Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2019 Dec 14.
Published in final edited form as: Anal Biochem. 2000 Jan 1;277(1):74–85. doi: 10.1006/abio.1999.4346

Background Suppression in Frequency-Domain Fluorometry

Joseph R Lakowicz *,1, Ignacy Gryczynski *, Zygmunt Gryczynski *, Michael L Johnson
PMCID: PMC6911226  NIHMSID: NIHMS1061828  PMID: 10610691

Abstract

Gated detection is often used in time-domain measurements of long-lived fluorophores for suppression of interfering short-lived autofluorescence. However, no direct method has been available for gated detection and background suppression when using frequency-domain fluorometry. We describe a direct method for real-time suppression of autofluorescence in frequency-domain fluorometry. The method uses a gated detector and the sample is excited by a pulsed train. The detector is gated on following each excitation pulse after a suitable time delay for decay of the prompt autofluorescence. Under the same experimental conditions a detectable reference signal is obtained by using a long lifetime standard with a known decay time. Because the sample and reference signals are measured under identical excitation, gating and instrumental conditions, the data can be analyzed as usual for frequency-domain data without further processing. We show by simulations that this method can be used to resolve single and multiexponential decays in the presence of short lifetime autofluorescence.


Fluorescent detection is being used throughout the biosciences for numerous applications including clinical chemistry, DNA sequencing, FISH, flow cytometry, high-throughput screening, and cellular imaging (18). In many instances the sensitivity is limited by interfering autofluorescence from the sample rather than detectability of the emission. This interference can be suppressed with gated detection (9, 10) in which the detector is gated off during an excitation pulse, and gated on following a suitable time delay which allows the scattered light and short-lived autofluorescence to decay. This method is frequently used with the long lifetime lanthanides in the so called “time-resolved immunoassays” (11).

At present, methods to directly suppress or off-gate the short-lived interference are not available when using the frequency-domain method. In frequency-domain fluorometry the sample is typically excited with sine-wave-modulated light, the detector is continuously on, so there is no useful temporal separation of the short and long lifetime emissions. Lack of a procedure for background suppression is disadvantageous because FD2 measurements are of current interest for developing simple and robust instruments for on-site measurements of a variety of analytes, blood chemistry, and immunoassays. The need for gated detection in FD fluorometry has been increased by the introduction of long lifetime metal-ligand complexes (MLCs) as luminescence probes. These probes display lifetimes ranging from 20 ns to 10 μs (1215), which is much longer than the typically nanosecond decay times for autofluorescence. Because of their long lifetimes these MLC probes can provide high sensitivity detection when used with gated detection.

In the present report we describe a method for frequency-domain measurement of the time-resolved intensity decays of long-lived luminophores with real-time suppression of the short-lived interfering autofluorescence. We note that our method allows recovery of the lifetimes and amplitudes, and is not just a measurement of the integrated intensity of the long lifetime emission which is the basis of the so-called “time-resolved immunoassays” (11). The intensity decay parameters are of interest because of the possibility of chemical sensing based on the decay times (1618).

There have been two previous reports of correcting for background fluorescence in FD fluorometry (19, 20). These methods require a separate measurement of a blank sample displaying background, as well as a measurement of the sample which also displays the background signal. The data from the sample are then corrected for the background using procedures appropriate for the FD data (19, 20). This approach allows correction for background signal even when its decay times are comparable to those displayed by the desired sample components. This is a more stringent requirement than is needed for suppression of prompt autofluorescence, and the method is more complex than necessary for suppression of short-lived components.

In the present report we describe a FD method which can be used to directly measure the long-lived decay times with simultaneous suppression of the short decay time components. The method depends on the use of a train of excitation pulses, rather than a sine-wave-modulated light source. It is known that a pulse train can be used for excitation in FD fluorometry based on the harmonic content of the pulses (2124). As an example, a 1-MHz pulse train with a pulse width near 0.3 ns has useful harmonic content at each integer multiple of 1 MHz up to about 1 GHz (22).

The use of pulse train excitation provides the opportunity to turn off the detector during and immediately after the excitation pulse, at which time when the emission contains the short-lived autofluorescence. This opportunity does not seem to be available with sine wave excitation. Methods to suppress single decay time components have been described (2527), but these methods cannot be used to suppress autofluorescence and still recover the time-resolved parameters.

In frequency-domain fluorometry one compares the phase shift and relative modulation of a sample and reference. The reference is typically a slightly turbid solution which scatters light and provides a measure of the phase and modulation of the incident light. The use of a scattering reference poses a dilemma because gating of the detector will prevent detection of the scattered light, precluding comparison of the sample and reference. If one uses gating on the sample, but not on the reference, then the sample and reference emission are no longer directly comparable. This comparison is fundamental to the measurements and used in all modern FD instruments. These difficulties can be avoided by the use of a long-lived luminophore as the reference. The use of short lifetime reference fluorophores was described previously as a method to correct for the wavelength-dependent time response of photomultiplier tubes (PMTs) (28, 29). If the decay time of the reference is known, the measured phase and modulation of the sample can be corrected to that which would have been observed with scattered light or with a zero decay time reference. However, such nanosecond lifetime standards cannot be used with the present background suppression method because their emission will not be detectable at longer times. In the present report we use a reference with an adequately long lifetime so that its emission is observable until the PMT is gated on. The sample and reference are observed with the same gated PMT, with the same gating time profile. The PMT is gated on after each excitation pulse, following a delay time suitable for decay of the autofluorescence. We show by simulations that the phase and modulation data can be used directly to recover the intensity decay parameters of the long-lived luminescence.

INTUITIVE DESCRIPTION OF BACKGROUND SUPPRESSION

Prior to a description of the theory for frequency-domain measurements with background suppression it is informative to have an intuitive understanding of the method. Assume the intensity decay of the sample following δ-function excitation is given by a sum of exponential components

I(t)=iαi0exp(t/τi). [1]

In this expression α10 are the amplitudes at t = 0 associated with each decay component. The superscript 0 is used to stress the fact that the α10 value refers to t = 0. As will be shown below, gated detection can alter the apparent values of αi, particularly when the sample (S) decay is a multiexponential.

To illustrate the usefulness of background suppression we assume that the intensity decay of the sample displays two decay times (τB and τS) associated with the emission from the background (B) and from the sample (S). The intensity decay is thus given by

I(t)=αB0exp(t/τB)+αS0exp(t/τS). [2]

The goal of gated detection is to measure the sample decay time τS without interference from the background component.

The usefulness of gated detection can be seen by examination of a simulated intensity decay (Fig. 1). In this simulation we assumed that the sample displays two decay times of τB = 10 ns and τS = 1 μs. Following δ-function excitation the observed intensity decay is given by

Iobs(t)=1000exp(t/10ns)+10exp(t/1000ns). [3]

FIG. 1.

FIG. 1.

Simulated time-dependent intensity decay according to Eq. [1], τB = 10 ns, τS = 1000 ns, αB0=1000,andαS0=10. The dashed line shows the gating function, with an on-time of tON = 100 ns.

In this expression αB0=1000 is the relative amplitude of the short lifetime background (B) component with τB = 10 ns, and αS0=10 the relative amplitude of the long lifetime sample(s) component with τS = 1000 ns. The experimental goal is to measure the longer decay time without interference from the short-lived autofluorescence.

It is important to recall the meaning of the αi values, which are amplitudes in the intensity decay. When measuring a steady-state intensity one usually wants to know the fractional contribution ( fi) of each component to the measured intensity. If the sample is excited continuously and the detector is always on, then the fractional intensities of the short-lived background ( fB) and long-lived sample ( fS) decay components are given by

fB0=αB0τBαB0τB+αS0τS=0.5 [4]
fS0=αSτSαB0τB+αS0τS=0.5. [5]

The simulated intensity decay shows that a component of interest, with a total intensity of 50% of the measure signal, will have a minor amplitude in the time-dependent decay (Fig. 1). Stated conversely, the time-zero amplitude of the background signal can be many-fold greater than the amplitude of the component of interest.

Now consider that this time-dependent decay is observed with a gated detector which has an on-off ratio of 105. Such ratios can be achieved and ratios as large as 7 × 105 have been reported (10). With a standard squirrel-cage PMT the rise time of the gating on signal is typically 1–5 ns, which we will assume to be instantaneous compared to the longer components in the decay. Suppose the gate is turned on at a delay time tON = 50 ns, and that the gating rise time is instantaneous. The fractional steady-state intensities of each component are given by the integral by Eq. [1] from tON = 50 to infinity. Integrating each decay component separately and normalizing by the sum reveal that the fractional amplitudes are now fB = 0.007 and fS = 0.993 (Fig. 2). The superscript zero has been dropped to reflect the use of gated detection and a change in the apparent αi and fi values. Still greater suppression of the background occurs with tON = 100 ns, which is the gating function shown as the dashed line in Fig. 1. In this case fB = 5 × 10−5 and fS = 0.99995 (Table 1). It is clear from these simulations that the amplitude (αB) and fractional intensity ( fB) of the background are progressively decreased as the delay time tON is increased. Hence, readily achievable PMT gating results in essentially complete elimination of the scattered light and/or autofluorescence from the sample. If measured in the time domain the data obtained for times greater than 100 ns can now be analyzed by the usual method of nonlinear least squares as applied to time-domain data (30). The use of gated detection would eliminate the signal due to the short-lived autofluorescence and result in faster data acquisition of the signal of interest. Hence gated detection can also be valuable for time-domain measurements by removal of the background signal rather than measurement and analysis of data which are corrupted by the background.

FIG. 2.

FIG. 2.

Fractional intensity of the background (B) and sample (S) for various on-gating times (tON). The intensity decay law is given by Eq. [1].

TABLE 1.

Effect of Off-Gating on the Fractional Intensities of the Background (B) and Sample (S)a

Gating time (ns) fB fS

None 0.50 0.50
10 ns 0.271 0.729
25 ns 0.078 0.922
50 ns 0.007 0.993
75 ns 0.0006 0.9994
  100 ns 0.00005 0.99995
  200 ns 2.5 × 10−9
a

The intensity decay parameters are given in Eq. [1].

It is informative to examine the effects of short-lived autofluorescence on the frequency-domain data measured without gated detection. Figure 3 shows the frequency response expected for a single exponential decay of the sample (τS = 1000 ns) with increasing amplitudes (αB0) of the 10-ns background signal when measured without gated detection. As the amplitude of the background increases the frequency responses (—) become distorted from the response expected from the sample itself (– – –). In principle one could analyze the frequency-domain data in terms of two decay times, and thereby recover τS as one of the components from the multiexponential analysis. In practice, the amplitude of the background can exceed that of the sample, making the signal of interest a minor component in the measured signal. Also, the fraction of the signal due to the component of interest contains more Poisson noise due to the higher intensity signal. For these reasons, it is preferable to eliminate the background signal prior to detection.

FIG. 3.

FIG. 3.

Distortion of the frequency response by increased background. For these simulations, τB = 10 ns and τS = 1000 ns and αB0 as shown in each panel. The goal of gated detection is to recover the background-free intensity decay (dashed line).

The gated concept can be applied to frequency-domain fluorometry. Consider an FD experiment in which the light source is a continuous pulse train (Fig. 4). A pulse train is known to be useful for frequency-domain measurements when using the harmonic content method (2124). When using such a light source the FD data can be measured at every integer multiple of the pulse repetition frequency up to a frequency near (tp)1, where tp is the pulse width of the incident light. The top panel shows the signal expected without gated detection. The vertical dotted lines (. . .) show the signal observed from the usual dilute scattering reference which displays a zero lifetime. The solid lines (—) show the intensity decay of the sample. The sample intensity decay is assumed to display a short-lived signal due to background, as well as a long decay time due to the sample. For these simulations we assumed the sample lifetime of interest was τS = 2000 ns. The short-lived component is the vertical region of the decay, and the decay of the sample of interest is the angled region of this line. The dashed line shows the decay of the reference fluorophore with τR = 1000 ns. For these simulations we assumed that the reference did not display a short-lived component, but this assumption is not necessary because gated detection is performed on both the sample and the reference.

FIG. 4.

FIG. 4.

Effect of gated detection on the emission resulting from pulse train excitation. Top, no gated detection; middle, the gating function; bottom, signal seen by the detection with gating. For these simulations the various parameters were as follows: pulse repetition rate 0.07 MHz, τB = 10 ns, αB0=100,τS=2000ns,αS0=1,τR=1000ns,αR0=1, tON = 100 ns, Δt = 100 ns, tOFF = 10,000 ns, on-off ratio 7 × 104. In the top and bottom panels the solid lines represent the sample signal and the dashed lines the reference signal.

When the phase and modulation are measured, the sample (—) and reference (– – –) signals are alternatively observed using the same detector. The phase and modulation are typically measured using cross-correlation electronics at the desired measurement frequency (ω in radians/s) and one obtains the same phase and modulation of the emission as if the excitation source was modulated as a pure sine wave at frequency ω (2124). Because the detector is continuously on, one observes the total emission from the short and long lifetime components, and the observed phase (ϕωobs) and modulation (mωobs) are distorted by the presence of the short-lived background.

Now assume the detector is gated off during the excitation pulse, and gated on after the autofluorescence has decayed, at about 100 ns after the excitation pulse. The gating function would be a sequence of rectangular gating pulses (Fig. 4, middle panel). In this case the short-lived components do not contribute to the signal seen by the PMT. However, the gating function will completely suppress the signal from the scattering reference, so one cannot measure the phase difference and modulation of the sample as compared to the scattering reference. While in principle one could determine the arrival time of the pulses by other means, much of the precision and freedom from arti-facts in FD measurements originates by comparing the scattering reference and the sample with the same detector under the same experimental conditions.

The difficulty caused by suppression of the reference signal by the gating function can be overcome using long lifetime reference luminophores. Suppose the scattering solution is replaced by a reference which displays a known single exponential lifetime (τR). The phase and modulation of the reference, relative to a scattering reference, are given by

ϕR=tan1(ωτR) [6]
mR=(1+ω2τR2)1/2. [7]

Suppose the sample is measured relative to this reference rather than to the scattering solution. Then the observed phase angle (ϕobs) for the sample is shorter than the true value by ϕR (28, 29). The actual phase angle (ϕ) of the sample is given by

φ=φobs+φR. [8]

Similarly, the observed modulation of the sample (mobs) is larger than the true value by a factor (mR1). The actual modulation (m ) is given by

m=mobs(1+ω2τR2)1/2=mobsmR. [9]

Correction of the measured phase and modulation values for a reference lifetime is a standard part of most FD data analysis programs. References with known lifetimes are used to correct for the color-dependent time response of PMTs.

Most fluorophores used as lifetime references have decay times of 1 to 10 ns (28, 29). Hence these standards cannot be used with this gating method because their emission will have decayed prior to on-gating of the detector. This difficulty can be solved by using longer decay time references luminophores. In particular, the transition metal-ligand complexes display usefully long decay times near 1 μs and frequently display single exponential decays in fluid solvents. Because of the long decay time the signal will persist after the detector is gated on at t = tON.

Examination of the bottom panel of Fig. 4 reveals a useful result. The detector does not see the arrival time of the light pulse, but only the rise of the photocurrent at t = tON. Hence, comparing the reference and sample, with the same detector and gating function, yields data comparable to those observed with a reference fluorophore without gating. Frequency-domain measurements can thus be performed with suppression of autofluorescence by using off-gating at times near the excitation pulse. For a single exponential decay the long lifetime can be obtained using Eqs. [8] and [9]. Alternatively, the resulting data can be directly used in currently available frequency-domain software to recover the multiple decay times without any modification. Minor changes in the software are needed to recover the true time-zero amplitudes.

THEORY

A detailed description of FD background suppression requires expressions which describe the time-dependent signals. Following δ-function excitation the observed (obs) time-dependent decay is given by

Iobs(t)=IB(t)+IS(t), [10]

where IB(t) describes the intensity decay of the background autofluorescence and IS(t) the decay law of the sample in the absence of autofluorescence. For simplicity with no loss of generality we assume that the autofluorescence decays with a single decay time τB, and the sample itself displays a multiexponential decay. Then

Iobs(t)=αBexp(t/τB)+iαiexp(t/τi), [11]

where τi are the background-free decay times and Σαi = 1.0. We choose to normalize the Σαi to 1.0 because the magnitude of the short-lived component is then easily seen as the excess amplitude of Iobs(t) over 1.0.

Analysis of the frequency-domain data is accomplished by comparing the measured phase (ϕω) and modulation (mω) at a given frequency with those calculated (ϕcωandmcω) for an assumed decay law (31, 32). The calculated values are found from the sine and cosine transforms of the intensity decay. For on-gating at t = tON and off-gating at t = tOFF these transforms are

Nω=1JtONtOFFIobs(t)sinωtdt [12]
Dω=1JONtOFFIobs(t)cosωtdt [13]
J=tONtOFFIobs(t)dt. [14]

The calculated phase (ϕcω) and modulation (mcω) are given by

φcω=Nω/Dω [15]
mcω=(Nω2+Dω2)1/2, [16]

where Nω,Dω, and J are evaluated with assumed parameters values in Eq. [11]. Equations [12][14] are comparable to the standard expressions, except for integration from tON to tOFF rather than t = 0 to infinity as is done without gated detection.

For simulation purposes we calculated the phase and modulation according to Eqs. [12][16], for assumed values of tON and the parameters in Eq. [11] (αB, τB, αi, and τi). These values were compared with the values expected without gated detection, with no background (αB = 0) calculated using Eqs. [10][14] with tON = 0. Simulations were performed to determine whether analysis of the data expected with background suppression could be used to recover the expected values of αi and τi from the background-free decay.

The simulations were performed with different assumed decay laws (I(t)), time delays (tD), and shapes of the on-gating function. After a suitable on-time the gate is turned off at t = toff. We assumed the shape of the on-gate was given by the error function complement shape, which is an integrated Gaussian. In this case the gating function is given by

g(t)=1G+(11G)(112erfc[ttONΔt])×(12erfc[ttOFFΔt]), [17]

where G is the on-off gain ratio. This equation which is inserted into Eqs. [12][14] for the integration. The value of Δt was varied to simulate on-gating with various rise times. For the long assumed sample decay times the values of Δt near 10 to 100 ns gave essentially a rectangular gating function where the signal is only detected between tON or tOFF.

MULTIEXPONENTIAL INTENSITY DECAYS

It is straightforward to recover of a single decay time from the sample when using gated detection. In this case the recovered decay time is the long decay time of the sample, and the amplitude assumed equal to 1.0. The situation is slightly more complex for a multiexponential decay. The amplitudes (αi) in a time-dependent decay (Eq. [11]) represent the amplitudes at t = 0. Since the detector is off at t = 0, the recovered time-zero amplitudes reflect the values at t = tON. Fortunately, it is straightforward to calculate the αi values at t=tON. using the recovered lifetimes and amplitudes. When using gated detection the observed amplitude (αiObs) will be given by the integrated intensity of this component between t = tON and t = tOFF. For a double exponential decay these amplitudes are proportional to

α1Obs=kα10[exp(tON/τ1)exp(tOFF/τ1)] [18]
α2Obs=kα20[exp(tON/τ2)exp(tOFF/τ2)], [19]

where k is the proportionally constant. Hence the ratio of α10toα20 can be calculated using

α10α20=α1Obsα2Obs[exp(t/ON/τ2)exp(tOFF/τ2)][exp(tON/τ1)exp(tOFF/τ1)]. [20]

The normalized values of α10andα20 are calculated by recalling α10+α20=1.0. It will be necessary to develop other expressions for nonexponential decays.

RESULTS

We simulated the frequency-domain data expected with gated detection (Fig. 5). For these simulations we chose a pulse repetition rate of 0.07 MHz and the decay law shown in Eq. [3], except the long sample decay time (τS) was varied from 500 to 5000 ns. The gating on and off times were 100 and 10,000 ns, respectively, and the reference lifetime was τR = 1000 ns. The solid lines shown in Fig. 5 are the least fits to a single exponential decay. Except for τS = 5000 ns (lowest panel), the data are well matched to the single exponential model, and the recovered lifetimes agree with the values assumed for the simulations. These results demonstrate that FD measurements can be performed with background suppression and that the data are essentially identical to those found in the absence of autofluorescence.

FIG. 5.

FIG. 5.

Simulated frequency-domain phase and modulation data with gated detection. The parameter values are the same as on Fig. 3, lowest panel, except that τS was varied from 500 to 5000 ns. From top to bottom the lifetimes recovered from the least-squares analysis were 492, 996, 1996, and 5022 ns.

Somewhat surprising results were found for τS = 5000 ns (Fig. 5, bottom panel). The phase angles appear to be “noisy.” By further simulations we found that for decay times very different from the reference lifetime, and comparable to the spacing between the pulses, there were oscillations in the phase and modulation values. This is more clearly seen in Fig. 6, which shows the result with τS = 5000 ns for a more appropriate range of frequencies and a larger number of data points. The phase and modulation values oscillate around the values expected for a single exponential decay with τS = 5000 ns. By variation of the assumed parameters we found that the amplitude and frequency of the oscillation depended on the τR, τS, tON, and tOFF. In practice, this does not seem to be a serious problem. The origin seems to mostly be truncation of the decay at t = tOFF, but the value of tON also has an effect. This effect can be minimized by selecting a pulse repetition rate and values of tON and tOFF which allow observation of most of the intensity decay of the reference and sample. In this case the oscillations are minimal, as seen in the upper three panels of Fig. 5.

FIG. 6.

FIG. 6.

Simulated frequency-domain phase and modulation data. The parameter values are the same as on Fig. 3 except for τS = 5000 ns and a pulse repetition rate of 0.01 MHz.

Effect of Incomplete Suppression

In highly scattering samples it is possible that some of the autofluorescence will be observed even with gated detection. Hence we simulated the results expected for incomplete background suppression. If the background amplitude is too large to be suppressed then the FD data will be distorted (Fig. 7). The distortion is typical of the presence of a short-lived component, which is a decreasing phase angle at high frequencies (19). For these simulations we assumed that the on/off gain ratio was 7 × 104. This value is 10-fold smaller than is readily achievable with gated PMTs (9), so that even high intensity autofluorescence can be suppressed.

FIG. 7.

FIG. 7.

Effect of incomplete suppression of the background signal due to a large amplitude of αB. The assumed on-off ratio was 7 × 104.

The autofluorescence can also distort the FD data if the gate-on time is too short compared to the decay time of the background (Fig. 8). Suppose the autofluorescence decay time is 10 ns. When the on time is delayed to 150 ns the background signal is not seen in the frequency response. As the on time is shortened to 100 or 90 ns, the presence of a short-lived component is visually evident. In practice the on time can be adjusted to longer times. It is straightforward to calculate the effect of the gating-on time on an assumed intensity decay. These calculations (Fig. 9) show that, for the assumed parameter values, the amplitude of a background with τB = 10 ns is not significant for on times larger than about 70 ns, even for a high intensity background.

FIG. 8.

FIG. 8.

Effect of incomplete background suppression because of an early gating-on time. For these simulations αB0=1000,τB=10ns,αS0=1,τS=1000ns, and the gating ratio was 7 × 104.

FIG. 9.

FIG. 9.

The dependence of observed fluorescence intensity on the on-gating time tON. B refers to a background fluorescence, S to sample, and T to total (B + S) fluorescence. The insert shows the dependence of the S/B ratio on tON. For tON = 180 ns the signal intensity is 106-fold higher than background.

Multiexponential Sample Decays

The situation is somewhat more complex if the sample displays more than a single decay time. In this case the decay times will be accurately recovered, but the time-zero amplitudes (αi0) will be distorted, more specifically, the amplitude of the shorter sample decay time to be attenuated by gating, relative to the attenuation of the longer decay time components. Following the usual normalization of the amplitudes, the apparent αi and fi values of the shorter decay components will be lower than the true time-zero value.

We simulated this effect of gating when measuring double exponential sample decays with decay times of τ1 and τ2. An intuitive presentation is shown in Fig. 10. In the absence of gating (top) the sample shows a sharp spike due to the short lifetime autofluorescence. Following this spike the log intensity plot is curved showing the presence of multiple decay times. The lower panel (Fig. 10) shows the effect of gating. The spike is removed from the sample decay. Importantly, the log intensity plot is still curved, showing that the data still contain information on the multiple decay times. We further showed that analysis of the simulated phase and modulation data with gating yields the expected decay times and amplitudes (Fig. 11).

FIG. 10.

FIG. 10.

Measurements of multiexponential intensity decay with gated detection. The intensity decay was assumed to be a double exponential with parameters τS1 = 500 ns, τS2 = 3000 ns, αS1 = 0.9, and αS2 = 0.1. The background lifetime was τB = 10 ns and amplitude αB = 100. The gating parameters were tON = 100 ns, Δt = 10 ns, tOFF = 9000 ns, on-off ratio 7 × 104, and pulse repetition 0.1 MHz. In the top and bottom panels the solid lines are the sample signal and the dashed lines the reference signals.

FIG. 11.

FIG. 11.

(Top) Simulated data for a double exponential decay with (—) and without background (– – –). (Bottom) The analysis of simulated data for a double exponential decay with background and with gated detection. The decay, background, and gating parameters are the same as in Fig. 10.

We questioned how the apparent amplitudes of the multiexponential sample decay would be affected by gating. Hence we performed simulations with a constant value of τ1 = 1000 ns. The value of τ2 was varied from 500 to 50 ns (Table 2). The detector was gated on at tON = 100 ns. The simulated data were then analyzed as usual by nonlinear least squares. As the shorter decay time was decreased the recovered amplitude (α2) also decreased below the simulated value of 0.5. For τ2 = 500 ns the amplitudes are distorted by about 1%. However, for τ2 = 100 ns the amplitude α2 is decreased to 0.20 (Table 2). Hence, correction procedures are needed if the measured decay times are comparable to the gate-on time. In practice, autofluorescence decays in about 5 ns, so gate-on times as long as 100 ns will rarely be needed.

TABLE 2.

Simulated and Recovered Parameters for Double Exponential Intensity Decays

Simulateda Recoveredb


τ1 (ns) α1 τ2 (ns) α2 τ1 (ns) α1 τ2 (ns) α2 α20 χR2

1000 0.5 500 0.5 994.7 0.509 522.4 0.491 0.487c 1.2 (10.8)d
1000 0.529 500 0.471 0.507 1.4
1000 0.5 300 0.5 1004.6 0.533 323.8 0.467 0.506 1.0 (79.6)
1000 0.553 300 0.447 0.501 1.6
1000 0.5 200 0.5 1002.1 0.592 220.1 0.408 0.513 1.3 (161.0)
1000 0.602 200 0.398 0.513 2.0
1000 0.5 150 0.5 1006.6 0.632 175.1 0.368 0.529 1.5 (183.6)
1000 0.639 150 0.361 0.515 2.5
1000 0.5 100 0.5 1002.0 0.715 117.7 0.285 0.561 1.2 (129.1)
1000 0.706 100 0.294 0.517 1.5
1000 0.5   70 0.5 982.2 0.807 70.3 0.193 0.560 1.0 (40.0)
1000 0.796 70 0.204 0.541 1.4
1000 0.5   50 0.5 996.7 0.910 90.8 0.090 0.804 0.9 (10.2)
1000 0.870 50 0.130 0.548 1.1
a

For simulations the various parameters were as follows: pulse repetition rate 0.07 MHz, τB = 10 ns, αB0=100,τR = 1000 ns, tON = 100 ns, tOFF = 10,000 ns. Δt = 10 ns, on-off ratio 7 × 104. Background was presented only in the sample. The Gaussain noise in the simulations was δϕ=0.3° and δm = 0.007.

b

The values in brackets were held fixed at the indicated values during least-squares analysis.

c

The value of amplitude at t = 0, calculated from Eq. [20].

d

Values for one-exponential fit.

Fortunately, it is relatively easy to correct the distorted amplitudes. The basic idea is to use the recovered decay times and amplitudes to extrapolate back to t = 0. This extrapolation is contained in Eq. [20]. The corrected values α20 are listed in Table 3. The calculated values of α20 are all accurately recovered except for τ2 values shorter than the gate-on time (70 and 50 ns in Table 2). In these cases the contribution of the short component to the data is not adequate to allow reliable recovery of α20.

TABLE 3.

Recovered Decay Parameters of a Double Two-Exponential Intensity Decay with Various Values of the On-Gating Timea

Recovered parameters

tON
(ns)
τ1
(ns)
τ2
(ns)
α1 α10

2000 441.1 2994.8 0.234b 0.929c
500 3000 0.248 0.893
1500 459.2 2943.6 0.390 0.903
500 3000 0.410 0.886
1000 466.8 2770.2 0.585 0.888
500 3000 0.619 0.889
  700 465.3 2613.7 0.696 0.883
500 3000 0.741 0.896
  500 482.2 2640.8 0.755 0.873
500 3000 0.787 0.889
  300 494.9 2724.7 0.833 0.887
500 3000 0.849 0.898
  100 501.1 2929.5 0.873 0.885
500 3000 0.876 0.888
a

For these simulations the various parameters were as follows: pulse repetition rate 0.01 MHz, τB = 10 ns, αB = 100, τR = 1000 ns, αR = 1, τ1 = 500 ns, α10=0.9,τ2 = 0.1, tOFF = 9.000, Δt = 10 ns, on-off ratio = 7 × 104.

b

α2 = 1 – α1.

C

The value of amplitude at t = 0, calculated from Eq. [20].

We also considered a double exponential sample decay with τ1 = 500 ns and τ2 = 3000 ns. In this case we varied the gate-on time from 100 to 2000 ns. As the gate-on time becomes longer the amplitudes of α1 decreased (Fig. 11). With gate-on times as long as 2000 ns we reliably recovered the amplitudes of the two decay times (Table 3). This suggests the procedure is somewhat robust if at least one of the decay times is longer than the gating time. Alternatively one can measure the intensity decays for several gate-on times and graphically extrapolate the αi values to t = 0 (Fig. 12).

FIG. 12.

FIG. 12.

Dependence of short component amplitude α1 on gating start time tON. The decay, background, and gating parameters are the same as in Figs. 10 and 11.

DISCUSSION

One may question why this method of frequency-domain autofluorescence suppression is presented without experimental verification. There is a growing need for such a procedure, but we are not planning to construct the needed electronics within the foreseeable future. During the past 15 years we have performed extensive simulations of FD data, for a wide variety of intensity decay models. In all cases the simulations matched our experimental data. Hence we are confident that the current simulations accurately predict the performance of FD measurements with real-time background suppression.

While we have not yet constructed the electronics for FD background suppression, this is not a difficult task. Pulsed laser sources are now routinely used for FD measurements (2224), particularly since the recent interest in multiphoton excitation (33, 34). Also, it is now known that light-emitting diodes (LEDs) can be modulated at frequencies in excess of 100 MHz (35, 36). Also, it is known that laser diodes can give pulse widths of 50 ps or less, and LEDs can provide pulse widths less than 2 ns (37). Hence it appears that FD background suppression can be accomplished with simple light sources and electronics.

FD background suppression is needed in a variety of analytical and clinical applications of time-resolved fluorescence. For instance, phase-modulation measurements with long-lived metal-ligand complexes are being developed for use in resonance energy transfer immunoassays, measurements of blood electrolytes and gases, bioprocess monitoring, and in high-throughput screening for drug discovery. In all these applications it would be valuable to make use of the high sensitivity of gated detection with the robustness of phase-modulation fluorometry. We believe that our description of background suppression will result in its near term use of gated FD measurements in these important applications.

ACKNOWLEDGMENTS

This work was supported by the National Institutes of Health, National Center for Research Resources, RR-08119 and NIGMS GM-35154.

Footnotes

2

Abbreviations used: FD, frequency-domain; TD, time-domain; bpy, 2,2′-bipyridyl; MLC, metal-ligand complex; PMT, photomultiplier tube; LED, light-emitting diodes.

REFERENCES

  • 1.Wolfbeis OS (Ed.) (1992) Proceedings of the 1st European Conference on Optical Chemical Sensors and Biosensors, Europ-t(R)ode I, Sensors and Actuators B, pp. 565. [Google Scholar]
  • 2.Prober JM, Trainor GL, Dam RJ, Hobbs FW, Robertson CW, Zagursky RJ, Cocuzza AJ, Jensen MA, and Baumeister K (1987) A system for rapid DNA sequencing with fluorescent chain-terminating dideoxynucleotides. Science 238, 336–343. [DOI] [PubMed] [Google Scholar]
  • 3.Speicher MR, Ballard SG, and Ward DC (1996) Karyotyping human chromosomes by combinatorial multi-fluor FISH. Nature Genet. 12, 368–378. [DOI] [PubMed] [Google Scholar]
  • 4.Shapiro HM (1988) Practical Flow Cytometry, 2nd ed., pp. 353 Liss R, New York. [Google Scholar]
  • 5.Pawley JB (Ed.) (1995) Handbook of Biological Confocal Microscopy, pp. 632 Plenum, New York. [Google Scholar]
  • 6.Wang XF, and Herman B (1996) Fluorescence Imaging Spectroscopy and Microscopy, pp. 483 Wiley, New York. [Google Scholar]
  • 7.Nederlof PM, van der Flier S, Wiegant J, Raap AK, Tanke HJ, Ploem JS, and van der Ploeg M (1990) Multiple fluorescence in situ hybridization. Cytometry 11, 126–131. [DOI] [PubMed] [Google Scholar]
  • 8.Schober A, Gunther R, Tangen U, Goldmann G, Ederhof T, Koltermann A, Wienecke A, Schwienhorst A, and Eigen M (1997) High throughput screening by multichannel glass fiber fluorimetry. Rev. Sci. Instrum. 68, 2187–2194. [Google Scholar]
  • 9.Barisas BG, and Leuther MD (1980) Grid-gated photomultiplier photometer with subnanosecond time response. Rev. Sci. Instrum. 51, 74–78. [Google Scholar]
  • 10.Hanselman D, Withnell R, and Hieftje GM (1991) Side-on photomultiplier gating system for thomson scattering and laser-excited atomic fluorescence spectroscopy. Appl. Spectrosc. 45, 1553–1560. [Google Scholar]
  • 11.Dickson EFG, Pollak A, and Diamandis EP (1995) Ultrasensitive bioanalytical assays using time-resolved fluorescence detection. Pharmacol. Ther. 66, 207–235. [DOI] [PubMed] [Google Scholar]
  • 12.Demas JN, and DeGraff BA (1992) Applications of highly luminescent transition metal complexes in polymer systems. Macromol. Chem. Macromol. Symp. 59, 35–51. [Google Scholar]
  • 13.DeGraff BA, and Demas JN (1994) Direct measurement of rotational correlation times of luminescent ruthenium (II) molecular probes by differential polarized phase fluorometry. J. Phys. Chem. 98, 12478–12480. [Google Scholar]
  • 14.Terpetschnig E, Szmacinski H, and Lakowicz JR (1997) Long-lifetime metal-ligand complexes as probes in biophysics and clinical chemistry. Methods Enzymol. 278, 295–321. [DOI] [PubMed] [Google Scholar]
  • 15.Szmacinski H, and Lakowicz JR (1994) in Topics in Fluorescence Spectroscopy, Vol. 4, Probe Design and Chemical Sensing (Lakowicz JR, Ed.), pp. 295–334. Plenum Press, New York. [Google Scholar]
  • 16.Lakowicz JR, and Szmacinski H (1992) Fluorescence lifetime-based sensing of pH, Ca2+, K+ and glucose. Sensors Actuators B 11, 133–143. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Guo X, Li L, Castellano FN, Szmacinski H, and Lakowicz JR (1997) A long-lived, high luminescent rhenium (I) metal-ligand complex as a bimolecular probe. Anal. Biochem. 254, 179–186. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Guo X-Q, Castellano FN, Li L, and Lakowicz JR (1998) Along-lifetime Ru(II) metal-ligand complex as a membrane probe. Biophys. Chem. 71, 51–62. [DOI] [PubMed] [Google Scholar]
  • 19.Lakowicz JR, Jayaweera R, Joshi N, and Gryczynski I (1987) Correction for contaminant fluorescence in frequency-domain fluorometry. Anal. Biochem. 160, 471–479. [DOI] [PubMed] [Google Scholar]
  • 20.Reinhart GD, Marzola P, Jameson DM, and Gratton E (1991) A method for on-line background subtraction in frequency domain fluorometry. J. Fluoresc. 1, 153–162. [DOI] [PubMed] [Google Scholar]
  • 21.Berndt K, Duerr H, and Palme D (1982) Picosecond phase fluorometry by mode-locked CW lasers. Opt. Commun. 42, 419–422. [Google Scholar]
  • 22.Gratton E, and Lopez-Delgado R (1980) Measuring fluorescence decay times by phase-shift and modulation techniques using the high harmonic content of pulsed light sources. Nuovo Cimento B56, 110–124. [Google Scholar]
  • 23.Lakowicz JR, Laczko G, and Gryczynski I (1986) 2-GHz frequency-domain fluorometer. Rev. Sci. Instrum. 57, 2499–2506. [Google Scholar]
  • 24.Laczko G, Gryczynski I, Gryczynski Z, Wiczk W, Malak H, and Lakowicz JR (1990) A 10-GHz frequency-domain fluorometer. Rev. Sci. Instrum. 61, 2331–2337. [Google Scholar]
  • 25.Lakowicz JR, and Cherek H (1981) Phase-sensitive fluorescence spectroscopy. A new mixture to resolve fluorescence lifetimes or emission spectra of components in a mixture of fluorophores. J. Biochem. Biophys. Methods 5, 19–35. [DOI] [PubMed] [Google Scholar]
  • 26.Lakowicz JR, Szmacinski H, Nowaczyk K, and Johnson ML (1992) Fluorescence lifetime imaging of free and protein-bound NADH. Proc. Natl. Acad. Sci. USA 89, 1271–1275. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Szmacinski H, Lakowicz JR, and Johnson ML (1994) in Methods in Enzymology, pp. 723–748. Academic Press, New York. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Lakowicz JR, Cherek H, and Balter A (1981) Correction of timing errors in photomultiplier tubes used in phase-modulation fluorometry. J. Biochem. Biophys. Methods 5, 131–146. [DOI] [PubMed] [Google Scholar]
  • 29.Barrow DA, and Lentz BR (1983) The use of isochronal reference standards in phase and modulation fluorescence lifetime measurements. J. Biochem. Biophys. Methods 7, 217–234. [DOI] [PubMed] [Google Scholar]
  • 30.Birch DJS, and Imhof RE (1991) in Topics in Fluorescence Spectroscopy: Techniques (Lakowicz JR, Ed.), Vol. 1, pp. 1–95. Plenum Press, New York. [Google Scholar]
  • 31.Lakowicz JR, and Gryczynski I (1991) in Topics in Fluorescence Spectroscopy, Vol. 1, Techniques (Lakowicz JR, Ed.), pp. 293–355. Plenum Press, New York. [Google Scholar]
  • 32.Lakowicz JR, Laczko G, Cherek H, Gratton E, and Limke-man M (1984) Analysis of fluorescence decay kinetics from variable-frequency phase shift and modulation data. Biophys. J. 46, 463–477. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Lakowicz JR, and Gryczynski I (1997) in Topics in Fluorescence Spectroscopy, Vol. 5, Nonlinear and Two-Photon Induced Fluorescence (Lakowiczz JR, Ed.), pp. 87–144. Plenum Press, New York. [Google Scholar]
  • 34.Watkins AN, Ingersoll CM, Baker GA, and Bright FV (1998) A parallel multiharmonic frequency-domain fluorometer for measuring excited-state decay kinetics following one-, two-, and three-photon excitation. Anal. Chem. 70, 3384–3396. [DOI] [PubMed] [Google Scholar]
  • 35.Sipior J, Carter GM, Lakowicz JR, and Rao G (1996) Single quantum well light emitting diodes demonstrated as excitation sources for nanosecond phase-modulation fluorescence lifetime measurements. Rev. Sci. Instrum. 67, 3795–3798. [Google Scholar]
  • 36.Sipior J, Carter GM, Lakowicz JR, and Rao G (1997) Blue light emitting diodes demonstrated as an ultraviolet excitation source for nanosecond phase-modulation fluorescence lifetime measurements. Rev. Sci. Instrum. 68, 2666–2670. [Google Scholar]
  • 37.IBH, Inc., (1999) Nano LED pulsed light source, product literature. Glasgow, UK. [Google Scholar]

RESOURCES