Skip to main content
Plastic and Reconstructive Surgery Global Open logoLink to Plastic and Reconstructive Surgery Global Open
. 2019 Oct 29;7(10):e2437. doi: 10.1097/GOX.0000000000002437

Use of Near-infrared Spectroscopy and Implantable Doppler for Postoperative Monitoring of Free Tissue Transfer for Breast Reconstruction: A Systematic Review and Meta-analysis

Melissa Berthelot *,, James Ashcroft , Piers Boshier , Judith Hunter , Francis Patrick Henry , Benny Lo *, Guang-Zhong Yang *, Daniel Leff *
PMCID: PMC6846307  PMID: 31772881

Supplemental Digital Content is available in the text.

Background:

Failure to accurately assess the perfusion of free tissue transfer (FTT) in the early postoperative period may contribute to failure, which is a source of major patient morbidity and healthcare costs. This systematic review and meta-analysis aim to evaluate and appraise current evidence for the use of near-infrared spectroscopy (NIRS) and/or implantable Doppler (ID) devices compared with conventional clinical assessment (CCA) for postoperative monitoring of FTT in reconstructive breast surgery.

Methods:

A systematic literature search was performed in accordance with the preferred reporting items for systematic reviews guidelines. Studies in human subjects published within the last decade relevant to the review question were identified. Meta-analysis using random-effects models of FTT failure rate and STARD scoring was then performed on the retrieved publications.

Results:

Nineteen studies met the inclusions criteria. For NIRS and ID, the mean sensitivity for the detection of FTT failure is 99.36% and 100% respectively, with average specificity of 99.36% and 97.63%, respectively. From studies with sufficient reported data, meta-analysis results demonstrated that both NIRS [OR = 0.09 (0.02–0.36); P < 0.001] and ID [OR = 0.39 (0.27–0.95); P = 0.04] were associated with significant reduction of FTT failure rates compared with CCA.

Conclusions:

The use of ID and NIRS provided equivalent outcomes in detecting FTT failure and were superior to CCA. The ability to acquire continuous objective physiological data regarding tissue perfusion is a perceived advantage of these techniques. Reduced clinical staff workload and minimized hospital costs are also perceived as positive consequences of their use.

INTRODUCTION

Free tissue transfer (FTT) for breast reconstruction following mastectomy has become a standard procedure on account of its superior aesthetics and durability compared with implant reconstruction. Vascular microanastomosis is a critical step for tissue survival. Anastomosis failure causes a lack of oxygen and nutrients to be perfused within the FTT: thrombus or bleeding of either the recipient artery or donor vein may lead to ischemia and congestion, respectively, which may contribute to tissue necrosis. Studies have described that venous thrombosis is the most common microsurgical complication followed by arterial thrombosis and bleeding.4 Most microsurgical complications have been reported to happen within the first 24–48 hours following surgery with higher rates within the first 4 hours.5 Close monitoring is therefore necessary to detect signs of vascular complications to salvage the FTT and decrease the failure rate.13 Although there is no standardization, conventional clinical assessments (CCA) usually consists of regular visual and kinetic evaluations of the FTT. As the final health status depends on the expertise of the clinical team, additional tools are often used. Hand-held acoustic Doppler sonography can also be used for assessment of the blood flow across pedicles. However, these assessments are discrete, prone to human error, and, cannot provide prompt, real time and systematic detection of possible microanastomotic complications. Due to the shortcomings of CCA, devices based on the biophysical and biochemical tissue properties have been developed for continuous monitoring of the FTT. Specifically, near-infrared spectroscopy (NIRS) and implantable ultrasound Doppler (ID) devices have been commonly used for continuous and objective assessment of the tissue health.68

In the context of growing interest in the use of NIRS and ID to aid early detection of FTT complication and to prevent adverse patient outcomes, it is pertinent that current evidence in regard to these technologies is reviewed. The purpose of this systematic review is to compare the clinical outcomes of NIRS and/or ID and CCA for FTT monitoring.

METHODS

Systematic Review

A systematic search of the literature (title and abstract of full papers and conference abstracts) was performed using the guidelines described by the preferred reporting items for systematic reviews (http://www.prisma-statement.org/) and meta-analysis statement to identify publications between 2008 and October 2018 regarding the use of (1) ID, (2) NIRS, and (3) combined NIRS and ID in the postoperative monitoring of FTT for immediate or delayed breast reconstruction following mastectomy.

Inclusion and Exclusion Criteria

Only studies published in English before October 2018 containing original data where ID, NIRS, combined NIRS/ID, combined NIRS/CCA and combined ID/CCA were used to monitor FTT for breast reconstruction in humans were included. Review articles, oral or poster presentations, conference abstracts, letters, comments, any study describing the validation of animal or cadaveric simulation in surgical training, unpublished data, and any article using nonoriginal data (ie, previously published) were excluded.

Search

We conducted a systematic search strategy, which combined the following search terms and their variations with AND and OR operators using the OVID (EMBASE/Medline) database: “reconstruction,” “autograft,” “surgical flap,” “implantable Doppler,” “monitoring,” “physical examination,” “near-infrared spectroscopy,” and “NIRS.” An additional search concerning specific devices (“Cook-Swartz,” “O2C,” “Synovis”) was performed. MeSH terms and truncation symbols were used where possible to widen the search.

Our systematic search strategy was applied the following online databases: Medline, Embase, PsycINFO, Global Health, HMIC, the Cochrane Libraries Database of Systematic Reviews, SCOPUS, NHS Evidence, the Transport Database. We used the OvidSP platform to search the Medline, EMBASE, PsychINFO, Global Health, and HMIC databases.

Selection Protocol and Data Extraction

Titles and abstracts of studies identified by the primary search were independently reviewed by M.B. and J.A. to identify potentially relevant articles. The full texts of potentially relevant articles were obtained to facilitate further review. In the cases were from the same patient population was presented in full or partially in separate publications either the most recent or relevant article was included. The reference lists of retrieved articles and relevant reviews were also hand-searched to identify other suitable publications. Any areas of disagreement between reviewers were resolved by P.B.

The articles satisfying the selection criteria were retrieved for independent data extraction by M.B. and J.A. Data include author, year of publication, country of design, patient number, characteristics of patient population, FTT overall survival rate, sensitivity, and, specificity of the monitoring method.

Statistical Analysis and Assessment of Quality Report

An additional meta-analysis is conducted using random effects, DerSimonian-Laird method (RevMan, Edition 5, Cochrane Library of Systematic Reviews) based on papers with adequate information to observe FTT failure results when respectively comparing ID and NIRS with CCA. Publication bias of included studies was assessed using funnel plots. The Egger test was used to assess funnel plot asymmetry. The I2 value was used to assess heterogeneity between studies, to determine the degree of variation not attributable to chance alone. Low, moderate, and high degrees of heterogeneity were ascribed to I2 values <25%, 25%–75%, and >75%, respectively. Statistical significance was assigned to P values <0.05. Methodological quality of the papers was independently assessed by M.B. and J.A. using the STARD checklist28 against 34 criteria (see appendices, Supplemental Digital Content 1, which displays STARD checklists for ID retrieved publications, http://links.lww.com/PRSGO/B226, and Supplemental Digital Content 2, which displays STARD checklists for NIRS retrieved publications, http://links.lww.com/PRSGO/B227).

RESULTS

The initial search identified 129 citations, of which 98 were excluded upon title and abstract screen. Thirty-one citations underwent further review, upon which 21 citations were removed. The remaining 15 citations were retrieved. An additional 4 citations were identified through hand searching of bibliographies (Fig. 1). Data were extracted from the 19 retrospective and prospective studies (Tables 1 and 2).

Fig. 1.

Fig. 1.

PRISMA diagram of the search of the last 10 years of original publications on clinical studies on the investigation of implantable Doppler, NIRS, combined NIRS/implantable Doppler, combined NIRS/conventional clinical assessment, and combined implantable Doppler/conventional clinical assessment for the postoperative monitoring of FTT for immediate or delayed breast reconstructive surgery following mastectomy. PRISMA indicates preferred reporting items for systematic reviews.

Table 1.

A Summary of Studies Using ID for Postoperative Monitoring of FTT for Immediate or Delayed Breast Reconstruction following Mastectomy

Authors Year Country Study Design No. Patients No. FTT No. Re-explanations Overall Survival (%) Measured Sensitivity (%) Measured Specificity (%) P Value at 5%* Device and Placement STARD Score References
Smit et al 2008 Sweden RCS 103 121 14 98 100 99 Cook-Swartz Vein 18/34 10
Rozen et al 2010 Sweden RCS 121 121 11 96.3 100 99 P < 0.01 Cook-Swartz Vein 18/34 13
Whitaker et al 2010 Denmark RCS 103 121 11 97.5 100 99 P = 0.93 Cook-Swartz Vein 20/34 27
Schmulder et al 2011 Germany RCS 37 37 6 97.3 P = 0.049 Cook-Swartz Vein 16/34 29
Levine et al 2013 United States RCS 84 134 4 97.8 Cook-Swartz Artery/Vein 9/34 30
Um et al 2014 United States RCS 76 109 11 98 100 99 Cook-Swartz Vein 21/34 18
Um et al 2014 United States RCS 74 111 5 99 100 98 Synovis flow coupler vein 21/34 18
Kempton et al 2014 United States RCS 50 85 6 95.3 100 94 Synovis flow coupler vein 19/34 15
Chang et al 2015 United States RCS 53 100 88 Cook-Swartz Artery/Vein 16/34 31

*P value obtained with a t test at 5% comparing the significant difference in the monitoring outcomes with the use of ID and conventional clinical monitoring methods.

RCS, retrospective case series.

Table 2.

A Summary of Studies Using NIRS Devices for Postoperative Monitoring of FTT for Immediate or Delayed Breast Reconstruction following Mastectomy

Authors Year Country Study Design No. Patients No. FTT No. Re-explorations Overall Survival (%) Measured Sensitivity (%) Measured Specificity (%) P Value at 5%* NIRS Device STARD Score References
Repez et al 2008 Slovenia PCS 48 50 13 94 100 100 InSpectra Model 325 20/34 32
Keller 2009 United States RCS 145 208 8 100 100 100 ViOptix 14/34 33
Lin et al 2010 United States RCS 164 234 16 93.8 100 100 P = 0.015 ViOptix 17/34 26
Pelletier et al 2011 United States PCS 25 25 4 96 100 100 ViOptix in ICU 19/34 34
Pelletier et al 2011 United States PCS 25 25 3 100 100 100 ViOptix in the ward 19/34 34
Whitaker et al 2012 United Kingdom PCS 10 10 4 99 100 100 InSpectra Model 650 17/34 35
Rothenberger et al 2013 Germany PCS 34 34 5 97 100 97 O2C Machine 17/34 36
Koolen et al 2015 United States RCS 451 670 29 96.6 96.5 99.8 P = 0.464 ViOptix 24/34 37
Vranken et al 2017 The Netherlands PCS 29 29 2 96.5 100 96.4 Invos 5000C Oximeter 16/34 38
Ricci et al 2017 United States RCS 595 900 32 99.7 96.5 99.8 ViOptix 19/34 39
Fox et al 2013 United States PCS 27 32 1 100 100 100 Spectros T-Stat 14/34 40
Mericli et al 2017 United States PCS 68 81 3 100 100 100 Spectros T-Stat 16/34 54

*P value obtained with a t test at 5% comparing the significant difference in the monitoring outcomes with the use of NIRS devices and conventional clinical monitoring methods.

PCS, prospective case series; RCS, retrospective case series.

STARD Methodological Quality Report

Assessment of reporting standards against the 34 criteria of the STARD checklist shows a mean of 17.1 (median: 18, range: 9–21) for ID and a mean of 19.2 (median: 17, range: 14–24) for NIRS. Tables 1 and 2 give the STARD score for each retrieved study.

Implantable Doppler

Of the studies identified, 8 used ID to monitor FTT. The extracted data from these studies are presented in Table 1.

Meta-analysis: FTT Failure Results

Comparisons between ID and clinical monitoring outcomes were made in 4 studies.13,15,27,29 Meta-analysis of outcomes presented within these studies identified that ID monitoring was associated with reduced odds of FTT failure [OR = 0.39 (0.27–0.95); P = 0.04) (Fig. 2).

Fig. 2.

Fig. 2.

A forest plot of the odds ratios (ID monitoring vs clinical monitoring).

ID versus Clinical Monitoring

Schmulder et al,29 Rozen et al,13 and Whitaker et al27 investigated the impact of using the Cook-Swartz ID (Cook Medical; Cook Ireland Ltd., Limerick, Ireland) compared with that of CCA (Table 1). Schmulder et al29 and Rozen et al13 found that using ID significantly improved the success rate of FTT surgery for breast reconstruction. However, Whitaker et al27 found similar success rate for both CCA and ID. In addition, although Rozen et al13 found that there are no statistical differences in the false-positive and re-exploration rates, Whitaker et al27 did in favor of ID. This difference might be due to the difference between the number of cases monitored with CCA and that with the ID. In Rozen et al13 study, about 3.5 times more FTT cases were monitored with CCA alone while about 2 and 1.1 times more in Whitaker et al27 and Schmulder et al29 studies, respectively. Additional differences can originate from CCA results which remain subjective and require expertise.

Sensitivity and Specificity of ID Probes

Focusing on the sensitivity and specificity of the use of Cook-Swartz ID, Chang et al31 found that they are both high despite 40.4% of negative findings on re-exploration. When looking at all types of surgery, Chang et al31 advocate that positioning the ID on the artery is recommended as sensitivity and specificity are greater than when placed on the vein (respectively, Npatient = 267 at 94.2% sensitivity and Npatient = 101 at 74.0% sensitivity). They also found no statistical difference in the results when an ID is placed only on the artery compared with the results when placing 2 ID probes on each the vein and artery (Npatient = 71).

ID Probe Placement

Unlike Chang et al,31 Swartz et al41,42 have demonstrated that when placed on the artery, while arterial occlusion is immediately detected, detection of venous occlusion is delayed for up to 6 hours due to persistent arterial pulse. Likewise, when placed on the vein, venous occlusion is immediately detected and arterial occlusion is detected after 6 minutes in average. The difference in the practical use of the probe might come from the characteristics of the FTT, such as the size of the pedicle, the patient’s vascular pattern, or the type and weight of the FTT.

Near-infrared Spectroscopy

Of the studies identified, 11 used NIRS devices to monitor FTT. The extracted data from these studies are in Table 2.

Meta-analysis: FTT Failure Results

Comparisons between NIRS and CCA were made in only 2 studies,26,37 with NIRS monitoring being greatly associated with reduced odds of FTT failure [OR = 0.09 (0.02–0.36); P < 0.001] (Fig. 3).

Fig. 3.

Fig. 3.

A forest plot of the odds ratios (NIRS monitoring vs clinical monitoring).

NIRS versus Conventional Monitoring

Lin et al26 and Koolen et al37 compared the outcomes of CCA with these of the T.Ox. (ViOptix Inc.). For Lin et al,26 the difference in the outcomes between the 2 methods is statistically significant, while it is not for Koolen et al.37 Similar studies were conducted by Fox et al40 with the T-Stat device (Spectros Corp., Portola Valley, Calif.), by Whitaker et al35 and by Repez et al32 with the InSpectra device (Novatech Resources Pte Ltd.), and Rothenberger et al36 using the O2C machine (LEA Medizintechnik GmbH). In general, results show an increase in the FTT salvage rate and an early detection of complications, even before clinical evidence. Differences in the results might come from subjectivity in CCA results as it requires expertise.

Sensitivity and Specificity of NIRS: Clinical Use

Keller33 and Mericli et al54 compared NIRS monitoring to CCA for early detection of FTT complications. Results show higher sensitivity and specificity with a detection of complications before clinical evidence using the NIRS device.

Ricci et al39 investigated the potential use of NIRS device as a mean to reduce intensive monitoring and lower consequent care costs. As high sensitivity and specificity were found using the T.Ox device, they advocate that the use of a NIRS device for postoperative monitoring can reduce the time patients spend in an intensive monitoring setting while saving hospital cost for similar outcomes compared with CCA. Similarly, Pelletier et al34 found concurring results and suggest that a NIRS device can completely replace specialized clinical staff for postoperative monitoring.

NIRS Placement

Vranken et al38 investigated the use of NIRS devices as a reliable mean for postoperative FTT monitoring and can provide early detection of a complication. Using the Invos 5000C (MedTronic Ltd.), which is primary designed for cerebral/somatic use, results were conclusive and the authors advocate that the comparison of the StO2 measurements pattern from “healthy” FTT with the measurements of the FTT under monitoring can help in the identification of complications at an early stage. However, the pattern was not fully determined and merely described as increasing or decreasing slopes. The authors also compared the measurements of the monitored FTT with another sensor placed on the native breast tissue which acts as a personalized reference value. They advocate that their comparison provides a deeper insight on the absolute value of the variation of StO2 within the FTT.

DISCUSSION

The main limitation of the ID lies in its positioning. Improper positioning can result in (1) false-positive alerts, especially if the probe has moved, which may induce negative re-explorations; (2) hematoma following anastomosis rupture, sometimes due to inadvertent pulling or removal of the wire; (3) blood vessel thrombosis, if the probe is too tightly secured onto the blood vessel. With experience, the number of complications due to the use of the ID generally reduces. Also, by combining the ID outcomes with CCA and other monitoring methods, false-positive alarms can be backed up to avoid negative re-explorations.44

NIRS devices have limitations that are intrinsic to (1) the device’s hardware, (2) the device’s algorithm for the analysis of the measurements, and (3) the patient’s demographic details.9,21 Components (light emitter and light detector) placement and light wavelengths induce hardware limitations which result in a specific range of penetration depth and a relative measurement of the concentration of blood compound or tissue components. Hardware limitations can be overcome with the design of a specific algorithm for the analysis of the measurements, for signal amplification and adjustment when investigating a given compound.21 Ozturk et al45 showed that surgical and clinical parameters, such as blood pressure, supplemental oxygen saturation, FTT type, perforator’s size and number, patient’s demographic details (such as skin tone and age), and the environment’s characteristics into which measurements are taken (such as ambient light) can also affect the final outcome of the NIRS device.9,21,25 Mostly due to these limitations, it is difficult to provide absolute StO2 measurements or to rigorously compare StO2 measurements between different NIRS devices.

To directly compare ID and NIRS devices, it is necessary to acknowledge their differences: (1) between what is measured, respectively blood flow perfusion, and, StO2 and its related parameters; (2) between their positions, respectively, implanted at the entrance/exit of the FTT (onto the pedicles) and at an external localized area.49 Although these differences are related because a change in blood flow perfusion will affect the overall StO2 independently to where the measurement is taken, they can affect the minimum time taken to detect a complication. To compensate for autonomic denervation, inflammatory reaction, and ischemia, which are normal reactions following FTT harvest, StO2 consumption increases. This induces an increased blood flow perfusion and a decreased vascular resistance that dilates the microcirculation within the tissue, which facilitate the dissolution of possible microthrombosis and the repair of micronecrosis.5052 These physical responses can vary according to the vascular pattern, nicotine or alcohol abuse, demographics, and comorbidities of the patient.45,49,51 Raittinen et al52 argued that although the readings might vary across patients, the StO2 trend remains the same. This trend has been described as a sharp increase followed by a decrease and ultimately a stabilization within the first 3 days following surgery.52,53

CONCLUSIONS

Most studies found the ID as an effective, efficient, and safe monitoring system that is a valuable adjunct to CCA. Specifically, Rozen et al,13 Schmulder et al,29 and Whitaker et al28 have reported that the system can improve salvage rates. However, Smit et al10 reported no such improvement in salvage rates, but encourage its use as it reduces the workload of the clinical staff and interruptions on the patients. Chang et al31 suggest that CCA should remain the gold standard for postoperative FTT monitoring, especially when a cutaneous paddle on the FTT is available. Using a weighted average analysis, based on the total number of FTT cases, the overall re-exploration rate, survival rate, sensitivity, and specificity were calculated. Overall studies investigating the results of the Cook-Swartz placed at the venous pedicle,10,13,18,27,29 with a total number of 509 FTT cases, results demonstrate a 10.41% re-exploration rate, 97.20% survival rate, 100% sensitivity, and 99.00% specificity. Similarly, with the Synovis Flow Coupler placed at the venous pedicle,15,18 with a total number of 196 FTT cases, results show 5.63% re-exploration rate, 97.39% survival rate, 100% sensitivity, and 96.26% specificity.

Most studies reported that NIRS is a highly sensitive, specific, and reliable technique that can improve the FTT salvage rate. Most studies have a 0% false-positive rate.26,27,3234 Ricci et al39 and Koolen et al37 studies have a false-positive rate at 0.1% and 0.15%, respectively. In these studies, no particular definition of a false-positive result was given. Vranken et al38 and Rothenberg et al36 did not provide false-positive results; however, their high specificities suggest low false-positive rates. In addition, Pelletier et al34 advocate that the use of NIRS devices can reduce the FTT monitoring cost structure. Although further research on the use of NIRS technique for buried FTT has been reported,24,43 at the moment NIRS devices are conventionally used when a cutaneous paddle is available. A weighted average analysis based on the number of FTT cases was performed. Overall studies investigating the results of the ViOptix,26,33,34,37,39 with a total number of 2,062 FTT cases, results show 4.41% re-exploration rate, 97.92% survival rate, 97.24% sensitivity, and 99.75% specificity.

Specific studies investigating either implantable or NIRS methods, compared with or without CCA have been conducted. With the included studies, the weighted average results show that NIRS, using the T.Ox machine, gives better outcomes than when using the Cook-Swartz or Synovis ID, with, respectively, 4 and 10.5 times more FTT cases recruited. However, large studies aiming to compare NIRS and ID on the same patients under the same clinical conditions would need to be conducted for a more rigorous comparison. Table 3 shows the prices of the most common devices used in the presented clinical studies.

Table 3.

Costs of the NIRS and ID Devices Which Are Currently Available and Used in the Reviewed Studies

Company Name Controller Machine Price Disposable Probe Price
Cook Medical Cook-Swartz 2,494.00 GBP 380.00 GBP
Synovis Micro Flow Coupler 2,305.00 GBP 714.00 GBP
ViOptix T.Ox 25,000.00 USD 1,450.00 USD
Spectros T-Stat 32,000.00 USD 629.00 USD

Many studies have demonstrated that both methods are useful and reliable as they reduce the staff burden and improve the overall FTT salvage rate. However, few studies have tried to decipher which method is the most appropriate taking into consideration the multiple variables across the range of microsurgical applications. In addition to high accuracy and robustness, wider and systematic use of such methods lies into their (1) ease of use, with a relatively short learning curve, (2) acceptance (based on the medical relevance, learning curve, and large studies outcomes), and (3) relative costs.

Supplementary Material

gox-7-e2437-s001.pdf (1.4MB, pdf)
gox-7-e2437-s002.pdf (1.9MB, pdf)

Footnotes

Published online 25 October 2019.

Supported by EPSRC Smart Sensing for Surgery (EP/LO14149/1) and the EPSRC CDT HiPEDS (EP/LO16796/1).

Disclosure: The authors have no financial interest to declare in relation to the content of this article.

Related Digital Media are available in the full-text version of the article on www.PRSGlobalOpen.com.

REFERENCES

  • 1.Hovius SE, van Adrichem LN, Mulder HD, et al. The predictive value of the laser Doppler flowmeter for postoperative microvascular monitoring. Ann Plast Surg. 1993;31:307–312. [DOI] [PubMed] [Google Scholar]
  • 2.Poder TG, Fortier PH. Implantable Doppler in monitoring free flaps: a cost-effectiveness analysis based on a systematic review of the literature. Eur Ann Otorhinolaryngol Head Neck Dis. 2013;130:79–85. [DOI] [PubMed] [Google Scholar]
  • 3.Ponticorvo A, Taydas E, Mazhar A, et al. Quantitative assessment of partial vascular occlusions in a swine pedicle flap model using spatial frequency domain imaging. Biomed Opt Express. 2013;4:298–306. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Bui DT, Cordeiro PG, Hu QY, et al. Free flap reexploration: indications, treatment, and outcomes in 1193 free flaps. Plast Reconstr Surg. 2007;119:2092–2100. [DOI] [PubMed] [Google Scholar]
  • 5.Chen KT, Mardini S, Chuang DC, et al. Timing of presentation of the first signs of vascular compromise dictates the salvage outcome of free flap transfers. Plast Reconstr Surg. 2007;120:187–195. [DOI] [PubMed] [Google Scholar]
  • 6.Cervenka B, Bewley AF. Free flap monitoring: a review of the recent literature. Curr Opin Otolaryngol Head Neck Surg. 2015;23:393–398. [DOI] [PubMed] [Google Scholar]
  • 7.Chao AH, Meyerson J, Povoski SP, et al. A review of devices used in the monitoring of microvascular free tissue transfers. Expert Rev Med Devices. 2013;10:649–660. [DOI] [PubMed] [Google Scholar]
  • 8.Smit JM, Zeebregts CJ, Acosta R, et al. Advancements in free flap monitoring in the last decade: a critical review. Plast Reconstr Surg. 2010;125:177–185. [DOI] [PubMed] [Google Scholar]
  • 9.Berthelot M, Chen CM, Yang GZ, et al. Wireless wearable self-calibrated sensor for perfusion assessment of myocutaneous tissue in 2016. 2016Paper presented at: IEEE 13th International Conference on Wearable and Implantable Body Sensor Networks (BSN):171–176IEEE. [Google Scholar]
  • 10.Smit JM, Whitaker IS, Liss AG, et al. Post operative monitoring of microvascular breast reconstructions using the implantable Cook-Swartz Doppler system: a study of 145 probes & technical discussion. J Plast Reconstr Aesthet Surg. 2009;62:1286–1292. [DOI] [PubMed] [Google Scholar]
  • 11.Rozen WM, Ang GG, McDonald AH, et al. Sutured attachment of the implantable Doppler probe cuff for large or complex pedicles in free tissue transfer. J Reconstr Microsurg. 2011;27:99–102. [DOI] [PubMed] [Google Scholar]
  • 12.Paydar KZ, Hansen SL, Chang DS, et al. Implantable venous Doppler monitoring in head and neck free flap reconstruction increases the salvage rate. Plast Reconstr Surg. 2010;125:1129–1134. [DOI] [PubMed] [Google Scholar]
  • 13.Rozen WM, Chubb D, Whitaker IS, et al. The efficacy of postoperative monitoring: a single surgeon comparison of clinical monitoring and the implantable Doppler probe in 547 consecutive free flaps. Microsurgery. 2010;30:105–110. [DOI] [PubMed] [Google Scholar]
  • 14.Smit JM, Klein S, de Jong EH, et al. Value of the implantable Doppler system in free flap monitoring. J Plast Reconstr Aesthet Surg. 2012;65:1276–1277. [DOI] [PubMed] [Google Scholar]
  • 15.Kempton SJ, Poore SO, Chen JT, et al. Free flap monitoring using an implantable anastomotic venous flow coupler: analysis of 119 consecutive abdominal-based free flaps for breast reconstruction. Microsurgery. 2015;35:337–344. [DOI] [PubMed] [Google Scholar]
  • 16.Kind GM, Buntic RF, Buncke GM, et al. The effect of an implantable Doppler probe on the salvage of microvascular tissue transplants. Plast Reconstr Surg. 1998;101:1268–1273. [PubMed] [Google Scholar]
  • 17.Wax MK. The role of the implantable Doppler probe in free flap surgery. Laryngoscope. 2014;124:S1–S12. [DOI] [PubMed] [Google Scholar]
  • 18.Um GT, Chang J, Louie O, et al. Implantable Cook-Swartz Doppler probe versus synovis flow coupler for the post-operative monitoring of free flap breast reconstruction. J Plast Reconstr Aesthet Surg. 2014;67:960–966. [DOI] [PubMed] [Google Scholar]
  • 19.Wukitsch MW, Petterson MT, Tobler DR, et al. Pulse oximetry: analysis of theory, technology, and practice. J Clin Monit. 1988;4:290–301. [DOI] [PubMed] [Google Scholar]
  • 20.Severinghaus JW. Takuo Aoyagi: discovery of pulse oximetry. Anesth Analg. 2007;105:S1–S4. [DOI] [PubMed] [Google Scholar]
  • 21.Berthelot M, Yang GZ, Lo B. A self-calibrated tissue viability sensor for free flap monitoring. IEEE J Biomed Health Inform. 2018;22:5–14. [DOI] [PubMed] [Google Scholar]
  • 22.Matsui A, Lee BT, Winer JH, et al. Quantitative assessment of perfusion and vascular compromise in perforator flaps using a near-infrared fluorescence guided imaging system. Plast Reconstr Surg. 2009;124:451. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Kagaya Y, Miyamoto S. A systematic review of near-infrared spectroscopy in flap monitoring: current basic and clinical evidence and prospects. J Plast Reconstr Aesthet Surg. 2018;71:246–257. [DOI] [PubMed] [Google Scholar]
  • 24.Berthelot M, Yang GZ, Lo B. Tomographic probe for perfusion analysis in deep layer tissue in 2018. 2018Paper presented at: IEEE 15th International Conference on Wearable and Implantable Body Sensor Networks (BSN)IEEE. [Google Scholar]
  • 25.Wassenaar EB, Van den Brand JG. Reliability of near-infrared spectroscopy in people with dark skin pigmentation. J Clin Monit Comput. 2005;19:195–199. [DOI] [PubMed] [Google Scholar]
  • 26.Lin SJ, Nguyen MD, Chen C, et al. Tissue oximetry monitoring in microsurgical breast reconstruction decreases flap loss and improves rate of flap salvage. Plast Reconstr Surg. 2011;127:1080–1085. [DOI] [PubMed] [Google Scholar]
  • 27.Whitaker IS, Rozen WM, Chubb D, et al. Postoperative monitoring of free flaps in autologous breast reconstruction: a multicenter comparison of 398 flaps using clinical monitoring, microdialysis, and the implantable Doppler probe. J Reconstr Microsurg. 2010;26:409–416. [DOI] [PubMed] [Google Scholar]
  • 28.Bossuyt PM, Reitsma JB, Bruns DE, et al. Towards complete and accurate reporting of studies of diagnostic accuracy: the STARD initiative. Family Practice. 2004;21:4–10. [DOI] [PubMed] [Google Scholar]
  • 29.Schmulder A, Gur E, Zaretski A. Eight-year experience of the Cook-Swartz Doppler in freeflap operations: microsurgical and reexploration results with regard to a wide spectrum of surgeries. Microsurgery. 2011;31:1–6. [DOI] [PubMed] [Google Scholar]
  • 30.Levine SM, Snider C, Gerald G, et al. Buried flap reconstruction after nipple-sparing mastectomy: advancing toward single-stage breast reconstruction. Plast Reconstr Surg. 2013;132:489e–497e. [DOI] [PubMed] [Google Scholar]
  • 31.Chang EI, Ibrahim A, Zhang H, et al. Deciphering the sensitivity and specificity of the implantable Doppler probe in free flap monitoring. Plast Reconstr Surg. 2016;137:971–976. [DOI] [PubMed] [Google Scholar]
  • 32.Repez A, Oroszy D, Arnez ZM. Continuous postoperative monitoring of cutaneous free flaps using near infrared spectroscopy. J Plast Reconstr Aesthet Surg. 2008;61:71–77. [DOI] [PubMed] [Google Scholar]
  • 33.Keller A. A new diagnostic algorithm for early prediction of vascular compromise in 208 microsurgical flaps using tissue oxygen saturation measurements. Ann Plast Surg. 2009;62:538–543. [DOI] [PubMed] [Google Scholar]
  • 34.Pelletier A, Tseng C, Agarwal S, et al. Cost analysis of nearinfrared spectroscopy tissue oximetry for monitoring autologous free tissue breast reconstruction. J Reconstr Microsurg. 2011;27:487–494. [DOI] [PubMed] [Google Scholar]
  • 35.Whitaker IS, Pratt GF, Rozen WM, et al. Near infrared spectroscopy for monitoring flap viability following breast reconstruction. J Reconstr Microsurg. 2012;28:149–154. [DOI] [PubMed] [Google Scholar]
  • 36.Rothenberger J, Amr A, Schaller HE, et al. Evaluation of a non-invasive monitoring method for free flap breast reconstruction using laser Doppler flowmetrie and tissue spectrophotometry. Microsurgery. 2013;33:350–357. [DOI] [PubMed] [Google Scholar]
  • 37.Koolen PG, Vargas CR, Ho OA, et al. Does increased experience with tissue oximetry monitoring in microsurgical breast reconstruction lead to decreased flap loss? Plast Reconstr Surg. 2016;137:1093–1101. [DOI] [PubMed] [Google Scholar]
  • 38.Vranken NPA, Weerwind PW, Onna MA, et al. Non-invasive tissue oximetry following unilateral DIEP-flap reconstruction: a pilot evaluation. JPRAS Open. 2017;12:59–65. [Google Scholar]
  • 39.Ricci JA, Vargas CR, Ho OA, et al. Evaluating the use of tissue oximetry to decrease intensive unit monitoring for free flap breast reconstruction. Ann Plast Surg. 2017;79:42–46. [DOI] [PubMed] [Google Scholar]
  • 40.Fox PM, Zeidler K, Carey J, et al. White light spectroscopy for free flap monitoring. Microsurgery. 2013;33:198–202. [DOI] [PubMed] [Google Scholar]
  • 41.Swartz WM, Jones NF, Cherup L, et al. Direct monitoring of microvascular anastomoses with the 20-MHz ultrasonic Doppler probe: an experimental and clinical study. Plast Reconstr Surg. 1988;81:149–161. [DOI] [PubMed] [Google Scholar]
  • 42.Swartz WM, Izquierdo R, Miller MJ. Implantable venous Doppler microvascular monitoring: laboratory investigation and clinical results. Plast Reconstr Surg. 1994;93:152–163. [DOI] [PubMed] [Google Scholar]
  • 43.Cai ZG, Zhang J, Zhang JG, et al. Evaluation of near infrared spectroscopy in monitoring postoperative regional tissue oxygen saturation for fibular flaps. J Plast Reconstr Aesthet Surg. 2008;61:289–296. [DOI] [PubMed] [Google Scholar]
  • 44.Arya R, Griffiths L, Figus A, et al. Post-operative assessment of perfusion of Deep Inferior Epigastric Perforator (DIEP) free flaps via Pulsatility Index (PI) using a portable colour Doppler sonogram device. J Plast Reconstr Aesthet Surg. 2013;66:931–936. [DOI] [PubMed] [Google Scholar]
  • 45.Ozturk CN, Ozturk C, Ledinh W, et al. Variables affecting postoperative tissue perfusion monitoring in free flap breast reconstruction. Microsurgery. 2015;35:123–128. [DOI] [PubMed] [Google Scholar]
  • 46.Steele MH. Three-year experience using near infrared spectroscopy tissue oximetry monitoring of free tissue transfers. Ann Plast Surg. 2011;66:540–545. [DOI] [PubMed] [Google Scholar]
  • 47.Yoshitani K, Kawaguchi M, Tatsumi K, et al. A comparison of the INVOS 4100 and the NIRO 300 near-infrared spectrophotometers. Anesth Analg. 2002;94:586–590. [DOI] [PubMed] [Google Scholar]
  • 48.Hyttel-Sorensen S, Hessel TW, Greisen G. Peripheral tissue oximetry: comparing three commercial near-infrared spectroscopy oximeters on the forearm. J Clin Monit Comput. 2014;28:149–155. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Jonas R, Schaal T, Krimmel M, et al. Monitoring in microvascular tissue transfer by measurement of oxygen partial pressure: four years experience with 125 microsurgical transplants. J Craniomaxillofac Surg. 2013;41:303–309. [DOI] [PubMed] [Google Scholar]
  • 50.H¨olzle F, Loeffelbein DJ, Nolte D, et al. Free flap monitoring using simultaneous non-invasive laser Doppler flowmetry and tissue spectrophotometry. J Craniomaxillofac Surg. 2006;34:25–33. [DOI] [PubMed] [Google Scholar]
  • 51.Schultze-Mosgau S, Wiltfang J, Birklein F, et al. Microlightguide spectrophotometry as an intraoral monitoring method in free vascular soft tissue flaps. J Oral Maxillofac Surg. 2003;61:292–297. [DOI] [PubMed] [Google Scholar]
  • 52.Raittinen L, Laranne J, Baer G, et al. How we do it: postoperative tissue oxygen monitoring in microvascular free flaps. Clin Otolaryngol. 2005;30:276–278. [DOI] [PubMed] [Google Scholar]
  • 53.Najefi A, Leff DR, Nicolaou M, et al. Monitoring of free flaps using near-infrared spectroscopy: a systematic review of the initial trials. Plast Reconstr Surg. 2010;125:182e–184e. [DOI] [PubMed] [Google Scholar]
  • 54.Mericli AF, Wren J, Garvey PB, et al. A prospective clinical trial comparing visible light spectroscopy to handheld Doppler for postoperative free tissue transfer monitoring. Plast Reconstr Surg. 2017;140:604–613. [DOI] [PubMed] [Google Scholar]

Articles from Plastic and Reconstructive Surgery Global Open are provided here courtesy of Wolters Kluwer Health

RESOURCES