Dear Editor,
Immunoassays (IAs) based on streptavidin-biotin binding are widely used in clinical laboratory testing owing to the high affinity and stable interaction between streptavidin and biotin (vitamin B7) and the development of various biotinylation methods [1]. Biotin is a water-soluble B-complex vitamin and a coenzyme responsible for carboxyl transfer in essential carboxylases. Circulating serum concentrations of biotin in the general population typically range from 0.1 ng/mL to 0.8 ng/mL [2]. Biotin is rapidly absorbed, reaches peak plasma concentrations within 1–2 hours, and has an effective serum half-life of 15 hours [3]. Oral administration of biotin doses of 10 mg resulted in peak plasma concentrations ranging from 53 ng/mL to 141 ng/mL [3,4]. Considering a maximum dosage of 10 mg once a day (q.d.; 10 mg is >300-fold the adequate daily intake) via over the counter (OTC) biotin products, the serum biotin concentration would drop below the in vitro interference threshold of ≤30 ng/mL after eight hours [3]. A high biotin concentration in the blood can interfere with IAs based on streptavidin-biotin binding and is known to cause false high results in competitive IAs and false low results in sandwich IAs [1]. Although the recommended daily intake of biotin is low (30 µg/day), the use of high-dose biotin supplements (up to 10 mg), which are available OTC, has increased in recent years due to unfounded claims that biotin exerts beneficial effects on hair, nails, and skin [5]. According to one survey, 7.7% (95% confidence interval [CI], 6.6–8.9%) of an outpatient population (N=1,944) indicated biotin use; measuring biotin in plasma samples from emergency department patients (N=1,442) showed that 7.4% (95% CI, 6.2–8.9%) had a concentration at or above the lowest known threshold (10 ng/mL) for biotin interference in a Roche Diagnostics IA (Indianapolis, IN, USA) [6].
Because the Roche Elecsys assay is based on streptavidin-biotin binding, it may yield incorrect results for samples with high biotin concentration. Measurement of free thyroxine (FT4) using a competitive IA from Roche Diagnostics revealed that high biotin concentrations led to false high results in the Elecsys FT4 II assay. Recently, Roche Diagnostics released the Elecsys FT4 III assay, which is claimed to perform better in the presence of biotin interference. This is the first study to evaluate biotin interference in FT4 assays by comparing the newly released Elecsys FT4 III assay with the previous Elecsys FT4 II assay using a Cobas e602 Analyzer (Roche Diagnostics). The Institutional Review Board of Korea University Anam Hospital, Seoul, Korea, approved this study and waived the need for informed consent (No. 2019AN0400).
Pooled serum samples from 10 subjects with three targeted FT4 concentrations (high: >3 ng/dL; normal: 0.89–1.8 ng/dL; low: <0.8 ng/dL) were prepared by mixing three or four serum samples remaining after routine clinical assays of T3 and FT4. Biotin was purchased from Sigma-Aldrich (St. Louis, MO, USA) to produce a stock solution (4,000 ng/mL) for spiking the pooled serum samples (4 mL for each level). High-, normal-, and low-concentration samples were mixed with the biotin stock solution (9:1, high-biotin pool). For each FT4 range sample, the zero- and high-biotin serum pools were serially diluted to produce biotin-spiked samples (biotin concentration, 12.5, 25, 50, 100, 200, and 400 ng/mL). For both FT4 assays, all samples were analyzed in duplicate using a Roche Cobas e602 Analyzer. All experiments were completed in less than eight hrs without storage. Relative bias was calculated as follows:
| bias (%)= ([biotin-treated analyte]−[untreated analyte])/[untreated analyte]). |
To determine the biotin interference concentration, a clinically significant bias was defined as a 10% change in the result. For the Elecsys FT4 II assay, a biotin concentration ≥50 ng/mL caused a positive bias in low- and high-concentration pooled samples, while the normal-concentration samples showed >10% bias starting at 100 ng/mL of biotin. For Elecsys FT4 III, low- and normal-concentration pooled samples showed a positive bias of 32.5% and 27.7% at 400 ng/mL, respectively. For the high-concentration samples, a positive bias of 39.2% and 43.8% was observed at biotin concentrations of 200 and 400 ng/mL, respectively. The Elecsys FT4 II data sheet indicates a biotin interference threshold of ≤25 ng/mL. Thus, our findings showed that Elecsys FT4 III is less sensitive to biotin interference and corroborate its revised biotin interference threshold (≤100 ng/mL) (Table 1).
Table 1. Comparison of relative bias in biotin concentration between Elecsys FT4 II and Elecsys FT4 III.
| Estimated biotin concentration (ng/mL) | Elecsys FT4 II | Elecsys FT4 III | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Low FT4 | Normal FT4 | High FT4 | Low FT4 | Normal FT4 | High FT4 | |||||||
| FT4 (ng/dL) | Relative bias (%) | FT4 (ng/dL) | Relative bias (%) | FT4 (ng/dL) | Relative bias (%) | FT4 (ng/dL) | Relative bias (%) | FT4 (ng/dL) | Relative bias (%) | FT4 (ng/dL) | Relative bias (%) | |
| 0 | 0.76 | 0.0 | 1.34 | 0.0 | 5.05 | 0.0 | 0.72 | 0.0 | 1.27 | 0.0 | 4.62 | 0.0 |
| 12.5 | 0.81 | 6.0 | 1.38 | 3.0 | 5.24 | 3.8 | 0.68 | -5.6 | 1.21 | -4.7 | 4.40 | -4.8 |
| 25 | 0.82 | 6.7 | 1.42 | 6.0 | 5.43 | 7.4 | 0.68 | -5.0 | 1.19 | -5.9 | 4.44 | -3.9 |
| 50 | 0.86 | 12.2 | 1.47 | 9.7 | 5.79 | 14.6 | 0.66 | -8.4 | 1.17 | -7.9 | 4.42 | -4.4 |
| 100 | 0.93 | 21.6 | 1.60 | 19.9 | 6.46 | 27.8 | 0.67 | -6.8 | 1.17 | -7.5 | 4.69 | 1.5 |
| 200 | 1.08 | 40.6 | 1.82 | 36.0 | > 7.77 | > 53.9 | 0.74 | 3.2 | 1.34 | 5.9 | 6.43 | 39.2 |
| 400 | 2.39 | 212.0 | 3.91 | 192.5 | > 7.77 | > 53.9 | 0.95 | 32.5 | 1.62 | 27.7 | 6.65 | 43.8 |
Relative biases >10% are in bold.
Abbreviation: FT4, free thyroxine.
In conclusion, Elecsys FT4 III showed decreased biotin interference compared with the previous version of the assay, but an appreciable number of false results were noted. Therefore, laboratories, clinicians, and patients should be aware of the potential for biotin interference in IAs based on streptavidin-biotin binding and take measures to minimize the risk of false results. If biotin interference is suspected, additional methods can be used to reduce the risk of false results, including analysis with an alternative assay that is not based on biotin, use of a sample diluent, or removal of excess biotin using streptavidin-coated beads [7,8]. Furthermore, based on population pharmacokinetics of exogenous biotin, we recommend that samples from patients receiving therapy with high doses of biotin (i.e., >5 mg/day) should be assayed at least eight hrs following final biotin administration [3,4,8].
ACKNOWLEDGEMENTS
None declared.
Footnotes
AUTHOR CONTRIBUTIONS: SGY designed the study. JC and SGY contributed to the implementation of the research, analysis of the results, and writing of the manuscript.
CONFLICTS OF INTEREST: None declared.
RESEARCH FUNDING: None declared.
References
- 1.Piketty ML, Polak M, Flechtner I, Gonzales-Briceño L, Souberbielle JC. False biochemical diagnosis of hyperthyroidism in streptavidin-biotin-based immunoassays: the problem of biotin intake and related interferences. Clin Chem Lab Med. 2017;55:780–788. doi: 10.1515/cclm-2016-0606. [DOI] [PubMed] [Google Scholar]
- 2.Livaniou E, Evangelatos GP, Ithakissios DS, Yatzidis H, Koutsicos DC. Serum biotin levels in patients undergoing chronic hemodialysis. Nephron. 1987;46:331–332. doi: 10.1159/000184381. [DOI] [PubMed] [Google Scholar]
- 3.Grimsey P, Frey N, Bendig G, Zitzler J, Lorenz O, Kasapic D, et al. Population pharmacokinetics of exogenous biotin and the relationship between biotin serum levels and in vitro immunoassay interference. Int J Pharmacokinet. 2017;2:247–256. [Google Scholar]
- 4.Peyro Saint Paul L, Debruyne D, Bernard D, Mock DM, Defer GL. Pharmacokinetics and pharmacodynamics of MD1003 (high-dose biotin) in the treatment of progressive multiple sclerosis. Expert Opin Drug Metab Toxicol. 2016;12:327–344. doi: 10.1517/17425255.2016.1136288. [DOI] [PubMed] [Google Scholar]
- 5.Lipner SR. Rethinking biotin therapy for hair, nail, and skin disorders. J Am Acad Dermatol. 2018;78:1236–1238. doi: 10.1016/j.jaad.2018.02.018. [DOI] [PubMed] [Google Scholar]
- 6.Katzman BM, Lueke AJ, Donato LJ, Jaffe AS, Baumann NA. Prevalence of biotin supplement usage in outpatients and plasma biotin concentrations in patients presenting to the emergency department. Clin Biochem. 2018;60:11–16. doi: 10.1016/j.clinbiochem.2018.07.004. [DOI] [PubMed] [Google Scholar]
- 7.Gifford JL, Sadrzadeh SMH, Naugler C. Biotin interference: underrecognized patient safety risk in laboratory testing. Can Fam Physician. 2018;64:370. [PMC free article] [PubMed] [Google Scholar]
- 8.Trambas C, Lu Z, Yen T, Sikaris K. Depletion of biotin using streptavidin-coated microparticles: a validated solution to the problem of biotin interference in streptavidin–biotin immunoassays. Ann Clin Biochem. 2018;55:216–226. doi: 10.1177/0004563217707783. [DOI] [PubMed] [Google Scholar]
