. 2019 May 21;24(5):056004. doi: 10.1117/1.JBO.24.5.056004

Photoacoustic properties of anterior ocular tissues

Kelsey P Kubelick a, Eric J Snider a, C Ross Ethier a,*, Stanislav Emelianov a,b,*
PMCID: PMC6992976  PMID: 31115200

Abstract.

Clinical imaging techniques for the anterior segment of the eye provide excellent anatomical information, but molecular imaging techniques are lacking. Molecular photoacoustic imaging is one option to address this need, but implementation requires use of contrast agents to distinguish molecular targets from background photoacoustic signals. Contrast agents are typically selected based on a priori knowledge of photoacoustic properties of tissues. However, photoacoustic properties of anterior ocular tissues have not been studied yet. Herein, anterior segment anatomy and corresponding photoacoustic signals were analyzed in brown and blue porcine eyes ex vivo. Measured photoacoustic spectra were compared to known optical absorption spectra of endogenous chromophores. In general, experimentally measured photoacoustic spectra matched expectations based on absorption spectra of endogenous chromophores reported in the literature, and similar photoacoustic spectra were observed in blue and brown porcine eyes. However, unique light–tissue interactions at the iris modified photoacoustic signals from melanin. Finally, we demonstrated how the measured PA spectra established herein can be used for one application of molecular PA imaging, detecting photoacoustically labeled stem cells in the anterior segment for glaucoma treatment.

Keywords: ophthalmology, anterior segment, ultrasound imaging, photoacoustic imaging, spectroscopic imaging, stem cell tracking

1. Introduction

The most-investigated approaches for noninvasive assessment of internal ocular structures have been ultrasound (US) and optical imaging techniques.19 Clinical use of ultrasound biomicroscopy and optical coherence tomography (OCT) are well established in ophthalmology.1,2,10 OCT provides excellent anatomical information by measuring changes in backscattered light.6 Advances in OCT now allow real-time imaging, wide-field imaging, higher resolution, and assessment of other tissue optical properties, such as birefringence.1014 Overall OCT research has drastically improved the quality of anatomical imaging in ophthalmology. However, molecular imaging techniques are lacking. Accordingly, development of molecular contrast OCT seeks to address this need1517 by manipulating backscattering properties to distinguish molecular targets.17 One approach is to introduce exogenous contrast agents that increase backscattering. However, high background signals from surrounding tissue still make it difficult to distinguish molecular targets. As a result, researchers have invented other clever solutions, including magnetomotive OCT and US, spectroscopic-OCT, pump-probe OCT, and photothermal OCT.1719

Significant effort has recently been dedicated to development of photothermal OCT for the retina.2023 In photothermal OCT, an optical absorber is irradiated with light. Heat deposition causes thermoelastic expansion of the surrounding tissue and ultimately changes the optical path length measured by OCT.20 Thus, photothermal OCT obtains molecular information by indirectly measuring changes in backscattering resulting from an absorption event. A more straightforward approach may be to use an imaging modality that directly measures optical absorption, such as photoacoustic (PA) imaging.

PA imaging is similar to photothermal OCT in that irradiation of an optical absorber results in heat deposition and thermoelastic expansion of surrounding tissues. However, conditions used in PA imaging produce transient acoustic waves detected by a traditional ultrasound transducer.2426 Thus, PA imaging combines advantages of optical and acoustic modalities to allow high resolution, high contrast, real-time, molecular imaging. A variety of endogenous and exogenous optical absorbers produce PA signals.27 These optical absorbers serve as PA contrast agents and are detected using multiwavelength spectroscopic PA imaging. Because PA contrast is based on absorption, molecular information complements anatomical information provided by scattering-based modalities such as OCT or ultrasound. Thus, ultrasound and photoacoustic (US/PA) imaging systems are frequently combined, which is advantageous due to the established use of ultrasound in clinical ophthalmology.5,9,28 Similarly, a multimodal PA/OCT system has also been demonstrated for molecular imaging at the posterior segment to assess the retina.29,30

Our work focuses on extending molecular PA imaging to the anterior segment of the eye, i.e., tissues anterior to the vitreous humor. Anterior segment imaging techniques, such as OCT, slit-lamp microscopy, and ultrasound biomicroscopy, primarily provide anatomical information.10 There is a need for molecular imaging techniques that could aid diagnosis, guide therapies, and improve understanding of anterior ocular diseases, including glaucoma, uveitis, and iridial pathologies.17 Implementation of molecular PA imaging requires endogenous or exogenous contrast agents that must be carefully selected based on a priori knowledge of PA properties of anterior segment tissues to ensure the contrast agent or interventional devices can be distinguished from background PA signals.3133 Although PA imaging has been demonstrated at a select few wavelengths in the anterior segment,3436 exploration of PA properties over a broad range of wavelengths is required to inform contrast agent selection for molecular PA imaging.

Thus, the goal of the work presented here was to establish PA spectral properties for the anterior segment over a range of wavelengths corresponding to the ideal tissue optical window.27 In some cases, PA spectra in biological tissues directly correspond to optical absorption spectra of endogenous chromophores, i.e., melanin, oxygenated, and deoxygenated hemoglobin. Alternatively, PA spectra can be modified by light–tissue interactions.37 We hypothesized that the latter scenario may occur in the anterior segment due to the role of anterior tissues to control light propagation. Differences in eye color, resulting from unique combinations of scattering and melanin absorption at the iris, may also modify PA spectra.38,39 For these reasons, we systematically measured the PA spectral properties of anterior segment structures in blue and brown porcine eyes, which are a relevant model of human anatomy and pigmentation. We also demonstrated how these PA spectra can aid development of molecular PA imaging strategies in the anterior eye, specifically by informing contrast agent selection to detect stem cells for glaucoma treatment.4042

2. Materials and Methods

2.1. Eye Preparation

All experiments used enucleated porcine eyes (Holifield Farms, Covington, Georgia). Blue and brown porcine eyes were studied to represent the two extremes in the range of eye color from light to dark. Blue or brown eye color was determined by visual inspection, and no other colors were observed in samples from the slaughterhouse. Fresh and fixed eye globes were imaged. Eye globes were fixed by submersion in 10% neutral buffered formalin overnight before imaging, then maintained in phosphate-buffered saline (PBS) prior to use. In total, 10 brown (three fresh and seven fixed) and 6 blue porcine eyes (two fresh and four fixed) were imaged. Minimal differences between PA spectra of fresh and fixed samples were observed (data not shown).

2.2. Imaging Protocol

A Vevo 2100/LAZR (VisualSonics, Inc., Toronto, Canada) imaging system, incorporating US/PA imaging modes, was used for all imaging experiments. US/PA images were acquired using a 256-element transducer array operating at 20-MHz (LZ250) or 40-MHz (LZ550) center frequencies. The laser source was a Q-switched Nd:YAG laser-pumped optical parametric oscillator system (pulse repetition frequency=20  Hz, 7-ns pulse duration). PA images were acquired within the 680- to 970-nm wavelength range at a frame rate of 5 Hz. Variations in laser pulse energy were corrected at the time of data acquisition using built in features of the Vevo 2100/LAZR.

In the first set of brown eye data, US/PA images were acquired at 1 deg increments as the eye was rotated 360 deg. Data were collected using both the 20- and 40-MHz transducers, and the wavelength was varied from 680 to 970 nm in increments of 10 nm. The transducer was oriented either on-axis or perpendicular to the trabecular meshwork for translimbal light delivery for a total of four datasets. Preliminary analysis confirmed ultrasound-determined anatomy. PA spectra were similar in all views, and data acquisition at every degree was unnecessary. Adjustments were thus made to permit acquisition of more wavelengths and compile more thorough spectroscopic information. Specifically, US/PA images were acquired every 45 deg at wavelengths from 680 to 970 nm in 5-nm increments. US/PA images were postprocessed in MATLAB to determine the PA spectrum at anatomical landmarks.

2.3. Calculating and Analyzing Measured PA Spectra

The average PA spectra for the anterior and posterior iris, trabecular meshwork, and sclera were determined to establish PA spectral properties specific to anatomy of the anterior segment. Each spectrum was determined by calculating the average PA signal over each region of interest. Each region of interest was isolated using spatial masks determined by manual segmentation based on anatomical information depicted in US/PA images. Variations in mask size and shape had minimal impact on PA spectra; specifically, the magnitude of the variation in the PA spectra between masks was comparable to the variations in the PA spectra between eyes and was thus not considered significant. The spectrum for each region of interest was then low-pass filtered using a 10-point moving average. The resulting spectra are henceforth referred to as “measured PA spectra.” Measured PA spectra were compared to established optical absorption spectra for deoxygenated hemoglobin (Hb), oxygenated hemoglobin (HbO2), and melanin.4349

2.4. PA Analysis in Phantom Experiments of Pigment and Scattering

The measured PA spectrum for the posterior iris required further analysis to evaluate effects of optical scattering. Following methods described elsewhere, melanin was isolated from dissected irises of fresh brown porcine eyes by collecting tissue in 10 mL of water and mechanically disrupting to release pigment.50,51 Pigment samples were size separated via centrifugation and isolated particles were resuspended in 2 mL of PBS for US/PA imaging.

A 10×10  cm plastic container held isolated pigment samples for US/PA imaging. Two small holes were drilled on opposite sides of the container and a silicone tube (Helix Medical) with an inner diameter of 0.062 in. and an outer diameter of 0.095 in. was fed through each hole and secured. The thickness of the iris was 1  mm based on anatomical ultrasound images, slightly smaller than the inner diameter of the silicone tube. The tube was loaded with the following samples for US/PA imaging: PBS, iris pigment alone, and the same iris pigment with a piece of sclera wrapped around the tube. In the PA images of the circular cross-section of the tube, three regions of interest (top, middle, and bottom of the tube) were identified. The average PA signal was determined for each region of interest to assess changes in the PA spectrum of pigment due to light attenuation as a function of depth.

2.5. Analysis of Local Tissue Composition Using Spectroscopic PA Processing

The PA signal at each pixel, i.e., pressure P(x,y,λ), is expressed as described elsewhere:25,5255

P(x,y,λ)=ΓF(x,y,λ)i=1nμai(x,y,λ), (1)

where λ is the wavelength, μa(λ) is the optical absorption coefficient, F(x,y,λ) is the fluence, and Γ is the Grüneisen parameter. For heterogeneous tissue, the total PA signal at each pixel is a superposition of each PA signal from n optical absorbers.

Spectroscopic PA datasets were preprocessed to correct for wavelength-dependent fluence attenuation in the iris. Fluence, F(z,λ), is expressed as37,56

F(z,λ)=Fo(λ)eμeff(λ)·z, (2)

where z is the imaging depth, and F0 is the fluence at the iris surface. Effective attenuation, μeff(λ)=3μa(λ)[μa(λ)+μs(λ)(1g(λ)], was calculated based on: (i) the coefficient of anisotropy, g(λ)=2.25·λ0.155,57 (ii) the optical scattering coefficient, μs(λ), of 10% intralipid,58 and (iii) μa(λ) of melanin at the retinal pigmented epithelium (RPE).44,45

Unfortunately, published iris-specific absorption and scattering values were not available for our wavelength range. The best alternatives to estimate μeff at the iris were to use generic scattering (intralipid) and melanin absorption in the RPE. For brown eyes, μa[λ(nm)]=8·1010λ3.48  (cm1), and for blue eyes, μa[λ(nm)]=6.5·1010λ3.48  (cm1).44,45 The exponent of λ from the expression for μa(λ) is the critical parameter to describe the melanin absorption spectrum and was taken directly from published data for the RPE.44 However, concentrations of melanin vary by orders of magnitude in ocular tissues, and thus the values of the leading coefficient in the expression μa(λ) were reduced to accommodate different melanin concentrations and distributions at the iris compared to the RPE.44 The leading coefficients were incrementally adjusted until artifacts at the posterior iris were minimized. However, processing was conservative and some artifacts were deliberately kept. Fluence correction is a complex technical issue and further work is required in the future.37,56 Using Eq. (2), F(z,λ) was calculated to produce a spatial distribution map of fluences in the iris. The PA signal, P, in Eq. (1) was multiplied by 1/F(z,λ) to correct for wavelength and depth-dependent fluence attenuation.37,56

After fluence correction, spectroscopic PA datasets were filtered using a 3×3  pixel (41  μm×45  μm) median filter to remove large noise spikes. Images were then low pass filtered using a 6×9×3  pixel (depth×width×wavelength) moving average filter, approximately corresponding to 252  μm×405  μm and a wavelength range of 15 nm. PA signals below the noise floor, defined by min(P)+0.025[max(P)min(P)], were eliminated.

After filtering, spectroscopic PA datasets were analyzed using the linear least squares (LLS) regression method. LLS regression creates a spatial distribution map of each optical absorber in tissue. Concentration, C, of each absorber was estimated by separating contributions of each absorber to the total PA signal at each pixel. According to LLS regression methods described elsewhere,25,59,60 the optical absorption coefficient, μa(λ), in Eq. (1) can be expressed as the product of the concentration of the absorber, Ci(x,y), and the optical cross-section of the absorber, σa. Equation (1) is rewritten as

P(x,y,λ)=ΓF(x,y,λ)i=1nCi(x,y)σai(λ). (3)

The optical cross-section describes the likelihood an absorption event will occur, and σai is a matrix whose rows and columns correspond to wavelength and the normalized absorption spectrum of each preselected optical absorber, respectively. For our studies, two matrices of optical cross-sections were determined a priori for LLS regression. The first matrix of optical cross-sections was obtained from the normalized absorption spectra of melanin, HbO2, and Hb, which are reported in the literature and represent optical properties of generic tissue.4349 The second matrix of optical cross-sections was obtained from our normalized measured PA spectra, which were specific to anterior ocular tissues. Thus, the same images were processed with LLS regression to produce two sets of concentration maps based on absorption spectra from the literature or based on our measured PA spectra.

Other research has established that it is theoretically ideal to use the maximum number of wavelengths for LLS regression of spectroscopic PA datasets.61,62 In practice, such an approach may introduce inaccuracies due to laser energy stability, noise, and overlap of optical spectra.61,62 Thus, a subset of acquired wavelengths, λ=720 to 970 nm, was used for LLS regression of spectroscopic PA datasets. This range captured a majority of wavelengths acquired, eliminated wavelengths with energy instability, and preserved distinct features of all spectra.

2.6. Stem Cell Tracking

As an example application, we detected human adipose-derived mesenchymal stem cells (MSCs, Lonza, Basel, Switzerland) in the anterior eye. The measured PA spectra established here were used to identify an appropriate contrast agent to label stem cells and detect them using molecular PA imaging. Based on the PA spectra of anterior ocular tissues, established herein, gold nanospheres (AuNSs) were a good contrast agent. AuNSs with a 20 nm diameter were synthesized in-house as described previously,63 sterilized under UV light for at least 12 h, and used to label MSCs following previous methods.6365 Briefly, MSCs were maintained in α-minimum essential medium containing 20% fetal bovine serum, penicillin, streptomycin, and l-glutamine. When MSCs reached 80% confluence, the media in the tissue culture flask was exchanged for AuNS-containing media. After 24 h of incubation, the AuNS-containing media was aspirated. MSCs were washed with PBS to remove residual AuNSs. The remaining AuNS-labeled MSCs were collected and suspended in PBS at a concentration of 1000  cells/μL for a 250-μL injection.

AuNS-labeled MSCs were injected through the cornea into the anterior chamber of a porcine eye ex vivo. The control eye received no injection. To mimic in vivo conditions using an organ perfusion system, all eyes were hydrostatically clamped to maintain physiological intraocular pressures, thereby inducing circulation throughout the anterior chamber due to the natural outflow of fluid through the trabecular meshwork. US/PA datasets were acquired before and 5 h after injection. LLS regression of spectroscopic photoacoustic (sPA) datasets from λ=680 to 710 nm was performed as described in Sec. 2.5. In this case, the matrix of optical cross-sections, described in Sec. 2.5, was modified to contain the spectrum of melanin and AuNS-labeled MSCs. The spectrum of AuNS-labeled MSCs was determined a priori in phantom experiments according to previous methods.66

3. Results

Brown porcine eyes were imaged ex vivo to evaluate PA response from constituent tissues. Ultrasound images resolved anatomical features in the anterior segment, such as the sclera, cornea, lens, and iris [Figs. 1(a) and 1(b)]. Histology aided proper identification of structures [Fig. 1(c)]. Heavy pigmentation was also noted at the posterior iris [Fig. 1(c)]. By ultrasound, the cornea appeared slightly darker compared to other tissues, consistent with its relative transparency, and the interface between the cornea and sclera was distinct, reflecting differences in composition and microstructural organization of the cornea relative to other tissues. As expected, other tissues, such as the trabecular meshwork (TM), could not be unequivocally identified due to similar scattering and ultrasound speckle patterns, confirming low ultrasound contrast between most tissue types.

Fig. 1.

Fig. 1

Visualization of anatomy and PA signals in the anterior segment. (a, b) Grayscale ultrasound images showing anatomy of the ex vivo porcine eye. S, sclera; AC, anterior chamber; C, cornea; L, lens; TM, trabecular meshwork; AI, anterior iris; PI, posterior iris; SCS, suprachoroidal space; CPE, ciliary pigmented epithelium. (c) Corresponding histologic image of the porcine anterior segment stained with hematoxylin and eosin. (d–g) Overlay of ultrasound (gray) and PA (red) images at different locations (scans at 0 deg, 45 deg, 90 deg, and 135 deg) at λ=700  nm. PA signals appeared at melanin-rich regions, such as the TM and throughout the iris. PA signals consistently appeared at the same anatomical landmark in all frames. Scale bar for ultrasound images=3  mm.

PA images were acquired at 700 nm wavelength and consecutive 45 deg cross-sections. PA signals were observed in melanin-rich tissues, namely the anterior and posterior iris and TM [Figs. 1(d)1(g)]. Although the ciliary pigmented epithelium (CPE) is melanin-rich, no PA signal was present in this tissue due to strong attenuation of light by the overlying iris and sclera, resulting in low fluence at the CPE. PA signal was observed in some views in the scleral region, possibly corresponding to the suprachoroidal space (SCS) [Figs. 1(e) and 1(g)]. Although the sclera and SCS are not pigmented, the PA signal at these locations may result from subsurface fluence.

Spectroscopic PA datasets were acquired in brown eyes at the anterior iris and posterior iris, scleral region, and TM from λ=680 to 970 nm. Representative PA spectra were plotted for fixed brown porcine eyes and distinct PA spectra were observed for each region [Fig. 2(a)]. To assess impact of fixation on the PA spectra, average PA spectra and standard deviations were calculated for multiple fresh brown porcine eyes [Figs. 2(b) and 2(c)]. PA spectra were similar between fresh and fixed eyes. The measured PA spectrum at the anterior iris closely followed the absorption spectrum of melanin reported in the literature [Fig. 2(b)]. Unexpectedly, the measured PA spectrum at the posterior iris did not follow the spectrum of melanin [Fig. 2(b)] and showed more similarities to the literature absorption spectrum of oxygenated hemoglobin (HbO2). Although HbO2 was not expected in ex vivo tissue, comparison with the measured PA spectrum at the posterior iris was important. Similarities between the absorption spectrum of HbO2 and the PA spectrum at the posterior iris [Fig. 2(c)] highlight a potential source of PA imaging artifacts and in turn impact use of these PA spectra to identify endogenous absorbers, select contrast agents, and facilitate molecular PA imaging.

Fig. 2.

Fig. 2

PA spectra in fixed and fresh brown porcine eyes. (a) Representative PA spectra for the scleral region (S), TM, anterior iris (AI), and posterior iris (PI) for fixed brown porcine eyes. Average PA spectra and standard deviations were also calculated for (b) the anterior iris and (c) posterior iris in multiple fresh brown porcine eyes. PA spectra appeared similar in fixed and fresh eyes. The measured PA spectrum (solid blue line) of the anterior iris was compared to the reported absorption spectrum44,45 of melanin (dashed blue line). (c) The measured PA spectrum (solid orange line) of the posterior iris was compared to the reported absorption spectrum4649 (dashed orange line) of HbO2 to highlight a potential source of PA imaging artifacts following LLS regression analysis of spectroscopic PA datasets.

We suspected the measured PA spectrum at the posterior iris actually represented a PA signal from melanin modified by wavelength-dependent fluence reaching the posterior iris. Therefore, fluence attenuation due to tissue absorption and scattering was studied in a silicone tube phantom. Cross-sectional US/PA images of the tube filled with purified pigment were acquired and divided into three regions of interest to study fluence attenuation and the effect on the PA spectrum of melanin at increasing depths [Fig. 3(a)]. At the top of the tube, the PA spectrum matched the literature absorption spectrum of melanin. As light was attenuated by overlying pigment, the spectrum of melanin shifted and a peak appeared at 850 nm at the bottom of the tube [Fig. 3(a)]. Additional scattering from scleral tissue further modified the PA spectrum of melanin [Fig. 3(b)]. Specifically, compared to attenuation primarily by optical absorption [Fig. 3(a)], the spectrum shifted at shallower depths, and a peak was observed in the middle of the tube [Fig. 3(b)]. The spectrum in the bottom region of interest further shifted with a peak closer to 900 nm. We suspected similar fluence attenuation occurred for light propagating from the anterior to posterior iris in situ. In other words, even though the measured PA spectrum at the posterior iris should match the literature absorption spectrum of melanin, the observed PA spectrum at the posterior iris [Fig. 2(c)], resulted from wavelength-dependent light–tissue interactions.

Fig. 3.

Fig. 3

PA spectrum of fresh pigment as a function of depth. A silicone tube was filled with purified pigment and cross-sectional US/PA images were acquired. The PA spectrum was analyzed within different regions of interest. “Top” and “bottom” indicate locations closest to, and further from, the transducer with the least and most light attenuation, respectively. (a) The PA spectrum of melanin was modified by depth due to light absorption of melanin, causing a significant change in the PA spectrum. (b) The experiment was repeated with the addition of a scattering layer (sclera), resulting in further modification of the measured PA spectrum of melanin.

The literature absorption spectra of melanin, HbO2, and Hb [Fig. 4(a)] were compared to the measured PA spectra [Fig. 4(e)] by processing spectroscopic PA datasets from brown eyes using the LLS regression method. The set of literature absorption spectra and measured PA spectra were used to obtain two different matrices of optical cross-sections. A spectrum for Hb could not be measured in brown eyes due to high melanin content; thus, the absorption spectrum of Hb from the literature was used instead [Fig. 4(e)]. The goal of this study was to compare each set of concentration maps, generated using the literature absorption spectra, serving as the “ground truth,” or our experimentally obtained measured PA spectra.

Fig. 4.

Fig. 4

sPA analysis using LLS regression in brown porcine eyes. (a) Absorption spectra from literature used for sPA analysis,4449 and resulting computed distributions of (b) melanin, (c) HbO2, and (d) Hb. (e–h) as in (a–d), but using measured PA spectra. Overlay of ultrasound (grayscale) and spectroscopic PA images showed localization of melanin (blue color scale), HbO2 (orange color scale), and Hb (pink color scale). Melanin was localized as expected. Absorption spectra from the literature or measured by PA imaging produced similar results. Some artifacts remained in the HbO2 and PI images (c and g). Scale bar=3  mm.

Using either set of spectra, melanin was localized as expected at the anterior iris, posterior iris, and TM [Figs. 4(b) and 4(f)]. Because HbO2 was not expected in post-mortem tissue, pixels indicating presence of HbO2 were determined to be image artifacts [Fig. 4(c)]. Agreement between the concentration maps for HbO2 [Fig. 4(c)] and the posterior iris [Fig. 4(g)] indicated the artifact occurred due to the measured spectrum at the posterior iris being most similar to that of HbO2 among the species we considered (melanin, Hb, and HbO2); however, the match is only very approximate [Fig. 2(c)]. Although the posterior iris is a melanin-rich tissue, the PA spectrum was modified by fluence attenuation. Without fluence correction, LLS regression predicted a prominent HbO2 distribution at the posterior iris (data not shown). By implementing fluence correction, melanin was accurately localized to the anterior iris [Figs. 4(b) and 4(f)] and the presence of HbO2 was minimized at the posterior iris [Figs. 4(c) and 4(g)]. Hb was primarily localized to the posterior iris, but some discrepancies were observed between distribution maps [Figs. 4(d) and 4(h)]. LLS regression of sPA datasets from multiple brown eyes produced similar results, indicating PA properties were similar across various colorations of brown eyes from light to dark. Agreement between both sets of concentration maps [Figs. 4(b)4(d) and 4(f)4(h)] impacts future use of measured PA spectra for molecular PA imaging. Similar concentration maps across multiple brown eyes indicated our measured PA spectra can be used to inform contrast agent selection for a particular application according to endogenous absorber distribution.

Characterization and analysis of PA properties was repeated in blue porcine eyes (Fig. 5). Each spectrum in brown and blue eyes was normalized according to its peak absorption to better visualize the PA properties in different eye colors. Results do not depict differences in magnitude of PA signals. The posterior iris was the only structure to consistently show a PA signal in the blue eye at the 700-nm wavelength [Fig. 5(a)]. Due to less melanin than in brown eyes, PA signals were not evident in other tissues. The PA signal from melanin at the posterior iris was modified by fluence attenuation, and thus a similar artifact was observed in the blue eye [Fig. 5(b)] as in brown eyes. Other types of PA spectra were also occasionally observed in blue eyes [Figs. 5(c) and 5(d)] but were not localized to a consistent location, i.e., the spectrum shown in Fig. 5(b) was most prevalent in blue eyes.

Fig. 5.

Fig. 5

Comparison of measured PA spectra in blue and brown eyes. (a) Overlay of ultrasound (gray) and phototacoustic (red) images in a blue porcine eye. PA signal was localized to the posterior iris. (b–d) Spectra were separated based on results from brown porcine eyes. (b) A modified melanin spectrum, resembling that of oxygenated hemoglobin, was isolated to the posterior iris in both eye colors and was the dominant spectrum observed in blue eyes. (c and d) Other types of spectra were observed in blue eyes, but were not localized to a consistent location. (d) The PA spectrum resembling melanin (b) was most prevalent in brown eyes. Scale bar=3  mm.

Spectroscopic PA analysis using the LLS regression method was repeated for blue eyes (Fig. 6). In the case of blue eyes, sPA analysis was conducted using the measured PA spectra from brown eye datasets, except for the spectrum of Hb, which was possible to isolate in a blue eye. Spectroscopic PA datasets from blue porcine eyes were similarly processed using either the literature absorption spectra [Fig. 6(a)] or the measured PA spectra [Fig. 6(e)] to identify absorbers. After correcting for wavelength-dependent fluence attenuation, melanin was accurately localized at the iris [Figs. 6(b) and 6(f)]. Some effects of fluence attenuation remained [Figs. 6(c) and 6(g)]. However, fluence correction increased melanin localization at the posterior iris [Figs. 6(b) and 6(f)] and minimized the presence of HbO2 [Fig. 6(c)] and the modified melanin spectrum at the posterior iris [Fig. 6(g)]. Hb was also identified [Figs. 6(d) and 6(h)]. sPA analysis in blue eyes accurately localized absorbers. In this case, agreement between concentration maps [Figs. 6(b)6(d) and 6(f)6(h)] indicated conservation of PA properties across drastically different eye colors.

Fig. 6.

Fig. 6

Spectroscopic PA (sPA) analysis using LLS regression in blue porcine eyes. (a) Literature absorption spectra used for sPA analysis.4449 (b–d) sPA analysis using the literature absorption spectra. (e) Measured PA spectra used for sPA analysis, which were compiled from brown eye datasets, except for the spectrum of Hb. (f–h) sPA analysis using the measured PA spectra. Overlay of ultrasound (grayscale) and spectroscopic PA images showed localization of melanin (blue colorscale), HbO2 (orange colorscale), and Hb (pink colorscale). Absorbers were localized as expected according to known anatomy. Literature absorption spectra or measured PA spectra produced similar results, and either can be applied for sPA analysis. These results show that the PA spectra can be used for sPA analysis in both brown and blue eyes. Scale bar=3  mm.

The PA properties of anterior segment tissues reported here can be used for a variety of applications to facilitate molecular PA imaging in the anterior eye. One such application is PA tracking of injected stem cells to regenerate the trabecular meshwork in glaucoma patients.40 In a preliminary study, MSCs were labeled with AuNSs, injected into the anterior chamber of a porcine eye, and allowed to circulate for 5 h. A separate eye, which was not injected with AuNS-labeled MSCs but otherwise followed the same experimental protocol, was used as a control. US/PA imaging and sPA analysis using LLS regression was conducted to separate signals from melanin (the primary endogenous absorber of interest in this case) and injected stem cells (Fig. 7). Prior to injection, melanin was accurately localized throughout the anterior segment [Fig. 7(a)] and no AuNS-labeled MSCs were detected [Fig. 7(b)]. This validates accuracy of the algorithm and confirms minimal overlap of the spectrum of the contrast agent with the spectrum of melanin. At 5 h postinjection, sPA analysis accurately localized melanin at the iris [Fig. 7(c)]. AuNS-labeled MSCs were detected lining the lens [Fig. 7(d)]. The measured PA spectra described here directly informed contrast agent selection to allow successful stem cell detection in an organ perfusion system that mimics in vivo conditions. Therefore, ex vivo analysis was applicable to a more clinically relevant organ perfusion system, and the measured PA spectra will continue to be used to improve upon this proof-of-concept stem cell tracking platform.

Fig. 7.

Fig. 7

Proof-of-concept demonstration of stem cell tracking in the anterior eye. Stem cells must be labeled with a contrast agent in culture, prior to injection, to allow detection with PA imaging. The measured PA spectra established here and concentration maps of endogenous absorbers were used to inform contrast agent selection. Accordingly, stem cells were labeled with AuNSs. To mimic in vivo conditions, enucleated brown porcine eyes were perfused to maintain physiological pressure and were perfused to maintain physiological pressure and were imaged before (a, b) and 5 h after (c, d) injection of AuNS-labeled stem cells. sPA analysis using LLS regression separated PA signals from (a, c) melanin and (b, d) AuNS-labeled stem cells. (b) Minor artifacts were observed at the sclera, but the location is outside of the anterior chamber and AuNS-labeled MSCs would not be present in this region. Scale bar=3  mm.

4. Discussion

Clinically available ophthalmic imaging modalities such as OCT and ultrasound provide excellent anatomical information for ocular tissues.9,10,67 However, development of molecular imaging techniques has not developed as rapidly, particularly for the anterior segment.16,17 PA imaging is one option to address the need for molecular imaging in the anterior eye. Implementation of molecular PA imaging requires use of endogenous or exogenous contrast agents.26,27 Contrast agents must be carefully selected based on the application of interest and PA properties of the surrounding tissues. Without prior knowledge of PA properties of anterior segment tissues, it is impossible to ensure a contrast agent can be distinguished from background PA signals. Thus, the first step to extend molecular PA imaging to the anterior eye is to establish PA spectral properties.

We here systematically analyzed PA properties of anterior segment tissues to produce a set of measured PA spectra in the first optical window between 680 and 970 nm to facilitate molecular PA imaging. Distinct PA spectra were observed at the TM, anterior iris, posterior iris, and scleral region in brown porcine eyes. Similar PA spectra were observed at the posterior iris in blue porcine eyes. These measured PA spectra were compared to literature absorption spectra by analyzing sPA datasets of blue and brown porcine eyes using LLS regression to produce concentration maps of each absorber. Good agreement was observed between absorber maps, which validates future use of our PA spectra to facilitate molecular PA imaging in vivo.

We hypothesized that each measured PA spectrum would match a known absorption spectrum of an endogenous absorber, specifically melanin. However, this was only true for the anterior iris in brown eyes. In other tissues, the PA spectrum was modified by wavelength- and depth-dependent fluence attenuation. These effects were most prominent at the posterior iris, especially in blue eyes. The impact of fluence attenuation on the PA spectrum at the iris was consistent with results from a simplified phantom experiment. Without fluence correction, sPA analysis using LLS regression contained many artifacts and mistakenly attributed PA signals to light absorption by HbO2, which coincidentally has a spectrum similar to the fluence-modified melanin spectrum. Wavelength- and depth-dependent fluence correction allowed better PA identification of melanin in the anterior segment and minimized artifacts. Further improvement to the fluence correction method is possible. Specifically, we assumed a uniform concentration of melanin, but concentration varies spatially throughout the iris. In addition, at the wavelengths of interest (680 to 970 nm), information on melanin absorption and scattering at the iris were unknown. Thus, optical properties of the iris were estimated based on available data for the RPE and a general model of scattering (intralipid) as substitutes.4345,57,58 This motivates further study so that parameters specific to the iris can be determined. We anticipate the effects of fluence attenuation observed here will impact molecular PA imaging strategies in the future.

Comparison between blue and brown porcine eyes indicated that the set of measured PA spectra were independent of eye color, at least in the near-infrared region. This result was fortuitous but not obvious a priori. Although blue and brown eyes have the same type of pigment, in the visible spectrum differences in melanin concentration and scattering produce variations in eye color. In spite of distinct optical properties in the visible regime, PA properties in the near-infrared regime were similar across eye colors. This result was important to confirm our measured PA spectra were applicable to both extremes in eye color (blue and brown). Thus, a single set of measured PA spectra can possibly inform molecular PA imaging strategies across the full spectrum of eye colors. Each eye color does not appear to require a unique set of spectra to predict endogenous absorber distribution and background PA signals to inform contrast agent selection. If this were the case, implementing molecular PA imaging in the anterior eye would likely be impractical.

Finally, we demonstrated how the measured PA spectra can be used for one application of molecular PA imaging, tracking stem cells in the anterior segment for glaucoma treatment.4042 In this application, stem cells were labeled with a contrast agent in culture, before injection, to demonstrate molecular PA imaging.63 Thus, prior knowledge of endogenous absorber distribution and background PA signals was required to ensure stem cells were labeled with an appropriate contrast agent for accurate detection.26,27 Fortunately, in this particular application, wavelength-dependent fluence attenuation observed at the posterior iris was irrelevant because stem cells will be localized more anteriorly. Based on our measured PA spectra and endogenous absorber concentration maps, AuNSs were a good option for the contrast agent. Results of this proof-of-concept study indicated stem cells labeled with AuNSs were successfully distinguished from background PA signals. Furthermore, the stem cell imaging study was conducted in a clinically relevant model system: a perfused porcine eye under physiological pressure, which closely mimics in vivo conditions. Use of organ culture systems is well established in ophthalmic research due to similarities between porcine, monkey, and human ocular anatomy, and organ culture systems provide a valuable intermediary step toward the development of in vivo imaging techniques.6870 The fact that the measured PA spectra established here allowed detection of stem cells in a perfusion setup further supports the impact of our results on future development of molecular PA imaging techniques for the anterior eye.

Although AuNSs were a good choice for stem cell tracking, other applications may have different requirements for molecular PA imaging. For example, if a researcher is interested in studying iridial pathologies, contrast agent options will vary according to the unique background PA signals at the iris. Beyond contrast agent selection and molecular PA imaging, the PA spectra could also serve as a baseline to evaluate spectral changes related to tissue health. In the future, we will use the measured PA spectra described herein to evaluate other contrast agents for stem cell tracking, aid development of light delivery systems for the anterior eye, and investigate therapeutic use of light to develop image-guided therapy platforms.

5. Conclusions

PA tissue properties were analyzed throughout the anterior segment. Measured PA spectra were compiled based on distinct PA spectra observed in the anterior and posterior iris, TM, and sclera. These spectra were compared to existing published absorption spectra. sPA analysis using both the measured PA spectra and published absorption spectra accurately localized absorbers throughout the anterior segment. In addition, the measured PA spectra from brown eyes were used for sPA analysis of datasets from blue eyes. Although there are many eye colors in the visible range, our comparison of blue and brown porcine eyes indicated anterior segments are photoacoustically similar in the near-infrared wavelength range. Results have important implications for development of molecular PA imaging strategies and other light-based imaging techniques and therapies in the anterior segment, including our particular application of interest, stem cell tracking for glaucoma treatment.

Acknowledgments

The authors acknowledge support from the Georgia Research Alliance (GRA) and the National Science Foundation (NSF).

Biographies

Kelsey P. Kubelick is a PhD candidate in the Coulter Department of Biomedical Engineering at the Georgia Institute of Technology and Emory University. Her research focuses on development of cell and particle tracking approaches using ultrasound, photoacoustic, and magnetic resonance imaging to expedite translation of novel therapies, particularly in the fields of regenerative medicine and immunology.

Eric J. Snider received his doctorate in biomedical engineering at Georgia Institute of Technology/Emory University, where his research took steps towards developing stem cell therapies for the glaucomatous trabecular meshwork. He is currently a postdoctoral fellow at the United States Army Institute of Surgical Research, focused on developing therapeutics and diagnostics for ocular trauma.

C. Ross Ethier holds the Lawrence L. Gellerstedt, Jr. chair in bioengineering and is a Georgia Research Alliance Eminent Scholar in the Coulter Department of Biomedical Engineering at Georgia Tech/Emory. His research is in the biomechanics of cells and whole organs, with specific emphasis on ocular biomechanics. He works on developing treatments for glaucoma, the second most common cause of blindness, and for SANS, a syndrome affecting astronauts that is a major NASA human health concern.

Stanislav Emelianov is a Joseph M. Pettit Endowed Chair, Georgia Research Alliance Eminent Scholar, and professor of electrical and computer engineering and biomedical engineering at the Georgia Institute of Technology and Emory University School of Medicine. Furthermore, he is director of the Ultrasound Imaging and Therapeutics Research Laboratory. Projects in his laboratory are focused on the discovery, development, and clinical translation of diagnostic imaging and therapeutic instrumentation, augmented with theranostic nanoagents.

Disclosures

None.

References

  • 1.Ramos J. L., Li Y., Huang D., “Clinical and research applications of anterior segment optical coherence tomography: a review,” Clin. Exp. Ophthalmol. 37(1), 81–89 (2009). 10.1111/ceo.2009.37.issue-1 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Zysk A. M., et al. , “Optical coherence tomography: a review of clinical development from bench to bedside,” J. Biomed. Opt. 12(5), 051403 (2007). 10.1117/1.2793736 [DOI] [PubMed] [Google Scholar]
  • 3.Bianciotto C., et al. , “Assessment of anterior segment tumors with ultrasound biomicroscopy versus anterior segment optical coherence tomography in 200 cases,” Ophthalmology 118(7), 1297–1302 (2011). 10.1016/j.ophtha.2010.11.011 [DOI] [PubMed] [Google Scholar]
  • 4.Dada T., et al. , “Comparison of anterior segment optical coherence tomography and ultrasound biomicroscopy for assessment of the anterior segment,” J. Cataract Refractive Surg. 33(5), 837–840 (2007). 10.1016/j.jcrs.2007.01.021 [DOI] [PubMed] [Google Scholar]
  • 5.Radhakrishnan S., et al. , “Comparison of optical coherence tomography and ultrasound biomicroscopy for detection of narrow anterior chamber angles,” Arch. Ophthalmol. 123(8), 1053–1059 (2005). 10.1001/archopht.123.8.1053 [DOI] [PubMed] [Google Scholar]
  • 6.Huang D., et al. , “Optical coherence tomography,” Science 254(5035), 1178–1181 (1991). 10.1126/science.1957169 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Hee M. R., et al. , “Optical coherence tomography of the human retina,” Arch. Ophthalmol. 113(3), 325–332 (1995). 10.1001/archopht.1995.01100030081025 [DOI] [PubMed] [Google Scholar]
  • 8.Swanson E. A., et al. , “In vivo retinal imaging by optical coherence tomography,” Opt. Lett. 18(21), 1864–1866 (1993). 10.1364/OL.18.001864 [DOI] [PubMed] [Google Scholar]
  • 9.Pavlin C. J., et al. , “Clinical use of ultrasound biomicroscopy,” Ophthalmology 98(3), 287–295 (1991). 10.1016/S0161-6420(91)32298-X [DOI] [PubMed] [Google Scholar]
  • 10.Konstantopoulos A., Hossain P., Anderson D. F., “Recent advances in ophthalmic anterior segment imaging: a new era for ophthalmic diagnosis?” Br. J. Ophthalmol. 91(4), 551–557 (2007). 10.1136/bjo.2006.103408 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Radhakrishnan S., et al. , “Real-time optical coherence tomography of the anterior segment at 1310 nm,” Arch. Ophthalmol. 119(8), 1179–1185 (2001). 10.1001/archopht.119.8.1179 [DOI] [PubMed] [Google Scholar]
  • 12.Cense B., et al. , “Thickness and birefringence of healthy retinal nerve fiber layer tissue measured with polarization-sensitive optical coherence tomography,” Invest. Opthalmol. Visual Sci. 45(8), 2606–2612 (2004). 10.1167/iovs.03-1160 [DOI] [PubMed] [Google Scholar]
  • 13.Grulkowski I., et al. , “Retinal, anterior segment and full eye imaging using ultrahigh speed swept source OCT with vertical-cavity surface emitting lasers,” Biomed. Opt. Express 3(11), 2733–2751 (2012). 10.1364/BOE.3.002733 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Potsaid B., et al. , “Ultrahigh speed 1050nm swept source/Fourier domain OCT retinal and anterior segment imaging at 100,000 to 400,000 axial scans per second,” Opt. Express 18(19), 20029–20048 (2010). 10.1364/OE.18.020029 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Diebold Y., Calonge M., “Applications of nanoparticles in ophthalmology,” Prog. Retinal Eye Res. 29(6), 596–609 (2010). 10.1016/j.preteyeres.2010.08.002 [DOI] [PubMed] [Google Scholar]
  • 16.Boppart S. A., et al. , “Optical probes and techniques for molecular contrast enhancement in coherence imaging,” J. Biomed. Opt. 10(4), 041208 (2005). 10.1117/1.2008974 [DOI] [PubMed] [Google Scholar]
  • 17.Oldenburg A. L., et al. , “Molecular optical coherence tomography contrast enhancement and imaging,” in Optical Coherence Tomography, Drexler W., Fujimoto J., Eds., pp. 1429–1454, Springer, Cham: (2015). [Google Scholar]
  • 18.Manapuram R. K., et al. , “In vivo estimation of elastic wave parameters using phase-stabilized swept source optical coherence elastography,” J. Biomed. Opt. 17(10), 100501 (2012). 10.1117/1.JBO.17.10.100501 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Mehrmohammadi M., et al. , “Pulsed magneto-motive ultrasound imaging using ultrasmall magnetic nanoprobes,” Mol. Imaging 10(2), 102–110 (2011). 10.2310/7290.2010.00037 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Lapierre-Landry M., et al. , “In vivo photothermal optical coherence tomography of endogenous and exogenous contrast agents in the eye,” Sci. Rep. 7(1), 9228 (2017). 10.1038/s41598-017-10050-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Lapierre-Landry M., et al. , “Imaging melanin distribution in the zebrafish retina using photothermal optical coherence tomography,” Transl. Vision Sci. Technol. 7(5), 4 (2018). 10.1167/tvst.7.5.4 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Lapierre-Landry M., et al. , “Photothermal optical coherence tomography of indocyanine green in ex vivo eyes,” Opt. Lett. 43(11), 2470–2473 (2018). 10.1364/OL.43.002470 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Lapierre-Landry M., et al. , “In vivo photothermal optical coherence tomography of gold nanorods in the mouse eye,” in Opt. Life Sci. Cong., OSA, Washington, D.C., p. BoM3A.2 (2017). [Google Scholar]
  • 24.Bell A. G., “LXVIII. Upon the production of sound by radiant energy,” London Edinburgh Dublin Philos. Mag. J. Sci. 11(71), 510–528 (1881). 10.1080/14786448108627053 [DOI] [PubMed] [Google Scholar]
  • 25.Xu M., Wang L. V., “Photoacoustic imaging in biomedicine,” Rev. Sci. Instrum. 77(4), 041101 (2006). 10.1063/1.2195024 [DOI] [Google Scholar]
  • 26.Emelianov S. Y., Li P.-C., O’Donnell M., “Photoacoustics for molecular imaging and therapy,” Phys. Today 62(5), 34–39 (2009). 10.1063/1.3141939 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Luke G. P., Yeager D., Emelianov S. Y., “Biomedical applications of photoacoustic imaging with exogenous contrast agents,” Ann. Biomed. Eng. 40(2), 422–437 (2012). 10.1007/s10439-011-0449-4 [DOI] [PubMed] [Google Scholar]
  • 28.Pavlin C. J., Harasiewicz K., Foster F. S., “Ultrasound biomicroscopy of anterior segment structures in normal and glaucomatous eyes,” Am. J. Ophthalmol. 113(4), 381–389 (1992). 10.1016/S0002-9394(14)76159-8 [DOI] [PubMed] [Google Scholar]
  • 29.Jiao S., et al. , “Photoacoustic ophthalmoscopy for in vivo retinal imaging,” Opt. Express 18(4), 3967–3972 (2010). 10.1364/OE.18.003967 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Song W., et al. , “Integrating photoacoustic ophthalmoscopy with scanning laser ophthalmoscopy, optical coherence tomography, and fluorescein angiography for a multimodal retinal imaging platform,” J. Biomed. Opt. 17(6), 061206 (2012). 10.1117/1.JBO.17.6.061206 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Bayer C. L., et al. , “Multiplex photoacoustic molecular imaging using targeted silica-coated gold nanorods,” Biomed. Opt. Express 2(7), 1828–1835 (2011). 10.1364/BOE.2.001828 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Homan K., et al. , “Silver nanosystems for photoacoustic imaging and image-guided therapy,” J. Biomed. Opt. 15(2), 021316 (2010). 10.1117/1.3365937 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Su J., et al. , “Photoacoustic imaging of clinical metal needles in tissue,” J. Biomed. Opt. 15(2), 021309 (2010). 10.1117/1.3368686 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Silverman R. H., et al. , “High-resolution photoacoustic imaging of ocular tissues,” Ultrasound Med. Biol. 36(5), 733–742 (2010). 10.1016/j.ultrasmedbio.2010.02.006 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Nam S. Y., Emelianov S. Y., “Array-based real-time ultrasound and photoacoustic ocular imaging,” J. Opt. Soc. Korea 18(2), 151–155 (2014). 10.3807/JOSK.2014.18.2.151 [DOI] [Google Scholar]
  • 36.de la Zerda A., et al. , “Photoacoustic ocular imaging,” Opt. Lett. 35(3), 270–272 (2010). 10.1364/OL.35.000270 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Maslov K., Zhang H. F., Wang L. V., “Effects of wavelength-dependent fluence attenuation on the noninvasive photoacoustic imaging of hemoglobin oxygen saturation in subcutaneous vasculature in vivo,” Inverse Probl. 23(6), S113–S122 (2007). 10.1088/0266-5611/23/6/S09 [DOI] [Google Scholar]
  • 38.Sturm R. A., Larsson M., “Genetics of human iris colour and patterns,” Pigment Cell Melanoma Res. 22(5), 544–562 (2009). 10.1111/pcr.2009.22.issue-5 [DOI] [PubMed] [Google Scholar]
  • 39.Imesch P. D., Wallow I. H., Albert D. M., “The color of the human eye: a review of morphologic correlates and of some conditions that affect iridial pigmentation,” Surv. Ophthalmol. 41, S117–S123 (1997). 10.1016/S0039-6257(97)80018-5 [DOI] [PubMed] [Google Scholar]
  • 40.Snider E. J., et al. , “Characterizing differences between MSCs and TM cells: toward autologous stem cell therapies for the glaucomatous trabecular meshwork,” J. Tissue Eng. Regener. Med. 12(3), 695–704 (2018). 10.1002/term.v12.3 [DOI] [PubMed] [Google Scholar]
  • 41.Manuguerra-Gagné R., et al. , “Transplantation of mesenchymal stem cells promotes tissue regeneration in a glaucoma model through laser-induced paracrine factor secretion and progenitor cell recruitment,” Stem Cells 31(6), 1136–1148 (2013). 10.1002/stem.v31.6 [DOI] [PubMed] [Google Scholar]
  • 42.Abu-Hassan D. W., et al. , “Induced pluripotent stem cells restore function in a human cell loss model of open-angle glaucoma,” Stem Cells 33(3), 751–761 (2015). 10.1002/stem.1885 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.Jacques S. L., “Optical properties of biological tissues: a review,” Phys. Med. Biol. 58(11), R37–R61 (2013). 10.1088/0031-9155/58/11/R37 [DOI] [PubMed] [Google Scholar]
  • 44.Jacques S. L., Glickman R. D., Schwartz J. A., “Internal absorption coefficient and threshold for pulsed laser disruption of melanosomes isolated from retinal pigment epithelium,” Proc. SPIE 2681, 468–477 (1996). 10.1117/12.239608 [DOI] [Google Scholar]
  • 45.Jacques S. L., McAuliffe D. J., “The melanosome: threshold temperature for explosive vaporization and internal absorption coefficient during pulsed laser irradiation,” Photochem. Photobiol. 53(6), 769–775 (1991). 10.1111/php.1991.53.issue-6 [DOI] [PubMed] [Google Scholar]
  • 46.Moaveni M. K., “A multiple scattering field theory applied to whole blood,” PhD Dissertation, Department of Electrical Engineering, University of Washington; (1970). [Google Scholar]
  • 47.Schmitt J. M., “Optical measurement of blood oxygenation by implantable telemetry,” Technical Report G5558-15, Stanford: (1986). [Google Scholar]
  • 48.Takatani S., Graham M. D., “Theoretical analysis of diffuse reflectance from a two-layer tissue model,” IEEE Trans. Biomed. Eng. BME-26(12), 656–664 (1979). 10.1109/TBME.1979.326455 [DOI] [PubMed] [Google Scholar]
  • 49.Prahl S., “Optical absorption of hemoglobin,” 1999, https://omlc.org/spectra/hemoglobin/index.html (11 July 2018).
  • 50.Porter K. M., Epstein D. L., Liton P. B., “Up-regulated expression of extracellular matrix remodeling genes in phagocytically challenged trabecular meshwork cells,” PLoS One 7(4), e34792 (2012). 10.1371/journal.pone.0034792 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51.Epstein D. L., et al. , “Experimental obstruction to aqueous outflow by pigment particles in living monkeys,” Invest. Ophthalmol. Visual Sci. 27(3), 387–395 (1986). [PubMed] [Google Scholar]
  • 52.Bowen T., et al. , “Some experimental results on the thermoacoustic imaging of tissue equivalent phantom materials,” in Ultrason. Symp., IEEE, pp. 823–827 (1981). 10.1109/ULTSYM.1981.197738 [DOI] [Google Scholar]
  • 53.Wang X., et al. , “Three-dimensional laser-induced photoacoustic tomography of mouse brain with the skin and skull intact,” Opt. Lett. 28(19), 1739–1741 (2003). 10.1364/OL.28.001739 [DOI] [PubMed] [Google Scholar]
  • 54.Kruger R. A., Liu P., “Photoacoustic ultrasound: pulse production and detection in 0.5% Liposyn,” Med. Phys. 21(7), 1179–1184 (1994). 10.1118/1.597399 [DOI] [PubMed] [Google Scholar]
  • 55.Oraevsky A. A., et al. , “Laser-based optoacoustic imaging in biological tissues,” Proc. SPIE 2134, 122–128 (1994). 10.1117/12.182927 [DOI] [Google Scholar]
  • 56.Cox B., et al. , “Quantitative spectroscopic photoacoustic imaging: a review,” J. Biomed. Opt. 17(6), 061202 (2012). 10.1117/1.JBO.17.6.061202 [DOI] [PubMed] [Google Scholar]
  • 57.Flock S. T., et al. , “Optical properties of intralipid: a phantom medium for light propagation studies,” Lasers Surg. Med. 12(5), 510–519 (1992). 10.1002/(ISSN)1096-9101 [DOI] [PubMed] [Google Scholar]
  • 58.van Staveren H. J., et al. , “Light scattering in lntralipid-10% in the wavelength range of 400–1100 nm,” Appl. Opt. 30(31), 4507–4514 (1991). 10.1364/AO.30.004507 [DOI] [PubMed] [Google Scholar]
  • 59.Wang X., et al. , “Noninvasive laser-induced photoacoustic tomography for structural and functional in vivo imaging of the brain,” Nat. Biotechnol. 21(7), 803–806 (2003). 10.1038/nbt839 [DOI] [PubMed] [Google Scholar]
  • 60.Wang L. V., Photoacoustic Imaging and Spectroscopy, CRC Press, Boca Raton, Florida: (2009). [Google Scholar]
  • 61.Luke G. P., Emelianov S. Y., “Optimization of in vivo spectroscopic photoacoustic imaging by smart optical wavelength selection,” Opt. Lett. 39(7), 2214–2217 (2014). 10.1364/OL.39.002214 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 62.Luke G. P., Nam S. Y., Emelianov S. Y., “Optical wavelength selection for improved spectroscopic photoacoustic imaging,” Photoacoustics 1(2), 36–42 (2013). 10.1016/j.pacs.2013.08.001 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 63.Nam S. Y., et al. , “In vivo ultrasound and photoacoustic monitoring of mesenchymal stem cells labeled with gold nanotracers,” PLoS One 7(5), e37267 (2012). 10.1371/journal.pone.0037267 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 64.Ricles L. M., et al. , “Function of mesenchymal stem cells following loading of gold nanotracers,” Int. J. Nanomed. 6, 407–416 (2011). 10.2147/IJN [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 65.Ricles L. M., et al. , “A dual gold nanoparticle system for mesenchymal stem cell tracking,” J. Mater. Chem. B 2(46), 8220–8230 (2014). 10.1039/C4TB00975D [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 66.Nam S. Y., et al. , “Nonlinear photoacoustic signal increase from endocytosis of gold nanoparticles,” Opt. Lett. 37(22), 4708–4710 (2012). 10.1364/OL.37.004708 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 67.Schuman J. S., et al. , Optical Coherence Tomography of Ocular Diseases, Slack, New Jersey: (2004). [Google Scholar]
  • 68.Bhattacharya S. K., et al. , “Cochlin expression in anterior segment organ culture models after TGFβ2 treatment,” Invest. Ophthalmol. Visual Sci. 50(2), 551–559 (2009). 10.1167/iovs.08-2632 [DOI] [PMC free article] [PubMed] [Google Scholar]