Abstract
Purpose: To investigate the correlation between anterior chamber volume (ACV) and corneal biomechanical properties in healthy and primary angle closure (PAC) eyes.
Methods: A total of 79 eyes from 55 participants were enrolled in this study, including 24 eyes from 17 PAC patients and 55 eyes from 38 normal subjects. Anterior chamber volume was detected via swept-source anterior segment optical coherence tomography (OCT), and the cornea biomechanical data were obtained via corneal visualization Scheimpflug technology (Corvis ST). A student's t-test, Chi-square test, Pearson's correlation coefficients, and linear regression were used in the statistical analysis.
Results: Anterior chamber volume was significantly associated with a lower deformation altitude (DA) (p = 0.033), higher stiffness parameter (SP-A1) (p = 0.005), younger age (p = 0.001), and higher biomechanical intraocular pressure (bIOP) (p = 0.001). PAC patients were suspected to have a much shallower anterior chamber than healthy participants. In the PAC group, the mean ACV (p < 0.01), SP-A1 (p < 0.01), and bIOP values (p = 0.01) were significantly reduced as compared to the normal group, and DA values (p = 0.02) and age (p = 0.01) were increased as compared to the normal group.
Conclusions: Bigger ACV was associated with lower DA values and age, as well as higher SP-A1 and bIOP values. Reduced corneal stiffness was more commonly found in the PAC suspects as compared to their normal counterparts, indicating a protective physiological mechanism for people with shallower anterior chambers that protects against potential elevations of IOP.
Keywords: cornea, biomechanical properties, anterior chamber, primary angle closure, imaging
Introduction
The cornea provides a proper anterior refractive surface for the eye and protects the intraocular contents from both infection and structural damage. The biomechanical properties of the cornea are important in maintaining corneal health. The measurement of these parameters helps in diagnosing certain diseases, such as keratoconus, Fuch's dystrophy, and glaucoma (del Buey et al., 2009; Medeiros et al., 2013; Elham et al., 2017). In addition, the therapeutic manipulation of corneal biomechanics has been introduced as a treatment strategy for ectatic corneal diseases (Wollensak et al., 2003) and presbyopia among others (Krueger, 2009). Thus, an understanding of corneal biomechanics is important in ophthalmology, both for diagnostics and therapeutics.
Measurements of human corneal biomechanics are traditionally destructive and performed after tissue excision (Lombardo et al., 2014). In vivo corneal biomechanics assessment began in 2005 with the introduction of the ocular response analyzer (ORA; Reichert Ocular Instruments, Depew, NY) (Luce, 2005). The ORA is a bidirectional applanation device that allows the indirect assessment of corneal deformation based on infrared light signals. Another method, called corneal visualization Scheimpflug technology (Corvis ST; Oculus Inc., Wetzlar, Germany), applies an additional high-speed Scheimpflug camera to detect changes in corneal shape (Naderan and Jahanrad, 2017; Chan et al., 2018).
Anatomically, the anterior chamber serves as the rear support for the cornea and may also have some effects on corneal biomechanics. Swept-source anterior segment OCT (Casia SS-1000 OCT; Tomey, Nagoya, Japan) has a high scan speed and distinct axial resolution, providing representative imaging of the cornea, iris, and lens for the measurement of anterior chamber volume (Mak et al., 2013). To date, there have been relatively few reports focused on the biomechanical properties of the cornea and their association with the depth or volume of the anterior chamber. Evidence suggests that anterior chamber depth has a negative correlation with corneal hysteresis (CH) (Chang et al., 2010). In contrast, other studies have demonstrated no such association (Hwang et al., 2013; Nemeth et al., 2017). To resolve this issue, proper sample selection and study design are required.
Therefore, the purpose of this study was to investigate the relationship between anterior chamber volume as obtained via Casia SS-1000 OCT and specific corneal deformation parameters determined by Corvis ST.
Methods
Subjects
To examine a wide range of ACV values, both healthy participants and primary angle closure (PAC) patients were chosen as the target in this study. The study was conducted in the Eye and ENT Hospital of Fudan University, Shanghai, China. All the subjects underwent a thorough examination of their visual acuity and intraocular pressure (IOP), as well as slit-lamp ophthalmic examination, fundus examination, optical coherence tomography (OCT), and Corvis-ST test (Oculus, Wetzlar, Germany). The details of the period are listed as following: visual acuity (15 min), intraocular pressure (10 min), slit-lamp ophthalmic examination (20 min), fundus examination (15 min), OCT (25 min), and Corvis-ST test (15 min). Usually, the interval between each examination is more than 10 min. The exclusion criteria were any ocular trauma, anterior segment disease, any intraocular, or refractive surgery. To eliminate the possible influence of refractive errors to anterior chamber, participants with refractive power beyond ±1.5 D were excluded. For the purposes of this study, PAC eyes were defined as eyes in which appositional contact between the peripheral iris and posterior trabecular meshwork (<180 of visible trabecular meshwork upon gonioscopy) without peripheral anterior synechia was detected during gonioscopy or OCT (Foster et al., 2002; Mak et al., 2013). The protocol was in accordance with the ethical standards stated in the 1964 Declaration of Helsinki and approved by our hospital ethics committee. After informed consent was obtained, a total of 79 eyes from 55 participants were enrolled, including 24 eyes from 17 PAC patients and 55 eyes from 38 normal subjects.
Swept-Source Anterior Segment Optical Coherence Tomography
Casia SS-1000 OCT (Tomey, Nagoya, Japan) with a swept-source laser wavelength of 1,310 nm was chosen to measure ACV in this study. This method uses a monochromatic, tunable, fast-scanning laser source, and a photodetector to detect wavelength-resolved interference signals (Yasuno et al., 2005; Liu et al., 2011; Mak et al., 2013). The ACV scan protocol contains 128 radial scans, each 16 mm in length and 6 mm in depth. The participants were asked to fixate on an internal fixation target during the scan. With the technician's help, they elevated the upper lid against the upper orbital rim and lowered the lower lid against the lower orbital rim to expose the entire limbus. After a scan duration of about 2.4 s, 360° of the anterior segment were automatically scanned and saved by the program. Each eye was imaged three times. A single masked observer was selected to measure ACV in all of the image series. The ACV value was automatically calculated by the instrument's software (Figure 1).
Corneal Visualization Scheimpflug Technology
The biomechanical properties of the cornea were measured via corneal visualization Scheimpflug technology, Corvis-ST. The device contains an ultra-high-speed Scheimpflug camera that captures 140 frames within 31 ms with a resolution of 640 × 480 pixels and covers 8.5 mm of the central cornea. While in operation, this device applies a precise air impulse with a Gaussian intensity distribution and a duration of 25 ms to the corneal apex. This air puff causes the cornea to move inwards, creating the first applanation point (applanation 1) and then a slight concavity. When the pressure from the air puff decreases, the cornea gradually returns to its normal convex curvature, passing through a second applanation point (applanation 2) (Salvetat et al., 2015). The Corvis-ST printout provides IOP and CCT measurements, and 13 numerical corneal deformation parameters are listed in Table 1 (Salvetat et al., 2015; Joda et al., 2016).
Table 1.
Parameters | Definition |
---|---|
A1L | Cord length of the first-degree applanation |
A1V | Corneal speed during the first-degree applanation |
A2L | Cord length of the second-degree applanation |
A2V | Corneal speed during the second-degree applanation |
PD | Distance of the 2 knee's at highest concavity |
Radius | Central concave curvature at highest concavity |
DA | Maximum amplitude at the apex of highest concavity |
CCT | Central corneal thickness |
IOP | Intraocular pressure (uncorrected non-contact tonometer) |
bIOP | biomechanically-corrected IOP |
DARatio | The ratio between the deformation/deflection amplitude at the apex and the average deformation/deflection amplitude measured at 1 or 2 mm from the center |
IR | Integrated radius, the inverse concave radius (1/R) between the first and second application events |
ARTh | Ambrosio's Relational Thickness to the horizontal profile, the devision between corneal thickness at the thinnest point and the Pachymetric Progression. |
SP-A1 | Stiffness parameter at first applanation, the resultant pressure (Pr) decided by deflection amplitude at first applanation (A1) |
CBI | Corvis Biomechanical Index, a combined biomechanical index based on different dynamic corneal response parameters, to aid the diagnosis of ectasia |
Statistical Analysis
The statistical analysis was performed using SPSS Version 20 (SPSS Inc. Chicago, IL, USA). Descriptive statistical results were written as mean ± standard deviations (SDs). A Pearson's correlation analysis was used to investigate the relationship between ACV and demographic/corneal biomechanical parameters. A multivariate linear regression analyses were performed using any significantly changed parameters as independent variables. A student's t-test and Chi-square test were used to compare the differences between PAC patients and healthy participants. The correlation coefficients were recorded, and a P-value below 0.05 was considered statistically significant.
Results
Our study enrolled 79 eyes from 55 participants, including 24 eyes from 17 PAC patients and 55 eyes from 38 normal subjects. The demographics, ACV values, and corneal biomechanical parameters of the participants (mean ± SD and range) were shown in Table 2. The average ACV value was 140.83 ± 38.11 mm3, and these values ranged from 79.58 to 256.93 mm3.
Table 2.
Total (n = 79) | PAC patients (n = 24) | Normal (n = 55) | P-value | |
---|---|---|---|---|
Mean ± SD (Range) | Mean ± SD | Mean ± SD | ||
Age (y) | 52.34 ± 14.34 (22–86) | 57.96 ± 10.92 | 49.89 ± 15.04 | 0.01** |
Sex (M/F) | 26/53 | 7 / 17 | 19 / 36 | 0.64 |
Height (cm) | 162.15 ± 8.38 (145–178) | 160.78 ± 6.27 | 162.72 ± 9.11 | 0.28 |
Weight (kg) | 62.63 ± 10.28 (45–92) | 61.26 ± 8.46 | 63.21 ± 10.97 | 0.45 |
ACV (mm3) | 140.83 ± 38.11 (79.58–256.93) | 106.00 ± 26.75 | 156.02 ± 31.92 | <0.01** |
A1L (mm) | 2.25 ± 0.32 (1.50–2.85) | 2.15 ± 0.37 | 2.30 ± 0.29 | 0.10** |
A1V (ms) | 0.14 ± 0.02 (0.10–0.18) | 0.15 ± 0.01 | 0.14 ± 0.02 | <0.01** |
A2L (mm) | 1.93 ± 0.37 (1.19–3.23) | 1.88 ± 0.28 | 1.95 ± 0.40 | 0.44 |
A2V (ms) | −0.28 ± 0.03 (−0.34 to −0.14) | −0.28 ± 0.02 | −0.27 ± 0.03 | 0.15 |
Peak Distance (mm) | 4.90 ± 0.21 (4.31–5.33) | 4.93 ± 0.19 | 4.89 ± 0.22 | 0.46 |
Radius (mm) | 7.18 ± 0.77 (5.56–9.25) | 7.02 ± 0.59 | 7.25 ± 0.82 | 0.21 |
DA (mm) | 1.05 ± 0.10 (0.77–1.29) | 1.09 ± 0.09 | 1.03 ± 0.09 | <0.01** |
CCT (μm) | 548 ± 33.86 (463–636) | 533.33 ± 34.80 | 554.16 ± 31.72 | 0.01* |
IOP (mmHg) | 15.43 ± 3.02 (9.50–21.50) | 14.91 ± 2.10 | 16.09 ± 2.55 | <0.01** |
bIOP (mmHg) | 14.30 ± 2.48 (9.60–20.50) | 14.15 ± 1.87 | 14.79 ± 2.56 | <0.01** |
DARatio | 4.33 ± 0.48 (3.20–5.60) | 4.55 ± 0.38 | 4.23 ± 0.49 | <0.01** |
IR (mm) | 8.28 ± 1.05 (6.10–10.50) | 8.65 ± 0.89 | 8.11 ± 1.08 | 0.04** |
ARTh (μm) | 430.12 ± 91.06 (195.30–714.50) | 419.17 ± 91.45 | 434.70 ± 91.35 | 0.50 |
SP-A1 | 126.04 ± 19.42 (86.40–175.20) | 114.61 ± 18.29 | 131.03 ± 17.86 | <0.01** |
CBI | 0.13 ± 0.26 (0.00–1.00) | 0.24 ± 0.33 | 0.08 ± 0.21 | 0.04** |
PAC, primary angle closure; SD, standard deviation; ACV, anterior chamber volume. Other abbreviations of corneal biochemical parameters were listed in Table 1.
P < 0.05 considered statistically significant.
Association Between ACV and Corneal Biomechanical Parameters
In the Pearson's correlation, only age (r = −0.506, p < 0.01) (and not most of the demographic factors, such as sex, height, and weight) was negatively correlated with ACV (Table 3). However, several characteristics of the cornea were associated with ACV. A larger ACV correlated with slower A1V (r = −0.363, p < 0.01), faster A2V (r = 0.278, p < 0.01); reduced DA (r = −0.406, p < 0.01), DARatio (r = −0.375, p < 0.01), IR (r = −0.371, p < 0.01), and CBI (r = −0.255, p = 0.02); and higher IOP (r = 0.510, p < 0.01), bIOP (r = 0.600, p < 0.01), and SP-A1 (r = 0.388, p < 0.01) (Table 3).
Table 3.
Variables | Pearson's correlation coefficient | P-value |
---|---|---|
Age (y) | −0.506 | <0.01** |
Sex (M/F) | −0.103 | 0.37 |
High (cm) | 0.211 | 0.06 |
Weight (kg) | 0.130 | 0.26 |
A1L (mm) | 0.134 | 0.24 |
A1V (ms) | −0.363 | <0.01** |
A2L (mm) | 0.118 | 0.30 |
A2V (ms) | 0.278 | <0.01** |
Peak Distance (mm) | −0.158 | 0.16 |
Radius (mm) | 0.069 | 0.54 |
DA (mm) | −0.406 | <0.01** |
CCT (μm) | 0.130 | 0.25 |
IOP (mmHg) | 0.510 | <0.01** |
bIOP (mmHg) | 0.600 | <0.01** |
DARatio | −0.375 | <0.01** |
IR (mm) | −0.371 | <0.01** |
ARTh (μm) | 0.137 | 0.233 |
SP-A1 | 0.388 | <0.01** |
CBI | −0.255 | 0.02** |
P < 0.05 considered statistically significant.
Next, all the parameters statistically correlated with ACV were thrown into a linear regression model using backward selection. For backward selection, we started with all variables in the model, then iteratively removed the variable with largest p-value, until all remaining variables had a significant p-value (p < 0.05). The results showed that ACV was negatively associated with age (coefficient = −0.334, p < 0.01) and DA (coefficient = −0.337, p = 0.03), but positively associated with bIOP (coefficient = 1.020, p < 0.01) and SP-A1 (coefficient = 0.448, p < 0.01; Table 4).
Table 4.
Variables | Coefficient | P-value |
---|---|---|
Age (y) | −0.334 | <0.01 |
DA (mm) | −0.337 | 0.03 |
bIOP (mmHg) | 1.020 | <0.01 |
SP-A1 | 0.448 | <0.01 |
The adjusted R2 value of this multivariable linear model was 0.482. P < 0.05 considered statistically significant.
Comparison Between PAC Patients and the Normal Group
The enrolled PAC patients were slightly older than the participants in the normal group (57.96 ± 10.92 and 49.89 ± 15.04 y, respectively, p = 0.01). There were no significant differences in terms of other demographics, such as sex, height, and weight, between the two groups. The mean ACV in PAC patients was 106.00 ± 26.75 mm3, significantly shallower than in normal participants (156.02 ± 31.92 mm3, p < 0.01, Table 2, Figure 2). Besides, it is worth mentioning that the SP-A1 was significantly reduced in the PAC patients as compared to the normal participants (114.61 ± 18.29 and 131.03 ± 17.86, respectively, p < 0.01, Figure 2), which was in accordance with the previous linear regression results (Table 2). The other differences between the two groups are listed in Table 2.
Discussion
This study reported the mean ACV values of PAC patients and normal subjects in a Chinese population, along with their correlated factors. To our knowledge, few such studies have been performed. Measuring these factors with swept-source anterior segment OCT and corneal visualization Scheimpflug technology, we showed that the mean ACV in PAC patients was 106.00 ± 26.75 mm3, while in normal participants, this increased to 156.02 ± 31.92 mm3. ACV was negatively correlated with age and DA but positively correlated with bIOP and SP-A1. In the group of PAC patients, SP-A1 was significantly reduced.
The correlation between ACV and corneal biomechanics has been discussed in few prior studies. Some showed that anterior chamber depth was negatively correlated with corneal hysteresis, but no significant correlation with corneal resistance factor was observed (Chang et al., 2010). Another study implied that corneal hysteresis was positively associated with the corneal volume, though the association between CH and anterior chamber volume was not significant (Hwang et al., 2013). Regarding corneal biomechanical factors, most of these studies chose to measure corneal hysteresis and corneal resistance factor via ORA (Krueger, 2009; Sedaghat et al., 2017). Similarly, Corvis ST offers parameters about ocular biomechanics as well as ORA and also provides a better understanding of dynamic corneal response (Vinciguerra et al., 2016). To our knowledge, only one paper has calculated the association between ACV and Corvis ST parameters, and this paper reported a negative result (Nemeth et al., 2017). The reason for these conflicting results remains unknown. One possible reason was that the subjects of the studies were different. Nemeth et al. enrolled 43 eyes from Hungarian people, while we collected 79 Chinese eyes. Another possible reason may be due to the different inclusion criteria. To enlarge the range the ACV values, PAC patients were enrolled in this study. Thus, in this study, ACV values ranged from 79.58 to 256.93 mm3. Finally, new Corvis ST parameters were discussed in this study which was absent in previous reports. Specifically, several updated parameters, such as DARatio, IR, ARTh, and SP-A1, were discussed, providing a more comprehensive examination of corneal biomechanics.
The average ACV in this study was 106.00 mm3 in PAC patients and 156.02 mm3 in normal participants. These numbers are in accordance with those found in other studies. Mark et al. investigated Hong Kong citizens, and ACV was found to be 87.6 mm3 in PAC patients and 133.0 mm3 in normal eyes (Mak et al., 2013). Lam et al. found ACV be 161.03 mm3 in 50 healthy middle-aged Chinese people (Sedaghat et al., 2017). In our study, results showed that ACV was negatively correlated with age. Decreases in ACV with age have been reported previously (Lam and Tse, 2013; Hashemi et al., 2016). The reason for this may be thickened crystalline lenses and relaxed accommodation in older people (Lam and Tse, 2013). Except for age, there were multiple corneal biomechanical parameters associated with ACV, as the findings of Pearson's correlation analyses revealed. However, these parameters may have correlations between one and another. To exclude the factors that may be confounding, we built a multivariable linear model to pick up the relevant factors, which may contribute to ACV variations, to strengthen the predicting power. Results showed that ACV was negatively correlated with DA while positively correlated with bIOP and SP-A1. To our knowledge, associations between DA, bIOP, SP-A1, and ACV have been rarely discussed before. Of these factors, SP-A1 shows a consistent and significant association with ACV, as confirmed by the fact that corneal stiffness is reduced in shallow anterior chambers. The novel stiffness parameter SP-A1 is defined as the adjusted pressure at A1 (adjusted AP1) minus a biomechanically corrected IOP value (bIOP) and then divided by A1 deflection amplitude (Vinciguerra et al., 2016). Liu et al. showed that in the stiffened eye, IOP elevations were significantly higher, even at small volume changes, which means that stiffer corneas may be associated with rapid and higher-magnitude IOP fluctuations (Liu and He, 2009).
We note that our study has several limitations. First, the sample selection was biased in some respects. Female subjects were more common than male participants, although the gender imbalance between PAC patients and normal subjects was not significant. Second, the subjects enrolled in the PAC group were older than their normal counterparts. A previous report showed that there was a decrease in corneal stiffness with an increase in age (Sharifipour et al., 2016), which may partly explain why corneal stiffness decreases in our PAC patients. Another limitation of the study is the combined use of the fellow eyes of some participants. Further investigations should be conducted to confirm our results and evaluate SP-A1 in PACG or primary open angle glaucoma (POAG) patients.
In conclusion, our study showed that smaller ACV was associated with decreased DA and age and increased SP-A1 and bIOP in the PAC patients and the healthy subjects. Reduced corneal stiffness was found in the PAC patients as compared to their normal counterparts, indicating a protective physiological mechanism against potential IOP elevations in people with shallower anterior chambers.
Data Availability Statement
The datasets supporting the conclusions of this article are available from the corresponding author on reasonable request.
Ethics Statement
The protocol was in accordance with the ethical standards stated in the 1964 Declaration of Helsinki and approved by our hospital ethics committee. After informed consent was obtained, a total of 79 eyes from 55 participants were enrolled, including 24 eyes from 17 PAC patients and 55 eyes from 38 normal subjects.
Author Contributions
JH, JX, and XC designed the experiment. XC, YY, YL, FH, YZ, HC conducted the experiment. After analyzing the data, XC wrote the initial draft. All authors discussed the analyzed data and the interpretations. JH, JX, and AM revised the manuscript and made the final version.
Conflict of Interest
The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Footnotes
Funding. This work was supported by grants from the National Natural Science Foundation of China (81970766, 81670818, 81670820, 81870630, 81900819, and 81700807); the Young Scientist Excellence Program, Shanghai (2017YQ055); Shanghai Rising-Star Program (18QA1401100); and the Guizhou Science and Technology Program (2016-2825). The sponsor or funding organization had no role in the design or conduct of this research.
References
- Chan T. C. Y., Wang Y. M., Yu M., Jhanji V. (2018). Comparison of corneal tomography and a new combined tomographic biomechanical index in subclinical keratoconus. J. Refract. Surg. 34, 616–621. 10.3928/1081597X-20180705-02 [DOI] [PubMed] [Google Scholar]
- Chang P. Y., Chang S. W., Wang J. Y. (2010). Assessment of corneal biomechanical properties and intraocular pressure with the Ocular Response Analyzer in childhood myopia. Br. J. Ophthalmol. 94, 877–881. 10.1136/bjo.2009.158568 [DOI] [PubMed] [Google Scholar]
- del Buey M. A., Cristóbal J. A., Ascaso F. J., Lavilla L., Lanchares E. (2009). Biomechanical properties of the cornea in Fuchs' corneal dystrophy. Invest. Ophthalmol. Vis. Sci. 50, 3199–3202. 10.1167/iovs.08-3312 [DOI] [PubMed] [Google Scholar]
- Elham R., Jafarzadehpur E., Hashemi H., Amanzadeh K., Shokrollahzadeh F., Yekta A., et al. (2017). Keratoconus diagnosis using Corvis ST measured biomechanical parameters. J. Curr. Ophthalmol. 29, 175–181. 10.1016/j.joco.2017.05.002 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Foster P. J., Buhrmann R., Quigley H. A., Johnson G. J. (2002). The definition and classification of glaucoma in prevalence surveys. Br. J. Ophthalmol. 86, 238–242. 10.1136/bjo.86.2.238 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Hashemi H., Khabazkhoob M., Mohazzab-Torabi S., Emamian M. H., Shariati M., Dadbin N., et al. (2016). Anterior chamber angle and anterior chamber volume in a 40- to 64-year-old population. Eye Contact Lens 42, 244–249. 10.1097/ICL.0000000000000192 [DOI] [PubMed] [Google Scholar]
- Hwang H. S., Park S. K., Kim M. S. (2013). The biomechanical properties of the cornea and anterior segment parameters. BMC Ophthalmol. 13:49. 10.1186/1471-2415-13-49 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Joda A. A., Shervin M. M., Kook D., Elsheikh A. (2016). Development and validation of a correction equation for Corvis tonometry. Comput. Methods Biomech. Biomed. Eng. 19, 943–953. 10.1080/10255842.2015.1077515 [DOI] [PubMed] [Google Scholar]
- Krueger R. R. (2009). Biomechanical manipulation: the next frontier in corneal refractive surgery. J. Refract. Surg. 25, 837–840. 10.3928/1081597X-20090917-04 [DOI] [PubMed] [Google Scholar]
- Lam A. K., Tse J. S. (2013). Pentacam anterior chamber parameters in young and middle-aged Chinese. Clin. Exp. Optom. 96, 85–91. 10.1111/j.1444-0938.2012.00795.x [DOI] [PubMed] [Google Scholar]
- Liu J., He X. (2009). Corneal stiffness affects IOP elevation during rapid volume change in the eye. Invest. Ophthalmol. Vis. Sci. 50, 2224–2229. 10.1167/iovs.08-2365 [DOI] [PubMed] [Google Scholar]
- Liu S., Yu M., Ye C., Lam D. S., Leung C. K. (2011). Anterior chamber angle imaging with swept-source optical coherence tomography: an investigation on variability of angle measurement. Invest. Ophthalmol. Vis. Sci. 52, 8598–8603. 10.1167/iovs.11-7507 [DOI] [PubMed] [Google Scholar]
- Lombardo G., Serrao S., Rosati M., Lombardo M. (2014). Analysis of the viscoelastic properties of the human cornea using Scheimpflug imaging in inflation experiment of eye globes. PLoS ONE 9:e112169. 10.1371/journal.pone.0112169 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Luce D. A. (2005). Determining in vivo biomechanical properties of the cornea with an ocular response analyzer. J. Cataract Refract. Surg. 31, 156–162. 10.1016/j.jcrs.2004.10.044 [DOI] [PubMed] [Google Scholar]
- Mak H., Xu G., Leung C. K. (2013). Imaging the iris with swept-source optical coherence tomography: relationship between iris volume and primary angle closure. Ophthalmology 120, 2517–2524. 10.1016/j.ophtha.2013.05.009 [DOI] [PubMed] [Google Scholar]
- Medeiros F. A., Meira-Freitas D., Lisboa R., Kuang T. M., Zangwill L. M., Weinreb R. N. (2013). Corneal hysteresis as a risk factor for glaucoma progression: a prospective longitudinal study. Ophthalmology 120, 1533–1540. 10.1016/j.ophtha.2013.01.032 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Naderan M., Jahanrad A. (2017). Topographic, tomographic and biomechanical corneal changes during pregnancy in patients with keratoconus: a cohort study. Acta Ophthalmol. 95, e291–e296. 10.1111/aos.13296 [DOI] [PubMed] [Google Scholar]
- Nemeth G., Szalai E., Hassan Z., Lipecz A., Flasko Z., Modis L. (2017). Corneal biomechanical data and biometric parameters measured with Scheimpflug-based devices on normal corneas. Int. J. Ophthalmol. 10, 217–222. 10.18240/ijo.2017.02.06 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Salvetat M. L., Zeppieri M., Tosoni C., Felletti M., Grasso L., Brusini P. (2015). Corneal deformation parameters provided by the corvis-ST pachy-tonometer in healthy subjects and glaucoma patients. J. Glaucoma 24, 568–574. 10.1097/IJG.0000000000000133 [DOI] [PubMed] [Google Scholar]
- Sedaghat M. R., Mohammad Zadeh V., Fadakar K., Kadivar S., Abrishami M. (2017). Normative values and contralateral comparison of anterior chamber parameters measured by Pentacam and its correlation with corneal biomechanical factors. Saudi J. Ophthalmol. 31, 7–10. 10.1016/j.sjopt.2016.11.006 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Sharifipour F., Panahi-Bazaz M., Bidar R., Idani A., Cheraghian B. (2016). Age-related variations in corneal biomechanical properties. J. Curr. Ophthalmol. 28, 117–122. 10.1016/j.joco.2016.05.004 [DOI] [PMC free article] [PubMed] [Google Scholar]
- Vinciguerra R., Ambrósio R., Jr., Elsheikh A., Roberts C. J., Lopes B., Morenghi E., et al. (2016). Detection of keratoconus with a new biomechanical index. J Refract. Surg. 32, 803–810. 10.3928/1081597X-20160629-01 [DOI] [PubMed] [Google Scholar]
- Wollensak G., Spoerl E., Seiler T. (2003). Riboflavin/ultraviolet-a-induced collagen crosslinking for the treatment of keratoconus. Am. J. Ophthalmol. 135, 620–627. 10.1016/S0002-9394(02)02220-1 [DOI] [PubMed] [Google Scholar]
- Yasuno Y., Madjarova V. D., Makita S., Akiba M., Morosawa A., Chong C., et al. (2005). Three-dimensional and high-speed swept-source optical coherence tomography for in vivo investigation of human anterior eye segments. Opt. Express 13, 10652–10664. 10.1364/OPEX.13.010652 [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The datasets supporting the conclusions of this article are available from the corresponding author on reasonable request.