Comparison of the performance of a chemiluminescence assay and an ELISA for detection of anti-GBM antibodies

Ying Tan a,b,c,d, Wei Pang a,b,c,d, Xiaoyu Jia a,b,c,d, Ming-hui Zhao a,b,c,d,e,
PMCID: PMC6968565  PMID: 31885301

Abstract

Objective

Autoantibodies to the α3 chain noncollagen 1 domain of type IV collagen (α3(IV)NC1) are a serological hallmark in the diagnosis of anti-glomerular basement membrane (GBM) disease. The objective of our study was to compare the performance of anti-glomerular basement membrane (GBM) antibody detection by chemiluminescence immunoassay (CIA) and by enzyme-linked immunosorbent assays (ELISAs).

Methods

Sera from outpatients who were suspected to have anti-GBM disease and 31 patients with biopsy-proven anti-GBM disease were collected. Thirty normal controls were also included. All samples were tested for anti-GBM antibodies by CIA and commercial ELISA. The anti-GBM antibody-positive samples were confirmed by a homemade ELISA coated with recombinant human α3(IV)NC1.

Results

Compared with detection of anti-GBM antibodies with ELISA, detection of anti-GBM antibodies with CIA showed a positivity agreement of 70% and a negativity agreement of 98.6%. Among the 4 patients with different results, the anti-GBM antibody detection by CIA was in agreement with the homemade ELISA coated with recombinant human α3(IV)NC1 and the clinical diagnosis. In 31 patients with anti-GBM disease, good agreement was achieved in the detection of anti-GBM antibodies with CIA, commercial ELISA and the homemade ELISA (100%, 100%). The AUC for CIA and commercial ELISA was 0.987 and 0.966, respectively.

Conclusions

The detection of anti-GBM antibodies with CIA demonstrated good sensitivity and specificity and was in good agreement with our homemade ELISA, which seems better than the commercial ELISA in suspected anti-GBM disease patients. The three assays performed in parallel in the diagnosis of anti-GBM disease patients.

Keywords: Autoantibodies, GBM, Goodpasture’s syndrome, Rapidly progressive crescentic glomerulonephritis (RPGN)

Introduction

Anti–glomerular basement membrane (anti-GBM) disease is a rare autoimmune disorder that is characterized by the production of autoantibodies directed to the GBM, rapidly progressive glomerulonephritis, and a high risk for alveolar hemorrhage [1]. Anti-GBM disease is a rare disease with a yearly incidence of 0.5–1 cases per million inhabitants, which can result in a rapid deterioration in renal function. The pathogenic role of anti-GBM antibodies has been demonstrated by their ability to transfer the disease to monkeys and by the recurrence of disease in human kidney allografts [2]. Early and accurate identification of anti-GBM antibodies are essential for renal prognosis since patients showed poor outcomes when the initial serum creatinine level was over 6.8mg/dL [3].

The target autoantigen of anti-GBM antibodies has been identified as the noncollagen domain 1 of the α3 chain of type IV collagen [α3(IV)NC1], with two major cryptic epitopes, EA and EB [4]. The confirmative feature of anti-GBM disease is the exhibition of linear deposits of immunoglobulin G (IgG) along the GBM on renal biopsy and detectable circulating antibodies against GBM by means of commercial enzyme-linked immunosorbent assay (ELISA) kits. However, there have been some cases reported with anti-GBM antibody deposition along the GBM but without positive findings for the autoantibodies by ELISA [5]. For these patients, diagnosis was difficult to confirm, and the indicator for plasmapheresis was deficient. Sometimes, the low sensitivity of anti-GBM detection may mislead the cessation of plasmapheresis even though anti-GBM antibodies are still not fully removed. Additionally, some patients are positive for anti-GBM antibodies by ELISA, but their autoantibodies do not recognize recombinant human α3 or demonstrate linear deposition in the GBM, which indicates false positive results that might partly be caused by the difference in coated bovine α3(IV)NC1 from the human homolog (unpublished data). Thus, more sensitive and specific methods need to be discovered.

A previous study demonstrated that the performance characteristics of an anti-GBM chemiluminescence immunoassay (CIA) had good sensitivity and specificity and were in good agreement with other methods [6]. Moreover, the CIA method takes 30min for one sample and 1min for the next sample.

The objective of this study was to compare the performance of an anti-GBM antibody CIA with an ELISA for clinical use.

Methods

Patients

Sera from patients who were suspected to have anti-GBM disease were collected at the Institute of Nephrology, Peking University from 5 June 2017 to 16 June 2017. The exclusion criteria were when the samples showed signs of hemolysis, lipemia, or bilirubinemia. Sera from 31 patients with biopsy-proven anti-GBM disease were collected from the Institute of Nephrology, Peking University from 2008–2015 and were preserved at –20 °C until use. Informed consent was obtained from each patient.

Preparation of recombinant human α3(IV)NC1

Recombinant proteins were produced as described previously [7]. Briefly, cDNA from the NC1 domain of human type IV collagen α3 (Supplementary table 1) was ligated to a type X collagen triple-helix leader sequence and subcloned into the pcDNA3 vector. The constructs were then stably transfected into a human embryonic kidney (HEK 293) cell line, and recombinant proteins were harvested and purified from the medium and designated rα3.

Anti-GBM assays

Polystyrene microtiter plates (Nunc Immunoplate, Roskilde, Denmark) were coated with 100μL of α3(IV)NC1 in coating buffer (50mM sodium carbonate [pH 9.6]) overnight at room temperature. rα3 was coated at 0.5μg/mL. The plates were then washed three times. A total of 100μL of human sera diluted 1:100 in PBS was added to each well. The plates were incubated at room temperature for 1h; after washing, alkaline phosphatase–conjugated goat anti-human IgG (Fc specific; Sigma, St. Louis, MO, USA) diluted 1:10 000 was added. Incubation resumed for 1h. P-nitrophenyl phosphate (1mg/mL; Sigma) in substrate buffer (1M diethanolamine and 0.5mM MgCl2 [pH 9.8]) was used as a substrate, and color development was measured spectrophotometrically at 405nm. All assays were run in duplicate, and when standard errors >10% were found, samples were reanalyzed. Plasma from 30 healthy blood donors was used to build a cutoff value as the mean +2 SD. The binding of the known positive control serum was set as 100%, and the blank control with PBS was set as 0%. The bindings of the tested sera were expressed as a percentage of that of the known positive sample.

Anti-GBM antibodies were also detected by a commercial CIA kit (INOVA, San Diego, USA) on a BIO-FLASH instrument and by a commercial ELISA kit (EUROIMMUN, Lübeck, Germany) on a EUROIMMUN Analyzer.

Precision studies

Precision of the CIA was verified by performing the required testing according to the Clinical and Laboratory Standards Institute (CLSI) guidelines. For the precision study, the within-run, between-day, between-run, and total precision were determined by running two aliquots of the precision samples twice a day in a random order, with a minimum of 2h between each run. The samples were run on the same instrument for each assay, and runs were repeated for at least 5days, according to CLSI guideline EP5-A2.

Statistical analyses

SPSS 16.0 statistical software was employed for statistical analysis. Spearman correlation analysis was carried out to analyze the correlation between groups, and p values <.05 were considered significant. Quantitative data were expressed as the mean ± SD or median with range (minimum, maximum). Receiver operating characteristic (ROC) analysis was carried out to analyze the discrimination between different methods and the homemade ELISA.

Results

General data of patients

Sera from 154 patients with suspected anti-GBM disease were collected. No samples showed signs of hemolysis, lipemia, or bilirubinemia.

Comparison of CIA and ELISA in the detection of anti-GBM antibodies in suspected anti-GBM disease

Compared with the detection of anti-GBM antibody with ELISA, the detection of anti-GBM antibody with CIA showed agreement of positivity of 63.6% and of negativity of 97.3% among the 154 outpatients (Figure 1).

Figure 1.

Figure 1.

Scatter plot of levels of anti-GBM antibodies with different assays.

Among the 4 patients with different results, the anti-GBM antibody detection by CIA was in agreement with the homemade ELISA coated with recombinant human α3(IV)NC1 and with the clinical diagnosis at a 2-year follow up (Table 1).

Table 1.

The clinical features of patients with different results of anti-GBM antibody by ELISA and CIA.

No. of patients Sex Age ELISA (RU/mL) CIA (cu) Home-made ELISA (%) Renal biopsy Clinical diagnosis
P1 M 50 Negative 42.1 47 Crescentic nephritis Anti-GBM disease
P2 F 48 80.81 Negative Negative Minimal change disease Nephrotic syndrome
P3 F 33 >200 Negative Negative Focal glomerulosclerosis Chronic kidney disease
P4 F 65 26.02 Negative Negative Not performed Latent nephritis

Comparison of CIA and ELISA in the detection of anti-GBM antibodies in biopsy-proven anti-GBM disease

The general data of all 31 biopsy-proven anti-GBM disease patients are listed in Table 2. All the sera from the 31 patients with anti-GBM disease were tested by CIA, commercial ELISA and homemade ELISA (Table 2). The agreement of positivity and of negativity was both 100% for the commercial ELISA and the CIA.

Table 2.

The clinical features of patients with biopsy-proven anti-GBM disease.

No. Sex Age (years) Pathological diagnosis ELISA (RU/mL) CIA (cu) Home-made ELISA (%)
1 M 75 Membranous nephropathy (Stage II) combined with Type I Crescentic nephritis 86 120.4 60
2 F 46 Membranous nephropathy (Stage II) combined with Type I Crescentic nephritis 33 600.1 92
3 M 48 Type I Crescentic nephritis 119 371.9 47
4 M 58 Type I Crescentic nephritis 338 >1437.8 83
5 F 20 Type I Crescentic nephritis 109 1021.2 101
6 F 58 Type I Crescentic nephritis 130 101.2 87
7 F 29 Type I Crescentic nephritis 40 67.3 68
8 M 50 Type I Crescentic nephritis 193 821.2 85
9 F 28 Type I Crescentic nephritis >200 >1437.8 107
10 M 68 Type I Crescentic nephritis 74 125.2 45
11 F 41 Type I Crescentic nephritis combined with TMA >200 593.6 114
12 F 49 Type I Crescentic nephritis combined with IgA nephropathy 59 113.2 58
13 F 73 Type I Crescentic nephritis 68 45.9 21
14 F 71 Type I Crescentic nephritis >200 934.3 53
15 F 52 Type I Crescentic nephritis >200 >1437.8 121
16 F 62 Type I Crescentic nephritis 95 >1437.8 113
17 M 60 Type I membranous nephropathy combined with Type I Crescentic nephritis >200 >1437.8 115
18 M 20 Type I Crescentic nephritis >200 820.4 131
19 M 58 Type I Crescentic nephritis 92 >1437.8 52
20 F 55 Type I Crescentic nephritis >200 784.7 108
21 M 67 Type I + Type III Crescentic nephritis 60 529.1 75
22 F 49 Type I Crescentic nephritis combined with idiopathic memebraneous nephropathy 45 49.4 101.9
23 F 22 Type I Crescentic nephritis 155 >1437.8 50
24 M 48 Type I Crescentic nephritis 51 92 15
25 M 38 Type I Crescentic nephritis 165 >1437.8 87
26 M 17 Type I Crescentic nephritis >200 >1437.8 90
27 F 35 Type I Crescentic nephritis 125 828 94
28 M 26 Type I Crescentic nephritis 79 191.7 87
29 M 27 Type I Crescentic nephritis 140 >1437.8 98
30 F 21 Type I Crescentic nephritis 172 >1437.8 94
31 M 34 Type I Crescentic nephritis Neg Neg Neg

The levels of anti-GBM antibodies according to CIA were significantly correlated with those according to commercial ELISA (r=0.458, p =.001) and homemade ELISA (r=0.297, p=.028).

The prediction of levels of anti-GBM antibodies detected by CIA or commercial ELISA

The sensitivity and specificity of CIA in the detection of anti-GBM antibodies were 97.4% and 100%, respectively, compared with 94.9% and 97.9%, respectively, by commercial ELISA. The AUC for CIA and commercial ELISA was 0.987 and 0.966, respectively (Figure 2).

Figure 2.

Figure 2.

ROC curves for predicting anti-GBM disease based on commercial ELISA and CIA.

Precision of the CIA

For the precision testing of the CIA, 2 samples, from 40.2 to 1304IU/mL, were tested. For the 2 samples, the max within-run variation was 4.2% and 5.6%, the between-day variation was 6.3% and 4.4%, and the between-run variation was 7.4% and 4.1%, respectively.

Discussion

Anti-GBM antibodies are a hallmark in the diagnosis of anti-GBM disease.

In recent decades, several novel technologies have been developed for anti-GBM antibody detection, including conventional ELISA and indirect immunofluorescence assay and, more recently, CIA. As ELISAs are only moderately fast, with assay times of approximately 2h, the focus has shifted toward a decrease in assay time and ease of use, as well as toward increased sensitivity since several patients were misdiagnosed until a renal biopsy was performed. CIA takes 30 min for the first sample and then 1min for the next. Thus, we evaluated the faster CIA method for the detection of anti-GBM antibodies.

The CIA method displayed good agreement with our homemade ELISA coated with recombinant human α3(IV)NC1 expressed in a HEK 293 cell line. Many studies have focused on identifying the epitopes of anti-GBM antibodies. It has been clearly shown that anti-GBM antibodies react with conformational epitopes of α3(IV)NC1, which limits the application of linear synthetic peptides for diagnosis strategies. Accordingly, chimeric molecules of human α3(IV)NC1 and α1(IV)NC1 expressed in a mammalian cell line have been used for epitope mapping [8,9]. However, the human α3(IV)NC1 expressed in the mammalian cell line achieved only low production and was highly expensive for clinical use. Assays using the recombinant antigen expressed in insects were produced, with different performances [6]. The good agreement achieved for CIA and our homemade ELISA may be partly because of the exposure of the target antigen for binding on the beads rather than in a well in the ELISA plate, which might provide more conformational epitope mimicry in vitro and expose more cryptic epitopes through the same coating antigen, bovine NC1 alpha 3(IV), than the ELISA plate, which showed better agreement with the clinical diagnosis.

Given the aggressive nature of anti-GBM disease, there is a compelling need for a rapid and sensitive test for the detection and monitoring of anti-GBM antibodies. Plasma exchange is one of the most important therapies for anti-GBM disease. In our study, there were also 2 biopsy-proven anti-GBM disease patients with low levels of anti-GBM antibodies, revealing discrepant results in ELISA assays (Supplementary table 2). Since the indication of cessation is the levels of anti-GBM antibody in plasma exchange, the negative result directed the administration of inadequate regimens.

Only a few studies have evaluated and compared the diagnostic performance of anti-GBM antibody immunoassays. However, we found consistently negative results obtained with all assays for 1 anti-GBM disease patient, which is consistent with the reported prevalence of anti-GBM-negative anti-GBM disease patients [10]. Approximately 2–8% of patients with anti-GBM disease have been reported to be anti-GBM antibody negative by enzyme immunoassays or Western blot [10]. The antibodies of sometimes patients may recognize highly conformational epitopes, which could be found by nonreducing Western blotting, and some may bind to α chains other than α3[5].

This study also had some shortcomings. First, the positivity is unusually low since anti-GBM disease is rare and because this report is a prospective study. Thus, we included another group of patients with biopsy-proven anti-GBM disease to further validate our findings. Second, this report is a single center study, although we have the largest cohort of anti-GBM disease patients in the world. Thus, multicenter studies may be needed.

Conclusion

The detection of anti-GBM antibodies with CIA demonstrated good sensitivity and specificity and was in good agreement with our homemade ELISA coated with recombinant human α3(IV)NC1, which seems to provide better performance than the commercial ELISA assay in suspected anti-GBM disease patients. The three assays performed in parallel for the diagnosis of anti-GBM disease patients.

Supplementary Material

Supplemental Material
Supplemental Material

Funding Statement

This work was supported by grants National Natural Science Foundation of China [No.81500526, No. 81670640 and No, 81670639], Beijing Natural Science Foundation [No. 7172215] and Clinical Medicine Plus X-Young Scholars Project of Peking University supported by the Fundamental Research Funds for the Central Universities.

Disclosure statement

No potential conflict of interest was reported by the authors.

References

  • 1.Hudson BG, Tryggvason K, Sundaramoorthy M, et al. . Alport’s syndrome, Goodpasture’s syndrome, and type IV collagen. N Engl J Med. 2003;348(25):2543–2556. [DOI] [PubMed] [Google Scholar]
  • 2.Almkuist RD, Buckalew VM, Jr., Hirszel P, et al. . Recurrence of anti-glomerular basement membrane antibody mediated glomerulonephritis in an isograft. Clin Immunol Immunopathol. 1981;18(1):54–60. [DOI] [PubMed] [Google Scholar]
  • 3.Cui Z, Zhao J, Jia XY, et al. . Anti-glomerular basement membrane disease: outcomes of different therapeutic regimens in a large single-center Chinese cohort study. Medicine. 2011;90(5):303–311. [DOI] [PubMed] [Google Scholar]
  • 4.Yang R, Hellmark T, Zhao J, et al. . Levels of epitope-specific autoantibodies correlate with renal damage in anti-GBM disease. Nephrol Dial Transplant. 2009;24(6):1838–1844. [DOI] [PubMed] [Google Scholar]
  • 5.Jia XY, Qu Z, Cui Z, et al. . Circulating anti-glomerular basement membrane autoantibodies against alpha3(IV)NC1 undetectable by commercially available enzyme-linked immunosorbent assays. Nephrology. 2012;17(2):160–166. [DOI] [PubMed] [Google Scholar]
  • 6.Mahler M, Radice A, Sinico RA, et al. . Performance evaluation of a novel chemiluminescence assay for detection of anti-GBM antibodies: an international multicenter study. Nephrol Dial Transplant. 2012;27(1):243–252. [DOI] [PubMed] [Google Scholar]
  • 7.Yang R, Hellmark T, Zhao J, et al. . Antigen and epitope specificity of anti-glomerular basement membrane antibodies in patients with goodpasture disease with or without anti-neutrophil cytoplasmic antibodies. J Am Soc Nephrol. 2007;18(4):1338–1343. [DOI] [PubMed] [Google Scholar]
  • 8.Netzer KO, Leinonen A, Boutaud A, et al. . The goodpasture autoantigen. Mapping the major conformational epitope(s) of alpha3(IV) collagen to residues 17–31 and 127–141 of the NC1 domain. J Biol Chem. 1999;274(16):11267–11274. [DOI] [PubMed] [Google Scholar]
  • 9.Borza DB, Netzer KO, Leinonen A, et al. . The goodpasture autoantigen. Identification of multiple cryptic epitopes on the NC1 domain of the alpha3(IV) collagen chain. J Biol Chem. 2000;275(8):6030–6037. [DOI] [PubMed] [Google Scholar]
  • 10.Hellmann MA, Gerhardt TM, Rabe C, et al. . Goodpasture’s syndrome with massive pulmonary haemorrhage in the absence of circulating anti-GBM antibodies? Nephrol Dial Transplant. 2006;21(2):526–529. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplemental Material
Supplemental Material

Articles from Renal Failure are provided here courtesy of Taylor & Francis

RESOURCES