Cardiovascular Diagnosis and Therapy logoLink to Cardiovascular Diagnosis and Therapy
. 2019 Oct;9(5):481–491. doi: 10.21037/cdt.2019.09.07

Accuracy of 3-dimensional and 2-dimensional quantitative coronary angiography for predicting physiological significance of coronary stenosis: a FAVOR II substudy

Daixin Ding 1,2,#, Junqing Yang 3,#, Jelmer Westra 4, Yundai Chen 5, Yunxiao Chang 1,2, Martin Sejr-Hansen 4, Su Zhang 1,2, Evald H Christiansen 4, Niels R Holm 4, Bo Xu 6,, Shengxian Tu 1,2,
PMCID: PMC6837915  PMID: 31737519

Abstract

Background

Three-dimensional quantitative coronary angiography (3D-QCA) enables reconstruction of a coronary artery in 3D from two angiographic image projections. This study compared the diagnostic accuracy of 3D-QCA vs. 2-dimensional (2D) QCA in predicting physiologically significant coronary stenosis, using fractional flow reserve (FFR) as the reference standard.

Methods

All interrogated vessels in the FAVOR II China study and the FAVOR II Europe-Japan study were assessed by 2D-QCA and 3D-QCA according to standard operating procedures in core laboratories. QCA analysts were blinded to the corresponding FFR values.

Results

A total of 645 vessels from 576 patients with 3D-QCA, 2D-QCA, and FFR were analyzed. Using the conventional cut-off value of 50% for percent diameter stenosis (DS%), 3D-QCA was more accurate in predicting FFR ≤0.80 than 2D-QCA [accuracy 74.0% (95% CI: 69.9–77.7%) vs. 64.9% (95% CI: 61.3–68.7%), difference: 9.1%, P<0.001]. Sensitivity was higher by 3D-QCA compared with 2D-QCA [69.1% (95% CI: 63.0–75.1%) vs. 47.1% (95% CI: 40.5–53.6%), difference: 22.0%, P<0.001] and specificity was similar [76.5% (95% CI: 72.5–80.6%) vs. 74.4% (95% CI: 70.2–78.6%), difference: 2.1%, P=0.40]. Area under the receiver operating characteristic curve was significantly higher for 3D-QCA than for 2D-QCA [0.81 (95% CI: 0.77–0.84) vs. 0.66 (95% CI: 0.62–0.71), P<0.001].

Conclusions

3D-QCA demonstrated better diagnostic performance in predicting physiologically significant coronary stenosis compared with 2D-QCA, when FFR was used as the reference standard.

Keywords: Coronary physiology, fractional flow reserve (FFR), ischemia, quantitative coronary angiography (QCA)

Introduction

The degree and extent of a luminal narrowing is frequently used to evaluate the severity of coronary artery stenosis in patients with coronary artery disease (CAD). Coronary artery lesions with ≥50% percent diameter stenosis (DS%) by quantitative coronary angiography (QCA) are traditionally considered to be significant based on study demonstrating that hyperemic flow starts to decrease with presence of at least 50% reduction in vessel diameter (1). However, novel diagnostic tools such as fractional flow reserve (FFR) have challenged the use of visual estimation or QCA-based evaluation of coronary stenosis (2-4).

Dedicated 3-dimensional (3D) angiographic reconstruction algorithms were developed to overcome some of the inherited limitations in 2D-QCA analysis such as foreshortening and out-of-plane magnification (5,6). Excellent correlation was reported between 3D-QCA and intravascular imaging including intravascular ultrasound (IVUS) (7) and optical coherence tomography (OCT) (5). Compared with 2D-QCA, 3D-QCA reported better diagnostic performance for predicting physiologically significant coronary stenosis when FFR was used as the reference standard (7-9). Nevertheless, these studies had a relatively small population size and were limited by their retrospective design. The aim of the present study was to compare the diagnostic accuracy of 3D- and 2D-QCA in a large and prospectively enrolled international population, using FFR as the reference standard.

Methods

This study reports an ad hoc subanalysis of the FAVOR II China (Functional Diagnostic Accuracy of Quantitative Flow Ratio in Online Assessment of Coronary Stenosis, NCT03191708) and FAVOR II Europe-Japan (Functional Assessment by Various Flow Reconstructions, NCT02959814) studies; two prospective, multicenter studies enrolling a total of 637 patients at 16 centers in the Europe and Asia. Both studies were designed to validate the in-procedure diagnostic accuracy of angiography-based quantitative flow ratio (QFR) (10,11). The study design, main endpoints and inclusion and exclusion criteria were reported elsewhere (10,11). Collection of patient outcome data was not planned in the trial designs.

All FFR traces and angiographic image data were analyzed in core laboratories (CCRF, Beijing, China, CardHemo, Med-X research Institute, Shanghai Jiao Tong University, Shanghai, China, and the Institute of Clinical Medicine, Aarhus University, Denmark) in a blinded fashion. 3D-QCA, 2D-QCA and FFR data were available for paired analysis in 645 vessels. The study protocol complies with the Declaration of Helsinki and was approved by the institutional review board (FAVOR II China study is 2017-861, for FAVOR II Europe/Japan study is 1-10-72-219-16). All patients provided written informed consent.

FFR, 2D-QCA and 3D-QCA analysis

FFR reading was performed in blinded fashion in the core laboratories using RadiView (St. Jude Medical) as specified in the main study (10,11). The angiographic view with better exposure of the stenosis severity was used for 2D-QCA analysis, using the QAngio XA software package (Medis Medical Imaging Systems, Leiden, The Netherlands). Catheter calibration or isocenter calibration was used.

3D-QCA analysis was performed based on two angiographic image views with ≥25° difference using AngioPlus (Pulse Medical Imaging Technology, Shanghai, China) or QAngio XA 3D (Medis Medical Imaging Systems, Leiden, The Netherlands) that both used the same algorithms for 3D angiographic reconstruction. Automated calibration was used for 3D angiographic reconstruction. For each projection, lumen contours were delineated using the same vessel edge detection algorithm as used in the 2D-QCA analysis. Based on the delineated lumen contours, coronary lumen and reference vessel were reconstructed and anatomical parameters were quantified in 3D. All the QCA measurements were performed by experienced users following standard operation procedures (10-12).

Statistical analysis

Continuous variables are reported as mean ± SD, and categorical variables are presented as counts (%). Normality of the data was tested using the Shapiro-Wilk and Q-Q plots. Bland-Altman plots were used to identify agreement between 2D- and 3D-QCA DS%. Pearson or Spearman correlation test was used to determine the correlation among variables. Pairwise comparisons between 2D- and 3D-QCA DS% were made by Student t-test, and Wilcoxon’s rank test was used for comparisons between 2D- and 3D-QCA lesion length, minimum luminal diameter (MLD) and reference diameter. Diagnostic performance estimates by 2D- and 3D-QCA were compared using McNemar’s test, generalized score statistics and density based logistic regression, as appropriate. Comparisons for accuracy, sensitivity and specificity among groups stratified by vessel were performed using chi-squared test. Area under the receiver-operating characteristic (ROC) curves (AUC) for QCA DS% to predict FFR ≤0.80 were compared using the Hanley and McNeil method. All statistical analyses were performed using the STATA 15 (StataCorp LLC, Texas, USA) and R software (R Foundation for Statistical Computing, Vienna, Austria). A two-sided P value of <0.05 was considered statistically significant.

Results

A total of 645 vessels from 576 patients with paired 2D-QCA, 3D-QCA and FFR data were available for analysis: 353 (55%) left anterior descending arteries (LAD), 131 (20%) left circumflex arteries (LCx) and 161 (25%) right coronary arteries (RCA). Median FFR was 0.85 [interquartile range (IQR): 0.77–0.90] in the overall population. A total of 223 (35%) vessels were hemodynamically significant as defined by FFR ≤0.80. Baseline patient characteristics are listed in Table S1.

Table S1. Baseline characteristics of the study population.

Characteristics Value (n=576)
Age, years 63.8±10.7
Male 73.1 [421]
BMI, kg/m2 26.1±4.1 (n=569*)
Current smoker 24.5 [141]
Hypertension 66.7 [384]
Hyperlipidemia 56.4 [325]
Diabetes mellitus 28.3 [163]
Family history of CAD 21.5 [124]
Previous PCI 30.0 [173]
Previous CABG 2.1 [12]
Clinical presentation
   Silent ischemia 3.3 [19]
   Stable angina pectoris 15.1 [87]
   Unstable angina pectoris 45.1 [260]
   Acute myocardial infarction 35.1 [202]
   Other 1.4 [8]

Values are presented as mean ± SD or % [n]. *, number of patients for whom continuous variables were calculated. BMI, body mass index; CABG, coronary artery bypass surgery; CAD, coronary artery disease; PCI, percutaneous coronary interventions.

Correlation and agreement between 2D- and 3D-QCA measurement

Baseline lesion characteristics by 2D- and 3D-QCA were shown in Table 1. Compared with 2D-QCA, 3D-QCA showed higher DS%, longer lesions, lower MLD but similar reference diameter in the overall population. Correlation was good between 2D-QCA DS% and 3D-QCA DS% (0.59, P<0.001) with modest scatter around the regression line (Figure 1A). The mean difference was 2.7%±9.3% (Figure 1B).

Table 1. Baseline lesion characteristics by 2-dimensional and 3-dimensional quantitative coronary angiography.

Index Stratification 2D-QCA 3D-QCA P value*
DS% Overall 45.7±10.8 48.4±9.6 <0.001
LAD 44.2±10.4 47.6±8.7 <0.001
RCA 45.8±9.9 48.5±9.4 <0.001
LCx 49.6±11.9 50.5±11.6 0.32
Lesion length (mm) Overall 12.4±6.5 23.2±13.4 <0.001
LAD 12.7±6.4 24.5±12.1 <0.001
RCA 13.1±7.3 24.6±15.4 <0.001
LCx 10.6±5.5 17.7±10.2 <0.001
MLD (mm) Overall 1.55±0.44 1.47±0.40 <0.001
LAD 1.55±0.41 1.47±0.36 <0.001
RCA 1.71±0.45 1.59±0.43 <0.001
LCx 1.37±0.44 1.30±0.43 0.01
Reference diameter (mm) Overall 2.84±0.55 2.84±0.50 0.13
LAD 2.76±0.51 2.82±0.44 <0.001
RCA 3.12±0.57 3.07±0.53 0.78
LCx 2.68±0.51 2.62±0.49 0.03

All variables are presented by mean ± SD. *, DS% by 3D-QCA and 2D-QCA was compared by paired Student’s t-test; lesion length, MLD and reference diameter by 3D-QCA and 2D-QCA were compared by Wilcoxon’s rank test. 2D-QCA, 2-dimensional quantitative coronary angiography; 3D-QCA, 3-dimensional quantitative coronary angiography; DS%, percent diameter stenosis; LAD, left anterior descending artery; LCx, left circumflex; MLD, minimum luminal diameter; RCA, right coronary artery.

Figure 1.

Figure 1

Correlation (A) and agreement (B) between 2D-QCA and 3D-QCA for DS%. 2D-QCA, 2-dimensional quantitative coronary angiography; 3D-QCA, 3-dimensional quantitative coronary angiography; DS%, percent diameter stenosis; SD, standard deviation.

Correlation between FFR and QCA measurements

3D-QCA correlated better than 2D-QCA with FFR (−0.56 vs. −0.32, P<0.001) (Figure 2). Subgroup analysis showed that the difference was statistically significant in LAD (−0.59 vs. −0.38, P<0.001), RCA (−0.66 vs. −0.37, P<0.001), and LCx (−0.66 vs. −0.43, P=0.01).

Figure 2.

Figure 2

Correlation between DS% versus FFR for 2D-QCA (A) and 3D-QCA (B). 2D-QCA, 2-dimensional quantitative coronary angiography; 3D-QCA, 3-dimensional quantitative coronary angiography; DS%, percent diameter stenosis; FFR, fractional flow reserve.

Performance of 2D- and 3D-QCA in predicting physiologically significant stenosis

Applying a discrimination limit of DS% ≥50% to predict FFR ≤0.80 resulted in significantly better diagnostic accuracy (74.0% vs. 64.9%, P<0.001) and sensitivity (69.1% vs. 47.1%, P<0.001) for 3D-QCA compared with 2D-QCA (Tables 2,S2,S3). Using FFR as the reference standard, 3D-QCA DS% had a higher AUC compared with 2D-QCA DS% [0.81 (95% CI: 0.77–0.84) vs. 0.66 (95% CI: 0.62–0.71), P<0.001] (Figure 3).

Table 2. Diagnostic performance of 50% diameter stenosis by 2-dimensional and 3-dimensional quantitative coronary angiography for predicting fractional flow reserve ≤0.80.

Index 3D-QCA 2D-QCA Difference P value
Accuracy (95% CI) (%) 74.0 (69.9–77.7) 64.9 (61.3–68.7) 9.1 <0.001
Sensitivity (95% CI) (%) 69.1 (63.0–75.1) 47.1 (40.5–53.6) 22.0 <0.001
Specificity (95% CI) (%) 76.5 (72.5–80.6) 74.4 (70.2–78.6) 2.1 0.40
LR(+) (95% CI) 2.94 (2.43–3.57) 1.84 (1.49–2.28) 0.63* <0.001
LR(−) (95% CI) 0.40 (0.33–0.50) 0.71 (0.62–0.81) 1.78* <0.001
PPV% (95% CI) 60.9 (55.0–66.9) 49.3 (42.6–56.0) 11.8 <0.001
NPV% (95% CI) 82.4 (78.6–86.2) 72.7 (68.5–76.9) 9.2 <0.001

*, ratio. Accuracy, sensitivity and specificity were compared by McNemar’s test, PPV and NPV were compared by generalized score statistics, and LR(+) and LR(−) were compared using DLR regression model. 2D-QCA, 2-dimensional quantitative coronary angiography; 3D-QCA, 3-dimensional quantitative coronary angiography; LR(+), positive likelihood ratio; LR(−), negative likelihood ratio; NPV, negative predictive value; PPV, positive predictive value.

Table S2. Diagnostic accuracy of 2D-QCA in predicting FFR ≤0.80.

2D-QCA DS% FFR Total
≤0.80 >0.80
≥50% 105 108 213
<50% 118 314 432
Total 223 422 645

2*2 table for DS% vs. FFR for 2D-QCA. 2D-QCA, 2-dimensional quantitative coronary angiography; DS%, percent diameter stenosis; FFR, fractional flow reserve.

Table S3. Diagnostic accuracy of 3D-QCA in predicting FFR ≤0.80.

3D-QCA DS% FFR Total
≤0.80 >0.80
≥50% 154 99 253
<50% 69 323 392
Total 223 422 645

2*2 table for DS% vs. FFR for 3D-QCA. 3D-QCA, 3-dimensional quantitative coronary angiography; DS%, percent diameter stenosis; FFR, fractional flow reserve.

Figure 3.

Figure 3

Receiver operating characteristic (ROC) curves of 2D-QCA and 3D-QCA to predict FFR ≤0.80. 2D-QCA, 2-dimensional quantitative coronary angiography; 3D-QCA, 3-dimensional quantitative coronary angiography; DS%, percent diameter stenosis; AUC, area under the receiver operating characteristic curve.

The relationship between FFR and DS% category is shown in Figure 4. Lesions were more likely to become physiologically significant with increasing DS%. Nevertheless, misclassification between physiological significance and anatomical obstruction occurred more with 2D-QCA than 3D-QCA. FFR indicated physiological significance in 21.6% of lesions with <40% DS by 2D-QCA. The false negative ratio decreased to 5.6% when DS% was evaluated by 3D-QCA. Likewise, in 27.3% of lesions with ≥70% DS by 2D-QCA, FFR showed physiological insignificance. This false positive ratio decreased to 6.7% when DS% was evaluated by 3D-QCA.

Figure 4.

Figure 4

Scatter plot showing relationship between angiographic stenosis grade and FFR for all vessels by 3D-QCA (A) and 2D-QCA (B). The dotted line represents FFR =0.80. Numbers in each category: (A) <40%, n=108; 40% to 49.9%, n=284; 50% to 59.9%, n=178; 60% to 69.9%, n=60; ≥70%, n=15. In lesions graded <40% DS, FFR ≤0.80 was present in 5.6%, thus in 5.6% of stenoses <40% DS, FFR indicated physiologically significant lesion. In lesions graded ≥70% DS, FFR ≤0.80 was present in 93.3%, thus in 6.7% of stenoses ≥70% DS, FFR indicated physiologically non-significant lesion. In lesions graded between 40% to 69.9% DS, FFR ≤0.80 was present in 38.9%. (B) <40%, n=194; 40% to 49.9%, n=238; 50% to 59.9%, n=144; 60% to 69.9%, n=58; ≥70%, n=11. In lesions graded <40% DS, FFR ≤0.80 was present in 21.6%, thus in 21.6% of stenoses <40% DS, FFR indicated physiologically significant lesion. In lesions graded ≥70% DS, FFR ≤0.80 was present in 72.7%, thus in 27.3% of stenoses ≥70% DS, FFR indicated physiologically non-significant lesion. In lesions graded between 40% to 69.9% DS, FFR ≤0.80 was present in 39.3%. 2D-QCA, 2-dimensional quantitative coronary angiography; 3D-QCA, 3-dimensional quantitative coronary angiography; DS%, percent diameter stenosis; FFR, fractional flow reserve.

Influence of lesion location on diagnostic performance

Median FFR was significantly lower in LAD lesions [0.81 (IQR: 0.75–0.87)] than in RCA lesions [0.88 (IQR: 0.83–0.94), P<0.001] or in LCx lesions [0.88 (IQR: 0.82–0.93), P<0.001]. For both 2D-QCA DS% and 3D-QCA DS%, significant variations existed in sensitivity and specificity among lesions in LAD, LCx and RCA (Table 3). The optimal cut-off values for DS% in predicting FFR ≤0.80 were 50% for 3D-QCA and 48% for 2D-QCA (Figure 5). Marked variations existed in optimal cut-off values and the associated diagnostic performance estimates after stratification by lesion location (Figure 5). Applying vessel-specific optimal cut-off values in DS% for both 3D-QCA (cut-off values: 49% for LAD, 56% for LCx and 52% for RCA) and 2D-QCA (cut-off values: 45% for LAD, 52% for LCx and 48% for RCA), the superiority of 3D-QCA over 2D-QCA remained in diagnostic accuracy (74.3% vs. 65.1%, P<0.001), sensitivity (73.1% vs. 56.1%, P<0.001) and specificity (74.9% vs. 69.9%, P=0.04).

Table 3. Stratified analysis of the diagnostic performance of 50% diameter stenosis by 2-dimensional and 3-dimensional quantitative coronary angiography for predicting fractional flow reserve ≤0.80 according to different lesion locations.

Index LAD LCx RCA P value
2D-QCA 50% DS
   Accuracy 61.8 (56.7–66.9) 66.4 (58.2–74.6) 70.8 (63.7–77.9) 0.13
   Sensitivity 39.6 (32.0–47.7) 75.0 (55.1–89.3) 58.3 (40.8–74.5) <0.001
   Specificity 79.9 (73.6–85.3) 64.1 (54.0–73.3) 74.7 (65.8–81.8) 0.01
3D-QCA 50% DS
   Accuracy 74.5 (70.0–79.1) 69.4 (61.5–77.5) 76.4 (69.8–83.0) 0.38
   Sensitivity 61.6 (53.6–69.2) 85.7 (67.3–96.0) 88.9 (73.9–96.9) 0.004
   Specificity 85.1 (79.2–89.8) 65.0 (55.0–74.2) 72.8 (64.1–80.4) <0.001

Values are n% (95% CI). Comparisons for accuracy, sensitivity and specificity among lesion groups of LAD, LCx and RCA were performed by chi-squared test. 2D-QCA, 2-dimensional quantitative coronary angiography; 3D-QCA, 3-dimensional quantitative coronary angiography; DS%, percent diameter stenosis; LAD, left anterior descending artery; LCx, left circumflex; RCA, right coronary artery.

Figure 5.

Figure 5

Comparisons of optimal cut-off values of DS% by 2D-QCA (A) and 3D-QCA (B) among lesions in the overall population, left anterior descending artery, left circumflex and right coronary artery. 2D-QCA, 2-dimensional quantitative coronary angiography; 3D-QCA, 3-dimensional quantitative coronary angiography; DS%, percent diameter stenosis.

Discussion

This is the largest study to compare the diagnostic accuracy of 2D-QCA and 3D-QCA for the physiological assessment of coronary stenosis in prospectively enrolled patient populations. We observed that (I) 3D-QCA had better diagnostic accuracy than 2D-QCA, with FFR as the reference standard; (II) when evaluated by 2D-QCA, 21.6% of coronary lesions with DS <40% were physiologically significant, and 27.3% with diameter stenosis ≥70% were physiologically non-significant. The diagnostic discordance with FFR was reduced when stenosis <40% and ≥70% DS were evaluated by 3D-QCA; (III) the optimal cut-off value of DS% by 3D-QCA to predict physiologically significant coronary stenosis is 50%; (IV) diagnostic performance and optimal diagnostic cut-off value varies in different vessels, and the improvement in diagnostic performance of 3D-QCA over 2D-QCA is greatest in LAD lesions.

The accuracy of 2D-QCA in the present study population is moderate due to a number of well-known limitations in 2D analysis, including vessel foreshortening, out-of-plane magnification, and imprecise representation of lumen cross-sections for eccentric lesions (13). Our findings are in line with previous studies showing limited diagnostic performance of 2D-QCA indices in predicting physiological significance of coronary stenosis (2,14). By reconstruction of lumen geometry from two angiographic projections, 3D-QCA partially resolved some of these limitations (13,15,16). Of note, Yong et al. previously compared the diagnostic accuracy of 3D-QCA vs. 2D-QCA (8) in predicting FFR and found that MLA by 3D-QCA had higher diagnostic accuracy than by 2D-QCA. However, DS% by 3D-QCA had much lower diagnostic accuracy in predicting FFR ≤0.80 than by 2D-QCA (AUC =0.63 vs. 0.74). On the contrary, we found that DS% by 3D-QCA had substantially higher accuracy in predicting FFR ≤0.80 than by 2D-QCA (AUC =0.81 vs. 0.66). The difference might be explained by the difference in the study design and the enrolled patient populations. While the study by Yong et al. was a retrospective study with a population of only 63 patients, the present study analyzed 576 unselected patients with indication for coronary physiological assessment who were prospectively enrolled in 16 international centers. The prospective study design and large patient population may also partly contribute to the result that the diagnostic accuracy of 3D-QCA in our study tends to be higher than in other studies which investigate its diagnostic performance (9). Using this large patient population, we also showed that the superiority in diagnostic accuracy by 3D-QCA was maintained in the three major epicardial coronary arteries. In addition, we demonstrated for the first time that 3D-QCA on average yielded a smaller MLD but similar reference vessel size compared with 2D-QCA. This resulted in a higher diameter stenosis by 3D-QCA than by 2D-QCA. Indeed, 2D-QCA allows assessment of lesion severity in a single projection that might not fully expose the lesion severity. On the contrary, 3D-QCA reconstructed the coronary lumen in 3D and hence increased the chance of exposing the smaller lumen diameter. This improved assessment of the narrowest cross-sections may have contributed to improved accuracy in predicting physiologically significant coronary stenosis.

Although previous studies demonstrated discrepancies between FFR and angiographic diameter stenosis (2,4,7,8), in daily practice, decisions about revascularization are mainly based on angiographic DS% in combination with other factors related to ischemia (17). Solely considering angiographic DS%, it is generally perceived by many interventional cardiologists that lesions of <40% DS are unlikely to reduce hyperemic blood flow, while lesions of >70% DS are believed to cause ischemia. The recent RIPCORD (4) study found that 13% of the lesions with <30% DS by visual assessment had an FFR ≤0.80, while 47% of the lesions with >70% DS had an FFR >0.80 (4). The present study was based on data from prospectively enrolled consecutive patients with coronary lesions between 30% and 90% DS. The diagnostic performance was improved by 3D-QCA compared to 2D-QCA. The increased rates of false negative predictions by 2D-QCA could be explained by several reasons mainly related to diffused atherosclerosis and proximal lesion subtending large myocardial territories in which blood flow can be limited even without critical stenosis by angiography (18). The lower false negative ratio in the RIPCORD study compared with that of 2D-QCA in our study is in line with some previous studies showing that visual assessment of angiography had a better diagnostic performance than 2D-QCA in predicting FFR ≤0.80 because physicians take additional clinical factors related to ischemia into account (17). The false positive ratio was much lower by 2D-QCA in this study than in the RIPCORD study, with that of 3D-QCA being even lower. This gap may be explained by the tendency towards overestimation of stenosis severity by physician’s visual assessment than by QCA (14,17,19,20). Of note, the false negative ratio and false positive ratio were much smaller for 3D-QCA than for 2D-QCA, again suggesting the superiority of 3D-QCA in identifying physiological significance of coronary stenosis.

Revascularization of physiologically non-significant stenosis introduces procedural risk without offering comparable benefits (3), while deferral of physiologically significant stenosis can result in detrimental clinical consequences (21). The discrepancies between FFR and QCA reveal the theoretical differences between these two techniques. While angiography depicts the degree and extent of epicardial luminal narrowing, FFR aims to reveal to what extent coronary flow is limited by an epicardial stenosis (22). Reliability of FFR may be limited as maximum hyperemia may be difficult to achieve due to complex patient-specific physiological conditions such as high venous pressure and the presence of microvascular dysfunction (23). There are situations in which different coronary flow patterns with identical anatomical components are not differentiated by QCA (24). Technically, both FFR and QCA have inherent limitations. The quality of FFR measurement may be hampered by several pitfalls (e.g., dampening, drift, waveform distortion) (25) while QCA is vulnerable to bias related to lumen border detection, especially for severe lesions. Therefore, one-to-one correspondence between FFR and QCA values is not expected. And this precludes the use of 2D-QCA or 3D-QCA alone to guide revascularization decision-making. Nevertheless, it is promising that 3D-QCA appears to identify physiologically significant coronary stenosis better than 2D-QCA.

In this study, the overall optimal cut-off value for 3D-QCA DS% was equal to the traditional cut-off value of 50% that thus appears appropriate. This is in accordance with a recent study by Toth et al. in which the overall diagnostic performance of QCA DS% decreased when adjusting the cut-off value from 50% to 70% (2). However, our subgroup analysis found that the optimal cut-off value was lowest for LAD lesions while highest for LCx lesions by both 2D- and 3D-QCA. Previous studies have found that FFR tends to be lower in segments subtending larger myocardial mass than in segments subtending smaller myocardium territories (2,22,26). This is further confirmed by our study where median FFR in LAD lesions was lower compared to non-LAD lesions, and may explain the low sensitivity and high specificity for 50% DS in LAD vs. non-LAD lesions. We also found that for both 2D- and 3D-QCA, no statistically significant differences exist in the diagnostic accuracy in predicting FFR ≤0.80 among lesions in LAD, LCx and RCA, despite significant variations in sensitivity and specificity. This indicates that although there may be dissimilarities in the physiological environment combined with variations in vessel length and bifurcation conditions in individual vessels (27), the overall discriminative power for physiologically significant lesions of QCA remains stable in major branch subgroups. Thus, vessel-specific cut-off values for QCA-based endpoints in clinical studies may exert limited benefits. This is further confirmed by our study findings: using vessel-specific cut-off values instead of the same cut-off value in DS% led to higher sensitivity (73.1% vs. 69.1%) but lower specificity (74.9% vs. 76.5%) for 3D-QCA in predicting physiologically significant stenosis. However, the overall diagnostic accuracy was similar (74.3% vs. 74.0%).

The anatomy derived from 3D-QCA may be combined with information of blood flow for predicting physiologically significant stenosis. QFR, a novel approach for fast computation of virtual FFR based on 3D-QCA and thrombolysis in myocardial infarction (TIMI) frame count, has been validated against FFR with good diagnostic accuracy in several studies (10,11,28,29). Similar concepts were also tested in other angiography-based FFR methods (30,31).

This study is not without limitation. Selection variation may exist in the 2D-QCA procedure. For every patient, the angiographic image for 2D-QCA analysis was chosen by operator’s preference from two angiographic image runs used for 3D-QCA analysis. Nevertheless, the chosen angiographic image runs were with minimum overlap or foreshortening, best image quality and exposure of target vessels, thus reducing the selection variation for 2D-QCA to minimum.

Conclusions

3D-QCA analysis demonstrated higher diagnostic accuracy in predicting physiologically significant coronary stenosis than 2D-QCA analysis. When FFR is not available, 3D-QCA may be used to assist evaluation of obstructive coronary stenosis.

Acknowledgments

Funding: This work was supported by the Natural Science Foundation of China (Grant Number 81871460); Shanghai Jiao Tong University (Grant Number YG2016ZD09); and by the Program of Shanghai Technology Research Leader.

Ethical Statement: The authors are accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved. The study protocol complies with the Declaration of Helsinki and was approved by the institutional review board (FAVOR II China study is 2017-861, for FAVOR II Europe/Japan study is 1-10-72-219-16). All patients provided written informed consent.

Footnotes

Conflicts of Interest: S Tu received a research grant from Medis medical imaging and Pulse medical imaging technology. N Holm received research grants from Medis medical imaging, Abbot, and Boston Scientific. The other authors have no conflicts of interest to declare.

References

  • 1.Gould KL, Lipscomb K, Hamilton GW. Physiologic basis for assessing critical coronary stenosis. Instantaneous flow response and regional distribution during coronary hyperemia as measures of coronary flow reserve. Am J Cardiol 1974;33:87-94. 10.1016/0002-9149(74)90743-7 [DOI] [PubMed] [Google Scholar]
  • 2.Toth G, Hamilos M, Pyxaras S, et al. Evolving concepts of angiogram: fractional flow reserve discordances in 4000 coronary stenoses. Eur Heart J 2014;35:2831-8. 10.1093/eurheartj/ehu094 [DOI] [PubMed] [Google Scholar]
  • 3.Tonino PA, De Bruyne B, Pijls NH, et al. Fractional flow reserve versus angiography for guiding percutaneous coronary intervention. N Engl J Med 2009;360:213-24. 10.1056/NEJMoa0807611 [DOI] [PubMed] [Google Scholar]
  • 4.Curzen N, Rana O, Nicholas Z, et al. Does routine pressure wire assessment influence management strategy at coronary angiography for diagnosis of chest pain? The RIPCORD study. Circ Cardiovasc Interv 2014;7:248-55. 10.1161/CIRCINTERVENTIONS.113.000978 [DOI] [PubMed] [Google Scholar]
  • 5.Tu S, Xu L, Ligthart J, et al. In vivo comparison of arterial lumen dimensions assessed by co-registered three-dimensional (3D) quantitative coronary angiography, intravascular ultrasound and optical coherence tomography. Int J Cardiovasc Imaging 2012;28:1315-27. 10.1007/s10554-012-0016-6 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Tu S, Echavarria-Pinto M, von Birgelen C, et al. Fractional flow reserve and coronary bifurcation anatomy: a novel quantitative model to assess and report the stenosis severity of bifurcation lesions. JACC Cardiovasc Interv 2015;8:564-74. 10.1016/j.jcin.2014.12.232 [DOI] [PubMed] [Google Scholar]
  • 7.Nishi T, Kitahara H, Fujimoto Y, et al. Comparison of 3-dimensional and 2-dimensional quantitative coronary angiography and intravascular ultrasound for functional assessment of coronary lesions. J Cardiol 2017;69:280-6. 10.1016/j.jjcc.2016.05.006 [DOI] [PubMed] [Google Scholar]
  • 8.Yong AS, Ng AC, Brieger D, et al. Three-dimensional and two-dimensional quantitative coronary angiography, and their prediction of reduced fractional flow reserve. Eur Heart J 2011;32:345-53. 10.1093/eurheartj/ehq259 [DOI] [PubMed] [Google Scholar]
  • 9.Chu M, Dai N, Yang J, et al. A systematic review of imaging anatomy in predicting functional significance of coronary stenoses determined by fractional flow reserve. Int J Cardiovasc Imaging 2017;33:1-16. 10.1007/s10554-017-1085-3 [DOI] [PubMed] [Google Scholar]
  • 10.Westra J, Andersen BK, Campo G, et al. Diagnostic performance of in-procedure angiography-derived quantitative flow reserve compared to pressure-derived fractional flow reserve: the FAVOR II Europe/Japan Study. J Am Heart Assoc. doi: . 10.1161/JAHA.118.009603 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Xu B, Tu S, Qiao S, et al. Diagnostic accuracy of angiography-based quantitative flow ratio measurements for online assessment of coronary stenosis. J Am Coll Cardiol 2017;70:3077-87. 10.1016/j.jacc.2017.10.035 [DOI] [PubMed] [Google Scholar]
  • 12.Tuinenburg JC, Koning G, Hekking E, et al. One core laboratory at two international sites, is that feasible? An inter-core laboratory and intra-observer variability study. Catheter Cardiovasc Interv 2002;56:333-40. 10.1002/ccd.10189 [DOI] [PubMed] [Google Scholar]
  • 13.Tu S, Koning G, Jukema W, et al. Assessment of obstruction length and optimal viewing angle from biplane X-ray angiograms. Int J Cardiovasc Imaging 2010;26:5-17. 10.1007/s10554-009-9509-3 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Park SJ, Kang SJ, Ahn JM, et al. Visual-functional mismatch between coronary angiography and fractional flow reserve. JACC Cardiovasc Interv 2012;5:1029-36. 10.1016/j.jcin.2012.07.007 [DOI] [PubMed] [Google Scholar]
  • 15.Ramcharitar S, Daeman J, Patterson M, et al. First direct in vivo comparison of two commercially available three-dimensional quantitative coronary angiography systems. Catheter Cardiovasc Interv 2008;71:44-50. 10.1002/ccd.21418 [DOI] [PubMed] [Google Scholar]
  • 16.Collet C, Onuma Y, Cavalcante R, et al. Quantitative angiography methods for bifurcation lesions: a consensus statement update from the European Bifurcation Group. EuroIntervention 2017;13:115-23. 10.4244/EIJ-D-16-00932 [DOI] [PubMed] [Google Scholar]
  • 17.Adjedj J, Xaplanteris P, Toth G, et al. Visual and quantitative assessment of coronary stenoses at angiography versus fractional flow reserve: the impact of risk factors. Circ Cardiovasc Imaging 2017;10:e006243. 10.1161/CIRCIMAGING.117.006243 [DOI] [PubMed] [Google Scholar]
  • 18.De Bruyne B, Hersbach F, Pijls NH, et al. Abnormal epicardial coronary resistance in patients with diffuse atherosclerosis but "Normal" coronary angiography. Circulation 2001;104:2401-6. 10.1161/hc4501.099316 [DOI] [PubMed] [Google Scholar]
  • 19.Nallamothu BK, Spertus JA, Lansky AJ, et al. Comparison of clinical interpretation with visual assessment and quantitative coronary angiography in patients undergoing percutaneous coronary intervention in contemporary practice: the Assessing Angiography (A2) project. Circulation 2013;127:1793-800. 10.1161/CIRCULATIONAHA.113.001952 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Zhang H, Mu L, Hu S, et al. Comparison of physician visual assessment with quantitative coronary angiography in assessment of stenosis severity in China. JAMA Intern Med 2018;178:239-47. 10.1001/jamainternmed.2017.7821 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.De Bruyne B, Pijls NH, Kalesan B, et al. Fractional flow reserve-guided PCI versus medical therapy in stable coronary disease. N Engl J Med 2012;367:991-1001. 10.1056/NEJMoa1205361 [DOI] [PubMed] [Google Scholar]
  • 22.De Bruyne B, Sarma J. Fractional flow reserve: a review. Heart 2008;94:949-59. 10.1136/hrt.2007.122838 [DOI] [PubMed] [Google Scholar]
  • 23.Meuwissen M, Chamuleau SA, Siebes M, et al. Role of variability in microvascular resistance on fractional flow reserve and coronary blood flow velocity reserve in intermediate coronary lesions. Circulation 2001;103:184-7. 10.1161/01.CIR.103.2.184 [DOI] [PubMed] [Google Scholar]
  • 24.Gould KL, Johnson NP, Bateman TM, et al. Anatomic versus physiologic assessment of coronary artery disease: role of coronary flow reserve, fractional flow reserve, and positron emission tomography imaging in revascularization decision-making. J Am Coll Cardiol 2013;62:1639-53. 10.1016/j.jacc.2013.07.076 [DOI] [PubMed] [Google Scholar]
  • 25.Matsumura M, Johnson NP, Fearon WF, et al. Accuracy of fractional flow reserve measurements in clinical practice: observations from a core laboratory analysis. JACC Cardiovasc Interv 2017;10:1392-401. 10.1016/j.jcin.2017.03.031 [DOI] [PubMed] [Google Scholar]
  • 26.Leone AM, De Caterina AR, Basile E, et al. Influence of the amount of myocardium subtended by a stenosis on fractional flow reserve. Circ Cardiovasc Interv 2013;6:29-36. 10.1161/CIRCINTERVENTIONS.112.971101 [DOI] [PubMed] [Google Scholar]
  • 27.Svanerud J, Ahn JM, Jeremias A, et al. Validation of a novel non-hyperemic index of coronary artery stenosis severity: the Resting Full-cycle Ratio (VALIDATE RFR) study. EuroIntervention 2018;14:806-14. 10.4244/EIJ-D-18-00342 [DOI] [PubMed] [Google Scholar]
  • 28.Tu S, Westra J, Yang J, et al. Diagnostic accuracy of fast computational approaches to derive fractional flow reserve from diagnostic coronary angiography: the international multicenter FAVOR Pilot Study. JACC Cardiovasc Interv 2016;9:2024-35. 10.1016/j.jcin.2016.07.013 [DOI] [PubMed] [Google Scholar]
  • 29.Westra J, Tu S, Winther S, et al. Evaluation of coronary artery stenosis by quantitative flow ratio during invasive coronary angiography: The WIFI II Study (Wire-Free Functional Imaging II). Circ Cardiovasc Imaging 2018;11:e007107. 10.1161/CIRCIMAGING.117.007107 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Pellicano M, Lavi I, De Bruyne B, et al. Validation study of image based fractional flow reserve during coronary angiography. Circ Cardiovasc Interv 2017. doi: . 10.1161/CIRCINTERVENTIONS.116.005259 [DOI] [PubMed] [Google Scholar]
  • 31.Morris PD, Ryan D, Morton AC, et al. Virtual fractional flow reserve from coronary angiography: modeling the significance of coronary lesions: results from the VIRTU-1 (VIRTUal Fractional Flow Reserve From Coronary Angiography) study. JACC Cardiovasc Interv 2013;6:149-57. 10.1016/j.jcin.2012.08.024 [DOI] [PubMed] [Google Scholar]

Articles from Cardiovascular Diagnosis and Therapy are provided here courtesy of AME Publications