Table 1.

Accuracy and precision of QC and select calibrator samples quantitated. With and without the voglibose internal standard.

Sample expected conc (ng/mL) Calculated with voglibose
Calculated without voglibose
Calc conc (ng/mL) Accuracy of mean (%) Precision (% CV) Calc conc (ng/mL) Accuracy of mean (%) Precision (% CV)
Within-day variation (n = 4)
7.48 7.96 106% 8.40% 7.84 105% 5.10%
75.9 72.3 95.0% 8.03% 74.3 98.0% 6.91%
Between-day variation (n = 20)
7.48 (QCL) 7.60 102% 6.33% 7.89 106% 11.50%
75.9 (QCH) 76.8 101% 8.04% 77.0 101% 4.27%
Calibration samples (n = 20)
2.80 2.76 98.5% 10.1% 2.91 104.0% 15.1%
52.3 49.6 94.8% 7.33% 50.2 96.0% 5.90%
249 248 99.5% 3.74% 248 99.4% 2.80%