Fragmented QRS Predicts the Arrhythmic Events in Patients with Heart Failure Undergoing ICD Implantation for Primary Prophylaxis: More Fragments More Appropriate ICD Shocks

Fırat Özcan 1, Osman Turak 1, Uğur Canpolat 1,, Sedat Avci 1, Derya Tok 1, Ahmet Isleyen 1, Muhammed Cebeci 1, Hüseyin Yuzgeçer 1, Özgül Malçok Gurel 1, Serkan Topaloglu 1, Dursun Aras 1, Fatma Nurcan Basar 1, Sinan Aydogdu 1
PMCID: PMC6932166  PMID: 24920012

Abstract

Background

Fragmented QRS complex (fQRS) is associated with cardiovascular outcomes in various patient populations. Although there were clinical studies investigating the association of fQRS with arrhythmic events in patients with systolic heart failure, the results were conflicting regarding the association of implantable cardioverter defibrillator (ICD) shocks and fQRS. In this study, we aimed to evaluate the association between the presence and extent of fQRS with appropriate ICD shocks and/or all‐cause mortality.

Methods

A total of 215 patients (age: 58.2 ± 11.6 years, 72.5 % male) with the diagnosis of left ventricular systolic heart failure in whom ICD had been implanted for primary prophylaxis were enrolled. Standard ECG evaluation revealed fQRS complex in 123 patients (57.2 %). The phenomenon of fQRS was defined as deflections at the beginning of the QRS complex, on top of the R wave, or in the nadir of the S wave similar to the definition in CAD.

Results

At mean 23.5 ± 12.1 months follow‐up, all‐cause mortality was observed in 45 (20.9 %) patients and 111 (51.6 %) patients experienced appropriate ICD shocks. Median number of ECG leads with fQRS were higher in patients with appropriate ICD shocks (3 [2–6] vs 1 [0–2], P < 0.001, respectively). The presence of fQRS (HR: 6.64, 95 % CI: 3.54–12.4, P < 0.001) and the number of leads with fQRS (HR: 1.35, 95% CI: 1.22–1.67) were found as independent predictors of appropriate ICD shocks. Additionally, there was a negative correlation between left ventricular ejection fraction and the number of leads with fQRS (r = −0.434, P < 0.001). Rates of all‐cause mortality did not differ between the fQRS(+) (29 [24 % ]) and fQRS(−) (16 [17 % ]) groups (P = 0.27).

Conclusion

Our findings suggest that the presence and extent of fQRS complex on standard 12‐lead ECG predicts appropriate ICD shocks in patients with left ventricular systolic heart failure who underwent ICD implantation for primary prophylaxis.

Keywords: ımplantable cardioverter defibrillator, appropriate shock, fragmented QRS complex


Implantable cardioverter defibrillator (ICD) implantation is recommended for primary prevention of sudden cardiac death to reduce total mortality in selected patients with ischemic or nonischemic cardiomyopathy.1 Although there were several studies investigating the predictors of appropriate ICD shocks in those patients, left ventricular ejection fraction (LVEF) seems to be significant major contributor in heart failure with either ischemic2 or nonischemic etiology.3 Fragmented QRS complex (fQRS), as a noninvasive prognostic marker on resting electrocardiography (ECG) emerged as a diagnostic and/or prognostic marker in several cardiovascular conditions.4, 5, 6, 7 However, clinical trials showed conflicting results regarding the association of arrhythmic events and appropriate ICD shocks with fQRS in patients with systolic heart failure.8, 9, 10, 11, 12, 13 So, in this study, we aimed to investigate the association of appropriate ICD shocks and all‐cause mortality with the presence and extent of fQRS in patients with systolic heart failure who underwent ICD implantation for primary prophylaxis previously.

METHODS

Study Population

A total of 215 eligible patients (mean age: 58.2 ± 11.6 years, 72.5 % male) who admitted to our Pacemaker Clinic for scheduled ICD interrogation as a part of routine follow‐up visits between June 2011 and April 2012 were enrolled in this study. All patients had primary prevention indications according to the current American College of Cardiology/American Heart Association Guidelines for ICD at the time of implantation. The study population included patients with >18 years old, with dilated cardiomyopathy due to ischemic or nonischemic etiology, chronic stable heart failure (NYHA class I–III) and with LVEF <35%. Patients in whom ICD implantation have been performed for secondary prophylaxis were excluded from the study. The study population divided into two subgroups according to the detection of appropriate ICD shocks (Group I; with appropriate ICD shocks and Group II; without any ICD shocks). Informed consent was taken from each patient before enrollment. The study was in compliance with the principles outlined in the Declaration of Helsinki and approved by Institutional Ethics Committee.

Electrocardiography

At the time of initial diagnosis, standard resting 12‐lead ECGs (filter range 0.15–100 Hz, AC filter 60 Hz, 25 mm/s , 10 mm/mV) were analyzed by two independent readers blinded to the patient clinical and echocardiographic findings. There was 96.5% concordance for the ECG signs including fQRS. When there was a disagreement, the final conclusion was achieved by mutual agreement.

Fragmentation of narrow QRS (<120 ms) is defined as presence of an additional R wave (R’) or notching in the nadir of the S wave, or the presence of >1 R’ in two contiguous leads, corresponding to a major coronary artery territory on the resting 12‐lead ECG (Fig. 1). Fragmentation of wide complex QRS (>120 ms) consists of various RSR patterns, with more than 2 R waves (R’’) or more than two notches in the R wave, or more than two notches in the downstroke or upstroke of the S wave.14

Figure 1.

Figure 1

(A) The electrocardiogram of a patient with left ventricular ejection fraction of 28 % and fQRS in inferolateral derivations (six leads with fQRS) in whom most frequent ventricular tachycardia episodes were determined during ICD interrogation (B).

Echocardiography

Standard echocardiographic imaging was performed in the left lateral decubitus position in the parasternal and apical views. Two‐dimensional, M‐mode, pulsed and color flow Doppler echocardiographic examinations of all subjects were performed by the same examiner with a commercially available machine (Vivid 7, GE Medical System, Horten, Norway; 3.5‐MHz phased array transducer). LVEF was calculated from apical 4‐chamber views, according to the modified Simpson's rule.15

Follow‐up and ICD Interrogation

The end points for the study were all‐cause mortality, appropriate ICD shocks due to ventricular tachyarrhythmias and combined endpoint of appropriate ICD shocks or all‐cause mortality. Outpatient follow‐up visits were performed for every 6 months or according to the patient's clinical status. All stored episodes of arrhythmia that occurred since the implantation of the device were retrospectively analyzed for accuracy of detection, diagnostic precision, and appropriateness of device therapy. Device interrogation was performed by an experienced electrophysiologist who was unaware of the presence of fQRS. The ICD programming included therapy for standard ventricular tachycardia (VT) with antitachycardia pacing combined with low‐energy shock therapy and for ventricular fibrillation (VF) shock therapy with a 300‐ms cutoff cycle interval. It was considered standard VT in the presence of sustained tachycardia with a cycle interval ranging from 300 to 400 ms, not discriminated as supra‐VT by specific algorithms. It was considered VF when the cycle interval was inferior to 300 ms. Appropriate ICD shock was defined as therapy for rapid sustained ventricular tachycardia or fibrillation episode.

Statistical Analysis

Continuous variables were expressed as mean ± SD or median (interquartle range) and categorical variables were expressed as percentages. Continuous variables were tested for normal distribution by the Kolmogorov–Smirnov test. Categorical variables were analyzed by chi‐square test or Fisher's exact test. Comparisons of continuous variables between the two groups were performed using the Student's t‐test. Kappa statistics was used to evaluate the interobserver agreement of ECG signs. Univariate and multivariable Cox proportional hazard regression was applied for assessment of the association of baseline clinical and noninvasive variables with occurrence of arrhythmic events. Kaplan–Meier estimator was generated to estimate survival curves and log‐rank test was used to compare survival probability among the f‐QRS (+) and fQRS (−) groups. Statistical analyses were performed using SPSS statistical software (version 20.0; SPSS Inc., Chicago, IL, USA). A P value <0.05 was considered statistically significant.

RESULTS

During a mean follow‐up of 23.5 ± 12.1 months, appropriate ICD shock was observed in 111 of 215 patients (51.6%). Study population divided into two subgroups due to presence of appropriate ICD shocks (group I: 111 patients and group II: 104 patients). The baseline characteristics of the study population according to the presence of ICD shocks were shown in Table 1. All clinical and laboratory parameters were similar among study groups except older age in group I (P < 0.05). While ECG findings were similar regarding the baseline rhythm, there were significant differences between groups due to QRS duration (100 [90–120] ms in group I vs 120 [100–130] ms in group II, P < 0.001), presence and number of leads with fQRS complex (P < 0.05).

Table 1.

Baseline Characteristics of the Study Population due to Development of Appropriate ICD Shocks (n = 215)

Shock (−) n = 104 Shock (+) n = 111 P
Clinical parameters
Age, years 56.6 ± 12.8 59.8 ± 12.6 0.001
Male gender, n (%) 79 (76 %) 77 (64 %) 0.279
Hypertension, n (%) 70 (67 %) 84 (76) 0.174
Dyslipidemia, n (%) 62 (60 %) 73 (66 %) 0.351
DM, n (%) 21 (21 %) 25 (23 %) 0.677
Smoking, n (%) 15 (14 %) 17 (15 %) 0.854
Ischemic etiology, n (%) 42 (41 %) 60 (59 %) 0.045
NYHA functional class
I 39 (51 %) 37 (49 %)
II 53 (47 %) 59 (53 %) 0.787
III 12 (44 %) 15 (56 %)
Laboratory findings
FPG, mg/dL 102(94–117) 104(93–134) 0.669
Serum creatinine, mg/dL 0.95(0.8–1.2) 1.0(0.9–1.2) 0.441
Free T3, pg/mL 2.7 ± 0.7 2.9 ± 0.8 0.461
Free T4, pg/mL 1.4 ± 0.6 1.3 ± 0.3 0.873
WBC, x103/uL 8.2 ± 1.7 8.3 ± 1.6 0.823
Hemoglobin, mg/dL 14.3 ± 1.6 13.8 ± 1.5 0.086
Platelet, range, x103/uL 218 (118–274) 216 (188–295) 0.232
CRP, mg/dL 9.6 ± 3.6 11.1 ± 4.0 0.645
Electrocardiographic findings
Sinus rhythm, n (%) 74 (71 %) 82 (74 %) 0.655
Atrial fibrillation, n (%) 30 (29 %) 29 (26 %)
LVEF,%, mean 28.1 ± 3.7 27.3 ± 3.2 0.324
QRS duration, msec 100 (90–120) 120 (100–130) <0.001
No. leads with fQRS 1 (0–2) 3 (2–6) <0.001
No. patients with fQRS, n (%) 35 (34 %) 88 (75 %) <0.001

CRP = C‐reactive protein; DM = diabetes mellitus; FPG = fasting plasma glucose; fQRS = fragmentation of QRS complex; ICD = implantable cardioverter defibrillator; LVEF = left ventricular ejection fraction; NYHA = New York Heart Association; WBC = white blood cell.

The predictors of appropriate ICD shocks were determined by univariable and multivariable regression analysis. The presence of fQRS (HR: 6.64, 95% CI: 3.54–12.4, P < 0.001), number of leads with fQRS (HR: 1.35, 95% CI: 1.22–1.67) and ischemic etiology (OR: 3.04, 95% CI: 2.93–3.15, P < 0.001) were found as independent predictors of appropriate ICD shocks after adjustment for other variables including age, gender, LVEF, and QRS duration (Table 2). Furthermore, Spearman correlation analysis revealed that there were significant negative correlation between LVEF and the number of leads with fQRS (r = −0.434, P < 0.001).

Table 2.

Univariate and Multivariate Regression Analysis for Evaluating the Predictors of Appropriate ICD Shocks

Univariate Multivariate
Parameters HR (CI 95 %) P HR (CI 95 %) P
Age, years 0.99 (0.97–1.21) 0.156
Male gender 1.04 (0.72–1.27) 0.244
No. leads with fQRS 1.58 (1.36–1.83) <0.001 1.35 (−1.22–1.67) 0.001
Presence of fQRS 7.54 (4.08–13.9) <0.001 6.64 (3.54–12.4) <0.001
QRS duration, msec 1.02 (1.0–1.03) 0.001 1.06 (0.99–1.02) 0.106
LVEF, % 0.98 (0.95–1.02) 0.258
Ischemic etiology 3.21 (2.85–3.58) <0.001 3.04 (2.93–3.15) <0.001

CI = confidence interval; fQRS = fragmentation of QRS complex; ICD = implantable cardioverter defibrillator; HR = hazard ratio; msec = milliseconds; No. = number of.

A total of 45 (20.9 %) patients died during study follow‐up, of which 29 (24 %) were in the fQRS(+) group and 16 (17 %) were in the fQRS (−) group (P = 0.27). Also, a total of 111 (51.6 %) patients experienced appropriate ICD shocks due to VT/VF, of which 88 (72 %) were in the fQRS (+) group and 23 (25 %) were in the fQRS(−) group (P < 0.001) (Table 3). Figure 2 represents the rates of event‐free survival on Kaplan–Meier plots between the fQRS (+) and fQRS (−) groups. Additionally, all‐cause mortality, appropriate ICD shocks due to VT/VF and combined end points were significantly higher in patients with ischemic etiology independent from the presence of fQRS complex (Table 4).

Table 3.

Summary of Study Outcomes According to the Presence of fQRS Complex

fQRS (−) (n = 92) fQRS (+) (n = 123) P
All‐cause mortality, n (%) 16 (17 %) 29 (24 %) 0.270
VT/VF, n (%) 23 (25 %) 88 (72 %) <0.001
Combineda, n (%) 25(27 %) 92(75 %) <0.001

fQRS = fragmented QRS complex; VF = ventricular fibrillation; VT = ventricular tachycardia

a

Combined end points included VT/VF or all‐cause mortality.

Figure 2.

Figure 2

Kaplan–Meier analysis for the ventricular tachyarrhythmias, all‐cause mortality and combined end point of all‐cause mortality or ventricular tachyarrhythmia in patients with fragmented QRS and nonfragmented QRS.

Table 4.

Summary of Study Outcomes According to the Etiology of the Systolic Heart Failure

Nonischemic (n = 113) Ischemic (n = 102) P
All‐cause mortality, n (%) 14 (12 %) 31 (30 %) 0.001
VT/VF, n (%) 51 (45 %) 60 (59 %) 0.045
Combineda, n (%) 55 (49 %) 83 (81 %) <0.001

fQRS = Fragmented QRS complex; VF = ventricular fibrillation; VT = ventricular tachycardia

a

Combined end points included VT/VF or all‐cause mortality.

DISCUSSION

Our study results showed that; (i) presence of fQRS and number of leads with fQRS were well predicted the appropriate ICD shocks in patients with systolic heart failure and, (ii) the LVEF was negatively correlated with the number of leads with fQRS.

The mechanism of fQRS on standard 12‐lead ECG is speculative. It has been suggested that fQRS has occurred as a result of inhomogeneous electrical activation of the ventricles due to myocardial scar and/or ischemia and also it was precisely predicted the arrhythmic events and sudden cardiac death in several studies.14, 16, 17, 18 The diagnostic accuracy of fQRS has been investigated in different studies in which they have showed that fQRS had higher sensitivity and negative predictive value than Q wave14 and fQRS was moderately sensitive but highly specific for myocardial scar.4 In imaging studies, fQRS was well signified the myocardial scar by delayed enhancement of gadolinium technique on cardiac MRI.19, 20 Also several studies have confirmed the association of fQRS with myocardial perfusion abnormality and scar tissue.4, 14, 21, 22, 23 Endocardial and epicardial mapping in dilated cardiomyopathy patients with ventricular arrhythmias have shown that fractionated electrograms over a wide area surrounding the myocardial scar.24, 25 So, we have postulated that fQRS precisely represents the myocardial scar which was a milieu for reentry and malignant ventricular arrhythmias.

In a study by Das et al.,8 the association of arrhythmic events and mortality with fQSR was investigated in 361 patients with ischemic or nonischemic cardiomyopathy who received ICD for primary or secondary prophylaxis. At 16.6 ± 10.2 months follow‐up, fQRS was found to be a predictor of arrhythmic events, but not mortality. Moreover, electrical storm or performing VT ablation was significantly higher in the group with fQRS. Cheema et al.9 evaluated the association of fQRS with all‐cause and arrhythmic mortality in 842 patients with moderate to severe left ventricular systolic dysfunction (ischemic or nonishemic etiology). During 40 ± 17 months follow‐up period, they have found no association between the presence of fQRS and all‐cause or arrhythmic mortality. Also the rate of appropriate ICD shocks were similar among non‐fQRS and fQRS groups. Additionally, in a retrospective study, Forleo et al.10 found that fQRS did not predict the arrhythmic events and mortality. The discussed study findings may be negatively resulted for fQRS due to heterogenous study populations in which ICD recipients consisted of primary and secondary prophylaxis groups. However, contrary to previous study results, Micheal et al.11 and Sha et al.12 have showed that there were significant association of fQRS with ventricular arrhythmias and mortality in patients with nonischemic cardiomyopathy. Also, Maskoun et al.13 reported that fQRS as a predictor of arrhythmic events in patients with ischemic cardiomyopathy. Similarly, in a recently published retrospective study by Brenyo et al.,26 they confirmed the relationship of fQRS with appropriate ICD shocks and sudden cardiac death in patients with ischemic cardiomyopathy. So, there were conflicting data regarding the prognostic role of fQRS in patients with systolic heart failure. Our study results were consistent with the significant prognostic role of fQRS in either ischemic or nonischemic cardiomyopathy patients who underwent ICD implantation for primary prophylaxis. We have showed that fQRS was associated with appropriate ICD shocks for malignant ventricular arrhythmias, but was not associated with all‐cause mortality. Additionally, the number of leads with fQRS found as significant predictor for appropriate ICD shocks. Because of well documented negative correlation of LVEF with the number of leads with fQRS, we proposed that those patients with higher number of leads with fQRS had greater myocardial scar tissue in our study. Also, we assumed that more leads with fQRS reflecting increased myocardial scar tissue was well correlated with appropriate ICD shocks.

Our study should be evaluated by some limitations. First, it is a cross‐sectional study with small sample size, so prospectively designed large scale studies may precisely highlight the association of fQRS with arrhythmic events. Second, quantification of the myocardial scar tissue by SPECT or cardiac MRI may well define the association of fQRS with arrhythmic events.

In conclusion, fQRS which is a simple, inexpensive and readily available standard 12‐lead ECG marker, predicts arrhythmic events and appropriate ICD shocks, but not all‐cause mortality in patients with systolic heart failure undergoing ICD implantation for primary prophylaxis. So, fQRS may be used as an additional noninvasive marker for risk stratification and patient selection before ICD implantation for primary prophylaxis.

This study was not funded by any institution. Authors do not have any conflict of interest.

Disclosure: None

REFERENCES

  • 1. Yancy CW, Jessup M, Bozkurt B, et al. 2013 ACCF/AHA Guideline for the management of heart failure: a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines. J Am Coll Cardiol 2013;62(16):e147–e239. [DOI] [PubMed] [Google Scholar]
  • 2. Chow T, Kereiakes DJ, Bartone C, et al. Microvolt T‐wave alternans identifies patients with ischemic cardiomyopathy who benefit from implantable cardioverter‐defibrillator therapy. J Am Coll Cardiol 2007;49:50–58. [DOI] [PubMed] [Google Scholar]
  • 3. Zipes DP, Camm AJ, Borggrefe M, et al.; American College of Cardiology /American Heart Association Task Force, European Society of Cardiology Committee for Practice Guidelines, European Heart Rhythm Association, Heart Rhythm Society . ACC/AHA/ESC 2006 Guidelines for Management of Patients With Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death: a report of the American College of Cardiology/American Heart Association Task Force and the European Society of Cardiology Committee for Practice Guidelines (writing committee to develop Guidelines for Management of Patients With Ventricular Arrhythmias and the Prevention of Sudden Cardiac Death): developed in collaboration with the European Heart Rhythm Association and the Heart Rhythm Society. Circulation 2006;114:e385–484. [DOI] [PubMed] [Google Scholar]
  • 4. Das MK, Suradi H, Maskoun W, et al. Fragmented wide QRS on a 12‐lead ECG: a sign of myocardial scar and poor prognosis. Circ Arrhythm Electrophysiol 2008;1:258–268. [DOI] [PubMed] [Google Scholar]
  • 5. Das MK, Saha C, El Masry H, et al. Fragmented QRS on a 12‐lead ECG: a predictor of mortality and cardiac events in patients with coronary artery disease. Heart Rhythm 2007;4:1385–1392. [DOI] [PubMed] [Google Scholar]
  • 6. Pietrasik G, Goldenberg I, Zdzienicka J, et al. Prognostic significance of fragmented QRS complex for predicting the risk of recurrent cardiac events in patients with Q‐wave myocardial infarction. Am J Cardiol 2007;100:583–586. [DOI] [PubMed] [Google Scholar]
  • 7. Canpolat U, Kabakçi G, Aytemir K, et al. Fragmented QRS complex predicts the arrhythmic events in patients with arrhythmogenic right ventricular cardiomyopathy/dysplasia. J Cardiovasc Electrophysiol 2013;24:1260–1266. [DOI] [PubMed] [Google Scholar]
  • 8. Das MK, Maskoun W, Shen C, et al. Fragmented QRS on twelve‐lead electrocardiogram predicts arrhythmic events in patients with ischemic and nonischemic cardiomyopathy. Heart Rhythm 2010;7:74–80. [DOI] [PubMed] [Google Scholar]
  • 9. Cheema A, Khalid A, Wimmer A, et al. Fragmented QRS and mortality risk in patients with left ventricular dysfunction. Circ Arrhythm Electrophysiol 2010;3:339–344. [DOI] [PubMed] [Google Scholar]
  • 10. Forleo GB, Della Rocca DG, Papavasileiou LP, et al. Predictive value of fragmented QRS in primary prevention implantable cardioverter defibrillator recipients with left ventricular dysfunction. J Cardiovasc Med 2011;12:779–784. [DOI] [PubMed] [Google Scholar]
  • 11. Michael M, Das MK. Fragmented QRS (fQRS) on 12‐lead EKG is a predictor of arrhythmic events and mortality in patients with dilated cardiomyopathy. Heart Rhythm 2006;(3 Suppl):S103. [Google Scholar]
  • 12. Sha J, Zhang S, Tang M, et al. Fragmented QRS is associated with all‐cause mortality and ventricular arrhythmias in patient with idiopathic dilated cardiomyopathy. Ann Noninvasive Electrocardiol 2011;16:270–275. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Maskoun W, Suradi H, Mahenthiran J, et al. Fragmented QRS complexes on a 12‐ lead ECG predict arrhythmic events in patients with ischemic cardiomyopathy who receive an ICD for primary prophylaxis. Heart Rhythm 2007;4:S211–S212. [Google Scholar]
  • 14. Das MK, Khan B, Jacob S, et al. Significance of a fragmented QRS complex versus a Q wave in patients with coronary artery disease. Circulation 2006;113:2495–2501. [DOI] [PubMed] [Google Scholar]
  • 15. Lang RM, Bierig M, Devereux RB, et al.; Chamber Quantification Writing Group, American Society of Echocardiography's Guidelines, Standards Committee, European Association of Echocardiography . Recommendations for chamber quantification: a report from the American Society of Echocardiography's Guidelines and Standards Committee and the Chamber Quantification Writing Group, developed in conjunction with the European Association of Echocardiography, a branch of the European Society of Cardiology. J Am Soc Echocardiogr 2005;18:1440–1463. [DOI] [PubMed] [Google Scholar]
  • 16. Mahenthiran J, Khan BR, Sawada SG, et al. Fragmented QRS complexes not typical of a bundle branch block: a marker of greater myocardial perfusion tomography abnormalities in coronary artery disease. J Nucl Cardiol 2007;14(3):347–353. [DOI] [PubMed] [Google Scholar]
  • 17. Varriale P, Chryssos BE. The RSR’ complex not related to right bundle branch block: diagnostic value as a sign of myocardial infarction scar. Am Heart J 1992;123:369–376. [DOI] [PubMed] [Google Scholar]
  • 18. Flowers NC, Horan LG, Thomas JR, et al. The anatomic basis for highfrequency components in the electrocardiogram. Circulation 1969;39:531–539. [DOI] [PubMed] [Google Scholar]
  • 19. Homsi M, Alsayed L, Das MK, et al. Fragmented QRS complexes on a 12‐lead ECG is a marker of greater myocardial scarring related to coronary artery disease by magnetic resonance imaging. J Am Coll Cardiol 2008;51:A31. [Google Scholar]
  • 20. Homsi M, Alsayed L, Das MK, et al. Fragmented QRS complexes on a 12‐ lead ECG is a marker of greater myocardial scarring related to non‐coronary artery diseases by magnetic resonance imaging. J Am Coll Cardiol 2009;53:A140. [Google Scholar]
  • 21. Basaran Y, Tigen K, Karaahmet T, et al. Fragmented QRS complexes are associated with cardiac fibrosis and significant intraventricular systolic dyssynchrony in nonischemic dilated cardiomyopathy patients with a narrow QRS interval. Echocardiography 2011;28:62–68. [DOI] [PubMed] [Google Scholar]
  • 22. Park SJ, On YK, Kim JS, et al. Relation of fragmented QRS complex to right ventricular fibrosis detected by late gadolinium enhancement cardiac magnetic resonance in adults with repaired tetralogy of fallot. Am J Cardiol 2012;109:110–115. [DOI] [PubMed] [Google Scholar]
  • 23. Homsi M, Alsayed L, Safadi B, et al. Fragmented QRS complexes on 12‐lead ECG: a marker of cardiac sarcoidosis as detected by gadolinium cardiac magnetic resonance imaging. Ann Noninvasive Electrocardiol 2009;14:319–326. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Stevenson WG, Soejima K. Catheter ablation for ventricular tachycardia. Circulation 2007;115:2750–2760. [DOI] [PubMed] [Google Scholar]
  • 25. Wiener I, Mindich B, Pitchon R. Fragmented endocardial electrical activity in patients with ventricular tachycardia: a new guide to surgical therapy. Am Heart J 1984;107:86–90. [DOI] [PubMed] [Google Scholar]
  • 26. Brenyo A, Pietrasik G, Barsheshet A, et al. QRS fragmentation and the risk of sudden cardiac death in MADIT II. J Cardiovasc Electrophysiol 2012;23:1343–1348. [DOI] [PubMed] [Google Scholar]