Skip to main content
Journal of Clinical Laboratory Analysis logoLink to Journal of Clinical Laboratory Analysis
. 2018 Nov 2;33(3):e22703. doi: 10.1002/jcla.22703

Comparison between DiaPlexQ™ STI6 and GeneFinder™ STD I/STD II multiplex Real‐time PCR Kits in the detection of six sexually transmitted disease pathogens

Hee Jae Huh 1,, Chang‐Seok Ki 2, Sun Ae Yun 3, Jungsoo Lee 4, Gwi Young Oh 4, Nam‐Sihk Lee 4, Young Ho Yoon 4, Nam Yong Lee 1
PMCID: PMC6818581  PMID: 30390337

Abstract

Background

The DiaPlexQ™ STI6 Detection Kit (DiaPlexQ; Solgent Co., Ltd., Daejeon, South Korea) is a multiplex real‐time PCR assay for the detection of the following sexually transmitted disease (STD) pathogens: Chlamydia trachomatis, Neisseria gonorrhoeae, Mycoplasma hominis, Trichomonas vaginalis, Ureaplasma urealyticum, and Mycoplasma genitalium. We compared the performance of the DiaPlexQ assay with the GeneFinder™ STD I (CT/NG/UU) and STD II (MG/MH/TV) Multiplex Real‐time PCR Kits (GeneFinder; Infopia Co., Ltd., Anyang, South Korea).

Methods

We evaluated the performance of the DiaPlexQ assay in comparison to that of GeneFinder using 1106 clinical specimens (542 genital swabs and 564 urine samples). The analytical performance of the DiaPlexQ assay, including the limit of detection (LOD) and analytical specificity, was evaluated using reference strains.

Results

The positive percent agreement, negative percent agreement, and kappa value between the two assays were 96.6%‐99.4%, 98.2%‐99.8%, and 0.93%‐0.99%, respectively. No cross‐reactivity was observed in a collection of 41 different microorganisms and the LOD of the DiaPlexQ assay ranged from 1 to 10 copies/reaction for each microorganism.

Conclusion

The DiaPlexQ assay showed comparable performance to that of the GeneFinder assay so that it can be used for the screening and diagnosis of non‐viral curable STD pathogens.

Keywords: multiplex real‐time PCR, sexually transmitted disease, validation

1. INTRODUCTION

Sexually transmitted diseases (STDs) impose substantial strain on the budgets of both households and national health systems, and have an adverse effect on the overall well‐being of individuals. The World Health Organization (WHO) estimated that there are 357 million new cases of gonorrhea, chlamydia, syphilis, and trichomoniasis annually among people aged 15‐49 years.1 Several methods are available for detecting non‐viral curable STD pathogens, including culture, microscopic identification, antigen detection testing, and nucleic acid amplification tests (NAATs).2, 3, 4 NAAT is a highly sensitive method with excellent specificity, and therefore it is the preferred method for detecting STD pathogens. Moreover, multiplex molecular assays have an additional advantage in screening and diagnosis since they enable simultaneous diagnosis of several sexually transmitted infections (STIs). Several commercial multiplex PCR kits have been developed, and a few studies have evaluated their performance.5, 6, 7, 8, 9, 10, 11, 12, 13

Recently, the DiaPlexQ™ STI6 Detection Kit (DiaPlexQ; Solgent Co., Ltd., Daejeon, South Korea), a European Conformity‐In Vitro Diagnostic marked multiplex real‐time PCR assay, was developed and received Korean Ministry of Food and Drug Safety (MFDS) approval. The assay comprises two reaction tubes and targets six STD pathogens: set 1 includes Chlamydia trachomatis (CT), Neisseria gonorrhoeae (NG), and Mycoplasma hominis (MH); set 2 includes Trichomonas vaginalis (TV), Ureaplasma urealyticum (UU), and Mycoplasma genitalium (MG).

The purpose of this study was to evaluate the analytical performance of the DiaPlexQ assay and compare the results with those of a comparative method, the GeneFinder™ STD I (CT/NG/UU) and STD II (MG/MH/TV) Multiplex Real‐time PCR Kit (GeneFinder; Infopia Co., Ltd., Anyang, South Korea).

2. MATERIALS AND METHODS

This retrospective study was approved by the Institutional Review Board of Samsung Medical Center. A total of 1100 patients who were suspected of having an STI and who visited gynecology or urology hospitals were enrolled. A total of 1106 clinical specimens were obtained from a total of 542 genital swabs (531 vaginal swabs and 11 urethral swabs) and 564 urine samples collected from Feb. to Nov. 2016. Patient ages ranged from 13 to 90 years old. DNA was extracted using a QIAamp DNA Mini Kit (Qiagen cat#51304; Qiagen, Hilden, Germany) according to the manufacturer's instructions.

For the DiaPlexQ assay, extracted DNA (5 μL) was added to a tube containing 15 μL of PCR premix (10 μL of 2X Multiplex Real‐Time PCR Smart mix and 5 μL of Primer & Probe Mixture). PCR was conducted using the 7500 Fast Real‐Time PCR System (Applied Biosystems, Foster City, CA, USA). The mixture was incubated at 50˚C for 3 minutes to activate the uracil‐DNA glycosylase (UDG) system, followed by pre‐denaturation at 95˚C for 15 minutes and 45 cycles of PCR (20 seconds at 95˚C and 40 seconds at 60˚C). A positive test result was defined as a cycle threshold (Ct) ≤43 for individual targets.

The procedures for the GeneFinder assay were performed according to the manufacturer's instructions. The GeneFinder assay is a Korean MFDS‐approved multiplex real‐time assay for the detection of STD pathogens. The GeneFinder assay targets six STD pathogens: STD I detects CT, NG, and UU; STD II detects MG, MH, and TV. PCR was performed in a total volume of 20 μL (10 μL PCR reaction mixture, 5 μL probe mixture, and 5 μL DNA) using the 7500 Fast Real‐Time PCR System. A positive test result was defined as a cycle threshold (Ct) ≤40 for individual targets.

Comparison of diagnostic performance between the DiaPlexQ and GeneFinder assays was assessed by measuring positive/negative percent agreement values and kappa coefficients using the VassarStats website (https://vassarstats.net/). Specimens with discordant results between the DiaPlexQ assay and the GeneFinder assay were further confirmed by monoplex PCR and sequencing. The PCR protocols and primers used are listed in Table S1. After detection of PCR products by gel electrophoresis and purification, sequencing was performed on an automated ABI Prism 3730 instrument using a BigDye terminator cycle sequencing kit (Applied Biosystems, Foster City, CA, USA). Positive and negative controls were included in each run.

Analytical sensitivity was determined by directly comparing dilutions of six type strains in cell culture and in the DiaPlex assay. Serial dilutions of quantified target organism were made from 100 to 104 copies/reaction by spiking negative genital swabs/urine samples. Serial dilutions were analyzed with 24 replicates per dilution. The lower detection limit was determined via Probit analysis. Analytical specificity of the DiaPlexQ assay was evaluated using 41 different microorganisms (Table 1).

Table 1.

Analytic specificity of the DiaPlexQ™ STI6 Detection Kit

No. Microbial species Strain Result
Set 1 Set 2
1 Gardnerella vaginalis ATCC‐49145D‐5 Negative Negative
2 Treponema pallidum ATCC‐BAA‐2642SD Negative Negative
3 Haemophilus ducreyi ATCC‐51620 Negative Negative
4 Candida albicans KCTC‐7678 Negative Negative
5 Ureaplasma parvum ATCC‐27815 Negative Negative
6 Neisseria meningitidis ATCC‐700532DQ Negative Negative
7 Proteus mirabilis KCTC‐2510 Negative Negative
8 Corynebacterium accolens KCTC‐3431 Negative Negative
9 Corynebacterium ammoniagenes KCTC‐1838 Negative Negative
10 Corynebacterium diphtheriae KCTC‐3075 Negative Negative
11 Corynebacterium glutamicum KCTC‐3017 Negative Negative
12 Enterobacter aerogenes KCTC‐2190 Negative Negative
13 Enterobacter cloacae KCTC‐2361 Negative Negative
14 Enterococcus faecalis KCTC‐3206 Negative Negative
15 Klebsiella oxytoca KCTC‐1686 Negative Negative
16 Klebsiella planticola KCTC‐1687 Negative Negative
17 Klebsiella pneumoniae KCTC‐2242 Negative Negative
18 Pseudomonas aeruginosa KCTC‐2004 Negative Negative
19 Sphingomonas paucimobilis ATCC‐29837 Negative Negative
20 Staphylococcus aureus ATCC‐49444 Negative Negative
21 Streptococcus alactolyticus KCTC‐3644 Negative Negative
22 Streptococcus criceti KCTC‐3292 Negative Negative
23 Streptococcus downei KCTC‐3634 Negative Negative
24 Streptococcus ferus KCTC‐3637 Negative Negative
25 Streptococcus gallinaceus KCTC‐3876 Negative Negative
26 Streptococcus hyointestinalis KCTC‐3660 Negative Negative
27 Streptococcus intermedius KCTC‐3268 Negative Negative
28 Streptococcus lutetiensis KCTC‐3877 Negative Negative
29 Streptococcus oralis KCTC‐13048 Negative Negative
30 Streptococcus parasanguinis KCTC‐13046 Negative Negative
31 Streptococcus vestibularis KCTC‐3650 Negative Negative
32 Streptococcus anginosus KCTC‐3983 Negative Negative
33 Streptococcus macacae KCTC‐3659 Negative Negative
34 Streptococcus mitis KCTC‐5650 Negative Negative
35 Streptococcus parauberis KCTC‐3651 Negative Negative
36 Streptococcus pneumoniae KCCM‐40410 Negative Negative
37 Streptococcus pyogenes KCCM‐11817 Negative Negative
38 Streptococcus suis KCTC‐3557 Negative Negative
39 Escherichia coli KCTC‐2593 Negative Negative
40 Herpes simplex virus 1 ATCC‐VR‐260D Negative Negative
41 Herpes simplex virus 2 ATCC‐VR‐734D Negative Negative

ATCC, American type culture collection; KCCM, Korean culture center of microorganisms; KCTC, Korean collection for type cultures.

3. RESULTS

At least one microorganism was detected in each of 877 specimens (433 genital swabs and 444 urine samples) by at least one of the two assays (the DiaPlexQ or the GeneFinder assay) after discrepancy resolution. A total of 643 specimens were positive for a single microorganism, while 234 specimens were positive for multiple microorganisms (184 with two microorganisms, and 50 with ≥3 microorganisms).

A total of 885 (80.0%) and 884 (80.0%) samples were DiaPlexQ‐ and GeneFinder‐positive, respectively. The positive percent agreement between the DiaPlexQ and GeneFinder assays ranged from 96.6% to 99.4% (Table 2).The negative percent agreement for the two assays ranged from 98.2% to 99.8%. The kappa value range was 0.93‐0.99. When stratified by specimen type, the range of kappa values was 0.93‐1.00 in genital swabs and 0.93‐0.99 in urine samples.

Table 2.

Comparison of the DiaPlexQ™ STI6 Detection Kit and the GeneFinder™ STD I (CT/NG/UU) and STD II (MG/MH/TV) Multiplex Real‐time PCR Kit for the detection of STD pathogens

Virus Specimen type Percent agreement Kappa value
Positive Negative Observed kappa 95% CI
% 95% CI % 95% CI
Chlamydia trachomatis Total 98.7 (234/237) 96.0‐99.7 99.5 (865/869) 98.7‐99.9 0.98 0.96‐0.99
Genital swab 98.3 (117/119) 93.5‐99.7 99.5 (419/423) 98.1‐99.9 0.97 0.94‐0.99
Urine 99.2 (117/118) 94.7‐99.9 100 (446/446) 98.9‐100 0.99 0.98‐1.00
Neisseria gonorrhoeae Total 99.4 (162/163) 96.1‐99.9 99.0 (934/943) 98.1‐99.5 0.96 0.94‐0.99
Genital swab 100 (75/75) 93.9‐100 99.1 (463/467) 97.7‐99.7 0.97 0.94‐0.99
Urine 98.9 (87/88) 92.9‐99.9 98.9 (471/476) 97.4‐99.6 0.96 0.93‐0.99
Mycoplasma hominis Total 98.3 (229/233) 95.4‐99.4 98.7 (862/873) 97.7‐99.3 0.96 0.94‐0.98
Genital swab 99.2 (126/127) 95.0‐99.9 99.8 (407/415) 98.4‐99.9 0.95 0.93‐0.98
Urine 97.2 (103/106) 91.3‐99.3 99.3 (455/458) 97.9‐99.8 0.97 0.94‐0.99
Trichomonas vaginalis Total 99.4 (154/155) 95.9‐99.9 99.8 (949/951) 99.2‐99.9 0.99 0.98‐1.00
Genital swab 100 (79/79) 94.2‐100 100 (463/463) 99.0‐100 1.00
Urine 98.7 (75/76) 91.9‐99.9 99.6 (486/488) 98.4‐99.9 0.98 0.95‐1.00
Ureaplasma urealyticum Total 96.6 (230/238) 93.2 ‐ 98.4 98.2 (852/868) 97.0 ‐ 98.9 0.94 0.91 ‐ 0.96
Genital swab 97.4 (111/114) 91.9‐99.3 97.7 (418/428) 95.6‐98.8 0.93 0.89‐0.97
Urine 96.0 (119/124) 90.4‐98.5 98.9 (434/440) 97.2‐99.6 0.94 0.91‐0.98
Mycoplasma genitalium Total 98.7 (155/157) 95.0‐99.8 98.2 (932/949) 97.1‐98.9 0.93 0.90‐0.96
Genital swab 98.7 (76/77) 92.0‐99.9 98.3 (457/465) 96.5‐99.2 0.93 0.89‐0.98
Urine 98.8 (79/80) 92.3‐99.9 98.1 (475/484) 96.4‐99.1 0.93 0.89‐0.97

A total of 6.8% (75/1106) samples yielded discrepant results. Among those, three samples had two discordant targets; thus, 78 discrepancies were produced. Fifty‐nine results were DiaPlexQ‐positive and GeneFinder‐negative. As shown in Table 3, two were confirmed to be positive on monoplex PCR and sequencing. Likewise, among 19 DiaPlexQ‐negative and GeneFinder‐positive results, one was confirmed to be positive by the monoplex PCR and sequencing.

Table 3.

Resolution of discordant results in the DiaPlexQ™ STI6 Detection Kit and the GeneFinder™ STD I (CT/NG/UU) and STD II (MG/MH/TV) Multiplex Real‐time PCR Kit

Virus Discordant results No. of positive results on monoplex PCR and sequencing (genital swab/urine)
DiaPlexQ assay/GeneFinder assay No. of samples (genital swab/urine)
Chlamydia trachomatis ± 4 (4/0) 0 (0/0)
−/+ 3 (2/1) 0 (0/0)
Neisseria gonorrhoeae ± 9 (4/5) 1 (0/1)
−/+ 1 (0/1) 0 (0/0)
Mycoplasma hominis ± 11 (8/3) 0 (0/0)
−/+ 4 (1/3) 0 (0/0)
Trichomonas vaginalis ± 2 (0/2) 0 (0/0)
−/+ 1 (0/1) 0 (0/0)
Ureaplasma urealyticum ± 16 (10/6) 1 (1/0)
−/+ 8 (3/5) 0 (0/0)
Mycoplasma genitalium ± 17 (8/9) 0 (0/0)
−/+ 2 (1/1) 1 (1/0)

The detection limits of the DiaPlexQ assay for CT and NG were one copy/reaction, whereas those for the other microorganisms (MH, TV, UU, and MG) were 10 copies/reaction. On the analytical specificity test, no false‐positive results were obtained for any of the 41 microorganisms (Table 1).

4. DISCUSSION

Recent advances in molecular testing for the detection of STD pathogens have led to changes in the diagnosis of STIs. NAATs have increasingly replaced conventional diagnostics due to their superior sensitivity and speed.2, 4, 5, 14 Variable molecular technologies for the detection of STD pathogens from clinical specimens have been introduced including strand displacement amplification, transcription‐mediated amplification, and real‐time PCR.4, 15, 16 Some of these technologies will allow for rapid near‐patient testing at the point of care and others have potential for high‐throughput batch testing.5, 6, 7, 8, 11, 17 Rapid testing using random‐access platforms such as the GeneXpert system (Cepheid, Sunnyvale, CA) may be suitable for in‐clinic testing handling of a low number of specimens, whereas high‐throughput batch testing including the DiaPlexQ assay is useful for laboratories performing a large number of assays.12, 17, 18, 19

Numerous studies have shown that multiplex real‐time PCR assays were sensitive and useful modality for the diagnosis of STIs.5, 8, 9, 11, 13 Chung et al recently evaluated the performance of two commercial multiplex real‐time PCR assays including the GeneFinder assay for six STD pathogens and reported excellent agreement between the two assays (κ = 0.87).13 In the present study, there was excellent concordance between the results of the DiaPlexQ assay and the GeneFinder assay for each microorganism regardless of specimen type. Discordant results were observed in 3‐24 of 1106 specimens for each microorganism. The vast majority of the discordant results were associated with a high Ct value, indicating low viral loads (data not shown). Among the discordant results, the majority (54/73) were DiaPlexQ‐positive and GeneFinder‐negative. This finding may suggest the DiaPlexQ assay detected more positive samples with low viral loads, considering the analytical sensitivity of the assays: the detection limit of the DiaPlexQ assay was 1‐10 copies/reaction in the present study, whereas that of the GeneFinder assay was 50 copies/reaction according to the package insert. However, almost all discordant cases yielded negative results in discrepancy resolution with monoplex PCR and sequencing. Although false‐positive reactions in real‐time PCR assays could lead to these discrepancies, we cannot exclude the possibility that differences in analytical sensitivity between real‐time PCR assays and monoplex PCR were the cause, since real‐time PCR is generally more sensitive than conventional PCR.

The main limitation of this study was the lack of a reference standard and the retrospective nature of the study. We only resolved discrepancies for specimens with discordant results between the DiaPlexQ assay and the GeneFinder assay using monoplex PCR and sequencing. Therefore, we could not assess the diagnostic accuracy, including the sensitivity and specificity, of the real‐time PCR assays. Furthermore, we resolved the discordant results with conventional PCR and sequencing, which is generally less sensitive than real‐time PCR. In addition, it was not possible to estimate the clinical incidence of the six STIs and positive/negative predictive values due to the retrospective nature of our study.

In conclusion, the DiaPlexQ assay produces results comparable to those of the GeneFinder assay. The DiaPlexQ assay will be a useful tool for STD pathogen detection in clinical laboratories.

Supporting information

 

Huh HJ, Ki C‐S, Yun SA, et al. Comparison between DiaPlexQ™ STI6 and GeneFinder™ STD I/STD II multiplex Real‐time PCR Kits in the detection of six sexually transmitted disease pathogens. J Clin Lab Anal. 2019;33:e22703 10.1002/jcla.22703

Huh and Ki contributed equally to this article.

REFERENCES

  • 1. World Health Organization . Global health sector strategy on sexually transmitted infections, 2016–2021. https://www.who.int/reproductivehealth/publications/rtis/ghss-stis/en/ (Updated on July 2016). Accessed September 24, 2018.
  • 2. Centers for Disease Control and Prevention . Sexually transmitted diseases treatment guidelines, 2015. https://www.cdc.gov/std/tg2015/tg-2015-print.pdf (Update on June 2015). Accessed September 24, 2018.
  • 3. Unemo M, Bradshaw CS, Hocking JS, et al. Sexually transmitted infections: challenges ahead. Lancet Infect Dis. 2017;17:e235–e279. [DOI] [PubMed] [Google Scholar]
  • 4. Persing DH, Tenover FC, Hayden R, et al. Molecular microbiology: diagnostic principles and practice. Washington DC: American Society for Microbiology; 2016. [Google Scholar]
  • 5. Van Der Pol B, Williams JA, Fuller D, Taylor SN, Hook EW 3rd. Combined testing for chlamydia, gonorrhea, and trichomonas by use of the BD Max CT/GC/TV Assay with genitourinary specimen types. J Clin Microbiol. 2017;55:155‐164. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. Muvunyi CM, Dhont N, Verhelst R, et al. Evaluation of a new multiplex polymerase chain reaction assay STDFinder for the simultaneous detection of 7 sexually transmitted disease pathogens. Diagn Microbiol Infect Dis. 2011;71:29‐37. [DOI] [PubMed] [Google Scholar]
  • 7. Dize L, West SK, Mkocha H, Quinn TC, Gaydos CA. Evaluation of pooled ocular and vaginal swabs by the Cepheid GeneXpert CT/NG assay for the detection of Chlamydia trachomatis and Neisseria gonorrhoeae compared to the GenProbe Aptima Combo 2 Assay. Diagn Microbiol Infect Dis. 2015;81:102‐104. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8. Rumyantseva T, Golparian D, Nilsson CS, et al. Evaluation of the new AmpliSens multiplex real‐time PCR assay for simultaneous detection of Neisseria gonorrhoeae, Chlamydia trachomatis, Mycoplasma genitalium, and Trichomonas vaginalis . Apmis. 2015;123:879‐886. [DOI] [PubMed] [Google Scholar]
  • 9. Ursi D, Crucitti T, Smet H, Ieven M. Evaluation of the Bio‐Rad Dx CT/NG/MG® assay for simultaneous detection of Chlamydia trachomatis, Neisseria gonorrhoeae and Mycoplasma genitalium in urine. Eur J Clin Microbiol Infect Dis. 2016;35:1159‐1163. [DOI] [PubMed] [Google Scholar]
  • 10. Peters R, de Vos L, Maduna L, et al. Laboratory validation of Xpert Chlamydia trachomatis/Neisseria gonorrhoeae and Trichomonas vaginalis testing as performed by nurses at three primary health care facilities in South Africa. J Clin Microbiol. 2017;55:3563‐3565. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11. Choe HS, Lee DS, Lee SJ, et al. Performance of Anyplex II multiplex real‐time PCR for the diagnosis of seven sexually transmitted infections: comparison with currently available methods. Int J Infect Dis. 2013;17:e1134–e1140. [DOI] [PubMed] [Google Scholar]
  • 12. Gaydos CA. Review of use of a new rapid real‐time PCR, the Cepheid GeneXpert® (Xpert) CT/NG assay, for Chlamydia trachomatis and Neisseria gonorrhoeae: results for patients while in a clinical setting. Expert Rev Mol Diagn. 2014;14:135‐137. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Chung HS, Lee M. Comparative evaluation of multiplex real‐time PCR assays for six pathogens of sexually transmitted infections. Ann Clin Microbiol. 2017;20:1‐6. [Google Scholar]
  • 14. Golparian D, Borang S, Sundqvist M, Unemo M. Evaluation of the new BD max GC Real‐Time PCR assay, analytically and clinically as a supplementary test for the BD ProbeTec GC Qx Amplified DNA assay, for molecular detection of Neisseria gonorrhoeae . J Clin Microbiol. 2015;53:3935‐3937. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15. Little MC, Andrews J, Moore R, et al. Strand displacement amplification and homogeneous real‐time detection incorporated in a second‐generation DNA probe system. BDProbeTecET. Clin Chem. 1999;45:777‐784. [PubMed] [Google Scholar]
  • 16. Chernesky M, Jang D, Gilchrist J, et al. Head‐to‐head comparison of second‐generation nucleic acid amplification tests for detection of Chlamydia trachomatis and Neisseria gonorrhoeae on urine samples from female subjects and self‐collected vaginal swabs. J Clin Microbiol. 2014;52:2305‐2310. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17. Dize L, West S, Williams JA, Van Der Pol B, Quinn TC, Gaydos CA. Comparison of the Abbott m2000 RealTime CT assay and the Cepheid GeneXpert CT/NG assay to the Roche Amplicor CT assay for detection of Chlamydia trachomatis in ocular samples from Tanzania. J Clin Microbiol. 2013;51:1611‐1613. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. Schwebke JR, Gaydos CA, Davis T, et al. Clinical evaluation of the Cepheid Xpert TV Assay for detection of Trichomonas vaginalis with prospectively collected specimens from men and women. J Clin Microbiol. 2018;56. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19. Gaydos CA, Van Der Pol B, Jett‐Goheen M, et al. Performance of the Cepheid CT/NG Xpert rapid PCR test for detection of Chlamydia trachomatis and Neisseria gonorrhoeae . J Clin Microbiol. 2013;51:1666‐1672. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

 


Articles from Journal of Clinical Laboratory Analysis are provided here courtesy of Wiley

RESOURCES