Skip to main content
PLOS One logoLink to PLOS One
. 2019 Dec 11;14(12):e0226263. doi: 10.1371/journal.pone.0226263

Swaying slower reduces the destabilizing effects of a compliant surface on voluntary sway dynamics

Dimitrios A Patikas 1,*, Anastasia Papavasileiou 1, Antonis Ekizos 2,3, Vassilia Hatzitaki 4, Adamantios Arampatzis 2,3
Editor: Yih-Kuen Jan5
PMCID: PMC6905565  PMID: 31826026

Abstract

The ability to control weight shifting (voluntary sway) is a crucial factor for stability during standing. Postural tracking of an oscillating visual target when standing on a compliant surface (e.g. foam) is a challenging weight shifting task that may alter the stability of the system and the muscle activation patterns needed to compensate for the perturbed state. The purpose of this study was to examine the effects of surface stability and sway frequency on the muscle activation of the lower limb, during visually guided voluntary postural sway. Seventeen volunteers performed a 2-min voluntary sway task in the anterior-posterior direction following with their projected center of pressure (CoPAP) a periodically oscillating visual target on a screen. The target oscillated at a frequency of 0.25 Hz or 0.125 Hz, while the participants swayed on solid ground (stable surface) or on a foam pad (unstable surface), resulting in four experimental conditions. The electromyogram (EMG) of 13 lower limb muscles was measured and the target–CoPAP coupling was evaluated with coherence analysis, whereas the difference in the stability of the system between the conditions was estimated by the maximum Lyapunov exponent (MLE). The results showed that slower oscillations outperformed the faster in terms of coherence and revealed greater stability. On the other hand, unstable ground resulted in an undershooting of the CoPAP to the target and greater MLE. Regarding the EMG data, a decreased triceps surae muscle activation at the low sway frequency compared to the higher was observed, whereas swaying on foam induced higher activation on the tibialis anterior as well. It is concluded that swaying voluntarily on an unstable surface results in reduced CoPAP and joint kinematics stability, that is accomplished by increasing the activation of the distal leg muscles, in order to compensate for this perturbation. The reduction of the sway frequency limits the effect of the unstable surface, on the head and upper body, improves the temporal component of coherence between CoP and target, whereas EMG activity is decreased. These findings might have implications in rehabilitation programs.

Introduction

Although bipedal stance is an essential and seemingly simple task, maintaining balance under various circumstances, as for example in the presence of external mechanical perturbations, is a rather complex issue [1,2]. The challenge for the central nervous system (CNS) is to integrate sensory input mainly from visual, proprioceptive and vestibular sources, and to create appropriate motor commands that take into account all environmental constraints and requirements. Challenging the sensorimotor system, including perturbations during different balance tasks, has been proposed as an efficient approach to understand neuromuscular control mechanisms for maintaining stability [35]. Understanding the adaptive responses of the motor system to cope with challenging balance conditions can improve our knowledge in order to develop successful exercise interventions to improve stability and to decrease risk of falls.

One of the most common strategies to introduce external perturbations is the use unstable surfaces with high viscoelasticity [6,7]. The viscoelastic properties of such surfaces reduce the effectiveness of transferring the ankle torque to the ground in order to adjust the body’s position [810], and have been frequently used as means to improve postural stability in older individuals [7,11] and patients [12,13]. Previous research supports the notion that standing on a compliant surface decreases the reliability of sensory input from the plantar mechanoreceptors [3,14,15] and changes the contribution of visual, vestibular and somatosensory information to control balance [16]. However, there is still much to be discovered on how the system reacts and adapts to this type of perturbation in terms of muscle activation. This knowledge, for example, has implications in the capacity of the CNS to control the center of pressure (CoP) accurately during voluntary tasks, such as forward/backward whole-body sway on stable or unstable surfaces.

Shifting body weight is necessary for everyday activities, such as gait initiation, getting up from a chair, and might be crucial under externally imposed sensory constraints (e.g. when avoiding an obstacle or stepping on a slippery surface). According to an observational study [17], incorrect weight shifting is the most prevalent cause of falling in older adults while reduced amplitude of voluntary sway is related to an increased risk of falling [18]. For this reason, the ability to perform fine and accurate adjustments of the CoP (voluntary body sway) has been used in the past as a rehabilitation tool for people with balance deficits [1922]. Previous studies have shown that the frequency of voluntary sway in the anterior-posterior direction affects stability [23] and the spatiotemporal variability of the oscillations [24]. More specifically, stability, as evaluated by the margins of stability [25], was greater when the sway frequency increased [23], whereas variability was higher with increasing frequency [24]. However, it is still unexplored how the system responds to external perturbations–by changing for example the compliance of the support surface–during voluntary sway at low frequencies, close to or even lower than the natural (i.e. self-selected), voluntary sway frequency during standing [26,27]. Thus, it is possible that there is an interaction of oscillation frequency with the superimposed perturbations that influences differently the neuromuscular responses of the sensorimotor system.

Considering the above, the main purpose of this study is to manipulate the surface stability (i.e. standing on firm ground or on foam) and the execution speed (i.e. two different sway frequencies) examine the effects on manipulating surface stability and execution speed during voluntary, visually guided voluntary postural sway. It is expected that swaying at a lower frequency may improve coherence between the target and the CoP when standing on unstable ground and this might be accompanied with changes in the activation of the lower limb muscles. Therefore, it is hypothesized that a lower sway frequency may compensate for the instability of the system when swaying on unstable ground and this behavior might be reflected in the EMG responses of the lower limb muscles and the ability to control more precisely the position of the CoP. The stability of the system will be quantified by the maximum Lyapunov exponent (MLE). The effects of the different sway conditions reflected on the CoP-target coupling, will be evaluated by means of coherence analysis. The aim of this study is to give useful information about how the system adapts to changes in the somatosensory input and to describe the compensatory strategies that the neuromuscular system develops when externally induced perturbations (standing on foam) are introduced. Furthermore, this study will describe the adaptation mechanisms employed when the time to process the sensory input is prolonged, i.e. during voluntary sway at a slower sway frequency. This information may be useful when applying voluntary sway in rehabilitation.

Materials and methods

Participants

The experiment was performed with the approval of the institution’s (Humboldt University) ethics committee (approval code: HU-KSBF-EK_2018_0013) in accordance with the Declaration of Helsinki. The consent was informed in written form, and the study did not include any minors. Seventeen healthy adults (10/7 males/females, mean±SD age 32.1±5.8 years, height 175±8 cm, body mass 69.8±12.9 kg) were recruited. All participants had normal or corrected to normal vision, wore no orthotic insoles and none of them had a history of neuromuscular impairments or balance-related dysfunctions. Prior to their inclusion participants were informed about the experimental protocol and gave their written informed consent.

Procedures

Prior to electrode placement, body mass and height were measured. Furthermore, foot length was calculated as the average distance between the tip of the toe and the calcaneus of both feet. Bipolar surface electrodes (sensor area 15 mm2, wet gel Ag/AgCl, N-00-S, Ambu A/S, Denmark) were placed over 13 superficial lower limb muscles of the right side. The muscles measured were the gluteus medius, gluteus maximus, tensor fasciae late, rectus femoris, vastus medialis, vastus lateralis (VL), semitendinosus, biceps femoris, tibialis anterior (TA), peroneus longus, medial gastrocnemius (MG), lateral gastrocnemius (LG) and soleus muscle (SO). The inter-electrode distance was set at 2 cm and the electrodes were positioned according to the recommendations of the SENIAM project [28]. The skin was carefully prepared (shaving, abrasion with sandpaper and cleaning with alcohol solution) to assure good electrode-skin contact. Manual tests (brief muscle contractions) with real-time visual inspection of the raw EMG were performed to verify the electrode placement and to ensure minimum amount of crosstalk.

During the assessment the participants stood on a force platform (60×90 cm, Kistler, Winterthur, Switzerland). The following experimental paradigm was used to create a visually guided weight-shifting task for voluntary controlling the CoP in the anterior-posterior direction. A monitor (47-inch diagonal) was placed at eye level, 1.5 m in front of them and displayed 2 dots (yellow and red) with black background (Fig 1A). The yellow dot (feedback) showed in real-time the position of CoP at the anterior-posterior direction (CoPAP). The red dot (target) represented the position of CoPAP that the participants were instructed to follow with the yellow dot. The movement of both dots was restricted to the vertical direction, in the middle of the screen. Upward or downward movement of the dots signified shifting of the CoP to the anterior or posterior direction, respectively.

Fig 1. Schematic illustration of the experimental setup.

Fig 1

A: Representation of a participant performing voluntary forward (blue) and backward (green) sway and watching on the monitor the target (red dot) that is moving up and down, and the feedback of his/her CoPAP (yellow dot), with the objective to match the two dots. B: Transverse view of foot placement on the force platform and the position of the CoP (yellow dot). Red shaded area indicates the oscillating range of the target (60% of foot length). Areas shaded in grey designates the position of the foam pads for the trials on unstable ground.

Postural tasks

The participants stood on both feet with their hands at their waist (akimbo position). The distance between the two medial malleoli was set at 10% of the body height. The postural tasks consisted of visually guided body sway in the anterior-posterior direction. The target (red dot) was a time-series sinusoidal signal, generated by a sine wave with a fixed time period for each condition (either 4 or 2 s, to create an oscillation at 0.25 or 0.125 Hz, respectively), with a sample frequency of 50 Hz (i.e. one data point every 20 ms, which resulted in 8000 and 4000 samples per cycle for the 0.25 and 0.125 Hz conditions, respectively). The target (red dot) moved vertically in a sinusoidal pattern and the participant was instructed to match the red dot with the yellow one, by swaying his/her torso forward or backward, without flexing the hips or knees. The amplitude of the target movement (red dot) was set at 60% of the foot length, with zero representing the midpoint of the CoPAP range when the participant was leaning as much as possible anteriorly and posteriorly without moving the feet from the ground (Fig 1B).

Four postural tasks varying in sway frequency and ground stability were performed in random order. Two sway frequencies were selected; one natural [26,27] with a period of 4 s (0.25 Hz) and one lower, with a longer of 8 s (0.125 Hz). For each frequency, voluntary sway was recorded on rigid ground and on foam surface. A familiarization session before the measurement was performed and prior to the measurement, room lights were dimmed. Each trial lasted 2 minutes with 2–3 minutes interval in-between. The session, including the subject preparation (15 minutes), lasted no longer than one hour, considering the set-up for ultrasound recordings and two additional conditions (data not presented here).

Data acquisition

The target signal was created and displayed on the monitor with custom made MATLAB scripts (version 2014b, Math Works Inc, USA), while an interface was created for triggering and synchronizing all devices with a single pulse. The force platform signal was digitized with a 14-bit resolution A/D card (NI USB-6009, National Instruments, USA) at 1000 Hz sampling rate and the anterior-posterior component was normalized to the foot length and was returned as input to the monitor for the vertical position of the yellow dot, with the full height of the screen representing 100% of foot length. The refresh rate of the dots was set at 50 fps.

The EMG signals were captured with a wireless EMG system (myon m320, myon AG, Schwarzenberg, Switzerland). The signal was pre-amplified (gain: 500, input impedance: 2 MΩ, bandwidth: 5–500 Hz) and transmitted at 12-bit resolution with 1000 Hz sampling frequency. All digitized signals were stored for further processing.

Data processing

CoPAP signal was filtered with a 4th order Butterworth low-pass filter with cutoff frequency at 25 Hz, and the first cycle was omitted from the analysis. The CoPAP-target coupling was evaluated using the spectral coherence analysis which represents the amount of correlation between the two signals on the frequency domain from 0 to 1 Hz. Both target and CoPAP signals were interpolated at a sampling frequency of 64 Hz which responded to a frequency resolution of 0.0625 Hz when assessing the fast-Fourier transform. Three variables were analyzed at the specific frequencies that the task was executed (i.e. either at 0.125 Hz or 0.25 Hz for the slow or natural sway, respectively): the spectral coherence as a measure of the correlation between the two signals (target and CoPAP) in the frequency domain, spectral phase as a temporal measure of the phase lag between the signals, and the spectral gain which reveals spatial information about the amplitude of the two signals (values over 1 designate CoPAP values above and below the target peaks and valleys, respectively). The spectral phase was expressed in % of the sway cycle, with 0 being interpreted as the absolute synchronization between the signals and negative values as a delayed CoPAP relative to the target signal.

Local dynamic stability represents the ability of a system to maintain its movement pattern despite intrinsic and extrinsic perturbations [2931]. The local dynamic stability of the system in the current study was assessed through the maximum finite-time Lyapunov exponent (MLE), which quantifies the rate of divergence of nearby trajectories in the reconstructed state space [32,33]. Our analysis followed the procedure as described in a previous study [34].

In the present study we calculated the MLE in the CoP and coordinate data of the markers placed in different parts of the body. For the CoP, data acquisition was performed at 2,000 Hz and MLE has been calculated on the norm of the anterio-posterior and mediolateral axes. The original time-series have been filtered using a 4th order Butterworth low-pass filter with a cut-off frequency of 20 Hz and were consequently down-sampled to 20,000 data points. The 3-dimensional coordinate data of the ankles (lateral malleoli), knees (lateral epicondyles), hips (greater trochanters), spine (7th cervical vertebra) and head (4 markers around the head placed on a headband) were acquired at 250 Hz and have been filtered in the same manner. All kinematic data were subsequently down-sampled to 15,000, and we calculated the midpoint between the two sides of the body and the midpoint of the 4 head markers. The 3-dimensional coordinates of these virtual markers in addition to the marker at spine were used for the MLE calculation. The norm of all axes has been used. Due to the standardized constant movement of the target signal and overall trial time (i.e. 120 seconds) the number of cycles (i.e. 30 cycles for the natural frequency and 15 for the slow frequency) was the same for all participants and no interpolation of the time-series was needed. To reconstruct the state space from the one dimensional time series, we used delay-coordinate embedding [35] as follows:

S(t)=[z(t),z(t+τ),,z(t+(m1)τ)], (1)

with S(t) being the m-dimensional reconstructed state vector, z(t) the input 1D coordinate series, τ the time delay and m the embedding dimension. Time delays were selected based on the first minimum of the Average Mutual Information function [36]. For these data m = 3 was sufficient to perform the reconstruction, similarly to previous studies examining human movement [29,34]. Individually selected time delays were chosen by averaging the outcome delays deriving from both trials performed by the participants [29,37]. Values of τ were slightly lower in the slow condition (CoPAP: 0.11–0.19, coordinates: 0.14–0.33 of one cycle) compared to the natural condition (CoPAP: 0.20–0.25, coordinates: 0.17–0.35 of one cycle).

Further, the average divergence of each point’s trajectory to its closest neighbor was calculated, using the Rosenstein algorithm [38]. The resulting MLE was calculated based on the delay of each participant. That ensured the standardization of the calculation for the MLE across individuals, due to the first pick in the resulting divergence curves corresponding to 0.5 delay. As such, the final MLE value was calculated as the slope of the average divergence curves’ linear fit corresponding to 0.5 of the individuals’ delay value (i.e. the most linear part of the curve).

The EMG signals were filtered (Butterworth 4th order bandpass filter from 10 to 450 Hz), fully rectified and smoothed using a low-pass filter (Butterworth 4th order low pass filter at 5 Hz). To remove the baseline activation of each muscle the minimum EMG was subtracted from the filtered signal and all values were normalized to the mean of the trial. The start and end of each sway cycle was identified when the CoPAP moved from negative (posterior) to positive (anterior) values and crossed the zero line (Fig 2). Each cycle was interpolated to 200 data points and the average of all cycles was calculated, after excluding the first sway cycle.

Fig 2. Data from typical subject.

Fig 2

Target (thick gray line) and CoPAP (thin black line) are shown during the four conditions of the voluntary postural sway on rigid ground or on foam and with slow or natural sway frequency. Target-CoPAP coherence, phase lag (% of sway cycle) and gain for each condition are displayed on the middle top of the graphs, and values on vertical axis are presented as percentage of foot length. The vertical dotted lines represent the start and end of each sway cycle. Grey shaded area designate the first cycle that has been excluded from the analysis.

Statistical analysis

Data are reported as mean±SD. For the spectral coherence analysis (coherence, phase lag and gain) and MLE, a two-way ANOVA design for repeated measures was assessed to detect the effects of ground surface and sway frequency. The Scheffé post-hoc test was performed when the level of significance was reached. To compare the effects of the different conditions on the CoPAP and EMG curves, paired t-test comparisons were assessed for all combinations using one-dimensional statistical parametric mapping [39]. For this purpose, the open-access SPM1D code for MATLAB was used (www.spm1d.org, v. 0.4). The level of significance α was set at 0.05 for all analyses.

Results

A typical example of the CoPAP with the target signal in all four conditions is shown in Fig 2. The MLE (Table 1) of all variables showed a significant main effect for both factors (p<0.01 for surface and frequency). The surface by frequency interaction was not significant for the CoP and the ankle joint (p>0.05). The significant (p<0.05) surface by frequency interaction in the rest of the variables and the respective post-hoc test, revealed increased MLE when standing on foam only for the natural sway frequency (p<0.01), whereas no statistically significant change was observed at the slower sway frequency (p>0.05).

Table 1. MLE values and statistical analysis for COP and kinematics.

Rigid
ground
Main effect for surface
Main effect for frequency
Interaction surface×frequency
Foam F-values p-values
CoP Natural (0.25 Hz) 13.7±1.7 14.4±1.4 F1,16 = 11.8
F1,16 = 56.2
F1,16 = 0.03
p = 0.003
p<0.001
p = 0.860
Slow (0.125 Hz) 10.0±1.4 10.8±1.9
Ankle Natural (0.25 Hz) 9.4±1.2 10.0±1.4 F1,16 = 24.7
F1,16 = 45.6
F1,16 = 1.8
p<0.001
p<0.001
p = 0.199
Slow (0.125 Hz) 6.7±1.3 7.7±1.4
Knee Natural (0.25 Hz) 9.3±1.1 11.2±1.1 *** F1,16 = 52.4
F1,16 = 200.1
F1,16 = 8.8
p<0.001
p<0.001
p = 0.009
Slow (0.125 Hz) 6.0±0.9 6.8±1.1 n.s.
Hip Natural (0.25 Hz) 9.6±1.0 10.7±1.1 ** F1,16 = 13.9
F1,16 = 243.4
F1,16 = 4.7
p = 0.002
p<0.001
p = 0.046
Slow (0.125 Hz) 6.4±0.9 6.9±1.1 n.s.
Spine Natural (0.25 Hz) 9.8±1.6 11.1±1.4 *** F1,16 = 16.2
F1,16 = 105.7
F1,16 = 12.8
p<0.001
p<0.001
p = 0.002
Slow (0.125 Hz) 6.3±1.0 6.7±1.0 n.s.
Head Natural (0.25 Hz) 9.3±2.3 10.9±2.3 ** F1,16 = 14.6
F1,16 = 57.0
F1,16 = 6.6
p = 0.002
p<0.001
p = 0.021
Slow (0.125 Hz) 6.2±0.1 6.5±1.6 n.s.

Mean±standard deviation of the MLE for COP and norm coordinates at ankle, knee, hip, spine and head. Results of the 2-way ANOVA for each variable (F- and p-values for the main effects and interaction) are shown in the last two columns. Asterisks demonstrate significant difference for the post-hoc Scheffé test (n.s.: non-significant difference, p>0.05

**: p<0.01

***: p<0.001) between the rigid ground and foam for each frequency, when the interaction reached the level of significance.

Coherence of the target and CoPAP signals (Fig 3A) was significantly higher in the slow sway compared to the natural (p = 0.001). The unstable surface did not affect coherence (p = 0.109) and the interaction between sway frequency and surface was not significant (p = 0.858). Regarding the phase lag between the two signals (Fig 3B) the slower sway resulted in values closer to zero, whereas the natural sway frequency revealed significantly lower values, which shows a delayed response of the CoPAP relative to the target motion (p<0.001). Similar to the coherence, the phase lag did not show any significant differences (p = 0.126) between stable and unstable condition and the interaction between the two factors was also not significant (p = 0.265). The gain (Fig 3C) was significantly lower (under-shooting of CoPAP relative to the target) in the slow compared to the natural sway (p<0.001) and when on the foam compared to the rigid ground (p<0.001). However, the interaction between the factors frequency and surface was not statistically significant (p = 0.065).

Fig 3. Results of the coherence analysis.

Fig 3

Group results of the CoPAP-target coherence (A), phase lag (B) and gain (C) for the slow and natural voluntary sway on stable (rigid) and unstable (foam) ground. Pink circles connected with continuous lines and blue triangles connected with dashed lines represent the condition of sway on rigid ground and foam, respectively. Small and large symbols (circles and triangles) represent data of each individual and group means, respectively. Vertical lines designate one standard deviation and asterisks express the presence of significant main effect (p<0.05).

As shown in Fig 4, when comparing the slow vs. the natural sway condition on rigid ground or the rigid ground vs. foam during natural sway, the difference between the target signals did not differ significantly (p>0.05). On the contrary, during the 0–13, 30–62, and 81.5–100% of the sway cycle, the target signals were significantly different between slow and natural sway on foam, due to a shift to the right for the natural relative to the slow sway condition. Similarly, during the 0–17.5 and 44.5–66.55 of the sway cycle, the target signals were significantly different between slow sway on rigid ground and foam, due to a shift to the left for the foam relative to the rigid ground surface.

Fig 4. Mean values of all participants for the target signal and CoPAP during the voluntary sway.

Fig 4

Each column of graphs depicts one comparison between conditions. Dashed lines represent one standard deviation and black stripes at the horizontal axes illustrate the time of the sway cycle that there was a significant difference (p<0.05) between the two conditions. Horizontal axis is normalized to the duration of each sway cycle of CoPAP and values for the vertical axis are expressed as percent of the foot length.

Regarding the EMG recordings, a visual representation of the EMG amplitude for the 13 examined muscles of a typical subject during the voluntary sway in all conditions, is shown in Fig 5. On average, the results of EMG showed that gluteus maximus and medius, as well as tensor fascia latae had no observable phasic activation during any of the tasks (Fig 6). The rest of the muscles demonstrated phasic behavior, with activation when the CoPAP was at the extreme anterior or posterior phase of the sway cycle (i.e. transition phase from on direction to the other). Most remarkable differences between the conditions were observed at the distal muscles. According to the SPM analysis, slow compared to natural sway on rigid ground showed lower EMG activity for the MG and SO during 96–98.5% and 93–97.5% of the sway cycle, respectively. These differences were present for longer portion of the sway cycle (MG: 0–1, and 95.5–99.5%; SO: 0–1, 3.5–4.5, 9–12, and 94.5–100%) and for more muscles when this comparison was made for the foam condition (LG: 0–3.5, 6–8, 8.5–13, and 94.5–100%; TA: 58–63.5, 64–68.5, and 71–76.5%; VL: 18–27.5, and 50.5–53.5%). Specifically, during slow voluntary sway, the LG (at 69–70, 75.5–82, 83–84, and 85.5–87%) and VL (at 89–91%) revealed significantly higher activation when standing on foam compared to standing on the rigid surface. For the natural sway frequency, the higher values on foam compared to the rigid surface were detected on the TA (at 71–87%), LG (at 0–4, 71.5–76.5, 78.5–88, and 95.5–100%) and SO (at 0–2.5, 8–24, 25–29.5, 60–63, 66–71, 72–86, and 98–100%) muscles.

Fig 5. Typical example of EMG signals at all four conditions of the 120-s voluntary sway.

Fig 5

Each of the four graphs represents one condition as described on the left side of the graph. The color of each cell represents the mean EMG amplitude for every 1/6 of the sway cycle. All muscles are shown at each row with the following abbreviations: gluteus medius (GMed), gluteus maximus (GMax), tensor fasciae late (TFL), rectus femoris (RF), vastus medialis (VM), vastus lateralis (VL), semitendinosus (ST), biceps femoris (BF), tibialis anterior (TA), peroneus longus (PL), medial gastrocnemius (MG), lateral gastrocnemius (LG) and soleus muscle (SO)].

Fig 6. Pairwise comparisons between the four voluntary sway conditions for the mean EMG amplitude.

Fig 6

See legend of Fig 5 for abbreviations of muscle names. Dashed lines represent one standard deviation and black stripes at the horizontal axes illustrate the time of the sway cycle that there was a significant difference (p<0.05) between the two conditions. Values for EMG are expressed in % of the mean EMG during the sway cycle and for time as % of the sway cycle.

Discussion

Voluntary sway on foam increases the instability of the system compared to rigid ground, making the system more unstable in both investigated frequencies. The increased instability resulted in reduced CoP-target gain (lower CoPAP values than the target) and in higher muscle activation of the distal muscles. On the other hand, decreasing the frequency of the voluntary sway resulted in lower muscle activation, better coupling of the CoPAP to the target, with stability differences (i.e. increased MLE) limited at the ankle and CoP level.

According to the current findings regarding the muscle activation patterns, it could be argued that during the voluntary visually guided sway the proximal muscles (i.e. muscles associated with the hip) demonstrate minimal or no phasic activity. This is supported by previous research arguing that the ankle strategy (implying no significant movement around the hip joint) can be retained at sway frequencies lower than 0.5 Hz [40]. Although it could not be excluded that the proximal muscles might be active even without movement on the hip to stabilize the trunk, our data give no evidence for such activation. On the other hand, the thigh and shank muscles were active during the first or second half of the cycle. Interestingly, ST, MG, LG and SO became active as the CoPAP moved forward (shortly before CoPAP crossed the midpoint), with the acting forces serving to decelerate the body’s inertia and to initiate the backward sway (moving to a more plantar flexed position), as soon as the maximum anterior position (dorsiflexion) is reached. Likewise, the RF, VM, VL and TA act in the same manner, but in the opposite direction during the second half of the sway cycle. Based on the EMG data of the current study, it is evident that the CNS activates the muscles primarily as dampening elements to control the movement and secondary as active components to produce force in the direction of movement. This dampening function is in agreement with previous findings during slow-frequency voluntary sway and has been attributed to the limited capacity of the passive stiffness components to stabilize the body [41]. Furthermore, it could be argued that the system seems to function with two basic muscle groups (i.e. muscles that act synergistically comprising a muscle synergy), and two motor primitives (i.e. activation patterns). Similar muscle synergies that act reciprocally have been previously reported, even during faster (1 Hz) voluntary body sway [42] or during voluntary sway towards one direction (forward or backward) [42,43]. This distinction of two muscle synergies, reduces the dimensionality and thus complexity, making the system easier to control, organize and manipulate [44]. However, altering the environmental constraints, may increase the task complexity that could introduce new muscle synergies to the system [45,46].

The experimental paradigm used in the present study was selected to challenge the stability of the human system and to introduce external perturbations during the voluntary sway task, by changing the ground compliance. Indeed, MLE increased when standing on foam revealing higher instability, with less stable trajectories in time for the CoP and the ankle, knee, hip, spine and head kinematics. This could be attributed to the premise that standing on a compliant surface impedes the direct force transfer to the ground and reduces the quality of sensorimotor information to the CNS [47,48]. However, the stability of the upper body (head and spine) as well as that of the hip and knee joints was not affected by the unstable surface at the low sway frequency. Earlier studies have supported the notion that the movement of the head and trunk during perturbations is controlled by vestibular, visual and proprioceptive inputs, independent of the ascending information from peripheral body segments [49]. This shows that the quasi-unpredictable situation on the ground (standing on foam), that tends to destabilize the body, has a reduced effect in the head and trunk in terms of stability, when the movement is executed slowly. This might have implications in rehabilitation programs or experimental setups that require more stable head movement during a task, in order to reduce the dependency on the visual and vestibular system.

The consequences of the increased instability of sway on the target–CoP coupling was the reduced gain shown on foam surface compared to the rigid ground condition, revealing that the participants were not capable to reach the target limits (peaks and valleys). The reduction in gain could be interpreted as a reduction in the limits of stability when standing on foam and could be attributed to the limited capacity to transfer forces to the ground [810] and the lower quality of somatosensory input [3,14,15]. Furthermore, there are indications of increased co-activation of the antagonist muscles when standing on foam and the CoP is at a posterior position (plantar flexion). This change is in agreement with previous studies that observed increased level of co-activation with increasing difficulty of postural tasks [50], which may act as compensatory mechanism, by increasing joint stiffness and thereby stability [51]. On the other hand, it counteracts towards the direction of the movement and thereby limits its range of motion.

An alternative explanation of the stronger temporal coupling during slow compared to natural sway and the stronger spatial coupling when standing on rigid ground compared to the foam surface could stem from the perceptual-motor reality of postural time-to-contact [52,53]. According to this concept, the temporal proximity to the margin of stability is reduced intuitively when a) sway velocity increases b) the base of support decreases and c) the degrees of freedom are reduced [54]. Regarding the findings of the present study, it could be argued that one reason for the weaker temporal coupling (greater phase lag) in the higher than the lower sway frequency might be the greater movement velocity that induced a reduction in the margins of stability. This is also reflected by the fact that MLE was increased only distally (CoP and ankle) when the surface of support became unstable at slower sway frequencies, whereas the consequences of the foam surface on the stability of the system emerged up to the head when swaying at higher frequencies.

The improved synchronization observed in the present study (lower phase lag between CoPAP and target) during slow oscillations, was characterized by reduced activation of the calf muscles. Although previous studies have shown that increasing the frequency of sway may involve the activation of more proximal muscles by moving the hip more actively [40], there is no evidence for such recruitment at least for the sway frequencies that have been tested. The improved temporal coupling is in agreement with studies performing voluntary periodic sway as fast as possible at a certain target range [55,56] and is in accordance with Fitts’ law, which addresses a trade-off between accuracy and movement speed, i.e. slower movements are more accurate [57]. Another explanation for the improved temporal coupling during the slow sway is the greater stability observed on the head, which may introduce less bias from visual and vestibular sources. Furthermore, it has been suggested that when movement is slow, proprioceptive input might have a more prominent role to control movement and achieve stability [58], in contrast to faster movements, when the intrinsic mechanical properties of the system are of greater importance than the peripheral feedback [59].

In general, although the voluntary sway task was more challenging on the foam, especially during the higher frequency, no systematic EMG increase was shown on the proximal muscles. On the other hand, changes in EMG have been observed during the second half of the cycle, when the RF, VM, VL and TA decelerate the body from moving backwards and initiate the forward motion. Therefore, it seems that there is a modular organization during voluntary sway that is retained when the system becomes more unstable in the perturbed condition. Furthermore, previous studies reported that distal muscles are more responsive to perturbations than proximal ones [58], possibly due to morphological and anatomical differences (i.e. large pennation angle, short fascicle and longer tendons) that reinforce sensitivity at low force levels [60]. The observed increased motor drive is in agreement with previous studies which support that unstable surfaces result in increased contraction speeds and higher motor output [61,62]. This supports the notion that muscle synergies are not simply defined as groups of muscles that act together, but as variables controlled by the CNS to co-vary depending on the task, in order to stabilize the body [43,63]. With the current experimental setup, we were not able to confirm any widening in the EMG activation time periods when changing the stability of the system. Recently it has been shown that when increasing unsteadiness during dynamic conditions (running or walking), muscle activation changes and creates a more “robust” motor output which results in developed strategies capable to cope with errors when required [59]. This discrepancy could be attributed to the fact that the task of voluntary sway, is less dynamic than walking or running, since the base of support is fixed and the range of motion of the involved joints during the movement is smaller. Therefore, under this condition, when fewer joints are involved with reduced degrees of freedom, the system may cope for stability with different strategies than the ones used during more dynamic movements such as walking or running. However, we found no indication that manipulation of the movement properties (i.e. changes in stability and sway frequency) created new synergies. It rather seems that the system modified the activation onset and amplitude of the already active muscles.

Conclusions

In conclusion, standing on foam results in a more unstable CoP trajectory and body movement, and this is reflected to a higher muscle activation especially in the distal muscles. Slower execution of the voluntary sway limits the effect of the unstable surface, on the head and upper body, reduces the phase-lag between CoP and target, and exhibits a reduced EMG activity. However, it remains to be examined whether this behavior changes during aging or whether is different in patient populations. These findings can have implications in rehabilitation programs, depending on the goal of intervention and on which parameters (stability, accuracy, muscle activation) is necessary to be changed.

Acknowledgments

The authors would to acknowledge the contribution of Dr. Arno Schroll for the creation of the interface for data collection and the hardware setup, Haris Sotirakis for contributing to the data analysis, as well as Dr. Sebastian Bohm, Dr. Alessandro Santuz and Victor Hugo Muñoz for their assistance in the recruitment of the participants.

Data Availability

All relevant data are within the paper and its Supporting Information files.

Funding Statement

This work was supported by the German Academic Exchange Service (DAAD) “GGP-Age”, 57339989 (https://www.daad.de/en/). The funder played no role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript.

References

  • 1.Taube W, Schubert M, Gruber M, Beck SC, Faist M, Gollhofer A. Direct corticospinal pathways contribute to neuromuscular control of perturbed stance. J Appl Physiol 2006;101:420–9. 10.1152/japplphysiol.01447.2005 [DOI] [PubMed] [Google Scholar]
  • 2.Ozdemir RA, Contreras-Vidal JL, Paloski WH. Cortical control of upright stance in elderly. Mech Ageing Dev 2018;169:19–31. 10.1016/j.mad.2017.12.004 [DOI] [PubMed] [Google Scholar]
  • 3.Perry SD, McIlroy WE, Maki BE. The role of plantar cutaneous mechanoreceptor in the control of compensatory stepping reactions evoked by unpredictable multi-directional perturbation. Brain Res 2000;877:401–6. 10.1016/s0006-8993(00)02712-8 [DOI] [PubMed] [Google Scholar]
  • 4.Creath R, Kiemel T, Horak FB, Peterka RJ, Jeka J. A unified view of quiet and perturbed stance: simultaneous co-existing excitable modes. Neurosci Lett 2005;377:75–80. 10.1016/j.neulet.2004.11.071 [DOI] [PubMed] [Google Scholar]
  • 5.Bardy BG, Oullier O, Lagarde J, Stoffregen TA. On perturbation and pattern coexistence in postural coordination dynamics. J Mot Behav 2007;39:326–34. 10.3200/JMBR.39.4.326-336 [DOI] [PubMed] [Google Scholar]
  • 6.Anderson K, Behm DG. Trunk muscle activity increases with unstable squat movements. Can J Appl Physiol 2005;30:33–45. 10.1139/h05-103 [DOI] [PubMed] [Google Scholar]
  • 7.Hamed A, Bohm S, Mersmann F, Arampatzis A. Exercises of dynamic stability under unstable conditions increase muscle strength and balance ability in the elderly. Scand J Med Sci Sports 2018;28:961–71. 10.1111/sms.13019 [DOI] [PubMed] [Google Scholar]
  • 8.Horak FB, Hlavacka F. Somatosensory loss increases vestibulospinal sensitivity. J Neurophysiol 2001;86:575–85. 10.1152/jn.2001.86.2.575 [DOI] [PubMed] [Google Scholar]
  • 9.MacLellan MJ, Patla AE. Adaptations of walking pattern on a compliant surface to regulate dynamic stability. Exp Brain Res 2006;173:521–30. 10.1007/s00221-006-0399-5 [DOI] [PubMed] [Google Scholar]
  • 10.Patel M, Fransson PA, Lush D, Gomez S. The effect of foam surface properties on postural stability assessment while standing. Gait Posture 2008;28:649–56. 10.1016/j.gaitpost.2008.04.018 [DOI] [PubMed] [Google Scholar]
  • 11.Hirase T, Inokuchi S, Matsusaka N, Okita M. Effects of a Balance Training Program Using a Foam Rubber Pad in Community-Based Older Adults. J Geriatr Phys Ther 2015;38:62–70. 10.1519/JPT.0000000000000023 [DOI] [PubMed] [Google Scholar]
  • 12.Page P. Sensorimotor training: A “global” approach for balance training. J Bodyw Mov Ther 2006;10:77–84. 10.1016/J.JBMT.2005.04.006 [DOI] [Google Scholar]
  • 13.Hung J-W, Chou C-X, Hsieh Y-W, Wu W-C, Yu M-Y, Chen P-C, et al. Randomized Comparison Trial of Balance Training by Using Exergaming and Conventional Weight-Shift Therapy in Patients With Chronic Stroke. Arch Phys Med Rehabil 2014;95:1629–37. 10.1016/j.apmr.2014.04.029 [DOI] [PubMed] [Google Scholar]
  • 14.Wu G, Chiang JH. The significance of somatosensory stimulations to the human foot in the control of postural reflexes. Exp Brain Res 1997;114:163–9. 10.1007/pl00005616 [DOI] [PubMed] [Google Scholar]
  • 15.Patel M, Fransson PA, Johansson R, Magnusson M. Foam posturography: Standing on foam is not equivalent to standing with decreased rapidly adapting mechanoreceptive sensation. Exp Brain Res 2011;208:519–27. 10.1007/s00221-010-2498-6 [DOI] [PubMed] [Google Scholar]
  • 16.Furman JM. Role of posturography in the management of vestibular patients. Otolaryngol Neck Surg 1995;112:8–15. 10.1016/S0194-59989570300-4 [DOI] [PubMed] [Google Scholar]
  • 17.Robinovitch SN, Feldman F, Yang Y, Schonnop R, Lueng PM, Sarraf T, et al. Video capture of the circumstances of falls in elderly people residing in long-term care: an observational study. Lancet 2013;381:778–82. 10.1016/S0140-6736(12)61263-X.Video [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Tucker MG, Kavanagh JJ, Morrison S, Barrett RS. What are the relations between voluntary postural sway measures and falls-history status in community-dwelling older adults? Arch Phys Med Rehabil 2010;91:750–8. 10.1016/j.apmr.2010.01.004 [DOI] [PubMed] [Google Scholar]
  • 19.Shumway-Cook A, Anson D, Haller S. Postural sway biofeedback: its effect on reestablishing stance stability in hemiplegic patients. Arch Phys Med Rehabil 1988;69:395–400. [PubMed] [Google Scholar]
  • 20.Hamman RG, Mekjavic I, Mallinson AI, Longridge NS. Training effects during repeated therapy sessions of balance training using visual feedback. Arch Phys Med Rehabil 1992;73:738–44. [PubMed] [Google Scholar]
  • 21.Hirvonen TP, Aalto H, Pyykkö I. Stability limits for visual feedback posturography in vestibular rehabilitation. Acta Otolaryngol Suppl 1997;529:104–7. 10.3109/00016489709124096 [DOI] [PubMed] [Google Scholar]
  • 22.Dault MC, de Haart M, Geurts ACHH, Arts IMPP, Nienhuis B. Effects of visual center of pressure feedback on postural control in young and elderly healthy adults and in stroke patients. Hum Mov Sci 2003;22:221–36. 10.1016/s0167-9457(03)00034-4 [DOI] [PubMed] [Google Scholar]
  • 23.Lorenzo TM, Vanrenterghem J. Effects of increased anterior-posterior voluntary sway frequency on mechanical and perceived postural stability. Hum Mov Sci 2015;39:189–99. 10.1016/j.humov.2014.11.012 [DOI] [PubMed] [Google Scholar]
  • 24.Danion F, Duarte M, Grosjean M. Variability of reciprocal aiming movements during standing: The effect of amplitude and frequency. Gait Posture 2006;23:173–9. 10.1016/j.gaitpost.2005.01.005 [DOI] [PubMed] [Google Scholar]
  • 25.Hof AL, Gazendam MGJ, Sinke WE. The condition for dynamic stability. J Biomech 2005;38:1–8. 10.1016/j.jbiomech.2004.03.025 [DOI] [PubMed] [Google Scholar]
  • 26.Sofianidis G, Hatzitaki V, Grouios G, Johannsen L, Wing A. Somatosensory driven interpersonal synchrony during rhythmic sway. Hum Mov Sci 2012;31:553–66. 10.1016/j.humov.2011.07.007 [DOI] [PubMed] [Google Scholar]
  • 27.Coste A, Salesse RN, Gueugnon M, Marin L, Bardy BG. Standing or swaying to the beat: Discrete auditory rhythms entrain stance and promote postural coordination stability. Gait Posture 2018;59:28–34. 10.1016/j.gaitpost.2017.09.023 [DOI] [PubMed] [Google Scholar]
  • 28.Hermens HJ, Freriks B, Merletti R, Stegeman DF, Blok JH, Raw G, et al. European recommendations for surface electromyography. Enshede, The Netherlands: Roessingh Research and Development; 1999. [Google Scholar]
  • 29.Ekizos A, Santuz A, Schroll A, Arampatzis A. The maximum Lyapunov exponent during walking and running: reliability assessment of different marker-sets. Front Physiol 2018;9:1101 10.3389/fphys.2018.01101 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Ihlen EAF, van Schooten KS, Bruijn SM, Pijnappels M, van Dieën JH. Fractional stability of trunk acceleration dynamics of daily-life walking: toward a unified concept of gait stability. Front Physiol 2017;8:516 10.3389/fphys.2017.00516 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Lyapunov AM. The general problem of the stability of motion. Int J Control 1992;55:531–4. 10.1080/00207179208934253 [DOI] [Google Scholar]
  • 32.Dingwell JB, Cusumano JP, Cavanagh PR, Sternad D. Local dynamic stability versus kinematic variability of continuous overground and treadmill walking. J Biomech Eng 2001;123:27–32. 10.1115/1.1336798 [DOI] [PubMed] [Google Scholar]
  • 33.England SA, Granata KP. The influence of gait speed on local dynamic stability of walking. Gait Posture 2007;25:172–8. 10.1016/j.gaitpost.2006.03.003 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Ekizos A, Santuz A, Arampatzis A. Transition from shod to barefoot alters dynamic stability during running. Gait Posture 2017;56:31–6. 10.1016/j.gaitpost.2017.04.035 [DOI] [PubMed] [Google Scholar]
  • 35.Packard NH, Crutchfield JP, Farmer JD, Shaw RS. Geometry from a time series. Phys Rev Lett 1980;45:712–6. 10.1103/PhysRevLett.45.712 [DOI] [Google Scholar]
  • 36.Fraser AM, Swinney HL. Independent coordinates for strange attractors from mutual information. Phys Rev A 1986;33:1134–40. 10.1103/PhysRevA.33.1134 [DOI] [PubMed] [Google Scholar]
  • 37.Ekizos A, Santuz A, Arampatzis A. Short- and long-term effects of altered point of ground reaction force application on human running energetics. J Exp Biol 2018;221:jeb176719 10.1242/jeb.176719 [DOI] [PubMed] [Google Scholar]
  • 38.Rosenstein MT, Collins JJ, De Luca CJ. Reconstruction expansion as a geometry-based framework for choosing proper delay times. Phys D Nonlinear Phenom 1994;73:82–98. 10.1016/0167-2789(94)90226-7 [DOI] [Google Scholar]
  • 39.Pataky TC, Vanrenterghem J, Robinson MA. Zero- vs. one-dimensional, parametric vs. non-parametric, and confidence interval vs. hypothesis testing procedures in one-dimensional biomechanical trajectory analysis. J Biomech 2015;48:1277–85. 10.1016/j.jbiomech.2015.02.051 [DOI] [PubMed] [Google Scholar]
  • 40.McCollum G, Leen TK. Form and exploration of mechanical stability limits in erect stance. J Mot Behav 1989;21:225–44. 10.1080/00222895.1989.10735479 [DOI] [PubMed] [Google Scholar]
  • 41.Loram ID, Maganaris CN, Lakie MD. Paradoxical muscle movement during postural control. Med Sci Sports Exerc 2009;41:198–204. 10.1249/MSS.0b013e318183c0ed [DOI] [PubMed] [Google Scholar]
  • 42.Wang Y, Asaka T, Zatsiorsky VM, Latash ML. Muscle synergies during voluntary body sway: Combining across-trials and within-a-trial analyses. Exp Brain Res 2006;174:679–93. 10.1007/s00221-006-0513-8 [DOI] [PubMed] [Google Scholar]
  • 43.Krishnamoorthy V, Latash ML, Scholz JP, Zatsiorsky VM. Muscle synergies during shifts of the center of pressure by standing persons. Exp Brain Res 2003;152:281–92. 10.1007/s00221-003-1574-6 [DOI] [PubMed] [Google Scholar]
  • 44.Nashner LM, McCollum G. The organization of human postural movements: A formal basis and experimental synthesis. Behav Brain Sci 1985;8:135 10.1017/S0140525X00020008 [DOI] [Google Scholar]
  • 45.Torres-Oviedo G, Ting LH. Muscle synergies characterizing human postural responses. J Neurophysiol 2007;98:2144–56. 10.1152/jn.01360.2006 [DOI] [PubMed] [Google Scholar]
  • 46.Chvatal SA, Ting LH. Common muscle synergies for balance and walking. Front Comput Neurosci 2013;7:1–14. 10.3389/fncom.2013.00001 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47.Shumway-Cook A, Horak FB. Assessing the influence of sensory interaction on balance. Phys Ther 1986;66:1545–50. 10.1093/ptj/66.10.1548 [DOI] [PubMed] [Google Scholar]
  • 48.Schut IM, Engelhart D, Pasma JH, Aarts RGKM, Schouten AC. Compliant support surfaces affect sensory reweighting during balance control. Gait Posture 2017;53:241–7. 10.1016/j.gaitpost.2017.02.004 [DOI] [PubMed] [Google Scholar]
  • 49.Keshner EA, Woollacott MH, Debu B. Neck, trunk and limb muscle responses during postural perturbations in humans. Exp Brain Res 1988;71:455–66. 10.1007/bf00248739 [DOI] [PubMed] [Google Scholar]
  • 50.Donath L, Kurz E, Roth R, Zahner L, Faude O. Leg and trunk muscle coordination and postural sway during increasingly difficult standing balance tasks in young and older adults. Maturitas 2016;91:60–8. 10.1016/j.maturitas.2016.05.010 [DOI] [PubMed] [Google Scholar]
  • 51.Baratta R V, Solomonow MR, Zhou B-H, Letson D, Chuinard R, D’Ambrosia RD. Muscular coactivation. The role of the antagonist musculature in maintaining knee stability. Am J Sports Med 1988;16:113–22. 10.1177/036354658801600205 [DOI] [PubMed] [Google Scholar]
  • 52.Slobounov SM, Moss SA, Slobounova ES, Newell KM. Aging and time to instability in posture. J Gerontol A Biol Sci Med Sci 1998;53:B71–8. 10.1093/gerona/53a.1.b71 [DOI] [PubMed] [Google Scholar]
  • 53.Slobounov SM, Slobounova ES, Newell KM. Virtual Time-to-Collision and Human Postural Control. J Mot Behav 1997;29:263–81. 10.1080/00222899709600841 [DOI] [PubMed] [Google Scholar]
  • 54.Riccio GE, Stoffregen TA. Affordances as constraints on the control of stance. Hum Mov Sci 1988;7:265–300. [Google Scholar]
  • 55.Danion F, Duarte M, Grosjean M. Fitts’ law in human standing: The effect of scaling. Neurosci Lett 1999;277:131–3. 10.1016/s0304-3940(99)00842-3 [DOI] [PubMed] [Google Scholar]
  • 56.Duarte M, Freitas SMSF. Speed-Accuracy Trade-Off in Voluntary Postural Movements. Motor Control 2005;9:180–96. 10.1123/mcj.9.2.180 [DOI] [PubMed] [Google Scholar]
  • 57.Fitts PM. The information capacity of the human motor system in controlling the amplitude of movement. J Exp Psychol 1954;47:381–91. 10.1037/h0055392 [DOI] [PubMed] [Google Scholar]
  • 58.Biewener AA, Daley MA. Unsteady Locomotion: integrating muscle function with whole body dynamics and neuromuscular control. J Exp Biol 2007;210:2949–60. 10.1242/jeb.005801 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 59.Santuz A, Ekizos A, Eckardt N, Kibele A, Arampatzis A. Challenging human locomotion: stability and modular organisation in unsteady conditions. Sci Rep 2018;8:2740 10.1038/s41598-018-21018-4 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 60.Azizi E, Brainerd EL, Roberts TJ. Variable gearing in pennate muscles. Proc Natl Acad Sci U S A 2008;105:1745–50. 10.1073/pnas.0709212105 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 61.Ihara H, Nakayama A. Dynamic joint control training for knee ligament injuries. Am J Sports Med 1986;14:309–15. 10.1177/036354658601400412 [DOI] [PubMed] [Google Scholar]
  • 62.Beard DJ, Dodd CA, Trundle HR, Simpson AH. Proprioception enhancement for anterior cruciate ligament deficiency. A prospective randomised trial of two physiotherapy regimes. J Bone Joint Surg Br 1994;76:654–9. [PubMed] [Google Scholar]
  • 63.Slijper H, Latash ML. The effects of instability and additional hand support on anticipatory postural adjustments in leg, trunk, and arm muscles during standing. Exp Brain Res 2000;135:81–93. 10.1007/s002210000492 [DOI] [PubMed] [Google Scholar]

Decision Letter 0

Yih-Kuen Jan

30 Aug 2019

PONE-D-19-17729

Swaying slower reduces the destabilizing effects of a compliant surface on voluntary sway dynamics

PLOS ONE

Dear Dr. Patikas,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

We would appreciate receiving your revised manuscript by Oct 14 2019 11:59PM. When you are ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter.

To enhance the reproducibility of your results, we recommend that if applicable you deposit your laboratory protocols in protocols.io, where a protocol can be assigned its own identifier (DOI) such that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). This letter should be uploaded as separate file and labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. This file should be uploaded as separate file and labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. This file should be uploaded as separate file and labeled 'Manuscript'.

Please note while forming your response, if your article is accepted, you may have the opportunity to make the peer review history publicly available. The record will include editor decision letters (with reviews) and your responses to reviewer comments. If eligible, we will contact you to opt in or out.

We look forward to receiving your revised manuscript.

Kind regards,

Yih-Kuen Jan, PhD

Academic Editor

PLOS ONE

Journal Requirements:

1. When submitting your revision, we need you to address these additional requirements.

Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

http://www.journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and http://www.journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. Please provide additional details regarding participant consent also in the ethics statement (currently you only describe consent given in the methods section). In the ethics statement in the online submission information, please ensure that you have specified (1) whether consent was informed and (2) what type you obtained (for instance, written or verbal, and if verbal, how it was documented and witnessed). If your study included minors, state whether you obtained consent from parents or guardians. If the need for consent was waived by the ethics committee, please include this information.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Partly

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: No

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: The study is interesting; however, some minor concerns might need to be clarified.

1.The manuscript mention that seventeen volunteers performed a 2-min voluntary sway task (page 2), and each trail lasted 2 minutes with 2-3 minutes interval in between…no longer than one hour (page 7). How many trails for each volunteer in the study? Is there any pre-trial for subjects to familiar the device?

2.What is the function of the 2 dots (yellow and red) mentioned in the page 6? By the way, what is the purpose of the experimental design showed in figure 1 (A)?

3.What is target signal and how to capture and estimate?

4.The values showed in table 1 by the joint kinematics (Ankle, knee, hip, spine and haed). How these values be measured and estimated? And what are the meanings for these joint and head?

5.In figure 2, what is the unit in vertical axis?

6.In figure 3, the pink circle and blue triangle mean the average values?

7.Figure 5 show the typical example of EMG signal by the figure legend (page 27), however, it needs more explanations to realize the four figures mean.

8.Can the authors quantify a clear threshold or standard to define a sway frequency range for rehabilitation programs in the conclusion by the results?

Reviewer #2: See attached PDF review. Please see attached review. Please see attached PDF, which is my review. Perhaps now I have reached 100 characters.

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files to be viewed.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email us at figures@plos.org. Please note that Supporting Information files do not need this step.

Attachment

Submitted filename: Review of PONE.pdf

Decision Letter 1

Yih-Kuen Jan

30 Oct 2019

PONE-D-19-17729R1

Swaying slower reduces the destabilizing effects of a compliant surface on voluntary sway dynamics

PLOS ONE

Dear Dr. Patikas,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please address the reviewer 2's comments and use the track changes to highlight the changes in the revised manuscript.  

We would appreciate receiving your revised manuscript by Dec 14 2019 11:59PM. When you are ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter.

To enhance the reproducibility of your results, we recommend that if applicable you deposit your laboratory protocols in protocols.io, where a protocol can be assigned its own identifier (DOI) such that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). This letter should be uploaded as separate file and labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. This file should be uploaded as separate file and labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. This file should be uploaded as separate file and labeled 'Manuscript'.

Please note while forming your response, if your article is accepted, you may have the opportunity to make the peer review history publicly available. The record will include editor decision letters (with reviews) and your responses to reviewer comments. If eligible, we will contact you to opt in or out.

We look forward to receiving your revised manuscript.

Kind regards,

Yih-Kuen Jan, PhD

Academic Editor

PLOS ONE

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #1: All comments have been addressed

Reviewer #2: (No Response)

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Partly

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: The manuscript has been revised and all comments have been addressed completely by the authors' responses.

Reviewer #2: The authors Response to Reviews is good. Unfortunately, not much of that actually appears in the revised manuscript. The paragraph beginning Line 82 offers a prediction about EMG, but makes no other predictions, and says nothing about why it is useful to combine data on EMG with the various types of kinematic data. The impression continues to be that of a fishing expedition.

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files to be viewed.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email us at figures@plos.org. Please note that Supporting Information files do not need this step.

Decision Letter 2

Yih-Kuen Jan

25 Nov 2019

Swaying slower reduces the destabilizing effects of a compliant surface on voluntary sway dynamics

PONE-D-19-17729R2

Dear Dr. Patikas,

We are pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it complies with all outstanding technical requirements.

Within one week, you will receive an e-mail containing information on the amendments required prior to publication. When all required modifications have been addressed, you will receive a formal acceptance letter and your manuscript will proceed to our production department and be scheduled for publication.

Shortly after the formal acceptance letter is sent, an invoice for payment will follow. To ensure an efficient production and billing process, please log into Editorial Manager at https://www.editorialmanager.com/pone/, click the "Update My Information" link at the top of the page, and update your user information. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to enable them to help maximize its impact. If they will be preparing press materials for this manuscript, you must inform our press team as soon as possible and no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

With kind regards,

Yih-Kuen Jan, PhD, University of Illinois at Urbana-Champaign

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #1: All comments have been addressed

Reviewer #2: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Partly

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: The manuscript has been revised and all comments have been addressed completely by the authors'

responses.

Reviewer #2: In revising, the authors have made the minimum possible changes. By this choice, they have minimized the contribution of their study to the literature. I suppose that is their choice.

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

Acceptance letter

Yih-Kuen Jan

3 Dec 2019

PONE-D-19-17729R2

Swaying slower reduces the destabilizing effects of a compliant surface on voluntary sway dynamics

Dear Dr. Patikas:

I am pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please notify them about your upcoming paper at this point, to enable them to help maximize its impact. If they will be preparing press materials for this manuscript, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

For any other questions or concerns, please email plosone@plos.org.

Thank you for submitting your work to PLOS ONE.

With kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Yih-Kuen Jan

Academic Editor

PLOS ONE

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Supplementary Materials

    Attachment

    Submitted filename: Review of PONE.pdf

    Attachment

    Submitted filename: Response1.docx

    Attachment

    Submitted filename: Response.docx

    Data Availability Statement

    All relevant data are within the paper and its Supporting Information files.


    Articles from PLoS ONE are provided here courtesy of PLOS

    RESOURCES