Abstract
Purpose:
Spin-echo functional MRI (SE-fMRI) has the potential to improve spatial specificity when compared with gradient-echo fMRI. However, high spatiotemporal resolution SE-fMRI with large slice-coverage is challenging as SE-fMRI requires a long echo time to generate blood oxygenation level-dependent (BOLD) contrast, leading to long repetition times. The aim of this work is to develop an acquisition method that enhances the slice-coverage of SE-fMRI at high spatiotemporal resolution.
Theory and Methods:
An acquisition scheme was developed entitled multisection excitation by simultaneous spin-echo interleaving (MESSI) with complex-encoded generalized slice dithered enhanced resolution (cgSlider). MESSI uses the dead-time during the long echo time by interleaving the excitation and readout of 2 slices to enable 2× slice-acceleration, while cgSlider uses the stable temporal background phase in SE-fMRI to encode/decode 2 adjacent slices simultaneously with a “phase-constrained” reconstruction method. The proposed cgSlider-MESSI was also combined with simultaneous multislice (SMS) to achieve further slice-acceleration. This combined approach was used to achieve 1.5-mm isotropic whole-brain SE-fMRI with a temporal resolution of 1.5 s and was evaluated using sensory stimulation and breath-hold tasks at 3T.
Results:
Compared with conventional SE-SMS, cgSlider-MESSI-SMS provides 4-fold increase in slice-coverage for the same repetition time, with comparable temporal signal-to-noise ratio. Corresponding fMRI activation from cgSlider-MESSI-SMS for both fMRI tasks were consistent with those from conventional SE-SMS. Overall, cgSlider-MESSI-SMS achieved a 32× encoding-acceleration by combining Rinplane × MB × cgSlider × MESSI = 4 × 2 × 2 × 2.
Conclusion:
High-quality, high-resolution whole-brain SE-fMRI was acquired at a short repetition time using cgSlider-MESSI-SMS. This method should be beneficial for high spatiotemporal resolution SE-fMRI studies requiring whole-brain coverage.
Keywords: functional imaging, GSlider, Hadamard encoding, slice interleaving, spin echo
1 |. INTRODUCTION
Functional MRI (fMRI) has been used as a powerful tool to investigate human brain function.1,2 It is well known that acquisition strategies can significantly affect the specificity and sensitivity of blood oxygenation level-dependent (BOLD) signals. Generally, gradient-echo (GE)-BOLD imaging is often used because of its ease of implementation and high contrast-to-noise ratio. However, GE-BOLD is highly sensitive to large draining veins3–6 and suffers from signal dropout due to susceptibility effects near air-tissue interfaces.7 On the other hand, spin-echo (SE)-BOLD results in reduced contrast-to-noise ratio compared with GE-BOLD, which limits the use of SE-BOLD in fMRI studies at lower fields. SE-BOLD at 3T has similar intravascular and extravascular contributions, which reduces tissue sensitivity compared with GE-BOLD.8,9 Despite these disadvantages of SE-BOLD, studies have shown the advantage of SE-BOLD over GE-BOLD at 3T8,10,11 in recovering the signal dropout near the regions of strong B0 inhomogeneity.
At ultrahigh magnetic field strengths (e.g., 7T), SE-fMRI has also been shown to provide improved spatial specificity when compared with GE-fMRI, as it enhances the relative sensitivity of the BOLD signal from the parenchyma.12–14 Thus, SE-fMRI with enhanced spatial specificity can be useful to study brain organization and function at the cortical laminar or columnar levels.15–20 However, high spatiotemporal resolution SE-fMRI is difficult due to the long echo time (TE) needed to generate BOLD contrast (TE ≈ T2 of gray matter)13,21 and associated long repetition times (TRs), along with higher specific absorption rate (SAR) from high flip-angle pulses. Nevertheless, achieving high spatiotemporal resolution as well as high spatial specificity is important in fMRI to investigate brain function at fine scales.
Although partial Fourier22 and parallel imaging23–25 techniques are very useful in reducing the number of phase-encoding steps in echo-planar imaging (EPI), these methods do not alleviate challenges in achieving whole-brain imaging with high spatiotemporal resolution. Recently, simultaneous multislice (SMS) has been introduced to increase the temporal resolution of fMRI26–30 while maintaining whole-brain slice-coverage, and the accelerated temporal sampling has been shown to be beneficial in several applications.31,32 Use of the CAIPIRINHA27 (controlled aliasing in parallel imaging results in higher acceleration) technique in the form of blipped-CAIPI30 for EPI can reduce the g-factor noise by shifting adjacent excited slices relative to each other in the phase-encoding direction and has been established as a standard technique in SMS-EPI.33 However, SE-SMS-EPI typically operates at low multiband (MB) factors (MB ≤ 3) due to peak power and SAR limitations, as well as T1 saturation effects.34 Higher MB accelerations also introduce significant g-factor noise, especially when combined with in-plane acceleration.35 Further slice-acceleration beyond conventional-SMS has been demonstrated with the principles of echo-shifting with a train of observations (PRESTO) technique,31,32 which has been used for fMRI acquisitions.31,36–38 Other echo-shifting techniques39–42 use the dead-time between excitation and readout, but these techniques are based on GE sequences. TE Interleaving imaging43 and simultaneous echo refocusing44 increase the number of slices per TR, up to 3. However, to the best of our knowledge, echo-shifting techniques have not been combined with SE-EPI.
In this work, we introduce 2 complementary technologies (i) complex-encoded generalized slice dithered enhanced resolution (cgSlider) and (ii) multisection excitation by simultaneous spin-echo interleaving (MESSI) to achieve higher slice-accelerations in SE-fMRI. With cgSlider, temporally modulated radiofrequency (RF)-encodings between spatially adjacent simultaneously acquired imaging subslices are used along with a phase-constrained reconstruction to achieve a 2× gain in slice-acceleration by taking advantage of the stable temporal background phase in SE. With MESSI, the dead-time during the long TE period in SE-fMRI40,45 is used to interleave the excitation and readout of 2 imaging slices to provide an additional 2× slice-acceleration. cgSlider and MESSI can be combined, which can also be used in conjunction with conventional SMS parallel imaging. The 4× increase in slice-acceleration provided by cgSlider and MESSI does not come with additional g-factor penalty or any significant increase in peak RF power.
We demonstrate that cgSlider-MESSI-SMS enables whole-brain SE-fMRI acquisition at a nominal isotropic spatial resolution of 1.5mm, with a high temporal resolution of 1.5 s and low image distortion and blurring (Rinplane = 4). A total encoding-acceleration of 32× was achieved in this acquisition using Rinplane × MB × cgSlider × MESSI = 4 × 2 × 2 × 2. SE-fMRI experiments at 3T using sensory stimulation and breath-hold tasks were used to demonstrate that the 4× enhancement in slice-coverage from cgSlider-MESSI-SMS can be achieved with minimal penalty when compared with conventional SE-SMS-EPI with the same temporal resolution.
2 |. THEORY
In this section, descriptions of the cgSlider and MESSI techniques are provided. Each of these methods can achieve a 2-fold slice-acceleration and can be used jointly, along with conventional-SMS acceleration, to achieve high slice-accelerations in SE-fMRI.
2.1 |. cgSlider
Two adjacent subslices are acquired together using cgSlider RF-encoding, where the excitation phase of 1 of the subslices (blue colored subslice in Figure 1A) is modulated across the time frames, as shown in Figure 1A, which can be described as:
| (1) |
where Scg is the cgSlider signal acquired at each time point (n) consisting of the combination of 2 simultaneously encoded adjacent subslices. SA and SB are the magnitudes of subslices A and B, ϕA and ϕB are the corresponding background phases of subslices A and B, respectively. Here, θ(n) denotes the temporally modulated RF-encoding phase of subslice B at the nth temporal frame. A Shinnar-Le Roux46 based cgSlider RF pulse34 was used for the 90° excitation pulse with a time-bandwidth-product (TBWP) of 9, in conjunction with a standard Shinnar-Le Roux pulse for the 180° refocusing with a TBWP of 5. This design provides no increase in the 180° peak voltage and approximately the same 90° peak-voltage when compared with standard single-slice acquisition. It is important to note the use of complex-valued signal modulation of subslice B and the addition of a time-varying phase modulation, which enables separation of the 2 subslices from the acquired slab-signal using the reconstruction method described below. Note that the division of the slab into subslices will result in different inflow effects for each of the 2 subslices due to inflowing spins from above and below the slab. In this initial work we will not investigate these effects but plan to investigate them more quantitatively in future work.
FIGURE 1.

A, Illustration of complex-valued gSlider RF-encoding. Temporally varying phase modulation of ±π/2 was applied to the second subslice (blue text). B, Illustration of the “sliding-window” reconstruction. By assuming signal magnitudes and background phases are slowly varying between time frames, signal magnitudes and background phases for each slice were estimated through a “sliding-window” reconstruction shown as black arrows. C, After estimating background phases from the sliding-window reconstruction, the “phase-constrained” reconstruction estimates signal magnitudes for subslices A and B at each time frame directly without temporal smoothing
2.2 |. cgSlider image reconstruction
2.2.1 |. “Sliding-window” reconstruction
The complex-encoding described above enables conventional Hadamard encoding reconstruction methods47 across 2 adjacent time frames. For example, with the 1st and 2nd time frames, it is assumed that the underlying subslice image magnitudes and phases SA, SB, ϕA, and ϕB are slowly varying between these 2 adjacent time frames in the SE-fMRI acquisition (ϕA(1) ≈ ϕA(2), ϕB(1) ≈ ϕB(2), SA (1) ≈ SA(2), SB(1) ≈ SB(2)). In other words, due to the refocusing of an SE, we assume that both the magnitude of the signal and, more critically, the phase of the signal are not changing in the brain over this short time frame, that is, physiological processes such as those driven by the cardiac or respiratory cycles as well as neuronal activity are assumed to not cause a substantial change in the image phase. Under this assumption, Equation 1 can be rewritten as:
| (2) |
The signal magnitude and phase of each subslice can then be obtained by adding or subtracting the cgSlider subslices signal of 2 adjacent time frames, which can be written as:
| (3) |
where the signals from the 2 subslices at the time point half-way between the 2 acquisitions (i.e., at n = 1.5) are expressed as linear combinations of the slab signals measured at time point 1 and time point 2. This expression can be applied to subsequent time points to provide a reconstruction of the subslice time-series data by a “sliding-window” reconstruction in which each reconstructed time-point results from the linear combination of the 2 surrounding time points, as illustrated in Figure 1B with black arrows. This method allows simple separation of the cgSlider signal while obtaining an image signal-to-noise ratio (SNR) benefit by a factor of due to noise averaging. However, this method also causes temporal blurring effects owing to the sharing of data across 2 adjacent time points.
2.2.2 |. “Phase-constrained” reconstruction
The sliding-window reconstruction makes a strong assumption about both the magnitude and phase of the acquired image data being slowly varying over time such that neither components of the complex-valued signal change appreciably from 1 time point to the next. However, this strong assumption can be relaxed for the signal magnitude. It is observed that the signal change in SE-fMRI is mostly confined to the image magnitude, and there is little change in the background phase (shown in Supporting Information Figure S1, which is available online). By taking advantage of the observed negligible temporal phase variations in SE-fMRI, the background phase reconstructed from the sliding window method (ϕA−SW and ϕB−SW in Figure 1B) can be used as an initialization for a phase-constrained reconstruction so that the number of unknown values in Equation 1 is reduced from 4 to 2; in this approach the sliding-window reconstruction provides a reference phase to enable the reconstruction of the magnitude of the 2 imaging subslices directly from each acquired time frame without the temporal blurring induced by the sliding-window approach. However, the time point of the estimated phase from the sliding-window approach is ϕA−SW (n + 0.5) or ϕA−SW (n + 0.5). To match the number of time points, n′ was set as (n − 0.5) and the last time point was repeated with (last time point − 0.5). Here, the phase-constrained reconstruction estimates the subslice signals SA−PC (n) and SB−PC (n) by solving the following linear system of equations through simple matrix inversion:
If n is odd,
| (4) |
If n is even,
| (5) |
where ΔϕSW is the phase difference calculated between the 2 subslices from the sliding-window reconstruction, i.e., ΔϕSW = ΔϕB−SW − ΔϕB−SW. Equations 4 and 5 are derived in Supporting Background Information.
2.3 |. MESSI
A schematic diagram of the MESSI pulse sequence is shown in Figure 2. To acquire 2 imaging slices jointly in an interleaved manner (denoted as MESSI-1 and MESSI-2), the following 4 sequence components were added to conventional SE-EPI sequence. First, an additional readout and 90° and 180° pulses for the MESSI-2 slice (Figure 2, blue-colored RF pulses and readout) with a TE matched to that of the MESSI-1 slice (Figure 2, red-colored RF pulses and readout) were added. Second, to separate the k-space signals of the 2 MESSI slices, dephasing gradients (Figure 2, green-colored gradients) that shift the signal of the different slices in-plane were added before the 180° pulse of the MESSI-2 slice. Gradient moment parameters α and β correspond to kmax/2 of frequency encoding and phase encoding, respectively, and kfactor is the integer-valued scaling parameter determining the distance in k-space between the 2 MESSI slices. As kfactor is increased by 1, the distance between the signal of the 2 slices is increased by kmax in frequency and phase encoding directions, which acts to prevent k-space signal leakage between slices. Third, before the data acquisition of the MESSI-1 slice, rephasing gradients (Figure 2, red-striped gradients) were inserted to rephase the signal for MESSI-1 slice. During data acquisition of MESSI-1 slice, the spins from MESSI-2 slice are dephased. For the same reason, rephasing gradients for MESSI-2 slice (Figure 2, blue-striped gradients) were inserted. Fourth, to avoid free induction decay signal from 180° RF pulse of MESSI-2 slice introduced by imperfect RF refocusing pulse, spoiler gradients were added (Figure 2, purple-striped gradients).
FIGURE 2.

MESSI sequence diagram, showing 2 interleaved slices (MESSI-1, red and MESSI-2, blue). The dephasing gradients separate the k-spaces of the 2 MESSI groups (green-striped gradients). α and β correspond to kmax/2 values along the readout and phase encoding directions, respectively. kfactor is the integer-valued scaling parameter that determines the distance of the k-space centers between the 2 MESSI groups. The rephasing gradients are for ensuring that the 0th moments of all gradients are zero before acquiring the MESSI-1 or MESSI-2 readouts (red- and blue-striped gradients). Spoiling gradients (purple-striped gradients) were added to avoid possible artifacts from free induction decay signals arising from imperfect RF refocusing pulses
2.3.1 |. kfactor optimization
The effect of the inserted MESSI echo-shifting (dephasing and rephrasing) gradients on the spins from the 2 MESSI slices in the cases where kfactor is set to 1 or 2 is illustrated in Figure 3A,B, respectively. Rephasing gradients for MESSI-1 slice (Figure 3, red-striped gradients) rephase spins in the readout for MESSI-1, while dephasing magnetization from MESSI-2 slice. The same is true for the rephasing gradients (Figure 3, blue-striped gradients) for MESSI-2 slice. Increasing the value of kfactor increases the signal dephasing between MESSI slice groups and reduces the potential for signal leakage between the slices for data at the edges of k-space.
FIGURE 3.

A,B, The MESSI sequence diagrams and phase evolutions along the readout gradient for spins from the 2 MESSI groups, for the cases where the integer-valued kfactor parameter is set to 1 or 2, respectively. The red solid line and blue dashed line correspond to the MESSI-1 and MESSI-2 groups, respectively. As kfactor is increased, the distance between 2 k-spaces of the 2 MESSI groups is increased, as shown with blue and red boxes. Numbers listed within the gradient lobes signify the relative value of the 0th moment
2.4 |. cgSlider-MESSI-SMS sequence
Figure 4 describes how the cgSlider, MESSI and conventional SMS techniques can be combined synergistically to provide high slice-accelerations in SE-fMRI. MESSI enables an increase in the slice-coverage by exploiting dead-time and exciting an additional slice group (Figure 4, red and blue slices) per TR, whereas cgSlider allows an increase in coverage by exciting complex-encoded spatially adjacent subslices (Figure 4, red slab). Combining these techniques (cgSlider-MESSI) enables simultaneous excitation of additional slice groups and their complex-encoded spatially adjacent slices (Figure 4, red and blue slabs). Inclusion of SMS further extends the slice coverage through exciting cgSlider-MESSI-1 and cgSlider-MESSI-2 groups (Figure 4, yellow and green slabs), spaced apart evenly across the field of view (FOV) in the z direction.
FIGURE 4.

Overview of the combination of cgSlider, MESSI, and conventional SMS techniques. MESSI enables an increase in the slice coverage by exciting an additional slice group (red and blue slices), whereas cgSlider allows an increase in coverage by exciting complex-encoded spatially adjacent slices (red slab). Combining the techniques (cgSlider-MESSI) enables simultaneous excitation of additional slice groups and their complex-encoded spatially adjacent slices (red and blue slabs). Inclusion of SMS extends the slice coverage through exciting cgSlider-MESSI-1 and cgSlider-MESSI-2 groups (yellow and green slabs), spaced apart evenly across the FOVz
2.5 |. Velocity-encoding phase correction and reference phase acquisition
There are 3 phase components in the image produced by cgSlider-MESSI-SMS: the background phase, the velocity-encoding phase from additional gradients for the MESSI sequence implementation, and the change in phase due to the fMRI activation. Large dephasing/rephasing gradients in the MESSI sequence can introduce non-negligible velocity-encoding, which can induce phase variations due to respiration/cardiac induced movement and head motion.48–51 Such phase variations can affect the cgSlider-MESSI reconstruction, causing striping artifacts in the reconstructed images along the slice-direction. An approach to remove this phase corruption was developed that takes advantage of the fact that this phase corruption is typically spatially smooth and should not vary substantially across the thin slab of the cgSlider-encoding. In this approach, first, the phase images (denoted as ∠(Scg(t))) were averaged separately for all odd- and even-numbered frames of the time series data of the cgSlider slab-encoded signal as shown in Figure 5 in the green box (avg ∠ (Scg)). Second, the phase difference between each time frame and the averaged phase (∠(difference) = ∠Scg(t) − avg ∠ (Scg)) was calculated separately for the odd and even time frames. Third, because of the background phase is well known to be smoothly varying in SE images, a spatial filter was applied to the phase difference image (∠(difference_filtered)), to reduce noise and more accurately estimate the velocity-encoding phase variation, and the estimated velocity-encoding phase was removed, leaving behind the background phase and the phase change related to fMRI activation. Finally, the phase-constrained reconstruction was performed after this velocity-encoding phase correction. However, this velocity phase removal process is not perfect, which can remain striping artifacts. An alternative approach was also examined, neglects the temporal phase changes related to fMRI activation, which should be relatively small. With this assumption, the phase of the cgSlider-MESSI-SMS was replaced by a reference phase from a cgSlider-SMS with matching sequence parameters that contains only the background phase with no velocity-encoding phase contamination.
FIGURE 5.

Overview of the phase correction process for cgSlider and cgSlider-MESSI acquisitions: First, the averaged phases for odd- and even-numbered time frames were calculated (green box). The phase difference between each time frame and averaged phases (∠(difference)) was found, before a Hamming filter was then applied (∠(difference_filtered)) to subtract the velocity encoding phase that is sensitive to physiological changes such as motion. After the velocity encoding phase correction, the “phase-constrained” reconstruction was performed
3 |. METHODS
3.1 |. Participants
Nine healthy subjects (5 male, 4 female), aged 25–39 years old, participated in this study. All procedures followed the guidelines of the Institutional Review Board of the Massachusetts General Hospital and Sungkyunkwan University. Procedures were fully explained to all subjects, and informed written consent was obtained before scanning in accordance with the Declaration of Helsinki.
3.2 |. MRI acquisition
All measurements were performed on a 3T scanner (MAGNETOM Prisma, Siemens Healthineers, Erlangen, Germany) with the vendor-supplied 32-channel head coil and the vendor supplied 64-channel head and neck coil. The developed sequence was combined with the blipped-CAIPI SMS technique30 at MB = 2 to further increase slice-coverage and capture the entire brain in a single repetition, and Rinplane = 4 was used to minimize image distortion and blurring. VERSE52 was also applied to the MB cgSlider RF pulses to reduce peak voltage and SAR. VERSE was applied to both 90° and 180° pulses to ensure that the slice profile degradations and shifts at off-resonance are similar across excitation and refocusing to achieve good signal level.34 The sequence parameters used here are as follows: TR/TE = 1500/75 ms, FOVxy = 210 × 210 mm2, partial Fourier = 6/8, 1.5-mm isotropic resolution, effective echo spacing (ESP) = 0.173 ms. The readout bandwidth parameter value was chosen here to minimize the echo spacing to minimize EPI blurring and distortion, as is commonly done for conventional EPI. The reference phase acquisition was collected using cgSlider-SMS with a TR of 2500 ms; this longer TR was required to match the number of slices between the reference data and the accelerated cgSlider-MESSI-SMS data. The reference data were acquired at the beginning of each run before fMRI data collection and, therefore, introduced a small increase in total scan duration (2.5 s per run).
3.3 |. kfactor optimization in MESSI sequence
To examine the level of signal leakage between MESSI slices, direct measurements of the signal leakage levels were obtained in the MESSI sequence by setting either MESSI-1 or MESSI-2 RF excitation pulse flip angles to 0° for acquisitions. Two kfactor settings of 1 or 2 were evaluated.
3.4 |. Velocity-encoding phase correction in both cgSlider and cgSlider-MESSI
For both cgSlider and cgSlider-MESSI cases, reconstructions were performed with and without velocity-encoding phase correction to assess temporal SNR (tSNR) level improvement (see below for description of tSNR comparisons).
3.5 |. tSNR comparisons
For tSNR analysis, 3 protocols were compared with 4 subjects: conventional-SMS, cgSlider-SMS, and cgSlider-MESSI-SMS. To achieve an unbiased comparison, the MR parameters and TR were kept constant and the total number of slices were adjusted accordingly to the net slice-acceleration factor of each protocol. In summary, we compared (i) conventional-SMS with Rinplane × MB = 4 × 2, 26 slices (100% slice gap), FOVz = 78 mm; (ii) cgSlider-SMS with Rinplane × MB × cgSlider = 4 × 2 × 2, 52 slices (no slice gap), FOVz = 78 mm; and (iii) cgSlider-MESSI-SMS with Rinplane × MB × cgSlider × MESSI = 4 × 2 × 2 × 2, 84 slices (no slice gap), FOVz = 126 mm (whole-brain coverage), kfactor = 2. The number of repetitions (NR) was 140 for each protocol, corresponding to a total acquisition time of 3 min 30 s per protocol. The tSNR maps were calculated from 100 NRs, excluding 20 NRs at both the beginning and at the end, by dividing the temporal mean of the time series by the temporal standard deviation. Additionally, average and standard deviation of the resulting tSNR were calculated in ROIs defined as brain in 4 subjects.
3.6 |. Visual/breath-hold fMRI activation
To assess the performance of cgSlider-MESSI-SMS compared with conventional SMS in SE-fMRI, fMRI data were acquired using the 3 protocols described above: conventional-SMS, cgSlider-SMS, and cgSlider-MESSI-SMS. For the visual stimulation session, 3 subjects were presented with a standard flashing scaled-checkboard stimulus (12 s on, 20 s off, 4 on-off blocks per run); each run lasted 210 s, and 4 runs were acquired for each protocol that were averaged together during the analysis. For the timed breath-hold task, the subject was cued to hold their breath for 12 s followed by 30 s of free breathing with 4-breath-holds/run, and 7 runs were acquired for each protocol averaged during analysis. FSL (http://www.fmrib.ox.ac.uk/fsl) was used to perform fMRI analysis; spatial smoothing (3-mm kernel) was applied for the breath-hold task to boost SNR but not for the visual stimulation task where SNR is sufficient, while MCFLIRT motion correction was applied to both.
4 |. RESULTS
Supporting Information Figure S1A shows the z-statistic maps for a visual stimulation task obtained using the time-series image magnitude and the time-series image phase of a single conventional-SMS SE-fMRI acquisition. The results demonstrate that the BOLD responses in conventional-SMS SE-fMRI is mostly confined to the image magnitude, and little/no change in the corresponding background phase was detected in response to activation. Supporting Information Figure S1B shows the estimated background phases for 2 adjacent subslices from the cgSlider acquisition. The phases of the adjacent subslices were similar to each other, which supports the feasibility of the proposed phase-constrained reconstruction approach.
To quantify signal leakage as a function of the value of kfactor, we acquired test data in 1 subject with different parameter values. Figure 6 shows the results of this analysis, including the reconstructed images and the signal leakage maps corresponding to signal from 1 slice in the MESSI slice group leaking into the other slice. Figure 6A,B shows the signal leakage between the MESSI slice groups for acquisitions with kfactor of 1 and 2. Upper and lower rows show the results from the cases, when MESSI-1 or MESSI-2 pulses were set to 0°. With kfactor = 2, the signal leakage between MESSI groups is negligible, whereas with kfactor = 1, the leakage is clearly seen. The leakage maps were all multiplied by a factor of 10 relative to the brain images to visualize the leakage pattern. To avoid signal leakage, a kfactor of 2 was used for MESSI acquisitions.
FIGURE 6.

Characterization of potential k-space signal leakage expected in the high spatial frequencies, as a function of the kfactor parameter value. To directly investigate the level of signal leakage between 2 MESSI groups, either MESSI-1 or MESSI-2 excitation pulses were set to 0°. A,B, Display maps of signal leakage between MESSI groups for the case of kfactor = 1 and 2, respectively. Red and blue tinted frames represent MESSI-1 and 2 groups, respectively. With kfactor = 2 the signal leakage between MESSI groups is negligible, whereas with kfactor = 1 leakage is clearly seen (white arrows). The leakage maps were all multiplied by a factor of 10 relative to the brain images to visualize the leakage pattern
The left column of Figure 7A shows tSNR maps without velocity-encoding phase correction, at varying kfactor from 1 to 4 to evaluate the effect of large dephasing/rephasing gradients in the MESSI sequence. Higher kfactor results in lower tSNR, which reflects that larger gradients induce higher sensitivity to potential image phase variations due to respiration/cardiac induced movement and head motion. To overcome tSNR deterioration, velocity-encoding phase correction was applied and showed comparable tSNR level, despite the stronger dephasing/rephasing gradients from the increased kfactor, as shown in right column of Figure 7A. Also, a comparison of the reconstruction without correction, with the velocity-encoding phase correction, and with reference phase is shown in Figure 7B, where sagittal reformats of axially acquired slices are presented. The striping artifact across slices, was substantially reduced by the velocity-encoding phase correction, but not perfectly removed. For example, the white arrows in Figure 7B point to an area that shows the reduced striping artifact both with velocity-encoding phase correction and with reference phase. However, the yellow arrows point to an area showed less striping with the reference phase than with the velocity-encoding phase correction, which reflects that the striping artifact is mostly originated from the velocity-encoding due to the additional gradient lobes in the MESSI sequence.
FIGURE 7.

A, tSNR maps without velocity encoding phase correction, varying kfactor from 1 to 4 to evaluate the effect of large dephasing and rephasing gradients in the MESSI sequence (left column). Corresponding tSNR maps after the application of velocity encoding phase correction in the reconstruction, at varying kfactor from 1 to 4 (right column). B, Reconstructed cgSlider-MESSI-SMS sagittal images with no correction (top), with velocity encoding phase correction (middle), and with reference phase (bottom)
The reconstructed conventional-SMS, cgSlider-SMS, and cgSlider-MESSI-SMS images were compared in terms of overall image quality as well as the resulting tSNR before and after velocity-encoding phase correction (see Figure 8). The cgSlider-SMS and cgSlider-MESSI-SMS tSNR maps, before and after velocity-encoding phase correction, are shown in Figure 8B,C. There is no apparent difference in tSNR maps among conventional SMS, cgSlider-SMS, and cgSlider-MES-SI-SMS with velocity-encoding phase correction. However, although the TBWP of RF pulses in conventional-SMS were matched with that of cgSlider-SMS, slightly higher tSNR values were seen for the cgSlider-SMS reconstruction when compared with that from conventional-SMS. This is likely caused by the expected small increase in signal level in cgSlider-SMS due to improved signal refocusing performance. In particular, the refocusing pulse for the cgSlider-SMS acquisition extends across the 2 subslices of cgSlider, with each subslice experiencing only 1 transition band with incomplete refocusing, rather than 2 in the conventional-SMS. For further assessment, average and standard deviations of tSNR obtained from 4 subjects were compared. Average values ± standard deviations for conventional-SMS, cgSlider-SMS, and cgSlider-MESSI-SMS were 9.2 ± 1.0, 10.6 ± 1.4, and 10.6 ± 1.2, respectively. Average tSNR values among different methods were not statistically significant. Our velocity-encoding phase correction resulted in comparable tSNR to the conventional-SMS (Figure 8C). However, tSNR maps from cgSlider-MESSI-SMS without velocity-encoding phase correction showed much lower tSNR than other protocols (Figure 8B) due to the increased velocity-encoding induced by the dephasing/rephasing gradients used for MESSI.
FIGURE 8.

Comparisons of image quality (A) and corresponding tSNR before (B) and after (C) velocity encoding phase correction for conventional SE SMS, cgSlider-SMS, cgSlider-MESSI-SMS. Similar image quality and tSNR levels were achieved among different methods with velocity encoding phase correction. For both cgSlider-SMS and cgSlider-MESSI-SMS, phase correction resulted in an improved tSNR. The improvement is particularly significant in cgSlider-MESSI-SMS, which contains more shot-to-shot image phase variations due to the increased velocity encoding induced by the echo shifting (dephasing and rephrasing) gradients. Intensity correction was performed for (A)
In particular, the proposed cgSlider-MESSI-SMS approach achieves whole-brain coverage at 1.5-mm isotropic resolution with 1.5-s temporal resolution. In the presented conventional-SMS there is a 100% gap imposed to allow for this acquisition, to have the same brain-coverage as cgSlider-SMS, albeit with half the number of reconstructed slices.
Finally, Figure 9 demonstrates the feasibility of SE-fMRI using cgSlider-MESSI-SMS by comparing the resulting BOLD activation maps with those values from conventional-SMS and cgSlider-SMS. For each acquisition, z-statistic maps (thresholded at P < 0.01) are overlaid on a single reconstructed image of the corresponding acquisition. Figure 9A shows z-statistical maps from 2 adjacent-slices, demonstrating the feasibility of the proposed cgSlider reconstruction to enable both anatomical details and functional activation patterns from 2 adjacent subslices. For both the visual stimulation and breath-hold fMRI experiments, cgSlider-MESSI-SMS maintains the same temporal resolution at 2× brain coverage when compared with that from cgSlider-SMS, while exhibiting comparable activation patterns. The similarity of activation maps between these acquisitions indicates that fMRI sensitivity is not compromised with the addition of MESSI. For further validation, z-statistic map comparisons among different methods were shown with 2 subjects in Supporting Information Figure S2, also mean and standard deviation of thresholded z values (z > 2), and the number of activated voxels (clusters of minimum size of 30 voxels) were calculated in Supporting Information Table S1. Especially in Figure 9B, activations were detected in the medial prefrontal cortex (yellow arrows) are known to be nearby regions with large susceptibility gradients, which are difficult to detect with GE-EPI sequence.8,11
FIGURE 9.

Functional activation maps, represented as z-statistics of the detected BOLD response, resulting from visual stimulation and breath-hold fMRI datasets (spatial smoothing was applied for the breath-hold task to boost SNR but not for the visual stimulation task where SNR was sufficient). The cgSlider-MESSI-SMS method maintains high-temporal resolution at 4×, and 2× brain coverage compared with conventional SE-SMS and cgSlider-SMS. Also, the cgSlider-MESSI-SMS achieves a comparable extent of activation compared with cgSlider-SMS for both the visual stimulation (A) and breath-hold task (B). Intensity correction was performed for all images
5 |. DISCUSSION
Here we proposed a new method for accelerating SE-fMRI using the cgSlider-MESSI-SMS acquisition. The cgSlider component enables a 2-fold increase in slice-acceleration using a phase-constrained reconstruction that uses the spatiotemporal smoothness of SE-image phase. The MESSI component provides a ~2× higher efficiency in slice acquisition by interleaving excitation and data collection within the sequence dead-time, taking advantage of the long TE of BOLD-weighted SE-fMRI. Finally, conventional-SMS was further combined with both cgSlider and MESSI, resulting in a total ~32× acceleration factor (Rinplane × MB × cgSlider × MESSI = 4 × 2 × 2 × 2). Our cgSlider-MESSI-SMS approach successfully demonstrated 1.5-mm isotropic whole-brain coverage at a temporal resolution of 1.5 s, which is not feasible with conventional SE-SMS-EPI. The feasibility of SE-fMRI with this new sequence was demonstrated for both sensory stimulation and breath-hold tasks in healthy subjects at 3T, which showed comparable z-statistic maps to those from conventional SE-SMS-EPI at the same temporal resolution but at 4 times greater slice-coverage. Large slice-coverage with high spatiotemporal resolution should be particularly useful in resting-state fMRI studies, where whole-brain acquisitions are required.53
When the echo-shift method is applied to SE-EPI, the target TE will limit the readout window. In this work, we have achieved echo-shifting factor of 2 by applying in-plane acceleration factor of 4, allowing us to shorten the readout window. With our current scheme, a higher echo-shifting factor is not achievable without a significant TE increase. The MB-factor was limited to 2 to limit the total acceleration factor to 8 when using a 32-channel coil array at 3T (we used Rinplane × MB = 4 × 2).54 At higher field strength, acceleration performance increases and so a higher acceleration factor can be achieved,55 which would benefit the reduction of readout window further.
It is worthwhile to note that, while the use of higher kfactor and/or higher spatial resolution acquisition can reduce the leakage of high frequency signal between the MESSI data groups as shown in Figure 6, the velocity encoding and, hence, the image phase corruption would also increase due to the stronger gradients, as demonstrated in Figure 7A. Moreover, a higher kfactor results in a longer TE, lower SNR, and additional diffusion effect, which is not desirable for fMRI applications. Due to these potential artifacts from increasing kfactor, it is necessary to optimize kfactor to minimize the strength of gradients while still avoiding leakage of high frequency signal between the MESSI groups.
The additional diffusion effect of cgSlider-MESSI-SMS sequence with kfactor = 2 was calculated, and it was found that the maximum b-value was 1.28 s/mm2, which should not significantly affect the fMRI signal. The detrimental effect of this on the cgSlider reconstruction can be mitigated by the velocity-encoding phase correction step, as shown in Figure 7B. With this correction, cgSlider-MESSI-SMS results in nearly the same tSNR as that of cgSlider-SMS while providing increased slice-acceleration as shown in Figure 8. However, there are remaining stripes across slices, as shown in Figure 7B, suggesting that velocity/motion artifacts were not perfectly corrected even though the velocity-encoding phase correction provides improved image quality.
As an alternative approach in correcting phase corruption from velocity-encoding, a reference phase that uses a cgSlider-SMS “prescan” is proposed and shown to significantly reduce the striping artifacts. Moreover, the fMRI z-statistic maps from the reconstruction using this reference phase were compared with those using the velocity-encoding phase correction in Supporting Information Figure S2. While the reference phase does not incorporate the phase changes due to the fMRI activation into the reconstruction, the estimated fMRI activation from such approach was found to be comparable to that of the velocity-encoding phase correction approach, suggesting that the phase changes due to the fMRI activation are small relative to the velocity-encoded phase variation.
Based on the timing of the sequence, the target BOLD-weighted protocol can achieve a minimal volume acquisition time of 1.2 s while maintaining whole brain coverage with 1.5-mm isotropic resolution; however, the current implementation is limited by SAR. Typical SAR level in the cgSlider-MESSI-SMS approach was ~95% of the 6-min SAR limits. However, the use of lower flip angles to avoid overflipping in the center of the brain, such as 78° for excitation and 160° for refocusing,56 reduce RF power deposition and, thus, could also be explored with cgSlider-MESSI-SMS. This should reduce SAR level to ~75.5%. On the other hand, when the TR value is short compared with the tissue T1 value (e.g., TR = 1.5 s), the Ernst angle will maximize signal; however, the Ernst angle for the excitation pulse of an SE acquisition is typically larger than 90°. If we consider the SAR resulting from this higher excitation flip angle, which here would be ~113° at 3T, the SAR level would increase to ~103.8%. Therefore, the optimization of RF flip angles must balance between maximizing signal levels and achieving signal uniformity over the tissue of interest while remaining within the SAR limits.
Even though the peak power of the RF pulses in the sequence does not increase when cgSlider and MESSI are used, the total SAR increases by ~4× from the 4× increase in the number of slices being excited and refocused. In this work, the VERSE algorithm was applied to the RF pulses to help reduce SAR, which can result in some compromise in image quality in regions of strong B0 inhomogeneity. Future work will explore the use of alternative pulse design approaches57 and parallel transmission to help reduce SAR.58,59 In particular, Power Independent Number of Slices (PINS) pulses60–63 can be used to reduce SAR at the high MB factor.
Future work will also focus on the application of this method to ultra-high field SE-fMRI at 7T, where the T2 weighting can provide enhanced microvascular specificity and higher spatial resolution imaging. At ultra-high fields, the optimal TE value for SE-BOLD are shorter, which requires high in-plane acceleration. However, a higher in-plane acceleration factor might be acceptable because of reduced g-factor penalties55 at 7T. A multishot approach would allow for reduced echo-train-lengths, at the cost of advanced reconstruction techniques to overcome artifacts due to phase variations across shots.64,65 A reduced FOV acquisition with outer volume suppression66,67 will also be explored to further reduce echo-train-lengths, which in theory should not greatly affect image quality.
Given the parameters in this work, the main source of BOLD signal would be thermal noise. The tSNR maps shown in Figure 8B reflects this, where the highest tSNR values are on the outer part of the brain closest to the receiver coil, reflecting a thermal noise dominated acquisition.68
6 |. CONCLUSIONS
Here we proposed a new method, cgSlider-MESSI-SMS, and demonstrated that this can provide whole-brain SE-fMRI acquisitions at 3T with a spatial resolution of 1.5 mm and temporal resolution of 1.5 s through achieving a total acceleration factor of 32-fold using Rinplane × MB × cgSlider × MESSI = 4 × 2 × 2 × 2. With this newly developed pulse sequence and associated image reconstruction approaches, SE-fMRI experiments at 3T using sensory stimulation and breath-hold tasks successfully demonstrated the 4× enhanced slice-coverage with minimal SNR penalties.
Supplementary Material
FIGURE S1 A, Z-statistical maps obtained from signal intensity (top) and phase data (bottom) from a visual stimulation task. Detected fMRI responses are found only in the signal magnitude reconstruction with little/no fMRI responses detected in the phase image. B, Estimated background phase from cgSlider acquisition and sliding-window reconstruction for 2 adjacent slices (A, B subslices depicted in Figure 1) and 3 time points (1.5, 2.5, 3.5) with little phase variation over time and between subslices
FIGURE S2 Z-statistic maps from visual stimulation with conventional SMS (A), cgSlider-SMS (B), cgSlider-MES-SI-SMS after velocity encoding phase correction (C), and cgSlider-MESSI-SMS after correction using the reference phase (D). Intensity correction was performed for all images
TABLE S1 Mean and standard deviation of z values (z > 2), and thenumber of activated voxels (clusters of minimum size of 30 voxels) from three subjects
Funding information
IBS, Grant/Award Number: IBS-R015-D1; National Institutes of Health, Grant/Award Number: P41-EB015896, R01-EB019437, R01-EB020613, R01-MH111419, R01-MH116173, S10-RR019254, S10-RR019307, S10-RR023043, S10-RR023401 and U01-EB025162
Footnotes
SUPPORTING INFORMATION
Additional supporting information may be found online in the Supporting Information section.
REFERENCES
- 1.Kwong KK, Belliveau JW, Chesler DA, et al. Dynamic magnetic resonance imaging of human brain activity during primary sensory stimulation. Proc Natl Acad Sci USA. 1992;89:5675–5679. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Ogawa S, Tank DW, Menon R, et al. Intrinsic signal changes accompanying sensory stimulation: functional brain mapping with magnetic resonance imaging. Proc Natl Acad Sci USA. 1992;89:5951–5955. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Boxerman JL, Bandettini PA, Kwong KK, et al. The intravascular contribution to fmri signal change: Monte Carlo modeling and diffusion-weighted studiesin vivo. Magn Reson Med. 1995;34:4–10. [DOI] [PubMed] [Google Scholar]
- 4.Duyn JH, Moonen CTW, van Yperen GH, de Boer RW, Luyten PR. Inflow versus deoxyhemoglobin effects in bold functional MRI using gradient echoes at 1.5 T. NMR Biomed. 1994;7:83–88. [DOI] [PubMed] [Google Scholar]
- 5.Frahm J, Merboldt KD, Hänicke W, Kleinschmidt A, Boecker H. Brain or vein–oxygenation or flow? On signal physiology in functional MRI of human brain activation. NMR Biomed. 2007;7:45–53. [DOI] [PubMed] [Google Scholar]
- 6.Kim SG, Hendrich K, Hu X, Merkle H, Uğurbil K. Potential pitfalls of functional MRI using conventional gradient-recalled echo techniques. NMR Biomed. 1994;7:69–74. [DOI] [PubMed] [Google Scholar]
- 7.Weiskopf N, Hutton C, Josephs O, Deichmann R. Optimal EPI parameters for reduction of susceptibility-induced BOLD sensitivity losses: a whole-brain analysis at 3 T and 1.5 T. NeuroImage. 2006;33:493–504. [DOI] [PubMed] [Google Scholar]
- 8.Norris DG, Zysset S, Mildner T, Wiggins CJ. An investigation of the value of spin-echo-based fMRI using a Stroop Color-Word Matching Task and EPI at 3 T. NeuroImage. 2002;15:719–726. [DOI] [PubMed] [Google Scholar]
- 9.Ragot DM, Chen JJ. Characterizing contrast origins and noise contribution in spin-echo EPI BOLD at 3 T. Magn Reson Imaging. 2019;57:328–336. [DOI] [PubMed] [Google Scholar]
- 10.Thulborn KR, Chang SY, Shen GX, Voyvodic JT. High-resolution echo-planar fMRI of human visual cortex at 3.0 tesla. NMR Biomed. 1997;10:183–190. [DOI] [PubMed] [Google Scholar]
- 11.Khatamian YB, Golestani AM, Ragot DM, Chen JJ. Spin-echo resting-state functional connectivity in high-susceptibility regions: accuracy, reliability, and the impact of physiological noise. Brain Connect. 2016;6:283–297. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Harel N, Lin J, Moeller S, Ugurbil K, Yacoub E. Combined imaging-histological study of cortical laminar specificity of fMRI signals. NeuroImage. 2006;29:879–887. [DOI] [PubMed] [Google Scholar]
- 13.Lee SP, Silva AC, Ugurbil K, Kim SG. Diffusion-weighted spin-echo fMRI at 9.4 T: microvascular/tissue contribution to BOLD signal changes. Magn Reson Med. 1999;42:919–928. [DOI] [PubMed] [Google Scholar]
- 14.Han S, Son JP, Cho H, Park J, Kim S. Gradient-echo and spin-echo blood oxygenation level-dependent functional MRI at ultrahigh fields of 9.4 and 15.2 Tesla. Magn Reson Med. 2019;81:1237–1246. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Yacoub E, Shmuel A, Logothetis N, Uğurbil K. Robust detection of ocular dominance columns in humans using Hahn Spin Echo BOLD functional MRI at 7 Tesla. NeuroImage. 2007;37:1161–1177. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Yacoub E, Van De Moortele PF, Shmuel A, Uǧurbil K. Signal and noise characteristics of Hahn SE and GE BOLD fMRI at 7 T in humans. NeuroImage. 2005;24:738–750. [DOI] [PubMed] [Google Scholar]
- 17.Olman CA, Yacoub E. High-field fMRI for human applications: an overview of spatial resolution and signal specificity. Open Neuroimag J. 2011;5:74–89. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Norris DG. Spin-echo fMRI: the poor relation? NeuroImage. 2012;62:1109–1115. [DOI] [PubMed] [Google Scholar]
- 19.Zhao F, Wang P, Hendrich K, Ugurbil K, Kim SG. Cortical layer-dependent BOLD and CBV responses measured by spin-echo and gradient-echo fMRI: insights into hemodynamic regulation. NeuroImage. 2006;30:1149–1160. [DOI] [PubMed] [Google Scholar]
- 20.Koopmans PJ, Yacoub E. Strategies and prospects for cortical depth dependent T2 and T2* weighted BOLD fMRI studies. NeuroImage. 2019;197:668–676. [DOI] [PubMed] [Google Scholar]
- 21.Uğurbil K, Adriany G, Andersen P, et al. Magnetic resonance studies of brain function and neurochemistry. Annu Rev Biomed Eng. 2000;2:633–660. [DOI] [PubMed] [Google Scholar]
- 22.Feinberg DA, Hale JD, Watts JC, Kaufman L, Mark A. Halving MR imaging time by conjugation: demonstration at 3.5 kG. Radiology. 1986;161:527–531. [DOI] [PubMed] [Google Scholar]
- 23.Sodickson DK, Manning WJ. Simultaneous acquisition of spatial harmonics (SMASH): fast imaging with radiofrequency coil arrays. Magn Reson Med. 1997;38:591–603. [DOI] [PubMed] [Google Scholar]
- 24.Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. SENSE: sensitivity encoding for fast MRI. Magn Reson Med. 1999;42:952–962. [PubMed] [Google Scholar]
- 25.Griswold MA, Jakob PM, Heidemann RM, et al. Generalized auto-calibrating partially parallel acquisitions (GRAPPA). Magn Reson Med. 2002;47:1202–1210. [DOI] [PubMed] [Google Scholar]
- 26.Larkman DJ, Hajnal JV, Herlihy AH, Coutts GA, Young IR, Ehnholm G. Use of multicoil arrays for separation of signal from multiple slices simultaneously excited. J Magn Reson Imaging. 2001;13:313–317. [DOI] [PubMed] [Google Scholar]
- 27.Breuer FA, Blaimer M, Heidemann RM, Mueller MF, Griswold MA, Jakob PM. Controlled aliasing in parallel imaging results in higher acceleration (CAIPIRINHA) for multi-slice imaging. Magn Reson Med. 2005;53:684–691. [DOI] [PubMed] [Google Scholar]
- 28.Norris DG. Principles of magnetic resonance assessment of brain function. J Magn Reson Imaging. 2006;23:794–807. [DOI] [PubMed] [Google Scholar]
- 29.Moeller S, Yacoub E, Olman CA, et al. Multiband multislice GE-EPI at 7 tesla, with 16-fold acceleration using partial parallel imaging with application to high spatial and temporal whole-brain fMRI. Magn Reson Med. 2010;63:1144–1153. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Setsompop K, Gagoski BA, Polimeni JR, Witzel T, Wedeen VJ, Wald LL. Blipped-controlled aliasing in parallel imaging for simultaneous multislice echo planar imaging with reduced g-factor penalty. Magn Reson Med. 2012;67:1210–1224. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Liu G, Sobering G, Duyn J, Moonen CTW. A functional MRI technique combining principles of echo-shifting with a train of observations (PRESTO). Magn Reson Med. 1993;30:764–768. [DOI] [PubMed] [Google Scholar]
- 32.Liu G, Sobering G, Olson AW, Van Gelderen P, Moonen CTW. Fast echo-shifted gradient-recalled MRI: combining a short repetition time with variable T2* weighting. Magn Reson Med. 1993;30:68–75. [DOI] [PubMed] [Google Scholar]
- 33.Barth M, Breuer F, Koopmans PJ, Norris DG, Poser BA. Simultaneous multislice (SMS) imaging techniques. Magn Reson Med. 2016;75:63–81. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Setsompop K, Fan Q, Stockmann J, et al. High-resolution in vivo diffusion imaging of the human brain with generalized slice dithered enhanced resolution: simultaneous multislice (gSlider-SMS). Magn Reson Med. 2018;79:141–151. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Setsompop K, Feinberg DA, Polimeni JR. Rapid brain MRI acquisition techniques at ultra-high fields. NMR Biomed. 2016;29:1198–1221. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Ramsey NF, van den Brink JS, van Muiswinkel A, et al. Phase navigator correction in 3D fMRI improves detection of brain activation: quantitative assessment with a graded motor activation procedure. NeuroImage. 1998;8:240–248. [DOI] [PubMed] [Google Scholar]
- 37.Klarhöfer M, Dilharreguy B, van Gelderen P, Moonen CTW. A PRESTO-SENSE sequence with alternating partial-Fourier encoding for rapid susceptibility-weighted 3D MRI time series. Magn Reson Med. 2003;50:830–838. [DOI] [PubMed] [Google Scholar]
- 38.Gibson A, Peters AM, Bowtell R. Echo-shifted multislice EPI for high-speed fMRI. Magn Reson Imaging. 2006;24:433–442. [DOI] [PubMed] [Google Scholar]
- 39.Loenneker T, Hennel F, Hennig J. Multislice interleaved excitation cycles (MUSIC): an efficient gradient-echo technique for functional MRI. Magn Reson Med. 1996;35:870–874. [DOI] [PubMed] [Google Scholar]
- 40.van Gelderen P, Duyn JH, Ramsey NF, Liu G, Moonen CTW. The PRESTO technique for fMRI. NeuroImage. 2012;62:676–681. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Moonen CTW, Liu G, van Gelderen P, Sobering G. A fast gradient-recalled MRI technique with increased sensitivity to dynamic susceptibility effects. Magn Reson Med. 1992;26:184–189. [DOI] [PubMed] [Google Scholar]
- 42.Boyacioğlu R, Schulz J, Norris DG. Multiband echo-shifted echo planar imaging. Magn Reson Med. 2017;77:1981–1986. [DOI] [PubMed] [Google Scholar]
- 43.Bishop JE, Plewes DB. TE interleaving: new multisection imaging technique. J Magn Reson Imaging. 1991;1:531–538. [DOI] [PubMed] [Google Scholar]
- 44.Günther M, Feinberg DA. Simultaneous spin-echo refocusing. Magn Reson Med. 2005;54:513–523. [DOI] [PubMed] [Google Scholar]
- 45.Bilgic B, Ye H, Wald LL, Setsompop K. Simultaneous Time Interleaved MultiSlice (STIMS) for rapid susceptibility weighted acquisition. NeuroImage. 2017;155:577–586. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Pauly J, Le Roux P, Nishimura D, Macovski A. Parameter relations for the Shinnar-Le Roux selective excitation pulse design algorithm (NMR imaging). IEEE Trans Med Imaging. 1991;10:53–65. [DOI] [PubMed] [Google Scholar]
- 47.Souza SP, Szumowski J, Dumoulin CL, Plewes DP, Glover G. SIMA: simultaneous multislice acquisition of MR images by Hadamard-encoded excitation. J Comput Assist Tomogr. 1988;12:1026–1030. [PubMed] [Google Scholar]
- 48.Maki JH, Macfall JR, Johnson GA. The use of gradient flow compensation to separate diffusion and microcirculatory flow in MRI. Magn Reson Med. 1991;17:95–107. [DOI] [PubMed] [Google Scholar]
- 49.Hänicke W, Merboldt K-D, Gyngell ML, Bruhn H, Frahm J, Deimling M. MRI of “diffusion” in the human brain: new results using a modified CE-FAST sequence. Magn Reson Med. 2007;9:423–429. [DOI] [PubMed] [Google Scholar]
- 50.Perman WH, Moran PR, Moran RA, Bernstein MA. Artifacts from pulsatile flow in MR imaging. J Comput Assist Tomogr. 1986;10:473–483. [PubMed] [Google Scholar]
- 51.Haacke EM, Patrick JL. Reducing motion artifacts in two-dimensional Fourier transform imaging. Magn Reson Imaging. 1986;4:359–376. [DOI] [PubMed] [Google Scholar]
- 52.Conolly S, Nishimura D, Macovski A, Glover G. Variable-rate selective excitation. J Magn Reson. 1988;78:440–458. [Google Scholar]
- 53.Feinberg DA, Moeller S, Smith SM, et al. Multiplexed echo planar imaging for sub-second whole brain FMRI and fast diffusion imaging. PLoS ONE. 2010;5:e15710. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54.Smith SM, Beckmann CF, Andersson J, et al. Resting-state fMRI in the Human Connectome Project. NeuroImage. 2013;80:144–168. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55.Wiesinger F, Van de Moortele P-F, Adriany G, De Zanche N, Ugurbil K, Pruessmann KP. Parallel imaging performance as a function of field strength? An experimental investigation using electrodynamic scaling. Magn Reson Med. 2004;52:953–964. [DOI] [PubMed] [Google Scholar]
- 56.Sotiropoulos SN, Jbabdi S, Xu J, et al. Advances in diffusion MRI acquisition and processing in the Human Connectome Project. NeuroImage. 2013;80:125–143. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57.Grissom WA, Setsompop K, Hurley SA, Tsao J, Velikina JV, Samsonov AA. Advancing RF pulse design using an open-competition format: report from the 2015 ISMRM challenge. Magn Reson Med. 2017;78:1352–1361. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58.Poser BA, Anderson RJ, Guérin B, et al. Simultaneous multislice excitation by parallel transmission. Magn Reson Med. 2014;71:1416–1427. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 59.Guérin B, Setsompop K, Ye H, Poser BA, Stenger AV, Wald LL. Design of parallel transmission pulses for simultaneous multislice with explicit control for peak power and local specific absorption rate. Magn Reson Med. 2015;73:1946–1953. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 60.Norris DG, Koopmans PJ, Boyacioğlu R, Barth M. Power independent of number of slices (PINS) radiofrequency pulses for low-power simultaneous multislice excitation. Magn Reson Med. 2011;66:1234–1240. [DOI] [PubMed] [Google Scholar]
- 61.Koopmans PJ, Boyacioǧlu R, Barth M, Norris DG. Whole brain, high resolution spin-echo resting state fMRI using PINS multiplexing at 7T. NeuroImage. 2012;62:1939–1946. [DOI] [PubMed] [Google Scholar]
- 62.Boyacioğlu R, Schulz J, Müller NCJ, Koopmans PJ, Barth M, Norris DG. Whole brain, high resolution multiband spin-echo EPI fMRI at 7T: a comparison with gradient-echo EPI using a color-word Stroop task. NeuroImage. 2014;97:142–150. [DOI] [PubMed] [Google Scholar]
- 63.Eichner C, Wald LL, Setsompop K. A low power radiofrequency pulse for simultaneous multislice excitation and refocusing. Magn Reson Med. 2014;72:949–958. [DOI] [PubMed] [Google Scholar]
- 64.Chen N, Guidon A, Chang H-C, Song AW. A robust multi-shot scan strategy for high-resolution diffusion weighted MRI enabled by multiplexed sensitivity-encoding (MUSE). NeuroImage. 2013;72:41–47. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 65.Mani M, Jacob M, Kelley D, Magnotta V. Multi-shot sensitivity-encoded diffusion data recovery using structured low-rank matrix completion (MUSSELS). Magn Reson Med. 2017;78:494–507. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 66.von Morze C, Kelley DAC, Shepherd TM, Banerjee S, Xu D, Hess CP. Reduced field-of-view diffusion-weighted imaging of the brain at 7 T. Magn Reson Imaging. 2010;28:1541–1545. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 67.Heidemann RM, Anwander A, Feiweier T, Knösche TR, Turner R. k-space and q-space: combining ultra-high spatial and angular resolution in diffusion imaging using ZOOPPA at 7T. NeuroImage. 2012;60:967–978. [DOI] [PubMed] [Google Scholar]
- 68.Wald LL, Polimeni JR. Impacting the effect of fMRI noise through hardware and acquisition choices - implications for controlling false positive rates. NeuroImage. 2017;154:15–22. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
FIGURE S1 A, Z-statistical maps obtained from signal intensity (top) and phase data (bottom) from a visual stimulation task. Detected fMRI responses are found only in the signal magnitude reconstruction with little/no fMRI responses detected in the phase image. B, Estimated background phase from cgSlider acquisition and sliding-window reconstruction for 2 adjacent slices (A, B subslices depicted in Figure 1) and 3 time points (1.5, 2.5, 3.5) with little phase variation over time and between subslices
FIGURE S2 Z-statistic maps from visual stimulation with conventional SMS (A), cgSlider-SMS (B), cgSlider-MES-SI-SMS after velocity encoding phase correction (C), and cgSlider-MESSI-SMS after correction using the reference phase (D). Intensity correction was performed for all images
TABLE S1 Mean and standard deviation of z values (z > 2), and thenumber of activated voxels (clusters of minimum size of 30 voxels) from three subjects
