Tractography Reproducibility Challenge with Empirical Data (TraCED): The 2017 ISMRM Diffusion Study Group Challenge

Vishwesh Nath 1, Kurt G Schilling 2, Prasanna Parvathaneni 3, Yuankai Huo 3, Justin A Blaber 3, Allison E Hainline 4, Muhamed Barakovic 5, David Romascano 5, Jonathan Rafael-Patino 5, Matteo Frigo 5, Gabriel Girard 5, Jean-Philippe Thiran 5, Alessandro Daducci 6, Matt Rowe 7, Paulo Rodrigues 7, Vesna Prchkovska 7, Dogu B Aydogan 8, Wei Sun 8, Yonggang Shi 8, William A Parker 9, Abdol A Ould Ismail 9, Ragini Verma 9, Ryan P Cabeen 10, Arthur W Toga 10, Allen T Newton 11, Jakob Wasserthal 12, Peter Neher 12, Klaus Maier-Hein 12, Giovanni Savini 13, Fulvia Palesi 14, Enrico Kaden 15, Ye Wu 16, Jianzhong He 16, Yuanjing Feng 16, Michael Paquette 17, Francois Rheault 17, Jasmeen Sidhu 17, Catherine Lebel 18, Alexander Leemans 19, Maxime Descoteaux 17, Tim B Dyrby 20, Hakmook Kang 4, Bennett A Landman 1,2,3,11
PMCID: PMC6900461  NIHMSID: NIHMS1029852  PMID: 31179595

Abstract

Background:

Fiber tracking with diffusion-weighted magnetic resonance imaging has become an essential tool for estimating in vivo brain white matter architecture. Fiber tracking results are sensitive to the choice of processing method and tracking criteria.

Purpose:

Phantom studies provide concrete quantitative comparisons of methods relative to absolute ground truths, yet do not capture variabilities because of in vivo physiological factors. Reproducibility does not assess validity of brain connections yet, is still of importance because it describes the variability for an algorithm in group studies. The ISMRM 2017 TraCED challenge was created to fulfill the gap.

Study Type:

Systematic review of algorithms and tract reproducibility study

Subjects:

Single healthy volunteer

Field Strength/Sequence:

3.0 T, two different scanners of the same manufacturer. The multi-shell acquisition included b-values of 1000, 2000 and 3000 s/mm2 with 20, 45 and 64 diffusion gradient directions per shell, respectively.

Assessment:

Nine international groups submitted 46 tractography algorithm entries each consisting 16 tracts per scan. The algorithms were assessed using intra-class correlation (ICC) and dice similarity measure.

Statistical Tests:

Containment analysis was performed to assess if the submitted algorithms had containment within tracts of larger volume submissions. This also serves the purpose to detect if spurious submissions had been made.

Results:

7The top five submissions had high ICC & Dice > 0.88. Reproducibility is high within top 5 submissions when assessed across sessions or across scanners: 0.87–0.97. Containment analysis shows that top 5 submissions are contained within larger volume submissions. From the total of 16 tracts as an outcome relatively the number of tracts with high, moderate and low reproducibility are 8, 4 and 4.

Conclusion:

The different methods clearly result in fundamentally different tract structures at the more conservative specificity choices. Data and challenge infrastructure remain available for continued analysis and provide a platform for comparison.

Keywords: Tractography, Reproducibility, in vivo, Challenge, DW-MRI, HARDI

INTRODUCTION

Diffusion weighted magnetic resonance imaging (DW-MRI) is a technique which allows for non-invasive mapping of the human brain’s micro-architecture at milli-metric resolution. Using voxel-wise fiber orientation reconstruction methods, tractography can provide quantitative and qualitative information for studying structural brain connectivity and continuity of neural pathways of the nervous system in vivo. There have been many algorithms, global, iterative, deterministic and probabilistic, that reconstruct streamlines using fiber reconstruction methods. Tractography was conceived[1, 2] using one of the first fiber reconstruction method, diffusion tensor imaging (DTI) [1]. However, DTI has a well-known limitation: it cannot resolve complex fiber configurations [3]. With the advancement in acquisitions protocols allowing for better resolution and higher number of gradient values new methods for reconstruction of local fiber have been developed. These methods are commonly referred to as high angular resolution diffusion imaging (HARDI), e.g., q-ball, constrained spherical deconvolution (CSD), persistent angular structure (PAS) [46]. HARDI methods enable characterization of more than a single fiber direction per voxel, but have been often shown to be limited when more than two fiber populations exist per voxel [79]. While there is definite gain in sensitivity when using HARDI methods, there remain critical questions of their reproducibility [10].

There have been many validation efforts that aim to assess the anatomical accuracy of tractography. Early studies investigated how well tractography followed large white matter trajectories through qualitative comparisons with dissected human samples [11], or previous primate histological tracings [12]. Later works on the macaque [13] or porcine [14] brains highlighted limitations and common errors in tractography. Recently, the sensitivity and specificity of tractography in detecting connections has been systematically explored against tracers in the monkey [1517], porcine [18], or mouse [19] brains. The main conclusions drawn from these are (1) that algorithms always show a tradeoff in sensitivity and specificity (i.e. those that find the most true connections have the most false connections) (2) short-range connections are more reliably detected than long-range, (3) connectivity predictions do better than chance and thus have useful predictive power, and (4) tractography performs better when assessing connectivity between relatively large-scale regions rather than identifying fine details or connectivity.

Despite the wide range of validation studies, there have been few reproducibility studies of tractography [2023]. Rather than ask how right (or wrong) tractography is, we ask how stable are the outputs of these techniques? Because tractography is an essential part of track segmentation, network analysis, and microstructural imaging, it is important that reproducibility is high, otherwise power is lost in group analyses or in longitudinal comparisons. In this study, given a standard, clinically realistic, diffusion protocol, we aim to assess how reproducible tractography results are between repeats, between scanners, and between algorithms.

Publicly organized challenges provide unique opportunities for research communities to fairly compare algorithms in an unbiased format, resulting in quantitative measures of the reliability and limitation of competing approaches, as well as potential strategies for improving consistency. In the diffusion MRI community, challenges have focused on recovering intra-voxel fiber geometries using synthetic data [24] and physical phantoms [20, 25]. Similarly, diffusion tractography challenges [21] have provided insights into the effects of different acquisition settings, voxel-wise reconstruction techniques, and tracking parameters on tract validity by comparing results to ground truth physical phantom fiber configurations [20, 22]. Recently, more clinically relevant evaluations have been put forth. For example, a recent MICCAI challenge benchmarked DTI tractography of the pyramidal tract in neurosurgical cases presenting with tumors in the motor cortex [26]. Towards this direction, the current challenge utilized a large-scale single subject reproducibility dataset, acquired in clinically feasible scan times. This challenge was intended to study reproducibility to describe the limitations for capturing physiological and imaging considerations prevalent in human data and evaluate the newest generation of tractography algorithms.

This paper is organized as follows. First, we present the analysis structure of this challenge to characterize which tracts are the most reproducible. Second, we characterize the variance across the tractography methods by design features and compare the potential containment of tracts on a per algorithm basis.

METHODS

DW-MRI Data Acquisition

Informed consent following Institutional Review Board (IRB) approval, was obtained to conduct this study.

The data were acquired with a multi-shell HARDI sequence on single healthy human subject. The two scanners were both Phillips, Achieva, 3T, Best, Netherlands. These are referred to as scanner ‘A’ and ‘B’. The three shells that were acquired: b=1000 s/mm2, 2000s/mm2 and 3000s/mm2 with 20, 48 and 64 gradient directions respectively (uniformly distributed over a hemi-sphere and independently per shell, this was done in consideration of scanner hardware.). The other parameters were kept consistent for all shells. They are as follows: Delta=~48ms, delta=~37ms, partial fourier=0.77, TE = 99 ms, TR ~= 2920 ms and voxel resolution=2.5mm isotropic. A total of 15 non-weighted diffusion volumes ‘b0’ images interspersed as 5 per shell were acquired. Additionally, for scanner A & B, 5 reverse phase-encoded b0 images and 3 diffusion weighted directions were acquired to aid in distortion correction. The additional 3 diffusion-weighted direction volumes were acquired for ease of acquisition from the scanner. They do not contribute to the pre-processing of the data in any way.

Additionally, a T1-weighted reference image (MPRAGE) was acquired for each session per scanner (4 volumes total). A single volume of T1 was used which was registered to the first session of scanner A where the session had already been registered to the MNI template. This was done using a 6 degree of freedom rigid body registration.

For the initial data release, a technical issue resulted in 5 non-reverse phase-encoded b0 images for scanner A. Note that at the end of the challenge, the scanner ‘A’ data were completely re-acquired for both sessions with 5 reverse phase-encoded b0 images and 3 diffusion weighted directions. These data were released as supplementary material, but not included in the presented challenge data. Following the protocol for tractography in [25], we delineated six tracts cingulum (CNG) Left/Right (L/R), inferior longitudinal fasciculus (ILF) (L/R), inferior fronto-occipital (IFO) (L/R). The mean intra-class correlation (ICC) inter-scanner values for the original challenge data and the updated challenge data were 0.86 and 0.89, respectively. The mean difference between methods was 0.15 in terms of ICC. As expected, the inclusion of full reverse phase encoding for Scanner ‘A’ introduced a small increased in consistency relative to much larger differences between methods.

DW-MRI Data Pre-processing as illustrated in Fig 1, the 5 repeated acquisitions from each of the four sessions (two repeated on scanner A and B) were concatenated and corrected with FSL’s eddy and topup [2729]. Intensity normalization was performed by dividing each diffusion weighted scan by the mean of all non-weighted diffusion volume (B0) per session. The average B0 from scanner A of the first sessions was rigidly (six degrees of freedom) registered [30] to a 2.5 mm T2 MNI template (this was done to ensure resampling from registration was done on both datasets). Next, the average B0 from the scanner A second session was rigidly registered to the average B0 of the registered scanner A first session B0 which had already been registered to the MNI space. Successively, the sessions from scanner B were registered to the sessions of scanner A. The b-vectors were rotated to account for the registration of the DW-MRI data [31].

Figure 1.

Figure 1

The acquisition per session included five repeats of a single b0 and successively at b-values of 3000, 2000 and 1000 s/mm2 using 64, 48 and 20 gradient directions respectively. Each session was individually corrected using topup, eddy and then normalized. All the sessions were registered using flirt to the first session of scanner A.

The T1 weighted MPRAGE was rigidly registered to the average registered b0 from the first session of scanner A. This transformation was applied to the T1 maintaining 1 mm isotropic resolution, thus providing a high-resolution segmentation that may be converted into diffusion space by performing a simple down-sampling. Multi-atlas segmentation with non-local spatial STAPLE fusion was used for the segmentation of the T1 volume to 133 different ROI’s [32, 33]. Finally, Multi-atlas CRUISE (MaCRUISE) was used to identify cortical surfaces [34]. These were provided for ease of algorithm implementations.

Challenge Rules and Metrics

For each of the 20 HARDI datasets (5 repetitions x 2 sessions x 2 scanners), participants were asked to submit a tractogram (i.e., “fiber probability membership function”) for each well-modeled fiber structures (uncinate (UNC) [L/R], fornix (FNX) [L/R], genu of the corpus callosum, cingulum (CNG) [L/R], corticospinal tract (CST) [L/R], splenium of the corpus callosum, inferior longitudinal fasciculus (ILF) [L/R], superior longitudinal fasciculus (SLF) [L/R], and inferior fronto-occipital (IFO) [L/R](1)). Each tractogram is a NIFTI volume at the field of view and resolution of the T1-weighted reference space where the floating-point value (32-bit single precision) of each voxel is in [0, 1] and indicates the probability of the voxel belonging to the specified fiber tract. Thus, participants submitted a total of 320 (5 × 2 × 2 × 16) NIFTI volumes using the acquisition of both the scanners. Assessment of fiber fractions was supported (i.e., the sum across all tracts is <=1 with the remainder as background). However, strict probabilities where each voxel may have a high probability of 2 or more fibers with a sum greater than 1 were permitted as well.

Tractograms within a submission were compared based on reproducibility of the tracts (intra-class correlation coefficient (ICC) statistics for continuous values and Dice similarity scores based on maximum probability assignment at 0.5). Intra-session, inter-session, same scanner, and inter-scanner scanner metrics have been reported for quantitative interpretation. The ICC and dice value of unique number of combinations of pairs of repeats were used as data points for violin plots depicting results of intra-session, inter-session and inter-scanner. The unique combinations of repetitions were 40, 50 and 100 respectively for the three levels of reproducibility.

Containment Analysis

A key question is whether the differences in tractography are driven by different considerations of the volume of the track, i.e., the larger the volume is, the more likely the track may include the underlying true track. For example, it is plausible that a set of tractography methods could see the same underlying probabilistic connection pattern and choose to threshold it based on different preferences for the volume of tracks. If the preference was driving the tractography differences, then tractograms would essentially be able to be nested from smallest to largest. To examine this hypothesis, we define the property containment index (CI) for two tracts where

CIA,B=A=0:1A0|andB=0:0otherwise:AB/B (1)

For the purposes of this discussion, we define the tractogram set to be the binary volume resulting at a 0.05 threshold of the mean of all results submitted for each algorithm. A visual understanding of containment index can be observed in Fig 3.

Fig 3.

Fig 3

A) Where the shape X is impeccably contained in Y and Y is contained in Z. The resulting containment CI(Y, X) = 1, CI(Z, X) = 1 and CI(Z, Y) = 1. B) Shape Y is a noisy representation of shape Z where CI(Y, Z) = 0.84. C) Shape Z is different from shape Y in a different orientation and the CI(Z,Y) = 0.17

Then, an optimal ordering (“nesting”) of tractogram entries can be computed by maximizing the containment energy (CE, i.e., sum of CI for all tracts versus the tracts earlier than the one under consideration):

argmaxoperm1EntryCE=argmaxoperm1EntryiNjiNCIEntryoi,Entryij (2)

Where perm denotes the permutation operator and Entry is a list of all entered tractograms. Conceptually, this procedure finds the ideal order to stack the tractograms inside each other where the first tract is “most inside” the subsequent ones and the last tract is “most outside” all others. We define <CI> as the average containment index of all nesting for the ordered entries that are smaller than or equal to an entry provides a quantitative way to examine “nesting” (note, this approach includes the self-containment index so that the first entry has a CI of 1). Then, we can see how the nesting holds up from the inner (#’1’) to the outer (#’46’) entry.

Tractography Reconstruction Methods

Team 1

A CSD-based probabilistic streamline tractography [35] seeding from the WM, with a step size of 0.2mm and a maximum deviation angle of 30 degrees (multi-shell data, 20 different datasets). The Tract Querier [36] was used to extract the bundles from anatomical constraints. The processing steps were identically applied to all dataset, without manual intervention any level of the processing. Automatic streamline filtering was applied using COMMIT [37]. The bundle volumes were also post-processed using a distance transform, resulting in weighted volumes with values corresponding to the distance to the closest boundary.

Team 2

Instead of using traditional tractography, which reconstructs every single fiber, the problem is viewed as an image segmentation task. A U-Net is used to segment each fiber bundle. As input features the peaks of the fiber orientation distribution functions were used. As ground truth data for training of the model manual segmentation was performed on the bundles of 30 HCP subjects. The approach is described in more detail in [38].

Team 3

Two shells at b-values of 1000 and 2000 s/mm2 were used and the data was upsampled to 2mm isotropic. The tractography approach relies on free water invariant estimation of tensor (FERNET). The free water correction has been adapted from the work described in [39]. FA thresholds of 0.15 and 0.20 were used with angular thresholds of 25 and 35 degrees.

Team 4

The diffusion data ( b=3000 s/mm2) was up-sampled from 2.5mm isotropic to 1mm isotropic. WM mask was computed from the T1-weighted image using FSL FAST [40]. The fiber orientation distribution (FOD) function was estimated using CSD [41] implemented in DiPy [35]. An in-house probabilistic tractography algorithm was used [42]. Tractography parameters were: step size = 0.2 mm, maximal deviation = 20 degrees, number of seeds per voxel = 3. Bundle segmentation was performed using a novel automatic atlas-based bundle segmentation method [43]. The atlas was built from a HCP tractogram using the definition provided by the organisers.

Team 5

A technique for the estimation of the FOD was combined with a more traditional method for tract selection. Specifically, the spherical mean technique (SMT) was used to fit a microscopic compartment model of brain tissue [44] to the pre-processed multi shell diffusion data which subsequently enabled the recovery of the FOD quantitatively using CSD. Resulting FOD’s were used for whole-brain probabilistic tractography. Anatomically constrained tractography and spherical deconvolution informed filtering of tractograms were used to discard spurious tracts. Streamlines belonging to the tracts were selected using both atlas-derived regions (JHU-DTI) and ROI’s manually drawn [45] on track density images.

Team 6

A pipeline was evaluated for diffusion tensor bundle modeling using an atlas-based approach [46], implemented using the Quantitative Imaging Toolkit [47]. First, reference bundles were produced in the IIT tensor atlas consisting of include, exclude, and seed masks, which were deformed to subject space using DTI-TK. The pipeline included denoising using nonlocal means filtering, up-sampling to 1mm3 voxels, weighted-least squares tensor estimation, and streamline integration with step size 1mm, angle threshold 40 degrees, and a FA threshold of 0.15. Fiber bundles were extracted from the streamlines using the atlas masks with outlier rejection and finally discretized to native space voxels.

Team 7

The method described in [48] was augmented into a 2 pass tractography approach. Broad WM waypoints are defined in MNI template space at anatomically significant locations for each fascicle [49]. A deterministic tractography algorithm (based on CSD [50]) generates 500,000 streamlines. These streamlines are filtered by groups of waypoints warped from MNI space into patient space. Waypoints are then refined, only retaining waypoint voxels intersected by at least 1 streamline from each relevant fascicle. 5,000,000 streamlines generated by a CSD based probabilistic tractography algorithm (iFOD2), are filtered by refined waypoints then converted to fiber probability maps for each fascicle.

Team 8

Multi-shell compartment analysis is used on the pre-processed images (v1.0) to compute FODs [51]. Streamlines are computed using the algorithm in [52]. For each of the 20 images, 1.25 million streamlines are obtained using the following parameters: FOD cut-off threshold=0.04, min. length=10mm, max. length=250 mm, step=0.005 mm, σT2 = 60°, σN2 = 0.8°, σB2 =0.8°, σK2=0.5, σt2=0.25, r=3.75 mm. No post-processing is done. Initially, bundles are obtained by filtering whole brain tractograms with manually drawn ROIs. Final bundles are obtained by removing dangling streamlines. After extracting the bundles, separate binary masks are generated on the 1 mm template.

Team 9

First, a morphological T1-weight image is used to segment the brain (estimate a new mask or tracts identified), identify different structures and different tissues. Then, whole-brain tractography with multi-tissue CSD is performed using four tractography methods. Moreover, multi-tractography-based tracts atlas is exploited to combine four tractography and to reduce the error from individuals tractography algorithms. Fiber bundles with multi-tractography-based tracts atlas were finally automatically segmented with a Tract Querier [36] method. Finally, tracts space will be converted to DWI space by affine transform to mapping tracts to binary images.

RESULTS

Table 1 presents a more detailed technical contribution of each of the works:

Table 1:

The table presents all the hyper-parameters of the different algorithms that were submitted and an overall evaluation of the algorithm in terms of ICC and Dice.

Synapse Submission id Algorithm ID ICC DICE b-value shells HARDI/Tensor Model Step size Threshold angle Additional Pre-Processing Post-Processing
syn8533598 1A 0.7753 0.6364 All shells CSD 0.2mm 30 degrees NA Distance transform of bundle volumes
syn8643780 1B 0.6857 0.6596 All shells CSD 0.2mm 30 degrees NA NA
syn8643793 1C 0.6343 0.6346 All shells CSD 0.2mm 30 degrees NA Distance transform of bundle volumes
syn8648608 1D 0.7707 0.5402 All shells CSD 0.2mm 30 degrees NA Distance transform of bundle volumes
syn8649314 1E 0.6498 0.6508 All shells CSD 0.2mm 30 degrees NA NA
syn8649322 1F 0.6192 0.6197 All shells CSD 0.2mm 30 degrees NA Distance transform of bundle volumes
syn8649611 1G 0.6324 0.6332 All shells CSD 0.2mm 30 degrees NA Automatic spurious fiber removal
syn8649618 1H 0.6494 0.6503 All shells CSD 0.2mm 30 degrees NA Automatic spurious fiber removal
syn8649622 1I 0.6517 0.6526 All shells CSD 0.2mm 30 degrees NA Automatic spurious fiber removal
syn8649650 1J 0.6662 0.6671 All shells CSD 0.2mm 30 degrees NA Automatic spurious fiber removal
syn8649652 1K 0.6616 0.6624 All shells CSD 0.2mm 30 degrees NA Automatic spurious fiber removal
syn8649654 1L 0.6362 0.637 All shells CSD 0.2mm 30 degrees NA Automatic spurious fiber removal
syn8649656 1M 0.7093 0.7103 All shells CSD 0.2mm 30 degrees NA Automatic spurious fiber removal
syn8649658 1N 0.6984 0.6994 All shells CSD 0.2mm 30 degrees NA Automatic spurious fiber removal
syn8555229 2A 0.8506 0.7918 All shells + 30 HCP subjects CSD + U-net 0.2mm 20 degrees NA Spurious Fiber Removal
syn8656474 3A 0.7379 0.7253 b1000 and b2000 Tensor Variant 0.2mm 25 degrees Data Upsampling NA
syn8656475 3B 0.6463 0.6341 b1000 and b2000 Tensor Variant 0.2mm 25 degrees Data Upsampling NA
syn8662707 4A 0.5285 0.5317 b3000 CSD 0.2mm 20 degrees Data Upsampling NA
syn8662708 4B 0.5822 0.3207 b3000 CSD 0.2mm 20 degrees Data Upsampling NA
syn8662709 4C 0.5881 NaN b3000 CSD 0.2mm 20 degrees Data Upsampling NA
syn8662710 4D 0.5285 0.5317 b3000 CSD 0.2mm 20 degrees Data Upsampling NA
syn8662711 4E 0.5781 0.3182 b3000 CSD 0.2mm 20 degrees Data Upsampling NA
syn8662712 4F 0.5835 NaN b3000 CSD 0.2mm 20 degrees Data Upsampling NA
syn8662713 4G 0.5285 0.5317 b3000 CSD 0.2mm 20 degrees Data Upsampling NA
syn8662714 4H 0.5291 0.4932 b3000 CSD 0.2mm 20 degrees Data Upsampling NA
syn8662715 4I 0.5302 NaN b3000 CSD 0.2mm 20 degrees Data Upsampling NA
syn8662716 4J 0.5285 0.5317 b3000 CSD 0.2mm 20 degrees Data Upsampling NA
syn8662717 4K 0.5596 0.5323 b3000 CSD 0.2mm 20 degrees Data Upsampling NA
syn8662718 4L 0.5616 NaN b3000 CSD 0.2mm 20 degrees Data Upsampling NA
syn8664905 5A 0.9738 0.8231 All shells CSD 1.25mm 45 degrees Additional Segmentation SIFT2
syn8666133 6A 0.7702 0.7708 All shells Tensor Variant 1mm 40 degrees Denoising, Upsampling Outlier Rejection
syn8666134 6B 0.8358 0.5742 All shells Tensor Variant 1mm 40 degrees Denoising, Upsampling Outlier Rejection
syn8666135 6C 0.8171 0.7595 All shells Tensor Variant 1mm 40 degrees Denoising, Upsampling Outlier Rejection
syn8666136 6D 0.817 0.7704 All shells Tensor Variant 1mm 40 degrees Denoising, Upsampling Outlier Rejection
syn8666137 6E 0.8586 0.571 All shells Tensor Variant 1mm 40 degrees Denoising, Upsampling Outlier Rejection
syn8666138 6F 0.8458 0.7646 All shells Tensor Variant 1mm 40 degrees Denoising, Upsampling Outlier Rejection
syn8667007 7A 0.8868 0.6187 b3000 CSD 1.25mm 40 degrees NA NA
syn8666587 8A 0.86 0.8672 All shells Compartment Model 0.005mm 60 degrees NA Spurious Fiber Removal
syn8666598 8B 0.8367 0.5166 All shells Compartment Model 0.005mm 60 degrees NA Spurious Fiber Removal
syn8666602 8C 0.8349 0.5287 All shells Compartment Model 0.005mm 60 degrees NA Spurious Fiber Removal
syn8666936 8D 0.8901 0.6409 All shells Compartment Model 0.005mm 60 degrees NA Spurious Fiber Removal
syn8667021 8E 0.8145 0.4983 All shells Compartment Model 0.005mm 60 degrees NA Spurious Fiber Removal
syn8667022 8F 0.8103 0.4773 All shells Compartment Model 0.005mm 60 degrees NA Spurious Fiber Removal
syn8698866 9A 0.6145 0.6015 All shells CSD 0.2mm 40 degrees Addtional Segmentation NA
syn8698867 9B 0.6968 0.6804 All shells CSD 0.2mm 40 degrees Addtional Segmentation NA
syn8698868 9C 0.2703 0.2572 All shells CSD 0.2mm 40 degrees Addtional Segmentation NA
  • Team 1, Team 5, Team 6, Team 8 and Team 9 used all three shells of b-values provided in the dataset. Team 2 used all shells with data from an additional 30 subjects from the Human Connectome Project. Team 3 used shells of b-values 1000 and 2000 s/mm2. Team 7 and Team 4 only used the shell of b-value 3000 s/mm2.

  • Additional pre-processing has been used by four teams. Team 4: Data was up-sampled to 1mm isotropic resolution. Team 6 used image de-noising techniques and up-sampled the data to 1.25mm. Team 5 and Team 9 used different styles of segmentation of the data presented for analysis.

  • In terms of the fiber detection model, Team 6 and Team 3 used variants of tensor models while the others have used different variants of constrained spherical deconvolution. Notably Team 8 used a compartment analysis model using spherical harmonics.

  • Considering the tractography parameters - the range of step sizes that have been used lie between 0.2–1.25mm. Threshold angle lies in the range of 20–40 degrees.

  • Single fiber assumptions were considered with the condition of FA > 0.7 by teams Team 1 and Team 4. A notable observation here is that a general assumption was made by Team 6 to reject voxels which were less than 0.15 FA.

  • Team 2, Team 6 and Team 8 post-processed the tractography results for removal of spurious fibers by defining different and specific constraints.

  • Of note, Team 2 treated the tractography problem as a segmentation problem and developed a U-net which was trained on the HCP data. While Team 9 used a multi-atlas approach to tractography. The other teams used the general approach of probabilistic or deterministic tractography.

An overlay of all 46 submissions, for all estimated fiber pathways can be observed (Fig 2 Column 1 & 3). Only the left side has been shown as the right side is a similar observation. There are vast differences that can be noticed in the estimated pathways. The volume of the brain occupied by each tract from different submissions varied dramatically. When all 46 submissions are overlaid, tracts occupy 14–53% of the brain volumetrically (average – 34%). Specifically, the union of all entries for FNX (L/R), CNG (L/R), IFO (L/R) and SLF (L/R) cover (30.7, 25.8), (40.9, 37.2), (42.4, 46.1), (50.6, 53.3) respectively, while CST (L/R), ILF (L/R), UNC (L/R) and Fminor and Fmajor cover (23.6, 25.4), (33.4, 33.6), (14.3, 17.4), 44.3 and 34.1. Note that individual submissions appear qualitatively reasonable (Fig 2 Column 2 & 4).

Figure 2.

Figure 2

Left: An overlay of all the 46 submissions from all sessions that were acquired using both scanners per tract Right: An overlay of a single submission using all sessions that were acquired using both scanners per tract A) Uncinate left B) Fornix left C) Cingulum eft D) Corticospinal tract left E) Inferior Longitudinal Fasciculus left F) Inferior Fronto-Occipital Fasciculus left G) Superior Longitudinal Fasciculus left H) Fminor.

The number of algorithmic submission’s team wise are Team 1: 14, Team 2: 1, Team 3: 2, Team 4: 12, Team 5: 1, Team 6: 6, Team 7: 1, Team 8: 6 and Team 9: 3. It can be observed that the ICC range for the set of algorithms on a per team basis does not show a lot of variance. The ICC range of algorithms per team are Team 1 (0.61 – 0.77), Team 4 (0.52 – 0.58), Team 6 (0.77 – 0.85), Team 8 (0.81 – 0.89), Team 9 (0.27 – 0.69), Team 3 (0.64, 0.73), Team 2 (0.85), Team 7 (0.88) and Team 5 (0.97). The teams that submitted more than 3 algorithms show an average difference of 0.1 in terms of ICC.

Violin plots (depict the probability density of the data) of ICC and Dice for intra-session reproducibility, inter-session, and inter-scanner measures of reproducibility are presented in Figures 4, 5 and 6, respectively. Since the observations are highly similar in the afore-mentioned figures we only present a detailed comment on Figure 4 which holds true for Figure 5 and 6 as well. This figure helps in identifying the low, moderate and high reproducibility tracts. The intra-session distributions (Figure 4B) across entries for UNC (L/R) and FNX (L/R) are bi-modal with a median of the lower mode less than 0.4 ICC. The CST (L) has a smaller fraction of the entries with ICC less than 0.4, while the remainder of the entries have only a few outlier entries less than 0.4 The inter-session (Fig. 5) and inter-scanner (Fig. 6) distributions were similar, with a slight increase in outlier entries for IFO (L/R). The patterns in the dice were similar when using a quality threshold of less than 0.4 dice.

Figure 4.

Figure 4

Violin plots of intra-session submissions across both the scanners per tract. A) Dice similarity coefficients B) Intra-class correlation coefficients. The top row depicts the median of the top five intra session submissions. The tracts are in the following order (L/R): a) Uncinate b) Fornix c) Fminor & Fmajor d) Cingulum e) Corticospinal tract f) Inferior longitudinal fasciculus g) Superior longitudinal fasciculus h) Inferior fronto-occipital tract

Figure 5.

Figure 5

Violin plots of inter-session submissions across both the scanners per tract. A) Dice similarity coefficients B) Intra-class correlation coefficients. The top row depicts the median of the top five inter session submissions. The tracts are in the following order (L/R): a) Uncinate b) Fornix c) Fminor & Fmajor d) Cingulum e) Corticospinal tract f) Inferior longitudinal fasciculus g) Superior longitudinal fasciculus h) Inferior fronto-occipital tract

Figure 6.

Figure 6

Violin plots of inter-scanner submissions across both the scanners per tract. A) Dice similarity coefficients B) Intra-class correlation coefficients. The top row depicts the median of the top five inter scanner submissions. The tracts are in the following order (L/R): a) Uncinate b) Fornix c) Fminor & Fmajor d) Cingulum e) Corticospinal tract f) Inferior longitudinal fasciculus g) Superior longitudinal fasciculus h) Inferior fronto-occipital tract

We define cutoffs for high, moderate, and low reproducibility on the inter-scanner reproducibility. High reproducibility was defined as a median ICC greater than 0.6 and less than 5% of entries less than 0.4 ICC. Moderate reproducibility was defined as median ICC greater than 0.4 and less than 25% of entries less than 0.4 ICC. Low reproducibility was defined as a median ICC less than 0.4 or more than 25% of entries less than 0.4 ICC. Hence, the high reproducibility tracts were Fminor, CST (/R), ILF (L/R), SLF (L/R) and IFO (L/R). The moderate reproducibility tracts were CST (L), Fmajor, CNG (L/R). The low reproducibility tracts were UNC (L/R) and FNX (L/R). This above is observed when looking at all submissions however when observing the top 5 submissions we see higher reproducibility.

When the analysis is restricted to only the top five submissions, we see a different picture that suggests substantively reproducible methods. The inter-scanner reproducibility among the top 5 entries in ICC (min-max, average) are shown in Fig 6.

Figure 7 illustrates the top five entries for the tracts with the lowest inter-scanner reproducibility alongside the volumetric median (median per voxel from five submissions) of the top five entries. Qualitatively, the volumetric profiles of the UNC (L/R) and FNX (L/R) are very different across the top five entries. The first submission has small “core” tracts labeled, while the second, third and fifth found much larger spatial extents and the fourth was mid-way between.

Figure 7.

Figure 7

First row shows the median of Uncinate (L/R) and the top five submissions. The second row shows the median and submissions of Fornix (L/R).

DISCUSSION

The most reproducible tracts were Fminor, CST (\R), ILF (L\R), SLF (L\R), IFO (L\R), while the moderately reproducible tracts were Fmajor, CNG (L\R) and CST (\L). Lowest reproducibility tracts are UNC (L\R), FNX (L\R). These tracts have a well-spread/broad probability distribution. Note that the reproducibility of these tracts was maintained across imaging sessions and change of scanner. It is evident that all the algorithms entered are not consistently identifying the same fiber structures given the extreme variance observed. While most of the individual submissions show a reasonable detection of the tracts if observed from a ROI point of view, the difference between tract volumes between methods is quite high.

Previous validation studies have assessed sensitivity and specificity with respect ground truth. For example, comparisons with phantoms, histology, and known anatomy have provided valuable insight into the limitations of tractography [53]. Fibercup shows limitations in terms of tuning of different parameters of tractography [22]. The ISMRM 2015 challenge shows false positives and ambiguity. The 3D Votem challenge shows that sensitivity and specificity are limited. However, rather than anatomical accuracy, this study focuses on the variability of the algorithms themselves, specifically, the volume of the pathways where it was assessed by the containment analysis to assess if there is an ordering for the containment of volumes.

The data acquisition is limited by the acquisitions of two scanners. Alternative approach would be similar to MUSHAC challenge where their ultimate goal was harmonization, but there were multiple datasets across multiple scanners. This study investigated the reproducibility and containment of volume overlap, and did not assess the trajectories of streamlines themselves. For example, streamline density, orientation, and trajectories would provide detailed analysis of the similarities and differences within and across algorithms on the scale of individual voxels.

This study presents a perspective of reproducibility as compared to prior studies which characterize specificity and sensitivity with reproducibility as a limitation. This study shows that reproducibility can be characterized with an indirect estimation of sensitivity and specificity analysis (containment analysis).

However, the study is limited by data acquisition of a single subject. For a better reproducibility validation multi-subject and multi-site (>2) acquisitions would be valuable. Another limitation is that there are quite a few hyper-parameters that can be tweaked for pre and post-processing. In general, the tweaking improves reproducibility but a deterministic study of which combinations maximize reproducibility without tradeoff’s on specificity and sensitivity would be valuable to the diffusion community.

In conclusion, there are 8, 4 and 4 tracts which are depict high, moderate and low reproducibility. The most reproducible algorithms are 5A, 8D, 7A, 6E and 6F as per criteria of ICC. The mentioned algorithms are not an example of a consistent null learning as they all lie with in a nested containment with the largest covered volume.

The 2017 ISMRM TraCED Challenge created a publicly available multi-scanner, multi-scan in-vivo reproducibility dataset and engaged nine groups with 46 algorithm entries. The TraCED Challenge dataset is freely available at www.synapse.org. Consistent with previous studies, reproducibility of tractograms was found to vary by anatomical tract. When viewed across all entries, reproducibility was concerning (ICC <0.5); however, the cluster of top performing methods resulting in reassuringly high results (ICC > 0.85). Variation in performance were seen across processing parameters, but the challenge design did not provide sufficient number of samples to identify uniformly preferred design choices. The key novel finding of this challenge is that variations in tractography methods can be largely attributed to larger/smaller volumetric difference tradeoffs for the larger tracts, especially among methods that are tuned towards volumetrically larger tractograms. Yet, the different methods clearly result in fundamentally different tract structures at the more conservative specificity choices (i.e., volumetrically smaller tractograms). The containment index, containment energy, and containment index framework provide a consistent approach to evaluate the nesting structure tractograms, and the freely available data and results from this challenge can be used to quantify new tractography approaches. There is still room for improvements and innovations in the challenge process itself. For example, combining the knowledge and skillsets of the tractography community with the microstructural community could lead to improvements in phantoms, ground truth evaluations, and potentially unique features that could increase tracking accuracy.

Figure 8.

Figure 8

A) Quantifies the number of algorithms that used a specific part of the dataset or added more from other sources. B) Quantifies the usage of HARDI/Tensor methods by different tractography algorithms as a pre-step. C & D) Quantifies the step size and threshold angle parameter for tractography algorithms. E & F) Quantify the number of additional pre-processing and post-processing techniques applied for the tractography algorithms.

Figure 9.

Figure 9

Ordering entries to minimize containment energy (CE) shows that containment index is generally lower for the volumetrically smaller tractograms (toward “inside” on each subplot) and increases for the larger tractograms (toward “outside” on each subplot). Variations in containment explained the least amount of entry variability for the UNC and Fornix, while the other tracts were more consistent. The containment between all methods (A) were more variable and lower than the containment for the top five methods (B).

ACKNOWLEDGEMENTS

This work was supported by R01EB017230 (Landman). This work was conducted in part using the resources of the Advanced Computing Center for Research and Education at Vanderbilt University, Nashville, TN. This project was supported in part by the National Center for Research Resources, Grant UL1 RR024975–01, and is now at the National Center for Advancing Translational Sciences, Grant 2 UL1 TR000445–06. The content is solely the responsibility of the authors and does not necessarily represent the official views of the NIH. This work is supported in part by China Scholarship Council Scholarship. This work was supported in part by the National Natural Science Foundation of China (Grant No. 61379020). An informed consent under the Vanderbilt University (VU) Institutional Review Board (IRB) was obtained to conduct this study.

REFERENCES

  • 1.Basser PJ, Mattiello J, and LeBihan D, MR diffusion tensor spectroscopy and imaging. Biophysical journal, 1994. 66(1): p. 259–267. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Jeurissen B, et al. , Diffusion MRI fiber tractography of the brain. NMR in Biomedicine, 2017. [DOI] [PubMed]
  • 3.Jeurissen B, et al. , Investigating the prevalence of complex fiber configurations in white matter tissue with diffusion magnetic resonance imaging. Human brain mapping, 2013. 34(11): p. 2747–2766. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Tuch DS, Q-ball imaging. Magnetic resonance in medicine, 2004. 52(6): p. 1358–1372. [DOI] [PubMed] [Google Scholar]
  • 5.Tournier J-D, et al. , Direct estimation of the fiber orientation density function from diffusion-weighted MRI data using spherical deconvolution. NeuroImage, 2004. 23(3): p. 1176–1185. [DOI] [PubMed] [Google Scholar]
  • 6.Jansons KM and Alexander DC, Persistent angular structure: new insights from diffusion magnetic resonance imaging data. Inverse problems, 2003. 19(5): p. 1031. [DOI] [PubMed] [Google Scholar]
  • 7.Schilling KG, et al. , Empirical consideration of the effects of acquisition parameters and analysis model on clinically feasible q-ball imaging. Magnetic Resonance Imaging, 2017. 40: p. 62–74. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Nath V, Empirical Estimation of Intra-Voxel Structure with Persistent Angular Structure and Q-ball Models of Diffusion Weighted MRI 2017. [DOI] [PMC free article] [PubMed]
  • 9.Nath V, et al. Comparison of multi-fiber reproducibility of PAS-MRI and Q-ball with empirical multiple b-value HARDI. in Medical Imaging 2017: Image Processing 2017. International Society for Optics and Photonics. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Smith SM, et al. , Tract-based spatial statistics: voxelwise analysis of multi-subject diffusion data. Neuroimage, 2006. 31(4): p. 1487–1505. [DOI] [PubMed] [Google Scholar]
  • 11.Lawes INC, et al. , Atlas-based segmentation of white matter tracts of the human brain using diffusion tensor tractography and comparison with classical dissection. Neuroimage, 2008. 39(1): p. 62–79. [DOI] [PubMed] [Google Scholar]
  • 12.Schmahmann JD, et al. , Association fibre pathways of the brain: parallel observations from diffusion spectrum imaging and autoradiography. Brain, 2007. 130(3): p. 630–653. [DOI] [PubMed] [Google Scholar]
  • 13.Dauguet J, et al. , Comparison of fiber tracts derived from in-vivo DTI tractography with 3D histological neural tract tracer reconstruction on a macaque brain. Neuroimage, 2007. 37(2): p. 530–538. [DOI] [PubMed] [Google Scholar]
  • 14.Dyrby TB, et al. , Validation of in vitro probabilistic tractography. Neuroimage, 2007. 37(4): p. 1267–1277. [DOI] [PubMed] [Google Scholar]
  • 15.Donahue CJ, et al. , Using diffusion tractography to predict cortical connection strength and distance: a quantitative comparison with tracers in the monkey. Journal of Neuroscience, 2016. 36(25): p. 6758–6770. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Thomas C, et al. , Anatomical accuracy of brain connections derived from diffusion MRI tractography is inherently limited. Proceedings of the National Academy of Sciences, 2014. 111(46): p. 16574–16579. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Azadbakht H, et al. , Validation of high-resolution tractography against in vivo tracing in the macaque visual cortex. Cerebral Cortex, 2015. 25(11): p. 4299–4309. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Knösche TR, et al. , Validation of tractography: comparison with manganese tracing. Human brain mapping, 2015. 36(10): p. 4116–4134. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Calabrese E, et al. , A diffusion MRI tractography connectome of the mouse brain and comparison with neuronal tracer data. Cerebral Cortex, 2015. 25(11): p. 4628–4637. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Côté M-A, et al. , Tractometer: towards validation of tractography pipelines. Medical image analysis, 2013. 17(7): p. 844–857. [DOI] [PubMed] [Google Scholar]
  • 21.Maier-Hein KH, et al. , The challenge of mapping the human connectome based on diffusion tractography. Nature communications, 2017. 8(1): p. 1349. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Neher PF, et al. , Fiberfox: facilitating the creation of realistic white matter software phantoms. Magnetic resonance in medicine, 2014. 72(5): p. 1460–1470. [DOI] [PubMed] [Google Scholar]
  • 23.Schilling KG, et al. , Challenges in diffusion MRI tractography–Lessons learned from international benchmark competitions 2018. [DOI] [PMC free article] [PubMed]
  • 24.Daducci A, et al. , Quantitative comparison of reconstruction methods for intra-voxel fiber recovery from diffusion MRI. IEEE transactions on medical imaging, 2014. 33(2): p. 384–399. [DOI] [PubMed] [Google Scholar]
  • 25.Ning L, et al. , Sparse Reconstruction Challenge for diffusion MRI: Validation on a physical phantom to determine which acquisition scheme and analysis method to use? Medical image analysis, 2015. 26(1): p. 316–331. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Pujol S, et al. , The DTI challenge: toward standardized evaluation of diffusion tensor imaging tractography for neurosurgery. Journal of Neuroimaging, 2015. 25(6): p. 875–882. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Andersson JL and Sotiropoulos SN, An integrated approach to correction for off-resonance effects and subject movement in diffusion MR imaging. Neuroimage, 2016. 125: p. 1063–1078. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Andersson JL, Skare S, and Ashburner J, How to correct susceptibility distortions in spin-echo echo-planar images: application to diffusion tensor imaging. Neuroimage, 2003. 20(2): p. 870–888. [DOI] [PubMed] [Google Scholar]
  • 29.Smith SM, et al. , Advances in functional and structural MR image analysis and implementation as FSL. Neuroimage, 2004. 23: p. S208–S219. [DOI] [PubMed] [Google Scholar]
  • 30.Jenkinson M and Smith S, A global optimisation method for robust affine registration of brain images. Medical image analysis, 2001. 5(2): p. 143–156. [DOI] [PubMed] [Google Scholar]
  • 31.Leemans A and Jones DK, The B-matrix must be rotated when correcting for subject motion in DTI data. Magnetic resonance in medicine, 2009. 61(6): p. 1336–1349. [DOI] [PubMed] [Google Scholar]
  • 32.Asman AJ and Landman BA, Non-local statistical label fusion for multi-atlas segmentation. Medical image analysis, 2013. 17(2): p. 194–208. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Huo Y, et al. Combining multi-atlas segmentation with brain surface estimation. in Proceedings of SPIE--the International Society for Optical Engineering 2016. NIH Public Access. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Huo Y, et al. , Consistent cortical reconstruction and multi-atlas brain segmentation. NeuroImage, 2016. 138: p. 197–210. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Garyfallidis E, et al. , Dipy, a library for the analysis of diffusion MRI data 2014. 8: p. 8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Wassermann D, et al. , The white matter query language: a novel approach for describing human white matter anatomy 2016. 221(9): p. 4705–4721. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Daducci A, et al. , COMMIT: convex optimization modeling for microstructure informed tractography 2015. 34(1): p. 246–257. [DOI] [PubMed] [Google Scholar]
  • 38.Wasserthal J, Neher P, and Maier-Hein KHJN, TractSeg-Fast and accurate white matter tract segmentation 2018. 183: p. 239–253. [DOI] [PubMed] [Google Scholar]
  • 39.Pasternak O, et al. , Free water elimination and mapping from diffusion MRI 2009. 62(3): p. 717–730. [DOI] [PubMed] [Google Scholar]
  • 40.Zhang Y, Brady M, and Smith S, Segmentation of brain MR images through a hidden Markov random field model and the expectation-maximization algorithm. IEEE transactions on medical imaging, 2001. 20(1): p. 45–57. [DOI] [PubMed] [Google Scholar]
  • 41.Tournier J-D, Calamante F, and Connelly A, Robust determination of the fibre orientation distribution in diffusion MRI: non-negativity constrained super-resolved spherical deconvolution. Neuroimage, 2007. 35(4): p. 1459–1472. [DOI] [PubMed] [Google Scholar]
  • 42.Girard G, et al. , Towards quantitative connectivity analysis: reducing tractography biases 2014. 98: p. 266–278. [DOI] [PubMed] [Google Scholar]
  • 43.Garyfallidis E, et al. , Recognition of white matter bundles using local and global streamline-based registration and clustering 2018. 170: p. 283–295. [DOI] [PubMed] [Google Scholar]
  • 44.Kaden E, et al. , Multi-compartment microscopic diffusion imaging 2016. 139: p. 346–359. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Wakana S, et al. , Reproducibility of quantitative tractography methods applied to cerebral white matter 2007. 36(3): p. 630–644. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Cabeen RP, Bastin ME, and Laidlaw DHJN, Kernel regression estimation of fiber orientation mixtures in diffusion MRI 2016. 127: p. 158–172. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47.Cabeen R, Laidlaw D, and Toga A. Quantitative imaging toolkit: software for interactive 3D visualization, data exploration, and computational analysis of neuroimaging datasets. in Proceedings 2018.
  • 48.Yeatman JD, et al. , Tract profiles of white matter properties: automating fiber-tract quantification 2012. 7(11): p. e49790. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Catani M, et al. , Virtual in vivo interactive dissection of white matter fasciculi in the human brain 2002. 17(1): p. 77–94. [DOI] [PubMed] [Google Scholar]
  • 50.Tournier J-D, Calamante F, and Connelly AJN, Robust determination of the fibre orientation distribution in diffusion MRI: non-negativity constrained super-resolved spherical deconvolution 2007. 35(4): p. 1459–1472. [DOI] [PubMed] [Google Scholar]
  • 51.Tran G and Y.J.I.t.o.m.i. Shi, Fiber orientation and compartment parameter estimation from multi-shell diffusion imaging 2015. 34(11): p. 2320–2332. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52.Aydogan DB and Shi Y. Probabilistic tractography for topographically organized connectomes. in International Conference on Medical Image Computing and Computer-Assisted Intervention 2016. Springer. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 53.Schilling KG, et al. , Limits to anatomical accuracy of diffusion tractography using modern approaches 2019. 185: p. 1–11. [DOI] [PMC free article] [PubMed] [Google Scholar]