Skip to main content
International Journal of Particle Therapy logoLink to International Journal of Particle Therapy
. 2016 Mar 24;2(4):544–554. doi: 10.14338/IJPT-15-00026.1

Improving Head and Neck Cancer Treatments Using Dynamic Collimation in Spot Scanning Proton Therapy

Alexandra Moignier 1,, Edgar Gelover 1, Dongxu Wang 1, Blake Smith 1, Ryan Flynn 1, Maura Kirk 2, Liyong Lin 2, Timothy Solberg 2, Alexander Lin 2, Daniel Hyer 1
PMCID: PMC6871640  PMID: 31772966

Abstract

Purpose:

Interest in using collimation for spot scanning proton therapy has recently increased in an attempt to improve the lateral penumbra. To investigate the advantages of such an approach for complex targets, a plan comparison between uncollimated and collimated beam spots was performed for patients with head and neck cancer.

Patients and Methods:

For 10 patients with head and neck cancer, previously treated with spot scanning proton therapy, uncollimated and collimated treatment plans were created using an in-house treatment-planning system capable of modeling asymmetric-beamlet dose distributions resulting from the use of a dynamic collimation system. Both uncollimated and collimated plans reproduced clinically delivered plans in terms of target coverage. A relative plan comparison was performed using both physical and radiobiological metrics on the organs at risk.

Results:

The dynamic collimation system improved dose-distribution conformity while preserving target coverage. The median reduction of the mean dose to the esophagus, uninvolved larynx, and uninvolved parotids were −11.9% (minimum to maximum, −6.4% to −24.1%), −7.2% (−0.8% to −60.1%), and −5.2% (−0.2% to −21.5%), respectively, and depended on the organ location relative to the target and radiation beam angle. The collimation did not improve dose to some organs at risk surrounded by the target or located upstream of Bragg peaks because of the priority on the target coverage.

Conclusion:

In spot scanning proton therapy, the dynamic collimation system generally affords better target conformity, which results in improvement in organ-at-risk sparing in the head and neck region while preserving target coverage. However, the benefits of collimation and the increased complexity should be considered for each patient. Patients with large bilateral targets or organs at risk surrounded by the target showed the least benefit.

Keywords: spot scanning proton therapy, collimation, head and neck

Introduction

Low-energy spot scanning proton therapy (SSPT) treatments suffer from poor conformity because of increased lateral spot size [13], which is further increased with patient scattering. Wang et al [1], van de Water et al [2], and Widesott et al [3] have found that, to improve plan quality over the state-of-the-art photon techniques for brain or head and neck cancers, the in-air spot size, defined as the sigma (σair) of the Gaussian function used to represent fluence in the beam's eye view of a proton pencil beam, must have a value of approximately 4 mm. Unfortunately, spots of this size at low energies do not seem achievable with current clinical equipment [46], even with the IBA-dedicated nozzle (IBA International, Louvain-La-Neuve, Belgium) which is known to have one of the smaller spot sizes commercially available [7]. However, collimation devices for SSPT, such as apertures or multileaf collimators, can potentially reduce the effective σ of a proton pencil beam and improve overall plan quality [8, 9]. Hyer et al [10, 11] recently proposed a theoretical novel dynamic collimation system (DCS) with a small footprint that can provide collimation to SSPT at each beam spot.

An earlier study [12] investigating the clinical benefits of the DCS for brain treatments found that the DCS was able to greatly improve target conformity and normal tissue sparing, while maintaining target coverage, with an average mean dose reduction to normal brain tissue of 25.1%. Although head and neck cancers often have much larger target volumes and more-complex shapes than do the brain tumors studied previously, we believe that these patients may also benefit from the use of the DCS, largely because target volumes are often located near many organs at risk (OARs). With the ability to reduce the effective spot size and further improve dose conformity, using the DCS in SSPT is expected to increase or maintain tumor control probability while decreasing acute and late toxicities [13]. The purpose of this study was to evaluate the use of the DCS for head and neck tumors and retrospectively quantify the benefits of the DCS in 10 patients previously treated with SSPT.

Materials and Methods

Patient Selection and Planning Method

Treatment planning images and data for 10 patients with head and neck cancer (Table 1) treated previously at the University of Pennsylvania Proton Therapy Center (Philadelphia) were obtained through an institutional review board–approved research protocol and data sharing agreement. The patients were originally planned using a commercial proton treatment planning system (Eclipse version 11, Varian Medical Systems, Palo Alto, California) and treated on an IBA ProteusPlus system equipped with a Universal Nozzle. The plan characteristics are summarized in Table 1. The beam characteristics of the Universal Nozzle are described by Lin et al [14]. The clinical plans used SSPT with a single-field optimization approach, except for 2 plans (P2 and P3) that utilized a boost planned with multifield optimization. The single-field optimization considers each beam individually, whereas the multifield optimization takes into account all beams simultaneously. The plans were optimized to pencil beam-scanning planning target volumes (PBSTV), which are defined as a 5-mm uniform expansion from the clinical target volume to account for setup and range uncertainties. Some plans had multiple PBSTVs with different dose prescriptions that were joined in a structure named the PBSTV composite for ease of display and analysis of the dose-volume histogram (DVH) curves. The PBSTV composite volumes for patients P1 to P10 were 500, 1004, 591, 822, 290, 436, 305, 200, 272, and 332 cm3, respectively. Other plan characteristics, such as individual PBSTV volumes, dose prescriptions, and beam incidences, are described in Table 1. The initial clinical plans were designed such that 95% of each PBSTV received the prescribed dose. The OAR dose constraints used for planning purposes were mean dose (Dmean) <  20 Gy for the esophagus, larynx, oral cavity, and contralateral parotid; Dmean < 26 Gy for the ipsilateral parotid, Dmean < 39 Gy for the contralateral submandibular gland; maximum dose (Dmax) < 54 Gy for the brain stem, Dmax < 45 Gy for the cord, and Dmax < 50 Gy for the 5-mm expansion of the cord.

Table 1.

Plan characteristics.

Patient No.
Target location
PBSTV dose prescription (Gy)
PBSTV volume (cm3)
Bean incidence
P1 Bilateral: SCC of the tongue base 54
60
235
254
Left posterior
Right posterior
P2 Bilateral: SCC of the left oral tongue 59.5
63
70
72.1
284
261
 23
365
Left posterior
Right posterior
Left posterior oblique
P3 Bilateral: SCC of the left oral tongue, recurred in bilateral neck 54
57
60
63
66
 73
231
145
 22
 45
Left posterior
Right posterior
Right posterior oblique
Anterior
P4 Bilateral: SCC of the right tonsil 54
60
408
452
Left posterior
Right posterior
P5 Unilateral: SCC of the right pyriform sinus 54
60
241
 43
Right posterior
Posterior
P6 Unilateral: SCC of the left supraglottic larynx 50.4 436 Left anterior oblique
Left posterior oblique
P7 Unilateral: carcinoma of the left parotid gland 60
63
256
 36
Left anterior oblique
Left posterior oblique
P8 Central: recurrent head and neck cancer involving the oropharynx 70 188 Left posterior oblique
Right posterior oblique
P9 Unilateral: recurrent basal cell cancer of the scalp, metastatic to the left neck lymph node 54
60
63
 71
 95
 36
Left anterior oblique
Left posterior oblique
P10 Unilateral: recurrent SCC of the skin, with spread to the left parotid 70.4 332 Left lateral
Left posterior oblique

Abbreviations: PBSTV, pencil beam scanning planning target volumes; SCC, squamous cell carcinoma.

The clinically delivered SSPT plans were used as a reference to which uncollimated plans generated by our in-house treatment-planning system RDX were matched, mainly in terms of target coverage. RDX has been previously used for other studies involving proton-dose calculation [15, 16] and is capable of modeling asymmetric beamlets resulting from collimation [17]. The ability of RDX to calculate both collimated and uncollimated treatment plans eliminates any variance from different planning systems and allows the benefits of the DCS to be directly quantified. The beam angles of uncollimated and collimated plans in RDX were chosen so that the beam entered parallel to the long axis of the target structure, as shown in Figure 1. These beams are consistent with recommendations of Frank et al [18], who suggested an angle selection based on the premise of placing the Bragg peaks lateral to the spinal cord and minimizing the uncertainties produced by density heterogeneities along the beam path by avoiding beams going through the mouth and teeth. The initial PBSTV outlines of bilateral targets were split into left and right sides using the middle line to benefit from the spot collimation for OARs in the middle of the neck surrounded by the bulky target. A left posterior beam was used to calculate the left target beamlets, whereas a right posterior beam was used for the right target. Then, both beamlet batches were merged for the optimization calculation of beamlet weights.

Figure 1.

Figure 1.

Two typical beam setups for 2 target geometries, 1 a bulky tumor surrounding the cord (grey circle): field geometries (blue) with target outlines (orange). Abbreviations: LPO, left posterior oblique; LAO, left anterior oblique; LPOST, left posterior; RPOST, right posterior.

A model of the IBA Universal Nozzle was used for the uncollimated plans in RDX. For the collimated plan, the effects of the DCS were added to the IBA Universal Nozzle. The modeling of the collimated beamlets for the clinical range of proton energy was validated and implemented in RDX by Gelover et al [17]. The spot spacing used for the plans was 3 mm, and energy layer separation was 5 mm. Both the spot spacing and energy layer separation were different from those used for the initial clinical plans. The rationale for using a reduced x to y spot spacing is that a smaller σ size must be compensated for by reducing the distance between spots to obtain adequate coverage without hot spots [3]. The energy layer separation was set to be constant in RDX, whereas it was varying with energy based on the incident energy spread of the beam in Eclipse. Once the uncollimated plans were generated, collimated plans were created in RDX using the same gantry angles, beam spots, and energy layers for comparison purposes, but future collimated plans outside the scope of this study might be planned using optimized beam angles. Although the DCS is theoretically able to collimate each beam spot, only spots around the target boundary were collimated in practice because the beam spots inside the target do not need collimation. As shown by Wang et al [19], the use of a collimation requires considering a target expansion for spot placement to ensure the target coverage: in this study, a 6-mm margin was chosen because that is the maximum displacement of the beamlet centroid by the DCS obtained through simulation by Gelover et al [17].

Evaluation Tools

The cumulative DVH curves of the target and OARs were used for initial evaluation. A focus on either the Dmean or the Dmax dose (D1%) was made depending on the OAR, with D1% corresponding to 1% of the organ volume on the cumulative DVH. The PBSTV coverage was assessed using D95% of each individual PBSTV volume. The homogeneity index (HI) and conformity index (CI 50%) were calculated for further evaluation of the treatment plans. The HI is defined as the difference between the dose received by at least 95% of the PBSTV volume and the dose received by at least 5% of the PBSTV, normalized to the PBSTV Dmean: (D95% − D5%)/Dmean. The CI 50% is the ratio between the patient volume receiving at least 50% of the prescribed dose and the PBSTV volume. The PBSTV coverage and HI were chosen to quantify the PBSTV dose distribution similarity between the uncollimated and collimated plans, whereas the CI 50% was chosen to quantify any improvement in dose distribution conformity. For OARs, Dmean or D1% are reported. In the analysis, an uninvolved structure means the part of the structure that was outside any planning target volume.

Normal tissue complication probability (NTCP) was calculated for the parotid glands, esophagus, oral cavity, and cord using the Lyman-Kutcher-Burman model [20]; the parameters for the computation were summarized by Luxton et al [20] and Jakobi et al [21].

The RayStation treatment planning system (version 3.99, RaySearch Laboratories, Stockholm, Sweden) was used to render dose distributions and compute the DVH for the both uncollimated and collimated plans.

Statistical significance was evaluated with a Student t test to provide a 95% confidence interval (95% CI) and a P value.

Results

Target Coverage

The equivalence of target coverage in the uncollimated and collimated plans was verified through the comparison of the PBSTV DVH, the 95% PBSTV coverage, and the HI. Although most patients have more than one target volume, for simplicity, only the comparison of the PBSTV composite DVH is shown in Figures 2 and 3. The PBSTV DVHs were nearly identical. Only small discrepancies were observed (maximum difference of 1.4%) for the PBSTV coverage and a close HI (maximum difference of 0.02) as reported in Tables 2 and 3 for all patients.

Figure 2.

Figure 2.

Dose-volume histograms of the pencil beam-scanning planning target volumes composite and relevant organs at risk for patients P1 to P6; the uncollimated plan is in solid lines and the collimated plan is in dashed line. * indicates that only the uninvolved portion of the organ at risk was considered.

Table 2.

Homogeneity index (HI) and conformity index based on 50% of the prescribed dose (CI 50%) for each target (pencil beam scanning planning target volumes [PBSTV]) of all patients and their relative difference to the uncollimated plan.

Patient No.
Treatment
PBSTV 1
PBSTV 2
PBSTV 3
PBSTV 4
PBSTV 5
HI
CI 50%
HI
CI 50%
HI
CI 50%
HI
CI 50%
HI
CI 50%
P1 Uncol 0.12 3.87 0.06 7.15
Col 0.11 3.59 0.07 6.64
P2 Uncol 0.19 2.75 0.18 3.88 0.09 6.34 0.07 6.65
Col 0.20 2.69 0.19 3.82 0.09 6.28 0.08 6.62
P3 Uncol 0.04 2.94 0.09 3.34 0.09 6.79 0.07 21.33 0.06 31.26
Col 0.06 2.71 0.10 3.09 0.09 6.31 0.08 19.92 0.06 29.37
P4 Uncol 0.13 3.93 0.09 7.22
Col 0.13 3.41 0.11 6.31
P5 Uncol 0.09 4.58 0.06 28.00
Col 0.09 4.38 0.06 26.76
P6 Uncol 0.06 2.22
Col 0.08 2.01
P7 Uncol 0.05 1.93 0.05 15.36
Col 0.06 1.68 0.05 13.49
P8 Uncol 0.11 2.65
Col 0.11 2.34
P9 Uncol 0.06 2.79 0.11 3.99 0.04 14.68
Col 0.07 2.48 0.10 3.48 0.04 12.74
P10 Uncol 0.16 1.92
Col 0.17 1.72

Abbreviations: Uncol, uncollimated; Col, collimated.

Dose-Conformity Improvement and Organ-at-Risk Dose Reduction

Dose distributions from the uncollimated and collimated plans and the difference for one patient (P5) are shown in Figure 4. The improvement in target conformity can be directly observed in the dose difference in Figure 4c. The values of CI 50% are reported in Tables 2 and 3 to demonstrate the improvement in conformity afforded by the DCS as well as the variability depending on the target volume and shape. The conformity improvement by the DCS was also observed through the significant reduction of the mean dose to the 10-mm ring surrounding the PBSTV, yielding an average decrease of 7.1% (95% CI, 4.4% to 9.8%; P value < .001).

Figure 4.

Figure 4.

(a) Uncollimated plan, (b) collimated plan, and (c) relative difference (%) between the 2 dose distributions for 1 patient (P5). The target is shown in black, the 10-mm normal tissue margin is shown in white, the right uninvolved parotid gland is shown in blue, and the uninvolved larynx is shown in green.

Figure 3.

Figure 3.

Dose-volume histograms of the pencil beam-scanning planning target volumes composite and relevant organs at risk for patients P7 to P10; the uncollimated plan is in solid lines and the collimated plan is in dashed line. * indicates that only the uninvolved portion of the organ at risk was considered.

The collimated and uncollimated plans were also compared in the DVHs of selected OARs, as shown in Figures 2 and 3. The collimated plans showed reduced Dmean or D1% to OARs adjacent to the target compared with the uncollimated plan. The values for selected OARs are reported in Tables 2 and 3. For each patient, a graphical summary of the OAR having a Dmean or D1% difference > ± 2% (arbitrary value) between the uncollimated and collimated plan is shown in Figures 5 and 6.

Table 3.

The mean or maximum dose to some relevant organs at risk for both the uncollimated (Uncol) and collimated (Col) plans and their relative difference (Rel diff) to the uncollimated plan. The median, minimum, and maximum of the relative difference are also shown.

Patient No.
Treatment
Mean dose (cGy)
Max dose (cGy)
Esophagus
Mandible*
Oral Cavity*
Larynx*
Parotid*
Submandibular gland*
Ear mid
Brain stem
Cord
Lt
Rt
Lt
Rt
Lt
Rt
P1 Uncol 2173 1600 465 2479 3152 858 3468 2917 3661 2435
Col 1986 1262 311 2449 2835 839 2238 2964 3202 1566
Rel diff (%) −8.6 −21.1 −33.1 −1.2 −1.8 −2.2 −35.5 1.6 −12.5 −35.7
P2 Uncol 2329 3200 2985 3065 2018 2950 2467 4713 4036
Col 2075 3050 2827 2889 1980 2993 2296 4254 3854
Rel diff (%) −10.9 −4.7 −5.3 −5.7 −1.9 1.5 −6.9 −9.7 −4.5
P3 Uncol 2525 3771 2102 2699 1395 1672 219
Col 2363 3628 1814 2563 1388 1585 37
Rel diff (%) −6.4 −3.8 −13.7 −5.0 −0.5 −5.2 −83.1
P4 Uncol 1110 2587 1111 2165 1439 3331 5689 4462 4691 3404
Col 845 2198 873 1979 1130 2978 4149 3646 4509 2870
Rel diff (%) −23.9 −15.0 −21.4 −8.8 −21.5 −10.6 −27.1 −18.3 −3.9 −15.7
P5 Uncol 958 164 3960 2285 4086 3371 2486 3899
Col 727 154 3896 2135 3505 2757 2443 3953
Rel diff (%) −24.1 −6.1 −1.6 −6.6 −14.2 −18.2 −1.7 1.4
P6 Uncol 1716 155 684 3504 205 1388
Col 1494 75 478 3496 116 903
Rel diff (%) −12.9 −51.6 −30.1 −0.2 −43.4 −34.9%
P7 Uncol 872 282 972 3654 4524 616
Col 795 195 388 3293 3887 125
Rel diff (%) −8.8 −30.9 −60.1 −9.9 −14.1 −79.7%
P8 Uncol 2631 2335 509 2767 3510 120 313 994
Col 2160 2325 505 2736 3256 73 269 844
Rel diff (%) −17.9 −0.4 −0.8 −1.1 −7.2 −39.2 −14.1% −15.1%
P9 Uncol 846 864 3685 1393 1360 1882
Col 681 645 3499 1160 1340 1795
Rel diff (%) −19.5 −25.3 −5.0 −16.7 −1.5% −4.6%
P10 Uncol 551 260 717 3850 7112 1676 626
Col 401 146 492 3173 7151 1208 148
Rel diff (%) −27.2 −43.8 −31.4 −17.6 0.5 −27.9% −76.4%
Median −11.9 −15.0 −32.0 −7.2 −5.2 −15.9 −15.4 −9.7 −25.3
Min −6.4 −3.8 −0.4 −0.8 −0.2 −5.0 1.6 −1.5 1.4
Max −24.1 −27.2 −51.6 −60.1 −21.5 −35.5 −43.4 −27.9 −83.1

Note: An asterisk (*) Indicates that only the uninvolved part of the organ at risk was considered.

Abbreviations: cGy, centigray; Max, maximum; Lt, left; Rt, right; mid, middle; min, minimum.

Figure 5.

Figure 5.

Absolute mean dose (Dmean) or maximum dose (D1%) to some relevant organs at risk (OARs) for both the uncollimated (circle) and collimated (triangle) plans, and the difference normalized to the uncollimated plan (grey bar) for patients P1 to P4. P1 to P4 are patients with bilateral target, the other with unilateral targets.* indicates that only the uninvolved part of the OAR was considered. Abbreviations: Esoph, esophagus; Mandib, mandible; OC, oral cavity; Subm, submandibular gland; Par, parotid; BS, brain stem; ON, optic nerve; TL, temporal lobe; EarM, ear mid; TMJ, temporomandibular joint; Lt, left; Rt, right.

Figure 6.

Figure 6.

Absolute mean dose (Dmean) or maximum dose (D1%) to some relevant organs at risk (OARs) for both the uncollimated (circle) and collimated (triangle) plans, and the difference normalized to the uncollimated plan (grey bar) for patients P5 to P10. P1 to P4 are patients with bilateral target, the other with unilateral targets. * indicates that only the uninvolved part of the OAR was considered. Abbreviations: Esoph, esophagus; Mandib, mandible; OC, oral cavity; Subm, submandibular gland; Par, parotid; BS, brain stem; ON, optic nerve; TL, temporal lobe; EarM, ear mid; TMJ, temporomandibular joint; Lt, left; Rt, right.

Toxicity Reduction through Radiobiological Modeling

The NTCP values for the toxicities affected by the DCS are presented in Table 4. The largest improvements of the parotid gland NTCP for the risk of xerostomia were observed in patients P1, P5, and P8. The largest improvements of the esophagus NTCP for acute toxicity were observed in patients P1 and P2. A noticeable improvement of the larynx NTCP for the risk of edema was only observed in patient P7. The largest improvements of the larynx NTCP for aspiration pneumonia were observed in patient P2, P4, and P7. Some NTCP values were increased after collimation, but none by > 0.7%.

Table 4.

Absolute normal tissue complication probability (NTCP; %) for both the uncollimated (Uncol) and collimated (Col) plans, and their absolute difference NTCPcollimated − NTCPuncollimated (relative difference to the uncollimated plan between parentheses) for relevant organs at risk.

Patient No.
Treatment
Parotid (%)
Esophagus (%)
Larynx (edema) (%)
Larynx (aspiration) (%)
Left
Right
P1 Uncol 92.7 0.23 10.4 25.3 17.8
Col 81.9 0.20 8.6 25.7 17.5
Difference −10.8 (−11.7) −0.03 (−13.5) −1.8 (−17.4) 0.4 (1.7) −0.3 (−1.9)
P2 Uncol 100 31.6 9.5 87.8 39.3
Col 100 28.5 7.5 88.1 37.1
Difference 0.0 (0.0) −3.1 (−9.8) −2.0 (−20.9) 0.3 (0.3) −2.3 (−5.7)
P3 Uncol 0.76 2.9 10.6 0.18 3.2
Col 0.74 2.0 9.5 0.03 2.6
Difference −0.02 (−2.6) −0.9 (−32.0) −1.1 (−10.3) −0.15 (−82.8) −0.6 (−21.3)
P4 Uncol 0.52 96.0 2.8 9.8 14.4
Col 0.08 87.6 1.9 12.7 12.7
Difference −0.44 (−84.3) −8.4 (−8.7) −0.9 (−30.7) 2.9 (29.2) −1.7 (−12.1)
P5 Uncol 41.9 1.4 81.6 55.3
Col 32.2 1.4 81.8 54.7
Difference −9.7 (−23.0) 0.0 (2.2) 0.2 (0.2) −0.6 (−1.0)
P6 Uncol 99.3 4.6 1.9 4.5
Col 99.3 3.6 1.5 3.7
Difference 0.1 (0.1) −0.9 (−20.5) −0.4 (−18.6) −0.8 (−17.2)
P7 Uncol 8.8 5.7
Col 3.6 3.4
Difference −5.2 (−59.4) −2.4 (−41.2)
P8 Uncol 44.3 90.5 0.3 74.2 12.7
Col 42.0 79.2 0.3 74.9 12.7
Difference −2.3 (−5.2) −11.3 (−12.5) 0.0 (−3.3) 0.7 (1.0) 0.1 (0.4)
P9 Uncol 100 0.27 4.4 5.2
Col 100 0.27 3.7 4.3
Difference 0.0 (0.0) 0.0 (0.0) −0.8 (−16.9) −0.9 (−17.8)
P10 Uncol 100 0.28 5.2 4.5
Col 100 0.27 4.5 3.7
Difference 0.0 (0.0) −0.01 (−3.6) −0.7 (−12.9) −0.9 (−18.8)

Discussion

The collimated plans maintained the target coverage of the uncollimated plans. Noticeable improvement of the OAR sparing was observed for most of the 10 head and neck plans. For parotid sparing, 5 of 10 patients benefitted from the addition of collimation. All patients benefitted from the addition of collimation in sparing of the uninvolved submandibular gland, because the Dmean was systematically high (> 34.7 Gy) in uncollimated plans and greatly reduced by the collimation (from −5.0% to −35.5%). Four of 10 patients benefitted from the addition of collimation in sparing the uninvolved mandible, oral cavity, and the larynx and esophagus because the Dmean values close to, or slightly beyond, the usual clinical constraints were reduced. Only 1 patient had a relatively minor benefit from collimation. In that case (P6), the dose to most OARs with the uncollimated plan were already sufficiently low relative to the clinical constraints used for the treatment planning; therefore, the small dosimetric improvement was not considered sufficient to advise the use of the DCS. Regarding the radiobiological metrics, Nguyen et al [22] showed that it is more reliable to focus on the relative difference of the NTCP values instead of the absolute difference because of NTCP model uncertainties. Based on this argument, the results provide evidence that the DCS affords a significant reduction in different complication probabilities: for instance, 6 of 9 plans had great relative reductions in xerostomia risk (NTCP to the parotids), suggesting potential clinical significance.

Results from the 10 patients also indicate that the therapeutic benefit of the DCS for the treatment of head and neck cancer by SSPT is dependent on beam orientation as well as properties of the target, such as its shape, volume, and location relative to the OAR. For instance, P9 had a great conformity improvement for each PBSTV, whereas P2 demonstrated little improvement in conformity. The major difference between P2 and P9 was the extent of the target; P2 had a bilateral PBSTV with a volume close to the full neck compared with the unilateral PBSTV of P9 with a volume nearly 4 times smaller than that for P2. The variability in OAR sparing depending on these parameters was also described by Jakobi et al [21] when comparing intensity-modulated photon therapy to intensity-modulated proton therapy to identify a subgroup of patients who would most benefit from proton therapy. In the present study, unilateral targets generally benefitted the most from collimation by normal tissue sparing; however, fewer OARs are likely to be spared with unilateral targets than with bilateral targets because of the lesser extent of the disease. These bilateral targets are also challenging because of target coverage robustness from the splitting of the target structure to benefit from the dynamic collimation for the OARs in the middle of the neck.

The benefit of DCS remains theoretical because this was an in silico study. In addition, this study evaluated only a subset of head and neck plans, which may not be representative of all head and neck tumors encountered in clinical practice or in other tumor locations. Many variables influence target conformity, such as the energy layer, spot spacing, spot size, and the use of range shifter. However, those values were chosen to be representative of clinical plans and were identical between the uncollimated and collimated plans, except for the spot size and spot spacing, because the collimation changes the spot geometry (sharper penumbra and asymmetry) requiring a slight spot-spacing reduction. A previous study based on brain tumor plans [12] can be used as a point of comparison to detail the behavior of a DCS-based plan, depending on the plan and target properties. For the head and neck plans, improvement in Dmean to the 10-mm ring of normal tissue surrounding the target was slightly more than one-half of that obtained for the brain plans (−7.1% versus −13.7%). The main difference was that the brain tumors were smaller and more regularly shaped. The principal challenges with the head and neck plans included large target volumes with little ability to spare normal tissue, bulky tumors wrapped around the OAR with part of the beam crossing through the OAR to reach the target, and OARs partially intersecting the target or directly at the target boundary (Supplementary Data 1, 10.14338/IJPT-15-00036.1.S1 (2.6MB, pdf) ). A minor challenge was related to a slight increase in the entrance dose by the DCS [17]. Given these challenges, the DCS did not yield uniform improvement in sparing for every OAR for each plan. Although general conclusions can be drawn from this study, the use of the DCS must still be considered for each individual plan by weighing the clinical benefit against the additional complexity. The DCS can be considered an optional collimation system offering plan improvement dependent on the target complexity and at a low-cost for treatment delivery time (theoretically, between 29 s and 84 s, depending on the target size and the number of beams, assuming a continuous motion of the trimmer blades at a velocity of 0.35 m/s).

Future developments with an adaptive method for spot placement using contour scanning and eliminating spots too far away from the target boundary are expected to further decrease the dose to normal tissue and to partially solve the limited improvement for OARs surrounded by the target in patients with head and neck cancer.

Conclusion

In SSPT, the DCS generally affords better target conformity, which results in an improvement in OAR sparing in the head and neck region while preserving target coverage. However, the benefits of collimation and increased complexity should be considered for each patient. Patients with large bilateral targets or OARs surrounded by the target showed the least benefit.

ADDITIONAL INFORMATION AND DECLARATIONS

Conflicts of interest: This research was supported by IBA (Louvain-la-Neuve, Belgium).

Acknowledgments: The authors acknowledge RaySearch Laboratories (Stockholm, Sweden) for providing the RayStation, a commercial treatment-planning system used for plan comparisons.

References

  • 1.Wang D, Dirksen B, Hyer DE, Buatti JM, Sheybani A, Dinges E, Felderman N, TenNapel M, Bayouth JE, Flynn RT. Impact of spot size on plan quality of spot scanning proton radiosurgery for peripheral brain lesions. Med Phys. 2014;41:121705. doi: 10.1118/1.4901260. [DOI] [PubMed] [Google Scholar]
  • 2.van de Water TA, Lomax AJ, Bijl HP, Schilstra C, Hug EB, Langendijk JA. Using a reduced spot size for intensity-modulated proton therapy potentially improves salivary gland-sparing in oropharyngeal cancer. Int J Radiat Oncol Biol Phys. 2012;82:e313–9. doi: 10.1016/j.ijrobp.2011.05.005. [DOI] [PubMed] [Google Scholar]
  • 3.Widesott L, Lomax AJ, Schwarz M. Is there a single spot size and grid for intensity modulated proton therapy? simulation of head and neck, prostate and mesothelioma cases. Med Phys. 2012;39:1298–308. doi: 10.1118/1.3683640. [DOI] [PubMed] [Google Scholar]
  • 4.Gillin MT, Sahoo N, Bues M, Ciangaru G, Sawakuchi G, Poenisch F, Arjomandy B, Martin C, Titt U, Suzuki K, Smith AR, Zhu XR. Commissioning of the discrete spot scanning proton beam delivery system at the University of Texas MD Anderson Cancer Center, Proton Therapy Center, Houston. Med Phys. 2010;37:154–63. doi: 10.1118/1.3259742. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Lin L, Kang M, Solberg TD, Ainsley CG, McDonough JE. Experimentally validated pencil beam scanning source model in TOPAS. Phys Med Biol. 2014;59:6859–73. doi: 10.1088/0031-9155/59/22/6859. [DOI] [PubMed] [Google Scholar]
  • 6.Lomax AJ. Intensity modulated proton therapy and its sensitivity to treatment uncertainties, 1: the potential effects of calculational uncertainties. Phys Med Biol. 2008;53:1027–42. doi: 10.1088/0031-9155/53/4/014. [DOI] [PubMed] [Google Scholar]
  • 7.Kralik JC, Xi L, Solberg TD, Simone CB, Ii, Lin L. Comparing proton treatment plans of pediatric brain tumors in two pencil beam scanning nozzles with different spot sizes. J Appl Clin Med Phys. 2015;16:5389. doi: 10.1120/jacmp.v16i6.5389. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Bues M, Newhauser WD, Titt U, Smith AR. Therapeutic step and shoot proton beam spot-scanning with a multi-leaf collimator: a Monte Carlo study. Radiat Prot Dosimetry. 2005;115:164–9. doi: 10.1093/rpd/nci259. [DOI] [PubMed] [Google Scholar]
  • 9.Dowdell SJ, Clasie B, Depauw N, Metcalfe P, Rosenfeld AB, Kooy HM, Flanz JB, Paganetti H. Monte Carlo study of the potential reduction in out-of-field dose using a patient-specific aperture in pencil beam scanning proton therapy. Phys Med Biol. 2012;57:2829–42. doi: 10.1088/0031-9155/57/10/2829. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Hyer DE, Hill PM, Wang D, Smith BR, Flynn RT. A dynamic collimation system for penumbra reduction in spot-scanning proton therapy: proof of concept. Med Phys. 2014;41:091701. doi: 10.1118/1.4837155. [DOI] [PubMed] [Google Scholar]
  • 11.Hyer DE, Hill PM, Wang D, Smith BR, Flynn RT. Effects of spot size and spot spacing on lateral penumbra reduction when using a dynamic collimation system for spot scanning proton therapy. Phys Med Biol. 2014;59:N187–96. doi: 10.1088/0031-9155/59/22/N187. [DOI] [PubMed] [Google Scholar]
  • 12.Moignier A, Gelover E, Wang D, Smith B, Flynn R, Kirk M, Lin L, Solberg T, Lin A, Hyer D. Theoretical benefits of dynamic collimation in pencil beam scanning proton therapy for brain tumors: dosimetric and radiobiological metrics. Int J Radiat Oncol Biol Phys. 2015. [DOI] [PubMed]
  • 13.Chan AW, Liebsch NJ. Proton radiation therapy for head and neck cancer. J Surg Oncol. 2008;97:697–700. doi: 10.1002/jso.21013. [DOI] [PubMed] [Google Scholar]
  • 14.Lin L, Ainsley CG, McDonough JE. Experimental characterization of two-dimensional pencil beam scanning proton spot profiles. Phys Med Biol. 2013;58:6193–204. doi: 10.1088/0031-9155/58/17/6193. [DOI] [PubMed] [Google Scholar]
  • 15.Flynn RT, Barbee DL, Mackie TR, Jeraj R. Comparison of intensity modulated x-ray therapy and intensity modulated proton therapy for selective subvolume boosting: a phantom study. Phys Med Biol. 2007;52:6073–91. doi: 10.1088/0031-9155/52/20/001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Flynn RT, Bowen SR, Bentzen SM, Rockwell Mackie T, Jeraj R. Intensity-modulated x-ray (IMXT) versus proton (IMPT) therapy for theragnostic hypoxia-based dose painting. Phys Med Biol. 2008;53:4153–67. doi: 10.1088/0031-9155/53/15/010. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Gelover E, Wang D, Hill PM, Flynn RT, Gao M, Laub S, Pankuch M, Hyer DE. A method for modeling laterally asymmetric proton beamlets resulting from collimation. Med Phys. 2015;42:1321–34. doi: 10.1118/1.4907965. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Frank SJ, Cox JD, Gillin M, Mohan R, Garden AS, Rosenthal DI, Gunn GB, Weber RS, Kies MS, Lewin JS, Munsell MF, Palmer MB, Sahoo N, Zhang X, Liu W, Zhu XR. Multifield optimization intensity modulated proton therapy for head and neck tumors: a translation to practice. Int J Radiat Oncol Biol Phys. 2014;89:846–53. doi: 10.1016/j.ijrobp.2014.04.019. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Wang D, Smith BR, Gelover E, Flynn RT, Hyer DE. A method to select aperture margin in collimated spot scanning proton therapy. Phys Med Biol. 2015;60:N109–19. doi: 10.1088/0031-9155/60/7/N109. [DOI] [PubMed] [Google Scholar]
  • 20.Luxton G, Keall PJ, King CR. A new formula for normal tissue complication probability (NTCP) as a function of equivalent uniform dose (EUD) Phys Med Biol. 2008;53:23–36. doi: 10.1088/0031-9155/53/1/002. [DOI] [PubMed] [Google Scholar]
  • 21.Jakobi A, Bandurska-Luque A, Stutzer K, Haase R, Lock S, Wack LJ, Monnich D, Thorwarth D, Perez D, Luhr A, Zips D, Krause M, Baumann M, Perrin R, Richter C. Identification of patient benefit from proton therapy for advanced head and neck cancer patients based on individual and subgroup normal tissue complication probability analysis. Int J Radiat Oncol Biol Phys. 2015;92:1165–74. doi: 10.1016/j.ijrobp.2015.04.031. [DOI] [PubMed] [Google Scholar]
  • 22.Nguyen J, Moteabbed M, Paganetti H. Assessment of uncertainties in radiation-induced cancer risk predictions at clinically relevant doses. Med Phys. 2015;42:81–9. doi: 10.1118/1.4903272. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from International Journal of Particle Therapy are provided here courtesy of Particle Therapy Cooperative Group (PTCOG)

RESOURCES