Abstract
Data provided with this article are relative to kinetic measures from standing posture trials in eye open and eye closed conditions of 15 healthy subjects, acquired from a dynamometric force plate and a Nintendo Wii Balance Board (NBB). Data have been originally collected for a research project aimed at evaluating the reliability of low-cost devices in clinical scenarios. Raw data from the force plate include three ground reaction force components, center of pressure trajectories and torque around the vertical axis. Raw data from the NBB consist of vertical component of the ground reaction force measured by each of the four device sensors. Processed data consist of synchronized center of pressure time-series from both devices, referred to the force plate reference frame. Data were acquired simultaneously from the devices, allowing a direct comparison between the kinetic measures provided by the gold-standard for posture analysis (dynamometric force plate) and a low-cost device (NBB). Utility of present data can be twofold: first they can be used to assess the overall quality of the NBB signals for posturographic analysis by a direct comparison with the same signals acquired from the gold-standard device for kinetic measurement. Secondly, data from the dynamometric force plate can be used per se to evaluate different kind of parameters useful to assess balance capabilities, also by comparing data from different sensorial conditions (eye open versus eye closed).
Keywords: Posture analysis, Balance, Center of pressure, Kinetic measurements, Nintendo Wii Balance Board
Specifications Table
| Subject | Biomedical Engineering |
| Specific subject area | Low cost devices for balance and posture analysis |
| Type of data | MATLAB data files (.mat) Excel file (.xlsx) |
| How data were acquired | Dynamometric platform (Bertec H4060) Nintendo Wii Balance Board |
| Data format | Raw data Processed data |
| Parameters for data collection | Data were acquired from 15 healthy young adults (8 males, 7 females, age 23.3 ± 3.0 years, weight 67.4 ± 9.7 kg, height, 1.68 ± 0.09 m) without any musculoskeletal or neurological disease which can affect posture and balance maintenance. |
| Description of data collection | Subjects maintained their posture upon a Nintendo Wii Balance board placed on a dynamometric force plate, mounted flush with the floor. Each subject performed two posture trials in an eyes open and eyes closed condition. Each trial lasted at least 30 s. Kinetic data were simultaneously acquired from the two devices. |
| Data source location | Movement Analysis Laboratory Department of Information Engineering Università Politecnica delle Marche Ancona Italy |
| Data accessibility | Data are within the article |
Value of the Data
|
1. Data
The dataset provided with this article contains data simultaneously acquired during unperturbed standing posture trials in 15 healthy and young subjects (male/female ratio 8/7, age 23.3 ± 3.0 years, weight 67.4 ± 9.7 kg, height, 1.68 ± 0.09 m) from a dynamometric force plate (FP, Bertec, H4060, Fig. 1A) and a Nintendo Wii Balance Board (NBB, Fig. 1B). All the 15 volunteers performed the standing trial in an eye-open and an eye-closed condition but for two subjects the data streamed from the NBB failed to be correctly saved and stored, hence resulting not available for further analysis. Therefore, for the eye-closed condition, data relative to a subset of 13 individuals can be found in the proposed dataset. The dataset consists of two different files: the first is a .xlsx file, named “Subjects_data”, where the anthropometric data of each subject (S_1, …, S_15) are reported, i.e. gender, age, weight and height. The second file is a .mat file, named “Dataset_posture.mat”, containing the posture trial data acquired from both devices.
Fig. 1.
Dynamometric platform (A) and Nintendo Wii Balance Board (B) reference frames. TL, TR, BL and BR indicate the Nintendo Wii Balance Board sensors. In panel C, the experimental setup is reported, where the subject maintains its upright stance while standing on the NBB, placed upon the FP.
The .mat file consists of four different structure files: RawDataEO, RawDataEC, SincDataEO and SincDataEC, where the suffix EO indicates structures containing data acquired during the eye open trials while EC stands for the structure where data acquired in the eye closed trials are reported. Each of the 4 structures contains another structure with data relative to each considered subject; therefore, as reported above, the eye open structures (RawDataEO and SincDataEO) are made by 15 different structures, while the eye closed structures (RawDataEC and SincDataEC) consist in 13 different structures. Each subject structure is named according with the nomenclature reported in the “Subjects_data.xlsx” file, from S_1 to S_15.
Each subject structure includes two MATLAB matrices, the first one, named “DP”, contains raw data acquired from the dynamometric force plate while in the second one, named “NWBB” there are the raw signals from the NBB. In RawData structures, the DP matrix contains six data columns: in the first one there are the time stamps (ms), while columns two, three and four contain the anterior-posterior (AP), medial-lateral (ML) and vertical components of the ground reaction force (N). Center of pressure (CoP) displacement (mm) in AP and ML directions is in the fifth and sixth columns, referred to the dynamometric force plate reference frame (Fig. 1A). The last column contains the torque (N⋅mm) around the platform vertical axis. The NWBB matrix has six columns, organized as follows: in the first one there are the time stamps of the acquisition (s). In the second column is reported the vertical ground reaction force component, corresponding to the average of the measures of the four NBB sensors. The third and fourth columns store the vertical force values (N) measured by the top left and right sensors (TL and TR, Fig. 1B), while the fifth and sixth columns contain vertical force values from bottom left and bottom right sensors (BL and BR, Fig. 1B). It is important to note that NBB sensors are able to measure only the vertical component of the ground reaction force and not also the two planar components, along the x and y axes (Fig. 1B).
In the SincData structures, the DP and NWBB matrices have two columns where the CoP displacement in the AP and ML directions is reported (first and second column, respectively). CoP data of both devices are temporally synchronized with a sampling rate equal to 500 Hz, low-pass filtered with a cut-off frequency of 15 Hz and referred to the FP reference frame (Fig. 1A).
2. Experimental design, materials, and methods
All testing procedures for each volunteer were performed within a single day. The dynamometric force plate (Bertec H4060, dimension 40 × 60 cm) was used as the gold-standard reference for kinetic measurements, mounted flush with the laboratory floor. The Nintendo Wii Balance Board (dimension 23.8 × 43.3 cm) was placed on top of the dynamometric FP and each subject stood barefoot on the NBB surface, with the feet within the rectangular borders imprinted on the NBB surface. Subjects were instructed to stand with their arms on the sides of the body and to remain as still as possible for the entire duration of the task (Fig. 1C). Each posture trials lasted at least 30 s. Each volunteer signed written informed consent prior the beginning of the test and the experimental procedures were approved by the local expert committee.
FP data were acquired by means of a professional movement analysis system (Elite, BTS-Bioengineering, Italy) with a sampling rate of 500 Hz, while NBB data were streamed through Bluetooth and collected by a free software (https://wiimotephysics.codeplex.com). NBB raw signals were also visually checked by the examiner and in case of glitches or artifacts the trial was aborted and repeated. The latter procedure resulted in completely clean data, without any amplitude spike or sample leakage, as also assessed by a further data examination before proceeding with the post-processing. NBB data presented time jitter and an inconsistent sampling rate for all the four sensors, close to 100 Hz [1,2] and therefore they were linearly interpolated in order to obtain proper time-series with 500 samples∙s−1, following what was reported in [3,4].
For each posture trial, the center of pressure was calculated and provided in SincData structures, representing a central feature for further analysis regarding balance capabilities and posture control [3,5,6]. CoP displacement for the FP were provided directly from the acquisition system and therefore they were reported in RawData structures (see Data section). However, due to the presence of the NBB upon the FP surface, the CoP trajectories of the FP needed to be recomputed by taking into account the thickness of the NBB. From raw FP data, the CoP displacement in both x and y directions was computed by the acquisition system from the following relations:
| (1) |
| (2) |
where Mx and My represent the torques on the x and y axis respectively; Fx, Fy and Fz represent the ground reaction force components on the x, y and vertical axis and h stands for the thickness (mm) above the top surface of any material covering the force plate. Therefore, by using (1), (2), the center of pressure trajectories for both axes have been recomputed with a value of h equal to the NBB height (55 mm). Due to the experimental setup, the x axis represents the AP direction and conversely the y axis the ML one (Fig. 1A). From raw NBB data, the center of pressure displacement was computed as follows:
| (3) |
| (4) |
where TL, TR, BL and BR are the vertical ground reaction force component measured by the four NBB sensors and Lx and Ly are the length of the NBB in the x and y dimensions (Fig. 1B). Due to the experimental setup, the x axis represents the ML direction and the y axis the AP one. All the center of pressure trajectories were expressed in mm. Onset of the measurements were synchronized following [3] and the NBB CoP trajectories were referred to the FP reference frame. It is worth noticing that the latter computation does not prevent to use present data to analyse CoP trajectories as time-series referenced to the mean CoP, as commonly performed in the majority of posture and balance related studies [4,5,7]. CoP time-series for both FP and NBB were low-pass filtered by a second-order, linear-phase Butterworth filter, with a cut-off frequency of 15 Hz.
Conflict of Interest
The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Footnotes
Supplementary data to this article can be found online at https://doi.org/10.1016/j.dib.2019.105028.
Appendix A. Supplementary data
The following are the Supplementary data to this article:
References
- 1.Bartlett H.L., Ting L.H., Bingham J.T. Accuracy of force and center of pressure measures of the Wii Balance Board. Gait Posture. 2014;39:224–228. doi: 10.1016/j.gaitpost.2013.07.010. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Mengarelli A., Verdini F., Cardarelli S., Di Nardo F., Burattini L., Fioretti S. Balance assessment during squatting exercise: a comparison between laboratory grade force plate and a commercial, low-cost device. J. Biomech. 2018;71:264–270. doi: 10.1016/j.jbiomech.2018.01.029. [DOI] [PubMed] [Google Scholar]
- 3.Huurnink A., Fransz D.P., Kingma I., van Dieën J.H. Comparison of a laboratory grade force platform with a Nintendo Wii Balance Board on measurement of postural control in single-leg stance balance tasks. J. Biomech. 2013;46:1392–1395. doi: 10.1016/j.jbiomech.2013.02.018. [DOI] [PubMed] [Google Scholar]
- 4.Mengarelli A., Cardarelli S., Strazza A., Di Nardo F., Fioretti S., Verdini F. Validity of the Nintendo Wii balance board for the assessment of balance measures in the functional reach test. IEEE Trans. Neural Syst. Rehabil. Eng. 2018;26:1400–1406. doi: 10.1109/TNSRE.2018.2843884. [DOI] [PubMed] [Google Scholar]
- 5.Prieto T.E., Myklebust J.B., Hoffman R.G., Lovett E.G., Myklebust B.M. Measures of postural steadiness: differences between healthy young and elderly adults. IEEE Trans. Biomed. Eng. 1996;43:956–966. doi: 10.1109/10.532130. [DOI] [PubMed] [Google Scholar]
- 6.Baratto L., Morasso P.G., Re C., Spada G. A new look at posturographic analysis in the clinical context: sway-density versus other parameterization techniques. Mot. Control. 2002;6:246–270. doi: 10.1123/mcj.6.3.246. [DOI] [PubMed] [Google Scholar]
- 7.Severini G., Straudi S., Pavarelli C., Da Roit M., Martinuzzi C., Di Marco Pizzongolo L., Basaglia N. Use of Nintendo Wii balance board for posturographic analysis of multiple sclerosis patients with minimal balance impairment. J. NeuroEng. Rehabil. 2017;14 doi: 10.1186/s12984-017-0230-5. [DOI] [PMC free article] [PubMed] [Google Scholar]