Abstract
Purpose
The aim of this study was to compare the metabolic load between adaptive support ventilation (ASV) and pressure support ventilation (PSV) modes in critically ill patients.
Methods
Sequential 20 min ventilation by PSV followed by 20 min ASV in critically ill patients was assessed. ASV was set for full support, i.e., with the minute volume control set at the same level as the minute volume observed during PSV. The trial started from PSV 8 cmH2O and continued with high (PSV 12 cmH2O) to low (PSV 0) conditions or low to high conditions, in random order. The oxygen consumption (VO2), production of carbon dioxide (VCO2), and energy expenditure (EE) were measured by indirect calorimetry (IC).
Results
Twenty-four patients with critical illness participated in the study. Comparing with the PSV mode, the EE in the ASV mode was lower in the level of PSV 0 cmH2O (1069 ± 73 vs. 1425 ± 76 kcal), PS 8 cmH2O (1116 ± 70 vs. 1284 ± 61 kcal), and PS 12 cmH2O (1017 ± 70 vs. 1169 ± 58 kcal) (p < 0.05). The VO2, VCO2, and P0.1 in PSV were significantly higher than those in ASV (p < 0.05). The VO2, VCO2, and P0.1 in PSV were significantly higher than those in ASV (
Conclusion
In patients with critical illness, the application of ASV set for full support was associated with a lower metabolic load and respiratory drive than in any of the studied PSV conditions.
1. Introduction
For patients with acute respiratory failure, mechanical ventilation (MV) is the cornerstone of management. Patients admitted to the intensive care unit (ICU) who require MV are expected to have higher mortality rates than those who do not require respiratory support [1]. The application of the mechanical ventilator is crucial for patients with critical illness, especially during the weaning process. Choosing a ventilator mode that maintains adequate oxygenation and ventilation without inducing pulmonary complication or delaying the weaning process is a critical concern for clinicians.
Pressure support ventilation (PSV) is a widely used mode of ventilation, especially during the weaning process for critically ill patients [2]. During ventilation, PSV provides a fixed end-inspiratory pressure, regardless of the variabilities of patient's demand for ventilator or gas exchange [3–6]. Thus, the preset values for ventilator parameters, such as a fixed end-inspiratory pressure, are unlikely to provide appropriate assist immediately.
Adaptive support ventilation (ASV) is one of the recently developed MV modes. It is a closed-loop pressure-controlled ventilation mode. ASV is a mode of ventilation which allows delivery of pressure cycles that may be assisted (PSV-like) or controlled (PC-like). The characteristics of the cycles delivered by the ventilator are related to a minute ventilation target set by the clinician and automated measurements of the patient's respiratory mechanics [7]. The working principles of this mode are based on the findings of Otis et al., suggesting that for a given level of alveolar ventilation and dead space, there is an optimum combination of respiratory rate and tidal volume delivered in terms of the respiratory work of breathing (WOB) [8]. In this context, “Optimal” means minimizing mechanical WOB. During this mode, any changes in respiratory mechanics and effort of the patient are accompanied by a dynamic pattern of breathing that gradually guides the patient to the new goal where optimal breath is achieved. Therefore, maybe more energy efficient to minimize the cumulative effects of elastic and resistive load is imposed on the respiratory system [4].
Measuring oxygen consumption (VO2) and carbon dioxide production (VCO2) can be useful for assessing the function of alveolar ventilation, pulmonary perfusion, and ventilation optimization [9]. In patients supported with MV, monitoring VO2 predicts the success or failure of weaning trials [10]. In addition, the VO2 of respiratory muscles is computed as the difference between VO2 measured during MV and spontaneous breathing. This approach assumes that the VO2 of respiratory muscles varied with and without ventilatory support [11, 12]. In addition, VCO2 and VO2 in mechanically ventilated patients can be measured using an indirect calorimeter [13]. A study reported significantly increased WOB in terms of energy expenditure (EE) and VO2 during T-piece breathing compared with those in the PSV mode in critically ill patients [14].
In ASV applied in passive patients, the optimal combination of Vt and RR should minimize the energy transferred by the ventilator to the respiratory system. Whether ASV applied in actively breathing patients results in a lower metabolic load, such as VO2 or EE when compared with a conventional ventilation mode like pressure support ventilation (PSV), remains unclear. The purpose of this study was to compare the metabolic load measured by indirect calorimetry between ASV set for full respiratory support and PSV set at different pressure levels in stable critically ill patients.
2. Methods
This prospective study was performed in a 24-bed medical ICU at Chang Gung Memorial Hospital, Taiwan. The inclusion criteria for this study were as follows: (1). patients who had been mechanically ventilated for more than 24 h with PSV levels of 8 cmH2O, PEEP ≤8 cmH2O, and FiO2 ≤40%. (2). Medical stability (arterial blood gas pH = 7.35–7.45, PaO2 ≥ 60 mmHg on FiO2 ≤40%, and PaCO2 40 ± 5 mm Hg; for patients with COPD who had been chronically hypercapnic and had well tolerated the ventilator support, a PaCO2 value up to 55 mmHg was acceptable, absence of signs and symptoms of infection, and hemodynamic stability). The exclusion criteria included acute lung or systemic infection, hemodynamic instability, previous or ongoing neuromuscular disease (e.g., myasthenia gravis and Guillain–Barré disease), and patients with chest tubes. The study was approved by the hospital's institutional review board. The study was performed in accordance with the Declaration of Helsinki. Written informed consent forms were obtained from subjects or their relatives prior to inclusion.
2.1. Study Protocol
First, the clinical characteristics were recorded, including age, sex, Acute Physiology and Chronic Health Evaluation II (APACHE II) score, and diagnosis. Patients were monitored until their discharge from the ICU or until death. All patients were ventilated with a microprocessor-controlled ventilator that had a closed-loop ventilation capability (Galileo Gold; Hamilton Medical, Rhäzüns, Switzerland). After enrollment, the bronchial hygiene therapy was performed first, followed by lying and the bed head at 45° for 5–10 minutes before starting the IC measurements. All measurements were performed at both PSV and ASV modes under 3 ventilator support levels: low (PSV 0 cmH2O, PSV0), baseline (PSV 8 cmH2O, PSV8), and high (PSV 12 cmH2O, PSV12). We conducted a randomized crossover controlled trial. Patients were randomized using a computer-generated randomization sequence in sealed envelopes to receive ventilator support levels in the sequences of low-to-high or high-to-low conditions. All patients started from baseline level (PSV 8). After the baseline period, patients were then randomized into groups with different sequences: low (PSV0) to high (PSV12) or high (PSV12) to low (PSV0) conditions (Figure 1). The pressure support level of PSV0 was set at zero. In each level, the ventilator mode started from PSV modes for 20 minutes. After confirming that patients were well tolerated with the minute volume of the PSV mode without signs/symptoms of respiratory distress, the ventilator mode was then switched to the ASV mode for 20 min. The minute volume of the last 5 min of each PSV level was recorded and averaged. The result was used to set the minute volume target (i.e. the %MinVol control) of the following ASV step of the study. At the beginning of each matched-ASV level, the %MinVol was adjusted until the measured minute volume (which is measured by the ventilator) of the patient was similar to that in the previous PSV mode. The rest of the ventilator parameter setting remained the same as those before enrolling the study. Besides IC parameters, the respiratory parameters such as mean airway pressure (MAP), peak pressure (Ppeak), airway occlusion pressure at 0.1 s (P0.1), spontaneous RR, Vt, and minute volume were recorded from the mechanical ventilator in the last 5 min of each condition.
2.2. Indirect Calorimetry (IC) Measurement
A calorimeter module (Engström Carestation, General Electric, Madison, WI) was used for indirect calorimetry. The module consists of a fast differential paramagnetic oxygen analyzer, an infrared analyzer for carbon dioxide, and a pneumotachograph to measure the inspired and expired volumes. The pneumotachograph and gas sampling ports were connected to a disposable connector, placed between the Y-piece of the ventilator circuit and the endotracheal tube. The signals from the pneumotachograph and gas analyzers were synchronized for breath-by-breath data on gas exchange. Under each condition, the data were collected for 30 min with the first 10 min being discarded for analysis purposes. We consider the steady state to be the point after 5 consecutive minutes measurement when VCO2 and VO2 vary by ±10%. During the last 5 min of each steady state, the following data were recorded: VCO2, VO2, EE, and respiratory quotient (RQ).
2.3. Statistical Analysis
The sample size was calculated according to a previous study [13], and assuming a middle effect size and an α error of 0.05, a sample size of 25 would have 80% power to detect the difference between the modes. Allowing for the 20% dropout rate, the sample size was increased to 30.
Analysis was conducted using SPSS v.17. The results were expressed as mean ± standard deviation. The variables between the PSV and ASV modes were compared using Student's t test. A repeated-measure analysis of variables was used to examine the difference among the three different levels in the PSV and ASV modes. The variables were correlated using Pearson correlation coefficient. p < 0.05 indicated statistical significance.
3. Results
3.1. Demographic Characteristics
From June 2017 to May 2018, 26 patients were enrolled in the study (Figure 1). During the study period, two subjects met the exclusion criteria (hemodynamic instability) due to their primary diagnosis and were excluded, leaving 24 subjects for the analysis.
The demographic and clinical characteristics of the subjects are summarized in Table 1. Most subjects were men (70.8%). The admission diagnosis was mostly respiratory system diseases (pneumonia 37.5%, COPD 8.3%, and other respiratory diseases 16.7%). The mean age was 70.6 ± 14.0 years, and the mean APACHE II score was 19.0 ± 4.0. During ICU admission, only one subject (4.1%) failed to extubate from the mechanical ventilator.
Table 1.
(n = 24) | ||
---|---|---|
Male/female, (n) | 17/7 | |
Age (year) | 70.6 ± 14.0 | |
APACHE II score | 19.0 ± 4.1 | |
Days on MV (day) | 9.6 ± 5.6 | |
ICU stay (day) | 10.3 ± 5.8 | |
Diagnosis, n (%) | ||
Pneumonia | 9 | 37.5% |
COPD AE | 2 | 8.3% |
Other respiratory diseases | 4 | 16.7% |
Congestive heart failure | 2 | 8.3% |
Cerebral vascular accident | 2 | 8.3% |
Cancer | 5 | 20.8% |
Ventilator setting | ||
Pressure support level (cmH2O) | 8 | |
PEEP (cmH2O) | 8 | |
FiO2 (%) | 34.5 ± 2.1 |
APACHE II, Acute Physiology and Chronic Health Evaluation; COPD AE, Chronic Obstructive Pulmonary Disease Acute Exacerbation.
3.2. Comparison of Lung Mechanics and Metabolic Load between the PSV and ASV Modes
The metabolic load was measured (Table 2), and no significant difference was observed in the minute volumes between the PSV and ASV modes in the PSV0, PSV8, and PSV12 levels (p < 0.05). The EE in PSV0 (1425 ± 76 kcal), PSV8 (1284 ± 76 kcal), and PSV12 (1169 ± 58 kcal) was significantly higher than that in the ASV mode with matched minute volumes (1069 ± 73, 1116 ± 70, and 1017 ± 70 kcal, respectively) (p < 0.001). VO2 in PSV0 (218 ± 11 mL/min), PSV8 (195 ± 9 mL/min), and PSV12 (178 ± 8 mL/min) was significantly higher than that in the ASV mode with matched minute volumes (161 ± 10, 169 ± 10, and 154 ± 10 mL/min, respectively) (p < 0.001). VCO2 was also significantly higher in the PSV mode than that in the ASV mode in each level (p < 0.001).
Table 2.
PS0 | ASV PS0 | p | PS8 | ASVPS8 | p | PS12 | ASVPS12 | p | |
---|---|---|---|---|---|---|---|---|---|
MV (L/min) | 9.1 ± 1 | 8.9 ± 0.9 | 0.51 | 9.3 ± 1.1 | 8.8 ± 1 | 0.49 | 8.7 ± 1 | 8.6 ± 1 | 0.51 |
EE (kcal) | 1425 ± 76 | 1069 ± 73∗ | <0.001 | 1284 ± 61 | 1116 ± 70∗ | <0.001 | 1169 ± 58 | 1017 ± 70∗ | <0.001 |
VO2 (ml) | 218 ± 11 | 161 ± 10∗ | <0.001 | 195 ± 9 | 164 ± 10∗ | <0.001 | 178 ± 8 | 154 ± 10∗ | <0.001 |
VCO2 (ml) | 143 ± 7 | 116 ± 9∗ | <0.001 | 135 ± 7 | 120 ± 8∗ | <0.001 | 121 ± 7 | 108 ± 8∗ | <0.001 |
RQ | 0.68 ± 0.1 | 0.69 ± 0.1 | 0.902 | 0.67 ± 0.1 | 0.71 ± 0.1 | 0.257 | 0.70 ± 0.1 | 0.70 ± 0.1 | 0.899 |
∗ p < 0.05, comparisons between two modes. MV, minute ventilation; EE, energy expenditure; VCO2, oxygen consumption; VCO2, carbon dioxide production.
In the measurement of lung mechanics (Table 3), a significant difference in Vt, RR, and P0.1 was observed between the PSV and ASV modes. In the PSV mode, Vt in PSV0 (442 ± 25 mL), PSV8 (498 ± 30 mL), and PSV12 (501 ± 30 mL) levels was significantly lower than that in the ASV mode (520 ± 23, 539 ± 27, 529 ± 26 mL, respectively) (p < 0.001). RR in the PSV mode was significantly higher than that in the ASV mode with matched minute volumes. P0.1 in PSV0 (−4.4 ± 0.4 cmH2O), PSV8 (−3.3 ± 0.4 cmH2O), and PSV12 (−2.7 ± 0.3 cmH2O) was significantly higher than that in the ASV mode (−2.1 ± 0.2, −2.0 ± 0.2, and −1.8 ± 0.2 cmH2O) (p < 0.001). Ppeak and MAP in PSV0 and PSV8 were significantly lower than those in the ASV mode with matched minute volumes (p < 0.001).
Table 3.
PS0 | ASV PS0 | p | PS8 | ASVPS8 | p | PS12 | ASVPS12 | p | |
---|---|---|---|---|---|---|---|---|---|
MV (L/min) | 9.1 ± 1 | 8.9 ± 0.9 | 0.51 | 9.3 ± 1.1 | 8.8 ± 1 | 0.49 | 8.7 ± 1 | 8.6 ± 1 | 0.51 |
Spont RR (bpm) | 21 ± 1 | 17 ± 1∗ | <0.001 | 19 ± 1 | 16 ± 1∗ | <0.001 | 18 ± 1 | 16 ± 1∗ | <0.001 |
Vt (ml) | 442 ± 25 | 520 ± 23∗ | <0.001 | 498 ± 30 | 539 ± 27∗ | <0.001 | 501 ± 30 | 529 ± 26∗ | <0.001 |
Ppeak (cmH2O) | 12 ± 1 | 21 ± 4∗ | <0.001 | 16 ± 1 | 21 ± 3∗ | <0.001 | 20 ± 1 | 21 ± 4 | 0.06 |
MAP | 9.1 ± 2 | 12 ± 1∗ | <0.001 | 10.8 ± 1 | 12.1 ± 0.1 | <0.001 | 11.8 ± 1 | 12.1 ± 1 | 0.07 |
P0.1 (cmH2O) | −4.4 ± 0.4 | −2.1 ± 0.2 | <0.001 | −3.3 ± 0.4 | −2.0 ± 0.2 | <0.001 | −2.7 ± 0.3 | −1.8 ± 0.2 | <0.001 |
∗ p < 0.05, comparisons between two modes. Vt, tidal volume; RR, respiratory rate; MV, minute ventilation; Ppeak, peak pressure; MAP, mean airway pressure; P0.1, airway occlusion pressure.
3.3. Comparison of Measurements among 3 PSV Levels and 3 ASV Trials
In the comparison of measurements among 3 different PSV levels (Table 4), there was a significant difference in the measurement of EE, VO2, and VCO2. The EE in PS12 (1169 ± 58 kcal) was significantly lower than that in PS8 (1284 ± 61 kcal) and PSV0 (1425 ± 76 kcal) (p < 0.001). In addition, the EE in PS8 was also significantly lower than that in PSV0. This indicated that a higher support level was associated with a lower metabolic load in the PSV mode. The Vt, Ppeak, and MAP were significantly higher in PS12 when compared with those in PS8 and PSV0 (p < 0.001). The Vt in PS12 (501 ± 30 ml) was significantly higher than that in PS8 (498 ± 30 ml) and PSV0 (442 ± 25 ml) (p < 0.001). The RR and P0.1 were significantly lower in PS12 than those in PS8 and PSV0 (p < 0.001). Among the 3 ASV trials, no significant difference was found in the %MinVol setting, as well as in the measurements of tidal volume, frequency, inspiratory pressure above PEEP, and P0.1 (Table 5). The corresponding metabolic measurements of EE, VO2, and VCO2 showed some differences, whose amounts were clinically irrelevant although statistically significant (Table 5).
Table 4.
PS0 cmH2O | PS8 cmH2O | PS12 cmH2O | p | ||
---|---|---|---|---|---|
EE (kcal) | 1425 ± 76 | 1284 ± 6 | 1169 ± 58 | <0.001∗ | c < a, b; b < c |
VO2 (ml/min) | 218 ± 11 | 195 ± 9 | 178 ± 8 | <0.001∗ | c < a, b; b < c |
VCO2 (ml/min) | 143 ± 7 | 135 ± 7 | 121 ± 7 | <0.001∗ | c < a, b |
MV (L/min) | 9.1 ± 1 | 9.3 ± 1.1 | 8.7 ± 1 | 0.21 | |
Spont RR (bpm) | 21 ± 1 | 19 ± 1 | 18 ± 1 | 0.01∗ | c < a, b; b < c |
Vt (ml) | 442 ± 25 | 498 ± 30 | 501 ± 30 | <0.001∗ | c > a, b; b > c |
Ppeak (cmH2O) | 12 ± 1 | 16 ± 1 | 20 ± 1 | <0.001∗ | c > a, b; b > c |
MAP (cmH2O) | 9.1 ± 2 | 10.8 ± 1 | 11.8 ± 1 | <0.001∗ | c > a, b; b > c |
P0.1 (cmH2O) | −4.4 ± 0.4 | −3.3 ± 0.4 | −2.7 ± 0.3 | <0.001∗ | c < a, b; b < c |
MV, minute ventilation; EE, energy expenditure; VCO2, oxygen consumption; VCO2, carbon dioxide production; Vt, tidal volume; RR, respiratory rate; MV, minute ventilation; Ppeak, peak pressure; MAP, mean airway pressure; P0.1, airway occlusion pressure; a, PS0 cmH2O; b, PS8 cmH2O; c, PS12 cmH2O. ∗Variables comparison over 3 levels of PS were analyzed by repeated-measures ANOVA, p < 0.05.
Table 5.
ASVPS0 | ASVPS8 | ASVPS12 | p | ||
---|---|---|---|---|---|
EE (kcal) | 1069 ± 73 | 1116 ± 70 | 1017 ± 70 | <0.001∗ | c < b |
VO2 (ml/min) | 161 ± 10 | 169 ± 10 | 154 ± 10 | <0.001∗ | c < b |
VCO2 (ml/min) | 116 ± 9 | 120 ± 8 | 108 ± 8 | 0.01∗ | c < a, b |
MV (%) | 145 ± 6 | 144 ± 6 | 143 ± 6 | 0.82 | |
MV (L/min) | 8.9 ± 0.9 | 8.8 ± 1 | 8.6 ± 1 | 0.52 | |
Spont RR (bpm) | 17 ± 1 | 16 ± 1 | 16 ± 1 | 0.25 | |
Vt (ml) | 520 ± 23 | 539 ± 27 | 529 ± 26 | 0.14 | |
Ppeak (cmH2O) | 21 ± 4 | 21 ± 3 | 21 ± 4 | 0.90 | |
MAP (cmH2O) | 12 ± 1 | 12.2 ± 1 | 12 ± 1 | 0.74 | |
Delta P (cmH2O) | 13 ± 4.2 | 13.4 ± 3.8 | 13.3 ± 4.3 | 0.94 | |
P0.1 (cmH2O) | −2.1 ± 0.2 | −2.2 ± 0.2 | −1.8 ± 0.2 | 0.12 |
MV, minute ventilation; EE, energy expenditure; VCO2, oxygen consumption; VCO2, carbon dioxide production; Vt, tidal volume; RR, respiratory rate; MV, minute ventilation; Ppeak, peak pressure; MAP, mean airway pressure; P0.1, airway occlusion pressure; Delta P, difference between Ppeak and PEEP. a, PS0 cmH2O; b, PS8 cmH2O; c, PS12 cmH2O. ∗Variables comparison over 3 levels of PS were analyzed by repeated-measures ANOVA, p < 0.05.
3.4. Correlation between Lung Mechanics and Metabolic Load in the PSV and ASV Modes
After pooling data from PSV and ASV modes, a significant correlation was observed among the mechanical load and metabolic load. P0.1 was positively correlated with EE (r = 0.442), VO2 (r = 0.424), and VCO2 (r = 0.457) (Table 6).
Table 6.
EE (kcal) | VO2 (ml/min) | VCO2 (ml/min) | |
---|---|---|---|
P0.1 (cmH2O) | |||
r | 0.442∗ | 0.424∗ | 0.457∗ |
p | <0.001 | <0.001 | <0.001 |
EE, energy expenditure; VO2, oxygen consumption; Vt, tidal volume; RR, respiratory rate; P0.1, airway occlusion pressure at 0.1 sec. ∗p < 0.05.
4. Discussion
This study compared the metabolic load of PSV and ASV. We discovered that when the %MinVol control of ASV was set to match the minute volume observed during PSV with PS from zero to 12 cmH2O, ASV was associated with a larger Vt and lower RR than PSV even when the latter was set at the top level of 12 cmH2O. This was due to the generation of PS levels by ASV larger than by PSV and was associated with lower EE, VO2, VCO2, and P0.1 during ASV.
In our study, EE and VO2 in the ASV mode were significantly lower than those in the PSV mode at similar minute volumes. In addition, P0.1 was lower in ASV which indicated patients maintain similar minute volumes with lower respiratory effort and respiratory muscle work in the ASV mode than those in the PSV mode. This could be part of the reasons that contribute to the lower metabolic load in the ASV mode. ASV is a complex, dual-controlled mode. After setting the target %MinVol, the ventilator calculates an optimal breath pattern and associated target values for tidal volume and respiratory rate according to the Otis equation [15]. The basic assumption of Otis equation is that the optimal breath pattern is identical to a patient would choose naturally in resting status [8]. In healthy individuals, the resting VO2 from the respiratory system is about only 1–2% of whole body oxygen consumption [16]. ASV monitors patients' respiratory mechanics breath by breath and automatically adjusts breathing pattern based on user inputs and the changes of respiratory mechanics. ASV ensures effective alveolar ventilation and leads patients to an optimal and comfortable breathing pattern [4]. Our result was consistent with previous studies.
P0.1 reflected both the respiratory drive and work of respiratory muscles. Our study showed that the P0.1 in the ASV mode was significantly lower than that in the PSV mode. The mechanisms for the lower metabolic work and P0.1 at similar minute volume were probably related to the higher Vt, as recommended by the ASV algorithm when ASV is set for full support, i.e., with the %MinVol control set close to the minute volume requirement of the patient. The higher Vt decreased the muscle force, and the patient has to generate to get the desired Vt. The higher Vt may also stimulate pulmonary stretch receptors, sending the impulse to central drive, to lower the respiratory rate [17–19]. This was further substantiated by the higher Vt and lower RR in ASV when compared with those in PSV in our study.
According to our inclusion criteria, only patients with stable condition (hemodynamically stable, acceptable ABG data) were recruited. This indicated that the original ventilator setting provided adequate support for patient's ventilation requirement. To reach similar VE, the %MinVol was set to an average of 144%MinVol at the presence of patients' full spontaneous breath. A similar finding has been reported in a previous study [18]. Wu et al. reported that, in patients with respiratory failure, when the % MinVol setting is increased gradually to a point where mandatory breath appeared (target point), the P0.1 was significantly decreased, compared with that in 100% MinVol [18]. The average %MinVol setting of the target point in Wu's study was 165%. Although all the measurements in our study were performed in the presence of spontaneous breath, both our and Wu's data showed that the %MinVol required to decrease WOB in patients with respiratory failure is greater than what operator manual recommended [15]. The ability of ASV to unload the patient's respiratory muscles also depends on the ASV + a given setting of (%MinVol) control of ASV. The inadequate %MinVol setting may result in increased work of breath and respiratory muscle fatigue in patients. However, an oversetting may result in highly unloading respiratory muscle and put patients under the risk of respiratory muscle atrophy. Thus, carefully adjusted setting to meet patients' requirement without inducing other complications is crucial for patients with a mechanical ventilator. The P0.1 and presence of spontaneous/mandatory breath may be the indicators when set the %MinVol. In addition, the data from IC measurements may also provide useful information to ensure appropriate ventilator setting.
The lung mechanics data reveal that Vt in PSV was significantly lower and RR was higher than those in ASV under 3 levels. In PSV, the level of pressure is fixed regardless of the ventilator demand of the patients. When the level of PSV cannot provide adequate Vt, patients must increase RR to match the demand, which may thus lead to increased WOB [20].
4.1. Limitations
Some limitations affected this study. First, the population was relatively small. The results reflect findings from a single institution and represent a heterogeneous collection of diseased state. However, the population reflects patients that are likely to be encountered in an ICU. Second, the data in our study were obtained by the version of ASV provided by the Galileo Gold ventilator (Hamilton Medical). The generalizability of our results to other versions requires further study. Third, we used a commercial device to measure; this system has the advantage of being noninvasive and of providing continuous measurements of VO2. We utilized a single device to ensure that device-device agreement was not a confounding factor in our study. However, the data may not specifically reflect the work of respiratory muscles, which may also change with the ventilatory mode. The measurements of esophageal pressure obtained through the placement of the esophageal balloon were proposed as a valid indicator of respiratory muscle work and may be considered in future studies.
5. Conclusion
Our study showed that, compared with PSV, ASV set for full support in actively breathing patients provides larger Vt and lower RR with lower VO2, VCO2, and EE at similar minute volume. The ventilatory drive as assessed by P0.1 was also lower during the ASV mode than during the PSV mode. Our study also shows the interest of metabolic measurements in patients with critical illness during ventilated or weaning stages.
Acknowledgments
This study was supported by a research grant from Ministry of Science and Technology, Taiwan (MOST 107-2314-182049).
Data Availability
The data that have been used are confidential currently and may be ready to share later.
Disclosure
The funding body has no input into the study design, data collection, and interpretation of the results, in the writing of the manuscript, and in the decision of submitting the manuscript for publication.
Conflicts of Interest
The authors declare that they have no conflicts of interest.
Authors' Contributions
Hsiu-Feng Hsiao and Yen-Huey Chen contributed equally to this work.
References
- 1.Fialkow L., Farenzena M., Wawrzeniak I., et al. Mechanical ventilation in patients in the intensive care unit of a general university hospital in southern Brazil: an epidemiological study. Clinics. 2016;71(3):145–151. doi: 10.6061/clinics/2016(03)05. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Bellani G., Foti G., Spagnolli E., et al. Increase of oxygen consumption during a progressive decrease of ventilatory support is lower in patients failing the trial in comparison with those who succeed. Anesthesiology. 2010;113(2):378–385. doi: 10.1097/aln.0b013e3181e81050. [DOI] [PubMed] [Google Scholar]
- 3.Coisel Y., Chanques G., Jung B., et al. Neurally adjusted ventilatory assist in critically ill postoperative patients: a crossover randomized study. Anesthesiology. 2010;113(4):925–935. doi: 10.1097/aln.0b013e3181ee2ef1. [DOI] [PubMed] [Google Scholar]
- 4.Fernández J., Miguelena D., Mulett H., Godoy J., Martinón-Torres F. Adaptive support ventilation: state of the art review. Indian Journal of Critical Care Medicine. 2013;17(1):16–22. doi: 10.4103/0972-5229.112149. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Zhu F., Gomersall C. D., Ng S. K., Underwood M. J., Lee A. A randomized controlled trial of adaptive support ventilation mode to wean patients after fast-track cardiac valvular surgery. Anesthesiology. 2015;122(4):832–840. doi: 10.1097/aln.0000000000000589. [DOI] [PubMed] [Google Scholar]
- 6.Ghodrati M., Pournajafian A., Khatibi A., et al. Comparing the effect of adaptive support ventilation (ASV) and synchronized intermittent mandatory ventilation (SIMV) on respiratory parameters in neurosurgical ICU patients. Anesthesiology and Pain Pedicine. 2016;6(6) doi: 10.5812/aapm.40368.e40368 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Brunner J. X., Iotti G. A. Adaptive support ventilation (ASV) Minerva Anestesiologica. 2002;68(5):365–368. [PubMed] [Google Scholar]
- 8.Otis A. B., Fenn W. O., Rahn H., et al. Mechanics of breathing in man. Journal of Applied Physiology. 1950;2(11):592–607. doi: 10.1152/jappl.1950.2.11.592. [DOI] [PubMed] [Google Scholar]
- 9.Mehta N. M., Compher C. A.S.P.E.N. clinical guidelines: nutrition support of the critically ill child. Journal of Parenteral and Enteral Nutrition. 2009;33(3):260–276. doi: 10.1177/0148607109333114. [DOI] [PubMed] [Google Scholar]
- 10.Hubmayr R. D., Loosbrock L. M., Gillespie D. J., Rodarte J. R. Oxygen uptake during weaning from mechanical ventilation. Chest. 1988;94(6):1148–1155. doi: 10.1378/chest.94.6.1148. [DOI] [PubMed] [Google Scholar]
- 11.Kirakli C., Ozdemir I., Ucar Z. Z., Cimen P., Kepil S., Ozkan S. A. Adaptive support ventilation for faster weaning in COPD: a randomised controlled trial. European Respiratory Journal. 2011;38(4):774–780. doi: 10.1183/09031936.00081510. [DOI] [PubMed] [Google Scholar]
- 12.Weyland W., Schuhmann M., Rathgeber J., et al. Oxygen cost of breathing for assisted spontaneous breathing modes: investigation into three states of pulmonary function. Intensive Care Medicine. 1995;21(3):211–217. doi: 10.1007/bf01701474. [DOI] [PubMed] [Google Scholar]
- 13.Smallwood C. D., Walsh B. K., Bechard L. J., Mehta N. M. Carbon dioxide elimination and oxygen consumption in mechanically ventilated children. Respiratory Care. 2015;60(5):718–723. doi: 10.4187/respcare.03605. [DOI] [PubMed] [Google Scholar]
- 14.dos Santos L. J., Hoff F. C., Condessa R. L., Kaufmann M. L., Vieira S. R. R. Energy expenditure during weaning from mechanical ventilation: is there any difference between pressure support and T-tube? Journal of Critical Care. 2011;26(1):34–41. doi: 10.1016/j.jcrc.2010.05.032. [DOI] [PubMed] [Google Scholar]
- 15.Arnal J. M., Wysocki M., Novotni D., et al. Hamilton manual. 2009. http://www.hamilton-medical.com.
- 16.Donahoe M., Rogers R. M., Wilson D. O., et al. Oxygen consumption of the respiratory muscles in normal and inmalnourished patients with chronic obstructive pulmonary disease. American Review of Respiratory Disease. 1989;140(2):385–391. doi: 10.1164/ajrccm/140.2.385. [DOI] [PubMed] [Google Scholar]
- 17.Tassaux D., Dalmas E., Gratadour P., Jolliet P. Patient-ventilator interactions during partial ventilatory support: a preliminary study comparing the effects of adaptive support ventilation with synchronized intermittent mandatory ventilation plus inspiratory pressure support. Critical Care Medicine. 2002;30(4):801–807. doi: 10.1097/00003246-200204000-00014. [DOI] [PubMed] [Google Scholar]
- 18.Wu C. P., Lin H. I., Perng W. C., et al. Correlation between the %MinVol setting and work of breathing during adaptive support ventilation in patients with respiratory failure. Respiratory Care. 2010;55(3):334–341. [PubMed] [Google Scholar]
- 19.Widdicombe J. G. Reflexes from the lungs and airways:historical perspective. Journal of Applied Physiology. 2006;101(2):628–634. doi: 10.1152/japplphysiol.00155.2006. [DOI] [PubMed] [Google Scholar]
- 20.Jaber S., Sebbane M., Verzilli D., et al. Adaptive support and pressure support ventilation behavior in response to increased ventilatory demand. Anesthesiology. 2009;110(3):620–627. doi: 10.1097/aln.0b013e31819793fb. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The data that have been used are confidential currently and may be ready to share later.