Table 3.

HPLC method validation results of unconjugated bile acids (u-BA) in mice plasma and eyes (n = 3) 1.

Bile Acid (Peak No.) Plasma Eyes
Conc. (μM) Precision (%) Accuracy (%) Conc. (μM) Precision (%) Accuracy (%)
u-BA 0.2 9.01 102 (10.82) 0.2 36.77 104 (20.36)
0.5 4.55 99 (4.57) 0.5 7.24 99 (4.74)
UDCA (3) 1 8.58 96 (7.31) 2 14.56 108 (19.22)
10 4.26 106 (2.80) 10 6.67 114 (2.77)
100 5.08 101 (1.37) 100 2.70 99 (0.91)
CA (6) 0.2 7.19 96 (13.40) 0.1 22.42 89 (13.90)
0.5 4.89 102 (3.64) 0.5 15.18 104 (12.17)
1 8.67 98 (7.43) 1 4.62 99 (2.99)
10 5.32 103 (2.90) 10 7.73 108 (4.22)
100 2.27 101 (1.20) 100 3.14 99 (0.79)
CDCA (11) 0.2 23.81 101 (10.02) 0.1 31.96 90 (6.72)
0.5 5.66 101 (7.62) 0.5 16.60 96 (7.28)
1 9.77 95 (7.66) 1 7.46 101 (1.25)
10 5.39 103 (2.86) 10 12.03 109 (8.10)
100 3.05 101 (1.25) 100 4.93 99 (1.27)
DCA (12) 0.2 5.73 104 (10.53) 0.1 27.87 102 (13.76)
0.5 4.25 100 (5.18) 0.5 19.08 96 (5.35)
1 8.22 96 (8.06) 1 5.91 101 (1.07)
10 4.74 103 (3.07) 10 11.26 111 (7.34)
100 4.28 101 (1.23) 100 4.51 99 (1.01)
LCA (16) 0.2 17.75 99 (9.72) 0.2 21.92 86 (8.31)
0.5 12.08 102 (11.57) 0.5 22.05 101 (12.31)
1 12.07 98 (12.86) 1 3.35 100 (2.75)
10 4.66 102 (2.89) 10 9.60 105 (6.72)
100 4.02 101 (1.17) 100 4.81 100 (1.51)

1 RSD (residual deviation, %) values are displayed in the parenthesis for the averaged accuracy (%) values.