Abstract
Motion of the myocardium deteriorates the quality and quantitative accuracy of cardiac PET images. We present a method for MR-based cardiac and respiratory motion correction of cardiac PET data and evaluate its impact on estimation of activity and kinetic parameters in human subjects. Three healthy subjects underwent simultaneous dynamic 18F-FDG PET and MRI on a hybrid PET/MR scanner. A cardiorespiratory motion field was determined for each subject using navigator, tagging and golden-angle radial MR acquisitions. Acquired coincidence events were binned into cardiac and respiratory phases using electrocardiogram and list mode-driven signals, respectively. Dynamic PET images were reconstructed with MR-based motion correction (MC) and without motion correction (NMC). Parametric images of 18F-FDG consumption rates (Ki) were estimated using Patlak’s method for both MC and NMC images. MC alleviated motion artifacts in PET images, resulting in improved spatial resolution, improved recovery of activity in the myocardium wall and reduced spillover from the myocardium to the left ventricle cavity. Significantly higher myocardium contrast-to-noise ratio and lower apparent wall thickness were obtained in MC versus NMC images. Likewise, parametric images of Ki calculated with MC data had improved spatial resolution as compared to those obtained with NMC. Consistent with an increase in reconstructed activity concentration in the frames used during kinetic analyses, MC led to the estimation of higher Ki values almost everywhere in the myocardium, with up to 18% increase (mean across subjects) in the septum as compared to NMC. This study shows that MR-based motion correction of cardiac PET results in improved image quality that can benefit both static and dynamic studies.
INTRODUCTION
Motion associated with respiratory and cardiac functions is a well-known source of image quality degradation in cardiac Positron Emission Tomography (PET) studies. The magnitude of regular heart displacements is indeed substantial, particularly when viewed in comparison to the ~4–5 mm intrinsic spatial resolution of PET cameras. For example, respiration typically shifts the heart by more than 10 mm along the superior-inferior direction [1], whereas cardiac contraction moves the base of the myocardium toward the apex by an average of ~12.8 mm [2] and at the same time thickens the myocardium wall by 4.0–6.1 mm [3]. The continuous movement of the heart during scanning introduces artifacts that alter quantification of tracer concentration in cardiac tissues and deteriorate the diagnostic quality of PET images. Two types of image artifacts directly result from motion: One presents as a spatially-dependent blurring of the reconstructed activity distribution that can lead to underestimation of the local tracer uptake and even impair the detection of small lesions in some cases. The second type of artifact, resulting from spatially inconsistent emission and attenuation distributions, manifests primarily in regions adjacent to large attenuation gradients (e.g. at heart/lung or liver/lung interfaces) and can introduce artefactual myocardial defects. Its impact on image diagnostic quality has been extensively documented in cardiac PET/CT [4], [5].
Motion correction of cardiac PET has been an area of active research for more than two decades [6]–[11] (see reviews in [12], [13]). It is hoped that artifact-free PET images will translate into more accurate assessment of myocardial perfusion, metabolism and cardiac function, leading to improved diagnosis and staging of cardiovascular diseases [14], [15]. For studies performed on standalone PET or PET/CT systems, motion has generally been tackled using the following multistep framework: First, the measured emission data are gated, i.e. assigned to specific cardiac and respiratory phases based on surrogate signals measured with external sensors (e.g. EKG, respiratory bellows, optical tracking) or the PET data themselves [16]. Second, an image volume is reconstructed for each motion phase. Third, motion vector fields (i.e. dense sets of 3-D displacements) mapping the voxel-to-voxel correspondence between every motion phase and a reference phase are estimated by registration of the gated PET volumes. Fourth, the motion fields are used to correct the PET data for motion by being either applied to the gated images directly [17] or modeled within motion-compensated image reconstruction [18]. A well-known limitation of this approach, however, is the fact that motion estimation relies on gated PET volumes, which are characterized by their low signal-to-noise ratio (SNR) especially when dual (i.e. cardiac + respiratory) gating is employed. Furthermore, image registration is generally complicated by the overall lack of anatomical landmarks in cardiac PET images: for instance, certain intramural components of contractile motion (e.g. torsion) are nearly impossible to track accurately due to the uniformity of myocardium in PET images [6]. Finally, low tracer accumulation in diseased cardiac tissues could further confound local estimation of the motion field with PET images.
In addition to its clinical potential in cardiovascular imaging [19], [20], a unique feature of hybrid PET and Magnetic Resonance Imaging (MRI) is that it provides a robust solution to the problem of motion in PET. Indeed, owing to its lack of ionizing radiation, excellent soft-tissue contrast, high SNR and good spatiotemporal resolution, MRI has ideal characteristics for measuring organ motion. In simultaneous PET/MR scanners, the acquired motion information can be leveraged to correct the PET data for motion. MR-based motion correction of PET has been investigated extensively over the past few years, with applications in imaging of malignancies [21]–[30], brain [31]–[33] and heart [14], [25], [34]–[41] (refer to [42]–[46] for comprehensive reviews on the topic). In their vast majority, these works have focused on static (i.e. single time-point) PET scanning and have evaluated the impact of motion correction on various semi-quantitative image metrics such as target-to-background ratio or SUV. However, much less attention has been given to evaluating MR-based motion correction in the context of fully quantitative cardiac PET studies based on dynamic imaging.
We have previously presented and evaluated a method for MR-based motion correction of dynamic PET data in a pre-clinical porcine model of myocardial perfusion [25]. Cardiac motion correction only was performed in that study, due to the small amplitude of respiration-induced myocardium motion in this animal model. Here, we extend the method to human subjects and perform MR-guided correction of both cardiac and respiratory motion of dynamic 18F-FDG PET data. The impact of the technique on PET image quality and estimates of myocardial 18F-FDG consumption rates was evaluated in healthy subjects.
MATERIALS AND METHODS
Overview of motion correction methodology and enabling assumptions
Dynamic 18F-FDG PET and MRI data were simultaneously acquired during free-breathing in three healthy human subjects. The PET events for each dynamic frame were reconstructed in two different ways: with the proposed MR-based cardiac/respiratory motion correction (MC) method and without motion correction (NMC). Improvement in cardiac PET image quality was assessed by quantifying the apparent thickness of the myocardium wall as well as the myocardium contrast-to-noise ratio (CNR) in a late dynamic frame. The Patlak analysis method [47], [48] was applied to both MC and NMC dynamic sets to estimate voxel-wise 18F-FDG uptake rate constants (Ki) in cardiac tissues.
Overview of proposed methodology
The motion correction method for dynamic scan data comprises four key components: (1) Surrogate signals for the cardiac and respiratory cycles that are acquired continuously during the experiment and are used to assign a cardiac and respiratory phase index to every measured coincidence event. (2) A deforming cardiorespiratory tissue motion model which assigns a given cardiorespiratory motion phase (i.e. paired cardiac and respiratory phase) index to a specific 3-D deforming motion vector field. The motion field for a specific phase encodes the movement of each voxel of tissue between said phase and a chosen reference one. As detailed hereinafter, the model is constructed by means of navigator, tagging and golden-angle radial MR acquisitions. Importantly, the MRI motion tracking measurements are performed while also collecting surrogate signals, allowing us to establish the correspondence between the surrogate signals and motion fields. (3) A motion-dependent attenuation model (i.e. one attenuation map per motion phase), generated by deforming an acquired attenuation map with the motion model. (4) A motion-compensated PET reconstruction algorithm which incorporates the motion fields and motion-dependent attenuation data to produce a motion corrected image volume for each frame. Importantly, all coincidences detected during a frame are used to form the final image and therefore the reconstruction method preserves the SNR in each frame. A detailed description of each component is provided afterwards.
Enabling assumptions
Several assumptions are made to simplify the measurement of heart motion and facilitate motion correction of dynamic scan data. One, we assume that myocardium motion due to cardiac contractions is superimposed on top of the respiration motion. Accordingly, one can measure each component separately and subsequently compose the two vector fields to obtain the complete cardiorespiratory motion model. Two, we assume that the positions of tissues are the same for a given state of both the cardiac and respiratory surrogate signals at any time during the dynamic experiment. Therefore, the motion model can be used to correct any acquired event throughout the scan based on the knowledge of surrogate signals only. This implies that cardiac and respiratory motion do not change significantly during the scan and that bulk body motion is absent or negligible.
Imaging data acquisition
Imaging protocol
Figure 1 presents a schematic timeline of the imaging protocol. All scans were acquired on a Siemens Biograph mMR, a whole-body integrated 3T PET/MR system. The characteristics and imaging performance of the system have been described elsewhere [49]. Approval for the studies was obtained from the Institutional Review Board at Massachusetts General Hospital and all the subjects provided written informed consent. Study subjects were instructed to fast for 6 hours prior to the start of the experiment. At about 45 min prior to tracer administration, the subjects were instructed to drink a glucose-rich beverage to promote 18F-FDG uptake in the myocardium. Subjects were prepared for PET/MR imaging, equipped with a 3-lead Electrocardiogram (EKG) system and positioned head first supine on the scanner’s bed. Subjects were then sent in the scanner gantry after a flexible receiving surface coil was placed on the chest. Acquisition of imaging data commenced with the vendor’s MR-based attenuation correction sequence (MRAC) during a breath-hold at end-exhalation to measure an attenuation map. Shortly afterwards, a bolus of 18F-FDG (average radioactivity: ~10 mCi) was administered intravenously and a 50-min list mode PET acquisition was initiated. Subjects were instructed to relax and breath freely during the scan. At about 10 min after the start of PET acquisition, MR pulse sequences (described afterwards) were played sequentially to track motion due to cardiac contraction and respiration. Four venous blood samples were drawn at 30, 35, 40 and 45 min after tracer administration. After the scan ended, concentrations of radioactivity in the whole-blood and plasma samples were measured on a well counter.
Measurement of cardiac contractile motion
We employed a navigated, cardiac-gated MRI tagging sequence [25], [37] to measure the contractile motion of the myocardium during free breathing. “Tagging” is a magnetization preparation technique which consists in perturbing the magnetization in a spatially-dependent manner before the acquisition of MR imaging data to create temporary fiducial markers (“tags”) within tissues. Because tags result from perturbations of the magnetization of the tissue itself, changes in the shape of tags during imaging reflect the movement of the underlying tissue [50], [51]. The tagging sequence acquires line-tagged, multi-slice cardiac cine MR data sets using SPAMM [52] and a gradient recalled echo acquisition method. Imaging data acquisition is performed at a chosen respiratory phase (end-exhalation in this case) using a pencil-beam navigator intersecting the right lung-diaphragm interface in the superior-inferior direction. Each time an R-wave is detected, a navigator is measured (acquisition time: ~20 msec), and the diaphragm position is calculated; if the position is within a predefined range, the sequence proceeds to acquire the imaging data in all cardiac phases; if not, acquisition remains on standby until the next R-wave. To track motion in 3D, acquisition of line-tagged data was repeated three times, each with a different line tag orientation, corresponding to the three orthogonal Cartesian axes. Two of the three tag acquisitions were performed with transverse slices and the other with sagittal slices, ensuring that tag lines orthogonally intersect the imaging plane. Ten volumes corresponding to ten cardiac phases were acquired for each orientation with the following parameters: TE = 1.97 ms, TR = 4.8–10.2 ms, flip angle = 5–7°, in-plane resolution = 2.3×2.3 mm2, acquisition matrix = 128×104, slice thickness = 8 or 10 mm, tag line distance = 8 mm and number of slices = 10. Data acquisition for each tag line orientation took ~1.5–2 min, depending on the subject’s cardiac and respiratory rate.
Measurement of respiratory motion
We used a free-breathing, navigated multi-slice radial sequence with a golden-angle spoke acquisition scheme [28], [53] to measure respiratory motion. For each TR, the sequence first acquires a 1D slice-projection navigator oriented along the superior-inferior direction and intersecting the right lung-diaphragm interface to track the diaphragm position. Afterwards, the same spoke (i.e. with the same in-plane orientation) is acquired in all the prescribed slices. From one TR to the next, spoke orientation is incremented according to the golden-angle formula [54]. The imaging parameters were: TE = 1.53 ms, TR = 3.49 ms, flip angle = 30°, in-plane resolution = 2.3×2.3 mm2, acquisition matrix = 128×128, slice thickness = 6 mm, image plane = coronal and number of slices = 24. We let the sequence ran until a total of 5120 spokes per slice were acquired for a total acquisition time of ~7.5 min. During data post-processing, the navigator, acquired for each TR, was used to assign a respiratory phase index to each acquired spoke. The golden-angle encoding scheme provides pseudo-uniform coverage of the k-space for an arbitrary number of acquired spokes, thereby enabling flexible a posteriori binning of spokes into respiratory bins.
Data processing
Tagged MRI
For each subject, the tag data sets comprise three image volumes (two transverse and one sagittal), each with orthogonal tag line directions, for every cardiac phase. To estimate cardiac motion in 3D, the three volumes for each phase were re-oriented and resampled onto the same spatial grid (isotropic 2×2×2 mm3 voxels) and then summed. This procedure yielded ten cardiac volumes tagged along all three Cartesian axes. Note that the tagged volumes depict tissues in the end-exhalation respiratory phase.
Golden-angle radial MRI
The following procedure was employed to process the golden-angle MR data: First, we used the slice-projection navigators to obtain a respiratory trace by tracking the diaphragm position over time. The trace was obtained by selecting a reference navigator signal and by calculating Pearson correlation coefficients between the reference and all other time points. We then applied a previously proposed histogram-guided navigator sorting technique [28] to the trace to assign the acquired spokes to five respiratory bins. The sorting method adapts the amplitude of each bin to obtain approximately the same number of spokes in each bin. Next, iterative reconstruction of the MR data was performed for each bin. Let denote the set of spokes corresponding to respiratory bin (or ‘phase’) rk, k ∈ [1; R] (R=5 in this study). The reconstruction of each slice in each phase was performed by preconditioned conjugate gradient-based minimization of a cost-function comprised of a data consistency term including the non-uniform fast Fourier transform (NUFFT) operator [55], and a total-variation regularization term that helps reduce streak artifacts and noise in the reconstructed images [56]:
Eq. 1 |
where denotes the 2-D MR image for respiratory phase rk, F is the NUFFT operator, C are the estimated 2-D coil sensitivity maps, ‖∗‖1 and ‖∗‖2 denote the L-1 and L-2 norms, respectively, S is the total-variation operator [57] and β is a regularization parameter (scalar). This procedure yielded five respiratory-resolved MR volumes that were subsequently used to determine the respiratory motion fields. Note that because cardiac gating was not applied during data reconstruction, the reconstructed MR volumes represented an average cardiac position.
Surrogate signals
As mentioned earlier, the motion correction method relies on surrogate signals to monitor the respiratory and cardiac cycles of the subject during the dynamic scan. EKG is a straightforward and robust approach to monitor the cardiac cycle. Throughout the PET acquisition, the scanner’s physiological monitoring unit inserts a trigger event in the list mode data stream each time an R-wave is detected by the EKG sensor. These trigger events can then be retrieved in the list mode file and a specific cardiac phase can be assigned to each event based on its relative time of arrival with respect to a directly preceding R-wave event. Monitoring of respiration, however, is not as straightforward: for instance, MR navigators are considered as one of the most accurate approaches to monitoring breathing, since unlike respiratory bellows or other external sensors, they can directly measure the position of internal organs (e.g. diaphragm) with high temporal resolution. The disadvantage of using navigators, however, is the fact that their measurement interferes with the MR acquisition and it is therefore impractical to use navigators for continuously monitoring respiration during a long PET acquisition. In this study, we opted for a previously proposed list mode-driven approach to monitor respiration during the dynamic scan [16], and directly compared the method to diaphragmatic navigator acquisitions. First, we obtained a “raw” respiratory curve by counting the total number of coincidences detected within a non-overlapping 100-ms sliding window. Because rapid changes in tracer distribution during the early frames of the scan can deteriorate the quality of the gating signal, this procedure was only applied to events detected after the first-pass delivery of 18F-FDG, i.e. starting from 1 min after tracer injection. The respiratory trace was obtained by filtering the raw curve using a band-pass filter with cutoff frequencies at [0.1–0.4] Hz. The cutoffs were selected to remove slow trends in the data (e.g. radioactive decay, tracer washout) as well as high frequency components (e.g. counting noise, cardiac contraction), while at the same time preserving the expected frequency range of respiratory fluctuations.
PET data
The list mode data for each subject were sorted into a series of 20 time frames (8 × 15 sec, 4 × 60 sec and 8 × 300 sec) covering a total of 46 min acquisition. The prompt events for each of the last sixteen frames (i.e. after 18F-FDG first-pass) were binned into 10 cardiac phases and 5 respiratory phases using the surrogate signals (i.e. EKG and list mode-driven respiratory curve), resulting in a total of 50 sinograms per frame. Cardiac phases were defined by dividing each cardiac cycle (i.e. R-R interval) in temporal bins of equal duration. Respiratory phases were defined by applying the histogram-guided sorting method to the list mode-driven respiratory trace [28]. A sinogram of delayed events for each frame was also obtained for correction of random coincidences.
Calculation of cardiorespiratory motion model
The tagging and golden-angle MR volumes were used to measure motion due to cardiac and respiratory functions, respectively. For both data sets, motion field estimation was performed using a nonrigid registration method based on a 3-D tensor product B-spline parametric model [58], [59]. Given two image volumes to register, the algorithm seeks to estimate a diffeomorphic nonrigid transformation by conjugate gradient-based optimization of a cost-function comprised of a standard sum-of-squared-differences data consistency term and a quadratic regularization term. To improve convergence speed and robustness, the algorithm employs a bilevel multiresolution scheme in both image and transformation domains. For both tag and golden-angle MR data sets, rather than directly registering the volume of each phase to the reference phase, we registered consecutive pairs of volumes for estimation of inter-phase (or “partial”) motion fields. The reference phases for cardiac and respiratory motion were chosen as end-diastole and end-exhalation, respectively. The complete cardiac and respiratory motion fields were computed by accumulating the displacements from the partial fields. Let cn, n ∈ [1; C] and rk, k ∈ [1; R] denote indices for cardiac and respiratory phases, respectively, and vector denote a cardiorespiratory phase. For simplicity, end-diastole and end-exhalation (i.e. reference phases for cardiac and respiration motion) are indexed with n, k = 1. The procedure described earlier yielded two distinct fields:
, the cardiac displacement field estimated using tagged MRI data, which encodes the 3-D displacement of each voxel between cardiorespiratory phases and .
, the respiratory displacement field estimated using the golden-angle MRI data, which encodes the 3-D displacement of each voxel between respiratory phases r1 and rk. Note that since no cardiac gating was used during data acquisition and reconstruction, these MR volumes represent an average cardiac position. However, we assume that the estimated respiratory motion field apply to any cardiac phase.
We obtain the complete cardiorespiratory motion model by first resampling the two fields on a common voxel grid and then sum the results:
Eq. 2 |
The combined motion field model Dresp+card describes the 3-D displacement of each voxel between the end-diastolic, end-exhalation phase and all other cardiorespiratory phases. The fields were then converted to a set of 3-D image warping operators {W1→m}, m ∈ [1, M = R × C] for incorporation within PET reconstruction.
Attenuation correction
The vendor’s MRAC method was used to acquire an attenuation map during an end-exhalation breath-hold. MRAC employs Dixon acquisition and tissue segmentation to produce a 4-classes attenuation map (air, lung, fat, soft-tissue). The following procedure was used to generate a motion-dependent attenuation map from breath-hold MRAC scans: First, we non-rigidly registered the water volume produced by MRAC to the end-exhalation golden-angle MR volume using mutual-information based B-spline registration. Second, the estimated transformation was applied to the 4-classes attenuation map to ensure spatial consistency between the motion model and the attenuation map. Third, the cardiorespiratory warping operators were used to deform the resulting attenuation map to all other motion phases.
PET reconstruction
The dynamic PET data for each subject were reconstructed with and without motion correction on a frame-by-frame basis. As mentioned earlier, the prompt events detected in each of the last sixteen dynamic frames were binned into cardiac and respiratory phases, yielding a total of M=50 sinograms per frame. Let denote the sinogram of prompt events with I bins in dynamic frame f and motion phase m ∈ [1; M]. It is assumed that the elements of are realizations of statistically independent Poisson random variables with expected values . Accordingly, the forward PET model for multi-phase sinogram data in a given temporal frame is:
Eq. 3 |
where , and all are sinograms of expected prompt, scattered and random coincidences, respectively, is the motion-corrected image (i.e. distribution of 18F-FDG at end-diastole/end-exhalation) with J voxels that we seek to estimate from the sinograms, is an image warping operator that deforms the PET image in the reference phase to phase m (note that W1→1 = I), is the geometric probability matrix, is a diagonal matrix containing the attenuation coefficients for phase m, is a diagonal matrix containing the sensitivity coefficient for each sinogram bin, and is a scalar corresponding to the ratio between the number of events detected in phase m and the total number of events detected in the frame. Image reconstruction for each frame was performed by maximizing the corresponding Poisson log-likelihood density function using ordered subset expectation maximization [60]. For reconstruction without motion correction (NMC), the forward model in Eq. 3 was modified by setting and Am = A1 (i.e. no motion warping and motion dependent attenuation correction performed). For both MC and NMC reconstructions, sinograms of scattered coincidences were calculated using the single scatter simulation technique [61] and sinograms of randoms using the smoothed delayed window approach. The distributions of scatters and randoms were assumed to be independent of motion, i.e. and , where and represent scatter and random sinograms estimated using all events in the frame regardless of motion. All PET images were reconstructed on a 2.08×2.08×2.03 mm3 voxel grid with 4 iterations and 12 subsets.
PET images analyses
Evaluation of image quality
For each subject, we evaluated the quality of MC and NMC images by comparing myocardium contrast-to-noise ratio (CNR) and apparent myocardial wall thickness in a late (40 min after injection) 5-min frame. Activity images were first re-sliced in short axis orientation and five evenly spaced planes sampling the heart from apex to base were selected. For each plane, four regions of interest (ROI) were manually positioned in the septal, lateral, anterior and inferior walls (see Figure 2 A). A spherical ROI (4 mm radius) was also drawn in the mid left-ventricular blood cavity. Next, CNR [9] was calculated for each ROI as:
Eq. 4 |
where Mean and SD are the voxel-wise mean and standard deviation of activity values in a ROI, MyocROI and LVROI refer to ROIs taken in the myocardium and left-ventricle blood bool respectively. To measure the apparent thickness of the myocardium in septal, lateral, anterior and inferior walls, we extracted line profiles in short axis (SA), horizontal long axis (‘HLA’, approximating the four-chamber view [62]) and vertical long axis (‘VLA’, approx. the two-chamber view) planes, as illustrated in Figure 2 B. More specifically, we extracted 20 profiles along the septal, lateral, anterior and inferior walls in 5 SA slices spread-out evenly over the length of the myocardium, 10 profiles in the septal and lateral walls in the selected HLA plane, and 10 profiles in the anterior and inferior walls in the VLA plane. The exact same slices were selected for MC and NMC images. This procedure yielded a total of 40 profiles for each volume. Thickness of the wall was quantified as the full-width at half maximum of a 1-D Gaussian function fitted to each profile.
Calculation of myocardial 18F-FDG uptake rates
Parametric images of net 18F-FDG influx rate (Ki, with units of min−1) were calculated by applying the Patlak plot method [63] to the last 40 min of the dynamic acquisition. The first frames were omitted during the analysis to obtain steady-state equilibrium between the reversible and plasma compartments. To compute input functions for kinetic modeling, we extracted a whole-blood time-activity curve (TAC) in a spherical ROI (4 mm radius) located in the base of the left ventricle blood pool in both MC and NMC image sets. Next, arterial plasma input functions were computed by scaling whole-blood TACs to the plasma activity concentrations measured in the four blood samples. Patlak analysis was then applied on a voxel-wise basis for both MC and NMC dynamic sets using the corresponding input functions. The operational equation for the Patlak model is defined according to:
Eq. 5 |
where CT(t) and Cp(t) are instantaneous concentrations of activity in myocardium tissue and arterial plasma, respectively, and Vb is the intercept. Parameters of the Patlak model were estimated for each voxel by linear least-square regression. The resulting parametric images were resliced to short-axis view using the transformation parameters applied to the activity images. Average Ki values were then calculated in the previously defined 20 myocardium ROIs. According to Eq. 5, an increase in reconstructed myocardium activity concentration following motion correction should translate into higher Ki values.
Statistical Analysis
All statistical tests were performed using a two-tailed paired t-test. A p-value lower than 0.05 was considered significant.
RESULTS
Figure 3 presents examples of images and displacement fields obtained using tagging and golden-angle radial MR acquisitions for various cardiac and respiratory phases. The motion fields estimated with tagged MRI (Figure 3 A) depict the movement of the myocardium between the reference end-diastolic phase (C1) and all other cardiac phases. Note that due to T1 recovery, the contrast of the tag lines progressively decreases towards the end of the cardiac cycle, which may impact the accuracy of motion estimation in the late cardiac phases. The displacement fields overlaid on top of the respiratory-gated radial MRI slices (Figure 3 B) show the movement of thoracic tissues, including heart and liver, between the reference end-exhalation phase (R1) and other respiratory phases. Note that non-rigid registration techniques, such as the one employed in this work cannot guarantee accurate estimation of motion in areas with uniform intensities such as the left-ventricle blood pool.
Figure 4 compares the traces obtained using the list mode-driven method and the MR navigator measured during the golden-angle radial scan for all three subjects. There was a strong correlation between the navigator and list mode-driven signals for all subjects, in agreement with findings from other groups [64]. Figures 5 and 6 display five evenly spaced short-axis and horizontal long-axis images of a late dynamic frame reconstructed with MC and NMC for subject 1 and 2, respectively. MC images showed visibly higher spatial resolution than NMC images as evidenced by the improved delineation of papillary muscles (red arrows), higher myocardium wall signal (white arrows), reduced spillover of activity from the myocardium to the left-ventricle blood pool (yellow arrows) and thinner wall. Visually, the impact of motion correction was most substantial in the septum, especially at the basal level.
Table 1 presents CNR and wall thickness values extracted in MC and NMC images for different sectors of the myocardium for all three subjects. Figure 7 plots percent differences in myocardium CNR and thickness between MC and NMC (i.e. ). For all subjects, MC yielded significantly higher myocardium CNR (all p<0.05) and lower apparent wall thickness than NMC (all p<0.05). CNR increased by an average of up to 20.1% in the septal wall and 16.4% in the anterior wall for MC compared to NMC. Apparent wall thickness decreased by an average of 18.9% in the inferior wall and 18.4% in the septal wall as compared to NMC. Overall, the impact of motion correction was lower in subject 3 than in the other two subjects. One possible reason for this is the lower quality of the list mode-based respiratory gating curve for this subject, which was noisier and exhibited a weaker correlation with the navigator-based signal. The lower performance of event-based gating may have been caused by the relatively lower apparent 18F-FDG myocardium uptake in subject 3 compared with the other subjects.
Table 1.
Subject no. | CNR | Thickness (mm) | |||
---|---|---|---|---|---|
MC | NMC | MC | NMC | ||
1 | inferior | 16.2±2.3 | 14.7±2.1 | 15.5±3.9 | 18.7±4.3 |
lateral | 21.0±3.7 | 19.2±4.3 | 14.7±2.7 | 15.4±2.6 | |
anterior | 17.8±2.8 | 14.9±2.6 | 11.8±2.9 | 14.1±3.4 | |
septal | 14.6±2.4 | 12.0±1.2 | 13.6±2.8 | 16. 2±4.5 | |
2 | inferior | 17.3±0.8 | 16.1±1.4 | 16.1±4.4 | 19.3±5.8 |
lateral | 22.9±3.5 | 21.1±4.9 | 14.8±4.5 | 15.7±4.5 | |
anterior | 19.9±2.9 | 16.9±4.9 | 11.5±1.6 | 13.6±2.8 | |
septal | 17.3±0.9 | 14.0±1.8 | 14.8±3.2 | 15.8±3.8 | |
3 | inferior | 6.9±0.6 | 6.4±0.7 | 16.2±3.1 | 20.2±7.1 |
lateral | 9.0±0.85 | 8.5±1.1 | 16.7±4.8 | 17.5±5.7 | |
anterior | 8.8±1.4 | 7.9±0.7 | 15.1±1.9 | 16.9±4.1 | |
septal | 5.1±0.7 | 4.5±1.2 | 15.6±3.4 | 19.4±6.7 |
Figure 8 shows the same transverse slice in the various dynamic frames reconstructed using NMC and MC methods for subject 2. Supplementary Figure 1 presents the same data for subject 1. In general, we found that for the frames acquired up to ~4 min after administration of 18F-FDG (frame #5 to 10), MC led to a slightly lower reconstructed myocardium activity concentration as compared to NMC, which is consistent with reduced spillover effects from the blood cavity to the myocardium. When the activity distribution is uniform (e.g. frame #11 and 12), with similar blood and myocardium signals, MC does not seem to have much effect, which is likewise consistent with expectations. The impact of the correction becomes more pronounced for the later frames (e.g. frame #13–20) as the tracer accumulates in the myocardium and the contrast between myocardium and blood increases.
Parametric images of 18F-FDG influx rate (Ki, with units of min−1) were calculated as described in Methods for all three subjects. Figures 9 and 10 display short and horizontal long-axis Ki slices computed with MC and NMC data for subjects 1 and 2, respectively. Note that the presented slices correspond to those shown in Figures 5 and 6. As can be seen, MC dynamic data yielded Ki maps with overall higher values in the myocardium and improved spatial resolution as compared to NMC images. Figure 11 shows examples of Patlak plots for MC and NMC in one ROI taken in the basal section of the septum, where motion correction was found to have the most impact.
Table 2 shows MC and NMC Ki values for the different myocardium regions in all three subjects. Figure 12 displays box plots of Ki percent differences between MC and NMC. Motion correction yielded significantly higher Ki values for all subjects as compared to NMC (all p<0.05). Higher Ki values were obtained everywhere, especially in the septum where average Ki increased by 18% as compared to NMC.
Table 2.
Ki (min−1) | |||
---|---|---|---|
Subject no. | MC | NMC | |
1 | inferior | 0.031±0.004 | 0.029±0.004 |
lateral | 0.038±0.007 | 0.035±0.007 | |
anterior | 0.031±0.005 | 0.027±0.005 | |
septal | 0.027±0.002 | 0.023±0.001 | |
2 | inferior | 0.034±0.002 | 0.029±0.006 |
lateral | 0.035±0.006 | 0.030±0.004 | |
anterior | 0.035±0.006 | 0.029±0.009 | |
septal | 0.033±0.006 | 0.027±0.010 | |
3 | inferior | 0.050±0.005 | 0.048±0.005 |
lateral | 0.063±0.007 | 0.061±0.010 | |
anterior | 0.064±0.011 | 0.060±0.006 | |
septal | 0.037±0.008 | 0.033±0.008 |
DISCUSSION
Motion of the myocardium deteriorates the spatial resolution and quantitative accuracy of cardiac PET images. We developed a methodology for MRI-based cardiac and respiratory motion correction of PET data and evaluated its impact on image quality as well as quantification of myocardial 18F-FDG uptake rates in three healthy subjects. The results obtained suggest that the proposed correction method alleviates motion artifacts in PET images, resulting in visibly improved spatial resolution, increased recovery of activity in the myocardium and reduced spillover of activity from the myocardium to the left ventricle cavity. As a result, significantly higher CNR (up to 20.1% mean CNR increase in septum across subjects) and lower apparent wall thickness (up to 18.9% mean thickness decrease in inferior wall) were obtained in the resulting motion corrected images as compared to non-corrected ones. Likewise, motion correction yielded parametric images of 18F-FDG uptake rates (Ki) with improved spatial resolution. The correction led to the estimation of higher Ki values almost everywhere in the myocardium, with up to 18% increase in septal Ki (mean across subjects) as compared to non-corrected data. Overall, the impact of motion correction was the greatest in the basal regions of the myocardium in all subjects.
A key component of the correction technique is the recording of respiration as a function of scanning time. In this study, we used a previously proposed sensitivity-based event-driven method to derive a surrogate signal for the respiratory cycle. In agreement with works from other groups [22], a strong correlation was obtained for all subjects between the obtained respiratory curves and diaphragm position readouts using navigators (see Figure 4). Therefore, the event-driven signal was considered accurate enough to bin PET events into the different respiratory phases. It is important to highlight that, in a dynamic PET scan, the raw signal extracted using the sensitivity method (i.e. number of counts as a function of time) will be correlated not only to respiratory and cardiac motion, but also to decay and tracer kinetics. The latter factor can produce either high or low frequency components in the signal depending on timing with respect to tracer injection: typically, it is expected that tracer kinetics will introduce high frequency changes in early frames (first-pass of the tracer) and lower frequency contributions (e.g. tracer washout) in later frames. In this study, to minimize the influence of tracer kinetics, event-driven gating was applied after the first-pass of 18F-FDG and band-pass Fourier filtering was used to isolate the respiratory component. Further investigations are needed to study the accuracy of event-driven gating methods in early frames, especially in the context of cardiac blood flow studies.
We employed a combination of sequential tag and golden-angle MR acquisitions to obtain a cardiorespiratory motion field for each subject. MR tagging is considered as one of the most accurate approaches to measure cardiac contractile motion in that it provides “features” in cardiac tissues that non-tagged MRI lacks and thereby facilitates estimation of intramural motion. Another advantage of tagging is that it naturally offers contrast between heart muscle and blood cavity, as tagging of blood protons rapidly vanishes with blood circulation within the ventricles during the cardiac cycle. A drawback of 3D MR tagging, however, is the long time needed to acquire the data. Here, to limit scan times, the acquisition of the three tagged MRI data sets was performed using a relatively coarse slice resolution (8–10 mm), which will necessarily impact the accuracy of the motion fields. Parallel imaging or compressed sensing could be employed to drastically accelerate data acquisition without significantly degrading motion field accuracy [37]. T1 recovery-induced tag line fading, which can be appreciated by inspecting the tagged images in Figure 3 A, is another well-known limitation of tagged MRI that can lead to a deterioration of the quality of the motion field estimates, especially in the late cardiac phases. Nevertheless, since the largest displacements occur during the first phases of the cardiac cycle which exhibit good tag contrast, the impact of fading on the motion corrected PET images should remain relatively small. Moreover, our tagged MRI sequence assumes that the subject’s respiratory phase, determined by a navigator readout prescribed immediately after each R-wave, does not change during the R-R interval length whose duration is typically ~0.7–1.0 sec in resting conditions. In reality, however, the respiratory phase may change during this time period, which can introduce respiratory motion artifacts, especially towards the end of the cardiac cycle. Lastly, in the current version of the sequence, the duration of the data acquisition window after R-wave detection is fixed and must be pre-defined by the operator based on the average cardiac cycle rate of the subject. Therefore, difficulties could arise when imaging a patient with a very irregular cardiac cycle.
Respiratory motion was tracked in each subject using a navigated, golden-angle based radial MR sequence. As mentioned earlier, we let the sequence run until a total of 5,120 spokes per slice (and thus 5,120 navigators) were acquired, leading to a ~7.5 min scan. All acquired spokes were then binned into 5 respiratory phases based on the navigator-measured diaphragm position. As a result, the reconstruction of each slice in each respiratory phase was performed using ~1,000 spokes, which is about 5-fold higher than the number of spokes prescribed by the Nyquist sampling criterion. Therefore, it is important to underline that while we deliberately chose to acquire a large number of spokes in this study in order to obtain a robust estimate of the correlation between the navigator and event-driven respiratory curves, the scan time for measuring respiratory motion using this sequence could be drastically shortened in practice. Furthermore, because the main components of respiratory motion are along head-feet and antero-posterior directions [65], the use of sagittally-oriented slices should be more optimal for capturing respiratory motion. Here, we chose to acquire coronal rather than sagittal orientation because it allows for imaging, and hence motion estimation, of the whole thorax using a relatively smaller number of slices, while keeping the primary motion component (head-feet) within the imaging plane. It is indeed important that the cardiorespiratory motion model includes not only the heart but also other moving tissues (e.g. liver) within the PET field-of-view for accurate motion correction of projection measurements. Finally, it should be noted that irregular breathing patterns and large variations in respiratory motion amplitude would make the measurement and correction of respiratory motion more challenging. Further investigations are needed, especially in patients with cardiovascular diseases, to understand how the current approach should be modified to account for such effects.
The proposed motion tracking and modeling method relies on the assumption that cardiac motion is superimposed on top of respiratory motion and that the two motions can therefore be measured separately. A limitation of this approach is that it does not account for possible variations in cardiac motion patterns in the different respiratory phases. Furthermore, it is known that respiratory motion can show hysteretic effects in some subjects [66], [67], whereby the same diaphragm position corresponds to different heart positions at inspiration and expiration. Although this has not been explored in the present study, it is in principle possible to account for hysteresis by assigning the PET and MR data to different phases according to the respiratory state (i.e. inspiration or expiration).
Although motion correction generally led to a decrease in apparent wall thickness and an increase in CNR and Ki as compared to no motion correction, it should be noted that this was not the case everywhere, as in some areas, MC images exhibited similar or even slightly lower CNR and Ki values as well as greater apparent wall thickness than NMC images (see box plots in Figure 7 and 12). A possible reason for this might be local errors in the cardiorespiratory motion field that may deteriorate the quality of the correction in relevant areas.
In general, by alleviating spillover from myocardium to blood and vice-versa in reconstructed PET images, motion correction (MC) is expected to reduce bias in estimates of activity concentration and kinetic parameters. In this population of healthy subjects, motion correction led to an increase in reconstructed myocardium activity in the frames used for kinetic modeling, which resulted in the estimation of overall higher Ki values as compared to non-corrected data. However, it should be noted that motion correction of dynamic cardiac PET scan data may not always translate into an increase in estimated Ki values. For instance, in diseased myocardium with low tracer uptake, one expects motion corrected images to exhibit lower activity concentration in the defect as compared to non-corrected images [35], which might lead to the estimation of lower Ki values in relevant segments [41].
To evaluate the relative impact of respiratory and cardiac motion correction, Supplemental Figures 2–5 compare images of activity and Ki obtained without motion correction to those obtained by applying respiratory-only, cardiac-only and ‘dual’ (i.e. respiratory and cardiac, as used throughout the paper) motion correction. Overall, we found that correcting respiratory or cardiac motion effects alone lead to noticeable improvement in image quality as compared to no motion correction. By visual inspection of the estimated activity and Ki images, it is clear that correction of both cardiac and respiratory motion had the most impact on image quality (e.g. compare delineation of papillary muscles and septal wall in left-most HLA slice in Supplemental Figures 2 and 4).
Long dynamic scans are associated with an increased likelihood of body movement during the acquisition, especially in patients with cardiac disease. Body motion is unpredictable and can occur for numerous reasons including subject discomfort, coughing or deep breathing. Such motion can have a severe impact on image quality and confound kinetic parameter estimation [68], [69]. In principle, estimates of body motion (e.g. obtained using registration of dynamic PET images [70]) could be employed to adapt the cardiorespiratory motion model to each dynamic frame based on patient position. This will be part of future technical developments.
Another potential issue that comes along with extended PET scans is the possibility of changes in cardiac and respiratory motion patterns that could result in inaccurate modeling and correction of motion for some portions of the scan. Further works are needed to investigate how the motion model could be modified to account for such variation. Nevertheless, the fact that motion correction led to estimation of parametric images with improved resolution suggests that the motion model was relatively accurate for the whole dynamic scan in this study.
Myocardial perfusion imaging protocols typically involve the acquisition of two PET scans, one in resting conditions and the other during pharmacological stress. The administration of vasodilators (e.g. regadenoson) for stress imaging will cause an increase heart rate and may also substantially change the breathing pattern of the patient [11]. Therefore, it is possible that distinct cardiorespiratory motion models need to be obtained for both scanning conditions, although this warrants further investigations.
The number of subjects in this study was small as we only scanned three subjects. Another limitation of this work is the lack of motion-free reference standard for validation of the motion correction methodology. Indeed, due to the drastic increase in noise associated to rejecting events in dynamic frames, we did not apply kinetic modeling to dual-gated PET images. Nevertheless, comparisons of motion corrected vs. non-corrected results could demonstrate the effectiveness of the method and its impact on activity and kinetic parameter quantification. However, further investigations in patients with cardiac diseases are needed to determine whether motion correction improves the diagnostic value of cardiac PET.
CONCLUSION
In this work, we presented a method for MR-based cardiac and respiratory motion correction of cardiac 18F-FDG PET data and evaluated its impact on quantification of activity and kinetic parameters in healthy volunteers. The method was shown to substantially improve the spatial resolution and contrast of cardiac 18F-FDG images. Motion correction yielded parametric images with improved spatial resolution and produced a significant increase in Ki estimates, consistent with increased reconstructed myocardial activity in the frames used during kinetic analyses.
Supplementary Material
ACKNOWLEDGMENTS
This work was supported in part by grants R01-HL118261, P41-EB022544 and T32-EB013180 from the National Institutes of Health. The authors would like to thank Julia-Ann Scotton, Nicolas Guehl, Shirley Hsu and Grae Arabasz for their help with the data acquisition.
REFERENCES
- [1].Klein GJ, Reutter RW, and Huesman RH, “Four-dimensional affine registration models for respiratory-gated PET,” IEEE Transactions on Nuclear Science, vol. 48, no. 3, pp. 756–760, 2002. [Google Scholar]
- [2].Rogers WJ et al. , “Quantification of and correction for left ventricular systolic long-axis shortening by magnetic resonance tissue tagging and slice isolation,” Circulation, vol. 84, no. 2, pp. 721–731, 1991. [DOI] [PubMed] [Google Scholar]
- [3].Fisher M, von Schulthess G, and Higgins C, “Multiphasic cardiac magnetic resonance imaging: normal regional left ventricular wall thickening,” American Journal of Roentgenology, vol. 145, no. 1, pp. 27–30, Jul. 1985. [DOI] [PubMed] [Google Scholar]
- [4].Gould KL, Pan T, Loghin C, Johnson NP, Guha A, and Sdringola S, “Frequent diagnostic errors in cardiac PET/CT due to misregistration of CT attenuation and emission PET images: a definitive analysis of causes, consequences, and corrections,” Journal of Nuclear Medicine, vol. 48, no. 7, pp. 1112–1121, 2007. [DOI] [PubMed] [Google Scholar]
- [5].Loghin C, Sdringola S, and Gould KL, “Common artifacts in PET myocardial perfusion images due to attenuation–emission misregistration: clinical significance, causes, and solutions,” Journal of Nuclear Medicine, vol. 45, no. 6, pp. 1029–1039, 2004. [PubMed] [Google Scholar]
- [6].Klein GJ and Huesman RH, “Four-dimensional processing of deformable cardiac PET data,” Medical Image Analysis, vol. 6, no. 1, pp. 29–46, 2002. [DOI] [PubMed] [Google Scholar]
- [7].Dawood M, Gigengack F, Jiang X, and Schäfers KP, “A mass conservation-based optical flow method for cardiac motion correction in 3D-PET,” Medical physics, vol. 40, no. 1, p. 12505, 2013. [DOI] [PubMed] [Google Scholar]
- [8].Lamare F et al. , “Evaluation of respiratory and cardiac motion correction schemes in dual gated PET/CT cardiac imaging,” Medical physics, vol. 41, no. 7, p. 72504, 2014. [DOI] [PubMed] [Google Scholar]
- [9].Slomka PJ et al. , “Dual-gated motion-frozen cardiac PET with flurpiridaz F 18,” Journal of Nuclear Medicine, vol. 56, no. 12, pp. 1876–1881, 2015. [DOI] [PubMed] [Google Scholar]
- [10].Le Meunier L et al. , “Motion frozen 18F-FDG cardiac PET,” Journal of nuclear cardiology, vol. 18, no. 2, pp. 259–266, 2011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [11].Yu Y et al. , “Event-by-Event Continuous Respiratory Motion Correction for Dynamic PET Imaging,” Journal of Nuclear Medicine, p. jnumed. 115.167676, 2016. [DOI] [PubMed] [Google Scholar]
- [12].Rahmim A, Tang J, and Zaidi H, “Four-dimensional image reconstruction strategies in cardiac-gated and respiratory-gated PET imaging,” PET Clinics, vol. 8, no. 1, pp. 51–67, 2013. [DOI] [PubMed] [Google Scholar]
- [13].Slomka PJ, Pan T, and Germano G, “Imaging moving heart structures with PET,” Journal of nuclear cardiology, pp. 1–5, 2015. [DOI] [PubMed] [Google Scholar]
- [14].Wang X, Rahmim A, and Tang J, “MRI-assisted dual motion correction for myocardial perfusion defect detection in PET imaging,” Medical Physics, vol. 44, no. 9, pp. 4536–4547, Sep. 2017. [DOI] [PubMed] [Google Scholar]
- [15].Tang J, Wang X, Gao X, Segars WP, Lodge MA, and Rahmim A, “Enhancing ejection fraction measurement through 4D respiratory motion compensation in cardiac PET imaging,” Phys. Med. Biol, vol. 62, no. 11, p. 4496, 2017. [DOI] [PubMed] [Google Scholar]
- [16].Büther F et al. , “List mode-driven cardiac and respiratory gating in pet,” Journal of Nuclear Medicine, vol. 50, no. 5, pp. 674–681, 2009. [DOI] [PubMed] [Google Scholar]
- [17].Bai W and Brady M, “Motion correction and attenuation correction for respiratory gated PET images,” Medical Imaging, IEEE Transactions on, vol. 30, no. 2, pp. 351–365, 2011. [DOI] [PubMed] [Google Scholar]
- [18].Qiao F, Pan TS, Clark JW Jr, and Mawlawi OR, “A motion-incorporated reconstruction method for gated PET studies,” Phys. Med. Biol, vol. 51, pp. 3769–3783, 2006. [DOI] [PubMed] [Google Scholar]
- [19].Robson PM et al. , “MR/PET Imaging of the Cardiovascular System,” JACC: Cardiovascular Imaging, vol. 10, no. 10 Part A, pp. 1165–1179, Oct. 2017.28982570 [Google Scholar]
- [20].Nappi C and El Fakhri G, “State of the art in cardiac hybrid technology: PET/MR,” Current cardiovascular imaging reports, vol. 6, no. 4, pp. 338–345, 2013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [21].Chun SY et al. , “MRI-based nonrigid motion correction in simultaneous PET/MRI,” Journal of Nuclear Medicine, vol. 53, no. 8, pp. 1284–1291, 2012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [22].Guerin B et al. , “Nonrigid PET motion compensation in the lower abdomen using simultaneous tagged-MRI and PET imaging,” Medical physics, vol. 38, p. 3025, 2011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [23].Furst S et al. , “Motion correction strategies for integrated PET/MR,” Journal of Nuclear Medicine, vol. 56, no. 2, pp. 261–269, 2015. [DOI] [PubMed] [Google Scholar]
- [24].Petibon Y et al. , “Relative role of motion and PSF compensation in whole-body oncologic PET-MR imaging,” Medical Physics, vol. 41, no. 4, 2014. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [25].Petibon Y et al. , “Impact of motion and partial volume effects correction on PET myocardial perfusion imaging using simultaneous PET-MR,” Physics in medicine and biology, vol. 62, no. 2, p. 326, 2016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [26].Rank CM et al. , “Respiratory motion compensation for simultaneous PET/MR based on highly undersampled MR data,” Medical Physics, vol. 43, no. 12, pp. 6234–6245, Dec. 2016. [DOI] [PubMed] [Google Scholar]
- [27].Fuin N et al. , “Concurrent Respiratory Motion Correction of Abdominal PET and Dynamic Contrast-Enhanced–MRI Using a Compressed Sensing Approach,” J Nucl Med, vol. 59, no. 9, pp. 1474–1479, Sep. 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [28].Dutta J, Huang C, Li Q, and El Fakhri G, “Pulmonary imaging using respiratory motion compensated simultaneous PET/MR,” Medical physics, vol. 42, no. 7, pp. 4227–4240, 2015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [29].Fayad H, Schmidt H, Wuerslin C, and Visvikis D, “Reconstruction-incorporated respiratory motion correction in clinical simultaneous PET/MR imaging for oncology applications,” Journal of Nuclear Medicine, vol. 56, no. 6, pp. 884–889. [DOI] [PubMed] [Google Scholar]
- [30].Grimm R et al. , “Self-gated MRI motion modeling for respiratory motion compensation in integrated PET/MRI,” Medical Image Analysis, vol. 19, no. 1, pp. 110–120, 2015. [DOI] [PubMed] [Google Scholar]
- [31].Huang C, Ackerman JL, Petibon Y, Brady TJ, El Fakhri G, and Ouyang J, “MR-based motion correction for PET imaging using wired active MR microcoils in simultaneous PET-MR: Phantom studya),” Medical physics, vol. 41, no. 4, p. 41910, 2014. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [32].Catana C et al. , “MRI-assisted PET motion correction for neurologic studies in an integrated MR-PET scanner,” Journal of Nuclear Medicine, vol. 52, no. 1, pp. 154–161, 2011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [33].Huang C et al. , “Motion compensation for brain PET imaging using wireless MR active markers in simultaneous PET–MR: Phantom and non-human primate studies,” Neuroimage, vol. 91, pp. 129–137, 2014. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [34].Munoz C et al. , “Motion-corrected simultaneous cardiac positron emission tomography and coronary MR angiography with high acquisition efficiency,” Magnetic Resonance in Medicine, vol. 79, no. 1, pp. 339–350, Jan. 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [35].Petibon Y et al. , “Cardiac motion compensation and resolution modeling in simultaneous PET-MR: A cardiac lesion detection study,” Physics in Medicine and Biology, vol. 58, no. 7, pp. 2085–2102, 2013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [36].Küstner T et al. , “MR-based respiratory and cardiac motion correction for PET imaging,” Medical Image Analysis, vol. 42, pp. 129–144, Dec. 2017. [DOI] [PubMed] [Google Scholar]
- [37].Huang C et al. , “Accelerated acquisition of tagged MRI for cardiac motion correction in simultaneous PET-MR: Phantom and patient studies,” Medical physics, vol. 42, no. 2, pp. 1087–1097, 2015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [38].Robson PM et al. , “Correction of respiratory and cardiac motion in cardiac PET/MR using MR-based motion modeling,” Phys. Med. Biol, vol. 63, no. 22, p. 225011, 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [39].Petibon Y, El Fakhri G, Nezafat R, Johnson N, Brady T, and Ouyang J, “Towards coronary plaque imaging using simultaneous PET-MR: a simulation study,” Physics in medicine and biology, vol. 59, no. 5, p. 1203, 2014. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [40].Kolbitsch C et al. , “Cardiac and Respiratory Motion Correction for Simultaneous Cardiac PET/MR,” J. Nucl. Med, vol. 58, no. 5, pp. 846–852, May 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [41].Guo R, Petibon Y, Ma Y, El Fakhri G, Ying K, and Ouyang J, “MR-based motion correction for cardiac PET parametric imaging: a simulation study,” EJNMMI Phys, vol. 5, no. 1, p. 3, Feb. 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [42].Catana C, “Motion Correction Options in PET/MRI,” Seminars in Nuclear Medicine, vol. 45, no. 3, pp. 212–223, May 2015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [43].Munoz C, Kolbitsch C, Reader AJ, Marsden P, Schaeffter T, and Prieto C, “MR-Based Cardiac and Respiratory Motion-Compensation Techniques for PET-MR Imaging,” PET Clinics, vol. 11, no. 2, pp. 179–191, Apr. 2016. [DOI] [PubMed] [Google Scholar]
- [44].Catana C, “PET/MRI: Motion Correction,” in PET/MRI in Oncology: Current Clinical Applications, Iagaru A, Hope T, and Veit-Haibach P, Eds. Cham: Springer International Publishing, 2018, pp. 77–96. [Google Scholar]
- [45].Lalush DS, “Magnetic Resonance–Derived Improvements in PET Imaging,” Magnetic Resonance Imaging Clinics, vol. 25, no. 2, pp. 257–272, May 2017. [DOI] [PubMed] [Google Scholar]
- [46].Rakvongthai Y and Fakhri GE, “Magnetic Resonance–based Motion Correction for Quantitative PET in Simultaneous PET-MR Imaging,” PET Clinics, vol. 12, no. 3, pp. 321–327, Jul. 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [47].Patlak CS and Blasberg RG, “Graphical evaluation of blood-to-brain transfer constants from multiple-time uptake data. Generalizations,” J. Cereb. Blood Flow Metab, vol. 5, no. 4, pp. 584–590, Dec. 1985. [DOI] [PubMed] [Google Scholar]
- [48].Gjedde A, “Calculation of cerebral glucose phosphorylation from brain uptake of glucose analogs in vivo: a re-examination,” Brain Res, vol. 257, no. 2, pp. 237–274, Jun. 1982. [DOI] [PubMed] [Google Scholar]
- [49].Delso G et al. , “Performance measurements of the Siemens mMR integrated whole-body PET/MR scanner,” Journal of Nuclear Medicine, vol. 52, no. 12, pp. 1914–1922, 2011. [DOI] [PubMed] [Google Scholar]
- [50].Ozturk C, Derbyshire JA, and McVeigh ER, “Estimating motion from MRI data,” Proceedings of the IEEE, vol. 91, no. 10, pp. 1627–1648, 2003. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [51].O’Dell WG, Moore CC, Hunter WC, Zerhouni EA, and McVeigh ER, “Three-dimensional myocardial deformations: calculation with displacement field fitting to tagged MR images,” Radiology, vol. 195, no. 3, pp. 829–835, 1995. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [52].Axel L and Dougherty L, “MR imaging of motion with spatial modulation of magnetization,” Radiology, vol. 171, no. 3, pp. 841–845, 1989. [DOI] [PubMed] [Google Scholar]
- [53].Huang C et al. , “A novel golden-angle radial FLASH motion-estimation sequence for simultaneous thoracic PET/MR,” 2013. [Google Scholar]
- [54].Jung H, Ye JC, and Kim EY, “Improved k-t BLAST and k-t SENSE using FOCUSS,” Phys. Med. Biol, vol. 52, 2007. [DOI] [PubMed] [Google Scholar]
- [55].Fessler JA and Sutton BP, “Nonuniform fast Fourier transforms using min-max interpolation,” IEEE Transactions on Signal Processing, vol. 51, no. 2, pp. 560–574, Feb. 2003. [Google Scholar]
- [56].Block KT, Uecker M, and Frahm J, “Undersampled radial MRI with multiple coils. Iterative image reconstruction using a total variation constraint,” Magn Reson Med, vol. 57, no. 6, pp. 1086–1098, Jun. 2007. [DOI] [PubMed] [Google Scholar]
- [57].Rudin LI, Osher S, and Fatemi E, “Nonlinear total variation based noise removal algorithms,” Physica D: Nonlinear Phenomena, vol. 60, no. 1, pp. 259–268, Nov. 1992. [Google Scholar]
- [58].Chun SY and Fessler JA, “A simple regularizer for B-spline nonrigid image registration that encourages local invertibility,” Selected Topics in Signal Processing, IEEE Journal of, vol. 3, no. 1, pp. 159–169, 2009. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [59].Ledesma-Carbayo MJ, Derbyshire JA, Sampath S, Santos A, Desco M, and McVeigh ER, “Unsupervised estimation of myocardial displacement from tagged MR sequences using nonrigid registration,” Magnetic resonance in medicine, vol. 59, no. 1, pp. 181–189, 2008. [DOI] [PubMed] [Google Scholar]
- [60].Hudson HM and Larkin RS, “Accelerated image reconstruction using ordered subsets of projection data,” Medical Imaging, IEEE Transactions on, vol. 13, no. 4, pp. 601–609, 1994. [DOI] [PubMed] [Google Scholar]
- [61].Watson CC, “New, faster, image-based scatter correction for 3D PET,” Nuclear Science, IEEE Transactions on, vol. 47, no. 4, pp. 1587–1594, 2000. [Google Scholar]
- [62].Cerqueira MD et al. , “Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart a statement for healthcare professionals from the cardiac imaging committee of the Council on Clinical Cardiology of the American Heart Association,” Circulation, vol. 105, no. 4, pp. 539–542, 2002. [DOI] [PubMed] [Google Scholar]
- [63].“Graphical Evaluation of Blood-to-Brain Transfer Constants from Multiple-Time Uptake Data. Generalizations - Patlak Clifford S., Blasberg Ronald G., 1985. ” [Online]. Available: https://journals.sagepub.com/doi/10.1038/jcbfm.1985.87. [Accessed: 10-Dec-2018]. [DOI] [PubMed] [Google Scholar]
- [64].Furst S et al. , “Motion correction strategies for integrated PET/MR,” Journal of Nuclear Medicine, vol. 56, no. 2, pp. 261–269, 2015. [DOI] [PubMed] [Google Scholar]
- [65].Scott AD, Keegan J, and Firmin DN, “Motion in cardiovascular MR imaging,” Radiology, vol. 250, no. 2, pp. 331–351, 2009. [DOI] [PubMed] [Google Scholar]
- [66].Dasari PKR et al. , “Correction of hysteretic respiratory motion in SPECT myocardial perfusion imaging: Simulation and patient studies,” Medical Physics, vol. 44, no. 2, pp. 437–450, Feb. 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [67].Nehrke K, Bornert P, Manke D, and Bock JC, “Free-breathing Cardiac MR Imaging: Study of Implications of Respiratory Motion—Initial Results 1,” Radiology, vol. 220, no. 3, pp. 810–815, 2001. [DOI] [PubMed] [Google Scholar]
- [68].Piccinelli M, Votaw JR, and Garcia EV, “Motion Correction and Its Impact on Absolute Myocardial Blood Flow Measures with PET,” Curr Cardiol Rep, vol. 20, no. 5, p. 34, 24 2018. [DOI] [PubMed] [Google Scholar]
- [69].Lu Y and Liu C, “Patient motion correction for dynamic cardiac PET: Current status and challenges,” J. Nucl. Cardiol, Nov. 2018. [DOI] [PubMed] [Google Scholar]
- [70].Lee BC et al. , “Automated dynamic motion correction using normalized gradient fields for 82rubidium PET myocardial blood flow quantification,” J. Nucl. Cardiol, Nov. 2018. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.