Abstract
Despite advances in metallurgy, fatigue failure of hardware in orthopaedics is common especially when a fracture fails to heal. Revision procedures can be difficult, usually requiring removal of intact or broken hardware. Several different methods are usually used to successfully remove the intact or broken hardware. The proximal blade fracture is rare and under-reported in the literature. In our tertiary non-union and limb reconstruction unit, we have developed a technique for an easy and quick removal of a broken proximal femoral blades.
Keywords: Implant failure, Broken implant, Implant removal retrieval of blade blade removal
1. Background
Despite improvements in metallurgy, implant fatigue is prevalent especially in cases of non-union. Revision is always challenging, and regularly demand the extraction of intact or broken implants. Various described approaches are required to remove broken implants. Hardware removal can profess a challengeable surgical problem in revision trauma surgery. The challenge of the retrieval of broken hardware can considerably prolong operative time and complexity.
The use of intramedullary nails is the gold standard for treatment of the proximal and diaphyseal femoral fractures. Nevertheless, the magnitude of bending forces in this region of the femur can often yield to implant failure before union. Broken proximal femoral blades after femoral interlocking nails are under-reported and a surgically challenging situation. There are different published techniques to remove broken locking bolts1,2 which is not the case for removal of broken blade; which is technically challenging. As yet, to the best of our knowledge, only one extraction method has been described to retrieve broken proximal blades,3 which is not successful in our experience. In other situations, a broken implant may be encountered surprisingly throughout regular hardware retrieval, this usually happens in the case of nonunion or delayed union of a fracture following a fracture. Moreover, Broken blades might happen after a tumor resection or reconstruction in the case of absence of healing. Also, intramedullary femoral nails are used for prophylactic fixation of impending pathologic fracture; the proximal blade might break because of improved patient survival.
Despite advances in metallurgy, fatigue failure of hardware is common when a fracture fails to heal. Revision procedures can be difficult, usually requiring removal of intact or broken hardware. Several different methods may need to be attempted to successfully remove intact or broken hardware. The proximal blade fracture is rare and under-reported in the literature. We describe an alternate technique; that we have utilised in our tertiary fracture non-union and limb reconstruction unit. We have developed a technique for an easy and quick removal of a broken proximal femoral blade. The following tips simplify the process and reduce the operative time required.
2. Methods
2.1. Case report
2.1.1. Surgical technique
Broken proximal blades often are identified on preoperative radiographs (Fig. 1), although it can also occurs during the primary surgery. The patient is positioned on a radiolucent table and standard draping was used.
Fig. 1.

Broken femoral blade held in the medial femoral cortex.
The original incision is utilised for the proximal blade hole; the incision is extended proximally and distally by two cm. Surgical dissection is extended down to the bone after splitting the vastus lateralis muscle; using the image intensifier the proximal blade is identified. Maintaining a stringently subperiosteal plane, the vastus intermedius and medialis origin are lifted subperiosteally and two curved Hohmann retractors are introduced to aid in retraction.
Under the guidance of the image intensifier, the lateral accessible fragment of the proximal blade is retrieved followed by the removal of the nail. An initial attempt to pull the remaining broken medial part of the proximal blade was undertaken without success especially that the fracture was not united.
A 4mm tap is used (Fig. 2) to engage in the central tunnel of the spiral blade. The tap is then rotated clockwise into the tunnel in the blade screw (Fig. 3) and it was advanced in the centre of the tunnel. Subsequently, the tap was pulled out with the broken medial part of the blade (Fig. 4a). As the fracture was not united, a proximal femoral blade plate was used to fix the fracture (Fig. 4a).
Fig. 2.
Size 4 mm standard tap is used.
Fig. 3.
The tap is rotated clockwise into the tunnel in the proximal blade and subsequently pulling on the drill bit retrieves the broken part.
Fig. 4.
A) Successful removal of the broken blade. B) revision fixation of the non-united fracture using proximal femoral blade plate.
3. Discussion
The breakage of the proximal blade in femoral nails is techniqually challenging.3 As yet, to the best of our knowledge, our technique has not been described elsewhere. We report a cost-effective, reproducible, bespoke and safe technique that is effective in our experience. Usually the proimal blade failure happens in cases of non union, thus, pulling on the broken medial part is unsuccessful. It is mandatory to remove the medial broken part so as to undertake revision fixation of the non-united fracture.
Compared with other implants, cephalomedullary nails provide biomechanical superiority due to their shorter lever arms and reduced deforming forces.4,5 Intramedullary nails are the preferred surgical option for the management of proximal femoral fractures, especially in the case of closed reduction procedure.4, 5, 6 Additionally, intramedullary nailing is correlated with reduced soft tissue injury, reduced blood loss, and decreased infection rates and wound complications.4, 5, 6 Ma et al.6 published decreased blood loss and decreased the length of hospitalisation. Besides, they reported no significant difference in the rate of fixation failure between the intramedullary nails to dynamic hip screw. Thus, femoral intramedullary nailing is regularly utilised to treat unstable geriatric trochanteric hip fractures.7,8
On the other hand, several complications might develop in the intraoperative and postoperative periods with the use of intramedullary nails. The reported complications include cut out/back out of the blade and non-union of the fracture.8, 9, 10, 11, 12 There are reported complications in the literature; these include cut out/back out of the blade and non-union of the fracture. The complications for utilising cephalomedullary nails’ fixation for proximal femoral fracture also include shortening, malrotation, malunion, non-union and implant failure or malposition. Proximal blade failures are under-reported in the literature.9,12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 22, 23 Hypothetically, this risk can be avoided by satisfactory reduction before nail insertion, precise assembly of the implant and frequent check of components using fluoroscopic views.
The complication rate of proximal femoral nails that requires revision ranges from 3% to 28% in the literature.3, 4, 5,8,9,16,19, 20, 21, 22,24, 25, 26, 27, 28, 29, 30
Paraschou et al.21 reported outcomes of 257 trochanteric fractures managed with intramedullary nailing. They reported two malunions, one nonunion, one screw cut out and one screw migrated medially. Also, they reported that 2 out of 275 distal locking screws were misplaced using a commercially prepared jig. Fogagnolo et al.5 found mechanical failures in 23.4% patients; Akan et al.31 reported 10% mechanical failure in a cohort of 80 patients, while Boldin et al.32 reported a complication rate of mechanical failure 21.8% in 55 patients treated with intramedullary nails compared to 4.6% only by Simmermacher et al..8 Although the rate of implant failure is variable in different studies4, 5, 6,8,20,26,31,32 but this might be because of the variability in the clarity of the definition of failure. It is reported to be around 5%, 2%–10% with Sliding Hip screws, 2%–12% with lag screw Cephalomedullary nails, and 1%–8% with blade Cephalomedullary nails.7, 8, 9,33,34 In the meantime, the implant-related complication rate of basicervical fracture in previous studies showed remarkable variations.24,28
The removal of broken blade can be technically challenging, yet, a crucial step in revision surgery, especially that they occur in cases of non union where refixation is planned. Preoperative planning is mandatory to make sure that the required removal instruments are available. Trauma surgeons should be knowledgeable of accessible instruments and possible techniques to remove broken blade in theatres as assuming the unexpected would prevent failure. There are multiple reports on the use of conversion to total hip arthroplasty for failed proximal fixations of intertrochanteric fractures, which would be another successful option.26,30,35,36 The removal of broken implants is a challenging task.2,3 There is a variation in the metal breakage rate, in orthopaedic surgery, between elective and trauma work – ranging from 0.3 to 7.9 per 1000 cases respectively.37 There are various techniques in the literature described for removal of broken locking bolts,2,37 broken drill bits, broken interlocking screws,38 broken cannulated screws39 and broken intramedullary nails.1,37
As yet, to the best of our knowledge, only one extraction method has been described to retrieve broken proximal blades,3 which may fail. Stover et al.3 reported a surgical technique that entails connecting a T handle to the reverse threaded conical extraction bolt, from the AO broken screw extraction bolt, after locking it to the spiral blade. This has failed in our experience as there is no room for striking the blade like the described case in this study especially when displaced medially.
In conclusion, there is not a single removal technique that is extensively successful, the surgeon should be aware of various different techniques to remove broken blades. The described technique is successful in our hands and reproducible in our experience.
Declaration of competing interest
All authors report no conflicts of interests.
References
- 1.Hak D.J., McElvany M. Removal of broken hardware. J Am Acad Orthop Surg. 2008;16:113–120. doi: 10.5435/00124635-200802000-00009. [DOI] [PubMed] [Google Scholar]
- 2.Imam M.A., Harb Z., Elliott D. An innovative bespoke technique for the retrieval of broken distal locking bolts. Ann R Coll Surg Engl. 2018:1–2. doi: 10.1308/rcsann.2018.0080. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Stover M.D., Lin I., Bosse M.J. Removal of a broken Synthes proximal spiral blade. J Orthop Trauma. 1998;12:190–191. doi: 10.1097/00005131-199803000-00010. [DOI] [PubMed] [Google Scholar]
- 4.Banan H., Al-Sabti A., Jimulia T., Hart A.J. The treatment of unstable, extracapsular hip fractures with the AO/ASIF proximal femoral nail (PFN)--our first 60 cases. Injury. 2002;33:401–405. doi: 10.1016/s0020-1383(02)00054-2. [DOI] [PubMed] [Google Scholar]
- 5.Fogagnolo F., Kfuri M., Jr., Paccola C.A. Intramedullary fixation of pertrochanteric hip fractures with the short AO-ASIF proximal femoral nail. Arch Orthop Trauma Surg. 2004;124:31–37. doi: 10.1007/s00402-003-0586-9. [DOI] [PubMed] [Google Scholar]
- 6.Ma K.L., Wang X., Luan F.J. Proximal femoral nails antirotation, Gamma nails, and dynamic hip screws for fixation of intertrochanteric fractures of femur: a meta-analysis. Orthop Traumatol Surg. Res.: OTSR. 2014;100:859–866. doi: 10.1016/j.otsr.2014.07.023. [DOI] [PubMed] [Google Scholar]
- 7.Aktselis I., Kokoroghiannis C., Fragkomichalos E. Prospective randomised controlled trial of an intramedullary nail versus a sliding hip screw for intertrochanteric fractures of the femur. Int Orthop. 2014;38:155–161. doi: 10.1007/s00264-013-2196-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Simmermacher R.K., Ljungqvist J., Bail H. The new proximal femoral nail antirotation (PFNA) in daily practice: results of a multicentre clinical study. Injury. 2008;39:932–939. doi: 10.1016/j.injury.2008.02.005. [DOI] [PubMed] [Google Scholar]
- 9.Barton T.M., Gleeson R., Topliss C., Greenwood R., Harries W.J., Chesser T.J. A comparison of the long gamma nail with the sliding hip screw for the treatment of AO/OTA 31-A2 fractures of the proximal part of the femur: a prospective randomized trial. J Bone Joint Surg Am. 2010;92:792–798. doi: 10.2106/JBJS.I.00508. [DOI] [PubMed] [Google Scholar]
- 10.Flores S.A., Woolridge A., Caroom C., Jenkins M. The utility of the tip-apex distance in predicting axial migration and cutout with the trochanteric fixation nail system helical blade. J Orthop Trauma. 2016;30:e207–e211. doi: 10.1097/BOT.0000000000000505. [DOI] [PubMed] [Google Scholar]
- 11.Li H., Wang Q., Dai G.G., Peng H. PFNA vs. DHS helical blade for elderly patients with osteoporotic femoral intertrochanteric fractures. Eur Rev Med Pharmacol Sci. 2018;22:1–7. doi: 10.26355/eurrev_201807_15346. [DOI] [PubMed] [Google Scholar]
- 12.Wang C., Wang Q. Helical blade compression failure occurred during PFNA implantation: a rare case and ingenious solution. Medicine (Baltim) 2019;98 doi: 10.1097/MD.0000000000016584. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Li H., Wang H., Zhang Y., Zhou J., Yuan F. The migration of helical blade and the tip apex distance value in cephalomedullary nail for geriatric intertrochanteric fractures. Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2019;33:1234–1238. doi: 10.7507/1002-1892.201904125. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.van Leur J.P.H., Jakma T.S.C., Willemsen S.P., Punt B.J. Trochanteric fixation nail(R) with helical blade compared with femoral neck screw for operative treatment of intertrochanteric femoral fractures. Hip Pelvis. 2019;31:48–56. doi: 10.5371/hp.2019.31.1.48. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Albareda J., Laderiga A., Palanca D., Paniagua L., Seral F. Complications and technical problems with the gamma nail. Int Orthop. 1996;20:47–50. doi: 10.1007/s002640050026. [DOI] [PubMed] [Google Scholar]
- 16.Heinz T., Vecsei V. [Complications and errors in use of the gamma nail. Causes and prevention] Chirurg. 1994;65:943–952. [PubMed] [Google Scholar]
- 17.Hesse B., Gachter A. Complications following the treatment of trochanteric fractures with the gamma nail. Arch Orthop Trauma Surg. 2004;124:692–698. doi: 10.1007/s00402-004-0744-8. [DOI] [PubMed] [Google Scholar]
- 18.Huang H., Xin J., Ma B. Analysis of complications of intertrochanteric fracture treated with Gamma 3 intramedullary nail. Int J Clin Exp Med. 2014;7:3687–3693. [PMC free article] [PubMed] [Google Scholar]
- 19.Lyddon D.W., Jr. The prevention of complications with the gamma locking nail. Am J Orthoped. 1996;25:357–363. [PubMed] [Google Scholar]
- 20.Pascarella R., Cucca G., Maresca A. Methods to avoid gamma nail complications. Chir Organi Mov. 2008;91:133–139. doi: 10.1007/s12306-007-0030-3. [DOI] [PubMed] [Google Scholar]
- 21.Pascarella R., Fantasia R., Maresca A. How evolution of the nailing system improves results and reduces orthopedic complications: more than 2000 cases of trochanteric fractures treated with the gamma nail system. Musculoskelet Surg. 2016;100:1–8. doi: 10.1007/s12306-015-0391-y. [DOI] [PubMed] [Google Scholar]
- 22.Rosenthal R.E. Prevention of complications with the gamma nail. Am J Orthoped. 1996;25:729. [PubMed] [Google Scholar]
- 23.Schupfner R., Kasmann L.T., Wagner W., Schulz A.P. Complications in treatment of 31-A fractures with trochanteric gamma nail (TGN) versus Gamma3 nail (G3N) - a review of 217 cases. Open Orthop J. 2016;10:389–395. doi: 10.2174/1874325001610010389. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Chen C.Y., Chiu F.Y., Chen C.M., Huang C.K., Chen W.M., Chen T.H. Surgical treatment of basicervical fractures of femur--a prospective evaluation of 269 patients. J Trauma. 2008;64:427–429. doi: 10.1097/01.ta.0000239255.47280.6f. [DOI] [PubMed] [Google Scholar]
- 25.Hak D.J., Bilat C. Avoiding varus malreduction during cephalomedullary nailing of intertrochanteric hip fractures. Arch Orthop Trauma Surg. 2011;131:709–710. doi: 10.1007/s00402-010-1182-4. [DOI] [PubMed] [Google Scholar]
- 26.Harrington P., Nihal A., Singhania A.K., Howell F.R. Intramedullary hip screw versus sliding hip screw for unstable intertrochanteric femoral fractures in the elderly. Injury. 2002;33:23–28. doi: 10.1016/s0020-1383(01)00106-1. [DOI] [PubMed] [Google Scholar]
- 27.Koyuncu S., Altay T., Kayali C., Ozan F., Yamak K. Mechanical failures after fixation with proximal femoral nail and risk factors. Clin Interv Aging. 2015;10:1959–1965. doi: 10.2147/CIA.S96852. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Massoud E.I. Fixation of basicervical and related fractures. Int Orthop. 2010;34:577–582. doi: 10.1007/s00264-009-0814-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Williams W.W., Parker B.C. Complications associated with the use of the gamma nail. Injury. 1992;23:291–292. doi: 10.1016/0020-1383(92)90169-s. [DOI] [PubMed] [Google Scholar]
- 30.Yang Z., Liu H., Xie X., Tan Z., Qin T., Kang P. Total hip arthroplasty for failed internal fixation after femoral neck fracture versus that for acute displaced femoral neck fracture: a comparative study. J Arthroplast. 2015;30:1378–1383. doi: 10.1016/j.arth.2015.02.037. [DOI] [PubMed] [Google Scholar]
- 31.Akan K., Cift H., Ozkan K., Eceviz E., Tasyikan L., Eren A. Effect of osteoporosis on clinical outcomes in intertrochanteric hip fractures treated with a proximal femoral nail. J Int Med Res. 2011;39:857–865. doi: 10.1177/147323001103900319. [DOI] [PubMed] [Google Scholar]
- 32.Boldin C., Seibert F.J., Fankhauser F., Peicha G., Grechenig W., Szyszkowitz R. The proximal femoral nail (PFN)--a minimal invasive treatment of unstable proximal femoral fractures: a prospective study of 55 patients with a follow-up of 15 months. Acta Orthop Scand. 2003;74:53–58. doi: 10.1080/00016470310013662. [DOI] [PubMed] [Google Scholar]
- 33.Stern R., Lubbeke A., Suva D., Miozzari H., Hoffmeyer P. Prospective randomised study comparing screw versus helical blade in the treatment of low-energy trochanteric fractures. Int Orthop. 2011;35:1855–1861. doi: 10.1007/s00264-011-1232-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Shin Y.S., Chae J.E., Kang T.W., Han S.B. Prospective randomized study comparing two cephalomedullary nails for elderly intertrochanteric fractures: zimmer natural nail versus proximal femoral nail antirotation II. Injury. 2017;48:1550–1557. doi: 10.1016/j.injury.2017.04.011. [DOI] [PubMed] [Google Scholar]
- 35.McKinley J.C., Robinson C.M. Treatment of displaced intracapsular hip fractures with total hip arthroplasty: comparison of primary arthroplasty with early salvage arthroplasty after failed internal fixation. J Bone Joint Surg Am. 2002;84:2010–2015. doi: 10.2106/00004623-200211000-00016. [DOI] [PubMed] [Google Scholar]
- 36.Zhang X., Liu Y., Ren K., Liu J., Zhu B., Sun Z. [Secondary total hip arthroplasty for osteonecrosis of femoral head after failed internal fixation of femoral neck fracture] Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2010;24:257–261. [PubMed] [Google Scholar]
- 37.Price M.V., Molloy S., Solan M.C., Sutton A., Ricketts D.M. The rate of instrument breakage during orthopaedic procedures. Int Orthop. 2002;26:185–187. doi: 10.1007/s00264-002-0344-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Chouhan D.K., Sharma S. "Push back" technique: a simple method to remove broken drill bit from the proximal femur. World J Orthop. 2015;6:847–849. doi: 10.5312/wjo.v6.i10.847. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Williams R.L., Kunzru K.M. A new technique for the removal of a broken cannulated screw. Injury. 1995;26:59–60. doi: 10.1016/0020-1383(95)90556-d. [DOI] [PubMed] [Google Scholar]