Skip to main content
Journal of Clinical Laboratory Analysis logoLink to Journal of Clinical Laboratory Analysis
. 1998 Dec 12;12(6):363–370. doi: 10.1002/(SICI)1098-2825(1998)12:6<363::AID-JCLA6>3.0.CO;2-X

Reference distributions for immunoglobulins A, G, and M: A practical, simple, and clinically relevant approach in a large cohort

Robert F Ritchie 1,, Glenn E Palomaki 1, Louis M Neveux 1, Olga Navolotskaia 1, Thomas B Ledue 1, Wendy Y Craig 1
PMCID: PMC6808031  PMID: 9850188

Abstract

Serum immunoglobulins are measured millions of times each year, yet clinical interpretations remain hampered by inadequate age‐ and gender‐specific reference limits. In order to provide more reliable and comprehensive reference distributions for IgA, IgG, and IgM measurements, we analyzed automated immunoassay values from 115,017 serum samples from northern New England patients (99% Caucasian) who were tested in our laboratory between 1986 and 1995. Measurements were standardized to reference material, CRM 470 (RPPHS). A simple, practical, and clinically relevant approach was used to determine reference distributions for the immunoglobulins over a wide range of ages for males and females. Levels of IgA and IgM varied considerably by age, and by gender for IgM. For each of the analytes, the observed 5th and 95th centiles were symmetric about the median and approximately constant over the entire age range. When immunoglobulin reference values are expressed as multiples of the age‐ and gender‐specific regressed medians, the resulting distributions fit a log‐Gaussian distribution well. This finding enables interpretation of serum immunoglobulin measurements using a common unit (multiples of the median) that is independent of age or gender. Insights gained from this study can help improve and simplify the interpretation of immunoglobulin measurements. J. Clin. Lab. Anal. 12:363–370, 1998. © 1998 Wiley‐Liss, Inc.

Keywords: serum proteins, reference ranges, Caucasians, CRM 470/RPPHS, multiple of the median

References

  • 1. Ritchie RF, Palomaki GE, Neveux LM, Navolotskaia O: Reference values for immunoglobulins A, G, and M: A comparison of the world's literature to a single large cohort. J Clin Lab Anal 12: 371–377, 1998. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2. National Committee for Clinical Laboratory Standards . Nomenclature and Definitions for use in NRSCL and other NCCLS Documents, 2nd ed. Proposed Guideline. NCCLS Document NRSCL8‐P3. NCCLS, Villanova, PA, 1993. [Google Scholar]
  • 3. Hudson GA, Poulin SE, Ritchie RF: Twelve‐protein immunoassay profile on the COBAS FARA. J Clin Lab Anal 1: 191–197, 1987. [Google Scholar]
  • 4. Rowe D, Anderson SG, Grab B: A research standard for human serum immunoglobulins IgG, IgA and IgM. Bull WHO 42: 535–552, 1970. [PMC free article] [PubMed] [Google Scholar]
  • 5. Reimer CB, Smith SJ, Hannon WH, et al.: Progress towards international reference standards for human serum proteins. J Biol Standard 6: 133–158, 1978. [DOI] [PubMed] [Google Scholar]
  • 6. Whicher JT, Ritchie RF, Johnson AM, et al.: New international reference preparation for proteins in human serum (RPPHS). Clin Chem 40: 934–938, 1994. [PubMed] [Google Scholar]
  • 7. Healy MJR: Outliers in clinical chemistry quality‐control schemes. Clin Chem 25: 675–677, 1979. [PubMed] [Google Scholar]
  • 8. Royston P: Constructing time‐specific reference ranges. Statistics Med 10: 175–190, 1991. [DOI] [PubMed] [Google Scholar]
  • 9. Cohn JS, McNamara JR, Schaefer EJ: Lipoprotein cholesterol concentrations in the plasma of human subjects as measured in the fed and fasted states. Clin Chem 34: 2456–2459, 1988. [PubMed] [Google Scholar]
  • 10. Statland BE, Bokelund H, Winkel P: Factors contributing to intra‐individual variation of serum constituents, 4: Effects of posture and tourniquet application on variation of serum constituents in healthy subjects. Clin Chem 20: 1513–1519, 1974. [PubMed] [Google Scholar]
  • 11. Winkel P, Statland BE, Bokelund H: The effect of time of venipuncture on variation of serum constituents. Am J Clin Pathol 64: 433–447, 1975. [DOI] [PubMed] [Google Scholar]
  • 12. Woodhouse PR, Khaw KT, Plummer M, Foley A, Meade TW: Seasonal variations of plasma fibrinogen and factor VII activity in the elderly: Winter infections and death from cardiovascular disease. Lancet 343: 435–439, 1994. [DOI] [PubMed] [Google Scholar]
  • 13. Miller M, Bachorik PS, Cloey TA: Normal variation of plasma lipoproteins: Postural effects on plasma concentrations of lipids, lipoproteins and apolipoproteins. Clin Chem 38: 569–574, 1992. [PubMed] [Google Scholar]
  • 14. Humphrey KR, Gruemer H‐D, Lott JA: Impact of posture on the “reference range” for serum proteins and calcium. Clin Chem 23: 1343–1346, 1977. [PubMed] [Google Scholar]
  • 15. Tan MH, Errol MD, Wilmshurst G, Gleason RE, Soeldner JS: Effect of posture on serum lipids. New Engl J Med 289: 416–419, 1973. [DOI] [PubMed] [Google Scholar]
  • 16. National Committee for Clinical Laboratory Standards . How to define and determine reference intervals in the clinical laboratory. Approved Guideline. NCCLS Document C28‐A. NCCLS, Villanova, PA, 1995. [Google Scholar]
  • 17. Grasbeck R, Siest G, Wilding P, Williams GZ, Whitehead TP: Provisional recommendation on the theory of reference values (1978). Part 1. The concept of reference values. Clin Chim Acta 87: 461F–465F, 1978. [PubMed] [Google Scholar]
  • 18. Dybkæker R: Approved recommendation (1987) on the theory of reference values. Part 6. Presentation of observed values related to reference values J Clin Chem Clin Biochem 25: 657–662, 1987. [Google Scholar]
  • 19. Harris EK, Boyd JC: Calculating reference limits In Statistical Bases of Reference Values in Laboratory Medicine. Marcel Decker, New York, 1995, pp. 1–61. [Google Scholar]

Articles from Journal of Clinical Laboratory Analysis are provided here courtesy of Wiley

RESOURCES