Toroidally focused sensor array for real-time laser-ultrasonic imaging: The first experimental study

Anton Bychkov a,b, Varvara Simonova c,d,*, Vasily Zarubin a,b, Igor Kudinov b, Elena Cherepetskaya a, Alexander Karabutov a,c,d
PMCID: PMC6957820  PMID: 31956490

Abstract

In this paper we report on the first toroidally focused 2D real-time laser-ultrasonic imaging system and a modified filtered back projection algorithm that can be used in the region near the waist of the astigmatic laser-ultrasonic probe beam. The system is capable of visualizing an acupuncture needle 0.2 mm in diameter located at ∼4 cm depth in water. The lateral spatial resolution is better than ∼0.32 mm and axial spatial resolution is ∼30 μm. The achieved frame rate is up to 30 Hz. The depth dependency of the sensitivity region width and lateral resolution are experimentally measured and discussed. The array is intended to be used as a part of combined real-time photoacoustic and laser-ultrasonic imaging system.

Keywords: Laser ultrasound, Laser ultrasound imaging, Tomography, Real-time, Back projection, Toroidal array, Focused array

1. Introduction

Focused sensor arrays can be used in photoacoustic (PA) and laser-ultrasonic (LU) imaging to increase sensitivity and spatial resolution in a limited spatial region [1]. Focusing a linear array in the elevation direction allows forming the image plane for 2D cross-sectional (selective-plane) imaging [2]. Additional focusing of a linear array in the lateral direction can improve the lateral resolution, sacrificing the size of the sensitivity region [1,[3], [4], [5], [6]]. Generally, the resulting shape of the array will be toroidal. Moreover, the larger the width of the spatial region with high sensitivity the lower the spatial resolution in the corresponding direction [3]. Focused arrays are of particular interest in hand-held PA imaging probes because of better lateral resolution, reduced imaging artifacts and an ability to image sloped surfaces [7]. In addition, only the sensitivity region should be illuminated, so the light delivery system can be optimized.

Combined PA and traditional ultrasound pulse-echo imaging can provide both functional and morphological information about the tissue [8]. In PA imaging, the laser radiation is absorbed in the object itself and the generated acoustic pulses are used to reconstruct the distribution and strength of optical absorbers [9,10]. In traditional immersion pulse-echo ultrasound, a short acoustic probe pulse is used to insonify the object and backscattered acoustic field is used to reconstruct the external boundary and internal acoustic inhomogeneities (scatterers). LU pulse-echo imaging is similar to the traditional ultrasound. However, instead of using piezoelectric transducers to both generate and receive a short acoustic probe pulse, it uses a short (typically <10 ns) laser pulse to generate short (<100 ns) and wideband [11] (1–9 MHz) bipolar acoustic probe pulses by PA effect. In this case strongly damped piezoelectric transducers can be utilized to receive the backscattered acoustic field in a wide frequency band [4,12]. This combination allows broadband acoustic inspection of objects and effectively decouples generation and detection of acoustic signals to increase signal-to-noise ratio. Wideband acoustic generation and detection is the main reason why LU inspection has almost no “dead zone” and allows visualizing complex materials such as CFRP composites [13]. Furthermore, the transverse profile of the acoustic beam replicates the smooth transverse profile of the illuminating laser beam, thus practically eliminating the side-lobes. Therefore, combined PA and LU imaging can be advantageous [14,15].

The first multichannel combined PA and LU imaging system was cylindrically focused [[14], [15], [16]]. It had 16 PVDF piezofilm sensors, a plane-parallel optically absorbing plate (PA generator) and a cylindrical acoustic lens. The lens formed a 0.4 mm thick image plane, and the longitudinal resolution in water was 0.1 mm. This setup was used for real-time imaging of a medical needle inside a model of a blood vessel [15] as well as the external and internal boundaries of duralumin and polymethylmethacrylate (PMMA) samples [17]. However, the lateral resolution of this system was limited by the width of a single transducer (1 mm). At the same time, the noise equivalent pressure of the transducer is inversely proportional to the square root of its surface area [4]. Thus, the width of a transducer also directly affects the sensitivity. The use of a toroidally focused array could improve the lateral resolution of 2D LU imaging maintaining the high sensitivity and could expand the potential applicability of the LU imaging technique. While the use of toroidally focused arrays in PA imaging is more or less understood, to the best of the authors’ knowledge, no experimental toroidally focused LU imaging systems have been presented. Thus, it is important to build and explore such systems before combining them with PA imaging.

Toroidally focused arrays require not only a complicated manufacturing process, but also a suitable image reconstruction algorithm capable of real-time operation. In PA imaging, the filtered back projection (FBP) algorithm was shown to provide good results with a moderate number of sensors in real time [3,16,18]. In LU imaging, the FBP algorithm can be used for small animal imaging [19], for visualization of the external [20] and internal [21] boundaries of various objects and approximating them with sub-PSF accuracy [17]. The possibility of treating a focused transducer as an effectively point-like virtual detector situated at the focus of the transducer has been studied in PA microscopy [22,23]. However, the toroidally focused arrays are expected to have the highest lateral resolution in the vicinity of the focus in the image plane [24], where the diffraction of the probe beam should be taken into account. In the case of a spherically focused array, all the virtual detectors will coincide, which will make imaging around them impossible without moving the array.

The model-based algorithms [25,26] were shown to be successful in PA imaging with arbitrary-shaped detectors [27]. Although some implementations of model-based algorithms were capable of operating in real time [28], such algorithms are much more computationally intensive than closed-form FBP algorithms [29]. Simulation of ultrasound scattering requires additional calculations of the probe beam field. Therefore, implementing model-based algorithms in real time oriented portable PA and LU systems can be challenging. Moreover, analytical inversion schemes can be used in theoretical analysis of spatial resolution [30] and imaging artifacts.

In this paper, we present the first real-time LU imaging system utilizing a toroidally focused sensor array. A modified FBP algorithm for real-time LU imaging of the region near the waist of the acoustic probe pulse is introduced. The achieved frame rate is up to 30 Hz. The spatial resolution of 30 μm x 0.32 mm in a region ∼4 mm wide is demonstrated. The achieved spatial resolution of the toroidal array is significantly higher than the spatial resolution of the cylindrical array (0.1 mm × 1.1 mm) of the combined PA and LU imaging system presented earlier in [15]. However, the sensitivity region of the toroidal array is much smaller than that of the cylindrical array (∼32 mm). The depth dependency of the sensitivity region width and lateral resolution are explored experimentally.

2. Material and methods

2.1. Experimental setup

Fig. 1 shows the schematic of the experimental setup and the photograph of the developed LU sensor array. Laser radiation from a Q-switched laser (1053 nm, 200 Hz pulse repetition rate, 500 μJ pulse energy, ∼8 ns pulse duration, model TECH-1053, Laser-Compact, Russia) is delivered by the fiberoptic cable (600 μm core diameter, Optofiber, Russia) to the PA generator inside the sensor array, where it is absorbed. The PA generator is made of a black polymer material acoustically matched to the PMMA lens; the optical absorption coefficient is ∼180 cm−1 [14]. The short generated acoustic pulse travels through the toroidal PMMA acoustic lens and undergoes scattering by the object. The back-scattered acoustic waves propagate through the lens, and then they are registered by 16 wideband piezoelectric PVDF transducers (1 mm × 19 mm × 52 μm) with 2 mm pitch. Analog electrical signals from the detectors are pre-amplified and fed to the high-speed 32-channel data acquisition and processing system based on NI FlexRIO architecture (National Instruments, USA): NI 5752 ADC (12 bit, 50 MHz) and NI PXIe-7962R FPGA module. It digitizes, stores, averages the acquired data over several laser pulses, and then transmits the averaged digital signals via high speed communication lines (PCIe) to a personal computer (Intel® Core™ i7-4770 CPU @ 3.4 GHz, Intel, USA). The LU image was reconstructed in real time using the FBP algorithm modified for toroidally focused arrays. The algorithm was accelerated on a GPU (NVIDIA GeForce GTX 770, NVIDIA, USA) using NVIDIA CUDA computing platform.

Fig. 1.

Fig. 1

The experimental real-time LU imaging setup: the schematic of the setup (left), the photograph of the toroidal LU array (right).

2.2. Filtered back projection algorithm for laser-ultrasonic imaging with toroidally focused arrays

The LU FBP algorithm is a delay-and-sum type algorithm: the strength of a scatterer or the reflection coefficient of a boundary εr at point r is considered to be proportional to the weighted sum of the signal values prm,t recorded by the sensors located at points rm at the appropriate full travel times of the probe acoustic pulse:

εr=m=0NsDrm,rFprm,t=Tgr+rm-r/c (1)

Here Ns is a number of sensors, Drm,r=G-1rrm-rΔΩm/Ω0 is a coefficient that accounts for 1/rm-r attenuation of the scattered waves and can account for the amplitude Gr of the probe pulse at point r, ΔΩm is the solid angle of the m -th sensor as seen from the point r, Ω0 is the total solid angle of the array, Tgr is the time it takes for the probe pulse to reach the point r from the PA generator, c is the speed of sound in the immersion fluid, F is the filtering operator. In PA imaging, Fp is proportional to -tp/t. In LU imaging, Fp may be taken as -p/t due to the bipolar waveform of the LU probe pulse. Here, in the experiments, it was assumed that Gr=1. It should be noted that the sensitivity region directly depends on the form of the correction factor G-1r, but the spatial resolution as a local characteristic of the system does not depend on G.

Both Drm,r and Tgr depend on the geometry of the PA generator and the transverse profile of the illuminating laser beam. After calculation at the preparation stage, they can be stored and used for image reconstruction until the parameters of the array or the positions of the image pixels are changed. However, the sensitivity region of a focused array as well as the object can be small. In experiments, it is often necessary to manipulate the image scale and position in order to find the object and get better results. In this scenario, it is advantageous to minimize the duration of the preparation stage. One of the possible ways is to approximate the probe beam with an astigmatic monochromatic Gaussian beam propagating along the x direction:

UX,Y,ZU0=w0Zw0YwZXwYX1/2exp-Z2wZ2X-Y2wY2Xexp-ikcTgX,Y,Z (2)
cTgX,Y,Z=X+Z22RZX+Y22RYX-12katanX-X0ZXRZ+atanX-X0YXRY (3)

This beam has six parameters: wavenumber (k), two Rayleigh lengths (XRZ and XRY), two positions of the foci (X0Z and X0Y), and the amplitude (U0). The beam widths are wZX=w0Z1+X-X0Z2/XRZ2 and wYX=w0Y1+X-X0Y2/XRY2, where w0Z and w0Y are the beam waists, which are connected to the Rayleigh lengths XRZ=kw0Z2/2 and XRY=kw0Y2/2. The radii of curvature of the wavefront in the perpendicular planes are RZX=X-X0Z1+XRZ2/X-X0Z2 and RYX=X-X0Y1+XRY2/X-X0Y2. The angular apertures are 2θZ=4/kw0Z and 2θY=4/kw0Y correspondingly.

Consider a toroidally focused array, which has curvature radii rf and R+rf and angular apertures 2Θ and 2Φ. Let the center of curvature in the XY plane coincide with the origin (see Fig. 2). The foci of the corresponding Gaussian beam (X0Z and X0Y) can be chosen so that they coincide with the two centers of curvature of the sensor array: X0Z=0 and X0Y=-R. The Rayleigh lengths can be determined approximately from the angular apertures: XRZ=2/kΦ2 and XRY=2/kΘ2. The effective wavenumber k and Rayleigh lengths should be adjusted experimentally. If the right wavenumber is chosen, then the 2D LU image of a reflective plane will be a straight line and the Gouy phase shift will be compensated correctly, otherwise it will be curved and shifted.

Fig. 2.

Fig. 2

The example of a toroidally focused LU array and the acoustic probe pulse field emitted by the PA generator in the Gaussian beam approximation for rf=30 mm, R=20 mm, Φ=22.5°, Θ=20°. Solid black lines are the isolines of Tgr.

The experimental array had rf=40 mm, R=4 mm, Φ=20°, Θ=15°. The array shown in Fig. 2 has different parameters than the experimental one to illustrate the complex structure of the sensitivity region in the case of Gaussian astigmatic probe beam. The foci do not coincide and the overall sensitivity is less than for a confocal array. Note that the curvature of the wavefront changes significantly near the origin. During the processing of the experimental data the parameters were set as follows: k = 16.8 mm−1 (frequency 4 MHz), Φ=12.8°, Θ=7.1° to account for the limited irradiation of the PA generator.

2.3. Image processing

The spatial resolution of PA and LU imaging systems can be determined using the point-spread function (PSF) [1,3,19,30]. A PSF is an image of a point-like object and it is determined by the parameters of the system. If the distance between the two point-like objects is less than the full widths at half-maximum (FWHM) of the PSF, the objects cannot be distinguished. That is why the FWHM of the PSF can be treated as spatial resolutions in the corresponding direction in PA and LU imaging. The three directions are considered in the case of a 2D transducer array: axial, lateral and elevation. In the case of a toroidal sensor array in Fig. 2 and a point-like object on the X axis the axial direction corresponds to the X axis, the lateral direction – to the Z axis, and the elevation direction – to the Y axis. If the object is off the X axis its image may be blurred and rotated. That is why in [3] the widths of the PSF were determined by fitting the ellipse to the binarized PSF.

The image processing procedure consisted of three steps. First, the LU image was reconstructed using the FBP algorithm (1) – (3). Second, the image was binarized – the values greater than the half of the maximum value were replaced by ones and the others were replaced by zeros. Third, the ellipse that had the same second-moments as the image was found using the MATLAB/Octave regionprops function. In addition, the bounds along the Z axis were found for a binarized image of a plane-parallel plate to determine the width of the sensitivity region.

3. Results and discussion

Fig. 3 shows the schematic of the experiment with a model of a point-like acoustic scatterer (an acupuncture needle 0.2 mm in diameter) and the reconstructed LU image. The needle was fixed in a duralumin holder that was mounted on an automated 3D translation stage (see Fig. 1). The LU image (1 mm × 1 mm, 400 × 400 pixels) was reconstructed using the developed FBP algorithm (1) – (3) in real-time (up to 30 Hz frame rate) in XZ plane, which was perpendicular to the needle. The needle was translated along the X axis, and the experimental signals were saved for post-processing.

Fig. 3.

Fig. 3

The schematic of the experiment (left). The LU image of the acupuncture needle 0.2 mm in diameter (right). The solid black line in the LU image corresponds to the half-maximum level of the image. The colorbar is linear scale.

It should be noted that the better representation of a point-like object would be a needle fixed parallel to the X axis. However, such a setup would require a precise alignment of a needle and an array. A misalignment of 1° at a scan length of 20 mm would give ∼0.35 mm shift off the image plane which would be close to the waist of the beam in the elevation direction (w0Y0.45mm). This shift of the needle along Y axis should be compensated. The needle placed approximately parallel to the Y axis (as in [19]) allows to obtain the correct sensitivity depth profile without this compensation. The axial spatial resolution can be slightly worse due to the integration of the wavefront of the probe beam along the Y axis in this case.

Fig. 4 shows the dependencies of normalized sensitivity (black), axial (blue) and lateral (red) resolutions (FWHM of the image) on the position of the needle. Fig. 4 shows the reconstructed image and the binarized image of an acupuncture needle as well as the fitted ellipse. The best lateral resolution achieved in the experiment was 0.31–0.33 mm in the region between the foci (–5 mm < X < 0 mm), and the best axial resolution is ∼30 μm in the same region. The high sensitivity is achieved in the region around the foci: –9.6 mm < X < 1.8 mm.

Fig. 4.

Fig. 4

The dependencies of maximum amplitude (normalized sensitivity – black, top left), axial (ΔX, blue, middle left) and lateral (ΔZ, red, bottom left) FWHM of the image (spatial resolutions) on the position of the acupuncture needle 0.2 mm in diameter – the model of the point-like scatterer. The LU image of the needle reconstructed using the proposed algorithm (top right) and its binarized form with the fitted ellipse (dashed black line, bottom right). The solid black line in the LU image corresponds to the half-maximum level of the image. The lengths of the major and minor axes of this ellipse correspond to the widths of the image and spatial resolutions. The presented LU images correspond to the position of the needle X = 4.9 mm (near the focus).

The experimental determination of the size of the sensitivity region normally requires translating the needle along the Z axis. This procedure can be sped up if one uses reflective plane surface instead of a point scatterer. This plane surface can be considered as a superposition of infinite number of point-like scatterers. The size of the image of a reflective plane surface along the Z axis corresponds to the size of the sensitivity region. Thus, in the second experiment, we used a duralumin plate with a flat bottom surface mounted on the holder instead of the needle to model a reflective plane surface. Again, the plate was translated along the X axis, and the experimental signals were saved for post-processing.

The post-processing procedure was similar to that used for the needle images. The only difference was that we determined the left and right boundaries of the binarized image instead of calculating the second moments. The full width of the sensitivity region is shown in Fig. 5.

Fig. 5.

Fig. 5

The dependency of the lateral size W of the glass plane-parallel plate image on the axial position of the plate (width of the sensitivity region, magenta, left). The LU images of the plate for the three positions along the X axis: 5 mm (top right), 0 mm (middle right), –5 mm (bottom right). The solid black line in the LU images corresponds to the half-maximum level of the image.

Fig. 5 clearly shows that the minimum width of the sensitivity region (magenta) is achieved at X ≈ 0.7 mm. The minimum width of the sensitivity region is located near the center of curvature of the sensor array in the XZ plane (X0Z,) while the highest spatial resolution is achieved near the center of curvature in the XY plane (X0Y). The sensitivity region (where high sensitivity is achieved) is between the two centers of curvature (X0Z and X0Y) (Fig. 2). Therefore, toroidal focusing of the LU sensor arrays can be used to expand its sensitivity region.

It can be clearly seen in Fig. 5 that the LU image of the plate at X ≈ –5 mm resembles the negated version of the LU at X ≈ 5 mm. This is the consequence of the Gouy phase shift (two terms with atan in (3)). The total Gouy phase shift along the X axis is π which is equivalent to negating the probe pulse. This should be taken into account during the interpretation of the LU images because the polarity of the image of the boundary also depends on the reflection coefficient of the material. Since the glass plate in water is almost an acoustic hard boundary and the Guoy phase shift is taken into account in the calculation of Tgr it is the positive part of the image that represents the correct position of the reflective boundary.

The high axial resolution (30–40 μm) can be attributed to the wide ultrasound frequency band 2–14 MHz at –6 dB. As it was noted in [16] the axial spatial resolution in LU imaging is ∼0.4 λ ≈ 44 μm at 14 MHz. In [31] it is noted that the axial resolution for a spherical ultrasound radiator is 0.5c/BW63μm, where BW=12MHz – bandwidth of the transducer, the lateral resolution for the spherical radiator is λ¯f-number=λ¯/2sinΦ292μm, where λ¯0.2mm is the average wavelength. It should be noted that the lateral resolution is actually less than 0.32 mm since the needle itself is 0.2 mm in diameter and its LU image is roughly the spatial convolution of the PSF and the cross-section of the needle. Therefore, the toroidal LU sensor array allows achieving the resolution that is close to the diffraction limit of the system near the focus in the image plane. If the diffraction of the probe beam is not taken into account by the modified algorithm, the images of the reflective plate will be curved, and imaging near the foci of the array will be impossible.

It should be noted that the acoustic lens had a thickness of h=2mm in the image plane which corresponded to the round-trip time of Δt=2h/cl1.8μs, cl – the speed of sound in the material of the lens. Therefore, there was a series of false artifact images of the external boundary of the object at distances ncwΔt=n2.6mm, where n=1,2, – the number of reflection above the object. Since the model of the point scatterer was 0.2 mm in diameter these reflections were outside the imaging region.

4. Conclusions

In this paper we presented a system for 2D LU imaging with improved lateral spatial resolution in the image plane due to the toroidal focusing of the sensor array. The width (1 mm) and number (16) of the transducers as well as the aperture (∼45°) of the array were the same as we had before in the cylindrical sensor array (see [14,15]) to maintain high sensitivity and to preserve the single-sided access to the inspected object. A modified BP algorithm for real-time LU imaging of the region near the waist of the acoustic probe pulse is presented and tested experimentally. The developed toroidally focused LU imaging system allowed us to visualize a sample located at ∼4 cm depth in water with the lateral spatial resolution of 0.32 mm and axial spatial resolution of 30 μm in real time. The achieved frame rate was up to 30 Hz. The minimum width of the sensitivity region is achieved near the focus of the sensor array in the image plane while the best spatial resolution is achieved near the focus in the plane perpendicular to the image plane. The high sensitivity is achieved between the foci of the array.

The developed array is intended to be used as a part of combined real-time PA and LU imaging system as in [14,15]. The manufacture of the toroidally focused imaging system is much more difficult than the manufacture of the cylindrically focused system. Therefore, it is important to explore experimentally the trade-off between the size of the sensitivity region and the lateral spatial resolution to optimize the setup to a particular tomographic application. The modified FBP algorithm might be improved and the advanced signal processing for combined real-time LU and PA imaging might be implemented in future studies. Current work was focused on the experimental study of the toroidally focused LU array. Thorough numerical analysis of the developed array will be the topic of our future work.

Declaration of Competing Interest

None.

Acknowledgements

This work is supported by NUST MISIS Competitiveness Program by the Ministry of Science and Higher Education of the Russian Federation [K2-2019-004] and within the Russian Federation State assignment of FSRC “Crystallography and Photonics” RAS.

Biographies

graphic file with name fx1.jpg

Anton Bychkov has graduated Lomonosov Moscow State University, Russia in 2016. In 2013 he started his work on theoretical, numerical and experimental research in combined real-time laser optoacoustic and laser-induced ultrasound imaging of biological objects in Laser Optoacoustic Laboratory at Lomonosov MSU. At present, he is a PhD student at Lomonosov MSU. In 2015 he joined Laser Ultrasonic Nondestructive Testing Laboratory at National University of Science and Technology “MISiS”, Russia, where he is doing research on possibilities of application of optoacoustic and laser-ultrasound tomography in material structure study and non-destructive testing of industrial products, and working on experimental research automation.

graphic file with name fx2.jpg

Varvara Simonova studied physics at the M.V. Lomonosov Moscow State University, Russia. She received his Ph.D. degree in quantum electronics at the Institute of Laser and Information Technologies, Shatura, Russia. She now works as a Research Assistant at the Institute of Laser and Information Technologies, Shatura, Russia. Her main research interests include optoacoustic and laser ultrasound and connected with it imaging techniques, nondestructive evaluation of materials. She currently develops novel transducers and transducers arrays for laser ultrasound method for non-destructive testing and biological needs.

graphic file with name fx3.jpg

Vasily Zarubin has graduated Physical Faculty at Lomonosov MSU, Moscow in 2016. Now he is a Ph.D. student at Physical Faculty. His current research is focused on optoacoustic and laser ultrasound tomography, laser ultrasound defectoscopy, non-destructive testing and evaluation and corresponding algorithms development.

graphic file with name fx4.jpg

Igor Kudinov is leading engineer at Chair of General Physics and Wave Processes of Physics Department of M.V. Lomonosov Moscow State University. His current research is focused on electronics, development and manufacture of printed circuit boards, microcircuits, power supplies, parallel real-time data acquisition.

graphic file with name fx5.jpg

Elena Cherepetskaya is Professor at National University of Science and Technology “MISiS” (Moscow, Russia). Her most research has focused on optoacoustic spectroscopy of heterogeneous media, including rocks, concrete, carbon composites, on application of laser ultrasound in NDT and material evaluation.

graphic file with name fx6.jpg

Alexander Karabutov is Professor at International Laser Center of M.V.Lomonosov Moscow State University (ILC MSU), head of Laboratory of laser optoacoustics of ILC MSU and Leading Scientist of Laser Ultrasonic Nondestructive Testing Laboratory at National University of Science and Technology “MISiS”. Since 1980 has been developing theoretical foundations and experimental research in the field of nonlinear acoustics and development Fundamentals of Laser-ultrasonic spectroscopy. The principles of laser ultrasonic diagnostic structures and properties of various materials were developed. Recently, pays special attention to the measurement of the parameters of supercritical conditions of various metals and alloys using laser generation of ultrasound and broadband ultrasonic spectroscopy of composite materials.

References

  • 1.Rosenthal A., Ntziachristos V., Razansky D. Acoustic inversion in optoacoustic tomography: a review. Curr. Med. Imaging Rev. 2013;9:318–336. doi: 10.2174/15734056113096660006. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Khokhlova T.D., Pelivanov I.M., Karabutov A.A. Optoacoustic tomography utilizing focused transducers: the resolution study. App. Phys. Lett. 2008;92 [Google Scholar]
  • 3.Bychkov A.S., Cherepetskaya E.B., Karabutov A.A., Makarov V.A. Toroidal sensor arrays for real-time photoacoustic imaging. J. Biomed. Opt. 2017;22 doi: 10.1117/1.JBO.22.7.076003. [DOI] [PubMed] [Google Scholar]
  • 4.Oraevsky A., Karabutov A. Ultimate sensitivity of time-resolved optoacoustic detection. Proc. SPIE. 2000;3916:228–239. [Google Scholar]
  • 5.Dima A., Burton N.C., Ntziachristos V. Multispectral optoacoustic tomography at 64, 128, and 256 channels. J. Biomed. Opt. 2014;19 doi: 10.1117/1.JBO.19.3.036021. 036021. [DOI] [PubMed] [Google Scholar]
  • 6.Neuschmelting V., Burton N.C., Lockau H., Urich A., Harmsen S., Ntziachristos V., Kircher M.F. Performance of a Multispectral Optoacoustic Tomography (MSOT) system equipped with 2D vs. 3D handheld probes for potential clinical translation. Photoacoustics. 2016;4:1–10. doi: 10.1016/j.pacs.2015.12.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Schellenberg M.W., Hunt H.K. Hand-held optoacoustic imaging: a review. Photoacoustics. 2018;11:14–27. doi: 10.1016/j.pacs.2018.07.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Oraevsky A.A., Clingman B., Zalev J., Stavros A.T., Yang W.T., Parikh J.R. Clinical optoacoustic imaging combined with ultrasound for coregistered functional and anatomical mapping of breast tumors. Photoacoustics. 2018;12:30–45. doi: 10.1016/j.pacs.2018.08.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Beard P. Biomedical photoacoustic imaging. Interface Focus. 2011;1:602–631. doi: 10.1098/rsfs.2011.0028. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Xu M., Wang L.V. Photoacoustic imaging in biomedicine. Rev. Sci. Instrum. 2006;77 [Google Scholar]
  • 11.Oraevsky A.A. Optoacoustic tomography: from fundamentals to diagnostic imaging of breast cancer. In: Vo-Dinh T., editor. Biomedical Photonics Handbook PM222. 2014. pp. 715–757. [Google Scholar]
  • 12.Karabutov A.A., Podymova N.B., Letokhov V.S. Time-resolved laser optoacoustic tomography of inhomogeneous media. Appl. Phys. B. 1996;63:545–563. doi: 10.1364/AO.34.001484. [DOI] [PubMed] [Google Scholar]
  • 13.Podymova N.B., Karabutov A.A., Kobeleva L.I., Chernyshova T.A. Laser optoacoustic method for measuring local porosity of dispersion strengthened metal-matrix composite materials. Inorg. Mater. Appl. Res. 2018;4:481–486. [Google Scholar]
  • 14.Simonova V.A., Savateeva E.V., Karabutov A.A. Novel combined optoacoustic and laser-ultrasound transducer array system. Moscow Univ. Phys. Bull. 2009;64:394–396. [Google Scholar]
  • 15.Bychkov A.S., Zarubin V.P., Karabutov A.A., Simonova V.A., Cherepetskaya E.B. On the use of an optoacoustic and laser ultrasonic imaging system for assessing peripheral intravenous access. Photoacoustics. 2017;5:10–16. doi: 10.1016/j.pacs.2017.01.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Bychkov A., Simonova V., Zarubin V., Cherepetskaya E., Karabutov A. The progress in photoacoustic and laser ultrasonic tomographic imaging for biomedicine and industry: a review. Appl. Sci. 2018;8:1931. [Google Scholar]
  • 17.Zarubin V., Bychkov A., Zhigarkov V., Karabutov A., Cherepetskaya E. Model-based measurement of internal geometry of solid parts with sub-PSF accuracy using laser-ultrasonic imaging. NDT E Int. 2019;105:56–63. [Google Scholar]
  • 18.Xu M., Wang L.V. Universal back-projection algorithm for photoacoustic computed tomography. Phys. Rev. E. 2005;71 doi: 10.1103/PhysRevE.71.016706. [DOI] [PubMed] [Google Scholar]
  • 19.Wurzinger G., Nuster R., Schmitner N., Gratt S., Meyer D., Paltauf G. Simultaneous three-dimensional photoacoustic and laser-ultrasound tomography. Biomed. Opt. Express. 2013;4:1380–1389. doi: 10.1364/BOE.4.001380. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Zarubin V., Bychkov A., Karabutov A., Simonova V., Cherepetskaya E. Laser-induced ultrasonic imaging for measurements of solid surfaces in optically opaque liquids [Invited] Appl. Opt. 2018;57:C70–C76. doi: 10.1364/AO.57.000C70. [DOI] [PubMed] [Google Scholar]
  • 21.Zarubin V., Bychkov A., Simonova V., Zhigarkov V., Karabutov A., Cherepetskaya E. A refraction-corrected tomographic algorithm for immersion laser-ultrasonic imaging of solids with piecewise linear surface profile. Appl. Phys. Lett. 2018;112 [Google Scholar]
  • 22.Li M.-L., Zhang H.F., Maslov K., Stoica G., Wang L.V. Improved in vivo photoacoustic microscopy based on a virtual-detector concept. Opt. Lett. 2006;31:474–476. doi: 10.1364/ol.31.000474. [DOI] [PubMed] [Google Scholar]
  • 23.Perekatova V.V., Kirillin M.Yu, Turchin I.V., Subochev P.V. Combination of virtual point detector concept and fluence compensation in acoustic resolution photoacoustic microscopy. J. Biomed. Opt. 2018;23 doi: 10.1117/1.JBO.23.9.091414. [DOI] [PubMed] [Google Scholar]
  • 24.Bychkov A.S., Cherepetskaya E.B., Karabutov A.A., Makarov V.A. Improvement of image spatial resolution in optoacoustic tomography with the use of a confocal array. Acoust. Phys. 2018;64:77–82. [Google Scholar]
  • 25.Dean-Ben X.L., Buehler A., Ntziachristos V., Razansky D. Accurate model-based reconstruction algorithm for three-dimensional optoacoustic tomography. IEEE Trans. Med. Imaging. 2012;31:1922–1928. doi: 10.1109/TMI.2012.2208471. [DOI] [PubMed] [Google Scholar]
  • 26.Rosenthal A., Razansky D., Ntziachristos V. Fast semi-analytical model-based acoustic inversion for quantitative optoacoustic tomography. IEEE Trans. Med. Imaging. 2010;29:1275–1285. doi: 10.1109/TMI.2010.2044584. [DOI] [PubMed] [Google Scholar]
  • 27.Rosenthal A., Ntziachristos V., Razansky D. Model-based optoacoustic inversion with arbitrary-shape detectors. Med. Phys. 2011;38:4285–4295. doi: 10.1118/1.3589141. [DOI] [PubMed] [Google Scholar]
  • 28.Ding L., Deán-Ben X.L., Razansky D. Real-time model-based inversion in cross-sectional optoacoustic tomography. IEEE Trans. Med. Imaging. 2016;35:1883–1891. doi: 10.1109/TMI.2016.2536779. [DOI] [PubMed] [Google Scholar]
  • 29.Lutzweiler C., Razansky D. Optoacoustic imaging and tomography: reconstruction approaches and outstanding challenges in image performance and quantification. Sensors. 2013;13:7345–7384. doi: 10.3390/s130607345. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Xu M., Wang L.V. Analytic explanation of spatial resolution related to bandwidth and detector aperture size in thermoacoustic or photoacoustic reconstruction. Phys. Rev. E. 2003;67 doi: 10.1103/PhysRevE.67.056605. [DOI] [PubMed] [Google Scholar]
  • 31.Foster F.S., Pavlin C.J., Harasiewicz K.A., Christopher D.A., Turnbull D.H. Advances in ultrasound biomicroscopy. Ultrasound Med. Biol. 2000;26:1–27. doi: 10.1016/s0301-5629(99)00096-4. [DOI] [PubMed] [Google Scholar]