Abstract
Purpose
In vivo 1H and 31P magnetic resonance spectroscopic imaging (MRSI) provide complementary information on the biology of prostate cancer. In this work we demonstrate the feasibility of performing multiparametric imaging (mpMRI) and 1H and 31P spectroscopic imaging of the prostate using a 31P and 1H endorectal radiofrequency coil (ERC) in combination with a multitransmit body array at 7 Tesla (T).
Methods
An ERC with a 31P transceiver loop coil and 1H receive (Rx) asymmetric microstrip (31P/1H ERC) was designed, constructed and tested in combination with an external 8‐channel 1H transceiver body array coil (8CH). Electromagnetic field simulations and measurements and in vivo temperature measurements of the ERC were performed for safety validation. In addition, the signal‐to‐noise (SNR) benefit of the 1H microstrip with respect to the 8CH was evaluated. Finally, the feasibility of the setup was tested in one volunteer and three patients with prostate cancer by performing T2‐weighted and diffusion‐weighted imaging in combination with 1H and 31P spectroscopic imaging.
Results
Electromagnetic field simulations of the 31P loop coil showed no differences in the E‐ and B‐fields of the 31P/1H ERC compared with a previously safety validated ERC without 1H microstrip. The hotspot of the specific absorption rate (SAR) at the feed point of the 31P/1H ERC loop coil was 9.42 W/kg when transmitting on 31P at 1 W. Additional in vivo measurements showed a maximum temperature increase at the SAR hotspot of 0.7°C over 6 min on 31P at 1.9 W transmit (Tx) power, indicating safe maximum power levels. When transmitting with the external 1H body array at 40W for 2:30 min, the temperature increase around the ERC was < 0.3°C. Up to 3.5 cm into the prostate the 1H microstrip of the ERC provided higher SNR than the 8CH. The total coil combination allowed acquisition of an mpMRI protocol and the assessment of 31P and 1H metabolites of the prostate in all test subjects.
Conclusion
We developed a setup with a 31P transceiver and 1H Rx endorectal coil in combination with an 8‐channel transceiver external body array coil and demonstrated its safety and feasibility for obtaining multiparametric imaging and 1H and 31P MRSI at 7T in patients with prostate cancer within one MR examination.
Keywords: 1H, 31P, endorectal coil, prostate cancer, spectroscopy
1. Introduction
Proton magnetic resonance spectroscopic imaging (MRSI) is used to study the in vivo metabolic profile of the prostate by detecting and localizing the levels of citrate, choline, creatine and polyamines.1, 2, 3, 4 An increase in the ratio of choline over citrate, which is partly attributed to increased levels of choline, has been described as a valuable biomarker to discriminate prostate cancer from healthy prostate tissue.5 It indicates prostate cancer specific changes in the phospholipid metabolism which plays an important role in cell membrane generation and degradation.5, 6, 7, 8 This phospholipid metabolism can be studied in more detail when using 31P‐MRSI to assess the levels of the phosphomonoesters (PMEs): phosphocholine (PC) and phosphoethanolamine (PE), and the phosphodiesters (PDEs): glycerophosphocholine (GPC) and glycerophosphoethanolamine (GPE).9, 10, 11 Parts of the PME and PDE signals, together with resonances of free choline, free ethanolamine, taurine, and myo‐inositol, comprise the total choline (tCho) peak at 3.2 ppm in 1H spectroscopy.11 Since 31P and 1H spectroscopy of the prostate provide complementary information on tumor biology, it would be of high interest to obtain both within one measurement session. Ideally, this would even be combined with T2‐weighted (T2W) and diffusion‐weighted imaging (DWI) of the prostate, as these are paramount in accurate localization and characterization of potential prostate cancer lesions.
The combination of 1H and 31P spectroscopy and 1H imaging requires a radiofrequency (RF) coil setup with transmit (Tx) and receive (Rx) capabilities on both nuclei,12 preferably one in which no coils need to be exchanged during the examination.
Because of the lower intrinsic magnetic resonance (MR) sensitivity of 31P compared to 1H and the low concentration of 31P metabolites in the human body, 31P spectroscopy of the prostate is challenging. 31P MRSI should preferably be performed at 7 Tesla with an endorectal radiofrequency coil (ERC) to optimize signal‐to‐noise (SNR) and to increase spectral resolution. However, for T2W imaging and DWI of the prostate at 7T, only using an endorectal coil is inadequate. Transmitting radiofrequency pulses with a small local surface coil results in substantial transition bands in T2W and DWI images due to its severely inhomogeneous transmit profile.12, 13 This can be solved by using a multichannel external body array in combination with B1 + shimming to provide a homogeneous transmit field.13, 14 Since an endorectal coil is needed for 31P, this ERC can also be equipped with 1H receive capabilities to enhance 1H imaging and spectroscopy of the prostate, increasing SNR for 1H spectroscopy and spatial resolution for T2W imaging and DWI. Moreover, the 1H transmit capabilities of the external body array allow the use of the nuclear Overhauser effect for an extra increase in 31P SNR.15
Therefore, we introduce and evaluate a coil setup with a 31P Tx/Rx – 1H Rx ERC in combination with an 8‐channel external multitransmit 1H array to enable multiparametric imaging and 1H and 31P spectroscopic imaging of the prostate at 7T. We perform electromagnetic field simulations, phantom measurements and in vivo temperature measurements to validate the safety of the setup. In addition, we evaluate its SNR performance and demonstrate the feasibility of the complete setup in three patients and a volunteer.
2. Methods
2.1. Subjects
One healthy volunteer (age 40 yr, weight 83 kg) and three patients (age: 60, 63, and 67 yr, weight: 97, 83, and 88 kg) with histopathologically proven prostate cancer were measured using the setup described below. An intramuscular injection of butylscopolamine bromide (Buscopan, Boehringer‐Ingelheim, Ingelheim, Germany) was used to suppress peristalsis. For all volunteer and patient measurements, informed consent was signed and the study was approved by the institutional review board.
2.2. Hardware
All MR measurements were performed on a 7‐Tesla whole‐body MRI system (MAGNETOM, Siemens Healthcare, Erlangen, Germany). Specific absorption rate (SAR) monitoring of the 1H and 31P measurements was performed with a custom build SAR supervision system, which measured the time‐averaged input power over 10 s and over 6 min. This input power was not to exceed the safety levels based on local SAR restrictions, otherwise the safety system would stop the measurement.
2.2.1. Combined 31P Tx/Rx and 1H Rx endorectal coil ( 31P/ 1H ERC)
The mechanical housing and conductors of a 3T endorectal 1H receive (Rx) coil (MEDRAD, Pittsburgh, PA) were disassembled and equipped with a 1H receive element and a 31P transceiver loop coil. The 1H Rx element was placed directly on the plastic rod16 and consisted of an asymmetric microstrip of length 53 mm, with a distance of 4 mm between the conducting line and the ground plane [Fig. 1(a)]. By rotating the conducting line with respect to the axis of the plastic rod, the B1 ‐ profile was oriented in the same direction as the receive profile of the 31P loop coil. The coil was tuned and matched to 297.2 MHz and a series PIN diode was used for detuning the microstrip during transmit with the external 1H coil. The loop coil was tuned and matched to 50 Ω at 120.3 MHz for 31P at 7T and was mounted on top of the inner balloon (Fig. 1).10 In every patient measurement, after the final coil placement, the tuning and matching was checked using a network analyzer, as it slightly depends on the inflation of the inner balloon. The inner balloon was inflated using inert perfluoro‐polyether liquid (Fomblin, Solvay Solexis, Bollate, Italy). To eliminate any common modes a 1H cable trap was placed directly at the cable exit of the plastic rod. Finally, a balun was added to the 1H Rx element at the final part of the 1H cabling in the plastic rod (Fig. 2). The 31P transceiver loop was connected via a coaxial cable to a Tx/Rx switch with integrated small‐band low‐noise preamplifier (Ar2 Communication Products, Burlington US). The received signals from the 1H microstrip were routed through a separate coaxial cable to its own interface with integrated 1H preamplifier (Ar2 Communication Products, Burlington US). These electronics were used to interface the elements in the ERC to the MR system.
2.2.2. 1H external array coil
The 1H external multitransmit RF body coil consisted of two arrays containing four microstrip elements with meanders.10, 17, 18 The arrays were put ventrally and dorsally of the body. The coil channels were powered by eight 1 kW amplifiers (LPPA 13080W, Dressler, Germany) and were connected using a separate Tx/Rx switch. The cable length between elements and switch box was adjusted to achieve preamplifier decoupling during reception.
2.2.3. In vivo 1H performance of 31P/ 1H endorectal coil
To evaluate the in vivo performance of the asymmetric microstrip 1H receive element, we performed in vivo SNR measurements within the prostate to compare the SNR of the ERC with the SNR of the 8‐channel external body array (8CH). Gradient echo (GRE) imaging in transversal and sagittal orientation (TR = 25 ms, TE = 4.1 ms, field of view (FOV) = 261 × 380 mm2, matrix = 176 × 256, 3 slices, one average, bandwidth 241 Hz/pixel) of the healthy volunteer was performed with the ERC in situ, using the external body array for transmit and either only the 8‐channel external body array or only the ERC for receive. In the 8CH measurements, the images of the individual coil channels were added using the sum of squares method, after normalizing the individual images to their noise level which was determined within a region of interest (ROI) of air. For both measurements exactly the same measurement parameters were used. The SNR in pixel (x,y) of the ERC () and 8CH ) measurements was determined by:
Here, is an estimate for the noise within the prostate. A region of air was used for this estimation to reduce the influence of physiological noise and variation. was determined by correcting the standard deviation within the air ROI for the Rayleigh distribution in single channel ERC imaging and noncentral chi distribution in sum‐of‐squares multichannel 8CH imaging.19, 20
The ROI placement for noise calculation was the same for both measurements. The ratio of the SNR of the ERC and the external body array was determined by dividing the respective SNRs:
The SNR ratio image was filtered by convoluting the image with a 3 × 3 voxel wide block function.
2.3. Safety testing of the 31P/1H endorectal coil
Before performing any patient measurements, the setup was thoroughly validated for safety. The SAR safety levels of the setup were in part based on the simulations and measurements of a previously validated design by Kobus et al.10 They determined the safety levels of a 31P‐only ERC in combination with the same external body array as used in this study. Here we added a 1H Rx‐only microstrip to their 31P ERC design, which could influence the 31P loop coil performance. Because of the close proximity of the ERC to tissue, this could introduce safety concerns with respect to the maximum allowed 31P transmit power. To demonstrate the safety of the 31P/1H ERC, we performed electromagnetic field simulations of an experimental 31P/1H ERC design which we compared with a design similar to that of Kobus et al.10 In addition, we performed electromagnetic field measurements of this experimental 31P/1H ERC to validate the simulation results. The safety characteristics of the 31P/1H ERC were further investigated by performing SAR simulations in a more realistic prostate model and by performing in vivo temperature measurements.
Additional phantom measurements using a network analyzer to assess coupling between the ERC and the external coil array are presented in the Supporting Information S.1.
2.3.1. Electromagnetic field simulations and phantom measurements
To promote reproducibility, the actual 31P loop coil was replaced by a stiff printed circuit board (PCB) plate with the same dimensions, in both simulations and measurements (Fig. 3). The electromagnetic fields of 31P/1H ERC were analyzed by performing Finite Integration Technique (FIT) simulations (CST Studio Suite 2011, Darmstadt, Germany) at 120.3 MHz of a three‐dimensional (3D) model (Autodesk Inventor, San Rafael, USA) of the 31P/1H endorectal coil and a similar ERC without the 1H channel (31P‐only ERC) resembling the design by Kobus et al.10 In the simulations of the 3D models of the coils, the coils were surrounded by a phantom fluid with a relative permittivity of ϵr = 79 and a conductivity of σ = 0.47 S/m at a frequency of 120.3 MHz. Additional measurements were performed of the 31P/1H ERC using a setup with field probes (Schmid & Partner Engineering AG (SPEAG), Zürich, Switzerland, probes H3DV7 & ES3DV2) to collect E‐ and B‐field measurements in the sagittal plane in a phantom filled with liquid (ϵr = 79, σ = 0.49 S/m, measured with an Agilent 85070E dielectric probe kit at 120 MHz). To stay within the calibration range of the probes, two different power settings were used: 0.01 W for the B‐field and 0.05 W for the E‐field measurements. The measurements were performed at 120 MHz in a 2D plane using a sampling density of one sample per 5 × 5 mm2 area. A sagittal plane of 150 × 30 mm2 was sampled in the midline of the coil, 10–40 mm above the conductors.
2.3.2. SAR simulations
Electromagnetic field simulations (FIT) of the 31P loop coil at 120.3 MHz in a more realistic prostate model were performed to determine the SAR levels at 1 W transmit power. The 31P/1H ERC was compared to the 31P‐only ERC in a model incorporating the different surrounding tissues (Fig. 4). Both were simulated with the actual 31P loop coil model (Fig. 1). The permittivity and electric conductivity values were obtained from literature.21
2.3.3. In vivo temperature measurements
In vivo temperature measurements were performed with the complete setup: the 31P/1H ERC in combination with the external 1H body array. Three thermocouples were attached to the ERC at different locations (Fig. 5). One temperature sensor was placed at the feed point of the coil, the location of the hotspot in previous SAR simulation by Kobus et al,10 and the other two sensors were attached to the cable to ensure no safety compromising common modes were present. The thermocouple at the feeding point was attached to the outer side of the external balloon and a thin condom was used to cover the ERC and thermocouples. The insulating effect of this thin latex layer was expected to be of minimal influence and was therefore neglected. Two different safety tests were performed. The ERC was applied as usual and for the first test a time‐averaged input power of 1.9 W10 was used for a duration of 6 min on the 31P channel. For the second test the external 1H body array was used with a power of 40 W for 2:30 min to test for any coupling between the body array and the ERC.
2.4. Patient measurements
To demonstrate the feasibility of the complete setup, all patient measurements were performed with the external body array coil for 1H transmit only and the 31P/1H ERC for 1H receive and 31P transmit and receive (see Figure S‐2 in Supporting Information S.3 in for a schematic overview of the in vivo setup). After a localizer and B0 shimming, B1 + shimming was performed by using the B1TIAMO technique 22 to acquire absolute B1 + maps and by maximizing the phase coherence of the individual coil channels in an ROI within the prostate. 1H T2‐weighted imaging and diffusion‐weighted imaging were then performed with an in‐plane spatial resolution of respectively 0.43 × 0.43 mm2 and 1.75 × 1.75 mm2.
Proton MRSI was performed using a PRESS‐like sequence with RF refocusing pulses that are both spectrally and spatially selective, such that the signal of a volume of interest (VOI) is received only from the spectral region of interest (2.3–3.3 ppm).23 31P MRSI was performed using a 3D phase‐encoded pulse‐acquire sequence with a nonselective BIR‐4 excitation pulse9 (see Table 1 for further imaging and spectroscopy details). The nuclear Overhauser effect was used with continuous low‐power irradiation of the water resonance to enhance 31P SNR.15 31P MRSI data was fitted using Metabolite Report (Siemens Healthineers Erlangen, Germany). PE, PC, GPE, and GPC were modeled as a triplet of three Gaussian peaks with a separation equal to the J‐coupling (6–7.1 Hz) and amplitudes in a 1:2:1 ratio. The PCr lineshape, which was fitted with a single Gaussian shape, was used as a constraint to the linewidth for all signals.9 The spectra shown in this work were corrected for constant and linear phase, an exponential time filter (40 ms for 31P and 150 ms for 1H) was applied and the spectra were zero‐filled to 2048 points. In the 31P spectra the PCr signal was referenced to 0.0 ppm and in the 1H spectra water was referenced to 4.7 ppm.
Table 1.
T2W | DWI | 31P MRSIb | 1H MRSI | |
---|---|---|---|---|
Sequence type | Turbo‐Spin‐Echo | EPI (RESOLVE); b0, b100, b400, and b800 | 3D MRSI with non‐selective BIR‐4 excitation | PRESS‐like with spectrally and spatially selective RF refocusing pulses |
FOV | 111 × 111 mm2 | 154 × 154 mm2 |
a): 120 × 100 × 100 mm3 b): 120 × 120 × 120 mm3 |
84 × 70 × 70 mm3 |
Matrix | 256 × 256 | 88 × 88 |
a): 12 × 10 × 10 b): 10 × 10 × 10 |
12 × 10 × 10 |
Voxel sizea | 0.43 × 0.43 × 3 mm3 | 1.75 × 1.75 × 3 mm3 |
a): 4.9 cc b): 9.1 cc |
1.4 cc |
Slices | 19 | 19 | n.a. | n.a. |
Slice thickness | 3 | 3 | n.a. | n.a. |
TR (ms) | 4250 | 4400 | 1500 | 1000 |
TE (ms) | 84 | 68 | n.a. | 135 |
Bandwidth (Hz) | 106 (per pixel) | 1015 | 10000 (spectral) | 2400 (spectral) |
Vector size | n.a. | n.a. | 2048 | 1024 |
Averages | 1 | 1 | 4 (weighted) | 1 |
Acquisition time (min:sec) | 1:59 | 4:34 | 13:09 | 7:01 |
Flip angle (⁰) | n.a. | n.a. | 45 | n.a. |
Filter | Prescan normalization filter | n.a. | 100% Hamming filter | 50% Hamming filter |
DWI, diffusion‐weighted imaging; MRSI, magnetic resonance spectroscopic imaging.
For the MRSI measurements, this was based on the true voxel size based on voxel volume within the 50% range of the maximum of the point spread function after correcting for spatial filter and sampling scheme.
Two different measurement parameter sets were used: set a) and set b).
3. Results
The 31P/1H ERC was constructed according to the circuit and model specifications (Figs. 1 and 2) and all validations and in vivo experiments were performed successfully.
Simulations of the B1 − profile of the 1H microstrip were performed and are shown in the Figure S‐1 in Supporting Information S.2.
3.1. Safety
The electromagnetic field simulations to test the safety when transmitting at 31P showed no differences between the 31P‐only ERC by Kobus et al.10 and the 31P/1H ERC proposed in this work (Fig. 3). The hotspot with respect to the electric field in both the 31P‐only and the 31P/1H ERC was identified at the same location at the feed point of the coil. Both the E‐ and B‐field simulations corresponded well with the actual measurements of the 31P/1H ERC, yielding similar E‐ and B‐field peak locations and intensities. The SAR simulations in the prostate tissue model also showed no substantial differences between the two coils. Maximum SAR10g values of 9.48 W/kg for the 31P‐only ERC and 9.42 W/kg for the 31P/1H ERC were found at the feed point of the coil when transmitting with the 31P loop coil at 1 W (Fig. 4).
When transmitting on 31P at 1.9 W power, the in vivo temperature measurements showed a maximum temperature increase of approximately 0.7°C over 6 min at the feed point of the coil, converging to a plateau in the second half of the 6‐min period. The maximum temperature increase measured at the sensors located on the cable was < 0.5°C. When transmitting with the external 1H body array at 40 W for 2:30 min, the temperature increase was less than 0.3°C in any of the temperature sensors, indicating that there is no significant coupling between the external array and the ERC (Fig. 5).
There was no evidence for any significant coupling between the 1H ERC channel, the 31P ERC channel and the external body array in the phantom measurements with the network analyzer. There was also no influence when manipulating the cables (Table S‐1 in Supporting Information S.1).
3.2. Volunteer and patient measurements
Using the external body array for 1H transmit resulted in a homogeneous flip angle distribution over an ROI in the prostate of the volunteer of 78.5 ± 3.9° (Figure S‐2 in Supporting Information S.3). In combination with the high local sensitivity of the ERC’s 1H Rx channel, this allowed high resolution T2W and DW imaging without artifacts (Fig. 7, Figure S‐3 in Supporting Information S.4). The ERC’s 1H asymmetric microstrip showed improved SNR in the prostate of the healthy volunteer when compared with the external array, up to 3.5 cm anterior to the ERC at the center of the microstrip and 3 cm when 20 mm superior or inferior to the center of the microstrip [(Fig. 6(d) and 6(f)]. In the midline through the ERC, 1 cm into the prostate, the SNR (mean ± standard error of the mean) over the center ten voxels was 316 ± 7 for the ERC microstrip and 59 ± 0.6 for the 8CH array. Due to the asymmetric microstrip design, the receive profile is symmetric with respect to and oriented toward the prostate [Fig. 6(c)], (Figure S‐1 in Supporting Information S.2). Thirty millimeters anterior to the ERC, the ERC had a higher SNR than the 8CH within a range of ≥ 30 mm in the left‐right direction [Fig. 6c and 6(e)].
We were able to detect prostate cancer lesions in the three patients with the imaging sequences, such that voxels with 1H and 31P spectra could be localized within these lesions. Prostate cancer lesions were verified by MR‐guided biopsy (one patient), MR‐TRUS guided biopsy (one patient) or TRUS biopsy and 3T clinical MRI results (one patient). The ERC’s sensitivity profile was wide enough to perform T2W and DW imaging of the seminal vesicles, apex of the prostate, and anterior regions of the prostate (Fig. 7), even in a relatively large prostate (Figure S‐3 in Supporting Information S.4). Due to the overlap between the 31P and 1H sensitivity profiles, 31P spectroscopy of those regions can be obtained as well with sufficient SNR to measure PE and PC in a healthy volunteer or patient (Fig. 7 , Fig. 8). The 1H spectral map shows adequate SNR for citrate and spermine from apex to base in the prostate. The line widths achieved after B0 shimming within a VOI around the prostate ranged from 30–60 Hz. Close to the ERC and in the seminal vesicles the spectral map shows wider peaks because those areas are generally harder to B0 shim properly. This was less noticeable in the 31P spectra due to the larger ppm difference of the 31P signals. For both 31P and 1H spectroscopy the signal drops anteriorly in prostate.
31P spectroscopic imaging was obtained successfully in the volunteer and in all patients. PE, PC, PCr, and γ‐ and αNTP were detectable in spectra that were located within the prostate. Similar to previous results9 the PDEs were occasionally within detectable ranges, and if so, they were well separable (Fig. 8). 1H spectroscopy showed well distinguishable tCho and Spm signals and also showed Cit which occasionally suffered from lipid contamination. High spermine signals were present in 1H spectra (Fig. 8, Figure S‐3 in Supporting Information S.4).
In one of the patients with a high‐grade Gleason 4 + 4 prostate cancer lesion the GPC, GPE, and PC levels were clearly identifiable and elevated in the spectra located within the cancer lesion, whereas these signals were almost absent in contralateral healthy prostate tissue (Fig. 8). PCr, which is mostly present in muscle tissue, could serve as an indicator for the 31P coil sensitivity profile due to its assumed symmetric distribution over the prostate and nearby smooth muscles. As PCr was highest on the healthy side of the prostate, the GPE, GPC, and PC elevations in the lesion were not caused by any asymmetry in the coil profile, but were due to real locally increased metabolite levels. The increase in 31P metabolites in tumor tissue corresponded with an elevated 1H tCho peak, which was absent on the contralateral side of the prostate. Other patients did not show detectable GPC or GPE within the cancer lesion (Figure S‐4 in Supporting Information S.4).
Please refer to the Supporting Information S.4 for additional patient measurements.
4. Discussion
31P spectroscopic imaging of the prostate is only possible with an ERC at an ultra‐high field strength. When adding a 1H receive channel to the ERC and combining it with an external body array coil for 1H transmit, 31P spectroscopic imaging can be performed within the same MR examination as high‐resolution prostate imaging and 1H spectroscopic imaging.
4.1. Safety
To validate the safety of our setup, we used the results of previous work by Kobus et al10 with a 31P Tx/Rx ERC combined with an external body coil. Their work showed no substantial coupling between the ERC and the external body array. Furthermore, when transmitting at 31P with a power of 1.9 W their simulations stayed within the SAR10g limit of the first level operation mode in the hotspot at the feed of the coil.
In our proposed 31P Tx/Rx – 1H Rx endorectal coil, we added a 1H receive microstrip to the 31P Tx/Rx ERC by Kobus et al.10 This addition may pose safety concerns for the 31P transmit safety levels. We performed electromagnetic field and SAR simulations of both the 31P‐only ERC by Kobus et al and our proposed design with an additional 1H microstrip and found no influence of the 1H microstrip on the simulation results or local hotspot locations. We also validated these results with experimental phantom measurements. These results indicate that we can use the SAR safety levels of Kobus et al10 for our design and that our design stays within the local SAR10g limit of first level operation mode when transmitting at 1.9 W on 31P. Moreover, new in vivo temperature measurements of the proposed 31P/1H ERC also showed a temperature increase < 1°C in the local hotspot when transmitting at this power level, again indicating safe power levels.
4.2. In vivo measurements
Kobus et al10 and Lagemaat et al9 used a measurement setup with an external body array in combination with a 31P‐only ERC, which allowed them to obtain imaging and 31P spectroscopy. No 1H spectroscopic imaging was performed, as this is challenging without a local receive coil due to SNR restrictions. In this work we equipped the ERC with a 1H Rx element to substantially increase local receive sensitivity. Not only did this facilitate the acquisition of 1H MRSI, it also enabled high resolution T2W imaging and DWI with a zoomed field‐of‐view of the prostate. The imaging resolution was substantially higher than the reported spatial resolution of 0.75 × 0.75 × 3 mm3 in earlier work at 7T without an ERC24 and similar to the resolution of 0.4 × 0.4 × 3 mm3 in earlier work at 3T with a dedicated 1H ERC.25 This resolution is possible due to the SNR increase that the ERC 1H microstrip offered in the prostate. Up to 3.5 cm anterior to the ERC, the ERC has a higher SNR than the 8‐channel external body array, which is comparable to previously published work on 1H imaging at 7T.26 When combined with the external body array for receive, the ERC would provide additional performance up to about 5 cm in the prostate (when taking the − 0.5 logarithmic SNR ratio criterium for added benefit in Fig. 6). The SNR measurements were performed in one healthy volunteer of average body composition and we corrected for differences in noise distribution between the reconstruction algorithms used, to perform a fair comparison. The SNR gain of the ERC with respect to a multichannel array surely also depends upon the exact body composition: in larger body circumferences the elements of the external coil array are further away from the prostate, decreasing their SNR within the prostate, increasing the relative SNR contribution from the ERC over a larger area. The opposite holds for smaller body compositions. The overlap in the 31P and 1H sensitivity profile of the ERC allowed combining 1H with 31P MRSI in the same patient, enabling us to correlate tCho signals to individual phosphorus containing choline signals. Because of the SNR constraints of spectroscopy, the ERC might not be optimal for assessing the anterior regions of (very) large prostates, as the sensitivity profile of an ERC drops significantly at large distances.
Because of the increased SNR due to the use of 7T MRI in combination with the nuclear Overhauser effect, 31P spectroscopic imaging was obtained with high quality spectra in all patients, using a spectroscopic imaging method to accurately localize signals. Although the true voxel size was 4.9 or 9.1 cc, it allows voxel localization with respect to the T2W and DWI images within or covering suspicious lesions, instead of relying on coil profile‐based localization with an unlocalized pulse sequence. Using a 31P/1H ERC without external multitransmit array coil also enables the acquisition of 31P and 1H MRSI. However, one would need to revert to the use of many adiabatic RF refocusing pulses and T2‐weighted MRI and DWI becomes particularly problematic with the use of only a small endorectal coil 12.
1H MRSI was also obtained with a high SNR and with well distinguishable tCho and Spm signals, but it suffered from lipid contamination in some cases. This is most likely caused by B0 inhomogeneities, the small chemical shift separation between the citrate and lipid signals and the spectral selectivity of the spectral‐spatial pulses of the 1H spectroscopy sequence. 1H MRSI was performed with a corrected voxel size of 1.4 cc, enabling voxel localization within or covering lesions. The spectral spatial refocusing pulses selectively inverted the 3.1 ppm resonance of Spm, refocusing its J‐coupling at the echo time,23 resulting in high spermine signals that might yield additional biological information.1, 27
In the seminal vesicles, an increased PC/PE ratio and a high tCho signal with an almost absent Spm peak were found (Figure S‐3 in Supporting Information S.4). In one particular case (Fig. 8) with an aggressive Gleason 4 + 4 prostate cancer with bone metastases, increased levels of total choline were observed in the cancer lesion. The elevated tCho correlated with increases in GPE, GPC, and PC levels and suggests an increased degradation of membrane phospholipids.7
Although there is a correlation between 31P and 1H choline signals, it appears as though the increase in 31P metabolites cannot completely account for the large difference in tCho levels observed between cancer and normal tissue. This indicates that also other overlapping signals, such as free choline, change significantly, but drawing conclusions about specific free choline levels may be difficult using the current setup. Due to the ERC profile, absolute quantification of the metabolites is difficult and necessitates the use of metabolite ratios or coil profile corrections to draw more quantitative conclusions. For 31P spectroscopy, relevant ratios would be PDE/PME, PC/PE, or PE/(PDE + PME).9 For 1H spectroscopy tCho/Spm could be used, as tCho tends to be elevated in cancer lesions, whereas spermine tends to decrease in cancer tissue.27
5. Conclusion
Using a 31P Tx/Rx 1H Rx endorectal coil in combination with an external multitransmit 1H body array enables high resolution multiparametric MR imaging and 1H and 31P spectroscopy of the prostate at 7T. The setup does not pose any safety issues and provides superior 1H SNR locally in the prostate in comparison with a 1H external body array alone. The clinical data shown in this study are anecdotal but they illustrate how the proposed setup would be particularly suitable to combine anatomically and functionally with metabolic imaging in prostate cancer.
Conflicts of interest
The authors have no relevant conflicts of interest to disclose.
Supporting information
Acknowledgements
Dutch Cancer Society: KUN 2014‐6624
References
- 1. Heerschap A, Jager GJ, van der Graaf M, et al. In vivo proton MR spectroscopy reveals altered metabolite content in malignant prostate tissue. Anticancer Res. 1997;17:1455–1460. [PubMed] [Google Scholar]
- 2. Kurhanewicz J, Vigneron DB, Hricak H, Narayan P, Carroll P, Nelson SJ. Three‐dimensional H‐1 MR spectroscopic imaging of the in situ human prostate with high (0.24–0.7‐cm3) spatial resolution. Radiology. 1996;198:795–805. [DOI] [PubMed] [Google Scholar]
- 3. Scheenen TW, Rosenkrantz AB, Haider MA, Futterer JJ. Multiparametric magnetic resonance imaging in prostate cancer management: current status and future perspectives. Invest Radiol. 2015;50:594–600. [DOI] [PubMed] [Google Scholar]
- 4. Johnson LM, Turkbey B, Figg WD, Choyke PL. Multiparametric MRI in prostate cancer management. Nat Rev Clin Oncol. 2014;11:346–353. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Scheenen TW, Futterer J, Weiland E, et al. Discriminating cancer from noncancer tissue in the prostate by 3‐dimensional proton magnetic resonance spectroscopic imaging: a prospective multicenter validation study. Invest Radiol. 2011;46:25–33. [DOI] [PubMed] [Google Scholar]
- 6. Ackerstaff E, Pflug BR, Nelson JB, Bhujwalla ZM. Detection of increased choline compounds with proton nuclear magnetic resonance spectroscopy subsequent to malignant transformation of human prostatic epithelial cells. Cancer Res. 2001;61:3599–3603. [PubMed] [Google Scholar]
- 7. Keshari KR, Tsachres H, Iman R, et al. Correlation of phospholipid metabolites with prostate cancer pathologic grade, proliferative status and surgical stage ‐ impact of tissue environment. NMR Biomed. 2011;24:691–699. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Awwad HM, Geisel J, Obeid R. The role of choline in prostate cancer. Clin Biochem. 2012;45:1548–1553. [DOI] [PubMed] [Google Scholar]
- 9. Lagemaat MW, Vos EK, Maas MC, et al. Phosphorus magnetic resonance spectroscopic imaging at 7 T in patients with prostate cancer. Invest Radiol. 2014;49:363–372. [DOI] [PubMed] [Google Scholar]
- 10. Kobus T, Bitz AK, van Uden MJ, et al. In vivo 31P MR spectroscopic imaging of the human prostate at 7 T: safety and feasibility. Magn Reson Med. 2012;68:1683–1695. [DOI] [PubMed] [Google Scholar]
- 11. Swanson MG, Keshari KR, Tabatabai ZL, et al. Quantification of choline‐ and ethanolamine‐containing metabolites in human prostate tissues using 1H HR‐MAS total correlation spectroscopy. Magn Reson Med. 2008;60:33–40. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Luttje MP, Italiaander MG, Arteaga de Castro CS, et al. (31) P MR spectroscopic imaging combined with (1) H MR spectroscopic imaging in the human prostate using a double tuned endorectal coil at 7T. Magn Reson Med. 2014;72:1516–1521. [DOI] [PubMed] [Google Scholar]
- 13. Metzger GJ, Snyder C, Akgun C, Vaughan T, Ugurbil K, Van de Moortele PF. Local B1+ shimming for prostate imaging with transceiver arrays at 7T based on subject‐dependent transmit phase measurements. Magn Reson Med. 2008;59:396–409. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Lagemaat MW, Philips BW, Vos EK, et al. Feasibility of multiparametric magnetic resonance imaging of the prostate at 7 T. Invest Radiol. 2017;52:295–301. [DOI] [PubMed] [Google Scholar]
- 15. Lagemaat MW, Maas MC, Vos EK, et al. (31) P MR spectroscopic imaging of the human prostate at 7 T: T1 relaxation times, Nuclear Overhauser Effect, and spectral characterization. Magn Reson Med. 2015;73:909–920. [DOI] [PubMed] [Google Scholar]
- 16. Arteaga de Castro CS, van den Bergen B, Luijten PR, et al. Improving SNR and B1 transmit field for an endorectal coil in 7 T MRI and MRS of prostate cancer. Magn Reson Med. 2012;68:311–318. [DOI] [PubMed] [Google Scholar]
- 17. Orzada S, Quick HH, Ladd ME, et al. A flexible 8‐channel transmit/receive body coil for 7 Thuman imaging Paper presented at: In Proceedings of the 17th Annual Meeting of ISMRM; Abstract Number 2999; 2009; Honolulu, Hawaii, USA.
- 18. Rietsch SH, Quick HH, Orzada S. Impact of different meander sizes on the RF transmit performance and coupling of microstrip line elements at 7 T. Med Phys. 2015;42:4542–4552. [DOI] [PubMed] [Google Scholar]
- 19. Larsson EG, Erdogmus D, Yan R, Principe JC, Fitzsimmons JR. SNR‐optimality of sum‐of‐squares reconstruction for phased‐array magnetic resonance imaging. J Magn Reson. 2003;163:121–123. [DOI] [PubMed] [Google Scholar]
- 20. Gilbert G. Measurement of signal‐to‐noise ratios in sum‐of‐squares MR images. J Magn Reson Imaging. 2007;26:1678–1678; author reply 1679. [DOI] [PubMed] [Google Scholar]
- 21. Gabriel C. Compilation of the Dielectric Properties of Body Tissues at RF and Microwave Frequencies. San Antonio, TX: Brooks AFB, Tech. Rep; 1996. [Google Scholar]
- 22. Brunheim S, Gratz M, Johst S, et al. Fast and accurate multi-channel B1+ mapping based on the TIAMO technique for 7T UHF body MRI. Magn Reson Med. 2018;79:2652–2664. [DOI] [PubMed] [Google Scholar]
- 23. Lagemaat MW, Breukels V, Vos EK, et al. (1)H MR spectroscopic imaging of the prostate at 7T using spectral‐spatial pulses. Magn Reson Med. 2016;75:933–945. [DOI] [PubMed] [Google Scholar]
- 24. Maas MC, Vos EK, Lagemaat MW, et al. Feasibility of T2 ‐weighted turbo spin echo imaging of the human prostate at 7 tesla. Magn Reson Med. 2014;71:1711–1719. [DOI] [PubMed] [Google Scholar]
- 25. Vos EK, Sambandamurthy S, Kamel M, et al. Clinical comparison between a currently available single‐loop and an investigational dual‐channel endorectal receive coil for prostate magnetic resonance imaging: a feasibility study at 1.5 and 3 T. Invest Radiol. 2014;49:15–22. [DOI] [PubMed] [Google Scholar]
- 26. Metzger GJ, van de Moortele PF, Akgun C, et al. Performance of external and internal coil configurations for prostate investigations at 7 T. Magn Reson Med. 2010;64(6):1625–1639. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27. Giskeodegard GF, Bertilsson H, Selnaes KM, et al. Spermine and citrate as metabolic biomarkers for assessing prostate cancer aggressiveness. PLoS ONE. 2013;8:e62375. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.