Skip to main content
The Journal of the Acoustical Society of America logoLink to The Journal of the Acoustical Society of America
. 2020 Mar 2;147(3):1323–1332. doi: 10.1121/10.0000809

Effects of motion on correlations of pulse-echo ultrasound signals: Applications in delay estimation and aperture coherence

Dongwoon Hyun 1,a),, Jeremy J Dahl 1,b)
PMCID: PMC7051867  PMID: 32237854

Abstract

The correlation between two pulse-echo ultrasound signals is used to achieve a wide range of ultrasound techniques, such as Doppler imaging and elastography. Prior theoretical descriptions of pulse-echo correlations were restricted to stationary scatterers. Here, a theory for the correlation of moving scatterers is presented. An expression is derived for the correlation of two pulse-echo signals with arbitrary transmit and receive apertures acquired from a medium undergoing bulk motion using the Fresnel approximation. The derivation is shown to coincide with prior derivations in the absence of scatterer motion. The theory was compared against simulations in applications of phase-shift estimation and aperture coherence measurements. The phase-shift estimate and jitter were accurately predicted under axial and transverse motion for focused transmit apertures and for sequential and interleaved synthetic transmit apertures. The theory also accurately predicted how motion affects the correlation coefficient between receive aperture elements for a synthetic transmit aperture. The presented theory provides a framework for analyzing the correlations of arbitrary pulse-echo configurations for applications in which scatterer motion is expected.

I. INTRODUCTION

In pulse-echo ultrasound, an aperture is used to transmit an ultrasonic pulse into a scattering medium, and the reflected pressure wave is received using a second (possibly the same) aperture. The pulse-echo signal is a radiofrequency trace whose amplitude in time corresponds to the echogenicity of the medium in depth. A raster scan of pulse-echo signals is used to reconstruct medical B-mode images, which display the echogenicity of the underlying tissue anatomy. Most medical pulse-echo signals contain speckle, a noise that arises due to diffuse unresolvable scatterers in the medium.1–4 Although speckle is a noise multiplicative with the echogenicity of tissue, it is used advantageously across a wide range of advanced medical ultrasound techniques that utilize the joint statistics of pairs of pulse-echo signals. In this work, we derive an expression for the complex correlation of two pulse-echo signals as a function of scatterer motion.

The complex correlation of two pulse-echo signals is a second-order statistic of particular interest. For instance, Doppler and speckle-tracking techniques measure the complex correlation between subsequent pulse-echo acquisitions of a moving medium; the displacement between acquisitions is proportional to the complex angle of the correlation,5–8 or to the time-shift that maximizes its normalized magnitude.9 Similarly, phase aberration correction and sound speed estimation techniques measure the complex angle of the correlation between neighboring receive elements to measure the relative time-of-arrival of the echo from a stationary medium.10–12 The Cramér-Rao lower bound of the displacement estimator variance is a function of the magnitude of the correlation coefficient between the two pulse-echo signals.1,7,9 Additionally, several adaptive beamforming methods based on spatial coherence have been proposed, such as generalized coherence factor,13 filter-delay-multiply-and-sum,14 phase coherence factor,15 and short-lag spatial coherence.16 The latter has been used to differentiate tissue signals from incoherent acoustical and electrical noise according to the correlation coefficients as a function of receive aperture spacing.16–18

More recently, correlation-based methods have been combined with a retrospective beamforming technique called synthetic aperture.19,20 Unlike traditional methods where a single tightly-focused pulse-echo signal is swept over the field of view (FOV), synthetic transmit apertures (STAs) are the retrospective coherent sums of multiple broad transmissions. STAs can be composed of different subapertures,19 virtual sources,21,22 or steered plane waves,23,24 and effectively image the medium using the aggregate transmit aperture. STAs provide transmit focusing throughout the entire FOV and enable significantly higher frame rates in excess of 1000 frames per second. The benefits of STAs are a balance between the quality of transmit focusing and the sensitivity to motion artifacts,20 the latter of which can be corrected using correlation-based techniques.25–27

STAs have augmented a wide range of correlation-based methods. For instance, STAs have enabled the acquisition of very long ensembles at high pulse-repetition frequencies (PRFs) for Doppler28 and vector velocity imaging29 by eliminating the need to raster-scan across the FOV. Long ensemble lengths in conjunction with eigen-based filtering30,31 have been used to detect extremely slow motion for applications such as functional ultrasound imaging of the brain.32 STAs have also been used for shear wave elastography imaging,33 intracardiac vector flow,34 aperture coherence imaging,35 phase aberration correction,36 and sound speed estimation.37,38 Each of these methods utilizes the correlation between two STA pulse-echo signals.

The second-order statistics of pulse-echo speckle signals (e.g., correlation, covariance, coherence) have been studied extensively in the literature. Goodman1,39 analyzed the statistics of speckle signals in general. Flax et al.40 and Wagner et al.4 described the correlations of speckle field points and their relationship to the system point spread function. Burckhardt,2 Trahey et al.,41 and Wagner et al.42 provided theoretical and experimental measurements of the correlation of translated receive apertures. Mallart and Fink43 provided a derivation for the correlation of two point receivers using an impulsional formulism44,45 based on the Fresnel approximation.46 Walker and Trahey47 extended the theory to large receive apertures in the presence of phase aberration.

In each of these descriptions of second-order statistics, the scattering function was assumed to remain stationary between the two pulse echo signals. However, motion is commonly encountered in medical ultrasonography, and in the case of motion estimation, is actually the target of interest. STAs are particularly susceptible to motion because they are acquired over a distributed period of time. There is a need for a motion-sensitive coherence theory in order to improve our understanding of how these techniques perform under scatterer motion.

Below, we present a general formulation for the correlation of two pulse-echo signals from diffuse scatterers undergoing bulk motion in Sec. II. The derivation is based on the Fresnel approximation, and predicts the correlation of backscatter from a given depth for arbitrary transmit and receive apertures. In Sec. III, we examine the theoretical consequences of several common types of motion. In Sec. IV, we demonstrate the predictive power of the theory in sample applications of phase-shift estimation and aperture coherence estimation, illustrating the impact of scatterer motion on correlation estimates.

II. THEORY

The imaging configuration described below is illustrated in Fig. 1. Below, we derive the correlation of two pulse-echo signals from scatterers that have undergone a uniform bulk motion.

FIG. 1.

FIG. 1.

A pair of pulse-echo configurations is depicted. The first is due to transmit T1, receive R1, and scattering function S. The second is due to transmit T2, receive R2, and scattering function S with a bulk uniform displacement. In the Fresnel approximation, the apertures and the scatterers are assumed to be at a distance z.

A. Transmitted pressure field

The pressure field transmitted by an aperture T is measured at field point Xs=[xs,ys,zs]T as a function of temporal frequency f as

H(Xs,f)=T(Xt,f)ej2πfrts/c2πrtsψ(θts)dXt, (1)

where Xt is a point on the transmitting aperture, rts is equal to ||XsXt||2 where ||·||2 denotes Euclidean distance, and ψ is an obliquity factor that depends on the angle θts between Xt and Xs.44–46

Assumption 1. Assume that the axial distance z=|zszt| is much larger than the transverse distance, i.e., zxsxt2, where the lower-case boldface xi=[xi,yi]T denotes the transverse coordinate vector. We can use a first order approximation for amplitude terms

rtsz, (2)

and a second order approximation for phase terms

rtsz(1+xsxt222z2). (3)

Additionally, Eq. (2) implies ψ(θts)1. These approximations are collectively referred to as the Fresnel or the “paraxial” approximation.46 Applied to Eq. (1), the transmitted pressure field is

H(xs,z,f)12πzT(xt,f)×exp[j2πzλ(1+xsTxs2xsTxt+xtTxt2z2)]dxt, (4)

where λ=c/f. This approximation is not valid for large steering angles. However, note that the approximation can be applied to steered phased arrays by rotating the coordinate system to align with the direction of wave propagation and by appropriately transforming the apertures, e.g., via projection/propagation onto the rotated z = 0 plane.

The quadratic term xtTxt can be incorporated as a parabolic focusing term into the transmit aperture definition as T(xt,f)ejπ(xtTxt)/(λz)T(xt,f), allowing the expression to be rewritten as

H(xs,z,f)12πzexp[j2πzλ]×T(xt,f)exp[jπλz(xsTxs2xsTxt)]dxt. (5)

In the case that the transmit aperture cannot be focused, Eq. (5) is equivalent to applying the stronger Fraunhofer or “far-field” approximation, where the quadratic term is ignored (i.e., xtTxt0).46 Below, we omit f and z as parameters for readability.

B. Backscattered pressure field

Consider a field of scatterers with a backscattering function S(Xs). When insonified by pressure field H(Xs), the backscattered pressure received by a receive aperture R with coordinates Xr is

P(T,S,R)=H(Xs)S(Xs)R(Xr)ej2πrsr/λ2πrsrdXsdXr. (6)

Let us assume that R has the same z coordinate as T. Plugging in Eq. (5) and applying the same approximations to the receive aperture sensitivity function, the total backscattered signal is approximated as

P(T,S,R)ej4πz/λ(2πz)2T(xt)S(xs)R(xr)×exp[j2πλz(xsT(xsxtxr))]dxrdxsdxt. (7)

C. Correlation of two backscattered signals

Consider two backscattered signals obtained with arbitrary transmit and receive apertures and scatterers, denoted P(T1,S1,R1) and P(T2,S2,R2). Their correlation is given as

Γ12=E[P(T1,S1,R1)P*(T2,S2,R2)] (8)
E[ej4π(z1z2)/λ(2π)4z12z22T1(xt1)T2*(xt2)×S1(xs1)S2*(xs2)R1(xr1)R2*(xr2)×exp[j2πxs1T(xs1xt1xr1)/(z1λ)]×exp[j2πxs2T(xs2xt2xr2)/(z2λ)]*×dxr1dxr2dxs1dxs2dxt1dxt2], (9)

where * denotes complex conjugation. From here, further assumptions are necessary to proceed with the analysis.

D. Correlation under bulk motion of diffuse scatterers

Assumption 2. Consider the case of spatially incoherent diffuse scatterers undergoing a constant and uniform bulk motion Xd=[xd,yd,zd]T, i.e.,

E[S1(X1)S2*(X2)]=|S0|2δ(X1X2+Xd). (10)

When S1 and S2 are the only random components in Eq. (9), the expectation can be taken inside the integral. Note in Eq. (9) that the scattering functions are weighted by the transmit pressure and receive sensitivity fields, meaning that assumption 2 only needs to be satisfied within the focal spot defined by the transmit and receive apertures.

Assumption 3. Assume that the axial displacement zd is small relative to z such that zd/(z+zd)0. The correlation can then be simplified by integrating over the delta function in Eq. (10) (see the Appendix for details),

Γ12(xd,zd)|S0|2(2πz)4ej4πzd/λej2πxdT(xd/(zλ)+2u)×T1~(u)R1~(u)[T2~(u+xdλz)R2~(u+xdλz)]*du, (11)

where X~(u) denotes the spatial two-dimensional Fourier transform of X(x) at spatial frequency vector u.

Equation (11) gives the monochromatic approximation of the correlation of two pulse-echo signals acquired from moving diffuse scatterers for arbitrary transmit and receive apertures for wavelength λ. The monochromatic expression can be extended to broadband (BB) apertures by incorporating the aperture frequency response and integrating over the frequency. In the case that all four apertures have the same response F(f),

Γ12broad(xd,zd)=|F(f)|4Γ12(xd,zd)df, (12)

where Γ12 is implicitly a function of f.

III. COMMON SCENARIOS

We illustrate the theoretical consequences of Eq. (11) in a few common scenarios of scatterer motion. For simplicity, we consider the monochromatic correlation Γ12.

A. Stationary scatterer case

In the case of no scatterer motion (i.e., xd=0 and zd = 0), the scatterer model reduces to

E[S1(X1)S2*(X2)]=|S0|2δ(X1X2). (13)

The correlation is then simplified as

Γ12(0,0)A2T1~(u)R1~(u)[T2~(u)R2~(u)]*du, (14)

where A=|S0|/(2πz)2.

B. Translating aperture case

In the case that the scatterers are stationary and the transmit and receive apertures are identical up to a translational offset, i.e., T(xt)=T1(xt)=T2(xt+Δt) and R(xr)=R1(xr)=R2(xr+Δr), the correlation reduces to

Γ12(0,0)A2|T~(u)|2|R~(u)|2ej2πuT(Δt+Δr)du. (15)

This is equivalent to the statement that the correlation between two backscattered signals is equal to the cross-correlation of the autocorrelations of the transmit and receive apertures, evaluated at lag Δt+Δr, which coincides with previous derivations for the correlation given stationary incoherent sources.43,47 This model applies to techniques that rely on coherence measurements between aperture elements, such as phase aberration correction, sound speed estimation, and aperture coherence imaging.

C. Estimation of purely axial scatterer motion

Consider the case of purely axial motion, i.e., xd=0 and zd0. The correlation simplifies to

Γ12(0,zd)Γ12(0,0)ej4πzd/λ. (16)

Note that axial motion effectively contributes a complex phase rotation to the correlation for stationary scatterers. For cases where Γ12(0,0) has no inherent complex phase (i.e., is real-valued), the displacement can be estimated using the phase-shift of the measured correlation Γ^12,

z^d=λ4πarg[Γ^12]. (17)

In the case where T1 = T2 and R1 = R2, the phase-shift estimator is sometimes called an “autocorrelator.” Phase-shift estimators are commonly employed in speckle tracking methods such as Color Doppler and elastography.

D. Estimation of jitter due to speckle decorrelation

Decorrelation of the underlying speckle fundamentally degrades the ability to estimate the true phase-shift between two backscattered signals. The amount of degradation depends on the correlation coefficient magnitude μ,1,6,9 which is defined as

μ=|Γ12(xd,zd)Γ11(0,0)Γ22(0,0)|. (18)

Decorrelation (i.e., μ<1) introduces jitter, causing the measurable phase-shift to deviate from the true phase-shift. The variance of the phase-shift error Δθ between two arbitrary speckle signals with correlation μ is1

σΔθ2=π23πarcsin(μ)+(arcsin(μ))212Li2(μ2), (19)

where Li2 is the dilogarithm function. The variance is zero when μ = 1 and grows rapidly as μ is reduced. Figure 2 plots the phase-shift estimator jitter as a function of μ for an example case of an 8 MHz signal with sound speed of 1540 m/s.

FIG. 2.

FIG. 2.

(Color online) The expected jitter of a phase-shift estimate is plotted as a function of the magnitude of the correlation coefficient between the two signals. For an 8 MHz signal and sound speed of 1540 m/s, the maximum detectable phase shift is 48 μm. The expected jitter is approximately 8 μm when μ=0.95.

In the monochromatic case for T1 = T2 and R1 = R2, Eqs. (16) and (19) predict that pure axial motion only rotates the complex phase (which has no effect on μ) and therefore does not contribute to estimator jitter.

E. Impact of transverse scatterer motion

In Eq. (11), transverse scatterer motion (i.e., xd0) induces the equivalent of both a transverse shift and modulation to one of the transmit-receive aperture pairs. Perfect signal correlation (i.e., μ = 1) can be achieved only when, for all u,

T1~(u)R1~(u)=T2~(u+xdλz)R2~(u+xdλz)ej4πuTxd. (20)

This equality is not satisfied in typical ultrasound aperture configurations (e.g., rectangular apertures), in which case μ<1. Thus transverse scatterer motion generally introduces jitter into axial phase-shift estimation.

F. Correlation of synthetic transmit apertures

In aperture synthesis, multiple acquisitions are coherently summed to retrospectively form an aggregate transmit aperture. Like traditional apertures, STAs can be used in correlation measurements. However, each STA component may be acquired from a different scattering function due to scatterer motion, which could cause decorrelation and increased jitter in STA-based phase-shift estimation.

Consider a STA synthesized from N transmit-receive aperture pairs (T1,R1),,(TN,RN). For two consecutive STA acquisitions, denote each component scattering function as S1,,S2N. The correlation of two STAs is the sum of the cross-correlations of each component,

ΓΣ1Σ2=E[p=1NP(Tp,Sp,R)q=1NP*(Tq,SN+q,R)]=p=1Nq=1NE[P(Tp,Sp,R)P*(Tq,SN+q,R)]. (21)

Assuming a constant uniform motion of Xd, i.e.,

E[Sp(X1)Sq*(X2)]|S0|2δ(X1X2+(qp)Xd), (22)

the correlation between two STA signals is expressed as

ΓΣ1Σ2=p=1Nq=1NΓpq((N+qp)xd,(N+qp)zd). (23)

For this model of scatterer motion, the autocorrelation of either synthetic signal is the same,

ΓΣ1Σ1=ΓΣ2Σ2=p=1Nq=1NΓpq((qp)xd,(qp)zd). (24)

Consequently, the correlation coefficient magnitude is simplified as

μ=|ΓΣ1Σ2ΓΣ1Σ1ΓΣ2Σ2|=|ΓΣ1Σ2ΓΣ1Σ1|. (25)

IV. NUMERICAL EXPERIMENTS

We demonstrate the presented theory in three applications of pulse-echo correlation with scatterer motion: (1) phase-shift estimation with focused apertures; (2) phase-shift estimation with plane wave STA using sequential or interleaved48 sequences; and (3) correlation coefficient measurements as a function of receive aperture spacing with STAs. The first two applications are examples of so-called autocorrelators in the sense that the same transmit and receive apertures are used for the two correlated signals; the third is an example of a cross-correlator, where different receive apertures are used. Each application was evaluated using numerical integration of Eqs. (11) and (12) and compared against matched pulse-echo simulations.

A. Numerical and simulation methods

The correlation Γ12 and the correlation coefficient magnitude μ were evaluated numerically using the matlab Symbolic Toolbox (The Mathworks, Natick, MA) with parameters for a Verasonics L12–3v transducer (128 elements, 200 μm pitch, c = 1540 m/s, z = 50 mm). The transducer was modeled with a center frequency of f = 8 MHz. For x within the domain of the aperture T, parabolic and plane wave transmit focusing was applied as

Tpar(x)=1, (26)
Tpln(x)=exp[j2πλ(xsinαxTx2z)], (27)

where α denotes the steering angle. [Note that these definitions take into account the parabolic focusing term ejπxTx/(λz) that was incorporated into the derivations of Eqs. (11) and (12).] Elevation displacement was not considered. The expected jitter was predicted using Eq. (19) based on the predicted value of μ. Monochromatic [narrowband (NB)] analysis was performed using Eq. (11), whereas BB analysis used Eq. (12) assuming a Gaussian frequency response with a fractional bandwidth of B = 60% about center frequency f0=8 MHz, i.e.,

F(f)=exp[π(ff0Bf0)2]. (28)

Paired simulations were executed using identical imaging parameters with Field II Pro,49–51 a C program that simulates the spatial impulse response of ultrasound transducers using linear systems theory. A small field of scatterers with a density of 30 scatterers per resolution cell was placed at a distance of z = 50 mm unless otherwise noted. The (x, y, z) extent of scatterers was roughly (8λ,13λ,4λ) to fully encompass the point spread function at the focal depth. Broadband simulations used a transducer fractional bandwidth of 60%, whereas NB simulations used a fractional bandwidth of 1%. A 2-cycle sinusoid excitation pulse was employed. For each experiment, 128 independent scatterer realizations were simulated to obtain the mean and standard deviation results. In the simulations, the transmit aperture was focused using the true geometric distance (i.e., without the Fresnel approximation),

Tgeo(x)=exp[j2πλ(xTx+z2xTx2z)]. (29)

B. Experiment 1: Focused transmit autocorrelator

A focused transmit aperture was used to acquire two consecutive pulse-echo signals with scatterer motion. Two cases were considered:

  • (1)

    Lateral displacement ranging from xd = 0 to 100 μm; fixed axial displacement of zd = 10 μm.

  • (2)

    Axial displacement ranging from zd = 0 to 100 μm; fixed lateral displacement of xd = 10 μm.

In the first case, the monochromatic theory was compared against BB simulations (60% bandwidth). (The BB theory and NB simulations are omitted for visual clarity; transducer bandwidth was observed to have minimal effect on sensitivity to lateral motion.) In the second case, monochromatic theory was compared against NB simulations (1% bandwidth), and BB theory against BB simulations. For reference, a 100 μm displacement observed at 10 kHz PRF corresponds to 1 m/s velocity.

Figure 3 plots the axial displacement estimates and jitter as a function of lateral and axial scatterer motions. Lateral scatterer motion did not bias the estimates but increased the jitter in both the theory and the simulations. Notably, the monochromatic theory showed good agreement with the BB simulations. Axial scatterer motion was accurately predicted and measured by both the NB and BB theories and simulations, with an aliasing artifact near zd48 μm, which corresponds to π radians for 1540 m/s sound speed at 8 MHz. Axial scatterer motion was found to increase the jitter only in the BB cases, whereas the NB apertures had no effect on jitter.

FIG. 3.

FIG. 3.

(Color online) Experiment 1: Focused transmits were used to measure axial displacements (left column) and jitter (right column) in the presence of lateral (top row) and axial (bottom row) scatterer motion. The theoretical results in Eqs. (11) and (12) were compared against NB and BB simulations. (a) Displacement estimates with lateral motion. (b) Estimator jitter with lateral motion. (c) Displacement estimates with axial motion. (d) Estimator jitter with axial motion.

The results of this numerical experiment suggest that decorrelation is likely caused by the re-weighting of scatterers passing through the focus of the imaging system. Axial motion through the tight axial focus of BB apertures caused decorrelation, whereas the same motion through the long and uniform focus of NB apertures had no effect. Similarly, lateral scatterer motion across the lateral focus of the transmit-receive beampattern (e.g., sinc-squared profile for rectangular apertures) caused decorrelation. These results indicate that a monochromatic analysis is insufficient for characterizing jitter under axial motion; by contrast the lateral effect appears to be independent of the bandwidth of the transducer, as shown by the good agreement between monochromatic theory and BB simulations. These findings are generally consistent with previous observations where tighter focusing and greater displacement magnitudes resulted in higher jitter in acoustic radiation force impulse (ARFI) elastography.52,53

C. Experiment 2: STA autocorrelator

The effects of scatterer motion were investigated for two types of STA autocorrelator sequences:

  • Sequential: [T1(1),,TN(1),T1(2),,TN(2)];

  • Interleaved: [T1(1),T1(2),,TN(1),TN(2)],

where Tn(m) denotes the nth transmit of the mth frame. The interleaved configuration has recently been shown to increase the maximum detectable phase-shift estimate N-fold for an N-pulse STA by reducing the effective PRF between the two STAs,48,54 and can be used to address aliasing artifacts like the one observed in Fig. 3(c). Plane wave STAs were formed using N = 3 angles steered at −3°, 0°, and 3° with either the sequential or interleaved sequence. Two consecutive frames were acquired. The scatterer displacement between frames was selected to match those of experiment 1; that is, a 10 μm displacement between frames was achieved using a 3.3 μm displacement between pulses. Here, only the more applicable BB theory and simulations were considered.

Figure 4 shows close agreement between the theoretical predictions and simulation measurements of axial motion. The results also match the BB results from experiment 1 (Fig. 3), both in trends and in absolute values. The sequential and interleaved STAs had similar performance in estimating the displacements, with lateral and axial motions causing increased jitter in both. Notably, interleaved STA eliminated the aliasing artifact caused by fast axial motion, but exhibited more jitter due to lateral motion than sequential STA.

FIG. 4.

FIG. 4.

(Color online) Experiment 2: Sequential and interleaved STAs were used to measure axial displacements (left column) and jitter (right column) in the presence of lateral (top row) and axial (bottom row) scatterer motion. The BB theoretical result is compared against BB simulations. Interleaved STA eliminates the aliasing artifact at the cost of increased jitter. (a) Displacement estimates with lateral motion. (b) Estimator jitter with lateral motion. (c) Displacement estimates with axial motion. (d) Estimator jitter with axial motion.

These results indicate that STAs perform similarly to ordinary focused transmits for the same effective displacements (i.e., per-frame for STAs versus per-pulse for focused transmits). Interestingly, intra-STA scatterer motion was not a significant source of jitter, an observation consistent with the evident widespread success of STA Doppler. Despite being composed of multiple distinct acquisitions, the results indicate that the correlation coefficient between two STAs remained similar to those of focused transmits, possibly due to the uniform nature of the motion and its presence in both signals.

D. Experiment 3: Receive aperture coherence with STA

In this experiment, the correlation coefficient was measured as a function of receive aperture spacing for a single STA. The STA was constructed using four non-overlapping transmit sub-apertures of 32 elements each. Scatterer motion was present between each transmission, and the resulting pulse-echo signals were summed coherently to form the STA. The receive apertures were selected to be single elements, separated by a lag 0Δr127d, where the element pitch was d = 0.2 mm. Lateral displacements of 0, 0.1, 1, and 10 mm per STA (i.e., per every 4 transmissions) and axial displacements of 0, 0.01, 0.1, and 1 mm per STA were considered. Lateral and axial motions were considered independent of one another. All analysis was performed with BB transducers.

Figure 5 plots the correlation coefficients as a function of receive aperture spacing for lateral and axial scatterer motion. Close agreement was observed between theoretical predictions and simulation measurements, including a piecewise behavior that was observed at multiples of the sub-aperture size of 32 elements. Small displacements less than 0.01 mm axially and 0.1 mm laterally showed small deviations from the predictions for stationary scatterers, which corresponds roughly to a triangle function for rectangular apertures.43,47 The robustness to minor intra-STA scatterer motion observed here is similar to that observed with STA Doppler in experiment 2, and is likely because both received signals observe the same intra-STA motion. Larger displacements degraded the correlation coefficient; such large decorrelations can likely be avoided in practice by using a sufficiently high PRF between STA components. Axial displacements caused greater decorrelation than equivalent transverse motions. A potential explanation for this disparity is that single-element receive apertures result in worse lateral resolution and thus exhibit a greater tolerance for motion across the widened lateral focus.

FIG. 5.

FIG. 5.

(Color online) Experiment 3: The theoretically predicted and simulation-measured correlation coefficients are plotted as a function of receive aperture spacing for lateral (top) and axial (bottom) scatterer motion. Large motions in either direction resulted in decorrelation throughout. (a) Lateral motion and (b) axial motion.

E. Limitations of the theory

The presented theory utilizes the Fresnel approximation to simplify the analysis. The range of validity of the Fresnel and Fraunhofer approximations are well-studied.46,55 We utilized a brief study (omitted here) to select an appropriate depth for analysis, and discovered that a focal depth of z = 5 cm or more was necessary to obtain reasonable match between theory and simulation. This depth corresponds to an f-number of approximately 2. However, many ultrasound imaging configurations utilize f-numbers smaller than this, possibly invalidating the Fresnel approximation (and possibly the small-displacement assumption) and leaving open-ended the utility of such a theoretical analysis.

The bulk uniform motion assumption is also commonly violated in pulse-echo imaging. For instance, blood flow is often modeled as having a parabolic displacement profile, and ARFI is modeled as producing a Gaussian displacement profile. In the latter case, rapid changes in the displacement profile within the focal beam (referred to as “shearing”) have been linked to increased jitter.52,53 Furthermore, real physiological motions are generally pulsatile and time-varying. The presented theory also does not consider other common decorrelating phenomena such as electronic noise from the system and acoustical noise from multiple reflections and strong off-axis scatterers, nor does it examine other implicit assumptions, such as a homogeneous speed of sound.

V. CONCLUSION

We have presented a theory for the correlation of motion in pulse-echo ultrasound. Equation (11) predicts the complex correlation between two pulse-echo ultrasound speckle signals, each due to an arbitrary transmit and receive aperture, with scattering functions that are simple translations of one another. The derivation relies on three primary assumptions: (1) the Fresnel approximation (|zszt|xsxt2); (2) bulk uniform motion of diffuse scatterers [Eq. (10)]; and (3) small axial displacements [zd/(z+zd)0]. The resulting expression describes correlation as a function of axial and transverse motion, and simplifies to previously derived formulations of correlation in the absence of motion.

The theory was then compared against simulations in three numerical experiments: focused transmit phase-shift estimation (Fig. 3), STA phase-shift estimation (Fig. 4), and receive aperture coherence with STAs (Fig. 5). A key theoretical result of this paper was that transverse scatterer motion causes decorrelation for typical ultrasound imaging aperture configurations, whereas axial scatterer motion causes decorrelation when using BB apertures. The theory successfully predicted that interleaved STAs increase the aliasing limit of phase-shift estimators at the cost of a slight increase in jitter. The theory also confirmed that intra-STA scatterer motion does not cause decorrelation when both STAs contain the identical motion, supporting the current widespread use of ultrafast STA Doppler techniques. Finally, the theory was used to predict receive-aperture correlation coefficients with STAs, which can be used for applications such as arrival time estimation and coherence-based imaging. The presented theory provides a framework for analyzing the correlations of arbitrary pulse-echo configurations for applications in which scatterer motion is expected.

ACKNOWLEDGMENTS

This research was supported by the National Institute of Biomedical Imaging and Bioengineering under Grant No. R01-EB013361. The authors would like to thank Dr. You Leo Li and Dr. Marko Jakovljevic for helpful discussions.

APPENDIX: CORRELATION OF BULK MOTION

Substituting Eq. (10) into Eq. (9) and integrating over the delta function such that xs=xs1=xs2xd and zd=z2z1, we have

Γ12|S0|2(2πz)4ej4πzd/λT1(xt1)T2*(xt2)×R1(xr1)R2*(xr2)exp[j2πλzxsT(xsxt1xr1)]×exp[j2πλz(xs+xd)T(xs+xdxt2xr2)]×exp[j2πλzzdz+zd(xs+xd)T(xs+xdxr2xt2)]×dxsdxt1dxt2dxr1dxr2, (A1)

where we approximate zz1z2 for amplitude terms.

By assumption 3, let us assume zd/(z+zd)0, allowing higher order phase terms to be ignored. Collecting terms containing xs,

Γ12|S0|2(2πz)4ej4πzd/λT1(xt1)T2*(xt2)×R1(xr1)R2*(xr2)exp[j2πλzxdT(xt2+xr2xd)]×exp[j2πλzxsT(xt2+xr2xt1xr12xd)]dxs×dxt1dxt2dxr1dxr2. (A2)

The inner integral evaluates to a delta function

Γ12|S0|2(2πz)4ej4πzd/λT1(xt1)T2*(xt2)×R1(xr1)R2*(xr2)exp[j2πλzxdT(xt2+xr2xd)]×δ(xt2+xr2xt1xr12xd)dxt1dxt2dxr1dxr2. (A3)

Letting xt=xt2+xr2xr12xd and integrating over the delta function, we have

Γ12|S0|2(2πz)4ej4πzd/λR1(xr1)R2*(xr2)×T1(xt)T2*(xtxr2+xr1+2xd)×exp[j2πλzxdT(xt+xr1+xd)]dxtdxr1dxr2. (A4)

Collecting xr2 terms inside the innermost integral and letting xr1=x1 and xr2=x2, the expression can be rearranged as

Γ12|S0|2(2πz)4ej4πzd/λR1(x1)T1(xt)×R2*(x2)T2*(xtx2+x1+2xd)dx2×exp[j2πλzxdT(xt+x1+xd)]dxtdx1. (A5)

Further, letting xt=xx1, we have

Γ12|S0|2(2πz)4ej4πzd/λej2πxdTxd/(λz)×exp[j2πλzxdTx](R1(x1)T1(xx1)dx1)×(R2(x2)T2(x+2xdx2)dx2)*dx. (A6)

Observe that the integral in Eq. (A6) is essentially the Fourier transform of two convolutions. At this point, we can utilize the convolution theorem to reduce the triple integral into a single integral using the Fourier transform. Letting Ai denote the convolution of Ti and Ri, the Fourier transform Ai~ can be expressed as

Ai~(u)=(Ti(xi)Ri(xxi)dxi)ej2πuTxdx=Ti(xi)Ri(x)ej2πuT(x+xi)dxdxi=Ti~(u)Ri~(u). (A7)

The integral in Eq. (A6) can be simplified as

exp[j2πλzxdTx]A1(x)A2*(x+2xd)dx=exp[j2πλzxdTx](A1~(u)ej2πuTxdu)×(A2~(v)ej2πvT(x+2xd)dv)*dx=A1~(u)A2~*(v)ej4πvTxd×exp[j2πxT(xdλz+uv)]dxdudv=A1~(u)A2~*(v)ej4πvTxdδ(xdλz+uv)dudv=A1~(u)A2~*(u+xdλz)ej4π(u+xd/(λz))Txddu. (A8)

Therefore,

Γ12|S0|2(2πz)4ej4πzd/λej2πxdT(xd/(zλ)+2u)×T1~(u)R1~(u)[T2~(u+xdλz)R2~(u+xdλz)]*du. (A9)

References

  • 1. Goodman J. W., Speckle Phenomena in Optics: Theory and Applications ( Roberts & Company, Englewood, CO, 2007). [Google Scholar]
  • 2. Burckhardt C. B., “ Speckle in ultrasound b-mode scans,” IEEE Trans. Sonics Ultrasonics 25(1), 1–6 (1978). 10.1109/T-SU.1978.30978 [DOI] [Google Scholar]
  • 3. Abbott J. G. and Thurstone F. L., “ Acoustic speckle: Theory and experimental analysis,” Ultrasonic Imag. 1(4), 303–324 (1979). 10.1177/016173467900100402 [DOI] [PubMed] [Google Scholar]
  • 4. Wagner R. F., Smith S. W., Sandrik J. M., and Lopez H., “ Statistics of speckle in ultrasound B-scans,” IEEE Trans. Sonics Ultrasonics 30(3), 156–163 (1983). 10.1109/T-SU.1983.31404 [DOI] [Google Scholar]
  • 5. Kasai C., Namekawa K., Koyano A., and Omoto R., “ Real-time two-dimensional blood flow imaging using an autocorrelation technique,” IEEE Trans. Sonics Ultrasonics 32(3), 458–464 (1985). 10.1109/T-SU.1985.31615 [DOI] [Google Scholar]
  • 6. Pinton G. F., Dahl J. J., and Trahey G. E., “ Rapid tracking of small displacements with ultrasound,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 53(6), 1103–1117 (2006). 10.1109/TUFFC.2006.1642509 [DOI] [PubMed] [Google Scholar]
  • 7. Walker W. F. and Trahey G. E., “ A fundamental limit on delay estimation using partially correlated speckle signals,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 42(2), 301–308 (1995). 10.1109/58.365243 [DOI] [Google Scholar]
  • 8. Loupas T., Powers J., and Gill R. W., “ An axial velocity estimator for ultrasound blood flow imaging, based on a full evaluation of the Doppler equation by means of a two-dimensional autocorrelation approach,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 42(4), 672–688 (1995). 10.1109/58.393110 [DOI] [Google Scholar]
  • 9. Carter G. C., “ Coherence and time delay estimation,” Proc. IEEE 75(2), 236–255 (1987). 10.1109/PROC.1987.13723 [DOI] [Google Scholar]
  • 10. Flax S. and O'Donnell M., “ Phase-aberration correction using signals from point reflectors and diffuse scatterers: Basic principles,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 35(6), 758–767 (1988). 10.1109/58.9333 [DOI] [PubMed] [Google Scholar]
  • 11. Liu D.-L. and Waag R. C., “ Time-shift compensation of ultrasonic pulse focus degradation using least-mean-square error estimates of arrival time,” J. Acoust. Soc. Am. 95(1), 542–555 (1994). 10.1121/1.408348 [DOI] [PubMed] [Google Scholar]
  • 12. Ivancevich N. M., Dahl J. J., and Smith S. W., “ Comparison of 3-d multi-lag cross-correlation and speckle brightness aberration correction algorithms on static and moving targets,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 56(10), 2157–2166 (2009). 10.1109/TUFFC.2009.1298 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Li P.-C. and Li M.-L., “ Adaptive imaging using the generalized coherence factor,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 50(2), 128–141 (2003). 10.1109/TUFFC.2003.1182117 [DOI] [PubMed] [Google Scholar]
  • 14. Matrone G., Savoia A. S., Caliano G., and Magenes G., “ The delay multiply and sum beamforming algorithm in ultrasound b-mode medical imaging,” IEEE Trans. Med. Imag. 34(4), 940–949 (2014). 10.1109/TMI.2014.2371235 [DOI] [PubMed] [Google Scholar]
  • 15. Camacho J., Parrilla M., and Fritsch C., “ Phase coherence imaging,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 56(5), 958–974 (2009). 10.1109/TUFFC.2009.1128 [DOI] [PubMed] [Google Scholar]
  • 16. Lediju M. A., Trahey G. E., Byram B. C., and Dahl J. J., “ Short-lag spatial coherence of backscattered echoes: Imaging characteristics,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 58(7), 1377–1388 (2011). 10.1109/TUFFC.2011.1957 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17. Pinton G. F., Trahey G. E., and Dahl J. J., “ Spatial coherence in human tissue: Implications for imaging and measurement,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 61(12), 1976–1987 (2014). 10.1109/TUFFC.2014.006362 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. Morgan M. R., Trahey G. E., and Walker W. F., “ Multi-covariate imaging of sub-resolution targets (mist),” IEEE Trans. Med. Imag. 38(7), 1690–1700 (2019). 10.1109/TMI.2019.2917021 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19. Karaman M., Li P.-C., and O'Donnell M., “ Synthetic aperture imaging for small scale systems,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 42(3), 429–442 (1995). 10.1109/58.384453 [DOI] [Google Scholar]
  • 20. Jensen J. A., Nikolov S. I., Gammelmark K. L., and Pedersen M. H., “ Synthetic aperture ultrasound imaging,” Ultrasonics 44, e5–e15 (2006). 10.1016/j.ultras.2006.07.017 [DOI] [PubMed] [Google Scholar]
  • 21. Frazier C. H. and O'Brien W. D., “ Synthetic aperture techniques with a virtual source element,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 45(1), 196–207 (1998). 10.1109/58.646925 [DOI] [PubMed] [Google Scholar]
  • 22. Bae M.-H. and Jeong M.-K., “ A study of synthetic-aperture imaging with virtual source elements in b-mode ultrasound imaging systems,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 47(6), 1510–1519 (2000). 10.1109/58.883540 [DOI] [PubMed] [Google Scholar]
  • 23. Montaldo G., Tanter M., Bercoff J., Benech N., and Fink M., “ Coherent plane-wave compounding for very high frame rate ultrasonography and transient elastography,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 56(3), 489–506 (2009). 10.1109/TUFFC.2009.1067 [DOI] [PubMed] [Google Scholar]
  • 24. Tanter M. and Fink M., “ Ultrafast imaging in biomedical ultrasound,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 61(1), 102–119 (2014). 10.1109/TUFFC.2014.2882 [DOI] [PubMed] [Google Scholar]
  • 25. Trahey G. E. and Nock L. F., “ Synthetic receive aperture imaging with phase correction for motion and for tissue inhomogeneities. II. Effects of and correction for motion,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 39(4), 496–501 (1992). 10.1109/58.148540 [DOI] [PubMed] [Google Scholar]
  • 26. Karaman M., Bilge H. S., and O'Donnell M., “ Adaptive multi-element synthetic aperture imaging with motion and phase aberration correction,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 45(4), 1077–1087 (1998). 10.1109/58.710591 [DOI] [PubMed] [Google Scholar]
  • 27. Jeong J.-S., Hwang J.-S., Bae M.-H., and Song T.-K., “ Effects and limitations of motion compensation in synthetic aperture techniques,” in Proceedings of the 2000 IEEE Ultrasonics Symposium. An International Symposium (Cat. No. 00CH37121) (2000), Vol. 2, pp. 1759–1762. [Google Scholar]
  • 28. Bercoff J., Montaldo G., Loupas T., Savery D., Mézière F., Fink M., and Tanter M., “ Ultrafast compound Doppler imaging: Providing full blood flow characterization,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 58(1), 134–147 (2011). 10.1109/TUFFC.2011.1780 [DOI] [PubMed] [Google Scholar]
  • 29. Jensen J. A., Nikolov S. I., Alfred C., and Garcia D., “ Ultrasound vector flow imaging—Part II: Parallel systems,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 63(11), 1722–1732 (2016). 10.1109/TUFFC.2016.2598180 [DOI] [PubMed] [Google Scholar]
  • 30. Gallippi C. M. and Trahey G. E., “ Adaptive clutter filtering via blind source separation for two-dimensional ultrasonic blood velocity measurement,” Ultrasonic Imag. 24(4), 193–214 (2002). 10.1177/016173460202400401 [DOI] [PubMed] [Google Scholar]
  • 31. Demené C., Deffieux T., Pernot M., Osmanski B.-F., Biran V., Gennisson J.-L., Sieu L.-A., Bergel A., Franqui S., Correas J.-M., Cohen I., Baud O., and Tanter M., “ Spatiotemporal clutter filtering of ultrafast ultrasound data highly increases Doppler and ultrasound sensitivity,” IEEE Trans. Med. Imag. 34(11), 2271–2285 (2015). 10.1109/TMI.2015.2428634 [DOI] [PubMed] [Google Scholar]
  • 32. Macé E., Montaldo G., Cohen I., Baulac M., Fink M., and Tanter M., “ Functional ultrasound imaging of the brain,” Nature Methods 8(8), 662–664 (2011). 10.1038/nmeth.1641 [DOI] [PubMed] [Google Scholar]
  • 33. Tiran E., Deffieux T., Correia M., Maresca D., Osmanski B.-F., Sieu L.-A., Bergel A., Cohen I., Pernot M., and Tanter M., “ Multiplane wave imaging increases signal-to-noise ratio in ultrafast ultrasound imaging,” Phys. Med. Biol. 60(21), 8549–8566 (2015). 10.1088/0031-9155/60/21/8549 [DOI] [PubMed] [Google Scholar]
  • 34. Fadnes S., Wigen M. S., Nyrnes S. A., and Lovstakken L., “ In vivo intracardiac vector flow imaging using phased array transducers for pediatric cardiology,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 64(9), 1318–1326 (2017). 10.1109/TUFFC.2017.2689799 [DOI] [PubMed] [Google Scholar]
  • 35. Bottenus N., Byram B. C., Dahl J. J., and Trahey G. E., “ Synthetic aperture focusing for short-lag spatial coherence imaging,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 60(9), 1816–1826 (2013). 10.1109/TUFFC.2013.2768 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36. Dahl J. J., Guenther D. A., and Trahey G. E., “ Adaptive imaging and spatial compounding in the presence of aberration,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 52(7), 1131–1144 (2005). 10.1109/TUFFC.2005.1503999 [DOI] [PubMed] [Google Scholar]
  • 37. Jaeger M., Held G., Peeters S., Preisser S., Grünig M., and Frenz M., “ Computed ultrasound tomography in echo mode for imaging speed of sound using pulse-echo sonography: Proof of principle,” Ultrasound Med. Biol. 41(1), 235–250 (2015). 10.1016/j.ultrasmedbio.2014.05.019 [DOI] [PubMed] [Google Scholar]
  • 38. Ali R. and Dahl J. J., “ Distributed phase aberration correction techniques based on local sound speed estimates,” in 2018 IEEE International Ultrasonics Symposium (IUS) (2018), pp. 1–4. [Google Scholar]
  • 39. Goodman J. W., Statistical Optics ( John Wiley & Sons, New York, 2015). [Google Scholar]
  • 40. Flax S. W., Glover G. H., and Pelc N. J., “ Textural variations in b-mode ultrasonography: A stochastic model,” Ultrasonic Imag. 3(3), 235–257 (1981). 10.1177/016173468100300302 [DOI] [Google Scholar]
  • 41. Trahey G. E., Smith S., and Von Ramm O., “ Speckle pattern correlation with lateral aperture translation: Experimental results and implications for spatial compounding,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 33(3), 257–264 (1986). 10.1109/T-UFFC.1986.26827 [DOI] [PubMed] [Google Scholar]
  • 42. Wagner R. F., Insana M. F., and Smith S. W., “ Fundamental correlation lengths of coherent speckle in medical ultrasonic images,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 35(1), 34–44 (1988). 10.1109/58.4145 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43. Mallart R. and Fink M., “ The van Cittert-Zernike theorem in pulse echo measurements,” J. Acoust. Soc. Am. 90(5), 2718–2727 (1991). 10.1121/1.401867 [DOI] [Google Scholar]
  • 44. Stepanishen P. R., “ Transient radiation from pistons in an infinite planar baffle,” J. Acoust. Soc. Am. 49(5B), 1629–1638 (1971). 10.1121/1.1912541 [DOI] [Google Scholar]
  • 45. Fink M. A. and Cardoso J.-F., “ Diffraction effects in pulse-echo measurement,” IEEE Trans. Sonics Ultrason. 31(4), 313–329 (1984). 10.1109/T-SU.1984.31512 [DOI] [Google Scholar]
  • 46. Goodman J. W., Introduction to Fourier Optics ( Roberts and Company Publishers, Greenwood Village, CO, 2005). [Google Scholar]
  • 47. Walker W. F. and Trahey G. E., “ Speckle coherence and implications for adaptive imaging,” J. Acoust. Soc. Am. 101(4), 1847–1858 (1997). 10.1121/1.418235 [DOI] [PubMed] [Google Scholar]
  • 48. Jensen J. A., “ Estimation of high velocities in synthetic aperture imaging—Part I: Theory,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 66(6), 1024–1031 (2019). 10.1109/TUFFC.2019.2906384 [DOI] [PubMed] [Google Scholar]
  • 49. Jensen J. A. and Svendsen N. B., “ Calculation of pressure fields from arbitrarily shaped, apodized, and excited ultrasound transducers,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 39(2), 262–267 (1992). 10.1109/58.139123 [DOI] [PubMed] [Google Scholar]
  • 50. Jensen J. A., “ Field: A program for simulating ultrasound systems,” Med. Biol. Eng. Comput. 34(1), 351–353 (1996).8945858 [Google Scholar]
  • 51. Jensen J. A., “ A multi-threaded version of Field II,” IEEE International Ultrasonics Symposium, IUS 2229-2232 (2014).
  • 52. McAleavey S. A., Nightingale K. R., and Trahey G. E., “ Estimates of echo correlation and measurement bias in acoustic radiation force impulse imaging,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 50(6), 631–641 (2003). 10.1109/TUFFC.2003.1209550 [DOI] [PubMed] [Google Scholar]
  • 53. Palmeri M. L., McAleavey S. A., Trahey G. E., and Nightingale K. R., “ Ultrasonic tracking of acoustic radiation force-induced displacements in homogeneous media,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 53(7), 1300–1313 (2006). 10.1109/TUFFC.2006.1665078 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54. Jensen J. A., “ Estimation of high velocities in synthetic aperture imaging—Part II: Experimental investigation,” IEEE Trans. Ultrason., Ferroelectr., Freq. Control 66(6), 1032–1038 (2019). 10.1109/TUFFC.2019.2906390 [DOI] [PubMed] [Google Scholar]
  • 55. Mezouari S. and Harvey A. R., “ Validity of Fresnel and Fraunhofer approximations in scalar diffraction,” J. Opt. A 5(4), S86–S91 (2003). 10.1088/1464-4258/5/4/360 [DOI] [Google Scholar]

Articles from The Journal of the Acoustical Society of America are provided here courtesy of Acoustical Society of America

RESOURCES