Epicardial left ventricular lead implantation in cardiac resynchronization therapy patients via a video‐assisted thoracoscopic technique: Long‐term outcome

Massimiliano Marini 1,, Stefano Branzoli 2, Paolo Moggio 1, Marta Martin 1, Giuseppina Belotti 3, Giulio Molon 4, Fabrizio Guarracini 1, Alessio Coser 1, Silvia Quintarelli 1, Carlo Pederzolli 2, Angelo Graffigna 2, Daniele Penzo 5, Sergio Valsecchi 6, Maria Caterina Bottoli 7, Patrizia Pepi 8, Roberto Bonmassari 1, Andrea Droghetti 7
PMCID: PMC7068064  PMID: 31837030

Abstract

Background

Epicardial placement of the left ventricular (LV) lead via a video‐assisted thoracoscopic (VAT) approach is an alternative to the standard transvenous technique.

Hypothesis

Long‐term safety and efficacy of VAT and transvenous LV lead implantation are comparable. To test it, we reviewed our experience and we compared the outcomes of patients who underwent implantation with the two techniques.

Methods

The VAT procedure is performed under general anesthesia, with oro‐tracheal intubation and right‐sided ventilation, and requires two 5 mm and one 15 mm thoracoscopic ports. After pericardiotomy at the spot of the epicardial target area, pacing measurements are taken and a spiral screw electrode is anchored at the final pacing site. The electrode is then tunneled to the pectoral pocket and connected to the device.

Results

105 patients were referred to our center for epicardial LV lead implantation. After pre‐operative assessment, 5 patients were excluded because of concomitant conditions precluding surgery. The remaining 100 underwent the procedure. LV lead implantation was successful in all patients (median pacing threshold 0.8 ± 0.5 V, no phrenic nerve stimulation) and cardiac resynchronization therapy was established in all but one patient. The median procedure time was 75 min. During a median follow‐up of 24 months, there were no differences in terms of death, cardiovascular hospitalizations or device‐related complications vs the group of 100 patients who had undergone transvenous implantation. Patients of both groups displayed similar improvements in terms of ventricular reverse remodeling and functional status.

Conclusions

Our VAT approach proved safe and effective, and is a viable alternative in the case of failed transvenous LV implantation.

Keywords: CRT, epicardial LV lead implantation, video‐assisted thoracoscopic approach

1. INTRODUCTION

Epicardial placement of left ventricular (LV) lead has been proposed as an alternative approach in the case of failure of the transvenous approach during cardiac resynchronization therapy (CRT) device implantation. In our center we developed a minimally invasive video‐assisted thoracoscopic (VAT) technique. We reviewed our experience in order to evaluate the safety and efficacy of this technique. We also assessed long‐term safety and efficacy by comparing the outcome of the study group with that of a control group of patients who underwent standard transvenous LV lead implantation.

2. METHODS

2.1. Patient selection

The study was approved by the Institutional Review Board and all subjects provided written consent. Patients underwent baseline evaluation, which included demographics and medical history, clinical examination, 12‐lead electrocardiogram, and echocardiogram. Patients were also evaluated by means of a CT scan to rule out any thoraco‐pulmonary disease, spirometry with diffusing capacity of the lung for carbon monoxide, and anesthesiology evaluation. In all patients, previous LV lead implantation had been unsuccessful, owing to unsuitable coronary or subclavian venous anatomy, LV lead dislodgement or failure, phrenic nerve stimulation not correctable by reprogramming, or lead extraction because of infection.

2.2. Surgical technique and approach

The technique has been previously described.1 In summary, after a general anesthetic had been administered, a double‐lumen endotracheal tube was inserted. With the patient in a right lateral decubitus position (90°) and on single‐lung ventilation, three ports (2 × 5 mm; 1 × 15 mm) were inserted in the left hemithorax. The pericardium was opened posterior to the phrenic nerve. A bare area of the heart, free from visible scarring, in the posterolateral region was visualized between the diagonal and obtuse marginal arteries. A MyoPore sutureless myocardial bipolar pacing lead (Greatbach Medical, New York) was anchored via a thoracoscopic approach by means of the FasTacTM Flex (Greatbatch Medical) steerable lead implantation tool. Finally, the patient was returned to the supine position and the electrode was tunneled to the pectoral pocket and connected to the device.

2.3. Patient management

Patients were extubated in the operating room or in the Intensive Care Unit. The chest tube was removed 12 to 24 hours after surgery. No redo surgery was necessary. After discharge, clinic visits were scheduled every 6 months.

2.4. Control group

A group of 100 consecutive patients who underwent successful transvenous CRT implantation in our center was used as a control group.

2.5. Endpoints

Primary end‐points were: time to death due to any cause and time to the combination of death and cardiovascular hospitalization. Additional endpoints were: device‐related complications, LV ejection fraction and volume change, NYHA class, and ventricular pacing parameters.

2.6. Statistical analysis

Descriptive statistics are reported as means ± SD for normally distributed continuous variables, or medians with 25th to 75th percentiles in the case of skewed distribution. Differences between mean data were compared by means of a T‐test for Gaussian variables. The Mann‐Whitney test and the Wilcoxon non‐parametric test were used to compare non‐Gaussian variables for independent and paired samples, respectively. Differences in proportions were compared by applying Chi‐square analysis or Fisher's exact test, as appropriate. Event rates were summarized by constructing Kaplan‐Meier curves. The log‐rank test was applied in order to evaluate differences between trends. A P < .05 was considered significant for all tests. All statistical analyses were performed by means of SPSS Statistics, software, version 20 (IBM Corp, New York, New York).

3. RESULTS

3.1. Study population

From January 2008 through February 2017, 105 consecutive patients with standard CRT indications were referred to our center for epicardial LV lead implantation. After pre‐operative assessment, 5 patients were excluded because of concomitant conditions precluding surgery. The remaining 100 patients underwent the surgical procedure.

An additional group of 100 consecutive patients who had undergone successful transvenous LV lead implantation in the period between 2012 and 2015 was analyzed in order to compare long‐term outcomes and event rates. Table 1 shows the clinical characteristics of the two groups.

Table 1.

Demographics and baseline characteristics of the study population

Characteristics Thoracoscopic (100) Transvenous (100) P‐value
Male sex 69 (69%)a 80 (80%)a .074b
Age (years) 73 (65‐77)c 74 (68‐76) § .578d
Ischemic cardiomyopathy 45 (45%)a 47 (47%)a .777b
Primitive cardiomyopathy 39 (39%)a 42 (42%)a .666b
NYHA III or IV 69 (69%)a 59 (59%)a .141b
QRS duration (ms) 160 (155‐190)c 160 (135‐173)c .021d
LBBB QRS morphology 51 (51%)a 62 (62%)a .117b
RBBB QRS morphology 6 (6%)a 17 (17%)a .015b
AF on implantation 28 (28%)a 22 (22%)a .327b
History of AF 58 (58%)a 41 (41%)a .016b
Arterial hypertension 70 (70%)a 61 (61%)a .181b
Diabetes 31 (31%)a 32 (32%)a .879b
COPD 25 (25%)a 13 (13%)a .031b
CKD 47 (47%)a 25 (25%)a .001b
Previous cardiac surgery 35 (35%)a 24 (24%)a .088b
ACE‐inh/ARB 76 (76%)a 70 (70%)a .339b
B‐blockers 91 (91%)a 85 (85%)a .192b
MRA 61 (61%)a 57 (57%)a .565b
Anticoagulant 53 (53%)a 40 (40%)a .065b
Antiarrhythmic drug 34 (34%)a 18 (18%)a .010b
LVEF (%) 28 (21‐32)c 28 (25‐33)c .308d
LVEDV (mL) 172 (138‐241)c 167 (142‐223)c .465d
LVESV (mL) 123 (102‐175)c 123 (95‐167)c .487d
Follow‐up (months) 24 (13–39)c 32 (24–46)c .003d

Abbreviations: AF, atrial fibrillation; CKD, chronic kidney disease; COPD, chronic obstructive pulmonary disease; LBBB, left bundle branch block; LVEDV, left ventricular end diastolic volume; LVEF, left ventricular ejection fraction; LVESV, left ventricular end systolic volume; MRA, mineralocorticoid receptor antagonists; RBBB, right bundle branch block.

a

Number of cases (percentage %).

b

Chi‐square test.

c

Median (Q1‐Q3).

d

Mann‐Whitney U test.

In the thoracoscopic group, LV lead implantation was successful in all patients and CRT was successfully established in all but one patient at the end of the procedure. The median pacing threshold was 0.8 V at 0.5 ms (IQR 0.6‐1.2) and no phrenic nerve stimulation was reported at the final electrode position. Fixation in a basal LV segment was achieved in 90 (90%) patients. The median total procedure time was 75 min (IQR 55‐95) and the time to epicardial lead fixation was 30 min (IQR 15‐40). The procedure time was significantly longer in the presence of pleural adherence (127 ± 47 min vs 71 ± 25 min, with vs without, P < .001) or pericardial adherence (115 ± 48 min vs 74 ± 30 min, with vs without, P = .004). No complications were reported, except for two cases of transitory peri‐electrode bleeding and three cases of ventricular fibrillation induced during the procedure. No sequelae were reported for all these events. During the post‐operative hospital stay, 12 complications occurred in nine patients: five cases of worsening heart failure, one followed by death, four pocket or chest‐wall hematomas (only one requiring intervention), one episode of ventricular tachycardia correctly interrupted by the ICD, one episode of atrial fibrillation, and one dislocation of the right ventricular defibrillation lead, causing inappropriate shocks. The median hospital stay after the procedure was 5 days (IQR 3‐7).

In the transvenous group, the median LV lead pacing threshold was 1.1 V at 0.5 ms. The final position of the LV lead was lateral‐anterolateral. Table 2 compares procedural data of the two groups.

Table 2.

Pacing threshold parameters and final position of the LV lead

Parameter Thoracoscopic (100) Transvenous (100) P‐value
LV lead pacing parameters Threshold (V at 0.5 ms) 0.8 (0.6‐1.2)c 1.1 (0.7‐1.6)c .019d
Impedance (Ω) 612 (503‐707)c 999 (683‐1254)c <.001d
Position of LV lead tip in LAO view Posterior 34 (34%)a 4 (4%)a <.001b
Postero‐lateral 65 (65%)a 9 (9%)a
Lateral 1 (1%)a 43 (43%)a
Antero‐lateral 0 (0%)a 38 (38%)a
Anterior 0 (0%)a 6 (6%)a
Position of LV lead tip in RAO view Basal 90 (90%)a 25 (25%)a <.001b
Mid 9 (9%)a 75 (75%)a
Apical 1 (1%)a 0 (0%)a
a

Number of cases (percentage).

b

Chi‐square test.

c

Median (Q1‐Q3).

d

Mann‐Whitney U test.

3.2. Follow‐up

In the thoracoscopic group, 21 deaths occurred during a median follow‐up period of 24 months (IQR, 13‐39); in the transvenous group, 27 deaths occurred during a median follow‐up of 32 months (IQR, 24‐46). The rates of death due to any cause were comparable between the groups (Figure 1; log‐rank test, P = .193). The rates of death due to any cause and cardiovascular hospitalization were also comparable between the groups (Figure 1; log‐rank test, P = .949). Moreover, the risk of device‐related complications was similar between the groups. Figure 2 shows the survival from device‐related complications in both groups (log‐rank test, P = .783). In the thoracoscopic group, we recorded five infections (requiring system revision only in one patient), four pocket hematomas (not requiring system revision), one right ventricular lead dislodgement, one system failure, and one replacement of the LV lead. In the transvenous group, six patients had LV lead dislodgement, three had pocket erosion, two had right ventricular lead dislodgement, one had right atrial lead dislodgement, and one system failure.

Figure 1.

Figure 1

A, Kaplan‐Meier estimate of time to death due to any cause in the thoracoscopic and transvenous groups (log‐rank test, P = .193). B, Kaplan‐Meier estimate of time to death due to any cause or cardiovascular hospitalization in the thoracoscopic and transvenous groups (log‐rank test, P = .949)

Figure 2.

Figure 2

Kaplan‐Meier estimate of time to the first episode of device‐related complication in the thoracoscopic and transvenous groups (log‐rank test, P = .783)

In both groups, significant reverse remodeling of the LV was observed in terms of increased LV ejection fraction and reduced LV volumes (Table 3). Patients of both groups similarly improved in their functional status, as evidenced by a reduction of patients in NYHA class III‐IV at 1‐year follow‐up. The pacing parameters remained satisfactory in both groups (Table 3).

Table 3.

Comparison of clinical, echocardiographic, and electrical parameters at baseline and follow‐up in the thoracoscopic and transvenous groups

Impedance Thoracoscopic (100) Transvenous (100)
Basal F‐U P‐value Basal F‐U P‐value P‐value F‐U
NYHA III or IV 69 (69%)a 29 (29%)a <.001c 59 (59%)a 18 (18%)a <.001c .172c
LVEF (%) 30 (22‐33)d 36 (31‐43)d <.001b 28 (24‐33)d 40 (29‐48)d <.001b .550e
LVEDV (mL) 180 (140‐224)d 170 (115‐224)d .029b 167 (143‐226)d 152 (123‐183)d .010b .359e
LVESV (mL) 117 (105‐160)d 104 (64‐160)d .002b 124 (94‐167)d 88 (63‐122)d <.001b .307e
Pacing threshold (V at 0.5 ms) 0.8 (0.6‐1.2)d 1.6 (1.1‐2.1)d .002b 1.0 (0.6‐1.8)d 1.2 (0.6‐2.9)d .100b .326e
Impedance (Ω) 645 (483‐705)d 383 (344‐430)d <.001b 1161 (937‐1272)d 669 (486‐913)d .005b <.001e
a

Number of cases (percentage %).

b

Wilcoxon matched‐pairs signed rank test.

c

Chi‐square test.

d

Median (Q1‐Q3).

e

Mann‐Whitney U test.

4. DISCUSSION

In clinical practice, the first‐line approach to LV lead implantation during a CRT procedure is the transvenous epicardial approach. The final position of the LV lead depends on the anatomy of the CS, on the performance and stability of the pacing lead and on the absence of phrenic nerve stimulation. Despite all the available technologies, the rate of failed LV lead implantation remains substantial (5‐10%) in transvenous CRT.2 Endocardial LV pacing via the trans‐septal route is the only transvenous alternative to the standard technique.3 In a recent sub‐analysis from the Alternate Site Cardiac Resynchronization (ALSYNC) study, Biffi et al found that prior non‐responders to CRT were able to improve on LV endocardial pacing via the trans‐septal route.4 Many experiences have been published on LV pacing via trans‐septal approach,3, 5, 6, 7, 8, 9, 10, 11, 12, 13, 14 but no randomized studies have compared its efficacy with that of the epicardial approach.

There are several open‐chest techniques to implanting the LV pacing lead: median sternotomy, fully left thoracotomy, mini‐thoracotomy, video‐assisted thoracoscopy (VAT), and robotically assisted surgery.15 These approaches offer the advantages of direct visual control, with the possibility of choosing the lead‐tip position, less fluoroscopy use and the avoidance of intravenous contrast material, whereas the disadvantages are the need for general anesthesia, and the presence of epicardial fat and adhesions.

Two randomized studies have demonstrated the non‐inferiority of open‐chest access vs transvenous approach,16, 17 in addition to several other non‐randomized studies.16, 17, 18, 19, 20, 21, 22, 23, 24

The VAT technique causes less postoperative pain and requires smaller incisions. The lateral approach enables the target area of the LV (basal postero‐lateral) to be reached more easily than in the supine position previously suggested.19 Singh et al25 found an association between basal positioning of the LV lead and a lower risk of death and worsening heart failure. In addition, it has been demonstrated that pacing from the site of maximum electrical delay further contributes to improving patient outcomes,26 and current guidelines advocate targeting the regions of latest activation.27 As the VAT procedure overcomes the limitations of the CS anatomy, it allows easier access to the optimal pacing site, according to the values of pacing parameters and, if possible, to the degree of the electrical delay.

Previous studies have been published on the VAT technique.19, 20, 21, 24, 28, 29, 30, 31, 32, 33, 34, 35, 36 We showed that the VAT approach is safe and effective. We included both patients who had undergone unsuccessful de novo transvenous LV lead implantation and patients who had responded positively to CRT but had experienced CRT discontinuation due to lead dislodgement.

LV lead implantation was successful in all patients, with few operative complications. The VAT technique resulted in optimal basal postero‐lateral positioning of the electrode, with adequate values of the electrical parameters. Over the long term, the system showed a good safety profile when compared with the transvenous group, with limited complications and stability of the pacing parameters. All‐cause death and cardiovascular hospitalizations were similar. In both groups, there were significant improvements in clinical and echocardiographic parameters, although VAT patients were clinically worse, owing to their comorbidities. Further studies are required in order to determine whether VAT should be limited to patients in whom effective CRT is discontinued, as in the present study, or whether it can also be considered an option for therapy optimization in patients who do not initially improve on CRT.

4.1. Study limitations

The main limitation is its lack of randomization. The study group consisted of consecutive patients who were referred to our center for epicardial LV lead implantation. The number of patients who initially underwent an attempt of transvenous implantation at the referring centers is unknown. The control group consisted of consecutive patients who underwent successful transvenous CRT implantation in our center. As data on operative complications in this group were not available for comparison, we limited our comparison to long‐term safety and outcome.

5. CONCLUSIONS

In patients in whom transvenous LV lead implantation has previously failed, or in patients who have responded positively to CRT but have experienced CRT discontinuation, the VAT epicardial procedure is a safe and effective technique for LV lead implantation.

CONFLICT OF INTEREST

S. Valsecchi is a Boston Scientific employee. The other authors declare no potential conflict of interests.

Marini M, Branzoli S, Moggio P, et al. Epicardial left ventricular lead implantation in cardiac resynchronization therapy patients via a video‐assisted thoracoscopic technique: Long‐term outcome. Clin Cardiol. 2020;43:284–290. 10.1002/clc.23300

REFERENCES

  • 1. Droghetti A, Branzoli S, Moggio P, et al. Minimally invasive thoracoscopic technique for LV lead implantation in CRT patients. J Clin Exp Cardiol. 2018;9:3. [Google Scholar]
  • 2. Gras D, Bocker D, Lunati M, et al. Implantation of cardiac resynchronization therapy systems in the CARE‐HF trial: Procedural success rate and safety. Europace. 2007;9:516‐522. [DOI] [PubMed] [Google Scholar]
  • 3. Morgan JM, Biffi M, Geller L, Leclercq C, Ruffa F, et al. Alternate Site Cardiac ResYNChronation (ALSYNC): a prospective and multicentre study of the left ventricular endocardial pacing for cardiac resynchronization therapy. Eur Heart J. 2016;37:2118‐2127. [DOI] [PubMed] [Google Scholar]
  • 4. Biffi M, Defaye P, Jais P, et al. Benefits of left ventricular endocardial pacing comparing failed implants and prior non responders to conventional cardiac resynchronization therapy: a subanalysis from the ALSYNC study. Int J Cardiol. 2018;259:88‐93. [DOI] [PubMed] [Google Scholar]
  • 5. Leclerq F, Hager XF, Macia JC, Mariottini CJ, Pasquie JL, et al. Left ventricular lead insertion using a modified transseptal catheterization technique: a totally endocardial approach for permanent biventricular pacing in end‐stage heart failure. PACE. 1999;22:1570‐1575. [DOI] [PubMed] [Google Scholar]
  • 6. Jais P, Takahashi A, Garrigue S, Yamane T, et al. Mid‐term follow‐up of endocardial biventricular pacing. PACE. 2000;23:1744‐1747. [DOI] [PubMed] [Google Scholar]
  • 7. Garrigue S, Jais P, Espil G, Labeque J, Hocini M, et al. Comparison of chronic biventricular pacing between epicardial and endocardial left ventricular stimulation using Doppler tissue imaging in patients with heart failure. Am J Cardiol. 2001;88:858‐862. [DOI] [PubMed] [Google Scholar]
  • 8. Pasquiè JL, Massin F, Macia JC, Gervasoni R, et al. Long‐term follow‐up of biventricular pacing using a totally endocardial approach in patients with end‐stage cardiac failure. PACE. 2007;30:S31‐S33. [DOI] [PubMed] [Google Scholar]
  • 9. Van Gelder MB, Scheffer GM, Meijer A, Bracke AF. Transseptal endocardial left ventricular pacing: an alternative technique for coronary sinus lead placement in cardiac resynchronization therapy. Heart Rhythm. 2007;4:454‐460. [DOI] [PubMed] [Google Scholar]
  • 10. Morgan JM, Scott AP, Turner GN, Yue MA, et al. Targeted left ventricular endocardial pacing using a steerable introducing guide catheter and active fixation pacing lead. Europace. 2009;11:502‐506. [DOI] [PubMed] [Google Scholar]
  • 11. Gelder MB, Houthuizen P, Bracke AF. Transseptal left ventricular endocardial pacing: preliminary experience from a femoral approach with subclavian pull‐through. Europace. 2011;13:1454‐1458. [DOI] [PubMed] [Google Scholar]
  • 12. Kutyfa V, Merkely B, Szilagyi S, Zima E, et al. Usefulness of electroanatomical mapping during transseptal endocardial left ventricular lead implantation. Europace. 2012;14:599‐604. [DOI] [PubMed] [Google Scholar]
  • 13. Morina ‐Vazquez P, Roa‐Garrido J, Fernandez Gomez MJ, Vengas‐Gamero J, et al. Direct left ventricular endocardial pacing: an alternative when traditional resynchronization via coronary sinus is not feasible or effective. PACE. 2013;36:699‐706. [DOI] [PubMed] [Google Scholar]
  • 14. Neuhoff I, Szilagyi S, Molnar L, Osztheimer I, et al. Transseptal leftventricular endocardial pacing is an alternative technique in cardiac resynchronization therapy. One year experience in a high volume center. Rom J Intern Med. 2016;54:121‐127. [DOI] [PubMed] [Google Scholar]
  • 15. Mihalcz A, Kassai I, Geller L, Szili‐Török T. Alternative techniques for left ventricular pacing in cardiac resynchronization therapy. PACE. 2014;37:255‐261. [DOI] [PubMed] [Google Scholar]
  • 16. Doll N, Plorkowki C, Czesla M, Kallenbach M, et al. Epicardial versus transvenous left ventricular lead placement in patients receiving cardiac resynchronization therapy: results from a randomized prospective study. Thorac Cardiovasc Surg. 2008;56:256‐261. [DOI] [PubMed] [Google Scholar]
  • 17. Giraldi F, Cattadori G, Roberto M, et al. Long‐Term effectiveness of cardiac resynchronization therapy in heart failure patients with unfavorable cardiac veins anatomy. JACC. 2011;58:483‐490. [DOI] [PubMed] [Google Scholar]
  • 18. Mair H, Jansens JL, Lattouf MO, Reichart B, Daebritz S. Epicardial lead implantation tchniques for biventricolar pacing via left lateral mini‐thoracotomy, video‐assisted thoracoscopy, and robotic approach. Heart Surg Forum. 2003;6:412‐417. [PubMed] [Google Scholar]
  • 19. Navia JL, Atik FA, Grimm RA, et al. Minimally invasive left ventricular epicardial lead placement: surgical techniques for heart failure resynchronization therapy. Ann Thorac Surg. 2005;79:1536‐1544. [DOI] [PubMed] [Google Scholar]
  • 20. Corvera JS, Puskas JD, Thourani VH, Guyton RA, Lattouf OM. Minimally invasive surgical cardiac resychronization therapy: an intermediate‐term follow‐up study. Innovations. 2007;2:40‐47. [DOI] [PubMed] [Google Scholar]
  • 21. Pojar M, Vojacek J, Tauchman M, et al. Thoracoscopic left ventricular lead implantation. Acta Cardiol. 2011;66:797‐801. [DOI] [PubMed] [Google Scholar]
  • 22. Kassai I, Friedrich O, Ratnatunga C, Betts RT, Mihalcz A, Szili‐Torok T. Feasibility of percutaneous implantation of transapical endocardial left ventricular pacing electrode for cardiac resynchronization therapy. Europace. 2011;13:1653‐1657. [DOI] [PubMed] [Google Scholar]
  • 23. Mihalcz A, Kassai I, Kardos A, Foldesi C, Theuns D, Szili‐Torok T. Comparison of the efficacy of two surgical alternatives for cardiac resynchronization therapy: trans‐apical versus epicardial left ventricular pacing. PACE. 2012;35:124‐130. [DOI] [PubMed] [Google Scholar]
  • 24. Nesher N, Ganiel A, Paz Y, et al. Thoracoscopic epicardial lead implantation as an alternative to failed endovascular insertion for cardiac pacing and resynchronization therapy. Innovations. 2014;9:427‐431. [DOI] [PubMed] [Google Scholar]
  • 25. Singh JP, Klein HU, Huang DT, et al. Left ventricular lead position and clinical outcome in the multicenter automatic defibrillator implantation trial‐cardiac resynchronization therapy (MADIT‐CRT) trial. Circulation. 2011;123:1159‐1166. [DOI] [PubMed] [Google Scholar]
  • 26. Gold MR, Birgersdotter‐Green U, Singh JP, et al. The relationship between ventricular electrical delay and left ventricular remodelling with cardiac resynchronization therapy. Eur Heart J. 2011;32:2516‐2524. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27. Brignole M, Auricchio A, Baron‐Esquivias G, et al. 2013 ESC guidelines on cardiac pacing and cardiac resynchronization therapy: the task force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association (EHRA). Europace. 2013;15:1070‐1118. [DOI] [PubMed] [Google Scholar]
  • 28. Fernandez AL, Garcia‐Bengochea JB, Ledo R, et al. Minimally invasive surgical implantation of left ventricular epicardial leads from ventricular resynchronization using video‐assisted thoracoscopy. Rev Esp Cardiol. 2004;57:313‐319. [PubMed] [Google Scholar]
  • 29. Gabor S, Prenner G, Wasler A, Schweiger M, Tscheliessnigg KH, Smolle‐Juettner FM. A simplified technique for implantation of left ventricular epicardial leads for biventricular resynchronization using video‐assisted thoracoscopy. Eur J Cardiothorac Surg. 2005;28:797‐800. [DOI] [PubMed] [Google Scholar]
  • 30. Jutley RS, Waller DA, Loke I, et al. Video‐assisted thoracoscopic implantation of the left ventricular pacing lead for cardiac resynchronization therapy. Pacing Clin Electrophysiol. 2008;31:812‐818. [DOI] [PubMed] [Google Scholar]
  • 31. Zhang HB, Meng X, Han J, Li Y, Zeng YP, Zhang Y. Minimally invasive video‐assisted epicardial lead cardiac resynchronization therapy for the dilated cardiomyopathy heart failure cases. Innovations. 2010;5:345‐348. [DOI] [PubMed] [Google Scholar]
  • 32. Ailawadi G, LaPar DJ, Swenson BR, et al. Surgically placed left ventricular leads provide similar outcomes to percutaneous leads in patients with failed coronary sinus lead placement. Heart Rhythm. 2010;7:619‐625. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33. Papiashvilli M, Haitov Z, Fuchs T, Bar I. Left ventricular epicardial lead implantation for resynchronazation therapy using a video‐assisted thoracoscopic approach. Heart Lung Circ. 2011;20:220‐222. [DOI] [PubMed] [Google Scholar]
  • 34. Nelson KE, Bates MGD, Turley AJ, Linker NJ, Owens WA. Video‐assisted thoracoscopic left ventricular pacing in patients with and without previous sternotomy. Ann Thorac Surg. 2013;95:907‐913. [DOI] [PubMed] [Google Scholar]
  • 35. Polasek R, Skalsky I, Wichterle D, et al. High‐density epicardial activation mapping to optimize the site for video‐thoracoscopic left ventricular lead implant. J Cardiovasc Electrophysiol. 2014;25:882‐888. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36. Schouwenburg JJ, Klinkenberg TJ, Maass AH, Mariani MA. Video‐assisted thoracic placement of epicardial leads. J Card Surg. 2014;29:286‐289. [DOI] [PubMed] [Google Scholar]