Speciation of chromium in waters using dispersive micro-solid phase extraction with magnetic ferrite and graphite furnace atomic absorption spectrometry

Ignacio López-García 1, Juan José Marín-Hernández 1, Manuel Hernández-Córdoba 1,
PMCID: PMC7093401  PMID: 32210320

Abstract

The combination of a solid-phase microextraction process with graphite furnace atomic absorption spectrometry provides a very sensitive determination method for determining chromium in waters. Freshly prepared ferrite particles are used to retain the chromium species, and then separated by a magnet without the need for a centrifugation step. The solid phase is suspended in water and directly introduced into the graphite furnace to obtain the analytical signal. The complexation of Cr(III) with ethylenediaminetetraacetate allows the selective retention of Cr(VI), and thus the speciation of the metal. The procedure is sensitive (0.01 µg L−1 detection limit when using a 10 mL sample aliquot) and reproducible (5% relative standard deviation for five consecutive experiments at the 0.3 µg L−1 level). The reliability of the procedure is verified by analysing five certified water samples.

Subject terms: Environmental chemistry, Hydrology

Introduction

The preparation of the sample in environmental analysis is a critical stage since it largely determines the quality of the results obtained and, consequently, methodologies are required that without losing efficiency and reliability be fast, affordable and sustainable1. The elemental trace analysis deals with the determination of metals, metalloids and non-metals that sometimes are present at very low concentrations. In biological or environmental samples, the determination is difficult because the levels of some elements may be even below the determination limit attainable in most conventional atomic techniques2,3. To this circumstance must be added the importance that elemental speciation has reached in recent years4,5, which has resulted in the development of metalomics6.

Within the different stages of sample preparation, the transfer of the analyte from a donor phase to another immiscible one (the acceptor phase) fulfil a double purpose since in addition the clean-up effect that avoids possible difficulties in the subsequent determination, allows a preconcentration of the analyte that facilitates measurement. The process should be carried out using simple, easily available reagents compatible with the analytical technique used for the final measurement and avoiding or minimizing the production of contaminated wastes. In most cases, the donor phase is already in a liquid state7 and the cleaning and separation stages are combined in a single stage. The acceptor phase may be a liquid immiscible with the donor phase, a supported or dispersed solid or a micellar phase8. The first two approaches are the most used in a large number of ways with their advantages and disadvantages. Liquid phase microextraction911 and the use of nanomaterials as the acceptor phase, especially when dispersed (dispersive solid phase microextraction, DSPME)12 have proven to be particularly useful for the purpose. The advantages of using nanomaterials in DSPME have aroused a great interest in recent years13,14. Due to its small particle size, the transfer of the analyte is rapid and, after the separation of the donor phase, the back-extraction of the analyte is carried out in a microvolume, that is then submitted to measurement in an instrument appropriate to deal with small volumes. When the adsorbent material has magnetic characteristics15 the separation of phases can be achieved by applying a magnetic field, which speeds up the overall process.

In the case of metallic species which are present at low concentrations, the final measurement stage is usually carried out using atomic absorption spectrometry (AAS), inductively coupled atomic emission spectrometry (ICP-AES) or inductively coupled plasma mass spectrometry (ICP-MS). When dealing with the extremely low concentrations of some toxic or hazardous metals in waters, ICP-MS is the best alternative since it allows very sensitive determinations of a large number of analytes. The supremacy of ICP-MS is undeniable, but this analytical technique is expensive both in terms of acquisition of the instruments and their maintenance, which sometimes put it beyond the reach of small or medium-sized laboratories. By contrast, AAS is a well-established technique in most laboratories; it is relatively cheap, consumes small amounts of gases and maintenance costs are low. The sensitivity attainable by AAS is good but below that possible with ICP-MS. However, the above mentioned modern microextraction techniques offer a way of boosting the analytical performance of AAS-based procedures by increasing sensitivity. As indicated, this methodology means the analytes can be transferred from a relatively large volume of sample to a few microlitres of extract, thus resulting in a preconcentration of the metal to be measured. In addition to liquid-liquid microextraction approaches11,1620 another interesting alternative is to use solid phase extraction or, better still, micro-solid phase extraction with an appropriate solid phase followed by releasing the analyte using a suitable reagent12,2023. Such a possibility is especially useful when combined with graphite furnace atomic absorption spectrometry (GFAAS) since, when using this atomization mode, only a small volume (10–20 µL) is required for the measurement. The interest of the approach further increases if, instead of separating the micro-solid phase from the liquid phase by a prolonged centrifugation step, a magnetic material is used as the solid phase since this obviates the need for centrifugation, and the application of a magnet allows the simple and rapid separation of phases2428.

Ferrite particles are suitable for the above purpose since they have excellent adsorptive characteristics, and their magnetic properties enable easy separation by a magnet. This methodology has been used for the separation of a number of metallic species2931 including the difficult case of chromium in waters3235 but, to the best of our knowledge, in all the analytical procedures reported to date, the ferrite particles are functionalized or mixed with other solid-phases to obtain nanocomposites with magnetic properties. Full benefit is not taken of the good adsorptive properties of the ferrite particles, which are merely used as a support to render the magnetic separation feasible. Recent experiments in our laboratory29,36 have demonstrated that freshly prepared ferrite particles are particularly effective for retaining small amounts of species, such as arsenic and silver, which can then be measured by GFAAS resulting in analytical determination procedures with a degree of sensitivity similar to that of ICP-MS. This manuscript reports the results obtained when using this approach (freshly prepared ferrite particles for micro-solid separation followed by GFAAS measurement) for the difficult case of determining low concentrations of chromium in waters. At the best of our knowledge, there are no previous reports using non-functionalized ferrite particles for the purpose. The procedure here studied is reliable, involves a non-expensive solid reagent which is easily synthesized and allows the two forms of chromium, trivalent and hexavalent, to be discriminated, which is of interest because of their different toxicity.

Methods

Chemicals

Chromium (VI) and chromium (III) stock solutions (1 g L−1) were prepared from K2Cr2O7 and Cr(NO3)3.9H2O (Fluka, Buchs SG,Switzerland), respectively, and diluted daily to obtain suitable standard working solutions. A 0.2 M Fe(II) solution was prepared from FeCl2.4H2O and a 0.1 M Fe(III) solution was prepared from FeCl3.6H2O, the solid reagents being provided by Sigma (St. Louis, MO, EE.UU.). Despite the high purity of these chemicals, and due to the extreme sensitivity of the analytical procedure, these solutions had to be purified to remove chromium traces that would have led to excessively high blank assays. Therefore, the Fe(III) solution was prepared in a 9 M hydrochloric acid medium and a 5 mL aliquot was shaken with the same volume of n-octanol. After centrifuging and discarding the aqueous phase, the extraction was repeated with a new aliquot of the organic solvent; the two organic extracts were mixed and then iron was back-extracted twice with 2.5 mL water. In this way, most of the chromium traces initially present were removed. In the case of the Fe(II) solution, purification was carried out by passing it through a minicolumn containing an anionic exchange resin (IRA-743), that retained most of the chromium while the Fe(II) concentration remained unchanged. Other chemicals used were obtained from Merck (Darmstadt, Germany).

Instrumentation

A Perkin-Elmer model 800 (Shelton, MA, USA) spectrometer was used for all the atomic absorption measurements. The spectrometer was equipped with a transversely heated electrothermal atomizer and a Zeeman-based correction device. The graphite atomizers as well as an automatic sampler were also obtained from Perkin-Elmer. The instrumental parameters and the heating program used are summarized in Table 1.

Table 1.

Instrumental parameters and heating program.

Parameter
Lamp current, mA 30
Wavelenght, nm 357.9
Slit, nm 0.7
Atomizer Transverse with L’Vov platform
Injected sample volume, µL 20
Chemical modifier none
Sample volumen, mL 10
Heating program
Step Temperature, °C Ramp, s Hold, s
1: Dry 110 10 20
2: Dry 130 15 30
3: Ashing 1500 10 20
4a: Atomization 2500 0 5
5: Cleaning 2550 1 3

aArgon flow 250 mL min−1 in all steps, except during atomization, where the gas flow was stopped.

The permanent magnet blocks (50 × 15 × 15 mm and 86 grams weight with a strength of 33 kg) composed of Nd-Fe-B that were used to carry out the magnetic separations were supplied by Supermagnete (Gottmadingen, Germany). A common ultrasonic bath and a vortex device were also used.

Samples and analytical procedure

Water samples

Six water samples were analyzed. Two of them were bottled mineral waters purchased in a local supermarket. A tap water sample was taken in the laboratory. Samples were also obtained from a natural spring and from the Segura river, which flows through Murcia, south-eastern Spain. A seawater sample was taken from a coastal marine lagoon in the same geographical area. All these samples were filtered and kept at 4 °C in plastic containers until the analyses were carried out.

Reference materials

In addition, five standard samples with a certified chromium content were used to verify the reliability of the results. These reference materials, namely SRM 1640a, NASS-6, SRM TM-23.4, SRM TM-25.4, and TMRain-04, were from the National Institute of Standards and Technology, the Research Council of Canada and Environment Canada.

Analytical procedures

The MNPs were obtained in situ as described elsewhere36, but the procedure is summarized here to help the reader. To 10 mL pure water, 0.1 mL of the 0.2 M Fe(II) solution and 0.1 mL of the 0.1 M Fe(III) solution were added. After heating at 60 °C, a small volume (40 µL) of concentrated ammonia solution was incorporated, and the mixture was submitted to ultrasounds for 4 min. The solid material was separated using a magnet and washed twice with 1 mL pure water. The MNPs thus obtained and remaining inside the tube were used directly. To determine the chromium total content, 10 mL of sample (0.3 M sodium hydrogen carbonate was incorporated to bring the pH close to neutrality, if necessary) was added to the tube containing the freshly prepared MNPs and, after shaking for a few seconds, the magnetic material was separated by applying a magnet to the external part of the tube. The supernatant was discarded, and the residue was washed twice with 1 mL water again using the magnet for the separation. Finally, a suspension was obtained by adding 0.1 mL water and homogenizing with the help of a vortex. A 20 µL aliquot was taken and introduced into the graphite atomizer before applying the heating program given in Table 1. The analytical signal (area obtained during the atomization stage) corresponded to the total chromium content. The measurement was always obtained in duplicate.

To calculate the Cr(VI) content, 0.3 mL of 1 M sodium hydrogen carbonate adjusted to pH = 7 and 0.1 mL of 0.01 M EDTA were incorporated in the sample, and the solution was heated at 60 °C for 15 min to achieve Cr(III) complexation. Next, the procedure described was repeated with another tube containing freshly prepared MNPs. The analytical signal finally obtained in the GFAAS instrument corresponded to Cr(VI). The concentration of Cr(III) was obtained by difference.

Results

Retention of chromium species by ferrite particles

All the experiments were carried out using freshly prepared ferrite particles for the preconcentration step. As was to be expected, the retention of chromium species by the solid particles strongly depended on the acidity of the medium. To study this parameter, a number of solutions containing 25 µg/mL Cr(III) or Cr(VI) were treated with the magnetic material and, after separating the solid by means of a magnet, the concentration of the metal remaining in the supernatant was measured. The results shown in Fig. 1 demonstrated that the retention of Cr(VI) was high in all the pH range studied, and that the trivalent form was practically totally retained for solutions close to neutrality, but less so as the acidity was increased to pH 4. At pHs below this value the solid phase was partially dissolved. This behavior agrees with the z-potential of the ferrite particles, as reported elsewhere36.

Figure 1.

Figure 1

Effect of pH on the retention of Cr (VI) and Cr (III) (curves a and b, respectively) using freshly synthesized ferrite. The shaded pH zone corresponds to a partial solubilization of the ferrite particles, and so the data have a greater variability.

Speciation of chromium

A large number of experiments were devoted to developing a strategy that allowed trivalent and hexavalent chromium to be discriminated, i.e., to achieve a reliable chromium speciation. Since both species are retained by the MNPs at pH values close to neutrality, several complexing agents for Cr(III) were assayed in the search for a robust complex that could avoid its retention. Excellent results were found when ethylenediaminetraacetate (EDTA) was used for the purpose. This chemical forms a very stable chelate with Cr(III) although the rate of formation is slow and requires mild heating and/or the presence of an auxiliary anion (carbonate) acting as a catalyst37. Once it was verified that the EDTA-Cr(III) complex was not retained on the MNPs at pH values close to 7, experiments were carried out to verify that the possible speciation would be reliable. To this effect it should be noted that there is a risk that Cr(VI) may oxidize EDTA, jeopardizing a correct speciation. However, in the case here considered it was experimentally verified that such a red-ox process did not take place because the pH was not acid and the solution was heated only gently. Both the temperature and time of heating as well as the EDTA concentration were optimized and found to be 60 °C during 15 min in the presence of 1 mM EDTA. The concentration of the carbonate incorporated in the solution to achieve a pH close to neutrality and to increase the rate of formation of the EDTA-Cr(III) complex was also optimized experimentally and a 0.03 M concentration was finally selected.

To summarize, chromium speciation can be achieved by means of two consecutive experiments, as detailed in the Experimental section. The first one allows the total concentration (Cr(VI) + Cr(III)) present in the sample to be calculated. The experiment is then repeated using another aliquot of sample but in the presence of 1 mM EDTA so that only Cr(VI) is retained by the MNPs, and then obtaining again the analytical signal. The concentration of the trivalent species is obtained by difference. The reliability of the strategy was checked by preparing a set of six solutions in which the Cr(VI)/Cr(III) ratio was varied from 50 to 0.02 using a concentration level for total chromium close to 5 µg/L The recoveries of the metal for five consecutive experiments, for each of the six solutions prepared were in the 98–102% range.

Calibration. Analytical figures of merit

Using 10 mL-sample aliquots as described in the Experimental section, calibration graphs were obtained by least-squares linear regression analysis of the analytical signal (peak area measured at the atomization stage) vs. chromium concentration and were linear in the 0.03–0.4 µg L−1 range (0.9983 for the regression coefficient of a typical six points calibration plot). A statistical test proved the absence of significant differences between the slopes of calibration graphs obtained from standard solutions prepared for trivalent or hexavalent chromium. The detection limit calculated on the basis of three times the standard error of the regression38 was found to be 0.01 µg L−1 chromium. The relative standard deviations for solutions containing 0.1 and 0.3 µg L−1 (five consecutive experiments in each case and measurements in duplicate) were 5.3 and 4.7%, respectively. It is of note that the enrichment factor, calculated as the ratio of a calibration graph divided by the slope of a calibration graph obtained from chromium solutions that were not submitted to the treatment with MNPs but directly analyzed was close to 100, which is the ratio of the sample volume used (10 mL) divided by the volume (0.1 mL) of the final solution in which the GFAAS measurement was carried out, thus confirming that chromium separation was practically total. The atomization profiles obtained when the heating program given in Table 1 was run were well-shaped, and the low background signals were easily corrected by the Zeeman device.

Table 2 summarizes the main characteristics of similar procedures reported for chromium determination at very low levels. The LOD of the procedure compares well with most of the other procedures, with the advantage of simplicity, low cost of reagents and easy preparation of the solid material. It is of note that the LOD indicated in this table (0.01 µg L−1 chromium) is based in a 10-mL sample aliquot. The LOD can be increased by increasing the volume of sample up to 50 mL but then reproducibility decreases. Since the limit of detection is enough low for all practical purposes the use of 10 mL for the volume of sample is recommended. The effect that the species commonly present in water samples have on the determination of chromium by application of the proposed procedure has been studied. Thus, it was experimentally verified that Na+, K+, Ca2+, Mg2+, NO3, Cl and SO42− ions are tolerated up to 5 g L−1. Metallic ions such as Cu (II), Co (II), Ni (II), Cd (II), Al (III), Fe (III), As (III) and Sb (III) are tolerated even in a 500:1 ratio. Other metallic ions that could also be retained in the ferrite are not interfering due to the selectivity of the detection technique, provided that the retention capacity of the adsorbent material is not exceeded.

Table 2.

Comparison of proposed procedures for the determination of Cr (VI) and/or Cr (III) using magnetic support.

Specie Adsorbent Reagent Desorption Detection VSample, mL LOD, µg/L
Cr(VI)/Cr(III)
EF, %
Cr(VI)/Cr(III)
Samples Ref.
Cr(III) CoFe2O4 PAN EDXRF 15 4 Etanol fuel 39
Cr(III), Cr(VI) CNTM-BGs DPC ethanol FO-LADS 50 0.1 318 Water 40
Cr(III), Cr(VI) Fe3O4@SiO2@Amino TAR HCl 2.5 M FAAS 45 1.1/3.2 16/12 Water and biological samples 41
Cr(III), Cr(VI) Fe3O4@GO@Trien NH4OH 2 M FAAS 50 1.4/1.6 10 Tannery wastewater, electroplating wastewater and river water 42
Cr(III), Cr(VI) Fe3O4@GO HNO3, 0.5 M + methanol + US FAAS 100 0.1 200 Environmental water 43
Cr(VI) Fe3O4@Cr(VI)IIPS HCl 1 M FAAS 500 0.3 98 Water 44
Cr(VI) Fe3O4@ADMPT DPC Vis-UV 10 2 Water and soils 45
Cr Fe3O4@decanoic PAN HCl 0.25 M + propanol FI-ICP-OES 47 0.5 120 Water 46
Cr(III) Fe3O4@En/MIL 101(Fe) HNO3 + EDTA FAAS 1000 0.5 238 SRM and agricultural samples 47
Cr(III) Fe3O4@ZrO2 HNO3, 0.5 M FAAS 75 0.7 25 Environmental and biological samples 48
Cr(III), Cr(VI) Fe3O4@Al2O3@Triton X-114 PAN HNO3, 0.5 M FAAS 200 1.4 120 Waters and soils 49
Cr(III), Cr(VI) Fe3O4@MnO2,Al2O3@AAPTMS HNO3, 2 M ICP-OES 0.02 94 SRM and river waters 32
Cr(III), Cr(VI) Fe3O4@En/MIL 101(Fe) /PAEDTC HNO3 + EDTA ETAAS 400 0.001 470 Water and tea 33
Cr(III), Cr(VI) Fe3O4@GO@Im HCl 2.2 M ETAAS 500 1.2/1.9 357 Water 50
Cr(III) Fe3O4@SiO2@MPA HNO3, 1 M FAAS 200 0.19 92 Biological and environmental samples 51
Cr(III), Cr(VI) Fe3O4@SiO2@Zincon HCl 2 M ETAAS 100 0.016/0.011 100/150 Water 52
Cr(III),Cr(VI) Fe3O4 (1) ETAAS 10 0.01 100 Water [*]

PAN: 1-(2-pyridylazo)-naphthol; EDXRF: energy dispersive X-ray fluorescence spectrometry; CNTM-BGs: carbon nanotube-based magnetic bucky gels; FO-LADS: fibre optic linear array detection spectrophotometer; DPC: 1,5-diphenylcarbazide; TAR: 4-(2-thiazolylazo)resorcinol; FAAS: flame atomic absorption spectrometry; Trien: triethylenetetramine; US: ultrasounds; Fe3O4@Cr(VI)IIPS: magnetic Cr (VI)-imprinted nanoparticles; ADMPT: 3-aminopropyltriethoxysilan-2,4-bis(3,5-dimethylpyrazol)triazine; FI-ICP-OES: flow injection inductively coupled plasma-optical emission spectrometry; Fe3O4@En/MIL 101(Fe): magnetic metal-organic framework nanocomposite; SRM: standard reference material; AAPTMS: [3-(2-aminoethylamino)propyl] trimethoxysilane; PAEDTC: 2-(propylamino-ethyl) dithiocarbamate; Im: imidazolium; MPA: 3-mercaptopropionic acid; (1): slurry in water; [*]: this work.

Results for water samples and certified reference materials

The optimized procedure was used to analyze six different water samples. All of them gave signals below the detection limit with the exception of a bottled mineral water sample that contained 0.1 µg L−1 total chromium (0.04 µgL−1 for the hexavalent species), a very low level without toxicological relevance. Table 3 shows details of the recovery tests used to confirm the results.

Table 3.

Analytical results obtained in the determination of Cr (III) and Cr (VI) in water samples.

Sample Added, ng/L Found, ng/L Recovery, %
Cr(III) Cr(VI) Cr(III) Cr(VI) Cr (total) Cr(III) Cr(VI)
Tap water

0

50

100

0

50

100

<LOD

53 ± 4

109 ± 5

<LOD

47 ± 5

92 ± 5

<LOD

100 ± 5

201 ± 5

106

109

94

92

Spring water

0

50

100

0

50

100

<LOD

51 ± 4

99 ± 5

<LOD

48 ± 4

98 ± 5

<LOD

99 ± 5

197 ± 7

102

99

96

98

Sea water

0

50

100

0

50

100

<LOD

52 ± 4

106 ± 5

<LOD

47 ± 5

93 ± 5

<LOD

99 ± 5

199 ± 6

104

106

94

93

River water

0

50

100

0

50

100

<LOD

47 ± 4

107 ± 6

<LOD

54 ± 5

94 ± 5

<LOD

101 ± 5

201 ± 6

94

107

108

94

Bottled water 1

0

50

100

0

50

100

<LOD

57 ± 4

105 ± 5

<LOD

52 ± 5

92 ± 5

<LOD

109 ± 5

197 ± 6

114

105

104

92

Bottled water 2

0

50

100

0

50

100

60 ± 4

108 ± 5

158 ± 6

35 ± 4

84 ± 5

132 ± 7

95 ± 4

187 ± 6

29 ± 7

96

92

98

97

aMean value of three determinations ± standard deviation.

The reliability of the results was checked by analyzing five standard reference materials with certified total chromium contents. It should be noted that, due to the sensitivity of the approach here presented, to obtain signals within the linear response range, four of these samples had to be diluted before analysis. The results given for total chromium and its speciation are given in Table 4.

Table 4.

Analytical results obtained in the determination of Cr (III) and Cr (VI) in reference materials.

Sample Dilution Certified Cr founda, µg L−1
Total, µg/L Cr(III) Cr(VI) Cr (total)
SRM 1640ab 1:500 40.22 ± 0.28 16.2 ± 0.1 26.6 ± 0.2 42.8 ± 0.1
SRM TM-23.4c 1:50 6.77 ± 0,63 6.11 ± 0.03 0.07 ± 0,01 6.28 ± 0.09
SRM TM-25.4d 1:100 24.0 ± 1.73 23.2 ± 0.1 0.09 ± 0.01 23.3 ± 0.1
NASS-6e 0.116 ± 0.008 0.05 ± 0.01 0.05 ± 0.01 0.114 ± 0.003
TMRain-04f 1:4 0.866 ± 0.165 0.90 ± 0.05 0.02 ± 0.01 0.92 ± 0.05

aMean value of three determinations ± standard deviation.

bTrace elements in natural water (an acidified spring water; details can be found in https://www-s.nist.gov/srmors/certificates/1640a.pdf).

cFortified (high level) and acidified Lake Ontario water; details can be found in https://topslide.net/document/certified-reference-material-tm-23-4-a-trace-element-fortified-sample.

dFortified (low level) and acidified Lake Ontario water.

eAcidified seawater (details can be found in https://nrc.canada.ca/en/certifications-evaluations-standards/certified-reference-materials/list/113/pdf/nass-6-en.pdf).

fSimulated rain sample for trace elements (details can be found in https://nwql.usgs.gov/Public/Performance/ECPT0098TE.pdf).

Conclusions

The determination of chromium at low concentrations in waters can be carried out by using graphite furnace atomic absorption spectrometry, GFAAS, an analytical technique available in most laboratories and sometimes underused, despite its advantages in terms of cost and maintenance compared with inductively coupled plasma mass spectrometry, ICP-MS. The combination of a modern microextraction process with the characteristics (sensitivity and selectivity) inherent in GFAAS provides a procedure involving low cost reagents, which makes such determinations feasible in laboratories with moderate budgets. In addition to the low cost and easy availability of the reagents used, the strength of the procedure lies in its simplicity, since the synthesis of the sorbent is quite simple, not requiring any immobilization of extractive groups on its surface. The approach allows the reliable non-chromatographic speciation of chromium, even at the low concentrations usually present in waters.

Acknowledgements

The authors acknowledge the financial support of the Comunidad Autónoma de la Región de Murcia (CARM, Fundación Séneca, Project 19888/GERM/15) the Spanish MICINN (PGC 2018-098363-B-100) and the European Commission (FEDER/ERDF).

Author contributions

I.L.G. and M.H.C. conceived the study. J.J.M.H. carried out the experiments and discussed with the other two authors the results. I.L.G. prepared the figure and together with M.H.C. wrote the manuscript. The three authors approved the final version of the manuscript.

Competing interests

The authors declare no competing interests.

Footnotes

Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Cerutti S, Pacheco PH, Gil R, Martinez LD. Green sample preparation strategies for organic/inorganic compounds in environmental samples. Curr. Opin. Green. Sustain. Chem. 2019;19:76–86. doi: 10.1016/j.cogsc.2019.08.007. [DOI] [Google Scholar]
  • 2.Butcher DJ. Recent highlights in graphite furnace atomic absorption spectrometry. Appl. Spectrosc. Rev. 2017;52:755–773. doi: 10.1080/05704928.2017.1303504. [DOI] [Google Scholar]
  • 3.He M, Huang LJ, Zhao BS, Chen BB, Hu B. Advanced functional materials in solid phase extraction for ICP-MS determination of trace elements and their species - A review. Anal. Chim. Acta. 2017;973:1–24. doi: 10.1016/j.aca.2017.03.047. [DOI] [PubMed] [Google Scholar]
  • 4.Werner J, Grześkowiak T, Zgoła-Grześkowiak A, Stanisz E. Recent trends in microextraction techniques used in determination of arsenic species. TrAC, Trends Anal. Chem. 2018;105:121–136. doi: 10.1016/j.trac.2018.05.006. [DOI] [Google Scholar]
  • 5.Hamilton EM, Young SD, Bailey EH, Watts MJ. Chromium speciation in foodstuffs: A review. Food Chem. 2018;250:105–112. doi: 10.1016/j.foodchem.2018.01.016. [DOI] [PubMed] [Google Scholar]
  • 6.Galazzi, R. M., de Jesus, J. R. & Arruda, M. A. Z. In Emerging Sample Treatments in Proteomics (ed José-Luis Capelo-Martínez) 161–185 (Springer International Publishing, 2019).
  • 7.Matusiewicz, H. In Handbook of Trace Analysis: Fundamentals and Applications (ed Irena Baranowska) 75–122 (Springer International Publishing, 2016).
  • 8.Samaddar P, Sen K. Cloud point extraction: A sustainable method of elemental preconcentration and speciation. J. Ind. Eng. Chem. 2014;20:1209–1219. doi: 10.1016/j.jiec.2013.10.033. [DOI] [Google Scholar]
  • 9.Basheer C, Kamran M, Ashraf M, Lee HK. Enhancing liquid-phase microextraction efficiency through chemical reactions. TrAC, Trends Anal. Chem. 2019;118:426–433. doi: 10.1016/j.trac.2019.05.049. [DOI] [Google Scholar]
  • 10.Hashemi B, et al. Recent advances in liquid-phase microextraction techniques for the analysis of environmental pollutants. TrAC, Trends Anal. Chem. 2017;97:83–95. doi: 10.1016/j.trac.2017.08.014. [DOI] [Google Scholar]
  • 11.Aguirre MA, Baile P, Vidal L, Canals A. Metal applications of liquid-phase microextraction. TrAC, Trends Anal. Chem. 2019;112:241–247. doi: 10.1016/j.trac.2018.11.032. [DOI] [Google Scholar]
  • 12.Ghorbani M, Aghamohammadhassan M, Chamsaz M, Akhlaghi H, Pedramrad T. Dispersive solid phase microextraction. TrAC, Trends Anal. Chem. 2019;118:793–809. doi: 10.1016/j.trac.2019.07.012. [DOI] [Google Scholar]
  • 13.Arain MB, Ali I, Yilmaz E, Soylak M. Nanomaterial’s based chromium speciation in environmental samples: A review. TrAC, Trends Anal. Chem. 2018;103:44–55. doi: 10.1016/j.trac.2018.03.014. [DOI] [Google Scholar]
  • 14.Pyrzynska K. Nanomaterials in speciation analysis of metals and metalloids. Talanta. 2020;212:120784. doi: 10.1016/j.talanta.2020.120784. [DOI] [PubMed] [Google Scholar]
  • 15.Filik H, Avan AA. Magnetic nanostructures for preconcentration, speciation and determination of chromium ions: A review. Talanta. 2019;203:168–177. doi: 10.1016/j.talanta.2019.05.061. [DOI] [PubMed] [Google Scholar]
  • 16.Campillo N, Viñas P, Sandrejova J, Andruch V. Ten years of dispersive liquid-liquid microextraction and derived techniques. Appl. Spectrosc. Rev. 2017;52:267–415. doi: 10.1080/05704928.2016.1224240. [DOI] [Google Scholar]
  • 17.Plotka-Wasylka J, Owczarek K, Namiesnik J. Modern solutions in the field of microextraction using liquid as a medium of extraction. TrAC, Trends Anal. Chem. 2016;85:46–64. doi: 10.1016/j.trac.2016.08.010. [DOI] [Google Scholar]
  • 18.Leong M-I, Fuh M-R, Huang S-D. Beyond dispersive liquid-liquid microextraction. J. Chromatogr. A. 2014;1335:2–14. doi: 10.1016/j.chroma.2014.02.021. [DOI] [PubMed] [Google Scholar]
  • 19.de la Calle I, Pena-Pereira F, Lavilla I, Bendicho C. Liquid-phase microextraction combined with graphite furnace atomic absorption spectrometry: A review. Anal. Chim. Acta. 2016;936:12–39. doi: 10.1016/j.aca.2016.06.046. [DOI] [PubMed] [Google Scholar]
  • 20.Bendicho C, Bendicho-Lavilla C, Lavilla I. Nanoparticle-assisted chemical speciation of trace elements. TrAC. Trends Anal. Chem. 2016;77:109–121. doi: 10.1016/j.trac.2015.12.015. [DOI] [Google Scholar]
  • 21.Płotka-Wasylka J, Szczepańska N, de la Guardia M, Namieśnik J. Miniaturized solid-phase extraction techniques. TrAC, Trends Anal. Chem. 2015;73:19–38. doi: 10.1016/j.trac.2015.04.026. [DOI] [Google Scholar]
  • 22.Xu CH, et al. Applications of solid-phase microextraction in food analysis. TrAC, Trends Anal. Chem. 2016;80:12–29. doi: 10.1016/j.trac.2016.02.022. [DOI] [Google Scholar]
  • 23.Souza-Silva EA, Jiang R, Rodriguez-Lafuente A, Gionfriddo E, Pawliszyn J. A critical review of the state of the art of solid-phase microextraction of complex matrices I. Environmental analysis. TrAC, Trends Anal. Chem. 2015;71:224–235. doi: 10.1016/j.trac.2015.04.016. [DOI] [Google Scholar]
  • 24.Hernandez-Hernandez AA, Alvarez-Romero GA, Contreras-Lopez E, Aguilar-Arteaga K, Castaneda-Ovando A. Food Analysis by Microextraction Methods Based on the Use of Magnetic Nanoparticles as Supports: Recent Advances. Food Anal. Meth. 2017;10:2974–2993. doi: 10.1007/s12161-017-0863-9. [DOI] [Google Scholar]
  • 25.Rios A, Zougagh M, Bouri M. Magnetic (nano)materials as an useful tool for sample preparation in analytical methods. A review. Anal. Methods-UK. 2013;5:4558–4573. doi: 10.1039/c3ay40306h. [DOI] [Google Scholar]
  • 26.Rios A, Zougagh M. Recent advances in magnetic nanomaterials for improving analytical processes. TrAC, Trends Anal. Chem. 2016;84:72–83. doi: 10.1016/j.trac.2016.03.001. [DOI] [Google Scholar]
  • 27.Chisvert A, Cardenas S, Lucena R. Dispersive micro-solid phase extraction. TrAC, Trends Anal. Chem. 2019;112:226–233. doi: 10.1016/j.trac.2018.12.005. [DOI] [Google Scholar]
  • 28.Hu, B., He, M. & Chen, B. In Solid-Phase Extraction (ed Colin, F. Poole) 235–284 (Elsevier, 2020).
  • 29.López-García I, Marín-Hernández JJ, Hernández-Córdoba M. Magnetic ferrite particles combined with electrothermal atomic absorption spectrometry for the speciation of low concentrations of arsenic. Talanta. 2018;181:6–12. doi: 10.1016/j.talanta.2017.12.086. [DOI] [PubMed] [Google Scholar]
  • 30.López-García I, Vicente-Martínez Y, Hernández-Córdoba M. Determination of ultratraces of mercury species using separation with magnetic core-modified silver nanoparticles and electrothermal atomic absorption spectrometry. J. Anal. At. Spectrom. 2015;30:1980–1987. doi: 10.1039/c5ja00213c. [DOI] [Google Scholar]
  • 31.López-García I, Rengevicova S, Muñoz-Sandoval MJ, Hernández-Córdoba M. Speciation of very low amounts of antimony in waters using magnetic core-modified silver nanoparticles and electrothermal atomic absorption spectrometry. Talanta. 2017;162:309–315. doi: 10.1016/j.talanta.2016.10.044. [DOI] [PubMed] [Google Scholar]
  • 32.Munonde TS, Maxakato NW, Nomngongo PN. Preconcentration and speciation of chromium species using ICP-OES after ultrasound-assisted magnetic solid phase extraction with an amino-modified magnetic nanocomposite prepared from Fe3O4, MnO2 and Al2O3. Microchim. Acta. 2017;184:1223–1232. doi: 10.1007/s00604-017-2126-2. [DOI] [Google Scholar]
  • 33.Saboori A. A nanoparticle sorbent composed of MIL-101(Fe) and dithiocarbamate-modified magnetite nanoparticles for speciation of Cr(III) and Cr(VI) prior to their determination by electrothermal AAS. Microchim. Acta. 2017;184:1509–1516. doi: 10.1007/s00604-017-2155-x. [DOI] [Google Scholar]
  • 34.Periyasamy S, Gopalakannan V, Viswanathan N. Fabrication of magnetic particles imprinted cellulose based biocomposites for chromium(VI) removal. Carbohydr. Polym. 2017;174:352–359. doi: 10.1016/j.carbpol.2017.06.029. [DOI] [PubMed] [Google Scholar]
  • 35.Wei W, Zhao B, He M, Chen B, Hu B. Iminodiacetic acid functionalized magnetic nanoparticles for speciation of Cr(iii) and Cr(vi) followed by graphite furnace atomic absorption spectrometry detection. RSC Adv. 2017;7:8504–8511. doi: 10.1039/C6RA27544C. [DOI] [Google Scholar]
  • 36.López-García I, Marín-Hernández JJ, Hernández-Córdoba M. Freshly prepared magnetic ferrite for the speciation of silver using dispersive micro-solid phase extraction and electrothermal atomic absorption spectrometry. J. Anal. At. Spectrom. 2019;34:2112–2118. doi: 10.1039/C9JA00241C. [DOI] [Google Scholar]
  • 37.Phatak GM, Bhat TR, Shankar J. Catalytic effects in reactions of chromium(III) — I: Kinetics of reaction of chromium(III) with ethylenediaminetetraacetate in the presence of carbonate, nitrite and sulphite catalysts. J. Inorg. Nucl. Chem. 1970;32:1305–1311. doi: 10.1016/0022-1902(70)80129-4. [DOI] [Google Scholar]
  • 38.Miller, J. N. & Miller, J. C. Statistics and Chemometrics for Analytical Chemistry, 6th Ed., (2010).
  • 39.Meira LA, et al. Multi-element determination of Cd, Pb, Cu, V, Cr, and Mn in ethanol fuel samples using energy dispersive X-ray fluorescence spectrometry after magnetic solid phase microextraction using CoFe2O4 nanoparticles. Microchem. J. 2018;142:144–151. doi: 10.1016/j.microc.2018.06.025. [DOI] [Google Scholar]
  • 40.Yousefi SM, Shemirani F. Carbon nanotube-based magnetic bucky gels in developing dispersive solid-phase extraction: application in rapid speciation analysis of Cr(VI) and Cr(III) in water samples. Int. J. Environ. Anal. Chem. 2017;97:1065–1079. doi: 10.1080/03067319.2017.1381236. [DOI] [Google Scholar]
  • 41.Diniz KM, Teixeira Tarley CR. Speciation analysis of chromium in water samples through sequential combination of dispersive magnetic solid phase extraction using mesoporous amino-functionalized Fe3O4/SiO2 nanoparticles and cloud point extraction. Microchem. J. 2015;123:185–195. doi: 10.1016/j.microc.2015.06.011. [DOI] [Google Scholar]
  • 42.Islam A, Ahmad H, Zaidi N, Kumar S. A graphene oxide decorated with triethylenetetramine-modified magnetite for separation of chromium species prior to their sequential speciation and determination via FAAS. Microchim. Acta. 2016;183:289–296. doi: 10.1007/s00604-015-1641-2. [DOI] [Google Scholar]
  • 43.Kazemi E, Shabani AMH, Dadfarnia S, Izadi F. Speciation and determination of chromium ions by dispersive micro solid phase extraction using magnetic graphene oxide followed by flame atomic absorption spectrometry. Int. J. Environ. Anal. Chem. 2017;97:1080–1093. doi: 10.1080/03067319.2017.1381693. [DOI] [Google Scholar]
  • 44.Qi X, Gao S, Ding G, Tang A-N. Synthesis of surface Cr (VI)-imprinted magnetic nanoparticles for selective dispersive solid-phase extraction and determination of Cr (VI) in water samples. Talanta. 2017;162:345–353. doi: 10.1016/j.talanta.2016.10.040. [DOI] [PubMed] [Google Scholar]
  • 45.Behzad SK, Amini MM, Balati A, Ghanbari M, Sadeghi O. Adsorption and determination of Cr(VI) in environmental samples using triazine-modified Fe3O4 nanoparticles: kinetics and equilibrium modeling. J. Sol-Gel Sci. Technol. 2016;78:446–456. doi: 10.1007/s10971-016-3965-8. [DOI] [Google Scholar]
  • 46.Faraji M, et al. A nanoparticle-based solid-phase extraction procedure followed by flow injection inductively coupled plasma-optical emission spectrometry to determine some heavy metal ions in water samples. Anal. Chim. Acta. 2010;659:172–177. doi: 10.1016/j.aca.2009.11.053. [DOI] [PubMed] [Google Scholar]
  • 47.Babazadeh M, Hosseinzadeh-Khanmiri R, Abolhasani J, Ghorbani-Kalhor E, Hassanpour A. Solid phase extraction of heavy metal ions from agricultural samples with the aid of a novel functionalized magnetic metal-organic framework. RSC Adv. 2015;5:19884–19892. doi: 10.1039/c4ra15532g. [DOI] [Google Scholar]
  • 48.Wu YW, et al. Fe3O4@ZrO2 nanoparticles magnetic solid phase extraction coupled with flame atomic absorption spectrometry for chromium(III) speciation in environmental and biological samples. Appl. Surf. Sci. 2012;258:6772–6776. doi: 10.1016/j.apsusc.2012.03.057. [DOI] [Google Scholar]
  • 49.Tavallali H, Deilamy-Rad G, Peykarimah P. Preconcentration and speciation of Cr(III) and Cr(VI) in water and soil samples by spectrometric detection via use of nanosized alumina-coated magnetite solid phase. Environ. Monit. Assess. 2013;185:7723–7738. doi: 10.1007/s10661-013-3130-6. [DOI] [PubMed] [Google Scholar]
  • 50.Sarikhani Z, Manoochehri M. Determination of Ultra Trace Cr(III) and Cr(VI) Species by Electrothermal Atomic Absorption Spectrometry after Simultaneous Magnetic Solid Phase Extraction with the Aid of a Novel Imidazolium-Functionalized Magnetite Graphene Oxide Nanocomposite. Bull. Chem. Soc. Jpn. 2017;90:746–753. doi: 10.1246/bcsj.20160407. [DOI] [Google Scholar]
  • 51.Mashhadizadeh MH, Amoli-Diva M. Atomic absorption spectrometric determination of Al3+ and Cr3+ after preconcentration and separation on 3-mercaptopropionic acid modified silica coated-Fe3O4 nanoparticles. J. Anal. At. Spectrom. 2013;28:251–258. doi: 10.1039/c2ja30286a. [DOI] [Google Scholar]
  • 52.Jiang H-m, Yang T, Wang Y-h, Lian H-z, Hu X. Magnetic solid-phase extraction combined with graphite furnace atomic absorption spectrometry for speciation of Cr(III) and Cr(VI) in environmental waters. Talanta. 2013;116:361–367. doi: 10.1016/j.talanta.2013.05.008. [DOI] [PubMed] [Google Scholar]