Abstract
In this work, we estimate the diagnostic threshold of the instantaneous wave‐free ratio (iFR) through the use of a one‐dimensional haemodynamic framework. To this end, we first compared the computed fractional flow reserve (cFFR) predicted from a 1D computational framework with invasive clinical measurements. The framework shows excellent promise and utilises minimal patient data from a cohort of 52 patients with a total of 66 stenoses. The diagnostic accuracy of the cFFR model was 75.76%, with a sensitivity of 71.43%, a specificity of 77.78%, a positive predictive value of 60%, and a negative predictive value of 85.37%. The validated model was then used to estimate the diagnostic threshold of iFR. The model determined a quadratic relationship between cFFR and the ciFR. The iFR diagnostic threshold was determined to be 0.8910 from a receiver operating characteristic curve that is in the range of 0.89 to 0.9 that is normally reported in clinical studies.
Keywords: coronary arteries, FFR, haemodynamic modelling, iFR
In this work we estimate the diagnostic threshold of iFR through the use of a one‐dimentional haemodynamic model. The 1D model is first validated through a comparison with invasive fractional flow reserve measurements, for which a diagnostic accuracy of 75.76% was achieved. The one‐dimentional model was then utilised to compare cFFR and ciFR on patient‐specific arterial geometries that were extracted via CCTA. The diagnostic threshold of iFR was determined to be 0.891 which is in the normal range reported from invasive clinical studies.
1. INTRODUCTION
Coronary heart disease (CHD) is the leading cause of mortality worldwide and has been attributed to approximately 15.5% of deaths per year. The prevalence of CHD creates a significant economic burden because of the direct health care costs, reduced productivity, and informal public aftercare caused by CHD and other cardiovascular diseases (CVD).1, 2 The risk factors of CHD and other types of CVD have been extensively studied and indicate the risk of developing a CVD is increased for individuals who smoke and those who are overweight.3 There have been efforts from health care systems aimed at encouraging populations to lead a healthier lifestyle4 as this has been shown to be effective at preventing many types of CVD and other diseases. However, even with this guidance, the prevalence of CHD (and other types of CVD) has continued to either remain constant or increase, with a striking finding that approximately 48% of all Americans have a CVD.5 Because of advances in medicine with improved diagnostic and treatment tools, the mortality rates caused by CHD are reducing in high‐income countries; however, this is not the case in middle‐income and low‐income countries where the number of deaths are increasing.6, 7 The economic burden is also expected to significantly increase as health projections have indicated there will be an increase in CHD and other types of CVD in part caused by ageing populations.8, 9 Thus, it is important to not only improve access to medical facilities that includes early diagnostic and treatment methods but also reduce the costs and increase the efficiency of the various patient treatment pathways for more sustainable health care systems.10
CHD is caused by a narrowing (stenosis) of a coronary artery and is often from the process of atherosclerosis. The current gold standard of care for assessing the functional significance of a stenotic lesion is fractional flow reserve (FFR).11, 12, 13 An FFR‐guided treatment strategy has been shown to reduce unnecessary stenting, reduce costs, and increase patient outcomes14, 15, 16, 17 compared with a strategy that utilises only quantitative coronary angiography. FFR is an invasive procedure that is performed during coronary angiography. The procedure requires a catheter and guide wire to be inserted into either the femoral artery or radial artery. The catheter is then guided to the ascending aorta where the coronary arteries are located and a pressure‐sensitive wire is then used to measure the pressure ratio from a point distal of the stenosis to a point proximal to the stenosis (aorta). Critical to the technique of FFR is that the measurement is performed under hyperaemic conditions and so requires a hyperaemic‐inducing drug to be administered to the patient. Hyperaemia increases the heart rate and cardiac output of the patient, which increases blood flow through the coronary arteries significantly.18, 19 However, a study showed that 28% of patients experience adverse reactions to adenosine, with 8% having such a severe reaction that the adenosine infusion had to be discontinued during the procedure.20 Other serious reactions to adenosine have also been reported, such as bronchospasms,21 tachyarrhythmia, and even cardiac arrest,22 although adverse effects as serious as these are rare.
Alternative diagnostic tools for CHD are continually being developed. One such technique is diastolic FFR (dFFR),23 which refers to a collection of variants to FFR. In the same way as the conventional measurement, dFFR requires a hyperaemic drug to be administered, and the main idea behind this variant is that the largest flow rates and lowest resistances in the coronary arteries are observed during diastole. The following are the main variants of dFFR: full dFFR, where the mean FFR over the entire diastolic phase is used; mid‐dFFR, where the FFR is measured at a single point in time (the middle of diastole); and end‐dFFR, where the FFR measurement is recorded only at end‐diastole by utilising an electrocardiography gating technique. End‐dFFR appears to be the most promising of these and has been shown to have a better correlation than conventional FFR to the FFR measured using flow probes,24 although few studies on dFFR currently exist.
Another promising alternative to FFR is the instantaneous wave‐free ratio (iFR).25, 26 The iFR procedure also requires a pressure sensitive catheter to be inserted in order to measure the pressure drop across stenotic lesion, but iFR is performed under resting conditions, so it does not require a hyperaemic‐inducing drug to be administered. The use of iFR was initially met with significant resistance, particularly by the community who strongly supported the use of conventional FFR27, 28 under maximum hyperaemia, but many studies have shown the viability and potential of iFR in diagnosing functionally significant stenotic lesions, showing similar diagnostic accuracy to conventional FFR29, 30, 31, 32, 33, 34, 35, 36, 37 and even showing a better repeatability than conventional FFR38 and a stronger correlation with the coronary velocity flow reserve.30 Questions still remain on whether iFR can reliably be used to replace FFR as several studies have shown mismatches in patient categorisation between iFR and the clinically trusted technique of conventional FFR,32 while the reasons for these disagreements were not fully understood. However, it has since been shown that the differences between the two indices was more likely related to FFR overestimating the severity of the stenosis, rather than iFR underestimating the severity.39 A hybrid iFR‐FFR approach has also been proposed,38, 40 which enhances the diagnostic accuracy while reducing the number of patients who require adenosine to be administered. In the DEFINE‐FLAIR study,41 iFR was shown to have noninferior patient outcomes when compared with conventional FFR and is considered a practical alternative to FFR that can avoid adverse patient reactions to hyperaemic drugs while costing less. The clinical cut‐off points of FFR and iFR are generally 0.8 and 0.89 respectively, although different cut‐off points have been proposed for iFR, ranging from 0.83 to 0.92,29, 36, 42 implying that the true cut‐off point for iFR is still uncertain. In the FFR‐iFR hybrid approach, a lesion with an iFR value in the grey‐zone of 0.86 to 0.93 also undergoes FFR, while an iFR value of less than 0.86 requires treatment and lesions with iFR values above 0.93 do not undergo surgical treatment.38, 40
Non‐invasive methods for estimating FFR are based on reconstructed and segmented coronary computed tomography angiography (CCTA) images. Techniques from computational fluid dynamics are then applied to the patient geometry to determine the FFR while avoiding the need for any invasive cardiac catheterisation. The technique is often called computed FFR (cFFR) and relies on the estimation of several important modelling components, which includes the patients' cardiovascular response to the hyperaemic drug infusion, which influences the elasticity of arteries, and boundary conditions of the model. Different computational methodologies have been proposed, which include computationally expensive three‐dimensional models,43, 44, 45, 46, 47, 48, 49, 50 and reduced‐order models.51, 52, 53, 54, 55, 56, 57, 58, 59, 60 One‐dimensional models have been shown to have excellent agreement with three‐dimensional models53, 60 but can be computed in seconds, rather than hours (for 3D). Computational models of iFR have been proposed but are generally compared with only invasive FFR measurements,61 in which the estimated iFR cut‐off point was 0.82; however, rigid‐wall conditions were assumed, which is known to significantly overestimate the pressure drop across a stenosis for cFFR62 and is likely to have a similar impact for ciFR or utilised in a hybrid‐approach,63 although a Monte Carlo simulation that utilises a lumped model64 without patient‐specific geometry has been implemented. Nevertheless, a comparison between the invasive iFR measurement and computational iFR was performed in Calmac et al.65 However, the computational methodology is not described in Calmac et al,65 requires images from invasive coronary angiogragraphy, and the cut‐off point for iFR is assumed to be 0.89.
The purpose of this work is to compare clinical FFR measurements with cFFR estimated via a one‐dimensional haemodynamic model in order to validate the cFFR strategy and to determine the diagnostic cut‐off point of iFR through the application of computational models on patient‐specific coronary arterial networks that have been extracted from non‐invasive CCTA. By using the same extracted patient‐specific coronary network geometry, the boundary conditions of the 1D model can be easily adjusted to change from hyperaemic conditions to resting physiological conditions, which essentially adapts the cFFR model into a ciFR model. The correlation between ciFR and cFFR and the clinically accepted FFR diagnostic threshold of 0.8 will then be used in order to estimate the diagnostic threshold for the iFR procedure. An advantage of using this computational framework is that variability of surgical techniques and of the patients' physiological conditions can be eliminated, which allows the iFR diagnostic threshold to be estimated in a more controlled environment. The patient data were collected retrospectively, and no patient specific measurements were used to aid the model parameter estimation.
The paper is organised into the following parts: in Section 2, the characteristics of the data utilised in this paper and the one‐dimensional computational framework are described; the results in Section 3 begin by comparing the cFFR predicted by the computational framework with the clinically invasive FFR measurements; this leads on to the comparison of the model predicted FFR and iFR values; in Section 4, the results are discussed and compared with previous findings in literature; and finally, the main conclusions of the work are presented in Section 5.
2. MATERIALS AND METHODS
2.1. Study characteristics
In this study, an anonymised retrospective data set of 52 patients with a total of 66 lesions was collected. The exact clinical FFR values were not available for a subset of 10 of these patients (15 stenosis). Only whether the FFR was positive or negative for these patients was available. An overview of the locations and characteristics of these stenotic lesions can be seen in Table 1. The majority of lesions (78.43%) considered in this work are in the intermediate FFR range of 0.7<FFR<0.9, with 68.2% of the total lesions having a negative FFR value and 31.8% having a positive FFR value. A stenosis location of the left anterior descending (LAD) artery was most prevalent (60.6%). Patients who had undergone previous coronary surgical interventions were included only if the intervention was performed within vessels that were not of interest for this study. Patients with serial stenoses and CCTA data images that contained motion artefacts and significant levels of calcification were also included in this study. Table 2 shows statistics on the characteristics of the CCTA images, which includes information on the percentage of cases with calcification and motion artefacts that are present.
Table 1.
Stenosis Location (n = 66) | n, % |
---|---|
LCA | 3 (4.5) |
LAD | 40 (60.6) |
LCX | 7 (10.6) |
DA | 3 (4.5) |
MA | 2 (3.0) |
IA | 1 (1.5) |
RCA | 10 (15.2) |
FFR Characteristics (n = 66) | n, % |
---|---|
FFR <0.8 | 21 (31.8) |
FFR ≥0.8 | 45 (68.2) |
Abbreviations: FFR, fractional flow reserve; LAD, left anterior descending.
Table 2.
Stenosis No. and Type | Single Focal | Single Diffuse | Multiple |
---|---|---|---|
50 | 16.67 | 33.33 | |
Stent no. | 0 | 1 | 2 |
87.88 | 10.61 | 1.52 | |
Occulsion no. | 0 | 1 | 2 |
83.33 | 15.15 | 1.52 | |
Bifurcation | No | Within 1‐2 mm | Within 1 mm |
50 | 30 | 20 | |
Calcification | None | Minor | Major |
22.73 | 48.48 | 28.79 | |
Motion artefacts | None | Minor | Major |
74.24 | 19.7 | 6.06 |
Note. Percentage of cases includes the following: single focal stenosis; single diffuse stenosis; or if there are multiple stenoses (either focal of diffuse), number of stents and vessel occlusions, number of cases with a stenosis near (between 2 and 5 mm), and at (within 2 mm) of a bifurcation; whether calcification is present, and to what severity (minor is less likely to affect segmentation accuracy, while major is likely to impact segmentation accuracy); and whether motion artefacts are present and to what severity. All values given as a percentage of the cohort.
Abbreviation: CCTA, coronary computed tomography angiography.
2.2. Computed FFR methodology
The segmentation of CCTA images, centreline and vessel geometry extraction, were all performed in the image segmentation software VMTKLab, (Orobix, Italy).
2.2.1. One‐dimensional haemodynamic model
The modelling methodology implemented in this work is described in Carson66 and involves the one‐dimensional haemodynamic equations in a pressure‐volumetric flow rate formulation66, 67 that is given by the conservation of mass:
(1) |
where C a is the vessel compliance, P is the mean hydrostatic pressure in a cross section, and Q is the volumetric flow rate; and the momentum equation:
(2) |
where ρ=1.05g/cm3 is the density of blood, A is the cross‐sectional area of the vessel, μ=0.04Poise is the blood viscosity, and the magnitude of the viscous coefficient of 22 is from Smith et al68 and was shown to be the best fit to experimental data for the coronary system. The vessel compliance is calculated as from the following viscoelastic constitutive law,69
(3) |
with
(4) |
where the reference wave speed c 0 is calculated from the vessel diameter using the empirical formula69, 70 as
(5) |
where D 0 is the reference diameter, and the fitting parameters are k 1=20 g/s2/cm, k 2=−22.5 cm−1, and k 3=86.5 g/s2/cm. The reference pressure P 0 is set to equal the diastolic pressure, and the collapsing pressure P collapse=−10 mmHg. The wall viscous coefficient Γ69 is calculated from the following:
(6) |
where D is the vessel lumen diameter.
The system of equations are solved using a subdomain collocation scheme66, 71 that is second‐order accurate in both time and space.
2.2.2. Boundary conditions
The inlet and outlet boundary conditions for the model are predicted by using a two‐tiered parameter estimation technique. The first tier utilises a general, nonpatient‐specific, closed‐loop 1D‐0D network to estimate the volumetric inflow rate for the coronary arteries, while the second tier uses the inflow rate determined from the first tier as the coronary inflow boundary condition and then estimates the vascular bed resistances for the patient‐specific coronary arterial network that was extracted from the CCTA image data.
2.2.2.1. First tier of the parameter estimation
The first‐tier of the parameter estimation utilising the closed‐loop cardiovascular model with an initial and adaptive parameter estimation technique that are described in the previous studies.66, 72 The purpose of this technique is to provide an estimate for the volumetric flow rate at the inlet of the coronary arteries and to provide an estimate for the left and right ventricular pressures, as they add an external pressure to the coronary vascular beds. The first tier is performed twice: once for hyperaemic conditions and once for resting conditions. As no clinical data on patient pressures or heart rates were available, only population averaged values from literature were used. The population averaged values for both hyperaemic and resting conditions (assumed the same for all patients) are described in Table 3 and are from population averages of previous coronary artery studies.18, 19 The assumption that the values of parameters such as the heart rate and blood pressures are consistent among patients is valid as FFR has been shown to be independent of heart rate, blood pressure, and heart contractility12, 73; the relationship between iFR and FFR is also independent of heart rate,26 and iFR has also been shown to be independent of heart rate, blood pressure, and heart contractility.25
Table 3.
Resting | Hyperaemic | |
---|---|---|
Systolic pressure, mmHg | 115 | 115 |
Diastolic pressure, mmHg | 74 | 70 |
Cardiac output, L/min | 5.19 | 7.6 |
Heart rate, BPM | 65 | 90 |
An additional change from the resting condition to the hyperaemic condition is performed by reducing the total coronary vascular bed resistance by 78%, this causes the mean flow rate in the coronary arteries for the hyperaemic condition to be 3.5 times larger than in resting conditions.18 This means that there are only two different defined inflow waveforms for the coronary arteries, which allow a more straightforward comparison between cFFR and cIFR.
2.2.2.2. Second tier of the parameter estimation
The inlet boundary condition for all simulations is a defined flow rate (one for hyperaemic conditions and one for resting conditions) that was generated by the closed‐loop cardiovascular model described in the first parameter estimation tier. The second tier uses the patient‐specific coronary geometry that was extracted via segmentation. The outlet boundary condition is a lumped‐parameter model that includes external pressure from the left and right ventricles66 (saved from the first tier of the parameter estimation). The total coronary resistance of each branch is calculated using the commonly used relation:
(7) |
where the subscript i represents the branch (left or right) of the coronary artery network, and Q i is the defined inflow into the coronary branch from the first tier. The resistance of each branch is then distributed to each terminal vessel using the proximal Murray's law with a power of 2.27.74 This means that the second tier of parameter identification relies solely on the patient specific geometry for the distribution of the vascular resistance.
2.2.2.3. Model cFFR and ciFR
Both conventional FFR and iFR are measurements that involve the ratio of the pressure proximal (P p) to a stenosis (usually aortic pressure) and the pressure distal (P d) to a stenosis. The ratio of P d/P p is used for all cases. The main differences between these two methods are as follows: conventional FFR is measured under maximal hyperaemic conditions, which requires a drug such as adenosine to be administered and is the mean of the pressure ratio P d/P p over one cardiac cycle (in clinical practice multiple cycles); iFR is performed under resting conditions and is the mean of the pressure ratio P d/P p during the wave‐free period, which is shown in Figure 1 as the grey shaded region. In this model, ciFR is assumed to begin 1/5th into diastole to the end of the cardiac cycle.
3. RESULTS
3.1. cFFR comparison with clinical invasive FFR
An overview of results for the reduced‐order cFFR model in comparison with the invasive clinical FFR measurement is shown in Table 4 with the correlation and Bland‐Altman graphs shown in Figure 2. The diagnostic accuracy of the model is 75.76%. The mean absolute difference between the cFFR and FFR values is 0.059, with a mean difference of −0.015 and a standard deviation of 0.0813. A Pearson coefficient of 0.484 shows a moderate linear correlation between the cFFR and invasive FFR values.
Table 4.
No. of true positive | 15 |
No. of false positive | 10 |
No. of true negative | 35 |
No. of false negative | 6 |
Sensitivity, % | 71.43 |
Specificity, % | 77.78 |
PPV, % | 60.00 |
NPV, % | 85.37 |
Diagnostic accuracy % | 75.76 |
Abbreviations: cFFR, computed fractional flow reserve; NPV, negative predictive value; PPV, positive predictive value.
3.2. Correlation between cFFR and ciFR
The main results are shown in Figure 3 with a Pearson correlation coefficient of ρ=0.95352 showing a strong linear correlation between cFFR and ciFR values. The Bland‐Altman plot shown in Figure 3B shows the mean difference between cFFR and ciFR to be 0.08402.
Lines of best‐fit can be used to aid determining the correct relationship between cFFR and ciFR, as many correlation measures are based on the assumption of a linear relation. Figure 3A compares ciFR to cFFR using the the following polynomials of best‐fit: the linear polynomial
(8) |
the quadratic polynomial
(9) |
and the cubic polynomial
(10) |
The polynomials of best fit allow the determination of the cut‐off point for iFR by comparing the iFR value to the diagnostic cut‐off point of FFR=0.8. The linear polynomial estimates a value of ciFR1=0.8863, the quadratic polynomial estimates ciFR2=0.8918, and the cubic polynomial estimates ciFR3=0.8916. The results indicate that the relationship between FFR and iFR is better described by a second‐order polynomial that provides an excellent fit for the data.
3.2.0.4. Receiver operating characteristic curve
It is common to use receiver operating characteristic curves (ROC) to determine the optimum threshold of a methodology. Figure 4 shows the ROC curve of ciFR for different threshold values, when compared with cFFR (with a threshold of 0.8), where the area under the curve is AUC=0.9971. The optimum threshold for ciFR via ROC analysis is determined to be ciFR=0.8910.
4. DISCUSSION
The performance of the reduced‐order cFFR methodology in comparison with the invasive FFR measurements is very satisfactory at this early stage of development and shows a similar level of accuracy to several three‐dimensional cFFR methodologies.75 The diagnostic accuracy of the reduced‐order methodology is 75.76%, which is in the region seen from other studies; however, the current methodology did not have patient information along with the CCTA data, and hence, a population average blood pressure, heart rate, and cardiac output from published studies18, 19 was chosen. The utilisation of routinely measured patient data may improve the models estimation of cFFR by integrating more patient‐specific parameters. Even with the lack of patient‐specific data the strategy performed well with a sensitivity of 71.43%, a specificity of 77.78%, a positive predictive value of 60% and a negative predictive value of 85.37%. Furthermore, the mean difference between cFFR and invasive FFR was −0.015314, which is lower in magnitude than the mean difference shown in the three‐dimensional methodology performed for different patient data in the DISCOVER‐FLOW43 (0.022), DeFACTO44 (0.058), and NXT45 (0.03) studies. The standard deviation of the reduced‐order model is 0.0813, which is close to the lowest standard deviation from the other studies of 0.074.45 The model‐predicted cFFR showed a lower Pearson correlation with the FFR than in other studies75; however, this could be attributed to the smaller cohort size and the low range of measured FFR values, as the majority (78.43%) of invasive cFFR measurements were in the range 0.7< FFR <0.9. Overall, the results indicate that the reduced‐order methodology presented here provides a high level of diagnostic accuracy for cFFR.
The framework presented does need to be tested on a larger cohort in fully blinded conditions and ideally using prospective data. In the current cohort, there are several patients that would have been excluded in other studies because of poor image quality75; however, the diagnostic performance of the methodology was not the only objective of this paper. The main reason for the comparison at this stage was to validate the methodology and provide confidence that the predicted cFFR values of the model are close to the invasive FFR measurements, which in turn will provide an indication that the iFR values predicted by this model can be trusted. The methodology also allows the computed iFR values to be implemented within the same framework as cFFR, and thus, the developed software could easily be adapted to automatically perform the hybrid iFR‐FFR technique.38, 40 The use of the same inlet volumetric flow rate and estimated heart rate also ensures that only the geometry and downstream resistance of the coronary network are the main variables that affect the FFR and iFR value predictions.
4.1. cFFR and ciFR comparison
The use of iFR in combination with or instead of FFR has received increased attention in recent years. However, many studies use different threshold values of iFR to determine whether a patient needs further treatment.36 The estimation of the iFR diagnostic threshold ranges from 0.92,76 0.9,29, 31, 77, 78 0.89,79, 80 0.88,81 and even 0.83,82 which obviously indicates that the best diagnostic threshold for iFR is still not known. The results from this paper indicate that the diagnostic threshold for iFR is close to 0.89 and that the relationship between ciFR and cFFR is quadratic in nature. This agrees with the observed behaviour from a comparison of invasive FFR and iFR measurements.39 Other studies have reported an approximately linear behaviour from measurements32, 83 and from a Monte Carlo simulation with a lumped parameter model64; however, there is a noticeable deviation from the linear line of best fit at lower FFR and iFR values, which indicates that a nonlinear relationship would provide a better fit between FFR and iFR measures.
Another interesting aspect to consider is whether the cardiovascular system is a pressure‐driven or flow‐driven system, which depends on whether the heart is a flow generator or pressure generator.84 The fact that the heart contracts and creates a pressure that acts on coronary capillary vessels complicates matters as there would be a forward‐propagating pressure wave originating from the heart that travels into the aorta and then through the coronary arteries from the proximal to distal location and also a backward‐propagating pressure wave that originates from the coronary capillaries as a result of the contracting heart that travels from the distal region of the coronary arteries to the proximal region.85 In reality ,it is more likely that the cardiovascular system is a complex combination of both pressure‐driven and flow‐driven phenomena.84 However, because of the forward and backward pressure waves during systole and at the start of diastole, which can be seen in Figure 5, a pressure index based on the diastolic phase could be more reliable as the lowest resistance to flow in the coronary arterial network is observed during diastole.
Many of the fundamental assumptions needed for FFR have since been proven incorrect. For example, the pressure and flow are assumed to be directly proportional (have a linear relationship) when the resistance is constant and minimal86 but has since been shown have a curvilinear relationship experimentally87 and mathematically. This essentially means that FFR relies on the assumption that the characteristics of flow are the same in both diseased and healthy vessels. The assumption of the microcirculatory resistance being constant and minimal during hyperaemia is also questionable and has been proven incorrect in the presence of several conditions including microvascular dysfunction.86 In hyperaemia, the resistance averaged over a cardiac period may be consistent during hyperaemia, but the microvascular resistance will vary significantly over a cardiac cycle as the heart contraction squeezes on the microcirculation during systole that increases resistance.88 These assumptions are not required for iFR; instead, the main assumption for iFR is that the resistance of the microcirculation is stable during the wave‐free period, which is during a period in diastole where no waves are generated, such as seen in Figure 5C from a normalised time of 0.6 to 1.
Many studies compare the diagnostic performance of iFR to FFR and consider the latter index to be the “perfect measure.” This generally means that iFR looks worse than FFR, but this is a rather unfair comparison. Further studies have indicated that iFR and FFR have a similar number of negative cardiovascular events after 1 year41 and thus is at least on‐par with FFR. Several studies have compared iFR and FFR29, 30, 31, 32, 33, 34, 35, 36, 37; however, only one attempted to explain the cause of the various diagnostic disagreements observed between iFR and FFR.39 The study concluded that it was actually the FFR that was likely overestimating the severity of the stenosis because of the hyperaemic condition rather than iFR underestimating the stenosis, and for the cases that had a positive FFR but negative iFR, the observed coronary flow characteristics were similar to that seen in angiographically unobstructed vessels, which indicates that iFR may be the more reliable and suitable measure.
From a non‐invasive standpoint for the determination of iFR and FFR, iFR has more advantages over FFR as it does not require an estimation of how the patient will react to the administered hyperaemic inducing drug. Many cardiac and haemodynamic parameters can be measured by non‐invasive means, including cardiac output estimations, brachial artery blood pressures, and heart rate. These parameters can be directly utilised by any non‐invasive iFR model predictions rather than attempting to predict the effects of a hyperaemic condition, which are variable between patients and are required for FFR estimations. In addition, in the majority, the patient CCTA scans are performed at resting conditions. This is particularly important as the inducing of hyperaemic conditions through the use of a drug, such as adenosine, was observed to increase coronary vessel diameters by up to 15%89; thus, there is even significant uncertainty for cFFR regarding the actual patient geometry that is extracted from CCTA data.
4.2. Limitations
The main limitations of this study are that the size of the cohort is relatively small with 52 patients and a total of 66 stenoses. Furthermore, we assume that there are no pressure losses at vessel junctions and do not have patient data such as age, blood pressures, heart rate, or gender, which may play a role in FFR prediction. We do not have any invasive iFR measurements; only FFR measurements and the patient CCTA images were available. As the data are retrospective, we do not know the exact location of the pressure measurement taken during invasive FFR, as only the general location is normally recorded in the clinic. However, the main target of this study was to determine the diagnostic threshold of iFR in a more controlled environment, and although the cFFR performance of the model is very good, it was not the focus of this paper, and thus, the additional patient data are not of importance here.
It has been observed that vessels with coronary artery disease90 can cause a long‐term auto‐regulatory response in the coronary micro‐vasculature resistance to preserve flow. This may impact the accuracy of the flow estimates. However, it is not known how or if this auto‐regulation affects flow rates in hyperaemia and would add an additional unknown and source of uncertainty in the model, which will make comparison between FFR and iFR more challenging. Thus, we have not included this auto‐regulatory response in this study.
5. CONCLUSIONS
The diagnostic performance of the cFFR with the invasive FFR measurement is very promising. The methodology not only showed a lower correlation between cFFR and FFR than in other studies but also showed the lowest magnitude of mean difference and one of the lowest standard deviations between the computed and measured FFR when compared with other studies. The lower correlation coefficient may be due to the lower patient numbers considered in this cohort and also on the nonselective nature in this study with regard to image quality such as the presence of significant motion artefacts, blooming artefacts, and high levels of calcification that can have a large impact on the accuracy of the segmentation process. The methodology must now be performed on a significantly larger cohort in a fully blinded fashion and ideally prospectively in order to improve the confidence in the 1D modelling methodology for both FFR and iFR predictions.
A comparison between cFFR and computed iFR was also performed by utilising the patient CCTA data. The model predicted an iFR diagnostic cut‐off point of 0.891 with the correlation and polynomial of best fit between cFFR and ciFR being quadratic in nature. Further studies involving comparisons between FFR and iFR must be performed in order to determine whether FFR or iFR is the more reliable measure. There is a significant advantage of iFR, as it does not require a hyperaemic drug infusion which can cause negative side affects in patients and would also be less expensive. When considering a computational methodology for prediction purposes, ciFR is also more attractive as a diagnosis index when compared with cFFR, as the patients non‐invasive measurements can be utilised directly without the need to predict hyperaemic conditions.
ACKNOWLEDGEMENTS
This work was supported by Health Data Research UK (MR/S004076/1), which is funded by the UK Medical Research Council, Engineering and Physical Sciences Research Council, Economic and Social Research Council, Department of Health and Social Care (England), Chief Scientist Office of the Scottish Government Health and Social Care Directorates, Health and Social Care Research and Development Division (Welsh Government), Public Health Agency (Northern Ireland), British Heart Foundation, and the Wellcome Trust.
Carson JM, Roobottom C, Alcock R, Nithiarasu P. Computational instantaneous wave‐free ratio (IFR) for patient‐specific coronary artery stenoses using 1D network models. Int J Numer Meth Biomed Engng. 2019;35:e3255 10.1002/cnm.3255
REFERENCES
- 1. Organization WH . Global Atlas on Cardiovascular Disease Prevention and Control. Geneva, Switzerland: WORLD HEALTH ORGN; 2012. [Google Scholar]
- 2. Benjamin EJ, Blaha MJ, Chiuve SE, et al. Heart disease and stroke statistics‐2017 update: a report from the american heart association. Circulation. 2017;135(10):e146‐e603. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Virtanen M, Vahtera J, Singh‐Manoux A, Elovainio M, Ferrie JE, Kivimäki M. Unfavorable and favorable changes in modifiable risk factors and incidence of coronary heart disease: the whitehall II cohort study. Int J Cardiol. 2018;269:7‐12. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Dunn AL. Effectiveness of lifestyle physical activity interventions to reduce cardiovascular disease. Am J Lifestyle Med. 2009;3(1_suppl):11S‐18S. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Benjamin EJ, Muntner P, Alonso A, et al. Heart disease and stroke statistics‐2019 update: a report from the american heart association. Circulation. 2019;139(10):e56‐e528. [DOI] [PubMed] [Google Scholar]
- 6. Finegold JA, Asaria P, Francis DP. Mortality from ischaemic heart disease by country, region, and age: Statistics from world health organisation and united nations. Int J Cardiol. 2013;168(2):934‐945. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7. Bhatnagar P, Wickramasinghe K, Wilkins E, Townsend N. Trends in the epidemiology of cardiovascular disease in the UK. Heart. 2016;102(24):1945‐1952. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8. Mathers CD, Loncar D. Projections of global mortality and burden of disease from 2002 to 2030. PLoS Med. 2006;3(11):e442. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Vilahur G, Badimon JJ, Bugiardini R, Badimon L. Perspectives: the burden of cardiovascular risk factors and coronary heart disease in europe and worldwide. Eur Heart J Suppl. 2014;16(suppl A):A7‐A11. [Google Scholar]
- 10. Roth GA, Johnson C, Abajobir A, et al. Global, regional, and national burden of cardiovascular diseases for 10 causes, 1990 to 2015. J Am Coll Cardiol. 2017;70(1):1‐25. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. Pijls NH, De Bruyne B, Peels K, et al. Measurement of fractional flow reserve to assess the functional severity of coronary‐artery stenoses. N Engl J Med. 1996;334:1703‐1708. [DOI] [PubMed] [Google Scholar]
- 12. Kolli KK, Banerjee RK, Peelukhana SV, et al. Influence of heart rate on fractional flow reserve, pressure drop coefficient, and lesion flow coefficient for epicardial coronary stenosis in a porcine model. Am J Physiol Heart Circ Physiol. 2011jan;300(1):H382‐H387. [DOI] [PubMed] [Google Scholar]
- 13. Lotfi A, Jeremias A, Fearon WF, et al. Expert consensus statement on the use of fractional flow reserve, intravascular ultrasound, and optical coherence tomography: a consensus statement of the society of cardiovascular angiography and interventions. Catheter Cardiovasc Interv. 2014;83:509‐518. [DOI] [PubMed] [Google Scholar]
- 14. Pijls NH, van Schaardenburgh P, Manoharan G, et al. Percutaneous coronary intervention of functionally nonsignificant stenosis: 5‐year follow‐up of the DEFER Study. J Am Coll Cardiol. 2007;49:2105‐2111. [DOI] [PubMed] [Google Scholar]
- 15. Tonino PA, Fearon WF, De Bruyne B, et al. Angiographic versus functional severity of coronary artery stenoses in the fame study fractional flow reserve versus angiography in multivessel evaluation. J Am Coll Cardiol. 2010;55:2816‐2821. [DOI] [PubMed] [Google Scholar]
- 16. Pijls NH, Fearon WF, Tonino PA, et al. Fractional flow reserve versus angiography for guiding percutaneous coronary intervention in patients with multivessel coronary artery disease: 2‐year follow‐up of the fame (fractional flow reserve versus angiography for multivessel evaluation) study. J Am Coll Cardiol. 2010;56:177‐184. [DOI] [PubMed] [Google Scholar]
- 17. Fearon WF, Bornschein B, Tonino PA, et al. Economic evaluation of fractional flow reserve‐guided percutaneous coronary intervention in patients with multivessel disease. Circulation. 2010;122:2545‐2550. [DOI] [PubMed] [Google Scholar]
- 18. Watzinger N, Lund GK, Saeed M, et al. Myocardial blood flow in patients with dilated cardiomyopathy: quantitative assessment with velocity‐encoded cine magnetic resonance imaging of the coronary sinus. J Magn Reson Imaging. 2005;21(4):347‐353. [DOI] [PubMed] [Google Scholar]
- 19. Wieneke H, von Birgelen C, Haude M, et al. Determinants of coronary blood flow in humans: quantification by intracoronary doppler and ultrasound. Eur J Appl Physiol. 2005;98(3):1076‐1082. [DOI] [PubMed] [Google Scholar]
- 20. Voigtländer T, Schmermund A, Bramlage P, et al. The adverse events and hemodynamic effects of adenosine‐based cardiac MRI. Korean J Radiol. 2011;12(4):424. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Pencharz D, Quigley A‐M, Hall M, Wagner TLJ. Rapid and severe adverse reaction to adenosine during a pharmacological stress test for a myocardial perfusion scan. Clin Nucl Med. 2013;38(9):758. [DOI] [PubMed] [Google Scholar]
- 22. El‐Menyar A, Gehani A. Adenosine‐induced tachyarrhythmia and cardiac arrest. Futur Cardiol. 2010;6(4):433‐436. [DOI] [PubMed] [Google Scholar]
- 23. Abe M, Tomiyama H, Yoshida H, Doba N. Diastolic fractional flow reserve to assess the functional severity of moderate coronary artery stenoses. Circulation. 2000;102(19):2365‐2370. [DOI] [PubMed] [Google Scholar]
- 24. Chalyan DA, Zhang Z, Takarada S, Molloi S. End‐diastolic fractional flow reserve. Circ Cardiovasc Interv. 2014;7(1):28‐34. [DOI] [PubMed] [Google Scholar]
- 25. Sen S, Escaned J, Malik IS, et al. Development and validation of a new adenosine‐independent index of stenosis severity from coronary wave‐intensity analysis: results of the ADVISE (ADenosine Vasodilator Independent Stenosis Evaluation) study. J Am Coll Cardiol. 2012;59:1392‐1402. [DOI] [PubMed] [Google Scholar]
- 26. Sen S, Asrress KN, Nijjer S, et al. Diagnostic classification of the instantaneous wave‐free ratio is equivalent to fractional flow reserve and is not improved with adenosine administration. J Am Coll Cardiol. 2013;61(13):1409‐1420. [DOI] [PubMed] [Google Scholar]
- 27. Rudzinski W, Waller AH, Kaluski E. Instantaneous wave‐free ratio and fractional flow reserve: Close, but not close enough! J Am Coll Cardiol. 2012;59(21):1915‐1916. [DOI] [PubMed] [Google Scholar]
- 28. Pijls NHJ, ‘t Veer MV, Oldroyd KG, et al. Instantaneous wave‐free ratio or fractional flow reserve without hyperemia. J Am Coll Cardiol. 2012;59(21):1916‐1917. [DOI] [PubMed] [Google Scholar]
- 29. Jeremias A, Maehara A, Généreux P, et al. Multicenter core laboratory comparison of the instantaneous wave‐free ratio and resting Pd/Pa with fractional flow reserve. J Am Coll Cardiol. 2014;63(13):1253‐1261. [DOI] [PubMed] [Google Scholar]
- 30. Petraco R, van de Hoef TP, Nijjer S, et al. Baseline instantaneous wave‐free ratio as a pressure‐only estimation of underlying coronary flow reserve. Circ Cardiovasc Interv. 2014;7(4):492‐502. [DOI] [PubMed] [Google Scholar]
- 31. Petraco R, Al‐Lamee R, Gotberg M, et al. Real‐time use of instantaneous wave–free ratio: Results of the ADVISE in‐practice: an international, multicenter evaluation of instantaneous wave–free ratio in clinical practice. Am Heart J. 2014;168(5):739‐748. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32. Pisters R, Ilhan M, Veenstra LF, et al. Instantaneous wave‐free ratio and fractional flow reserve in clinical practice. Neth Hear J. 2018;26(7‐8):385‐392. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33. van’t Veer M, Pijls NHJ, Hennigan B, et al. Comparison of different diastolic resting indexes to iFR. J Am Coll Cardiol. 2017;70(25):3088‐3096. [DOI] [PubMed] [Google Scholar]
- 34. Götberg M, Christiansen EH, Gudmundsdottir IJ, et al. Instantaneous wave‐free ratio versus fractional flow reserve to guide PCI. N Engl J Med. 2017;376(19):1813‐1823. [DOI] [PubMed] [Google Scholar]
- 35. Lee JM, Shin E‐S, Nam C‐W, et al. Clinical outcomes according to fractional flow reserve or instantaneous wave‐free ratio in deferred lesions. J Am Coll Cardiol Intv. 2017dec;10(24):2502‐2510. [Google Scholar]
- 36. Maini R, Moscona J, Katigbak P, et al. Instantaneous wave‐free ratio as an alternative to fractional flow reserve in assessment of moderate coronary stenoses: a meta‐analysis of diagnostic accuracy studies. Cardiovasc Revasc Med. 2018jul;19(5):613‐620. [DOI] [PubMed] [Google Scholar]
- 37. Verardi R, Fioravanti F, Barbero U, et al. Network meta‐analysis comparing iFR versus FFR versus coronary angiography to drive coronary revascularization. J Interv Cardiol. 2018;31(6):725‐730. [DOI] [PubMed] [Google Scholar]
- 38. Härle T, Bojara W, Meyer S, Elsässer A. Comparison of instantaneous wave‐free ratio (iFR) and fractional flow reserve (FFR)—first real world experience. Int J Cardiol. 2015;199:1‐7. [DOI] [PubMed] [Google Scholar]
- 39. Cook CM, Jeremias A, Petraco R, et al. Fractional flow reserve/instantaneous wave‐free ratio discordance in angiographically intermediate coronary stenoses. JACC Cardiovasc Interv. 2017dec;10(24):2514‐2524. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40. Rivero F, Cuesta J, Bastante T, et al. Diagnostic accuracy of a hybrid approach of instantaneous wave‐free ratio and fractional flow reserve using high‐dose intracoronary adenosine to characterize intermediate coronary lesions: results of the PALS (practical assessment of lesion severity) pros. Catheter Cardiovasc Interv. 2017;90(7):1070‐1076. [DOI] [PubMed] [Google Scholar]
- 41. Davies JE, Sen S, Dehbi H‐M, et al. Use of the instantaneous wave‐free ratio or fractional flow reserve in PCI. N Engl J Med. 2017;376(19):1824‐1834. [DOI] [PubMed] [Google Scholar]
- 42. Escaned J, Echavarría‐Pinto M, Garcia‐Garcia HM, et al. Prospective assessment of the diagnostic accuracy of instantaneous wave‐free ratio to assess coronary stenosis relevance. JACC Cardiovasc Interv. 2015;8(6):824‐833. [DOI] [PubMed] [Google Scholar]
- 43. Koo BK, Erglis A, Doh JH, et al. Diagnosis of ischemia‐causing coronary stenoses by noninvasive fractional flow reserve computed from coronary computed tomographic angiograms. Results from the prospective multicenter DISCOVER‐FLOW (diagnosis of ischemia‐causing stenoses obtained Via noninvasive fractional flow reserve) study. J Am Coll Cardiol. 2011;58:1989‐1997. [DOI] [PubMed] [Google Scholar]
- 44. Min JK, Leipsic J, Pencina MJ, et al. Diagnostic accuracy of fractional flow reserve from anatomic ct angiography. JAMA. 2012;308:1237‐1245. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45. Norgaard BL, Leipsic J, Gaur S, et al. Diagnostic performance of noninvasive fractional flow reserve derived from coronary computed tomography angiography in suspected coronary artery disease: the NXT trial (Analysis of Coronary Blood Flow Using CT Angiography: Next Steps). J Am Coll Cardiol. 2014;63:1145‐1155. [DOI] [PubMed] [Google Scholar]
- 46. Morris PD, Ryan D, Morton AC, et al. Virtual fractional flow reserve from coronary angiography: modeling the significance of coronary lesions: results from the VIRTU‐1 (VIRTUal Fractional Flow Reserve From Coronary Angiography) study. JACC Cardiovasc Interv. 2013;6:149‐157. [DOI] [PubMed] [Google Scholar]
- 47. Papafaklis MI, Muramatsu T, Ishibashi Y, et al. Fast virtual functional assessment of intermediate coronary lesions using routine angiographic data and blood flow simulation in humans: comparison with pressure wire fractional flow reserve. EuroIntervention. 2014;10:574‐583. [DOI] [PubMed] [Google Scholar]
- 48. Tu S, Barbato E, Köoszegi Z, et al. Fractional flow reserve calculation from 3‐dimensional quantitative coronary angiography and TIMI frame count: a fast computer model to quantify the functional significance of moderately obstructed coronary arteries. JACC Cardiovasc Interv. 2014;7:768‐777. [DOI] [PubMed] [Google Scholar]
- 49. Zhang J‐M, Zhong L, Luo T, et al. Simplified models of non‐invasive fractional flow reserve based on CT images. PLoS ONE. 2016;11(5):e0153070. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50. Shi C, Zhang D, Cao K, et al. A study of noninvasive fractional flow reserve derived from a simplified method based on coronary computed tomography angiography in suspected coronary artery disease. Biomed Eng Online. 2017;16:43. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51. Renker M, Schoepf UJ, Wang R, et al. Comparison of diagnostic value of a novel noninvasive coronary computed tomography angiography method versus standard coronary angiography for assessing fractional flow reserve. Am J Cardiol. 2014;114:1303‐1308. [DOI] [PubMed] [Google Scholar]
- 52. Boileau E, Nithiarasu P. One‐dimensional modelling of the coronary circulation. Application to noninvasive quantification of fractional flow reserve (FFR) In: Tavares JMRS, Jorge RN, eds. Computational and Experimental Biomedical Sciences: Methods and Applications, LN Comput. Vis. Biomech., vol. 21: Azores, Portugal: Springer International; 2015:137‐155. [Google Scholar]
- 53. Boileau E, Pant S, Roobottom C, et al. Estimating the accuracy of a reduced‐order model for the calculation of fractional flow reserve (FFR). Int J Numer Methods Biomed Eng. 2017aug;34(1):e2908. [DOI] [PubMed] [Google Scholar]
- 54. Baumann S, Wang R, Schoepf UJ, et al. Coronary CT angiography‐derived fractional flow reserve correlated with invasive fractional flow reserve measurements–initial experience with a novel physician‐driven algorithm. Eur Radiol. 2015;25:1201‐2017. [DOI] [PubMed] [Google Scholar]
- 55. Coenen A, Lubbers MM, Kurata A, et al. Fractional flow reserve computed from noninvasive CT angiography data: Diagnostic performance of an on‐site clinician‐operated computational fluid dynamics algorithm. Radiology. 2015;274:674‐683. [DOI] [PubMed] [Google Scholar]
- 56. Wang R, Renker M, Schoepf UJ, et al. Diagnostic value of quantitative stenosis predictors with coronary CT angiography compared to invasive fractional flow reserve. Eur J Radiol. 2015;84:1509‐1515. [DOI] [PubMed] [Google Scholar]
- 57. Itu L, Rapaka S, Passerini T, et al. A machine‐learning approach for computation of fractional flow reserve from coronary computed tomography. J Appl Physiol. 2016;121:42‐52. [DOI] [PubMed] [Google Scholar]
- 58. Tröbs M, Achenbach S, Röther J, et al. Comparison of fractional flow reserve based on computational fluid dynamics modeling using coronary angiographic vessel morphology versus invasively measured fractional flow reserve. Am J Cardiol. 2016;117:29‐35. [DOI] [PubMed] [Google Scholar]
- 59. Ko BS, Cameron JD, Munnur RK, et al. Noninvasive CT‐derived FFR based on structural and fluid analysis: a comparison with invasive FFR for detection of functionally significant stenosis. JACC Cardiovasc Imaging. 2017;10(6):663‐673. [DOI] [PubMed] [Google Scholar]
- 60. Blanco PJ, Bulant CA, Müller LO, et al. Comparison of 1D and 3D models for the estimation of fractional flow reserve. Sci Rep. 2018;8(1):17275 https://doi.org/10.1038 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 61. Ma Y, Liu H, Hou Y, et al. Instantaneous wave‐free ratio derived from coronary computed tomography angiography in evaluation of ischemia‐causing coronary stenosis. Medicine. 2017;96(4):e5979. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62. Yong ASC, Javadzadegan A, Fearon WF, et al. The relationship between coronary artery distensibility and fractional flow reserve. PLOS ONE. 2017jul;12(7):e0181824. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 63. Itu L, Passerini T, Badila E, et al. Image‐based computation of instantaneous wave‐free ratio from routine coronary angiography: evaluation of a hybrid decision making strategy with FFR. J Am Coll Cardiol. 2016;67(13):328. [Google Scholar]
- 64. Johnson NP, Kirkeeide RL, Asrress KN, et al. Does the instantaneous wave‐free ratio approximate the fractional flow reserve? J Am Coll Cardiol. 2013;61(13):1428‐1435. [DOI] [PubMed] [Google Scholar]
- 65. Calmac L, Niculescu R, Badila E, et al. TCT‐40 image‐based computation of instantaneous wave‐free ratio from routine coronary angiography—initial validation by invasively measured coronary pressures. J Am Coll Cardiol. 2015;66(15):B17‐B18. [Google Scholar]
- 66. Carson J. Development of a cardiovascular and lymphatic network model during human pregnancy. Ph.D. Thesis: Swansea University; 2018.
- 67. Bessems D, Rutten M, van de Vosse F. A wave propagation model of blood flow in large vessels using an approximate velocity profile function. J Fluid Mech. 2007;580:145‐168. [Google Scholar]
- 68. Smith N, Pullan A, Hunter P. An anatomically based model of transient coronary blood flow in the heart. SIAM J Appl Math. 2002;62:990‐1018. [Google Scholar]
- 69. Mynard JP, Smolich JJ. One‐dimensional haemodynamic modeling and wave dynamics in the entire adult circulation. Ann Biomed Eng. 2015;43(6):1443‐1460. [DOI] [PubMed] [Google Scholar]
- 70. Olufsen MS. Structured tree outflow condition for blood flow in larger systemic arteries. Am J Physiol. 1999;276:H257‐H268. [DOI] [PubMed] [Google Scholar]
- 71. Carson J, Loon RV. An implicit solver for 1D arterial network models. Int J Numer Methods Biomed Eng. 2016;33(7):e2837. [DOI] [PubMed] [Google Scholar]
- 72. Carson J, Lewis M, Rassi D, Loon RV. A data‐driven model to study utero‐ovarian blood flow physiology during pregnancy. Biomechanics and Modeling in Mechanobiology. 2019;18(4):1155‐1176. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 73. de Bruyne B, Bartunek J, Sys SU, Pijls NHJ, Heyndrickx GR, Wijns W. Simultaneous coronary pressure and flow velocity measurements in humans. Circulation. 1996;94(8):1842‐1849. [DOI] [PubMed] [Google Scholar]
- 74. van der Giessen AG, Groen HC, Doriot P‐A, et al. The influence of boundary conditions on wall shear stress distribution in patients specific coronary trees. J Biomech. 2011;44(6):1089‐1095. [DOI] [PubMed] [Google Scholar]
- 75. Budoff M, Nakansihi R. Noninvasive FFR derived from coronary CT angiography in the management of coronary artery disease: technology and clinical update. Vasc Health Risk Manag. 2016;12:269. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 76. Indolfi C, Mongiardo A, Spaccarotella C, et al. The instantaneous wave‐free ratio (iFR) for evaluation of non‐culprit lesions in patients with acute coronary syndrome and multivessel disease. Int J Cardiol. 2015;178:46‐54. [DOI] [PubMed] [Google Scholar]
- 77. Kobayashi Y, Johnson NP, Berry C, et al. The influence of lesion location on the diagnostic accuracy of adenosine‐free coronary pressure wire measurements. JACC Cardiovasc Interv. 2016;9(23):2390‐2399. [DOI] [PubMed] [Google Scholar]
- 78. Hennigan B, Oldroyd KG, Berry C, et al. Discordance between resting and hyperemic indices of coronary stenosis severity. Circ Cardiovasc Interv. 2016;9(11). [DOI] [PubMed] [Google Scholar]
- 79. Fede A, Zivelonghi C, Benfari G, et al. iFR‐FFR comparison in daily practice. J Cardiovasc Med. 2015;16(9):625‐631. [DOI] [PubMed] [Google Scholar]
- 80. Kanaji Y, Murai T, Lee T, et al. Efficacy of pressure parameters obtained during contrast medium‐induced submaximal hyperemia in the functional assessment of intermediate coronary stenosis. Int J Cardiol. 2016;202:207‐213. [DOI] [PubMed] [Google Scholar]
- 81. Meimoun P, Clerc J, Ardourel D, et al. Assessment of left anterior descending artery stenosis of intermediate severity by fractional flow reserve, instantaneous wave‐free ratio, and non‐invasive coronary flow reserve. Int J Cardiovasc Imaging. 2016;33(7):999‐1007. [DOI] [PubMed] [Google Scholar]
- 82. Berry C, van't Veer M, Witt N, et al. VERIFY (VERification of instantaneous wave‐free ratio and fractional flow reserve for the assessment of coronary artery stenosis severity in EverydaY practice). J Am Coll Cardiol. 2013;61(13):1421‐1427. [DOI] [PubMed] [Google Scholar]
- 83. Yamanaka F, Shishido K, Ochiai T, et al. Instantaneous wave‐free ratio for the assessment of intermediate coronary artery stenosis in patients with severe aortic valve stenosis. JACC Cardiovasc Interv. 2018;11(20):2032‐2040. [DOI] [PubMed] [Google Scholar]
- 84. Mitchell JR. Is the heart a pressure or flow generator? possible implications and suggestions for cardiovascular pedagogy. Adv Physiol Educ. 2015;39(3):242‐247. [DOI] [PubMed] [Google Scholar]
- 85. Davies JE, Whinnett ZI, Francis DP, et al. Evidence of a dominant backward‐propagating “suction” wave responsible for diastolic coronary filling in humans, attenuated in left ventricular hypertrophy. Circulation. 2006;113(14):1768‐1778. [DOI] [PubMed] [Google Scholar]
- 86. Briceno N, Lumley M, Perera D. Fractional flow reserve: conundrums, controversies and challenges. Interv Cardiol. 2015;7(6):543‐552. [Google Scholar]
- 87. van de Hoef TP, Nolte F, Rolandi MC, et al. Coronary pressure‐flow relations as basis for the understanding of coronary physiology. J Mol Cell Cardiol. 2012;52(4):786‐793. [DOI] [PubMed] [Google Scholar]
- 88. Schelbert HR. Anatomy and physiology of coronary blood flow. J Nucl Cardiol. 2010;17(4):545‐554. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 89. Lupi A, Buffon A, Finocchiaro ML, Conti E, Maseri A, Crea F. Mechanisms of adenosine‐induced epicardial coronary artery dilatation. Eur Heart J. 1997;18(4):614‐617. [DOI] [PubMed] [Google Scholar]
- 90. Ge X, Yin Z, Fan Y, Vassilevski Y, Liang F. A multi‐scale model of the coronary circulation applied to investigate transmural myocardial flow. Int J Numer Methods Biomed Eng. 2018;34:e3123. [DOI] [PubMed] [Google Scholar]