Skip to main content
Annals of Noninvasive Electrocardiology logoLink to Annals of Noninvasive Electrocardiology
. 2014 Mar 6;19(5):471–476. doi: 10.1111/anec.12147

Influence of Atrioventricular Optimization on Hemodynamic Parameters and Quality of Life in Patients with Dual Chamber Pacemaker with Ventricular Lead in Right Ventricular Outflow Tract

Artur Klimczak 1,, Adam S Budzikowski 2, Marcin Rosiak 1, Marzenna Zielińska 3, Bożena Urbanek 1, Karol Bartczak 4, Michał Chudzik 1, Jerzy K Wranicz 1
PMCID: PMC6932620  PMID: 24597906

Abstract

Background

The aim of this study was to ascertain whether individual atrioventricular delay (AVD) optimization using impedance cardiography (ICG) offers beneficial hemodynamic effects as well as improved exercise tolerance and quality of life in patients with requiring constant right ventricular pacing.

Methods

There were 37 patients with advanced AV block included in the study. Several examinations were performed at the beginning. Next, the optimization of AVD by ICG was done. The next step of the study patients have been randomized into optimal AVD group (AVDopt) or factory setting group (AVDfab). After 3 months, the follow‐up all data were collected again and crossover was performed. After another 3 months, during the final follow‐up all these measures were repeated.

Results

In 87.5% patients, AVDopt were different than factory value. Cardiac output (CO), cardiac index (CI), and stroke volume (SV) were significantly (P < 0.001) higher in AVDopt group than in AVDfab group (CO: 6.0 ± 1.4 L/minute vs. 5.3 ± 1.2 L/minute; SV: 85.8 ± 25.7 mL vs.76.9 ± 22.5 mL; CI: 3.2 ± 0.7 L/minute/m2 vs. 2.7 ± 0.6 L/minute/m2). There was a statistical significant (P < 0.05) reduction of proBNP and NYHA class in patients with AVDopt compared to AVDfab (proBNP: 196.4 ± 144.7pg/mL vs. 269.4 ± 235.8 pg/mL; NYHA class: 1.7 ± 0.5 vs. 2.3 ± 0.6). Six‐minute walking test was significantly (P < 0.05) higher in AVDopt group (409 ± 90 m) than in AVDfab group (362 ± 93 m). There were no statistically significant differences in echocardiographic parameters between AVDopt and AVDfab settings.

Conclusion

Our study results suggest that AVD optimization in patients with DDD pacemaker with ICG improves hemodynamic when compared to the default factory settings. Furthermore, optimally programmed AVD reduces BNP and improves exercise tolerance and functional class.

Keywords: pacing, optimization, atrioventricular delay, impedance cardiography, atrioventricular block


Cardiac pacing over the past decades allowed saving numerous lives of patients with bradycardia and complete heart block. Nevertheless, long observation of patients with implanted pacemaker showed unfavorable electrical and structural remodeling of the heart muscle.1 Although clear benefits of alternative right ventricular pacing sites remain unresolved,2 the pacing site should allow for the depolarization to mimic that achieved via native conduction system. The right ventricular outflow tract (RVOT) probably carries the smaller risk of unfavorable hemodynamic effect and electrical remodeling.3, 4

Although longterm structural remodeling and dyssynchrony may be avoided with optimal right ventricular pacing site, correct programming of atrioventricular delay (AVD) may offer additional hemodynamic benefits. The optimal AVD (AVDopt) differs among patients and requires individual setting.5, 6 Programming of AVD influences contractile function of left ventricle through preload modulation,7 and if done incorrectly, can have unfavorable hemodynamic results.8

The problem of the individual AVD optimization was noticed already in the 1980s.6 Various methods for determination of the AVDopt have been developed during last few decades (right heart catheterization, ventriculography, peak endocardial acceleration, echocardiography, impedance cardiography [ICG]), however, only echocardiography and ICG (reocardiography) are currently in general use.9 ICG is noninvasive, repetitive, undemanding of qualified personnel, and cheaper than the echocardiography and, therefore, can be readily used in selection of AVDopt in patients with dual‐chamber pacemakers.10, 11

The aim of this study was to ascertain whether individual AVD optimization offers beneficial hemodynamic effects as well as improved exercise tolerance and quality of life (QoL) in patients with requiring constant right ventricular pacing.

MATERIAL AND METHODS

There were 37 patients with advanced AV block (all the patients have percent of ventricular pacing >95% after pacemaker implantation) included in the study (28 men, nine women). All patients were implanted with a dual‐chamber pacemaker at least 4 weeks before the beginning of the study. All patients were implanted with Biotronik dual chamber pacemakers (manufactured in Berlin, Germany). Atrial lead was implanted in the right atrial appendage and ventricular lead was implanted in the RVOT. To verify the RVOT position of ventricular lead the fluoroscopic images during implantation and associated ECG patterns were used.12 Ventricular lead was placed approximately two‐thirds of the distance from the apex to the pulmonary valve in the posteroanterior fluoroscopic view and pointing toward the septum in the left‐anterior oblique view. The 12‐lead ECG helped identify the pacing site. Pacing at the RVOT septum was manifested by a negative QRS morphology in lead aVL and an upright QRS in leads II, III, and aVF.

All devices were programmed in the DDD mode with a lower rate limit of 50 beats per minute, to avoid atrial pacing.13

The exclusion criteria were: moderate to severe left ventricular dysfunction (LVEF < 40%), history of atrial fibrillation, significant valvular heart disease, sick sinus syndrome, or/and percent of atrial pacing >5%, abnormal (>120 ms) duration of P wave.

Before randomization all patients enrolled in the study underwent 6 minute walking test (6MWT), assessment of proBNP concentration in plasma, as well as had transthoracic echocardiogram. Next, the optimization of AVD by ICG was done. ICG was performed with a commercially available NICCOMO system (the CardioScreen®‐device, Medizinische Messtechnik GmbH, Ilmenau, Germany) used for noninvasive hemodynamic measurements and the monitoring of hemodynamic parameters. Changes in volume and velocity of blood in the aorta cause variations in the thoracic impedance which is measured and displayed as the ICG waveform. With each cardiac cycle, thoracic fluid volume changes, which affects the impedance of the electrical signal transmitted by the sensors. This signal is processed via physiological adaptive signal analysis to provide key hemodynamic parameters noninvasively and continuously.14 Four pairs of electrodes (one pair at the left and right neck, one pair on the left and right side of the lower thoracic aperture) were used for recording the thoracic impedance field. Cardiac index (CI) and other hemodynamic parameters were calculated beat‐to‐beat from the transthoracic impedance signal.15

AVD was set successively to the following values (in 5‐minute intervals): 100, 120, 140, 160, 180, 200, 250, and finally 300 ms. The ICG recording was performed for each AVD value and cardiac output (CO), stroke volume (SV) and CI were recorded. The AVD, for which CO was the highest value, was considered as the AVDoptAVDopt. The next step of the study was randomization. Pacemakers with even serial number were programmed to AVDopt and pacemakers with odd serial number were programmed to factory value of AVD (AVDfab). After 3 months, the follow‐up transthoracic echocardiogram was repeated as well as assessment of functional class (NYHA), 6MWT, and proBNP concentration and patients filled out QoL questionnaire—The Minnesota Living with Heart Failure. At this stage crossing‐over was done: pacemakers with even serial number were programmed to AVDfab and pacemakers with odd serial number were programmed to AVDopt. After another 3 months during the final follow‐up all these measures were repeated.

The medication remained unchanged when ever clinically feasible.

Statistica 5.0 PL software was used for statistical analysis. Shapiro‐Wilk W test was used for gathering evidence about the “nonnormality” of a sample. Paired t‐test was used for variables with normal distribution and Wilcoxon's signed‐rank test was used for nonparametric data. P < 0.05 was regarded as statistically significant.

Study protocol is shown in Figure 1. The study protocol was approved by the local Ethics Committee and informed consent was obtained from all patients.

Figure 1.

Figure 1

Study protocol.

AVD—atrio‐ventricular delay, ICG—impedance cardiography, AVDopt—optimal setting of AVD, AVDfab—factory setting of AVD.

RESULTS

During the study, five patients dropped out from the study because of various causes (new onset of atrial fibrillation, upgrade to cardiac resynchronization therapy, myocardial infarction, significant modification of pharmacotherapy, or missed follow‐up visits). Finally, 32 patients were completed all follow‐up visits. Clinical characteristic of these patients is shown in Table 1.

Table 1.

Clinical Characteristic of Patients

Number of Patients 32
Age 67.8 ± 11.3 years
Males 24 (75%)
Hypertension 21 (66%)
Diabetes type 2 8 (25%)
History of myocardial infarction 5 (16%)
LVEF 59.7 ± 6.2%
Mean NYHA class 2.3

LVEF = left ventricular ejection fraction; NYHA = New York Heart Association class of heart failure.

In 28 (87.5%) patients, AVDopt was different than factory value. Optimization of AVD by ICG resulted in better hemodynamic (measured during ICG optimization) as compared to the factory settings (Table 2).

Table 2.

Impendence Cardiography Parameters (Measured during ICG Optimization) in AVDfab and AVDopt Settings

Variables Cl fab Cl opt SV fab SV opt CO fab CO opt
Mean 2.67 3.18 76.89 85.84 5.26 6.00
SD 0.55 0.65 22.54 25.70 1.23 1.36
Median 2.55 3.1 78 85.5 5.4 6.1
Lower quartile 2.25 2.8 62 69.25 4.35 5.08
Upper quartile 3.08 3.48 89.75 102.5 6.1 6.75
P < 0.0001 P < 0.0001 P < 0.0001

CI = Cardiac index (L/minute/m2); SV = stroke volume (mL); CO = cardiac output (L/minute).

During 3 months of observation, there were no statistically significant differences in echocardiographic parameters (EDV, ESV, EF, TEI) between AVDopt and AVDfab settings.

There was a statistical significant reduction of pro‐BNP in patients with AVDopt settings compared to AVDfab settings (Table 3).

Table 3.

proBNP in AVDfab and AVDopt Settings

Variables proBNP fab proBNP opt
Mean 269.38 196.42
SD 235.81 144.67
Median 148.3 189.6
Lower quartile 107.08 94.58
Upper quartile 342.2 269.35
P = 0.016

proBNP = level of proBNP in plasma [pg/mL].

Patients with AVDopt settings were able to walk a longer distance (in 6MWT) compared to AVDfab settings (Table 4).

Table 4.

Six‐Minute Walking Time Test Results in AVDfab and AVDopt Settings

Variables 6MWT fab 6MWT opt
Mean 361.84 409.75
SD 92.72 90.38
Median 332.5 420
Lower quartile 280 319.5
Upper quartile 446.25 473.75
P = 0.0013

6MWT = six‐minute walking test (m).

Patients with AVDopt settings have also a significant lower NYHA class compared to AVDfab settings (Table 5).

Table 5.

NYHA Class in AVDfab and AVDopt Settings

Variables NYHA fab NYHA opt
Mean 2.31 1.69
SD 0.64 0.54
P < 0.001

NYHA = New York Heart Association class of heart failure.

Finally, QoL in patients with AVDopt settings were significant better than in AVDfab settings (Table 6).

Table 6.

QoL in AVDfab and AVDopt Settings

Variables QoL fab QoL opt
Mean 15.15 4.35
SD 10.90 4.25
Median 12.5 3
Lower quartile 7.75 0.75
Upper quartile 21.25 8.25
P < 0.001

QoL = quality of life (points).

Summary of the comparison of the main parameters are shown in Table 7.

Table 7.

Summary of the Effects of Optimal AV Delay Programming

graphic file with name ANEC-19-471-g002.jpg

proBNP = level of proBNP in plasma; CI = cardiac index; 6MWT = six‐minute walking test; QoL = quality of life; LVEF = left ventricle ejection fraction.

DISCUSSION

In this study, optimization of AVD by ICG provides better hemodynamics (CO, SV, CI) in the majority of patients as compared to the factory settings. Furthermore, optimal programmed AVD causes a statistically significant reduction in BNP, NYHA functional class and improves 6MWT.

Negative consequences of long‐term right ventricular pacing have been known for a long time16, 17 and may result in mechanical remodeling and increase in BNP. Lengthening of the programmed AVD or automated algorithms can allow for native depolarization of ventricles and potentially minimalize the unfavorable results of long‐term right ventricular pacing. However, excessive lengthening the delay in patients with atrioventricular block can paradoxically lead to the fall of CO, which by itself may be more detrimental than the right ventricular pacing.18, 19 On the other hand, programming too short of AVD will shorten the filling time and may provoke retrograde conduction.18, 19 When AVD is programmed optimally, it may improve CO by 13–40%.8

Patients with second and third degree atrioventricular block are inevitably destined to depend on the right ventricular pacing. In these patients, ventricular lead placement in other position than right ventricular apex (RVA) may significantly reduce negative influence of permanent ventricular stimulation. Many data suggest that RVOT pacing may offer significant benefits over RVA pacing.3, 20

Because the percent of ventricular stimulation in patients with atrioventricular block is always close or equal to 100%, the role of appropriate AVD optimization is particularly important. Echocardiography and ICG are two noninvasive methods that are commonly used to AVD optimization. ICG is free of some of the limitations compared to echocardiography—it is replicable, undemanding of qualified personnel, and cheaper than the echocardiography, and therefore it is good alternative to echocardiography in AVD optimization.10, 11 The measurement of the transthoracic impedance during cycle of the heart beat is the essence of the ICG. Thanks to the changes of the resistance of the blood pool in the aorta during the cycle of the heart beat it is possible to calculate CO and other hemodynamic parameters.11

With one notable exception,15 the majority of studies comparing the effects of the optimization of AVD by ICG with different methods collaborate the utility of this modality.21, 22, 23, 24 That is the reason why the ICG is applied more and more often as the helpful tool finding the most optimal setting of the AV delay. Furthermore, the ease of use makes possible finding of AVDopt in about 20 minutes. It seems that the combination of RVOT ventricular pacing and AVD optimization is the key method of management of patients with atrioventricular block.

Atrioventricular optimization in this study was performed only at rest similarly to other published data.25, 26 There are very little data regarding the hemodynamic effects of AVD optimization during exercise.27 It is still not clear as to whether AVD optimization performed during exercise may provide further hemodynamic benefits in comparison with only resting optimization. Most modern DDD pacemakers offer the function called “Dynamic AVD,” which provide dynamic AVD shortening with the increase of the heart rate. These algorithms were designed to approximate natural changes in AV conduction during exercise.9 However, recently published data28 seems to suggest that shortening or lengthening of the AVD during exercise has no considerable impact on CI hence programming of dynamic AVD may not offer additional benefits.

Our study was performed in patients with the intrinsic sinus rhythm (percentage of atrial pacing was <5%). When patient needs atrial pacing, e.g., because of significant bradycardia, AVD optimization is much more difficult. Optimal length of AVD differs whether the atrial event is paced (AP) or sensed (AS). Those two types AVD (AsVp and ApVp) differ on average by 16–134 ms, therefore some investigators consider that in patients with relevant percentage of atrial pacing, AVD should be optimized separately for sensed and paced P waves.29, 30

In our study, no significant difference were seen in echocardiographic parameters. It is plausible that the 3 month follow‐up may be too short to develop significant remodeling28, 31, 32 or the RVOT location of the lead had “protective” effect (carries the smaller risk of unfavorable hemodynamic effect compared to right ventricle apex pacing). There has been much confusion among investigators defining optimal site for the right ventricular pacing.12 True RVOT (especially right ventricular septum) pacing can be difficult to achieve. The technique for RVOT pacing, fluoroscopic appearances and ECG patterns have been elegantly reviewed recently.33

CONCLUSIONS

Successful AV sequential pacing involves not only appropriate right ventricular lead placement but also involves programming of an individual, AVDopt. Our study results suggest that AVD optimization in patients with DDD pacemaker with ICG improves hemodynamic when compared to the default factory settings. Furthermore, optimally programmed AVD reduces BNP and improves exercise tolerance and functional class.

REFERENCES

  • 1. Wecke L, Rubulis A, Lundahl G, et al. Right ventricular pacing‐induced electrophysiological remodeling in the human heart and its relationship to cardiac memory. Heart Rhythm 2007;4(12):1477–1486. [DOI] [PubMed] [Google Scholar]
  • 2. Gammage MD, Marsh AM. Randomized trials for selective site pacing: Do we know where we are going? Pace 2004;27:878. [DOI] [PubMed] [Google Scholar]
  • 3. Lewicka‐Nowak E, Dabrowska‐Kugacka A, Tybura S, et al. Right ventricular apex versus right ventricular outflow tract pacing: prospective, randomised, long‐term clinical and echocardiographic evaluation. Kardiol Pol 2006; 64(10):1082–1091. [PubMed] [Google Scholar]
  • 4. de Cock CC, Giudici MC, Twisk JW. Comparison of the haemodynamic effects of right ventricular outflow‐track pacing with right ventricular apex pacing: A quantitative review. Europace 2003;5:275–278. [DOI] [PubMed] [Google Scholar]
  • 5. Chirife R, Ortega DF, Salazar AI. A method for optimalization of left heart atrioventricular interval in patients with dual‐chamber pacemakers. Pace 1993;16(5 Pt II):A185. [Google Scholar]
  • 6. Wish M, Fletcher RD, Gottdiener JS, et al. Importance of left atrial timing in the programming of dual‐chamber pacemakers. Am J Cardiol 1987;60:566–571. [DOI] [PubMed] [Google Scholar]
  • 7. Chevalier S, Basta M, Leitch JW. The importance of the left atrioventricular interval during atrioventricular sequential pacing. Pace 1997; 20:2958–2966. [DOI] [PubMed] [Google Scholar]
  • 8. Daubert C, Mabo P, Gras D. Physiological cardiac pacing: An individual objective. The importance of AV synchrony optimization . In Aubert AE, Stroobandt R. (eds.): Cardiac Pacing and Electrophysiology. Dordrecht: Kluwer Academic Publishers, 1994, pp. 238–247. [Google Scholar]
  • 9. Klimczak A, Chudzik M, Zielinska M, et al. Optimization of atrio‐ventricular delay in patients with dual‐chamber pacemaker. Int J of Cardiol 2010;141:222–226. [DOI] [PubMed] [Google Scholar]
  • 10. Ovsyshcher I, Gross J, Blumberg S, et al. Variability of cardiac output as determined by impedance cardiography in pacemaker patients. Am J Cardiol 1993;72:183–187. [DOI] [PubMed] [Google Scholar]
  • 11. Sodolski T, Kutarski A. Impedance cardiography: A valuable method of evaluating haemodynamic parameters. Cardiology Journal 2007;14:115–126. [PubMed] [Google Scholar]
  • 12. Lieberman R, Grenz D, Mond HG, Gammage MD. Selective site pacing: Defining and reaching the selected site. Pacing Clin Electrophysiol 2004;27:883–886. [DOI] [PubMed] [Google Scholar]
  • 13. Bernheim A, Ammann P, Sticherling C, et al. Right atrial pacing impairs cardiac function during resynchronization therapy: Acute effects of DDD pacing compared to VDD pacing. J Am Coll Cardiol 2005;45:1482–1487. [DOI] [PubMed] [Google Scholar]
  • 14. http://www.medis-de.com. 2012.
  • 15. Ovsyshcher I, Furman S. Impedance cardiography for cardiac output estimation in pacemaker patients. Pacing Clin Electrophysiol 1993; 16:1412–1422. [DOI] [PubMed] [Google Scholar]
  • 16. Szili‐Torok T, Thornton A. The effects of right ventricular apical pacing on left ventricular function. Stimulation of the right ventricular apex: should it still be the gold standard? Indian Pacing Electrophysiol J 2003;3(2):74–80. [PMC free article] [PubMed] [Google Scholar]
  • 17. Ichiki H, Oketani N, Hamasaki S, et al. Effect of right ventricular apex pacing on the Tei index and brain natriuretic peptide in patients with a dual‐chamber pacemaker. Pacing Clin Electrophysiol 2006;29(9):985–990. [DOI] [PubMed] [Google Scholar]
  • 18. Ovsyshcher I, Zimlichman R, Katz A, et al. Measurement of cardiac output by impedance cardiography in pacemaker patients at rest: Effects of various AV delay. J Am Coll Cardiol 1993;21:761–767. [DOI] [PubMed] [Google Scholar]
  • 19. Brecker SJD, Gibson DC. What is the role of pacing in dilated cardiomyopathy? Eur Heart J 1996;17:819–824. [DOI] [PubMed] [Google Scholar]
  • 20. de Cock CC, Giudici MC, Twisk JW. Comparison of the haemodynamic effects of right ventricular outflow‐tract pacing with right ventricular apex pacing: A quantitative review. Europace 2003: 5(3):275–278. [DOI] [PubMed] [Google Scholar]
  • 21. Ovsyshcher I, Gross JN, Blumberg S, et al. Variability of cardiac output as determined by impedance cardiography in pacemaker patients. Am J Cardiol 1993;72:183–187. [DOI] [PubMed] [Google Scholar]
  • 22. Kindermann M, Frohlig G, Doerr T, et al. Optimizing AV delay in DDD pacemaker patients with high degree AV block: Mitral valve Doppler versus impedance cardiography. Pace 1997;20:2453–2462. [DOI] [PubMed] [Google Scholar]
  • 23. Ovsyshcher E: Toward physiological pacing: Optimization of cardiac hemodynamics by AV delay adjustment. Pace 1997;20:861–865. [DOI] [PubMed] [Google Scholar]
  • 24. Braun MU, Schnabel A, Rauwolf T, et al. Impedance cardiography as a noninvasive technique for atrioventricular interval optimization in cardiac resynchronization therapy. J Interv Card Electrophysiol 2005;13:223–229. [DOI] [PubMed] [Google Scholar]
  • 25. Auricchio A, Stellbrink C, Block M, et al. Effect of pacing chamber and atrioventricular delay on acute systolic function of paced patients with congestive heart failure. The pacing therapies for congestive heart failure study group. The Guidant congestive heart failure research group. Circulation 1999;99:2993–3001. [DOI] [PubMed] [Google Scholar]
  • 26. Kerlan JE, Sawhney NS, Waggoner AD, et al. Prospective comparison of echocardiographic atrioventricular delay optimization methods for cardiac resynchronization therapy. Heart Rhythm 2006;3:148–54. [DOI] [PubMed] [Google Scholar]
  • 27. Melzer C, Bondke H, Korber T, et al. Should we use the rate‐adaptive AV delay in cardiac resynchronization therapy‐pacing? Europace 2008;10:53–58. [DOI] [PubMed] [Google Scholar]
  • 28. Prochnau D, Forberg T, Kuhnert H, et al. Optimization of the atrioventricular delay during cardiac resynchronization therapy using a device for non‐invasive measurement of cardiac index at rest and during exercise. Europace 2012;14:249–253. [DOI] [PubMed] [Google Scholar]
  • 29. Janosik DL, Pearson AC, Buckingham TA, et al. The hemodynamic benefit of differential atrioventricular delay intervals for sensed and paced atrial events during physiologic pacing. J Am Coll Cardiol 1989;14:499–507. [DOI] [PubMed] [Google Scholar]
  • 30. Pearson AC, Janosik DL, Redd RR, et al. Doppler echocardiographic assessment of the of the effect of varying atrioventricular delay and pacemaker mode of left ventricular filling. Am Heart J 1988;115:611–621. [DOI] [PubMed] [Google Scholar]
  • 31. Bogaard MD, Doevendans PA, Leenders GE, et al. Can optimization of pacing settings compensate for a non‐optimal left ventricular pacing site?. Europace 2010;12:1262–1269. [DOI] [PubMed] [Google Scholar]
  • 32. Morales MA, Startari U, Panchetti L, et al. Atrioventricular delay optimization by Doppler‐derived left ventricular dP/dt improves 6‐month outcome of resynchronized patients. Pace 2006;29:564–568. [DOI] [PubMed] [Google Scholar]
  • 33. Hillock RJ, Mond HG. Pacing the right ventricular outflow tract septum: time to embrace the future. Europace 2012;14:28–35. [DOI] [PubMed] [Google Scholar]

Articles from Annals of Noninvasive Electrocardiology : The Official Journal of the International Society for Holter and Noninvasive Electrocardiology, Inc are provided here courtesy of International Society for Holter and Noninvasive Electrocardiology, Inc. and Wiley Periodicals, Inc.

RESOURCES