Abstract
One of the major challenges during polarimetric determination of glucose concentration is the spectral superposition with other optically active molecules, especially proteins like albumin. Since each of those substances has a characteristic optical rotatory dispersion (ORD), we developed a broadband polarimeter setup to distinguish between glucose and albumin. A partial least squares (PLS) regression with 5 components was applied to the polarimeter signal in the wavelength range of . To verify the efficacy of the proposed method, different glucose levels of were spiked with varying albumin concentrations up to . A standard error of prediction of was achieved compared to using a two-wavelength system with and under the same conditions.
1. Introduction
The proportion of people afflicted with diabetes in the adult population worldwide nearly doubled from in 1980 to in 2014 [1]. Delayed diagnosis or wrong treatments of the disease can lead to serious damage to the eyes, nerves, heart, kidneys and blood vessels [2], and in extreme cases may require limb amputation [1]. Therefore, early stage diagnosis and treatment are crucial to minimize health complications [1]. In this context, a number of research papers have reported various approaches for continuous glucose monitoring (CGM). These include optical coherence tomography (OCT), fluorescence [3], refractometry [4] and Raman spectroscopy [5,6].
A large portion of optical-based blood component analysis is represented by absorbance spectroscopy, including mid-infrared (MIR) and near-infrared (NIR) wavelength ranges. Approaches utilizing liquid nitrogen cooled Fourier transform infrared spectrometers (FTIRs) [7,8] and tunable quantum cascade lasers (QCL) [9,10] exploit the MIR absorbance spectra of glucose. However, NIR and MIR measurements suffer from the hardware’s low sensitivity in this wavelength range and low absorbance, overlapped by strong absorption bands of water and other blood components such as proteins, urea or lactate [11–13]. Consequently, Partial least squares (PLS) regression has been widely explored and demonstrated to be able to separate overlapped absorbance spectra of several blood components [6,11,14].
Polarimetry is another optical measurement method suitable for glucose determination in fluids. This technique takes advantage of glucose’s optical rotation of polarized light, that depends on the concentration, wavelength, path length, temperature and pH-value [15]. The challenges of this method are the scattering effects caused by erythrocytes and tissues [16], as well as the low optical rotation values (approximately millidegrees per glucose at a wavelength of for a sample cell with path length [17]). Another big obstacle is the high protein concentration, approximately [18], which impedes a reliable glucose prediction due to its optical activity [16]. Therefore, the aqueous humor of the eye has been repeatedly proposed for glucose measurements, as the influence of albumin [19] on the polarization signal can be neglected, and no erythrocytes are present to disturb polarization measurements by scattering. Nonetheless, time delays between [20] and [21] have been reported for glucose levels in blood as compared to measurements in the eye for anesthetized rabbits [22], which were under ideal laboratory conditions that prove to be unrealistic in a clinical application. A two-wavelength system with and was investigated by Cameron et. al [19,23] to determine glucose in the presence of varying albumin concentrations. This showed a significant improvement over single wavelength systems and enabled the polarimetric distinction of glucose and albumin for in vitro measurements. However, the authors used albumin levels between that are far below the physiological variation, which is reported to be [18,24]. Since the optical rotatory dispersions (ORD) of different optically active blood components are quite similar, multispectral regression models have been applied to help distinguish between different analytes [15,25,26]. In this paper we show that a broadband polarimeter in combination with PLS regression is able to reliably determine glucose in the presence of physiologically varying albumin.
With the goal of a long-term stable and robust polarimetry for medical applications, we already demonstrated that our setup is able to measure independent from fluctuations in absolute light intensity and slight scattering effects [27]. Since the spectral ORD overlap of glucose with other substances remains a major source of error, we present a broadband polarimeter setup, similar to the system proposed by Cote et al. [28]. The aim is to achieve a better signal separation between glucose and the most abundant blood protein albumin, as well as a substantial improvement of the prediction accuracy. Such a system could potentially be used beyond in-vitro diagnostics. For example, an application in intensive care settings would be feasible. In this case, blood plasma could be derived from the patient using a miniaturized erythrocyte filter to implement a continuous glucose measurement. A long-term implementation might also be an implantable system for diabetes patients.
2. Theoretical background
2.1. Optical activity of blood components
The optical activity of a given substance is the result of its molecules chirality, which can exist with different spatial arrangements but equal molecular weights. The ORD can be calculated for the wavelength using Drude’s equation [15,18]:
| (1) |
where defines the strength and the direction of rotation, and represents the absorption center wavelength [16], which is unique to each chiral component [29]. The ORD describes the total rotation at a specific temperature , pH-value and wavelength . At constant and pH, the absolute sample rotation for a given path length is proportional to the analyte concentration :
| (2) |
Table 1 shows the parameters and for glucose and the most relevant blood proteins. According to the literature [18,24], the typical physiological protein concentrations are for albumin, for globulin and for fibrinogen.
Table 1. Coefficient defining strength and direction of rotation and center wavelength for the most prevalent blood proteins and glucose [18,24].
| Analyte | A in | in |
|---|---|---|
| Glucose | 150 | |
| Albumin | 264 | |
| Globulin | 211 | |
| Fibrinogen | 260 |
Glucose exhibits the shortest absorption center wavelength at , while the longest corresponds to albumin at . Figure 1 shows the ORD for glucose, albumin, globulin and fibrinogen in blood as well as the ORD ratio for proteins referenced to glucose to show their linear independence. The spectral differences become larger at shorter wavelengths, especially in the UV range.
Fig. 1.
(left) ORD of common blood components glucose, albumin, globulin and fibrinogen [18]. (right) Ratio of ORD for proteins referenced to glucose. The spectral differences between glucose and proteins become larger the closer the measurement approaches the UV located absorption center wavelength.
Polarization measurements in the UV region comprise several obstacles. One is a high water absorption below approximately . Another is that most polarimeters for glucose determination usually require an optical modulation by a Faraday rotator [17,19,22,30]. The high absorption below of most optical glasses used in these Faraday rotators as well as the strong absorption of the used Glan polarizers, which absorb strongly below , limit the effective wavelength. The ORDs for glucose, proteins and other optically active molecules are very close to each other for wavelengths in the visible range (see Fig. 1). Therefore, ambient conditions and setup parameters need to be very stable during the measurements. Another obstacle is the additional absorbance of proteins below , which leads to a variation in sample transmittance.
3. Material and methods
3.1. Measurement setup
To acquire the spectra of spectrally superimposed ORD, we built a broadband polarimeter as illustrated in Fig. 2 that includes a broadband light source, two crossed polarizers, a cuvette, a spectrometer and a Faraday rotator for modulation.
Fig. 2.
Measurement setup: (1) Light source, (2) First polarizer, (3) Faraday rotator, (4) Sample cell, (5) Second Polarizer, (6) Spectrometer, (7) Data processing.
In order to perform highly sensitive broadband measurements, a high power light source with a continuous emission is required. In this case, a laser driven light source (EQ-99X, Energetiq Inc., USA) was chosen, providing a broad and continuous emission between and with a typical output power of . The light was linearly polarized by a Glan-Thompson Polarizer (GTH10, Thorlabs GmbH, Germany) with an extinction ratio of over a broad wavelength range of to .
For signal improvement, the light from the first polarizer was then modulated by a Faraday rotator. The rotator consists of a SF6-rod (Schott GmbH, Germany). The material was chosen due to its high and continuous transmission above because commonly used terbium gallium garnet (TGG) rotators show an increased absorption around [31]. The length of originates from stock availability of the manufacturer. UV transparent -BBO polarizers are available, but they are hygroscopic, which limits their long term stability. For this reason, they were not used in this setup. The rotator’s solenoid has turns and was driven at by a current driver. The reference signal for the coil driver was created by a function generator (33120A, Hewlett Packard, USA) with a frequency stability of and . A modulation frequency of was chosen in order to avoid any overlap with harmonics of . The current amplitude fluctuations were below ().
A flow-through cuvette for the sample was placed behind the Faraday rotator. Since a greater path length leads to an enhanced rotation, a path length of was used to better resolve ORDs from glucose and albumin as already successfully used in previous studies [19,26]. A second Glan-Thompson polarizer (GTH10, Thorlabs GmbH, Germany) was placed behind the cuvette in perpendicular orientation to the first one.
Measuring a modulated multispectral signal requires a spectrometer with a low timing jitter in order to identify frequency components of the transmitted intensity. For this purpose, the light transmitted by the second polarizer was measured by a grating spectrometer (AvaSpec-ULS2048L-EVO UV-VIS, Avantes BV, Netherlands), which has a fast internal memory to ensure a proper synchronization between time stamps and intensity. The spectrometer mean timing jitter was determined to be at an integration time of . A partial least squares (PLS) regression was applied on the spectral data using MATLAB 2019a (Mathworks Inc., USA) to distinguish between glucose and albumin in the wavelength domain. During the measurements, the whole setup was kept inside a temperature-controlled chamber where fluctuations stayed below . An average temperature of was chosen to maximize the protein stock solution life time.
3.2. Data processing
Based on the sinusoidal Faraday modulation at frequency , the detected signal has to be processed to obtain its respective frequency components. The resulting signal consists of three frequency components [23,27]. One DC-part, one at the modulation frequency and the last at the double frequency . Using the modulation depth , the sample rotation , modulation frequency , incoming light intensity and transmission , the measured intensity at a certain time can be calculated as
| (3) |
where the frequency component amplitudes are:
-
•
-
•
-
•
As demonstrated in a previous work [27], all frequency components in Eq. (3) are proportional to the absolute light intensity. Thus, the ratio
| (4) |
is independent from any kind of variations in light intensity. Using the ratio of Eq. (4) influences of the light source intensity, sample absorbance and slight scattering are greatly reduced, improving the reproducibility and robustness of polarimeter-setups without the need of additional hardware or reference measurements [27]. Therefore, this method was used in the present work. The spectral ORD overlap of multiple optically active substances in a sample remains as the main source of error, which is the focus of the investigation in this paper.
3.3. Sample preparation
For all measurements used in this work, samples of glucose and albumin in distilled water were prepared from stock solutions. The pH value was stabilized at using phosphate buffer (Morphisto GmbH, Germany). of D-glucose powder (Sigma Aldrich, Germany) were weighted with a high precision balance (Kern EW 120-4NM, Kern & Sohn GmbH, Germany) with an uncertainty of and then diluted in of distilled water. This resulted in a glucose concentration of . The stock solution uncertainty was . To achieve full mutarotation, the stock solution was kept at for , which has been reported to be sufficient [25], before the measurements were carried out.
In order to determine the effect of confounding proteins, albumin was chosen to model interference since it is the most abundant protein in blood [18]. Due to its similarity [29], bovine serum albumin (BSA) (Sigma Aldrich, Germany) was used for the samples and other polarimetric studies [17]. One liter of stock solution was created for each respective measurement series by adding BSA to water, which resulted in a concentration uncertainty of .
Samples with different concentrations of glucose and albumin were automatically generated by mixing the stock solutions using three high precision syringe pumps (neMESYS 290N, Cetoni GmbH, Germany) with three glass syringes (ILS GmbH, Germany). A sample volume of was chosen in order to keep contamination with prior sample remains as small as possible. After injecting the solutions into the cuvette, the syringes were automatically refilled. Additionally, the cuvette was rinsed with of distilled water after each measurement. Measurements were performed after the flow had stopped to ensure stationary conditions [27]. The stock solutions and syringe pumps were also maintained inside the temperature controlled chamber at .
The syringe pump accuracy was evaluated gravimetrically using distilled water and dosing volumes of , which was the smallest volume used in the sample generation. As result, an absolute combined uncertainty of of delivered volume (bias) + (precision) was determined.
3.4. Measurement procedure
With the purpose to investigate the influence of albumin concentration on glucose predictability, two sets of measurements were carried out as summarized in Table 2.
Table 2. Overview of performed measurements. Each glucose level was measured 6 times for calculation of standard deviations within measurement . For , each glucose concentration was combined with a different albumin level resulting in a 6x6 matrix. Each of these cases was measured 3 times.
| Glucose |
Albumin |
|||||
|---|---|---|---|---|---|---|
| Set No. | Matrix dimensions | Repetitions | Range in | Increments in | Range in | Increments in |
| 6x6 | 3 | |||||
| 6x6 | 3 | |||||
We used six different glucose levels () measured 6 times each. To evaluate the influence of albumin, each glucose solution was spiked with albumin, leading to a concentration matrix with 36 unique combinations. These samples were measured in a random order to avoid time-correlated effects, like for example those caused by temperature fluctuations, as described and suggested by other authors [17,25,32]. For each measurement, the broadband intensities and were extracted from the raw spectra and the ratio was used to obtain stabilized measurements of spectrally superimposed ORD as described in subsection 3.2. This procedure was repeated three times, resulting in three identical 6x6 matrices and a total of 108 measurements. The measurement set was later split in two groups: one for calibration and the other for validation datasets.
Previous works have shown that glucose concentration in the presence of albumin can be determined with a precision of utilizing a dual-wavelength system [19]. However, the authors used an albumin concentration range representing only of the typical physiological variation () and a maximum concentration of (), which was lower than the reported blood levels ( [18,24]). Since the absolute rotation depends linearly on the concentration (see Eq. (2)), an albumin level offset introduces a constant shift of the rotation signal, which is eliminated by the regression model. As demonstrated earlier, absorption effects which may be caused by albumin or other substances are compensated for by using according to Eq. (4) [27]. Consequently, we focused on the physiological variation in order to investigate the ORD interference of glucose and albumin. In this work, albumin concentrations between and were used.
In measurement the system prediction accuracy without any confounder (i.e. no albumin) was evaluated. For measurement albumin concentrations of were added to the matrix.
For each sample, 1000 spectra were recorded at an integration time of and saved to the internal memory of the spectrometer. The frequency components and were extracted at () and () for each wavelength by applying a Fourier transform. The ratio was then used to obtain the corrected ORD dependent spectra, as described in subsection 3.2.
The results were analyzed using a multivariate regression model, as suggested by other authors [15,24,26] to distinguish between different optically active molecules. For each measurement series, the data from the 3 concentration matrices (36 samples each) was split into calibration and validation data. The first 2 data sets (72 spectra) were used for calibration, while the last one (36 spectra) was used for validation. For the model calibration, a PLS regression was performed on the of corresponding data sets using Matlab 2019a (statistics toolbox). The created regression model was then applied to the validation data to obtain the corresponding glucose concentrations. The deviation between predicted and reference concentration was computed as the standard error of prediction (SEP).
4. Results and discussion
4.1. Glucose determination without albumin
We first evaluated the system’s overall glucose prediction accuracy without albumin to later classify the results for varying confounding albumin levels in subsection 4.2. For both cases, the results for a broadband determination were compared with those obtained by a two-wavelength prediction with and as used by other authors [19,33,34]. For this purpose, we extracted the intensities at these wavelengths from the same data sets as used for the broadband determination.
The results for pure glucose solutions in the absence of albumin can be found in Fig. 3 for dual-wavelength analysis at and as well as for broad range.
Fig. 3.
Prediction of glucose within the range of in the absence of albumin with PLS regression. (left) using and , (right) using broadband spectra between with 5 components.
The best prediction accuracy was achieved with 5 PLS components and a range of . Since no confounders were used in this measurement and only the glucose concentration had a major contribution to the signal, broadband and dual-wavelength range deliver similar results ( vs. ), which is about one third of the reported in the literature for a similar scenario [25]. The broadband data performs slightly better, probably due to the ORD being identified, which provides for distinguishing the sample signal from minor disturbance effects like drift.
4.2. Glucose determination with confounding albumin
In measurement #2, the glucose samples were spiked with albumin concentrations between . Consequently, the ORDs of glucose and albumin are superimposed. The ratio is therefore depicted in Fig. 4 for and as well as for broadband wavelength data. The combination of 6 different clockwise rotating glucose concentrations with 6 different counterclockwise rotating albumin concentrations within the 6x6 matrix leads to 36 unique overlapped ORDs. As shown in Fig. 1, the broadband data provides ORD-based spectral information, which was used in the past to distinguish between glucose and albumin with a two-wavelength system [19,33,34].
Fig. 4.
Raw signal for the glucose and albumin matrix from measurement #2. (left) for and , (right) for .
The influence of confounding albumin on the prediction accuracy was then evaluated as for the pure glucose solutions. The comparison for broadband and dual-wavelength evaluation can be found in Fig. 5.
Fig. 5.
Glucose prediction in the range of randomly mixed with albumin in the range . (left) The dual-wavelength data analysis at and (right) Results of the broadband analysis.
Compared to the results shown in Fig. 3, the SEP increases significantly to when using dual-wavelength analysis in the presence of albumin. The reason is that both, glucose and albumin, have considerable contributions to the sample’s overall rotation. Dual-wavelength analysis is partially able to distinguish between them, but due to the low amount of spectral information only in a limited way. Under consideration of the albumin concentration range, the achieved accuracy of for corresponds to those of Cameron et. al, who reported for a tenth of our albumin level () [19].
The prediction with the broadband data from performs much better and still delivers an SEP of . Since glucose and albumin have their own characteristic ORD as shown in Fig. 1, the ORD overlap of glucose and albumin generates unique combinations of their ORDs depending on their concentrations. These superimposed ORDs can then be identified by the PLS evaluated broadband data. This substantiates the results obtained by Cameron et. al [19] who achieved a prediction accuracy of utilizing a two-wavelength system. However, the confounding albumin concentration range in our investigations was 10 times higher. A summary of the results is given in Table 3.
Table 3. Standard error of prediction (SEP) and regression coefficient of measurement #2 from the acquired data set for the extracted laser wavelengths 532 nm and 635 nm as well as for the broadband range.
| Albumin range in | Prediction accuracy | Wavelength range |
|
|---|---|---|---|
| SEP in | |||
|
| |||
| SEP in | |||
5. Conclusions
Since the fluctuations in absolute light intensity were already compensated for by using [27], the spectral overlap of other optically active substances, especially proteins remained as a major source of error for glucose prediction. To overcome this obstacle, we built a broadband polarimeter as suggested by Baba, Clarke and King [15,25,26] and applied a PLS regression on the extracted spectrally superimposed ORD. In order to investigate the influence of proteins as confounder, glucose was combined with different levels of albumin, the most abundant protein in blood. To compare these results with those for dual-wavelength analysis from other studies, we extracted the intensity at and from the broadband data.
We showed that the prediction accuracy via PLS regression and 5 components between without albumin is similar to that obtained by the dual-wavelength analysis at and and is as low as . By adding varying albumin concentrations in the range , the prediction error increases to for broadband data, while the dual-wavelength analysis leads to an extended SEP of .
We showed, that a PLS regression applied on broadband rotation data is able to distinguish between different optically active substances by their characteristic superimposed ORD. This could be used for a substantially improved glucose prediction in the presence of confounding albumin.
Although we showed that the method leads to an improvement in glucose prediction, the influence of other optically active substances like globulin and fibrinogen must and will be investigated in the future in order to estimate how the broadband ORD based method would perform for blood plasma samples. For later application as an implantable sensor for continuous glucose monitoring the system needs to be miniaturized including a reduction of cuvette size for less sample volume and shielded from further potential interferences like mechanical torsion of the setup.
Acknowledgments
The authors would like to thank Dr. Christian Damiani for manuscript editing.
Disclosures
None of the authors has relevant financial interests in this manuscript.
References
- 1.Chan D. M., “Global Report on Diabetes,” Tech. rep., World Health Organization, ISBN 978 92 4 156525 7 (2016). [Google Scholar]
- 2.American Diabetes Asssociation , “American Diagnosis and Classification of Diabetes Mellitus,” Diabetes Care 33(Supplement_1), S62–S69 (2010). 10.2337/dc10-S062 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Esposito R., Ventura B. D., Nicola S. D., Altucci C., Velotta R., Mita D. G., Lepore M., “Glucose sensing by time-resolved fluorescence of sol-gel immobilized glucose oxidase,” Sensors 11(4), 3483–3497 (2011). 10.3390/s110403483 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Zirk K., Poetzschke H., “A refractometry-based glucose analysis of body fluids,” J. Biomed. Eng. 29(4), 449–458 (2007). 10.1016/j.medengphy.2006.02.011 [DOI] [PubMed] [Google Scholar]
- 5.Rohleder D., Kocherscheidt G., Gerber K., Kiefer W., Köhler W., Möcks J., Petrich W., “Comparison of mid-infrared and raman spectroscopy in the quantitative analysis of serum,” J. Biomed. Opt. 10(3), 031108 (2005). 10.1117/1.1911847 [DOI] [PubMed] [Google Scholar]
- 6.Arnold M. A., Liu L., Olesberg J. T., “Selectivity assessment of noninvasive glucose measurements based on analysis of multivariate calibration vectors,” J. Diabetes Sci. Technol. 1(4), 454–462 (2007). 10.1177/193229680700100402 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Petibois C., Melin A.-M., Perromat A., Cazorla G., Déléris G., “Glucose and lactate concentration determination on single microsamples by fourier-transform infrared spectroscopy,” J. Lab. Clin. Med. 135(2), 210–215 (2000). 10.1067/mlc.2000.104460 [DOI] [PubMed] [Google Scholar]
- 8.Liakat S., Bors K. A., Huang T.-Y., Michel A. P. M., Zanghi E., Gmachl C. F., “In vitro measurements of physiological glucose concentrations in biological fluids using mid-infrared light,” Biomed. Opt. Express 4(7), 1083 (2013). 10.1364/BOE.4.001083 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Brandstetter M., Volgger L., Genner A., Jungbauer C., Lendl B., “Direct determination of glucose, lactate and triglycerides in blood serum by a tunable quantum cascade laser-based mid-IR sensor,” Appl. Phys. B: Lasers Opt. 110(2), 233–239 (2013). 10.1007/s00340-012-5080-z [DOI] [Google Scholar]
- 10.Yu S., Li D., Chong H., Sun C., Yu H., Xu K., “In vitro glucose measurement using tunable mid-infrared laser spectroscopy combined with fiber-optic sensor,” Biomed. Opt. Express 5(1), 275 (2014). 10.1364/BOE.5.000275 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Hazen K. H., Arnold M. A., Small G. W., “Measurement of glucose and other analytes in undiluted human serum with near-infrared transmission spectroscopy,” Anal. Chim. Acta 371(2-3), 255–267 (1998). 10.1016/S0003-2670(98)00318-3 [DOI] [Google Scholar]
- 12.Jensen P. S., Bak J., “Near-infrared transmission spectroscopy of aqueous solutions: Influence of optical pathlength on signal-to-noise ratio,” Appl. Spectrosc. 56(12), 1600–1606 (2002). 10.1366/000370202321115878 [DOI] [Google Scholar]
- 13.LaFrance D., “Near infrared determination of lactate in biological fluids and tissues,” Ph.D. thesis, McGill University; (2003). [Google Scholar]
- 14.Ren M., Arnold M. A., “Comparison of multivariate calibration models for glucose, urea, and lactate from near-infrared and raman spectra,” Anal. Bioanal. Chem. 387(3), 879–888 (2007). 10.1007/s00216-006-1047-4 [DOI] [PubMed] [Google Scholar]
- 15.Baba J. S., Cameron B. D., Theru S., Cote G. L., “Effect of temperature, pH, and corneal birefringence on polarimetric glucose monitoring in the eye,” J. Biomed. Opt. 7(3), 321 (2002). 10.1117/1.1484163 [DOI] [PubMed] [Google Scholar]
- 16.Tuchin V. V., Wang L., Zimnyakov D. A., Optical Polarization in Biomedical Applications, vol. 2 of 5–16 (Springer-Verlag Berlin Heidelberg, 2006). [Google Scholar]
- 17.Cameron B. D., Cote G. L., “Noninvasive glucose sensing utilizing a digital closed-loop polarimetric approach,” IEEE Trans. Biomed. Eng. 44(12), 1221–1227 (1997). 10.1109/10.649993 [DOI] [PubMed] [Google Scholar]
- 18.Wood M. F. G., Cote D., Vitkin I. A., “Combined optical intensity and polarization methodology for analyte concentration determination in simulated optically clear and turbid biological media,” J. Biomed. Opt. 13(4), 044037 (2008). 10.1117/1.2968198 [DOI] [PubMed] [Google Scholar]
- 19.Cameron B. D., Baba J. S., Cote G. L., “Optical polarimetry applied to the development of a noninvasive in-vivo glucose monitor,” in “Optical Diagnostics of Biological Fluids V,”, vol. 66 Priezzhev A. V., Asakura T., eds. (SPIE, 2000) vol. 66, pp. 66–77. [Google Scholar]
- 20.Cameron B. D., Baba J. S., Coté G. L., “Measurement of the glucose transport time delay between the blood and aqueous humor of the eye for the eventual development of a noninvasive glucose sensor,” Diabetes Technol. Ther. 3(2), 201–207 (2001). 10.1089/152091501300209552 [DOI] [PubMed] [Google Scholar]
- 21.Chou C., Han C.-Y., Kuo W.-C., Huang Y.-C., Feng C.-M., Shyu J.-C., “Noninvasive glucose monitoring in vivo with an optical heterodyne polarimeter,” Appl. Opt. 37(16), 3553–3557 (1998). 10.1364/AO.37.003553 [DOI] [PubMed] [Google Scholar]
- 22.Purvinis G., Cameron B. D., Altrogge D. M., “Noninvasive polarimetric-based glucose monitoring: An in vivo study,” J. Diabetes Sci. Technol. 5(2), 380–387 (2011). 10.1177/193229681100500227 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Cameron B. D., “The Application of Polarized Light to Biomedical Diagnostics,” Ph.D. thesis, Texas A&M University Texas; (2000). [Google Scholar]
- 24.Wood M. F. G., Rohani A., Ghazalah R., Vitkin I. A., Pawluczyk R., “Multivariate analysis methods for spectroscopic blood analysis,” in “Biomedical Vibrational Spectroscopy V: Advances in Research and Industry,”, vol. 8219 Mahadevan-Jansen A., Petrich W., eds. (SPIE, 2012). [Google Scholar]
- 25.Clarke B. W., Cameron B. D., “The development of an integrated faraday modulator and compensator for continuous polarimetric glucose monitoring,” in “Optical Diagnostics and Sensing XIII: Toward Point-of-Care Diagnostics,” vol. 8591 Coté G. L., ed. (SPIE, 2013), pp. 1–11. [Google Scholar]
- 26.King T. W., Cote G. L., McNichols R. J., Marcel J. Goetz J., “Multispectral polarimetric glucose detection using a single pockels cell,” Opt. Eng. 33(8), 2746 (1994). 10.1117/12.173556 [DOI] [Google Scholar]
- 27.Stark C., Behroozian R., Redmer B., Fiedler F., Müller S., “Real-time compensation method for robust polarimetric determination of glucose in turbid media,” Biomed. Opt. Express 10(1), 308 (2019). 10.1364/BOE.10.000308 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Cote G., Ansari R., “A noninvasive glucose sensor based on polarimetric measurements through the aqueous humor of the eye,” in “Handbook of Optical Sensing of Glucose in Biological Fluids and Tissues,” Tuchin Valery V., ed. (Taylor & Francis, 2008), chap. 15, pp. 457–485. [Google Scholar]
- 29.Jirgensons B., “Optical Rotatory Dispersion of Proteins and Other Macromolecules,” Science 168, 962 (1969). 10.1007/978-3-662-12641-7 [DOI] [Google Scholar]
- 30.Malik B. H., Pirnstill C. W., Cote G. L., “Polarimetric glucose sensing in an artificial eye anterior chamber,” in “Optical Diagnostics and Sensing XII: Toward Point-of-Care Diagnostics and Design and Performance Validation of Phantoms Used in Conjunction with Optical Measurement of Tissue IV,” Nordstrom R. J., Cote G. L., eds. (SPIE, 2012). [Google Scholar]
- 31.Yoshida H., Tsubakimoto K., Fujimoto Y., Mikami K., Fujita H., Miyanaga N., Nozawa H., Yagi H., Yanagitani T., Nagata Y., Kinoshita H., “Optical properties and faraday effect of ceramic terbium gallium garnet for a room temperature faraday rotator,” Opt. Express 19(16), 15181 (2011). 10.1364/OE.19.015181 [DOI] [PubMed] [Google Scholar]
- 32.Riley M. R., Rhiel M., Zhou X., Arnold M. A., Murhammer D. W., “Simultaneous measurement of glucose and glutamine in insect cell culture media by near infrared spectroscopy,” Biotechnol. Bioeng. 55(1), 11–15 (1997). 10.1002/(SICI)1097-0290(19970705)55:1<11::AID-BIT2>3.0.CO;2-# [DOI] [PubMed] [Google Scholar]
- 33.Malik B. H., Pirnstill C. W., Coté G. L., “Dual-wavelength polarimetric glucose sensing in the presence of birefringence and motion artifact using anterior chamber of the eye phantoms,” J. Biomed. Opt. 18(1), 017007 (2013). 10.1117/1.JBO.18.1.017007 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Pirnstill C. W., Malik B. H., Gresham V. C., Coté G. L., “In vivo glucose monitoring using dual-wavelength polarimetry to overcome corneal birefringence in the presence of motion,” Diabetes Technol. Ther. 14(9), 819–827 (2012). 10.1089/dia.2012.0070 [DOI] [PMC free article] [PubMed] [Google Scholar]





