Skip to main content
Scientific Reports logoLink to Scientific Reports
. 2019 Dec 27;9:20178. doi: 10.1038/s41598-019-56239-8

Measurement of Retinal Microvascular Blood Velocity Using Erythrocyte Mediated Velocimetry

Breanna M Tracey 1, Lakyn N Mayo 2, Christopher T Le 1, Victoria Y Chen 1, Julian Weichsel 3, Corinne Y Renner 1, Jiaqi Li 1, Joby J Tsai 4, Sachin P Kalarn 1, Michael T Ou 5, Luke Y Chang 6, Taariq K Mohammed 1, Isa S K Mohammed 1, Khelly A Shah 7, Ginger M Thompson 1, Anja M K Jones 7, Lily T Im 1, Mona A Kaleem 1, Osamah J Saeedi 1,
PMCID: PMC6934793  PMID: 31882799

Abstract

Changes in retinal blood flow may be involved in the pathogenesis of glaucoma and other ocular diseases. Erythrocyte mediated velocimetry (EMV) is a novel technique where indocyanine green (ICG) dye is sequestered in erythrocyte ghosts and autologously re-injected to allow direct visualization of erythrocytes for in vivo measurement of speed. The purpose of this study is to determine the mean erythrocyte speed in the retinal microvasculature, as well as the intravisit and intervisit variability of EMV. Data from 23 EMV sessions from control, glaucoma suspect, and glaucoma patients were included in this study. In arteries with an average diameter of 43.11 µm ± 6.62 µm, the mean speed was 7.17 mm/s ± 2.35 mm/s. In veins with an average diameter of 45.87 µm ± 12.04 µm, the mean speed was 6.05 mm/s ± 1.96 mm/s. Intravisit variability, as measured by the mean coefficient of variation, was 3.57% (range 0.44–9.68%). Intervisit variability was 4.85% (range 0.15–8.43%). EMV may represent reliable method for determination of retinal blood speed, potentially allowing insights into the effects of pharmacologic agents or pathogenesis of ocular diseases.

Subject terms: Optic nerve diseases, Translational research, Imaging and sensing

Introduction

Numerous systemic and ocular diseases have been associated with alterations in retinal blood flow (RBF)1,2. Local changes in RBF are implicated in the pathogenesis of major causes of blindness such as diabetic retinopathy and glaucoma3,4. Furthermore, retinal vasculature is readily accessible for imaging, allowing a window into alterations of blood flow associated with systemic and neurodegenerative diseases such as Alzheimer’s dementia5. While structural changes in retinal vasculature, such as change in vessel caliber or increased tortuosity, are established markers of hypertension6, diabetes7, vasculitis8, and neurodegenerative diseases9, changes in ocular blood flow have also been linked to these diseases10. Dynamic changes in RBF in ocular and systemic disease11 may precede these known structural changes, making RBF an important physical biomarker.

Numerous methods to measure RBF have been developed, but there is no established gold standard12 for the measurement of RBF in individual vessels in vivo that is accurate, precise, and capable of determining flow of multiple vessels over a wide field of view. Color doppler imaging, laser speckle imaging13, and more recently OCT angiography14 allow for the measurement of relative blood velocity and flow, but not absolute flow. Non-invasive techniques that record RBF in absolute terms include laser doppler imaging and the retinal function imager (RFI), whose coefficients of variation range from 10–11% for the RFI15,16 to 20% for laser doppler imaging using the Canon Laser Blood Flow Meter17. Adaptive Optics - Scanning Laser Ophthalmoscopy (AO-SLO) also allows for quantification of flow in the microvasculature, but in a relatively small field of view18. Furthermore, there are few comparison studies of different devices that measure RBF. The current gold standard for RBF measurement uses fluorescent and radioactive microspheres in animal models19. However, this method cannot be easily transferred to humans in vivo, and radioactive microspheres similarly have a coefficient of variation of 15–50%20. There is a critical need for an accurate and highly reproducible gold standard for human RBF quantification to validate devices and to establish the use of ocular microvascular blood flow as a robust biomarker for ocular and systemic disease.

Erythrocyte mediated angiography (EMA) is a technique in which indocyanine green (ICG) dye is sequestered in the erythrocytes to allow direct visualization of erythrocytes in vivo21. Prior work has emphasized its role in characterizing vasomotion in the eye22. Here, we describe methods used to quantify microvascular erythrocyte speed in vivo using a novel variant: Erythrocyte Mediated Velocimetry (EMV). We assess the precision of erythrocyte speed through intravisit and intervisit variability of EMV speed measurements in small arterioles and venules using a commercially available scanning laser ophthalmoscope, which was specifically modified for this purpose.

Methods

Participants

We enrolled glaucoma, glaucoma suspect, and control patients at least 40 years of age. The diagnosis of each patient was determined by a fellowship-trained glaucoma specialist. The diagnosis for glaucoma was made using the preferred practice patterns of the American Academy of Ophthalmology, specifically evidence of characteristic optic nerve damage characterized by retinal nerve fiber layer structural abnormalities or reliable and reproducible visual field abnormalities23. Glaucoma suspect was assigned to patients with normal visual field testing and either consistently elevated intraocular pressure or a suspicious-appearing optic disc24. Twenty-six EMV sessions were conducted on 35 eyes of 18 individual patients from May 19th, 2016 to October 19th, 2018. As shown in Fig. 1, we included data from 23 EMV sessions of 29 eyes of 16 individual patients. EMV sessions were considered ungradable if erythrocytes could not be resolved in the vessels for tracking. This was due to media opacity such as dry eye or cataract, or excessive eye movement which precluded image registration. Intravisit variability included only data from control patients. Intervisit variability included data from the 5 subjects who underwent multiple EMV sessions. Blood pressure (BP), pulse (HR), pulse oximetry, intraocular pressure (IOP), keratometry, ocular exam, ocular medications, and co-morbid disease were recorded. Patient demographics are shown in Table 1. Informed consent was obtained from all patients after discussion of the risks and benefits of the study. The study was conducted in accordance with the Declaration of Helsinki, and the study protocol was approved by the institutional review board of the University of Maryland Baltimore.

Figure 1.

Figure 1

Flowchart showing inclusion and exclusion of patients.

Table 1.

Patient Demographics.

Patient Session (if multiple) Age Sex Race Diagnosis Eyes HR BP IOP
OD OS
1 a 46 M Black Control OD,OS 64 115/73 14 12
b 47 M Black Control OD,OS 61 120/69 15 15
2 a 55 M Black Control OD,OS 79 143/78 19 20
b 55 M Black Control OD,OS 80 134/91 21 21
3 a 61 M Black Glaucoma OD 66 130/84 15
b 62 M Black Glaucoma OD 61 131/84 12
4 65 F White Control OD, OS 82 144/93 17 15
5 56 F White Glaucoma OD,OS 62 122/68 14 17
6 69 F Black Glaucoma Suspect OD,OS 60 160/86 17 19
7 57 F Black Glaucoma OD,OS 64 126/70 21 12
8 a 53 M Black Glaucoma OD 75 109/75 11
b 54 M Black Glaucoma OD 69 116/69 11
c 54 M Black Glaucoma OD 71 117/73 14
9 a 55 F White Glaucoma Suspect OD,OS 50 111/73 16 13
b 55 F White Glaucoma Suspect OD,OS 54 126/73 15 16
c 56 F White Glaucoma Suspect OD,OS 62 112/76 11 13
10 56 F Black Glaucoma Suspect (OD), Glaucoma (OS) OD,OS 50 103/60 15 38
11 60 F White Glaucoma Suspect OD 54 97/50 11
12 57 F White Glaucoma OD,OS 93 137/86 13 14
13 66 F White Glaucoma Suspect OD,OS 60 139/82 30 17
14 64 F Black Glaucoma OD,OS 71 138/89 20 18
15 61 F Black Glaucoma Suspect OD,OS 63 118/56 15 15
16 64 F White Glaucoma Suspect OD,OS 68 128/76 18 20

Erythrocyte preparation

Cell preparation was conducted as described by Flower et al.21,22. Up to 34 mL of blood was drawn from each subject. Using sterile procedures, erythrocyte ghosts produced from the blood were loaded with ICG dye. Following pupillary dilation with Tropicamide 0.5%, up to 1 mL of autologous ICG-loaded erythrocytes were injected intravenously. These cells persisted and were used to measure erythrocyte speed for up to three hours for one session. As previously described, ICG-loaded erythrocytes have comparable physical properties to native erythrocytes22.

Image acquisition

A Heidelberg Retinal Angiograph (HRA) 2 (Heidelberg Engineering GmbH, Germany) was used to acquire 10–20 second angiograms of subjects undergoing EMV. Angiograms of the disc, macula, and peripapillary retina were obtained in both eyes of all subjects except for one monocular patient. Angiograms were acquired with a custom scan pattern at 24.6 frames per second with a 15-degree horizontal and 7.5-degree vertical field of view. Conventional ICG or fluorescein angiography images were obtained concurrently or at the conclusion of each EMV session.

Image registration

To account for eye motion, images were registered by aligning the entire angiogram sequence to a reference image. We used a custom MATLAB (MathWorks, version 2018a) script that performed spatial domain registration with algorithms utilizing the overall similarities of the images. The registration script allows for rotation and translation to align the image with a user-selected reference image. No scale change was permitted to preserve the calculated scale. Both the mean squares and Mattes mutual information metrics were used independently to register the images, with the final compiled registered sequence consisting of the optimally registered frames based on either of the two metrics, as selected by the user. This effectively eliminated eye motion and generated a consistent set of coordinates to allow for accurate speed measurement.

Image registration validation

In addition to visual inspection of registered images, registration was quantitatively validated by comparing the stability between frames in unregistered and registered EMV angiograms. The stationary RBC closest to the optic disc and clearly visualized in all angiogram images was used as a reference point. Motion between frames was quantified by computing the straight-line distance between the stationary RBC on adjacent frames. A straight-line distance of more than three pixels between the stationary RBC coordinates on two adjacent frames was classified as poor stability. The number of frames with poor stability was compared between corresponding unregistered and registered angiogram sequences. To validate image registration, we chose five representative sequences chosen for their varying degrees of eye motion, one with minimal eye motion, one with moderate eye motion, and three with substantial eye motion, as judged by visual inspection. We compared the percentage of frames with poor stability before and after registration.

Erythrocyte speed measurement

A protocol for speed measurement was developed that consists of manual erythrocyte tracking followed by erythrocyte speed measurement using a custom MATLAB script. Individual erythrocytes were tracked as they flowed in a vessel over multiple EMV image frames (Fig. 2). Tracking was only done when there was clearly one ICG labelled erythrocyte flowing in a vessel segment in a given time without any other ICG labelled erythrocytes that would create uncertainty. To further standardize erythrocyte tracking, cells were eligible for tracking only if the same cell was visible on at least three consecutive frames within the vessel of interest. Only erythrocytes visible on the retina were tracked, and those in the optic disc were excluded. After erythrocyte tracking, speed was determined in MATLAB using a semi-automated custom script. For standardization, the center of the brightest pixel was recorded as the cell location. After recording all cell locations, the user defined the vessel path (Fig. 3). The distance along the vessel course in pixels was converted to microns using the conversion scale provided by the Heidelberg HRA2 software, accounting for keratometry25. Speed was calculated using the vessel distance traveled by an erythrocyte over two adjacent frames and dividing by the time between frames. Mean speed was obtained by averaging the individual erythrocyte speed measurements. At least 30 individual measurements were required. Angiograms were only eligible for inclusion if they included at least 85 frames.

Figure 2.

Figure 2

Example of erythrocyte tracking from erythrocyte mediated velocimetry images. Red circles show an individual erythrocyte flowing in a small vein along the retina (ae).

Figure 3.

Figure 3

EMV image showing coordinates of many individual erythrocytes from an angiogram sequence (blue) overlaid on user-drawn vessel path (red).

For determination of erythrocyte speed, one vessel chosen at random was analyzed per eye for each included EMV session. Due to concerns about the temporal resolution of imaging at 24.6 frames per second, we conservatively chose to exclude arteries over 60 µm and veins over 80 µm. While our imaging technique allows for measurement of erythrocyte speeds up to 25 mm/s, we conservatively chose this criteria for vessel caliber to place our mean speeds less than 15 mm/s26. For calculations of intravisit and intervisit variability, to account for variations in the field of view in each angiogram, we ensured the same region of the vessel was analyzed. This vessel region was chosen adjacent to the optic disc and limited to no more than 100 pixels in the horizontal direction, approximately 1 mm. In the case of one macular vessel, the region was chosen to include the terminal end of the vessel, which was best visualized on the angiogram.

Vessel diameter measurement

The diameter of each vessel was determined from registered sequences of conventional ICG or fluorescein angiograms obtained concurrently or immediately after the EMV imaging session. Conventional ICG images were obtained at 24.6 frames per second with a 15° horizontal and 7.5° vertical field of view. Fluorescein angiography images were obtained at 4.7 frames per second with a 30° horizontal and 30° vertical field of view. Diameter measurements were obtained using a custom MATLAB script to trace vessel boundaries and determine the average diameter over the course of the vessel segment. To account for possible variation in vessel diameter with the cardiac cycle27, diameter measurements were obtained on five separate images, each approximately 200 msec apart, and averaged.

Statistical analysis

Interrater reliability was evaluated with the intraclass correlation coefficient (ICC), two-way mixed, absolute agreement28, using 380 speed measurements of two independent graders on a standard set of angiograms. Intravisit variability, or the variability of erythrocyte speed within a single EMV session, was evaluated with 10 vessels (3 arteries and 7 veins) in six eyes of three control patients imaged twice within the same EMV session. We also evaluated intravisit variability in 10 vessels (3 arteries and 7 veins) of five eyes of four glaucoma patients. Intervisit variability, or the variability of erythrocyte speed across time, was evaluated with 10 vessels (3 arteries and 7 veins) of eight eyes of five patients who underwent two EMV sessions separated by at least 60 days. For calculation of intervisit variability, if two angiograms of a given vessel from the same visit were available, the measurements from both angiograms were combined into a single mean speed. If patients had angiograms from three EMV sessions, the two visit days with the highest quality angiograms were used for comparison. The coefficient of variation (CV) was used for determination of intravisit and intervisit variability. The CV is defined as the ratio of the standard deviation (SD) to the mean, as shown in Eq. (1).

CV=SDmean 1

The CV was calculated for each vessel individually, using mean speeds obtained from the two angiograms. Mean ocular perfusion pressure (MOPP) was computed for each EMV session using the patient’s systolic blood pressure (SBP), diastolic blood pressure (DBP) and IOP, as shown in Eq. (2)29.

MOPP=23(13SBP+23DBP)IOP 2

Overall mean speed in retinal arteries and veins was determined by averaging mean speeds obtained from individual angiograms from all included EMV sessions. Data is presented as mean ± SD. All statistical analysis was performed in SPSS Statistics (IBM Corporation, version 24.0).

Results

Image registration

The five sequences used for quantitative image registration validation exhibited minimal, average, and high eye motion, judged by visual inspection. In the unregistered images, the sequences had a mean of 12.5% frames with poor stability. In the corresponding registered image sequences, the angiograms had a mean of 0.9% frames with poor stability (p < 0.01), effectively validating the MATLAB image registration script.

Variability of speed

The interrater variability, defined as the ICC for two independent graders calculating erythrocyte speed, is 0.983. The intravisit variability, defined as the average CV between two angiograms obtained in the same EMV session, is 3.57% (range 0.44–9.68%) in control subjects (Tables 2) and 6.66% in glaucoma patients (P > 0.10). The intervisit variability, defined as the average CV between two EMV sessions, is 4.85% (range 0.15–8.43%), as shown in Table 3. The mean time between visits was 220 days. Graphical representation of the intravisit and intervisit variability is shown in Fig. 4.

Table 2.

Intravisit Speed – Comparison of speed from angiograms obtained within the same EMV session.

Patient Eye Vessel Diameter, µm (SD) Angiogram #1 Angiogram #2 CV
Velocity, mm/s (SD) Velocity, mm/s (SD)
1a OD Vein 39.74 (7.27) 4.14 (1.54) 4.30 (1.78) 2.69%
1a OS Vein 35.79 (5.33) 3.17 (0.76) 3.46 (0.73) 6.24%
1b OD Vein 34.69 (5.74) 4.54 (1.98) 4.56 (1.37) 0.44%
1b OD Vein 40.82 (6.57) 4.85 (1.24) 4.23 (1.18) 9.68%
2a OD Artery 34.34 (6.90) 7.22 (2.19) 7.07 (2.47) 1.45%
2a OD Vein 36.35 (6.23) 4.63 (1.36) 4.97 (1.82) 4.87%
2a OS Vein 53.00 (9.77) 5.14 (1.02) 5.25 (1.18) 1.46%
2b OD Artery 35.05 (6.02) 7.98 (2.83) 7.48 (1.89) 4.54%
4 OD Vein 29.88 (5.13) 7.19 (1.87) 7.29 (1.61) 1.02%
4 OS Artery 44.30 (4.51) 10.87 (3.80) 11.39 (5.52) 3.30%
Mean CV 3.57%

Superscripts denote patient imaging session (if multiple). CV is defined as the ratio of the standard deviation (SD) to the mean.

Table 3.

Intervisit Speed – Comparison of speed from angiograms obtained from two separate EMV sessions.

Patient Eye Vessel Diameter, µm (SD) Visit #1 Visit #2 CV
Velocity, mm/s (SD) Velocity, mm/s (SD)
1a,b OD Vein 39.74 (7.27) 4.08 (1.63) 4.49 (1.23) 6.80%
1a,b OS Vein 35.79 (5.33) 3.35 (0.76) 3.22 (1.46) 2.72%
2a,b OD Artery 34.34 (6.90) 7.12 (2.38) 7.74 (2.43) 5.92%
2a,b OD Vein 36.35 (6.23) 4.79 (1.60) 4.53 (1.72) 3.88%
2a,b OS Vein 53.00 (9.77) 5.58 (1.05) 5.96 (1.43) 4.70%
3a,b OD Vein 38.59 (6.27) 6.60 (1.46) 7.43 (1.70) 8.43%
8a,b OD Vein 42.92 (7.93) 7.15 (2.42) 7.14 (1.69) 0.15%
8b,c OD Artery 39.00 (7.80) 7.57 (3.01) 8.38 (2.76) 7.21%
9a,b OS Vein 67.75 (13.75) 8.29 (3.06) 8.40 (3.75) 0.98%
9a,c OD Artery 49.62 (9.48) 7.06 (3.86) 6.33 (2.92) 7.76%
Mean CV  4.85%

Superscripts denote which patient imaging sessions were compared. CV is defined as the ratio of the standard deviation (SD) to the mean.

Figure 4.

Figure 4

Difference in velocity as a function of the mean velocity between angiograms acquired in the same session (a) or separate sessions. (b) The center line represents the mean difference in velocity between sessions. The outer lines represent the mean difference ±1.96(SD).

Erythrocyte speed

Of 16 patients, 3 (18.8%) patients had diabetes, 9 (56.3%) had hypertension, 8 (50%) had dyslipidemia, and 6 (37.5%) of patients were on topical glaucoma medications. Using one vessel from each eye, mean erythrocyte speed was calculated in small retinal arteries and veins, as shown in Table 4. MOPP was also calculated for each eye. Eighteen arteries less than 60 µm in diameter and 22 veins less than 80 µm were analyzed. Mean arterial erythrocyte speed was 7.17 mm/s ± 2.35 mm/s in an artery with an average diameter of 43.11 µm ± 6.62 µm. Mean venous erythrocyte speed was 6.05 mm/s ± 1.96 mm/s in a vein with an average diameter of 45.87 µm ± 12.04 µm.

Table 4.

Erythrocyte speed in small retinal arteries and veins.

Patient Eye MOPP Vessel Diameter, µm (SD) Velocity, mm/s (SD)
Arteries
2a OD 47.44 Artery 34.34 (6.90) 7.10 (2.36)
4 OS 58.33 Artery 44.30 (4.51) 10.96 (4.68)
5 OD 43.33 Artery 42.73 (9.27) 7.19 (2.92)
5 OS 40.33 Artery 43.17 (5.50) 6.70 (2.77)
6 OD 54.78 Artery 41.53 (10.36) 7.58 (2.40)
7 OD 38.11 Artery 40.58 (6.62) 7.14 (2.25)
8b OD 45.44 Artery 39.00 (7.80) 7.82 (3.24)
8c OD 44.44 Artery 39.99 (4.27) 8.30 (2.76)
9a OD 41.11 Artery 46.97 (7.53) 5.18 (2.39)
9b OD 45.44 Artery 52.72 (10.96) 6.47 (2.66)
9c OD 47.67 Artery 45.73 ((6.11) 6.21 (2.87)
10 OS 11.56 Artery 49.33 (7.72) 3.77 ((3.81)
11 OD 36.78 Artery 34.81 (5.94) 4.57 (2.75)
12 OD 55.67 Artery 38.50 (6.05) 7.76 (2.95)
12 OS 54.67 Artery 42.78 (7.46) 11.02 (4.47)
13 OD 37.33 Artery 31.18 (1.63) 2.19 (0.89)
14 OD 50.44 Artery 54.72 (12.05) 10.72 ((7.41)
16 OD 44.22 Artery 53.62 (4.65) 8.32 (3.59)
Average Arterial Velocity, mm/s (SD) 7.17 (2.35)
Average Arterial Diameter, µm (SD) 43.11 (6.62)
Veins
1a OD 44 Vein 39.74 (7.27) 4.14 (1.62)
1a OS 46 Vein 35.79 (5.33) 3.32 (0.78)
1b OD 42.33 Vein 40.82 (6.57) 4.82 (1.47)
1b OS 42.33 Vein 38.12 (5.73) 3.21 (1.45)
2a OS 46.44 Vein 53.00 (9.77) 4.42 (1.35)
2b OD 49.22 Vein 34.69 (5.74) 4.53 (1.72)
2b OS 49.22 Vein 49.06 (10.88) 4.49 (1.71)
3a OD 51.22 Vein 38.59 (6.27) 7.51 (1.74)
3b OD 54.44 Vein 38.82 (7.60) 6.61 (1.41)
4 OD 56.33 Vein 29.88 (5.13) 7.15 (1.71)
6 OD 56.78 Vein 43.19 (8.66) 4.79 (1.47)
7 OS 47.11 Vein 42.13 (8.02) 8.49 (3.94)
8a OD 46.56 Vein 42.92 (7.93) 7.20 (2.38)
9a OS 44.11 Vein 67.75 (13.75) 8.01 (3.14)
9b OS 44.44 Vein 71.44 (7.68) 8.42 (3.50)
9c OS 45.67 Vein 68.10 (8.52) 9.00 (3.74)
10 OD 34.56 Vein 44.49 (8.18) 4.50 (2.08)
13 OS 50.33 Vein 40.02 (4.00) 4.17 (1.35)
14 OS 52.44 Vein 68.30 (6.29) 8.57 (4.13)
15 OD 36.11 Vein 40.63 (2.85) 9.01 (4.53)
15 OS 36.11 Vein 40.85 (3.13) 5.61 (1.64)
16 OS 42.22 Vein 40.85 (2.05) 5.18 (2.34)
Average Venous Velocity, mm/s (SD) 6.05 (1.96)
Average Venous Diameter, µm (SD) 45.87 (12.04)

Superscripts denote patient imaging session (if multiple).

Discussion

We present the results of the first study utilizing EMV, a novel technique, to determine microvascular retinal blood speeds. EMV is a promising and highly precise method for determination of retinal microvascular blood velocity. Variability of quantitative absolute retinal blood flow and blood velocity measurements using other imaging techniques have ranged from 10–11% (RFI)15,16 to a mean of 20% (Canon Laser Blood Flowmeter)17, and our values were comparable or better.

Retinal velocity measurements obtained using EMV are comparable to previously published velocities using laser doppler26 in vessels of similar caliber. For a 42 µm arteriole, Riva, et al.26 found a mean velocity of approximately 7.5 mm/s, which is similar to 7.17 mm/s found in our study. For a 45 µm venule, mean velocity was approximately 6 mm/s, which is similar to our 6.05 mm/s. A relative advantage of EMV is that while Riva, et al. measured velocities in vessels larger than 40 µm in diameter, EMV allows for velocity determination in smaller vessels. Furthermore, by directly measuring the speed of individual erythrocyte ghosts, EMV offers the advantage of quantifying absolute speed. AO-SLO offers a similar ability to quantify flow in the retinal microvasculature, but it does so over a small field a view using equipment that is not commercially available30. EMV in comparison is more invasive, but it allows for visualization over a larger field of view with a modified commercially available device.

Despite the more invasive nature of this procedure, we have now imaged 26 patients and shown that it is relatively safe. It may, in fact, be safer than traditional angiography as the aggregate concentration of sequestered ICG dye using EMV is approximately 1/700th that of conventional ICG. Given that the cells are autologously reinjected there is a concern for risk of infection. Sterile technique is strictly adhered to minimize this potential risk. To date, only one patient had an adverse event when they experienced a vasovagal episode after injection of the conventional ICG dye.

This study had broad inclusion criteria, and hence we imaged some patients who did not have gradable angiograms due to media opacity such as cataract or dry eye or due to the learning curve of imaging patients with this new technology. Our broad inclusion criteria resulted in a cohort of patients demographically similar to those at risk for ocular disease with an average age of 57.8 years, and 25% with diagnoses of dry eye or ocular surface disease.

While we chose to assess mean retinal erythrocyte speeds, we note that other metrics such as median, minimum, maximum, range, and distribution may be as important if not more important given the natural physiologic variability of blood flow. Physiologic variability of erythrocyte speed and flow comes from the pulsatile nature of flow, higher in systole and lower in diastole, as well as the distribution of speeds of individual erythrocytes flowing through the vasculature, where cells travelling in the vessel periphery are slower than the cells traveling in the center of the lumen. The potential difference in speed with time and luminal position is why we chose to report mean speeds. The mean speed for each vessel was determined by averaging many individual erythrocyte speeds (mean of 70 individual speed measurements per angiogram) to account for these differences. Given the time-consuming nature of manual speed determination, we have since developed a more automated method of determining erythrocyte speeds from EMV (automated tracking videos available)31.

We focused this manuscript on erythrocyte speed as opposed to overall retinal blood flow due to potential differences in vessel caliber in measuring vessel diameter using fluorescein and using ICG angiography, which were both used to determine vessel caliber in this study. Vessel diameter is an important component of blood flow determination, and future studies will be designed to take this into account. EMV sessions were generally conducted at the same time of day and it is possible that measurements could have been affected by diurnal variation. For individuals with glaucoma, it is possible that worsening of the disease in between imaging sessions could affect erythrocyte velocity measurements. Other limitations include the relatively small sample size and the varied diagnoses of this cohort that may affect retinal blood speed. While we did compare the intravisit variability in controls versus glaucoma and found no statistically significant difference, this sample was not large enough to study those differences in a robust manner. Future studies will explore changes in retinal blood flow with different conditions and disease states, such as glaucoma.

With the continued development of noninvasive methods of determining ocular blood flow, EMV may represent a reliable method for determination of retinal blood flow. The potentially accurate and reproducible measurement of erythrocyte speed further allows for determining the effect of disease states or pharmacologic agents on ocular blood flow – even if small in magnitude.

Acknowledgements

This research was supported in part by Fight for Sight and the Program for Research Initiated by Students and Mentors (PRISM), University of Maryland School of Medicine Office of Student Research. Dr. Saeedi is supported by NIH Grant K23 EY025014.

Author contributions

B.T. and O.S. wrote the manuscript. L.M., J.T., M.O., L.C., A.J., G.T., S.K., and O.S. performed cell preparation, imaged patients, and provided administrative support for the study. B.T., L.M., C.L., V.C., C.R., J.L., S.C., T.M., I.M., and K.S. performed data analysis. J.W. provided custom software modifications for the imaging platform. L.I., M.K., an O.S. provided clinical care for and enrolled patients. All authors critically reviewed the manuscript.

Data availability

The dataset generated and analyzed during the current study is available from the corresponding author on reasonable request.

Competing interests

J.W. is employed by Heidelberg Engineering GmbH, who provided technical support and imaging devices used in the study. O.S. received occasional travel support from Heidelberg Engineering GmbH. The remaining authors have no competing interests to declare.

Footnotes

Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Gameiro GR, et al. Retinal tissue hypoperfusion in patients with clinical Alzheimer’s disease. Eye Vis (Lond) 2018;5:21. doi: 10.1186/s40662-018-0115-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Lee B, et al. En Face Doppler Optical Coherence Tomography Measurement of Total Retinal Blood Flow in Diabetic Retinopathy and Diabetic Macular Edema. JAMA Ophthalmol. 2017;135:244–251. doi: 10.1001/jamaophthalmol.2016.5774. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Mozaffarieh M, Grieshaber MC, Flammer J. Oxygen and blood flow: players in the pathogenesis of glaucoma. Mol Vis. 2008;14:224–233. [PMC free article] [PubMed] [Google Scholar]
  • 4.Pemp B, Schmetterer L. Ocular blood flow in diabetes and age-related macular degeneration. Can. J. Ophthalmol. 2008;43:295–301. doi: 10.3129/i08-049. [DOI] [PubMed] [Google Scholar]
  • 5.Jiang H, et al. Altered Macular Microvasculature in Mild Cognitive Impairment and Alzheimer Disease. J Neuroophthalmol. 2018;38:292–298. doi: 10.1097/WNO.0000000000000580. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Cheung CY-L, et al. Retinal vascular tortuosity, blood pressure, and cardiovascular risk factors. Ophthalmology. 2011;118:812–818. doi: 10.1016/j.ophtha.2010.08.045. [DOI] [PubMed] [Google Scholar]
  • 7.Bek T. Diameter Changes of Retinal Vessels in Diabetic Retinopathy. Curr. Diab. Rep. 2017;17:82. doi: 10.1007/s11892-017-0909-9. [DOI] [PubMed] [Google Scholar]
  • 8.Liew G, et al. Retinal vessel caliber changes in vasculitis. Retina (Philadelphia, Pa.) 2015;35:803–808. doi: 10.1097/IAE.0000000000000369. [DOI] [PubMed] [Google Scholar]
  • 9.Ikram MK, et al. Retinal vascular calibers associate differentially with cerebral gray matter and white matter atrophy. Alzheimer Dis Assoc Disord. 2013;27:351–355. doi: 10.1097/WAD.0b013e31829344ed. [DOI] [PubMed] [Google Scholar]
  • 10.Tayyari F, et al. Retinal Blood Flow and Retinal Blood Oxygen Saturation in Mild to Moderate Diabetic Retinopathy. Invest. Ophthalmol. Vis. Sci. 2015;56:6796–6800. doi: 10.1167/iovs.15-17481. [DOI] [PubMed] [Google Scholar]
  • 11.Ciulla L, et al. The role of hypertension in retinal blood flow alterations in open-angle glaucoma patients. Acta Ophthalmol. 2017;95:e794–e795. doi: 10.1111/aos.13365. [DOI] [PubMed] [Google Scholar]
  • 12.Wei X, et al. Assessment of flow dynamics in retinal and choroidal microcirculation. Surv Ophthalmol. 2018;63:646–664. doi: 10.1016/j.survophthal.2018.03.003. [DOI] [PubMed] [Google Scholar]
  • 13.Rege A, et al. Noninvasive Assessment of Retinal Blood Flow Using a Novel Handheld Laser Speckle Contrast Imager. Transl Vis Sci Technol. 2018;7:7. doi: 10.1167/tvst.7.6.7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Spaide RF, Fujimoto JG, Waheed NK, Sadda SR, Staurenghi G. Optical coherence tomography angiography. Prog Retin Eye Res. 2018;64:1–55. doi: 10.1016/j.preteyeres.2017.11.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Burgansky-Eliash Z, Bartov E, Barak A, Grinvald A, Gaton D. Blood-Flow Velocity in Glaucoma Patients Measured with the Retinal Function Imager. Curr. Eye Res. 2016;41:965–970. doi: 10.3109/02713683.2015.1080278. [DOI] [PubMed] [Google Scholar]
  • 16.Harris, A., Jonescu-Cuypers, C., Kagemann, L., Ciulla, T. & Krieglstein, G. Atlas of Ocular Blood Flow: Vascular Anatomy, Pathophysiology, and Metabolism. (2003).
  • 17.Guan K, Hudson C, Flanagan JG. Variability and repeatability of retinal blood flow measurements using the Canon Laser Blood Flowmeter. Microvasc. Res. 2003;65:145–151. doi: 10.1016/S0026-2862(03)00007-4. [DOI] [PubMed] [Google Scholar]
  • 18.Gu B, et al. Noninvasive in vivo characterization of erythrocyte motion in human retinal capillaries using high-speed adaptive optics near-confocal imaging. Biomed. Opt. Express, BOE. 2018;9:3653–3677. doi: 10.1364/BOE.9.003653. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Shih Y-YI, et al. Quantitative Retinal and Choroidal Blood Flow During Light, Dark Adaptation and Flicker Light Stimulation in Rats Using Fluorescent Microspheres. Curr Eye Res. 2013;38:292–298. doi: 10.3109/02713683.2012.756526. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Nilsson, S. F. E. & Alm, A. A Determination of Ocular Blood Flow with the Microsphere Method. In Ocular Blood Flow 29 (Springer, 2012).
  • 21.Flower R, et al. Observation of erythrocyte dynamics in the retinal capillaries and choriocapillaris using ICG-loaded erythrocyte ghost cells. Invest. Ophthalmol. Vis. Sci. 2008;49:5510–5516. doi: 10.1167/iovs.07-1504. [DOI] [PubMed] [Google Scholar]
  • 22.Flower RW, Kling R. Observation and characterization of microvascular vasomotion using erythrocyte mediated ICG angiography (EM-ICG-A) Microvasc. Res. 2017;113:78–87. doi: 10.1016/j.mvr.2017.02.006. [DOI] [PubMed] [Google Scholar]
  • 23.Primary Open-Angle Glaucoma PPP - 2015. American Academy of Ophthalmology Available at, https://www.aao.org/preferred-practice-pattern/primary-open-angle-glaucoma-ppp-2015 (Accessed: 9th September 2019) (2015).
  • 24.Primary Open-Angle Glaucoma Suspect PPP - 2015. American Academy of Ophthalmology Available at, https://www.aao.org/preferred-practice-pattern/primary-open-angle-glaucoma-suspect-ppp-2015 (Accessed: 8th June 2019) (2015).
  • 25.Ctori I, Gruppetta S, Huntjens B. The effects of ocular magnification on Spectralis spectral domain optical coherence tomography scan length. Graefes Arch. Clin. Exp. Ophthalmol. 2015;253:733–738. doi: 10.1007/s00417-014-2915-9. [DOI] [PubMed] [Google Scholar]
  • 26.Riva CE, Grunwald JE, Sinclair SH, Petrig BL. Blood velocity and volumetric flow rate in human retinal vessels. Invest. Ophthalmol. Vis. Sci. 1985;26:1124–1132. [PubMed] [Google Scholar]
  • 27.Chen HC, Patel V, Wiek J, Rassam SM, Kohner EM. Vessel diameter changes during the cardiac cycle. Eye (Lond) 1994;8(Pt 1):97–103. doi: 10.1038/eye.1994.19. [DOI] [PubMed] [Google Scholar]
  • 28.Koo TK, Li MY. A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability Research. Journal of Chiropractic Medicine. 2016;15:155–163. doi: 10.1016/j.jcm.2016.02.012. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Keer KV, Breda JB, Pinto LA, Stalmans I, Vandewalle E. Estimating Mean Ocular Perfusion Pressure Using Mean Arterial Pressure and Intraocular Pressure. Invest. Ophthalmol. Vis. Sci. 2016;57:2260–2260. doi: 10.1167/iovs.16-19375. [DOI] [PubMed] [Google Scholar]
  • 30.Joseph, A., Guevara-Torres, A. & Schallek, J. Imaging single-cell blood flow in the smallest to largest vessels in the living retina. Elife8 (2019). [DOI] [PMC free article] [PubMed]
  • 31.Wang D, Haytham A, Mayo L, Tao Y, Saeedi O. Automated retinal microvascular velocimetry based on erythrocyte mediated angiography. Biomed. Opt. Express, BOE. 2019;10:3681–3697. doi: 10.1364/BOE.10.003681. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The dataset generated and analyzed during the current study is available from the corresponding author on reasonable request.


Articles from Scientific Reports are provided here courtesy of Nature Publishing Group

RESOURCES