Graphical abstract
Keywords: In situ 3D bio-printing, Cartilage regeneration, Tissue engineering, Bio-ink crosslinking, Robot
Abstract
The concept of in situ 3D bio-printing was previously reported, while its realization has still encountered with several difficulties. The present study aimed to report robotic-assisted in situ 3D bio-printing technology for cartilage regeneration, and explore its potential in clinical application. A six-degree-of-freedom (6-DOF) robot was introduced in this study, and a fast tool center point (TCP) calibration method was developed to improve printing accuracy. The bio-ink consisted of hyaluronic acid methacrylate and acrylate-terminated 4-armed polyethylene glycol was employed as well. The in vitro experiment was performed on a resin model to verify the printing accuracy. The in vivo experiment was conducted on rabbits to evaluate the cartilage treatment capability. According to our results, the accuracy of the robot could be notably improved, and the error of printed surface was less than 30 μm. The osteochondral defect could be repaired during about 60 s, and the regenerated cartilage in hydrogel implantation and in situ 3D bio-printing groups demonstrated the same biomechanical and biochemical performance. We found that the cartilage injury could be treated by using this method. The robotic-assisted in situ 3D bio-printing is highly appropriate for improving surgical procedure, as well as promoting cartilage regeneration.
Introduction
Over the past decades, three-dimensional (3D) bio-printing technology has been extensively applied in the field of tissue engineering. To date, several types of 3D bio-printing were developed to fabricate scaffolds for cell culture and tissue regeneration, including stereolithography, inkjet bio-printing, laser-assisted bio-printing, extrusion-based bio-printing, and electrospinning-based bio-printing [1]. Briefly, the process is carried consisting of bio-printing a scaffold in vitro, cultivation of cells using bioactive factors for few days or weeks, and finally in vivo implantation to repair the damaged region. The whole process typically lasts for a couple of weeks to months. However, the waiting period is a bit long for utilization of 3D bio-printing in clinical treatment. In order to eliminate the existing gap between laboratory and clinical findings in tissue engineering, Campbell and Weiss defined bio-printing as the selective deposition of ‘bio-inks’ of biologically active components, involving proteins, peptides, DNA, cells, hormones (e.g., cytokines, growth factors, synthetic hormonal signaling peptides, etc.), extracellular matrix (ECM) molecules, and native or synthetic biopolymers. They demonstrated that Bio-printing holds a great promise for tissue engineering, while these technologies are still in relatively early stages of development and have numerous hurdles to overcome [2]. Based on the previous achievements, almost all the elements in 3D bio-printing are obstacle for in situ additive manufacturing. For instance, the 3D bio-printing equipment is massive and the working space is small, the crosslinking of bio-inks is difficult in humoral microenvironment, and there is a remarkable uncertainly in accuracy of bio-printing. To overcome these shortcomings, a number of solutions have been recently reported to facilitate the implementation of in situ 3D bio-printing technology.
The execution mode of in situ 3D bio-printing can be divided into robotic and handheld modes. The robotic mode is similar to the conventional 3D bio-printing technology. The motion of the bio-printing unit is driven by a three-axis portable system, and the printing path is generated using a 3D bio-printing slicer software [3]. The shape of a printing objective depends on the predefined 3D geometry, and complex internal structures can be accordingly obtained. The handheld mode is entirely different from robotic mode, and the bio-printing unit is syringe-like and driven by human. In addition, predefined printing path and geometry are unnecessary, because the shape of a bio-printing objective is completely based on the motion of operator’s hand [4]. This mode is highly appropriate for fabrication of simple structures with superficial position and limited damage. The two modes both possess advantages and disadvantages. The robotic mode is highly essential for precise reconstruction of tissues, while that is costly and there are a number of ethical challenges. The handheld mode is easier for sterilization and clinical location, and the ethical challenges can be avoided as well [5]. However, it is difficult to achieve multi-biomaterial 3D bio-printing, restricting its application in the repair of complex tissues.
In 2010, a number of French scholars employed the laser-assisted bio-printing to repair the calvarial defect in situ, and they selected nano-hydroxyapatite (n-HA) as the bio-ink [6]. Although the regeneration effect was not reported to be satisfactory, their research was carried out as the first in vivo experiment of in situ 3D bio-printing. In their next research, mesenchymal stromal cells (MSCs) were utilized in the preparation of bio-ink, in which the bone repairing effect was notably improved [6]. Cohen et al. proposed the concept of the in situ repair of cartilage and osteochondral defects using in situ additive manufacturing technology, expanding the applications of 3D printing in cartilage tissue engineering [7]. In the mentioned study, a commercial 3D bio-printer was employed to complete the 3D bio-printing process on the calf femur. The alginate and demineralized bone matrix were applied to eliminate the bone defects. The whole of the 3D bio-printing process was completed by a modified commercial 3D bio-printer. Another study reported a satisfied osteochondral repairing effect that was accomplished by a robotic arm-based 3D bio-printer [8]. In addition to bone defects, the skin defects were reported to be repaired in situ using multi-nozzle inkjet print-heads driven by the three-axis movement system [9], [10]. With presenting the wound scanner, geometry of the defect could be appropriately achieved, and the injured skin could be fully repaired with the bio-ink, which provided a novel solution for early treatment and rapid closure of acute or chronic skin wounds. In contrast to the robotic-assisted 3D bio-printing system mentioned above, a number of Australian scholars focused on the developing in situ handheld 3-D bio-printing for cartilage regeneration. They presented a hand-held 3D printing device (BioPen) that allows the simultaneous coaxial extrusion of bioscaffold and cultured cells directly into the cartilage defect in vivo in a single-session surgery [4], [11]. The co-axial design of the BioPen can fabricate tubular structure with a bio-ink shell [12]. A uniform cell distribution and high cell viability could be accordingly achieved by the BioPen. In their in vivo experiment, a standardized critical-sized full-thickness chondral defect was created in the weight-bearing surface of the lateral and medial condyles of both femurs of six sheep. At 8 weeks after surgery, macroscopic, microscopic and biomechanical tests were performed, and it was unveiled that real-time, in vivo bio-printing with cells and scaffold is a feasible means of delivering a regenerative medicine strategy in a large animal model to regenerate articular cartilage. The above-mentioned studies demonstrated that the realization of in situ 3D bio-printing is highly feasible in clinical practice.
Although some optimistic progresses have been made, numerous difficulties are still existed. In our previous study, we found that the accuracy of 3D bio-printing can meet the requirements of the bone and cartilage repair [13]. However, the limited workspace of a 3D bio-printer restricts the application area, and the damaged region must be placed in the workspace. This issue is the major technical obstacle for the translation of in situ 3D bio-printing from benchside to bedside. A robust device for clinical application has mainly two typical sizes: either large enough to wrap the whole body (e.g., computed tomography (CT) or nuclear magnetic resonance (NMR)), or small in size, while large in workspace (e.g., Da Vinci surgical system or MAKOplasty robotic surgery). These two types can easily scan or complete surgery on each patient’s position. Considering the occasion and purpose of application, a 3D bio-printer with a large workspace and small size of equipment is highly convenient for surgeons. Herein, we introduced a six-degree-of-freedom (6-DOF) robot to solve the aforementioned challenge. For this purpose, a fast tool center point (TCP) calibration method was employed to improve the accuracy of 3D bio-printing. Besides, we used hyaluronic acid methacrylate (HAMA) as a bio-ink due to its excellent performance in cartilage regeneration [14], [15]. To enhance the mechanical properties and speed of photopolymerization, the acrylate-terminated four-armed polyethylene glycol (4-Armed PEG-ACLT) was introduced in the bio-ink to act as a cross-linker [16], [17]. Eventually, an in vivo study was conducted on knee joint of a rabbit to verify the practicability of the proposed approach in repair of cartilage. Our approach may enhance in situ 3D bio-printing in clinical treatment.
Materials and methods
Synthesis of bio-ink
In the present research, HAMA was synthesized as previously reported [18]. Hyaluronic acid (HA) modified with a double-bond was synthesized by reacting with the methacrylic acid (MA). In addition, 2 g HA was dissolved in 100 mL distilled (DI) water, and then, stirred in a cold room overnight, which followed by addition of 1.6 mL MA into the HA solution. The pH of the reaction was maintained between 8 and 9 by addition of 5 N NaOH and kept at 4 °C under continuous stirring for 24 h. Subsequently, HAMA was precipitated in acetone, washed with ethanol, and then dissolved in deionized water (DI) water. After doing dialysis against DI water for 48 h, the purified HAMA with a yield of 87.5% was obtained by lyophilization. The purified HAMA and 4-Armed PEG-ACLT were dissolved in the phosphate-buffered saline (PBS) solution to a final concentration of 2% w/v and 5% w/v before 3D bio-printing. The 4-Armed PEG-ACLT was purchased from Sinopec Shanghai Petrochemical Co., Ltd. (Shanghai, China), and other regents were purchased from Sigma-Aldrich (St. Louis, MO, USA).
Fabrication of robotic-assisted 3D bio-printing system
The robotic-assisted 3D bio-printing system was consisted of a 6-DOF robot, an extrusion-based bio-printing unit, an air compressor, a control cabinet, and a workbench (Figure S1). The robot had a reach of 1 m and handling capacity of 10 kg, which fully met the requirement of 3D bio-printing. According to the functions provided by the robot, we developed an off-line programming software for this system. The main interface is shown in Figure S2. In the developed software, the relevant 3D bio-printing parameters (e.g., diameter of nozzle, speed of printing, etc.) were set to generate the bio-printing process. The program was also debugged in the non-print mode to improve the bio-printing process.
The workspace of the robot was computed by the Monte Carlo method. The limited value of each joint angle was imported into the MATLAB 2016 software (MathWorks, Natick, MA, USA) to generate the random values of the joint angle using the Rand (N, 1) function. A total of 6 groups of numerical values (30000 values for each group) were substituted into the kinematic equation of the robot in an order, in which they were generated to obtain the end effector relative to the base coordinate system.
Fast calibration of the 3D bio-printing system
A traditional robot calibration technology requires expensive and complicated equipment, thus some calibration methods have been developed that only adopt internal encoder data of a robot. Briefly, in the current study, the robot was set to touch a number of special points using TCP, and the touch accuracy was judged by human eyes, which resulted in a large error in the calibration process. Thus, we utilized a fast calibration method in the present study. The calibration method accurately calculated the TCP according to the kinematic model of the robot, the distance constraints, and the measurement of the laser tracker. It was unveiled that the calibration error was reduced by the proposed approach.
The kinematic model of the robot was established using the D-H method [19], and the coordinate systems were set as illustrated in Fig. 1.
where ai represents the length of link, αi denotes the link twist, di is the link offset, and θi represents the joint angle. The link-based parameters are presented in Table S1.
According to the coordinate systems and the link-based parameters, the relationship between adjacent coordinate systems is given by
(1) |
Then, the kinematic model of the robot is formulated as
(2) |
where represents the transformation equation from {0} to {6}.
Based on the coordinate systems for calibration (Figure S3), {B} is the base coordinate system of the robot, {E} represents the coordinate system of the flange at the end of the robot, {L} is the coordinate system of the laser tracker, and C denotes the center point of the target ball. According to Eqs. (1–2), Eq. (3) is given by
(3) |
where is the transformation matrix from {B} to {E}.
The position of point C can be written as
(4) |
where EP denotes the transformation matrix from {E} to C, and (EPx, EPy , EPz) represents the coordinate value of C in {E}.
At this point, C can be represented in {B} as
(5) |
where BTC denotes the transformation equation from {B} to C, and C is the matrix obtained by substituting the link-based parameters of point C into Eq. (3).
As mentioned above, it can be concluded that EPx, EPy and EPz are unknown parameters that require solutions. If two different measuring positions M and N are chosen arbitrarily, the difference between the two points is calculated as follows:
(6) |
where ΔBPMN is the difference between the coordinates of the two points M and N in {B}, BTM and BTN are the transformation equations from {B} to M and N, respectively, and are the matrices obtained by substituting the link-based parameters of M and N into Eq. (3), respectively.
Simultaneously, the position of the two points measured by the laser tracker is
(7) |
where ΔLPMN is the difference between the coordinates of the two points M and N in {L}, and LTM and LTN are the transformation matrices from {L} to M and N, respectively.
Although the coordinate values of the points in {B} and {L} are different, the distance between M and N in the two coordinate systems is equal, that is
(8) |
It can be written as
(9) |
In Eq. (9), and can be calculated by the link-based parameters corresponding to points M and N, respectively. LTM and LTN are obtained by the laser tracker. Thus, the three unknown parameters contained in EP require solutions for the calibration. Besides, any two points in the space can formulate an equation according to Eq. (9). Therefore, at least three points are required to complete the calibration. We used the laser tracker to collect six points and the control cabinet to record the link-based parameters of the points. Consequently, three equations were formulated, and were solved to obtain EP. Eventually, the reference TCP value (-88.6049, 105.9695, 90.9429) was achieved.
According to the geometric size and the reference TCP value, the TCP value of the nozzle was computed by the off-line programming software.
In order to verify accuracy of calibration for TCP, the transformation matrix from {L} to {B} was calculated. Theoretically, C in {L} can be expressed as
(10) |
where LTC represents the coordinate value of C in {L}.
In the solution of , LTC could be obtained by laser tracker, and BTC could be calculated by Eq. (5). Using LTC=·BTC and four measuring points, the non-homogeneous linear equations with twelve unknowns could be formulated, and then, was attained.
In the analysis of accuracy of TCP, we selected LTC=· ·EP in Eq. (10). At this time, and were found as known numbers. The theoretical value of LTC was achieved by substituting the TCP value obtained by the fast calibration method into EP. The actual value of LTC was obtained by the laser tracker. By subtracting these two values, the accuracy of the fast TCP calibration method was calculated. Analogously, the accuracy of the traditional TCP calibration method could be obtained by substituting EP in the traditional TCP calibration method into the equation.
The 3D bio-printing path planning
Herein, the 3D bio-printing target was taken as a cylinder into account, and the printing path was designed to fill this specific geometry. The coordinate system was set in the target. As illustrated in Figure S4a, point A and point B represent the positions of 3D bio-printing nozzle, and point C shows any point in the bottom.
The center of the user coordinate system U was set as U = [Ux, Uy, Uz], where Uz could be obtained from the point C, which meant that the Z coordinate of point C in the coordinate system {B} was Uz. Besides, Ux and Uy were calculated by the relationship between point A and point B using the following equation
(11) |
where (xA, yA) and (xB, yB) denote the coordinate values of X-axis and Y-axis for point A and point B in the coordinate system {B}, respectively. R represents the radius of the defect, and r is the radius of the 3D bio-printing nozzle.
Then, the origin position of the user coordinate system could be achieved. The orientation of each axis of user coordinate system was the same as {B}.
After the establishment of the user coordinate system, the bio-printing path was planned based on the parametric line method, and the diagram of the path is schematically shown in Figure S4b. In this algorithm, the user coordinate system was set as the coordinate system of path planning, and the number of vertical lines N in the trajectory could be calculated by Eq. (12).
(12) |
The abscissa sequence V(x) of each target point could be generated as
(13) |
The ordinate of each vertical line could be attained by Eq. (14),
(14) |
where the positive (negative) of y(x) represents the target point of the vertical line that is above (below) the X-axis.
The number of printed layers is given by
(15) |
where H denotes the height of the damaged region.
Then, the coordinate sequence of Z-axis can be obtained as follows
(16) |
According to Eqs. (12), (13), (14), (15), (16), the coordinate value of each target point can be calculated. With connecting the points with straight lines, then, the desired 3D bio-printing path could be generated.
In vitro experiment
To verify the validity of the 3D bio-printing path, an in vitro experiment was performed on a resin femoral condyle model. The model was established with the help of micro CT images of a 6-month-old female New Zealand rabbit, and fabricated using a digital light processing (DLP) 3D bio-printer (Prism, Shanghai, China). A cylinder defect with a diameter of 5 mm and a height of 4 mm was made on the model to mimic an osteochondral injury.
The defect was printed using the robotic-assisted in situ 3D bio-printing method with the bio-ink. A 3D scanner (Shining 3D Tech Co., Ltd., Hangzhou, China) was employed to obtain the post-printed geometry. A “3D Samples Comparison” operation was conducted using the Geomagic Control software (3D Systems Inc., Rock Hill, SC, USA) to evaluate the accuracy of bio-printing as described previously [8], [13]. The robot calibrated using a traditional method was considered as control group in the current test.
In vivo experiment
The in vivo test was performed on female New Zealand rabbits. The animal study was carried out in compliance with the regulations and guidelines of the Ethics Committee of Drum Tower Hospital affiliated to the Medical School of Nanjing University (Nanjing, China), and conducted according to the Institutional Animal Care and Use Committee (IACUC) guidelines.
A total of 12 rabbits (body weight, 2.5 kg) were enrolled in this study and randomly divided into 3 groups, including control group, hydrogel implantation group, and in situ bio-printing group. A lateral para-patellar incision was made on the articular surface, and an osteochondral defect with a height of 4 mm and a diameter of 5 mm was made in the center of the trochlear groove. The defects in the in situ bio-printing group were directly repaired through the 6-DOF robot using the bio-ink; the defects in the hydrogel implantation group were implanted; and the defects in the control group remained blank. Cefuroxime sodium was injected intramuscularly for 3 days after operation to avoid infection. All animals were postoperatively sacrificed at week 12 to evaluate the chondrogenic capacity.
The surface of the defective region was assessed according to color, integrity, contour, and smoothness, and the repair quality was consequently recorded. The International Cartilage Repair Society (ICRS) macroscopic scoring system was used to assess the macroscopic appearance of the repaired tissue, and the scoring criteria are presented in Table S2 [20].
The biomechanical properties of the regenerated region and the correlated healthy part were evaluated using a nanoindenter (Nanomechanics, Oak Ridge, TN, USA) according to a previous study [21]. The force was set at 1.5 mN, which maintained as long as 5 s after reaching the maximum loading force, and then removed to reach the recovery curve. Each sample had 5 indentation test points.
The surface tomography of the regenerated cartilage and healthy cartilage was observed using a scanning electron microscope (SEM; Hitachi, Tokyo, Japan). After nanoindentation testing, the samples were dehydrated using ethanol solution and freeze-dried. Finally, they were sputter-coated with platinum and imaged with SEM.
The samples for biochemical testing were fixed in 10% formalin for 24 h, and then decalcified in 15% ethylenediaminetetraacetic acid (EDTA) for 28 days. The samples were then embedded into paraffin and cut into 5 μm thick sections. The sections were then stained with toluidine blue, Safranin O, and collagen II. All the sections were observed using a microscope equipped with a charge-coupled device (CCD) camera (Olympus, Tokyo, Japan). The histological results were assessed using the O’Driscoll scoring system [22]. The scoring criteria are shown in Table S3.
Statistical analysis
The macroscopic and histological results were analyzed by 3 investigators who were blinded to the groups. The statistical analysis and exponential curve fitting were performed using SPSS 19.0 software (IBM, Armonk, NY, USA) and Igor Pro 6.12 software (WaveMetrics Inc., OR, USA). Data were presented as mean ± standard deviation, and evaluated using an unpaired Student’s t-test. P less than 0.05 was considered statistically significant.
Results
Accuracy of the 6-DOF robot
After the calibration process was completed, 16 points were selected as the measurement objects. The results of the movement error are depicted in Fig. 2. Compared with traditional calibration method, the fast TCP calibration method notably reduced the volatility of curves in all axes. The average errors of position accuracy for the X, Y, and Z axes in traditional calibration method were 0.0224 ± 0.5009, 0.1574 ± 0.6900, and −0.2626 ± 0.6144 mm, respectively. These values were decreased respectively to −0.0089 ± 0.1702, 0.0025 ± 0.2773, and −0.0128 ± 0.2143 mm using the fast TCP calibration method. The runout values (subtraction of minimum error from maximum error) of various axes were reduced from 1.6747, 2.0578, and 1.8771 mm to 0.6380, 0.8108, and 0.6538 mm, respectively. The average movement error of these points decreased from 1.050 ± 0.1525 mm to 0.3719 ± 0.0226 mm, indicating improved accuracy and stability of the robot.
The workspace of the 6-DOF robot is displayed in Figure S5. It could reach a maximum distance of 1.188 m and a maximum height of 1.454 m.
Evaluating the feasibility of in situ 3D bio-printing
Next, we quantitatively measured the bio-printing accuracy in vitro once the fast TCP calibration and path planning were conducted. The whole process is shown in Figure S6. The air pressure applied in the present study was 0.2 MPa, the speed was set to 6 mm/s, the thickness of layer was 320 μm, and the diameter of the nozzle was 400 μm. It was revealed that the defect region was filled by the bio-ink according to the pre-defined bio-printing path. The general view from three directions unveiled that the objective fabricated by in situ 3D bio-printing could match the geometry of defect area. To further explore the accuracy of in situ 3D bio-printing, the “3D Sample Comparison”was used and the results are illustrated in Fig. 3. The colors on the surface were corresponding to the error magnitude, which meant that the 3D error gradually increased in positive direction as the color transferred from green to red, and gradually increased in negative direction as the color transferred from green to deep blue.
As shown in Fig. 3a, the bio-printed area in control group (traditional calibration) was mainly covered by blue (11.5770 mm2), and the green and yellow areas were 4.0180 and 2.5470 mm2, respectively. The distribution of color was changed in Fig. 3b, in which the green area was notably increased to 9.8820 mm2, and the blue and yellow areas were 3.9970 mm2 and 4.2120 mm2, respectively. This phenomenon demonstrated that the most of the surface geometry can completely fit that of the healthy cartilage. The statistical results presented in Table 1 confirmed this conclusion as well. With the help of the fast TCP calibration method, the average error was significantly decreased and the majority of the points were in the tolerance interval, and the dispersion was also decreased. Thus, it can be concluded that the stability and accuracy of 6-DOF robot are highly appropriate for in situ 3D bio-printing, and the effect of bio-printing path plan is accordingly satisfactory.
Table 1.
Calibration Method | Average (mm) | Standard deviation (mm) | Dispersion | Within tolerance (%) |
---|---|---|---|---|
Traditional | −0.1550 | 0.1704 | 0.0291 | 30.4617 |
Fast TCP | −0.0292* | 0.1447 | 0.0209 | 54.4188* |
In situ 3D bio-printing for cartilage regeneration
Finally, we explored the effects of direct in situ 3D bio-printing on living animals to evaluate its treatment efficacy. The bio-printing parameters were the same as to those in the in vitro experiment (Fig. 4). The osteochondral defect (ICRS grade IV) was made by mosaicplasty. The bio-printing nozzle was dragged to the starting point of the bio-printing path, and the in situ 3D bio-printing was carried out according to the planned path. Due to the effects of ultraviolet (UV) light on the camera, the light was turned off during video and photograph shootings, and the photopolymerization was undertaken at the end of the direct bio-printing procedure. The whole process was completed during 60 s, and the experiment was repeated for three times to verify its feasibility.
All the samples were postoperatively harvested for 12 weeks. The surface tomography in healthy group, implantation group, and in situ bio-printing group was similar according to the SEM images (Figure S7a). Those images all exhibited a smooth appearance. The biomechanical test demonstrated that the mechanical properties in implantation group and in situ bio-printing group were similar, which were lower in both groups than those in the normal cartilage (Figure S7b). The Young’s modulus for these three aforementioned groups was 10.0216 ± 1.0601, 6.5859 ± 0.8391, and 6.9115 ± 0.8380 GPa, respectively (Figure S7c).
The gross view in blank control group, hydrogel implantation group, and in situ bio-printing group is illustrated in Fig. 5a. It was disclosed that the cartilage defects were repaired in the hydrogel implantation group and in situ bio-printing group. The newly-born region in control group was irregular and pale. Compared with blank control group, the regenerated tissue in these two groups exhibited glossy and smooth appearance, which was found similar to native cartilage.
The results of ICRS scoring system are displayed in Fig. 5b, in which the hydrogel implantation group and in situ bio-printing group represented approximate points in all items. The score in control group was significantly lower than that in other two groups.
The histological results are presented in Fig. 6a–c. The regenerated tissue in control group did not fill the defect region, and an obvious gap could be observed in this area. In hydrogel implantation group and in situ bio-printing group, the defects were fully filled by newly-born tissue. The surface was uniform and smooth. In addition, ECM showed a strong staining with toluidine blue and Safranin O, indicating that the main component was glycosaminoglycan (GAG). The immunohistochemistry unveiled the existence of collagen II in ECM. It was noted that the regenerated tissue in hydrogel implantation group and in situ bio-printing group was hyaline cartilage. As depicted in Figure S8, the arrangement of chondrocytes in hydrogel implantation group and in situ bio-printing group was closer to native cartilage. They arranged in parallel on the surface layer, and the vertical cells were found in the deep layer, while the cell arrangement in control group was disorganized.
The O’Driscoll scoring system confirmed this conclusion as well (Fig. 6d), in which the rate in hydrogel implantation group and in situ bio-printing group was significantly higher than that in control group. Both groups outperformed in cellular morphology, Safranin O staining, structural integrity, cartilage thickness, and cellularity. However, no significant difference was found between these two groups, which was the same as the results of biomechanical test and ICRS scoring system.
Discussion
It is noteworthy that 3D bio-printing technology is highly appropriate, accompanying by a strong potential for the regeneration of complex tissues and organs. However, the realization of this technology still requires further exploration. We, in the present study, used the robotic-assisted 3D bio-printing to repair ICRS grade IV cartilage defects in living animals. To the best of our knowledge, we, for the first time, applied this technology in vivo. By introducing a 6-DOF robot, we could achieve a larger workspace as well as satisfied printing accuracy. The bio-ink could quickly react through the irradiation of UV light in the defect region. More importantly, the therapeutic efficacy was remarkable. The biomechanical and biochemical features in in situ bio-printing group and hydrogel implantation group were similar, indicating that the process of fabricating hydrogel in vitro can be omitted under certain conditions. Our results suggested that the in situ 3D bio-printing technology is highly appropriate to repair injuries, which can directly obtain autogenous MSCs from surrounding microenvironment (e.g., bone and cartilage damage).
Traditional non-surgical therapeutic methods for cartilage injuries include non-steroidal anti-inflammatory drugs (NSAIDs) and opioids [23]. The intra-articular injection and surgery, including mosaic arthroplasty and joint replacement, are highly efficient for patients with severe symptoms that cannot attain adequate pain relief and functional improvement using non-surgical treatments [24]. In the present study, a modified surgical procedure was adopted for mosaic arthroplasty, which was taken as a well-established surgical treatment into account for focal chondral and osteochondral defects [25]. The defect region should be remodeled to the cylinder shape for the purpose of filling graft plugs harvested from non-weight-bearing areas. The geometry of the transplanted plug is highly vital for the successfulness of this surgery [26]. Except for these early complications mentioned above, the late degradation of grafts is another reason for the implantation failure [27].
Thus, in situ 3D bio-printing technology is highly appropriate for such risk factors. Firstly, geometry of surgical area is of great importance for execution of 3D bio-printing technology. No extra supporting structures are required for the bio-printing target due to the flat bottom in the defect. The size and curvature of the defect region can be determined according to imaging examinations, including magnetic resonance imaging (MRI) and CT. In addition, 3D scanning is another option to achieve the geometry of defect region. Based on our previous study, the handheld 3D scanner can obtain the required 3D models in a short period of time. We also previously found that resolution (∼5 μm) was sufficient for reestablishing the superficial zone, and had only about 200 μm of thickness, which was almost impossible to repair manually [13]. These specific conditions ensured the practicability of the in situ 3D bio-printing. In the present study, in vitro and in vivo experiments mimicked the operative environment of mosaic arthroplasty. The surface error was found to be lower than 30 μm, which was remarkably higher than the accuracy of manual operation.
Secondly, the accuracy of bio-printing can be easily ameliorated by the TCP calibration method. Compared with traditional ball-bar instrument, the fast TCP calibration demonstrated a noticeable improvement in measuring space and operation process [28]. The points used for measurement can be arbitrary in space, and the calibration can be completed as soon as the positions of the points are read. The whole procedure can be accomplished by nonprofessionals, including physicians and nurses. In addition, the computation load of the above-mentioned method is small due to the simply procedure and calibration algorithm. A complex calibration principle generally results in sophisticated calibration algorithm, and the accuracy may decrease in each step of the operation process [29]. Meanwhile, that sophisticated method may be extremely time-consuming and has poor calibration efficiency. However, the proposed fast calibration method can be applied to a robotic-assisted 3D bio-printer with multiple printing units. The required TCP for each bio-printing unit can be automatically generated after the key geometric data can be imported into the software. Using the proposed calibration method, the movement accuracy of the 6-DOF robot used in the present study was increased. The whole calibration process could be completed in 10 min as well.
Finally, it should be noted that massive bleeding in the defect region generally leads to dilution of bio-ink, causing difficulty in terms of adhesion and polymerization, as well as hydrogel swelling and bulging due to damaged region. In contrast to microfracture surgery, the amount of haemorrhage in the mosaic arthroplasty is typically insignificant, which is highly proper for the crosslinking of bio-ink. The bio-ink in the current study consisted of HAMA and 4-Armed PEG-ACLT, which could provide both ECM-like components and acceptable biomechanical properties. The HAMA contains GAG, which can dissipate energy during the load bearing process by adjusting for the included water content [30]. The HA can lubricate cells, regulate cell movement on the viscoelastic matrix, stabilize reticular fibers, and also protect the cells from mechanical damages [31]. The 4-Armed PEG-ACLT was served as the covalent cross-linkers, and the photopolymerization rate of HAMA could be improved in the present research. It generally improves the mechanical properties due to the higher crosslinking efficiency and density [16]. A fast crosslinking reaction can be accomplished in the defect region during the 3D bio-printing process. Although the bio-ink was cell/drug-free, it could induce the regeneration of hyaline-like cartilage in the injured area. However, the biomechanical properties of the regenerated cartilage were lower than the native cartilage, and an obvious regenerated area could be observed in the center of the defect region. However, the microstructure and the components were the same as native cartilage, indicating that a remarkable improvement space was existed for the bio-ink. The addition of bioactive factors into bio-ink may ameliorate the mechanical properties and the treatment efficacy as well.
Therefore, the in situ 3D bio-printing technology can improve the surgical procedure for cartilage injury. Due to the limitation of donation region, the size of mosaic arthroplasty is generally limited to less than 4 cm2 [32]. Large size of cartilage defect typically requires more than one surgical region, indicating severe trauma in the non-weight-bearing areas. The application of in situ 3D bio-printing technology can also decrease the damage in knee joint and increase the graft accuracy.
Conclusions
In the present study, we employed a 6-DOF robot to act as a 3D bio-printer to achieve in situ 3D bio-printing. A fast TCP calibration method was developed for improving the robot’s movement and bio-printing accuracy. After performing in vitro experiment, the in situ 3D bio-printing process was conducted on rabbits to repair the ICRS grade IV cartilage defect. The bio-ink could fill the defect region with appropriate geometry and the cartilage injury could be repaired after 12 weeks. This study indicated the feasibility of the mentioned technology for clinical application.
Declaration of Competing Interest
The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Acknowledgments
This study was supported by the National Natural Science Foundation of China (Grant Nos. 51705259, 81420108021, 81730067, 81802135, and 51875101), National Key Research and Development Project (Grant No. 2018YFF0301100), Postgraduate Research & Practice Innovation Program of Jiangsu Province (Grant No. KYCX18_0065), the Key Research and Development Project of Jiangsu Province (Grant No. BE2018010-3), Jiangsu Provincial Key Medical Center Foundation, and Jiangsu Provincial Medical Outstanding Talent Foundation.
Footnotes
Peer review under responsibility of Cairo University.
Supplementary data to this article can be found online at https://doi.org/10.1016/j.jare.2020.01.010.
Contributor Information
Lan Li, Email: lanl17@163.com.
Qing Jiang, Email: qingj@nju.edu.cn.
Xingsong Wang, Email: xswang@seu.edu.cn.
Appendix A. Supplementary data
The following are the Supplementary data to this article:
References
- 1.Murphy S.V., Anthony A. 3D bioprinting of tissues and organs. Nat Biotechnol. 2014;32:773–785. doi: 10.1038/nbt.2958. [DOI] [PubMed] [Google Scholar]
- 2.Campbell P.G., Weiss L.E. Tissue engineering with the aid of inkjet printers. Expert Opin Biol Ther. 2007;7:1123–1127. doi: 10.1517/14712598.7.8.1123. [DOI] [PubMed] [Google Scholar]
- 3.Mironov V., Kasyanov V., Markwald R.R. Organ printing: from bioprinter to organ biofabrication line. Curr Opin Biotechnol. 2011;22:667–673. doi: 10.1016/j.copbio.2011.02.006. [DOI] [PubMed] [Google Scholar]
- 4.Di Bella C., Duchi S., O'Connell C.D., Blanchard R., Augustine C., Yue Z. In situ handheld three-dimensional bioprinting for cartilage regeneration. J Tissue Eng Regener Med. 2018;12:611–621. doi: 10.1002/term.2476. [DOI] [PubMed] [Google Scholar]
- 5.Singh S., Choudhury D., Yu F., Mironov V., Naing M.W. In situ bioprinting - Bioprinting from benchside to bedside? Acta Biomater. 2019 doi: 10.1016/j.actbio.2019.08.045. [DOI] [PubMed] [Google Scholar]
- 6.Keriquel V., Oliveira H., Remy M., Ziane S., Delmond S., Rousseau B. In situ printing of mesenchymal stromal cells, by laser-assisted bioprinting, for in vivo bone regeneration applications. Sci Rep. 2017;7:1778. doi: 10.1038/s41598-017-01914-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Cohen D.L., Lipton J.I., Bonassar L.J., Lipson H. Additive manufacturing for in situ repair of osteochondral defects. Biofabrication. 2010;2 doi: 10.1088/1758-5082/2/3/035004. [DOI] [PubMed] [Google Scholar]
- 8.Lipskas J., Deep K., Yao W. Robotic-Assisted 3D Bio-printing for Repairing Bone and Cartilage Defects through a Minimally Invasive Approach. Sci Rep. 2019;9:3746. doi: 10.1038/s41598-019-38972-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Skardal A., Mack D., Kapetanovic E., Atala A., Jackson J.D., Yoo J. Bioprinted amniotic fluid-derived stem cells accelerate healing of large skin wounds. Stem Cells Transl Med. 2012;1:792–802. doi: 10.5966/sctm.2012-0088. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Albanna M., Binder K.W., Murphy S.V., Kim J., Qasem S.A., Zhao W. In Situ Bioprinting of Autologous Skin Cells Accelerates Wound Healing of Extensive Excisional Full-Thickness Wounds. Sci Rep. 2019;9:1856. doi: 10.1038/s41598-018-38366-w. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.O'Connell C.D., Di Bella C., Thompson F., Augustine C., Beirne S., Cornock R. Development of the Biopen: a handheld device for surgical printing of adipose stem cells at a chondral wound site. Biofabrication. 2016;8 doi: 10.1088/1758-5090/8/1/015019. [DOI] [PubMed] [Google Scholar]
- 12.Duchi S., Onofrillo C., O'Connell C.D., Blanchard R., Augustine C., Quigley A.F. Handheld Co-Axial Bioprinting: Application to in situ surgical cartilage repair. Sci Rep. 2017;7:5837. doi: 10.1038/s41598-017-05699-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Li L., Yu F., Shi J., Shen S., Teng H., Yang J. In situ repair of bone and cartilage defects using 3D scanning and 3D printing. Sci Rep. 2017;7:9416. doi: 10.1038/s41598-017-10060-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Mouser V.H., Abbadessa A., Levato R., Hennink W.E., Vermonden T., Gawlitta D. Development of a thermosensitive HAMA-containing bio-ink for the fabrication of composite cartilage repair constructs. Biofabrication. 2017;9 doi: 10.1088/1758-5090/aa6265. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Abbadessa A., Mouser V.H.M., Blokzijl M.M., Gawlitta D., Dhert W.J.A., Hennink W.E. A Synthetic Thermosensitive Hydrogel for Cartilage Bioprinting and Its Biofunctionalization with Polysaccharides. Biomacromolecules. 2016;17:2137–2147. doi: 10.1021/acs.biomac.6b00366. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Sun W., Xue B., Ying L., Meng Q., Wei W. Polymer-Supramolecular Polymer Double-Network Hydrogel. Adv Funct Mater. 2016 [Google Scholar]
- 17.Wang J., Zhang F., Tsang W.P., Wan C., Wu C. Fabrication of injectable high strength hydrogel based on 4-arm star PEG for cartilage tissue engineering. Biomaterials. 2017;120:11–21. doi: 10.1016/j.biomaterials.2016.12.015. [DOI] [PubMed] [Google Scholar]
- 18.Loebel C., Rodell C.B., Chen M.H., Burdick J.A. Shear-thinning and self-healing hydrogels as injectable therapeutics and for 3D-printing. Nat Protoc. 2017;12:1521–1541. doi: 10.1038/nprot.2017.053. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Jin J., Gans N. Parameter identification for industrial robots with a fast and robust trajectory design approach. Robotics Comput Integr Manuf. 2015;31:21–29. [Google Scholar]
- 20.van den Borne M.P., Raijmakers N.J., Vanlauwe J., Victor J., de Jong S.N., Bellemans J. International Cartilage Repair Society (ICRS) and Oswestry macroscopic cartilage evaluation scores validated for use in Autologous Chondrocyte Implantation (ACI) and microfracture. Osteoarthr Cartil. 2007;15:1397–1402. doi: 10.1016/j.joca.2007.05.005. [DOI] [PubMed] [Google Scholar]
- 21.Zhang K., Zhang Y., Yan S., Gong L., Wang J., Chen X. Repair of an articular cartilage defect using adipose-derived stem cells loaded on a polyelectrolyte complex scaffold based on poly(l-glutamic acid) and chitosan. Acta Biomater. 2013;9:7276–7288. doi: 10.1016/j.actbio.2013.03.025. [DOI] [PubMed] [Google Scholar]
- 22.O'Driscoll S.W., Keeley F.W., Salter R.B. Durability of regenerated articular cartilage produced by free autogenous periosteal grafts in major full-thickness defects in joint surfaces under the influence of continuous passive motion. A follow-up report at one year. J Bone Joint Surg Am. 1988;70:595–606. [PubMed] [Google Scholar]
- 23.McAlindon T.E., Bannuru R.R., Sullivan M.C., Arden N.K., Berenbaum F., Bierma-Zeinstra S.M. OARSI guidelines for the non-surgical management of knee osteoarthritis. Osteoarthritis Cartil. 2014;22:363–388. doi: 10.1016/j.joca.2014.01.003. [DOI] [PubMed] [Google Scholar]
- 24.Zhang W., Moskowitz R.W., Nuki G., Abramson S., Altman R.D., Arden N. OARSI recommendations for the management of hip and knee osteoarthritis, Part II: OARSI evidence-based, expert consensus guidelines. Osteoarthritis Cartil. 2008;16:137–162. doi: 10.1016/j.joca.2007.12.013. [DOI] [PubMed] [Google Scholar]
- 25.Kunz M., Waldman S.D., Rudan J.F., Bardana D.D., Stewart A.J. Computer-assisted mosaic arthroplasty using patient-specific instrument guides. Knee Surg Sports Traumatol Arthrosc Official J ESSKA. 2012;20:857–861. doi: 10.1007/s00167-011-1638-2. [DOI] [PubMed] [Google Scholar]
- 26.Hangody L., Vasarhelyi G., Hangody L.R., Sukosd Z., Tibay G., Bartha L. Autologous osteochondral grafting–technique and long-term results. Injury. 2008;39(Suppl 1):S32–S39. doi: 10.1016/j.injury.2008.01.041. [DOI] [PubMed] [Google Scholar]
- 27.Koh J.L., Wirsing K., Lautenschlager E., Zhang L.O. The effect of graft height mismatch on contact pressure following osteochondral grafting: a biomechanical study. Am J Sports Med. 2004;32:317–320. doi: 10.1177/0363546503261730. [DOI] [PubMed] [Google Scholar]
- 28.Theissen N.A., Laspas T., Archenti A. Closed-force-loop elastostatic calibration of serial articulated robots. Rob Comput Integr Manuf. 2019;57:86–91. [Google Scholar]
- 29.Lin H., Sun L. Searching globally optimal parameter sequence for defeating Runge phenomenon by immunity genetic algorithm. Appl Math Comput. 2015;264:85–98. [Google Scholar]
- 30.Ge Z., Li C., Heng B.C., Cao G., Yang Z. Functional biomaterials for cartilage regeneration. J Biomed Mater Res Part A. 2012;100:2526–2536. doi: 10.1002/jbm.a.34147. [DOI] [PubMed] [Google Scholar]