Studying the Optical 3D Accuracy of Intraoral Scans: An In Vitro Study

Pokpong Amornvit 1, Sasiwimol Sanohkan 1,, Chaimongkon Peampring 1
PMCID: PMC7042536  PMID: 32148745

Abstract

There are various scanners available in dental practice with various accuracies. The aim of this study was to compare the 3D capturing accuracy of scans obtained from Trios 3 and Dental Wings scanner. A reference mandibular model was printed from FormLab with reference points in three axes (X, Y, and XY and Z). The printed model was scanned 5 times with 3 scans: normal scan by Trios 3 (Trios 3A), high-resolution scan by Trios 3 (Trios 3B), and normal scan by Dental Wings. After scan, the stereolithography (stl) files were generated. Then, the measurements were made from the computer software using Rhinoceros 3D (Rhino, Robert McNeel & Associates for Windows, Washington DC, USA). The measurements made with digital caliper were taken as control. Statistical analysis was done using one-way ANOVA with post hoc using Sheffe (P < 0.01). Trios 3 presented higher accuracy than Dental Wings and high resolution showed better results. The Dental Wings showed less accuracy at the measurements >50 mm of length and >30 mm in width. There was no significant difference (P > 0.05) of control with the Trios 3A and Trios 3B. Similarly, for the measurements in Z-axis, there was no significant difference of control with each scan (Trios 3A, Trios 3B, and Dental Wings). Accuracy of the scan is affected by the length of the scanning area and scanning pattern. It is less recommended to Dental Wings scan >3-unit prosthesis and that crosses the midline.

1. Introduction

There has been massive advancement in digital dentistry in the recent decade, especially since the invention of computer-aided design/ computer-aided manufacturing (CAD/CAM) system, milling systems, rapid and automated prototyping, and three-dimensional (3D) printing of dental biomaterials, and these have revolutionized and created a new modality in dentistry [1]. Currently, CAD/CAM is widely used in virtual occlusal records, full-mouth reconstruction, and orthodontics [24]. Moreover, they are extensively used in both the dental laboratory and the dental clinic for the fabrication of various prosthesis, such as inlays, onlays, veneers, crowns, fixed partial dentures, orthodontic aligners, surgical guides, and implant abutments [3]. A 3D scanning is a process that is used to capture the shape of an object using a 3D scanner. After scanning, a 3D file of an object is created which can be edited, and 3D printed. An intraoral scan (IOS) can be based on many different technologies, each with its own limitations, advantages, and costs [5]. Many limitations in the kind of objects that can be digitized are still present. For example, optical technology may encounter many difficulties with shiny, reflective, or transparent objects.

Teeth, especially the anterior, play in the esthetics of face [6]. For a successful esthetic dental restoration, a good dental impression is important [7, 8]. With the use of digital dentistry, the intraoral conditions can be transferred digitally and printed. When dental laboratories receive a digital impression, they create a model from the data and either continue with the traditional fabrication procedure or rescan the model and fabricate the prosthesis. The dental technician can do all the design restorative works directly on the computer based on the digital file received. Hence, the digital impression plays an important in the fabrication of all digital works and the accuracy of the impression is very important.

There are various 3D scanners technologies, such as image capturing or video capturing type contact scanning or noncontact scanning [912]. Contact canners probe the subject through physical touch while the object is in contact with or resting on a precision flat surface plate, ground and polished to a specific maximum of surface roughness. Noncontact scanners emit some kind of radiation or light and detect its reflection or radiation passing through the object in order to probe an object or environment [11]. Nowadays, the noncontact scanning technique is recommended widely. Types of scanning technology: the 3D scanning technologies rely on different physical principles and are explained in following categories [5]:

  1. Laser triangulation 3D scanning technology uses either a laser line or a single laser point to scan across an object.

  2. Structured light 3D scanning technology uses trigonometric triangulation but not the laser.

  3. Photogrammetry 3D scan scanning technology (photography) reconstructs 3D from 2D captures with computer vision and computational geometry algorithms.

  4. Contact-based 3D scanning technology is based on contact form of 3D data collection and uses a contact probe.

Accuracy comprises precision and trueness (ISO 5725-1) [13]. Precision describes how close repeated measurements are to each other [14]. The higher the precision, the more predictable the measurement is. Trueness describes how far the measurement deviates from the actual dimensions of the measured object. A high trueness delivers a result that is close or equal to the actual dimensions of the measured object. Many factors influence the accuracy of the IOS such as [1517]:

  1. Scanner: ability to record details and its accuracy

  2. Operator: scanning principles and span of scanning

  3. Scanning area: size of scanning area, arch length, and surface irregularities

  4. Intraoral environmental factors: temperature, relative humidity, and illumination

The IOS accuracy is enhanced by reducing the span of scanning, and ensuring the scanned surfaces exhibit minimal irregularities [10]. The problem with IOS is that it can be difficult to detect deep margin lines in prepared teeth and/or in case of deep margins or bleeding [18]. In addition, various studies done in evaluating the digital impression highlights several issues such as distortion of the digital models, problems with the intraoral conditions, and lower precision compared to conventional impressions [19, 20]. In addition, digital scanners with high accuracy are currently limited to small measurement fields such as single teeth or quadrants [19, 21]. The aim of this study was to compare the 3D capturing accuracy of scans obtained from Trios 3 and Dental Wings scanner in an in vitro study design.

2. Materials and Methods

A method modified from the American National Standard/ American Dental Association (ANS)/ADA) Standard No. 132 for the scanning accuracy was used in this study [22]. The study consists of fabrication of dental model, scanning, and measurements. The details of the study are shown in (Figure 1).

Figure 1.

Figure 1

Details of the study overview.

2.1. Fabrication of Dental Model

A digital mandibular model is made in the computer. Various points were marked on the digital model where the measurements can be measured in three axes (X, Y, and Z) (Figure 2).

Figure 2.

Figure 2

Digital dental model with various points marked in the three axes (X, Y, and Z).

The model was printed using FormLab following manufacturing recommendations. From 3 dental models, the best model was selected for this study as shown in Figure 3.

Figure 3.

Figure 3

Printed mandibular dental model.

2.2. Scanning

The printed model was scanned 5 times each with 3 Shape Trios 3A: normal scan, 3 Shape Trios 3B: high resolution (3 Shape Trios A/S 2018, Copenhagen, Denmark), and Dental Wings (Dental Wings Inc., Montreal QC, Canada) (Figure 4) according to the manufacturer's recommendation. After the scan, the scanned files were saved as stereolithography (STL) files.

Figure 4.

Figure 4

Intraoral scanners (IOS): (a) Trios 3 and (b) Dental Wings.

2.3. Measurements

Then, for the scanned files, the measurements were made from computer software using the Rhinoceros 3D modeling software (Rhino, Robert McNeel & Associates for Windows, Washington, DC, USA). The measurements were done in 3 axes (X, Y, and XY and Z) of various lengths as follows (Table 1 and Figures 5 and 6). The measurements made on the printed model with digital caliper were taken as the control (Figure 5).

Table 1.

Measurements in three axes (X, Y, XY, and Z) of various lengths.

X-axis Y-axis XY-axis Z-axis
X1: Mesiodistal width on teeth #21 (2 mm) Y1: 2 mm buccolingual width on teeth #27 (2 mm) AR: Diagonal distance from #12 to #27 (65 mm) Z1: Buccal notch on #17 (2 mm)
X2: Distance from #11 to #21 (10 mm) Y2: Buccolingual width on teeth #26 (10 mm) Z2: Buccal notch on #27 (4 mm)
X3: Distance from #12 to #22 (30 mm) Z3: Buccal notch on #16 (6 mm)
X4: Distance from #13 to #31 (40 mm) Y3: Buccolingual width from #25 to #27 (20 mm) AR: Diagonal distance from #22 to #17 (65 mm) Z4: Buccal notch on #27 (8 mm)
X5: Distance from #14 to #41 (50 mm)
X6: Distance from #16 to #61 (60 mm) Y4: Buccolingual width from #23 to #27 (30 mm)

Figure 5.

Figure 5

Reference points on the model and various measurements measured on model in different axes: X-axis (a), Y-axis (b), XY-axis (c), and Z-axis (d).

Figure 6.

Figure 6

Reference points on the model and various measurements in the X-axis (length) and Y-axis (length) from software of 1 scan.

In addition, the quality of the scans and capturing details of the both scanners were also evaluated.

2.4. Statistical Analysis

Microsoft Excel 2010 and SPSS version 20 (IBM Company, Chicago, USA) were used for the descriptive statistics and expressed as mean and standard deviation. Multiple comparison was done using one-way ANOVA with post hoc using Sheffe to see the significant difference (P < 0.01) between the control (dental model) and scans.

3. Results

Tables 24 shows the descriptive statistics of the scans of various lengths in the three axes (X, Y, XY, and Z).

Table 2.

Descriptive statistics of measurements of various groups in the X-axis.

Measurements Groups Mean SD 95% CI for mean Min Max
Lower bound Upper bound
X1 Control 1.938 0.039 1.888 1.987 1.90 2.00
Trios 3A 1.876 0.032 1.836 1.916 1.85 1.93
Trios 3B 1.886 0.036 1.841 1.931 1.84 1.94
Dental Wings 1.876 0.057 1.805 1.946 1.83 1.96
Control 9.938 0.022 9.911 9.964 9.90 9.95

X2 Trios 3A 9.864 0.027 9.835 9.892 9.84 9.90
Trios 3B 9.968 0.028 9.933 10.002 9.92 9.99
Dental Wings 9.866 0.071 9.778 9.954 9.79 9.97
Control 29.852 0.06 29.77 29.926 29.78 29.93

X3 Trios 3A 29.734 0.022 29.706 29.761 29.70 29.76
Trios 3B 29.914 0.033 29.872 29.955 29.87 29.95
Dental Wings 29.806 0.172 29.592 30.019 29.60 30.03
Control 39.910 0.054 39.842 39.978 39.82 39.96

X4 Trios 3A 39.716 0.011 39.702 39.73 39.70 39.73
Trios 3B 39.698 0.248 39.389 40.007 39.39 39.93
Dental Wings 40.112 0.177 39.892 40.332 39.88 40.28
Control 50.03 0.035 49.986 50.074 49.98 50.07

X5 Trios 3A 49.86 0.118 49.713 50.006 49.75 50.02
Trios 3B 50.21 0.136 50.04 50.379 49.98 50.31
Dental Wings 50.832 0.423 50.306 51.357 50.43 51.55
Caliper 60.616 0.052 60.552 60.68 60.54 60.67
Control 60.068 0.271 59.732 60.404 59.66 60.42

X6 Trios 3A 60.368 0.514 59.728 61.007 59.77 60.83
Dental Wings 61.952 0.374 61.487 62.417 61.55 62.41

SD = standard deviation; CI = confidence interval for mean; min = minimum; max = maximum.

Table 3.

Descriptive statistics of measurements of various groups in Y-axis and XY-axis.

Measurements Groups Mean SD 95% CI for mean Min Max
Lower bound Upper bound
Y1 Control 1.932 0.008 1.921 1.942 1.92 1.94
Trios 3A 1.856 0.035 1.812 1.899 1.80 1.89
Trios 3B 1.854 0.077 1.758 1.949 1.78 1.97
Dental Wings 1.828 0.017 1.805 1.85 1.80 1.85

Y2 Control 9.876 0.021 9.85 9.901 9.86 9.91
Trios 3A 9.806 0.149 9.619 9.992 9.54 9.89
Trios 3B 9.81 0.082 9.708 9.912 9.74 9.93
Dental Wings 9.8 0.074 9.707 9.892 9.71 9.90

Y3 Control 19.934 0.047 19.875 19.992 19.89 20.01
Trios 3A 19.71 0.12 19.561 19.859 19.56 19.84
Trios 3B 19.81 0.062 19.733 19.886 19.75 19.90
Dental Wings 19.636 0.518 18.993 20.278 18.81 20.02

Y4 Control 30.162 0.339 29.748 30.575 29.86 30.70
Trios 3A 29.974 0.391 29.488 30.459 29.38 30.35
Trios 3B 30.186 0.593 29.45 30.921 29.60 31.05
Dental Wings 29.754 0.425 29.225 30.282 29.17 30.37

AR Control 65.126 0.037 65.079 65.173 65.10 65.19
Trios 3A 64.872 0.119 64.724 65.019 64.78 65.07
Trios 3B 65.124 0.173 64.908 65.339 64.90 65.32
Dental Wings 66.044 0.624 65.268 66.819 65.17 66.88

AL Control 65.094 0.053 65.027 65.16 65.03 65.15
Trios 3A 65.43 0.484 64.828 66.031 64.88 65.81
Trios 3B 65.408 0.326 65.002 65.813 64.97 65.89
Dental Wings 65.942 0.133 65.776 66.107 65.84 66.14

SD = standard deviation; CI = confidence interval for mean; min = minimum; max = maximum.

Table 4.

Descriptive statistics of measurements of various groups in the Z-axis.

Measurements Groups Mean SD 95% CI for mean Min Max
Lower bound Upper bound
Z1 Control 2.014 0.023 1.985 2.042 1.99 2.04
Trios 3A 2.018 0.037 1.971 2.064 1.96 2.06
Trios 3B 1.998 0.047 1.938 2.057 1.96 2.08
Dental Wings 1.990 0.063 1.911 2.068 1.88 2.04

Z2 Control 3.986 0.011 3.972 4.001 3.97 4.00
Trios 3A 3.988 0.072 3.898 4.077 3.88 4.05
Trios 3B 3.952 0.073 3.861 4.042 3.90 4.08
Dental Wings 3.97 0.054 3.902 4.037 3.89 4.02

Z3 Control 6.15 0.137 5.979 6.32 5.97 6.32
Trios 3A 5.906 0.209 5.645 6.166 5.71 6.19
Trios 3B 5.978 0.011 5.964 5.991 5.96 5.99
Dental Wings 6.142 0.155 5.949 6.335 5.92 6.31

Z4 Control 8.046 0.04 7.995 8.096 8.01 8.11
Trios 3A 7.968 0.085 7.861 8.074 7.86 8.04
Trios 3B 7.978 0.136 7.808 8.147 7.84 8.14
Dental Wings 7.996 0.158 7.799 8.192 7.72 8.11

SD = standard deviation; CI = confidence interval for mean; min = minimum; max = maximum.

The multiple comparisons between the measurements of dental model and the scan are shown in Tables 57. It was seen that there was significant difference (P value <0.01) of the measurements X1, X2, Y1, Y2, Y3, AR, AL, Z1, Z3, and Z4 between the dental model and the scans.

Table 5.

Multiple comparison of the various measurements in the X-axis of control with other scan groups (Trios 3A, Trios 3B, and Dental Wings).

Measurements Comparison groups Mean difference P value
X1 Control Trios 3A 0.062 0.190
Trios HD 0.052 0.322
Dental Wings 0.062 0.190

X2 Control Trios 3A 0.074 0.082
Trios HD −0.03 0.727
Dental Wings 0.072 0.093

X3 Control Trios 3A 0.118 0.299
Trios HD −0.062 0.777
Dental Wings 0.046 0.893

X4 Control Trios 3A 0.194 0.309
Trios HD 0.212 0.240
Dental Wings −0.202 0.277

X5 Control Trios 3A 0.17 0.719
Trios HD −0.18 0.683
Dental Wings −0.802 0.001

X6 Control Trios 3A 0.548 0.143
Trios HD 0.248 0.737
Dental Wings −1.336 <0.001

Significant difference at P value <0.05.

Table 6.

Multiple comparison of the various measurements in the Y-axis and the XY-axis of control with other scan groups (Trios 3A, Trios 3B, and Dental Wings).

Measurements Comparison groups Mean difference Sig.
Y1 Control Trios 3A 0.076 0.092
Trios 3B 0.078 0.081
Dental Wings 0.104 0.015

Y2 Control Trios 3A 0.07 0.711
Trios 3B 0.066 0.746
Dental Wings 0.076 0.657

Y3 Control Trios 3A 0.224 0.637
Trios 3B 0.124 0.910
Dental Wings 0.298 0.407

Y4 Control Trios 3A 0.188 0.929
Trios 3B −0.024 1.000
Dental Wings 0.408 0.567

AR Control Trios 3A 0.254 0.692
Trios 3B 0.002 1.000
Dental Wings −0.918 0.005

AL Control Trios 3A −0.336 0.402
Trios 3B −0.314 0.459
Dental Wings −0.848 0.004

Significant difference at P value <0.05.

Table 7.

Multiple comparisons of the various measurements in the Z-axis of control with other scan groups (Trios 3A, Trios 3B, and Dental Wings).

Measurements Comparison groups Mean difference Sig.
Z1 Control Trios 3A −0.004 0.999
Trios 3B 0.016 0.957
Dental Wings 0.024 0.872

Z2 Control Trios 3A −0.002 1.000
Trios 3B 0.034 0.836
Dental Wings 0.016 0.979

Z3 Control Trios 3A 0.244 0.119
Trios 3B 0.172 0.366
Dental Wings 0.008 1.000

Z4 Control Trios 3A 0.078 0.766
Trios 3B 0.068 0.831
Dental Wings 0.050 0.923

Significant difference at P value <0.05.

For the measurements in the X-axis, X1–X4, there was no significant difference of each scan (Trios 3A, Trios 3B, and Dental Wings) compared to the control as shown in Table 5. But, X5–X6, Dental Wings showed there was significant difference (P value <0.01) from the dental model (control). Hence, Dental Wings showed less accuracy at the measurement length 50 mm and 60 mm.

Similarly, for the measurements in the Y-axis, Y2–Y4, there was no significant difference of each scan (Trios 3A, Trios 3B, and Dental Wings) compared to the control as shown in Table 6. But, for Y1, Dental Wings showed there was significant difference (P value = 0.015) from the dental model (control). Furthermore, for the measurements in the XY-axis, AR and AL, there was significant difference (P value <0.005) of Dental Wings compared to the control. But, there was no significant difference (P value >0.05) of control with the Trios 3A and Trios 3B for the AR and AL. Hence, Dental Wings showed less accuracy in the measurements.

Similarly, for the measurements in the Z-axis, Z1–Z4, there was no significant difference (P=0.05) of control with each scan (Trios 3A, Trios 3B, and Dental Wings) as shown in Table 7.

Regarding the quality and capturing details, Trios 3A showed the best results followed by Trios 3B and Dental Wings.

4. Discussion

Digital impressions reduce the patient discomfort; intraoral scanners (IOS) are time-efficient and simplify clinical procedures for the dentist and the laboratory technician, eliminating plaster models and allowing better communication with the dental technician and with patients. The accuracy of which influences the fit of the restorations, an important factor in the longevity of the final restoration [7, 8]. Renne et al. [23] compared 7 different IOS and they found that the Planscan had the best accuracy (trueness and precision) while the 3Shape Trios was found to be the poorest for sextant scanning. The order of trueness for complete arch scanning was as follows: 3Shape D800 > iTero > 3Shape TRIOS 3 > Carestream 3500 >Planscan > CEREC Omnicam > CEREC Bluecam. The order of precision for complete-arch scanning was as follows: CS3500 > iTero > 3Shape D800 > 3Shape TRIOS 3 >CEREC Omnicam > Planscan > CEREC Bluecam. For the secondary outcome evaluating the effect time has on trueness and precision, the complete-arch scan time was highly correlated with both trueness and precision. They concluded that for complete-arch scanning, the 3 Shape Trios was found to have the best balance of speed and accuracy. Park et al. [15] designed an intraoral environment simulator to assess the accuracy of 2 IOS using the simulator and found no difference due to the intraoral environment. The simulator contributes to the higher accuracy of IOS.

Mutwalli et al. [24] studied the trueness and precision of different IOS when scanning fully edentulous arch with multiple implants. They found that there were significant differences between all IOS. For the implant measurements, Trios 3 had the lowest trueness, followed by Trios 3 mono and Itero element. Trios had the lowest precision, followed by Itero element and Trios 3 mono. Regarding the interarch distance measurements, Trios 3 had the lowest trueness, followed by Trios 3 mono and Itero element. Trios 3 had the lowest precision, followed by Itero element and Trios 3 mono. But, in our study, Trios 3 presented higher accuracy than Dental Wings and high resolution showed better results. There can be minor errors in the measurements by the IOS in various steps. While scanning 1 arch, generally, the IOS captures around 1200 images. Errors in scanning may be due to overlapping of the partial images, especially in the anterior region [19, 25]. The occurrence of more errors of digital impression in the anterior regions is due to the less structured tooth surface and steep inclines. The superimposition process leads to the deviation. These errors might be reduced or avoided with further software improvements. In addition, there can be errors in computer processing, which may be due to filter algorithms and calibration errors of the scanner [25]. There can be errors in the x-axis, y-axis, and z-axis, but in our study, there were more errors in the z-axis (depth of scanning). The errors can be avoided by a longitudinal measurement of a calibrated length specimen.

The dimension measured can be implemented as follows. In the anterior region, X1 (2 mm) represents a scan body or an onlay, X2 (10 mm) represents 1-unit restoration or prosthesis, X3 (30 mm) represents the 4-unit restoration, X4 (40 mm) represents the 6-unit restoration, X5 (50 mm) represents the 10-unit restoration, and X6 (60 mm) represents the 14-unit restoration or full-arch restoration. X1–X4 represent the dimensions in one quadrant in the same arch (upper or lower). X5 and X6 extend to 2 quadrants in the same arch. Trios 3 allows us to record full-arch and high-resolution scans that are more accurate than normal scans, but there was no significant difference. In addition, Dental Wings allow for maximum 6-unit restorations.

In the posterior region, Y1 (2 mm) represents a scan body or an onlay, Y2 (10 mm) represents 1-unit restoration or prosthesis, Y3 (20 mm) represents the 2-unit restoration, and Y4 (30 mm) represents the 4-unit restoration. Similarly, AR and AL (65 mm) represent the full-arch restorations. For posterior measurements, all scanners showed acceptable accuracy. But more than 3-unit restoration showed less accuracy (P value <0.05). If we want to fabricate prosthesis with model less, >3-unit restoration is not recommended. For Dental Wings, the accuracy is less for the restoration that crosses the midline; hence, there is a need to be careful. It is less recommended to scan >3-unit prosthesis.

5. Conclusion

Within the limitations of this study, Trios 3 presented higher accuracy, better quality, and captured more details than Dental Wings, and high resolution showed better results. Accuracy of the scanners is affected by the length of the scanning area and scanning pattern. It is less recommended to use Trios 3 for scanning >3-unit prosthesis (50 mm) and that crosses the midline.

Acknowledgments

This research was supported by the Maxillofacial Prosthodontics and Dental Materials Research Unit, Faculty of Dentistry, Prince of Songkla University, and the Scholarship Award of Thai Ph.D. Student under Thailand's Education Hub for Southern Region of ASEAN Countries, Prince of Songkla University.

Data Availability

The data used to support the findings of this study are available from the corresponding author upon reasonable request.

Conflicts of Interest

The authors declare no conflicts of interest.

References

  • 1.Beuer F., Schweiger J., Edelhoff D. Digital dentistry: an overview of recent developments for CAD/CAM generated restorations. British Dental Journal. 2008;204(9):505–511. doi: 10.1038/sj.bdj.2008.350. [DOI] [PubMed] [Google Scholar]
  • 2.Solaberrieta E., Garmendia A., Brizuela A., Otegi J. R., Pradies G., Szentpétery A. Intraoral digital impressions for virtual occlusal records: section quantity and dimensions. BioMed Research International. 2016;2016:7. doi: 10.1155/2016/7173824.7173824 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Davidowitz G., Kotick P. G. The use of CAD/CAM in dentistry. Dental Clinics of North America. 2011;55(3):559–570. doi: 10.1016/j.cden.2011.02.011. [DOI] [PubMed] [Google Scholar]
  • 4.Richert R., Goujat A., Venet L., et al. Intraoral scanner technologies: a review to make a successful impression. Journal of Healthcare Engineering. 2017;2017:9. doi: 10.1155/2017/8427595.8427595 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Trebuòa P. 3D Scanning Technologies and the 3D Scanning Process. France: Aniwaa Pte. Ltd.; 2019. [Google Scholar]
  • 6.Rokaya D., Kitisubkanchana J., Wonglamsam A., Santiwong P., Srithavaj T., Humagain M. Nepalese esthetic dental (NED) proportion in Nepalese population. Kathmandu University Medical Journal. 2017;13(3):244–249. doi: 10.3126/kumj.v13i3.16816. [DOI] [PubMed] [Google Scholar]
  • 7.Perakis N., Belser U., Magne P. Final impressions: a review of material properties and description of a current technique. Journal of Periodontics and Restorative Dentistry. 2004;24(2):109–117. [PubMed] [Google Scholar]
  • 8.Persson A. S. K., Odén A., Andersson M., Sandborgh-Englund G. Digitization of simulated clinical dental impressions: virtual three-dimensional analysis of exactness. Dental Materials. 2009;25(7):929–936. doi: 10.1016/j.dental.2009.01.100. [DOI] [PubMed] [Google Scholar]
  • 9.Meguru Y., Yu K., Atsufumi M. Comparison of digital intraoral scanners by single-image capture system and full-color movie system. Bio-Medical Materials and Engineering. 2017;28(3):305–314. doi: 10.3233/bme-171676. [DOI] [PubMed] [Google Scholar]
  • 10.Abduo J., Elseyoufi M. Accuracy of intraoral scanners: a systematic review of influencing factors. European Journal of Prosthodontics and Restorative Dentistry. 2018;26(3):101–121. doi: 10.1922/EJPRD_01752Abduo21. [DOI] [PubMed] [Google Scholar]
  • 11.Miyazaki T., Hotta Y., Kunii J., Kuriyama S., Tamaki Y. A review of dental CAD/CAM: current status and future perspectives from 20 years of experience. Dental Materials Journal. 2009;28(1):44–56. doi: 10.4012/dmj.28.44. [DOI] [PubMed] [Google Scholar]
  • 12.Amornvit P., Sanohkan S. The accuracy of digital face scans obtained from 3D scanners: an in-vitro study. International Journal of Environmental Research and Public Health. 2019;16(24):p. 5061. doi: 10.3390/ijerph16245061. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.DIN Deutsches Institut für Normung. Accuracy (Trueness and Precision) of Measurement Methods and Results—Part 1: General Principles and Definitions. Berlin, Germany: International Organization for Standardization; 1997. [Google Scholar]
  • 14.Ziegler M. Digital impression taking with reproducibly high precision. International Society of Computerized Dentistry. 2009;12:159–163. [PubMed] [Google Scholar]
  • 15.Park H.-N., Lim Y.-J., Yi W.-J., Han J.-S., Lee S.-P. A comparison of the accuracy of intraoral scanners using an intraoral environment simulator. The Journal of Advanced Prosthodontics. 2018;10(1):58–64. doi: 10.4047/jap.2018.10.1.58. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Matta R. E., Adler W., Wichmann M., Heckmann S. M. Accuracy of impression scanning compared with stone casts of implant impressions. The Journal of Prosthetic Dentistry. 2017;117(4):507–512. doi: 10.1016/j.prosdent.2016.07.026. [DOI] [PubMed] [Google Scholar]
  • 17.Nedelcu R. G., Persson A. S. K. Scanning accuracy and precision in 4 intraoral scanners: an in vitro comparison based on 3-dimensional analysis. The Journal of Prosthetic Dentistry. 2014;112(6):1461–1471. doi: 10.1016/j.prosdent.2014.05.027. [DOI] [PubMed] [Google Scholar]
  • 18.Mangano F., Gandolfi A., Luongo G., Logozzo S. Intraoral scanners in dentistry: a review of the current literature. BMC Oral Health. 2017;17(1):p. 149. doi: 10.1186/s12903-017-0442-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Mehl A., Ender A., Mörmann W., Attin T. Accuracy testing of a new intraoral 3D camera. International Journal of Computerized Dentistry. 2009;12(1):11–28. [PubMed] [Google Scholar]
  • 20.Luthardt R. G., Loos R., Quaas S. Accuracy of intraoral data acquisition in comparison to the conventional impression. International Journal of Computerized Dentistry. 2005;8(4):283–294. [PubMed] [Google Scholar]
  • 21.Luthardt R., Kuhmstedt P., Walter M. A new method for the computer-aided evaluation of three-dimensional changes in gypsum materials. Dental Materials. 2003;19(1):19–24. doi: 10.1016/s0109-5641(02)00013-1. [DOI] [PubMed] [Google Scholar]
  • 22.ANSI/ADA. Standard No. 132: Scanning Accuracy of Dental Chairside and Laboratory CAD/CAM Systems. Chicago, IL, USA: American Dental Association; 2015. [Google Scholar]
  • 23.Renne W., Ludlow M., Fryml J., et al. Evaluation of the accuracy of 7 digital scanners: an in vitro analysis based on 3-dimensional comparisons. The Journal of Prosthetic Dentistry. 2017;118(1):36–42. doi: 10.1016/j.prosdent.2016.09.024. [DOI] [PubMed] [Google Scholar]
  • 24.Mutwalli H., Braian M., Mahmood D., Larsson C. Trueness and precision of three-dimensional digitizing intraoral devices. International Journal of Dentistry. 2018;2018:10. doi: 10.1155/2018/5189761.5189761 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Ender A., Mehl A. Accuracy of complete-arch dental impressions: a new method of measuring trueness and precision. The Journal of Prosthetic Dentistry. 2013;109(2):121–128. doi: 10.1016/s0022-3913(13)60028-1. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The data used to support the findings of this study are available from the corresponding author upon reasonable request.