Ultrasensitive and label-free biosensor for the detection of Plasmodium falciparum histidine-rich protein II in saliva

Gita V Soraya 1,5, Chathurika D Abeyrathne 1,2,3, Christelle Buffet 1,4, Duc H Huynh 2,3, Shah Mukim Uddin 1,2, Jianxiong Chan 1, Efstratios Skafidas 2,3, Patrick Kwan 1,3,✉,#, Stephen J Rogerson 1,4,✉,#
PMCID: PMC6877566  PMID: 31767887

Abstract

Malaria elimination is a global public health priority. To fulfil the demands of elimination diagnostics, we have developed an interdigitated electrode sensor platform targeting the Plasmodium falciparum Histidine Rich Protein 2 (PfHRP2) protein in saliva samples. A protocol for frequency-specific PfHRP2 detection in phosphate buffered saline was developed, yielding a sensitivity of 2.5 pg/mL based on change in impedance magnitude of the sensor. This protocol was adapted and optimized for use in saliva with a sensitivity of 25 pg/mL based on change in resistance. Further validation demonstrated detection in saliva spiked with PfHRP2 from clinical isolates in 8 of 11 samples. With a turnaround time of ~2 hours, the label-free platform based on impedance sensors has the potential for miniaturization into a point-of-care diagnostic device for malaria elimination.

Subject terms: Medical research, Biomarkers

Introduction

Malaria caused by intraerythrocytic Plasmodium parasites remains a significant public health threat. P. falciparum is responsible for most severe malaria illness and almost all deaths1,2 which occur mainly in young children in the World Health Organization’s African Region3.

In recent years the burden of malaria has decreased, through improvements in treatment and prevention. From 2010 to 2015, global malaria incidence and numbers of deaths decreased by 21% and 29% respectively3. With the reduction of transmission rates and the malaria burden, elimination agendas have been pushed with the aim to end local transmission of the disease in at least 35 countries by the year 20304.

Clinical malaria diagnosis relies on light microscopy (LM) for visual confirmation of parasites or rapid diagnostic tests (RDTs) to detect parasite antigens using lateral-flow technology5. A common RDT target is the P. falciparum histidine-rich protein II (PfHRP2) antigen, a multiplet protein6,7 produced exclusively in the P. falciparum parasite cytoplasm and exported to the parasitized erythrocyte membrane8. The PfHRP2 protein is readily detectable in whole blood, serum, plasma, urine9 and saliva10,11 samples of infected patients.

For malaria elimination, current diagnostics need to be adapted to detect increasing numbers of asymptomatic parasite carriers, with the essential and desirable target sensitivity in the context of population screening at 20 parasites/µL blood and ≤5 parasites/µL blood respectively12. Current RDTs however have a relatively low sensitivity. In a study assessing the relationship between antigen concentration and parasite density, a minimum of 4 ng/mL of PfHRP2 was required to obtain 95% positive results in a panel of malaria-infected blood samples with a parasite density of 200 parasites/µL13. A more recent study14 assessing the current best-in-class PfHRP2 RDTs according to the WHO-Foundation for Innovative New Diagnostics panel15 found an analytical sensitivity of 0.8 ng/mL. It can be inferred from these studies that to achieve the target sensitivity for elimination, PfHRP2 diagnostic tests need to be 1–2 logs lower than achievable by current RDTs.

While PfHRP2 ELISAs1618, PCR19,20 and loop-mediated isothermal amplification (LAMP)21 assays have shown superior sensitivity to RDTs for detection of low-density infections, the platforms are slow and technically complex assays typically performed by highly-skilled technicians in centralized laboratories. Implementation of these methods is therefore impractical in the low-resource settings of many countries striving for malaria elimination.

Additionally, detection of PfHRP2 in saliva is gaining substantial interest due to ease of collection, lower biohazard risk, and higher likelihood of testing compliance particularly in communities where blood collection poses cultural objections22,23. Although salivary PfHRP2 has been evaluated for the detection of low to high-density P. falciparum parasitemias10 no RDTs have been developed for this purpose, due to the lack of sensitivity. An in-house ELISA assay developed for detection of salivary PfHRP2 had a sensitivity of 0.17 pg/mL7, but the laboratory-based nature of the platform hinders effective field implementation.

Impedimetric biosensors are promising options to help close current diagnostic gaps, due to their high sensitivity, low cost, and amenability to miniaturization. Table 1 summarises previous reports on the development of ultrasensitive PfHRP2 biosensors. The biosensors were used in direct or sandwich immunoassay formats to target PfHRP2 protein in various sample matrices. To achieve high-level sensitivity, most of these sensors required additional labeling and signal amplification, which in the long run may incur additional costs during miniaturization.

Table 1.

Comparison between previously developed PfHRP2 immunosensors.

PfHRP2 Immunosensors
Assay Format Assay Time Labelling or Signal Amplification Immunosensor Type Lowest concentration of PfHRP2 detected Reference
Sandwich immunoassay ~ 30 minutes ALP SPEs with MWCNT and AuNPs characterized with cyclic voltammetry 8 ng/mL in DEA buffer 43
Direct immunoassay ~ 1 hour Label-free Piezoelectric sensor characterized with cyclic voltammetry and EIS 12 ng/mL in Tris buffer 37
Direct immunoassay NA Redox couple Faradaic EIS using Cu-doped ZnO electrospun nanofibers 6 ag/ml in PBS buffer 44
Sandwich immunoassay 1 minute CNF grown on NMBs Immunochromatography + resistivity measurements 0.01 ng/mL in PBS buffer 45
Sandwich immunoassay >2 hours HRP Amperometric; Electrochemical magneto immunosensor coupled with magnetic nanoparticles 0.36 ng/mL in spiked serum 46
Sandwich immunoassay >2 hours HRP and AuNP Cyclic voltammetry on SPEs 36 pg/mL in milk PBS buffer; 40 pg/mL in spiked Serum 38
Sandwich immunoassay ~1 hour MB, Ru(NH3)63+ and TCEP ECC redox cycling scheme for signal amplification on 3-electrode sensor 10 fg/mL in spiked plasma; 18 fg/mL in spiked whole blood 39
Direct Immunoassay 1–2 hours Label-free Non-faradaic IDE Sensors 2.5 pg/mL in PBS buffer; 25 pg/mL in spiked saliva This work

ALP = alkaline phosphatase, SPE = screen-printed electrodes, MWCNT = multiwall carbon nanotube, DEA = diethanolamine buffer, EIS = electrochemical impedance spectroscopy, Cu = copper, ZnO = zinc oxide, PBS = phosphate buffered saline, CNF = carbon nanofiber, NMBs = glass microballoons, HRP = horseradish peroxidase, AuNP = gold nanoparticles, MB = methylene blue, Ru(NH3)63+ = hexaamineruthenium(III) chloride, TCEP = tris(2-carboxyethyl)phosphine hydrochloride, ECC = electrochemical-Chemical, IDE = interdigitated electrodes.

In this study, we aimed to develop an interdigitated electrode (IDE) sensor for impedimetric detection of PfHRP2 at low concentrations suitable for malaria elimination, with a focus on the utilization of saliva as sample medium. Compared with other technologies, the IDE sensor geometry has demonstrated high level of sensitivity and specificity for label-free detection of various targets including nucleic acids2427, cells2830, and proteins3133. Here, the detection platform utilized anti-PfHRP2 monoclonal antibodies (MAbs) immobilized on sensor surface as capture probes towards circulating PfHRP2 protein. The application of periodic small AC voltage allowed measurement of the sensor impedance, defined as a complex of the circuitry’s resistance and capacitance. Specific detection of surface-bound PfHRP2 on the sensor was achievable through frequency-dependent characterization of the impedimetric changes. The platform demonstrates promising ultrasensitive detection of PfHRP2 protein in saliva. Fabricated using low-cost techniques, the platform is amenable to future automatization and miniaturization for point-of-care applications.

Results

Sensor preparation and platform development

IDE sensors were fabricated using UV-lithography on borosilicate glass wafers as previously described27, with modification. An additional SiO2 layer was deposited at a thickness of 25 nm. Each sensor consisted of paired electrode arrays with finger length (L) of 980 µm, finger width (W) of 8 µm and gap width (G) of 8 µm (Fig. 1a). The sensors were functionalized according to our established protocol30,34 by covalent immobilization of anti-PfHRP2 antibodies on the surface, followed by blocking to reduce non-specific binding to the antibodies. Optimization results of capture antibody concentration are provided in Supplementary Figure 1.

Figure 1.

Figure 1

Sensor preparation and platform development. (a) Fabricated IDE sensor array and schematic of the IDE sensing area geometry. L = length, W = width, and G = gap of the working electrode. (b) Circuit model and measurement setup used for characterization of PfHRP2 capture. The sensor was set up as a resistor (R) in series with a capacitor (C), with the associated input voltage (Vin), output voltage (Vout) and reference resistor (Rref) labeled accordingly. The cross-sectional view of the measurement setup depicts detection of PfHRP2 protein in a sample medium with the sensing area designated by pressure-sensitive adhesive (PSA) tape. The IDE sensor is composed of interdigitated electrodes with adjoining electrode contact pads, and borosilicate glass substrate. Probe electrodes were placed on the contact pads to deliver excitation current to, and to measure electrical signals from, the sensors. (c) The two phases of platform development.

Impedance of the sensors was measured using our previously established circuit setup30,34 shown in Fig. 1b, in which the sensor is represented as a resistor (R) in series with a capacitor (C) connected in series with a 1 kΩ reference resistor (Rref). A function generator provided the input sinusoidal AC excitatory signal (Vin) at 20 mV peak-to-peak voltage (Vpp) at multiple frequencies (1,10, 20, and 40 kHz). Sensors were measured in wet-state (in PBS 1×) at time points T1 (baseline measurement obtained before sample incubation) and T2 (after sample incubation and washing) to detect binding associated change occurring between the two points. A lock-in amplifier recorded the amplitude of the output voltage (Vout) and phase across the Rref, which was used for the acquisition of frequency (ω) dependent impedimetric parameters including impedance magnitude (Z), capacitance (C), and resistance (R) using the following equations34

Vout/Vin=Rref/(Z+Rref) 1
Z=Rj/(wC) 2

Development of the sensor platform was divided into two phases (Fig. 1c). In Phase I, phosphate buffered saline (PBS) samples spiked with PfHRP2 were used. The optimal detection frequency was determined, and parameters for specific PfHRP2 detection were established and used to obtain the analytical sensitivity. The protocol was then applied in Phase II in which saliva samples spiked with PfHRP2 derived from cultured cell lines were used. Optimal parameters for saliva-based detection were determined to obtain the sensitivity. Specificity was then assessed using a panel of saliva samples spiked with isolate-derived PfHRP2.

Phase I - Detection of PfHRP2 protein in PBS buffer

Among the range of excitation frequencies applied (1 kHz to 40 kHz), 20 kHz was found to elicit the most significant difference in the change in impedance magnitude between PBS samples with and without spiked PfHRP2 (Fig. 2a). This frequency was used for all subsequent experiments. The impedimetric parameters (impedance magnitude, impedance phase, resistance and capacitance) were also compared in their ability to differentiate PBS samples with and without spiked PfHRP2, and the change in impedance magnitude was found to be the optimal parameter for this purpose (Fig. 2b).

Figure 2.

Figure 2

Detection of PfHRP2 in PBS buffer using the sensors. (a) Optimal frequency for the detection of PfHRP2 (2.5 pg/mL) in circuit setup. (b) Parameters for PfHRP2 (2.5 pg/mL) detection in PBS buffer at 20 kHz frequency. (c) Sensor response towards various concentrations of PfHRP2 protein in PBS buffer. Change in impedance associated with serial dilution of PfHRP2 from 0.25 pg/mL to 250 ng/mL was measured at the 20 kHz optimal frequency. Figures show mean ± standard error of the mean (SEM), (a,b) n = 3 sensors per group, (c) n = 4–5 sensors per concentration, p-values calculated with Welch’s two-tailed t-test (a,b) or by one-way ANOVA followed by Dunnett’s multiple comparisons test with blank PBS used as the comparator (c). *p ≤ 0.05, **p ≤ 0.01, ***p ≤ 0.001, ****p ≤ 0.0001, ns = not significant. Red bars = sensors incubated in PBS buffer, blue bar = sensors incubated in PfHRP2. All incubations were performed for 1 hour within a parafilm-wrapped petri dish to avoid sample evaporation.

The finding that frequency-dependent characterization of PfHRP2 binding is optimal within the frequency range of 10–100 kHz is consistent with similar previous studies30,35, and supports the notion that the signal is primarily due to the changes in bulk medium conductivity as a result of dipole accumulation36 from the surface-bound PfHRP2.

The optimal excitation frequency and impedimetric parameter were then used to determine the limit of PfHRP2 detection in PBS medium. The limit of detection, defined as the concentration at which the test sample exhibited a signal change significantly distinct from the blank, was found to be 2.5 pg/mL (Fig. 2c). Starting at that PfHRP2 concentration, a significant increase in the impedance signal was observed, indicating the sensor response towards the solid-state capture. No signs of saturation or prozone effect were observed at the higher end of the tested concentrations. This detection limit was concordant with our previous finding that the IDE sensors were able to detect another protein at the minimum concentration of 2.9 pg/mL33.

The Phase I results provided proof-of-concept data for ultrasensitive PfHRP2 protein detection using the label-free electrical biosensor platform. The detection limit was 3 logs lower than that of the average RDTs (~ 4 ng/mL)13, and is also lower than previously developed label-free electrical biosensor for PfHRP2, which reported a sensitivity of 12 ng/mL37.

Phase II - Detection of PfHRP2 protein in saliva

Next, the ability of the platform to detect culture-derived PfHRP2 in human saliva samples was evaluated to support future real-world applications. Saliva can be collected non-invasively, making it an ideal biospecimen for population screening. The protocol in Phase I was modified for use in saliva samples owing to differences in the ionic composition and protein content between PBS and saliva, and the need to reduce degradation of PfHRP2 by protease enzymes present in saliva7. The modification included more stringent blocking protocol (Supplementary Figure 2), longer incubation time of 2 hours (Supplementary Figure 3), sample pretreatment with protease inhibitor and re-characterization of the impedimetric detection parameters. The optimized parameters were then used to identify the limit of detecting PfHRP2 in saliva and to validate the protocol using saliva samples spiked with PfHRP2 expressed by different P. falciparum isolates.

Unlike findings in Phase I that used PBS, change in sensor resistance was found to be a more specific parameter in differentiating saliva samples with and without spiked PfHRP2 (Fig. 3b). The finding that optimal differentiation is parameter-specific is one of the advantages of impedance-based measurements, as it permits optimal detection despite differences in ionic composition across sample types.

Figure 3.

Figure 3

Detection of PfHRP2 protein in saliva. (a) Optimization of protease inhibitor (PI) pretreatment for saliva samples. (b) Change in sensor resistance is the optimal impedimetric parameter for detection of PfHRP2 (25 pg/mL) protein in saliva. Red bar = sensors incubated in unspiked saliva, blue bar = sensors incubated in 2.5 pg PfHRP2 per mL of saliva. Bars represent mean ± SEM, n = 7 sensors per group, p-values calculated with Welch’s two-tailed t-test. (c) Sensor response towards various concentrations of PfHRP2 protein in saliva. A serial dilution of 2.5 pg/mL to 2.5 ng/mL PfHRP2 protein in saliva was measured. Bars represent mean ± SEM, n = 7–8 sensors per concentration. Statistical analysis by one-way ANOVA followed by Dunnett’s multiple comparisons test with blank saliva used as the comparator. *p ≤ 0.05, **p ≤ 0.01, ns = not significant. All incubations were performed for 2 hours within a parafilm-wrapped petri dish to avoid sample evaporation, and all measurements were performed at 20 kHz frequency.

The adapted parameters were used to determine the lowest limit of PfHRP2 detection, which was found to be 25 pg/mL (Fig. 3c). Starting at this concentration of PfHRP2, the change in sensor resistance was significantly different between the positive sample and the negative (unspiked) sample. In contrast to the PBS-based dynamic range assessment, signal saturation was observed earlier in the saliva concentration curve, at 2.5 ng/mL PfHRP2. This saturation, which resulted in a less optimal concentration-dependent response, is most likely due to the presence of interfering proteins in the saliva matrix compared to the PBS, which may have been the cause of reduction in binding magnitude. The presence of multitude of substances in saliva can result in failure of antibody and ligand complex formation on the sensor surface, which may contribute to the reduced dose-dependent effect.

To further validate the protocol for real-world implementation, 11 saliva samples spiked with PfHRP2 (samples P1–9 are PfHRP2 present in supernatants of clinical parasite isolates, P10–11 are PfHRP2 isolated from culture supernatants of 3D7 and CS2 laboratory adapted parasites lines) at a concentration of 25 pg/mL were incubated with the sensor platform. Among the 11 samples tested, 8 resulted in significantly higher change in resistance (%ΔR) compared to the un-spiked saliva sample and the sample spiked with Plasmodium lactate dehydrogenase (PLDH) (Fig. 4). The variations among the sensor signals from the different PfHRP2 samples might be attributed to the differences in PfHRP2 protein size since previous characterizations have shown the protein to be of multiple bands ranging from 50–80 kDa7, although this assumption will need to be assessed further.

Figure 4.

Figure 4

IDE sensor performance for PfHRP2 detection in spiked-saliva samples. Saliva samples were spiked with culture media and PfHRP2 (blue bars, n = 3–4 sensors per group) or PLDH (grey bar, n = 3 sensors) at a uniform concentration of 25 pg/mL, and blank (un-spiked) (red bar, n = 11 sensors). P1–9 were saliva samples spiked with PfHRP2 in culture supernatants of clinical malaria isolates, P10–11 were spiked with PfHRP2 isolated from supernatants of laboratory lines 3D7 and CS2. Applied frequency 20 kHz, 2-hour incubation within a parafilm-wrapped petri dish to avoid sample evaporation.. Bars represent mean ± SEM. Statistical analysis by one-way ANOVA followed by Dunnett’s multiple comparisons test with blank saliva used as the comparator. *p ≤ 0.05, **p ≤ 0.01, ***p ≤ 0.001, ****p ≤ 0.0001.

Discussion

The sensor platform developed in this study represents a highly sensitive impedimetric sensor capable of label-free detection of surface-bound PfHRP2. Promising sensitivity was demonstrated with a PfHRP2 detection limit of 2.5 pg/mL in PBS, an order of 3 logs lower than current PfHRP2 RDTs. Furthermore, the study also demonstrated the feasibility of PfHRP2 detection in saliva with a detection limit of 25 pg/mL, in 8 out of 11 tested samples. Specific detection was achieved using an optimized, constant excitation frequency of 20 kHz across both PBS and saliva samples.

Differences in assay sensitivity are commonly observed in the development of PfHRP2 immunosensors as previous studies have also reported the variations in assay performance among various matrices38,39, with few studies examining saliva samples. Adaptation of the platform for other types of biospecimens, such as whole blood or plasma, would require re-optimization of impedimetric parameters, incubation time, and blocking protocol due to the differences in the ionic properties and protein content of different specimen types.

Saliva-based PfHRP2 detection is an attractive and potentially cost-effective testing approach. To date, quantitative data on salivary PfHRP2 in cases of low-density parasitemia is lacking, as previous salivary PfHRP2 quantification using commercial10 or in-house developed7 ELISAs has been limited to testing of symptomatic individuals. Although there is currently no consensus regarding the correlation between saliva PfHRP2 and parasitemia, the sensitivity shown by the platform is adaptable towards the range of salivary PfHRP2 detected using the laboratory-based ELISA (17–1167 pg/mL)7. To futher guide the development of salivary PfHRP2 diagnostics for elimination and screening purposes, more research is required to study the concentration of salivary PfHRP2 protein in asymptomatic versus symptomatic individuals, and in low versus high density parasitemias. Additionally, more work is required on assessing the effect of patient conditions such as dehydration or hormonal fluctuations on the sensor readings, as minor physiological fluctuations can affect the saliva sample matrix.

Future work on the platform will focus on the miniaturization of sensors and integration of the platform with microfluidics. The lock-in amplifier readout technique has good miniaturization potential, with setup possible within small-scale portable devices40,41. Miniaturization and integration of the sensor platform has the potential to (1) reduce variation due to automatization of the electrical reading and washing steps, (2) reduce the required incubation time by avoiding the diffusion limited process of static incubation, and (3) improve efficiency due to reduction in the reagent and sample volume. Additionally, future work can also be geared towards adaptation of the IDE sensor platform to detect other proteins relevant for malaria diagnostics, including PLDH, and aldolase.

Ultrasensitive diagnostics are considered to play an important role in the global effort towards malaria eradication. Efforts continue to strive for a balance between assay performance, cost-effectiveness, and practicality. We have developed the PfHRP2 IDE sensor platform with these factors in mind. Our study has provided proof of concept that the platform may be a potential technology to help achieve this goal.

Methods

Sample preparation

In Phase I, recombinant PfHRP2 protein (CTK Biotech, California, USA) was suspended in PBS 1× buffer. Antigen quantification was performed using commercial PfHRP2 ELISA (Cellabs Pty. Ltd., Brookvale, New South Wales, Australia). An additional negative control was prepared by suspending 2.5 pg of PLDH protein (CTK Biotech, California, USA) in 1× buffer.

In Phase II, detection was performed in saliva samples spiked with PfHRP2 antigen harvested from in vitro culture supernatants. Culture specimens used included the P. falciparum laboratory lines 3D7 and CS2, in addition to 9 clinical isolates from Papua New Guinean (PNG) and Malawian children with malaria. The clinical isolates were collected as part of projects approved by the PNG Institute of Medical Research Institutional Review Board (IRB Number 136 1103) and the Medical Research Advisory Committee of the PNG Health Department (MRAC 137 Number 11.12) or by the College of Medicine Research Ethics Committee in Malawi (11/14.1566). Parents or guardians of infected children gave informed consent before venous blood was collected. The studies complied with the ethical standards of the Helsinki Declaration.

All specimens were cultured for 36 hours, to obtain samples at 6% parasitemia at mature trophozoite stage. Spent culture medium supernatants were collected. Control medium was prepared similarly by incubating medium with uninfected erythrocytes. Supernatants were stored at −80 °C and used to quantify PfHRP2 by ELISA, and to detect HRP2 using the sensors. Control medium was used to spike negative saliva controls.

To prepare spiked saliva samples, unstimulated fresh saliva was collected and subjected to mechanical filtration to remove residues and mucus (Corning® 0.2 µm filter). Protease inhibitor 100× (P8340 Protease Inhibitor Cocktail, Sigma Aldrich, Missouri, USA) was added immediately followed by the culture-harvested PfHRP2 antigens. The PfHRP2 samples (n = 11) were diluted using saliva to the final concentration of 25 pg/mL. Negative control blank samples (n = 11) were prepared by diluting the control medium in saliva to match the dilution of each PfHRP2 sample. An additional negative sample was prepared using PLDH protein spiked saliva at 25 pg/mL. Spiked samples were immediately added to sensor after baseline electrical readings.

Sensor fabrication

Fabrication was performed at the Melbourne Centre for Nanofabrication (MCN). Wafers of BOROFLOAT® borosilicate glass (UniversityWafer, Massachusetts, USA) were cleaned with isopropyl alcohol, dried, and then coated with hexamethyldisilazane (MicroChemicals GmbH, Ulm, Germany) and AZ1512HS (MicroChemicals GmbH, Ulm, Germany) photoresist. A chrome mask of the sensor design was applied on the substrate followed by UV exposure (75 mJ/cm2). After development, a thin film of chrome (5 nm), gold (100 nm), and titanium (5 nm) was deposited on the substrate, followed by a lift-off process to reveal the IDE pattern. This was followed by an addition of a thin layer (25 nm) of SiO2 using e-beam evaporation (Intlvac NanochromeTM II, Colorado, USA).

Sensor functionalization and antibody immobilization

The sensors were cleaned using a wash of acetone, isopropyl alcohol, and H2O, and then dried under N2 gas. Organic contaminants were eliminated using plasma treatment (PE-25 Plasma Etch, Nevada, USA) with argon (75%) and oxygen (25%) for 5 min at 50 W power and 30 cc/min flow rate. Sensors were silanized in 2% APTES (Sigma Aldrich, Missouri, USA) solution for 1 hour, followed by 3× 5-min washes in 100% ethanol with gentle shaking. Sensors were then immersed in 2.5% glutaraldehyde solution (Sigma Aldrich, Missouri, USA) for 2 hours to allow development of the bifunctional cross linker. After washing for 3× 5-min in PBS 1× with gentle shaking, sensors were dried under N2 gas. Both APTES and glutaraldehyde solutions were filtered prior to use with Corning® 0.2 μm pore size filters to remove impurity. Following functionalization, a 2 mm diameter incubation area was established using medical grade pressure sensitive adhesive tape (Adhesive Research, Pennsylvania, USA).

Anti-PfHRP2 IgG MAb receptor (AB-0445, Vista Diagnostics International, Washington, USA) at a volume of 15 µl (50 µg/mL) was applied on the modified sensing area and incubated in a humid chamber at 4 °C overnight. Sensors were then washed with PBS 1× for 5 min and blocked for 30 min using either 5% ethanolamine (Sigma Aldrich, Missouri, USA) for Phase I or 5% ethanolamine and 2.5% normal goat serum mix (Sigma Aldrich, Missouri USA) for Phase II.

Data processing and statistics for PfHRP2 detection

Acquisition of impedance properties from the lock-in amplifier Vout and phase output was calculated based on Eqs. 1 and 2 using MATLAB. The baseline measurement obtained before sample incubation (T1) were first evaluated for assessment of sensor quality, then T1 and T2 (after sample incubation and washing) values for each parameter were processed in Microsoft Excel to obtain the percentage changes in impedance magnitude (%∆Z = ABS ZT2/ZT1), impedance phase (%∆Zθ), resistance (%∆R), and capacitance (%∆C). The absolute percentage change in impedance (|%∆Z|), resistance (|%∆R|), and capacitance (|%∆C|) were then used to assess the sensor performance statistically.

GraphPad PRISM 7 was used to generate all plots and perform statistical analysis, and p-values ≤ 0.05 were considered statistically significant. To implement quality control on the sensors, a baseline outlier test was performed using the regression and outlier test (ROUT)42 on all derived T1 values, with the maximum false discovery rate (Q) set to 1%. An example of the PRISM ROUT test performance can be seen in Supplementary Table 1 and Supplementary Figure 4. Sensors with baselines excluded using the ROUT test were not included in the rest of the analysis.

In Phase I, Welch’s two-tailed t-test was used to compare the percentage change in impedance magnitude (%∆Z) between blank and test sensors. A serial dilution of PfHRP2 in PBS 1× buffer was incubated on the sensors, and the %∆Z obtained was assessed using one-way ANOVA with Dunnett’s multiple comparisons test. Detection limit is defined as the lowest tested concentration showing statistically significant difference from blank sensor reading.

In Phase II, Welch’s two tailed t-test was used to determine the optimal sample pretreatment and detection parameter in saliva. The optimized parameters were used to determine the detection limit in the same manner as in Phase I, using the optimized parameter for saliva (%∆R). Platform performance was then assessed in a panel of PfHRP2-spiked saliva using the Dunnett’s multiple comparisons test to determine the degree of differentiation against the un-spiked saliva control.

Supplementary information

41598_2019_53852_MOESM1_ESM.docx (2.3MB, docx)

Supplementary Information on the Development of an ultrasensitive and label-free biosensor for the detection of Plasmodium falciparum histidine-rich protein II in saliva

Acknowledgements

This work was funded by the Bill & Melinda Gates Foundation through the Grand Challenges Explorations Initiative (OPP1151367) and by a Program Grant (1092789) from the National Health and Medical Research Council of Australia. The work was performed in part at the Melbourne Centre for Nanofabrication (MCN) in the Victorian Node of the Australian National Fabrication Facility (ANFF). E.S. thanks the generous support of Sue and Leigh Clifford in establishing the Chair in Neural Engineering. G.V.S. was supported by the Indonesian Endowment Fund for Education (LPDP). P.K. is supported by a Practitioner Fellowship from the National Health and Medical Research Council of Australia (APP1136427).

Author contributions

S.R., P.K., E.S., G.V.S. and J.C. conceived the project. G.V.S. and C.D.A. implemented the electronics and performed the experiments. C.B. prepared the clinical isolates and performed all ELISAs. D.H.H. and S.M.U. fabricated the sensors. All the authors designed the experiments and contributed to writing and editing the manuscript.

Data availability

Relevant data are available from the authors on request.

Code availability

The codes used for impedimetric calculations are available on request from the authors.

Competing interests

The authors declare no competing interests.

Footnotes

Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

These authors contributed equally: Stephen Rogerson and Patrick Kwan.

Contributor Information

Patrick Kwan, Email: patrick.kwan@monash.edu.

Stephen J. Rogerson, Email: sroger@unimelb.edu.au

Supplementary information

is available for this paper at 10.1038/s41598-019-53852-5.

References

  • 1.Trampuz A, Jereb M, Muzlovic I, Prabhu RM. Clinical review: Severe malaria. Critical care. 2003;7:315–323. doi: 10.1186/cc2183. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.World Health Organization. Guidelines for the treatment of malaria. World Health Organization (2015).
  • 3.World Health Organization. World malaria report 2015. World Health Organization (2016).
  • 4.World Health Organization. Global technical strategy for malaria 2016–2030. (ed^(eds). World Health Organization (2015).
  • 5.Mouatcho JC, Goldring JD. Malaria rapid diagnostic tests: Challenges and prospects. Journal of Medical Microbiology. 2013;62:1491–1505. doi: 10.1099/jmm.0.052506-0. [DOI] [PubMed] [Google Scholar]
  • 6.Lee N, et al. Effect of sequence variation in Plasmodium falciparum histidine-rich protein 2 on binding of specific monoclonal antibodies: Implications for rapid diagnostic tests for malaria. Journal of Clinical Microbiology. 2006;44:2773–2778. doi: 10.1128/JCM.02557-05. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Fung AO, et al. Quantitative detection of PfHRP2 in saliva of malaria patients in the Philippines. Malaria Journal. 2012;11:175. doi: 10.1186/1475-2875-11-175. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Murray CK, Gasser RA, Jr., Magill AJ, Miller RS. Update on rapid diagnostic testing for malaria. Clinical Microbiology Reviews. 2008;21:97–110. doi: 10.1128/CMR.00035-07. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Oguonu T, et al. The performance evaluation of a urine malaria test (UMT) kit for the diagnosis of malaria in individuals with fever in south-east Nigeria: Cross-sectional analytical study. Malaria Journal. 2014;13:403. doi: 10.1186/1475-2875-13-403. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Wilson NO, Adjei AA, Anderson W, Baidoo S, Stiles JK. Detection of Plasmodium falciparum histidine-rich protein II in saliva of malaria patients. The American Journal of Tropical Medicine and Hygiene. 2008;78:733–735. doi: 10.4269/ajtmh.2008.78.733. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Sutherland CJ, Hallett R. Detecting malaria parasites outside the blood. The Journal of Infectious Diseases. 2009;199:1561–1563. doi: 10.1086/598857. [DOI] [PubMed] [Google Scholar]
  • 12.The malERA Consultative Group on Diagnoses and Diagnostics A research agenda for malaria eradication: Diagnoses and diagnostics. PLoS Medicine. 2011;8:e1000396. doi: 10.1371/journal.pmed.1000396. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Gatton ML, et al. Pan-Plasmodium band sensitivity for Plasmodium falciparum detection in combination malaria rapid diagnostic tests and implications for clinical management. Malaria journal. 2015;14:115. doi: 10.1186/s12936-015-0629-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Jimenez A, et al. Analytical sensitivity of current best-in-class malaria rapid diagnostic tests. Malaria Journal. 2017;16:128. doi: 10.1186/s12936-017-1780-5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.World Health Organization. Malaria rapid diagnostic test performance: Results of WHO product testing of malaria RDTs: Round 6 (2014–2015). World Health Organization (2015).
  • 16.Kifude CM, et al. Enzyme-linked immunosorbent assay for detection of Plasmodium falciparum histidine-rich protein 2 in blood, plasma, and serum. Clinical and Vaccine Immunology. 2008;15:1012–1018. doi: 10.1128/CVI.00385-07. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.PATH. Enzyme linked immunosorbent assays (ELISA) for histidine-rich protein 2 (HRP2). In: Project DIAMETER (Diagnostics for Malaria Elimination Toward Eradication) (ed^(eds) (2014).
  • 18.Markwalter CF, Ricks KM, Bitting AL, Mudenda L, Wright DW. Simultaneous capture and sequential detection of two malarial biomarkers on magnetic microparticles. Talanta. 2016;161:443–449. doi: 10.1016/j.talanta.2016.08.078. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Mangold KA, et al. Real-time PCR for detection and identification of Plasmodium spp. Journal of Clinical Microbiology. 2005;43:2435–2440. doi: 10.1128/JCM.43.5.2435-2440.2005. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Mharakurwa S, Simoloka C, Thuma PE, Shiff CJ, Sullivan DJ. PCR detection of Plasmodium falciparum in human urine and saliva samples. Malaria Journal. 2006;5:103. doi: 10.1186/1475-2875-5-103. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Poon LLM, et al. Sensitive and inexpensive molecular test for falciparum malaria: Detecting Plasmodium falciparum DNA directly from heat-treated blood by loop-mediated isothermal amplification. Clinical Chemistry. 2006;52:303–306. doi: 10.1373/clinchem.2005.057901. [DOI] [PubMed] [Google Scholar]
  • 22.Nokes DJ, et al. Has oral fluid the potential to replace serum for the evaluation of population immunity levels?: a study of measles, rubella and hepatitis B in rural Ethiopia. Bulletin of the World Health Organization. 2001;79:588–595. [PMC free article] [PubMed] [Google Scholar]
  • 23.Formenty P, et al. Detection of Ebola virus in oral fluid specimens during outbreaks of Ebola virus hemorrhagic fever in the Republic of Congo. Clinical Infectious Diseases. 2006;42:1521–1526. doi: 10.1086/503836. [DOI] [PubMed] [Google Scholar]
  • 24.Berdat D, Martin Rodríguez AC, Herrera F, Gijs MAM. Label-free detection of DNA with interdigitated micro-electrodes in a fluidic cell. Lab on a Chip. 2008;8:302–308. doi: 10.1039/B712609C. [DOI] [PubMed] [Google Scholar]
  • 25.Lai W-A, Lin C-H, Yang Y-S, Lu MSC. Ultrasensitive and label-free detection of pathogenic avian influenza DNA by using CMOS impedimetric sensors. Biosensors and Bioelectronics. 2012;35:456–460. doi: 10.1016/j.bios.2012.02.049. [DOI] [PubMed] [Google Scholar]
  • 26.Wang L, et al. A sensitive DNA capacitive biosensor using interdigitated electrodes. Biosensors and Bioelectronics. 2017;87:646–653. doi: 10.1016/j.bios.2016.09.006. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Soraya GV, et al. An interdigitated electrode biosensor platform for rapid HLA-B*15:02 genotyping for prevention of drug hypersensitivity. Biosensors and Bioelectronics. 2018;111:174–183. doi: 10.1016/j.bios.2018.01.063. [DOI] [PubMed] [Google Scholar]
  • 28.Radke SM. Design and fabrication of a microimpedance biosensor for bacterial detection. IEEE Sensors Journal. 2004;4:434–440. doi: 10.1109/JSEN.2004.830300. [DOI] [Google Scholar]
  • 29.Varshney M, Li Y. Review: Interdigitated array microelectrodes based impedance biosensors for detection of bacterial cells. Biosensors and Bioelectronics. 2009;24:2951–2960. doi: 10.1016/j.bios.2008.10.001. [DOI] [PubMed] [Google Scholar]
  • 30.Abeyrathne CD, et al. Lab on a chip sensor for rapid detection and antibiotic resistance determination of Staphylococcus aureus. Analyst. 2016;141:1922–1929. doi: 10.1039/C5AN02301G. [DOI] [PubMed] [Google Scholar]
  • 31.Qureshi A, Niazi JH, Kallempudi S, Gurbuz Y. Label-free capacitive biosensor for sensitive detection of multiple biomarkers using gold interdigitated capacitor arrays. Biosensors and Bioelectronics. 2010;25:2318–2323. doi: 10.1016/j.bios.2010.03.018. [DOI] [PubMed] [Google Scholar]
  • 32.Mihailescu C-M, Stan D, Iosub R, Moldovan C, Savin M. A sensitive capacitive immunosensor for direct detection of human heart fatty acid-binding protein (h-FABP) Talanta. 2015;132:37–43. doi: 10.1016/j.talanta.2014.08.067. [DOI] [PubMed] [Google Scholar]
  • 33.Abeyrathne CD, et al. GFAP antibody detection using interdigital coplanar waveguide immunosensor. IEEE Sensors Journal. 2016;16:2898–2905. doi: 10.1109/JSEN.2016.2525809. [DOI] [Google Scholar]
  • 34.Soraya GV, et al. A label-free, quantitative fecal hemoglobin detection platform for colorectal cancer screening. Biosensors. 2017;7:1–11. doi: 10.3390/bios7020019. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Ma H, et al. An impedance-based integrated biosensor for suspended DNA characterization. Scientific Reports. 2013;3:2730. doi: 10.1038/srep02730. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Fang X, Tan OK, Tse MS, Ooi EE. A label-free immunosensor for diagnosis of dengue infection with simple electrical measurements. Biosensors and Bioelectronics. 2010;25:1137–1142. doi: 10.1016/j.bios.2009.09.037. [DOI] [PubMed] [Google Scholar]
  • 37.Sharma M, et al. A novel piezoelectric immunosensor for the detection of malarial Plasmodium falciparum histidine rich protein-2 antigen. Talanta. 2011;85:1812–1817. doi: 10.1016/j.talanta.2011.07.008. [DOI] [PubMed] [Google Scholar]
  • 38.Hemben A, Ashley J, Tothill I. Development of an immunosensor for PfHRP 2 as a biomarker for malaria detection. Biosensors. 2017;7:28. doi: 10.3390/bios7030028. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Dutta G, Lillehoj PB. An ultrasensitive enzyme-free electrochemical immunosensor based on redox cycling amplification using methylene blue. Analyst. 2017;142:3492–3499. doi: 10.1039/C7AN00789B. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Haitao L, et al. CMOS electrochemical instrumentation for biosensor microsystems: A review. Sensors. 2017;17:1–26. doi: 10.1109/JSEN.2017.2761499. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Yang C, Jadhav SR, Worden RM, Mason AJ. Compact low-power impedance-to-digital converter for sensor array microsystems. IEEE Journal of Solid-State Circuits. 2009;44:2844–2855. doi: 10.1109/JSSC.2009.2028054. [DOI] [Google Scholar]
  • 42.Motulsky HJ, Brown RE. Detecting outliers when fitting data with nonlinear regression - a new method based on robust nonlinear regression and the false discovery rate. BMC Bioinformatics. 2006;7:1–20. doi: 10.1186/1471-2105-7-123. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.Sharma MK, et al. Highly sensitive amperometric immunosensor for detection of Plasmodium falciparum histidine-rich protein 2 in serum of humans with malaria: Comparison with a commercial kit. Journal of Clinical Microbiology. 2008;46:3759–3765. doi: 10.1128/JCM.01022-08. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.Paul KB, et al. A highly sensitive self assembled monolayer modified copper doped zinc oxide nanofiber interface for detection of Plasmodium falciparum histidine-rich protein-2: Targeted towards rapid, early diagnosis of malaria. Biosensors and Bioelectronics. 2016;80:39–46. doi: 10.1016/j.bios.2016.01.036. [DOI] [PubMed] [Google Scholar]
  • 45.Gikunoo E, Abera A, Woldesenbet E. A novel carbon nanofibers grown on glass microballoons immunosensor: A tool for early diagnosis of malaria. Sensors. 2014;14:14686–14699. doi: 10.3390/s140814686. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.de Souza Castilho M, Laube T, Yamanaka H, Alegret S, Pividori M. Magneto immunoassays for plasmodium falciparum histidine-rich protein 2 related to malaria based on magnetic nanoparticles. Analytical Chemistry. 2011;83:5570–5577. doi: 10.1021/ac200573s. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

41598_2019_53852_MOESM1_ESM.docx (2.3MB, docx)

Supplementary Information on the Development of an ultrasensitive and label-free biosensor for the detection of Plasmodium falciparum histidine-rich protein II in saliva

Data Availability Statement

Relevant data are available from the authors on request.

The codes used for impedimetric calculations are available on request from the authors.