Abstract
Purpose:
To investigate direct imaging of trabecular bone using a three-dimensional adiabatic inversion recovery prepared ultrashort echo time Cones (3D IR-UTE-Cones) sequence.
Methods:
The proposed 3D IR-UTE-Cones sequence employed a broadband adiabatic inversion pulse together with a short TR/TI combination to suppress signals from long T2 tissues such as muscle and marrow fat, followed by multispoke UTE acquisition to detect signal from short T2 water components in trabecular bone. The feasibility of this technique for robust suppression of long T2 tissues was first demonstrated through numerical simulations. Then, the proposed IR-UTE-Cones sequence was applied to a hip agarose bone phantom and to six healthy volunteers for morphological and quantitative T2* and proton density mapping of trabecular bone.
Results:
Numerical simulation suggests that the IR technique with a short TR/TI combination provides sufficient suppression of long T2 tissues with a wide range of T1s. High contrast imaging of trabecular bone can be achieved ex vivo and in vivo, with fitted T2* values of 0.3–0.45 ms and proton densities of 5–9 mol/L.
Conclusion:
The 3D IR-UTE-Cones sequence with a short TR/TI combination provides robust suppression of long T2 tissues and allows both selective imaging and quantitative (T2* and proton density) assessment of short T2 water components in trabecular bone in vivo.
Keywords: trabecular bone, ultrashort echo time, adiabatic inversion recovery
Introduction
Trabecular bone is highly responsive to metabolic stimuli and has a turnover rate about eight times higher than that of cortical bone (1,2), making it a prime target for detecting bone loss in early osteoporosis (OP). Areal bone mineral density (BMD) of trabecular bone in the spine and/or hip using dual energy X-ray absorptiometry (DEXA) is the most commonly used clinical diagnostic test for assessing skeletal status and fracture risk (3,4). However, a number of clinical studies have demonstrated the limitations of BMD measurements. It has been recognized that BMD can only account for about 60% of bone strength (5).
For the last two decades, quantitative magnetic resonance imaging (MRI) has been used to assess the properties of trabecular bone, including T2* or T2’ of bone marrow (6,7) and high resolution imaging of bone microstructure (2,4), which are helpful in predicting osteoporotic fracture risk. Quantification of T2* or T2’ takes advantage of the fact that local field is perturbed due to the difference in susceptibility between trabecular bone and marrow, which is affected by the density and structure of trabecular bone (6,7). High resolution imaging can directly visualize dark trabecular bone due to its low water content and short T2 relaxation. With image postprocessing, it is possible to obtain 3D architecture and corresponding structural parameters of the trabecular bone, which are highly related to the bone strength (2,4).
Direct MR imaging of trabecular bone is technically challenging due to its fast signal decay, rendering it “invisible” with conventional clinical MRI pulse sequences (8). Recently, ultrashort echo time (UTE) sequences and their variants (e.g. water- and fat-suppressed proton projection MRI (WASPI) and zero echo time (ZTE) sequences) have been developed to directly capture the fast decaying signals of phosphorus (9–11) or proton (12–14) in trabecular bone. These techniques have been used successfully for in vivo quantitative imaging of phosphorus density and its relaxations in both cortical and trabecular bone, strong indicators for characterizing bone quality (9–11). However, the hardware for phosphorus imaging is not available in most clinical scanners, posing a limitation for further optimization of the technique and for translation to clinical practice.
Recently, researchers have also investigated direct proton imaging of trabecular bone using WASPI and fat-suppressed UTE techniques (12–14). In order to create a high contrast for trabecular bone, it is critical to suppress signals from long T2 tissues since trabecular bone has a much lower proton density than long T2 tissues. WASPI uses two hard pulses with narrow frequency bands to selectively excite water and fat signals. Strong gradient crushers were used following the two hard pulses to saturate water and fat signals before data acquisition (15). Alternative scans using 180° pulses have been considered to further null residual water and fat signals. Wurnig et al. employed the UTE sequence to measure T2* of trabecular bone samples with a SPIR (spectral presaturation with inversion recovery) module to suppress marrow fat (14). These two techniques show promising results for trabecular bone imaging both qualitatively and quantitatively. However, both techniques are sensitive to B1 and B0 inhomogeneities, which may limit their clinical applications, especially for in vivo imaging of trabecular bone in the spine and hip.
In this study, we propose a broadband adiabatic inversion recovery prepared three-dimensional UTE Cones (3D IR-UTE-Cones) sequence for direct volumetric imaging of trabecular bone in the human spine and hip (16,17). To the best of our knowledge, this is the first proton imaging study to directly image trabecular bone in the human spine and hip on a clinical 3T scanner. The combination of a short repetition time (TR) and inversion time (TI) is chosen in order for the IR-UTE-Cones sequence to obtain robust suppression of a variety of long T2 tissues with different T1s. Using the adiabatic full passage (AFP) pulse with a relatively wide bandwidth (~1.6 kHz), the proposed IR preparation is insensitive to both B1 and B0 inhomogeneities (18). Furthermore, multispoke acquisition per IR preparation can be incorporated, allowing time-efficient volumetric imaging and T2* quantification of trabecular bone (17,19). Proton density can also be quantified by comparing 3D IR-UTE-Cones signal of trabecular bone with that of a calibration phantom. Numerical simulations, ex vivo studies, and in vivo studies were conducted to validate the feasibility of the proposed IR-UTE-Cones sequence to directly image and quantify trabecular bone.
Theory
Figure 1 shows the features of the 3D IR-UTE-Cones pulse sequence used in this study (16,17). The adiabatic IR pulse with a specific inversion recovery time of TI is repeated every TR period (Figure 1A). Following the IR pulse are Nsp separate k-space acquisitions with an equal time interval τ for fast data acquisition. TI is defined as the time from the center of the adiabatic IR pulse to the center of the multispoke acquisition. A short rectangular pulse with a duration of 30–60 μs is used for signal excitation in each spoke (Figure 1B), followed by spiral trajectories with 3D conical view ordering (Figure 1C).
Figure 1.
The 3D IR-UTE-Cones sequence employs an adiabatic inversion pulses for long T2 suppression, followed by 3D UTE-Cones data acquisition (A). In the basic 3D UTE-Cones sequence, a short rectangular pulse is used for signal excitation followed by 3D spiral sampling with a minimal nominal TE of 32 μs (B). The spiral trajectories are arranged with conical view ordering (C). To speed up data acquisition, multiple spokes can be sampled after each long T2 preparation (A).
The adiabatic IR pulse can effectively invert the longitudinal magnetizations of long T2 tissues, such as marrow fat and muscle. However, the longitudinal magnetizations of short T2 tissues such as bone are typically saturated, not inverted, by the relatively long adiabatic inversion pulse (20,21). Here, we introduce an inversion efficiency factor Q for the adiabatic IR pulse with a range of −1 (signifying full inversion) to 1 (signifying no disturbance to the z-magnetization) (21). Q is equal to zero in the condition of complete saturation.
To simplify the signal equation, a rectangular pulse was considered for excitation. At steady state, the longitudinal magnetization of the jth spoke is expressed as follows (17,22):
| [1] |
where
| [2] |
| [3] |
| [4] |
using the following definitions: , , E1 = exp{−[TI – τ(Nsp − 1)/2]/T1}, E2 = exp{−[TR − TI – τ(Nsp − 1)/2]/T1}, and e1 = exp(−τ/T1). M0 is the signal intensity in the equilibrium state. Mp is the longitudinal magnetization after the IR pulse; its explicit derivation can be found in the Appendix section. fz is the longitudinal magnetization mapping function that describes the response of the magnetization to the RF pulse, with . and are defined as the longitudinal magnetizations before and after RF excitation. fz is introduced to account for the signal loss during the RF excitation when tissue T2 is close to or less than RF pulse duration. The expression of the longitudinal magnetization mapping function is shown as follows (23):
| [5] |
where α is the excitation flip angle and d is the pulse duration. For the tissue with a T2 >> d, the T2 decaying during excitation can be neglected; thus, fz can be simplified to the conventional cos(α).
For short T2 tissues (e.g. T2 < 1 ms), both Q and Mp approach 0. Therefore, Eq. [1] can be simplified to . The signal of the jth acquisition from the short T2 component can be expressed as follows:
| [6] |
Long T2 Signal Suppression
It is difficult to completely suppress all the long T2 tissues with different T1s using a single IR pulse. However, when the TR is shorter, the nulling TIs for all the tissues get closer. Moreover, for long T2 tissues with longer T1 relaxation times, it is easier to achieve sufficient signal suppression with a broader range of TIs. More details can be found in the simulation section. When several spokes near the nulling point are acquired, the excited transverse magnetizations before the nulling point are of opposite polarity to those acquired after the nulling point. Then, long T2 signal suppression can be achieved because these transverse magnetizations cancel each other out in the regridding process during image reconstruction.
The magnetizations of short T2 tissues (such as trabecular bone) are not inverted, but instead largely saturated by the adiabatic IR pulse. They typically have a short T1 (24) and quickly recover to positive longitudinal magnetizations at TI. The signal intensities of short or long T2 tissues are both proportional to the magnetization averaging of the multispoke acquisitions:
| [7] |
A general framework to minimize signals from long T2 tissues for the IR-prepared sequence is expressed as follows:
| [8] |
where is the number of long T2 tissues. T1i (i = 1, 2, …, ) is the T1 value of the ith long T2 tissue. When TR, α, τ, and Nsp are given, TI can be determined by Eq. [8] to achieve optimized long T2 suppression. This framework can apply for suppressing either a single tissue component with an individual T1 or a group of long T2 tissues with a range of T1s.
Methods
The 3D IR-UTE-Cones sequence (Figure 1) was implemented on a 3T clinical scanner (MR750, GE Healthcare Technologies, Milwaukee, WI). The Cones sequence sampled data along evenly spaced twisting paths in the shape of multiple cones (16). Data sampling started from the center of k-space as soon as possible after the RF excitation with a minimal nominal TE of 32 μs (17,22). The adiabatic IR pulse with a pulse shape of commonly used hyperbolic secant function, duration of 6.048 ms, bandwidth of 1.643 kHz, and maximum B1 amplitude of 17 μT was used to invert or saturate tissues (25). The adiabatic IR pulse was centered on −220Hz in the middle of the water and fat peak at 3T.
Simulation
Numerical simulation was performed to investigate the efficiency of the IR preparation scheme in the suppression of long T2 signals with different TRs (i.e. 50, 100, 150, 200, 250, and 1000 ms). The simulated T1 values of the long T2 tissues ranged from 200 to 2000 ms. α, τ and Nsp were set to 20°, 4 ms, and 5, respectively, for all simulations. The excitation pulse duration d was 60 μs, which is much shorter than typical bone T2* (i.e. around 300 μs). Thus, the longitudinal magnetization mapping function fz can be simplified to cos(α).
Additionally, the contrast between trabecular bone and long T2 tissues was also investigated for the IR-UTE-Cones sequence with different TRs. The TI was determined by Eq. [8] in order to minimize the marrow fat signal since it is relatively difficult to suppress due to its relatively short T1 and since it is a dominant component in trabecular bone. The T1 values of marrow fat are assumed to be in the range of 320–350 ms, and the T1 value of trabecular bone is set to 140 ms (17,24). The proton density of trabecular bone is assumed to be 12 percent of the long T2 tissues (26).
Trabecular Bone Sample Study
Two hip bone samples (65-year-old female and 71-year-old male donors) were individually embedded in agarose gel (3 w/v %) to simulate human tissues. An 8-channel transmit/receive knee coil was used for both RF transmission and signal reception. Both clinical T1-FSE and the proposed IR-UTE-Cones sequences were used for the hip-agarose phantom experiment; the sequence parameters are listed as follows: 1) 2D T1-FSE: TR/TE = 550/8.1 ms, FOV = 15×15 cm2, slice thickness = 5 mm, acquisition matrix = 320×256, slice number = 32, and scan time = 59 s; 2) 3D IR-UTE-Cones: TR/TI = 82/37 ms, flip angle = 20°; Nsp = 3; τ = 7 ms; FOV = 16×16×21 cm3; matrix = 160×160×42; and five separate scans with TEs = 0.032/3.3, 0.2/4.4, 0.4, 0.8, and 1.1 ms to measure T2*, each with scan time = 4 min 20 sec.
In Vivo Trabecular Bone Study
In vivo spine imaging was performed on six healthy volunteers (24–38 years of age, 5 males and 1 female). Informed consent was obtained from all subjects in accordance with guidelines of the institutional review board. A rubber band with a T2* around 0.3 ms was placed between the volunteer and the spine coil during scanning to serve as a reference to calibrate the proton density of trabecular bone. The proton density of trabecular bone can be calculated by the following equation (11,27):
| [9] |
with
| [10] |
where fxy is the mapping function that describes the response of the transverse magnetization to a constant-amplitude RF pulse. It is a function of T2 and pulse duration (23). fz is expressed in Eq. [5]. If the pulse duration and tissue T2 are known, fxy and fz can be calculated directly. Ibone and Iref are the signal intensities of trabecular bone and rubber band, respectively. To measure the proton density of the rubber band used in this study, an H2O-D2O phantom was made with 20% H2O and 80% D2O by volume. It was doped with MnCl2 to achieve a T2* of 0.34 ms and a T1 of 6.5 ms. The T2* and T1 of the rubber band are 0.38 ms and 200 ms, respectively. The T1 relaxation was measured with our previously developed 3D UTE AFI-VTR method (28). The H2O-D2O phantom and rubber band were put together and scanned with the proposed IR-UTE-Cones sequence with TR/TI = 150/64 ms. Together with the measured T1 and T2* values of the H2O-D2O phantom and rubber band, the proton density of the used rubber band calculated by Eq. [9] was around 19 mol/L. In addition, T1 values of the trabecular bone was set to 140 ms (24,29).
To correct the signal intensity bias due to the coil sensitivity inhomogeneity of the spine coil, the regular 3D UTE-Cones sequence was applied twice using spine and body coils, respectively, for signal reception. Then, with the assumption that the body coil has a homogeneous reception profile, the coil sensitivity map of the spine coil was calculated by dividing UTE-Cones images acquired with the spine coil by UTE-Cones images acquired with the body coil. The final spine trabecular bone images were generated by dividing the 3D IR-UTE-Cones images by the obtained coil sensitivity map. The sequence parameters used for imaging of the spine were as follows: 1) T2-FSE: TR/TE = 4370/103 ms, FOV = 34×34 cm2, slice thickness = 3.2 mm, matrix = 360×270, slice number = 14, and scan time = 1 min 5 sec; 2) 3D IR-UTE-Cones: TR/TI = 150/64 ms, TE = 0.032 ms, flip angle = 18°, Nsp = 5, τ = 3.8 ms, FOV = 34×34×16 cm3, matrix = 160×160×32, oversampling factor = 4 (the center of k-space was oversampled by a factor of 4 to reduce artifacts), and scan time = 10 min; 3) 3D UTE-Cones: TR = 6 ms, TE = 0.032 ms, flip angle = 2°, FOV = 34×34×16 cm3, matrix = 160×160×32, oversampling factor = 4, and scan time = 1 min. T2* measurement was used to evaluate the efficiency of long T2 suppression in 3D IR-UTE-Cones imaging of trabecular bone. A single-component short T2* means a sufficient suppression of pore water and fat in trabecular bone is achieved, and that only bound water in trabecular bone is detected by the 3D IR-UTE-Cones sequence. Four separate scans with TEs = 0.032/2.2, 0.2, 0.4, and 0.8 ms were employed to measure T2* in three of the volunteer experiments.
In vivo hip imaging was performed on six healthy volunteers (24–36 years of age, three males and three females). A routine cardiac coil was used for the hip scan. The sequence parameters used for hip imaging were as follows: 1) T2-FSE: TR/TE = 10000/92 ms, FOV = 38×38 cm2, slice thickness = 3 mm, matrix = 320×320, slice number = 24, nex = 2, scan time = 3 min; 2) 3D IR-UTE-Cones: TR/TI = 150/64 ms, TE = 0.1 ms, flip angle = 18°, Nsp = 5, τ = 4.1 ms, FOV = 38×38×20 cm3, matrix = 160×160×40, oversampling factor = 3.2, and scan time = 9 min 32 sec.
Data Analysis
The trust-region-reflective algorithm was used to solve the non-linear minimization of Eq. [8] (30). A single exponential function was employed for T2* fitting of the multiple-TE IR-UTE-Cones data. The 3D UTE-Cones images acquired with both spine and body coils were smoothed using a 3D Gaussian kernel with standard deviation of 2 before the coil sensitivity calculation. All analysis algorithms were written in Matlab (The MathWorks Inc., Natick, MA, USA) and were executed offline on the DICOM images obtained by the acquisition protocols described above.
Results
Numerical simulations of the signal variations in the IR-UTE sequence (i.e. |Mz|) for a wide range of T1s (i.e., [200, 2000] ms) are shown in Figure 2. The TI ranges from 0 to TR for each T1. The best signal null point for each T1 is located in the dark blue spot. The dark blue region becomes wider when T1 is longer, demonstrating that the signal suppression for long T1 tissues is less sensitive to the choice of TI. Thus, sufficient signal suppression of longer T1 tissues can be achieved with a wider range of TIs. If there are several long T2 tissues to be suppressed, it is a good choice to set the TI at the null point of the shortest T1 tissue. Moreover, the null points get closer for all the T1s when the TR is shorter. Therefore, it is much easier to sufficiently suppress long T2 tissues with a wide range of T1s when a short TR is used in the single IR type sequences.
Figure 2.
Numerical simulation to investigate the effects of TR and TI on signal suppression of long T2 tissues with a wide range of T1s (i.e. [200, 2000] ms) in IR-UTE imaging. The signal intensities of the long T2 tissues in IR-UTE imaging vary with both TI and T1. TI ranges from 0 to TR. TRs are chosen at 50 (A), 100 (B), 150 (C), 200 (D), 250 (E), and 1000ms (F).
Figure 3 shows the simulation results of the contrast between bone and long T2 tissues in IR-UTE imaging. The TIs are calculated by optimizing the Eq. [8] to null signal from marrow fat. Similar to the findings in Figure 2, the contrast between bone and long T2 tissues is higher when a shorter TR is used. Even with a TR of 250 ms, we still obtain moderate bone contrast. In the case of long TR (e.g., 1000 ms), only a small range of T1s could be well suppressed, suggesting that the signal suppression efficiency is very sensitive to the choice of TI.
Figure 3.
Investigation of the trabecular bone contrast generated by the IR-UTE sequence with different TRs. The Sratio is defined as the signal intensity ratio between trabecular bone and long T2 tissue. The T1s of long T2 tissues ranged from 200 to 2000 ms and the T1 of trabecular bone was assumed to be 140 ms. The Sratio curves with relatively short TRs of 50, 100, 150, 200, 250 ms are shown in A and the curve with a much longer TR of 1000 ms is shown in B. The optimal TI for each TR was determined by minimizing Eq. [8] to null marrow fat. Improved trabecular bone contrast is achieved when a shorter TR is used in the IR-UTE sequence.
Figure 4 shows representative images selected from 3D IR-UTE-Cones imaging of a hip agarose phantom. Both agarose and marrow fat are bright with the conventional T1-FSE sequence, but almost fully suppressed with the IR-UTE-Cones sequence. Fitting of the 3D IR-UTE-Cones images with different TEs ranging from 0.032 to 4.4 ms demonstrates a short T2* of 0.41 ± 0.02 ms for hip trabecular bone, which is similar to the T2* of bound water in cortical bone (26). The single-component decay with R2 = 0.998 and the short T2* value suggest that marrow fat and free water of the hip sample are well suppressed by the adiabatic inversion pulse. Only water bound to the organic matrix of trabecular bone in the hip is selectively imaged with the 3D IR-UTE-Cones sequence.
Figure 4.
3D IR-UTE-Cones imaging of a hip-agarose phantom. The long T2 agarose and marrow fat are bright in the clinical T1-FSE image (A). In comparison, signals of agarose are almost fully suppressed in the IR-UTE-Cones images with TEs of 0.032 ms (B), 0.2 ms (C), 0.4 ms (D), 0.8 ms (E), and 4.4 ms (F). Single-component fitting is achieved for trabecular bone in the hip sample with a short T2* of 0.41 ± 0.02 ms (G), which demonstrates that long T2 water and marrow fat are sufficiently suppressed in the IR-UTE-Cones images.
Figure 5 shows in vivo imaging of the spine of a 36-year-old male volunteer. Similar to the hip agarose study, the 3D IR-UTE-Cones sequence shows high contrast between the vertebrae and surrounding soft tissues. These results confirm that the IR technique with a short TR allows nulling of both muscle and fat despite their large differences in T1 relaxation times. Exponential fitting of the 3D IR-UTE-Cones images with different TEs demonstrates a short T2* of 0.31 ± 0.01 ms for trabecular bone in the vertebrae.
Figure 5.
In vivo imaging of the spine of a 36-year-old male volunteer using the 3D IR-UTE-Cones sequence with TEs of 0.032 ms (A), 0.2 ms (B), 0.4 ms (C), 0.8 ms (D), and 2.2 ms (E). The surrounding soft tissues such as muscle and fat are well-suppressed, leading to high contrast imaging of vertebrae and ligaments in the spine. Single-component fitting is achieved for vertebrae with a short T2* of 0.31 ± 0.01 ms, which demonstrates that long T2 water and marrow fat are sufficiently suppressed in the IR-UTE-Cones images.
Figure 6 shows in vivo images of the hip of a 24-year-old female volunteer. In contrast to the conventional T2-weighted FSE images, soft tissues are well-suppressed, but cortical bone is bright in the corresponding IR-UTE-Cones images. Trabecular bone of the hip shows lower proton density compared with cortical bone.
Figure 6.
In vivo imaging of the hip of a 24-year-old female volunteer with a clinical 2D T2-weighted FSE (first column) and 3D IR-UTE-Cones (second column) sequences. The long T2 muscle and fat are bright in the clinical T2-FSE images. In comparison, the soft tissues are well-suppressed in the 3D IR-UTE-Cones images, demonstrating a high contrast for cortical and trabecular bone in the hip.
Figure 7 shows the bound water proton density map of vertebrae in a 31-year-old male volunteer. Since the coil sensitivity of the spine coil is quite inhomogeneous, signal intensity correction is critical for accurate quantitative proton density mapping. It can be found that images of the vertebrae are much more uniform after coil sensitivity correction. Proton densities of the vertebrae calculated by Eq. [9] range from 5 to 9 mol/L.
Figure 7.
In vivo qualitative and quantitative imaging of the spine of a 31-year-old male volunteer using the 3D IR-UTE-Cones sequence. The long T2 muscle and fat are bright in the clinical T2-FSE image (A). The original 3D IR-UTE-Cones image (B) show non-uniform signal intensity distribution due to the inhomogeneous coil sensitivity (C) of the spine coil. After the coil sensitivity correction, the spine bone image demonstrates a more uniform signal intensity distribution (D). The proton density map of the spine trabecular bone (E) is calculated based on the coil sensitivity corrected 3D IR-UTE-Cones image (D).
Discussion
We demonstrated in this study that the 3D IR-UTE-Cones sequence with a short TR/TI combination can suppress signals from long T2 water and fat simultaneously, and can provide high image contrast for short T2 trabecular bone. Our simulation study suggests that the TI should be selected close to the null point of short T1 tissues since the long T1 tissue suppression is less sensitive to the selection of TI. We observed that the shorter the TR of the IR-UTE sequence, the better we were able to suppress long T2 tissues with a wide range of T1s since their signal null points were getting closer. Our ex vivo and in vivo studies demonstrated the robustness of the 3D IR-UTE-Cones sequence in suppressing long T2 water and fat signals in the spine and hip. Furthermore, the 3D IR-UTE-Cones sequence allowed quantitative proton density mapping and T2* measurement of the short T2 water component in trabecular bone.
UTE techniques can provide direct imaging of short T2 bone, which is invisible with conventional sequences. The majority of bone studies using qualitative and quantitative UTE imaging are focused on cortical bone (31–34). However, evaluation of trabecular bone may be even more valuable since most osteoporotic fractures occur at locations that are rich in trabecular bone (35). WASPI and UTE with SPIR preparation have been proposed for trabecular bone imaging. However, these two techniques are sensitive to B1 and B0 inhomogeneities, making them perhaps unsuitable for in vivo spine and hip imaging. In comparison, an adiabatic IR pulse with a relatively broad spectral coverage of 1.643 kHz is used in the proposed 3D IR-UTE-Cones sequence, and the long T2 suppression is less sensitive to B1 and B0 inhomogeneities (18). Together with a short TR and a short TI, the proposed IR-UTE-Cones sequence is more robust in suppressing both water and fat. In this study, a TR of 150 ms was used in 3D IR-UTE-Cones imaging of the spine and hip in vivo to balance the effectiveness of long T2 suppression and specific absorption rate (SAR) limitation. Compared with DEXA, which is a 2D projection imaging technique that cannot distinguish between cortical and trabecular bone, the proposed 3D IR-UTE-Cones MR imaging technique can provide volumetric information of cortical and trabecular bone separately. Since the bound water proton density in cortical bone is typically much higher than that in trabecular bone (36,37), a threshold-based method can potentially be used to separate cortical and trabecular bone in 3D IR-UTE-Cones images. Therefore, the proposed 3D IR-UTE-Cones MR imaging technique may have significant advantages over the current gold standard, DEXA, which is a 2D projection imaging technique that cannot distinguish between cortical and trabecular bone.
The T2* values measured in both ex vivo and in vivo studies were in the range of 0.3–0.45 ms, which are similar to the T2*s of bound water in cortical bone (26,29). Thus, the proton density measured by the proposed 3D IR-UTE-Cones technique is likely collagen-bound water proton density with effective suppression of pore water components (19,24,38). The collagen bone matrix provides tensile strength and elasticity in bone (39). Thus, it would be useful to obtain information from the collagen bone matrix to evaluate bone quality. MR imaging of collagen matrix-bound water has been studied by several groups in recent years as a possible surrogate measure of collagen bone matrix (36,40,41). For example, the bound water proton density is highly correlated with collagen matrix density, with R2 = 0.74, as reported in a previous study of cortical bone samples (41). A lower bound water proton density may indicate a more degenerative collagen matrix with less tensile strength/elasticity in bone. We expect that the measured volumetric proton density in this study can be used as a potential biomarker to evaluate the bone quality in early osteoporosis and osteoporotic fracture risk.
Long T2 signal contamination is typically a major source of error in quantitative UTE imaging (26). Since the IR-UTE-Cones sequence allows for selective imaging of short T2 tissues, it can also be used for quantitative evaluation of proton density and T2*, as well as for T1 relaxation times (24,26). In this study, the IR-UTE-Cones acquisition together with a reference rubber phantom provided volumetric mapping of proton density for the trabecular bone, which may be a useful biomarker to evaluate the bone quality. A relatively high data oversampling factor was used in both in vivo spine and hip imaging with the IR-UTE-Cones sequence, which can increase the image signal to noise ratio (SNR) and simultaneously reduce the motion (as a result of motion averaging) and aliasing artifacts. Respiratory gating is also an effective strategy to reduce motion artifacts due to breathing. Since the marrow fat has a relatively short T2* between 5 and 15 ms (42), the inversion efficiency Q may not reach −1. To account for this imperfect inversion and to achieve a sufficient nulling of marrow fat, a smaller TI with 1–2 ms less than the TI calculated by the Eq. [8] was used in our study. The proposed 3D IR-UTE-Cones sequence can also be used for cortical bone imaging both morphologically and quantitatively (e.g., T2* and proton density).
Both the rubber band and the manganese-doped water (which has an extremely short T2) can serve as the reference to calibrate the trabecular bone proton density. The rubber band has closer T1 and T2* relaxations to the bound bone water than the manganese-doped water (which has a T1 that is much shorter than that of bound bone water). Thus, the rubber band has a contrast more similar to the bound bone water. Using the rubber band as the reference may be more resistant to the error in bound bone water quantification than using the manganese-doped water within a wide range of sequence parameters. On the other hand, there is a problem with the rubber band (a polybutadiene type elastomer), which may have more than one resonance (due to chemical shift), leading to errors in bound bone water quantification. We will compare the robustness and accuracy of bound bone water quantification using the two references in our future investigations.
The IR-UTE-Cones sequence with a short TR/TI combination can be readily used for imaging of other short T2 species, such as for direct imaging of myelin protons in the white matter of the brain. There are several groups of water protons, such as those in cerebrospinal fluid, extra- and intra-cellular water, and water trapped in the myelin bilayers, which may have different T1s (43); therefore, efficient suppression of all kinds of water protons is essential for selective imaging of myelin protons (13,20,44). The 3D IR-UTE-Cones sequence with a short TR/TI combination can largely suppress various water groups with different T1s as shown in Figure 2, and may thusly be used for more robust imaging of myelin where water components with various T1s may exist in white matter of the brain (43).
This study has several limitations. First, the SNR and contrast to noise ratio (CNR) of the 3D IR-UTE-Cones sequence have not been evaluated or compared with existing trabecular bone imaging techniques such as WASPI or UTE with SPIR preparation. The current protocol for quantitation of T2* relaxation is too long (i.e. 40 min) for clinical utility. We would expect the qualitative assessment of 3D IR-UTE-Cones images together with quantitative proton density mapping (12 min scan in total) to be the more suitable option for clinical use. Advanced fast imaging techniques, such as compressed sensing or deep learning, could be used to accelerate the scan (45–47). Third, improved long T2 suppression can be achieved with a shorter TR for 3D IR-UTE-Cones imaging of trabecular bone. The minimal TR used in the proposed 3D IR-UTE-Cones sequence is constrained by the SAR limitation, especially for body imaging (e.g., spine imaging). A dedicated transmission RF coil with a higher transmit efficiency than that of the used body coil would be helpful to alleviate the SAR limitation. Fourth, the correlation of 3D IR-UTE-Cones measured proton density with organic matrix density in trabecular bone is of high interest and will be investigated in future studies (36,40,41). Fifth, although the 3D IR-UTE-Cones sequence is ready for in vivo study of healthy volunteers and patients, the clinical value of this technique remains to be investigated.
Conclusion
The 3D IR-UTE-Cones sequence with a short TR/TI combination provides robust suppression of long T2 tissues such as muscle and marrow fat, thus allowing selective imaging of short T2 trabecular bone. The 3D IR-UTE-Cones sequence also provides quantitative T2* and proton density measurement for trabecular bone with little long T2 contamination.
Supplementary Material
Supporting information Figure S1 Steady state magnetization and timing for the 3D IR-UTE-Cones sequence. The short and long arrows represent the excitation and inversion pulses respectively. Q is the inversion efficiency of the used inversion pulse. t1 and t2 are the duration from the center of the IR pulse to the first excitation pulse and the duration from the last excitation pulse to the center of the IR pulse, respectively. Mp, Mz,1, and Mz,2 are the longitudinal magnetizations after the IR pulse, after the last excitation pulse and before the IR pulse, respectively.
Acknowledgments
The authors acknowledge grant support from GE Healthcare and NIH (1R21 AR073496, 1R01 AR062581, 1R01 AR068987 and 1R01 NS092650) and the VA Clinical Science R&D Service (I01CX001388 and I01RX002604).
Appendix
Supporting information Figure S1 shows the element timing of the IR multispoke sequence. t1 is the duration from the center of the IR pulse to the first excitation pulse. t2 is the duration from the last excitation pulse to the center of the IR pulse. Mp, Mz,1 and Mz,2 are the longitudinal magnetizations after the IR pulse, after the last excitation pulse and before the IR pulse, respectively. The relations of above three magnetizations according to Bloch equations, which are expressed as follows:
| [A1] |
| [A2] |
| [A3] |
where t1 = TI – 0.5τ(Nsp − 1) and t2 = TR − TI – 0.5τ(Nsp − 1). and can be found in the Theory section. Mp is determined from Eq. [A1]–[A3] and its final expression is shown in Eq. [4].
References
- 1.Frost HM. Dynamics of bone remodelling. In: Boston: Little Brown; 1964. pp. 315–334. [Google Scholar]
- 2.Majumdar S. Magnetic resonance imaging of trabecular bone structure. Top. Magn. Reson. Imaging TMRI 2002;13:323–334. [DOI] [PubMed] [Google Scholar]
- 3.Wahner HW, Fogelman I. The evaluation of osteoporosis: dual energy X-ray absorptiometry in clinical practice. In: Cambridge: Cambridge University Press; 1994. p. 296. [Google Scholar]
- 4.Wehrli FW. Structural and functional assessment of trabecular and cortical bone by micro magnetic resonance imaging. J. Magn. Reson. Imaging JMRI 2007;25:390–409 doi: 10.1002/jmri.20807. [DOI] [PubMed] [Google Scholar]
- 5.Wehrli FW, Saha PK, Gomberg BR, et al. Role of magnetic resonance for assessing structure and function of trabecular bone. Top. Magn. Reson. Imaging TMRI 2002;13:335–355. [DOI] [PubMed] [Google Scholar]
- 6.Majumdar S, Thomasson D, Shimakawa A, Genant HK. Quantitation of the susceptibility difference between trabecular bone and bone marrow: experimental studies. Magn. Reson. Med 1991;22:111–127. [DOI] [PubMed] [Google Scholar]
- 7.Ford JC, Wehrli FW. In vivo quantitative characterization of trabecular bone by NMR interferometry and localized proton spectroscopy. Magn. Reson. Med 1991;17:543–551. [DOI] [PubMed] [Google Scholar]
- 8.Robson MD, Gatehouse PD, Bydder M, Bydder GM. Magnetic resonance: an introduction to ultrashort TE (UTE) imaging. J. Comput. Assist. Tomogr 2003;27:825–846. [DOI] [PubMed] [Google Scholar]
- 9.Robson MD, Gatehouse PD, Bydder GM, Neubauer S. Human imaging of phosphorus in cortical and trabecular bone in vivo. Magn. Reson. Med 2004;51:888–892 doi: 10.1002/mrm.20055. [DOI] [PubMed] [Google Scholar]
- 10.Wu Y, Reese TG, Cao H, et al. Bone mineral imaged in vivo by 31P solid state MRI of human wrists. J. Magn. Reson. Imaging JMRI 2011;34:623–633 doi: 10.1002/jmri.22637. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Zhao X, Song HK, Wehrli FW. In vivo bone 31 P relaxation times and their implications on mineral quantification. Magn. Reson. Med 2018;80:2514–2524 doi: 10.1002/mrm.27230. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Wu Y, Dai G, Ackerman JL, et al. Water- and fat-suppressed proton projection MRI (WASPI) of rat femur bone. Magn. Reson. Med 2007;57:554–567 doi: 10.1002/mrm.21174. [DOI] [PubMed] [Google Scholar]
- 13.Weiger M, Stampanoni M, Pruessmann KP. Direct depiction of bone microstructure using MRI with zero echo time. Bone 2013;54:44–47 doi: 10.1016/j.bone.2013.01.027. [DOI] [PubMed] [Google Scholar]
- 14.Wurnig MC, Calcagni M, Kenkel D, et al. Characterization of trabecular bone density with ultra-short echo-time MRI at 1.5, 3.0 and 7.0 T--comparison with micro-computed tomography. NMR Biomed 2014;27:1159–1166 doi: 10.1002/nbm.3169. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Wu Y, Ackerman JL, Chesler DA, Graham L, Wang Y, Glimcher MJ. Density of organic matrix of native mineralized bone measured by water- and fat-suppressed proton projection MRI. Magn. Reson. Med 2003;50:59–68 doi: 10.1002/mrm.10512. [DOI] [PubMed] [Google Scholar]
- 16.Gurney PT, Hargreaves BA, Nishimura DG. Design and analysis of a practical 3D cones trajectory. Magn. Reson. Med 2006;55:575–582 doi: 10.1002/mrm.20796. [DOI] [PubMed] [Google Scholar]
- 17.Carl M, Bydder GM, Du J. UTE imaging with simultaneous water and fat signal suppression using a time-efficient multispoke inversion recovery pulse sequence. Magn. Reson. Med 2016;76:577–582 doi: 10.1002/mrm.25823. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Tannús A, Garwood M. Adiabatic pulses. NMR Biomed 1997;10:423–434 doi: 10.1002/(SICI)1099-1492(199712). [DOI] [PubMed] [Google Scholar]
- 19.Li C, Magland JF, Zhao X, Seifert AC, Wehrli FW. Selective in vivo bone imaging with long-T2 suppressed PETRA MRI. Magn. Reson. Med 2017;77:989–997 doi: 10.1002/mrm.26178. [DOI] [PubMed] [Google Scholar]
- 20.Horch RA, Gore JC, Does MD. Origins of the Ultrashort-T21H NMR Signals in Myelinated Nerve: A Direct Measure of Myelin Content? Magn. Reson. Med 2011;66:24–31 doi: 10.1002/mrm.22980. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Larson PEZ, Conolly SM, Pauly JM, Nishimura DG. Using adiabatic inversion pulses for long-T2 suppression in ultrashort echo time (UTE) imaging. Magn. Reson. Med 2007;58:952–961 doi: 10.1002/mrm.21341. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Ma Y-J, Zhu Y, Lu X, Carl M, Chang EY, Du J. Short T2 imaging using a 3D double adiabatic inversion recovery prepared ultrashort echo time cones (3D DIR-UTE-Cones) sequence. Magn. Reson. Med 2018;79:2555–2563 doi: 10.1002/mrm.26908. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Sussman MS, Pauly JM, Wright GA. Design of practical T2-selective RF excitation (TELEX) pulses. Magn. Reson. Med 1998;40:890–899. [DOI] [PubMed] [Google Scholar]
- 24.Chen J, Chang EY, Carl M, et al. Measurement of bound and pore water T1 relaxation times in cortical bone using three-dimensional ultrashort echo time cones sequences. Magn. Reson. Med 2017;77:2136–2145 doi: 10.1002/mrm.26292. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Silver MS, Joseph RI, Hoult DI. Highly selective π2 and π pulse generation. J. Magn. Reson. 1969 1984;59:347–351. [Google Scholar]
- 26.Du J, Bydder GM. Qualitative and quantitative ultrashort-TE MRI of cortical bone. NMR Biomed 2013;26:489–506 doi: 10.1002/nbm.2906. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Techawiboonwong A, Song HK, Leonard MB, Wehrli FW. Cortical bone water: in vivo quantification with ultrashort echo-time MR imaging. Radiology 2008;248:824–833 doi: 10.1148/radiol.2482071995. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Ma Y-J, Lu X, Carl M, et al. Accurate T1 mapping of short T2 tissues using a three-dimensional ultrashort echo time cones actual flip angle imaging-variable repetition time (3D UTE-Cones AFI-VTR) method. Magn. Reson. Med 2018;80:598–608. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Seifert AC, Wehrli SL, Wehrli FW. Bi-component T2* analysis of bound and pore bone water fractions fails at high field strengths. NMR Biomed 2015;28:861–872. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Coleman T, Li Y. An Interior Trust Region Approach for Nonlinear Minimization Subject to Bounds. SIAM J. Optim 1996;6:418–445 doi: 10.1137/0806023. [DOI] [Google Scholar]
- 31.Jerban S, Lu X, Jang H, et al. Significant correlations between human cortical bone mineral density and quantitative susceptibility mapping (QSM) obtained with 3D Cones ultrashort echo time magnetic resonance imaging (UTE-MRI). Magn. Reson. Imaging 2019. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Jerban S, Ma Y, Dorthe EW, et al. Assessing cortical bone mechanical properties using collagen proton fraction from ultrashort echo time magnetization transfer (UTE-MT) MRI modeling. Bone Rep 2019:100220. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Ma Y-J, Chang EY, Carl M, Du J. Quantitative magnetization transfer ultrashort echo time imaging using a time-efficient 3D multispoke Cones sequence. Magn. Reson. Med 2018;79:692–700. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Jerban S, Ma Y, Wong JH, et al. Ultrashort echo time magnetic resonance imaging (UTE-MRI) of cortical bone correlates well with histomorphometric assessment of bone microstructure. Bone 2019;123:8–17. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Wehrli FW, Song HK, Saha PK, Wright AC. Quantitative MRI for the assessment of bone structure and function. NMR Biomed 2006;19:731–764 doi: 10.1002/nbm.1066. [DOI] [PubMed] [Google Scholar]
- 36.Manhard MK, Horch RA, Harkins KD, Gochberg DF, Nyman JS, Does MD. Validation of quantitative bound- and pore-water imaging in cortical bone. Magn. Reson. Med 2014;71:2166–2171 doi: 10.1002/mrm.24870. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Jerban S, Ma Y, Li L, et al. Volumetric mapping of bound and pore water as well as collagen protons in cortical bone using 3D ultrashort echo time cones MR imaging techniques. Bone 2019. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Li S, Ma L, Chang EY, et al. Effects of inversion time on inversion recovery prepared ultrashort echo time (IR-UTE) imaging of bound and pore water in cortical bone. NMR Biomed 2015;28:70–78. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Mehta SS, Öz OK, Antich PP. Bone elasticity and ultrasound velocity are affected by subtle changes in the organic matrix. J. Bone Miner. Res 1998;13:114–121. [DOI] [PubMed] [Google Scholar]
- 40.Cao H, Ackerman JL, Hrovat MI, Graham L, Glimcher MJ, Wu Y. Quantitative bone matrix density measurement by water-and fat-suppressed proton projection MRI (WASPI) with polymer calibration phantoms. Magn. Reson. Med. Off. J. Int. Soc. Magn. Reson. Med 2008;60:1433–1443. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Seifert AC, Li C, Wehrli SL, Wehrli FW. A Surrogate Measure of Cortical Bone Matrix Density by Long T2-Suppressed MRI. J. Bone Miner. Res 2015;30:2229–2238. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Kühn J-P, Hernando D, Meffert PJ, et al. Proton-density fat fraction and simultaneous R2* estimation as an MRI tool for assessment of osteoporosis. Eur. Radiol 2013;23:3432–3439. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Deoni SCL, Rutt BK, Arun T, Pierpaoli C, Jones DK. Gleaning multicomponent T1 and T2 information from steady-state imaging data. Magn. Reson. Med 2008;60:1372–1387 doi: 10.1002/mrm.21704. [DOI] [PubMed] [Google Scholar]
- 44.Sheth V, Shao H, Chen J, et al. Magnetic Resonance Imaging of Myelin Using Ultrashort Echo Time (UTE) Pulse Sequences: Phantom, Specimen, Volunteer and Multiple Sclerosis Patient Studies. NeuroImage 2016;136:37–44 doi: 10.1016/j.neuroimage.2016.05.012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Lustig M, Donoho D, Pauly JM. Sparse MRI: The application of compressed sensing for rapid MR imaging. Magn. Reson. Med 2007;58:1182–1195. [DOI] [PubMed] [Google Scholar]
- 46.Zhu B, Liu JZ, Cauley SF, Rosen BR, Rosen MS. Image reconstruction by domain-transform manifold learning. Nature 2018;555:487–492 doi: 10.1038/nature25988. [DOI] [PubMed] [Google Scholar]
- 47.Wu Y, Ma Y, Liu J, Du J, Xing L. Self-attention convolutional neural network for improved MR image reconstruction. Inf. Sci 2019;490:317–328. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Supporting information Figure S1 Steady state magnetization and timing for the 3D IR-UTE-Cones sequence. The short and long arrows represent the excitation and inversion pulses respectively. Q is the inversion efficiency of the used inversion pulse. t1 and t2 are the duration from the center of the IR pulse to the first excitation pulse and the duration from the last excitation pulse to the center of the IR pulse, respectively. Mp, Mz,1, and Mz,2 are the longitudinal magnetizations after the IR pulse, after the last excitation pulse and before the IR pulse, respectively.







