Abstract
Timely diagnosis of acute diseases improves treatment outcomes and saves lives, but it requires fast and precision quantification of biomarkers. Here we report a time-resolved digital immunoassay based on plasmonic imaging of binding of single nanoparticles to biomarkers captured on a sensor surface. The real-time and high contrast of plasmonic imaging lead to fast and precise counting of the individual biomarkers over a wide dynamic range. We demonstrated the detection principle, evaluated the performance of the method using procalcitonin (PCT) as an example, and achieved a limit of detection of ~2.8 pg/mL, dynamic range of 4.2–12500 pg/mL, for a total detection time of ~25 mins.
Keywords: time-resolved digital immunoassay, procalcitonin, plasmonic imaging, rapid and sensitive quantification, single nanoparticles analysis
Detection and quantification of molecular biomarkers are critical to disease diagnosis and progression monitoring.1–3 Various approaches have been developed,4–7 but the most well-established technology is ELISA (enzyme-linked immunosorbent assay), which amplifies antibody-biomarker binding via enzymatic reactions and converts reaction products into an optical signal (e.g., color changes).8,9 Although ELISA is widely used in clinical and research labs, its detection limit, total test time (sample to result) and dynamic range are often insufficient for clinical applications. Recent technological advances have made it possible to detect single molecules, which have been used to measure the binding of a single biomarker molecule to a capture antibody with improved detection limit. This single-molecule approach has been referred to as digital immunoassay.10–13
Several digital immunoassay methods have been developed or proposed. One is to replace the large reaction well in the traditional ELISA with an array of microwells.10–13 The volume of the microwells is small, such that each microwell has either no biomarker present or a single biomarker molecule that binds to a capture antibody conjugated with a magnetic bead in the microwell. Another digital immunoassay platform combines single molecule fluorescent detection with flow cytometry.14 The third platform uses metal nanoparticles as the signal readout mechanism.15–18 The nanoparticles can be imaged individually via dark field, interference and plasmonic optical imaging techniques.19,20 These digital immunoassay technologies provide single molecule detection capability, which significantly improves the detection limit of the traditional ELISA. However, the improvement is often achieved at the expense of test time (e.g., > 1 hr) and dynamic range (e.g., < 2 ng/mL), both are crucial for disease diagnosis and treatment.2 An important example is sepsis, which develops rapidly with a survival rate of 79.9% from the first hour and a sharp drop of 7.6% per hour.21 Timely detection of biomarkers of sepsis is thus needed to improve treatment outcome and save lives.
Here we report a time-resolved digital immunoassay (TD-Immunoassay) based on plasmonic imaging of nanoparticles for rapid detection of biomarkers with a wide dynamic range. The plasmonic imaging offers high contrast and fast imaging of nanoparticles, allowing detection of single molecule binding on a sensor surface via detection of antibody-conjugated nanoparticles. It features real-time counting of the nanoparticles as they bind to the biomarker molecules captured by antibody, which provides accurate assessment of the biomarker concentration without the need of reaching thermal equilibrium via lengthy incubation. The real-time counting together with super-localization tracking of each nanoparticle allows resolving two binding events within a distance smaller than the diffraction limit,22 which enhances the dynamic range and minimizing the counting error. Using TD-Immunoassay, we have achieved a limit of detection of ~2.8 pg/mL, dynamic range of 4.2–1.25×104 pg/mL, and total detection time of ~25 mins for procalcitonin (PCT), an important biomarker for sepsis.23
RESULTS AND DISCUSSION
Principle and setup.
A gold-coated glass slide was used as the sensor, on which a capture antibody for PCT (procalcitonin) was immobilized via a Polyethylene Glycol (PEG) linker (Fig. 1a). A sample was introduced to allow binding of PCT in the sample to the capture antibody. A detection antibody, biotinylated anti-PCT, was then added to bind with the PCT bound to the capture antibody and form a capture antibody-PCT-detection antibody complex, followed by introducing streptavidin-coated gold nanoparticles (GNP), which bound to the antibody complex via the biotin-streptavidin interaction. The surface density of the capture antibody was adjusted such that each gold nanoparticle could bind only with one PCT molecule (see supporting information S8 for more details), providing readout of a PCT binding event on the surface.
Sensitive detection of single GNPs was achieved with a plasmonic imaging platform, where p-polarized light with a proper incident angle from a superluminescence (SLED) diode was directed onto the sensor surface via a 60× high numerical aperture oil immersion objective to excite plasmons on the gold surface. The scattered and reflected light was collected with the same objective and imaged with a CCD camera. The dynamic binding process of the individual GNPs was captured by the time sequence of the plasmonic images with a temporal resolution of ~37.5 ms (Fig. 1b). Each GNP was revealed as a bright spot with a parabolic pattern arising from the scattering for the plasmonic wave on the sensor surface by the GNP (colored inset at lower right of Fig. 1), which provides high image contrast and facilitates precise tracking of single GNPs.24,25 Using an automated imaging processing algorithm, we tracked the position of each individual GNPs and counted the individual GNPs binding to the surface over time, from which a standard curve was obtained for quantitative PCT concentration measurement (Fig. 1b).
The automated particle counting algorithm consists of following steps. First, noise in the image sequence is reduced by performing spatial Fast Fourier transform, and then differential images are obtained by subtracting the first image frame from the rest of the frames, followed by time average over multiple frames.26 These procedures remove common noise in the optical system and provides high contrast images of single GNPs (Fig. 2a). Second, the algorithm uses the distinct parabolic pattern of a single GNP plasmonic image as a template to search and identify all the GNPs on the sensor surface with an autocorrelation pattern recognition algorithm (see Supporting Information for more details). Because both the binding and unbinding of the GNPs take place on the sensor surface dynamically, two opposite patterns (inverted image contrast) are observed for the binding (top of Fig. 2b) and unbinding events (bottom of Fig. 2b), respectively. The former is: GNP image subtract background, and the latter is: background subtract GNP image, so the images of binding and unbinding events are inverted in contrast. This procedure allows the algorithm to differentiate and track both binding and unbinding processes over time (Fig 2c). Third, the spatial location of each GNP was determined and tracked with a procedure described in Supporting Information, S2. Tracking of binding events within a distance smaller than the diffraction limit was achieved with super-resolution fluorescent microscopy by Gooding et al. 20 This was achieved here with the dynamic tracking capability of the plasmonic imaging, which resolves multiple binding events within an area smaller than the optical diffraction limit in time domain, as long as no two or more GNPs bind to the area at the same moment (defined by the frame rate). This capability is critical for improving the counting accuracy and expanding the dynamic range.
Validation of TD-immunoassay with IgG/anti-IgG binding.
To validate the capability of TD-immunoassay for detecting antigen and quantifying antigen concentration, we first applied it to study the binding of IgG binding to anti-IgG. IgG was immobilized onto the sensor surface, followed by incubation with anti-IgG conjugated GNPs at different concentrations and the binding of the anti-IgG to the IgG on the sensor surface was tracked by counting the individual GNP binding events on the surface with a temporal resolution of ~37.5 ms. The number of GNPs counted with the algorithm varies with concentration, but for each concentration it increases linearly with time (Fig. 2d). This indicates that binding and unbinding of GNPs within the time interval is far from reaching saturation of the binding sites (IgG) on the sensor surface by the anti-IgG, and also far from reaching thermal equilibrium.
From the GNP counts obtained at 20 min, we obtained a standard curve of IgG detection (Fig. 2e). We repeated the experiment 3 times and found less than 20% variability for each concentration. From the standard deviation of the triplicate tests, error bars were determined and marked in Fig. 2e. The limit of detection was determined by the mean concentration measured for the blank solutions plus three times of standard deviations,27 which is 64.6 fg/mL (or 0.43 fM, blue dashed line of Fig. 2e). The detectable dynamic range was from 1.3 pg/mL to 50 μg/mL, covering 7 orders of magnitude. These results demonstrate the feasibility and performance of TD-immunoassay.
PCT detection.
Detection of PCT was based on sandwich assay, where a sample containing PCT was introduced and incubated for 10 min to allow binding of PCT to the capture antibody immobilized on the sensor surface (see supporting information for detailed procedure). Biotinylated detection antibody was then introduced followed by introducing streptavidin-GNPs, which bind to the detection antibody via the biotin-streptavidin interaction. The dynamic binding process was tracked over time to determine PCT concentration in each sample. Fig. 3a shows the GNP counts for different concentrations of PCT in reagent diluent buffer over 20 min recorded with a frame rate of 26.6 fps. The PCT concentration ranges from 0 (blank) to 1.25×104 pg/mL, which covers the concentration range of sepsis patients (0–104 pg/mL). However, the dynamic range of the present TD-immunoassay estimated from the maximum density of GNPs is significantly higher than demonstrated here (see discussion). Fig. 3b shows the standard curve of the PCT test at the 20-minute time point, showing a linear response of 4 logs ranging from 4 pg/mL to 1.25×104 pg/mL (R2 = 0.9981). The error bars in the standard curve are the standard deviations of triplicate tests. The limit of detection determined by the mean concentration measured for the blank solutions plus three times of standard deviation is 2.8 pg/mL, which is marked by the horizontal dashed red line in Fig. 3b. The limit of quantification defined by the mean concentration measured for the blank solutions plus ten times of the standard deviation is 4 pg/mL. This performance is sufficient to cover sepsis detection, where the PCT level varies from 50 to 10000 pg/mL. The practical limit of detection is affected by several factors, such as nonspecific binding levels, counting statistics and antibody quality.
For comparison, the standard curve (Fig. S4) obtained by the conventional ELISA for PCT shows a logarithmic response of optical density with from 31.3 pg/mL to 2000 pg/mL and a limit of detection of 31.3 pg/mL. The standard curve from the present plasmonic imaging-based TD-immunoassay (Fig. 3b) has a broader dynamic range (4 logs vs. 2 logs), a lower limit of detection (2.8 pg/mL vs. 31.3 pg/mL) and a lower limit of quantitation (4 pg/mL vs. 31.3 pg/mL).
The time-resolved detection of single GNPs helps improve the dynamic range and minimizing detection error associated with the binding of multiple GNPs to an area smaller than the spatial resolution of optical imaging. We show below that the real-time detection also helps to optimize the detection time required to achieve a desirable detection limit and precision. We studied the detection limit and precision by measuring GNP count vs. PCT concentration with GNP counting time of 1, 2, 5 and 15 min, respectively (Fig. 4). Three replicates were carried out for each concentration and for each GNP counting time, which are shown as black dots in Fig. 4. As expected, shorter GNP counting time leads to more scattered data due to smaller number of GNPs counted, but the GNP count is proportional to the logarithm of PCT concentration in each case. To evaluate the detection limit and concentration precision, we used a statistical model (see materials and methods, statistical analysis) to determine 95% prediction intervals (marked by the blue and red shaded regions in Fig. 4). The analysis shows that the error decreases with increasing GNP counting time, but the 5-min error is about the same as the 15-min error. This indicates that 5-min GNP counting is sufficient for precise PCT quantification.
We further examined the detection limit and precision of TD-immunoassay by comparing the GNP counts over time at different PCT concentrations, including blank solutions (Figs. 5a-d). The data shows once again that the detection limit and precision improve with GNP counting time. The data also confirms that 5 min detection time can lead to detection of PCT concentration of 3.9 pg/mL. We evaluated the limit of quantification and prediction accuracy vs. time, which shows high precision data can be obtained within 5 mins, and longer time does not significantly improve the precision (Figs. 5e-f). For this reason, we selected 5 min as the optimized counting time. At 5 min, the limit of detection is ~2.8 pg/mL, and the limit of quantification is 4.2 pg/mL. Considering that the experiment involved 10 min incubation for PCT binding to the capture antibody and 10 min incubation for detection antibody binding to the PCT bound to the capture antibody, the total detection time is 25 min. The detection time achieved here compares favorably with other high-performance detection techniques (Table 2) and may be further shortened by eliminating the second incubation step using detection-antibody conjugated GNPs. It is worth mentioning that the higher concentration of the biomarker the shorter the overall detection time will be. This provides a possibility of detecting a biomarker using only the time required to reach a fixed precision, rather than using a fixed time interval regardless the biomarker concentration.
Table 2.
TD-immunoassay (This work) | Convention al ELISA1 | ELECSYS BRAHMS PCT (Roche)2 | Quanterix SiMoA28 | SMC×PRO™ (MilliporeSigma)3 | |
---|---|---|---|---|---|
Sample volume (μL) | 100 | 100 | 30 | 100 | 1.6–3 |
Incubation time of PCT | 10 min | 2 hour | 18 min | 20 min | 1 h |
Total assay time 4 | 25 min | 1–2 days | 18 min | > 3 h | 1 day |
LoD (pg/mL) | 2.8 | 30 | 20 | 0.44 | Not available |
LoQ (pg/mL) | 4.2 | >30 | 60 | 1.23 | 0.05–2 |
Dynamic range (pg/mL) | 4–12500 | 31.3–2000 | 20–10000 | 1.23–900 | > 4 logs |
Data obtained with the ELISA kit.
Based on the specs of SMC×PRO™ (http://www.emdmillipore.com).
Time from sample (e.g., serum and plasma) to result, and preparation time, such as pre-coating of the capture antibody, reagent preparation, and surface blocking, is excluded.
Comparison of TD-Immunoassay with conventional ELISA method using PCT spiked serum samples.
To compare the present TD-Immunoassay method with the conventional ELISA method, PCT spiked serum samples with concentrations of 1.9, 3.9, 7.8, 20, 31.3, 200, 1000, 2000, 6250, and 12500 pg/mL were measured by the both methods, with each measurement repeated three times. The TD-immunoassay measurement was performed on a new gold chip for each replicate. The TD-immunoassay results were obtained with 5 min GNP counting time using the PCT standard curve shown in Fig. S5. Fig. S11 shows PCT spiked in serum can be measured precisely by the PCT TD-immunoassay standard curve. As shown in Table 1, both the TD-Immunoassay and conventional ELISA methods can accurately measure the PCT concentrations between 31.3 and 2000 pg/mL. However, the conventional ELISA does not have sufficient limit of detection to measure samples with PCT concentrations lower than 31.3 pg/mL. In contrast, the present method can accurately measure PCT in serum with concentrations as low as 2.1 pg/mL (Fig. S11a, Supporting Information).
Table 1.
PCT Concentration (pg/mL) |
TD-immunoassay* (pg/mL) (This work) |
Recovery (%) |
Replicates | Conventional ELISA (pg/mL) |
Recovery (%) |
Replicates |
---|---|---|---|---|---|---|
1.9 | 1.8±0.3 | 95.7±19.7 | 3 | Not quantifiable | Not quantifiable | 3 |
3.9 | 4.0±0.5 | 103.4±11.8 | 3 | Not quantifiable | Not quantifiable | 3 |
7.8 | 8.6±0.5 | 110.7±7.2 | 3 | Not quantifiable | Not quantifiable | 3 |
20 | 21.4±0.4 | 107.0±2.1 | 3 | Not quantifiable | Not quantifiable | 3 |
31.3 | 36.3±4.0 | 115.9±12.8 | 3 | 29.6±6.5 | 94.6±14.3 | 3 |
200 | 215.7±30.0 | 95.5±15.0 | 3 | 197.6±7.8 | 98.5±3.9 | 3 |
1000 | 886.3±89.3 | 88.6±8.9 | 3 | 927±97.0 | 92.7±9.7 | 3 |
2000 | 2010.0±284.8 | 100.5±14.2 | 3 | 1923.2±114.4 | 96.2±5.7 | 3 |
6250 | 6132.2±790.5 | 98.1±12.6 | 3 | Not quantifiable | Not quantifiable | 3 |
12500 | 12463.1±1962 | 99.7±15.7 | 3 | Not quantifiable | Not quantifiable | 3 |
Results for TD-immunoassay were obtained the standard curve at 5 min (Supporting Information). Recovery (%) is the measured PCT concentration over the actual PCT concentration, which describes the accuracy of each technique.
Comparison of TD-Immunoassay with different the state-of-the-art immunoassays.
Table 2 compares the performance of different immunoassays for PCT detection in terms of sample volume, incubation time, total detection time, limit of detection, limit of quantitation, and dynamic range. The TD-immunoassay platform presented here can detect PCT in 100 μL of serum samples with a detection limit of 2.8 pg/mL and a limit of quantitation of 4.2 pg/mL for a total of 25 min, and dynamic range of ~4 logs. This performance is excellent comparted other sensitive PCT detection technologies listed in the table. In addition to incubation time, one needs to consider time required to prepare samples and reagents, calibrate instrument and generate a standard curve. In our method, the sensors can be pre-fabricated similar to the commercial SPR sensor chips, and digital counting of nanoparticles does not involve calibration of the optical setup. SiMoA has a low limit of detection, but also a small dynamic range, making it unsuitable for high concentration PCT detection. SMC×PRO™ is also sensitive, but it involves long incubation time and total assay time. The most significant improvement of TD-immunoassay is the reduction of the total assay time from hours to ~25 min. This is particularly important for acute diseases, such as sepsis.
TD-immunoassay is based on counting of GNPs binding to a sensor surface. The relative error of counting scales with , where N is the GNP count, so the detection precision increases with N. The higher concentration of a biomarker, the faster is the binding rate of GNPs, which means shorter detection time to reach a desired precision. This provides a possibility to detect biomarker by tracking the time required to reach a required precision, rather than by counting GNPs over a fixed time interval regardless the biomarker concentration. Considering that patients with positive sepsis tend to have higher biomarker concentrations than normal people, this time resolved detection approach provides fast diagnosis of the patient conditions. In addition to shorten the detection time, time-resolved tracking of single GNPs also helps expand the dynamic range. This is because for highly concentrated biomarker, one can choose to shorten the detection time to avoid saturation associated with too many GNPs binding to the sensor surface.
Although we focus here on 150 nm-diameter GNPs, we also tested 40 nm-diameter GNPs, and found results in agreement with the 150 nm-GNPs (Supporting Information, S9). This indicates that the mass transport is not the rate-limiting step of GNP binding to the detection antibody. The plasmonic imaging contrast of 150 nm-diameter GNPs is better than that of 40 nm-diameter GNPs, which help minimize errors in the counting of the GNPs.
CONCLUSIONS
We have developed a time-resolved digital immunoassay by combining plasmonic imaging and automated counting of single gold nanoparticles. The high contrast plasmonic imaging allows accurate tracking of each nanoparticle. This capability, together with real-time imaging, enables resolving multiple nanoparticle binding to an area within the diffraction limit, and leads to highly precise detection of single binding events with a wide dynamic range. We validated the principle and demonstrated the performance of the immunoassay using PCT in controlled buffer and in sera with a limit of detection of 2.8 pg/mL, limit of quantification of 4.2 pg/mL, dynamic range of 4–1.25×104 pg/mL, and detection time of less than 25 mins for low concentration samples (a few pg/mL). Our data also shows that the real-time counting can significantly shorten the detection time for high concentration samples (relevant to sepsis patients) and increase the detection dynamic range. Further studies underway include testing of clinical samples. We anticipate that this time-resolved digital immunoassay is particularly useful for diagnosing and tracking progression of acute diseases (e.g., sepsis and cardiovascular diseases), where rapid and precise biomarker quantification is needed.
METHODS
Chemicals.
N-hydroxysuccinimide (NHS), N-ethyl-N’-(3-dimethylaminopropyl) carbodiimide hydrochloride (EDC), sodium acetate (NaOAc), acetic acid (AcOH), Tween 20, rabbit Immunoglobulin G (IgG) and bovine serum albumin (BSA) were purchased from Sigma-Aldrich (St. Louis, MO). 150 nm and 40 nm gold nanoparticles coated with goat anti-rabbit IgG (C11–150-TGARG-50), and with streptavidin (C11–150-TS-DIH-50) were from Nanopartz™. Dithiolalkanearomatic-PEG3-OH (Dithiol-PEG-OH) and dithiolalkanearomatic-PEG6-COOH (Dithiol-PEG-COOH) were purchased from SensoPath Technologies (Bozeman, MT). Human PCT ELISA kit (catalog number: dy8350–05), and DuoSet® ancillary reagent kit 2 (catalog number: DY008) were purchased from R&D system, USA. This kit contains mouse anti-human procalcitonin capture antibody, biotinylated sheep anti-human procalcitonin detection antibody, and recombinant human procalcitonin standard which could be used to sandwich ELISA assay. Human serum (from human male AB plasma, USA origin, sterile-filtered) was purchased from Sigma-Aldrich, USA. The kits and reagents mentioned above were sub-packed and stored according to the instructions.
Plasmonic imaging.
Plasmonic imaging was implemented on an inverted optical microscope (IX-81, Olympus, Shinjuku, Tokyo, Japan) with a 60× high numerical aperture (NA 1.49) oil immersion objective. A collimated p-polarized light beam (1 mW) from a 680 nm fiber coupled super luminescence diode (from Qphotonics, Ann Arbor, MI) was directed onto a sensor surface via the objective to excite surface plasmons. The spatial resolution was ~0.61λ/NA=278 nm. The plasmonic images were collected by a CCD camera (Pike, F-032B, Allied Vision Technologies, Newburyport, MA) at a frame of 106 frames per second with 4 frames binning (effective frame rate, 26.6 fps) with a view area of 640 × 480 pixels and a pixel size of 7.4 μm. The sensor chips were prepared by coating BK-7 glass coverslips with 1.5–2 nm chromium, followed by 47 nm gold. These chips were batch produced and stored until use. A Flexi-Perm silicone solution cell (SARSTEDT, Germany) was placed on top of the sensor to hold the sample solution. Automated particle counting from the images were performed with an imaging processing algorithm described in the supporting information (section S2).
Sample preparation.
50 μg/mL rabbit IgG solution was prepared by dissolving rabbit IgG in NaOAc/AcOH buffer (10 mM pH=5.0 NaOAc/AcOH). The solution was diluted via multiple serial dilution, 10× each, to reach different concentrations of rabbit IgG, from 5×10−9 to 5×10−1 μg/mL. PCT capture antibody (240 μg/mL) stock solution was diluted with NaOAc/AcOH buffer (10 mM pH=5.0 NaOAc/AcOH) to reach 2 μg/mL. PCT standard was diluted with reagent diluent (solution from DuoSet® ancillary reagent kit 2) using double ratio dilution method, to reach final of 12500, 6250, 2000, 1000, 200, 31.3, 7.83, 3.9, and 1.9 pg/mL. PCT spiked serum samples were prepared by diluting the PCT standard into human serum to reach 12500, 6250, 2000, 1000, 200, 31.3, 7.83, 3.9, and 1.9 pg/mL. PCT detection antibody (3 μg/mL) stock solution was diluted with reagent diluent to reach 50.00 ng/mL. 150 nm goat anti-rabbit IgG coated gold nanoparticle solution was prepared by diluting the stock solution 1000 times with DI water and then sonicated for 5 min. Streptavidin coated gold nanoparticle solution was prepared by adding 15 μL of gold nanoparticle solution to 135 μL of PBS buffer and then sonicated for 5 min.
Sensor surface modification.
The sensor chips were cleaned with deionized water (Milli-Q, Millipore Corp.) and then ethanol, followed by hydrogen flaming. After cleaning, the sensors were immediately soaked into 1 mM of 50:1 PEG-OH/PEG-COOH mixed dithiol ethanol solution for 24 hours in dark. The sensors were then rinsed with deionized water and ethanol, and then dried with nitrogen gas. Each sensor was activated for immobilization of a receptor molecule to its surface by adding 100 μL NHS/EDC aqueous solution (containing 100 mM NHS and 400 mM EDC), followed by gradient washing (gradually diluted the NHS/EDC solution by DI water).
IgG binding to anti-IgG.
IgG was immobilized to an activated sensor surface by incubating the sensor in 100 μL of prepared rabbit IgG solution at room temperature for 10 min, followed by washing it with 250 μL of PBST buffer (0.05% Tween 20 in PBS buffer, Corning Cellgro) three times. Residual activated binding sites were blocked by 250 μL BSA solution (1% w/v) for 5 min and then washed with 250 μL of PBST buffer three times. 100 μL solution of gold nanoparticles coated with goat anti-rabbit IgG was added, and the binding of the gold nanoparticles to IgG was tracked in real time.
Detection of PCT.
PCT capture antibody was immobilized to an activated sensor by incubation with 100 μL PCT capture antibody solution for 10 min at room temperature, followed by washing with 250 μL of PBST three times. Residual activated binding sites were blocked by 250 μL of reagent diluent buffer (containing 1% of BSA solution w/v) for 5 min and then washed with 250 μL of PBST buffer three times (Fig. S10 indicated that 5 min is enough for blocking). The capture antibody-functionalized sensor was either used immediately or stored at 4 °C prior to PCT assay. The detailed protocol of PCT immune-sandwich assay was described in the Supporting Information (Fig. S1).
Conventional ELISA assay for PCT.
Following the manufacturer’s protocol, 100 μL capture antibody was added to each well of a microplate and incubated overnight at room temperature. Next, the microplate was washed with PBST buffer for three times, and then blocked with 300 μL reagent diluent for at least 1 hour. After washing three times with PBST buffer, 100 μL standard or sample was added and incubated for 2 hr at room temperature, and then washed three times. Next, 100 μL detection antibody was added and incubated for 2 hr, and then washed three times. Next, 100 μL of streptavidin-HRP was added and incubated for 20 min followed by three times washing. Next, 100 μL substrate solution was added to the well and incubated for 20 min, and finally 50 μL stop solution was added. The optical density at 450 nm was measured using a microplate reader (model: Varioskan Flash, ThermoFisher Scientific).
Statistical Analysis.
We use linear regression model for PCT counting error prediction and 95% prediction interval to evaluate the prediction accuracy from the counting numbers (see supporting information for details). Data is processed by custom-written Matlab scripts.
Supplementary Material
ACKNOWLEDGMENT
The authors are grateful of support from NIH (1R01GM124335) and NSFC (#21773117) (YW).
Footnotes
ASSOCIATED CONTENT
This work has been previously submitted to a pre-print server: Jing, W.; Wang, Y.; Yang, Y.; Wang, Y.; Ma, G.; Wang, S.; Tao, N. Time-resolved digital immunoassay for rapid and sensitive quantitation of procalcitonin with plasmonic imaging. 2018, Doi: https://doi.org/10.1101/483347. Biorxiv. https://www.biorxiv.org/content/early/2018/11/29/483347.full.pdf.
Supporting Information.
The Supporting Information is available free of charge on the ACS Publications website.
Protocol of TD-immunoassay; time-resolved tracking of single nanoparticles with plasmonic imaging; TD-immunoassay detection of IgG/anti-IgG binding; evaluation of time dependence of TD-immunoassay; standard curve of conventional ELISA for PCT detection; dynamic range of TD-immunoassay; statistical analysis; molecular spacing on the sensor surface; comparison of two different sizes of GNPs; AFM image of the gold nanoparticles onto the surface; monitoring of blocking with SPR; PCT detection in serum; comparison between PCT-spiked human serum and PCT-spiked bovine serum (non-human PCT); dependence of particle counts on capture antibody concentration; confirmation of SPR binding events with SEM; PCT level of blank human serum; PCT standard curve with a highly sensitive ELISA kit; PCT level of blank human serum (PDF).
The authors declare no competing financial interest.
REFERENCES
- (1).Diaconu CC; Bălăceanu A; Moroşan E Sepsis Biomarkers. Past, Present and Future. Farmacia 2015, 811–815. [Google Scholar]
- (2).Mordente A; Meucci E; Martorana GE; Silvestrini A Cancer Biomarkers Discovery and Validation: State of the Art, Problems and Future Perspectives. Adv. Exp. Med. Biol 2015, 9–26. [DOI] [PubMed] [Google Scholar]
- (3).Solier C; Langen H Antibody-Based Proteomics and Biomarker Research-Current Status and Limitations. Proteomics 2014, 774–783. [DOI] [PubMed] [Google Scholar]
- (4).Uludag Y; Tothill IE Cancer Biomarker Detection in Serum Samples Using Surface Plasmon Resonance and Quartz Crystal Microbalance Sensors with Nanoparticle Signal Amplification. Anal. Chem 2012, 5898–5904. [DOI] [PubMed] [Google Scholar]
- (5).Satija J; Punjabi N; Mishra D; Mukherji S Plasmonic-ELISA: Expanding Horizons. RSC Adv. 2016, 85440–85454. [Google Scholar]
- (6).Cheng Z; Choi N; Wang R; Lee S; Moon KC; Yoon SY; Chen L; Choo J Simultaneous Detection of Dual Prostate Specific Antigens Using Surface-Enhanced Raman Scattering-Based Immunoassay for Accurate Diagnosis of Prostate Cancer. ACS Nano 2017, 4926–4933. [DOI] [PubMed] [Google Scholar]
- (7).Zhao Y; Cao M; McClelland JF; Shao Z; Lu M A Photoacoustic Immunoassay for Biomarker Detection. Biosens. Bioelectron 2016, 261–266. [DOI] [PubMed] [Google Scholar]
- (8).Tighe PJ; Ryder RR; Todd I; Fairclough LC ELISA in the Multiplex Era: Potentials and Pitfalls. Proteomics Clin. Appl 2015, 406–422. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (9).Gan SD; Patel KR Enzyme Immunoassay and Enzyme-Linked Immunosorbent Assay. J. Invest. Dermatol 2013, E10–E12. [DOI] [PubMed] [Google Scholar]
- (10).Desmet C; Blum LJ; Marquette CA Multiplex Microarray ELISA versus Classical ELISA, a Comparison Study of Pollutant Sensing for Environmental Analysis. Environ. Sci. Process. Impacts 2013, 1876–1882. [DOI] [PubMed] [Google Scholar]
- (11).Rissin DM; Kan CW; Campbell TG; Howes SC; Fournier DR; Song L; Piech T; Patel PP; Chang L; Rivnak AJ; Ferrell EP; Randall JD; Provuncher GK; Walt DR; Duffy DC Single-Molecule Enzyme-Linked Immunosorbent Assay Detects Serum Proteins at Subfemtomolar Concentrations. Nat. Biotechnol 2010, 595–599. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (12).Chen S; Svedendahl M; Duyne RPV; Käll M Plasmon-Enhanced Colorimetric ELISA with Single Molecule Sensitivity. Nano Lett. 2011, 1826–1830. [DOI] [PubMed] [Google Scholar]
- (13).Cohen L; Walt DR Single-Molecule Arrays for Protein and Nucleic Acid Analysis. Annu. Rev. Anal. Chem 2017, 345–363. [DOI] [PubMed] [Google Scholar]
- (14).Fischer SK; Joyce A; Spengler M; Yang T-Y; Zhuang Y; Fjording MS; Mikulskis A Emerging Technologies to Increase Ligand Binding Assay Sensitivity. AAPS J. 2015, 93–101. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (15).Nam JM; Thaxton CS; Mirkin CA Nanoparticle-Based Bio-Bar Codes for the Ultrasensitive Detection of Proteins. Science (80-. ). 2003, 1884–1886. [DOI] [PubMed] [Google Scholar]
- (16).Thaxton CS; Elghanian R; Thomas AD; Stoeva SI; Lee J-S; Smith ND; Schaeffer AJ; Klocker H; Horninger W; Bartsch G; Mirkin CA Nanoparticle-Based Bio-Barcode Assay Redefines “Undetectable” PSA and Biochemical Recurrence after Radical Prostatectomy. Proc. Natl. Acad. Sci. U. S. A 2009, 18437–18442. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (17).Liang K; Liu F; Fan J; Sun D; Liu C; Lyon CJ; Bernard DW; Li Y; Yokoi K; Katz MH; Koay EJ; Zhao Z; Hu Y Nanoplasmonic Quantification of Tumour-Derived Extracellular Vesicles in Plasma Microsamples for Diagnosis and Treatment Monitoring. Nat. Biomed. Eng 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (18).Zhu L; Li G; Sun S; Tan H; He Y Digital Immunoassay of a Prostate-Specific Antigen Using Gold Nanorods and Magnetic Nanoparticles. RSC Adv. 2017, 27595–27602. [Google Scholar]
- (19).Liao T; Yuan F; Yu H; Li Z An Ultrasensitive ELISA Method for the Detection of Procalcitonin Based on Magnetic Beads and Enzyme-Antibody Labeled Gold Nanoparticles. Anal. Methods 2016, 1577–1585. [Google Scholar]
- (20).Zhao M; Nicovich PR; Janco M; Deng Q; Yang Z; Ma Y; Böcking T; Gaus K; Gooding JJ Ultralow- and Low-Background Surfaces for Single-Molecule Localization Microscopy of Multistep Biointerfaces for Single-Molecule Sensing. Langmuir 2018, 10012–10018. [DOI] [PubMed] [Google Scholar]
- (21).Rhodes A; Evans LE; Alhazzani W; Levy MM; Antonelli M; Ferrer R; Kumar A; Sevransky JE; Sprung CL; Nunnally ME; Rochwerg B; Rubenfeld GD; Angus DC; Annane D; Beale RJ; Bellinghan GJ; Bernard GR; Chiche JD; Coopersmith C; De Backer DP et al. Surviving Sepsis Campaign: International Guidelines for Management of Sepsis and Septic Shock: 2016. Crit. Care Med 2017, 486552. [DOI] [PubMed] [Google Scholar]
- (22).Yang Y; Yu H; Shan X; Wang W; Liu X; Wang S; Tao N Label-Free Tracking of Single Organelle Transportation in Cells with Nanometer Precision Using a Plasmonic Imaging Technique. Small 2015, 2878–2884. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (23).Liu Y; Hou J. huan; Li Q; Chen K. jun; Wang SN; Wang J. min. Biomarkers for Diagnosis of Sepsis in Patients with Systemic Inflammatory Response Syndrome: A Systematic Review and Meta-Analysis . Springerplus 2016. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (24).Wang S; Shan X; Patel U; Huang X; Lu J; Li J; Tao N Label-Free Imaging, Detection, and Mass Measurement of Single Viruses by Surface Plasmon Resonance. Proc. Natl. Acad. Sci. U. S. A 2010, 16028–16032. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (25).Yu H; Shan X; Wang S; Chen H; Tao N Molecular Scale Origin of Surface Plasmon Resonance Biosensors. Anal. Chem 2014, 8992–8997. [DOI] [PubMed] [Google Scholar]
- (26).Young G; Hundt N; Cole D; Fineberg A; Andrecka J; Tyler A; Olerinyova A; Ansari A; Marklund EG; Collier MP; Chandler SA; Tkachenko O; Allen J; Crispin M; Billington N; Takagi Y; Seller JR; Eichmann C; Selenko P; Frey L et al. Quantitative Mass Imaging of Single Biological Macromolecules. Science (80-. ). 2018, 423–+. [DOI] [PMC free article] [PubMed] [Google Scholar]
- (27).Shrivastava A; Gupta V Methods for the Determination of Limit of Detection and Limit of Quantitation of the Analytical Methods. Chron. Young Sci 2011, 21–25. [Google Scholar]
- (28).Carcamo Yañez V; Göpfert J; Otto M; Tumani H; Peter A; Joos T Development and Validation of an Ultrasensitive Procalcitonin Sandwich Immunoassay. High-Throughput 2017. [DOI] [PMC free article] [PubMed] [Google Scholar]