Diagnostic accuracy of flat-panel computed tomography in assessing cerebral perfusion in comparison with perfusion computed tomography and perfusion magnetic resonance: a systematic review

Ernst L Stille 1,, Ilaria Viozzi 2, Mark ter Laan 2, Frederick JA Meijer 1, Jurgen J Futterer 1, Maroeska M Rovers 3
PMCID: PMC6848034  PMID: 31523757

Abstract

Purpose

Flat-panel computed tomography (FP-CT) is increasingly available in angiographic rooms and hybrid OR’s. Considering its easy access, cerebral imaging using FP-CT is an appealing modality for intra-procedural applications. The purpose of this systematic review is to assess the diagnostic accuracy of FP-CT compared with perfusion computed tomography (CTP) and perfusion magnetic resonance (MRP) in cerebral perfusion imaging.

Methods

We performed a systematic literature search in the Cochrane Library, MEDLINE, Embase, and Web of Science up to June 2019 for studies directly comparing FP-CT with either CTP or MRP in vivo. Methodological quality was assessed using the QUADAS-2 tool. Data on diagnostic accuracy was extracted and pooled if possible.

Results

We found 11 studies comparing FP-CT with CTP and 5 studies comparing FP-CT with MRP. Most articles were pilot or feasibility studies, focusing on scanning and contrast protocols. All patients studied showed signs of cerebrovascular disease. Half of the studies were animal trials. Quality assessment showed unclear to high risks of bias and low concerns regarding applicability. Five studies reported on diagnostic accuracy; FP-CT shows good sensitivity (range 0.84–1.00) and moderate specificity (range 0.63–0.88) in detecting cerebral blood volume (CBV) lesions.

Conclusions

Even though FP-CT provides similar CBV values and reconstructed blood volume maps as CTP in cerebrovascular disease, additional studies are required in order to reliably compare its diagnostic accuracy with cerebral perfusion imaging.

Electronic supplementary material

The online version of this article (10.1007/s00234-019-02285-y) contains supplementary material, which is available to authorized users.

Keywords: Cerebral perfusion imaging, Flat-panel detector computed tomography, Perfusion computed tomography, Perfusion magnetic resonance

Introduction

Over the last two decades, flat-panel detector (FD) technology has been introduced into many (neuro-) interventional angiographic suites for diagnostic and therapeutic purposes. Flat-panel detector computed tomography (FP-CT) uses FD technology mounted on a rotational C-arm to generate volumetric imaging with a high spatial resolution [1]. It allows the acquisition of high-quality 3D-vascular images and CT comparable images to study brain parenchyma [13].

More recently, FP-CT has been introduced in the hybrid operating room (OR). Given its easy access, FP-CT is an appealing imaging modality for intra-operative applications. In neurosurgery, magnetic resonance (MR) imaging is already used intra-operatively to control the extent of tumor resection in procedures involving craniotomy [4]. The use of cerebral perfusion imaging during these procedures could yield relevant information on the extent of tumor resection, as well as perfusion changes in the area of interest. Using FP-CT for assessing cerebral perfusion could provide a valuable and pragmatic way of intra-operative feedback in surgical procedures, complementing MR. Outside the hybrid OR, cerebral perfusion is commonly assessed with computed tomography (CTP) or magnetic resonance (MRP). Both CTP and MRP have been reported to allow quantitative assessment of (relative) perfusion parameters like cerebral blood volume (CBV) and cerebral blood flow (CBF), showing good agreement with correlation coefficients of up to r = 0.95 [5]. Because of its wide accessibility and speed, CTP is part of the diagnostic workup for acute stroke patients at many institutions [6]. However, CTP requires the use of iodinated contrast and exposes patients to radiation.

MRP comprises two contrast-dependent sequences (dynamic susceptibility contrast (DSC), dynamic contrast enhanced (DCE)) and one non-contrast sequence (arterial spin labeling (ASL)). MRP can be applied in evaluating the ischemic penumbra in acute ischemic stroke, selecting patients for reperfusion treatment [7]. MRP is more commonly used for the assessment of perfusion in cerebral gliomas in estimating tumor grade pre-operatively and to distinguish tumor recurrence from radiation necrosis post-treatment [8].

Given the growing availability of FP-CT in the hybrid OR, our aim was to study the diagnostic accuracy of FP-CT regarding cerebral perfusion as compared with the more established techniques of CTP and MRP [9]. Even though FP-CT is already in use to study cerebral perfusion, a systematic review to compare its accuracy with CTP and MRP has never been conducted. We therefore performed a systematic literature review, including all published information on FP-CT imaging compared directly with either CTP or MRP.

Materials and methods

The review protocol can be accessed at the website of PROSPERO, the International Prospective Register of Systematic Reviews (https://www.crd.york.ac.uk/PROSPERO/). The protocol was registered under number CRD42017048828. We performed this systematic review and meta-analysis according to the Preferred Reporting Items for Systematic reviews and Meta-Analysis (PRISMA) statement [10].

Search of the literature

We performed a systematic literature search in Medline, Embase, Cochrane, and Web of Science for studies evaluating the accuracy of different perfusion techniques in assessing cerebral hemodynamics in vivo. Our search strings combined “Cerebral circulation” synonyms with synonyms for “Perfusion imaging” and “Flat Panel detector CT” and either “Computed Tomography” or “Magnetic Resonance Imaging” (see Appendix for complete search strategies). Searches were performed from inception up to June 2019. Endnote X8 (Clarivate Analytics, Boston, MA, USA) was used to filter duplicate articles.

Study selection

We selected studies that compared the diagnostic accuracy of FP-CT directly with either conventional CTP or MRP. We included articles in all languages and contacted the authors of conference abstracts asking for any full-text publications reporting the complete results of their study. Considering the recent introduction of FP-CT, only a limited amount of human studies has been conducted. Therefore, animal studies are included as well but these studies are presented and discussed separately. Studies were excluded if no direct comparison between the two imaging modalities was performed, or if the same patient data had already been published in other articles.

In both CT-based techniques, iodinated contrast was used. We selected MRI studies that used perfusion weighted imaging (PWI), DSC, DCE, or ASL. Selected studies reported on absolute or relative perfusion parameters were used in quantifying cerebral hemodynamics: cerebral blood volume in ml/100 g (CBV), cerebral blood flow in ml/100 g/min (CBF), mean transit time in seconds (MTT), or time to peak in seconds (TTP).

Covidence web-based SR software (Veritas Health Innovation, Melbourne, Australia) was used to assist in the process of screening and selection of abstracts and articles. Two reviewers (ES and IV) independently assessed the eligibility of the identified papers. Any disagreements were resolved by discussion with a third reviewer (MR).

Data extraction

We extracted data on patient, study, imaging, and contrast characteristics for all studies included. Study population characteristics comprised human or animal (including species), age, whether general anesthesia was given (including used medication) and the pathology explored. Study characteristics included author and year of publication, study design, imaging methods compared, and method of analysis. Imaging characteristics included machines used, number and duration of sweeps indicating static or dynamic imaging, angle of gantry rotation, total amount of images, and radiation dose. Contrast protocol characteristics focus on FP-CT and consist of the concentration and total amount used, route of administration, and X-ray delay. Two reviewers extracted data for all studies using a standardized data extraction form.

Quality assessment

Both reviewers (ES and IV) independently assessed the risk of bias and concerns regarding applicability to the research question at study level using the validated Quality Assessment of Diagnostic Accuracy Studies (QUADAS-2) tool [11]. Four domains were scored as follows: (1) patient selection, which describes the method for patient selection and the patients included; (2) index test, which describes the test being studied and how it was conducted and interpreted; (3) reference standard, which describes the reference standard test used and how it was conducted and interpreted; and (4) flow and timing, which describe the flow of patient inclusion and exclusion, and the interval between index test and reference standard. Any disagreements were resolved by discussion with a third reviewer (MR).

Data synthesis and analysis

Data on diagnostic accuracy was extracted and pooled if possible. We gathered qualitative data on sensitivity and specificity and inter- and intra-observer variability in Cohen’s kappa coefficient. Quantitative data on absolute and relative perfusion values was gathered in CBV, CBF, MTT, or TTP. Correlations between techniques are shown using Pearson’s r correlation coefficient and agreement between tests is reported in Bland-Altman mean difference test. If extracted data were too heterogeneous to allow for pooling, data will be presented descriptively.

Results

Search of the literature

Our systematic searches for studies that compared FP-CT with CT yielded 458 records and FP-CT with MRI 121 records, which were reduced to 378 and 86 after removing duplicates (see also Fig. 1a and b respectively). Based on title and abstract, 323 and 67 articles were excluded due to incompatibility with our eligibility criteria. Full texts of the remaining 55 and 19 studies were screened and another 44 and 14 articles were excluded, mostly due to a lack of perfusion imaging or no direct comparison between techniques. A total of 11 articles could be included in this systematic review on FP-CT versus CT and 5 articles on FP-CT versus MR.

Fig. 1.

Fig. 1

Fig. 1

a Prisma flowchart FP-CT vs. CTP. b Prisma flowchart FP-CT vs. MRP

Study and patient characteristics

Tables 1 and 2 provide an overview of the included articles and their study design and population for the comparison of FP-CT versus CT and FP-CT versus MR, respectively.

Table 1.

Study characteristics: FP-CT vs. CTP

Author Study characteristics Population characteristics
Design Population Diagnosis Age
Mean Range
Human studies Fiorella (2014) Multicenter pilot study 56 patients CVA 62.9 25–86
Struffert (2012) Cohort 16 patients MCA occlusion, unilateral 69.6
Struffert (2010) Cohort 25 patients CVA 58.3
Zhang (2013) Cohort 12 patients CVA 52 20–76
Zhang (2012) Cohort 20 patients CVA 55.2 30–70
Animal studies Ahmed (2009) Animal trial 12 dogs Healthy subjects
Beuing (2014) Animal trial 7 sheep MCA occlusion, unilateral
Bley (2010) Animal trial 20 dogs MCA occlusion, unilateral
Ganguly (2011) Animal trial 5 pigs Healthy subjects
Royalty (2013) Animal trial 7 dogs ICA/MCA occlusion, unilateral
Yasuda (2012) Animal trial 12 dogs ICA/MCA occlusion, unilateral

CVA, cerebrovascular accident; MCA, middle cerebral artery; ICA, internal carotid artery

Table 2.

Study characteristics: FP-CT vs. MRP

Author Study characteristics Population characteristics
Design Population Diagnosis Age
Mean Range
Human studies Garcia (2017) Cohort 5 patients Arteriovenous malformation
Chen (2018) Cohort 13 patients Carotid stenosis > 70% 67.1 45–78
Kamran (2015) Cohort 26 patients DCI after SAH 55
Kuriyama (2018) Retrospective cohort 10 patients Unruptured aneurysm 58.1 40–71
Struffert (2015) Cohort 12 patients LVO 72

DCI, delayed cerebral ischemia; SAH, subarachnoid hemorrhage; LVO, large vessel occlusion

For the comparison of FP-CT versus CT, four prospective cohort studies [1215], one multicenter pilot study [16], and six animal studies could be included [1722]. The number of included participants or animals per study varied between 12 and 56 patients (median 20), and 5 and 12 animals (median 7). All five human studies only included patients with signs of cerebrovascular disease. Although a wide variety of diagnoses were studied (including aneurysms, vasospasms, and dural fistulas), most participants suffered from large vessel occlusion or stenosis. All examinations were done in the (angio-) intervention room and none of the patients was anesthetized. All six animal trials were performed prospectively using large adult mammals, mainly canines and sheep. The animals were either healthy or suffered from cerebrovascular disease, e.g., artificially created occlusions of the middle cerebral artery or the internal carotid artery. All animals were under general anesthesia to reduce motion artifacts.

For the comparison of FP-CT versus MR, four prospective cohort studies [2326] and one retrospective study [27] could be included. Numbers of included participants per study ranged from 5 to 26 (median 12). All included patients suffered from symptoms of cerebrovascular disease: arteriovenous malformations (n = 5) [26], unruptured cerebral aneurysms (n = 10) [27], large vessel occlusion (n = 12) [23], carotid stenosis (n = 13) [25], and delayed cerebral ischemia after subarachnoid hemorrhage (n = 26) [24]. All studies were conducted in the (angio-) intervention room and none of the patients was under general anesthesia.

Risk of bias assessment

Figure 2a and b show the results of the quality assessment. Risk of bias was scored medium to high in four different domains and there was little or low concern regarding applicability to the research question. Overall, the quality of studies was moderate to poor.

Fig. 2.

Fig. 2

a and b Quality assessment FP-CT vs. CTP/MRP

FP-CT vs. CTP

Of the human studies comparing FP-CT versus CT, one study had a high risk of bias regarding the patient selection due to uncertainties in this selection and heterogeneity in the final study population [16]. Regarding the index test, three studies had an unclear risk of bias due to lack of reporting, or the use of different machines [12, 13, 16]. One study was given a high risk of bias because observers were aware of results of the reference standard while interpreting the index test [14]. In the reference standard domain, two studies had an unclear risk of bias and one study had a high risk of bias due to the use of different machines [12, 14, 16]. In the flow and timing domain, three studies had an unclear risk and two studies a high risk of bias [12, 16] due to the use of different machines, inappropriate time between tests, and cases being excluded from the analyses without specification [1315].

Of the animal studies comparing FP-CT and CT, two studies had an unclear risk of bias in the patient selection domain due to inappropriate exclusions [20, 21]. One of these had concerns regarding applicability as well because of significant differences in (vascular) anatomy and the subsequent need to perform surgery in order to create an artificial stroke. For the index test domain, two studies had an unclear risk of bias because of the use of unspecified thresholds [18, 22]. In the reference standard domain, one study [22] had a high risk of bias due to using different reference standards. In the flow and timing domain, two studies had a high risk of bias due to inappropriate intervals between tests of up to 5 days [18, 22].

FP-CT vs. MRP

Of the five human studies comparing FP-CT with MR, two studies had an unclear risk of bias or concerns regarding applicability in the domain of patient selection [26, 27]. Four studies had an unclear risk of bias for the index test domain [2427], two of which also had unclear concerns in applicability due to the interpretation of test results [24, 25]. Two studies had an unclear risk of bias in flow and timing due to inappropriate intervals between tests [25, 26].

Scanning dose and contrast protocols

Tables 3 and 4 show FP-CT imaging details of all studies. Scan and contrast infusion protocols are described separately for the comparison of FP-CT with CTP and FP-CT with MRP. The most relevant findings are summarized below, with a focus on FP-CT.

Table 3.

Scan protocol specifics: FP-CT vs. CTP

Author Scan protocol Contrast protocol
Machine Sweeps Sec/sweep Static/dynamic Conc. (mg/ml) Route Total Dilution
Human studies Fiorella (2014) Artis Zee 2 8 Static 370 iv 60 ml
Struffert (2012) Artis dBA/Zeego 2 8 Static 350 iv 60 ml
Struffert (2010) Artis dBA 2 8 Static 350 iv 80 ml
Zhang (2013) Artis Zeego 2 8 Static 300 ia-AA 48 ml 50%
Zhang (2012) Artis Zee 2 8 Static iv
Animal studies Ahmed (2009) Artis dBA 2 5 Static 300–375 iv
Beuing (2014) Artis Zeego 7

4.43

2.83

Dynamic 400 iv 42 ml
Bley (2010) Artis dBA 2 5 Static 370 iv 25 ml
Ganguly (2011) Artis dTA 6 4.3 Dynamic 350 ia 40–50 ml 50–100%
Royalty (2013) Artis Zeego 9 2.8 Dynamic 370 iv 28 ml
Yasuda (2012) Artis Zeego 2 5 Static 370

iv

ia-AA

ia-CCA

ia-VA

25 ml

11–15 ml

4.5–15 ml

3–9.9 ml

100%

15–20%

10–30%

10–30%

iv, intravenous; ia, intra-arterial; AA, aortic arch; CCA, common carotid artery; VA, vertebral artery

Table 4.

Scan protocol specifics: FP-CT vs. MRP

Author Scan protocol Contrast protocol
Machine Sweeps Sec/sweep Static/dynamic Conc. (mg/ml) Route Total
Human studies Garcia (2017) Artis dBA 2 8 Static 370 iv 80 ml
Chen (2018) Artis 10 5 Dynamic 370 iv 60 ml
Kamran (2015) Artis dBA 2 8 Static 370 iv 80 ml
Kuriyama (2017) Artis Zee 2 6 Static 300 ia Variable
Struffert (2015) Artis dBA 9 5 Dynamic 350 iv 60 ml

iv, intravenous; ia, intra-arterial; AA, aortic arch

FP-CT vs. CTP

All human FP-CT scans were performed on Siemens (Erlangen, Germany) “Artis” models (dBA, dTA, Zee, or Zeego). All five studies performed two C-arm rotations of 8 s each, the first without contrast (mask run) and the second with contrast (fill run). Images are post-processed to produce colored blood volume maps, mainly using commercially available software (syngo X Workplace/Dyna PBV neuro). Four studies used contrast intravenously and one study used a catheter in the aortic arch for intra-arterial contrast administration [13]. All studies using iv contrast administered 60 or 80 ml with concentrations varying from 350 to 370 mg iodine per ml. Zhang et al. administered 48 ml of 300 mg/ml contrast intra-arterially, diluted by 50% [13]. Radiation exposure was reported in either Gray or Sievert, in dose per frame or total dose. Only Fiorella et al. actually measured radiation, averaging 219 mGy for FP-CT [16]. All other studies mentioned 0.36 μGy/frame or 2.3 mSv in total, as reported by manufacturers.

Although the same machines are used in six animal studies, scanning protocols show considerable differences. Static imaging uses only 2 rotations and provides CBV in ml/100 g. Three studies used multiple sweeps varying in duration between 2.8 and 4.43 s, providing flow-based parameters like CBF, MTT, and TTP as well [19, 21, 22]. As rotational speed improves, the number of total acquired images decreases. While 8-s C-arm rotations yielded 400 images, a rotation of 2.8 s allowed for a total of 133 images. Four studies administered contrast intravenously, two studies intra-arterially and one study did both. Contrast load ranged from 25 to 42 ml for intravenous use and from 5 to 50 ml for intra-arterial use. Two studies described and compared several arterial injection protocols with small variations in dilution of contrast (range 10–67%), location (aortic arch, common carotid artery, or vertebral artery), and infusion rate (range 1–6 ml/s). Two studies reported radiation dose, both at 1.2 μGy/projection.

FP-CT vs. MRP

All five human studies used scanners from the Siemens “Artis” models. Two studies used dynamic imaging with 9 or 10 sweeps of 5 s each, both using 60 ml of undiluted contrast administered intravenously [23, 28]. Two studies reported a protocol of two 8-s sweeps with 80 ml of undiluted iv contrast [24, 26]. One study used two 6-s sweeps and tried several dilutions of intra-arterial contrast. With concentrations ranging from 12 to 28%, a contrast dilution of 20% provided the best images [27]. Four studies reported on radiation dose, all at levels of 0.36 μGy/frame or 4.6 mSv in total.

Study results

Due to the large heterogeneity of included studies, a meta-analysis could not be performed. We therefore will describe the results of the individual studies. Tables 5 and 6 show quantitative results for all human studies.

Table 5.

Study results: FP-CT vs. CTP in humans

Author Population Methods Parameter(s) Results
Diagnostic accuracy
  Fiorella (2014) 56 CVA patients Detection of PBV deficits: any or > 1/3 of vascular territory CTP-CBV vs FD-PBV

Sens 100%

Spec 63–81%

  Zhang (2013) 3 patients: CVA, ICH, Moyamoya IA contrast dose 70%↓ in visual detection of perf. disorders IVCT-CBV vs IAFD-CBV

Sens 100%

Similar results in all patients

Bland-Altman mean differences
  Struffert (2010) 25 CVA patients CBV in 6 ROI’s CTP-CBV vs FD-CBV − 0.077 ± 0.48 ml/100 g
  Zhang (2012) 20 CVA patients CBV in 6 ROI’s CTP-CBV vs FDCT-CBV − 0.25 ± 2.79 ml/100 g
Correlation coefficients
  Struffert (2010) 25 CVA patients CBV in 6 ROI’s CTP-CBV vs FD-CBV r = 0.79
  Zhang (2012) 20 CVA patients CBV in 6 ROI’s

CTP-CBV vs FDCT-CBV

CTP-rCBV vs FDCT-rCBV

Pearson r = 0.68

Pearson r = 0.76

  Struffert (2012) 16 CVA patients Post-MCA treatment CBV deficits CTP-CBV vs FP-CT-CBV r = 0.90

CVA, cerebrovascular accident; ICH, intracranial hemorrhage

Table 6.

Study results: FP-CT vs. MRP in humans

Author Population Methods Parameter(s) Results
Bland-Altman mean differences
  Kamran (2015) 26 patients: DCI after SAH

Cortical VOI’s

Subcortical VOI’s

CCT-rPBV vs weighted

MR-rCBF/rCBV

0.015 ± 0.119 ml/100 g

0.043 ± 0.118 ml/100 g

Correlation coefficients
  Chen (2018) 13 patients: carotid stenosis MCA territory ROI’s

FD-CTP-rCBF vs MRP-rCBF

FD-CTP-rMTT vs MRP-rMTT

FD-CTP-rTTP vs MRP-rTTP

r = 0.73

r = 0.42

r = 0.58

  Kamran (2015) 26 patients: DCI after SAH

Cortical VOI’s

Subcortical VOI’s

CCT-rPBV vs weighted

MR-rCBF/rCBV

r = 0.91

r = 0.88

  Struffert (2015) 12 CVA patients Qualitative ASPECTS score

CBV

CBF, MTT, and TTP

Pearson r = 0.49

Pearson r = 0.95–0.98

  Garcia (2017) 5 AVM patients Perinidal perfusion

FPD-CT vs DSC-CBV

FPD-CT vs DSC-CBF

FPD-CT vs ASL-CBF

Pearson r = 0.36

Pearson r = 0.47

Pearson r = 0.60

DCI, delayed cerebral ischemia; SAH, subarachnoid hemorrhage; AVM, arteriovenous malformation

FP-CT vs. CTP

For the comparison of FP-CT versus CTP, two human studies reported on diagnostic accuracy. The largest prospective study by Fiorella et al. included 56 patients with ischemic CVA. The study reported a sensitivity and specificity of FP-parenchymal blood volume (PBV) images of 100 and 63% for the detection of CBV deficits larger than one-third of a vascular territory, whereas the sensitivity and specificity to detect any CBV deficit were 100 and 81% [16]. Zhang et al. showed no loss of diagnostic accuracy in 3 out of 12 patients when comparing intra-arterial FP-CT with intravenous CTP in assessing CBV perfusion deficits, while effectively reducing contrast load by 70% [13]. Other articles only compared perfusion parameters and reported on agreement or correlations between (r)CBV values of both techniques. Bland-Altman mean difference analysis was performed in two studies. Both used 6 regions of interest (ROI) per hemisphere, partially corresponding. Results between FP-CTP versus CTP showed mean differences of − 0.077 ± 0.48 ml/100 g in 25 patients and − 0.25 ± −2.79 ml/100 g in 20 patients [14, 15]. Three studies correlated measured perfusion values between FP-CT and CTP. Struffert et al. showed a correlation coefficient of 0.79 in absolute CBV values for 6 ROI’s and Zhang et al. showed moderate correlations in 25 ROI’s, ranging from 0.68 in CBV values to 0.76 in rCBV values [14, 15]. In another study, Struffert et al. correlated CBV deficits measured after recanalization therapy in the angiography suite. Post-treatment FP-CT-CBV deficit volumes showed good correlations with follow-up CT (r = 0.9, p < 0.05) and multi-slice CT (r = 0.9, p < 0.05) [12]. Zhang et al. compared the symmetry of contrast distribution in healthy brains of intravenous CTP-rCBV with intra-arterial FP-CT-rCBV and measured similar values in all regions, with total values of 1.01 ± 0.14 for ivCTP and 0.94 ± 0.18 for iaFP-CT (1.0 for perfect symmetry) [13].

Three animal studies compared diagnostic accuracy between techniques. Beuing et al. reported an average sensitivity and specificity of 94% and 83% in the detection of large vessel occlusions in 7 sheep by two observers (Cohen kappa = 0.86) [21]. Bley et al. studied 50 canine CBV maps and compared FP-CT with CTP in detecting MR-DWI confirmed infarctions. Three observers showed average true positive (TP) readings of 84% for FP-CT and 63% for CTP and false negative (FN) readings of 12% for FP-CT and 25% for CTP [17]. Yasuda et al. compared FP-CT with CTP using three different contrast injection protocols and reported similar TP and FN rates from two observers, albeit in a mere 6 animals [19]. Royalty et al. measured interobserver variability in detecting perfusion lesions on 24 CBV and CBF maps, showing a kappa coefficient of 0.74 for CTP and perfect agreement for FP-CT [20]. Other articles mostly performed exploratory analyses regarding the feasibility of FP-CT perfusion measurements, focusing on the development of robust scanning and contrast injection protocols [18, 19, 22].

FP-CT vs. MRP

For the comparison of FP-CT versus MRP, no studies reported diagnostic accuracy. Chen et al. compared FP-CT and MRP (DSC) in seven different perfusion parameters and showed moderate correlations ranging from 0.42 to 0.73 [25]. Struffert et al. [23] performed a qualitative comparison of FP-CT and MRP using the ASPECTS scoring system for perfusion deficits [29], showing a low Pearson correlation between observers of 0.40–0.49 in CBV and high correlations (0.95–0.98) in CBF, MTT, and TTP [23]. Garcia et al. showed variation coefficients for several techniques among which FP-CT (21.20%), DSC-rCBF (21.30%), and ASL-CBF (11.00%). FP-CT datasets of five patients showed moderate to poor correlations ranging from 0.60 (ASL-CBF) to 0.36 (DSC-CBV) [26]. Kamran et al. compared FP-rPBV with weighted perfusion parameters combining MR-CBV and MR-CBF. Correlation coefficients ranging from 0.72 to 0.91 were reported for cortical ROI’s, with mean differences in Bland-Altman plots as low as 0.015 ± 0.119 ml/100 g. Subcortical ROI’s showed correlations from 0.69 up to 0.88 with a mean difference of 0.043 ± 0.118 ml/100 g [24].

Discussion

Summary of evidence

This systematic review includes 11 studies comparing FP-CTP with CTP and 5 studies comparing FP-CTP with MRP. Quality assessment showed low concerns regarding applicability but moderate to high risks of bias, predominantly in the domains of both the index and reference test and the domain of flow and timing. Performing a meta-analysis was not possible due to both the quality of the studies and the large heterogeneity of study populations and designs, the majority being feasibility studies focusing on scanning and contrast protocols. Eleven studies used 2 C-arm sweeps and five studies investigated flow-based dynamic imaging with 6 to 10 gantry sweeps. Most studies (10 human and 4 animal trials) included subjects with cerebrovascular disease and two animal trials studied healthy subjects [18, 30]. Two human studies [13, 16] and three animal studies [17, 19, 21], all comparing FP-CT with CTP, reported on diagnostic accuracy. Although small by design, all five studies reported good to excellent sensitivity and moderate to good specificity for the detection of perfusion deficits in subjects with ischemic stroke. The largest study by Fiorella et al. showed a sensitivity and specificity of 100% and 81% in 56 patients [16], but with high risks of bias. In addition, good correlations [12, 14, 15] and small mean differences [14, 15] have been reported indicating that FP-CT imaging in assessing cerebral perfusion is feasible and provides blood volume images of similar quality as CTP in patients with cerebrovascular disease.

Strengths and limitations

The strength of our systematic review is that we provide an extensive evaluation of available evidence of FP-CT applications in assessing cerebral perfusion while revealing existing knowledge gaps in current literature.

Some limitations should also be discussed. First, most trials were feasibility or pilot studies showing a large heterogeneity of study population and designs. Imaging and contrast injection protocols were still being developed and fine-tuned, making a meaningful comparison nearly impossible. Only 5 out of 16 articles studied the diagnostic accuracy of the techniques compared, answering our main research question. In addition, risks of bias were high or unclear, making the summarized evidence limited at best.

Second, considering their technical and mathematical background, even directly comparing perfusion parameters like CBV between (FP-)CTP and MRP is not feasible. The relation between contrast-concentration and signal intensity is non-linear in MRI and different mathematical models are used to calculate perfusion parameters. Values can differ significantly between measurements, especially when comparing different techniques. Parameters like blood volume in ml/100 g can suggest (non-existing) calibrated measurements and should be interpreted with caution.

Third, half of the included studies were animal studies. Designs of animal trials in general can introduce selection bias or performance bias [31]. In addition, general anesthesia and differences in vascular anatomy like the presence of a rete mirabile in dogs, sheep, and swines may influence flow and distribution of contrast material [21]. Not necessarily of importance when exploring technical feasibility, these factors do make the transferability from animal to humans unclear.

Lastly, only patients or animals with cerebrovascular disease were studied in the included papers, leaving other applications unexplored. FP-CT scanners are mostly available in angiographic suites where many patients are diagnosed and/or treated, providing plenty of opportunity for clinical studies. In addition, cerebrovascular diseases are particularly suitable for exploring perfusion imaging, as large perfusion deficits are easy to detect. Even though FP-CT scanners are more and more common in hybrid OR’s, evidence for other (intra-operative) applications of FP-CT cerebral perfusion imaging is not available yet.

Clinical implications

Current literature is of insufficient quality to fully support the use of FP-CT instead of CTP or MRP in cerebral perfusion imaging. Studies so far do indicate comparable preliminary results in specific applications and some recommendations for daily clinical practice can be made. FP-CT provides similar values and reconstructed blood volume maps as CTP in cerebrovascular diseases [12, 32]. Sensitivity for perfusion lesions was high in all studies, but specificity was inferior to CTP and MRP. Multiple studies indicated a slight overestimation of perfusion deficits, which has to be taken into account when using FP-CT. Royalty et al. measured CBV and CBF with both techniques and FP-CT showed values two or three times as high as CTP [20]. Fiorella et al. showed a slight overestimation of CBV lesions with all sensitivities at 100% while specificities dropped to 56% [16]. As clinicians assess perfusion deficits in patients with ischemic stroke, a potential fixed bias and possible outliers must be anticipated in order to prevent false positives.

Some technical considerations can be added as well. In accordance with the assumptions of the indicator dilution theory, intra-arterial administration of a short and concentrated bolus of contrast material can reduce contrast load while maintaining image quality [13, 19, 27]. While most authors used the aortic arch, contrast administered through the vertebral artery does not distribute evenly, making it unfit for whole brain imaging [19]. Depending on administration route, the scanning delay after contrast administration differed from 8 to 19 s in the included studies. Several methods were used to establish the optimal moment to start the scan, number one being maximal opacification of venous structures on fluoroscopic images. Reaching a steady state of contrast with optimal scanning delay times could improve image quality and provide the most reliable results.

High spatial resolution is one of the strengths of FP-CT, as reflected in applications implemented so far [33]. High spatial resolution in cerebral FP-CT imaging enables clear visualization of small vessels and reduces partial volume effects. However, the main technical limitation of present-day FP-CT scanners is low temporal resolution. Temporal sampling > 0.5–3 s can result in unreliable quantification of dynamic perfusion parameters in patients with ischemic stroke [3437]. The two 8-s sweeps covering ~ 200° used in human studies are insufficient to reliably quantify CBF and MTT [16, 26]. New developments in dynamic imaging with multiple faster sweeps demonstrate the feasibility of flow-based perfusion imaging using FP-CT, although initial results show moderate correlations and qualitative outcomes [23, 25].

One study measured radiation exposure for both techniques and saw similar values of 219 mGy for FP-CT and 204 mGy for CTP [16]. All other studies reported doses per frame of 0.36 μGy/image as indicated by manufacturers. Most imaging protocols reach total effective doses of > 2 mSv per acquisition, which is considerable, especially in procedures requiring repeated imaging and younger patients. Radiation exposure must be closely monitored and should always be proportioned to clinical benefits to ensure safety in future applications, in particular when adding sweeps or images in order to increase temporal resolution.

Since its introduction in the angiographic suite and more recently in the hybrid OR, FP-CT is increasingly used to assess cerebral hemodynamics. While the use of FP-CT seems feasible and provides blood volume images comparable with CTP, additional human studies of adequate size must be conducted to validate the use of FP-CT in assessing cerebral perfusion. Scanning and injection protocols should be standardized, and terminology of techniques and parameters must be used in appropriate and consistent ways. New studies should report on diagnostic accuracy and study populations must be carefully selected with future applications in mind.

Conclusion

Even though FP-CT provides similar values and reconstructed blood volume maps as CTP in patients with cerebrovascular disease, its diagnostic accuracy remains unclear. Additional methodological sound studies are required in order to reliably determine the diagnostic accuracy of FP-CT in assessing cerebral perfusion.

Electronic supplementary material

ESM 1 (17.1KB, docx)

(DOCX 17.0 kb)

Abbreviation

CTP

Computed tomography perfusion

CVA

Cerebrovascular accident

DCE

Dynamic contrast enhanced

DSC

Dynamic susceptibility contrast

FD

Flat-panel detector

FP-CT

Flat-panel computed tomography

MRP

Magnetic resonance perfusion

OR

Operating room

Funding information

This study was funded by a research grant from Siemens Healthineers (Erlangen, Germany) to the Radboud University Medical Center. Funders had no role in conducting the study or the decision to submit for publication.

Compliance with ethical standards

Conflict of interest

MtL is a board member of the Dutch Society of Neuro-Oncology. All other authors declare that they have no conflict of interest.

Ethical approval

All procedures performed in the studies involving human participants were in accordance with the ethical standards of the institutional and national research committee and with the 1964 Helsinki Declaration and its later amendments or comparable ethical standards.

Informed consent

Informed consent was obtained from all individual participants included in the study.

Footnotes

Publisher’s note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Doerfler A, Gölitz P, Engelhorn T, Kloska S, Struffert T. Flat-panel computed tomography (DYNA-CT) in neuroradiology. From High-Resolution Imaging of Implants to One-Stop-Shopping for Acute Stroke. Clin Neuroradiol. 2015;25:291–297. doi: 10.1007/s00062-015-0423-x. [DOI] [PubMed] [Google Scholar]
  • 2.Frolich AM, Buhk JH, Fiehler J, Kemmling A. Voxel-based sensitivity of flat-panel ct for the detection of intracranial hemorrhage: comparison to multi-detector CT. PLoS ONE. 2016;11(11):1–10. doi: 10.1371/journal.pone.0165794. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Struffert T, Richter G, Engelhorn T, Doelken M, Goelitz P, Kalender WA, et al. Visualisation of intracerebral hemorrhage with flat-detector CT compared to multislice CT: Results in 44 cases. Eur Radiol. 2009;19(3):619–625. doi: 10.1007/s00330-008-1183-7. [DOI] [PubMed] [Google Scholar]
  • 4.Lindner T, Ahmeti H, Lübbing I, Helle M, Jansen O, Synowitz M, et al. Intraoperative resection control using arterial spin labeling - proof of concept, reproducibility of data and initial results. NeuroImage Clin. 2017;15(April):136–142. doi: 10.1016/j.nicl.2017.04.021. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Wintermark M, Reichhart M, Cuisenaire O, Maeder P, Thiran JP, Schnyder P, et al. Comparison of admission perfusion computed tomography and qualitative diffusion- and perfusion-weighted magnetic resonance imaging in acute stroke patients. Stroke. 2002;33(8):2025–2031. doi: 10.1161/01.STR.0000023579.61630.AC. [DOI] [PubMed] [Google Scholar]
  • 6.Connor SEJ, Flis C. The contribution of high-resolution multiplanar reformats of the skull base to the detection of skull-base fractures. Clin Radiol. 2005;60(8):878–885. doi: 10.1016/j.crad.2005.04.003. [DOI] [PubMed] [Google Scholar]
  • 7.Grigoryan M, Tung CE, Albers GW. Role of diffusion and perfusion MRI in selecting patients for reperfusion therapies. Neuroimaging Clin N Am. 2011;21:247–257. doi: 10.1016/j.nic.2011.01.002. [DOI] [PubMed] [Google Scholar]
  • 8.Fouke SJ, Benzinger T, Gibson D, Ryken TC, Kalkanis SN, Olson JJ. The role of imaging in the management of adults with diffuse low grade glioma: a systematic review and evidence-based clinical practice guideline. J Neuro-Oncol. 2015;125:457–479. doi: 10.1007/s11060-015-1908-9. [DOI] [PubMed] [Google Scholar]
  • 9.Provenzale J.M., Shah K., Patel U., McCrory D.C. Systematic Review of CT and MR Perfusion Imaging for Assessment of Acute Cerebrovascular Disease. American Journal of Neuroradiology. 2008;29(8):1476–1482. doi: 10.3174/ajnr.A1161. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Liberati Alessandro, Altman Douglas G., Tetzlaff Jennifer, Mulrow Cynthia, Gøtzsche Peter C., Ioannidis John P.A., Clarke Mike, Devereaux P.J., Kleijnen Jos, Moher David. The PRISMA statement for reporting systematic reviews and meta-analyses of studies that evaluate health care interventions: explanation and elaboration. Journal of Clinical Epidemiology. 2009;62(10):e1–e34. doi: 10.1016/j.jclinepi.2009.06.006. [DOI] [PubMed] [Google Scholar]
  • 11.Whiting PF, Rutjes AWS, Westwood ME, Mallett S, Deeks JJ, Reitsma JB, et al. Quadas-2: a revised tool for the quality assessment of diagnostic accuracy studies. Ann Intern Med. 2011;155:529–536. doi: 10.7326/0003-4819-155-8-201110180-00009. [DOI] [PubMed] [Google Scholar]
  • 12.Struffert T, Deuerling-Zheng Y, Engelhorn T, Kloska S, Golitz P, Kohrmann M, et al. Feasibility of cerebral blood volume mapping by flat-panel detector CT in the angiography suite: first experience in patients with acute middle cerebral artery occlusions. AJNR Am J Neuroradiol. 2012;33(4 PG-618-25):618–625. doi: 10.3174/ajnr.A2839. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Zhang Q, Wang B, Han J, Zhang Z, Sun Q, Beilner J, et al. Feasibility study of perfusion imaging using flat detector CT with an intra-arterial injection protocol compared to conventional multi-slice perfusion CT with an intravenous injection protocol. Interv Neuroradiol. 2013;19(4 PG-409-15):409–415. doi: 10.1177/159101991301900402. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Zhang YW, Hong B, Huang QH, Xu Y, Zhao WY, Yang ZG, et al. The feasibility and clinical application of flat-panel detector computer tomography in evaluating cerebral blood volume: An initial prospective study. Vol. 21. J Interv Radiol. 2012;21(4):265–269. [Google Scholar]
  • 15.Struffert T, Deuerling-Zheng Y, Kloska S, Engelhorn T, Boese J, Zellerhoff M, et al. Cerebral blood volume imaging by flat detector computed tomography in comparison to conventional multislice perfusion CT. Eur Radiol. 2011;21(4 PG-882-9):882–889. doi: 10.1007/s00330-010-1957-6. [DOI] [PubMed] [Google Scholar]
  • 16.Fiorella D, Turk A, Chaudry I, Turner R, Dunkin J, Roque C, et al. A prospective, multicenter pilot study investigating the utility of flat detector derived parenchymal blood volume maps to estimate cerebral blood volume in stroke patients. J NeuroInterventional Surg. 2014;6(6):451–456. doi: 10.1136/neurintsurg-2013-010840. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Bley T, Strother CM, Pulfer K, Royalty K, Zellerhoff M, Deuerling-Zheng Y, et al. C-arm CT measurement of cerebral blood volume in ischemic stroke: An experimental study in canines. Am J Neuroradiol. 2010;31(3):536–540. doi: 10.3174/ajnr.A1851. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Ahmed AS, Zellerhoff M, Strother CM, Pulfer KA, Redel T, Deuerling-Zheng Y, et al. C-arm CT measurement of cerebral blood volume: an experimental study in canines. AJNR Am J Neuroradiol. 2009;30(5 PG-917-22):917–922. doi: 10.3174/ajnr.A1513. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Yasuda R, Royalty K, Pulfer K, Consigny D, Strother CM. C-arm CT measurement of cerebral blood volume using intra-arterial injection of contrast medium: an experimental study in canines. AJNR Am J Neuroradiol. 2012;33(9 PG-1696-701):1696–1701. doi: 10.3174/ajnr.A3077. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Royalty K, Manhart M, Pulfer K, Deuerling-Zheng Y, Strother C, Fieselmann A, et al. C-arm CT measurement of cerebral blood volume and cerebral blood flow using a novel high-speed acquisition and a single intravenous contrast injection. AJNR Am J Neuroradiol. 2013;34(11 PG-2131-8):2131–2138. doi: 10.3174/ajnr.A3536. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Beuing O, Boese A, Kyriakou Y, Deuerling-Zengh Y, Jollenbeck B, Scherlach C, et al. A novel technique for the measurement of CBF and CBV with robot-arm-mounted flat panel CT in a large-animal model. AJNR Am J Neuroradiol. 2014;35(9 PG-1740-5):1740–1745. doi: 10.3174/ajnr.A3973. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Ganguly A, Fieselmann A, Marks M, Rosenberg J, Boese J, Deuerling-Zheng Y, et al. Cerebral CT perfusion using an interventional C-arm imaging system: cerebral blood flow measurements. AJNR Am J Neuroradiol. 2011;32(8 PG-1525-31):1525–1531. doi: 10.3174/ajnr.A2518. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Struffert T, Deuerling-Zheng Y, Kloska S, Engelhorn T, Lang S, Mennecke A, et al. Dynamic angiography and perfusion imaging using flat detector ct in the angiography suite: a pilot study in patients with acute middle cerebral artery occlusions. Am J Neuroradiol. 2015;36(10):1964–1970. doi: 10.3174/ajnr.A4383. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Kamran M, Byrne JV. C-arm flat detector computed tomography parenchymal blood volume imaging: the nature of parenchymal blood volume parameter and the feasibility of parenchymal blood volume imaging in aneurysmal subarachnoid haemorrhage patients. Vol. 57. Neuroradiology. 2015;57(9):937–949. doi: 10.1007/s00234-015-1545-1. [DOI] [PubMed] [Google Scholar]
  • 25.Chen LW, Lin CJ, Guo WY, Hung SC, Lee HJ, Chen KK, et al. Evaluating cerebral hemodynamics using quantitative digital subtraction angiography and flat-detector computed tomography perfusion imaging: a comparative study in patients with carotid stenosis. J Chin Med Assoc. 2018;81(11):977–984. doi: 10.1016/j.jcma.2018.01.016. [DOI] [PubMed] [Google Scholar]
  • 26.Garcia M, Okell TW, Gloor M, Chappell MA, Jezzard P, Bieri O, et al. Feasibility of flat panel detector CT in perfusion assessment of brain arteriovenous malformations: initial clinical experience. Am J Neuroradiol. 2017;38(4):735–739. doi: 10.3174/ajnr.A5091. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Kuriyama T, Sakai N, Beppu M, Sakai C, Imamura H, Kojima I, et al. Optimal dilution of contrast medium for quantitating parenchymal blood volume using a flat-panel detector. J Int Med Res. 2018;46(1):464–474. doi: 10.1177/0300060517715165. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Lin C-J, Guo W-Y, Chang F-C, Hung S-C, Chen K-K, Yu D-Z, et al. Using flat-panel perfusion imaging to measure cerebral hemodynamics. Medicine (Baltimore) 2016;95(20):e3529. doi: 10.1097/MD.0000000000003529. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Barber PA, Demchuk AM, Zhang J, Buchan AM. Validity and reliability of a quantitative computed tomography score in predicting outcome of hyperacute stroke before thrombolytic therapy. ASPECTS Study Group. Alberta Stroke Programme Early CT Score. Lancet. 2000;355(9216):1670–1674. doi: 10.1016/S0140-6736(00)02237-6. [DOI] [PubMed] [Google Scholar]
  • 30.Fieselmann A, Ganguly A, Deuerling-Zheng Y, Zellerhoff M, Rohkohl C, Boese J, et al. Interventional 4-D C-arm CT perfusion imaging using interleaved scanning and partial reconstruction interpolation. IEEE Trans Med Imaging. 2012;31(4 PG-892-906):892–906. doi: 10.1109/TMI.2011.2181531. [DOI] [PubMed] [Google Scholar]
  • 31.Hooijmans CR, Rovers MM, Vries RBMD, Leenaars M, Ritskes-hoitinga M, Langendam MW. SYRCLE’ s risk of bias tool for animal studies. BMC Med Res Methodol. 2014;14(1):1–9. doi: 10.1186/1471-2288-14-43. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Struffert T, Deuerling-Zheng Y, Kloska S, Engelhorn T, Strother CM, Kalender WA, et al. Flat detector CT in the evaluation of brain parenchyma, intracranial vasculature, and cerebral blood volume: a pilot study in patients with acute symptoms of cerebral ischemia. AJNR Am J Neuroradiol. 2010;31(8 PG-1462-9):1462–1469. doi: 10.3174/ajnr.A2083. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Gupta R, Cheung AC, Bartling SH, Lisauskas J, Grasruck M, Leidecker C, et al. Flat-panel volume CT: fundamental principles, technology, and applications. Radiogr A Rev Publ Radiol Soc North Am Inc. 2008;28(7):2009–2022. doi: 10.1148/rg.287085004. [DOI] [PubMed] [Google Scholar]
  • 34.Abels B, Klotz E, Tomandl BF, Villablanca JP, Kloska SP, Lell MM. CT perfusion in acute ischemic stroke: a comparison of 2-second and 1-second temporal resolution. Am J Neuroradiol. 2011;32(9):1632–1639. doi: 10.3174/ajnr.A2576. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Kämena A, Streitparth F, Grieser C, Lehmkuhl L, Jamil B, Wojtal K, et al. Dynamic perfusion CT: optimizing the temporal resolution for the calculation of perfusion CT parameters in stroke patients. Eur J Radiol. 2007;64(1):111–118. doi: 10.1016/j.ejrad.2007.02.025. [DOI] [PubMed] [Google Scholar]
  • 36.Wiesmann M, Berg S, Bohner G, Klingebiel R, Schöpf V, Stoeckelhuber BM, et al. Dose reduction in dynamic perfusion CT of the brain: effects of the scan frequency on measurements of cerebral blood flow, cerebral blood volume, and mean transit time. Eur Radiol. 2008;18(12):2967–2974. doi: 10.1007/s00330-008-1083-x. [DOI] [PubMed] [Google Scholar]
  • 37.Wintermark M, Smith WS, Ko NU, Quist M, Schnyder P, Dillon WP. Dynamic perfusion CT: optimizing the temporal resolution and contrast volume for calculation of perfusion CT parameters in stroke patients. Am J Neuroradiol. 2004;25(5):720–729. [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

ESM 1 (17.1KB, docx)

(DOCX 17.0 kb)


Articles from Neuroradiology are provided here courtesy of Springer

RESOURCES