Diagnostic Performance of Three-Dimensional Endothelium/Descemet’s Membrane Complex Thickness Maps in Active Corneal Graft Rejection

Taher K Eleiwa 1,2, Jane C Cook 1, Amr S Elsawy 1,3, Vatookarn Roongpoovapatr 1, Vincent Volante 1, Sonia Yoo 1, Mohamed F Abou Shousha 1,3,4
PMCID: PMC7002262  NIHMSID: NIHMS1059596  PMID: 31678558

Abstract

Purpose:

To evaluate the performance of three-dimensional (3D) Endothelium/Descemet’s membrane complex thickness (En/DMT) maps, versus total corneal thickness (TCT) maps in the diagnosis of active corneal graft rejection.

Design:

Cross-sectional study.

Methods:

81 eyes (32 clear grafts and 17 with active rejection, along with 32 age-matched control eyes) were imaged using high-definition optical coherence tomography (HD-OCT), and a custom-built segmentation algorithm was used to generate 3D color-coded maps of TCT and En/DMT of the central 6-mm cornea. Regional En/DMT and TCT were analyzed and compared between the studied groups. Receiver operating characteristic curves were used to determine the accuracy of En/DMT and TCT maps in differentiating between studied groups.

Main Outcome Measures:

We evaluated regional En/DMT and TCT.

Results:

Both regional TCT and En/DMT were significantly greater in actively rejecting grafts compared to both healthy corneas and clear grafts (P<0.001). Using 3D thickness maps, central, paracentral and peripheral En/DMT achieved 100% sensitivity and 100% specificity in diagnosing actively rejecting grafts (optimal cut-off value, OCV, of 19 μm, 24 μm and 26 μm, respectively), versus only 82% sensitivity and 96% specificity for central TCT, OCV of 587 μm. Moreover, central, paracentral and peripheral En/DMT correlated significantly with graft rejection severity (r = 0.972, r = 0.729, and r = 0.823, respectively; P < 0.001).

Conclusion:

3D-En/DMT maps can diagnose active corneal graft rejection with excellent accuracy, sensitivity and specificity. Future longitudinal studies are required to evaluate the predictive and prognostic role of 3D-En/DMT maps in corneal graft rejection.

Keywords: Optical coherence tomography, Descemet’s membrane, corneal endothelium, graft rejection

Introduction

Corneal transplantation is the most common type of organ transplantation, with approximately 185,000 corneas transplanted annually in a survey of 116 countries, and 1 in 70 of the needs are covered worldwide.1 Corneal graft rejection is the leading cause of graft failure in the late postoperative period,2,3 and up to 68% of penetrating keratoplasties are affected with at least one episode of rejection.47 In full-thickness corneal transplants, graft failure rate secondary to a rejection ranges from 5% in low-risk grafts after five years to 35% in high-risk grafts at three years.8 These failures impose a heavy burden on the health care system and on patients’ quality of life.9 Hence, early detection of graft rejection is crucial to enhance corneal grafts survival and maintain patients’ productivity.10

Clinicians rely on slit-lamp findings of conjunctival injection, corneal edema, epithelial and endothelial rejection lines, sub-epithelial infiltrates, anterior chamber cells and flare, and Descemet folds to diagnose graft rejection,7,11 but these findings are only clinically apparent after irreversible loss of endothelial cells has already happened, which are critical for maintenance of corneal transparency.12 Pachymetry, in-vivo confocal microscopy, specular microscopy and anterior segment optical coherence tomography (AS-OCT) have been investigated to help diagnose a rejection earlier than clinical diagnosis.3,1318 However, these methods analyzed the changes only in the central area of corneal graft and could thus miss an early rejection in the peripheral parts of the corneal graft. In our previous work, we disclosed that manual measurement of central endothelial/Descemet membrane complex thickness (En/DMT) is a novel quantitative index that correlates accurately with the severity of rejection in both ex-vivo and in-vivo studies.14 Also, we have shown that central En/DMT has better diagnostic performance than endothelial cell density and central corneal thickness in characterizing the immunological status of corneal grafts.13,14 However, these measurements were two-dimensional (2D) and thus more susceptible to missing minor changes in the optical scan.

In this study, we compared the performance of three-dimensional (3D) Endothelium/Descemet’s membrane complex thickness (En/DMT) maps, versus total corneal thickness (TCT) maps in the diagnosis of active corneal graft rejection. A custom-built segmentation algorithm was used to generate 3D color-coded En/DMT maps from the captured HD-OCT images of the central 6-mm cornea. Compared with manual segmentation, this algorithm has been validated to be capable of segmenting all corneal layers of healthy eyes with similar accuracy, albeit with significantly better repeatability as well as significantly less running-time per image.19 It utilizes automated segmentation algorithms to generate 3D thickness maps of corneal layers.19,23 This quasi-histological visualization can enhance evaluation and diagnosis of various corneal pathologies such as keratoconus,24,25 Fuchs’ endothelial corneal dystrophy,26 and corneal graft rejection.27 Additionally, we analyzed central, paracentral and peripheral En/DMT parameters that we found highly sensitive and specific in diagnosing active corneal graft rejection. We also present data demonstrating that regional En/DMT is a much better indicator than central corneal thickness in assessing the immunological status of the corneal graft.

Materials and Methods:

Study population

This study was approved by the University of Miami Institutional Review Board. All participants provided written informed consent before enrollment. The study design followed the tenets of the Declaration of Helsinki for biomedical research.

Forty-nine eyes of 49 patients and 32 eyes of 32 normal controls were prospectively and consecutively recruited from December 2016 to January 2019 at Bascom Palmer Eye Institute, University of Miami. Inclusion criteria for participation included uneventful Penetrating keratoplasty (PK) or Descemet Stripping Automated Endothelial Keratoplasty (DSAEK) surgery performed greater than 1 month prior. Exclusion criteria included corneal grafts with microbial infection or past history of a rejection episode. For the un-operated control eyes, a best corrected visual acuity (BCVA) better than 20/25 on the Snellen scale, and no corneal abnormalities as detected by slit-lamp examination were required to be included in the study. Individuals with recent contact lens use, ocular diseases, previous ocular surgery, and systemic diseases with ocular involvement were excluded. Slit lamp examination was performed on each eye by a masked cornea specialist (either MA or SY) in order to assign the examined cornea into either a clear graft, or actively rejecting category. Active endothelial graft rejection was diagnosed by detecting new keratic precipitates (KPs) or a Khodadoust line in the presence of anterior chamber cells and new persistent corneal graft edema on two consecutive visits in a graft that was previously clear.7, 11 Clinical grading corneal grafts rejection based on the ejection severity was not attempted.

Image Acquisition and Analysis

Each participant received anterior segment HD-OCT (Envisu R2210, Bioptigen, Buffalo Grove, IL, USA) with 6 mm radial cuts centered on the corneal vertex. This device uses a super-luminescent diode light source with a central wavelength of 840 nm and it has an axial resolution of 3 μm, and a scanning speed of 32,000 A-scans per second with 36 frames per scan. Each participant was asked to look at a central fixation target and the presence of a visible specular reflection in all images confirmed optimal centration. In decentered grafts and post-PK eyes with high irregular astigmatism, the patient was asked to look at an optimized fixation target to maintain the geometric centration of the scan upon the graft. Custom-built segmentation software was used to segment the corneal epithelium, Descemet’s membrane and endothelium automatically with manual editing if needed (Figure 1). Manual editing was executed by two masked operators when the automatic segmentation failed to delineate the irregularities on the endothelial layer in the rejecting corneal grafts. The segmentation software used the Random Sample Consensus (RANSAC) method with a polynomial model to estimate the corneal layer boundaries from potential points obtained from thresholding the OCT image.42 Then, the estimates are refined by searching locally in the original image for the best points. The segmentation method is robust to the specular reflection as each boundary is fitted to a polynomial model which depends on all points and does not get affected that much with the specular reflection, as shown in Figure 1. Then, the segmentation of different frames of the scan is mapped into 3D and interpolated using bi-cubic interpolation to obtain the corneal surfaces. After that, 3D ray tracing is applied iteratively at each interpolated surface to correct for the refraction in the OCT imaging light by applying the vector form of Snell’s law at the refractive interface between each 2 successive layers (Figure 2-a).43 Finally, the inter-surface distances are measured as the shortest axial distance between each consecutive surface to generate the thickness maps. We used the refractive index of 1.376 for the corneal layers.44 Thickness maps were regionally divided into a central 2 mm circle, surrounded by 2 mm-wide paracentral, and 2 mm-wide peripheral concentric rings (Figure 2-b). The average central, paracentral and peripheral thickness parameters were calculated and used for further analysis. Validation studies of our automated image processing techniques were published in our previous work and the results were comparable to the manual operators.19

Figure 1:

Figure 1:

High-definition optical coherence tomography (HD-OCT) images of a healthy unoperated cornea (First row, left), clear corneal graft (Second row, left) and an actively rejecting graft (Third row, left). Presets (Right column) show magnified images of the posterior part of the corresponding cornea on the left. The red dashed line represents the anterior boundary of the corneal epithelium (Epi), the green dashed line indicates the corneal endothelium (En), and the blue dashed line for Descemet’s membrane (DM). In the presets, the thick arrows indicate the segmentation of the endothelium/Descemet’s membrane complex (En/DM). Endothelium/Descemet’s membrane complex thickness (En/DMT) is measured as the inter-surface distance between the segmented En and DM layers. The specular reflection (SR) confirmed optimal centration, however, the segmentation method is robust, as each boundary is fitted to a polynomial model, and thus does not get affected that much. In both healthy corneas and clear grafts, corneal En and DM were visualized as a band formed by 2 smooth hyper-reflective lines with a translucent space in between (First and second rows, right). In actively rejecting grafts (third row), En/DM appeared as a thickened band bounded by 2 hyper-reflective lines. The anterior line and the translucent space in between the lines were like those of the healthy corneas and clear grafts, while the posterior line had a broader hyper-reflectivity (third row, right) and occasional nodular excrescences (third row, right, yellow arrows). Bars are 100 μm.

Figure 2-a:

Figure 2-a:

A diagram showing the concept of thickness measurement using three-dimensional ray tracing. The axial distances between the uncorrected surfaces (red arrows) represent the optical path length and it is converted to geometric distance (blue arrows) using the layer refractive index. Three-dimensional ray tracing is applied iteratively at each surface to correct for the refraction in the OCT imaging light by applying the vector form of Snell’s law at the refractive interface between each 2 successive layers. Then, the thickness is measured as the shortest distance (black arrows) between each consecutive surfaces.

Figure 2-b:

Figure 2-b:

The scheme of arrangement of the corneal regions for quantitative evaluation of the layer thickness: Central region (C1, C2) lies within a 1 mm radius, surrounded by 2 concentric paracentral (M1, M2, M3, M4, M5, M6) and outer (O1, O2, O3, O4, O5, O6) rings, each with a 2 mm width. N: Nasal; S: Superior; T: Temporal; I: Inferior.

Diagnostic indices

We used three diagnostic indices to describe the regional microstructural characteristics of the En/DM complex; central, paracentral and peripheral En/DMT. We previously reported the 2 interfaces of the En/DM as the two most posterior hyper-reflective bands of the cornea on HD-OCT images.14,26

Statistical Analysis

Statistical analyses were performed using SPSS software version 22.0 (SPSS, Chicago, IL, USA) to calculate descriptive statistics for all eyes. The obtained TCT measurements were verified to have a normal distribution by assessment of histograms and a Shapiro-Wilk’s test of normality, while En/DMT was not. Therefore, mean ± standard deviations (SD) was used to characterize the distribution of the corneal thickness values, while the median values were used to characterize both En/DMT. In addition, one-way analysis of variance (ANOVA) with post-hoc comparisons were performed to account for the differences in central, paracentral, and peripheral corneal thickness, while a Kruskal-Wallis test with pairwise comparisons was used for regional En/DMT. Additionally, factorial-ANOVA was performed to verify if the changes in the corneal thickness and En/DMT depend on the type of graft (Penetrating Keratoplasty; PK versus Descemet Stripping Automated Endothelial Keratoplasty; DSAEK). The sensitivity and specificity of regional En/DMT and regional TCT in differentiating between studied groups were determined by generating receiver operating characteristic curves (ROC). In order to determine if central, paracentral and peripheral En/DMT would be descriptive of graft rejection severity, coefficient of correlation (r-value) of those indices and rejection severity based on central TCT was computed.1416,41 Two-sided p-values less than 0.05 were considered to be statistically significant. Moreover, intraclass correlation coefficients (ICC) were used to assess the inter-operator reliability of the manual measurements in the selected eyes. The ICC is defined as the ratio of the between-subjects variance to the sum of the pooled within-subjects variance and the between-subjects variance.20 The ICC interpretation that was used considered the reliability of the values as poor for values less than 0.2, fair for values from 0.21 to 0.40, moderate for values between 0.41 and 0.60, good for values from 0.61 to 0.80, and excellent for values higher than 0.80.21,22

Results:

Our study included 49 eyes of 49 patients; the breakdown included 32 clear grafts (21 PK and 11 DSAEK), 17 actively rejecting grafts (13 PK and 4 DSAEK), and 32 eyes of 32 age- and gender-matched healthy controls. Table 1 summarizes the different characteristics of all groups. The type of graft (PK versus DSAEK) had no statistically significant effect on the changes in the mean regional corneal thickness and En/DMT values between the studied groups (Figure 3).

Table 1:

Characteristics of our study groups

Healthy control Clear grafts Actively rejecting grafts p value
Number of eyes Unoperated cornea 32 —— —— ——
PK —— 21 13
DSAEK —— 11 4
Gender Female 14 13 10 p = 0.06**
Male 18 19 7
Age (years) 46±18 56±18 59±22 p = 0.06**
Postoperative time (Months) —— 8±5 10±4 p = 0.11**
Central corneal thickness (μm) 516±28 518±39 697±119 p <0.001**
Paracentral corneal thickness (μm) 527±27 552±41 766±149 p <0.001**
Peripheral corneal thickness (μm) 543±29 569±37 744±119 p <0.001**
Central DMT (μm)* 16 15 28 p <0.001***
Paracentral DMT (μm)* 17 17 35 p <0.001***
Peripheral DMT (μm)* 18 19 38 p <0.001***

DMT: Endothelial/Descemet membrane complex (En/DM) thickness; PK: Penetrating keratoplasty; DSAEK: Descemet Stripping Automated Endothelial Keratoplasty Values are presented as means ± standard deviation.

DMT is presented as median values.

**

p value is generated using one-way ANOVA test.

***

p value is generated using Kruskal-Wallis test.

Figure 3:

Figure 3:

Box-plot distributions showing that there was no statistically significant effect of the type of graft-Penetrating keratoplasty (PK) versus Descemet Stripping Automated Endothelial Keratoplasty (DSAEK)-on the changes in the mean central, paracentral and peripheral total corneal thickness (upper raw), and central, paracentral and peripheral endothelium/Descemet’s membrane complex (En/DM) thickness (lower raw) values between the studied groups.

Using HD-OCT images, the En/DM in graft rejection was visualized as two smooth hyper-reflective lines with a translucent space in between in both healthy corneas and clear grafts.13 In actively rejecting grafts, it appears as a thickened band formed by two hyper-reflective lines with occasional granular hyper-reflectivity (Figure 1). The segmentation software was robust in delineating these granules automatically in most of the eyes except in three cases that needed manual segmentation. Using 3D thickness maps, we were able to quantify both the regional differences in TCT and En/DMT (Figure 4).

Figure 4:

Figure 4:

Graph showing the discrimination between two actively rejecting corneal grafts (case 1 and case 2), a clear corneal graft (case 3) and healthy unoperated cornea (case 4) using color-coded and bulls-eye three-dimensional thickness maps of the total corneal thickness (TCT, 1st and 2nd raws) and endothelial/Descemet’s membrane complex thickness (En/DMT, 3rd and 4th raws). En/DMT was higher in case 1 and 2 compared to case 3 and 4. Interestingly, case 2 showed an active rejection in a thin corneal graft with high En/DMT compared to a healthy thicker graft (case 3) with normal En/DMT. Note the striking difference between the presented cases using the self-explanatory color-coded En/DMT maps.

Regional analysis of mean central, paracentral, and peripheral En/DMT as well as TCT revealed no statistically significant differences between healthy controls and clear grafts (Table 2). However, there was a statistically significant increase in all parameters in the actively rejecting group compared to healthy eyes and clear grafts (P<0.001). With regards to the evaluation of the inter-operator reliability for the manual measurements, the ICC for TCT was 0.999 (95% Confidence Interval; CI, 0.997–1.00) and the ICC for En/DMT was 0.962 (95% CI: 0.842–0.991).

Table 2:

Pairwise comparisons of central, paracentral and peripheral Endothelial/Descemet thickness (En/DMT), and total corneal thickness (TCT) means in studied groups.

Dependent Variable (I) Studied groups (J) Studied groups Mean Difference (I-J), μm Standard Error P value 95% Confidence Interval
Upper Bound Lower Bound
Central corneal thickness control Clear graft −2 15 0.993 −38 35
Active rejection −181* 18 0.000 −225 −136
Clear graft Active rejection −179 * 18 0.000 −223 −134
Paracentral corneal thickness control Clear graft −25 18 0.376 −69 19
Active rejection −239 * 22 0.000 −292 −186
Clear graft Active rejection −214 * 22 0.000 −267 −161
Peripheral corneal thickness control Clear graft −25 15 0.222 −62 10
Active rejection −201 * 18 0.000 −245 −157
Clear graft Active rejection −175 * 18 0.000 −219 −131
Central En/DMT control clear graft 1 0.6 0.223 −0.5 2
Active rejection −12* 0.7 0.000 −14 −10
Clear graft Active brejection −13 * 0.7 0.000 −15 −11
Paracentral En/DMT control Clear graft −1 0.5 0.249 −2 0.4
Active rejection −18 * 0.6 0.000 −19 −16
Clear graft Active rejection −17 * 0.6 0.000 −18 −15
Peripheral En/DMT control Clear graft −2 0.7 0.056 −3 −0.9
Active rejection −21* 0.9 0.000 −23 −19
Clear graft Active rejection −20* 0.9 0.000 −21.6752 −17.4167
*.

The mean difference is significant at the 0.05 level.

Table 3 summarizes the diagnostic accuracy of regional TCT, versus En/DMT. Central, paracentral and peripheral En/DMT achieved 100% sensitivity and 100% specificity (optimal cut-off value (OCV) of 19 μm, 24 μm, and 26 μm, respectively), whereas central graft thickness achieved only 88% sensitivity and 90% specificity (OCV of 564 μm) in differentiating actively rejecting grafts from clear grafts (Figure 5).

Table 3:

Receiver operating characteristic (ROC) curve data which represent the diagnostic performance of regional En/DMT, and regional total corneal thickness (TCT) in diagnosing actively rejecting corneal grafts.

Central TCT Paracentral TCT Peripheral TCT Central En/DMT Paracentral En/DMT Peripheral En/DMT
P value <0.001 <0.001 <0.001 <0.001 <0.001 <0.001
AUC 0.922 0.926 0.926 1.0 1.0 1.0
Sensitivity 88% 88% 88% 100% 100% 100%
Specificity 90% 90% 84% 100% 100% 100%
Cutoff value 564 μm 583 μm 592 μm 19 μm 24 μm 26 μm

AUC = area under the curve; En/DMT= endothelial/Descemet thickness; TCT= total corneal thickness. Specificity, sensitivity, and cutoff values are chosen to maximize total diagnostic accuracy (minimize total number of errors).

Figure 5:

Figure 5:

Combined receiver operating characteristics (ROC) graphs of central, paracentral and peripheral corneal thickness; regional endothelial/Descemet’s membrane complex thickness (En/DMT) in differentiating clear corneas from actively rejecting grafts. Regional En/DMT provided excellent discrimination (area under the curve, AUC, 1), whereas the central, paracentral and peripheral corneal thickness achieved an AUC of 0.922, 0.926 and 0.926, respectively.

To test whether central, paracentral and peripheral En/DMT could be used to quantify the severity of graft rejection, we correlated each of them with central corneal thickness as an objective diagnostic index of graft rejection. Central, paracentral and peripheral En/DMTs did not show significant correlations with central corneal thickness in controls and clear grafts. On the other hand, central, paracentral, and peripheral En/DMT of actively rejecting grafts showed a significant linear correlation with central corneal thickness (r = 0.972, P < 0.001; r = 0.729, P < 0.001; and r = 0.823; P<0.001, respectively; figure 6).

Figure 6:

Figure 6:

Scatter plots showing a significant linear correlation between central corneal thickness and regional endothelial/Descemet’s membrane complex thickness (En/DMT) values in actively rejecting grafts.

Discussion:

Corneal graft rejection can start as a localized rejection before emanating throughout the entire cornea, hence regional analysis of corneal grafts would provide more substantial results.28 Corneal surgeons depend on clinical signs of limbal injection, aqueous cells, keratic precipitates in the graft, and segmental or diffuse edema of the graft to diagnose a rejection;7,11 however, these findings coincide with permanent endothelial cell injury.12 This fact highlights the importance of the early and accurate diagnosis of graft rejection to preserve corneal grafts and prolong their survival.2931 Corneal thickness has been used to predict rejection, as increasing central corneal thickness can be a sign of significant endothelial cell dysfunction; however, this is not sensitive for graft rejection.5 Decompensation of corneal grafts is usually present before central corneal thickness increases beyond 600 μm.32,33 The Cornea Donor Study showed that both central corneal thickness and endothelial cell density were predictive of graft failure from all causes, not only graft rejection.15

Previous studies had selected the central area of the corneal graft for consistent analysis1218, which could miss an ongoing rejection in the peripheral parts. Increased central corneal thickness measurements may offer an early warning of rejection, endothelial cell loss, inflammation, or other causes of endothelial cell dysfunction. However, central corneal thickness alone is not a reliable indicator of graft health or decompensation.6 The variability of normal corneal thickness between patients also presents a complication, because a mildly thickened graft undergoing rejection may still be within the range of normal.6, 15 For instance, acute graft rejection can present early with relatively normal corneal thickness (Figure 4, case 2); meanwhile, thicker grafts do not necessarily indicate active rejection (Figure 4, case 3).

Aqueous humor analysis has been reported to evaluate the increased cytokine levels during rejection of penetrating keratoplasty,34 but it is an invasive means of diagnosing rejection. Contact confocal microscopy has also been studied, and reports show that increased active keratocytes as well as immune cell densities in the central area can diagnose graft rejection via manual counting.18,35 Using specular microscopy, Musch et al. reported the significant decrease in the central endothelial cell density in severe graft rejection, whereas mild rejection episodes were not associated with a loss in cell density exceeding the expected.3 Using anterior segment OCT, Abou Shousha et al. reported a higher sensitivity and specificity of central two-dimensional En/DMT in evaluating the immunological status of the corneal graft, versus both central corneal thickness,10 and endothelial cell density.13

Our previous work only used manual measurements to take the 2D thickness measurements in the central region and we were not able to highlight peripheral localized changes that could be signs of an early rejection. In this study, our custom-built segmentation algorithm allows for non-contact in-vivo 3D objective quantification of En/DM complex from images captured using the commercially available anterior segment HD-OCT devices; thus, it is less susceptible to missing minor changes in the optical scan. Briefly, the algorithm works by automatically segmenting all the B-scans, remapping the segmentation into 3D points, interpolating the points to generate the 3D corneal surfaces using bicubic interpolation, correcting the surfaces from light refraction and lastly, measuring the inter-surface distances and subsequently generating the thickness maps.19 Elsawy A. et al had validated the automatic segmentation of corneal layers using this algorithm, compared to manual operators.19 Our study reveals excellent inter-operator reliability for the TCT and En/DMT measurements using the manual segmentation for severely distorted corneal layers. Our current data supports that imaging En/DMT has the highest sensitivity and specificity in diagnosing corneal graft rejection.

In our cross-sectional study, we report that the qualitative and quantitative regional analysis of the En/DM in corneal grafts using 3D color-coded thickness maps can accurately differentiate active rejection from healthy grafts; however, longitudinal studies showing some change in En/DM prior to the occurrence of actual active rejection is still required to validate that this diagnostic tool can detect rejections earlier than the current techniques. En/DM in actively rejecting grafts is characterized by diffuse or sectorial thickening of the posterior hyper-reflective layer with occasional nodular excrescences, giving it a saw-tooth appearance. These hyper-reflective granules may represent a tomographic visualization of the clinically detected KPs. However, the clinical and tomographic chronological changes in the En/DM of actively rejecting grafts were not examined due to the cross-sectional nature of our study. Hence, future longitudinal study is required to correlate the clinically detected KPs with these excrescences seen using OCT, and to evaluate the sensitivity of the 3D thickness maps to detect KPs earlier than slit-lamp examination. Quantitatively, the differences between the central, paracentral and peripheral En/DMT can diagnose acute rejection more accurately than corneal thickness measurements. The contemporaneous increase in the regional En/DMT in active graft rejection may potentially interact with endothelial function and subsequently play a part in the development of graft failure. Futher studies are required to evaluate the potential utility of these indices to remotely diagnose corneal graft rejection instead of in office exams. Furthermore, we investigated the objective relationship between the regional En/DMT and central TCT obtained from HD-OCT scans to test whether regional En/DMT could quantify the severity of graft rejection. We found the strongest positive correlation for the central En/DMT, followed by peripheral En/DMT and paracentral En/DMT. Prospective longitudinal study is required to identify the consecutive changes in the En/DMT of corneal allografts and to determine if that correlation could indicate that these changes occur before, and contribute to the edema in the rejecting corneal grafts.

Central En/DMT has been reported to be up to 18 μm in healthy subjects.13,14,26,36 Our results in healthy grafts and un-operated corneas match the reported measurements by Pekel et al.37, López de la Fuente et al.38, Alberto et al.39, and Hutchings et al40. Compared to the aforementioned studies, we demonstrated not only the central thickness of En/DM, but also the 3D mapping of the central 6 mm region in healthy corneas, clear grafts, and actively rejecting grafts using our automated segmentation algorithm.

Our study is not without limitations. First, imaging was limited to the central 6 mm of the cornea as the tele-centric probe has diminished axial resolution and signal intensity in peripheral regions. Nevertheless, in the context of keratoplasty, the graft is frequently 1–2.5 mm larger than central 6-mm zone. Second, we used 36 radial cuts on each eye to create a composite three-dimensional model of the cornea. Hence, a noncontiguous model was converted into a contiguous map, which may not account for minor gaps in measurement that may nonetheless be clinically significant. In addition, in severely distorted En/DM, manual editing of the automatically segmented endothelial boundary was necessitated to delineate the irregular excrescences in the En/DM complex. Moreover, the study was not correlating the clinically graded severity of graft rejection with regional En/DMT indices. However, we used the central TCT as an objective indicator of rejection severity.14,16,41 Lastly, the longitudinal changes in En/DMT with progressive corneal allograft rejection weren’t examined owing to the cross-sectional nature of this study. Thus, a future prospective longitudinal studies that uses an HD-OCT technology with wider scanning field and a more robust automatic segmentation algorithm are therefore warranted to address the aforementioned limitations and quantify long-term changes in the entire graft to more optimally explore the predictive and prognostic role of En/DMT in corneal graft rejection.

In conclusion, our study has disclosed, for the first time in the literature, the in-vivo 3D color-coded thickness maps of En/DM in corneal grafts. 3D-En/DMT maps have shown excellent sensitivity and specificity in detecting active corneal graft rejection that strongly correlate with the severity of graft rejection. Further studies are required to evaluate the potential utility of 3D-En/DMT maps in prediction and guidance of therapeutic decisions in corneal allograft rejection especially in high-risk corneal transplants.

Acknowledgments / Disclosure:

a. Conflicts of Interest and Source of Funding: Taher Eleiwa, Jane Cook, Amr Elsawy, Vincent Volante and Vatookarn Roongpoovapatr - None to declare. Sonia Yoo and Mohamed Abou Shousha: United States Non-Provisional Patents (Application No. 8992023 and 61809518), and PCT/US2018/013409. Patents and PCT are owned by University of Miami and licensed to Resolve Ophthalmics, LLC. Mohamed Abou Shousha and Sonia Yoo are equity holders and sit on the Board of Directors for Resolve Ophthalmics, LLC. Sonia H. Yoo is a consultant of CARL Zeiss.

b. Financial Support: This study was supported by a NEI K23 award (K23EY026118), NEI core center grant to the University of Miami (P30 EY014801), and Research to Prevent Blindness (RPB). The funding organization had no role in the design or conduct of this research.

c. Other Acknowledgments: None

d. Declarations:

1) Ethics approval and consent to participate: This study was approved by the University of Miami Institutional Review Board.

2) Availability of data and material: The datasets used and/or analyzed during the current study available from the corresponding author on reasonable request.

3) Authors’ contributions: All authors attest that they meet the current ICMJE criteria for Authorship. Each of the coauthors had seen and agreed with each of the changes made to this manuscript in this revision and to the way his or her name is listed.

List of Abbreviations:

En/DM

Endothelium-Descemet’s complex

En/DMT

Endothelium-Descemet’s complex thickness

TCT

total corneal thickness

HD-OCT

High-definition Optical Coherence Tomography

AS-OCT

Anterior segment Optical Coherence Tomography

ROC

Receiver Operating Characteristics

AUC

Area under curve

OCV

optimal cut-off value

PK

Penetrating keratoplasty

DSAEK

Descemet Stripping Automated Endothelial Keratoplasty

SD

standard deviation

ANOVA

analysis of variance

2D

two-dimensional

3D

three-dimensional

Epi

epithelium

KPs

keratic precipitates

References

  • 1.Gain P, Jullienne R, He Z et al. Global Survey of Corneal Transplantation and Eye Banking. JAMA Ophthalmol 2016; 134(2):167–173. [DOI] [PubMed] [Google Scholar]
  • 2.Wilson SE, Kaufman HE, Graft failure after penetrating keratoplasty. Surv Ophthalmol 1990; 34(5): 325–356. [DOI] [PubMed] [Google Scholar]
  • 3.Musch DC, Schwartz AE, Fitzgerald-Shelton K, Sugar A, Meyer RF. The effect of allograft rejection after penetrating keratoplasty on central endothelial cell density. 1991; 111(6): 739–742. [DOI] [PubMed] [Google Scholar]
  • 4.Alldredge OC, Krachmer JH. Clinical types of corneal transplant rejection. Their manifestations, frequency, preoperative correlates, and treatment. Arch Ophthalmol 1981; 99(4): 599–604. [DOI] [PubMed] [Google Scholar]
  • 5.Williams KA, Muehlberg SM, Lewis RF et al. How successful is corneal transplantation? A report from the Australian Corneal Graft Register. Eye 1995; 9 (2): 219–227. [DOI] [PubMed] [Google Scholar]
  • 6.Ing JJ, Ing HH, Nelson LR et al. Ten-year postoperative results of penetrating keratoplasty. Ophthalmology 1998; 105(10): 1855–1865. [DOI] [PubMed] [Google Scholar]
  • 7.Panda A, Vanathi M, Kumar A, Dash Y, Priya S. Corneal graft rejection. Surv Ophthalmol. 2007. Jul-Aug 52 (4):375–96 [DOI] [PubMed] [Google Scholar]
  • 8.Bartels MC, Doxiadis II, Colen TP et al. Long-term outcome in high-risk corneal transplantation and the influence of HLA-A and HLA-B matching. Cornea 2003; 22(6):552–6. [DOI] [PubMed] [Google Scholar]
  • 9.Wilson SE, Kaufman HE. Graft failure after penetrating keratoplasty. Surv Ophthalmol, 34 (5) (1990), pp. 325–356 [DOI] [PubMed] [Google Scholar]
  • 10.Claerhout, Beele H, De Bacquer D et al. Factors influencing the decline in endothelial cell density after corneal allograft rejection. Invest Ophthalmol Vis Sci 2003; 44 (11): 4747–6. [DOI] [PubMed] [Google Scholar]
  • 11.Dua HS, Azuara-Blanco A. Corneal allograft rejection: risk factors, diagnosis, prevention, and treatment. Indian J Ophthalmol. 1999, 47(1), 3. [PubMed] [Google Scholar]
  • 12.Lee HS, Kim MS. Influential factors on the survival of endothelial cells after penetrating keratoplasty. European journal of ophthalmology 2009, 19(6), 930–935. [DOI] [PubMed] [Google Scholar]
  • 13.Smith C, Kaitis D, Winegar J. Comparison of endothelial/Descemet’s membrane complex thickness with endothelial cell density for the diagnosis of corneal transplant rejection. Ther Adv Ophthalmol. 2018. December 3; 1–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Abou Shousha M, Yoo S, Sayed et al. In vivo characteristics of corneal endothelium/Descemet membrane complex for the diagnosis of corneal graft rejection. Am J Ophthal 2017; 178: 27–37. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Verdier DD, Sugar A, Baratz K et al. Corneal thickness as a predictor of corneal transplant outcome. Cornea 2013. 32(6): 729–36. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Liu YC, Lwin NC, Chan NS, et al. Use of anterior segment optical coherence tomography to predict corneal graft rejection in small animal models. Invest Ophthalmol Vis Sci. 2014; 55: 6736–6741. [DOI] [PubMed] [Google Scholar]
  • 17.McDonnell PJ, Enger C, Stark WJ, et al. Corneal thickness changes after high-risk penetrating keratoplasty. Collaborative Corneal Transplantation Study Group. Arch Ophthalmol. 1993; 111: 1374–1381. [DOI] [PubMed] [Google Scholar]
  • 18.Chirapapaisan C, Abbouda A, Jamali A et al. In Vivo Confocal Microscopy Demonstrates Increased Immune Cell Densities in Corneal Graft Rejection Correlating with Signs and Symptoms. Am J Ophthalmol. 2019, ePublication. [DOI] [PubMed] [Google Scholar]
  • 19.Elsawy A, Abdel-Mottaleb M, Sayed I-O, et al. Automatic Segmentation of Corneal Microlayers on Optical Coherence Tomography Images. Trans. Vis. Sci. Tech 2019;8(3):39. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Koo TK, Li MY. 2016. A guideline of selecting and reporting intraclass correlation coefficients for reliability research. J Chiropr Med. 15(2):155–163. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Lopez de la Fuente C, Sanchez-Cano A, Segura F, Hospital EO, Pinilla l. 2016. Evaluation of total corneal thickness and corneal layers with spectral-domain optical coherence tomography. J Refract Surg. 32(1):27–32. [DOI] [PubMed] [Google Scholar]
  • 22.Koo TK, Li MY. A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability Research [published correction appears in J Chiropr Med. 2017 Dec;16(4):346]. J Chiropr Med. 2016;15(2):155–163. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Abou Shousha M, Elsawy A. U.S. Patent No. US20180192866A1 Washington, DC: Inventor; Method and system for three-dimnsional thickness mapping of corneal micro-layers and corneal diagnoses. US patent US20180192866A1 (2018). [Google Scholar]
  • 24.Roongpoovapatr V, Elsawy A, Wen D, et al. Three-Dimensional Bowmans Microlayer Optical Coherence Tomography for the Diagnosis of Subclinical Keratoconus. Invest Ophthalmol Vis Sci. 2018; 59(9):5742. [Google Scholar]
  • 25.Abou Shousha M, Wen D, Yoo SH, et al. Three-Dimensional Corneal Collagen Crosslinking Demarcation Band to Evaluate Corneal Flattening effect from Treatment: Pilot Study. American Academy of Ophthalmology (AAO) Annual Meeting; 2017; New Orleans. [Google Scholar]
  • 26.Abou Shousha MA, Perez VL, Wang J, et al. Use of ultra-high-resolution optical coherence tomography to detect in vivo characteristics of Descemet’s membrane in Fuchs’ dystrophy. Ophthal. 2010; 117(6):1220–1227. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Roongpoovapatr V, Amr Elsawy A, et al. Diagnosis of Corneal Graft Rejection Using 3-Dimensional Endothelial Descemet Membrane Optical Tomography Poster presented at ASCRS•ASOA Annual Meeting 2018; Walter E. Washington Convention Center. [Google Scholar]
  • 28.Alldredge OC, Krachmer JH. Clinical Types of Corneal Transplant Rejection Their Manifestations, Frequency, Preoperative Correlates, and Treatment. JAMA Ophthal 1981; 99(4):599–604. [DOI] [PubMed] [Google Scholar]
  • 29.Hill JC, Maske R, and Watson P. Corticosteroids in corneal graft rejection. Oral versus single pulse therapy Ophthalmology, 98 (3) (1991), pp. 329–333. [DOI] [PubMed] [Google Scholar]
  • 30.Yamazoe K, Shimazaki-Den S, Shimazaki J. Prognostic factors for corneal graft recovery after severe corneal graft rejection following penetrating keratoplasty BMC Ophthalmol, 13 (2013), p. 5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Claerhout H, Beele D, De Bacquer P. Kestelyn Factors influencing the decline in endothelial cell density after corneal allograft rejection Invest Ophthalmol Vis Sci, 44 (11) (2003), pp. 4747–4752. [DOI] [PubMed] [Google Scholar]
  • 32.Afshari NA, Pittard AB, Siddiqui A, et al. Clinical study of Fuchs corneal endothelial dystrophy leading to penetrating keratoplasty: a 30-year experience. Arch Ophthalmol. 2006; 124: 777–780. [DOI] [PubMed] [Google Scholar]
  • 33.Seitzman GD, Gottsch JD, Stark WJ. Cataract surgery in patients with Fuchs’ corneal dystrophy: expanding recommendations for cataract surgery without simultaneous keratoplasty. Ophthalmology. 2005; 112: 441–446. [DOI] [PubMed] [Google Scholar]
  • 34.Maier P, Heizmann U, Böhringer D. Predicting the risk for corneal graft rejection by aqueous humor analysis. Mol Vis. 2011. April 25; 17:1016–23. [PMC free article] [PubMed] [Google Scholar]
  • 35.Kocaba V, Colica C, Rabilloud M et al. Predicting Corneal Graft Rejection by Confocal Microscopy. Cornea 2015. October; 34 Suppl 10:S61–4. [DOI] [PubMed] [Google Scholar]
  • 36.Bizheva K, Haines L, Mason E, et al. In Vivo Imaging and Morphometry of the Human Pre-Descemet’s Layer and Endothelium with Ultrahigh-Resolution Optical Coherence Tomography. Invest Ophthalmol Vis Sci. 2016; 57(6):2782–2787. [DOI] [PubMed] [Google Scholar]
  • 37.Pekel G, Yagc R, Acer S, Ongun GT, Çetin EN, Simavl H. Comparison of Corneal Layers and Anterior Sclera in Emmetropic and Myopic Eyes. 2015; 34(7):786–790. [DOI] [PubMed] [Google Scholar]
  • 38.Lopez de la Fuente C, Sanchez-Cano A, Segura F, Hospital EO, Pinilla I. Evaluation of Total Corneal Thickness and Corneal Layers With Spectral-Domain Optical Coherence Tomography. J Refract Surg. 2016; 32(1):27–32. [DOI] [PubMed] [Google Scholar]
  • 39.Alberto D, Garello R. Corneal Sublayers Thickness Estimation Obtained by High-Resolution FD-OCT. Int J Biomed Imaging. 2013; 2013:989624. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Hutchings N, Simpson TL, Hyun C, et al. Swelling of the Human Cornea Revealed by High-Speed, Ultrahigh-Resolution Optical Coherence Tomography. Invest. Ophthalmol. Vis. Sci 2010; 51(9):4579–4584. [DOI] [PubMed] [Google Scholar]
  • 41.Flynn TH, Ohbayashi M, Dawson M et al. Use of ultrasonic pachymetry for measurement of changes in corneal thickness in mouse corneal transplant rejection. Br J Ophthalmol. 2010; 94(3):368–71. [DOI] [PubMed] [Google Scholar]
  • 42.Fischler MA, & Bolles RC Random sample consensus: a paradigm for model fitting with applications to image analysis and automated cartography. Communications of the ACM, 1981; 24(6), 381–395. [Google Scholar]
  • 43.Artal P Handbook of Visual Optics, Two-Volume Set: CRC Press; 2017. [Google Scholar]
  • 44.Yadav R, Kottaiyan R, Ahmad K, Yoon G-Y. Epithelium and Bowman’s layer thickness and light scatter in keratoconic cornea evaluated using ultrahigh resolution optical coherence tomography. J Biomed Opt 2012; 17:116010. [DOI] [PMC free article] [PubMed] [Google Scholar]

RESOURCES