Abstract
Introduction
Management of the subscapularis during shoulder arthroplasty is controversial. The purpose of this study was to compare the biomechanical performance of subscapularis peel (SP) and lesser tuberosity osteotomy (LTO) in a cadaveric model.
Methods
The subscapularis and proximal humerus were dissected from all soft tissues in 21 fresh-frozen human cadaveric shoulders and randomized to undergo SP, LTO, or standard subscapularis tenotomy (ST, control). For SP and LTO, six #5 sutures were passed through eyelets in the implant (on lateral border and through drill holes in bicipital groove [2] and under trunion [4]). Double-row repair was performed using two lateral row transosseous sutures and four medial row sutures through the tendon (SP) or osseotendinous junction (LTO). Biomechanical properties and mode of failure were tested.
Results
There were no significant differences in elongation amplitude, cyclic elongation, or maximum load to failure between the three groups (P > 0.05). Mean stiffness was significantly higher in LTO (P = 0.009 vs. SP and ST). In the ST group, 7/7 specimens failed at the tendon-suture interface. For SP, 4/7 failed at the tendon-suture interface, one at the suture-bone interface, one fractured around the implant stem, and one at the knots. For LTO, 3/7 failed at the tendon-suture interface, two at the suture-bone interface and two fractured around the implant stem.
Conclusions
In this cadaveric model, subscapularis repair via ST, SP, and LTO techniques was biomechanically equivalent. Additional studies are needed to confirm these findings and determine the influence of biologic healing on healing rates and clinical outcomes.
Level of evidence
N/a, biomechanical laboratory study
Keywords: Total shoulder arthroplasty, Biomechanical analysis, Subscapularis tenotomy, Subscapularis peel, Lesser tuberosity osteotomy, Subscapularis failure
Introduction
The standard deltopectoral approach for total shoulder arthroplasty (TSA) requires compromise of the subscapularis tendon for insertion of prosthetic components, necessitating repair of the tendon at the conclusion of the procedure. Postoperative dysfunction of this repair is an increasingly recognized postoperative complication [2, 5, 7, 14] and complete rupture can result in significantly worsens clinical outcomes [14, 16]. While a number of techniques have been described including subscapularis tenotomy, subscapularis peel and lesser tuberosity osteotomy [6, 9, 11, 19], few studies have tested the biomechanical strengths of these repairs [8, 12, 22, 24]. Knowing which of the most commonly used repair technique is the strongest would inform the shoulder surgeon in his/her attempts to prevent this complication.
In a biomechanical study, Ahmad et al. [1] compared the strength of fixation under cyclic loading between a tendon-to-bone repair and a combined tendon-to-tendon/tendon-to-bone technique. Using six matched pairs of cadaveric shoulders, these authors demonstrated improved pull-out strengths with the combined repair. While informative, these results have not been duplicated elsewhere and do not consider a third commonly used subscapularis repair technique, lesser tuberosity osteotomy, which is popularized and elegantly described by Gerber et al. [6] where the lesser tuberosity is osteotomized and repaired as a single bone/tendon unit. In the lesser tuberosity osteotomy repair technique, the opportunity for exclusive bone-to-bone fixation has been shown to heal reliably in the clinical setting [6, 7, 11] and may offer greater pull-out strength. In another biomechanical study Van Thiel and colleagues [25] evaluated three different methods of subscapularis repair following TSA. The authors compared lesser tuberosity osteotomy, tendon-to-bone repair with bone tunnels (where the subscapularis was first removed periosteally), and finally a combined method of repair. Interestingly, the authors found no significant differences in elongation amplitude, cyclic elongation, or ultimate load to failure among the groups, and concluded that biomechanically, all constructs offered similar outcomes.
In 2012, Lapner and colleagues [11] performed a randomized controlled trial comparing patients undergoing TSA with lesser tuberosity osteotomy to those undergoing TSA with subscapularis peel. The authors found that at 24 months follow-up, there was no significant difference in subscapularis muscle strength (P = 0.131) or final subjective outcomes. In the peel group, the subscapularis tendon was peeled off the lesser tuberosity at the intertubercular groove. In the osteotomy group, a 5- to 10-mm-thick and 3-cm-long fragment of the lesser tuberosity was elevated along with the subscapularis insertion as described by Gerber et al. [6]. After implantation of the prosthesis, the tendon was repaired identically in both groups, with three nonabsorbable mattress transosseous sutures passed within the bicipital groove and tied over a mini-plate placed on the lateral aspect of the greater tuberosity.
Recently, a new humeral system for TSA was introduced, and it attempts to use the actual implant to aid in the subscapularis repair. The Univers APEX (Arthrex, Inc., Naples, FL) is a humeral component composed of titanium and consisting of a stem, trunion, inclination block, pins, and screws. The humeral component is implanted with a series of sutures placed through the actual implant. Two of these sutures are subsequently passed through drill holes made in the bicipital groove, while the remaining are passed through the subscapularis with free needles to allow for an anatomic subscapularis repair. Lederman et al [12] evaluated this novel technique when compared to lesser tuberosity osteotomy however no biomechanical study analyzing this suture technique has compared multiple subscapularis repair techniques including tenotomy. The purpose of the proposed study is to compare cyclic loading strength, slippage, and ultimate load-to-failure strengths of three tendon repair techniques (tenotomy, peel, and “sliver” lesser tuberosity osteotomy) in a cadaveric arthroplasty model utilizing the Univers APEX (Arthrex, Inc., Naples, FL) humeral stem.
Methods
This study was labeled as exempt from our university’s institutional review board. A total of 21 fresh-frozen human cadaveric shoulders with an average age of 60.1 ± 7.9 years were included for the final analysis. Computed tomography (CT) scans were performed prior to specimen dissection and preparation. Specimens with bony lesions were excluded (n = 1). Bone mineral density (BMD) was calculated from CT scans of the proximal humerus as described previously [22, 25].
Following thawing, the soft tissues were dissected to the level of the rotator cuff musculature. Another surgeon performed all dissections and visually inspected specimens for rotator cuff tears. Specimens with partial- or full-thickness rotator cuff tears (subscapularis, supraspinatus, or infraspinatus) were excluded. Eleven shoulders were excluded because of rotator cuff tears (4), bone lesion (1), inadequate fixation (4), and technical errors during biomechanical testing (2). The subscapularis was reflected off the scapula, and the humerus was detached from the glenoid with the subscapularis tendon intact. All tendons except the subscapularis and 1 cm of the anterior leading edge of the supraspinatus insertion were released sharply from their humeral insertions (Fig. 1). The specimens were subsequently randomized into three groups based on repair technique:
-
I.
Subscapularis tenotomy (ST; N = 7); control
-
II.
Subscapularis peel (SP; N = 7)
-
III.
Lesser tuberosity osteotomy (LTO; N = 7)
Subscapularis release from the humerus was performed based on the assigned repair technique. For specimens in group I, the tendon was tenotomized 1 cm medial to its insertion [25].
For group II, the tendon was peeled directly off its footprint as published previously [11]. In group III, the lesser tuberosity was osteotomized as described by Gerber et al. [6].
Humerus preparation and subscapularis repair
For each specimen, humeral preparation proceeded as for standard anatomic TSA. The natural inclination and version of the humeral head was marked with an indelible marker. A wide-blade oscillating saw was used to perform the humeral head osteotomy along the native retroversion and inclination followed by canal reaming and broaching. Trialing was performed to determine the optimum press fit of the humeral component. At this point, the type of repair determined the remaining part of the procedure.
In the ST group, three 2-mm drill holes were made through the lateral part of the subscapularis footprint (equally spaced) and three #5 high strength non-reabsorbable sutures (Arthrex Inc, Naples, FL) were passed. The implant was placed into the canal, and the medial ends of the sutures were passed through the subscapularis tendon in a Mason-Allen configuration. A rotator interval suture was passed through the anterior leading edge of the supraspinatus tendon and then through the upper rolled edge of the subscapularis tendon using a #2 high strength non-reabsorbable suture (Arthrex Inc, Naples, FL) and tied first to get the subscapularis tendon reduced anatomically. The medial and lateral ends of the three sutures passed through the subscapularis tendon were tied next. After this, three #5 high strength non-reabsorbable sutures (Arthrex Inc, Naples, FL) were passed in simple configuration through the lateral and medial part of the subscapularis tenotomy and tied. At the conclusion of the repair, there were a total of six sutures (alternating Mason-Allen and simple sutures) across the subscapularis tenotomy and one suture across the rotator interval.
In the SP and LTO groups, after trialing for stem size, two unicortical drill holes (2 mm) were drilled in the biceps groove. The proximal drill hole was made at the junction of the upper and middle third of the subscapularis footprint, and the distal drill hole was made at the junction of the middle and distal third of the footprint. A short-stemmed humeral stem component (APEX; Arthrex Inc., Naples, FL) was used in a press-fit fashion. Prior to implanting the humeral component, two high strength #5 nonabsorbable (Arthrex Inc., Naples, FL) lateral row sutures (blue and white color) were passed through the eyelets in the proximal lateral part of the body and shuttled through the transosseous holes in the bicipital groove yielding four suture limbs labeled A through D. The four high strength #5 nonabsorbable medial row sutures (Arthrex Inc., Naples, FL) were passed through the four eyelets located under the trunion of the humeral component yielding eight suture limbs labeled 1 through 8 (suture colors are blue (1), white (2, 3), blue (4, 5), white (6, 7) from proximal to distal, Fig. 2). The lateral row suture limbs were passed transosseously through the bicipital groove (A and B through proximal hole; C and D through the distal hole). At this point, all suture strands were held out to tension as the stem was implanted in the humeral canal. The medial row sutures, exiting transosseously at the calcar (limbs 1 through 8) were then passed through the subscapularis tendon (or at the junction tendon to bone in LTO) in a Mason Allen configuration. To complete the subscapularis repair, the sutures were then tied in the following sequence:
-
A.
1 to A (blue to blue)
-
B.
8 to D (tiger white to tiger white)
-
C.
4 to C (tiger white to tiger white)
-
D.
5 to B (blue to blue)
-
E.
2 to 3 (blue to tiger white)
-
F.
6 to 7 (blue to tiger white; these sutures compress the tendon repair to the lesser tuberosity)
In suture-tying process, one lateral row limb is tied to the one medial row limb from each of the four medial row sutures respectively to form a double-row construct (from superior to inferior). The remaining four medial row limbs are tied to each other (top two and bottom two) to eliminate any slack in the repair and create a dynamic compression of the subscapularis tendon or osteotomy fragment over the lesser tuberosity (Fig. 3a, b). The different suture color pattern aids in suture management and avoids mismatch of suture tying. The first four knots between lateral and medial row are tied between similar colors, and the last two knots (for dynamic compression) are tied between different colors in the medial row.
Biomechanical testing
Following subscapularis repair, the proximal humerus for each specimen was potted within a plastic cylinder using acrylic cement (Isocryl, Lang Dental, Wheeling, IL) and mounted to a custom alignment fixture secured to the base of an electromechanical materials testing system (MTS Insight 5, Eden Prairie, MN). The subscapularis muscle-tendon units were rigidly secured to an aluminum plate with the direction of loading in line with the natural subscapularis tendon vector. The arm was maintained in neutral rotation, abduction, and flexion. The subscapularis muscle belly was fixed in a customized cryogenic clamp. Three rows of three spatial markers were placed, two on either side of the repair construct as well as one on the upper grip. Markers were numbered 1 to 3 from superior to inferior. High-resolution digital imaging equipment was used to visualize the movement of the markers (1000 × 1000 × 48 fps, Imperx IPX-1M48-L, Boca Raton, FL), and digital motion analysis software (Spica Technology Corp, Maui, HI) was used to optically record the marker displacements to assure no slippage at the grip.
Biomechanical testing (Fig. 4) was then performed via the tensile apparatus, using cyclic loading under force control, with each specimen being maintained in a moist state throughout the testing. To start, each specimen was preloaded to 10 N for 1 min and then loaded for 150 cycles from 10 to 100 N at 0.5 Hz. After cyclic testing, each specimen was pulled to failure at a rate of 1 mm/s. Throughout cyclic and failure testing, load, actuator displacement, and time were recorded synchronously with the optical data using dedicated MTS Test Works software. Construct failure mode was visually classified as occurring within the muscle, tendon, repair site, or bone.
During cyclical loading, repairs were assessed for (1) first cycle excursion, which is peak displacement in the first cycle; (2) cyclical elongation, defined as the increase in segment length from the peak load of the first cycle to the peak load of the final cycle; (3) cyclic creep, defined as peak displacement at the last cycle; and (4) elongation amplitude, defined as the peak-valley measurement of the segment elongation for the final cycle. During pull-to-failure testing, repairs were assessed for (1) maximum load; (2) extension at maximum load; (3) linear stiffness, which was calculated as the steepest slope of the load-displacement curve between initiation of the load-to-fail test; and (4) energy to failure at maximum load.
Statistical analysis
The anatomical and biomechanical data was reported as mean ± standard error of mean. The data was compared between the two groups using ANOVA test (Graphpad Prism 5®, Graph Pad, LaJolla, CA). The categorical data was compared using chi-square test. The level of significance was set at P ≤ 0.05.
Results
Demographic data
A total of 21 specimens were included in the analysis. There were 7 specimens per group and the demographic, BMD, and geometric data of subscapularis tendon in the ST, SP, and LTO groups are summarized in Table 1. There were no significant differences between the 3 groups with respect to the mean age and sex distribution. The bone mineral density was not significantly different between the 3 groups (P = 0.48). The subscapularis tendon thickness (upper half) and tendon width were not significantly different (P = 0.54 and 0.39 respectively) in the 3 groups (Table 1).
Table 1.
Variable* | Tenotomy | Peel | Lesser tuberosity osteotomy |
---|---|---|---|
Number of specimens | 7 | 7 | 7 |
Age (year) | 61.6 ± 6.3 | 58 ± 8.8 | 60.9 ± 8.7 |
Sex (M:F) | 5:2 | 6:1 | 4:3 |
BMD (g/cm2) | 145.3 ± 35.1 | 169.7 ± 60.7 | 168.1 ± 56.9 |
Subscapularis tendon thickness (mm) | 5.5 ± 1.7 | 6.5 ± 1.6 | 6.1 ± 1.4 |
*Mean ± SD, except for the sex proportion
Cyclical loading and pull-to-failure data
Cyclic testing of repaired specimens in the three groups did not show any significant differences as shown in Table 2 (P > 0.05 for all). No statistically significant differences were noted in cyclic creep (P = 0.25), first cycle excursion (P = 0.52), elongation amplitude (P = 0.06), and cyclic elongation (P = 0.47) among the three repair techniques.
Table 2.
Testing | Variable | Tenotomy | Peel | LTO | P value |
---|---|---|---|---|---|
Cyclic | First cycle excursion (mm) | 3.8 ± 0.42 | 3.4 ± 0.53 | 3.2 ± 0.13 | 0.52 |
Elongation amplitude (mm) | 2.7 ± 0.35 | 2.4 ± 0.51 | 2.1 ± 0.29 | 0.06 | |
Cyclic elongation (mm) | 2.9 ± 0.76 | 2.5 ± 1.2 | 2.2 ± 1.2 | 0.46 | |
Cyclic creep (mm) | 6.7 ± 1.8 | 5.5 ± 1.8 | 5.3 ± 1.1 | 0.25 | |
Failure | Energy to failure (N*mm) | 7057 ± 2673 | 7959 ± 3208 | 8765 ± 3785 | 0.62 |
Max Load (N) | 596.5 ± 88 | 701 ± 156 | 750.7 ± 226 | 0.23 | |
Stiffness (N/mm) | 46.9 ± 8.7 | 49.4 ± 8.7 | 63.6 ± 11.2a | 0.009 | |
Extension at maximum load (mm) | 24 ± 2 | 25 ± 2.3 | 23 ± 2.1 | 0.81 |
avs. Tenotomy and Peel
Pull-to-failure testing demonstrated significantly higher mean stiffness in the LTO repair group (P = 0.009) compared to the ST and SP groups. However, there were no significant differences between the three groups with respect to maximum load (P = 0.23), maximum extension (P = 0.81), and energy to failure (P = 0.62).
Modes of failure
The modes of failure during pull-to-failure testing are shown in Table 3. All specimens in the ST group eventually failed due to suture pulling through the tendon during load-to-failure testing. In the SP group, four specimens failed due to suture pulling through the tendon, and in one specimen, the knot unraveled. In two specimens in the SP group, the bone fractured at the level of repair pulling the humerus stem out along with the repair. In the osteotomy group, three specimens failed due to suture pulling through the tendon. In two specimens in the LTO group, the bone fractured at the level of repair pulling the humerus stem out along with the repair, and in two remainder specimens, the suture cut through bone.
Table 3.
Groups | Failure mechanisms | |||
---|---|---|---|---|
Tendon-suture interface | Knot failure | Suture-bone interface | Fracture around the component | |
Tenotomy | 7 | 0 | 0 | 0 |
Peel | 4 | 1 | 0 | 2 |
LTO | 3 | 0 | 2 | 2 |
Discussion
The management of subscapularis tendon in anatomic total shoulder arthroplasty is an ongoing debate. The results of this cadaveric study demonstrate that there is no significant difference between the repair displacement during cyclic loading, maximum load, and load-to-failure mechanical properties between the tenotomy, peel, and osteotomy repair techniques using the Arthrex APEX TSA components. The stiffness of the repair construct was significantly higher in the LTO technique compared to the peel and tenotomy groups.
In this study, we used a double-row subscapularis repair in the peel and osteotomy group. In this technique, four medial row sutures pass through the eyelets of the humeral component before they are passed through the tendon to provide strong anchorage. The two lateral row sutures pass through the eyelets in the stem and exit through the two holes in the bicipital groove. The four medial and lateral row sutures are tied to each other to achieve static compression, and the remaining four medial sutures are tied to each other to achieve dynamic compression across the repair site. Studies have demonstrated that double-row fixation in regard to rotator cuff repairs results in an increased area of contact between the repair footprint and bone. Biomechanical data in a cadaveric model demonstrate that double-row fixation results in increased ultimate failure strength [4]. This study demonstrates that despite using the dynamic compression subscapularis repair technique, we did not find significant differences between the three repair techniques during cyclic biomechanical testing. Load-to-failure testing demonstrated that LTO repair technique has higher stiffness compared to the peel and tenotomy techniques. Other investigators have reported similar results when comparing the tenotomy, peel, and LTO using standard repair techniques [12, 14, 24, 25]. Van Thiel et al. [25] reported no significant differences in cyclic elongation, cyclic amplitude, maximum load, stiffness, and load to failure between the tenotomy, peel, and LTO repair techniques in a cadaveric experiment.
The outcome of total shoulder arthroplasty is adversely affected by postoperative subscapularis insufficiency [15, 16, 21]. Compromise of subscapularis repair can present as a complete insufficiency (failure of repair) or more commonly as loss of function (loss of terminal internal rotation, weak belly press, and poor lift-off) [9, 14, 15, 20, 21]. Clinical and radiographic studies have demonstrated higher healing rates and lower clinical signs of subscapularis dysfunction when using lesser tuberosity osteotomy compared to the tenotomy and/or transosseous repair [3, 10, 11, 17–19]. However, other studies and recent study do not show any significant difference in the outcomes between the three repair techinques [11, 13] . The osteotomy has demonstrated failure due to nonunion, which can occur due to failure of fixation. We hypothesized that improved time equal zero fixation will minimize or prevent this failure of poor fixation. In this technique, we demonstrated that fixation with osteotomy technique had higher stiffness, higher mean energy to failure, and maximum strength although statistical significance was only achieved for the stiffness parameter.
The three techniques, ST, SP, and LTO, heal via three different mechanisms, namely, tendon to tendon, tendon to bone, and bone to bone respectively. However, in this controlled experimental study, only time zero fixation was assessed. Proponents of tendon-to-tendon repair claim that it is a quick and simple repair technique and avoids the possibility of nonunion of osteotomy. However, MRI and USG data has shown a higher failure compared to osteotomy. Fortunately, this compromised subscapularis function has not translated into poor postoperative outcomes [9, 20, 23]. In actual practice, repair techniques using subscapularis suturing like the peel and tenotomy are more common than lesser tuberosity osteotomy [19].
Strengths
The strengths of this study include the comprehensiveness of the biomechanical testing and the inclusion of three different commonly used techniques of subscapularis tendon repair in total shoulder arthroplasty spanning three different mechanisms of fixation: tendon to tendon, tendon to bone, and bone to bone. All standard forms of biomechanical testing were performed including cyclic testing and pull-to-failure. In addition, these modes of testing were expanded to test first cycle excursion, cyclical elongation, cyclic creep, and elongation amplitude in cyclic testing and maximum load, extension at maximum load, linear stiffness, and energy to failure at maximum load in the pull-to-failure testing. Additionally, this study is the first to test a suturing technique that utilizes suture integration into the humeral stem in the repair of the subscapularis tendon.
Limitations
One of the known limitations of biomechanical studies is they assess repair integrity at time zero, which has several disadvantages when correlating clinically. These studies are generally performed on cadaveric models and are therefore not appropriate for determining ultimate healing and strength of the repaired tendon. Additionally, they lack chronicity, which is a critical factor in healing rates and clinical outcome. There are advantages to comparing various repair configurations using a time equal zero model including the ability to study failure modality, interface motion, gap formation, and repair surface area, all of which are important factors in a successful repair. There is also a hypothetical concern for increased risk of infection and tissue necrosis with the use of nonabsorbable sutures as described in this paper. Nonabsorbable sutures in conjunction with double-row repairs are often used for rotator cuffs as they are made of strong materials that allow for physiologic loading conditions. The alternative would be to use slow absorbable sutures to lower the risk of infection and tissue necrosis.
Conclusions
In this cadaveric shoulder arthroplasty model, subscapularis repair via tenotomy, SP, and LTO techniques was biomechanically equivalent with respect to elongation amplitude, cyclic elongation, and maximum load to failure. Additional studies are needed to confirm these biomechanical findings and determine the influence of biologic healing on overall healing rates and clinical outcomes.
Acknowledgements
Not applicable
Illustrations
No color illustrations are needed
Disclosures
Implants and sutures were donated by Arthrex Inc (Naples FL); the senior author is a consultant for Arthrex Inc (Naples, FL).
Abbreviations
- BMD
Bone mineral density
- CT
Computed tomography
- LTO
Lesser tuberosity osteotomy
- SP
Subscapularis peel
- ST
Subscapularis tenotomy
- TSA
Total shoulder arthroplasty
Authors’ contributions
MV, SA, and RMF made substantial contributions to the acquisition and interpretation of data as well as drafting of the work. CM made substantial contributions to the acquisition of data as well as drafting of the work. SA made substantial contributions to the revision of the work. EFS and VMW made substantial contributions to the interpretation of data and management of machinery and software use in acquiring data. AAR and MV made substantial contributions to the conception and design of the work as well as the acquisition, analysis, and interpretation of data. All authors read and approved the final manuscript.
Funding
No funding was provided for this project
Availability of data and materials
All data generated or analyzed during this study are included in this published article.
Ethics approval and consent to participate
This research was reviewed and approved by the IRB at RUSH University Medical Center.
Consent for publication
Not applicable
Competing interests
Anthony A. Romeo receives research support and is a consultant for Arthrex Inc. Mandeep S. Virk is a consultant for Exactech Inc. All other authors declare that they have no competing interests. No financial benefits were received for the production of this study.
Footnotes
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Mandeep S. Virk and Saleh S. Aiyash are co-first authors.
Contributor Information
Saleh S. Aiyash, Phone: (630) 392-3115, Email: salehaiyash@gmail.com
Anthony A. Romeo, Phone: (267) 339-3500, Email: anthony.romeo@rothmanortho.com
References
- 1.Ahmad CS, Wing D, Gardner TR, Levine WN, Bigliani LU. Biomechanical evaluation of subscapularis repair used during shoulder arthroplasty. J Shoulder Elbow Surg. 2007;16(3 Suppl):S59–S64. doi: 10.1016/j.jse.2006.09.002. [DOI] [PubMed] [Google Scholar]
- 2.Brems JJ. Complications of shoulder arthroplasty: infections, instability, and loosening. Instr Course Lect. 2002;51:29–39. [PubMed] [Google Scholar]
- 3.Buckley Taylor, Miller Richard, Nicandri Gregg, Lewis Richard, Voloshin Ilya. Analysis of subscapularis integrity and function after lesser tuberosity osteotomy versus subscapularis tenotomy in total shoulder arthroplasty using ultrasound and validated clinical outcome measures. Journal of Shoulder and Elbow Surgery. 2014;23(9):1309–1317. doi: 10.1016/j.jse.2013.12.009. [DOI] [PubMed] [Google Scholar]
- 4.Burkhart Stephen S., Fischer Scott P., Nottage Wesley M., Esch James C., Barber F.Alan, Doctor David, Ferrier James. Tissue fixation security in transosseous rotator cuff repairs: A mechanical comparison of simple versus mattress sutures. Arthroscopy: The Journal of Arthroscopic & Related Surgery. 1996;12(6):704–708. doi: 10.1016/S0749-8063(96)90174-7. [DOI] [PubMed] [Google Scholar]
- 5.EDWARDS T. BRADLEY, BOULAHIA AZIZ, KEMPF JEAN-FRANÇOIS, BOILEAU PASCAL, NÉMOZ CHANTAL, WALCH GILLES. THE INFLUENCE OF ROTATOR CUFF DISEASE ON THE RESULTS OF SHOULDER ARTHROPLASTY FOR PRIMARY OSTEOARTHRITIS. The Journal of Bone and Joint Surgery-American Volume. 2002;84(12):2240–2248. doi: 10.2106/00004623-200212000-00018. [DOI] [PubMed] [Google Scholar]
- 6.Gerber C, Pennington SD, Yian EH, Pfirrmann CA, Werner CM, Zumstein MA. Lesser tuberosity osteotomy for total shoulder arthroplasty. Surgical technique. J Bone Joint Surg Am. 2006;88:170–177. doi: 10.2106/JBJS.F.00407. [DOI] [PubMed] [Google Scholar]
- 7.Gerber C, Yian EH, Pfirrmann CA, Zumstein MA, Werner CM. Subscapularis muscle function and structure after total shoulder replacement with lesser tuberosity osteotomy and repair. J Bone Joint Surg Am. 2005;87(8):1739–1745. doi: 10.2106/JBJS.D.02788. [DOI] [PubMed] [Google Scholar]
- 8.Giuseffi SA, Wongtriratanachai P, Omae H, Cil A, Zobitz ME, An K, et al. Biomechanical comparison of lesser tuberosity osteotomy versus subscapularis tenotomy in total shoulder arthroplasty. J Shoulder Elbow Surg. 2012;21(8):1087–1095. doi: 10.1016/j.jse.2011.07.008. [DOI] [PubMed] [Google Scholar]
- 9.Jackson JD, Cil A, Smith J, Steinmann SP. Integrity and function of the subscapularis after total shoulder arthroplasty. J Shoulder Elbow Surg. 2010;19(7):1085–1090. doi: 10.1016/j.jse.2010.04.001. [DOI] [PubMed] [Google Scholar]
- 10.Jandhyala S, Unnithan A, Hughes S, Hong T. Subscapularis tenotomy versus lesser tuberosity osteotomy during total shoulder replacement: a comparison of patient outcomes. J Shoulder Elbow Surg. 2011;20(7):1102–1107. doi: 10.1016/j.jse.2011.03.019. [DOI] [PubMed] [Google Scholar]
- 11.Lapner PL, Sabri E, Rakhra K, Bell K, Athwal GS. Comparison of lesser tuberosity osteotomy to subscapularis peel in shoulder arthroplasty: a randomized controlled trial. J Bone Joint Surg Am. 2012;94(24):2239–2246. doi: 10.2106/JBJS.K.01365. [DOI] [PubMed] [Google Scholar]
- 12.Lederman E, Streit J, Idoine J, Shishani Y, Gobezie R. Biomechanical study of a subscapularis repair technique for total shoulder arthroplasty. Orthopedics. 2016;39(5):e937–e943. doi: 10.3928/01477447-20160623-09. [DOI] [PubMed] [Google Scholar]
- 13.Levine William N., Munoz Julianne, Hsu Stephanie, Byram Ian R., Bigliani Louis U., Ahmad Christopher S., Kongmalai Pinkawas, Shillingford Jamal N. Subscapularis tenotomy versus lesser tuberosity osteotomy during total shoulder arthroplasty for primary osteoarthritis: a prospective, randomized controlled trial. Journal of Shoulder and Elbow Surgery. 2019;28(3):407–414. doi: 10.1016/j.jse.2018.11.057. [DOI] [PubMed] [Google Scholar]
- 14.Miller BS, Joseph TA, Noonan TJ, Horan MP, Hawkins RJ. Rupture of the subscapularis tendon after shoulder arthroplasty: diagnosis, treatment, and outcome. J Shoulder Elbow Surg. 2005;14(5):492–496. doi: 10.1016/j.jse.2005.02.013. [DOI] [PubMed] [Google Scholar]
- 15.Miller SL, Hazrati Y, Klepps S, Chiang A, Flatow EL. Loss of subscapularis function after total shoulder replacement: a seldom recognized problem. J Shoulder Elbow Surg. 2003;12(1):29–34. doi: 10.1067/mse.2003.128195. [DOI] [PubMed] [Google Scholar]
- 16.Moeckel BH, Altchek DW, Warren RF, Wickiewicz TL, Dines DM. Instability of the shoulder after arthroplasty. J Bone Joint Surg Am. 1993;75(4):492–497. doi: 10.2106/00004623-199304000-00003. [DOI] [PubMed] [Google Scholar]
- 17.Ponce BA, Ahluwalia RS, Mazzocca AD, Gobezie RG, Warner JJ, Millett PJ. Biomechanical and clinical evaluation of a novel lesser tuberosity repair technique in total shoulder arthroplasty. J Bone Joint Surg Am. 2005;87(Suppl 2):1–8. doi: 10.2106/JBJS.E.00441. [DOI] [PubMed] [Google Scholar]
- 18.Qureshi S, Hsiao A, Klug RA, Lee E, Braman J, Flatow EL. Subscapularis function after total shoulder replacement: results with lesser tuberosity osteotomy. J Shoulder Elbow Surg. 2008;17(1):68–72. doi: 10.1016/j.jse.2007.04.018. [DOI] [PubMed] [Google Scholar]
- 19.Sanchez-Sotelo J. Subscapularis tenotomy in anatomic total shoulder arthroplasty. J Shoulder Elbow Surg. 2017;1:247154921774360. doi: 10.1177/2471549217743605. [DOI] [Google Scholar]
- 20.Scalise JJ, Ciccone J, Iannotti JP. Clinical, radiographic, and ultrasonographic comparison of subscapularis tenotomy and lesser tuberosity osteotomy for total shoulder arthroplasty. J Bone Joint Surg Am. 2010;92(7):1627–1634. doi: 10.2106/JBJS.G.01461. [DOI] [PubMed] [Google Scholar]
- 21.Scheibel M, Habermeyer P. Subscapularis dysfunction following anterior surgical approaches to the shoulder. J Shoulder Elbow Surg. 2008;17(4):671–683. doi: 10.1016/j.jse.2007.11.005. [DOI] [PubMed] [Google Scholar]
- 22.Schreiber JJ, Anderson PA, Rosas HG, Buchholz AL, Au AG. Hounsfield units for assessing bone mineral density and strength: a tool for osteoporosis management. J Bone Joint Surg Am. 2011;93(11):1057–1063. doi: 10.2106/JBJS.J.00160. [DOI] [PubMed] [Google Scholar]
- 23.Sperling JW, Potter HG, Craig EV, Flatow E, Warren RF. Magnetic resonance imaging of painful shoulder arthroplasty. J Shoulder Elbow Surg. 2002;11(4):315–321. doi: 10.1067/mse.2002.124426. [DOI] [PubMed] [Google Scholar]
- 24.Van den Berghe GR, Nguyen B, Patil S, D’Lima DD, Mahar A, Pedowitz R, et al. A biomechanical evaluation of three surgical techniques for subscapularis repair. J Shoulder Elbow Surg. 2008;17(1):156–161. doi: 10.1016/j.jse.2007.04.016. [DOI] [PubMed] [Google Scholar]
- 25.Van Thiel GS, Wang VM, Wang FC, Nho SJ, Piasacki DP, Bach BR, et al. Biomechanical similarities among subscapularis repairs after shoulder arthroplasty. J Shoulder Elbow Surg. 2010;19(5):657–663. doi: 10.1016/j.jse.2010.01.014. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
All data generated or analyzed during this study are included in this published article.