Abstract
Background: Pacemaker marker annotations facilitate the interpretation of device behavior in addition to ECG recordings. However, they are only available in conjunction with a programmer. We studied the diagnostic value of a prototype Telemetry Holter Decoder (THD), providing continuous marker annotations on a conventional Holter.
Methods: The study included 20 patients with VDD or DDDR pacemakers. A 24‐hour Holter was performed using the THD. Marker annotations are transmitted from the pacemaker to the THD, which transforms them into analog signals, which are recorded on one of the Holter channels.
Results: During a total recording time of 458 hours, high quality marker annotations were retrieved for every patient. Artefacts disturbed the recordings during 184 min (0.67%). The THD provided information not discernible on the ECG: intermittent atrial undersensing during sinus rhythm (1096 times). Atrial tachycardias, not visible on the ECG, were detected in 2 patients. The activation of tachycardia response algorithms was clearly annotated in 11,516 events. A total of 8875 PVC's occurred, 57.8% of which were classified incorrectly in the event counters as conducted or fusion beats. Atrial far‐field sensing or VA conduction was demonstrated 4294 times. Electromagnetic interferences, not visible on the ECG, could be seen three times.
Conclusion: Recording of continuous high‐quality marker annotation on a conventional Holter is feasible. The THD provides important information on device behavior, even in patients assumed to have regular device function, and shows to be clearly superior to ECG interpretation alone. Such data can be used for improved programming, troubleshooting and for the validation of new algorithms. A.N.E. 2002;7(1):22–28
Keywords: Holter recording, marker annotations, telemetry
REFERENCES
- 1. Warnowicz MA, Goldschlager N. Apparent failure to sense (undersensing) caused by oversensing: Diagnostic use of a noninvasively obtained intracardiac electrogram. PACE 1983;6:1341–1343. [DOI] [PubMed] [Google Scholar]
- 2. Ganz LI, Couper GS, Friedman PL, et al. Use of telemetered permanent pacemaker intracardiac electrograms to diagnose ventricular tachycardia. Am J Cardiol 1997;80:1511–1513. [DOI] [PubMed] [Google Scholar]
- 3. Wiegand UK, Bode F, Schneider R, et al. Diagnosis of atrial undersensing in dual chamber pacemakers: Impact of autodiagnostic features. PACE 1999;22:894–902. [DOI] [PubMed] [Google Scholar]
- 4. Waktare JE, Malik M. Holter, loop recorder, and event counter capabilities of implanted devices. PACE 1997;20: 2658–2669. [DOI] [PubMed] [Google Scholar]
- 5. Cazeau S, Ritter P, Nitzsche R, et al. Diagnostic of atrial arrhythmias using the Holter function of a new DDD pacemaker. PACE 1994;17:2106–2113. [DOI] [PubMed] [Google Scholar]
- 6. Lascault G, Frank R, Himbert C, et al. Diagnosis of ventricular tachycardia using a pacemaker Holter function. PACE 1994;17:1316–1319. [DOI] [PubMed] [Google Scholar]
- 7. Garrigue S, Cazeau S, Ritter P, et al. Incidence of atrial arrhythmias in patients with long term dual‐chamber pacemakers. Contribution of the Holter function of pacemakers. Arch Mal Coeur Vaiss 1996;89: 873–881. [PubMed] [Google Scholar]
- 8. Nowak B. Taking advantage of sophisticated pacemaker diagnostics. Am J Cardiol 1999;83:172D–179D. [DOI] [PubMed] [Google Scholar]
- 9. van Gelder LM, Bracke FA, el Gamal MI. Fusion or confusion on Holter recording. PACE 1991;14:760–763. [DOI] [PubMed] [Google Scholar]
- 10. Lesh MD, Langberg JJ, Griffin JC, et al. Pacemaker generator pseudomalfunction: An artifact of Holter monitoring. PACE 1991;14:854–857. [DOI] [PubMed] [Google Scholar]
- 11. Nowak B, Middelodorf T, Housworth C, et al. Holter recordings with continuous marker annotations: A new tool in pacemaker diagnostics. PACE 1996;19:1791–1795. [DOI] [PubMed] [Google Scholar]
- 12. Israel CW, Böckenförde JB. Pacemaker event counters: Possible sources of error in calculation of AV synchrony in VDD single lead systems as an example for present limitations. PACE 1998;21:489–493. [DOI] [PubMed] [Google Scholar]