and Gleason, R. L. , 2009, “ Fundamental Role of Axial Stress in Compensatory Adaptations by Arteries,” J. Biomech., 42(1), pp. 1–8. 10.1016/j.jbiomech.2008.11.011 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [49]. Vande Geest, J. P. , Sacks, M. S. , and Vorp, D. A. , 2004, “ Age Dependency of the Biaxial Biomechanical Behavior of Human Abdominal Aorta,” ASME J. Biomech. Eng., 126(6), pp. 815–822. 10.1115/1.1824121 [DOI] [PubMed] [Google Scholar]
  • [50]. Nagai, Y. , Fleg, J. L. , Kemper, M. K. , Rywik, T. M. , Earley, C. J. , and Metter, E. J. , 1999, “ Carotid Arterial Stiffness as a Surrogate for Aortic Stiffness: Relationship Between Carotid Artery Pressure-Strain Elastic Modulus and Aortic Pulse Wave Velocity,” Ultrasound Med. Biol., 25(2), pp. 181–188. 10.1016/S0301-5629(98)00146-X [DOI] [PubMed] [Google Scholar]
  • [51]. Bramwell, J. C. , McDowall, R. J. S. , and McSwiney, B. A. , 1923, “ The Variation of Arterial Elasticity With Blood Pressure in Man (Part I),” Proc. R. Soc. Lond. B. Biol. Sci., 94(663), pp. 450–454. 10.1098/rspb.1923.0015 [DOI] [Google Scholar]
  • [52]. Spronck, B. , Heusinkveld, M. H. , Vanmolkot, F. H. , Roodt, J. O. , Hermeling, E. , Delhaas, T. , Kroon, A. A. , and Reesink, K. D. , 2015, “ Pressure-Dependence of Arterial Stiffness: Potential Clinical Implications,” J. Hypertens., 33(2), pp. 330–338. 10.1097/HJH.0000000000000407 [DOI] [PubMed] [Google Scholar]
  • [53]. Spronck, B. , Delhaas, T. , Butlin, M. , Reesink, K. D. , and Avolio, A. P. , 2017, “ Options for Dealing With Pressure Dependence of Pulse Wave Velocity as a Measure of Arterial Stiffness: An Update of Cardio-Ankle Vascular Index (CAVI) and CAVI0,” Pulse, 5(1–4), pp. 106–114. 10.1159/000479322 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [54]. Shirai, K. , Song, M. , Suzuki, J. , Kurosu, T. , Oyama, T. , Nagayama, D. , Miyashita, Y. , Yamamura, S. , and Takahashi, M. , 2011, “ Contradictory Effects of Beta1- and Alpha1- Aderenergic Receptor Blockers on Cardio-Ankle Vascular Stiffness Index (CAVI)–CAVI Independent of Blood Pressure Computer Methods and Programs in Biomedicinere,” J. Atheroscler. Thromb., 18(1), pp. 49–55. 10.5551/jat.3582 [DOI] [PubMed] [Google Scholar]
  • [55]. Hayashi, K. , Handa, H. , Nagasawa, S. , Okumura, A. , and Moritake, K. , 1980, “ Stiffness and Elastic Behavior of Human Intracranial and Extracranial Arteries,” J. Biomech., 13(2), pp. 175–184. 10.1016/0021-9290(80)90191-8 [DOI] [PubMed] [Google Scholar]
  • [56]. Spronck, B. , Avolio, A. P. , Tan, I. , Butlin, M. , Reesink, K. D. , and Delhaas, T. , 2017, “ Arterial Stiffness Index Beta and Cardio-Ankle Vascular Index Inherently Depend on Blood Pressure But Can Be Readily Corrected,” J. Hypertens., 35(1), pp. 98–104. 10.1097/HJH.0000000000001132 [DOI] [PubMed] [Google Scholar]
  • [57]. Xiao, N. , Alastruey, J. , and Alberto Figueroa, C. , 2014, “ A Systematic Comparison Between 1‐D and 3‐D Hemodynamics in Compliant Arterial Models,” Int. J. Numer. Meth. Bio., 30(2), pp. 204–231. 10.1002/cnm.2598 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [58]. Cuomo, F. , Ferruzzi, J. , Agarwal, P. , Li, C. , Zhuang, Z. W. , Humphrey, J. D. , and Figueroa, C. A. , 2019, “ Sex-Dependent Differences in Central Artery Haemodynamics in Normal and Fibulin-5 Deficient Mice: Implications for Ageing,” Proc. R. Soc. A, 475(2221), p. 20180076. 10.1098/rspa.2018.0076 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [59]. Demiray, H. , 1992, “ Wave Propagation Through a Viscous Fluid Contained in a Prestressed Thin Elastic Tube,” Int. J. Eng. Sci., 30(11), pp. 1607–1620. 10.1016/0020-7225(92)90129-5 [DOI] [Google Scholar]
  • [60]. Chuong, C. J. , and Fung, Y. C. , 1984, “ Compressibility and Constitutive Equation of Arterial Wall in Radial Compression Experiments,” J. Biomech., 17(1), pp. 35–40. 10.1016/0021-9290(84)90077-0 [DOI] [PubMed] [Google Scholar]
  • [61]. Hayenga, H. N. , Trache, A. , Trzeciakowski, J. , and Humphrey, J. D. , 2011, “ Regional Atherosclerotic Plaque Properties in ApoE-/- Mice Quantified by Atomic Force, Immunofluorescence, and Light Microscopy,” J. Vasc. Res., 48(6), pp. 495–504. 10.1159/000329586 [DOI] [PubMed] [Google Scholar]
  • [62]. Leloup, A. J. , Van Hove, C. E. , Kurdi, A. , De Moudt, S. , Martinet, W. , De Meyer, G. R. , Schrijvers, D. M. , De Keulenaer, G. W. , and Fransen, P. , 2016, “ A Novel Set-Up for the Ex Vivo Analysis of Mechanical Properties of Mouse Aortic Segments Stretched at Physiological Pressure and Frequency,” J. Physiol., 594(21), pp. 6105–6115. 10.1113/JP272623 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [63]. Boutouyrie, P. , Bézie, Y. , Lacolley, P. , Challande, P. , Chamiot-Clerc, P. , Benetos, A. , Renaud de la Faverie, J. F. , Safar, M. , and Laurent, S. , 1997, “ In Vivo/In Vitro Comparison of Rat Abdominal Aorta Wall Viscosity. Influence of Endothelial Function,” Arterioscler. Thromb. Vasc. Biol., 17(7), pp. 1346–1355. 10.1161/01.ATV.17.7.1346 [DOI] [PubMed] [Google Scholar]
  • [64]. Bellini, C. , Caulk, A. W. , Li, G. , Tellides, G. , and Humphrey, J. D. , 2017, “ Biomechanical Phenotyping of the Murine Aorta: What Is the Best Control?” ASME J. Biomech. Eng., 139(4), p. 044501. 10.1115/1.4035551 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [65]. Haskett, D. , Speicher, E. , Fouts, M. , Larson, D. , Azhar, M. , Utzinger, U. , and Vande Geest, J. , 2012, “ The Effects of Angiotensin II on the Coupled Microstructural and Biomechanical Response of C57BL/6 Mouse Aorta,” J. Biomech., 45(5), pp. 772–779. 10.1016/j.jbiomech.2011.11.017 [DOI] [PubMed] [Google Scholar]
  • [66]. Leloup, A. J. , Fransen, P. , Van Hove, C. E. , Demolder, M. , De Keulenaer, G. W. , and Schrijvers, D. M. , 2014, “ Applanation Tonometry in Mice: A Novel Noninvasive Technique to Assess Pulse Wave Velocity and Arterial Stiffness,” Hypertension, 64(1), pp. 195–200. 10.1161/HYPERTENSIONAHA.114.03312 [DOI] [PubMed] [Google Scholar]
  • [67]. Fujikura, K. , Luo, J. , Gamarnik, V. , Pernot, M. , Fukumoto, R. , Tilson, M. D., 3rd , and Konofagou, E. E. , 2007, “ A Novel Noninvasive Technique for Pulse-Wave Imaging and Characterization of Clinically-Significant Vascular Mechanical Properties In Vivo,” Ultrason. Imaging, 29(3), pp. 137–154. 10.1177/016173460702900301 [DOI] [PubMed] [Google Scholar]
  • [68]. Trachet, B. , Fraga-Silva, R. A. , Londono, F. J. , Swillens, A. , Stergiopulos, N. , and Segers, P. , 2015, “ Performance Comparison of Ultrasound-Based Methods to Assess Aortic Diameter and Stiffness in Normal and Aneurysmal Mice,” PLoS One, 10(5), p. e0129007. 10.1371/journal.pone.0129007 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [69]. Saphirstein, R. J. , and Morgan, K. G. , 2014, “ The Contribution of Vascular Smooth Muscle to Aortic Stiffness Across Length Scales,” Microcirculation, 21(3), pp. 201–207. 10.1111/micc.12101 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [70]. Swift, J. , Ivanovska, I. L. , Buxboim, A. , Harada, T. , Dingal, P. C. , Pinter, J. , Pajerowski, J. D. , Spinler, K. R. , Shin, J. W. , Tewari, M. , Rehfeldt, F. , Speicher, D. W. , and Discher, D. E. , 2013, “ Nuclear Lamin-A Scales With Tissue Stiffness and Enhances Matrix-Directed Differentiation,” Science, 341(6149), p. 1240104. 10.1126/science.1240104 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [71]. Truesdell, C. , and Noll, W. , 1965, The Non-Linear Field Theories of Mechanics, Springer, New York. [Google Scholar]
  • [72]. Fung, Y. C. , 1973, “ Biorheology of Soft Tissues,” Biorheology, 10(2), pp. 139–155. 10.3233/BIR-1973-10208 [DOI] [PubMed] [Google Scholar]
  • [73]. Dajnowiec, D. , and Langille, B. L. , 2007, “ Arterial Adaptations to Chronic Changes in Haemodynamic Function: Coupling Vasomotor Tone to Structural Remodelling,” Clin. Sci., 113(1), pp. 15–23. 10.1042/CS20060337 [DOI] [PubMed] [Google Scholar]
  • [74]. Humphrey, J. D. , 2001, “ Stress, Strain, and Mechanotransduction in Cells,” ASME J. Biomech. Eng., 123(6), pp. 638–641. 10.1115/1.1406131 [DOI] [PubMed] [Google Scholar]
  • [75]. Bersi, M. R. , Ferruzzi, J. , Eberth, J. F. , Gleason, R. L., Jr. , and Humphrey, J. D. , 2014, “ Consistent Biomechanical Phenotyping of Common Carotid Arteries From Seven Genetic, Pharmacological, and Surgical Mouse Models,” Ann. Biomed. Eng., 42(6), pp. 1207–1223. 10.1007/s10439-014-0988-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [76]. Humphrey, J. D. , Dufresne, E. R. , and Schwartz, M. A. , 2014, “ Mechanotransduction and Extracellular Matrix Homeostasis,” Nat. Rev. Mol. Cell Biol., 15(12), pp. 802–812. 10.1038/nrm3896 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [77]. Greve, J. M. , Les, A. S. , Tang, B. T. , Draney Blomme, M. T. , Wilson, N. M. , Dalman, R. L. , Pelc, N. J. , and Taylor, C. A. , 2006, “ Allometric Scaling of Wall Shear Stress From Mice to Humans: Quantification Using Cine Phase-Contrast MRI and Computational Fluid Dynamics,” Am. J. Physiol. Heart Circ. Physiol., 291(4), pp. H1700–H1708. 10.1152/ajpheart.00274.2006 [DOI] [PubMed] [Google Scholar]
  • [78]. Fung, Y. , 2002, “ Celebrating the Inauguration of the Journal: Biomechanics and Modeling in Mechanobiology,” Biomech. Model. Mechanobiol., 1(1), pp. 3–4. 10.1007/s10237-002-0006-z [DOI] [Google Scholar]
  • [79]. Green, A. E. , and Adkins, J. E. , 1970, Large Elastic Deformations and Non-Linear Continuum Mechanics, Clarendon Press, Oxford, UK. [Google Scholar]
  • [80]. Oden, J. , 1972, Finite Elements of Nonlinear Continua, McGraw-Hill, New York. [Google Scholar]

Articles from Journal of Biomechanical Engineering are provided here courtesy of American Society of Mechanical Engineers