Table 1.

Comparison of different postprocessing techniques in the context of CW-STED microscopy.

  CW-STED gCW-STED gCW-STED filt. MID CW-STED SPLIT CW-STED
Principle PSF shrinking by stimulated emission depletion PSF shrinking by stimulated emission depletion and time-gating detection PSF shrinking by stimulated emission depletion and time-gating detection; filtering for background removal PSF shrinking by stimulated emission depletion and MID (background is taken into account and removed) PSF shrinking by stimulated emission depletion and SPLIT (background is estimated and removed)
Lateral resolution 20 to 50 nm3 42  nm17 60  nm20 60  nm35 60  nm24
Speed High High High Middle Middle
Power required for the sample (mW) 600 200 50 40 40
Advantages Simplicity and cost Enhanced on–off contrast using moderate light intensities Enhanced on–off contrast using moderate light intensities Enhanced SNR Enhanced SNR
Drawbacks High photobleaching Low SNR may negate benefit Low SNR may negate benefit Complex data processing with possible artifacts Complex data processing with possible artifacts