Abstract
Since the introduction of the radial head prosthesis (RHP) in 1941, many designs have been introduced. It is not clear whether prosthesis design parameters are related to early failure. The aim of this systematic review is to report on failure modes and to explore the association between implant design and early failure.
A search was conducted to identify studies reporting on failed primary RHP. The results are clustered per type of RHP based on: material, fixation technique, modularity, and polarity. Chi-square tests are used to compare reasons for failure between the groups.
Thirty-four articles are included involving 152 failed radial head arthroplasties (RHAs) in 152 patients. Eighteen different types of RHPs have been used.
The most frequent reasons for revision surgery after RHA are (aseptic) loosening (30%), elbow stiffness (20%) and/or persisting pain (17%). Failure occurs after an average of 34 months (range, 0–348 months; median, 14 months).
Press-fit prostheses fail at a higher ratio because of symptomatic loosening than intentionally loose-fit prostheses and prostheses that are fixed with an expandable stem (p < 0.01).
Because of the many different types of RHP used to date and the limited numbers and evidence on early failure of RHA, the current data provide no evidence for a specific RHP design.
Cite this article: EFORT Open Rev 2019;4:659-667. DOI: 10.1302/2058-5241.4.180099
Keywords: elbow, failure, radial head prosthesis, removal, replacement, revision
Introduction
Since the introduction of the radial head prosthesis (RHP) in 1941,1 many alterations in designs and materials have been proposed and tried that have varied in terms of material, fixation technique, modularity, and polarity. Radial head arthroplasty (RHA) is predominantly used to treat comminuted radial head fractures and other, less common, chronic posttraumatic sequels as nonunion, posttraumatic arthritis and elbow instability.2,3
During the past 75 years, moderate to good results have been reported for both primary4,5 and revision surgery of RHA.6 Implant revision and removal rates up to 8% at four years have been described.4 More recent studies showed conflicting 10-year survival numbers ranging from 61% to 97%.5,7
This raises questions of whether implant- or fixation-related factors may be related to early failure. Except in the case of silicone RHPs, that have previously proved to be biologically and biomechanically insufficient, with a substantial risk of fragmentation of the implant8–10 resulting in silicone synovitis, it is unclear which type of metallic RHP is superior. Taking the enormous discrepancies in failure rates into account, the aim of the current study was to report on failure modes of RHPs in recent years and to explore the association between implant design and early failure.
Materials and methods
Search strategy
This systematic review was based on Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines.11,12 A comprehensive literature search was conducted with the assistance of a clinical librarian using the following terms: radius[MeSH], radius fractures[MeSH], arthroplasty, replacement[MeSH], joint prosthesis[MeSH], radial head[tiab], replacement[tiab], arthroplasty[tiab], prosthesis implantation[tiab], and prosthesis[tiab]. The PubMed/MEDLINE and Embase databases were searched using the filters “English” and “humans” for the period from January 1941 to the date of search (10 September 2018). The start date was chosen as the first documentation of a radial head replacement by Speed dating back to 1941.1
Inclusion and exclusion criteria
This review was intended to include patients with a minimum age of 18 years who underwent revision surgery of their metallic of pyrocarbon RHA for any reason. Articles written in English and evaluating original clinical data on primary pyrocarbon or metallic RHPs requiring revision surgery were considered, regardless of the level of evidence. Only articles including at least five cases with a minimum of follow-up of two years were considered. No minimum of failed RHPs per article was set.
A study was excluded if the type of prosthesis and/or the mode of failure was not reported and was not provided by the author on request. Moreover, silicone RHPs were excluded, since these prostheses have been shown to be inferior and presumably would not be implanted nowadays.8
Study selection
Three authors independently assessed all titles and abstracts and identified eligible articles (IFK, JV and AH). Two authors (IFK and JV) assessed the full text of all eligible studies and made the final decision regarding inclusion. Disagreements were settled by discussion. With the use of this strategy, 952 articles were identified. After the screening of the title, abstract, methodology and results, 72 articles were found to be potentially eligible for inclusion. The full text of these studies was analysed, and the reference lists of all eligible publications were manually checked for additional studies potentially meeting the inclusion criteria. After application of the inclusion and exclusion criteria, 34 studies were finally included. The additional 38 articles were excluded for various reasons (Fig. 1).
Fig. 1.
Flowchart.
Note. RHP, radial head prosthesis.
Outcome parameters
The primary outcome of the current study is the failure mode of the RHP as defined in the included articles. Secondary outcome measures are 1 the time between primary surgery and failure (i.e. time to failure) and 2 the type of revision surgery (i.e. removal of the prosthesis, replacement with another RHP or revision surgery to a total elbow arthroplasty (TEA) or radiocapitellar prosthesis).
Data analysis
To summarize the data, descriptive statistics were used. Only 18 of the 34 included studies reported data on individual patients. The other 16 studies reported only pooled data on age at time of primary surgery. As a consequence, analyses covering all 34 studies had to be performed on the aggregated study level, with the data on the individual patients pooled per study. In 14 articles (67 patients), no data are available on the time to failure. In the remaining patients (n = 85), the Kruskal–Wallis test is used to compare time to failure per type of fixation. To compare the reason of failure related to polarity, Fisher’s exact test is used. In addition, for comparison of the failure mode related to the type of fixation, Chi-square tests are used, followed by a post hoc analysis.
Results
Thirty-four articles involving a total of 152 failed RHAs in 152 individual patients were included. The number of failed RHAs per study ranged from 1 to 22. All studies were case series (level-IV therapeutic studies). The oldest article was published in 1993, and the most recent article was published in 2018.
Population characteristics
Mean age at time of primary surgery was 50 years (SD = 10). The studies included 18 different types of RHPs. Most frequent used prostheses, representing 53% of all prostheses, included the Evolve Modular Radial Head (Wright Medical Group, Arlington, Tennessee) (n = 45, 30%), the MoPyC radial head (Tornier, Montbonnot, France) (n = 23, 15%) and the Guepar (De Puy Synthes, Johnson & Johnson, West Chester, Pennsylvania) (n = 15, 10%). Most prostheses were either intentionally loose-fit (n = 53, 35%) or press-fit (n = 47, 31%). A smaller proportion was placed with use of cement (n = 29, 19%) or had an expandable stem (n = 23, 15%). Regarding implant material, 127 implants (84%) were made of non-specified metal (including titanium and cobalt chromium), 23 (15%) were made of pyrocarbon and 2 (1%) were made of Vitallium. Regarding polarity, 106 prostheses (70%) were monopolar and 46 (30%) were bipolar (Table 1).
Table 1.
Included studies.
| Study number | First author | Year | Study design | Failed prosthesis (n) | Type of prosthesis | Material | Polarity | Modularity | Fixation | Removals (n) |
Replacement by RHP (n) | Revision to TEA or RC (n) | Mean follow-up (mo) |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 1 | Ricón13 | 2018 | Case series | 3 | MoPyC (Tornier) | PC | Mono | Modular | Expansion stem | 3 | 0 | 0 | 72 |
| 2 | Kachooei14 | 2018 | Case series | 3 | Mixed | CC | Bi | Modular | Mixed | 0 | 0 | 3 to RC | 28 |
| 3 | Sershon7 | 2018 | Case series | 1 | Katalyst (Integra) | CC | Bi | Modular | Int. Loose | 0 | 1 | 0 | 1 |
| 4 | Viveen6 | 2017 | Case series | 8 | Mixed | Mixed | Mixed | Mixed | Mixed | 0 | 8 | 0 | 19 |
| 5 | Strelzow15 | 2017 | Case series | 2 | Evolve (Wright) | CC | Mono | Modular | Int. Loose | 0 | 2 | 0 | NA |
| 6 | Hackl16 | 2017 | Case series | 5 | MoPyC (Tornier) | PC | Mono | Modular | Expansion stem | 0 | 4 | 1 to TEA | 25 |
| 7 | Laumonerie5 | 2017 | Case series | 19 | Mixed | CC | Mixed | Modular | Mixed | 19 | 0 | 0 | NA |
| 8 | Laflamme17 | 2017 | Case series | 1 | ExploR (Biomet) | CC | Bi | Modular | Press-fit | 0 | 1 | 0 | 6 |
| 9 | Kachooei18 | 2016 | Case series | 22 | Mixed | CC | Mono | Modular | Mixed | 19 | 3 | 0 | 22 |
| 10 | Van Hoecke19 | 2016 | Case series | 2 | Judet CRF II (Tornier) | CC | Bi | Modular | Cemented | 1 | 0 | 1 to TEA | 94 |
| 11 | Lópiz20 | 2016 | Case series | 4 | MoPyC (Tornier) | PC | Mono | Modular | Expansion stem | 1 | 3 | 0 | NA |
| 12 | Heijink21 | 2016 | Case series | 1 | RHS (Tornier) | Metal | Bi | Modular | Cemented | 1 | 0 | 0 | 24 |
| 13 | Kodde22 | 2016 | Case series | 3 | RHS (Tornier) | Metal | Bi | Modular | Press-fit | 0 | 2 | 1 to RC | 62 |
| 14 | Moghaddam23 | 2016 | Case series | 7 | Evolve (Wright) | CC | Mono | Modular | Int. Loose | 4 | 3 | 0 | NA |
| 15 | Levy24 | 2016 | Case series | 2 | Acumed | CC | Mono | Modular | Press-fit | 0 | 2 | 0 | 12 |
| 16 | Yan25 | 2015 | Case series | 1 | Radius Head Comp. (Link) | CC | Mono | Monoblock | Int. Loose | 0 | 1 | 0 | NA |
| 17 | Neuhaus26 | 2015 | Case series | 13 | Mixed | CC | Mixed | Mixed | Mixed | 13 | 0 | 0 | 12 |
| 18 | Schnetzke27 | 2014 | Case series | 6 | Evolve (Wright) | CC | Mono | Modular | Int. Loose | 4 | 2 | 0 | NA |
| 19 | Allavena28 | 2014 | Case series | 5 | Guepar (DePuy) | CC | Bi | Modular | Cemented | 4 | 1 | 0 | 28 |
| 20 | Watters29 | 2014 | Case series | 3 | Evolve (Wright) | CC | Mono | Modular | Int. Loose | 0 | 0 | 3 to TEA | NA |
| 21 | Katthagen30 | 2013 | Case series | 1 | Corin Radial Head (Corin) | CC | Mono | Monoblock | Press-fit | 1 | 0 | 0 | NA |
| 22 | Sarris31 | 2012 | Case series | 2 | MoPyC (Tornier) | PC | Mono | Modular | Expansion stem | 2 | 0 | 0 | 1 |
| 23 | Flinkkilä32 | 2012 | Case series | 9 | Mixed | CC | Mono | Modular | Press-fit | 9 | 0 | 0 | NA |
| 24 | Rotini33 | 2012 | Case series | 2 | rHead (Sbi) | CC | Mixed | Modular | Press-fit | 2 | 0 | 0 | 18 |
| 25 | Zunkiewicz34 | 2012 | Case series | 1 | Katalyst (Integra) | CC | Bi | Modular | Int. Loose | 0 | 1 | 0 | NA |
| 26 | Ricón35 | 2012 | Case series | 3 | MoPyC (Tornier) | PC | Mono | Modular | Expansion stem | 3 | 0 | 0 | 38 |
| 27 | Lamas36 | 2011 | Case series | 5 | MoPyC (Tornier) | PC | Mono | Modular | Expansion stem | 5 | 0 | 0 | NA |
| 28 | Burkhart37 | 2010 | Case series | 2 | Judet CRF II (Tornier) | CC | Bi | Modular | Cemented | 0 | 2 | 0 | 0 |
| 29 | Doornberg38 | 2007 | Case series | 2 | Evolve (Wright) | CC | Mono | Modular | Int. Loose | 2 | 0 | 0 | NA |
| 30 | Wretenberg39 | 2006 | Case series | 5 | Radius Head Comp. (Link) | CC | Mono | Monoblock | Int. Loose | 5 | 0 | 0 | NA |
| 31 | Brinkman40 | 2005 | Case series | 2 | Judet CRF II (Tornier) | CC | Bi | Modular | Cemented | 0 | 2 | 0 | 9 |
| 32 | Harrington41 | 2001 | Case series | 4 | Richards (Smith & Nephew) | Titanium | Mono | Monoblock | Press-fit | 4 | 0 | 0 | 237 |
| 33 | Smets42 | 2000 | Case series | 1 | Predecessor of Judet CRF II (Tornier) | CC | Bi | Modular | Cemented | 1 | 0 | 0 | 8 |
| 34 | Knight43 | 1993 | Case series | 2 | Osteonics radial head prosthesis (Stryker Howmedica) | Vitallium | Mono | Monoblock | Press-fit | 2 | 0 | 0 | NA |
| Total | 152 | 105 | 38 | 9 | 36 |
Note. NA, Not applicable; RHP, radial head prosthesis; TEA, total elbow arthroplasty; RC, radiocapitellar prosthesis; CC, cobalt-chromium; PC, pyrocarbon; Mono, monopolar; Bi, bipolar; Int. Loose, intentionally loose-fit.
Primary outcome – failure mode
The most prevalent failure mode was symptomatic aseptic loosening, occurring in 46 (30% of all failures) patients (Fig. 2). Of these 46 prostheses, 25 were placed press-fit, in 11 cement was used, six were intentionally loose-fit, and four had an expandable stem. Post hoc analyses revealed that symptomatic aseptic loosening was significantly more frequently the reason for revision in press-fit prostheses (25/47 press-fit prostheses) compared to intentionally loose-fit prostheses (11/29 intentionally loose-fit prostheses) and prostheses with an expandable stem (4/23 prostheses with an expandable stem) (p < 0.01) (Fig. 2).
Fig. 2.
Failure modes divided per type of fixation.
*Press-fit prostheses fail more often because of symptomatic loosening compared to intentionally loose-fit prostheses and prostheses with an expandable stem (p < 0.01).
§Intentionally loose-fit prostheses fail more often because of stiffness compared to press-fit protheses (p < 0.01).
#Cemented prostheses fail more often because of ulnohumeral arthritis compared to press-fit prostheses and intentionally loose-fit prostheses (p < 0.01).
A second failure mode was elbow stiffness (n = 30, 20% of all failures). Intentionally loose-fit prostheses were more frequently revised for elbow stiffness than press-fit prostheses (20/53 loose-fit prostheses versus 3/47 press-fit prostheses, p < 0.01) (Fig. 2). Among the 20 intentionally loose-fit prostheses, monoblock designs failed more often than modular designs (6/6 intentionally loose-fit monoblock prostheses versus 14/47 intentionally loose-fit modular prostheses, p < 0.01).
Other modes of failure were persistent pain (n = 26, 17% of all failures), overstuffing (n = 13, 9% of all failures) and dissociation of the prosthesis (n = 8, 5% of all failures). Of the eight dissociated prostheses, five were bipolar and three were monopolar (Fig. 3). Ulnohumeral arthritis was the reason for revision in six cases. Cemented prostheses (5/29 cemented prostheses) were more often revised for ulnohumeral arthritis than press-fit (0/47 press-fit prostheses) and intentionally loose-fit prostheses (5/29 versus 0/47 versus 0/53 versus, p < 0.01) (Fig. 2). All failures due to instability involved bipolar prostheses (p < 0.01) (Fig. 3). In one case with instability, only the head of the prosthesis was revised because of under sizing.
Fig. 3.
Failure modes divided into monopolar and bipolar.
*Bipolar prostheses fail more often because of instability compared to monopolar prostheses (p < 0.01).
Secondary outcomes
Time to failure
Time to failure was reported in 85 patients and ranged from 0–348 months (mean 34 months; median 14 months) (Table 1). Mean time to failure was 53 months for press-fit prostheses (n = 29), 36 months for prostheses with an expandable stem (n = 14), 27 months for cemented prostheses (n = 16), and 17 months for intentionally loose-fit prostheses (n = 26). Intentionally loose-fit prostheses failed earlier compared to press-fit prostheses (p < 0.01).
Type of revision surgery
Sixty-nine per cent (n = 105) of the revision surgeries involved removal of the prosthesis. In another 25% (n = 38) the prosthesis was removed and a new RHP was implanted. In addition, five RHPs were revised to TEAs (3%) and four RHPs were revised to radiocapitellar prostheses (3%) (Table 1). In only two out of five revisions to a TEA the indication for revision was ulnohumeral arthritis.
Discussion
This systematic review shows that the most frequent failure modes of RHAs are symptomatic aseptic loosening (30%), stiffness (20%) and persistent pain (17%) at an average time to failure of 34 months. Post hoc analyses revealed that press-fit RHPs failed more often because of symptomatic aseptic loosening (25/47 prostheses) compared to intentionally loose-fit prostheses (5/43 prostheses) and prostheses with an expandable stem (4/23 prostheses). In addition, intentionally loose-fit prostheses failed earlier compared to press-fit prostheses (17 versus 53 months, respectively).
Aseptic loosening is a frequently encountered problem. Radiolucencies around the prosthesis are frequently reported and seem to occur mostly shortly after implantation. Whether these radiolucencies also mean that a prosthesis is loose, is not always not clear. Subcollar resorption is often reported with press-fit prostheses, but seems to be stationary after one to two years, without progression to loosening and without clinical symptoms.22 In cases of progressive radiographic signs of loosening, a poor clinical outcome could be expected. In those cases, an additional computed tomography scan (CT scan) with or without a bone scan could be performed to investigate whether loose bodies are present and to assess the chondral condition of the capitellum and the ulnohumeral joint, in order to plan the appropriate treatment. Our analysis showed that 27% of the monopolar implants and 37% of the bipolar implants failed because of symptomatic aseptic loosening (p = 0.3). In contrast, van Riet et al. had observed more loosening with monopolar compared to bipolar prostheses.44 It has been hypothesized that poor bone ingrowth onto the stem of the press-fit prosthesis due to micromotion of the prosthesis within the medullary canal is one of the causes of aseptic loosening in monopolar implants.32,45 Possibly the bipolar design results in reduced stress and micromotion at the implant–bone interface.32,45
Obviously, most indications for revision of RHA are associated with pain in the elbow or forearm. Pain is the symptom, not the cause and pain can have many reasons other than a failed prosthesis. Interestingly, 26 patients (17% of all failures) underwent revision surgery solely for persisting pain. The question is what the underlying pathology (i.e. true failure mode) in these cases had been. O’Driscoll and Herald suggested that pain in the proximal forearm in patients with a press-fit RHP is a strong indicator for symptomatic loosening, even in the absence of radiographic signs of loosening.46 In the analysis 11/26 revisions for pain involved press-fit prostheses. This could imply that the prostheses could have been loose in this group. However, the remaining 15/26 prostheses were cemented in place or intentionally loose-fit. Further studies on this phenomenon are needed.
Moreover, this study revealed that intentionally loose-fit protheses failed earlier compared to press-fit protheses (a mean time to failure of 17 versus 53 months, respectively). A possible explanation for this from our data could be that intentionally loose-fit protheses failed more often because of stiffness compared to press-fit protheses. In general, stiffness is a problem encountered early on after elbow trauma and/or surgery47-49 and could have different underlying problems in the case of RHAs: over sizing of the head, stiffness because of the (surgically) injured soft tissues around the elbow joint or a loose stem followed by migration of the implant. A clear explanation in the cases of the patients included in this study remains unknown, since no additional data were available.
The strengths of the current review are the selection criteria for our studies that were set to include series with enough patients and follow-up time of the implants. As far as we know, there has been only one other review on revisions of RHPs.3 The primary objective of that review was to determine the incidence of revision or removal after RHPs placed for acute fractures. According to that review, the main reason for revision surgery was heterotopic ossification (HO). However, in the current study, there were no cases of HO at all. This discrepancy is likely the result of the fact that Kachooei et al.3 included a radiographic outcome study by Ha et al. that described nearly all cases having HO (33 patients).50 The study by Ha et al. was excluded in the current review because the follow-up was too short and the types of RHPs used unclear.
Several limitations are recognized. Due to the small numbers it was not possible to perform a meta-analysis on the extracted data. Since there are many different types of RHP included (n = 18), it is, with the relatively small numbers of primary implantations and revision cases, not possible to draw firm statistical conclusions. Moreover, prosthesis polarity, material, and fixation technique are not independent of each other. Thus, there are only eight combinations in practice, instead of the maximum of 32 possible combinations (two different polarities, four different materials, and four different techniques of fixation). These eight possible combinations reflect the true spectrum of available prostheses.4
Other limitations are the lack of reports in some studies on perioperative findings and individual time to failure of primary RHPs. In only 85 of the 152 patients the individual time to failure was reported. Then, although intentionally loose-fit prostheses were shown to fail earlier than press-fit prostheses, most other possible statistical comparisons of times to failure between the different fixation methods seemed to be underpowered. Also, studies regarding RHA are mostly mid-term follow-up. There are only a few studies with long-term follow-up (more than 10 years) available in the literature.7,19
In order to make a meta-analysis possible, a more uniform way of reporting indications for revision surgery and results is important. We think that the development of guidelines for standardized patient reported outcome measures (PROMs) and registration of clinical and radiographic outcomes is essential. National arthroplasty registries should play a leading role in this. Moreover, the outcomes after removal or revision of an RHP should be registered as well, as it is currently unknown which one is the preferred treatment for failed RHA and this choice seems to be more dependent on the preference of the hospital or surgeon, rather than on some level of evidence.18
Conclusions
In conclusion, the most frequent reasons for revision surgery after primary RHA are symptomatic (aseptic) loosening, elbow stiffness and/or persisting pain. Other, less common indications are technical failures such as overstuffing and dissociation of the implant. Failure occurs after an average of 34 months and the majority of the failed prostheses are removed. Taking into account the many different types of RHPs used and the various indications for revision surgery, the current data do not support a preference for a specific RHP design over one other. Guidelines for standardized follow-up are needed to improve our understanding of why RHPs fail.
Footnotes
OA licence text: This article is distributed under the terms of the Creative Commons Attribution-NonCommercial 4.0 International (CC BY-NC 4.0) licence (https://creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial use, reproduction and distribution of the work without further permission provided the original work is attributed.
ICMJE Conflict of interest statement: KLMK’s research institution receives financial support from ZimmerBiomet, Stryker and Mathys related and unrelated to specific research projects, but not related to the current work.
MPJB reports a research grant from Smith and Nephew and Wright/Tornier for this work. DE reports a research grant and a consulting fee or honorarium from Zimmer Biomet for this work; an Isakos grant and payment for lectures including service on speakers’ bureaus from AO International outside the submitted work.
The other authors declare no conflict of interest relevant to this work.
Funding statement
Although none of the authors has received or will receive benefits for personal or professional use from a commercial party related directly or indirectly to the subject of this article, benefits have been or will be received but will be directed solely to a research fund, foundation, educational institution, or other non-profit organization with which one or more of the authors are associated.
References
- 1. Speed K. Ferrule caps for the head of the radius. Surg Gynecol Obstet 1941;73:845–850. [Google Scholar]
- 2. Shore BJ, Mozzon JB, MacDermid JC, Faber KJ, King GJ. Chronic posttraumatic elbow disorders treated with metallic radial head arthroplasty. J Bone Joint Surg Am 2008;90:271–280. [DOI] [PubMed] [Google Scholar]
- 3. Kachooei AR, Baradaran A, Ebrahimzadeh MH, van Dijk CN, Chen N. The rate of radial head prosthesis removal or revision: a systematic review and meta-analysis. J Hand Surg Am 2018;43:39–53. [DOI] [PubMed] [Google Scholar]
- 4. Heijink A, Kodde IF, Mulder PG, et al. Radial head arthroplasty: a systematic review. JBJS Rev 2016;4:01874474-201610000-00001. [DOI] [PubMed] [Google Scholar]
- 5. Laumonerie P, Reina N, Ancelin D, et al. Mid-term outcomes of 77 modular radial head prostheses. Bone Joint J 2017;99-B:1197–1203. [DOI] [PubMed] [Google Scholar]
- 6. Viveen J, Kodde IF, Koenraadt KL, Beumer A, The B, Eygendaal D. Clinical and radiographic outcome of revision surgery of radial head prostheses: midterm results in 16 patients. J Shoulder Elbow Surg 2017;26:394–402. [DOI] [PubMed] [Google Scholar]
- 7. Sershon RA, Luchetti TJ, Cohen MS, Wysocki RW. Radial head replacement with a bipolar system: an average 10-year follow-up. J Shoulder Elbow Surg 2018;27:e38–e44. [DOI] [PubMed] [Google Scholar]
- 8. Duckworth AD, Wickramasinghe NR, Clement ND, Court-Brown CM, McQueen MM. Radial head replacement for acute complex fractures: what are the rate and risks factors for revision or removal? Clin Orthop Relat Res 2014;472:2136–2143. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Maghen Y, Leo AJ, Hsu JW, Hausman MR. Is a silastic radial head still a reasonable option? Clin Orthop Relat Res 2011;469:1061–1070. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Morrey BF, Askew L, Chao EY. Silastic prosthetic replacement for the radial head. J Bone Joint Surg Am 1981;63:454–458. [PubMed] [Google Scholar]
- 11. Moher D, Liberati A, Tetzlaff J, Altman DG, PRISMA Group. Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement. BMJ 2009;339:b2535. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Moher D, Shamseer L, Clarke M, et al. ; PRISMA-P Group. Preferred reporting items for systematic review and meta-analysis protocols (PRISMA-P) 2015 statement. Syst Rev 2015;4:1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13. Ricón FJ, Lajara F, Fuentes A, Aguilar ML, Boix A, Lozano JA. Pyrocarbon arthroplasty in acute unreconstructable radial head fractures: mid-term to long term results. J Orthop Traumatol 2018;19:13. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Kachooei AR, Heesakkers NAM, Heijink A, The B, Eygendaal D. Radiocapitellar prosthetic arthroplasty: short-term to midterm results of 19 elbows. J Shoulder Elbow Surg 2018;27:726–732. [DOI] [PubMed] [Google Scholar]
- 15. Strelzow JA, Athwal GS, MacDermid JC, et al. Effect of concomitant elbow injuries on the outcomes of radial head arthroplasty: a cohort comparison. J Orthop Trauma 2017;31:e327–e333. [DOI] [PubMed] [Google Scholar]
- 16. Hackl M, Wegmann K, Koslowsky TC, Zeifang F, Schoierer O, Müller LP. Rare implant-specific complications of the MoPyC radial head prosthesis. J Shoulder Elbow Surg 2017;26:830–837. [DOI] [PubMed] [Google Scholar]
- 17. Laflamme M, Grenier-Gauthier PP, Leclerc A, Antoniades S, Bédard AM. Retrospective cohort study on radial head replacements comparing results between smooth and porous stem designs. J Shoulder Elbow Surg 2017;26:1316–1324. [DOI] [PubMed] [Google Scholar]
- 18. Kachooei AR, Claessen FM, Chase SM, Verheij KK, van Dijk CN, Ring D. Factors associated with removal of a radial head prosthesis placed for acute trauma. Injury 2016;47:1253–1257. [DOI] [PubMed] [Google Scholar]
- 19. Van Hoecke E, Van De Vijver A, Van Glabbeek F, Gielen J. Long term results after bipolar radial head arthroplasty. Acta Orthop Belg 2016;82:382–388. [PubMed] [Google Scholar]
- 20. Lópiz Y, González A, García-Fernández C, García-Coiradas J, Marco F. Comminuted fractures of the radial head: resection or prosthesis? Injury 2016;47:S29–S34. [DOI] [PubMed] [Google Scholar]
- 21. Heijink A, Kodde IF, Mulder PGH, Van Dijk CN, Eygendaal D. Cemented bipolar radial head arthroplasty: midterm follow-up results. J Shoulder Elbow Surg 2016;25:1829–1838. [DOI] [PubMed] [Google Scholar]
- 22. Kodde IF, Heijink A, Kaas L, Mulder PG, van Dijk CN, Eygendaal D. Press-fit bipolar radial head arthroplasty, midterm results. J Shoulder Elbow Surg 2016;25:1235–1242. [DOI] [PubMed] [Google Scholar]
- 23. Moghaddam A, Raven TF, Dremel E, Studier-Fischer S, Grutzner PA, Biglari B. Outcome of radial head arthroplasty in comminuted radial head fractures: short and midterm results. Trauma Mon 2016;21:e20201. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24. Levy JC, Formaini NT, Kurowicki J. Outcomes and radiographic findings of anatomic press-fit radial head arthroplasty. J Shoulder Elbow Surg 2016;25:802–809. [DOI] [PubMed] [Google Scholar]
- 25. Yan M, Ni J, Song D, Ding M, Liu T, Huang J. Radial head replacement or repair for the terrible triad of the elbow: which procedure is better? ANZ J Surg 2015;85:644–648. [DOI] [PubMed] [Google Scholar]
- 26. Neuhaus V, Christoforou DC, Kachooei AR, Jupiter JB, Ring DC, Mudgal CS. Radial head prosthesis removal: a retrospective case series of 14 patients. Arch Bone Jt Surg 2015;3:88–93. [PMC free article] [PubMed] [Google Scholar]
- 27. Schnetzke M, Aytac S, Deuss M, et al. Radial head prosthesis in complex elbow dislocations: effect of oversizing and comparison with ORIF. Int Orthop 2014;38:2295–2301. [DOI] [PubMed] [Google Scholar]
- 28. Allavena C, Delclaux S, Bonnevialle N, Rongières M, Bonnevialle P, Mansat P. Outcomes of bipolar radial head prosthesis to treat complex radial head fractures in 22 patients with a mean follow-up of 50 months. Orthop Traumatol Surg Res 2014;100:703–709. [DOI] [PubMed] [Google Scholar]
- 29. Watters TS, Garrigues GE, Ring D, Ruch DS. Fixation versus replacement of radial head in terrible triad: is there a difference in elbow stability and prognosis? Clin Orthop Relat Res 2014;472:2128–2135. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30. Katthagen JC, Jensen G, Lill H, Voigt C. Monobloc radial head prostheses in complex elbow injuries: results after primary and secondary implantation. Int Orthop 2013;37:631–639. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31. Sarris IK, Kyrkos MJ, Galanis NN, Papavasiliou KA, Sayegh FE, Kapetanos GA. Radial head replacement with the MoPyC pyrocarbon prosthesis. J Shoulder Elbow Surg 2012;21:1222–1228. [DOI] [PubMed] [Google Scholar]
- 32. Flinkkilä T, Kaisto T, Sirniö K, Hyvönen P, Leppilahti J. Short- to mid-term results of metallic press-fit radial head arthroplasty in unstable injuries of the elbow. J Bone Joint Surg Br 2012;94:805–810. [DOI] [PubMed] [Google Scholar]
- 33. Rotini R, Marinelli A, Guerra E, Bettelli G, Cavaciocchi M. Radial head replacement with unipolar and bipolar SBi system: a clinical and radiographic analysis after a 2-year mean follow-up. Musculoskelet Surg 2012;96:S69–S79. [DOI] [PubMed] [Google Scholar]
- 34. Zunkiewicz MR, Clemente JS, Miller MC, Baratz ME, Wysocki RW, Cohen MS. Radial head replacement with a bipolar system: a minimum 2-year follow-up. J Shoulder Elbow Surg 2012;21:98–104. [DOI] [PubMed] [Google Scholar]
- 35. Ricón FJ, Sánchez P, Lajara F, Galán A, Lozano JA, Guerado E. Result of a pyrocarbon prosthesis after comminuted and unreconstructable radial head fractures. J Shoulder Elbow Surg 2012;21:82–91. [DOI] [PubMed] [Google Scholar]
- 36. Lamas C, Castellanos J, Proubasta I, Dominguez E. Comminuted radial head fractures treated with pyrocarbon prosthetic replacement. Hand (NY) 2011;6:27–33. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37. Burkhart KJ, Mattyasovszky SG, Runkel M, et al. Mid- to long-term results after bipolar radial head arthroplasty. J Shoulder Elbow Surg 2010;19:965–972. [DOI] [PubMed] [Google Scholar]
- 38. Doornberg JN, Parisien R, van Duijn PJ, Ring D. Radial head arthroplasty with a modular metal spacer to treat acute traumatic elbow instability. J Bone Joint Surg Am 2007;89:1075–1080. [DOI] [PubMed] [Google Scholar]
- 39. Wretenberg P, Ericson A, Stark A. Radial head prosthesis after fracture of radial head with associated elbow instability. Arch Orthop Trauma Surg 2006;126:145–149. [DOI] [PubMed] [Google Scholar]
- 40. Brinkman JM, Rahusen FT, de Vos MJ, Eygendaal D. Treatment of sequelae of radial head fractures with a bipolar radial head prosthesis: good outcome after 1–4 years follow-up in 11 patients. Acta Orthop 2005;76:867–872. [DOI] [PubMed] [Google Scholar]
- 41. Harrington IJ, Sekyi-Otu A, Barrington TW, Evans DC, Tuli V; V HIS-OABTEDT. The functional outcome with metallic radial head implants in the treatment of unstable elbow fractures: a long-term review. J Trauma 2001;50:46–52. [DOI] [PubMed] [Google Scholar]
- 42. Smets S, Govaers K, Jansen N, Van Riet R, Schaap M, Van Glabbeek F. The floating radial head prosthesis for comminuted radial head fractures: a multicentric study. Acta Orthop Belg 2000;66:353–358. [PubMed] [Google Scholar]
- 43. Knight DJ, Rymaszewski LA, Amis AA, Miller JH. Primary replacement of the fractured radial head with a metal prosthesis. J Bone Joint Surg Br 1993;75:572–576. [DOI] [PubMed] [Google Scholar]
- 44. van Riet RP, Sanchez-Sotelo J, Morrey BF. Failure of metal radial head replacement. J Bone Joint Surg Br 2010;92:661–667. [DOI] [PubMed] [Google Scholar]
- 45. Pilliar RM, Lee JM, Maniatopoulos C. Observations on the effect of movement on bone ingrowth into porous-surfaced implants. Clin Orthop Relat Res 1986;208:108–113. [PubMed] [Google Scholar]
- 46. O’Driscoll SW, Herald JA. Forearm pain associated with loose radial head prostheses. J Shoulder Elbow Surg 2012;21:92–97. [DOI] [PubMed] [Google Scholar]
- 47. King GJ, Faber KJ. Posttraumatic elbow stiffness. Orthop Clin North Am 2000;31:129–143. [DOI] [PubMed] [Google Scholar]
- 48. Mellema JJ, Lindenhovius AL, Jupiter JB. The posttraumatic stiff elbow: an update. Curr Rev Musculoskelet Med 2016;9:190–198. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49. Veltman ES, Doornberg JN, Eygendaal D, van den Bekerom MP. Static progressive versus dynamic splinting for posttraumatic elbow stiffness: a systematic review of 232 patients. Arch Orthop Trauma Surg 2015;135:613–617. [DOI] [PubMed] [Google Scholar]
- 50. Ha AS, Petscavage JM, Chew FS. Radial head arthroplasty: a radiologic outcome study. AJR Am J Roentgenol 2012;199:1078–1082. [DOI] [PubMed] [Google Scholar]