Performance of primary diagnostic monitors (PDMs) over time

Andres E Ruuge 1,, Yiming Gao 1, Yusuf E Erdi 1
PMCID: PMC6909116  PMID: 31833641

Abstract

In this work, we evaluated the change of primary monitor characteristics in two consecutive years. Sixty‐six primary monitors were included in the analysis. The monitors were located at radiology physicians' offices and radiology reading rooms. All primary monitors were equipped with the manufacturer's built‐in photometers and connected to the BarcoMediCalQA web service for manual and automatic quality control measurements. External photometer/illuminance meter (RaySafe Solo Light) was used to measure the luminance values. Measured luminance values of the TG18LN1‐18 and TG18UNL80 test patterns were used to evaluate the primary monitors performance. In a comparison of the quality assurance (QA) measurement results for the same monitors that were performed within 2 years, the luminance of 25 displays remained statistically the same (P > 0.01). The luminance of 17 displays decreased (P < 0.01) in 2017 when compared with 2016, the luminance of 24 displays increased (P < 0.01) in 2017 when compared with 2016. For the annual measurements of the MLD in 2016 and 2017, 25 out of 66 displays showed a decrease of MLD values in 2017 compared with the same measurements in 2016 and 41 displays showed an increase of MLD in 2017. All tested primary displays had the MLD value less than 17.2%. The mean value of illuminance measured in 2016 was 5.8 lux ± 3.1 lux. In 2017, the mean value of illuminance measured was 8.7 lux ± 5.3 lux. Although it is expected that monitors luminance values will decrease over time, we found displays with increased luminance. This is possibly due to the multiple monitor calibrations that were performed between two annual monitor QA tests. Based on the findings of this work, more efficient display QA programs with a shorter time interval than 1 year are needed.

Keywords: grayscale display function, primary diagnostic monitor, quality assurance

1. INTRODUCTION

Medical imaging plays an important role in a patient’s treatment. The ability of radiologists to recognize abnormal features on radiographic images depends on the display used to read the image. It is important that the image acquired using any modality is interpreted on a medical display that meets the set of requirements for image quality, medical regulations, and quality assurance.1, 2

The American Association of Physicists in Medicine (AAPM) TG‐18 report defines two classes of displays: primary displays are those used for interpretation of medical images and secondary displays include all other displays used in medicine.2 The ACR–AAPM–SIIM Technical Standard for Electronic Practice of Medical Imaging has provided basic recommendations regarding display performance characteristics for different display classifications.3 The Digital Imaging and Communication in Medicine (DICOM) Part 14: Grayscale Standard Display Function defines the calibration of primary displays.4

Quality assurance programs for displays used for the primary interpretation of medical images are required by some state and city regulators.5 Quality control (QC) measurements typically include an evaluation of display luminance characteristics such as the Grayscale Standard Display Function (GSDF), the luminance uniformity, and ambient light conditions of the display environment.

The GSDF curve is a luminance response curve of the display, which indicates that the display provides perceptual linearization for the human eyes.4 The characterization of the GSDF curve was performed by luminance measurements in steps of 18 gray levels both at acceptance and at annual testing of the display.5, 6 The luminance uniformity is assessed to ensure that the luminance in the corners and at the center of the display agree within a designated percentage. The luminance measurement is usually performed with an external photometer, which measures the intensity of visible light emitted from a unit area of the display’s surface. The results of the luminance measurements are part of the QC of the medical display. The illuminance on a display is the total luminous flux from all other light sources in the room and is evident on the viewing surface of the display. Illuminance is measured by an external photometer that that has been converted to an illuminance meter.

When assessing a display’s performance parameters, it is important to note that its luminance performance can change over time. This change may decrease a display’s overall performance, depending on its age and hours of usage.6, 7, 8, 9 The primary interest for this work is to determine how much the luminance level changes after 1 year of display usage.

2. METHODS

Measurements were performed on 66 primary displays of various models: MDCC‐6130 (35), MDCC‐6230 (5), MDCG‐10130 (2), MFGD 3420 (6), and MFGD 5421 (18) (Barco; Duluth, Georgia) as part of annual QA testing following New York City regulations.5 All measured displays were liquid crystal displays equipped with the manufacturer’s built‐in photometer (Barco, I‐Guard), which was connected to Barco MediCal QAWeb software for automatic built‐in QC service. A calibrated RaySafe Solo Light photometer (RaySafe; Billdal, Sweden) was used to perform manual luminance measurements using the TG‐18 test patterns. All displays were warmed up for at least 30 min before the luminance measurements were taken. The display output was left to stabilize after each switch of patterns, and then a stable reading on the luminance meter was obtained. To verify stability, several repeated recordings were made for each display to ensure that there was no drift in the photometer readings. The display performance data of the 66 primary displays was collected and analyzed for 2 years (2016 and 2017). The effect of changes in temperature and ambient light conditions on display luminance was assumed to be negligible because the locations of the primary displays in the diagnostic reading rooms remained the same over the year and ambient light conditions were controlled in those rooms.

2.1. Luminance measurements

Characterization of the GSDF curve was performed in steps of 18 gray levels, resulting in a total of 18 luminance measurements per year for each display. TG18LN1 through TG18LN18 test patterns were used to perform luminance measurements of the display.2, 3, 5 Each test pattern image in the series displayed a uniform background at a pixel value of 20% of the display’s maximum luminance, with a central square test region of specific luminance level occupying 10% of the image, allowing the repeatable positioning of the contact photometer during each of the 18 measurements. For data analysis, all displays were sorted by the year of initial use. For each display, the 18 luminance measurements in 2016 were compared with the 18 luminance measurements in 2017 using a paired t‐test. In addition, linear regression analysis was performed on all 1188 pairs of measurements to check whether human error or another unexpected error deviated any measurement from normal performance over the 2 years.

2.2. Display Uniformity

The Maximum Luminance Deviation (MLD) is described as the quantitative measure of display uniformity in the TG‐18 report.2 Luminance values were measured at five locations on the diagnostic display: the center and at each of the four corners using the TG18UNL80 test pattern. The MLD of a display was calculated using the following equation:

MLD%=200×L2-L1/(L2+L1) (1)

where L 2 is the highest recorded luminance, and L 1 is the lowest recorded luminance, regardless of the location on the display.2 MLD is a characteristic of a display that cannot be directly adjusted using the display calibration. The data of annual MLD changes for each studied display has been presented in this work. The difference between 2016 and 2017 MLD values was estimated for statistical significance by a Wilcoxon signed‐rank test, which is a suitable statistical test for dependent samples and is as efficient as a student t‐test for normal distributions.12

2.3. Illuminance measurements

In addition to diagnostic display luminance measurements, ambient illuminance was measured on each workstation display to confirm that the display’s ambient illuminance did not exceed 50 Lux, as is recommended in the ACR‐AAPM‐SIIM standard.3

The illuminance was measured using the same RaySafe Solo Light photometer configured for both luminance and illuminance measurements. The illuminance sensor was positioned in the center of the display facing outward while the primary display was turned off. The results of illuminance measurements for 2016 and 2017 were compared to evaluate the possible change in reading conditions. The difference between 2016 and 2017 was estimated for statistical significance by a Wilcoxon signed‐rank test.

3. RESULTS

3.1. Luminance measurements

All diagnostic displays analyzed in this study were sorted by the in‐service date and are presented in Table 1. All the displays were less than 8 years old at the time of the study.

Table 1.

List of all measured primary display models (BARCO; Duluth, Georgia) included in the study and sorted by year of first use.

# Model In use since # Model In use since # Model In use since
1 MFGD 5421 09/2009 23 MFGD 5421 03/2012 45 MDCC‐6130 01/2014
2 MFGD 5421 09/2009 24 MDCC‐6130 03/2012 46 MDCC‐6130 01/2014
3 MFGD 5421 10/2009 25 MDCC‐6130 05/2012 47 MDCC‐6130 01/2014
4 MFGD 5421 10/2009 26 MDCC‐6130 05/2012 48 MDCC‐6130 01/2014
5 MFGD 5421 10/2009 27 MDCC‐6130 05/2012 49 MDCC‐6130 01/2014
6 MFGD 5421 10/2009 28 MDCC‐6130 05/2012 50 MDCC‐6130 01/2014
7 MFGD 3420 11/2009 29 MFGD 3420 06/2012 51 MDCC‐6130 01/2014
8 MFGD 3420 11/2009 30 MFGD 3420 06/2012 52 MDCC‐6130 02/2014
9 MFGD 5421 11/2009 31 MDCC‐6130 09/2012 53 MDCC‐6130 03/2014
10 MFGD 5421 11/2009 32 MDCC‐6130 11/2012 54 MDCC‐6130 03/2014
11 MFGD 5421 11/2009 33 MDCC‐6130 11/2012 55 MDCC‐6130 03/2014
12 MFGD 5421 11/2009 34 MDCC‐6130 08/2013 56 MDCC‐6130 03/2014
13 MFGD 5421 11/2009 35 MDCC‐6130 08/2013 57 MDCC‐6130 03/2014
14 MFGD 5421 11/2009 36 MDCC‐6130 08/2013 58 MDCC‐6130 03/2014
15 MFGD 5421 11/2009 37 MDCC‐6130 08/2013 59 MDCC‐6130 03/2014
16 MFGD 5421 11/2009 38 MDCC‐6130 08/2013 60 MDCG‐10130 05/2014
17 MFGD 3420 06/2011 39 MDCC‐6130 08/2013 61 MDCG‐10130 05/2014
18 MFGD 3420 06/2011 40 MDCC‐6130 08/2013 62 MDCC‐6230 06/2014
19 MDCC‐6130 03/2012 41 MDCC‐6130 08/2013 63 MDCC‐6230 11/2014
20 MFGD 5421 03/2012 42 MDCC‐6130 12/2013 64 MDCC‐6230 06/2015
21 MFGD 5421 03/2012 43 MDCC‐6130 12/2013 65 MDCC‐6230 06/2015
22 MFGD 5421 03/2012 44 MDCC‐6130 12/2013 66 MDCC‐6230 06/2015

Twenty‐four displays were calibrated for Lmax = 500 cd/m2, and 42 displays (bold) were calibrated for Lmax = 400 cd/m2.

Luminance measurements for each of the display paired t‐tests of 2016 and 2017 are presented in Table 2. Of the 66 displays, the luminance of 25 displays remained statistically the same (P > 0.01). The luminance of 17 displays decreased (P < 0.01) in 2017 when compared with 2016. Finally, the luminance of 24 displays increased (P < 0.01) in 2017 when compared with 2016.

Table 2.

Paired t‐test (2016 luminance measurements vs 2017 measurements) results for all 66 monitors.

No change in Luminance P > 0.01 Luminance decreased in 2017 than in 2016, P < 0.01 Luminance increased in 2017 than in 2016, P < 0.01
Monitor # P Value Monitor # P Value Monitor # P Value
2 0.017 6 0.0003 1 0.0020
7 0.049 14 0.0095 3 0.0003
8 0.015 24 0.0005 4 0.0079
9 0.492 26 0.0023 5 0.0044
10 0.077 31 0.0001 15 0.0003
11 0.295 33 0.0002 17 0.0022
12 0.019 35 0.0032 20 0.0016
13 0.160 36 0.0023 21 0.0004
16 0.012 37 0.0090 22 0.0010
18 0.021 42 0.0027 23 0.0042
19 0.045 43 0.0066 27 0.0005
25 0.038 46 0.0074 28 0.0011
29 0.015 47 0.0028 34 0.0061
30 0.351 49 0.0002 41 0.0002
32 0.016 50 0.0006 44 0.0004
38 0.142 58 0.0011 45 0.0002
39 0.038 59 0.0002 51 0.0003
40 0.014 Count: 17 52 0.0013
48 0.316     53 0.0007
54 0.808     60 0.0007
55 0.096     61 0.0003
56 0.049     64 0.0038
57 0.029     65 0.0011
62 0.026     66 0.0003
63 0.117     Count: 24
Count: 25        

The 2016 and 2017 luminance measurement performance of the physicists was evaluated using simple linear regression. The scatterplot is presented in Fig. 1 where 1188 pairs of luminance observations are drawn as circles, showing that the measurements from the 2 years are highly correlated and almost fall perfectly on the 45‐degree line, with the Pearson correlation coefficient = 0.999 (P‐value < 0.0001).

Figure 1.

Figure 1

The linear regression with Pearson correlation coefficient = 0.999 (P‐value < 0.0001), where 1188 pairs of luminance observations (circles) are observed, shows measurements from 2016 and 2017 which are highly correlated and fall almost perfectly on the 45° line.

3.2. Evaluation of Display Uniformity

Figure 2 represents the results of annual measurements of the MLD in 2016 and 2017. Twenty‐five out of 66 displays showed a decrease of MLD values in 2017 compared with the same measurements in 2016, and 41 displays showed an increase of MLD in 2017. The maximum change in uniformity was 0.5% between 2016 and 2017. All tested primary displays had a MLD value less than 17.2%, well below the primary display QA standard of 30% or less.5

Figure 2.

Figure 2

The results of annual measurements of the maximum luminance deviation (MLD) performed in 2016 (circles) and 2017 (columns). Twenty‐five out of 66 displays showed a decrease in MLD values in 2017 when compared with 2016, and 41 displays showed an increase in MLD in 2017. The maximum change in MLD was 11.7% in 2016 and 17.2% in 2017.

The Wilcoxon signed‐rank test with a significance level of P‐value = 0.01 showed the statistically insignificant differences in MLD measurements performed in 2016 and 2017 (P‐value = 0.56).

3.3. Illuminance measurements

The mean value of illuminance measured in 2016 was 5.8 lux ± 3.1 lux. In 2017, the mean value of illuminance measured was 8.7 lux ± 5.3 lux. The levels for both years met the American College of Radiology recommended less than 50 lux for a nonmammography diagnostic display radiology reading room.3

4. DISCUSSION

In recent years, there has been an increase in the number of primary displays deployed in our radiology clinics. The quality and accuracy of a radiologist’s reading depends on the performance of the display that he/she uses to review images and to perform diagnosis. Problems in display luminance, display uniformity, or ambient light could negatively affect image interpretation. To ensure that primary displays are performing according to their specifications, display performance parameters must be checked regularly as has been described in various publications.7, 8, 9, 10, 11

One of the objectives of a well‐established QC program for the primary displays is to follow and quantify any change of performance parameters over time, thereby detecting the problems that arise with the displays before they lead to clinical issues.

In this study, the changes in primary display performance parameters, such as luminance, were inconsistent across displays: they either stayed the same, decreased, or increased after 1 year of usage. Measurements of illuminance or luminance and their corresponding accuracies, are influenced by various parameters such as operational conditions and properties of light sources as well as characteristics of the applied photometers. However, the systemic error of the photometer (0.3%) could not fully account for all the different changes in the display performance.

The high correlation between 2016 and 2017 luminance measurements precluded human error or any random error from affecting the measurements. This finding suggests the performance of the physicists in the primary display tests was consistent through the 2 years.

The light sources around the primary displays were reviewed and modified to comply with the ambient light requirements of the standard at acceptance testing and were maintained over the 2 years.3 In addition, all measurements were performed after at least a 30‐min warm‐up of the displays based on the results of our previous work, which was performed to estimate the stability of luminance measurements for all tested primary display models.9 The results from that study confirmed that it is important to perform any luminance measurements after the display has been turned on for at least 30 min.9

Another source of inaccuracy in QC measurements of primary displays is that all displays in the program undergo automated calibration biannually. Using an internal photometer reading, display software calibrates the luminance level according to company specifications — in this case, 400 cd/m2 or 500 cd/m2, depending on the display model. During this process, display luminance levels are adjusted to meet the specifications. We found if annual QA was performed after this calibration, the external photometer would measure the adjusted level, which in turn would show changed luminance values and adjusted uniformity of the primary display in our preliminary investigations. However, the measurements included in this study did not control for the effect of the internal calibration of the displays. We suspect the automatic calibration of the displays is one of the sources of changes in the luminance measurements over the 2 years. Thus, future work will be to perform measurements before and after the automatic calibration of the displays to determine the effect of such calibration on luminance measurements.

The Wilcoxon signed‐rank test results of a display’s MLD (P‐value = 0.56) showed that the one‐year change in uniformity performance was not statistically significant. This finding suggests that the uniformity of the displays was not as sensitive as luminance to the automatic calibration of the displays.

We propose shorter time intervals of QC measurements based on the preliminary findings of the potential effect of automatic display calibration on the measurements for QA program improvement and research proposes if the resources allow such actions. Ideally, primary display QC programs can be performed at the time of the internal display calibration. As a result, fluctuation between the display calibration date and the display QC date would not affect the measured primary display parameters.

5. CONCLUSION

From this 2‐year investigation, it has been shown that the performance of several primary displays changes over time. Although it is unclear how much these performance changes impact the accuracy of clinical diagnosis, it is essential for radiology departments to develop QA programs and follow‐up with primary display performance at least annually and ideally even in shorter time intervals.

CONFLICT OF INTEREST

The authors have no relevant conflicts of interest to disclose.

ACKNOWLEDGMENTS

The authors have no acknowledgements to declare.

REFERENCES

  • 1. Flynn MJ, Kanicki J, Badano A, Meyer, WR . High fidelity electronic display of digital radiographs. Radiographics 1999;19:1653Y1669. [DOI] [PubMed] [Google Scholar]
  • 2. Samei E, Badano A, Chakraborty D, et al. Assessment of display performance for medical imaging systems [Internet]. College Park, MD: American Association of Physicists in Medicine; 2005. 155 p. Available from: https://www.aapm.org/pubs/reports/OR_03.pdf.
  • 3. ACR‐AAPM‐SIIM Technical Standard for Electronic Practice of Medical Imaging , [Internet]. American College of Radiology (ACR), the American Association of Physicists in Medicine (AAPM), and the Society for Imaging Informatics in Medicine (SIIM); 2014. [Amended 2014 (Resolution39)]. 18 p. Available from: http://www.acr.org/~/media/AF1480B0F95842E7B163F09F1CE00977.pdf.
  • 4. Part 14: Grayscale Standard Display Function. In: Digital Imaging and Communications in Medicine (DICOM). Rosslyn (VA): National Electrical Manufacturers Association; 2004 55. Report No.: PS 3.14‐2004.
  • 5. NYC Quality Assurance Guidelines for Primary Diagnostic Monitors 2015. New York (NY): New York City Health Department’s Office of Radiological Health (ORH); 2015 Jun 8.
  • 6. Pollard BJ, Chawla AS, Delong DM, Hashimoto N, Samei E. Object detectability at increased ambient lighting conditions Med Phys. 2008:35;2204–2213. [DOI] [PubMed] [Google Scholar]
  • 7. Silosky M, Marsh RM. Constancy of built‐in luminance meter measurements in diagnostic displays. Med Phys. 2013;40:121902. [DOI] [PubMed] [Google Scholar]
  • 8. Ruuge AE, Mahmood UA, Erdi YE. The assessment and characterization of the built‐in internal photometer of primary diagnostic monitors. Appl Clin Med Phys. 2017;18:170–175. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9. Ruuge AE, Erdi YE. Comparison of the external luminance meter performance as a part of diagnostic monitor QC. Med Phys. 2017;44:p3207. [Google Scholar]
  • 10. Barten P.Physical model for the contrast sensitivity of the human eye. SPIE Human Vision, Visual Processing and Digital Display III, 57 (August 27, 1992).
  • 11. Barten P. Contrast Sensitivity of the Human Eye and Its Effects on Image Quality. Bellingham: SPIE Optical Engineering Press; 1999. [Google Scholar]
  • 12. Richard L. "Concepts & Applications of Inferential Statistics". http://vassarstats.net/textbook/ch12a.html. Retrieved 5 November 2018.