Skip to main content
NIHPA Author Manuscripts logoLink to NIHPA Author Manuscripts
. Author manuscript; available in PMC: 2020 Nov 1.
Published in final edited form as: Phys Med. 2019 Oct 10;67:1–8. doi: 10.1016/j.ejmp.2019.10.018

Electron modulated arc therapy (EMAT) using photon MLC for postmastectomy chest wall treatment I: Monte Carlo-based dosimetric characterizations

Chaoqiong Ma a,b, David Parsons b, Mingli Chen b, Steve Jiang b, Qing Hou a, Xuejun Gu b, Weiguo Lu b,*
PMCID: PMC6925626  NIHMSID: NIHMS1060014  PMID: 31606657

Abstract

Purpose:

To study the dosimetric properties of electron arc beams delivered by photon-beam multi-leaf collimators (pMLC) in electron modulated arc therapy (EMAT) for postmastectomy chest wall treatments.

Methods:

Using the Monte Carlo method, we simulated a 2100EX Varian linear accelerator and verified the beam models in a water tank. Dosimetric characterizations were performed on cylindrical water phantoms of elliptical bases with various field sizes, arc ranges and source-to-surface distances (SSDs) for 6, 9 and 12 MeV beam energy.

Results:

The arc beam has a higher bremsstrahlung dose than the static beam at the isocenter due to crossfire, but choosing a field size greater than 5 cm effectively reduces the bremsstrahlung dose. The depths of the 90% maximum dose located at 1.7, 2.8 and 4.1 cm for 6, 9 and 12 MeV, respectively, are similar to those of the static beams and independent of the field size and arc range.

Conclusion:

Based on the study, we recommend using the 5 cm field width for electron arc beams considering both bremsstrahlung dose at the isocenter and the arc profile penumbra. To ensure sufficient PTV edge coverage, we recommend a field length extension of at least 4 cm from PTV’s edge for all beam energies and an arc extension of around 7°, 5°, and 5° for beam energies 6, 9, and 12 MeV, respectively. These dosimetric characterizations are the basis of pMLC-delivered EMAT treatment planning for postmastectomy chest wall patients.

Keywords: electron arc therapy (EAT), photon MLC (pMLC), electron modulated arc therapy (EMAT), Monte Carlo (MC)

1. Introduction

Electron therapy has played an important role in radiation therapy for the past several decades because of the advantage of the well-defined superficial dose range compared to photon therapy. With the average energy loss of ~2 MeV·cm2/g in the entire beam range in water-like tissues [1] [2], clinical electron beam energies are typically selected from 6 to 20 MeV when used in treating superficial targets (within 5 cm of the patients’ surface), including head and neck, postmastectomy chest wall, ribs and limbs [3]. These treatment objectives are to cover the planning target volume (PTV) within the 90% isodose while minimizing the dose to adjacent and distal normal tissues [3].

In conventional electron conformal therapy, single or multiple static electron beams are typically applied by limiting the field shape to the beam’s eye view of the target [3]. However, for large curved superficial PTVs, such as in the case of postmastectomy chest wall patients, a single static electron field is insufficient to cover the target; multiple fields are used to provide target coverage but may result in unacceptable dose gaps and non-uniformity at the field junctions [4]. Thus, techniques, such as electron arc therapy (EAT) [4] and modulated electron radiotherapy (MERT) [5] have been developed to improve target dose conformity and homogeneity.

Electron arc therapy (EAT), which rotationally delivers a collimated narrow beam around the PTV, was developed to reduce the hot, cold spots from adjoining static fields and dose to normal tissues for these cases [4, 6-8]. However, due to potential collisions, standard applicators for static beams were replaced by secondary collimators, such as aluminum blocks, custom blocks, and multi-vane collimators, inserted on the accessory tray as well as tertiary collimators laying on the patient for edge sharpening [6], [9]. These custom collimations require laborious manufacturing and time-consuming, physics-demanding treatment planning; thus these techniques have been implemented only in few clinics despite the report of favorable treatment outcomes [8].

Rodrigues et al [10] proposed dynamic arc radiation therapy (DEAR), a non-coplanar version of EAT, which uses synchronized gantry rotation, couch motion, and dose rate modulation to improve dose conformity; however, with a more complicated source trajectory and still with an electron applicator, the challenge of collision may be even greater.

In contrast to EAT, modulated electron radiotherapy (MERT) combines multiple segmented fields at fixed beam angles to create intensity modulation [5, 11]; dose conformality in the depth direction is achieved by modulating electron incident energy [12-14] or using a custom bolus to modify beam penetration [15-19]. Except using the bolus for energy modulation, MERT and EAT are analogous to intensity modulated radiotherapy and volumetric modulated arc therapy in photon therapy. Multiple cutouts, dedicated electron multi-leaf collimators (eMLC), or few-leaf electron collimator (FLEC), have been implemented to create segmented fields in MERT [5, 20-22], but the involved workload, dosimetry, and treatment planning remain a challenge. The built-in photon MLC (pMLC) of a linear accelerator (LINAC), on the other hand, is an attractive workaround since no custom molding to create segmented fields or independent hardware control as in eMLC or FLEC is needed. pMLC has been investigated for its dosimetric feasibility [23, 24] and characterizations [25] in MERT, including treatment planning for various treatment sites [12, 13, 26].

The use of pMLC in EAT has also been investigated for segmented arcs [27] but with inferior dose uniformity and penumbra compared to applicator collimation due to adjoining segments and air gaps. However, using pMLC in a continuous arc may still be dosimetrically feasible, in particular, when combined with bolus modulation. Despite the interest in using pMLC, systematic dosimetric characterizations of pMLC-delivered electron arc beams are lacking. The purpose of this study is to characterize the dosimetry of electron arc beams delivered through pMLC for various beam energies, field sizes, arc ranges and source-to-surface distances (SSDs), using Monte Carlo simulations. Such characterizations allow us to identify the capabilities and limitations of pMLC-delivered electron arc beams and are the basis for selecting these parameters in electron modulated arc therapy (EMAT) treatment planning.

2. Methods and materials

2.1. Monte Carlo simulations

Electron dose distributions were simulated using the EGSnrc Monte Carlo software toolkit [28]. A 2100EX linac with a 120 leaf MLC (NDS120, Varian Medical Systems, Inc., Palo Alto, CA) was modeled in BEAMnrc [29] with exact geometric and material specifications provided by the manufacture. The model was previously validated for 6 MV photons [30-32] and updated to include 6, 9 and 12 MeV electron beams. The default parameters were used for EGSnrc with a global electron and photon cut-off energy of 521 and 10 keV, respectively. Approximately 108 particles were used such that the dose uncertainty was less than 1% at the depth of the maximum dose for the chosen voxel size. The models were validated through central-axis percent depth dose (PDD) and off-axis dose ratio profile measurements acquired using a 50 × 50 × 50 cm3 water tank (Blue Phantom 2, IBA Dosimetry GmbH, Schwarzenbruck, Germany) and a diode (60017, PTW Freiburg GmbH, Freiburg, Germany) and compared to simulated PDD using DOSXYZnrc [33] with an xyz voxel resolution of 4 × 4 × 2 mm2. The xy - dimensions composed the transverse planes, on which the depth dose profiles were evaluated, and the z-dimension was in the longitudinal direction, along which the in-plane profiles were evaluated. The electron beam measurements were performed by overriding the applicator interlock [27]. The electron energy was tuned to match PDD, and the scattering foils was tuned to match profiles [34] for the 30 × 30 cm2 field, without deviating too far from manufacturer’s specification, and validated by multiple rectangular fields of static beams.

Electron arc treatments were simulated using source 20 in DOSXYZnrc developed by Lobo et al [31]. The arcs were simulated using one-degree resolution for the calculation purpose; physical delivery would be continuous rotation. EGSPHANT files were created to simulate various regular and elliptical cylinders with an xyz voxel resolution of 1 × 1 × 4 mm3. For arc delivery, both the depth dose and arc profiles are evaluated on the central transverse plane.

2.2. Electron arc beams collimated by pMLC

The electron arc beam dose was calculated on cylindrical water phantoms, including circular and elliptical-based cylinders, for dosimetric characterizations. The arc delivery and circular-based cylindrical phantom are illustrated in figure 1. The radiation isocenter was selected at the cylinder center, and the field width and length were defined at the isocenter plane.

Figure 1.

Figure 1

Schematic illustration of beam geometry on a transverse plane of a cylindrical phantom with radius r and field width w.

Photon MLCs were used to shape rectangular fields, and the coordinates were defined using the IEC 61217 system. Unless stated otherwise, all simulations used the 180° electron arc clockwise from gantry angle 270° to 90° (figure 1), the 5 × 30 cm2 field, the cylinder with radius of 15 cm and length of 40 cm, and the fixed source-to-axis distance (SAD) of 100 cm. Depth dose in the radial direction at the arc center (radial PDD), arc dose on the transverse plane, and longitudinal dose in the cylinder axes direction were analyzed for combinations of beam energies, field sizes, arc ranges and cylinder radii. Table 1 lists the parameters of combinations. As clinical electron treatment plans are required to cover the PTV by the 90% isodose, all arc and longitudinal dose profiles were evaluated at R90, the depth of 90% maximum dose, of the radial PDD and normalized to the maximum dose. Here Rn denotes the depth of n% maximum dose on the electron percent depth dose (PDD) curve. For example, R100 is the depth of (100%) maximum dose.

Table 1.

Parameters for the dosimetric characteristic study. Combinations of these parameters were used to study the effect of beam energy, field size, arc range, and SSD. The value pairs of elliptical cylinders indicate the major and minor axes.

Beam energy (MeV) 6, 9, 12
Field width (cm) 3, 5, 7
Field length (cm) 10, 20, 30
Arc range (°) 60, 120, 180
Cylinder radius (cm) 10, 15, 20
Elliptical cylinder (cm) (20,10), (20,15)

2.2.1. Beam energy effect

We chose 6, 9 and 12 MeV beam energies because they are close to the energy for chest wall treatments and because higher energy beams are associated with higher bremsstrahlung dose and longer beam ranges causing adjacent OARs to receive more radiation dose. We investigated the beam energy effect on the depth dose at the center of the 180° arc by varying the beam energy but fixing the field size and phantom dimensions.

2.2.2. Field size effect

The dosimetric effect of field size was studied using rectangular fields, where the field width is the collimation width along the arc and the field length is determined by the superior-inferior target coverage. To reduce the effect of oblique electron beam incidence that results in large dose deviation from normal incidence [35], small field widths, 3, 5, and 7 cm, were selected for electron arc beams. Radial PDDs and arc dose profiles were studied for the field width effect. The effect of field length was studied by comparing longitudinal dose profiles of three field lengths, 10, 20, and 30 cm.

2.2.3. Arc range effect

We varied the arc range from 60° to 180° to simulate a range of irradiation from ipsilateral chest wall to bilateral chest wall. The effect of arc range on the radial PDDs and arc dose profiles was analyzed.

2.2.4. Source-to-surface distance (SSD) effect

The effect of chest wall SSD was investigated on radial PDDs, arc and longitudinal dose profiles using various cylinder radii, 10, 15, and 20 cm, and elliptical cylinders with a major axis of 20 cm and minor axis of 10 and 15 cm in three beam energies, 6, 9, and 12 MeV. Since the SAD was fixed at 100 cm, the cylinder radii of 10, 15, and 20 cm corresponded to the SSD of 90, 85, and 80 cm, respectively. We compared the dose distributions of regular and elliptical cylinders to create a more contrasting view of the SSD effect.

3. Results

3.1. Monte Carlo validation

The measured and simulated PDDs at the center of the field 30 × 30 cm2 agreed within 2% for energies 6, 9 and 12 MeV (figure 2). The profiles at R90 (figure 3) similarly agreed within 2% in the central region and within 2.5 mm in the gradient, corresponding to an energy uncertainty of approximately 0.5 MeV. Similar agreement was found for the validation fields. Figure 4 shows the results of the 30 × 3 cm2.

Figure 2.

Figure 2

Measured and simulated PDDs with a field size of 30 × 30 cm2 for energies 6, 9 and 12 MeV.

Figure 3.

Figure 3

Measured and simulated in-plane profiles at R90 with a field size of 30 × 30 cm2 for energies 6, 9 and 12 MeV.

Figure 4.

Figure 4

Measured and simulated PDDs and in-plane profiles with a field size of 30 × 3 cm2 for the 6 MeV electron beam.

The simulation time with EGSnrc was ~30 min and ~1% statistical uncertainty on a 14-core computer.

3.2. Dosimetric characteristics of electron arc beams collimated by pMLC

3.2.1. Beam energy effect

The radial PDDs of different beam energies are shown in figure 5. The surface doses were 68%, 65%, and 62%, for 6 MeV, 9 MeV and 12 MeV, respectively. This contrasted with static beams: the surface dose for static beams of the same field 5 × 30 cm2 were 74%, 78%, and 85% for 6 MeV, 9 MeV and 12 MeV, respectively. This contrast was due to the increased irradiation time (dose rate) with respect to the increased depth in arc delivery, also known as the β angle effect [36]. The depths R100, R90 and R20, like those of the static beam, increased as the energy increased. They also increased slightly relative to those of the static beam due to the β angle effect. The falloff distance increased, or equivalently, the gradient decreased between R90 and R20 (R90–20) with increased energy: R90–20 increased from 0.9 to 1.7 cm as the beam energy increased from 6 MeV to 12 MeV (table 2). Finally, the bremsstrahlung dose peaked near the isocenter were 2%, 4% and 8% for 6, 9, and 12 MeV electron arc beams, respectively.

Figure 5.

Figure 5

Radial PDDs of 180° electron arc beams with the 5 × 30 cm2 field for 6, 9 and 12 MeV beam energies.

Table 2.

Depth dose positions R100, R90 and R20 in unit of cm for a 5 × 30 cm2 field, 180° arc, and 6, 9 and 12 MeV beam energy. In parentheses are those of static beams of the same field size.

Beam energy (MeV) R100 R90 R20
6 1.3 (1.4) 1.7 (1.8) 2.6 (2.7)
9 2.2 (2.1) 2.8 (2.8) 4.0 (4.0)
12 3.3 (3.1) 4.1 (4.0) 5.8 (5.7)

3.2.2. Field size effect

The radial PDDs of field widths 3, 5, and 7 cm for each of the three beam energies, 6, 9, and 12 MeV are shown in figure 6. For the same energy, there was no significant difference in PDDs of different field widths until the depth reaches 10 cm, where the bremsstrahlung dose becomes prominent. The bremsstrahlung dose (%) at the isocenter increases as the field width decreases (table 3) due to the crossfire effect and is consistent with the work of Leavitt et al [6] and Pla et al [37]. Here the crossfire effect refers to that the isocenter is always under irradiation since it is the focus of all beams in rotational delivery. In fact, all points that are within half the field width from the isocenter will be irradiated all the time in rotational delivery, and as the field width decreases, the crossfire effect increases.

Figure 6.

Figure 6

Radial PDDs of 180° electron arc beams with field widths of 3, 5, and 7 cm for each of the three beam energies, 6 MeV (left), 9 MeV (middle), and 12 MeV (right).

Table 3.

Dose (%) at the isocenter for field width 3, 5 and 7 cm, beam energy 6, 9 and 12 MeV, and electron arc 180°.

Beam energy (MeV)Field width (cm) 3 5 7
6 3.9 2.3 1.8
9 6.0 3.8 2.9
12 12.5 8.1 5.9

The effect of field width on arc dose is shown in figure 7 for 9 MeV. The ramp (dose falloff) is centered at the 90 degree angle and extends about 15 degree angle on each side of the center, which can potentially impacts dose conformity and homogeneity. There is a slight effect of field width and beam energy on the falloff gradient. The smaller field and higher energy correspond to the slightly steeper dose falloff shown in table 4 as seen by the angle of the 80% maximum dose, A80, and the degree arc from the 80% to 20% maximum dose, A80–20.

Figure 7.

Figure 7

Arc dose profiles at R90 for field width of 3 cm, 5 cm and 7 cm for 180° arc beams of energy 9 MeV. Only half the profiles are shown due to symmetry about 0°. The x-axis is the angle (°) from the center of the arc.

Table 4.

A80 and A80-20 (in parentheses) of the arc dose profiles at widths of 3 cm, 5 cm and 7 cm for 180° electron arc beams of energies 6 MeV, 9 MeV and 12 MeV. All values are in units of degree (°).

Beam energy (MeV)Field width (cm) 3 5 7
6 78 (20) 76 (23) 75 (24)
9 79 (18) 78 (21) 76 (24)
12 80 (18) 78 (20) 76 (23)

The longitudinal dose profiles at the depth R90 along the arc center are shown in figure 8 for the 10, 20 and 30 cm field length for the 9 MeV beam. The dose falloff measured at the 80% maximum dose, L80, and distance between 80% and 20% maximum dose, L80–20, are listed in table 6 for different field lengths for each beam energy.

Figure 8.

Figure 8

Longitudinal dose profiles at R90 (defined at the arc center) with field lengths of 10 cm, 20 cm and 30 cm for 180° electron arcs at 9 MeV. Only half the profiles are shown due to symmetry.

Table 6.

L80 and L80-20 (in parentheses) for cylindrical phantoms of radiuses 10 cm, 15 cm and 20 cm (corresponding to SSD 90, 85, and 80 cm, respectively) for 180° electron arc beams of energies of 6 MeV, 9 MeV and 12 MeV. All values are in units of cm.

6 MeV 9 MeV 12 MeV
10 cm 12.0 (4.6) 12.2 (4.3) 12.1 (4.3)
15 cm 11.1 (4.3) 11.5 (3.8) 11.6 (3.9)
20 cm 10.5 (3.8) 11.0 (3.4) 11.0 (3.5)

As the energy increased from 6 MeV to 12 MeV, L80 increased slightly from 2.3 cm to 2.7, from 6.4 cm to 7.3 cm, and from 11.1 cm to 11.7 cm, for field length 10 cm, 20 cm, and 30 cm, respectively. Higher energy yielded slightly sharper penumbra, which enables higher dose homogeneity for the same normal tissue toxicity constraint. L80-20 increased by approximately 4 mm as field length increased from 10 cm to 30 cm for the same energy; that is, dose fall-off along the longitudinal direction was marginally sharper with smaller field lengths.

3.2.3. Arc range

No significant difference was observed in the PDD curves, except for the bremsstrahlung dose near the isocenter. Figure 9 illustrates the variation in bremsstrahlung dose component for various arc ranges. Dose at the isocenter increased by 1.3%, 2.4% and 5.0% for 6 MeV, 9 MeV and 12 MeV, respectively, as the arc range increased from 60° to 180°.

Figure 9.

Figure 9

Bremsstrahlung radial PDD component at arc ranges of 60°, 120° and 180° for 6 MeV (left), 9MeV (middle) and 12 MeV (right).

3.2.4. Source-to-surface distance

Figure 10 shows the radial PDDs of the 12 MeV beam for different phantom radii (therefore, different SSD). In figure 10a, each curve is normalized by its own maximum. There was no significant difference in R100, R90 and R90–20, but surface dose increased as the radius increased, with a maximum increase of 13% between the 10 cm and 20 cm phantom radii. Also, the bremsstrahlung dose at the isocenter increased with increased phantom radii. In contrast, the curves in figure 10b are normalized by the maximum dose of the phantom of 10 cm radius. The increased dose with decreased radius is due to the increased irradiation time toward the isocenter, while the increased bremsstrahlung dose with decreased radius is due to decreased attenuation.

Figure 10.

Figure 10

(a) Radial PDDs (normalized by each curve’s own maximum) of cylindrical phantoms of radii 10, 15 and 20 cm, corresponding to SSD of 90, 85 and 80 cm, respectively, at a fixed field size 5 × 30 cm2 for 180° electron arc beams of energy 12 MeV; (b) Similar to (a) but normalization was by dmax for the phantom of radius 10 cm.

Figure 11 (a) and (b) illustrate arc and longitudinal dose profiles at R90, respectively, for phantom radii 10 cm, 15 cm and 20 cm at 12 MeV, with specific values of A80, A80–20 listed in table 5 and L80, L80–20 in table 6. The penumbra increases as the phantom radius decreases, due to the divergence effect.

Figure 11.

Figure 11

(a) Arc dose profiles at the central transverse plane and (b) longitudinal dose profiles at the arc center, both at d90, for cylindrical phantoms of radii 10, 15, and 20 cm, corresponding to the SSD of 90, 85, and 80 cm, respectively, of the 12 MeV beam. Only half the profiles are shown due to symmetry. In (a), The x-axis is the angle (°) from the center of the arc.

Table 5.

A80 and A80-20 (in parentheses) for cylindrical phantoms of radii 10, 15, and 20 cm (corresponding to SSD 90, 85, and 80 cm, respectively) for 180° electron arc beams of energies of 6 MeV, 9 MeV and 12 MeV. All values are in units of degree (°).

6 MeV 9 MeV 12 MeV
10 cm 68 (40) 68 (40) 65 (42)
15 cm 78 (26) 78 (25) 77 (27)
20 cm 83 (15) 84 (14) 85 (14)

Figure 12 illustrates the dose distributions at the central transverse planes of elliptical cylindrical phantoms with the same major axis 20 cm but different minor axes 10 cm, 15 cm and 20 cm, for 9 MeV beam with a fixed field size 5 × 30 cm2 and arc range 360°. Surface dose increased as surface-to-isocenter distance decreased.

Figure 12.

Figure 12

Normalized dose distributions at the central transverse planes of the elliptical cylinders of the same major axis 20 cm but different minor axes: 10 (left), 15 (middle), and 20 cm (right) for a fixed field size 5 × 30 cm2 and 360° electron arc beams in 9 MeV beam energy. The white cross in each image marks the normalization reference point.

4. Discussion

In this work, we characterized pMLC-delivered electron arc beam dosimetry through investigating three dose profiles, radial PDD, arc profiles, and longitudinal profiles, on various circular and elliptical-based cylindrical phantoms, for various beam parameters based on Monte Carlo (MC) simulations validated for statics beams. Though we did not measure the electron arc beams, the MC-simulated radial PDD were similar to those measured by Leavitt et al [6] (fig. 5 therein). For the same geometry and setup as ours (5-cm field width, 100-cm SAD, 15-cm cylinder radius, 180° arc), except that they used aluminum secondary collimators mounted on an acrylic accessory tray insert instead of pMLC, their measured surface dose were 70%, 67%, and 68%, and R100 were 1.3 cm, 2.2 cm, and 3.3 cm, approximately, for 6, 9, and 12 MeV, respectively (by visual inspection of their graphs). Our simulated R100 agreed with their measurements, but surface dose was slightly lower at 68%, 65%, and 62%, for 6, 9, and 12 MeV, respectively (figure 5). The slightly lower surface dose in our simulations may be accounted for by the greater clearance between pMLC and patient surface that fewer low energy scattered electrons reach the phantom surface [6].

The electron arc beam can be regarded as a superposition of static beams, and difference of dosimetric characteristics between the arc and static beams can be largely attributed to the radial dose rate property. The dose rate increases as the radial depth increases due to the crossfire effect and saturates at depth within the field from all beam angles, i.e. where the distance from the rotation axis is less than one half field width. Such radial dose rate property has been observed by Khan et al [4] and Pla et al [36]. Khan et al explained this as a mechanism of angular velocity, in which as depth increases, linear velocity decreases, thereby increasing the apparent exposure time to the beam. Pla et al described this by the β-angle concept, i.e. the arc range of radiation exposure, and the inverse square law of the divergence effect; however, the latter has only a second order effect for typical treatment SADs. The β-angle ratio (dose rate ratio) against radial position for multiple field widths are illustrated in figure 13

Figure 13.

Figure 13

Dose rate ratio against the radial position for field widths of 1, 3, 5, and 7 cm in electron arc beams of 180° arc.

The radial dose rate property causes the surface dose of the electron arc beam to be lower than that of the static beam when all other beam parameters are fixed. Similarly, R100 is increased due to the increased dose rate along the depth. These effects are more substantial at higher energy since the R100 of higher energy beam is at a deeper location, causing the surface dose (%) to decrease as the beam energy (6, 9, and 12 MeV) increases, which exactly reverses the order of surface dose (%) with static beams (figure 2).

The higher surface doses near the minor axis than those near the major axis for elliptical cylinders can be explained by the same reason, which is commonly referred to as the SSD effect but can be more directly expressed as the effect of the distance to the rotation axis. The doses of elliptical cylinders, like those of the chest wall, were also affected by electron beam incidence obliquity, but the obliquity effect was small since the selected field widths were small. Furthermore, with a properly selected isocenter, the obliquity effect can also be reduced [38].

In addition to the increased practical range with energy, the bremsstrahlung component, which occurs within 3 cm from the rotation axis in all simulations, was amplified. This could be problematic for chest wall treatments, where the PTV distal edges are adjacent to the lung tissue. In these cases, a large amount of lung volume would be exposed to radiation with high energy (≥12 MeV) arc beams due to the low density of the lung tissue (approximately one-fifth of water). Therefore, the use of high energy (≥12 MeV) electron arcs for chest wall irradiations requires extra attention to avoid excessive dose to lungs.

The arc and longitudinal dose profiles were studied at R90, and the profile penumbra were evaluated using 80% and 20% maximum dose. The profile penumbra is relevant in treatment planning since the steeper penumbra enables higher dose conformity and homogeneity. We found that the beam energy has a slight effect on the penumbra of the arc and longitudinal dose profiles: the higher energy, the sharper penumbra with an effect of ~5 mm or less.

In the investigation of the field width effect, there was no significant difference in the build-up regions, R100, R90 and R20 of PDDs for all three investigated field widths, 3, 5, and 7 cm, for fixed beam energies. Thus, the common R90, 1.7 cm, 2.8 cm, and 4.1 cm, for 6 MeV, 9 MeV, and 12 MeV, respectively, can be regarded as the standard R90 when selecting the beam energy in pMLC-based electron arc treatment planning.

On the other hand, the field width effect on the bremsstrahlung dose is prominent. The 3 cm field width yields twice as high bremsstrahlung dose as that of the 7 cm field width. In fact, the bremsstrahlung dose increases as the field width decreases, again, due to the crossfire effect, since the relative dose rate of R100 to the center is lower for the larger field (figure 13). These results are consistent with those using the x and y jaws to collimate electron arc beams, where 4 to 8 cm field widths were recommended to avoid markedly higher bremsstrahlung dose of smaller field widths [6, 37].

There is also a clear effect of the field size on the penumbra of the arc and longitudinal dose profiles: the penumbra increases as the field size increase with increasing effect for increasing beam energy.

Non-rectangular fields may be used in arc delivery though not studied in this work. Leavitt et al [6] used custom collimation to create trapezoid fields for electron arc beams to generate non-uniform dose rates at different longitudinal positions. Dynamically opening/closing fields in arc delivery may be used to sharpen the penumbra and would entail the use of in-transit small fields. Klein et al [23] and Mihaljevic et al [39] verified Monte Carlo simulations for small fields (1 × 1, 2 × 2, 2.6 × 2.6 cm2) with static beams but did not recommend them for electron beams since the percentage of bremsstrahlung dose is high beyond the beam range. It would be much easier to create non-rectangular or dynamically changing fields with pMLC then with custom collimation, and their dosimetric feasibility will be studied in the future.

The arc range has minimal dosimetric effects except for the bremsstrahlung dose, which increases with the increasing arc range.

The effect of SSD is prominent in both bremsstrahlung dose and profile penumbra. Besides the radial dose rate property, the SSD also has an effect on in-air scatter. The larger penumbra was observed for the smaller phantom radius due to larger scatter with the larger SSD. The extended shoulder (L80) of the longitudinal dose profile with the small phantom radius, however, is attributed to the divergence effect.

The absolute dose was not calibrated in this work since we used the relative dose for treatment planning. To measure planning dose, the absolute dose calibration would be required, and it is possible [40]. The output needs be measured against a standard beam; currently, the standard beam is the one with 10 × 10 cm2 cutout and SSD 105 cm. Mueller et al [25] showed that with the same field size 9.8 × 9.8 cm2, the output of pMLC with SSD 100 cm is similar to that of the cut-out with SSD 105 cm, and the output of the pMLC was nearly 3x as high when the SSD was 70 cm. Thus, the delivery time would be comparable to that of comprehensive chest wall treatments, such as Electron Arc Therapy (EAT) [7] [8].

Based on the dosimetric characterizations, we recommend placing the isocenter such that the average SSD is less than 80 cm to reduce the penumbra due to air scatter and to minimize surface distance variation responsible for the hot/cold spots. Small SSD may be prohibitive because of the risk of collision between the patient and gantry. The isocenter selection is further investigated in the EMAT treatment planning study [38]. We recommend using the 5 cm field width because it marginally increases bremsstrahlung dose and penumbra compared to the 7 and 3 cm field widths, respectively. We also recommend using a field length that is at least a 4 cm extension from each end of the PTV as seen on SSD to ensure PTV end coverage by 80% isodose. Similarly, the arc extensions by 7, 5, and 5 degree from each end of PTV for 6, 9, and 12 MeV, respectively, are recommended. The dose falloff regions associated with the extensions will be accounted for by the bolus design in the EMAT treatment planning study [38].

5. Conclusion

In this work, we investigated the dosimetric characteristics of the pMLC-delivered electron arc beam. It has the potential to produce homogeneous dose distributions in the PTVs while maintaining sharp dose fall-offs with properly selected parameters. This work provides the basis of selecting arc beam parameters, including beam energy, field size, isocenter position, and arc range for chest wall treatment planning in pMLC-delivered EMAT.

Acknowledgment

This work was partially supported by China Scholarship Council (CSC) and by NIH R01 CA218402 and R01 CA235723.

References

  • [1].ICRU, “Report 90,” Journal of the International Commission on Radiation Units and Measurements, vol. 14, pp. NP-NP, 2016. [DOI] [PubMed] [Google Scholar]
  • [2].Strydom W, Parker W, and Olivares M, “Chapter 8 Electron Beams: Physical and Clinical Aspects,” in Radiation oncology physics: a handbook for teachers and students, Podgoršak EB, Ed., ed Vienna: International Atomic Energy Agency, 2005, p. 657 p. [Google Scholar]
  • [3].Hogstrom KR and Almond PR, “Review of electron beam therapy physics,” Phys Med Biol, vol. 51, pp. R455–89, July 7 2006. [DOI] [PubMed] [Google Scholar]
  • [4].Khan FM, Fullerton GD, Lee JMF, Moore VC, and Levitt SH, “Physical Aspects of Electron-Beam Arc Therapy,” Radiology, vol. 124, pp. 497–500, 1977/August/01 1977. [DOI] [PubMed] [Google Scholar]
  • [5].Ma CM, Pawlicki T, Lee MC, Jiang SB, Li JS, Deng J, et al. , “Energy- and intensity-modulated electron beams for radiotherapy,” Physics in Medicine & Biology, vol. 45, p. 2293, 2000. [DOI] [PubMed] [Google Scholar]
  • [6].Leavitt DD, Peacock LM, Gibbs FA, and Stewart JR, “Electron arc therapy: Physical measurement and treatment planning techniques,” International Journal of Radiation Oncology*Biology*Physics, vol. 11, pp. 987–999, 1985/May/01/ 1985. [DOI] [PubMed] [Google Scholar]
  • [7].Leavitt DD and Stewart JR, “Electron arc therapy of the postmastectomy prosthetic breast,” International Journal of Radiation Oncology*Biology*Physics, vol. 28, pp. 297–301, 1994/January/01/ 1994. [DOI] [PubMed] [Google Scholar]
  • [8].Gaffney DK, Leavitt DD, Tsodikov A, Smith L, Watson G, Patton G, et al. , “Electron arc irradiation of the postmastectomy chest wall with CT treatment planning: 20-year experience,” International Journal of Radiation Oncology*Biology*Physics, vol. 51, pp. 994–1001, 2001/November/15/ 2001. [DOI] [PubMed] [Google Scholar]
  • [9].Leavitt DD, Stewart JR, Moeller JH, and Earley L, “Optimization of electron arc therapy doses by multi-vane collimator control,” Int J Radiat Oncol Biol Phys, vol. 16, pp. 489–96, February 1989. [DOI] [PubMed] [Google Scholar]
  • [10].Rodrigues A, Yin FF, and Wu Q, “Dynamic electron arc radiotherapy (DEAR): a feasibility study,” Phys Med Biol, vol. 59, pp. 327–45, January 20 2014. [DOI] [PubMed] [Google Scholar]
  • [11].Zackrisson B and Karlsson M, “Matching of electron beams for conformal therapy of target volumes at moderate depths,” Radiother Oncol, vol. 39, pp. 261–70, June 1996. [DOI] [PubMed] [Google Scholar]
  • [12].Salguero FJ, Palma B, Arrans R, Rosello J, and Leal A, “Modulated electron radiotherapy treatment planning using a photon multileaf collimator for post-mastectomized chest walls,” Radiotherapy and Oncology, vol. 93, pp. 625–632, 2009. [DOI] [PubMed] [Google Scholar]
  • [13].Salguero FJ, Arrans R, Palma BA, and Leal A, “Intensity- and energy-modulated electron radiotherapy by means of an xMLC for head and neck shallow tumors,” Phys Med Biol, vol. 55, pp. 1413–27, March 7 2010. [DOI] [PubMed] [Google Scholar]
  • [14].Henzen D, Manser P, Frei D, Volken W, Neuenschwander H, Born EJ, et al. , “Beamlet based direct aperture optimization for MERT using a photon MLC,” Med Phys, vol. 41, p. 121711, December 2014. [DOI] [PubMed] [Google Scholar]
  • [15].Low DA, Starkschall G, Bujnowski SW, Wang LL, and Hogstrom KR, “Electron bolus design for radiotherapy treatment planning: Bolus design algorithms,” Medical Physics, vol. 19, pp. 115–124, 1992/January/01 1992. [DOI] [PubMed] [Google Scholar]
  • [16].Kudchadker RJ, Hogstrom KR, Garden AS, McNeese MD, Boyd RA, and Antolak JA, “Electron conformal radiotherapy using bolus and intensity modulation,” International Journal of Radiation Oncology • Biology • Physics, vol. 53, pp. 1023–1037, 2002. [DOI] [PubMed] [Google Scholar]
  • [17].Kudchadker RJ, Antolak JA, Morrison WH, Wong PF, and Hogstrom KR, “Utilization of custom electron bolus in head and neck radiotherapy,” Journal of Applied Clinical Medical Physics, vol. 4, pp. 321–333, 2003/September/01 2003. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [18].Su S, Moran K, and Robar James L, “Design and production of 3D printed bolus for electron radiation therapy,” Journal of Applied Clinical Medical Physics, vol. 15, pp. 194–211, 2014/July/01 2014. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [19].Lukowiak M, Jezierska K, Boehlke M, Wiecko M, Lukowiak A, Podraza W, et al. , “Utilization of a 3D printer to fabricate boluses used for electron therapy of skin lesions of the eye canthi,” J Appl Clin Med Phys, vol. 18, pp. 76–81, January 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [20].Ravindran BP, Singh IR, Brindha S, and Sathyan S, “Manual multi-leaf collimator for electron beam shaping--a feasibility study,” Phys Med Biol, vol. 47, pp. 4389–96, December 21 2002. [DOI] [PubMed] [Google Scholar]
  • [21].Hogstrom KR, Boyd RA, Antolak JA, Svatos MM, Faddegon BA, and Rosenman JG, “Dosimetry of a prototype retractable eMLC for fixed-beam electron therapy,” Med Phys, vol. 31, pp. 443–62, March 2004. [DOI] [PubMed] [Google Scholar]
  • [22].Al-Yahya K, Schwartz M, Shenouda G, Verhaegen F, Freeman C, and Seuntjens J, “Energy modulated electron therapy using a few leaf electron collimator in combination with IMRT and 3D-CRT: Monte Carlo-based planning and dosimetric evaluation,” Med Phys, vol. 32, pp. 2976–86, September 2005. [DOI] [PubMed] [Google Scholar]
  • [23].Klein EE, Vicic M, Ma CM, Low DA, and Drzymala RE, “Validation of calculations for electrons modulated with conventional photon multileaf collimators,” Phys Med Biol, vol. 53, pp. 1183–208, March 7 2008. [DOI] [PubMed] [Google Scholar]
  • [24].du Plessis FC, Leal A, Stathakis S, Xiong W, and Ma CM, “Characterization of megavoltage electron beams delivered through a photon multi-leaf collimator (pMLC),” Phys Med Biol, vol. 51, pp. 2113–29, April 21 2006. [DOI] [PubMed] [Google Scholar]
  • [25].Mueller S, Fix MK, Henzen D, Frei D, Frauchiger D, Loessl K, et al. , “Electron beam collimation with a photon MLC for standard electron treatments,” Phys Med Biol, vol. 63, p. 025017, January 11 2018. [DOI] [PubMed] [Google Scholar]
  • [26].Jin L, Ma CM, Fan J, Eldib A, Price RA, Chen L, et al. , “Dosimetric verification of modulated electron radiotherapy delivered using a photon multileaf collimator for intact breasts,” Phys Med Biol, vol. 53, pp. 6009–25, November 7 2008. [DOI] [PubMed] [Google Scholar]
  • [27].Klein EE, Li Z, and Low DA, “Feasibility study of multileaf collimated electrons with a scattering foil based accelerator,” Radiother Oncol, vol. 41, pp. 189–96, November 1996. [DOI] [PubMed] [Google Scholar]
  • [28].Kawrakow I, “Accurate condensed history Monte Carlo simulation of electron transport. I. EGSnrc, the new EGS4 version,” Medical Physics, vol. 27, pp. 485–498, 2000/March/01 2000. [DOI] [PubMed] [Google Scholar]
  • [29].Rogers DW, Faddegon BA, Ding GX, Ma CM, We J, and Mackie TR, “BEAM: a Monte Carlo code to simulate radiotherapy treatment units,” Med Phys, vol. 22, pp. 503–24, May 1995. [DOI] [PubMed] [Google Scholar]
  • [30].Berbeco RI, Detappe A, Tsiamas P, Parsons D, Yewondwossen M, and Robar J, “Low Z target switching to increase tumor endothelial cell dose enhancement during gold nanoparticle-aided radiation therapy,” Med Phys, vol. 43, p. 436, January 2016. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [31].Lobo J and Popescu IA, “Two new DOSXYZnrc sources for 4D Monte Carlo simulations of continuously variable beam configurations, with applications to RapidArc, VMAT, TomoTherapy and CyberKnife,” Phys Med Biol, vol. 55, pp. 4431–43, August 21 2010. [DOI] [PubMed] [Google Scholar]
  • [32].Parsons D and Robar JL, “Beam generation and planar imaging at energies below 2.40 MeV with carbon and aluminum linear accelerator targets,” Med Phys, vol. 39, pp. 4568–78, July 2012. [DOI] [PubMed] [Google Scholar]
  • [33].Walters B, Kawrakow I, and Rogers DW, “DOSXYZnrc Users Manual NRCC Report PIRS-794revB,” National Research Council, Ottawa, Canada: 2018. [Google Scholar]
  • [34].Bieda MR, Antolak JA, and Hogstrom KR, “The effect of scattering foil parameters on electron-beam Monte Carlo calculations,” Med Phys, vol. 28, pp. 2527–34, December 2001. [DOI] [PubMed] [Google Scholar]
  • [35].Ekstrand KE and Dixon RL, “The problem of obliquely incident beams in electron-beam treatment planning,” Med Phys, vol. 9, pp. 276–8, Mar-Apr 1982. [DOI] [PubMed] [Google Scholar]
  • [36].Pla M, Pla C, and Podgorsak EB, “The influence of beam parameters on percentage depth dose in electron arc therapy,” Med Phys, vol. 15, pp. 49–55, Jan-Feb 1988. [DOI] [PubMed] [Google Scholar]
  • [37].Pla M, Podgorsak EB, and Pla C, “Electron dose rate and photon contamination in electron arc therapy,” Med Phys, vol. 16, pp. 692–7, Sep-Oct 1989. [DOI] [PubMed] [Google Scholar]
  • [38].Ma C, Chen M, Parsons D, Jiang S, Hou Q, Gu X, et al. , “Electron modulated arc therapy (EMAT) using photon MLC for postmastectomy chest wall treatment II: 3D bolus design “ under consideration by PMB, 2019. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • [39].Mihaljevic J, Soukup M, Dohm O, and Alber M, “Monte Carlo simulation of small electron fields collimated by the integrated photon MLC,” Phys Med Biol, vol. 56, pp. 829–43, February 7 2011. [DOI] [PubMed] [Google Scholar]
  • [40].Popescu IA, Shaw CP, Zavgorodni SF, and Beckham WA, “Absolute dose calculations for Monte Carlo simulations of radiotherapy beams,” Phys Med Biol, vol. 50, pp. 3375–92, July 21 2005. [DOI] [PubMed] [Google Scholar]

RESOURCES