Harmonization of serum 25‐hydroxycalciferol assay results from high‐performance liquid chromatography, enzyme immunoassay, radioimmunoassay, and immunochemiluminescence systems: A multicenter study

Bahareh Nikooyeh 1, Siamak M Samiee 2, Marjan R Farzami 2, Hamid Alavimajd 3, Maliheh Zahedirad 1, Ali Kalayi 1, Nastaran Shariatzadeh 1, Nasrin Boroumand 3, Elham Golshekan 2, Yalda Gholamian 2, Tirang R Neyestani 1,
PMCID: PMC6817016  PMID: 28169461

Abstract

Background

Remarkable disagreement among different systems of 25‐hydroxy vitamin D 25(OH)D assay makes decision making for both clinical and community interventions very difficult. This study aimed to harmonize the results obtained from different 25(OH)D assay systems.

Methods

A total of 275 serum samples were analyzed for 25(OH)D using DIAsource‐enzyme immunoassay (EIA), DIAsource‐radioimmunoassay (RIA), Roche‐electrochemiluminescence (ECL), Diasorin‐chemiluminescent immunoassay (CLIA), and high‐performance liquid chromatography (HPLC), as the reference method. Serum intact parathyroid hormone (iPTH) was also measured in all samples. Between‐system agreement and harmonization were evaluated using Bland–Altman analysis, receiver operating characteristic (ROC), and regression analysis.

Results

Mean serum 25(OH)D concentrations and frequency distribution of vitamin D status showed a significant difference among the studied systems (P<.001 for both). Serum 25(OH)D assay results from all systems correlated with those from HPLC. As compared with HPLC, ECL showed a positive bias (+3.8 nmol/L), whereas CLIA had a negative bias (−11.9 nmol/L). Both EIA and RIA showed a more or less similar positive bias (8.0 and 8.1 nmol/L, respectively). Using serum iPTH‐based 25(OH)D cutoff points, only ECL results became comparable to and without significant difference with HPLC. However, when system‐specific cutoffs were defined based on HPLC results using regression equations, mean 25(OH)D and frequency distribution of vitamin D status were more harmonized compared with the other methods.

Conclusion

Our findings showed that with adjustment of circulating 25(OH)D based on HPLC, frequency distribution of vitamin D status, as judged by different methods, can be well harmonized with no statistically significant inter‐system difference.

Keywords: 25‐hydroxycalciferol, harmonization, high‐performance liquid chromatography, immunoassay, vitamin D


Abbreviations

AUC

area under the ROC curve

CI

confidence interval

CLIA

chemiluminescent immunoassay

CLSI

Clinical and Laboratory Standards Institute

DBP

vitamin D‐binding protein

DEQAS

Vitamin D External Quality Assurance Scheme

ECL

electrochemiluminescence

EIA

enzyme immunoassay

FDA

Food and Drug Administration

GC

gas chromatography

HOPRM

high‐order primary reference material

HRL

Health Reference Laboratories

HPLC

high‐performance liquid chromatography

iPTH

intact parathyroid hormone

LC

liquid chromatography

MS

mass spectrometry

NNFTRI

National Nutrition and Food Technology Research Institute

NPV

negative predictive value

PPV

positive predictive value

RAM

reference assay method

RIA

radioimmunoassay

RM‐ANOVA

repeated‐measures analysis of variance

RT

room temperature

SD

standard deviation

UVB

ultra violet beam

VDR

vitamin D receptor

VUS

volume under the ROC surface

1. Introduction

Vitamin D has two major vitamers, D2 or ergocalciferol and D3 or cholecalciferol. Ergocalciferol has a plant origin, whereas cholecalciferol is a secosteroid hormone synthesized in the skin following direct exposure to the solar ultra violet beam (UVB) in the spectrum of 290‐315 nm and consequent activation of the precursor 7‐dehydrocholesterol.1 Dietary and endogenous D vitamers have a similar fate in the body. Upon two steps activation in liver and kidney, 25‐hydroxycalciferol (25(OH)D or calcidiol) and then 1, 25‐dihydroxycalciferol (1, 25(OH)2D or calcitriol) are formed, respectively. It is believed that calcidiol is the major circulating form reflecting the body storage, while calcitriol is the functional isoform of the vitamin.2

Despite the presence of an eternal source of natural vitamin D, i.e., solar beam, vitamin D deficiency has become a global health problem for many sociocultural and environmental reasons.3 Poor vitamin D status is accompanied by higher circulating parathyroid hormone (PTH) concentrations and consequent lesser bone mass and muscular weakness.4 Detection of vitamin D receptor (VDR) in many tissues and cells led to definition of many so‐called noncalcemic functions of this vitamin.5 As a result, contribution of vitamin D deficiency has been documented in many human disorders, such as multiple sclerosis, diabetes, and various types of malignancies.6 Early detection and proper treatment of vitamin D deficiency has, therefore, attracted a great deal of concern both in clinical practice and at the community level.7, 8 However, there is no general agreement on definition of vitamin D deficiency at the time. While 25(OH)D concentrations above 50 nmol/L (20 ng/mL) are considered sufficient by some scientific bodies,9 it is debated by some experts.10, 11 The other important issue is determination of circulating calcidiol itself. Though many commercial kits using various analytical techniques have been introduced to the market, measurement of serum 25(OH)D is still not as easy as many other serum analytes.12 Remarkable disagreement among different systems of 25(OH)D assay has been shown by several studies.13, 14, 15 Large variances among the results obtained from different assay systems make clinical follow‐up of patient's status and also national and international comparisons of the prevalence of vitamin D deficiency very problematic, if not impossible.16, 17 For this reason, definition of method‐specific cutoff points18 and standardization of assay results based on a reference method19 have been both suggested.

Lack of standard assay method for 25(OH)D is another issue.20 Methods employing gas chromatography (GC) in conjunction with mass spectrometry detection and recently liquid chromatography (LC)‐tandem mass spectrometry with on‐line solid phase extraction have been proposed as reference methods.21, 22 Despite their high accuracy and precision, these methods cannot be routinely used in either diagnostic or community research laboratories because of their complexity and low throughput.23 Harmonization of 25(OH)D assay results may potentially improve the agreement among different methods and laboratories.24

The term “standardization” denotes to traceability of the results obtained from different systems of assaying an analyte by using a high‐order primary reference material (HOPRM) and/or a reference assay method (RAM).25 In this study, we used HPLC as a reference method. When HOPRM or RAM is not available, harmonization may be employed.25 The Clinical and Laboratory Standards Institute (CLSI) has defined “harmonization” as “the process of recognizing, understanding, and explaining differences while taking steps to achieve uniformity of results, or at minimum, a means of conversion of results such that different groups can use the data obtained from assays interchangeably.”26

In this study, we attempted to harmonize the results obtained from different assay systems of serum calcidiol. To do this, we tried several methods including adjustment of different assay results according to HPLC, as a reference method, and definition of method‐specific cutoff points using either serum iPTH changes or regression analysis model. Finally, the results of all harmonization methods were evaluated and the most proper one was proposed.

2. Materials and Methods

A total of 275 adult subjects aged 20‐60 years were enrolled in the study. Before blood sampling, the aims of the study were fully described for the subjects and then an informed written consent was signed by the participants. This study was approved by the Ethics Committee of the National Nutrition and Food Technology Research Institute (NNFTRI).

2.1. Blood sampling and handling

Ten milliliters fasting blood sample was drawn from antecubital vein. Following centrifugation at 800 g at room temperature (RT) for 15 minutes, sera were recovered and aliquoted in several microtubes which were then kept at −80°C until the day of analysis. Replicates of the serum samples in cryo‐boxes were transferred to both Health Reference Laboratories (HRL) and Laboratory of Day Hospital, accredited by HRL, while preserving the cold chain.

2.2. Determination of circulating 25(OH)D

Serum 25(OH)D concentration of each sample was determined using enzyme immunoassay (EIA), radioimmunoassay (RIA), chemiluminescence, and high‐performance liquid chromatography (HPLC), as the reference method.

2.2.1. EIA and RIA

We used DIAsource kits (Louvain‐la‐Neuve, Belgium) for both EIA and RIA. These kits are HRL approved. EIA tests were performed at HRL, whereas RIA assay was done at the Laboratory of Nutrition Research, NNFTRI. As claimed by the manufacturers of EIA and RIA kits, limits of detection (LOD) were 3.75 and 1 nmol/L, intra‐assay variations were <7.8% and <5.2%, and inter‐assay variations were <9.2% and <9.8%, respectively.

2.2.2. HPLC

Concentrations of 25(OH)D in serum samples were determined by the method described elsewhere27 at the Laboratory of Nutrition Research, NNFTRI. In our hands, LOD was 10 nmol/L and intra‐ and inter‐assay variations were 8.1% and 12.6%, respectively. This laboratory has been participating in Vitamin D External Quality Assurance Scheme (DEQAS) since 2012.

2.2.3. Chemiluminescence

Two chemiluminescence systems were applied for 25(OH)D assay including Elecsys electrochemiluminescence (ECL, Roche, Basel, Switzerland) and Liaison chemiluminescent immunoassay (CLIA, Diasorin, Stillwater, MN, USA). According to the manufacturers, for Elecsys‐ECL (Roche), LOD was 10 nmol/L and intra‐ and inter‐assay variations were <5.7% and <9.9%, respectively. As for Liaison‐CLIA (Diasorin), LOD was 10 nmol/L and intra‐ and inter‐assay variations were <4.8% and <12.2%, respectively. All 25(OH)D assays using these two systems were done at the Laboratory of Day Hospital.

2.3. Intact parathyroid hormone (iPTH) assay

Determination of serum iPTH was performed using Elecsys‐ECL system at the Laboratory of Day Hospital. According to the manufacturer, limit of detection was 1.20 pg/mL. In a multicenter study, intra‐ and inter‐assay variations were 3.1%‐6.6% and 3.4%‐15.6%, respectively. The analytical sensitivity was below 2.70 pg/mL.28

2.4. Statistical analyses

Quantitative or qualitative data were expressed as mean±standard deviation (SD) or absolute and proportional frequencies, respectively. Means were compared using repeated‐measures analysis of variance (RM‐ANOVA) and variances were evaluated by Levene's test. To evaluate agreement among the methods, sensitivity, specificity, and positive and negative predictive values were calculated and Bland–Altman analysis was also used.29, 30 Correlations between continuous data were evaluated using Pearson's or Spearman's correlation coefficient. To compare correlation coefficients, Fisher r‐to‐z test was employed. To adjust different analytical methods based on HPLC, single‐variable regression analysis was used.

Receiver operating characteristic (ROC) analysis was used to determine method‐based cutoff points for 25(OH)D according to serum iPTH concentrations. First, the area under the ROC curve (AUC) was used to evaluate the markers. The ROC curve analysis was then generalized to allow the tests to have more than two classes. The generalized ROC curve leads to a surface. In this case, the AUC changes to the volume under the ROC surface (VUS). In this study, we used the ROC surface analysis to determine two cutoff points, simultaneously. Vitamin D sufficiency was defined as a concentration of circulating 25(OH)D, wherein iPTH concentration attains a plateau.31 Maximum Youden index was considered to set proper cutoff points.32 In this study, P<.05 was considered as significant. All statistical tests were performed using Statistical Package for Social Sciences (SPSS, version 21; SPSS Inc., Chicago, IL, USA).

3. Results

A total of 275 blood samples were taken from the subjects (129 males and 146 females) aged 40.6±10.7 years (males 41.5±10.0 years, females 40.2±10.4 years). Comparison of mean serum 25(OH)D concentrations obtained by HPLC, ECL, CLIA, EIA, and RIA showed a significant difference (43.6±25.8, 46.3±28.5, 34.1±24.4, 51.2±26.5, and 54.8±37.7, respectively, P<.001). Further analysis using paired t‐test followed by Bonferroni correction revealed a significant difference in 25(OH)D concentrations obtained from HPLC with those from ECL, CLIA, EIA, and RIA (P<.001 for all). Comparison of the variance of 25(OH)D concentrations using Levene's test showed a significant difference between HPLC and RIA (P<.001), but HPLC did not differ significantly with ECL (P=.802), CLIA (P=.052), or EIA (P=.605).

Frequency distribution of vitamin D status based on HPLC results significantly differed with the results obtained from other systems with the exception of RIA which surprisingly showed no significant difference (P=.760). Overall, ECL had the most acceptable sensitivity and specificity. CLIA showed highest sensitivity (99.3%) at the cost of its very poor specificity (47.3%).

Serum 25(OH)D assay results from all systems correlated well with those from HPLC, with the strongest correlation with CLIA (r=.883, P<.001), followed by ECL (r=.855, P<.001), EIA (r=.799, P<.001), and RIA (r=.739, P<.001). The correlation coefficient of CLIA was significantly bigger than EIA and RIA as judged by Fisher r‐to‐z test (P<.001 for both). A significant inverse correlation was observed between serum iPTH and 25(OH)D concentrations obtained from HPLC (r=−.221, P<.001), ECL (r=−.289, P<.001), CLIA (r=−.201, P<.001), EIA (r=−.278, P<.001), and RIA (r=−.227, P<.001). Fisher r‐to‐z test showed no significant difference among these correlation coefficients.

To evaluate agreement between assay systems, Bland–Altman analysis was employed. As compared with HPLC, ECL showed a positive bias (+3.8 nmol/L), whereas CLIA had a negative bias (−11.9 nmol/L). Both EIA and RIA showed almost similar positive bias (8.0 and 8.1 nmol/L, respectively), but especially for RIA, several samples demonstrated huge differences.

Using regression analysis, an equation was developed to adjust the results from each method according to HPLC as follows:

  1. ECL‐Roche: 25(OH)DAdj nmol/L=(ECL‐3.55)/0.99; 95% CI: 3.08‐3.15

  2. CLIA‐Diasorin: 25(OH)DAdj nmol/L=(CLIA+4.55)/0.84; 95% CI: −12.36 to −11.27

  3. EIA‐DIAsource: 25(OH)DAdj nmol/L=(EIA‐14.12)/0.856; 95% CI: 7.44‐8.47

  4. RIA‐DIAsource: 25(OH)DAdj nmol/L=(RIA+2.81)/1.219; 95% CI: 6.65‐8.44

wherein 25(OH)DAdj is adjusted serum calcidiol concentration.

RM‐ANOVA revealed no statistical significant difference in mean harmonized circulating 25(OH)D concentrations among HPLC, ECL, CLIA, EIA, and RIA (43.6±25.8, 43.2±28.8, 46.0±29.1, 42.7±29.6, and 42.7±29.6 nmol/L, respectively, P=.514). Comparison of HPLC and other systems using paired t test followed by Bonferroni adjustment revealed no significant difference between HPLC results and those of adjusted values obtained from equations for ECL (P=.219), CLIA (P=.219), EIA (P=.249), or RIA (P=.248). When occurrence of vitamin D deficiency and insufficiency was calculated using adjusted values, no significant difference was observed between HPLC and other systems except for RIA. Table 1 shows vitamin D status of the studied population based on predefined cutoff points (deficiency <27.5, insufficiency: 27.5‐50, and sufficiency >50 nmol/L) before and after harmonization. Table 2 demonstrates pre‐ and post‐harmonization sensitivity, specificity, and positive and negative predictive values of different methods as compared with the reference method, i.e., HPLC, to diagnose undesirable vitamin D status (<50 nmol/L).

Table 1.

Vitamin D status of the studied subjects based on different systems of 25(OH)D analysis [n (%)] before and after harmonization based on HPLC results

System Before harmonization P valuea After harmonization P valuea
Deficient Insufficient Sufficient Deficient Insufficient Sufficient
HPLC 97 (35.8) 81 (29.9) 93 (34.3) 97 (35.8) 81 (29.9) 93 (34.3)
ECL‐Roche 87 (32.7) 77 (28.9) 104 (38.3) .016 106 (39.8) 69 (25.9) 91 (34.2) .368
CLIA‐Diasorin 120 (48.4) 83 (33.5) 45 (18.1) <.001 76 (30.6) 89 (35.9) 83 (33.5) .172
EIA‐DIAsource 40 (15.0) 110 (41.4) 116 (43.6) <.001 79 (29.7) 109 (41.0) 78 (29.3) .614
RIA‐DIAsource 68 (30.4) 59 (26.3) 97 (43.3) .760 87 (38.8) 54 (24.1) 83 (37.1) .010

Vitamin D status definitions based on serum 25(OH)D concentration (nmol/L): deficiency: <27.5; insufficiency: 27.6‐50.0; sufficiency: >50.

a

Comparison of each method with HPLC was done by Wilcoxon's test.

Table 2.

Sensitivity, specificity, and positive and negative predictive values of different systems compared with HPLC before and after harmonization

System Before harmonization After harmonization
ECL‐Roche CLIA‐Diasorin EIA‐DIAsource RIA‐DIAsource ECL‐Roche CLIA‐Diasorin EIA‐DIAsource RIA‐DIAsource
Sensitivity 86.4 CI: 80.2‐90.9 99.3 CI: 95.9‐99.9 78.1 CI: 71.1‐83.7 82.7 CI: 74.9‐88.5 90.3 CI: 84.8‐94.1 93.5 CI: 88.1‐96.6 91.4 CI:86.0‐94.9 87.2 CI: 80.0‐92.1
Specificity 87.9 CI: 78.9‐93.5 47.3 CI: 36.9‐57.8 87.9 CI: 78.9‐93.5 81.7 CI: 72.0‐88.6 83.1 CI: 73.4‐89.9 78.4 CI: 68.5‐86.0 69.2 CI: 58.5‐78.2 72.5 CI: 62.0‐81.1
PPV 93.2 CI: 88.0‐96.4 75.8 CI: 69.2‐81.4 92.6 CI: 86.9‐96.0 86.6 CI: 79.1‐91.7 91.4 CI: 86.0‐94.9 87.8 CI: 81.6‐92.2 85.1 CI: 79.0‐89.7 82.2 CI: 74.7‐87.9
NPV 76.9 CI: 67.4‐84.3 97.7 CI: 86.7‐99.8 67.2 CI: 57.9‐75.3 76.7 CI: 67.0‐84.4 81.3 CI: 71.4‐88.4 87.9 CI: 78.5‐93.7 80.7 CI: 69.9‐88.4 79.5 CI: 68.9‐87.2

NPV, negative predictive value; PPV, positive predictive value.

Vitamin D status was defined based on serum 25(OH)D concentrations (nmol/L): desirable >50, undesirable <50.

In another attempt to harmonize calcidiol assay results, serum concentrations of 25(OH)D and iPTH were applied in ROC analysis model. By considering circulating 25(OH)D concentrations wherein serum concentration of iPTH attains a plateau (28 ng/L) and the proposed serum level of PTH indicating hyperparathyroidism (>65 ng/L),33 new serum 25(OH)D cutoff points were determined for the reference HPLC method. Thus, the cutoff points for vitamin D deficiency, insufficiency, and sufficiency were then defined based on serum 25(OH)D concentrations (nmol/L) as follows:

  • HPLC: <19.0; 19.0‐38.6; >38.6

  • ECL‐Roche: <19.5; 19.5‐43.5; >43.5

  • CLIA‐Diasorin: <13.4; 13.4‐30.0; >30.0

  • EIA‐DIAsource: <34.2; 34.2‐51.2; >51.2

  • RIA‐DIAsource: <22.6; 22.6‐51.4; >51.4

By using these system‐specific cutoff points, proportion of the subjects diagnosed as deficient decreased dramatically, while proportion of insufficient subjects increased (Table 3). This was especially the case with HPLC, ECL, and CLIA. However, only ECL results became comparable to and without significant difference with HPLC results. Again here, both EIA and RIA showed less specificity than other systems (Table 4).

Table 3.

Vitamin D status of the studied subjects according to serum iPTH‐based system‐specific cutoff points for 25(OH)D assay [n (%)]

System Deficient Insufficient Sufficient P value
HPLC 37 (13.7) 105 (38.7) 129 (47.6)
ECL‐Roche 39 (14.7) 110 (41.4) 117 (44.0) .174
CLIA‐Diasorin 30 (12.1) 101 (40.7) 117 (47.2) .012
EIA‐Diasource 58 (21.8) 98 (36.8) 110 (41.4) <.001
RIA‐Diasource 44 (19.6) 87 (38.8) 93 (41.5) <.001

Table 4.

Sensitivity, specificity, and positive and negative predictive values of different systems compared with HPLC based on system‐specific cutoff points

ECL‐Roche CLIA‐Diasorin EIA‐DIAsource RIA‐DIAsource
Sensitivity 92.1 CI: 86.1‐95.8 92.4 CI: 85.7‐96.2 90.1 CI: 83.7‐94.3 91.8 CI: 84.0‐96.1
Specificity 84.6 CI: 76.8‐90.2 83.5 CI: 75.7‐89.3 77.2 CI: 68.6‐84.1 67.2 CI: 58.1‐75.1
PPV 87.2 CI: 80.5‐91.9 83.9 CI: 76.3‐89.5 82.0 CI: 74.9‐87.5 68.7 CI: 59.9‐76.3
NPV 90.5 CI: 83.2‐94.9 92.2 CI: 85.3‐96.1 87.1 CI: 79.0‐92.5 91.3 CI: 83.1‐95.8

Vitamin D status based on circulating 25(OH)D concentration (nmol/L) for HPLC: sufficiency >38.55, insufficiency ≤38.55; ECL‐Roche: sufficiency >43.5, insufficiency ≤43.5; CLIA‐Diasorin: Sufficiency >30.0; insufficiency ≤30.0; EIA‐Diasource: sufficiency >51.2; insufficiency ≤51.2; RIA‐Diasource: Sufficiency >51.4; insufficiency ≤51.4.

To develop HPLC‐specific cutoff points, changes of serum iPTH in different concentrations of serum 25(OH)D were used. Using ROC analysis, these cutoffs were then employed to develop cutoff points for other systems.

Since vitamin D functions are by no means confined just to the bones and hard tissues, determination of proper cutoff points for circulating 25(OH)D has been extremely challenging and controversial. With the assumption that IOM proposed cutoff points are based on a reference method, another set of system‐specific cutoffs were defined using the above regression equations. Based on serum concentrations of 25(OH)D (nmol/L), the cutoff points for vitamin D deficiency, insufficiency, and sufficiency were then defined as follows:

  • HPLC: <27.5; 27.5‐50; >50.0

  • ECL‐Roche: <30.8; 30.8‐53.0; >53.0

  • CLIA‐Diasorin: <18.6; 18.6‐37.4; >37.4

  • EIA‐DIAsource: <37.7; 37.7‐59.9; >59.9

  • RIA‐DIAsource: <30.7; 30.7‐58.1; >58.1

Using these cutoff points, distribution of vitamin D status in our subjects and diagnostic characteristics of the systems were the same as in Table 1, after harmonization (data not shown). To further evaluate this method actually, 25(OH)D concentration in 50 additional serum samples obtained from Day Hospital was determined using the studied systems. Mean 25(OH)D concentrations showed a significant between‐system difference (P<.001). Post hoc comparisons of each system with HPLC revealed that, except for ECL‐Roche, mean calcidiol concentration from all other systems differed significantly from that of HPLC. However, this between‐system difference disappeared after standardization (P=.624). The significant difference in distribution of vitamin D status according to different systems also removed following standardization (data not shown).

4. Discussion

We found that harmonization of the results of different assay systems caused an agreement in means of serum 25(OH)D concentrations and also in occurrence rates of vitamin D deficiency/insufficiency. The only exception was RIA based on which distribution of vitamin D status in our population following harmonization differed significantly with that of HPLC. Re‐evaluation of data revealed that the major difference in vitamin D status categories between HPLC and RIA was in deficiency and insufficiency. In other words, RIA had more efficiency in discrimination of undesirable from desirable status (i.e., calcidiol concentrations below and above 50 nmol/L, respectively). Harmonization caused a relative improvement of RIA sensitivity at the cost of a more decrease in its specificity.

Wide agreement range of RIA and HPLC, as judged by Bland–Altman analysis, indicates a wide scattering of RIA results around real values; thus, actually no acceptable agreement could be made between RIA and HPLC by any methods.

It has been proposed that for serum 25(OH)D, like other blood analytes, there must be just a normal range of 32‐100 ng/mL (80‐250 nmol/L) and categorization of vitamin D status to insufficiency and deficiency is unnecessary and confusing.34 Obviously, raising desirable limit of serum calcidiol to this level will inevitably lead to a dramatic increase in prevalence rates of vitamin D deficiency and also to changes of sensitivity and specificity of assay systems. Notwithstanding, there is no general agreement on this proposed normal range.35, 36 Classification of undesirable vitamin D status to “deficiency” and “insufficiency” bears a clinical implication. Vitamin D deficiency is accompanied by a high risk of bone problems, whereas vitamin D insufficiency is associated with an increased risk of noncalcemic complications.36 Consequently, identification of different states of undesirable vitamin D status and further prediction of its potential outcomes can have a determining role in management of the problem at both clinical and community settings.

In this study, cutoff points for circulating 25(OH)D concentrations were determined according to serum iPTH concentrations. Based on these cutoffs, only ECL results (including frequency distribution of vitamin D status) became almost similar to those of HPLC. In a study on 214 serum samples collected between February 2005 and December 2011 from children aged 0.1‐19.2 years, vitamin D deficiency, based on elevation of serum PTH concentrations above 50 ng/L, was defined as circulating 25(OH)D <34 nmol/L. However, in each time period during the study and in accord with introduction of new systems to the market, one type of assay system for 25(OH)D was employed. Consequently, during 6 years of serum sample collection, three different systems, including Nichols‐chemiluminescence (7 months), IDS‐RIA (~2.5 years) and finally LC‐tandem MS (~3 years), with different performance characteristics were used to measure circulating calcidiol.37 The major critique of PTH‐based serum 25(OH)D cutoff points is that only calcemic effects of vitamin D are considered in this way. While a desirable level of circulating calcidiol is a concentration at which both calcemic and noncalcemic effects of vitamin D are exerted for a long‐term health. In British National Diet and Nutrition Survey, some elderly subjects (85 years plus) had high serum PTH concentrations despite having desirable vitamin D status,38 indicating that PTH‐based serum calcidiol diagnostic limits may not be applicable for some subpopulations.

To harmonize 25(OH)D assay results from various systems, we proposed system‐specific cutoff points by using regression analysis. In this case, diagnostic characteristics of the systems actually did not differ with those from standardization of the results. Notwithstanding, using defined cutoff points is much easier for both medical and research laboratories.

During the recent decades, a remarkable improvement has been made in measurement of circulating 25(OH)D. According to DEQAS report, inter‐laboratory variations have decreased from 32% in 1994 to 15.3% in 2009.14 Nevertheless, high inter‐method and inter‐laboratory variations are still challenging. There are several reasons for these variations. Vitamin D is a hydrophobic and matrix‐sensitive compound. However, vitamin D analytes are stable for 2 weeks at 30°C and for 1 year (and even longer) at −20°C and are not affected by up to four freeze‐thaw cycles of serum samples.23 Ultra violet beam also does not influence calcidiol content of a serum sample.34 Both D2 and D3 isoforms may be found in the body. Binding molecules (antibodies or vitamin D‐binding protein [DBP]) commonly used in immunoassays have usually more affinity to D3 isoform.20 However, this issue does not seem to be problematic in our study. Because of very limited food sources, ergocalciferol is actually undetectable in the circulation of most, if not all, Iranians.27

Some limitations of this study are acknowledged. Results from a commercial assay kit of a certain method cannot necessarily be extended to all kits of that method. Though we used HRL‐approved kits, performance characteristics of different manufacturers may vary. Immunoassay kits from Diasorin and IDS both have FDA approval, but performance characteristics of Diasorin were shown to be actually more satisfactory.39 Harmonization, by any method, is mostly applicable to minimize the systematic errors. Assay systems with inherent possibility of random errors may not be harmonized efficiently. In previous generation of commercial immunoassay kits, for instance, there was one step of solvent extraction for DBP removal and releasing calcidiol. This step was highly prone to random errors. In newer generations, this step is performed in situ using a denaturing agent which is the manufacturer's secret so there is no available data on its compound and efficacy.39 Finally, though the results of our evaluation study on additional serum samples seem promising, the number of the tested samples was limited and the analyses were performed in the same laboratory in which the other tests were done. Therefore, further studies on a large number of serum samples in different laboratories are needed.

5. Conclusion

Our findings showed that with adjustment of circulating 25(OH)D based on HPLC, frequency distribution of deficiency, insufficiency, and sufficiency, as judged by different methods, can be well harmonized with no statistically significant inter‐system difference. We were unable to harmonize RIA‐DIAsource results by any methods because of its very wide agreement range with HPLC. Further researches are needed to evaluate this method in large population‐based studies.

Authors Contributions

This study was designed and supervised by TRN with intellectual assistance of BN. All statistical analyses were done by BN, HA, and NB. SMS and MRF supervised the tests performed at HRL and Day Laboratory. MZ performed HPLC tests. Other laboratory bench works were performed by AK, EG, YG, NS, BN, and TRN. The manuscript was written by TRN and BN.

Nikooyeh B, Samiee SM, Farzami MR, et al. Harmonization of serum 25‐hydroxycalciferol assay results from high‐performance liquid chromatography, enzyme immunoassay, radioimmunoassay, and immunochemiluminescence systems: A multicenter study. J Clin Lab Anal. 2017;31:e22117 10.1002/jcla.22117

References

  • 1. Nikooyeh B, Neyestani TR. Cholesterol and vitamin D: how the ‘mother’ and ‘daughter’ molecules interact In: Watson RR, De Meester F, eds. Handbook of Cholesterol. Wageningen: Wageningen Academic Publishers; 2016:427–450. [Google Scholar]
  • 2. Grant WB, Wimalawansa SJ, Holick MF, et al. Emphasizing the health benefits of vitamin D for those with neurodevelopmental disorders and intellectual disabilities. Nutrients. 2015;7:1538–1564. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Hossein‐nezhad A, Holick MF. Vitamin D for health: a global perspective. Mayo Clin Proc. 2013;88:720–755. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4. Girgis CM, Clifton‐Bligh RJ, Hamrick MW, Holick MF, Gunton JE. The roles of vitamin D in skeletal muscle: form, function, and metabolism. Endocr Rev. 2013;34:33–83. [DOI] [PubMed] [Google Scholar]
  • 5. Nezbedova P, Brtko J. 1alpha,25‐dihydroxyvitamin D3 inducible transcription factor and its role in the vitamin D action. Endocr Regul. 2004;38:29–38. [PubMed] [Google Scholar]
  • 6. Saggese G, Vierucci F, Boot AM, et al. Vitamin D in childhood and adolescence: an expert position statement. Eur J Pediatr. 2015;174:565–576. [DOI] [PubMed] [Google Scholar]
  • 7. Cannell JJ, Hollis BW, Zasloff M, Heaney RP. Diagnosis and treatment of vitamin D deficiency. Expert Opin Pharmacother. 2008;9:107–118. [DOI] [PubMed] [Google Scholar]
  • 8. Calvo MS, Whiting SJ. Public health strategies to overcome barriers to optimal vitamin D status in populations with special needs. J Nutr. 2006;136:1135–1139. [DOI] [PubMed] [Google Scholar]
  • 9. Institute of Medicine (IOM) . Dietary Reference Intakes for Calcium and Vitamin D. Washington, DC: National Academies Press; 2011. [PubMed] [Google Scholar]
  • 10. Heaney RP, Holick MF. Why the IOM recommendations for vitamin D are deficient. J Bone Miner Res. 2011;26:455–457. [DOI] [PubMed] [Google Scholar]
  • 11. Rosen CJ, Abrams SA, Aloia JF, et al. IOM committee members respond to Endocrine Society vitamin D guideline. J Clin Endocrinol Metab. 2012;97:1146–1152. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12. Hollis BW. Editorial: the determination of circulating 25‐hydroxyvitamin D: no easy task. J Clin Endocrinol Metab. 2004;89:3149–3151. [DOI] [PubMed] [Google Scholar]
  • 13. Carter GD, Carter R, Jones J, Berry J. How accurate are assays for 25‐hydroxyvitamin D? Data from the international vitamin D external quality assessment scheme. Clin Chem. 2004;50:2195–2197. [DOI] [PubMed] [Google Scholar]
  • 14. Carter G, Berry J, Gunter E, et al. Proficiency testing of 25‐hydroxyvitamin D (25‐OHD) assays. J Steroid Biochem Mol Biol. 2010;121:176–179. [DOI] [PubMed] [Google Scholar]
  • 15. Carter GD. Accuracy of 25‐hydroxyvitamin D assays: confronting the issues. Curr Drug Targets. 2011;12:19–28. [DOI] [PubMed] [Google Scholar]
  • 16. Binkley N, Krueger D, Cowgill CS, et al. Assay variation confounds the diagnosis of hypovitaminosis D: a call for standardization. J Clin Endocrinol Metab. 2004;89:3152–3157. [DOI] [PubMed] [Google Scholar]
  • 17. Sempos CT, Vesper HW, Phinney KW, Thienpont LM, Coates PM. Vitamin D status as an international issue: national surveys and the problem of standardization. Scand J Clin Lab Invest Suppl. 2012;243:32–40. [DOI] [PubMed] [Google Scholar]
  • 18. Glendenning P, Taranto M, Noble JM, et al. Current assays overestimate 25‐hydroxyvitamin D3 and underestimate 25‐hydroxyvitamin D2 compared with HPLC: need for assay‐specific decision limits and metabolite‐specific assays. Ann Clin Biochem. 2006;43(Pt 1):23–30. [DOI] [PubMed] [Google Scholar]
  • 19. Thienpont LM, Stepman HC, Vesper HW. Standardization of measurements of 25‐hydroxyvitamin D3 and D2. Scand J Clin Lab Invest Suppl. 2012;243:41–49. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20. Su Z, Narla SN, Zhu Y. 25‐Hydroxyvitamin D: analysis and clinical application. Clin Chim Acta. 2014;433:200–205. [DOI] [PubMed] [Google Scholar]
  • 21. Knox S, Harris J, Calton L, Wallace AM. A simple automated solid‐phase extraction procedure for measurement of 25‐hydroxyvitamin D3 and D2 by liquid chromatography‐tandem mass spectrometry. Ann Clin Biochem. 2009;46(Pt 3):226–230. [DOI] [PubMed] [Google Scholar]
  • 22. Thibeault D, Caron N, Djiana R, Kremer R, Blank D. Development and optimization of simplified LC‐MS/MS quantification of 25‐hydroxyvitamin D using protein precipitation combined with on‐line solid phase extraction (SPE). J Chromatogr B Analyt Technol Biomed Life Sci. 2012;883–884:120–127. [DOI] [PubMed] [Google Scholar]
  • 23. Wootton AM. Improving the measurement of 25‐hydroxyvitamin D. Clin Biochem Rev. 2005;26:33–36. [PMC free article] [PubMed] [Google Scholar]
  • 24. Binkley N, Sempos CT. Standardizing vitamin D assays: the way forward. J Bone Miner Res. 2014;29:1709–1714. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25. Plebani M. Harmonization in laboratory medicine: the complete picture. Clin Chem Lab Med. 2013;51:741–751. [DOI] [PubMed] [Google Scholar]
  • 26. Clinical and Laboratory Standards Institute (CLSI) . Harmonized Terminology Database. 2012. http://www.clsi.org/source/custom/termsall.cfm?Section=Harmonized_Terminology_Database. Accessed June 29, 2012.
  • 27. Neyestani TR, Gharavi A, Kalayi A. Determination of serum 25‐hydroxy cholecalciferol using high‐performance liquid chromatography: a reliable tool for assessment of vitamin D status. Int J Vitam Nutr Res. 2007;77:341–346. [DOI] [PubMed] [Google Scholar]
  • 28. Hermsen D, Franzson L, Hoffmann JP, et al. Multicenter evaluation of a new immunoassay for intact PTH measurement on the Elecsys System 2010 and 1010. Clin Lab. 2002;48:131–141. [PubMed] [Google Scholar]
  • 29. Akobeng AK. Understanding diagnostic tests 1: sensitivity, specificity and predictive values. Acta Paediatr. 2007;96:338–341. [DOI] [PubMed] [Google Scholar]
  • 30. Dewitte K, Fierens C, Stockl D, Thienpont LM. Application of the Bland‐Altman plot for interpretation of method‐comparison studies: a critical investigation of its practice. Clin Chem. 2002;48:799–801; author reply ‐2. [PubMed] [Google Scholar]
  • 31. Vieth R, Ladak Y, Walfish PG. Age‐related changes in the 25‐hydroxyvitamin D versus parathyroid hormone relationship suggest a different reason why older adults require more vitamin D. J Clin Endocrinol Metab. 2003;88:185–191. [DOI] [PubMed] [Google Scholar]
  • 32. Fluss R, Faraggi D, Reiser B. Estimation of the Youden Index and its associated cutoff point. Biom J. 2005;47:458–472. [DOI] [PubMed] [Google Scholar]
  • 33. Holick MF. Vitamin D status: measurement, interpretation, and clinical application. Ann Epidemiol. 2009;19:73–78. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34. Hollis BW. Assessment of vitamin D status and definition of a normal circulating range of 25‐hydroxyvitamin D. Curr Opin Endocrinol Diabetes Obes. 2008;15:489–494. [DOI] [PubMed] [Google Scholar]
  • 35. Saintonge S, Bang H, Gerber LM. Implications of a new definition of vitamin D deficiency in a multiracial us adolescent population: the National Health and Nutrition Examination Survey III. Pediatrics. 2009;123:797–803. [DOI] [PubMed] [Google Scholar]
  • 36. Thacher TD, Clarke BL. Vitamin D insufficiency. Mayo Clin Proc. 2011;86:50–60. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37. Atapattu N, Shaw N, Hogler W. Relationship between serum 25‐hydroxyvitamin D and parathyroid hormone in the search for a biochemical definition of vitamin D deficiency in children. Pediatr Res. 2013;74:552–556. [DOI] [PubMed] [Google Scholar]
  • 38. Bates CJ, Carter GD, Mishra GD, O'Shea D, Jones J, Prentice A. In a population study, can parathyroid hormone aid the definition of adequate vitamin D status? A study of people aged 65 years and over from the British National Diet and Nutrition Survey. Osteoporos Int. 2003;14:152–159. [DOI] [PubMed] [Google Scholar]
  • 39. Wallace AM, Gibson S, de la Hunty A, Lamberg‐Allardt C, Ashwell M. Measurement of 25‐hydroxyvitamin D in the clinical laboratory: current procedures, performance characteristics and limitations. Steroids. 2010;75:477–488. [DOI] [PubMed] [Google Scholar]

Articles from Journal of Clinical Laboratory Analysis are provided here courtesy of Wiley

RESOURCES