Endpoint selection for noninferiority percutaneous coronary intervention trials: a methodological description

Matthias Waliszewski 1,2,, Mark Rosenberg 3, Harald Rittger 4, Viktor Breul 5, Florian Krackhardt 6
PMCID: PMC7074513  PMID: 32168991

Abstract

Background:

The objective of this review is to provide a practical update on endpoint selection for noninferiority (NI) studies in percutaneous coronary intervention studies.

Methods:

A PubMed search was conducted for predefined terms to explore the use of NI designs and intrapatient comparisons to determine their current importance. Sample size calculations for the most frequently used endpoints with NI hypotheses were done to increase statistical awareness.

Results:

Reported NI trials, with the most frequently chosen clinical endpoint of major adverse cardiac events (MACE), had NI margins ranging from 1.66% to 5.00%, resulting in patient populations of 400–1500 per treatment group. Clinical study endpoints comprising of MACE complemented with rates of bleeding complications and stent thrombosis (ST) are suggested to conduct a statistically and clinically meaningful NI trial. Study designs with surrogate endpoints amenable to intrapatient randomizations, are a very attractive option to reduce the number of necessary patients by about half. Comparative clinical endpoint studies with MACE and ST/bleeding rates to study a shortened dual antiplatelet therapy (DAPT) in coronary stent trials are feasible, whereas ST as the sole primary endpoint is not useful.

Conclusions:

Expanded composite clinical endpoints (MACE complemented by ST and bleeding rates and intrapatient randomization for selected surrogate endpoints) may be suitable tools to meet future needs in device approval, recertification and reimbursement.

Keywords: clinical endpoints, coronary interventions, surrogate endpoints, trial design

Introduction

Clinical studies are essential to demonstrate the safety and efficacy of drugs or medical devices. In these studies, the primary endpoints are the measures of success by which treatment groups can be compared. Based on these primary endpoints, the size of the study population will be calculated. To design a meaningful study with clinically relevant endpoints, properly powered with a sufficient number of patients, and with the potential to impact guideline recommendations can be a cumbersome task. Even experienced trialist are not protected from designing a trial that may be inclusive despite its large study population. Very recently, a large randomized controlled trial (RCT) with a patient recruitment of over 3000 patients was reported to use an incorrect noninferiority (NI) margin, thereby impacting the main conclusion of the trial.1 (SORT OUT IX trial, ClinicalTrials.gov identifier: NCT02623140). This gives us reason to believe that there is a paucity of advice for the clinical scientist planning NI trial designs.

Furthermore, given the upcoming European Medical Device Regulation (EU MDR, article 61),2 which will be effective in May 2020, a stronger focus on clinical evidence with proprietary data for medical devices will be enforced for all device manufacturers. Therefore, a comparison of clinical safety, efficacy and usability between new and predicate devices, or existing and competitor devices, based on well-designed clinical studies appear to be essential to gain or maintain regulatory access to the European market.

Owing to the fact that device-related event rates are typically in the single-digit percentage range, NI study designs have been popular. Byrne and Kastrati commented that large NI margins may be statistically sound but may not represent the clinical scenario.3 However, their call for increased resources to properly study and compare devices may be limited by small differences in event rates leading to very large patient populations. These, in turn, would require logistics that would facilitate rapid patient inclusion to avoid long recruitment windows, which may affect data quality or may even introduce bias into the data set.

Based on a previous review,4 the objective of this collaborative work is to provide guidance on sample size calculations for NI studies in the field of percutaneous coronary angioplasty. In particular, we will elaborate on the following topics which seem to be of current interest.

  1. Recommendations for acceptable NI margins in angiographic and clinical endpoint trials.

  2. Sample size calculations for the most prominent endpoints such as major adverse cardiac events (MACE) and late lumen loss (LLL) with commonly accepted NI margins.

  3. Statistical methods to reduce sample sizes or to use existing data sets.

Methods

Literature searches were conducted on PubMed/Medline and EMBASE with prespecified search protocols. Inputs for sample size calculations were obtained from recently published studies investigating drug-eluting stents (DES) or drug-coated balloons (DCB).

Sample size calculations

Study designs for clinical endpoints rely on binomial proportions and their difference between two independent groups. The corresponding test hypotheses are as follows4:

  • Ho: The event rate π1 in the treatment group is higher than, or equal to, the event rate π2 in the control group plus a NI margin Δ.

  • Ha: Event rate π1 in the treatment group is lower than π2 in the control group plus a NI margin Δ.

The basic principle of the sample size calculation is that the probability of a positive test for a given study sample (i.e. power) is greater if the study sample is larger. In fact, there is a mathematical relationship between expected binomial proportions π1 and π2, the NI margin Δ, the significance level alpha, the number of patients n, and the power. So when π1, π2, and Δ are known, fixing alpha to 5% and the power to, for example, 80%, we can calculate the minimum sample size n.

Likewise, test hypotheses for angiographic endpoints such as LLL can be formulated4:

  • Ho: LLL in the treatment group is higher than, or equal to, LLL in the control group plus a NI margin ΔLLL.

  • Ha: LLL in treatment group is lower than LLL in the control group plus a NI margin ΔLLL.

Assuming normally distributed variables, the number of patients are estimated based on the significance level alpha, desired power, standard deviations (SD), and the NI limit. For all sample size calculations, nQuery Advisor version 7.0 (Cork, Ireland) was used.

Literature search

The search terms listed in Table 1 were used to explore current interests as outlined in the Introduction.

Table 1.

Search terms for endpoints in coronary interventions with a focus on coronary stents (until August 1, 2019).

No. Keywords
#1 Noninferiority margin drug-eluting stent
#2 Intrapatient comparison AND drug-eluting stent
#3 Intrapatient comparison AND drug coated balloon

Results

The prespecified literature searches (search #1) were conducted and can be summarized in Figure 1. Literature searches #2 and #3 revealed a total of two references. The results of the literature searches can be divided in NI for clinical endpoints (Table 2) and NI for angiographic endpoints (Table 3).

Figure 1.

Figure 1.

PRISMA flow chart for ‘non inferiority margin drug-eluting stent’.

Table 2.

NI margins of clinical endpoint trials.

No. Clinical endpoint NI margin and comparators Primary endpoints
01 (Kehdi)5 1.66%.
Single antiplatelet therapy from 6 to 12 months
versus double antiplatelet therapy from 6 to 12 months
Composite of all cause mortality, any MI, any revascularisation, stroke, and thrombolysis in MI major bleeding at 18 months
02 (Hahn)6 2.00%.
Dual antiplatelet therapy 6 months versus 12 months
Composite of all-cause death, MI, or stroke at 18 months
03 (Christiansen)7 2.00%.
BES versus SES
Composite of safety (cardiac death, MI, definite stent thrombosis) and efficacy (TVR) at 9 months
04 (Raungaard)8 2.50%.
ZES versus BES
Composite of safety (cardiac death and MI not clearly attributable to a nontarget lesion) and efficacy (TVR) at 12 months
05 (von Birgelen)9 2.50%.
ZES versus SES
Target vessel failure at 1 year, a composite of cardiac death, target-vessel-related MI, and TVR
06 (Zocca)10 2.70%.
Clopidogrel versus ticagrelor
NACCE: all-cause death, any MI, stroke or major bleeding
07 (Stone)11 2.90%
BVS versus EES
Target lesion failure (cardiac death, target vessel MI, or ischaemia-driven TLR) at 30 days
08 (Turco)12 3.00%.
PES versus PES of different stent designs
9-month TVR
09 (von Birgelen)13 3.50%.
SES versus EES
Composite of safety (cardiac death or target vessel-related MI) and efficacy (TVR) at 12 months
10 (Lam)14 3.50%.
SES versus EES
Composite endpoint target vessel failure at 1 year, consisting of cardiac death, target vessel-related MI, or clinically driven TVR
11 (Pilgrim)15 3.50%.
SES versus EES
Target lesion failure, was a composite of cardiac death, target vessel MI, and clinically indicated TLR at 12 months
12 (Landsky)16 3.50%.
SES versus EES
Target lesion failure at 12 months, a composite of cardiac death, target vessel MI, or ischaemia-driven TLR
13 (von Birgelen)17 3.60%.
ZES versus EES
Target-vessel failure as composite of safety (cardiac death or target-vessel-related MI) and efficacy (TVR) at 12 months
14 (Kandzari)18 3.85%.
SES versus EES
12-month target lesion failure
15 (Zaman)19 4.00%.
SES versus EES
Composite endpoint-cardiac death, target-vessel MI, or clinically indicated TLR-between groups at 12 months
16 (Jeger)20 4.00%.
DCB versus EES
MACE, i.e. cardiac death, nonfatal MI, and TVR after 12 months.
17 (de Winter)21 4.00%.
SES versus EES
DOCE, i.e., cardiac death, target-vessel MI, or clinically indicated TLR, between the groups at 12 months
18 (Chang)22 4.50%.
BVS versus EES
TLF, a device-oriented composite endpoint (cardiac death, target vessel MI and clinically indicated TLR)
19 (Hahn)23 5.00%.
Aggressive versus conservative angiographic guidance
Composite of all-cause death, MI, or stroke at 18 months

BES, biolimus-eluting stent; BVS, bioresorbable scaffold; DCB, drug coated balloon; DOCE, device-oriented composite endpoint; EES, everolimus-eluting stent; MI, myocardial infarction; NI, noninferiority; PES, paclitaxel-eluting stent; SES, sirolimus-eluting stent; TLF, target lesion failure; TLR, target lesion revascularisation; TVR, target vessel revascularization; ZES, zotarolimus-eluting stent.

Table 3.

NI margins of surrogate endpoint trials.

No Surrogate endpoint NI margin and study groups Primary endpoints
01 (Serruys)24 ΔLLL = 0.14 mm.
BVS versus EES
Angiographic vasomotor reactivity after administration of intracoronary nitrate at 3 years
02 (Han)25 ΔLLL = 0.15 mm.
EES versus EES of different coating technologies
In-stent late loss at 9 months
03 (Mehilli)26 ΔLLL = 0.16 mm.
PES versus SES
In-stent late luminal loss at 6 months
04 (Franzone)27 ΔLLL = 0.17 mm.
EES versus SES
In-stent late lumen loss at 9 months
05 (Fajadet)28 ΔLLL = 0.20 mm.
SES versus PES
In-stent late lumen loss
06 (Costa)29 ΔLLL = 0.24 mm.
BES versus PES
In-stent late lumen loss
07 (Park)30 ΔBR = 15%.
ZES versus SES
In-segment binary restenosis rate
08 (Byrne)31 ΔDS = 5.00%.
BVS versus EES
Diameter stenosis at follow-up angiography at 6–8 months
09 (Byrne)32 ΔDS = 7.00%.
DCB versus PES
Diameter stenosis at follow-up angiography at 6–8 months
10 (Elghamaz)33 ΔFFR = 0.05 and SD of 0.09.
Intracoronary adenosine versus intravenous adenosine
FFR, intra-interventional
11 (Oi)34 ΔFFR = 0.03.
Nicorandil versus adenosine triphosphate
Intrapatient difference in FFR between drug groups
12 (Götberg)35 ΔFFR = 0.032.
FFR versus iFR
Composite of death, nonfatal MI, unplanned revascularization
13 (Davies)36 ΔFFR = 0.034.
FFR versus iFR
Composite of death, nonfatal MI, unplanned revascularization at one year

BR, binary restenosis rate; BVS, bioresorbable scaffold; DCB, drug coated balloon; DS, diameter stenosis; EES, everolimus-eluting stent; FFR, fractional flow reserve; iFR, instantaneous wave free ratio; LLL, late lumen loss; MI, myocardial infarction; PES, paclitaxel-eluting stent; SES, sirolimus-eluting stent; ZES, zotarolimus-eluting stent.

NI trials with clinical endpoints

A total of 29 recently published trials were available for analysis, with 20 references reporting clinical endpoint NI margins ranging from 1.66% to 5.00% (Table 2). The lowest NI margin of 1.66% was reported by Kedhi and colleagues.5 They investigated different antiplatelet therapy (DAPT) strategies after event-free survival at 6 months with a composite endpoint of all cause mortality, myocardial infarction (MI), any revascularization including thrombolysis, stroke, and major bleeding. Their treatment group consisted of 432 (single antiplatelet therapy from 6 to 12 months) and 438 (DAPT from 6 to 12 months) patients.

The recently published BASKET SMALL II trial investigated DCB angioplasty and DES in small vessel de novo lesions.20 The sample size calculations in this study were based on a NI margin of 4%, a power of 90%, and expected MACE rates of 7% in the DCB and 10% in the DES treatment groups. Lansky and coworkers compared the target vessel failure (TVF) rates in an all-comers population receiving either everolimus eluting stents (EES) or a low dose sirolimus-eluting stent (SES).16 Their NI margin was set at 3.5%, assuming a 7% target vessel revascularization (TVR) rate in the control group. A NI rate of 4% and event rates of 8.3% in both treatment arms were used by de Winter and coworkers to study different polymer and drug coatings.21

In a prospective RCT comparing zotarolimus-eluting stents (ZES) with SES,9 the NI margin was set at 2.5% while assuming TVF rates at 12 months of 6.0%.

As Byrne and Kastrati pointed out,3 NI margins that are too large have limited value for clinical ramifications, and, therefore, future treatment recommendations. We ventured into sample size calculations with a 10% clinical event rate in the control group, various event rates in the treatment group, and a range of NI margins. Figure 2 details the patient group sizes with typical MACE/TVF rates and a range of NI margins. For a typical NI RCT, a minimal treatment group size starts at around 500 patients. A range of 250–1500 patients per group can therefore be expected (Figure 2).

Figure 2.

Figure 2.

Number of patients per treatment group for various NI trial designs with MACE as primary endpoints for different expected MACE rates in the treatment group (top panel) and different NI margins (bottom panel).

MACE, major adverse cardiac events; NI, noninferiority.

NI trials with surrogate endpoints

Based on a previous review,4 we would like to briefly present the most commonly used surrogate endpoints such as LLL, optical coherence tomography (OCT) and fractional flow reserve (FFR)/instantaneous wave-free ratios (iFR). Moreover, the concept of intrapatient randomization to eliminate patient-level bias appears to be of importance.

Late Lumen Loss

NI margins for LLL were reportedly used from 0.14 mm to 0.24 mm.2429 Based on these previously published NI margins with the primary endpoint LLL, a number of common sample sizes were calculated given the expected SD, which may range from 0.40 mm to 0.70 mm.37

Assuming a frequently reported SD of 0.45 mm, and NI margins of 0.05–0.20 mm, a minimum of 108 patients per group would have to be included given a 90% power (Figure 3). The patient population size may increase to the 1200–1700 range with a NI margin of 0.05 mm.

Figure 3.

Figure 3.

Number of patients per treatment group for various NI trial designs with late lumen loss as the primary endpoint.

NI, noninferiority.

Intrapatient randomization

There were two reports on intrapatient randomization. It was hinted that one way to increase statistical power while keeping patient numbers moderate is the use of intrapatient randomization. Intrapatient comparisons of angiographic changes in treated and untreated lesions have been previously conducted by Kleber and coworkers.38 They studied lumen enlargement after DCB angioplasty, and compared the corresponding mean lumen diameters with those in untreated nontarget vessels. This methodology was recently applied for comparative DES studies in the FRIENDLY OCT trial (ClinicalTrials.gov Identifier: NCT02785237), in which patients with two distinct lesions received both study devices, thus patient risk factors were identical in both device groups.

The use of paired t tests to test for differences was investigated and illustrated in Figure 4. With LLL SD in the 0.40–0.45 mm range, LLL means of 0.200 mm (treatment group) and 0.250 mm to 0.325 mm in the control group, the number of patients can be reduced by half.

Figure 4.

Figure 4.

Number of patients per treatment group for unpaired and paired t-tests with late lumen loss (LLL) as the primary endpoint. Assumptions: LLL = 0.200 mm in the treatment group, various LLL estimates in the control group (0.250–0.400 mm) and LLL SD of 0.400 mm in both groups with a power of 80%.

New surrogate endpoints

Without an exhaustive literature search, it seems that OCT- and FFR-derived endpoints are becoming more popular and are worth mentioning. Since OCT endpoints are not the focus of this review, only a brief excursion is intended to illuminate this surrogate endpoint which is increasingly being used.3943 Three different criteria for a ‘covered stent strut’, namely >0 µm, >10 µm and >20 µm (Table 4) have been reported in the literature.

Table 4.

Stent strut coverage criteria.

Strut coverage criterion Percentage of covered struts Follow-up time Reference Devices
>0 µm 95.2 ± 5.2% 3 months DISCOVERY 1 TO 339 BP-SES
(Ultimaster®, Terumo)
90.4% (ticagrelor) ;
88.3% (clopidogrel)
3 months Kim40 BP-SES
87.1% (3 months);
98.6% (6 months)
3 months and 6 months Nano+ OCT41 PF-SES (Nano+®, LEPU Medical)
>10 µm 99.6% 3 months Yano42 PP-EES
(Xience®, Abbott Vascular)
>20 µm 84.2% (BP-SES);
84.6% (PP-EES)
6 months Koppara43 BP-SES vs PP-EES

BP-SES, bioresorbable polymer sirolimus eluting stent; PP-EES, permanent polymer everolimus eluting stent; PF-SES, polymer-free sirolimus eluting stent.

With a range of 85–95% of covered struts at follow up, a 5% SD, sample size calculations can be made accordingly (Figure 5).

Figure 5.

Figure 5.

Number of patients per treatment group to test for differences in stent coverage rate using OCT with 85% in the control group and 87.5–95.0% in the treatment group with 6.0% SD in both groups.

OCT, optical coherence tomography; SD, standard deviation.

Chevalier and colleagues, Kim and colleagues, and Suwannasom and colleagues used the criterion of >0 µm strut coverage with a 3-month follow-up.3941 Yano and colleagues proposed a >10 µm cut-off value,42 whereas Koppara and colleagues utilized the criterion of >20 µm.43 Figure 5 reveals that, for strut coverage rates of 85% in the control group and, for example, 90% in the treatment group, 32 patients are necessary to detect a differences in stent coverage rates (power = 90%, SD = 6%).

FFR measurements translate into pressure drops across lesions. The higher the degree of stenosis, the higher the pressure drop. While this relationship is nonlinear, it serves as an accepted diagnostic tool to determine the severity of the lesion. Needless to say, the amount of myocardium affected by a stenosed segment is of cardinal importance to determine the need for revascularization. Shin and coworkers investigated FFR at 9 months after DES implantation or DCB angioplasty.44 They used an unpaired t test design to demonstrate that the FFR values between treatment groups were not different (0.86 ± 0.06 versus 0.83 ± 0.08, p = 0.105). Alternative study designs such as a loss in FFR analogous to a loss in lumen diameter beyond their statistical theory were not reported in clinical study reports so far.4 Recently, iFR have been used,41 which do not require intracoronary administration of pharmacological vasodilators.

Table 3 lists NI margins for FFR and iFR trials. The range from 0.030 to 0.050 was used to conduct sample size calculations for this hemodynamic endpoint.3336

Very common SD for iFR and FFR are 0.09–0.10 with clinically relevant cut-off values depending on the type of FFR.33,45 Based on the aforementioned SD, which are more on the conservative side, and NI margins, sample sizes range between 64 and 235 per group (Figure 6). Comprehensive reviews on iFR and FR were recently provided by Baumann and coworkers.46,47

Figure 6.

Figure 6.

Number of patients per treatment group to test for differences in iFR and FFR with a common SD of 0.05 and various NI margins of 0.020, 0.025, 0.030, 0.035, and 0.040 and power of 80% and 90%.

FFR, fractional flow reserve; iFR, instantaneous wave-free ratios; NI, noninferiority; SD, standard deviation.

New clinical endpoints

Despite the fact that, to the best of the authors’ knowledge, there have been no reports with ST rates as primary endpoints, this potential endpoint is worth mentioning. There is sufficient interest for some novel DES to reduce the DAPT. If one ventures to delve into sample size calculations with the primary endpoint rate of ST with a comparator rate of 0.5% and expected rates in the treatment group between 0.4% and 0.2%, patients will be typically in the ten thousands to single digit thousands per treatment group (Figure 7). In this scenario, a composite endpoint including TVR and nonfatal MI, potentially complemented with rates of bleeding complications, may be the only feasible option to reduce the sample size to manageable numbers. Christiansen and colleagues studied the safety and efficacy of biolimus-eluting stents (BES) versus sirolimus-eluting stents (SES) by adding the rates of ST to their MACE endpoint.7 Compared with other designs for device trials, their NI margin was only 2.0% with patient group sizes of 1208 and 1193, respectively.

Figure 7.

Figure 7.

Number of patients per treatment group for various trial designs to test for differences in the rate of ST with 0.5% in the control group and rates from 0.20% to 0.40% in the treatment group.

ST, stent thrombosis.

Discussion

An important aspect before feasible study designs are discussed, one ought to critically reflect whether differences in event rates or morphometric variables translate to clinical benefits for the patient. If we measure LLL in hundredths of millimeters with SD in the range or even higher than their respective mean values, what is the measurable benefit for the patient? We strongly believe that there needs to be a balance between statistical and clinical significance.

NI designs

The majority of clinical NI study designs use the classical MACE endpoint in most device trials.4,7,9,1018 The concept of MACE is complemented by bleeding and stroke rates whenever treatment groups involve different DAPT regimens (drug regimens or durations).56,10 With the exception of Hahn and coworkers,23 who investigated different procedures for revascularizations by choosing the combination of MACE and ST, there seem to be two clusters of study endpoints, namely MACE for device trials, and the combination of MACE, bleeding, and ST rates for DAPT studies. This appears very reasonable given the challenge to counterbalance the risks of ischemic versus bleeding events. Our assessments reveal that, with the addition of bleeding and ST rates to MACE rates, NI margins can be reduced to the 2.0% range56,10 with acceptable patient group sizes of 432 patients5, 1355 patients,6 and 1009 patients.10

An aspect of paramount importance is justification of the NI margin, which has been adequately addressed in the excellent review papers by Head and colleagues and Rehal and colleagues.48,49 While our sample size estimates (Figure 2) are based on a clinical event rate of 10% with NI margins from 1.5% to 5.0%, this may be interpreted as overly simplistic, since most clinical event rates for a composite endpoint such as MACE at 12 months may be as low as 4.5% with a NI margin of 2.5%.9 The relevant United States Food and Drug Administration (FDA) guideline cautions the trialist to choose a NI margin greater than the actual treatment effect.50 This will lead to a false conclusion that the new treatment is effective as compared with the control group.

Intrapatient randomization

As previously reported, intrapatient randomizations enable the clinical trial designer to eliminate all bias-introducing factors relative to patient demographics. However, in terms of lesion morphological factors, there may still be some differences. The use of the paired t test may provide some additional power, while the number of patients could be greatly reduced for proof of concept studies.4 Overall, as can be seen in Figure 4, the patient population could be reduced to half if the study design permits two treatments per patient, for example two different stents in one patient.

Surrogate endpoints

A simplified overview of commonly used endpoints is given in Figure 8. This review did not focus on hemodynamic endpoints such as FFR and iFR. Nevertheless, we used common estimates to provide a range of study population sizes. As mentioned earlier, these are pressure gradient measurements across lesions that can either guide clinical decision making or could be even used as an endpoint measure.35 Shin and coworkers used FFR as a surrogate endpoint at 9 months after DES implantation or DCB angioplasty.44 They found FFR measurements after predilatation with an uncoated balloon useful as a decision facilitator. Shin and colleagues also concluded that FFR may be an option to avoid DES stenting in case long-term antiplatelet therapy is not well tolerated by the patient.

Figure 8.

Figure 8.

Overview of commonly used endpoints in coronary intervention trials and patient group sizes per treatment group.

BARC, Bleeding Academic Research Consortium; FFR, fractional flow reserve; iFR, instantaneous wave-free ration; LLL, late lumen loss; MACE, major adverse cardiac events; MI, myocardial infarction; OCT, optical coherence tomography; ST, stent thrombosis; TLR, target lesion revascularization; TVF, target vessel failure; TVR, target vessel revascularization.

OCT endpoints may be an attractive option for morphometric measurements of stent healing. From a statistical viewpoint, the >20 µm criterion appear to be advantageous since the absolute stent coverage rates are somewhat lower as compared with the other criteria. Hence, the total number of patients will be somewhat lower given a similar SD of 6%. This remains debatable if true tissue coverage and the precision of OCT are in agreement with stent coverage of less than 20 µm.

Conclusion

Acceptably low NI margins can be used with expanded composite clinical endpoints, which are MACE complemented by ST and bleeding rates. In addition to expanded clinical endpoints, intrapatient randomization may be a suitable tool to meet future needs in device approval, recertification, and reimbursement.

Study limitations

The presented sample size estimates provide the junior trialist with a first ‘stepping stone’. They do not replace calculations for a given study design. Moreover, study endpoints, that is long-term safety versus proof of concept, are not interchangeable. It was not the intention to delve into all aspects of NI trial designs, and critical aspects such as missing data, type I error rates, and sensitivity analyses were not within the scope of this review.4850

Footnotes

Funding: The authors received no financial support for the research, authorship, and/or publication of this article.

Conflict of interest statement: The authors declare that there is no conflict of interest.

Ethics approval: Not applicable.

ORCID iD: Matthias Waliszewski Inline graphic https://orcid.org/0000-0002-8689-4505

Contributor Information

Matthias Waliszewski, B. Braun Melsungen AG, Medical Scientific Affairs, Sieversufer 8, Berlin, 12359, Germany; Department of Internal Medicine and Cardiology, Charité – Universitätsmedizin Berlin, Campus Virchow, Berlin, Germany.

Mark Rosenberg, Klinikum Aschaffenburg-Alzenau, Medizinische Klinik 1, Aschaffenburg, Germany.

Harald Rittger, Medizinische Klinik 1, Klinikum Fürth, Fürth, Germany.

Viktor Breul, Medical Scientific Affairs, Aesculap AG, Tuttlingen, Germany.

Florian Krackhardt, Department of Internal Medicine and Cardiology, Charité – Universitätsmedizin Berlin, Campus Virchow, Berlin, Germany.

References

  • 1. Jensen LO, Maeng M, Raungaard B, et al. Comparison of the polymer-free biolimus-coated BioFreedom stent with the thin-strut biodegradable polymer sirolimus-eluting Orsiro stent in an all-comers population treated with percutaneous coronary intervention: rationale and design of the randomized SORT OUT IX trial. Am Heart J 2019; 213: 1–7. [DOI] [PubMed] [Google Scholar]
  • 2. EUR-Lex. Official Journal of the European Union, https://eur-lex.europa.eu/legal-content/EN/TXT/?uri=OJ:L:2017:117:TOC (accessed 11 December 2018).
  • 3. Byrne RA, Kastrati A. Drug-eluting stent trials: too much non-inferiority, too little progress? Lancet 2014; 383: 386–388. [DOI] [PubMed] [Google Scholar]
  • 4. Waliszewski M, Rittger H. Surrogate and clinical endpoints in interventional cardiology: are statistics the brakes? Ther Adv Cardiovasc Dis 2016; 10: 314–326. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. Kedhi E, Fabris E, van der Ent M, et al. Six months versus 12 months dual antiplatelet therapy after drug-eluting stent implantation in ST-elevation myocardial infarction (DAPT-STEMI): randomised, multicentre, non-inferiority trial. BMJ 2018; 363: k3793. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6. Hahn JY, Song YB, Oh JH, et al. ; SMART-DATE Investigators. 6-month versus 12-month or longer dual antiplatelet therapy after percutaneous coronary intervention in patients with acute coronary syndrome (SMART-DATE): a randomised, open-label, non-inferiority trial. Lancet. 2018; 391: 1274–1284. [DOI] [PubMed] [Google Scholar]
  • 7. Christiansen EH, Jensen LO, Thayssen P, et al. ; Scandinavian Organization for Randomized Trials with Clinical Outcome (SORT OUT) V investigators. Biolimus-eluting biodegradable polymer-coated stent versus durable polymer-coated sirolimus-eluting stent in unselected patients receiving percutaneous coronary intervention (SORT OUT V): a randomised non-inferiority trial. Lancet 2013; 381: 661–669. [DOI] [PubMed] [Google Scholar]
  • 8. Raungaard B, Jensen LO, Tilsted HH, et al. ; Scandinavian Organization for Randomized Trials with Clinical Outcome (SORT OUT). Zotarolimus-eluting durable-polymer-coated stent versus a biolimus-eluting biodegradable-polymer-coated stent in unselected patients undergoing percutaneous coronary intervention (SORT OUT VI): a randomised non-inferioritytrial. Lancet. 2015; 385: 1527–1535. [DOI] [PubMed] [Google Scholar]
  • 9. von Birgelen C, Zocca P, Buiten RA, et al. Thin composite wire strut, durable polymer-coated (Resolute Onyx) versus ultrathin cobalt-chromium strut, bioresorbable polymer-coated (Orsiro) drug-eluting stents in allcomers with coronary artery disease (BIONYX): an international, single-blind, randomised non-inferiority trial. Lancet 2018; 392: 1235–1245. [DOI] [PubMed] [Google Scholar]
  • 10. Zocca P, van der Heijden LC, Kok MM, et al. Clopidogrel or ticagrelor in acute coronary syndrome patients treated with newer-generation drug-eluting stents: CHANGE DAPT. EuroIntervention 2017; 13: 1168–1176. [DOI] [PubMed] [Google Scholar]
  • 11. Stone GW, Ellis SG, Gori T, et al. ; ABSORB IV Investigators. Blinded outcomes and angina assessment of coronary bioresorbable scaffolds: 30-day and 1-year results from the ABSORB IV randomised trial. Lancet 2018; 392: 1530–1540. [DOI] [PubMed] [Google Scholar]
  • 12. Turco MA, Ormiston JA, Popma JJ, et al. Polymer-based, paclitaxel-eluting TAXUS Liberté stent in de novo lesions: the pivotal TAXUS ATLAS trial. J Am Coll Cardiol. 2007; 49: 1676–1683. [DOI] [PubMed] [Google Scholar]
  • 13. von Birgelen C, Kok MM, van der Heijden LC, et al. Very thin strut biodegradable polymer everolimus-eluting and sirolimus-eluting stents versus durable polymer zotarolimus-eluting stents in allcomers with coronary artery disease (BIO-RESORT): a three-arm, randomised, non-inferiority trial. Lancet 2016; 388: 2607–2617. [DOI] [PubMed] [Google Scholar]
  • 14. Lam MK, Sen H, Tandjung K, et al. Comparison of 3 biodegradable polymer and durable polymer-based drug-eluting stents in all-comers (BIO-RESORT): rationale and study design of the randomized TWENTE III multicenter trial. Am Heart J 2014; 167: 445–451. [DOI] [PubMed] [Google Scholar]
  • 15. Pilgrim T, Heg D, Roffi M, et al. Ultrathin strut biodegradable polymer sirolimus-eluting stent versus durable polymer everolimus-eluting stent for percutaneous coronary revascularisation (BIOSCIENCE): a randomised, single-blind, non-inferiority trial. Lancet 2014; 384: 2111–2122. [DOI] [PubMed] [Google Scholar]
  • 16. Lansky A, Wijns W, Xu B, et al. ; TARGEt all Comers Investigators. Targeted therapy with a localised abluminal groove, low-dose sirolimus-eluting, biodegradable polymer coronary stent (TARGEt all Comers): a multicentre, open-label, randomised non-inferiority trial. Lancet 2018; 392: 1117–1126. [DOI] [PubMed] [Google Scholar]
  • 17. von Birgelen C, Sen H, Lam MK, et al. Third-generation zotarolimus-eluting and everolimus-eluting stents in all-comer patients requiring a percutaneous coronary intervention (DUTCH PEERS): a randomised, single-blind, multicentre, non-inferiority trial. Lancet 2014; 383: 413–423. [DOI] [PubMed] [Google Scholar]
  • 18. Kandzari DE, Mauri L, Koolen JJ, et al. ; BIOFLOW V Investigators. Ultrathin, bioresorbable polymer sirolimus-eluting stents versus thin, durable polymer everolimus-eluting stents in patients undergoing coronary revascularisation (BIOFLOW V): a randomised trial. Lancet 2017; 390: 1843–1852. [DOI] [PubMed] [Google Scholar]
  • 19. Zaman A, de Winter RJ, Kogame N, et al. ; TALENT trial investigators. Safety and efficacy of a sirolimus-eluting coronary stent with ultra-thin strut for treatment of atherosclerotic lesions (TALENT): a prospective multicentre randomised controlled trial. Lancet. 2019; 393: 987–997. [DOI] [PubMed] [Google Scholar]
  • 20. Jeger RV, Farah A, Ohlow MA, et al. ; BASKET-SMALL 2 Investigators. Drug-coated balloons for small coronary artery disease (BASKET-SMALL 2): an open-label randomised non-inferiority trial. Lancet 2018; 392: 849–856. [DOI] [PubMed] [Google Scholar]
  • 21. de Winter RJ, Katagiri Y, Asano T, et al. A sirolimus-eluting bioabsorbable polymer-coated stent (MiStent) versus an everolimus-eluting durable polymer stent (Xience) after percutaneous coronary intervention (DESSOLVE III): a randomised, single-blind, multicentre, non-inferiority, phase 3 trial. Lancet 2018; 391: 431–440. [DOI] [PubMed] [Google Scholar]
  • 22. Chang CC, Onuma Y, Achenbach S, et al. ; COMPARE ABSORB trial investigators. Absorb bioresorbable scaffold versus xience metallic stent for prevention of restenosis following percutaneous coronary intervention in patients at high risk of restenosis: rationale and design of the COMPARE ABSORB trial. Cardiovasc Revasc Med 2019; 20: 577–582. [DOI] [PubMed] [Google Scholar]
  • 23. Hahn JY, Song YB, Oh JH, et al. ; SMART-DATE investigators. 6-month versus 12-month or longer dual antiplatelet therapy after percutaneous coronary intervention in patients with acute coronary syndrome (SMART-DATE): a randomised, open-label, non-inferiority trial. Lancet 2018; 391: 1274–1284. [DOI] [PubMed] [Google Scholar]
  • 24. Serruys PW, Chevalier B, Sotomi Y, et al. Comparison of an everolimus-eluting bioresorbable scaffold with an everolimus-eluting metallic stent for the treatment of coronary artery stenosis (ABSORB II): a 3 year, randomised, controlled, single-blind, multicentre clinical trial. Lancet 2016; 388: 2479–2491. [DOI] [PubMed] [Google Scholar]
  • 25. Han Y, Liu H, Yang Y, et al. A randomised comparison of biodegradable polymer- and permanent polymer-coated platinum-chromium everolimus-eluting coronary stents in China: the EVOLVE China study. EuroIntervention 2017; 13: 1210–1217. [DOI] [PubMed] [Google Scholar]
  • 26. Mehilli J, Dibra A, Kastrati A, et al. ; Intracoronary Drug-Eluting Stenting to Abrogate Restenosis in Small Arteries (ISAR-SMART 3) Study Investigators. Randomized trial of paclitaxel- and sirolimus-eluting stents in small coronary vessels. Eur Heart J 2006; 27: 260–266. [DOI] [PubMed] [Google Scholar]
  • 27. Franzone A, Zaugg S, Piccolo R, et al. A randomized multicenter trial comparing the XIENCE everolimus eluting stent with the CYPHER sirolimus elutingstent in the treatment of female patients with de novo coronary artery lesions: the SPIRIT WOMEN study. PLoS One 2017; 12: e0182632. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28. Fajadet J, Abizaid A, Ormiston J, et al. Six-month follow-up of RES-ELUTION NEVO-I, a randomized multi-center comparison of the nevo reservoir-based sirolimus-eluting stent with the TAXUS liberte paclitaxel-eluting stent. Euro Heart J 2009; 30(Suppl. 1): 461. [Google Scholar]
  • 29. Costa RA, Abizaid A, Mehran R, et al. De novo coronary lesions treated with the novel polymer-free Biolimus-A9 coated stents: 12-Month angiographic results from the prospective, randomized, multicenter BioFreedom clinical trial. JACC Cardiovasc Interv. 2016; 9(1): 51–64. [DOI] [PubMed] [Google Scholar]
  • 30. Park HJ, Kim HY, Lee JM, et al. Randomized comparison of the efficacy and safety of zotarolimus-eluting stents vs. sirolimus-eluting stents for percutaneous coronary intervention in chronic total occlusion–CAtholic Total Occlusion Study (CATOS) trial. Circ J. 2012; 76: 868–875. [DOI] [PubMed] [Google Scholar]
  • 31. Byrne RA, Alfonso F, Schneider S, et al. Prospective, randomized trial of bioresorbable scaffolds vs. everolimus-eluting stents in patients undergoing coronary stenting for myocardial infarction: the Intracoronary Scaffold Assessment a Randomized evaluation of Absorb in Myocardial Infarction (ISAR-Absorb MI) trial. Eur Heart J 2019; 40: 167–176. [DOI] [PubMed] [Google Scholar]
  • 32. Byrne RA, Neumann FJ, Mehilli J, et al. ; ISAR-DESIRE 3 investigators. Paclitaxel-eluting balloons, paclitaxel-eluting stents, and balloon angioplasty in patients with restenosis after implantation of a drug-eluting stent (ISAR-DESIRE 3): a randomised, open-label trial. Lancet. 2013; 381: 461–467. [DOI] [PubMed] [Google Scholar]
  • 33. Elghamaz A, Myat A, de Belder A, et al. A continuous intracoronary versus a standard intravenous infusion of adenosine for fractional flow reserve assessment - the HYPEREMIC trial. EuroIntervention. Epub ahead of print 9 July 2019. DOI: 10.4244/EIJ-D-18-01067. [DOI] [PubMed] [Google Scholar]
  • 34. Oi M, Toyofuku M, Matsumura Y, et al. Utility of nicorandil for the measurement of coronary fractional flow reserve. Cardiovasc Interv Ther 2014; 29: 24–30. [DOI] [PubMed] [Google Scholar]
  • 35. Götberg M, Christiansen EH, Gudmundsdottir IJ, et al. iFR-SWEDEHEART Investigators. Instantaneous wave-free ratio versus fractional flow reserve to guide PCI. N Engl J Med 2017; 376: 1813–1823. [DOI] [PubMed] [Google Scholar]
  • 36. Davies JE, Sen S, Dehbi HM, et al. Use of the instantaneous wave-free ratio or fractional flow reserve in PCI. N Engl J Med 2017; 376: 1824–1834. [DOI] [PubMed] [Google Scholar]
  • 37. Jensen CJ, Richardt G, Tölg R, et al. Angiographic and clinical performance of a paclitaxel-coated balloon compared to a second-generation sirolimus-eluting stent in patients with in-stent restenosis: the BIOLUX randomised controlled trial. EuroIntervention. 2018; 14: 1096–1103. [DOI] [PubMed] [Google Scholar]
  • 38. Kleber FX, Schulz A, Waliszewski M, et al. Local paclitaxel induces late lumen enlargement in coronary arteries after balloon angioplasty. Clin Res Cardiol 2015; 104: 217–225. [DOI] [PubMed] [Google Scholar]
  • 39. Chevalier B, Smits PC, Carrié D, et al. Serial assessment of strut coverage of biodegradable polymer drug-eluting stent at 1, 2, and 3 months after stent implantation by optical frequency domain imaging: the DISCOVERY 1 to 3 study (evaluation with OFDI of strut coverage of terumo new drug eluting stent with biodegradable polymer at 1, 2, and 3 months). Circ Cardiovasc Interv 2017; 10(12). [DOI] [PubMed] [Google Scholar]
  • 40. Kim C, Kim BK, Hong SJ, et al. Randomized comparison of strut coverage between ticagrelor and clopidogrel in acute myocardial infarction at 3-month optical coherence tomography. Yonsei Med J 2018; 59: 624–632. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41. Suwannasom P, Onuma Y, Benit E, et al. Evaluation of vascular healing of polymer-free sirolimus-eluting stents in native coronary artery stenosis: a serial follow-up at three and six months with optical coherence tomography imaging. EuroIntervention 2016; 12: e574–e583. [DOI] [PubMed] [Google Scholar]
  • 42. Yano H, Horinaka S, Ishikawa M, et al. Early vascular responses after everolimus-eluting stent implantation assessed by serial observations of intracoronary optical coherence tomography. Heart Vessels 2017; 32: 804–812. [DOI] [PubMed] [Google Scholar]
  • 43. Koppara T, Tada T, Xhepa E, et al. Randomised comparison of vascular response to biodegradable polymer sirolimus eluting and permanent polymer everolimus eluting stents: an optical coherence tomography study. Int J Cardiol 2018; 258: 42–49. [DOI] [PubMed] [Google Scholar]
  • 44. Shin ES, Ann SH, Balbir Singh G, et al. Fractional flow reserve-guided paclitaxel-coated balloon treatment for de novo coronary lesions. Catheter Cardiovasc Interv 2016; 88: 193–200. [DOI] [PubMed] [Google Scholar]
  • 45. Tanigaki T, Emori H, Kawase Y, et al. QFR Versus FFR derived from computed tomography for functional assessment of coronary artery stenosis. JACC Cardiovasc Interv 2019; 12: 2050–2059. [DOI] [PubMed] [Google Scholar]
  • 46. Baumann S, Mueller KSE, Hetjens S, et al. One-year clinical outcome of angiography, fractional flow reserve and instantaneous wave-free ratio guided percutaneous coronary intervention: a PRISMA-compliant meta-analysis. Exp Ther Med 2019; 17: 1939–1951. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47. Baumann S, Chandra L, Skarga E, et al. Instantaneous wave-free ratio (iFR®) to determine hemodynamically significant coronary stenosis: a comprehensive review. World J Cardiol 2018; 10: 267–277. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48. Head SJ, Kaul S, Bogers AJ, et al. Non-inferiority study design: lessons to be learned from cardiovascular trials. Eur Heart J 2012; 33: 1318–1324. [DOI] [PubMed] [Google Scholar]
  • 49. Rehal S, Morris TP, Fielding K, et al. Non-inferiority trials: are they inferior? A systematic review of reporting in major medical journals. BMJ Open. 2016; 6: e012594. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50. Food and Drug Administration, HHS. Non-Inferiority Clinical Trials to Establish Effectiveness. Draft Guidance for Industry, https://www.fda.gov/media/78504/download (2010).