Skip to main content
Netherlands Heart Journal logoLink to Netherlands Heart Journal
. 2020 Jan 17;28(3):153–160. doi: 10.1007/s12471-019-01362-4

The Absorb bioresorbable vascular scaffold in real-world practice: long-term follow-up of the AMC Single Centre Real World PCI Registry

R Y G Tijssen 1, M E Annink 1, R P Kraak 1,2, K T Koch 1, J Baan Jr 1, M M Vis 1, J J Piek 1, J P S Henriques 1, R J de Winter 1, M A M Beijk 1, J J Wykrzykowska 1,
PMCID: PMC7052095  PMID: 31953774

Abstract

Background

Bioresorbable scaffolds have been introduced to overcome the shortcomings of drug-eluting stents. Higher rates of device thrombosis, however, have been reported up to 3 years after implantation of the Absorb bioresorbable vascular scaffold (BVS). In the current article, we therefore report long-term clinical outcomes of the AMC Absorb Registry.

Methods and results

In the AMC Absorb Registry, all patients who underwent a percutaneous coronary intervention with Absorb BVS implantation between 30 August 2012 and 5 August 2013 at the Amsterdam University Medical Centre—Academic Medical Centre were included. The composite endpoint of this analysis was target-vessel failure (TVF). The median follow-up of the study cohort of the AMC Absorb Registry was 1534 days. At the time of the cross-sectional data sweep the clinical status at 4 years was known in 124 of 135 patients (91.9%). At long-term follow-up, the composite endpoint of TVF had occurred in 27 patients. The 4‑year Kaplan-Meier estimate of TVF was 19.8%. At 4 years cardiac death had occurred in 4 patients (3.2%) and target-vessel myocardial infarction in 9 (6.9%) patients. Definite scaffold thrombosis occurred in 5 (3.8%) patients. We found 1 case of very late scaffold thrombosis that occurred at 911 days after device implantation in a patient who was not on dual anti-platelet therapy.

Conclusion

In a patient population reflecting routine clinical practice, we found that cases of TVF continued to accrue beyond 2 years after Absorb BVS implantation.

Keywords: Bioresorbable scaffolds, Target-vessel failure, Scaffold thrombosis

What’s new?

  • Our study provides the first long-term follow-up data on the use of Absorb BVS in a patient population reflecting daily clinical practice with regard to percutaneous coronary intervention.

  • It is also the first study that reports long-term follow-up data without a previous intervention in dual anti-platelet therapy (DAPT) strategies (prolonging or re-starting).

  • In this patient population, we found that cases of target-vessel failure continued to accrue beyond 2 years after Absorb BVS implantation.

  • Moreover, we found 1 case of very late scaffold thrombosis that occurred 911 days after device implantation in a patient who was not on DAPT.

Introduction

Coronary bioresorbable scaffolds have been developed to overcome the shortcomings of drug-eluting stents (DES). They are designed to provide temporary coronary scaffolding, in order to prevent acute recoil, and allow for vessel healing, and fully resorb over time [1]. The most widely used bioresorbable scaffold is the Absorb bioresorbable vascular scaffold (BVS) (Abbott Vascular, Santa Clara, CA, USA), which received CE approval in 2011 and FDA approval in 2016. Initial short-term results of studies conducted with the Absorb BVS were promising, with similar low rates of target-vessel failure (TVF) when compared to metallic DES [2]. Soon after the Absorb BVS became commercially available, the Amsterdam University Medical Centre—Academic Medical Centre (UMC-AMC) started to implant the Absorb BVS in a ‘real-world’ population, and started to follow these patients within the context of the AMC Single Centre Real World PCI Registry (hereafter referred as AMC Absorb Registry) [3]. The 2‑year results of the AMC Absorb Registry showed that the use of Absorb BVS in a patient registry reflecting daily clinical practice was associated with good procedural safety and acceptable clinical outcomes at mid-term (2-year) follow-up [4]. Higher rates of device thrombosis, however, have been reported up to 3 years after Absorb BVS implantation [5, 6]. Long-term follow-up after Absorb BVS implantation is therefore necessary in order to examine whether the annual event rates will decline after scaffold dismantling and resorption has been completed. In the current article, we therefore report long-term clinical outcomes of the AMC Absorb Registry.

Methods

The design of the AMC Absorb Registry, the baseline, the procedural characteristics, the 6‑month clinical outcomes and the 2‑year clinical outcomes, have been reported previously [3, 4]. Briefly, in the AMC Absorb Registry, all patients who underwent percutaneous coronary intervention (PCI) with Absorb BVS implantation between 30 August 2012 and the 5 August 2013 at the Amsterdam UMC—AMC were included. The decision whether to implant the Absorb BVS was left to the discretion of the operator. We included patients with a wide range of indications, from presentation with stable angina pectoris to presentation with acute coronary syndrome (ACS). The necessity to obtain written informed consent from the included patients was waved by the institutional review committee. All patients received dual anti-platelet therapy (DAPT) for at least 12 months. Clinical follow-up was conducted through telephone contact, and if not possible by live status examination. All reported events were adjudicated by experienced interventional cardiologists (Y. Onuma (Erasmus MC, Rotterdam), P. Suwannasom (Cardialysis B.V., Rotterdam), and M. Beijk (Amsterdam UMC, Amsterdam)).

Definitions

The composite endpoint of this analysis was TVF, which was defined as a composite of cardiac death, target-vessel myocardial infarction (TV-MI) and target-vessel revascularisation (TVR). Secondary endpoints were MI, TVR, target-lesion revascularisation and scaffold thrombosis (ScT). All events were defined in accordance with the definitions of the Academic Research Consortium.

Statistical analysis

Continuous data are expressed as mean ± standard deviation or as median (interquartile ranges). Dichotomous data are summarised as frequencies (%). Cumulative event rates were estimated using the Kaplan-Meier method. All statistical analyses were performed with SPSS software, version 23 (IBM Corp., Armonk, NY, USA).

Results

The median follow-up of the study cohort of the AMC Absorb Registry was 1534 days. At the time of the cross-sectional data sweep, the clinical status at 2 years was known in 132 of 135 (97.8%) patients, the clinical status at 3 years in 127 of 135 (94.1%) patients, and the clinical status at 4 years in 124 of 135 patients (91.9%). The baseline characteristics of the population are shown in Tab. 1. We enrolled 135 patients, in whom a total of 159 lesions were treated. Patients were predominantly male (73%); stable angina was the most common indication for PCI (47%). A total of 43 (40%) patients presented with ACS, of whom 17 (13%) presented with ST-elevation myocardial infarction.

Table 1.

Baseline characteristics of the study population

Patient characteristics (n=135)
Age (years)  59 ± 11
Male sex  98 (73%)
Diabetes mellitus  27 (20%)
Hypertension  67 (50%)
Hypercholesterolaemia  58 (43%)
Current smoker  39 (29%)
Renal dysfunction  11 (8%)
Previous myocardial infarction  34 (25%)
Previous PCI  35 (26%)
Previous CABG   3 (2%)
Multivessel disease  64 (47%)
Syntax Score  11.5  (IQ range: 6–17.5)
DAPT at discharge 135 (100%)
– Acetylsalicylic acid and clopidogrel  42 (31%)
– Acetylsalicylic acid and prasugrel  19 (14%)
– Acetylsalicylic acid and ticagrelor  74 (55%)
Indication for PCI STEMI  17 (13%)
NSTEMI  36 (27%)
Unstable angina  13 (10%)
Stable angina  63 (47%)
Other   6 (4%)
Lesion characteristics (n=159)
Vessels treated LMCA   2 (1%)
LAD  96 (60%)
RCx  24 (15%)
RCA  37 (23%)
Lesion type A  27 (17%)
B1  25 (16%)
B2  67 (42%)
C  40 (25%)
Bifurcation lesions  24 (15%)
Ostial lesions   5 (3%)
Calcified lesions  18 (11%)
Chronic total occlusion  13 (8%)
Thrombus present  14 (9%)

PCI percutaneous coronary intervention, CABG coronary artery bypass graft, IQ interquartile range, DAPT dual antiplatelet therapy, STEMI ST-elevation myocardial infarction, NSTEMI non-STEMI, LMCA left main coronary artery, LAD left anterior descending artery, RCx ramus circumflex artery, RCA right coronary artery

An extensive description of the procedural and lesion characteristics has been published previously [7]. Briefly, most of the lesions treated with Absorb BVS (67%) were classified as type B2 or C (American Heart Association/American College of Cardiology classification); pre-dilatation was performed in 98% and post-dilatation in 55% of the treated lesions. The Syntax score ranged from 1 to 50, with a medium of 11.5 (interquartile range: 6–17.5). The clinical outcomes of all patients are shown in Tab. 2. At long-term follow-up, the composite endpoint of TVF had occurred in 27 patients. The 4‑year Kaplan-Meier estimate of TVF was 19.8% (Fig. 1). At 4 years cardiac death had occurred in 4 patients (3.2%) and TV-MI in 9 (6.9%) patients. Definite ScT had occurred in 5 (3.8%) patients. We found 1 case of very late ScT that occurred 911 days after device implantation in a patient who was not on DAPT. A detailed description of the cases with definite ScT is shown in Tab. 3. We found no cases of probable or possible device thrombosis.

Table 2.

Long-term clinical outcomes of the study population

Outcome Patients with an event 4‑year cumulative event rate
All-cause mortality 10  7.1%
Cardiac death  4  3.2%
Myocardial infarction 14  9.1%
Target-vessel myocardial infarction  9  6.9%
Target-vessel revascularisation 23 16.7%
Target-lesion revascularisation 18 14.6%
Definite scaffold thrombosis  5  3.8%
Probable/possible scaffold thrombosis  0  0%
Target-vessel failure 27 19.8%

Fig. 1.

Fig. 1

Kaplan-Meier curve of the outcome of target-vessel failure

Table 3.

Detailed description of the cases of scaffold thrombosis

Case Initial PCI indication Treated vessel Lesion type Calcification Pre-dilatation Pre-dilatation balloon Absorb size inflation pressure Post-dilatation Post-dilatation balloon On DAPT Timing (days) Treatment Possible reason
1 OHCA Mid LAD A No Yes 2.5 × 15 mm 3.0 × 18 mm (12) Yes 2.5 × 15 mm NC (8) Yes   4 Thrombus aspiration Distal edge dissection
Distal LAD B2 No Yes 2.5 × 15 mm 3.0 × 28 mm (14) Yes 2.5 × 15 mm NC (10)

Xience 3.0 × 38 mm

Xience 3.0 × 18 mm

2 NSTEMI Proximal LAD B2 No No 2.5 × 18 mm (16) No Yes   8

Thrombus aspiration

Xience 2.5 × 28 mm

Incomplete expansion distal part of scaffold
3 Unstable AP Distal LAD B1 No Yes 2.5 × 15 mm 2.5 × 28 mm (24) Yes 2.5 × 15 mm NC (24) No  24

Thrombus aspiration

Xience 2.5 × 28 mm

DAPT cessation
4 NSTEMI Proximal RCx B1 No Yes 2.5 × 15 mm 3.5 × 28 mm (10) Yes 3.5 × 16 mm NC (16) No 112 Thrombus aspiration multiple balloon inflations DAPT cessation
5 Stable AP Proximal RCA A No Yes 2.5 × 15 mm 3.0 × 18 mm (10) No No 911 Azule 3.0 × 23 mm DAPT cessation, intraluminal scaffold dismantling

OHCA out-of-hospital cardiac arrest, LAD left anterior descending artery, NC non-compliant, NSTEMI non-ST-elevation myocardial infarction, AP angina pectoris, DAPT dual antiplatelet therapy, RCx ramus circumflex artery, RCA right coronary artery

Discussion

In the AMC Absorb Registry, the composite endpoint of TVF continued to accrue beyond 2 years after Absorb BVS implantation. Additionally, we found 1 case of very late ScT that occurred 911 days after device implantation in a patient who was not on DAPT. Absorb BVS was initially expected to be fully resorbed at 2 years after device implantation. Several studies have, however, reported that (very) late ScT also frequently occurs between the 2nd and 3rd year after device implantation, and even beyond 3 years, with the latest case of ScT reported as late as 44 months after Absorb BVS implantation [5, 6, 8, 9]. In these studies, the leading mechanisms of very late ScT were associated with (disintegrated) strut material that protruded into the coronary lumen, most likely as a consequence of intraluminal scaffold dismantling or late (acquired) malapposition [10]. DAPT is hypothesised to be an important factor in the prevention of thrombotic events after coronary stent implantation. When the preliminary safety report of the Amsterdam Investigator-initiateD Absorb Strategy All-comers (AIDA) trial (an international, multicentre, randomised trial in which the BVS was compared to DES) was published due to safety concerns regarding device thrombosis, the AIDA steering committee recommended cardiologists to consider re-starting or prolonging DAPT in patients treated with Absorb BVS up to 3 years after device implantation [11]. At that time point, all patients in the AMC Absorb Registry were already beyond 3 years after Absorb BVS implantation, so there was no additional medical intervention in this group, unlike in the AIDA trial. At 4‑year follow-up, in the AMC Absorb Registry, the estimated definite ScT rate was 3.8%, whereas in the comparable AIDA population the definitive ScT rate was 3.3% at 3 years. To date, it remains uncertain whether prolonged DAPT actually prevents (very late) ScT. Notably, in a large meta-analysis, 92% of the cases of very late ScT occurred in patients that were not on DAPT at the time of the event [12].

One of the causes hypothesised to be a (major) contributor to the adverse outcomes of Patients treated with Absorb BVS was suboptimal Absorb BVS implantation techniques. Optimised Absorb BVS implantation techniques, generally based on a pre-dilatation, sizing and post-dilatation (PSP) strategy were thereafter hypothesised to optimise outcomes after Absorb BVS implantation. The most correct definition of correct PSP implantation remains unknown, however, and its effect on the improvement of outcomes remains cumbersome and varies between studies [1316]. Moreover, while suboptimal implantation techniques might explain (acute and early) events up to 3 years after implantation, it is difficult to stipulate that implantation techniques at the initial procedure impact outcomes at 4–5 years after device implantation, since the complex, and sometimes unpredictable and irregular, resorption process of the device occurs during this period.

In the ABSORB II trial, a downturn of events has been reported between the 3rd and 4th year of follow-up, thereby positively impacting the difference between the target-lesion failure rates of Absorb BVS and Xience EES at 4‑year follow-up [17]. Moreover, within the ABSORB III trial, in contrast to the pattern observed before 3 years, the event rates were similar between the Absorb BVS and Xience EES groups after 3 years [18]. The downturn in device-related events beyond 3 years after implantation in Absorb BVS patients within the ABSORB II and ABSORB III trials is encouraging. However, in the ABSORB II and ABSORB III trials, the patient populations were selected. In ABSORB II, patients that presented with an MI were excluded; and unstable patients and those with complex lesions were excluded from ABSORB III [19, 20].

Long-term follow-up, with precisely documented and investigated DAPT regimens/strategies, of large randomised studies such as the AIDA and the ABSORB IV trial, is therefore necessary in order to establish whether the annual rates of device-related-events in patients treated with Absorb BVS in routine clinical practice will decline after the period of complete scaffold dismantling and resorption.

Study limitations

This study is a registry, and therefore a control group is lacking. Second, the decision whether to implant an Absorb BVS was left to the discretion of the operator, and therefore potential patient selection bias has been introduced. Finally, routine intracoronary imaging has not been performed, and therefore potential information on the mechanism of scaffold failure might have been missed.

Conclusion

In a patient population reflecting routine clinical practice, we found that cases of TVF continued to accrue beyond 2 years after Absorb BVS implantation.

Conflict of interest

J.P.S. Henriques receives research grants from Abbott Vascular. J.J. Wykrzykowska receives consultancy fees and research grants from Abbott Vascular. R.Y.G. Tijssen, M.E. Annink, R.P. Kraak, K.T. Koch, J. Baan Jr., M.M. Vis, J.J. Piek, R.J. de Winter and M.A.M. Beijk declare that they have no competing interests.

References

  • 1.Serruys PW, Garcia-Garcia HM, Onuma Y. From metallic cages to transient bioresorbable scaffolds: change in paradigm of coronary revascularization in the upcoming decade? Eur Heart J. 2012;33(1):16–25b. doi: 10.1093/eurheartj/ehr384. [DOI] [PubMed] [Google Scholar]
  • 2.Stone GW, Gao R, Kimura T, Kereiakes DJ, Ellis SG, Onuma Y, et al. 1-year outcomes with the Absorb bioresorbable scaffold in patients with coronary artery disease: a patient-level, pooled meta-analysis. Lancet. 2016;387(10025):1277–1289. doi: 10.1016/S0140-6736(15)01039-9. [DOI] [PubMed] [Google Scholar]
  • 3.Kraak RP, Hassell ME, Grundeken MJ, Koch KT, Henriques JP, Piek JJ, et al. Initial experience and clinical evaluation of the Absorb bioresorbable vascular scaffold (BVS) in real-world practice: the AMC Single Centre Real World PCI Registry. EuroIntervention. 2015;10(10):1160–1168. doi: 10.4244/EIJY14M08_08. [DOI] [PubMed] [Google Scholar]
  • 4.Kraak RP, Grundeken MJ, Hassell ME, Elias J, Koch KT, Henriques JP, et al. Two-year clinical outcomes of Absorb bioresorbable vascular scaffold implantation in complex coronary artery disease patients stratified by SYNTAX score and ABSORB II study enrolment criteria. EuroIntervention. 2016;12(5):e557–e565. doi: 10.4244/EIJV12I5A95. [DOI] [PubMed] [Google Scholar]
  • 5.Serruys PW, Chevalier B, Sotomi Y, Cequier A, Carrie D, Piek JJ, et al. Comparison of an everolimus-eluting bioresorbable scaffold with an everolimus-eluting metallic stent for the treatment of coronary artery stenosis (ABSORB II): a 3 year, randomised, controlled, single-blind, multicentre clinical trial. Lancet. 2016;388(10059):2479–2491. doi: 10.1016/S0140-6736(16)32050-5. [DOI] [PubMed] [Google Scholar]
  • 6.Kerkmeijer LSM, Tijssen RYG, Hofma SH, van der Schaaf RJ, Arkenbout KE, Kraak RP, et al. Comparison of an everolimus-eluting bioresorbable scaffold with an everolimus-eluting metallic stent in routine PCI: three-year clinical outcomes from the AIDA trial. EuroIntervention. 2019;15:603. doi: 10.4244/EIJ-D-19-00325. [DOI] [PubMed] [Google Scholar]
  • 7.Kraak RP, Hassell ME, Grundeken MJ, et al. Initial experience and clinical evaluation of the Absorb bioresorbable vascular scaffold (BVS) in real-world practice: the AMC Single Centre Real World PCI Registry. EuroIntervention. 2015;10(10):1160–1168. doi: 10.4244/EIJY14M08_08. [DOI] [PubMed] [Google Scholar]
  • 8.Yamaji K, Ueki Y, Souteyrand G, Daemen J, Wiebe J, Nef H, et al. Mechanisms of very late bioresorbable scaffold thrombosis: the INVEST registry. J Am Coll Cardiol. 2017;70(19):2330–2344. doi: 10.1016/j.jacc.2017.09.014. [DOI] [PubMed] [Google Scholar]
  • 9.Raber L, Brugaletta S, Yamaji K, O’Sullivan CJ, Otsuki S, Koppara T, et al. Very late scaffold thrombosis: intracoronary imaging and histopathological and spectroscopic findings. J Am Coll Cardiol. 2015;66(17):1901–1914. doi: 10.1016/j.jacc.2015.08.853. [DOI] [PubMed] [Google Scholar]
  • 10.Kraak RP, Kajita AH, Garcia-Garcia HM, Henriques JPS, Piek JJ, Arkenbout EK, et al. Scaffold thrombosis following implantation of the ABSORB BVS in routine clinical practice: Insight into possible mechanisms from optical coherence tomography. Catheter Cardiovasc Interv. 2018;92(2):E106–E114. doi: 10.1002/ccd.27475. [DOI] [PubMed] [Google Scholar]
  • 11.Wykrzykowska JJ, Kraak RP, Hofma SH, van der Schaaf RJ, Arkenbout EK, IJsselmuiden AJ, et al. Bioresorbable scaffolds versus metallic stents in routine PCI. N Engl J Med. 2017;376(24):2319–2328. doi: 10.1056/NEJMoa1614954. [DOI] [PubMed] [Google Scholar]
  • 12.Collet C, Asano T, Miyazaki Y, et al. Late thrombotic events after bioresorbable scaffold implantation: a systematic review and meta-analysis of randomized clinical trials. Eur Heart J. 2017;38(33):2559–2566. doi: 10.1093/eurheartj/ehx155. [DOI] [PubMed] [Google Scholar]
  • 13.Ortega-Paz L, Capodanno D, Gori T, Nef H, Latib A, Caramanno G, et al. Predilation, sizing and post-dilation scoring in patients undergoing everolimus-eluting bioresorbable scaffold implantation for prediction of cardiac adverse events: development and internal validation of the PSP score. EuroIntervention. 2017;12(17):2110–2117. doi: 10.4244/EIJ-D-16-00974. [DOI] [PubMed] [Google Scholar]
  • 14.Stone GW, Abizaid A, Onuma Y, Seth A, Gao R, Ormiston J, et al. Effect of technique on outcomes following bioresorbable vascular scaffold implantation: analysis from the ABSORB trials. J Am Coll Cardiol. 2017;70(23):2863–2874. doi: 10.1016/j.jacc.2017.09.1106. [DOI] [PubMed] [Google Scholar]
  • 15.Tijssen RYG, Kerkmeijer LSM, Katagiri Y, Kraak RP, Takahashi K, Kogame N, et al. The relationship of pre-procedural Dmax based sizing to lesion level outcomes in Absorb BVS and Xience EES treated patients in the AIDA trial. Int J Cardiovasc Imaging. 2019;35(7):1189–1198. doi: 10.1007/s10554-019-01576-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Tijssen RYG, Kraak RP, Elias J, van Dongen IM, Kalkman DN, Nassif M, et al. Implantation techniques (predilatation, sizing, and post-dilatation) and the incidence of scaffold thrombosis and revascularisation in lesions treated with an everolimus-eluting bioresorbable vascular scaffold: insights from the AIDA trial. EuroIntervention. 2018;14(4):e434–e442. doi: 10.4244/EIJ-D-17-01152. [DOI] [PubMed] [Google Scholar]
  • 17.Chevalier B, Cequier A, Dudek D, Haude M, Carrie D, Sabate M, et al. Four-year follow-up of the randomised comparison between an everolimus-eluting bioresorbable scaffold and an everolimus-eluting metallic stent for the treatment of coronary artery stenosis (ABSORB II Trial) EuroIntervention. 2018;13(13):1561–1564. doi: 10.4244/EIJ-D-17-00873. [DOI] [PubMed] [Google Scholar]
  • 18.Kereiakes DJ, Ellis SG, Metzger DC, Caputo RP, Rizik DG, Teirstein PS, et al. Clinical outcomes prior to and following complete everolimus-eluting bioresorbable scaffold resorption: five-year follow-up from the ABSORB III trial. Circulation. 2019;140:1903. doi: 10.1161/CIRCULATIONAHA.119.042584. [DOI] [PubMed] [Google Scholar]
  • 19.Diletti R, Serruys PW, Farooq V, Sudhir K, Dorange C, Miquel-Hebert K, et al. ABSORB II randomized controlled trial: a clinical evaluation to compare the safety, efficacy, and performance of the Absorb everolimus-eluting bioresorbable vascular scaffold system against the XIENCE everolimus-eluting coronary stent system in the treatment of subjects with ischemic heart disease caused by de novo native coronary artery lesions: rationale and study design. Am Heart J. 2012;164(5):654–663. doi: 10.1016/j.ahj.2012.08.010. [DOI] [PubMed] [Google Scholar]
  • 20.Kereiakes DJ, Ellis SG, Popma JJ, Fitzgerald PJ, Samady H, Jones-McMeans J, et al. Evaluation of a fully bioresorbable vascular scaffold in patients with coronary artery disease: design of and rationale for the ABSORB III randomized trial. Am Heart J. 2015;170(4):641–51 e3. doi: 10.1016/j.ahj.2015.07.013. [DOI] [PubMed] [Google Scholar]

Articles from Netherlands Heart Journal are provided here courtesy of Springer

RESOURCES