Abstract/Summary
The implantable cardioverter defibrillator (ICD) is a life-saving therapy in various patient populations. Although data on the outcomes of initial ICD implants are abundant, data on ICD replacements, especially in patients with improved left ventricular function, are scarce. Therefore, it is not known when it is safe to not replace an ICD that has reached the end of battery life. In this paper, we review data on patients with primary prevention ICDs who have improvement in the left ventricular ejection fraction during follow-up and provide some guidance, based on the available evidence, related to circumstances in which replacement of an ICD may be forgone.
Keywords: sudden cardiac death, implantable cardioverter defibrillator, cardiac resynchronization therapy, left ventricular ejection fraction recovery
Several randomized clinical trials have demonstrated the efficacy of the implantable cardioverter defibrillator (ICD) in reducing sudden cardiac death and all-cause mortality in survivors of cardiac arrest, patients with sustained ventricular tachycardia and structural heart disease, and patients with significant systolic dysfunction.1–7 This robust evidence of the survival benefit of ICDs has not only informed professional guidelines,8 but it has transformed clinical practice with about 100,000 new ICDs being implanted in patients in the United States annually.9 The continued use of primary prevention ICDs in clinical practice has been further supported by analyses of the National Cardiovascular Data Registry ICD Registry.10,11
Current guidelines provide recommendations regarding when to implant a primary prevention or a secondary prevention ICD in a given patient.8 However, guidance on when it is safe not to replace an ICD is missing from these guidelines. In fact, this issue was identified by the most recent guidelines that were published in 2012 as a gap in knowledge in need of further research. Specifically, the guidelines state that: “Indicators should be identified that provide direction about when it is safe to not replace an ICD that has reached the end of its effective battery life.”8
The appropriate use criteria document generated by the Heart Rhythm Society (HRS) and the American College of Cardiology (ACC) also reflects the current uncertainty regarding the best treatment strategy for older patients whose ICD reaches end of life in the setting of an improved left ventricular ejection fraction (LVEF) and no prior ICD shocks for life-threatening arrhythmias. The document states that among patients with improved heart function and no prior ICD shocks, generator replacement may be “at times an appropriate option.”12 This recommendation was supported mainly by expert opinion and not by high-quality studies. In this article, we will review guideline recommendations regarding ICDs, summarize published data on the outcomes of patients with primary prevention ICDs with or without cardiac resynchronization therapy (CRT) who have improvement in the left ventricular (LV) function, provide some guidance regarding situations in which it may be safe to not replace an ICD, and identify gaps in the existing evidence on this topic that should be addressed by future research.
Current Indications for ICDs:
The most recent guideline document on device-based therapy of cardiac rhythm abnormalities lists the following class I indications for the ICD:8
“Survivors of cardiac arrest due to ventricular fibrillation or hemodynamically unstable sustained ventricular tachycardia (VT) after evaluation to define the cause of the event and to exclude any completely reversible causes.
Patients with structural heart disease and spontaneous sustained VT, whether hemodynamically stable or unstable.
Patients with syncope of undetermined origin with clinically relevant, hemodynamically significant sustained VT or ventricular fibrillation induced at electrophysiological study.
Patients with LV dysfunction due to prior myocardial infarction (MI), ≥ 40 days post-MI, LVEF ≤ 35%, New York Heart Association (NYHA) class II or III.
Patients with non-ischemic cardiomyopathy and LVEF ≤ 35%, NYHA class II or III
Patients with LV dysfunction due to prior MI who are at least 40 days post-MI, have an LVEF less than or equal to 30%, and are in NYHA functional Class I.
Patients with nonsustained VT due to prior MI, LVEF less than or equal to 40%, and inducible ventricular fibrillation or sustained VT at electrophysiological study.”8
Multiple class IIa indications for ICDs are also listed in the guideline document. These class IIa indications involve milder presentations of class I scenarios such as unexplained syncope with significant systolic dysfunction due to non-ischemic cardiomyopathy or sustained VT in patients with normal or near-normal ventricular function.8 All recommendations on ICDs stipulate that patients be on optimal medical therapy and have a reasonable expectation of survival for more than 1 year.8 As stated previously, the guidelines do not offer any guidance on when not to re-implant an ICD.
Data on ICDs:
During an ICD generator’s typical lifetime of approximately 7–10 years, about 65% of patients do not receive ICD shocks and up to 45% of patients have a substantial improvement in their LVEF to > 35%, such that they would no longer meet accepted indications for a new ICD.13,14 In a study of 195 patients, 50 (25.6%) had 10% improvement in LVEF reaching a median LVEF during follow-up of 41% (25th, 75th percentiles 37%, 49%).15 Studies have shown that LVEF improvement is associated with a significant long-term reduction in heart failure hospitalizations, ICD shocks, and death.13,16–18 Among patients with ICDs, one of the strongest predictors of ICD shocks is a prior heart failure hospitalization.19 The close correlation between the LVEF, worsening heart failure, and the risk of ventricular arrhythmias has raised questions about the benefits of ICD in patients with substantial improvement in the LVEF.20,21
However, a few retrospective studies with relatively small sample sizes have shown a non-trivial rate of appropriate ICD shocks even in patients with significant LVEF improvement.22–24 In a secondary analysis of the Defibrillators in Non-Ischemic Cardiomyopathy Treatment Evaluation (DEFINITE), patients whose LVEF improved had lower mortality risk compared with patients whose LVEF decreased, but appropriate shocks did not significantly correlate with LVEF improvement during follow-up.22 Another study of 91 patients with primary prevention ICDs found that at generator replacement, 25 patients (27%) had improved LVEF. During a mean follow-up of 6 years, 9 patients (36%) with improved LVEF versus 19 (29%) with unchanged LVEF had appropriate ICD shocks.23 In an analysis of 154 primary prevention ICD patients needing replacement because of battery depletion, 114 (74%) patients had not received appropriate ICD therapy. Following ICD generator replacement, the 3-year incidence of appropriate therapy in these patients was 14%; however, this study did not include information on the incidence of LVEF improvement.24 In a similar study, 275 of 403 patients (68%) needing a primary prevention ICD replacement had not received an appropriate therapy from the initial device. After ICD generator replacement, the 3-year risk of appropriate ICD therapy was 13.7%, but no information on follow-up LVEF was provided.25 So, based on these data, it seems appropriate to replace the ICD in patients with a primary prevention ICD who continue to meet the guideline recommendations at the time of device replacement.
Given the conflicting data on the outcomes of ICD patients with an improved LVEF and the paucity of data from prospective studies on the best management practices, it is not known whether ICD generator replacement should be performed in patients who no longer meet current indications for an ICD (due to an improved LVEF to > 35% and in the absence of prior ICD shocks).20,26 This uncertainty is amplified in patients ≥65 years of age in whom the burden of comorbid diseases and competing modes of death reduce ICD benefit.26,27 Although many studies have defined the risks associated with ICD generator replacement,28,29 there are currently no studies that have examined whether or not this invasive procedure leads to any benefit among these lower-risk patients.
The critical importance of understanding the best care strategy for individuals who no longer meet current indications for an ICD (due to an improved LVEF > 35% and absence of prior ICD shocks) has been emphasized by patients, care partners, physicians, and professional organizations. In early 2015, as part of its prioritization process, the Patient Centered Outcomes Research Institute (PCORI) tasked the Duke University Evidence Synthesis Group (ESG) to work with diverse stakeholders to identify and prioritize the future research that is most needed by patients and other decision makers on the topic of ICDs in older patients. The results identified “clinical effectiveness and safety of ICD replacements in older patients who have not had an appropriate shock or those with improved LVEF” as a key research priority for PCORI.26
ICD generator replacement is associated with a non-trivial risk of complications; within 6 months after the procedure, the risk of major complications is 4% and the risk of minor complications is 7.4%.28 Major complications associated with ICD generator replacement include lead dislodgement or damage, device infection requiring removal of the entire ICD system, significant bleeding requiring ICD pocket evacuation, and prolonged hospitalization.28 Minor complications include nerve injury, minor bleeding, skin infection, and pain. The risk of device infection is particularly troubling as it does not only appear to be increasing, but it generally requires removal of the whole ICD system. In a study of the National Hospital Discharge Survey database from 1996 through 2006, the risk of device infection was noted to have increased from 2004 to 2006 by 57%.30 Factors that could account for this rise in device infection included an increase in the rate of comorbid illnesses like end-organ failure and diabetes mellitus and an increase in the use of CRT devices.30 Removal of the entire ICD system (generator and leads) is associated with substantial risks including cardiac perforation with or without tamponade, vascular damage, and even death.31,32
Therefore, until more data become available, the following algorithm can be used to make decisions regarding ICD generator replacements:
In reviewing this algorithm, it is important to note that the LVEF measurement is subjective with an inter- and intra-observer variability that approaches 5–8%, so it may be better to use a cutoff for the improved LVEF at battery depletion of > 40% or > 45% to allow for that variability. Also, although the annual risk of receiving ICD therapy among patients with LVEF improvement is relatively small, low to moderate quality studies have shown that over 5 years, >10% (sometimes > 20%) of patients end up receiving an appropriate therapy. This information should be considered and shared with the patient during clinical decision making. Finally, if ICD replacement is forgone, it may be important to obtain periodic follow-up assessments of the LVEF to look for worsening to ≤ 35% at which point replacement of the ICD may be appropriate.
Data on Cardiac Resynchronization Therapy-Defibrillator
CRT has become one of the mainstay therapies in patients with heart failure with a reduced LVEF (≤ 35%) who have a wide QRS and symptomatic heart failure despite optimal medical therapy.8 Many studies have shown that this therapy improves patients’ survival, quality of life, and functional status and reduces heart failure hospitalization.33–37 CRT devices are either pacemakers (CRT-P) or defibrillators (CRT-D). Both types of devices resynchronize the heart through biventricular pacing; CRT-D can also deliver high-energy shocks or anti-tachycardia pacing to terminate life-threatening ventricular arrhythmias. Patients who receive a primary prevention CRT-D as their initial device may have significant improvement in their LVEF due to CRT and as such their risk of ventricular arrhythmias may decrease substantially. In such patients, when the battery reaches end of life, it is not known whether a CRT-P may suffice. The HRS/ACC Appropriate Use Criteria document considered these types of scenarios and determined that it may be appropriate to replace a CRT-D with a CRT-P device in patients who received a CRT-D device when the LVEF ≤ 35% and the LVEF at end of battery life is ≥ 36%.12
A few studies have reported on the risk of appropriate ICD shocks from CRT-Ds in patients with an improved LVEF. In a landmark analysis of 270 patients with post-CRT-D LVEF assessment and no ICD therapy within 1 year of device implantation, the 2-year risk of appropriate ICD therapy was 3.0%, 2.1%, and 1.5% for post-CRT-D LVEF of 45%, 50% and 55%, respectively. This study concluded that when the LVEF near normalizes to ≥ 45%, the incidence of ICD therapies for ventricular arrhythmias decreases appreciably.38 In another study of 270 patients, the LVEF improved to >35% in 21% of patients during a follow-up period of 12 months. Patients with an LVEF improvement to >35% received fewer ICD therapies during follow-up than those who did not (23 vs.38%; P-value 0.03). Limiting the analysis to patients with a primary prevention ICD showed that the rate of ICD therapies was 31% in patients with no LVEF improvement versus 6% in patients with an LVEF improvement to > 35%.39
Another study reported on the risk of appropriate ICD shocks in 110 patients with CRT-D who had an absolute increase in the LVEF of ≥ 10%, a decrease in LV end-systolic volume of ≥ 30%, or a decrease in LV end-diastolic volume of ≥ 20%. They found these 3 groups of patients received ICD shocks ((3-year risk: 64% vs 82% for LVEF ≥ 10%; 63% vs 92% for end-systolic volume ≥ 30%; and 62% vs 94% for end-diastolic volume ≥ 20%; all p<0.001) during a mean follow-up duration of 25 months; however, this study was limited by the relatively small sample size and short follow-up.40 In another study of 142 patients with a primary prevention CRT-D and a baseline LVEF< 35%, 42 had an improvement in the LVEF to > 35% over a median follow-up of 3 years. None of these 42 patients had appropriate ICD therapies during follow-up.40
A secondary analysis of the Multicenter Automatic Defibrillator Implantation Trial With Cardiac Resynchronization Therapy (MADIT-CRT) trial focused on 752 patients who survived and had echocardiographic data at enrollment and at 12 months. During an average follow-up period of 2.2 years, 7.3% patients had LVEF normalization. The risk of ventricular arrhythmias was significantly lower in patients with LVEF >50% (hazard ratio [HR], 0.24; 95% confidence interval [CI], 0.07–0.82; p=0.023) and LVEF of 36% to 50% (HR, 0.44; 95% CI, 0.28–0.68; p<0.001). Because the risk of inappropriate ICD therapy was non-trivial, the authors concluded that patients with LVEF improvement could be considered for downgrade from CRT-D to CRT-P when the battery reaches end of life if no ventricular arrhythmias had occurred.41
Finally, a meta-analysis was conducted to determine the association between LVEF recovery following CRT and the risk of appropriate ICD therapy.43 Six retrospective cohort studies were included (n=1740). The risk of appropriate ICD therapy was significantly lower in patients with post-CRT LVEF ≥35% (5.5/100 person-years) than patients with post-CRT LVEF<35% [incidence rate difference −6.5/100 person-years, 95% confidence interval (95% CI): −8.8 to −4.2, p=0.001]. Focusing the analysis on patients with a post-CRT LVEF of ≥45% showed a lower risk of ICD therapy (2.3/100 person-years) compared with patients without such recovery (−5.8/100 person-years, 95% CI: −7.6 to 24.0, p=0.001). Patients with a primary prevention ICD had very low rates of ICD therapy (0.4 to 0.8/100-person years) irrespective of LVEF recovery. The investigators concluded that LVEF recovery post-CRT is associated with significantly reduced appropriate ICD therapy, and this risk is lowest in patients with improvement of LVEF ≥45% and those with a primary prevention indication.43
Based on the aforementioned studies, it is not known how to best manage patients with a CRT-D device with no prior ICD therapies who have significant improvement in the LVEF at the time of battery depletion. Therefore, randomized clinical trials of CRT-P vs. CRT-D especially at the time of device replacement are needed. Until such trials are completed, the following algorithm can be used to make decisions regarding CRT-D generator replacements:
Conclusions:
A large number of patients with a primary prevention ICD or a CRT-D have significant improvement in their left ventricular function during follow-up. Patients with an ICD and improved LVEF to > 45% appear to be at a low risk of ventricular arrhythmias; however, there are no data from prospective studies on whether it is safe to not replace the ICD in such patients. In patients with a CRT-D who have significant improvement in their left ventricular function, evidence from retrospective studies suggests that the risk of future ventricular arrhythmias is very low in patients with a primary prevention indication whose LVEF improves to ≥ 45% with CRT. However, there are no data on whether replacing the CRT-D with a CRT-P in such patients is safe. It is noteworthy that the absolute LVEF value during follow-up may not be enough in assessing the risk of ventricular arrhythmias; it is possible that the magnitude of improvement is more important than whether the LVEF is < or > 35%. Future studies should look at the predictive power of the LVEF as a continuous variable. In addition, future randomized clinical trials should compare the outcomes of ICD replacement versus not and CRT-D versus CRT-P at the time of battery depletion. Until data from clinical trials are available to guide clinical recommendations, shared decision making is essential taking into account the uncertainties in the data and the known pros and cons of replacing versus not replacing the ICD.
Fig. 1.
ICD generator replacement decision algorithm. a The LVEF measurement is subjective and the variability may be up to 8%, so it may be better to use a cutoff of 40% or 45% to allow for that variability.
Fig. 2.
CRT-D generator replacement decision algorithm. a LVEF cutoff of 45% was used based on data from Chatterjee et al.43
(Data from Chatterjee NA, Roka A, Lubitz SA, et al. Reduced appropriate implantable cardioverter-defibrillator therapy after cardiac resynchronization therapy-induced left ventricular function recovery: a meta-analysis and systematic review. Eur Heart J 2015;36(41):2780–9.)
Key points.
Available data suggest that up to 45% of recipients of primary prevention ICDs have significant improvement in their left ventricular function during follow-up. Although prospective data on the outcomes of patients with a primary prevention ICD and substantial improvement in left ventricular function are scarce, such patients appear to be at a low risk of ventricular arrhythmias.
Prospective data are limited on the outcomes of patients with a cardiac resynchronization therapy defibrillator (CRT-D) who have significant improvement in their left ventricular function; however, the cumulative evidence from retrospective studies suggests that the risk of future ventricular arrhythmias is very low in patients with a primary prevention indication whose left ventricular ejection fraction (LVEF) improves to ≥ 45% with CRT.
It is not known when it is safe to not re-implant an ICD in patients with a primary prevention ICD with or without CRT.
Until more data are available, the 2 situations in which it may be safe not to re-implant an ICD are: (1) patients with a primary prevention ICD who have significant improvement in left ventricular function (LVEF ≥ 45%), no prior ICD therapies and no bradycardia pacing indication and (2) patients with a primary prevention CRT-D who have an improvement of the LVEF to ≥ 45% with CRT. If ICD replacement is forgone, it may be important to periodically assess the LVEF to look for worsening to ≤ 35% at which point replacement of the ICD may be appropriate.
Randomized clinical trials are needed on ICD replacement versus not and on CRT-D versus CRT-P at the time of battery depletion.
Acknowledgments
Disclosures: Dr. Al-Khatib reports none. Dr. Friedman has received modest research grants from Boston Scientific, significant research grants from the National Cardiovascular Data Registry, modest educational grants from St. Jude Medical and Boston Scientific, and salary support from the NIH T-32 training grant HL069749. Dr. Sanders reports none.
References:
- 1).Moss AJ, Hall WJ, Cannom DS, et al. Improved survival with an implanted defibrillator in patients with coronary disease at high risk for ventricular arrhythmia. Multicenter Automatic Defibrillator Implantation Trial Investigators. N Engl J Med 1996;335(26):1933–1940. [DOI] [PubMed] [Google Scholar]
- 2).A comparison of antiarrhythmic-drug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmias. Antiarrhythmics versus Implantable Defibrillators (AVID) Investigators. N Engl J Med 1997;337(22):1576–1583. [DOI] [PubMed] [Google Scholar]
- 3).Buxton AE, Lee KL, Fisher JD, Josephson ME, Prystowsky EN, Hafley G. A randomized study of the prevention of sudden death in patients with coronary artery disease. Multicenter Unsustained Tachycardia Trial Investigators. N Engl J Med 1999;341(25):1882–1890. [DOI] [PubMed] [Google Scholar]
- 4).Connolly SJ, Hallstrom AP, Cappato R, et al. Meta-analysis of the implantable cardioverter defibrillator secondary prevention trials. AVID, CASH and CIDS studies. Antiarrhythmics vs Implantable Defibrillator study. Cardiac Arrest Study Hamburg. Canadian Implantable Defibrillator Study. Eur Heart J 2000;21(24):2071–2078. [DOI] [PubMed] [Google Scholar]
- 5).Moss AJ, Zareba W, Hall WJ, et al. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med 2002;346(12):877–883. [DOI] [PubMed] [Google Scholar]
- 6).Kadish A, Dyer A, Daubert JP, et al. Prophylactic defibrillator implantation in patients with nonischemic dilated cardiomyopathy. N Engl J Med 2004;350(21):2151–2158. [DOI] [PubMed] [Google Scholar]
- 7).Bardy GH, Lee KL, Mark DB, et al. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med 2005;352(3):225–237. [DOI] [PubMed] [Google Scholar]
- 8).Epstein AE, DiMarco JP, Ellenbogen KA, Estes NAM, Freedman RA, Gettes LS, Gillinov AM, Gregoratos G, Hammill SC, Hayes DL, Hlatky MA, Newby LK, Page RL, Schoenfeld MH, Silka MJ, Stevenson LW, Sweeney MO. 2012 ACCF/AHA/HRS Focused Update Incorporated Into the ACCF/AHA/HRS 2008 Guidelines for Device-Based Therapy of Cardiac Rhythm Abnormalities: a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines and the Heart Rhythm Society Circulation 2013;127(3):e283–352. [DOI] [PubMed] [Google Scholar]
- 9).Kremers MS, Hammill SC, Berul CI, et al. The National ICD Registry Report: version 2.1 including leads and pediatrics for years 2010 and 2011. Heart Rhythm 2013;10(4):e59–65. [DOI] [PubMed] [Google Scholar]
- 10).Al-Khatib SM, Hellkamp A, Bardy GH, Hammill S, Hall WJ, Mark DB, Anstrom K, Curtis J, Al-Khalidi H, Curtis L, Heidenreich P, Peterson ED, Sanders G, Clapp-Channing N, Lee KL, Moss AJ. Survival of patients receiving a primary prevention ICD in clinical practice versus clinical trials. JAMA 2013;309:55–62. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11).Al-Khatib SM, Hellkamp AS, Fonarow GC, Mark DB, Curtis LH, Hernandez A, Anstrom KJ, Peterson ED, Sanders GD, Al-Khalidi HR, Hammill BG, Heidenreich PA, Hammill S. Association between prophylactic implantable cardioverter defibrillators and survival in patients with left ventricular ejection fraction between 30% and 35%. JAMA 2014;311:2209–2215. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12).Russo AM, Stainback RF, Bailey SR, et al. ACCF/HRS/AHA/ASE/HFSA/SCAI/SCCT/SCMR 2013 appropriate use criteria for implantable cardioverter-defibrillators and cardiac resynchronization therapy: a report of the American College of Cardiology Foundation appropriate use criteria task force, Heart Rhythm Society, American Heart Association, American Society of Echocardiography, Heart Failure Society of America, Society for Cardiovascular Angiography and Interventions, Society of Cardiovascular Computed Tomography, and Society for Cardiovascular Magnetic Resonance. Heart Rhythm 2013;10(4):e11–58. [DOI] [PubMed] [Google Scholar]
- 13).Kini V, Soufi MK, Deo R, et al. Appropriateness of primary prevention implantable cardioverter-defibrillators at the time of generator replacement: are indications still met? J Am Coll Cardiol. 2014;63(22):2388–2394. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14).Adabag S, Patton KK, Buxton AE, Rector TS, Ensrud KE, Vakil K, Levy WC, Poole JE. Association of implantable cardioverter defibrillators with survival in patients with and without improved ejection fraction: secondary analysis of the Sudden Cardiac Death in Heart Failure Trial. JAMA Cardiol. Published online. 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15).Friedman DJ, Overton R, Shaw L, Fudim M, Patel D, Velazquez EJ, Al-Khatib SM. The role of baseline versus long-term follow-up ejection fraction in predicting adverse events among primary prevention implantable cardioverter defibrillator recipients. J Am Coll Cardiol 2017. [abstract]. [Google Scholar]
- 16).Kalogeropoulos AP, Fonarow GC, Georgiopoulou V, et al. Characteristics and Outcomes of Adult Outpatients With Heart Failure and Improved or Recovered Ejection Fraction. JAMA Cardiol. 20161(5):510–8. [DOI] [PubMed] [Google Scholar]
- 17).Madhavan M, Waks JW, Friedman PA, et al. Outcomes After Implantable Cardioverter-Defibrillator Generator Replacement for Primary Prevention of Sudden Cardiac Death. Circ Arrhythm Electrophysiol. 2016;9(3):e003283. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18).Zhang Y, Guallar E, Blasco-Colmenares E, et al. Changes in Follow-Up Left Ventricular Ejection Fraction Associated With Outcomes in Primary Prevention Implantable Cardioverter-Defibrillator and Cardiac Resynchronization Therapy Device Recipients. J Am Coll Cardiol. 2015;66(5):524–531. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19).Singh JP, Hall WJ, McNitt S, et al. Factors influencing appropriate firing of the implanted defibrillator for ventricular tachycardia/fibrillation: findings from the Multicenter Automatic Defibrillator Implantation Trial II (MADIT-II). J Am Coll Cardiol. 2005;46(9):1712–1720. [DOI] [PubMed] [Google Scholar]
- 20).Merchant FM, Quest T, Leon AR, El-Chami MF. Implantable Cardioverter-Defibrillators at End of Battery Life: Opportunities for Risk (Re)-Stratification in ICD Recipients. J Am Coll Cardiol. 2016;67(4):435–444. [DOI] [PubMed] [Google Scholar]
- 21).Kramer DB, Buxton AE, Zimetbaum PJ. Time for a change--a new approach to ICD replacement. N Engl J Med 2012;366(4):291–293. [DOI] [PubMed] [Google Scholar]
- 22).Schliamser JE., Kadish AH, Subacius H, et al. , DEFINITE Investigators (2013) Significance of follow-up left ventricular ejection fraction measurements in the Defibrillators in Non-Ischemic Cardiomyopathy Treatment Evaluation trial (DEFINITE). Heart Rhythm 2013;10:838–846. [DOI] [PubMed] [Google Scholar]
- 23).Naksuk N, Saab A, Li JM, et al. (2013) Incidence of appropriate shock in implantable cardioverter-defibrillator patients with improved ejection fraction. J Card Fail. 19:426–430. [DOI] [PubMed] [Google Scholar]
- 24).Van Welsenes GH, Van Rees JB, Thijssen J, et al. Primary prevention implantable cardioverter defibrillator recipients: the need for defibrillator back-up after an event-free first battery service-life. J Cardiovasc Electrophysiol. 2011;22:1346–1350. [DOI] [PubMed] [Google Scholar]
- 25).Yap SC, Schaer BA, Bhagwandien RE, et al. Evaluation of the need of elective implantable cardioverter-defibrillator generator replacement in primary prevention patients without prior appropriate ICD therapy. Heart. 2014;100:1188–1192. [DOI] [PubMed] [Google Scholar]
- 26).Al-Khatib SM, Gierisch JM, Crowley MJ, et al. Future Research Prioritization: Implantable Cardioverter-Defibrillator Therapy in Older Patients. J Gen Intern Med. 2015;30(12):1812–1820. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27).Steinberg BA, Al-Khatib SM, Edwards R, et al. Outcomes of implantable cardioverter-defibrillator use in patients with comorbidities: results from a combined analysis of 4 randomized clinical trials. JACC Heart Fail. 2014;2(6):623–629. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28).Poole JE, Gleva MJ, Mela T, et al. Complication rates associated with pacemaker or implantable cardioverter-defibrillator generator replacements and upgrade procedures: results from the REPLACE registry. Circulation 2010;122(16):1553–1561. [DOI] [PubMed] [Google Scholar]
- 29).Prutkin JM, Reynolds MR, Bao H, et al. Rates of and factors associated with infection in 200 909 Medicare implantable cardioverter-defibrillator implants: results from the National Cardiovascular Data Registry. Circulation 2014;130(13):1037–1043. [DOI] [PubMed] [Google Scholar]
- 30).Voigt A, Shalaby A, Saba S. Continued rise in rates of cardiovascular implantable electronic device infections in the United States: temporal trends and causative insights. Pacing Clin Electrophysiol. 2010;33(4):414–419. [DOI] [PubMed] [Google Scholar]
- 31).Wazni O, Epstein LM, Carrillo RG, et al. Lead extraction in the contemporary setting: the LExICon study: an observational retrospective study of consecutive laser lead extractions. J Am Coll Cardiol. 2010;55(6):579–586. [DOI] [PubMed] [Google Scholar]
- 32).Wilkoff BL, Love CJ, Byrd CL, Bongiorni MG, Carrillo RG, et al. Transvenous lead extraction: Heart Rhythm Society expert consensus on facilities, training, indications, and patient management: this document was endorsed by the American Heart Association (AHA). Heart Rhythm. 2009;6(7):1085–104 [DOI] [PubMed] [Google Scholar]
- 33).Bristow MR, Saxon LA, Boehmer J, et al. Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure. N Engl J Med 2004;350(21):2140–50. [DOI] [PubMed] [Google Scholar]
- 34).Cleland JG, Daubert JC, Erdmann E, et al. The effect of cardiac resynchronization on morbidity and mortality in heart failure. N Engl J Med 2005;352:1539–1549. [DOI] [PubMed] [Google Scholar]
- 35).Linde C, Abraham WT, Gold MR, et al. Randomized trial of cardiac resynchronization in mildly symptomatic heart failure patients and in asymptomatic patients with left ventricular dysfunction and previous heart failure symptoms. J Am Coll Cardiol. 2008;52(23):1834–1843. [DOI] [PubMed] [Google Scholar]
- 36).Moss AJ, Hall WJ, Cannom DS, et al. Cardiac-resynchronization therapy for the prevention of heart failure events. N Engl J Med 2009;361(14):1329–38. [DOI] [PubMed] [Google Scholar]
- 37).Tang AS, Wells GA, Talajic M, et al. Cardiac-resynchronization therapy for mild-to-moderate heart failure. N Engl J Med 2010;363(25):2385–95. [DOI] [PubMed] [Google Scholar]
- 38).Manfredi JA, Al-Khatib SM, Shaw LK, Thomas L, Fogel RI, Padanilam B, Rardon D, Vatthyam R, Gemma LW, Golden K, Prystowsky EN. Association between left ventricular ejection fraction post-cardiac resynchronization treatment and subsequent ICD therapy for sustained ventricular tachyarrhythmias. Circ: Arrhythmia and Electrophysiol. 2013;6:257–264. [DOI] [PubMed] [Google Scholar]
- 39).Schaer BA, Osswald S, Di Valentino M, Soliman OI, Sticherling C, ten Cate FJ, Jordaens L, Theuns DA. Close connection between improvement in left ventricular function by cardiac resynchronization therapy and the incidence of arrhythmias in cardiac resynchronization therapy-defibrillator patients. Eur J Heart Fail 2010;12:1325–1332. [DOI] [PubMed] [Google Scholar]
- 40).Steffel J, Milosevic G, Hurlimann A, Krasniqi N, Namdar M, Ruschitzka F, Luscher TF, Duru F, Holzmeister J, Hurlimann D. Characteristics and long-term outcome of echocardiographic super-responders to cardiac resynchronization therapy: ‘real world’ experience from a single tertiary care centre. Heart 2011; 97:1668–1674. [DOI] [PubMed] [Google Scholar]
- 41).Van Boven N, Bogaard K, Ruiter J, Kimman G, Theuns D, Kardys I, Umans V. Functional response to cardiac resynchronization therapy is associated with improved clinical outcome and absence of appropriate shocks. J Cardiovasc Electrophysiol. 2013;24:316–322. [DOI] [PubMed] [Google Scholar]
- 42).Ruwald MH, Solomon SD, Foster E, et al. Left ventricular ejection fraction normalization in cardiac resynchronization therapy and risk of ventricular arrhythmias and clinical outcomes: results from the Multicenter Automatic Defibrillator Implantation Trial With Cardiac Resynchronization Therapy (MADIT-CRT) trial. Circulation 2014;130:2278–2286. [DOI] [PubMed] [Google Scholar]
- 43).Chatterjee NA, Roka A, Lubitz SA, Gold MR, Daubert C, Linde C, Steffel J, Singh JP, Mela T. Reduced appropriate implantable cardioverter-defibrillator therapy after cardiac resynchronization therapy-induced left ventricular function recovery: a meta-analysis and systematic review Eur Heart J. 2015; 36(41): 2780–2789. [DOI] [PMC free article] [PubMed] [Google Scholar]


